Yokogawa GD40 User Manual

User’s Manual
Model GD402 Gas Density Meter
IM 11T3E1-01E
R
IM 11T3E1-01E
7th Edition (YK)
<Introduction>

Introduction

The GD402 gas density meter and the GD40 detector not only provide continuous measurement of gas density, but also several other valuable parameters, including specic gravity and molecu­lar weight. The GD40 detector is designed for intrinsically safe and ame-proof, explosion pro­tected applications. It is designed to be virtually maintenance free for all accepted applications.
The GD402 is a rugged microprocessor-based converter designed in two versions to meet both general area and explosion-proof application requirements. In addition to the display of several key data items, the converter also provides the choice of three different means for calibration: automatic; semi-automatic and one-touch manual operation.
Before use, read this manual thoroughly in order to fully realize the capabilities of the Meter. The topics and information that require special attention when handling the product are given in the text of this manual along with cautionary notes - WARNING or CAUTION - which reect the importance of the information.
For safety reasons or to avoid possible damage to your equipment, strictly adhere to every cautionary note that appears in this manual. Cautionary notes, such as a WARNING, that also appear on the product itself are indicated together with the alert symbol shown below:
Throughout this user’s manual, you will nd several different types of symbols are used to identify different sections of text. This section describes these icons.
WARNING
Identies instructions that must be observed in order to avoid physical injury and electric shock or death of the operator.
CAUTION
Identies instructions that must be observed in order to prevent the software or hardware from being damaged or the system from becoming faulty.
NOTE
Identies important information required to understand operations or functions.
TIP
Identies additional information.
SEE ALSO
Identies a source to be referred to.
Clicking a reference displayed in green can call up its source, while clicking a reference
displayed in black cannot.
Media No. IM 11T3E1-01E (CD) 7th Edition : June 2012 (YK) All Rights Reserved Copyright © 1998, Yokogawa Electric Corporation
IM 11T3E1-01E
<Introduction>
l Cross-Checking the Specications
CAUTION
The GD402 gas density meter is shipped after adjusting the both of detector and converter in pairs. When installation, conrm whether the serial number on both of converter and detector are in pairs or not. If mismatched in pairs, converter is to be out of order. When converter or detector supplied individually, enter the detector constants, described on inside the lid of the GD40, into converter so that the GD402 is going to be well.
In detail, please refer to Figure 5.30, Figure 6.31 or Figure 7.28.
Upon delivery of the purchased product, unpack it carefully and make sure it is completely free from damage that may have occurred during transport. It must be shipped in strict conformance to the purchaser's specications. By way of precaution, conrm that the equipment is the exact model you ordered. Also check that all accessory components are included, when conrming the specications, refer to the model and sufx codes indicated on the nameplate on the equipment. For a description of the model and sufx codes, refer to Chapter1,''Specications.''
l Consideration of Operation Parameters
ii
The GD402 operates with the same parameters set when it was delivered (default data) , when it is put into operation under these conditions.
Before starting measurement, check whether or not the default data meets your operating condi­tions. If necessary, re-set the parameters to suit your operating requirements. To check the defaults, make use of the sheet, “Records of GD402’s Operation Parameter Set­tings,” in the back of this manual. It is advisable that, if any of the operation parameter settings have been changed, the new data be noted in this record.
l Information Covered in This Manual
This manual covers all of the information for handling the GD402 converter and the GD40 detec­tor, including instructions on installation, operation, setting of operation parameters, inspection and maintenance.
CAUTION
Modication and adaptation of the instrument’s parts, wiring and construction are prohibited.
IM 11T3E1-01E
<Introduction>
[Notational Conventions Specic to This Manual]
The following notational conventions apply to the representations of operation keys, information shown on the display, and information indicated on the product itself when they are discussed specically in the text of this manual.
l Operation Keys
Indicated with brackets [ ] or【 】as: [YES] Key or 【YES】key
l Information Shown on Display
Indicated with braces 『 』as:HOLD』, meaning the statusYES』, meaning the indicator on an operation keyCALIB』, meaning a message 205(lit) or205(blinking), meaning the data item shown along with its state
l Information Indicated on Product
Indicated with angle brackets < > as: <n> (lit) or <o> (unlit), meaning the contact output indicator
lamp along with its state
iii
<MEASURE> mode, meaning the mode of measurement
l Information on the State of Blinking
Indicated in shaded typography as:
, in contrast to (lit) (blinking)
[How to dispose the batteries]
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This direc-
tive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.ofce. Do not
dispose them as domestic household waste.
Battery type: silver oxide battery
Notice : The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in DIRECTIVE 2006/66/EC.
IM 11T3E1-01E
<Introduction>

u After-sales Warranty

n Do not modify the product.
n During the warranty period, for repair under warranty carry or send the prod-
uct to the local sales representative or service ofce. Yokogawa will replace
or repair any damaged parts and return the product to you. Before returning a
product for repair under warranty, provide us with the model name and serial
number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
l If we replace the product with a new one, we won’t provide you with a repair report.
l Yokogawa warrants the product for the period stated in the pre-purchase quotation
Yokogawa shall conduct dened warranty service based on its standard. When the custom­er site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
n In the following cases, customer will be charged repair fee regardless of war-
ranty period.
iv
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa Electric did not supply.
• Failure due to improper or insufcient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa does not authorize.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare, radiation and other natural changes.
n Yokogawa does not warrant conformance with the specic application at the
user site. Yokogawa will not bear direct/indirect responsibility for damage due to a specic application.
n Yokogawa Electric will not bear responsibility when the user congures the
product into systems or resells the product.
n Maintenance service and supplying repair parts will be covered for ve years
after the production ends. For repair for this product, please contact the near­est sales ofce described in this instruction manual.
IM 11T3E1-01E
<CONTENTS>
Model GD402 Gas Density Meter
IM 11T3E1-01E 7th Edition

CONTENTS

Introduction ..............................................................................................................i
u After-sales Warranty ...................................................................................................... iv
1. SPECIFICATIONS ..................................................................................... 1-1
1.1 Specications ....................................................................................................1-1
1.1.1 General Specications .......................................................................1-1
1.1.2 GD40G, T, V, R Detector ....................................................................1-5
1.1.3 GD402G, T, V, R Converter................................................................ 1-5
1.2 Model and Sufx Codes ...................................................................................1-7
1.2.1 Gas Density Converter ......................................................................1-7
1.2.2 Gas Density Detector .........................................................................1-7
1.2.3 Hardware for Connection with External Cables
(For Explosion-proof use) ..................................................................1-8
1.2.4 Two-core, Double-Shielded Cable ..................................................... 1-8
1.2.5 Brain Terminal (Optional) ................................................................... 1-9
1.2.6 Pressure transmitter (Optional)..........................................................1-9
1.3 External Views and Dimension ......................................................................1-10
1.3.1 GD402G Converter (General purpose) ...........................................1-10
1.3.2 Pipe- and Wall-Mounting Hardware (Optional) ................................ 1-11
1.3.3. GD402T, V, R Converter (Explosion-proof) .....................................1-12
1.3.4 GD40 Detector .................................................................................1-13
1.3.5 Detector Unit (Intrinsically Safe, Explosion-proof) ...........................1-14
1.3.6 Surge protector: K9215LP ...............................................................1-14
Toc-1
2. INSTALLATION, WIRING AND PIPING .................................................... 2-1
2.1 Installing the Detector ......................................................................................2-1
2.1.1 Selecting the Location ........................................................................ 2-1
2.1.2 GD40T (FM Explosion-proof and Intrinsically Safe Approval) ...........2-2
2.1.3 GD40V (CSA Explosion-proof and Intrinsically Safe Approval) ........2-3
2.1.4 Mounting the Detector........................................................................2-4
2.2 Installing the Converter ....................................................................................2-5
2.2.1 Selecting the Location ........................................................................ 2-5
2.2.2 GD402T (FM Explosion-proof Approval) ...........................................2-6
2.2.3 GD402V (CSA Explosion-proof Approval) .........................................2-7
2.2.4 Mounting the Converter .....................................................................2-8
IM 11T3E1-01E
<CONTENTS>
2.3 Piping ...............................................................................................................2-10
2.4 Wiring ...............................................................................................................2-12
2.4.1 Wiring Procedure .............................................................................2-12
2.4.2 Cables Wired to Power Supply ........................................................2-16
2.4.3 Cables Wired to Outputs ..................................................................2-17
2.4.4 Cables Wired to Contact I/Os ..........................................................2-18
2.4.5 Cable Wired to GD40 Detector .....................................................2-19
2.4.6 Cables Wired to the Ground ............................................................2-20
3. OPERATION .............................................................................................. 3-1
3.1 Preparation for Operation ................................................................................3-1
3.1.1 Inspecting Installation, Piping and Wiring Workmanship .................. 3-1
3.1.2 Supplying Power ................................................................................ 3-2
3.1.3 Display on Operation Panel and Operation Keys ..............................3-3
3.1.4 Basic Key Operation ..........................................................................3-4
3.1.5 Checking the Setting Parameters .....................................................3-6
3.1.6 Calibrating the Analyzer for Correct Readings ..................................3-6
3.1.7 Checking the Analyzer for Performance ............................................3-7
3.2 Normal Operation .............................................................................................. 3-8
3.2.1 Starting Operation ..............................................................................3-8
3.2.2 Corrective Actions in Case of Failure ................................................. 3-8
3.2.3 Inspection and Maintenance ..............................................................3-9
3.3 Shutdown and Restart ......................................................................................3-9
3.3.1 Measures for Shutdown .....................................................................3-9
3.3.2 Measures for Restarting ....................................................................3-9
Toc-2
4. FUNCTIONS .............................................................................................. 4-1
4.1 Summary of Setting Operations ...................................................................... 4-1
4.1.1 Measurement, Operation, Setting and Service .................................4-1
4.1.2 Key Operations ..................................................................................4-1
4.1.3 Points to Be Noted When Making Settings ........................................4-2
4.2 Setting Lists .......................................................................................................4-2
5. DENSITY / PARAMETER SETTING ......................................................... 5-1
5.1 Setting Parameters ...........................................................................................5-2
5.1.1 Setting Parameters at Measurement Level .......................................5-2
5.1.2 Setting Parameters at Operation Level ............................................. 5-3
5.1.3 Setting Parameters at Setting Level ..................................................5-4
5.1.4 Setting Parameters at Service Level .................................................5-5
5.2 Parameter Setting .............................................................................................5-6
5.2.1 Parameter Setting at Operation Level ...............................................5-6
5.2.2 Parameter Setting at Setting Level ..................................................5-10
5.2.3 Parameter Setting at Service Level .................................................5-16
6. CALORIE / PARAMETER SETTING ........................................................ 6-1
6.1 Setting Parameters ...........................................................................................6-2
IM 11T3E1-01E
<CONTENTS>
6.1.1 Setting Parameters at Measurement Level .......................................6-2
6.1.2 Setting Parameters at Operation Level ............................................. 6-3
6.1.3 Setting Parameters at Setting Level ..................................................6-4
6.1.4 Setting Parameters at Service Level .................................................6-5
6.2 Parameter Setting .............................................................................................6-7
6.2.1 Parameter Setting at Operation Level ...............................................6-7
6.2.2 Parameter Setting at Setting Level ..................................................6-12
6.2.3 Parameter Setting at Service Level .................................................6-16
7. HYDROGEN PURITY/PARAMETER SETTING ....................................... 7-1
7.1 Setting Parameters ...........................................................................................7-2
7.1.1 Setting Parameters at Measurement Level .......................................7-2
7.1.2 Setting Parameters at Operation Level ............................................. 7-2
7.1.3 Setting Parameters at Setting Level ..................................................7-3
7.1.4 Setting Parameters at Service Level .................................................7-4
7.2 Parameter Setting .............................................................................................7-6
7.2.1 Parameter Setting at Measurement Level ......................................... 7-6
7.2.2 Parameter Setting at Operation Level ...............................................7-7
7.2.3 Parameter Setting at Setting Level ..................................................7-10
7.2.4 Parameter Setting at Service Level .................................................7-14
Toc-3
8. DENSITY /CALIBRATION PROCEDURE ................................................ 8-1
8.1 Basic Calibration Procedure ............................................................................8-1
8.2 Preparation for Calibration ..............................................................................8-1
8.2.1 Setting Type of Calibration and Checking Valve Operation ...............8-1
8.2.2 Setting Calibration Data ..................................................................... 8-3
8.3 Calibration .........................................................................................................8-4
8.3.1 Semi-automatic Calibration ............................................................... 8-4
8.3.2 Manual Calibration .............................................................................8-5
8.3.3 Automatic Calibration .........................................................................8-6
9. CALORIE /CALIBRATION PROCEDURE ................................................ 9-1
9.1 Basic Calibration Procedure ............................................................................9-1
9.2 Preparation for Calibration ..............................................................................9-1
9.2.1 Setting Type of Calibration and Checking Valve Operation ...............9-1
9.2.2 Setting Calibration Data ..................................................................... 9-3
9.3 Calibration .........................................................................................................9-5
9.3.1 Semi-automatic Calibration ............................................................... 9-5
9.3.2 Manual Calibration .............................................................................9-6
9.3.3 Automatic Calibration .........................................................................9-7
10. HYDROGEN PURITY /CALIBRATION PROCEDURE .......................... 10-1
10.1 Basic Calibration Procedure ..........................................................................10-1
10.2 Preparation for Calibration ............................................................................10-1
10.3 Calibration .......................................................................................................10-2
IM 11T3E1-01E
<CONTENTS>
11. INSPECTION AND MAINTENANCE ...................................................... 11-1
11.1 Routine Inspection and Maintenance ........................................................... 11-1
11.1.1 Checking Readings and Calibrating the Analyzer ........................... 11-1
11.1.2 Checking the Flowrate of Sample Gas ............................................ 11-1
11.1.3 Periodic Replacement of the Detector's O-rings ............................. 11-1
11.1.4 Replacing the Fuse .......................................................................... 11-2
11.1.5 Cleaning ........................................................................................... 11-2
11.2 Inspection In Case of Failure ......................................................................... 11-3
11.2.1 Inspecting the Analyzer in an Alarm Status ..................................... 11-3
11.2.2 Inspecting the Analyzer in a FAIL Status ......................................... 11-4
Customer Maintenance Parts List ...........................................CMPL 11T3E1-01E
Customer Maintenance Parts List ...........................................CMPL 11T3E1-02E
Customer Maintenance Parts List ...........................................CMPL 11T3E1-03E
Revision Information ...............................................................................................i
Toc-4
IM 11T3E1-01E
<1. Specications>

1. SPECIFICATIONS

1.1 Specications
The specications of the GD402 gas density meter as described below.
Certicate is approved for model GD402G, GD40G.
1.1.1 General Specications
l System Components
(1) GD40G, T, V, R Detector:Rainproof for outdoor use (equivalent to IP65/NEMA4X)
(see note under “1.1.2 Ambient conditions” on page1-5.)
Ambient Temperature : -10 to 60°C
Ambient Humidity : 5 to 95%RH
GD40G : General purpose detector.
1-1
(Non Explosion-proof)
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40T : FM Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Enclosure : NEMA Type 4X
Temperature Code : T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40V : CSA Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Enclosure : Type 4X
Temperature Code : T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40R : TIIS Explosion-proof and Intrinsically safe Approval.
Explosion-proof code : Exd [ia] IIB+H2T5
Temperature Code : T5
Electrical connection : G1/2 female
Process connection : Rc1/4
IM 11T3E1-01E
<1. Specications>
(2) GD402G, T, V, R Converter : Rainproof for outdoor use (equivalent to IP65 / NEMA 4X)
Ambient Temperature : -10 to 55°C
Ambient Humidity : 5 to 95%RH
GD402G : General purpose converter. (Non Explosion-proof).
Electrical connection : 21mm (0.9inch) in diameter. Pg 13.5 cable glands includ-
ed
GD402T : FM Explosion-proof Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
Enclosure : NEMA Type 4X
Temperature Code : T6
Electrical connection : 1/2 NPT female
GD402V : CSA Explosion-proof Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
1-2
Enclosure : Type 4X
Temperature Code : T6
Electrical connection : 1/2 NPT female
GD402R : TIIS Explosion-proof Approval.
Explosion-proof code : Exd IIB+H2T6
Temperature Code : T6
Electrical connection : G3/4 female
(3) EJA310 Absolute pressure transmitter (optional)
FM Explosion-proof Approval:
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
Hazardous(classied locations, indoors and outdoors (NEMA 4X)
Temperature Code : T6
Ambient Temperature : -40 to 60°C
Ambient Humidity : 5 to 100%RH (at 40°C)
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
CSA Explosion-proof Approval:
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.
Division2 ‘SEALS NOT REQUIRED’
Enclosure : Type 4X
Temperature Code : T4, T5, T6
Max. Process Temp.: T4 120°C, T5 100°C, T6 85°C
Ambient Temperature : -40 to 80°C
IM 11T3E1-01E
<1. Specications>
Ambient Humidity : 5 to 100%RH (at 40°C)
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
TIIS Explosion-proof Approval:
Explosion-proof code : Ex do IIC T4X
Temperature Code : T4
Ambient Temperature : -20 to 60 °C
Ambient Humidity : 5 to 100%RH (at 40°C)
Electrical connection : G1/2 female
Process connection : Rc1/4
l Sample Gas conditions
Sample gas: All gases except for corrosive gas and acetylene gas.
Temperature: -10 to 50 °C (non-condensing)
Pressure: Max. 588.5 kPa (abs)
Gas ow: 0.1 to 1 l/min
1-3
l Output Signals
Output 1: 4-20 mA DC
Isolated from inputs; load resistance: 600 W maximum (Load resistance of 250-
550 W required when in the BRAIN communication mode.)
Output 2: 4-20 mA DC
Isolated from inputs; load resistance: 600 W maximum
l Power Supply
Rated voltage range: 100 to 240 V AC, 24 V DC
Allowable voltage range: 85 to 264 V AC, 21.6 to 26.4 V DC
Rated frequency: 50 or 60 Hz
Allowable frequency range: 47 to 63 Hz
l Power Consumption
Approximately 12 W.
l Safety and EMC Compliance
Safety Standards: EN61010-1
EMC Standards:
(Applied only when GD402G converter is used with GD40G detector.)
Emission: EN61326 Class A
EN61000-3-2
EN61000-3-3
AS/NZS CISPR11
Immunity: EN61326 Annex A and F
KC Marking: Korea Electromagnetic Conformity Standard
IM 11T3E1-01E
<1. Specications>
l Characteristics
GD402 specication list
Item Density kg/m3 Density lb/ft3 Specic Gravity Molecular Weight Concentration vol%
Range
0 - 6 (compensated) 0 - 60 (physical)
Minimum Range 0.1 0.01 0.1 4
Response Time 90% approx. 5 sec approx. 5 sec approx. 5 sec approx. 5 sec approx. 5 sec
Linearity +/-1 % FS +/-1 % FS +/-1 % FS +/-1 % FS +/- 1
Repeatability
+/- 0.001 or +/-0.5%FS *
Long term stability +/- 0.003/month +/- 0.002/month +/- 0.003/month +/- 0.07/month
*: Whichever is greater
Density is the basic measurement, the other representations are derived from the Density data.
Item H2 in Air vol% H2 in CO2 vol% Air in CO2 vol% Caloric value MJ/m
Range 85 - 100 0 - 100 0 - 100 0 - 130 0 - 3.5
Minimum Range
Response Time 90%
Linearity +/- 1 +/- 1 +/- 1 +/- 1 % FS +/- 1 % FS
Repeatability +/- 0.5 +/- 0.5 +/- 0.5 +/- 0.5%FS or Caloric value
Drift +/- 0.5/month +/- 0.5/month +/- 0.5/month Caloric value equivalent to
approx. 5 sec approx. 5 sec approx. 5 sec approx. 5 sec approx. 5 sec
0 - 0.4 (compensated) 0 - 4 (physical)
+/- 0.0001 or +/-0.5%FS*
0 - 5 0 - 140 0 - 100
Concentration equivalent to 100 kg/m3
+/- 0.001 or +/-0.5%FS*
+/- 0.02 or +/-0.5%FS*
+/- 0.5% or Concentration equivalent to +/-0.001kg/m3 *
Concentration equivalent to +/- 0.003 kg/m3/month
3
British Thermal Unit KBTU/ft
Caloric value equivalent to
0.100 kg/m
3
Caloric value equivalent to
0.100 kg/m
+/- 0.5%FS or Caloric value
equivalent to 0.001kg/m3*
equivalent to 0.001kg/m3 *
Caloric value equivalent to
+/- 0.003kg/m3 /month
+/- 0.0025/month
3
1-4
3
Caloric Value and BTU are possible representations of the Density. GD402 does not contain table information, only a single mathematical equation.
*: Whichever is greater
IM 11T3E1-01E
<1. Specications>

1.1.2 GD40G, T, V, R Detector

l Material exposed to gas
SUS 316 stainless steel, Acrylonitrile Butadiene Rubber and Fluorine-contained Rubber
(O-ring)
l Ambient conditions
Temperature: -10 to 60 °C (14 to 140 °F)
Humidity: 5 to 95 %RH
Installation: Pipe-mounted or on panel
Construction: Explosion-proof and Intrinsically safe. (See 1.1.1 “System Components”)
Though the detector construction makes it relatively insensitive to sudden changes in
the gas temperature, extra precision can be achieved by keeping ambient tempera­ture conditions as constant as possible. In measurements where optimum precision is required it is therefore not recommended to install the detector in an outdoors environ­ment, especially not if such installation is prone to direct sunlight.
l Finish
Cover: Equivalent to Munsell 0.6GY3.1/2.0
1-5
Case: Equivalent to Munsell 2.5Y8.4/1.2
l Weight
Approximately 7 kg (with pipe mounting bracket)
l Detector unit
When the system is ordered to be used as a hydrogen purity meter, an optional pressure
transmitter is required for pressure compensation.
• If /EJAJ1, /EJAF2, /EJAF3 or, /EJAF4 are ordered, the detector unit and the
pressure transmitter and the tubing in between will all be integrated on a single mounting plate. This allows the space where the pressure transmitter is normally mounted to be used effectively for other purposes.

1.1.3 GD402G, T, V, R Converter

l Display
Reading: Digital (6 digits maximum; resolution: 0.0001 kg/m3)
Data items shown:
Measured value: Always on display.
Alarm indications: Abnormal concentration, abnormal pressure range of input and abnormal
value of calibration.
Parameters for calibration: Time of calibration, settling time, starting time of calibration and
calibration cycle
Self-diagnostic indications: Sensor oscillation shutdown, abnormal oscillation frequency of
sensor, failure in sensor temperature detection, failure in A/D conversion stage and memory failure.
Alarm settings: The contact state can be set to either “normally open (NO)” or “normally
closed (NC)” depending on the needs of the application.
Temperature: Temperature of gas being measured
IM 11T3E1-01E
<1. Specications>
l Contact Outputs/Input
Contact output: Signals for Maintenance, Fail, Hi/Lo alarms
Contact capacity: 250 V AC at 3 A or 30 V DC at 3 A
Contact input: Signal for switching between the Hydrogen Purity meter and the Replace-
ment meter
l Calibration
Manual (one touch), Semi automatic, Automatic calibration
l Communication
Protocol: BRAIN communication
Data items that can be transmitted by the hand-held terminal are numerical data, such as concentration, temperature and pressure, alarm setpoint and self-diagnostic parameters.
l Ambient Conditions
Temperature: -10 to 55 °C (14 to 131°F)
Humidity: 5 to 95 %RH
l Installation
1-6
Non-explosion-proof models: Pipe-, panel- or wall-mounted
Explosion-proof models: Pipe- or panel-mounted
l Finish
GD402G (general purpose):
Front cover: Equivalent to Munsell 0.6GY3.1/2.0
Case: Equivalent to Munsell 2.5Y8.4/1.2
GD402T, V, R (explosion-proof):
Equivalent to Munsell 0.6GY3.1/2.0
l Weight
GD402G (general purpose): Approx. 3 kg (including mounting hardware)
GD402T, V, R (explosion-proof): Approx. 15 kg (including mounting hardware)
l Fuse
250 V 1 A quick acting type authorized VDE/SEMKO 100 to 240 V AC Model
250 V 2 A quick acting type authorized VDE/SEMKO 24 V DC Model
IM 11T3E1-01E
<1. Specications>
1.2 Model and Sufx Codes

1.2.1 Gas Density Converter

Model Sufx code Option code Description
GD402G - - - - - - - - - - - - - - - - - - - - - - - - - General purpose model, 6 cable glands included.
GD402T - - - - - - - - - - - - - - - - - - - - - - - - - FM certied explosion - proof model. Gland threads 1/2 NPT. No cable glands
GD402V - - - - - - - - - - - - - - - - - - - - - - - - - CSA certied explosion - proof model. Gland threads 1/2 NPT. No cable glands
GD402R - - - - - - - - - - - - - - - - - - - - - - - - - TIIS certied explosion - proof model. Gland threads G3/4. No cable glands included.
Power supply-D-A
Label and approval
Instruction Manual -E - - - - - - - - - - - - English
Options (only GD402G)
Note: Explosion -proof models, GD402T, V, R have only pipe mounting hardware as standard.
-E
-J
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
- - - - - - - - - - - -
/PA /U /H
included.
included.
24 V DC 100-240 V AC
English label TIIS approval, English label (only GD402R)
Panel mounting Universal (Pipe and Wall) Mounting Mood
1-7
GD402 Standard Accessory List
Model Item Qty Part Number
GD402G
GD402R GD402T GD402V
Fuse 1 A1109EF (Power Supply: 100-240 V AC)
Universal Mount Set 1 K9171SS
Panel Mount Set 1 K9171ST
Surge Protector 1 K9215LP
Fuse 1 A1109EF (Power Supply: 100-240 V AC)
Blacket 1 K9214HD
Blacket 1 K9214HE
U-Bolt Assy 2 D0177XL-A
Bolt 1 Y9835NU
Bolt 2 Y9820NU
A1111EF (Power Supply: 24 V DC)
A1111EF (Power Supply: 24 V DC)

1.2.2 Gas Density Detector

Model Sufx code Option code Description
GD40G - - - - - - - - - - - - - - - - - - - - - - - - - General purpose detector. 1/4 NPT gas threads and 1/2 NPT gland threads. No cable
GD40R - - - - - - - - - - - - - - - - - - - - - - - - - TIIS certied explosion - proof detector. Rc1/4 gas threads and G1/2 gland threads.
Label approval-E-J
Options /EJAJ1
- - - - - - - - - - - -
- - - - - - - - - - - -
/EJAJ1T
/EJAF2
/EJAF2T
gland included. Mounting hardware included.
Cable gland included. Mounting hardware included.
English label, no approval (only GD40G) TIIS approval, English label (only GD40R)
TIIS certied EJA mounted with detector on mounting plate. Rc1/4 gas threads and G1/2 gland thread. Cable gland included.(only GD40R) TIIS certied EJA mounted with detector on mounting plate. Rc1/4 gas threads and G1/2 gland thread. Cable gland included. EJA with TAG (only GD40R) EJA mounted with detector on mounting plate.1/4 NPT gas threads and 1/2 NPT gland threads. No cable gland included.(only GD40G) EJA mounted with detector on mounting plate.1/4 NPT gas threads and 1/2 NPT gland threads. No cable gland included. EJA with TAG (only GD40G)
IM 11T3E1-01E
<1. Specications>
Model Sufx code Option code Description
GD40T - - - - - - - - - - - - - - - - - - - - - - - - - FM certied explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT gland thread.
GD40V - - - - - - - - - - - - - - - - - - - - - - - - - CSA certied explosion - proof detector. 1/4 NPT gas threads and 1/2 NPT gland
Options /EJAF3
/EJAF3T
/EJAF4
/EJAF4T
No cable gland included. Mounting hardware included
thread. No cable gland included. Mounting hardware included.
FM certied EJA mounted with detector on mounting plate.1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included. (only GD40T) FM certied EJA mounted with detector on mounting plate.1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included. EJA with TAG (only GD40T)
CSA certied EJA mounted with detector on mounting plate. 1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included. (only GD40V) CSA certied EJA mounted with detector on mounting plate. 1/4 NPT gas threads and 1/2NPT gland thread. No cable gland included. EJA with TAG (only GD40V)
GD40 Standard Accessory
Item Qty Part Number
U-Bolt Assy*
Blacket*
Blacket*
Gland*
*1: Not supplied when option code “/EJAJ1”, “/EJAF2”, “/EJAF3” or “/EJAF4” is specied. *2: Supplied only for GD40R.
1
4 D0177XL-A
1
1
2
1 K9214HD
1 K9214HE
1 G9601AM
1-8

1.2.3 Hardware for Connection with External Cables (For Explosion-proof use)

Part No. Description
L9811LL G3/4 explosion-proof cable gland.
Cable’s outside diameter 8 to 16 mm. Used for the GD402R converter.
Note: Specify the number of cable glands for converter in hazardous area

1.2.4 Two-core, Double-Shielded Cable

Normally two conductor shielded cable can be used, but when failure arises from noises disturbance, this cable is recommended for connection between the GD402 converter and GD40 detector.
Model Basic code Description
GDW Two - core, double-shielded cable,
both ends nished with cable pins.
Length
-L
Length in meters, 500 meter maximum.
IM 11T3E1-01E
<1. Specications>

1.2.5 Brain Terminal (Optional)

Model Sufx code Option code Description
BT200 - - - - - - - - - - - - - - - - - - - - - - - - - Brain terminal [Note]
Printer -N
-P
00 - - - - - - - - - - - - Always 00
Options
Note: The BT200 has following accessories, two communication cables, one with IC clips and another with alligator clips, handy
carrying case and ve AA 1.5 V dry batteries.
OPTIONS FOR BT200
Options Description Option codes
Communication cable (Note 1)
Intrinsically safe type (Note 1) (Note 2)
(Note 1) Optional code /C1 can not be combined with /CS1. (Note 2) Applicable only for Model BT200-N00.
See GS 01C00A11-00EN for “BT200” brain terminal in detail.
- - - - - - - - - - - -
- - - - - - - - - - - -
/
With a 5-pin connector (for the signal conditioner)
CSA Intrinsically safe approval Class I, Groups A, B, C and D Temp. Code : T4
Standard type (without printer) With printer
/C1
/CS1
1-9

1.2.6 Pressure transmitter (Optional)

/EJAJ1 means TIIS certied EJA310.
/EJAF2 means general purpose model EJA310.
/EJAF3 means FM certied EJA310.
/EJAF4 means CSA certied EJA310.
See GS 01C22D01-00E for “EJA310” pressure transmitter in detail if a different
selection from pre-selected options seems necessary.
IM 11T3E1-01E
<1. Specications>

1.3 External Views and Dimension

1.3.1 GD402G Converter (General purpose)

1-10
Unit: mm(in.)
184 (7.2)
144(5.7)
144 (5.7)
A B
D F
36 36
Grounding terminal (4 mm screws)
C
E
(1.4)(1.4)
Hood (Option code: /H)
220 (8.7)
72 (2.8)
20 (0.8)
112 (4.4)
23
(0.9)
Cable inlet (21 mm in dia.) Compatible with a PG 13.5 cable gland
Cable gland Connection
36
38
(1.4)
(1.5)
A, B
C D, E
F
Four holes, 6 mm (0.24) in dia., 8 mm deep M6
80 (3.1)
80 (3.1)
• Pressure transmitter
• Analog Output • Contact Input
• Detector
• Contact Output
• Power Supply
Weight : Approx. 3 kg (6.6 pounds) (including mounting hardware)
F0101.ai
IM 11T3E1-01E
<1. Specications>

1.3.2 Pipe- and Wall-Mounting Hardware (Optional)

1-11
• Hardware for Pipe Mounting (Option code /U)
188 (7.4)
174 (6.9)
(2)
50
Mounting pipe of JIS 50 A (60.5 mm in outer dia.)
Optional hardware
4 - M6 screws
• Hardware for Wall Mounting (Option code /U)
135 (5.3)
13
(0.51)
Unit : mm (in.)
Weight : Approx. 3 kg
200 (7.9)
100 (3.9)
4 - M6 screws
224 (8.8)
Optional hardware
3-Ø10 holes
• Hardware for Panel Mounting (Option code /PA)
23
(0.9)
max. 12 (panel thickness)
100 (3.9)
4 - M6 screws
Optional hardware
178 (7)
200 (7.9)
35
(1.4)
(0.6)
70 (2.8)
100 (3.9)
185 (7.3)
15
Dimensions of panel cutout
+1
139
0
+0.08
(5.47 )
0
139
+1
0+1 0
(5.47 )
F0102.ai
195 (7.7)
IM 11T3E1-01E
<1. Specications>

1.3.3. GD402T, V, R Converter (Explosion-proof)

Weight : approx. 15 kg (33.1 pounds)
Pipe-mounting hardware
Mounting pipe of JIS 50 A nominal size (Φ60.5)
1-12
Unit : mm (in.)
Φ212 (8.3)
Side view from arrow A direction
Cable glands not included
4 - M8 depth 20
140 (5.5)
24 (0.9)
39
(1.5)
54
(2.1)
200 (7.9)
182 (7.2)
A
200 (7.9)
A B
D ECF
209 (8.2)
173 (6.8)
2 - M8 depth 16 (one on each side) (Used to fix the pipe-mounting hardware)
36 (1.4)
242 (9.5)
Grounding terminal (M5)
56
56
(2.2)
(2.2)
Cable gland Connection
A • Power Supply
B, C • Contact Output
D • Detector
E, F
• Pressure Transmitter
• Analog Output • Contact Input
Six wiring holes, G3/4 threaded (GD402R) 1/2 NPT threaded (GD402V, T)
F0103.ai
IM 11T3E1-01E
<1. Specications>

1.3.4 GD40 Detector

• Model GD40- /EJAJ1, /EJAF2, /EJAF3, /EJAF4
Weight: approx. 15 kg (27.8 pounds)
1-13
Unit: mm (in.)
GD40 Detector wiring port See Table
Cable gland is included only in EJAJ1
Four panel-mounting holes, 12 mm in diameter
Gas out See Table
abcdf m
abc cd nmv sc
abc vdfbjkosblm
xs xdcvbg vf
sbdgjklodenjhl;pk
ad xs cdnmn
nbdvf kggswba
z x d cdf vfg m
nscbxfk
150 (5.9)100 (3.9)
93 (3.7) 320 (12.6)
EJA wiring port See Table
4-ø12 (0.47)
300 (11.8)
5
350 (13.8)
Gas in See Table
120 (4.7)
Cable gland is included only in /EJAJ1
40
(1.6)
196 (7.7)
Option code GD40 wiring EJA wiring Gas out/in /EJAJ1 G1/2 G1/2 Rc1/4 /EJAF2 1/2 NPT 1/2 NPT 1/4 NPT /EJAF3 1/2 NPT 1/2 NPT 1/4 NPT /EJAF4 1/2 NPT 1/2 NPT 1/4 NPT
F0104.ai
IM 11T3E1-01E
<1. Specications>

1.3.5 Detector Unit (Intrinsically Safe, Explosion-proof)

• Hardware for Pipe Mounting : GD40
1-14
Unit: mm (in.)
Pipe-mounting hardware
Mounting pipe (JIS 50 A (60.5 mm in outer dia.) nominal size)
Approx. 191 (7.5)
Approx. 92 (3.6)
Sample gas outlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT
Grounding terminal (3 mm (0.11) screw)
Wiring hole, GD40R: G1/2 GD40V: 1/2 NPT GD40T: 1/2 NPT GD40G: 1/2 NPT
Approx. 264 (10.4)
Approx. 193 (7.6)
Approx. 117 (4.6)
Sample gas inlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT
Note: Cable gland is included only in GD40R.
F0105.ai

1.3.6 Surge protector: K9215LP

M4 SCREW TERMINAL
Ø4.2
(0.17)
1128.5
(0.43)
5.5
(0.22)
(1.12)
21
41
(1.61)
20028
(7.86)
(1.1)
Unit: mm (in.)
4.5
F0106.ai
(0.18)
IM 11T3E1-01E
<2. Installation, Wiring and Piping>
2-1

2. INSTALLATION, WIRING AND PIPING

The GD402 Gas Density Meter is thoroughly inspected at the factory and carefully packed to ensure the equipment does not suffer any damage during transportation. The package should also be handled with care when unpacking to prevent the equipment from undergoing severe mechanical shock. After unpacking, visually check the equipment to ensure that it is free from any damage.
Although the detector has no controls on it, it may need to be accessed for inspection or for other reasons. Install the detector in a location as close as possible to where the gas is sampled and where maintenance can be easily carried out. The converter has a display with controls on it; thus, you should install it so that the keys are positioned directly in front of you when working.
Note that there must be a clearance of at least 400 mm at the back of the converter GD402G, GD402T, GD402V and GD402R since cables are wired to the back by removing the screwed rear cover.

2.1 Installing the Detector

The GD40 detector is an explosion-proof with intrinsically safe sensor instrument. Install the detector in a location where the following conditions are satised.

2.1.1 Selecting the Location

l Explosion-proof construction
Before installing the detector in an explosion-hazardous area, ensure that the area conforms to the explosion-proof code noted above.
l No corrosive gases
Corrosive gases are not desirable because they may damage the electrical components in the detector.
l Slight mechanical vibration
Although the detector is vibration-resistant, vibration may loosen the connections of the external wiring.
l No exposure to direct sunlight
Exposure to direct sunlight may raise the temperature in the detector to abnormal levels, and should therefore be avoided. Note that such abnormal temperature levels can also result from heat irradiated from high-temperature equipment around the detector.
l Humidity maintained between 5% and 95 %RH
Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a prolonged period. It is recommended that the detector be used at a humidity between 25% and 85 % RH.
l No exposure to rain water
Even though the detector is rainproof, whenever possible install it where it is protected from water splashes. The reasoning for this is that the detector cover may need to be removed for mainte­nance or for other reasons.
l Altitude of installation site is lower than 2,000 m.
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.1.2 GD40T (FM Explosion-proof and Intrinsically Safe Approval)

2-2
NON-HAZARDOUS LOCATIONS
Sealing Fitting
Figure 2.1
HAZARDOUS LOCATIONS
475 mm (18") Max.
Conduit
GD40T
F0201.ai
1. GD40T Gas density detector for use in hazardous (classied) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D; Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
• Enclosure Rating: NEMA 4X
• Temperature Code: T5
• Ambient Temperature: -10 to 60°C
2. Wiring
• All wiring shall comply with National Electric Code ANSI/NAPA 70 and Local Electrical
Codes.
WARNING
Seal all conduits within 475 mm (18 inches). Refer to Figure 2.1.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the approval of Factory Mutual Research Corporation.
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.1.3 GD40V (CSA Explosion-proof and Intrinsically Safe Approval)

2-3
NON-HAZARDOUS LOCATIONS
Sealing Fitting
Figure 2.2
HAZARDOUS LOCATIONS
500 mm (20") Max.
Conduit
GD40V
F0203.ai
1. GD40V Gas density detector for use in hazardous (classied) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D; Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
• Enclosure Rating: Type 4X
• Temperature Code: T5
• Ambient Temperature: -10 to 60°C
2. Wiring
• All wiring shall comply with Canadian Electrical and Local Electrical codes.
• Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure. Refer to Figure 2.2.
3. Operation
• Note a warning label worded as follows.
WARNING
• Open circuit before removing cover.
• Substitution of components may impair instrinsic safety.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certication of CSA International.
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.1.4 Mounting the Detector

The detector is designed for pipe mounting.
Pipe-mounting hardware
Approx. 191 (7.5)
Approx. 92 (3.6)
Grounding terminal (3 mm (0.11) screw)
Wiring hole, GD40R: G1/2 GD40V: 1/2 NPT GD40T: 1/2 NPT GD40G: 1/2 NPT
Approx. 117 (4.6)
Sample gas inlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT
Approx. 264 (10.4)
Approx. 193 (7.6)
2-4
Unit: mm (in.)
Mounting pipe (JIS 50 A (60.5 mm in outer dia.) nominal size)
Figure 2.3 Vertical Mounting
Sample gas outlet GD40R: Rc1/4 GD40V: 1/4 NPT GD40T: 1/4 NPT GD40G: 1/4 NPT
Note: Cable gland is included only in GD40R.
F0205.ai
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.2 Installing the Converter

The GD402 converter comes in either the non-explosion-protected Model GD402G or explosion­protected Model GD402T, V, R (explosion-protection code: Refer to 1.1.1). Choose the Model GD402T or V or R if the converter is to be used in a hazardous area.

2.2.1 Selecting the Location

l Ease of operation
Select a location where you can easily view the readings on the display and work with the keys. Installing the converter closer to the detector will ease your maintenance work, including calibra­tion.
l Explosion-proof construction
When installing the converter in an explosion-hazardous area, ensure that the area conforms to the explosion-proof code noted above.
l No corrosive gases
Corrosive gases are not desirable because they may damage the electrical components in the converter.
l Slight mechanical vibration
2-5
Although the converter is vibration-resistant, vibration may loosen the connections of the external wiring.
l No exposure to direct sunlight
Exposure to direct sunlight may raise the temperature in the converter to abnormal levels, and should therefore be avoided. Note that such abnormal temperature levels can also result from heat irradiated from high-temperature equipment around the converter. If the converter is likely to receive direct sunlight and the temperature in the converter will likely exceed the operating limits, attach a hood (optional).
l Humidity maintained between 5% and 95% RH.
Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a prolonged period. It is recommended that the converter be used at a humidity between 25% and 85 % RH.
l No exposure to rainwater
Even though the converter is rainproof, whenever possible install it where it is protected from water splashes. The reasoning for this is that the converter cover may need to be removed for maintenance or for other reasons
l Altitude of installation site is lower than 2,000 m.
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.2.2 GD402T (FM Explosion-proof Approval)

GD402T
2-6
HAZARDOUS LOCATIONS
NON-HAZARDOUS LOCATIONS
Figure 2.4
Conduit
475 mm (18") Max.
Sealing Fitting
F0202.ai
1. GD402T Gas density converter for use in hazardous (classied) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D; Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G
• Enclosure Rating: NEMA 4X
• Temperature Code: T6
• Ambient Temperature: -10 to 55°C
2. Wiring
• All wiring shall comply with National Electrical Code ANSI/NFPA 70 and Local Electrical
Codes.
WARNING
Seal all conduits within 475 mm (18 inches). Refer to Figure 2.4.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the approval of Factory Mutual Research Corporation.
IM 11T3E1-01E
<2. Installation, Wiring and Piping>

2.2.3 GD402V (CSA Explosion-proof Approval)

GD402V
HAZARDOUS LOCATIONS
Conduit
500 mm (20") Max.
Sealing Fitting
2-7
NON-HAZARDOUS LOCATIONS
Figure 2.5
F0204.ai
1. GD402V Gas density converter for use in hazardous (classied) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D; Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G
• Enclosure Rating: Type 4X
• Temperature Code: T6
• Ambient Temperature: -10 to 55°C
2. Wiring
• All wiring shall comply with Canadian Electrical Code and Local Electrical Codes.
• Seal all conduits within 500 mm of the enclosure. Refer to Figure 2.5.
• Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure.
3. Operation
• Note a warning label worded as follows.
WARNING
Open circuit before removing cover.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
• The instrument modication or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certication of CSA Inter­national.
IM 11T3E1-01E
Loading...
+ 148 hidden pages