We are grateful for your purchase of Yokogawa’s Infrared Gas Analyzer, Model:IR400.
• First read this instruction manual carefully until an adequate understanding is acquired, and
then proceed to installation, operation and maintenance of the analyzer. Wrong handling
may cause an accident or injury.
• The specifications of this analyzer are subject to change without prior notice for further
product improvement.
• Modification of this analyzer is strictly prohibited unless a written approval is obtained
from the manufacturer. Yokogawa will not bear any responsibility for a trouble caused by
such a modification.
• This instruction manual shall be stored by the person who actually uses the analyzer.
• After reading the manual, be sure to store it at a place easier to access.
• This instruction manual should be delivered to the end user without fail.
Manufacturer:Yokogawa Electric Corporation
Type:Described in Yokogawa’s company nameplate on main frame
Date of manufacture: Described in Yokogawa’s company nameplate on main frame
Product nationality:Japan
Request
• It is prohibited to transfer part or all of this manual without
Yokogawa’s permission in written format.
• Description in this manual is subject to change without prior
notice for further improvement.
IM 11G02N01-01E
i
CAUTION ON SAFETY
First of all, read this “Caution on safety” carefully, and then use the analyzer in the correct way.
• The cautionary descriptions listed here contain important information about safety, so they should
always be observed. Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and
“PROHIBITION”.
DANGER
CAUTION
PROHIBITION
Caution on installation and transport of gas analyzer
DANGER
CAUTION
Improper handling may cause dangerous situations that may
result in death or serious injury.
Improper handling may cause dangerous situations that may
result in medium-level troubles, minor injury, or property
damage.
Items which must not be done are noted.
• The unit is not of explosion-proof specifications. Do not use it
in an atmosphere of explosive gases. Otherwise serious
accidents such as explosion or fire may result.
• Install the analyzer, observing the rules provided in this
manual, in a place that endures the weight of the analyzer.
Installation in an inadequate place may cause turnover or fall,
resulting in injury.
• Be sure to wear protective gloves when lifting the analyzer.
Lifting it with bare hands may result in injury.
• Be sure to fix the casing before transporting the analyzer.
Transportation in unstable state may result in injury.
• The gas analyzer is heavy. Two or more persons should carry
it, while exercising due care. Otherwise unexpected harm to
your body or injury may result.
• Take care not to let cable chips and other foreign objects enter
the unit during installation work. Otherwise fire, failure, or
malfunction may result.
ii
IM 11G02N01-01E
DANGER
Caution on piping
Be sure to observe the following precautions while installing
piping. Improper piping may result in gas leakage.
If the leaking gas contains a toxic component, serious
accidents may result. If it contains combustible gases,
explosion or fire may result.
• Connect pipes correctly referring to the instruction manual.
• Discharge the exhaust gas outdoors to prevent it from
remaining within the sampling device or indoors.
• Relieve the exhaust gas from the analyzer to the atmospheric
pressure to prevent buildup of undesirable pressure to the
analyzer. Otherwise piping within the analyzer may be
disconnected, resulting in gas leakage.
• Use pipes and pressure reducing valves to which no oil/grease
is attached for piping. Otherwise, fire may result.
Caution on wiring
CAUTION
DANGER
CAUTION
• Be sure to turn off the power before installing wiring.
Otherwise electric shock may result.
• Be sure to perform class D grounding work. Otherwise,
electric shock or failure may result.
• Select a proper wiring material that satisfies the ratings of the
instrument. Otherwise, electric shock or fire may result.
• Be sure to connect a power supply of correct rating.
Otherwise, fire may result.
Caution on use
• Be sure to read the instruction manual for reference gases before
handling reference gases such as calibration gas to use them
properly.
• Leaving the analyzer unused for a long time or restarting it after
long-term suspension requires procedures different from normal
operation or suspension procedures. Be sure to follow the
instructions in each instruction manual. Otherwise, intended
performance may not be achieved, or accidents or injury may
result.
• Do not operate the analyzer for a long time with its door left
open. Otherwise, dust, foreign matter, etc. may stick on internal
walls, thereby causing faults.
IM 11G02N01-01E
iii
Caution on use
• Do not touch the input/output terminals with metal or finger.
PROHIBITION
Caution on maintenance and check
DANGER
CAUTION
Otherwise, electric shock or injury may result.
• Do not smoke or use flames near the analyzer. Otherwise, fire
may result.
• Do not allow water to enter the analyzer. Otherwise, electric
shock or internal fire may result.
• Before performing work with the cover of the analyzer kept
open for maintenance and check, be sure to purge completely
not only within the analyzer but also measuring gas lines with
nitrogen or air. Otherwise, poisoning, fire, or explosion may
result due to gas leakage.
Be sure to observe the following to perform work safely,
avoiding electric shock or injury.
• Remove the watch and other metallic objects before work.
• Do not touch the instrument wet-handed.
CAUTION
• If the fuse is blown, eliminate the cause and replace it with the
one of the same capacity and type. Otherwise, electric shock
or accidents may result.
• Do not use replacement parts other than those specified by the
manufacturer. Otherwise, intended performance may not be
achieved, or accidents or failures may result.
• Dispose replacement parts such as maintenance parts as
incombustibles according to the local waste disposal
regulations.
Others
• If the cause of a fault cannot be identified by referring to the
instruction manual, be sure to contact your dealer or Yokogawa
technician in charge of adjustment. Disassembling the
instrument carelessly may result in electric shock or injury.
iv
IM 11G02N01-01E
r After - Sales Warranty
d Do not modify the product.
d During the warranty period, for repair under warranty carry or send the product to
the local sales representative or service office. Yokogawa will replace or repair any
damaged parts and return the product to you.
d Before returning a product for repair under warranty, provide us with the model
name and serial number and a description of the problem. Any diagrams or data
explaining the problem would also be appreciated.
d If we replace the product with a new one, we won’t provide you with a repair report.
d Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the
customer site is located outside of the service area, a fee for dispatching the
maintenance engineer will be charged to the customer.
d In the following cases, customer will be charged repair fee regardless of warranty
period.
• Failure of components which are out of scope of warranty stated in instruction
manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which
Yokogawa did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to misoperation, misuse or modification which Yokogawa does not
authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or
abnormal.
• Failure caused by any usage out of scope of recommended usage
• Any damage from fire, earthquake, a storm and flood, lightning, disturbance, riot,
warfare, radiation and other natural changes.
d Yokogawa does not warrant conformance with the specific application at the user
site. Yokogawa will not bear direct/indirect responsibility for damage due to a
specific application.
d Yokogawa will not bear responsibility when the user configures the product into
systems or resells the product.
d Maintenance service and supplying repair parts will be covered for five years after
the production ends. For repair this product, please contact the nearest sales office
described in this instruction manual.
IM 11G02N01-01E
v
Blank Page
≈
CONTENTS
PREFACE.............................................................................................................................. i
Caution on safety ................................................................................................................. ii
r After - Sales Warranty ................................................................................................... v
Contents ............................................................................................................................... vi
Customer Maintenance Parts List ................................................ CMPL 11G02N01-01E
Revision Record .................................................................................................................... i
1.OVERVIEW
This infrared gas analyzer (type: IR400) measures the concentration of NO, SO2, CO2, CO and CH
4
contained in sampling gas on the principle that different atomic molecules have an absorption spectrum in
the wave band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law.
Since this instrument incorporates a compact paramagnetic O
components simultaneously by using the built-in O
sensor (up to 4 components if O2 sensor is excluded).
2
sensor, it allows measuring up to 5
2
Furthermore, use of a microprocessor or large sized liquid crystal display realizes improvement of
operability, accuracy and multi-functions.
This instrument is optimum for measuring combustible gas exhausted from boilers or incinerators, and it
is effective for steel gas analysis (blast furnace, steel converter, thermal treatment furnace, sintering
(Pellet equipment), coke furnace), storage and maturity of vegetable and fruit, biochemistry (microbe),
[fermentation], air pollution [incinerator, exhaust gas desulfurization, denitration], automotive emission
(excluding tester), protection against disasters [detection of explosive gas and toxic gas, combustion gas
analysis of new building material], growth of plants, chemical analysis [petroleum refinery plant,
petroleum chemistry plant, gas generation plant], environment [landing concentration, tunnel
concentration, parking lot, building management] and various physical and chemical experiments.
IM 11G02N01-01E
1 - 1
Blank Page
≈
2.NAME AND DESCRIPTION OF EACH UNIT
2.1 Name and description of main unit
Front panel
POWER
(1) Handle
(2) Power switch
(3) Display/operation panel
Back panel
(6) Sector motor
(7) Light source cover
Name
(1) Handle
Used for withdrawing the main
unit from the panel.
Description
Fig. 2-1
(6) Sector motor
(4) Sampling gas inlet
(5) Sampling gas outlet
(4) Sampling gas inlet
(5) Sampling gas outlet
(or purge gas inlet)
(10) Protective cover
(9) Power inlet
(8) Input/output terminal connector
Name
For measuring
unit 1
For measuring
unit 2
Description
For driving the rotation of
sector
(2) Power switch
(3) Display/opera tion
(4) Sampling gas
inlet
(5) Sampling gas
outlet
IM 11G02N01-01E
Used for ON/OFF the analyzer.
Liquid crysral diaplay and keys
for setting various functions
For connecting to the measuring
gas tube
Connect to the exhaust line. (A
pair of sampling gas inlet/outlet is
provided for each measuring unit.
When ordered with purge, the
piping to measuring unit 2 is built
inside. In this case, the sample gas
outlet for measuring unit 2 is used
for purge gas inlet.)
(7) Light source
cover
(8) Input/output ter minal connector
(9) Power inlet
(10) Protective cover
Infrared light source is
arranged in the cover.
For connecting to the external
input/output terminal module
For connecting the power cable
Protective cover for the light
source and the motor. May be
removed during operation.
2 - 1
2.2 Input/Output terminal module
This analyzer provides input/output of various signals from the supplied input/outpt terminal
module by connecting the instrument to this module.
Input/output terminal for
signals of analog output, range
identification contact, alarm
contact output, etc.
Used for connecting the analyzer
main unit and the input/output
terminal module (4).
(5) Calibration solenoid valve drive signal
output connector (CN3)
Name
(4) Input/output
terminal module
connection cable
(5) Calibration
solenoid valve
drive signal
output connector
(6) Communication
connector
Description
Used for connecting the
analyzer main unit to the input/
output terminal module.
Cable connector for connecting
the analyzer to the relay board
for automatic calibration.
Connect communication cable.
*Please refer to another manual
(IM 11G02P01-01E) about
communication function.
2 - 2
IM 11G02N01-01E
3.INSTALLATION
DANGER
This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent
explosion, fire or other serious accidents.
CAUTION
•Entrust the installation, movement or re-installation to a specialist or the supplier. A poor
installation may cause accidental tipover, shock hazard, fire, injury, etc.
•The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tip
over or drop, for example, causing accident or injury.
•For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an
injury.
•This unit should be installed in a place which conforms to the conditions noted in the
instruction manual. Otherwise, it may cause electric shocks, fire or malfunction of the unit.
•During installation work, care should be taken to keep the unit free from entry of cable chips
or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
3.1 Installation conditions
To install the analyzer for optimum performance, select a location that meets the following
conditions;
(1) This instrument is system built in type. This instrument should be used while embedded
in a panel, locker, or enclosure of steel sheet.
(2) Use this instrument indoors.
(3) A vibration-free place
(4) A place which is clean around the analyzer.
(5) Power supply
Rated voltage: 100 V to 240 V AC
Operating voltage: 85 V to 264 V AC
Rated frequency: 50/60 Hz Power consumption: 250 VA max.Inlet: Comformity to EN60320 class I type 3-pin inlet
(6) Operation conditions
Ambient temperature: -5˚ to 45˚Ambient humidity: 90% RH or less, no condensation
C
IM 11G02N01-01E
3 - 1
3.2 Installation
3.2.1 Installation of analyzer main frame
Installation methods for the analyzer main unit are divided into 2 types;
Note 1 Check and maintenance of the analyzer main unit may be carried out with the top cover detached.
The guide rail method may be used if a space accessible for maintenance is provided at the top of
the main unit. If maintenance space is not provided specially, it is recommended to use the slide
rail method.
Recommended slide rail: Product No.: 305A-24 manufactured by Accuride International Co.
Note 2 For 19 inch rack mounting, the weight of the analyzer is supported with the bottom of the case
(with the side of the case in case of slide rail method). For mounting dimensions of the slide rail,
see “Item 9.3 External diagram”.
Don’t install the analyzer at a place which is exposed to direct sunlight.
The analyzer should be installed at a place where ambient temperature is within -5 to 45˚C, and
temperature fluctuation during use is minimum.
)
3 - 2
IM 11G02N01-01E
3.2.2 Mounting input/output terminal module
Mount the input/output terminal module on the panel; observing the following method.
(Note) To avoid the effect of noise generated from external units, mount the I/O terminal module
mounting plate on the panel for continuity at the mounting surface and connect the panel
to the same ground as the analyzer main unit.
Note) How to ground analyzer main unit and I/O terminal module
To avoid the effect of noises, etc. from external units, it is
recommended to ground them by the procedure described below.
Bring I/O terminal module sheet metal and
panel into continuity at ///// portion.
Bring analyzer main unit and panel into continuity
at ///// portion or rail mounting portion.
(No grounding is required at the power terminal).
Terminal
142142
Screwed to panel
Panel plate
Connection cable
To analyzer rear panel
IM 11G02N01-01E
Mount the analyzer and I/O terminal module on the
same panel.
Ground the panel casing.
Don’t separate the analyzer and I/O terminal module, and be sure to
*
ground them together.
3 - 3
3.3 Piping
Piping should be connected to the gas inlets and outlets of the front panel of the analyzer.
•Use a corrosion resistant tube of Teflon, stainless or polyethylene to connect the instrument to
a sampling system. Even if there is a danger of corrosion, refrain from using a tube of rubber
or soft vinyl. The instrument provides inaccurate indication due to gas absorption by piping
materials.
•Pipe connection port is Rc1/4 female thread (or 1/4 NPT). Piping should be cut as short as
possible for a quick response. About 4 mm inner diameter is recommended.
•Entry of dust into the instrument may result in defective operation. Use a clean piping or
coupling.
Connect the gas tube by the following method.
Sampling gas outlet
For measuring
unit 1
Sampling gas inlet
Sampling gas inlet
Sampling gas outlet
(Purge gas inlet)
For measuring
unit 2
Sampling gas inlet:Attach the gas tube to introduce gas to be measured such as one that has
completed dehumidification process and standard gases for zero and span
calibration to this inlet.
Gas flow to be introduced should be constant within the range of 0.5 ± 0.2
L/min.
Sampling gas outlet: Exhaust measured gas through the outlet. Attach the tube to exhaust mea-
sured gas outdoors or to the atmosphere.
Purge gas inlet:It is used for purging the inside of the total gas analyzer . When the ana-
lyzer must be purged, refer to Item 3.3.4 Purging of instrument inside.
Use dry gas N
or instrumentation air for purge gas. (flow rate of 1L/min or
2
more should be used and no dust or mist is contained).
3 - 4
IM 11G02N01-01E
Internal piping diagram
Measuring unit 2
Measuring unit 1
Built-in
sensor
O
2
Built-in
O
sensor
2
OUTLET INLETOUTLET INLET
Sampling cell
Note)
2 cells may be used
by combination of
Sampling gas inlet
Sampling gas outlet
Sampling gas inlet
Sampling gas outlet
For measuring
unit 2
For measuring
unit 1
range.
Note) When the purge gas inlet is provided, the piping to measuring unit 2 is built inside.
Correspondence of measured components and measuring units
Measuring components
1-component Analyzer for
NO, SO
, CO2, CO and CH
2
2-component Analyzer for
NO-SO
and CO2-CO
2
2-component Analyzer for
NO-CO
3-component Analyzer for
NO-SO2-CO
4-component Analyzer for
NO-SO
-CO2-CO
2
Measuring unit 1
Each component
4
NO-SO
CO2-CO
NO
NO-SO
NO-SO
2
2
2
Note) When there are two measuring units, the built-in O2 sensor must
be connected to the measuring unit 2.
Measuring unit 2
None
None
CO
CO
CO
-CO
2
IM 11G02N01-01E
3 - 5
Example of connecting each measuring unit
?
One pair of gas inlet/outlet
Sampling gas
Flow meter
0.5L/min
?
Two pair of gas inlet/outlet - (1)
Sampling gas
NO2/NO
converter
Flow meter
0.5L/min
Filter
Filter
Flow meter
0.5L/min
Filter
IR400
Sampling gas inlet
Sampling gas outlet
Release to atmosphere.
IR400
Sampling gas
inlet
Sampling gas
outlet
Sampling gas
inlet
Sampling gas
outlet
Exhaust
For
measuring
unit 1
For
measuring
unit 2
Note) The NO
when NO measurement is used for NOx measurement.
?
Two pair of gas inlet/outlet - (2)
Sampling gas
Note) The NO
when NO measurement is used for NOx measurement.
/NO converter is used
2
Filter
Flow meter
0.5L/min
/NO converter is used
2
/NO
NO
2
converter
Filter
Sampling gas
inlet
Sampling gas
outlet
Sampling gas
inlet
Sampling gas
outlet
Exhaust
Release to atmosphere.
IR400
For
measuring
unit 1
For
measuring
unit 2
Exhaust
Release to atmosphere.
3 - 6
IM 11G02N01-01E
3.4 Sampling
3.4.1 Conditions of sampling gas
1.Dust contained in the sampling gas should be completely removed with a filter. For the final
stage filter, use a filter that allows removing dust particles of 0.3mm.
2.Dew point of sampling gas must be lower than the ambient temperature to avoid occurrence
of drain in the gas analyzer. If vapor is contained in the sampling gas, dew point should be
lowered to 0°C by using a dehumidifier.
3.If SO
4.Corrosive gases such as Cl
5.Temperature of sampling gas should be within 0 to 50°C. Provide a means that prevents
3.4.2 Sampling gas flow
Flow of sampling gas should be 0.5 ± 0.2L/min.
Avoid flow fluctuation during measurement.
Observe the flow reading by a flowmeter provided as shown in the example of the sampling
system configuration (Item 3.4.6).
3.4.3 Preparation of standard gas
mist is contained in the sampling gas, use a mist filter or cooler to remove SO3 mist.
3
Other mists should be removed by using a mist filter or cooler.
2
erable amounts, will shorten the life of instruments.
entry of hot gas directly into the instrument.
, F2 and HCl, if they are contained in the sampling gas in consid-
Routine calibration is required by standard gas for keeping this instrument under normal operation
condition (once a week). Prepare a standard gas cylinder for zero calibration and span calibration.
Zero gas
Span gas other
than for O
measurement
Span gas for O
measurement
2
Analyzer without O
measurement
N2 gas
Gas with concentration of 90% or more
of full scale
2
Analyzer with built-in O
2
sensor
N2 gas
Gas with concentration of
90% or more of full scale
Gas with concentration of
90% or more of full scale or
atmospheric air (21%)
3.4.4 Reduction of moisture interference
NO and SO
2 measurement is subject to moisture interference.
As shown by the configuration example on the next page, provide a device for humidifying zero
calibration gas, thus controlling the moisture content at a constant level (moisture content in
sample gas should also be controlled here) in configuring a sampling system. That allows the
same moisture content as in the case of measurement to be contained in zero gas for calibration.
3.4.5 Purging of instrument inside
Analyzer with external zirconia O
2
sensor
Dry air or atmospheric air (Note)
Gas with concentration of 90% or
more of full scale
1 to 2% O
2
2
The inside of instrument need not be purged generally except for the following cases.
1. A combustible gas component is contained in sample gas.
2. Corrosive gas is contained in the atmospheric air at the installation site.
3. The same gas as the sample gas component is contained in the atmospheric air at the installa-
tion site.
In such cases as above, the inside of analyzer should be purged with the air for instrumentation
or N2. Purging flow rate should be about 1L/min.
If dust or mist is contained in purging gas, it should be eliminated completely in advance.
IM 11G02N01-01E
3 - 7
3.4.6 Pressure at sampling gas outlet
Pressure at the sampling gas outlet should be adjusted to atmospheric pressure.
3.4.7 Example configuration of gas sampling system
The following illustrates a typical system configuration for five component gas measurement for
monitoring combustion exhaust gas from boiler, refuse incinerator, etc.
Contact Yokogawa for system configuration matching the particular use or further information.
(1) Gas extractor
φ10/φ8
Teflon
tube
15
or larger
(2) Mist filter
(4) Gas aspirator
2-way
solenoid
valve
(5)
Electronic
gas cooler
(8) Flowmeter
Infrared gas analyzer
main unit CO,CO
(7) Membrane filter
(O2)
2
(11)
NO
converter
(7)
Membrane filter
Infrared gas analyzer
main unit NO, SO
IR400
/NO
2
2
Name
(1) Gas extractor
(2)
Mist filter
(3) Safety drain trap
(4) Gas aspirator
(5) Electronic gas
cooler
(6)
Solenoid valve
(7) Membrane filter
(3) Safety
drain
trap
Drain
Description
Gas extractor with a heating
type stainless steel filter of standard mesh 40 m
Removes drain, mist, and dust.
The safety drain trap divided
into two rooms for positive and
negative pressure. It monitors
and adjusts the sample gas pressure.
For aspiration of sample gas
Dries the moisture in sample
gas to a dew point of approx.
2˚C.
Used for introducing calibration gas.
PTFE filter used to eliminate
fine dust particles and permit
monitoring of dust adhering
condition on the front panel of
the gas analyzer.
ZERO NOSO
Air
Name
(8) Flowmeter
(9) Standard gas
(10)Zirconia O
2
sensor
(11)NO2/NO con-
verter
NO
/N
2
Pressure
reducing valve
O
2
/N
2
(6)
Solenoid
valve
2
CO O
CO
2
2
SO
CO
2
/N
2
(9) Standard gas
CO
/N
2
/N
2
2
Description
Adjusts and monitors the flow
rate of sample gas.
Reference gas used for calibrating zero and span of the analyzer. Total 6 cylinders required
for zero gas air, span gas NO,
, CO, CO2 and O2.
SO
2
External zirconia oxygen
sensor used for measuring the
oxygen concentration in
sample gas.
(This is not necessary in case
when O
sensor is built-in.)
2
Added to NOx analyzer.
A special catalyst material for
efficient conversion of NO
gas to NO is used.
(10)
Zirconia
O
senor
2
(ZX8D)
Exhaust
2
3 - 8
IM 11G02N01-01E
3.5 Wiring
3.5.1 Power inlet
The power inlet is provided at the rear panel.
Connect supplied power cable to this power inlet.
Rear panel
Power inlet
Input/output terminal connector
Grounding 2-pole plug
When noise source is in the vicinity
•Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder)
that generates much electrical noise. If using the
instrument near such a noise generating unit is
unavoidable, use a different power line to avoid noise.
Main unit
power supply
Varistor or
spark killer
•Mount a noise suppressor such as varister or spark killer
as shown at right figure to the noise generating unit when
noise is generated from relays or solenoid valves.
Mount the suppressor near the noise generating source,
or it will have no effect.
3.5.2 Input/output terminal module
This analyzer should be connected to the input/output terminal module by supplied exclusive
cable.
Plug this cable connector into the receptacle at the rear panel of the analyzer and the receptacle on
the PC board of the input/output module.
Connect the exclusive cable so that the ferrite core attached to the cable comes to the analyzer
side.
Ferrite core
(to analyzer side)
Install (connect)
near the source.
Noise
generating
source
IM 11G02N01-01E
Analyzer
Exclusive cable
(1 meter long)
Input/output
terminal module
3 - 9
(1) Analog output signal (AO): terminal block 1, 1 to 20, terminal block 2, 3 to 6.
Output signal:4 to 20 mA DC or 0 to 1 V DC (selected when ordering)
Non-insulated output
Allowable load: 4 to 20 mA DC, 550Ω or less
0 to 1 V DC, 100kΩ or more
•Analog output is provided from each terminal corresponding to the channel displayed in the
measurement screen.
All of analog output signals for the instrument are not isolated. It is recommended to isolate
signals individually to prevent interference from unnecessary signals or to prevent external
interference, especially leading the cable of more than 30 meters or to outdoor.
(2) O
sensor input: terminal block 2, 1 to 2.
2
Input signal:
External zirconia O
External O
analyzer:0 to 1 V DC (DC input resistor of 1MΩ or more)
2
•It is used when the external zirconia O
•To connect to the output of the external Zirconia analyzer or external O
analyzer:Zirconia O2 sensor signal (ZX8Dp C output)
2
analyzer or external O2 analyzer is specified as order.
2
analyzer prepared
2
separately.
•In case of an external O
analyzer, input a signal of 0 to 1 V DC with respect to O2 full scale
2
of the analyzer.
•In case of built-in O
O
sensor input is not isolated. It is recommended to isolate when an external O2 analyzer is
2
installed apart from this analyzer. Zirconia O
analyzer, do not use the terminals.
2
sensor Yokogawa make IR400 should be
2
installed at a location that is as close to this instrument as possible.
(3) Contact input (DI): terminal block 2, 1 to 20, terminal block 3, 5 to 10.
•It is for a contact input at no voltage. An input is provided when switching to short circuit
•An output is for a relay contact output. An output is provided when switching to conductive
(on) or open (off).
3 - 10
Wiring of analog output signal, O
sensor input and contact input should be fixed separately
2
from the wiring of power supply and contact output.
Note) To avoid the effect of noise generated from external units, be sure to ground the
analyzer main unit. Continue between the I/O module mounting plate and the panel
and connect the panel casing to the same ground as the analyzer.
IM 11G02N01-01E
(5) List of terminal blocks
Ch5 output
(Ch5_OUT)
Ch4 output
(Ch4_OUT)
Ch3 output
(Ch3_OUT)
Ch2 output
(Ch2_OUT)
Ch1 output
(Ch1_OUT)
Unassigned
Unassigned
Remote hold input
(R_HOLD)
Average value reset
input (RESET)
Auto calibration
remote start
input (R_CAL)
Terminal block 1
<TN1>
11
1
—
+
—
+
—
+
—
+
—
+
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
—
+
—
+
—
+
Terminal block 3
<TN3>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Ch10 output
(Ch10_OUT)
Ch9 output
(Ch9_OUT)
Ch8 output
(Ch8_OUT)
Ch7 output
(Ch7_OUT)
Ch6 output
(Ch6_OUT)
Ch5 range identification
contact output (RNG_ID Ch5)
Ch4 range identification
contact output (RNG_ID Ch4)
Ch3 range identification
contact output (RNG_ID Ch3)
Ch2 range identification
contact output (RNG_ID Ch2)
Ch1 range identification
contact output (RNG_ID Ch1)
Note 1
sensor input
O
2
(O
_IN)
2
Ch12 output
(Ch12_OUT)
Ch11 output
(Ch11_OUT)
Unassigned
Unassigned
Note 1 : For external O
Peak count alarm
output (PEAK_ALM)
Auto calibration
status/
contact
output (CAL)
Pump ON/OFF
contact output (PUMP)
Calibration error
contact output
(CAL_ALM)
Instrument error
contact output
(FAULT)
Terminal block 2
<TN2>
+
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
sensor input.
2
11
1
—
Terminal block 4
<TN4>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Ch5 remote range
switch input
(R_RNG_Ch5)
Ch4 remote range
switch input
(R_RNG_Ch4)
Ch3 remote range
switch input
(R_RNG_Ch3)
Ch2 remote range
switch input
(R_RNG_Ch2)
Ch1 remote range
switch input
(R_RNG_Ch1)
Unassigned
Unassigned
Unassigned
Unassigned
Unassigned
Unassigned
Alarm 3 output
(ALM_3)
Alarm 2 output
(ALM_2)
Alarm 1 output
(ALM_1)
IM 11G02N01-01E
Terminal block 5
<TN5>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Unassigned
Alarm 6
output
(ALM_6)
Alarm 5
output
(ALM_5)
Alarm 4
output
(ALM_4)
Connector
<CN3>
Solenoid valve drive signal output for calibration
Contact output for
sample gas selection
Contact output for
zero calibration
Contact output for Ch1
span calibration
Contact output for Ch2
span calibration
(Transister output)
1
2
3
4
5
<D-sub 9P>
Contact output for Ch3
6
span calibration
Contact output for Ch4
7
span calibration
Contact output for Ch5
8
span calibration
9
Power for relay drive
5V DC
3 - 11
(6) Description on terminal block
Terminal block 1
<TN1>
11
Ch5 output
(Ch5_OUT)
Ch4 output
(Ch4_OUT)
Ch3 output
(Ch3_OUT)
Ch2 output
(Ch2_OUT)
Ch1 output
(Ch1_OUT)
1
—
+
2
3
—
+
4
5
—
+
6
7
—
+
8
9
—
+
10
(M3.5 screw)
12
13
14
15
16
17
18
19
20
Terminal block 2
O
Note 1
sensor input
2
(O
_IN)
2
Ch12 output
(Ch12_OUT)
Ch11 output
(Ch11_OUT)
Unassigned
Unassigned
Note 1:For external O2 sensor input.
<TN2>
1
—
+
2
3
—
+
4
5
—
+
6
7
8
9
10
(M3.5 screw)
11
12
13
14
15
16
17
18
19
20
Ch10 output
—
+
(Ch10_OUT)
Ch9 output
—
+
(Ch9_OUT)
Ch8 output
—
+
(Ch8_OUT)
Ch7 output
—
+
(Ch7_OUT)
Ch6 output
—
+
(Ch6_OUT)
Ch5 remote range
changeover input
(R_RNG_Ch5)
Ch4 remote range
changeover input
(R_RNG_Ch4)
Ch3 remote range
changeover input
(R_RNG_Ch3)
Ch2 remote range
changeover input
(R_RNG_Ch2)
Ch1 remote range
changeover input
(R_RNG_Ch1)
Terminal block 1 <TN1>
Terminal block for analog output (non-isolated output)
Between 1–2:Ch5 output
Between 3–4:Ch4 output
Between 5–6:Ch3output
Between 7–8:Ch2 output
Between 9–10:Ch1 output
Between 11–12:Ch10 output
Between 13–14:Ch9 output
Between 15–16:Ch8 output
Between 17–18:Ch7 output
Between 19–20:Ch6 output
Terminal block 2 <TN2>
Between 1–2:O
sensor input
2
(For input of Fuji’s zirconia
oxygen sensor or externally
oxygen sensor. Must not be used
unless external O
sensor is
2
provided.)
Between 3–4:Ch12 output
Between 5–6:Ch11 output
Between 7–10For internal connection. Must not
be wired. (Must not be used as
junction terminal).
Between 11–12:Ch5 remote range switch input
Between 13–14:Ch4 remote range switch input
Between 15–16:Ch3 remote range switch input
Between 17–18:Ch2 remote range switch input
Between 19–20:Ch1 remote range switch input
3 - 12
Action of remote range switch
High range is selected when open.
Low range is selected when short-
circuited. For details of action, see
“Item 6.1 Switch of range.”
IM 11G02N01-01E
Unassigned
Unassigned
Remote hold input
(R_HOLD)
Average value reset
input (RESET)
Auto calibration
remote start
input (R_CAL)
Terminal block 3
<TN3>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Ch5 range identification
contact output (RNG_IDCh5)
Ch4 range identification
contact output (RNG_IDCh4)
Ch3 range identification
contact output (RNG_IDCh3)
Ch2 range identification
contact output (RNG_IDCh2)
Ch1 range identification
contact output (RNG_IDCh1)
Terminal block 3 <TN3>
Between 1–4:For internal connection. Must
not be wired. (Must not be used
as junction terminal.)
Between 5–6:Remote hold input. No hold
when open. Output hold when
short-circuited.
For details, refer to “Item 6.7
Parameter setting, Output Hold”.
Between 7–8:Average value reset input. short-
circuitting the contact input (for
at 1.5 sec min.) resets O
and O
corrected average simulta-
2
average
2
neously. Opening it restarts the
average value.
For details, refer to “Item 6.7
Parameter setting, Average Value
Resetting”
Between 9–10:Automatic calibration remote
start input
After shorting for 1.5 sec. or
more, automatic calibration is
started by the opening input
whether the automatic calibration
setting is ON/OFF.
For details, refer to “Item 6.4
Setting of auto calibration”
Between 11–12:Ch5 range identification contact
output
Between 13–14:Ch4 range identification contact
output
Between 15–16Ch3 range identification contact
output
Between 17–18:Ch2 range identification contact
output
Between 19–20:Ch1 range identification contact
output
IM 11G02N01-01E
Action of range identification signal
Range identification contact is
conductive at low range and open
at high range.
3 - 13
Peak count alarm
output (PEAK_ALM)
Auto calibration status/contact
contact output (PUMP)
output (CAL)
Pump ON/OFF
Calibration error
contact output
(CAL_ALM)
Instrument error
contact output
(FAULT)
Terminal block 4
<TN4>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Unassigned
Unassigned
Unassigned
Unassigned
Unassigned
Terminal 4 <TN4>
Between 1–2:Peak count alarm contact output
It is conductive when peak count
exceeds the setting time. It
remains open below the setting
time. For setting and operation,
refer to “Item 6.6 Peak alarm
setting”.
Between 3–4:Contact output of auto calibration
status
When the auto calibration is
carried out and remote hold is
ON, it is conductive. Remains
open otherwise.
Between 5–6:Pump ON/OFF contact output
Used when turning ON/OFF the
pump. It is open during auto and
manual calibration status and
conductive during measurement.
Between 7–8:Calibration error contact output
It is conductive when an error
occurs during zero calibration or
span calibration. It is normally
open.
Between 9–10:It is conductive when an error
occurs to the analyzer unit. It is
normally open.
Between 11–20:For internal connection, wiring is
not allowed. (Do not use it as
junction terminal).
3 - 14
IM 11G02N01-01E
Unassigned
Ch3 alarm output
(ALM_Ch3)
Ch2 alarm output
(ALM_Ch2)
Ch1 alarm output
(ALM_Ch1)
Terminal block 5
<TN5>
11
1
12
2
13
3
14
4
15
5
16
6
17
7
18
8
19
9
20
10
(M3.5 screw)
Unassigned
Power disconnection
alarm output
(POWER_OFF)
Ch5 alarm output
(ALM_Ch5)
Ch4 alarm output
(ALM_Ch4)
Terminal 5 <TN5>
Between 2, 3 and 4:
Ch3 alarm output
When the output exceeds the set value,
it is conductive between 2 and 3, and
open between 3 and 4. Otherwise, it is
open between 2 and 3 and conductive
between 3 and 4.
Between 5, 6 and 7:
Ch2 alarm output
When the output exceeds the set value,
it is conductive between 5 and 6, and
open between 6 and 7. Otherwise, it is
open between 5 and 6, and conductive
between 6 and 7.
Between 8, 9 and 10:
Ch1 alarm output
When the output exceeds the set value,
it is conductive between 8 and 9, and
open between 9 and 10. Otherwise, it
is open between 8 and 9.
Between 12, 13 and14:
Analyzer unit power OFF output
When the analyzer unit is turned ON,
it is conductive between 12 and 13,
and open between 13 and 14. When
the analyzer unit is turned OFF, it is
open between 12 and 13, and conductive between 13 and 14.
Between 15, 16 and 17:
Ch5 alarm output
When the output exceeds the set value,
it is conductive between 15 and 16,
and open between 16 and 17. Otherwise, it is open between 15 and 16,
and conductive between 16 and 17.
Between 18, 19 and 20:
Ch4 alarm output
When the output exceeds the set value,
it is conductive between 18 and 19,
and open between 19 and 20. Otherwise, it is open between 18 and 19,
and conductive between 19 and 20.
For detailed action of the alarm
contact, refer to “Item 6.3 Alarm
setting”.
IM 11G02N01-01E
3 - 15
Connector <CN3>
Solenoid valve drive signal output for calibration
Connector
<CN3>
(Transister output)
Contact output for
sample gas selection
Contact output for
zero calibration
Contact output for Ch1
span calibration
Contact output for Ch2
span calibration
1
2
3
4
5
<D-sub 9P>
Contact output for Ch3
6
span calibration
Contact output for Ch4
7
span calibration
Contact output for Ch5
8
span calibration
Power for relay drive
9
5V DC
Connector <CN3> provides outputs in combination with calibration action during auto calibration
and manual calibration.
An output is from a transistor (ratings: 5 V/50 mA).
A transistor is turned ON before starting each calibration.
Sample selection output is ON during measurement and OFF during calibration.
If calibration is not performed, the other transistors are OFF.
In case of auto calibration, sequential output is ON/OFF according to the setting.
Refer to “Item 6.4 Setting of auto calibration”.
Note) No. 9 pin is for solenoid valve ON/OFF relay drive power (5 V DC/0.5 A, max).
Use No. 9 with reference to the diagram.
Example of using solenoid valve drive signal output for calibration
<Electrical system>
SV4
Ch1
span
Contact relay
Relay board
Ch2
span
5V
I/O terminal module
<Gas sampling system>
Electronic
gas cooler
9
1
3
4
5
Cable
CN3
Flowmeter
Membrane filter
SV3
SV2SV1
Zero
Standard gas for calibration
SV1
SV2
SV3
SV4
Solenoid valve
drive power
SV1 to SV4: solenoid valves
Refer to Item 3.4.7 Example configuration
of gas sampling system .
3 - 16
Relay board and exclusive cable (D-sub 9p straight cable: 1.5 meters)
are available on request.
IM 11G02N01-01E
¥ Zero calibration
¥ Span calibration
Zero calibration
output
Zero calibration
output
Pump ON/OFF contact
Pump ON/OFF contact
Sample selection
output
Ch1 to 5 span
calibration output
Ch1 span
calibration
Output hold function
Sample selection
output
(with hold ON setting)
ZEROENT
ENT
Hold extension time.
Calibration gas flow
Calibration gas flow
Output signal hold
Output signal hold
Calibration end
Calibration end
on
off
on
off
on
off
off
Output hold function
(with hold ON setting)
SPANENT
ENT
Hold extemsion time.
Note)
When selecting
Ch2 using
and
keys.
Ch2 span
calibration
The hold extension time
depends on the gas
flow time of the
automatic calibration
settings.
(7) Timing of solenoid valve drive signal for calibration
1. Manual calibration (see “Item 6.9 Calibration”.)
IM 11G02N01-01E
3 - 17
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.