This user’s manual consists of the chapters listed below. For information about the
communication features and the accompanying software DAQSTANDARD, read the
respective manual.
ChapterTitle and Contents
1Feature Overview
The MV features.
2Installation and Wiring
How to install the MV, how to wire input terminals, and how to connect the power
cable.
3Measurement Channels and Alarms
How to set measurement conditions and how to set alarms.
4Measurement and Recording
How to record and store measured values and other values. This chapter also
explains how to load measured data or setup data from a CF card or USB flash
memory.
5Screen Operations
How to use the operation screen.
6Display Configuration
How to change the displayed information on the operation screen and how to write
a message.
7Event Action
How to make the MV perform a specific action in response to an event, a remote
control signal input, or a pressing of the USER key.
8Security Features
How to use the key lock feature and the feature that allows only registered users to
operate the MV.
9Environment Settings
How to configure the time settings and how to operate the MV by using a keyboard.
10Computation and Report Functions (/M1 and /PM1 options)
How to use computation channels and how to create various reports, such as
hourly, daily, weekly, and monthly reports.
11External Input Channels (/MC1 option)
How to use external input channels.
12Troubleshooting and Maintenance
Explains error messages and troubleshooting measures.
13Specifications
Lists the MV specifications.
Appendix Describes measured data file sizes, text file formats, etc.
Index
• This user’s manual covers information about MVs that have a suffix code for language “-2”
(English).
• For language configuration instructions, see section 9.4, “Changing the Language.”
i
The following symbols are used in this manual.
Unit
Denotes 1024. Example: 768 KB (file size)
Denotes 1000.
Markings
WARNING
Procedure
Display
Settings
Explanation
CAUTION
Note
Indicates character types that can be used.
Carry out the procedure according to the step numbers.
All procedures are written with inexperienced users in mind;
depending on the operation, not all steps need to be taken.
Explanation gives information such as limitations related the
procedure.
Indicates the display and explains the settings. A detailed
description of the function is not provided in this section.
For details on the function, see chapter 1.
Subheadings
A a # 1
A a
#
1
Uppercase alphabet, lowercase alphabet, symbols,
numbers.
Bold characters
Denotes key or character strings that appear on the screen.
Example: Volt
K
k
Improper handling or use can lead to injury to the user or damage
to the instrument. This symbol appears on the instrument to
indicate that the user must refer to the user's manual for special
instructions. The same symbol appears in the corresponding
place in the user's manual to identify those instructions. In the
manual, the symbol is used in conjunction with the word
“WARNING” or “CAUTION.”
Calls attention to actions or conditions that could cause serious
or fatal injury to the user, and precautions that can be taken to
prevent such occurrences.
Calls attention to actions or conditions that could cause light
injury to the user or damage to the instrument or user’s data, and
precautions that can be taken to prevent such occurrences.
Calls attention to information that is important for proper
operation of the instrument.
This mark is used to indicate a reference to a related procedure or
explanation.
Procedural Explanations
This manual mainly describes the MV1000 procedures. Where procedures differ between
the MV2000 and MV1000, the MV2000 procedures (menu operations) are also provided.
High-Speed and Medium-Speed Input Model Groupings
This manual uses high-speed input model and medium-speed input model to distinguish
between MV models as follows:
Appendix 3 Text File Formats...............................................................................................App-7
Index
vi
M-4660
1
Chapter 1 Feature Overview
Ethernet
(Communication Interface User’s Manual)
USB port
Alarm output*
Sections 3.5 to 3.8
Serial communication*
(Communication Interface User’s Manual)
PC
PC
MV
RecorderTemperature
regulator
FAIL/status output
Section 9.6
Remote input*
Section 7.4
Keys
Keyboard
Section 9.7
CF card
First Step Guide
Transmitter power supply*
Section 2.3
Pulse input*
Section 3.10
Measurement
input
Section 3.3
USB flash memory
Section 4.10
* Options
1.1 System Overview
System Configuration
The MV can be used to construct a system like the one shown below. The MV can
make voltage and temperature measurements by itself. It can also transfer data to other
devices via Ethernet or a serial interface, and it can acquire data from multiple channels
using the Modbus protocol.
Feature Overview
Terminology
• Memorysampling
Recording measured data.
• Memorystart
A command to start memory sampling.
• Memorystop
A command to stop memory sampling.
• Displaydata
The waveform data displayed on the MV screen. It is essentially measured data that is
recorded at the display data sampling rate.
• Eventdata
Measured data that is recorded at a set sampling rate that differs from that of the
display data.
M-4660
1-1
Power ON
Basic Setting Mode
Select Basic setting mode from the menu,
or hold down FUNC for three seconds.
Select End, or press ESC then DISP/ENTER.
Setting Mode
Operation Mode
MENU
MENU or
ESC
Mode Transition Diagram
MENU or ESC
DISP/ENTER
Configuration
screen
DISP/ENTER
ESC
Configuration
screen
Settings menu
Basic settings
menu
Operation
screen
1.1 System Overview
Modes
The MV has an Operation Mode, a Setting Mode, and a Basic SettingMode.
ModeDescription
Operation ModeFor performing measurements.
Setting ModeFor configuring settings, such as the input range and the
measurement method. You can change most of the settings in this
mode while the MV is engaged in memory sampling.
Basic Setting ModeFor configuring fundamental settings, such as the scan interval and
the measured data save method. You cannot switch to this mode
while the MV is engaged in memory sampling.
*For a list of the contents of Setting Mode and Basic Setting Mode, see the First Step Guide.
1-2
M-4660
1
1.2 Input
Measurement Channels
• Number of Channels vs. Scan Interval
The MV acquires data by sampling measurement channel input signals at the set scan
interval. The following is a list of the scan intervals available with different numbers of
channels.
The MV uses an A/D converter to convert sampled analog signals to digital signals.
You can effectively eliminate the inuence of power supply noise by setting the A/
D converter’s integration time to the same length as or to an integral multiple of the
power source period.
• Because 100 ms is an integral multiple of both 16.7 ms and 20 ms, you can effectively
eliminate the influence of power supply noise from both 50- and 60-Hz frequencies by
setting the integration time to 100 ms.
• Power supply noise elimination is less effective in fast sampling mode than it is in normal
mode. When measuring in an environment susceptible to power supply noise, we
recommend that you measure in normal mode.
For conguration instructions, see section 3.1.
Input Types and Computation Functions
You can measure the following types of input.
Input TypeDescription
DC voltageYou can measure DC voltages in the range of ±20 mV to ±50 V.
M-4660
DC currentYou can measure a DC current signal by converting it to a voltage signal using a shunt
resistor attached to the input terminal.
The converted signal can be measured within the DC voltage range (see above).
Thermocouple You can measure temperatures using these thermocouple types: R, S, B, K, E, J, T, N,
W, L, U, and WRe3-25. It is also possible to measure using other thermocouples, such as
PR40-20 and PLATINEL.
RTDYou can measure temperatures using RTD types Pt100 and JPt100. It is also possible to
measure using other RTD types such as Cu10 and Cu25,3 and Pt50 and Ni100.
ON/OFF inputYou can display contact input or voltage input signals correlated to 0% or 100% of the
display range.
Contact input: A closed contact is ON (1). An open contact is OFF (0).
Voltage input: Less than 2.4 V is OFF (0). 2.4 V or more is ON (1).
Pulse input
1 Shunt resistor sold separately. For example, you can use a 250-Ω shunt resistor to convert a 4- to 20-mA
signal to a 1- to 5-V signal.
2 /N3 option
3 /N1 option
4 /PM1 option
4
You can count pulses.
2
1
2
1-3
DC voltage
DC current
Shunt resistor
1 - 5 V
Burnout detection
Burnout detection
RJC
Square root computation
Thermocouple
RTD
Pulse
(/PM1 option)
ON/OFF input
Convert to number of pulses per second
Moving average Linear
scaling
Calibration
correction
(/CC1 option)
Difference
computation
Low cut Measured
value
Dotted lines indicate that the function can be selected.
√
Measured
value
Moving Average
External reference junction
Internal
reference
junction
Performs reference junction compensation on
a thermocouple.
Reference Junction Compensation
Burnout Detection
Detects and indicates a sensor burnout.
1 - 5 V
Detected on the
1-5 V value.
Burnout
For configuration instructions, see section 3.2.For configuration instructions, see section 3.2.
For configuration instructions, see section 3.4.
Burnout
Indicated as
Burnout.
Indicated as
Burnout.
Thermoelectromotive
force
1
2
3
10.0 mV
5.0 mV
0.0 mV
5.0 mV
15.0 mV
10.0 mV
5.0 mV
10.0 mV
10.0 mV
15.0 mV
10.0 mV
11.7 mV
Clear
n
th
time n + 1th time n + 2th time
Clear
(Moving average)
New sampled
data
New sampled
data
Operation example when the number
of moving average data points is 3
Number of moving
average data points: 2 to 400
Sampling data
in the buffer
Sampling
Eliminates noise.
MV
Thermocouple
Thermocouple
1.2 Input
The following input processing and math operations are available.
1-4
M-4660
1
Difference Computation
Low-cut
Square Root Computation
Calibration Correction (/CC1 Option)
Input
value
Measured
value
Measured
value
Input
value
Input
value
Input
value
Input
value
Measured
value
Measured
value
Measured value on the reference channel
–
The measured value of the channel is set to the
difference with respect to the measured value of the
reference channel.
Converts the unit to obtain the measured value.
Takes the square root of the input value and converts
the unit to obtain the measured value.
Corrects the input value with the characteristics
specified by segments to obtain the measured value.
Number of break
points: Up to 16
Pulse Input (/PM1 Option)
Linear Scaling
10 V
0 V
300 °c
–100 °c
a
b
x
a
b
B
B
X
A
A
X = (B – A)
+ A
x – a
b – a
The contact changes from open to closed.
The signal level at the input terminal changes from high
to low.
5 ms or more
Contact
or
Open
collector
Pulses of 100 Hz or less
can be counted
Counts pulses.
Input to the MV
Count on the MV
Measured
value
Output
value
For square root computation, measured values below
the specified value are cut.
For 1-5 V input, values below 0 % are cut.
Result of square root
computation
Low-cut value
MV
For configuration instructions, see section 3.3.For configuration instructions, see section 3.3.
For configuration instructions, see section 3.9.For configuration instructions, see section 3.3.
For configuration instructions, see section 3.3.For configuration instructions, see section 3.10.
( )
1.2 Input
Feature Overview
Note
You can use difference computation even if the reference channel and difference computation
channel’s input types and ranges are different. Difference computation takes place using the
difference computation channel’s decimal place and unit.
Example 1: If the input value of the difference computation channel is 10.00 and the measured
value of the reference channel is 100.0, the result is 10.00–100.0=–90.00.
Example 2: If the input value of the difference computation channel is 10.00 V and the
measured value of the reference channel is 5.00 mV, the computed result is 10.00
V – 5.00 mV = 5.00 V.
M-4660
1-5
1.3 Alarms
High Limit Alarm (H)
Alarm value
Alarm output ON Alarm output ON
Alarm output ON Alarm output ON
Alarm value
Hysteresis Measured value Measured value Hysteresis
Low Limit Alarm (L)
Alarm output ON Alarm output ON
Delay time Delay time
Alarm value
Alarm value
Measured value Measured value
Delay High Limit Alarm (T) Delay Low Limit Alarm (t)
High Limit on Rate-of-Change Alarm (R) Low Limit on Rate-of-Change Alarm (r)
Alarm value
Alarm value
Hysteresis Hysteresis
Difference in the measured values of two channels Difference in the measured values of two channels
Difference High Limit Alarm (h)Difference Low Limit Alarm (l)
Change in the
measured value
Change in the measured value
Measured value
Measured value
Time Time
Interval (t2-t1)
Interval (t2-t1)
Amount of change in
the setting
| |
T2
T1
t1 t2
T2
T1
t1 t2
T2-T1
Amount of change in
the setting
|
T2-T1
|
Alarm Types
This function generates an alarm when the measured data meets a set condition. You
can set a maximum of four different alarms on each channel.
You can use the types of alarms listed below. The letters contained in parentheses are
used as abbreviations for the alarms.
For alarm conguration instructions, see section 3.7.
• Hysteresis
You can set a difference between the value that activates an alarm and the value that
deactivates it.
For conguration instructions, see section 3.5.
• Delay High Limit Alarm and Delay Low Limit Alarm
If the measured value remains above or below the set alarm value for the set period of
time (the delay time), an alarm is activated.
• High Limit on Rate-of-Change Alarm and Low Limit on Rate-of-Change
1-6
Alarm
The MV checks the rate-of-change of the measured values over a set period of time (the
interval) and activates an alarm if the rate-of-change in the rising or falling direction is
greater than or equal to the set value.
The alarm value of the rate-of-change alarm is set using an absolute value. The
interval is derived from the sampling count using the following equation:
Interval = scan interval × sampling count
For instructions on how to set the interval, see section 3.5.
• Difference High Limit Alarm and Difference Low Limit Alarm
The MV activates an alarm when the difference between the values of two channels
goes above or below the set value. You can use this alarm on measurement channels
set to difference computation.
M-4660
1
Alarm display
Alarm
Alarm output relay
(reflash ON)
Alarm output relay
(reflash OFF)
Channel 1
Channel 3
Channel 2
Approx.
500 ms
Approx.
500 ms
Energize or De-energize
Non-HoldHold
AND/OR
Alarm
Channel 1
AND
Channel 2
OR
Alarm output relay
or
Internal switch
NO
Energize
De-energize
C NCNO C NCNO C NC
NO C NCNO C NCNO C NC
When power is shut down
NO: Normally Opened
C: Common
NC: Normally Closed
Normal operationWhen alarm is occurring
Alarm
Alarm occurrence
Alarm occurrence
Alarm release
Alarm release
Alarm
output
relay
Alarm
Alarm
output
relay
AlarmACK
AlarmACK
or
or
or
Activated
Activated
Deactivated
Deactivated
Activated
Deactivated
AlarmACK
Activated
Deactivated
Activated
Deactivated
AlarmACK
Activated
Deactivated
AlarmACK
Activated
Deactivated
Reflash
(You can set the alarm output relay and internal
switch. For details on the internal switch, see the
next page.)
(when a relay is set to “OR” logic)
Relay is activated
at the next scan
interval
Relay is activated
at the next alarm
occurrence
Normal
Reset
Relay Action on AlarmACK
Alarm conditions are indicated in operation windows (trend, digital, bar graph, overview,
etc.) and by icons in the status display section. The alarm summary displays detailed
alarm information.
• Hold/NonholdofIndication
When alarm conditions cease, their indicators can be set to either:
• Stop as soon as the alarm condition ceases (Nonhold).
• Continue until the execution of an alarm output release (AlarmACK) operation (Hold).
The default setting is Nonhold.
For conguration instructions, see Section 3.5.
• Alarm Hide Function
If this function is enabled, the MV will not display alarm condition information or record
it in the alarm summary. However, the MV will output alarm information to a relay (/A
options) or internal switch. You can set this function for each channel and alarm.
For conguration instructions, see Section 3.6.
Alarm Output Relay Operation
The MV can generate a contact signal from an alarm output relay (/A options) when an
alarm occurs. The alarm output relay operation can be changed.
For conguration instructions, see Section 3.5.
1.3 Alarms
Feature Overview
M-4660
1-7
Alarm
Alarm occurrence
Alarm release
Internal switch
1
0
1.3 Alarms
• Reflash
This function enables a single alarm output relay with multiple alarms assigned to it to
indicate all alarm occurrences. After the rst alarm, subsequent alarms are indicated
by the brief release (approximately 500 ms) of the output relay.
The reash function affects the rst three output relays.*
* I01 to I03 or I11 to I13. With the /A1 option, I01 and I02.
Note
When reflash is enabled, the first three output relays are used exclusively as reflash relays.
The first three output relays are set to OR logic and to the Nonhold operation regardless of the
AND/OR or Nonhold/Hold settings explained below.
• AND/OR
When multiple alarms are assigned to one output relay, you can choose to activate the
alarm in one of the following ways (you can also use AND/OR with internal switches):
• AND: Activate the relay when all of the alarms assigned to it occur simultaneously.
• OR: Activate the relay when any of the alarms assigned to it occur.
• EnergizeorDe-energizeOperation
You can choose whether to energize or de-energize alarm output relays when an
alarm occurs. If you select de-energize, the alarm output relays will be in the same
state when the power is shut off as they are when an alarm occurs. This setting
applies to all alarm output relays.
Internal Switch
• Nonhold/Hold
When an alarm condition is no longer met, alarm relays can be set to:
• Turn OFF immediately (Nonhold).
• Remain ON until the execution of an alarm output release (AlarmACK) operation (Hold).
This setting applies to all alarm output relays.
• Alarm Output Release Operation
The alarm output release operation (AlarmACK) stops the display and the relay output
of all alarms. For information about how alarm display and output relay operations
respond to an AlarmACK operation, see the previous page.
Note
When you enter Basic Setting Mode, the alarm output relay remains at its current active/
released state. In Basic Setting Mode, alarms are not detected and the AlarmACK operation is
invalid.
The MV can transmit the alarm status to software switches (30 internal switches).
Internal switch values are shown below. As with AlarmACK, you can apply an AND or OR
operation (see the previous page).
1-8
Internal switches can be used as events with the event action function (for details, see
section 1.7). Internal switches can also be written into computation channel (/M1 and /
PM1 options) expressions.
M-4660
1
1.4 Display
Status display section
Data display section
Common Display functions
• LCD Display and Its Layout
The RD-MV1000 has a 5.5-inch (240 × 320 dot resolution) TFT color LCD. The RD-
MV2000 has a 10.4-inch (480 × 640 dot resolution) TFT color LCD. The display
consists of a status display section and a data display section.
The RD-MV1000 Display
• StatusDisplaySection
The status display section indicates the display name, date/time, batch name
(when using the batch function), user name (when using the login function), internal
memory and CF card usage, alarm occurrence, computation status (/M1 or /PM1
option), key lock status, and e-mail transmission.
• DataDisplaySection
The data display section shows measured data using numbers, waveforms, and
bar graphs. It displays a conguration screen when you are conguring a function.
Feature Overview
• Group Display
On the trend, digital, and bar graph displays, channel data is displayed in preset
groups. With the RD-MV1000, you can register up to 10 groups, each with up to 6
channels. With the RD-MV2000, you can register up to 36 groups, each with up to 10
channels. The same groups are used for the trend, digital, and bar graph displays.
Displayed groups can be switched automatically at specied intervals (5 s to 1 min).
For conguration instructions, see section 6.1.
• ChannelNumberDisplayandTagNameDisplay
You can choose to label displayed channels according to their tag names or according
to their channel numbers. This setting applies to all channels.
For conguration instructions, see section 6.2.
• Update Interval of Measured Values
Measured values are updated every second. However, if the scan interval is longer
than 1 s, measured values are updated at the scan interval.
For conguration instructions, see section 6.3.
• AlarmIndication
The MV regularly checks for the alarms set to each channel and indicates alarms with
the following symbols:
Alarm TypeSymbolAlarm TypeSymbol
High limit alarmHHigh limit on rate-of-change alarmR
Low limit alarmLLow limit on rate-of-change alarmr
Difference high limit alarmhDelay high limit alarmT
Difference low limit alarmlDelay low limit alarmt
M-4660
1-9
Memory Sampling Status
At least one alarm condition has been met.
This icon blinks when an alarm condition is
met and AlarmACK has not been executed.
After an alarm condition has been met, all
alarms have been released, but there is still
at least one alarm for which an AlarmACK
operation has not been performed.
The condition set to the status output
relay (/F1 option) has been met.
CF Card Icons
Key lock is on.
E-mail notification has been activated.
White icon: Math in progress
Yellow icon: Math data dropout
Waiting.
Light blue icon: There is a CF card in the slot,
but it is not connected properly.
Remove the CF card, and then reinsert it. There
is a CF card error. To return the CF card icon to
normal, perform the following:
• Remove the CF card, and then reinsert it.
• Exchange it with a functional CF card.
• Use the MV to format the CF card
(the CF card data will be deleted).
The green bar indicates the amount of used CF card
memory. If Media FIFO is switched off, the bar
color will change to red when the amount of available
CF card memory falls below 10%.
If the batch and lot numbers exceed 20 characters,
the date/time area will also be used to display the
batch and lot.
* With the MV2000, the date/time and batch name are displayed
on different lines.
Batch name (displayed
alternately with the display
name)
Current user name
Batch name (displayed
alternately with the display
name and date/time*)
Date and time
If the function that gradually adjusts the time is enabled,
the time is displayed in yellow.
Display or group name
If all channels are being displayed with trend display,
the group name display shows “ALL”.
When using the login function
When using the batch function
Alarm Icons
Status Icon
(Displayed alternately on models with the /M1
or /PM1 option)
Math icon (/M1 and /PM1 options)
Date/time*
Current user name
When using the login and batch functions
Date/time*
(displayed alternately with the
display name)
(Red)
(Green)
Accessing the CF card.
Internal memory error. Contact your nearest Omega dealer for
repairs.
Memory sampling icon
No memory sampling
Memory sampling
in progress
Display paused
(memory sampling
continues)
Data type
DISP: Display data
EVENT: Event data
The green bar indicates the memory sampling progress.
The frame represents either the length of one file save interval (with display
data) or the length of one data segment (with event data).
The remaining memory sampling time.
Indicates that the MV is waiting for an event data trigger.
When an event data pre-trigger is set, the part of the bar that
represents the length of the pre-trigger section will be displayed
in orange.
1.4 Display
Status Display Section
The status display section contains the following information when the MV is in Operation
or Setting Mode.
1-10
M-4660
1
Vertical display
Display layout
Horizontal split display
(displays two groups)
Horizontal display
Current value mark
Alarm mark
Alarm type
The tag name or channel number,
measured value, unit, and the set
alarm
Displays the maximum and
minimum values sampled
in the time corresponding
to one dot.
Maximum value
Minimum value
1 dot
Waveform (displayed with the set channel color)
To change channel colors, see section 6.5
To change waveform line width, see section 6.10
To display every channel's waveform, see section 5.2
Color scale band
Section 6.8
Current value
Section 6.7
Section 6.10
Trend space function
Inserts a division-wide
space here.
Section 5.2
Time at the grid location
Either displays the time, or
the date and time.
Sections 6.3 and 9.1
Grid
Section 6.10
You can display a fine grid.
Fine grid
Section 5.2
Trend update
interval
Section 6.3
Trip line
(up to four lines)
Section 6.1
Measured
value scale
Section 6.7
Alarm value mark
Section 6.8
Numeric display
section
Section 5.2
1 division
Trend Display (T-Y)
1.4 Display
Displays measured data using waveforms. (For operating instructions, see section 5.2.)
Feature Overview
M-4660
1-11
Trend display
Preset messages
Free messages
You can enter non-preset messages.
1
2
3
4
Start
Material 1
List display
Vertical display
1.4 Display
• Waveform Updating
Every 30 dots makes up one division (also referred to as DIV) on the time axis.
The waveform update interval is the time corresponding to one dot. The time
corresponding to one dot depends on the time set to one division (the trend interval).
The relationship between the trend update interval and the speed of waveform
movement in the display is as follows:
Trend interval (per DIV)5 s
Time corresponding to one dot (in seconds)0.1250.250.512
Speed of waveform movement (approximation in mm/h) 10000500025001250625
Trend interval (per DIV)2 min5 min 10 min 15 min 20 min
Time corresponding to one dot (in seconds)410203040
Speed of waveform movement (approximation in mm/h) 312156784231
Trend interval (per DIV)30 min1 h2 h4 h10 h
Time corresponding to one dot (in seconds)601202404801200
Speed of waveform movement (approximation in mm/h) 21105.22.61.0
1 40 dots per division. Available on high-speed input models of the MV.
2 Available on medium-speed input models of the MV when in fast sampling mode .
Switching Trend Update Intervals
1
10 s115 s230 s1 min
(For instructions on how to set the second
interval, see section 6.3.)
You can press T/DIV to change the trend update interval. You can also switch from the
ordinary trend update interval to the secondary trend update interval, and vice versa,
while the MV is collecting data.
• Writing Messages
Preset Messages
You can choose preset messages to be written on the waveform.
Max. number of messages: 100 (Messages 1 through 10 are free messages.)
Free Messages
You can enter non-preset messages.
Max. number of free messages: 10
Automatically written messages (For configuration instructions, see
sections 6.3 and 6.17.)
• The MV will automatically write a message to indicate when the trend update rate has
been switched during memory sampling.
• The MV will automatically write a message to indicate when power has been
restored after a power failure that occurs during memory sampling.
Message Display Options
• When using horizontal or horizontal split display, messages can be displayed either
vertically or horizontally. (For conguration instructions, see section 6.10.)
• Messages can be displayed as a group in the upper left corner of the screen (list
display). (For operating instructions, see section 5.2.)
(For configuration and operating instructions, see section 6.4.)
(For configuration and operating instructions, see section 6.4.)
1-12
M-4660
1
• Zone Display
Zone 2
Zone 1
Zone 3
100%
Time axis
30%
60%
100%
0%0%
Time axis
Without zone display
With zone display
CH3
CH2
CH1
Percentage of
display span %
100
Percentage of
display span %
Measured value
Boundary
value
Boundary
value
shifted
location
Measured value
100
30
0
500
6 V6 V
0–6 V–6 V0
Compressed
area
Time axis
Time axis
With partial expansionWithout partial expansion
Expanded
area
Alarm
Release
RedNoneRed
NoneRedRedNoneNoneNone NoneNone
AlarmACKAlarmACK
Occurrence
Indication not heldIndication held
Alarm type
RedBlueRed
BlueRedRedBlueBlueBlueBlueBlue
Measured
value
Blinking
red
Blinking
green
Blinking
red
GreenRedRedGreenGreenGreenGreenGreen
Alarm mark
You can display channels in specied zones. This allows you to separate waveforms
so that they don’t overlap.
In the following example, channel 1 is set to 0-30%, channel 2 is set to 30-60%, and
channel 3 is set to 60-100%.
For conguration instructions, see section 6.6.
1.4 Display
Feature Overview
Auto Zone (
For operating instructions, see section 5.2.)
You can divide the trend display area evenly between each channel in a group.
• Partial Expansion
This function enables you to compress part of the display and expand the rest of it.
In the following example, the display below 0 V (the boundary value) is shifted to the
bottom 30% of the screen. The bottom 30% of the screen displays –6 V to 0 V, while
the top 70% displays 0 V to 6 V.
For conguration instructions, see section 6.9.
• AlarmDisplay
The alarm mark, alarm type, and measured values are displayed in the following
ways, depending on the alarm condition.
M-4660
1-13
Unit
Tag name or channel number
Alarm mark
Measured value
Alarm
Release
AlarmACKAlarmACK
Occurrence
Indication not heldIndication held
RedBlueRed
BlueRedRedBlueBlueBlueBlueBlue
Measured
value
Blinking
red
Blinking
green
Blinking
red
GreenRedRedGreenGreenGreenGreenGreen
Alarm mark
1.4 Display
Digital Display
Displays measured data using large numeric symbols.
For operating instructions, see section 5.2.
Note
• MeasurementChannelNumeric Value Display
When the MV detects a range-out (see below), the display is either “+Over” or “-Over.”
When the MV detects a burnout on a channel that has been set to be checked for burnouts,
the display is “Burnout”. All other values are displayed numerically.
Range-outs are detected on measurement channels in the following circumstances.
• If the DC voltage input value goes above the measurable range ± 5%, a range-out is
detected. For example, when the measurement range is 2 V, the measurable range is
–2.000 to 2.000 V. If the voltage exceeds 2.200 V, a positive range-out is detected. If it
goes below –2.200 V, a negative range-out is detected.
• If the input type is thermocouple or RTD, a range-out is detected when the temperature
goes 10°C above or below the measurable range. For example, when the measurement
range is set to R, the measurable range is 0.0 to 1760.0°C. If the temperature exceeds
1770.0°C, “+Over” will be displayed. If the temperature goes below -10.0°C, “-Over” will
be displayed.
• On channels that use linear scaling, the range-out values, ignoring the decimal point,
are above 30000 and below –30000. However, you can also set the range-out values to
105% or greater, and –5% or less of the scale width, as long as those values fall within
The range of displayable values, ignoring the decimal point, is –30000 to 30000. The
decimal place is determined by the location of the decimal point on the external input
channel’s minimum span value. Regardless of maximum and minimum span value settings,
all values within the range of –30000 to 30000 will be displayed. If the value exceeds 30000,
“+Over” will be displayed. If the value goes below –30000, “-Over” will be displayed.
• AlarmDisplay
The alarm mark and measured values are displayed in the following ways depending
on the alarm condition.
1-14
M-4660
1
Bar Graph Display
Horizontal
Vertical
Alarm value mark
Scale
Bar
(displayed with the set channel color)
Measured value
Unit of measurement and minimum value
Maximum value
Tag name or channel number
The bar graph base position is set to Normal,
Lower, or Upper.
Section 6.11
The bar graph base position is set to Normal, Lower, or Upper.
Section 6.11
The bar graph base position is set to Center.
Section 6.11
Section 6.7
Section 6.11
Section 6.10
The bar graph base position is set to Center.
Section 6.11
Alarm mark
Alarm
Release
RedGreenRed
GreenRedRedGreen GreenGreen GreenGreen
AlarmACKAlarmACK
Occurrence
Indication not heldIndication held
Value mark
RedBlueRed
BlueRedRedBlueBlueBlueBlueBlue
Measured
value
Blinking
red
Blinking
green
Blinking
red
GreenRedRedGreenGreenGreenGreenGreen
Alarm mark
Displays measured data using bar graphs.
For operating instructions, see section 5.2.
1.4 Display
Feature Overview
• BarGraphDisplayUpdating
The bar graph display is updated at the same rate as numeric values.
• AlarmDisplay
The alarm mark, alarm value mark, and measured values are displayed in the
M-4660
following ways, depending on the alarm condition.
1-15
Cursor
Message
Measured values (maximum and minimum
values for the entire display*)
Measured values (maximum and minimum
values at the cursor position*)
* You can also just display the digital values at the cursor position.
Waveform (channel display color)
Background
color (
Changeable)
Location of the most recent data
Date and time at the left end of the time axis**
Trip line
Date and time at the right end of the time axis** or the
time at the cursor position
(The time at the cursor position is surrounded by a yellow square.)
** You can also set this area to display the time relative to the start of recording.
Unit
Tag name or channel number
(channel display color)
Historical trend
Present trend
Half screen display
1.4 Display
Historical Trend
The MV can display the waveforms of past measured data (display or event data) stored
in internal or external memory. This function is called historical trend.
• HistoricalTrendDisplayOptions
There are ve ways that you can display measured data stored in internal memory:
• From the alarm summary display.
• From the message summary display.
• From the memory summary display.
• From the display selection menu.
See section 5.7.
See section 5.8.
See section 5.9.
See section 5.3.
• By pressing HISTORY.
You can also display measured data stored to external memory.
See section 4.8.
• DisplayContents
1-16
ItemDescription
Alarm summaryDisplays a summary of the displayed data’s alarms.
Message summaryDisplays a summary of the displayed data’s messages.
Memory summaryDisplays the properties (file name, sample start time, end time,
etc.) of the displayed data.
• Add Message
You can enter added messages.
For operating instructions, see section 6.4.
M-4660
1
• AutoSpan
0
Time axis
Measured valueMeasured value
Auto span
0
Time axis
0
Time axis
Measured valueMeasured value
Auto span
0
Time axis
The MV can automatically adjust the display span of the selected channel.* It sets
the span based on the maximum and minimum displayed historical data values. Auto
span is deactivated when you switch to another group.
* Auto span affects channels that are in the same scale position as the selected channel.
When the maximum or minimum values are outside of the set display span
When the maximum or minimum values are within the set display span
1.4 Display
Feature Overview
If the maximum or minimum data value falls outside the maximum selectable display
span, the MV adjusts the display span to the maximum or minimum possible value.
The MV responds in the same way when it encounters overow data.
• TopChannel
Displays the selected channel’s historical trend waveform in front of all of the others.
This setting is deactivated when you switch to another group.
M-4660
1-17
Tag name or channel number
Cursor
Measured value and unit
Alarm type
Channels with active alarms are red.
Alarm
Release
Blinking
white
Blinking
black
Blinking
white
Black WhiteWhiteBlack BlackBlack BlackBlack
Alarm ACK
Alarm ACK
Occurrence
When indication is
set to non-hold
When indication is
set to hold
Tag/Channel
WhiteNoneWhite
None WhiteWhiteNone
None
None NoneNone
Alarm type
WhiteBlackWhite
BlackWhiteWhiteBlack
Black
Black BlackBlack
Measured
value
RedGreenRed
GreenRedRedGreenGreenGreenGreenGreen
Channel
area
1.4 Display
Overview
Displays a list that shows the conditions of all channels.
You can select a channel to access the trend, digital, or bar graph display of its group.
For operating instructions, see section 5.4.
• AlarmIndication
The channel display area, channel number, tag name, alarm type, and measured
value all change in appearance based on alarm conditions in the following ways.
1-18
M-4660
1
Alarm Summary
To the historical trend display
The alarm information number of the alarm on the screen’s bottom
line and the number of alarm entries in internal memory.
Cursor (selects an alarm)
: Alarm occurrence
( blinks until the AlarmACK operation is carried out if Indicator
is set to Hold.)
Channel
Alarm number (1, 2, 3, 4) and type (H, L, h, l, R, r, T, t)
Date/time of alarm occurrence/release
: Alarm release
: Alarm output release (when blinking is cleared through the
AlarmACK operation)
You can list up-to-date alarm information.
• The data of up to 1000 alarms can be displayed.
• You can select alarm information to access the historical trend of the display data or
event data that contains the alarm.
For operating instructions, see section 5.7.
1.4 Display
Feature Overview
M-4660
1-19
To the historical trend display
The message number displayed on the screen’s bottom
line and the number of message entries in internal memory.
Cursor (used to select a message)
Message
Added message (blue)
Date and time when the message was written
The groups on which the message was written
Displays a group number, or A for all groups
1.4 Display
Message Summary
You can display a list of messages and when they were written.
• Up to 450 messages can be displayed.
• Up to 50 added messages can be displayed.
• You can select message information to access the historical trend of the display data
or event data that contains the message.
For operating instructions, see section 5.8.
• SwitchingDisplayedItems
You can switch between two sets of displayed items.
• Message, date and time, and group
• Message and the name of the user who wrote the message
1-20
M-4660
1
Memory Summary
To the historical trend display
Cursor (selects a file)
Memory start date and time
Memory stop date and time
Sampling count
Status
Internal memory data files/
maximum number of recordable files
Date and time of the most recently
recorded data
Data type
• Display data
• Event data
You can display a list of the display and event data that is stored in internal memory.
• You can select a display data or event data entry to access its historical trend display.
• The MV displays the number of manually sampled data and report data (/M1 and /
PM1 options) samples.
For operating instructions, see section 5.9.
1.4 Display
Feature Overview
• SwitchingDisplayedItems
You can switch between two lists of information.
• Start and stop times
• File names
• DataStorage
You can save the data in the internal memory to a CF card or to USB ash memory.
M-4660
1-21
Displayed report data number/number of report data entries in internal memory
Start: Start date and time
Timeout: Report date and time
Channel number
Unit
Report data status
Indicates whether any of the following
occurred during the report interval.
E: Error data
O: Overflow data
P: Power failure
C: Time adjustment
B: Burnout
Average, maximum, minimum, sum, or instantaneous value
Report type
Daily sum of all channels and the daily
sums of all report groups
If you select a bar graph, the sum of the hour
selected with the cursor will be displayed.
Hourly transition of report group sums (bar graph).
Daily sums (bar graph)
Example: Hourly + daily display
Single graph displayDual graph display
1.4 Display
Report Data (/M1 and /PM1 options)
You can display report data stored in the internal memory.
For operating instructions, see section 5.5.
Stacked Bar Graph (/M1 and /PM1 options)
You can display the report data of each report group in a stacked bar graph.
For operating instructions, see section 5.6.
For information on report groups, see section 10.5.
• DisplayedDataType
The type of data that is displayed is determined by the report type, which is set using
the report function.
Displayed DataReport Type
Hourly + dailyHourly, or hourly + daily
Daily + weeklyDaily + weekly
Daily + monthlyDaily, or daily + monthly
• DisplayModes
You can switch the bar graph between single graph and dual graph display.
1-22
M-4660
1
Status Display
Log Display
1.4 Display
These are the available status displays.
For operating instructions, see section 5.5.
• Relay Status Display
Displays the status of the alarm output relay and the internal switch.
• Modbus Client Status Display and Modbus Master Status Display
Display the command status.
Displays all logs (operation histories).
For operating instructions, see section 5.10.
Log TypeDescription
LoginLogin and logout history, time setting history, power failure history
ErrorError message history
CommunicationCommunication command history
FTPFTP transfer history
WEBWeb operation history
E-mailE-mail transfer history
SNTPSNTP server access history
DHCPDHCP server access history
MODBUSModbus master and client transfer history
Feature Overview
M-4660
1-23
Loading...
+ 271 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.