Yokogawa RIO4000 Data sheet

RIO4000 Sensor Model
General Specifications
The RIO4000 is the RTD measure and simulate module for the Yokogawa accurately measures or simulates the resistance of 22 different RTD sensor types with nominal resistance to 4000 ohms. 2-, 3-, and 4­wire RTDs are supported and RTD material may be platinum, copper, nickel or iron. Several alpha coefficients are supported per Table 1 on page 2. The module can simulate temperature to all conventional or smart RTD transmitters and signal conditioners including those utilizing pulsed excitation current.
The YPC converts measured resistance to a corresponding temperature value and displays the value in the user-selected temperature unit (°F, °C, °R or °K). Ohms can also be displayed for direct reading or troubleshooting assistance. Pt 100 RTDs are available from Yokogawa for use with the module. These 4-wire, DIN Class A probes come complete with handle, 5’ coiled cable, and connector. Accessory connectors are available to connect to any RTD or RTD transmitter for temperature measurement or simulation needs. Simulations are supported through setup controls that allow the user to enter a specific temperature, a range of temperatures stepped through with manual or automatic advance or a temperature range with automatic ramping advance for a selected time period.
YPC 4000 Series Multi-Function Tester. The module
RTD Measurement & Simulation Module for Use with
YPC4000 & YPC 4010
RIO4000 Sensor Module with
Connected Pt 100 Probe
Field recalibration of the module is provided by the YPC’s “Rcal” feature. “Rcal” can be used to re-calibrate the module alone or the module with dedicated RTD for “match set” accuracy. The module + RTD option provides the best possible measurement accuracy for the most demanding measurement applications.
The RIO4000 and a dedicated DIN Class A RTD makes an excellent temperature reference standard for comparison to RTDs installed in process applications. The RIO4000 and user cable
provides convenient RTD measurement and simulation capability.
RIO4000 Sensor Module with
Connected Pt 100 Probe
Specifications for RIO4000 Module
0-400Ω measurement accuracy: ±0.01% of R + 0.075Ω 0-400Ω simulation accuracy: See Table 2 400-4000Ω meas. accuracy: ±0.01% of R + 0.237Ω 400-4000Ω simulation accuracy: See Table 3 Temperature:
Effect: 0.001 ohm/ ºC Storage: -40° F to 140° F (-40° C to 60° C)
Operating: 23° F to 122°F (-5° C to 50° C) Mating Connector: Switchcraft #TA6FL Pulsed excitation current support
Certifications Available
• CE Mark (standard)
• NIST traceability certificate (standard)
• Intrinsically Safe, MET Laboratories per CSA C22.2 & UL Class I Division I (standard)
Units: °F, °C, °R or °K, Ohms Resolution: ± 0.1° or ± 1° for RTDs Resistance range: 0 – 4000 ohms Input impedance: > 10 megohms Weight: 3 oz
Yokogawa Corporation of America GS 61A YPCRIO-01E-A Dart Road, Newnan, GA 30265, USA ©Copyright November 2004 Tel: 1-800-888-6400 1st Edition November 2004
Material
R0
ALPHA
Rs (ID)
Temp Range (deg C)
Platinum
25.5
0.00392
17.8
-200 to +630° C Platinum
98.129
0.00392
21.5
-200 to +600° C Platinum
100
0.00385
100
-200 to +850° C Platinum
100
0.00392
46.4
-200 to +630° C Platinum
100
0.00391
56.2
-200 to +630° C Platinum
100
0.00393
68.1
-200 to +962° C Platinum
100
0.00389
82.5
-200 to +630° C Platinum
200
0.00385
178
-200 to +630° C Platinum
200
0.00392
215
-200 to +630° C Platinum
470
0.00392
261
-200 to +630° C Platinum
500
0.00392
316
-200 to +630° C Platinum
500
0.00391
383
-200 to +630° C Platinum
500
0.00385
464
-200 to +630° C Platinum
1000
0.00385
1000
-200 to +630° C Platinum
1000
0.00375
1210
- 50 to +500° C Copper
9.035
0.00427
12.1
-100 to +260° C Copper
100
0.00427
121
-100 to +260° C Nickel
100
0.00618
31.6
- 60 to +160° C Nickel
120
0.00672
147
- 80 to +260° C Iron
604
0.00518
562
-100 to +200° C Iron
908.4
0.00527
681
-100 to +200° C Iron
1816.81
0.00527
1470
-100 to +200° C
Table 1: Supported RTD Materials / Nominal Resistance
The adjacent table lists the RTDs supported by the RIO4000 module and YPC 4000 Series. Materials, types and alpha coefficients are based on popular RTDs used in the field.
Yokogawa offers Pt 100 accessory probes that are automatically recognized by the RIO4000 module. No set up is necessary on the YPC. Users can configure their own reference RTDs similarly by using the accessory connector and Rs resistor (Probe ID resistor listed in adjacent table) to connect their RTD. This technique allows the module to automatically recognize the connected RTD and recall its characteristics. If no Rs resistor is used, a copper wire must be installed in its place. This causes an RTD selection menu to automatically appear on the YPC display, enabling the user to configure the YPC for use with any supported RTD type.
Accessories
P/N A900028-90200 connector solders to a customer’s RTD wiring to
facilitate connection to the RIO4000 module
P/N A900028-90500 general-purpose connector/cable to connect the
RIO4000 module to RTD wires or transmitter terminals
P/N A900028-90201 Pt 100 RTD, DIN Class "A", 0.00385 alpha,
4 leads, 1/4" O.D., 304SS probe, 10" probe length, 3.5" handle and 5 ft coiled cable with connector for RPT4000 module
All Rights Reserved ©Copyright November 2004, Yokogawa Corporation of America GS 61A YPCRIO-01E-A Subject to change without notice.