Limitation of Warranty ...................................................................................................... 22
Contact Information ......................................................................................................... 22
1 INTRODUCTION
The YSI 9300 and 9500 direct-read photometers are designed to give long and
trouble-free operation. To ensure the best results, please read this manual
carefully and follow the procedures recommended. This manual covers both
the 9300 and 9500 photometers. Therefore, some of the information only
pertains to the 9500 as is noted in the appropriate sections.
The Photometers feature digital electronics and built-in filters. It is lightweight
and portable for field or laboratory use. The instruments are rugged, durable
and IP-67 rated. Additionally, the photometers are direct-reading, have
automatic blank setting, automatic wavelength selection, and automatic power
cut-off.
The following pages describe the use of the photometers, and give
instructions for the wide range of water tests which can be performed using
these instruments.
Keep the photometer clean and in good working order by adhering to the
following recommendations:
Do not pour out samples or prepare the tests directly over the instrument.
Always cap the test tubes before inserting into the instrument for readings.
Wipe test tubes with a clean tissue to remove drips or condensation before
placing in the photometer.
Do not leave tubes standing in the photometer test chamber. Remove the
tubes immediately after each test.
Immediately wipe up any drips or spills on the instrument or in the test
chamber with a clean tissue.
Keep the instrument clean. Clean the test chamber regularly using a
moistened tissue or cotton ball.
Keep the instrument away from all chemicals and cleaning materials.
Keep the instrument in a clean, dry place when it is not in use. Keep it on a
clean, dry bench away from chemicals, place it in a storage cupboard or
keep it in a carrying case.
Keep the carrying case in a clean, dry condition. Make sure that the carrying
case is dry before the case is closed up and the instrument is put away.
1
Test and
Number Keys
Multi-Size Cell
OK (Enter) Key
Scroll and Selection Keys
On/Off Key
USB Interface
(9500 only)
Graphic Display
Instrument Layout
2
Features and Technical Specification
Application For application in general water testing using Palintest
tablet reagent systems and Palintest Tubetests reagents.
Instrument Type Single-beam colorimeter with built-in colour filters and pre-
5nm and 655 5nm
Range 1 - 100%T
Accuracy 1%T
Display Large graphic display with option of backlight.
Language Test identification and prompts in English, French,
German, Spanish and Italian.
Timer Clock and timer feature to log test results and audible
alarm for timing test procedure.
Units Direct-reading of test results in mg/l, ppm, g/l or molar
units (mmol/l or µmol/l).
User Selectable Options 10 digit sample number entry, dilution factor, time/date,
date format, system lock and rounding of results.
Date Format Date format selectable as day/month/year or month/day/
year.
Zeroing Automatic zeroing on blank tube and hold blank facility for
series of tests. Continuation test facility without the need
for reblanking.
Internal Memory Stores 500 previous readings with option to view logged
results on screen, or download to computer.
USB Interface USB 1.1 full-speed, bus-powered device. Software selectable
between either emulation of a removable hard-drive or
emulation of a serial device connected via a virtual COM port.
Power 3 x 1.5V ‘AA’ alkaline batteries or via USB interface. Power
management system with variable length auto switch-off
or 'continuous' operation.
Size 250 x 150 x 70 mm
Weight 985 g (2.1 lbs)
Test Tubes For tablet reagents - 10 ml glass test tubes, 20 mm OD
(YPT 595).
Cell Holder Multi-size tube holder accepts test tubes from 12 –
20 mm OD and centres the tube for optimum optical
performance.
3
2 OPERATING PRINCIPLE
The YSI photometers are instruments that measure color intensity. Light is
passed through a test tube containing the sample solution, and then through a
colored filter onto a photodetector. Filters have been chosen so that light of a
specific wavelength is selected. When the solution is completely colorless, all of
the light passes through the sample. With colored samples, light is absorbed
and the light which passes through the sample is proportionately reduced.
In the following test procedures, the photometer is used to measure the color
which is produced when chemical reagents are reacted with the water sample.
In these tests, the color intensity produced is proportional to the concentration
of the parameter being tested.
The photometer is pre-programmed with calibrations for each test parameter.
Different test procedures are carried out at different wavelengths to optimize
the sensitivity of each test. The required wavelength is selected automatically
by the instrument.
The calibrations are accessed by entering a unique program number at the start
of each test procedure. This enables the instrument to select the appropriate
wavelength filter automatically and allows the photodiode response to be
converted to a concentration reading. The instrument thus displays a directreading of the test result.
The photometer is ideally suited for general analytical applications. The
instrument can be used as a laboratory or field photometer with user-generated
calibration graphs for standard analytical methods or for comparison of colored
solutions.
For general analytical applications, Transmittance (test program 0), or Absorbance
(test program 1) can be chosen.
Powering the Photometer
The Photometer is powered by (3) AAA batteries. The photometer features a
battery indicator – see ‘System Mode’ functions. A minimum voltage of 3.0V is
needed to operate the photometer. As a power-saving measure, in normal
use, both the 9300 and 9500 photometers automatically switch off five
minutes after the last key is pressed. The switch off period may be adjusted
for the 9500 in the System mode.
In addition to the above feature, a battery-warning message will appear
automatically on the display when the battery voltage becomes low. The batteries
should be replaced as soon as possible after the warning message appears.
Stored data in the instrument memory will not be lost during battery replacement.
4
Replacing the Batteries
The battery compartment in the base of the instrument is secured by four
screws. To replace the batteries, remove the cover and install the batteries,
observing the correct polarity as indicated. Use 3 x 1.5V 'AA’ alkaline batteries
or equivalent. To avoid corrosion damage through leakage, remove batteries
from the instrument if it is to be stored or left unused for a long period of time
(> 30 days).
Power Supply (Model 9500 only)
The 9500 photometer can be powered either from alkaline batteries or via the
USB socket. To use mains power, the instrument is connected using the USB
Connection Cable (YPT 284) plugged to the Mains Adapter (YPT283).
Alternatively, the USB connection cable can be plugged into a computer to
power the 9500 from the computer.
5
3 GENERAL PHOTOMETER OPERATION
The photometer is controlled by a simple intuitive menu system:
The highlight indicates the active line or section of the screen
The and keys move the highlight through the menu choices
The and keys allow selection of options
The flashing cursor in the ‘Options’ menu at the bottom of the screen
indicates the action which will occur if the [OK] button is pressed.
Operating Modes
The photometer has two distinct operating modes - the PHOTOMETER mode
and the SYSTEM mode.
The PHOTOMETER mode is the normal operating mode for taking photometer
readings. This mode is engaged automatically when the instrument is turned
on by pressing the key.
As a power-saving measure, in normal use, both the 9300 and 9500 photometers
automatically switch off five minutes after the last key is pressed. This may be
adjusted for the 9500 in the System mode.
The SYSTEM mode is used to set the system options. This mode is engaged
when the photometer is turned on using the key and then selecting 'System'
using the and keys and pressing [OK].
Scroll through the menu box to view all the options available.
System - Quick Start
When the instrument is first used, the SYSTEM mode should be used to set
the preferred operating options:
Use the and keys to scroll through the features.
Use the and keys to select the options.
Press [OK] to accept the selections and return to PHOTOMETER mode.
Select the desired language from English, French, German, Spanish or Italian.
Select the desired display units from mg/l, ppm, mmol/l, µmol/l and g/l.
Set the sample number option to ‘On’ to allow the entry of a sample
number during normal photometer operation (model 9500 only).
Set the sample increment option to ‘On’ to automatically increase the
sample number (model 9500 only).
6
Set the dilution factor to ‘On’ or ‘Off’. If the dilution factor option is set to
‘On’, the instrument will allow the entry of a numerical factor which will be
used in the calculation of the result displayed on the instrument (model
9500 only).
Select the preferred date format. The date may be shown in either Day/
Month/Year or Month/Day/Year (9500 only).
To change the date and time, select the date and time line then key in
correct setting using the numeric keys. To correct an error, use the
keys to move the cursor then key in the correct data (9500 only).
System - Full Options
The Photometer features a wide range of options which may be explored at
leisure to get the best results from the instrument. An explanation of the
application of these options is as follows:
View Log (9500 Only)
The 9500 photometer has an internal memory which can hold up to 500 test
results. Once the memory is full, each new result overwrites the oldest entry.
Select ‘View Log’ to view stored results on screen. The
used to scroll through the list of stored results. The ‘Options’ menu offers
several choices.
Select ‘Clear’ to empty the memory. Confirmation is requested to avoid
accidentally erasing the data. Select ‘Exit’ to return to SYSTEM mode. Select
‘Download’ to transmit stored data to a PC. This option only appears if the
USB mode is set to ‘COM Port’. Refer to ‘Interface Connection and Data
Memory’ for further information.
Back Light
The graphic display features a backlight to enhance the display contrast. This
may be switched off to conserve power when working on battery power.
Language Options
The photometer can be operated in a number of different languages. When a
particular language is selected, the test names and operating commands will
appear in that language. Certain tests and unit options are provided in
accordance with the conventions of particular countries and are only available
when the photometer is switched to the language that particular country.
7
Units
The photometer offers the choice of result expressed in mg/l, ppm, mmol/l,
µmol and g/l.
Sample Number (9500 Only)
A unique number may be associated with each result record to identify it in
the log. If Sample Number ‘On’ is selected, the user is offered the choice of
entering a number of up to 10 digits for each sample reading. If this function
is set to ‘Off’, a sample number is automatically allocated.
Sample Number Increment (9500 Only)
The sample number increment option may be used to determine whether the
instrument does or does not automatically increment the sample number after
each test. Incrementation of the sample number may be used when the
instrument is used for carrying out a series of similar tests. Alternatively it
may be preferable not to increment the number if typical use involves carrying
out a number of different tests on the same sample.
Dilution Factor (9500 Only)
When samples are out of range for the test, a dilution procedure may be
used. If the dilution factor option is set to ‘On’, the instrument will allow entry
of a numerical factor which will be used in the calculation of the result
displayed and stored in the log.
Date and Time (9500 Only)
The instrument records the date and time of each reading taken and
associates this with the data record in the log. To correct the date and time
on the internal clock, select the date and time display line.
Date Format (9500 Only)
The option of day/month/year or month/day/year date format is available.
Battery Level
A battery level indicator shows the power available. At least 3.0V is required
for successful operation of the instrument.
8
Locking System Mode Settings (9500 Only)
It is possible to 'lock’ the system settings so that these cannot be tampered
with or altered accidentally during use. This is important, for example, where
it is necessary to verify that tests have actually been carried out at a particular
time or date, or where procedures always require the use of a sample number
or dilution factor.
The instructions for locking the settings are not included in this manual; these
are provided to photometer owners or system administrators on formal
request to YSI’s Technical Support department (environmental@ysi.com).
Rounding (9500 Only)
In the normal default setting, the photometer will round test results
appropriately for the resolution of the test. The rounding applied differs for each
parameter depending on the shape of the calibration curve. This ensures the
optimum precision and accuracy of each test procedure. For normal purposes it
is strongly recommended that the instrument be left in the default setting.
However, for certain analytical applications, it may be useful to switch off the
rounding to display the result in unrounded form. This may be the case, for
example, when carrying out statistical evaluations of test methods where it is
necessary to use the data in calculation of standard deviation or distribution
data.
Time-Out (9500 Only)
As a power-saving measure, in normal use, both the 9300 and 9500 photometers
automatically switch off five minutes after the last key is pressed.
The 9500 photometer may be switched to ‘Long’ time-out which allows 15
minutes before shut-down or ‘Off’ which allows continuous use. This is
particularly useful when powering the instrument through the USB interface.
Edit User Defined Tests (9500 Only)
Users may wish to develop their own test methods and store calibrations on
the photometer. The 9500 has the facility to store up to 30 user-defined
calibrations. See ‘User Defined Tests’ below for full instructions.
9
USB (9500 Only)
The USB interface allows communication between the instrument and a PC.
There is a choice of two operating modes – Hard Drive and COM Port.
In Hard Drive mode, the instrument appears as a removable hard drive when
connected to a PC. No additional software is required on computers running
Windows 2000, ME or XP. A driver to use this option with Windows 98SE is
available from YSI Technical Support Department (environmental@ysi.com).
In COM Port mode, the instrument behaves as if connected to the PC serial port via
RS232. In this mode, the PC requires installation of a USB virtual COM Port driver,
available from YSI Technical Support Department (environmental@ysi.com).
See the section on ‘Interface Connections and Data Memory’ below for full
instructions.
10
4 INTERFACE CONNECTIONS AND DATA MEMORY (9500 ONLY)
Stored data can be accessed by recall to the instrument display (see ‘View
Log’). Alternatively, data can be accessed using a PC:
Connect the instrument to the computer via the USB port, using any
suitable USB cable, ie YPT 746
Turn the instrument ON and select SYSTEM mode from the ‘Options’ menu
Scroll to ‘USB’ and select either ‘Hard Drive’ or ‘COM Port’.
‘Hard Drive’ – Once this option is selected, simply turning the instrument ON
while it is connected to a PC will cause an extra hard drive containing the
instrument files to appear on the PC. The log of test results is in text file:
‘9500_LOG.txt’. The other files shown on screen contain calibration and operating
systems for use when upgrading the instrument and should not be accessed.
The log file can be copied from the instrument by dragging between windows.
Once copied, the file can be opened with many text editors, word processors
or spreadsheet programs.
Note: Deleting this file from the instrument’s hard drive will clear the data
from the instrument memory.
‘COM Port’ – Once this option is selected, data can be downloaded from the
instrument to the PC:
Open the ‘Virtual COM Port – HyperTerminal’ window on the computer
In the instrument SYSTEM mode, scroll to ‘View Log’ and select ‘Download’.
The data from the log will appear on the PC screen and can be transferred
to other PC applications or printed.
‘Unplugged’ – Note that the ‘Hard Drive’ or ‘Com Port’ may only be selected
while the instrument is being powered via its USB port. If the instrument is
running on batteries and is not connected to either a PC or a YPT783 external
power supply, an ‘Unplugged’ message will be displayed instead of either
‘Hard Drive’ or ‘COM Port’.
11
5 TAKING PHOTOMETER READINGS
The photometer is very simple to use. Screen prompts guide the user
towards the test result. The following sections describe how to get the best
results from the instrument.
Program Numbers and Test Instructions
Each test is identified by a separate program number or named key. Program
numbers are shown in the test instruction sheets supplied in this manual. For
some tests, a choice of different programs is offered in order to get the result
in different forms (for example, for Nitrate - NO3 or Nitrate Nitrogen - NO3-N).
In certain tests, such as free chlorine and total chlorine, the test can be
continued to a further stage. This is allowed for in the programming of the
photometer. In these tests, once the result of the first stage is obtained, the
‘Follow-On’ option may be selected to progress the test to the next test stage
or stages. The result will be calculated automatically.
These continuation programs have their own program number for reference
purposes although direct access to these programs may be restricted.
Sample Dilution (9500 Only)
The photometer has a sample dilution option. This enables a factor to be
entered when samples have been diluted to bring them within the measuring
range of the test. For example, if a five times dilution of a sample has been
made, then a dilution factor of x5 should be entered. The photometer will
multiply the observed result by this factor so that the correct result for the
original sample is displayed.
This option may be used in conjunction with the YSI Dilution Tube (YPT 512)
which enables dilutions of x2, x3, x4, x5 and x10 to be made. Higher dilution
factors may be entered but are subject to the limitation of the number of
digits available of the result display for each test. When the display capabilities
are exceeded, the symbol [xxx] will appear on the result display. The sample
should not be diluted prior to carrying out a pH test, or a Transmittance or
Absorbance reading.
Blank and Sample Tubes
A BLANK TUBE is needed each time the photometer is used. This enables the
instrument to be set automatically and compensates for any inherent color in
the test sample. It is important therefore to understand the meaning of the
term ‘BLANK TUBE'.
12
The BLANK TUBE is a test tube filled only with the water being tested only. It
is important to use the actual water to be tested to provide a true comparison
for the test results.
The term 'SAMPLE TUBE' is used to describe the tube containing the water
sample to which the reagents have been added in accordance with the
appropriate test instructions. This tube is used to take the photometer
reading.
Light Cap
A light cap is provided with the photometer. This cap fits over the test
chamber and prevents stray light reaching the photodiode.
It is NOT necessary to use the light cap when using the photometer indoors or
under shaded outdoor light. The light cap should be used when working outside
in strong sunlight. The light cap is also recommended when carrying out
turbidity-based tests such as the cyanuric acid test, under bright or variable
lighting conditions. Test instructions indicate when the light cap should be used.
Getting the Best Results
Success in obtaining accurate and consistent test results will depend on the
care with which test procedures are carried out. Always follow the test
instructions carefully and observe the stated standing periods and
temperature conditions where applicable.
Wipe test tubes free from condensation before placing in the photometer.
Test tubes should always be kept in a clean condition. Wash and dry the test
tubes carefully after use. Dirty tubes may be soaked in a mild detergent
solution if necessary. Tubes which become stained or scratched should be
discarded and replaced.
Taking Test Readings
Press key. The instrument displays the ‘Choose a Test’ menu box, with the
last test program used highlighted as the active line.
The cursor will flash on the [OK] symbol of the ‘options menu' at the bottom
of the screen. Press [OK] to accept this program.
To choose a different test program, either use the and keys to scroll
through the menu options, or use the numeric keys to enter the Phot number
of the desired test. The four most recently used tests are listed at the top of
the ‘Choose a Test’ screen for convenience. Press [OK] to accept the selected
program.
13
1024
Sam p l e N u m ber
OK
x 1
D ilu t i o n F a c t o r
OK
C hl o rin e-F ree / 5
P h ot 0 0 7
In se r t B la n k
Ch oos e a Te st
OK
If the sample number option is pre-selected, then the following display will
appear, for example (9500 only):
Enter or confirm the sample number (up to 10 digits), then press [OK].
1 If the dilution factor option is pre-selected, then the following display will
appear (9500 only):
Press [OK] to accept the default value (x1, no dilution), or key in new
dilution factor then press [OK].
2 The following display will now appear:
Place a BLANK TUBE in the test chamber, then press [OK].
NOTE: The instrument is designed to hold the blank setting as long as the
instrument is switched on. This stage will be omitted when further tests are
being carried out. However, when changing to a test which requires a
colored or reagent blank, or uses a tube of a different diameter, a new
blank reading is required. The ‘Insert Blank’ prompt will be displayed
automatically.
If the instrument is in used continuously, it is advisable to re-blank from
time to time.
14
C hl orin e -F ree / 5
P h ot 0 0 7
In se r t S am p l e
Bla nk - C hoo se a T es t - T im e
OK
Chlorine-Free / 5
Phot 007
1.00
mg/l Cl
2
Choose a Test
Read Blank Follow-On
3 The instrument will be set automatically. After a few seconds the following
display will appear:
Place SAMPLE TUBE in the test chamber, then press [OK].
4 The instrument will take the reading and display the result as follows, for
example:
The following symbols indicate the result is out of test range:
5 The ‘options menu' offers the choice to:‘Choose a Test’ - return to the menu of test programs and select another
‘Read’- read further sample tubes of the currently selected test
‘Blank’- re-blank the instrument
‘Follow-On’- carry out a continuation test if available.
Result is higher than range: >>
Result is lower than range: <<
test
15
Chlorine-Total / 5
Phot 008
1.50
mg/l Cl
2
Choose a Test Read Blank
Return
Continuation Tests (Certain Tests Only)
1 Select ‘Follow-On’ and press [OK] while the result is displayed of the
currently running test. The photometer applies the previously entered
sample number and dilution factor, and the ‘Insert Sample’ screen will
appear.
Place SAMPLE tube in the test chamber, then press [OK].
2 The instrument will take the reading and calculate the result from the
combination of readings (where appropriate). The result will be displayed
as follows, for example:
3 While the test result is displayed, similar options are available as at the
end of a normal test program. In order to run more samples for the same
parameters, select ‘Return’ from the ‘options menu' to take the program
back to the start of the first stage of a multiple test procedure.
Note: some continuation test procedures involve a standing period. The
photometer may switch off automatically during this time. To avoid the
instrument switching off, set for continuous operation or use the timer
function to time any standing period. See Timer section. The timer will
over-ride the auto switch off function.
Favorite Tests List
The four most recently used tests are listed at the top of the 'Choose a Test'
screen for convenience.
Expressing Different Chemical Forms
If the test result can be expressed in different chemical forms, the chemical
symbol will have flashing and to indicate this. Use the and keys to
step through the options available.
Note that the log stores the result in the primary form.
16
T i m er
00 : 00
EXIT
Reading in Transmittance and Absorbance
When taking readings in Transmittance or Absorbance mode, use the and
keys to step through the wavelengths until the required wavelength is
reached.
Timer
The photometer features a countdown timer with alarm as an aid to carrying
out test procedures. The timer can be accessed at any time by selecting
‘Timer’ from the ‘Options’ menu.
The following display will appear:
Key in the time required in minutes and seconds using the numerical keys,
then press [OK] to start the timer. The maximum time is 29 minutes and
59 seconds. Use the and keys to reposition the cursor and re-enter the
time if it is keyed in incorrectly.
The timer will count down, giving an audible alarm at the end of the timed
period. Press [OK] to stop the alarm.
During the timer countdown period, an ‘Options’ menu is available :
Stop - used to abort the timing operation or to stop the alarm at
the end of the timed period
Exit - used to return to the program screen to take readings. The
timer will continue to run and give an audible alarm at the
end of the period.
Exit and Read - used to return to the program screen with the timer counting
down on screen. The instrument will automatically take a
reading at the end of the timed period and no alarm will
sound.
User Defined Tests (9500 Only)
Users may wish to develop their own test methods and store the calibration
data on the 9500. This will allow direct reading of user tests. The 9500 has
the facility to store up to 30 user-defined calibrations.
17
To program user-defined calibrations:
Turn the instrument ‘ON’, select ‘System’ menu and press [OK]. Scroll through
the options to the USB entry and make sure the option is set to ‘Com Port’.
At the PC, open the HyperTerminal connection for the 9500 (contact YSI
Technical Support Department to receive the virtual comport drivers for
installation to PC).
At the instrument, in the ‘System’ menu, select 'Edit User Defined Tests' and
press [OK]. The instrument will display the tests already downloaded, or show
‘LIST EMPTY’.
In the ‘Options’ menu, select [Add] to add a new test, or [Edit] to edit the test
which is currently highlighted. Change the highlighted test with
Press [OK].
The instrument displays a message box instructing the user to download the
new or edited test file.
At the PC, download the calibration file from HyperTerminal using 'Transfer',
'Send Text File' and select the file to be downloaded.
The instrument will check the downloaded data. If it is acceptable, it will
display a message box ‘Accepted’ over the downloaded data. If there are
errors in the file, a list of errors will be displayed. The user should edit the
calibration file to correct the errors then re-send it.
Press [OK] to accept the test. The instrument will change to the ‘User Test
List’ screen (See 3), with a summary of the test displayed. Press [OK] to
accept the test and write to memory. Select [Cancel] at any time to reject the
calibration.
The format of the user calibration file is as follows:-
The file must start with ‘USER CALIBRATION’ and end with ‘END’.
Test Number must be between 900-929 (30 tests).
Test name, up to 18 characters.
Units. Must be one of the following - mg/l, ppm, mmol/l, µmol/l or g/l.
Wavelength. Must be one of the following – 450 nm, 500 nm, 550 nm,
570 nm, 600 nm or 650 nm.
Chemical symbols: up to 8 characters.
Data pairs. Up to 10 pairs of data in the form :-
ABSORBANCE x.xxx, CONCENTRATION
(Concentration may be up to five digits).
18
An example is shown below:
USER CALIBRATION
900
Chlorine
mmol/l
500nm
Cl2
0,0
0.174,0.50
0.481,1.50
0.733,2.50
0.854,3.00
1.022,4.00
1.086,4.50
1.187,5.00
END
19
6 CARE AND MAINTENANCE
The photometer is designed to give long and trouble-free operation. Care
must be taken, however, to avoid test solutions being spilt over the
instrument, and to prevent contamination of the instrument. Spillages or
moisture should be wiped off immediately with a dry cloth. Never use solvents
or abrasive materials to clean the instrument. Care should be taken to keep
the test chamber clean.
Cleaning the Optics
Any build-up of dirt or deposits may interrupt light transmission and affect
readings.
To clean the optics, undo the two screws to remove the optics base plate.
Gently clean the internal surfaces of the optics with a soft, non-abrasive cloth.
Deposits may be removed with a slightly dampened cotton bud. Replace the
optics base plate and re-fasten the screws.
The photometer is fitted with long-life light sources and contains no userserviceable components. If the instrument requires servicing or repair, contact
YSI Technical Support Department.
Service Requirement
Servicing of the photometer is essential to ensure optimum performance. To
arrange service of the instrument, contact YSI Technical Support Department
or the authorized distributor who supplied the instrument. The YSI standard
photometer service includes cleaning of the optical assembly, replacement of
any worn parts and checking/recalibration of the instrument.
Error Messages
The photometer will display an error message in the unlikely event of a
malfunction. These error messages are mainly designed to assist service staff
in diagnosing instrument faults. In the event an error message appears on the
photometer, contact YSI Technical Support Department for advice.
Error messages are coded 7 and 9 and both relate to blanking the instrument.
If you see one of these error messages, check the operating technique and
sample clarity. If you continue to get error messages, try the following:
Error 7 indicates too much light – remove the instrument from bright light and
use the light cap.
Error 9 indicates not enough light – follow ‘Cleaning the Optics’ routine.
20
Photometer Up-Grade (9500 only)
It is now possible to upgrade the photometer with new test calibrations using
a computer. This will ensure that users can always keep the instrument up-todate with the latest tests. Contact YSI to request an update at
environmental@ysi.com. No special computer software is required. Full
instructions will be supplied with the upgrade data.
Computer Controlled Operation (9500 Only)
The photometer can be controlled from a computer using suitable control
software. Such software is available from software houses or from water
treatment specialists to cover specific applications. These software programs
typically instruct the photometer to go through a predetermined series of tests
specific to that application, and then automatically receive data from the
photometer and process the test results. The internal software of the
photometer is able to receive computer commands to start new sample,
receive test program number, receive sample number and instigate
continuation test. Programmers requiring further details should contact YSI
Technical Support.
Warranty
The YSI photometers are warranted for one (1) year from date of purchase by
the end user against defects in materials and workmanship, exclusive of
batteries and any damaged caused by defective batteries. Within the warranty
period, YSI will repair or replace, at its sole discretion, free of charge, any
product that YSI determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or
contact YSI Customer Service at +1 937-767-7241 (800-897-4151). Send the
product and proof of purchase, transportation prepaid, to the Authorized
Service Center selected by YSI. Repair or replacement will be made and the
product returned, transportation prepaid. Repaired or replaced products are
warranted for the balance of the original warranty period, or at least 90 days
from date of repair or replacement.
21
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by:
1 Failure to install, operate or use the product in accordance with YSI's
written instructions;
2 Abuse or misuse of the product;
3 Failure to maintain the product in accordance with YSI's written instructions or
standard industry procedure;
4 Any improper repairs to the product;
5 Use by you of defective or improper components or parts in servicing or
repairing the product;
6 Modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE. YSI's LIABILITY UNDER THIS WARRANTY IS
LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL
BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT
COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING
FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.
Contact Information
For Customer Service, Technical and Ordering Support:
Telephone: (800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Natural and treated waters may contain a variety of dissolved alkaline
substances such as carbonates, bicarbonates, hydroxides and, to a lesser
extent, borates, phosphates and silicates. In water at neutral pH the alkalinity
derives mainly from the presence of bicarbonates.
Total alkalinity is an important test in determining the aggressiveness or scale
forming tendency of the water. If the total alkalinity is low the water may be
aggressive and cause corrosion to pipe work and structures; if the total
alkalinity is high the water may more readily promote scale formation. Alkalinity
control is therefore an important part of many water treatment programmes.
The YSI Alkaphot test uses a colorimetric method and covers the total alkalinity
range 0 - 500 mg/l CaCO3. The test is particularly suitable for checking natural
and drinking waters, swimming pool water, boiler water, etc.
Method
The YSI Alkaphot test is based on a unique colorimetric method and uses a
single tablet reagent. The test is simply carried out by adding a tablet to a
sample of the water. Under the conditions of the test, a distinctive range of
colors from yellow, through green, to blue is produced over the alkalinity range
0 - 500 mg/l CaCO3. The color produced in the test is indicative of the alkalinity
of the water and is measured using a YSI Photometer.
Reagents and Equipment
YSI Alkaphot Tablets
YSI 9300 OR 9500 Photometer
Round Test Tubes, 10 ml glass (PT 595)
Test Procedure
1 Fill test tube with sample to the 10 ml mark.
2 Add one Alkaphot tablet, crush and mix until all of the particles have dissolved.
3 Stand for one minute then remix.
4 Select Phot 2 on photometer.
5 Take photometer reading in usual manner (see photometer instructions).
6 The result is displayed as mg/l CaCO3.
Note: To convert Total Alkalinity as CaCO3 to Total Alkalinity as HCO
result by 1.22.
____________________________
multiply
3-
Loading...
+ 113 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.