Limiting-current type
Oxygen Analyzer
OX100, OX102
OX100, OX102
Limiting-current type
Oxygen Analyzer
• One Instrument measures 100 ppm to 25% Oxygen
• Simple "One Touch" Calibration
• Self Diagnostics Assure High Reliability
• "One Touch" Sensor Replacement
• Compact Size
• Accessories to Assure Accuracy
• Dedicated Sampling Unit (Separate from Oxygen Analyzer)
(Removes Dust, Mist and Organic Solvents)
• Calibration Kit
Simple On-Site Calibration
(Using 1 liter Gas Cylinder)
Bulletin 11M10A01-03E
www.yokogawa.com/an/
Low range 0 to 100 ppm
Autoranging in four ranges from 0-100 ppm to 0-25%
High and Low Alarms
Spans 1000 ppm to 25%
Autoranging in two ranges, 0-1000 ppm and 0-25%
High Alarm or Low Alarm (depends on specification)
Oxygen
concentration
FAIL lamp
High, Low
Alarm lamps
Alarm
Fixed-range
4-20 mA / 1-5 V,
Autorange 4-20 mA,
keys
Calibration keys
Sensor holder: 17 mm
(* 1)
dia. or G3/8
Sensor
(* 2)
Plastic sensor holder
(Specify Option code /J
or L9811LA
(* 3)
)
Sensor plus cable: 5 m (OX100), 5, 10 or 30 m (OX102)
OX100 and OX102 sensors are not interchangeable
1 : When mount in 17 mm hole in panel, panel thickness can be up to 4 mm. 2 : Within 5 cm of sensor tip, can reach 250°C.
3 : Specify Option code /S, or Model K9424DB, for metal nipple.
Lit when range
is 1.0 to 25%
between 0 to 1.0% and
0 to 25% other than
0 to 10,000ppm
Lit when range is
0 to 10,000 ppm
Validates
parameter entry
High alarm or
Low alarm lamp
(depends on specification)
Oxygen
concentration
Calibration air
flow lamp
Lit during
alarm setting
Alarm setting key
Mode switching key
Main applications
Globe box (processing crystal parts)
Globe box (in research lab.)
2
reflow furnace / auto soldering
N
Clean room
Aging machine
Monitoring for deficiency of O
Foodstuffs
Measurement cautions * 1 : Sensor should not be exposed to silicone vapor, silicon gas.
2
* 2 : Corrosive or flammable gases, halogens, Freon, or organic solvents may adversely affect sensor.
Box is to be filled with inert gas
If organic solvents in box, use sampling unit
Use sampling unit, remove alcohol / flux from furnace
Use sampling unit if organic solvents present
High temperatures, so use metal nipple K9424DB
Use OX102 with Hi/Lo Alarms, or OX100-3 (Lo)
Use sampling unit if disinfectant is alcohol-based
Fail lamp
Units (% or ppm)
Lit when range is
0 to 10,000 ppm
not when range is
0 to 25%
Lit during 0 to
1000 ppm calibration
Entry setting key
Usage cautions
Example: N2 reflow furnace
Temperature : 120-130°C
Gas flow : 1m/sec.
Environment : Alcohol / Rosin / Flux,
Oxygen 100 ppm
Compared with oxygen
meter it replaced,
Response was faster, so changes were
quickly detected
Easy maintenance for Sampling unit
(easy replacement of filter, activated carbon)
10005000
Oxygen ppm
Fast response
Solder flux flow start
Approx. 6 sec.
A
OX102 oxygen analyzer
Installed oxygen analyzer
B
A
Difference
B
Example: Measuring oxygen
concentration in welding furnace
Temperature : Room temperature,
atmospheric pressure
Gas flow : 30 Nl/min. (N
Environment : Oxygen about 500 ppm in N
Low oxygen concentration in furnace:
safety precautions required
OX100 can detect atmospheric air leakage into
furnace which causes oxygen to rise to 200 ppm,
so can help with welding quality assurance.
2
)
2
200ppm
10005000
Air leak
Oxygen ppm
20mm/hr
B
A
A
OX100 oxygen analyzer
Dew point measurement
B