Yokogawa GC1000 User Manual

User's Manual
GC1000 Mark II Process Gas Chromatograph Overview
IM 11B03A03-01E
IM 11B03A03-01E
3rd Edition
Notice
Regarding This Manual
1. This Manual should be passed on to the end user.
2. Read this manual carefully and fully understand how to operate this product before you start operation.
3. Y okogawa makes no warranty of any kind with regard to this material, but not limited to, implied warranties of merchantability for particular purpose.
4. All rights reserved. No part of this manual may be reproduced in any form without Y okogawa’ s written permission.
5. Great effort has been expended to ensure that the descriptions in this manual are correct. Should you, however, come across a questionable area or note an inconsis­tency , a telephone call or letter to Yokogawa.co.,ltd. noting the questionable area would be highly appreciated.
6. The contents of this manual are subject to change without prior notice.
i
Regarding Protection, Safety , and Prohibition Against Unauthorized Modifica-
tion.
1. For the protection and safe use of the product and the system controlled by it, be sure to follow the instructions on safety described in this manual when handling the prod­uct. In addition, if you handle the product in contradiction to these instructions, our company does not guarantee safety .
2. The following safety symbol marks are used on the product concerned or in this Manual :
WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls attention to a procedure, practice, condi­tion or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product.
IMPORT ANT :
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
NOTE:
Draws attention to information essential for understanding the operation and features.
TIP:
Gives information that complements the present topic.
All Rights Reserved Copyright © 2001, Y okogawa Electric Corporation
IM 11B03A03-01EMedia No. IM 11B03A03-01E 3rd Edition : Aug. 2006 (YK)
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
See Also:
Gives the reference locations for further information on the topic.
Protective ground terminal:
In order to provide protection against electrical shock in case of a fault. This symbol indicates that the terminal must be connected to ground prior to operation of equip­ment.
Function ground terminal:
In order to provide protection against noise. This symbol indicates that the terminal must be connected to ground prior to operation of equipment.
Alternating current
ii
3. If protection / safety circuits are to be used for the product or the system controlled by
4. When you replace parts or consumables of the product, use those specified by our
5. Do not modify the product.
Indicates the power switch state “ON”.
Indicates the power switch state “Stand - by”.
Indicate the power switch state “OFF”.
it, they should be installed outside of the product.
company .
Exemption from Responsibility
1. Y okogawa Electric Corporation does not make any warranties regarding the product except those mentioned in the WARRANTY that is provided separately.
2. Y okogawa Electric Corporation assumes no liability to any party for any loss or dam­age, direct or indirect, caused by the use or any unpredictable defect of the product.
Regarding Software and set including Software Supplied by Y OKOGA WA
1. Y okogawa makes no other warranties expressed or implied except as provided in its warranty clause for software supplied Y okogawa.
2. Use this software with one specified computer only. Y ou must purchase another copy of the software for use with each additional com-
puter.
3. Copying this software for purposes other than backup is strictly prohibited.
4. Store the streamer or floppy disk (original medium) in a secure place.
5. Reverse engineering such as the disassembly of software is strictly prohibited.
6. No portion of the software supplied by Y okogawa may be transferred, exchanged, or sublet or leased for use by any third party without prior permission by Y okogawa.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
Warning/Caution Labels
T o ensure safety operation of this equipment, warning/caution labels are at-
tached on the equipment as follows. Check these labels for your safety opera­tion.
D
A
C
B
E
iii
[Left side] [Front] [Right side] [Back]
F0001.EPS
AB
C
The inside of the enclosure is high temperature after turning off the power. Don’t touch the inside of the enclosure or the compponents, the protective gas should be suoolied for more than one hour after turning off the power.
D
E
F0002.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
Introduction
Thank you for purchasing the GC1000 Mark II process gas chromatograph. This manual describes the technical information on overview of Model GC1000D /
GC1000S / GC1000T / GC1000E / GC1000W / GC1000C (Hereafter , it is abbreviated as GC1000 Mark II) Process Gas Chromatograph.
Please lead the following respective documents before installing and using the GC1000 Mark II system.
Documents Related to the GC1000 Mark II Process Gas Chromatograph
1. Instruction manuals
The product comes with the following instruction manuals.
Instruction manuals that do not depend upon the specifications of the product
(1) Overview (IM 1 1B03A03-01E) (2) Basic Operation and Startup (IM 1 1B03A03-02E) (3) Maintenance and Inspection Manual (IM 1 1B03A03-04E)
iv
(4) LCD Panel Operation Manual (IM 1 1B03A03-05E) (5) Alarm Message Manual (IM 1 1B03A03-06E) (6) Password Manual (IM 1 1B03A03-07E)
and Installation Manual (TI 1 1B03A03-13E)
Instruction manuals that depend upon the specifications of the product
(1) GCMT Gas Chromatograph Maintenance Terminal Software Package Operation
Guide (IM 1 1B03G03-03E)
(2) Capture It! Manual (IM 1 1B3G1-02E)
Instruction manuals for related products
(1) PCAS PC Analyzer Server Software (IM 11B06B01-01E) (2) ASET Analyzer Server Engineering Terminal Software (IM 1 1B06C01-01E) (3) GCET GC Engineering Terminal Software (IM 1 1B06D01-01E) (4) ASGW Analyzer Server Gateway Software (IM 11B06E01-01E) (5) ASIU Analyzer Server Interface Unit Software (IM 11B06F01-01E)
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
2. Operation Data
Operation data is supplied with the operation manuals in the delivered package and con­tains the following required to use the gas chromatographs.
Process conditions and measurement range
Instrument specifications and operating conditions
Standard sample for calibration
Column system and column
Miscellaneous data
Chromatogram, base line, repeatability , power supply voltage variation, etc.
Analyzer flow diagram and installation
Parts composition table
General connection diagram
Sampling system diagram (only if supplied by Yokogawa)
Is the System Ready?
Before reading this manual, the following preparations must have been completed.
The system must be unpacked and installed in the correct place.
v
The piping for the utility gases such as carrier and calibration gases must be com­pleted, followed by leak checking.
The wiring for the power supply and others must be completed.
If these have not been completed yet, see the Installation Manual (TI 1 1B03A03-13E). After completion, return to this manual and do the following:
If the system power is on, turn off the power .
Shut off all the gases at the flow control units.
Please read the following respective cautions (General Precautions, Caution of using Explosion-Protection Instruments, on Piping Construction, and on Piping Work) before installing and using the GC1000 system.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
General Precautions
WARNING
(1) In order to analyze gases, process gas chromatographs use a sample of the process
gas and utility gases.
Since these gases are typically combustible, combustion-sustaining, toxic, odorous, resolvable, polymerizing, or corrosive, refer to the “Safety Informa­tion” in our approval drawings and others to ensure safety thoroughly before using these analyzers.
(2) Up to two protection systems, each of which weighs approximately 10 kg, are installed
on top of the GC1000. Therefore, the center of gravity is higher than the center of the analyzer body .
Take great care when carrying and installing (piping- wiring) the GC1000. The GC1000 must be carried and installed very carefully (including piping and wiring) by more than one person (at least four people are recommended).
(3) Since the GC1000 are precision instruments, take care when handling not to jolt
of knock them.
(4) Use the GC1000 within the range of your purchase specifications.
vi
Y okogawa assumes no responsibility for problems resulting from use by the customer outside the purchase specifications.
If the GC1000 need to be modified or repaired, please contact your nearest Y okogawa representative. Yokogawa assumes no responsibility for results where the customer or any third party has attempted to modify or repair these products.
(5) When touching LCD Panel switches
When touching LCD Panel switches, first, discharge Electro Static Charge of the body . Then, touch the LCD Panel switches. If not, LCD display may be changed by Electro
Static Discharge.
IMPORT ANT :
(1) Read the attached instruction manual before operating the GC1000 (2) The instruments must be installed and operated according to the installation manual,
instruction manual, approval drawings, and operation data.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
CA UTIONS OF USING EXPLOSION-PRO TECTION INSTR UMENTS
The GC1000 Process Gas Chromatographs are designed to protect against explosion. When these analyzers are used in a hazardous area, observe the following precautions.
Since the applicable standard differs depending on the specifications of the analyzer to be used, check the specifications of your analyzer.
(1) Kinds of explosion protection
To assure explosion protection, the GC1000 Process Gas Chromatographs have a pres­surized and flameproof construction, or type X purging and explosionproof construction meeting the following standards :
<GC1000D / GC1000S>
JIS Expd IIB + H2 T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid-sample valve 250˚C max.)
JIS Expd IIB + H2 T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid-sample valve 225˚C max.)
JIS Expd IIB + H2 T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid-sample valve 145˚C max.)
vii
JIS Expd IIB + H2 T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid-sample valve 95˚C max.)
<GC1000W / GC1000C>
CENELEC (A TEX directive) certified : Group II Category 2G
EEx pd II B +H2 T1 (programmed-temperature oven 320°C max, isothermal oven 225°C max., liquid-sample valve 250°C max.)
EEx pd II B +H2 T2 (programmed-temperature oven 225°C max, isothermal oven 225°C max., liquid-sample valve 225°C max.)
EEx pd II B +H2 T3 (programmed-temperature oven 145°C max, isothermal oven 145°C max., liquid-sample valve 145°C max.)
EEx pd II B +H2 T4 (programmed-temperature oven 95°C max, isothermal oven 95°C max., liquid-sample valve 95°C max.)
<GC1000T / GC1000E>
FM Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, NEMA3R.
Type Y purging and Type X purging for CLI, DIV1, GPS.B, C & D, NEMA3R.
T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid­sample valve 250˚C max.)
T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid­sample valve 225˚C max.)
T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid­sample valve 145˚C max.)
T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid­sample valve 95˚C max.)
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
CSA Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, NEMA3R.
Type Y purging and T ype X purging for CLI, DIV1, GPS.B, C & D, NEMA3R.
T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid­sample valve 250˚C max.)
T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid­sample valve 225˚C max.)
T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid­sample valve 145˚C max.)
T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid­sample valve 95˚C max.)
(2) Precautions for Explosionpr oof section (The analyzer with optional code "FM/CSA T ype Y purging" does not have the e xplosionproof section.)
When handling the screws on the cover of the Protection system, note the following to avoid damaging the screws since they cannot be repaired.
(1) The enclosure is pressurized. Before removing the cover, reduce the internal pres-
sure by loosening the sealing plug for wiring on the enclosure or relevant means.
viii
(2) When removing the cover, prevent any dirt or foreign matter from contaminating the
screw part.
(3) When installing the cover, tighten the screws by hand ; never use tools. (4) Since the screws are coated with MOL YKOTE, do not lubricate them.
(3) Precautions when using hydrogen gas
When using hydrogen gas as the carrier gas, the FID or FPD combustion gas, to ensure safety , install the analyzer in a location equipped with a ventilator or where there is suffi­cient ventilation. Make sure there are no gas leaks from the pipe joints and inspect for leaks.
(4) Installation site and environment
The analyzer specifications allow it to be used in hazardous areas as defined by Zone1 IIB + H
, T2, T3, T4 (JIS / CENELEC) or DIV1, GPS B, C & D, T1, T2, T3, T4 (FM / CSA).
2T1
However, never install the analyzer in an area where the density of explosive gas persists for a long time.
(5) Wiring works
Model GC1000D / GC1000S, analyzer obtains explosion proof authorization by the com­plete set including metal fittings of the attachment.
When performing wiring, always use the attached sealing fittings and flameproof packing adapter.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
(6) Maintenance and inspection
During usual maintenance and inspection, it is not necessary to check the explosion­protected section.
Before opening the door of the explosion-protected section for maintenance and inspec­tion, be sure to turn off the power . After completing maintenance and checks, close the door completely then turn on the power after checking that the specified explosion protec­tion performance is guaranteed. The parts to be checked are described in the Maintenance and Inspection Manual (IM 1 1B03A03 - 04E).
If any of the following damage occurs, contact a Yokogawa sales representative or the Y okogawa sales division
(1) If the screws securing the Protection System (explosionproof construction) are dam-
aged
(2) If the exterior or light transmission section of the enclosures is damaged (3) If packings are cracked or conspicuously deformed
(7) Override function (The analyzer with optional code "FM/CSA T ype Y purg ing" does not have this function.)
ix
WARNING
When the override function is used, Analyzer becomes an ignition source and the high temperature and the high voltage part will be exposed.
Please confirm that in the ambient atmosphere, the concentration of explosive gases is less than the allowable limit, by using a gas detector.
To return to the normal operation, turn off “the override switch” and then close the door as it was before turning on power.
In this analyzer, if the pressure of the pressurized / type X purged enclosure system (oven, electronic section) drops while the power is on, the pressurized explosion protection sec­tion is activated to stop power supply . Therefore, in case of opening the door of the oven or of the electronic section inadvertently , for maintenance, while the power is on, the protec­tion system is activated to cut off the power .
The “override function” intensively releases this function of protection system.
The override switch is installed in section.
(8) Replacing parts
Always use parts specified by Yokogawa when replacing parts, for replacement, refer to the Maintenance and Inspection Manual (IM 1 1B03A03 - 04E).
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
(9) Operation
WARNING
<CENELEC>
* Only trained persons may use this instrument in a hazardousl location. * Do not open when energized.
<FM>
For type X purging:
* This equipment contains components that operate at high temperature. The equip-
ment shall be deenergized for 60 minutes to permit those components to cool before the enclosure is opened unless the area is demonstrated to be nonhazardous at the time.
* Enclosure shall not be opened unless the area is known to be nonhazardous, or
unless all devices within have been de-energized.
x
* Power shall not be restored after enclosure has been opened until enclosure has
been purged for 12 minutes. (When the internal pressure is restored, the system automatically purges over the 12 minutes, then turns on the power again.)
For explosionproof enclosure:
* Seal all conduits within 18 inches * Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within have been de-energized. Power shall not be restored after enclosure has been opened until enclosure has been purged for 12 minutes at speci­fied pressure indicated by the pressure gauge labeled “EL.BOX” in the pressure and flow control section.
* Alarm shall be provided and connected to alarm contact output.
(a) The alarm shall generate a visual or audible signal that attracts attention (b) The alarm shall be located at constantly attended location. (c) Electrical alarms shall be approved for the location in which they are installed.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<CSA>
For type X purging:
* This equipment contains components that operate at high temperature. The equip-
ment shall be deenergized for 60 minutes to permit those components to cool before the enclosure is opened unless the area is demonstrated to be nonhazardous at the time.
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within the enclosure have been de-energized. Power must not be restored after enclosure has been opened until enclosure has been purged for 12 minutes at a flow rate of 0.05m
* Power will automatically be removed when purge pressure falls below 40 mm (1.6 in)
of water column.
For explosionproof enclosure:
* A seal shall be installed within 50 cm of the enclosure. * Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within the enclosure have been de-energized. Power must not be restored after enclosure has been opened until enclosure has been purged for 12 minutes at a flow rate of 0.05m
3
/min.
3
/minute min.
xi
* Remove power below 40mm (1.6in) of water column.
Take care not to generate mechanical spark when accessing to the instrument and periph­eral devices in hazardous locations.
Do not press prick the keyboard of LCD panel (operation and display section) using such as knives and sticks.
(10) Maintenance and Repair
The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation is prohibited and will void the approval of Factory Mutual Research Corporation and CSA certification and CENELEC certification .
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <Rev> <Introduction>
Precautions Against Electrostatic Pr oblems
The GC1000 system uses numerous IC components. When handling cards with IC components mounted on them for maintenance or setting changes, take full precau­tions against electrostatic problems.
These precautions are summarized below. (a) When storing or carrying cards, enclose them in a conductive bag or antistatic
bag. (Cards as shipped by Yokogawa are enclosed in a conductive bag or antistatic bag labeled with cautions against electrostatic problems.)
(b) Whenever mounting or demounting cards into or from a product, wear a wrist
strap grounded via a 1 M resistance. Connect the wrist strap to any ground terminal near the ground wire or to any unpainted part of the grounded frame.
xii
List strap
Sheet
1M‰
Using a wrist strap and conductive sheet Using a Conductive Sheet
Gard
List strap
F0003.EPS
(c) When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 M resistance, wearing a wrist strap as in (2) above. Keep easily-chargeable plastic materials away from the bench.
(d) Never touch components mounted on the cards, the pattern side, connectors,
pin components, etc. with bare hands, unless using a wrist strap and a conduc­tive sheet.
(e) Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Int> <Ind> <Rev>
GC1000 Mark II Process Gas Chromatograph Overview
CONTENTS
Notice .......................................................................................................... i
W arning/Caution Labels............................................................................iii
Introduction ...............................................................................................iv
1. Principle of Gas Chr omatograph ........................................................... 1-1
1.1 Sampling Mechanism ..................................................................................... 1-1
1.2 Component Separation Using Column.......................................................... 1-2
1.3 Detector ........................................................................................................... 1-3
2. Terminology ............................................................................................ 2-1
Toc-1
IM 11B03A03-01E 3rd Edition
2.1 Operation Terminology ................................................................................... 2-1
2.2 Instrument Terminology.................................................................................. 2-2
3. System Configuration ............................................................................ 3-1
3.1 T ype and Appearance..................................................................................... 3-1
3.2 Components and Their Functions.................................................................. 3-2
3.2.1. Protection System ............................................................................ 3-2
3.2.2 Electronic Section............................................................................. 3-2
3.2.3 Pressure and Flow Control Section................................................... 3-2
3.2.4 Isothermal Oven ............................................................................... 3-2
3.2.5 Programmed-temperature Oven ....................................................... 3-3
3.2.6 Analyzer Base Sampling Section ...................................................... 3-3
3.3 Block Diagram................................................................................................. 3-4
3.4 Internal Piping System Diagram .................................................................... 3-7
3.5 External Input and Output Signals............................................................... 3-12
4. Outline of Software................................................................................. 4-1
4.1 Status and Operation Mode............................................................................ 4-1
4.1.1 Process ............................................................................................ 4-2
4.1.2 Manual ............................................................................................. 4-5
4.1.3 Lab ................................................................................................... 4-5
4.2 Stream ............................................................................................................. 4-6
4.3 Method............................................................................................................. 4-7
4.4 Description of Actions.................................................................................... 4-9
4.4.1 Actions of Stream Sequence........................................................... 4-10
4.4.2 Actions of Stream (Continuous) .......................................................4-11
4.4.3 Actions of Stream (1 cycle) ............................................................. 4-13
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Int> <Ind> <Rev>
5. Actions of External Input and Output Signals ....................................... 5-1
Toc-2
4.4.4 Actions of Calibration...................................................................... 4-14
4.4.5 Actions of V alidation........................................................................ 4-17
4.5 Computation and Processing ...................................................................... 4-20
4.5.1 Peak Processing............................................................................. 4-20
4.5.2 Deviation processing ...................................................................... 4-20
4.5.3 Additional processing...................................................................... 4-20
4.5.4 Signal Processing........................................................................... 4-21
4.6 Alarm Processing ......................................................................................... 4-22
5.1 Analog Hold Output ........................................................................................ 5-1
5.1.1 When Actual Stream is Specified for Stream Number (without Stream
Identification Signal) ......................................................................... 5-2
5.1.2 When Actual Stream is Specified for Stream Number (with Stream
Identification Signal) ......................................................................... 5-3
5.1.3 When "99" is Specified for Stream Number (with Stream Identification
Signal) .............................................................................................. 5-4
5.2 Contact Output................................................................................................ 5-5
5.2.1 Stream Sequence............................................................................. 5-5
5.2.2 Stream.............................................................................................. 5-6
5.2.3 Operation Mode................................................................................ 5-6
5.2.4 Alarm................................................................................................ 5-6
5.2.5 Timing............................................................................................... 5-6
5.3 Contact Input................................................................................................... 5-7
5.4 Communication Input and Output ................................................................. 5-8
5.4.1 GC6 T ype Output Data Format (Fixed to 45 Characters)................... 5-8
5.4.2 GC6 T ype Input Data Format ............................................................ 5-9
5.4.3 GC8/GC1000 Type Output Data Format (Fixed to 45 Characters)....5-1 1
5.4.4 GC8/GC1000 T ype Input Data Format............................................ 5-12
5.4.5 MODBUS Communication Data Specification................................. 5-14
Revision Record ......................................................................................... i
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <1. Principle of Gas Chromatograph >

1. Principle of Gas Chromatograph

A gas chromatograph is an analyzer which first sends a fixed volume of the sampled multi­component gas mixture to a column, separates it in the column, then measures the con­centrations of the components with a detector. The process gas chromatograph analyzes intermittently , allowing periodic analysis in a specified sequence, thus automatic sampling is possible.
This chapter explains the measurement principle of the GC1000 Process Gas Chromato­graphs.

1.1 Sampling Mechanism

The process gas chromatograph consists of a sampling mechanism, a column and a detector.
Sampling is carried out by switching a sampling valve. When separating components or detecting concentrations, the sampling valve is set to allow the gas (liquid) to be measured to flow through the sample column. When sampling, the sampling valve leads the gas (liquid) to be measured to a column on a carrier gas. (See Figure1.1)
There are two important points regarding sampling: a regular volume is sampled since repeated sampling is required; and samples are taken quickly and securely . The volume is fixed by measuring a specific gas (liquid) of controlled temperature and pressure using a sample measurement tube. Samples are taken quickly and securely by ensuring that the gas to be measured always flows without interrupt.
1-1
Sample
Carrier gas
Sample
Carrier gas
Component separation and
concentration detection
Measuring tube
Sample valve
Vent
Column Detector
Electric signal
Status of sampling
Measuring tube
Sample valve
Vent
Column Detector
Electric signal
Figure 1.1 Basic Configuration of Gas Chromatograph
F0101.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <1. Principle of Gas Chromatograph >

1.2 Component Separation Using Column

Three types of columns are available for the GC1000 Process Gas Chromatographs: the packed column, the mega-bore column and the capillary column.
The packed column consists of a stainless pipe, 2 mm in diameter and 0.2 to 2.0 m in length and filled with a bulking agent called a stationary phase.
The mega-bore and capillary columns, of approximate diameter 0.5 mm and 0.3 mm respectively , are coated inside a certain phase called a stationary phase.
The components in the multi-component gas mixture sample with carrier gas, which called mobil phase move through the column, repeatedly absorbing the stationary phase and dissolving from the stationary phase at a certain cyclic rate conforming to a fixed partition coefficient * that is unique to each component.
Since the transfer rates differ depending on the partition coefficient, a multi-component gas mixture gradually separates into discrete components and is separated in the order of the transfer rates.
* Partition coefficient : The concentration ratio of the components, calculated by dividing the component concentration which
is in equilibrium in the stationary phase by the concentration which is in equilibrium in the mobile phase.
Figure 1.2 shows a diagram of how the multi-component gas mixture is led to a column and separated into its discrete components over time.
1-2
Multi-component gas mixture
Sampling
(1st round)
Time
(2nd round)
Figure 1.2 Separating Components Using a Column
(Intake) (Outlet)Column
Injection
:Component A, :Component B, :Component C
Detector Concentration signal
Carrier gas
Carrier gas
A
B
C
F0102.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <1. Principle of Gas Chromatograph >

1.3 Detector

The components separated in the column are led to the detector where the concentration of each component is measured.
The GC1000 Process Gas Chromatographs can be fitted with thermal conductivity detec­tors (TCD), flame ionization detectors (FID) or flame photometric detectors (FPD). The thermal conductivity detector can measure almost all non-corrosive components but sensitivity is relatively low. On the other hand, the hydrogen flame ionization detector can measure hydrocarbon and the flame photometric detector can measure sulfur compounds, respectively with high sensitivity .
(1)Thermal Conductivity Detector (TCD)
The TCD utilizes the difference in the thermal conductivity between the measured gas and the carrier gas and detects the unbalanced voltage produced in a bridge circuit as a mea­sure of concentration.
Figure 1.3 shows the fundamental principle of the TCD. As shown, there are two streams, each having two filaments. One stream passes the carrier gas only and the other, con­nected to the column outlet, allows the measured gas to pass during analysis. The fila­ments in the two streams form a bridge circuit such that the filament in one stream is adjacent to the filament in the other stream. The unbalanced voltage in the bridge is propor­tional to the concentration of the measured gas (liquid) component.
1-3
The TCD is frequently used to measure the component concentration of the measured gas (liquid).
Carrier gas
Figure 1.3 Fundamental Principle of Thermal Conductivity Detector
Comparison filament
Measurment filament
Carrier gas
+
Sampling gas
Z1 Z2
Z4 Z3
Constant voltage
Measurment filament
Comparison filament
Output
F0103.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <1. Principle of Gas Chromatograph >
(2)Flame Ionization Detector (FID)
The FID utilizes the phenomenon that carbon molecules in the measured component (hydrocarbon) are ionized in a hot hydrogen flame. That is, it detects the ionization current which flows between electrodes to which a high voltage is applied. The ionization current is proportional to the concentration of the measured component.
The FID is used to measure the component concentration of gases containing low concen­trations of hydrocarbons.
Ion collector coil
+
-
-
+
+
Hydrogen flame
Jet pipe (nozzle)
+
-
+
-
Output
1-4
Carrier gas
+
Sampling gas
Figure 1.4 Fundamental Principle of Flame Ionization Detector
(3)Flame Photometric Detector (FPD)
Figure 1.5 shows the structure of the FPD. As the sample gas containing a sulfur compo­nent is led into the excess hydrogen flame, the component containing the sulfur atoms is excited. The FPD detects the luminous intensity of the light emitted when this excited component return to its base state using multiplier phototube and converts it to a voltage. This voltage represents the concentration of the sulfur component in the measured gas.
The FPD can measure on the sulfur component with a high sensitivity of 1 ppm.
Hydrogen flame
Jet pipe (nozzle)
Hydrogen gas for combustion
Carrier gas
+
Sampling gas
Figure 1.5 Basic Configuration of Flame Photometric Detector
Photomultiplier tube
Hydrogen gas for combustion
F0104.EPS
Hydrogen gas for combustion
Air fo combustion
F0105.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <2. T erminology >

2. Terminology

2.1 Operation Terminology

Term Description Notes
Remote Mode
Local Mode
Status
Process Manual Lab
Operation Mode
Run Pause Stop
Measurement Status
Stream Sequence Stream (continuous) Stream (1 cycle) Calibration
Validation Method Analysis Cycle Warming Up Time Peak Detection Stop Time Pause Time Purging
Status accessible without password, LCD/key: used only for reference GCMT: connectable Status accessible by entering password LCD/key: used for changing settings
Normal measurement and calibration Manual operation Measurement with automatic peak detection, like a lab GC
Mode in which measurement is running Mode in which measurement pauses Mode in which measurement stops
Continuously measures streams in order specified in Stream Sequence Continuously measures the specified stream Measures the specified stream once Performs calibration of the specified number Performs validation of the specified number Configures action timing of various valves and other parameters Time from start (0 second) to stop of analysis Time for displacement in sample streams Time to stop peak detection of chromatogram compulsorily Time for measurement pause Displacement of the gas in the pressurized/Type X purged enclosures with a protection gas
2-1
T0201.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <2. T erminology >

2.2 Instrument Terminology

Term Explanation Notes
TCD FID FPD LSV Restrictor Methane converter Sampling valve Back-flush valve Column switching valve Atmospheric-pressure balancing
valve Protection gas
Flame arrester Splitter Pressurized enclosure Temperature protection circuit Protection device
EPC
A thermal conductivity detector A flame ionization detector A flame photometric detector A liquid sampling valve A variable resistor A methane reaction system : Methanizer A valve for inputting samples A switching valve for back flush A valve for switching between columns A balancing valve for sampling gases
Air A device for protecting against "flame runaway" A flow splitter An enclosure whose internal pressure is kept high with protection gas A circuit for turning off the heater to prevent overheating A device for detecting a pressure drop in the analyzer to shut down the power supply
Electric pressure controller
2-2
T0202.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >

3. System Configuration

3.1 T ype and Appearance

The GC1000 Process Gas Chromatographs consist of (A) a protection system, (B) an electronic section, (C) a pressure and flow control section, (D) an isothermal oven, (E) a programmed-temperature oven (GC1000D / GC1000T / GC1000W) and (F) an analyzer base sampling section (see Figure 3.1).
Note: There are two types of analyzer base sampling section, either the built-in type within GC1000 (the sample processor is
embedded), or the external type (the sample processor is separate). Select the type best suited to the intended usage.
Display Explosion proof enclosures
(number depends on specifications)
(A) Protection system
(B) Electronic section
Terminal section
3-1
(D) Isothermal oven (E) Programmed-temperature oven
(F) Analyzer base
sampling section
(Note)
Figure 3.1Analyzer Components
(C) Pressure and flow control section
EPC box (in the case that EPC is installed)
F0301.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >

3.2 Components and Their Functions

3.2.1. Protection System

The protection system has a explosion-proof construction and is equipped with a built-in protection circuit. The power relay , pressure switch, timer , relays, override switch, etc. are internal to the instrument. The override function is particularly important for maintenance, since it allows the power to be turned on even if there is an internal pressure loss. The system monitors the internal pressures of the electronic section, isothermal oven and the programmed-temperature oven (GC1000D / GC1000T / GC1000W), and if any of them indicates an internal pressure lower than 392 Pa, it shuts down the power supply to those components. After a power shut-down, when the internal pressure is restored, the system automatically purges over the 12 minutes, then turns on the power again. Analyzers of non­explosion-proof type and with FM/CSA T ype Y purging do not have this protection system.

3.2.2 Electronic Section

The electronic section has a pressurized protection/Type X purged structure and is de­signed for control of a detector, an isothermal oven, various valves and others, for process­ing and computation, and for output of the results. The LCD and operation keys on the front of the electronic section allow manual operation of the GC1000 Mark II.
3-2

3.2.3 Pressure and Flow Control Section

The pressure and flow control sections control and indicate the pressures of sample gases, standard gases, carrier gases, FID or FPD hydrogen (or nitrogen) for combustion, or FID or FPD air for combustion. Regulator values or EPC is installed. It also contains pressure­reducing valves for controlling purge gases, the air for actuating valves or the vortex tube, an air-actuated valve for balancing atmospheric pressure, a vortex tube and a hydrogen restriction system.

3.2.4 Isothermal Oven

The isothermal oven has an pressurized protection / T ype X purged structure. The tem­perature is set at a fixed level from 55 to 225°C (setting by 1°C unit). The isothermal oven contains valves such as the sample valve which is air-activated, the back flush valve, the column for separating a multi-component gas mixture into its individual components and leading the components to the detector in sequence, the detector for detecting the compo­nents, and the restrictor for controlling the gas flow rate, etc.
There are three types of detector, the thermal conductivity detector (TCD), the flame ionization detector (FID) and the flame photometric detector (FPD); either one or two of these detectors can be used simultaneously (however, the FPD can only be installed in the combustion chamber).
The component signals picked up by the detector are led to the electronic section for signal processing.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >

3.2.5 Programmed-temperature Oven

The programmed-temperature oven has an internal pressure protection / T ype X purged structure. It contains a column that separates multi-component mixture samples into individual components and leads them to the detector in sequence. The temperature can be set to a fixed setting or it can be programmed. The allowable temperature range is 60 to 320°C without a cooling system, or 5 to 320°C with a cooling system, and the temperature can be set to rise from between 0.1 to 30°C/min (setting by 0.1°C unit). Up to three tem­perature-rise patterns can be programmed.

3.2.6 Analyzer Base Sampling Section

The analyzer base sampling section is equipped with sample and standard gas streams, and controls the sample pressure and flow rate. It also selectively sends the sample to be analyzed in a stream, and can switch the standard gas to the isothermal or the pro­grammed-temperature oven by a valve depending on the signal sent from the electronic section.
If the analyzer does not contain the analyzer base sampling section, then samples can be processed externally by supplying air to a separate sampling section.
3-3
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >

3.3 Block Diagram

(1) GC1000D/GC1000T/GC1000W T ype X Purging
3-4
SHEETKEY LCD
LCD.KEY
BOARD
AO MOTHER BOARD
AO BOARD
AO BOARD
AO BOARD
BACKBOARD
AI CARD
DET2 CARD
DET1 CARD
P .SW P.SWP .SW
P.M B O ARD
AC/DC 24VAC/DC 24V
AO BOARD
AO BOARD
AO BOARD
AC/DC 15VAC/DC 3.3V
COM BOARD
COM BOARD
COM BOARD
MARSHALING.X BOARD
Terminal
RELAY BOARD
Relay Relay
Terminal
POWER UNIT
SSRSSR
Terminal
Ether (*A) or (*B) is selected.
AO TERMINAL BOARD
HUB TP BOARD
(*A)
OPT CONVERTER
(*B)
TERMINAL
CPU CARD
TEMP CARD
ETHERNET CARD
DET1DET2
I.Oven Heter
C.Oven Heter
FID2 Heter
I.Oven PT100
I.Oven PT100
C.Oven PT100
C.Oven PT100
FID1 Heter
FID2 PT100
FID2 PT100
LSV2 Heter
FID1 PT100
FID1 PT100
LSV2 PT100
EV IF
BOARD
TEMP IF BOARD
LSV1 Heter
LSV1 PT100
LSV2 PT100
LSV1 PT100
EPC BOARD
P .SW
P .SW
F0302.EPS
(2) GC1000T Type Y Purging
AO MOTHER BOARD
AO BOARD
AO BOARD
AO BOARD
AO BOARD
BACKBOARD
AI CARD
DET1 CARD
LCDSHEETKEY
LCD.KEY
BOARD
DET2 CARD
ETHERNET CARD
DET2 DET1
Ether (*A) or (*B) is selected.
P.SW
AC/DC 3.3V AC/DC 15V
I.Oven PT100
C.Oven Heter
C.Oven PT100
P.SW
P.SW
C.Oven PT100
FID2 Heter
FID2 PT100
FID2 PT100
FID1 Heter
AO BOARD
AO BOARD
AC/DC 24V AC/DC 24V
CPU CARD
TEMP CARD
I.Oven Heter
I.Oven PT100
FID1 PT100
FID1 PT100
COM BOARD
COM BOARD
COM BOARD
MARSHALING.Y BOARD
LSV2 Heter
LSV1 Heter
LSV2 PT100
LSV2 PT100
EV IF BOARD
LSV1 PT100
LSV1 PT100
Terminal
POWER UNIT
SSR SSR
SSR SSR
Terminal
TEMP IF
BOARD
AO TERMINAL BOARD
HUB TP BOARD
OPT CONVERTER
TERMINAL
EPC BOARD
(*A)
(*B)
P.SW
P.SW
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >
(3) GC1000S/GC1000E/GC1000C Type X Pur ging, with One Protection
System
3-5
SHEETKEY LCD
LCD.KEY
BOARD
AO MOTHER BOARD
AO BOARD
AO BOARD
AO BOARD
AO BOARD
BACKBOARD
DET2 CARD
DET1 CARD
ETHERNET CARD
FPD AMP BOARD
MC MC
P .SW P.SW
P.M B O ARD
AC/DC 24VAC/DC 24V
AO BOARD
AO BOARD
AC/DC 15VAC/DC 3.3V
COM BOARD
COM BOARD
COM BOARD
MARSHALING.X
BOARD
Terminal
RELAY BOARD
Relay Relay
Terminal
POWER UNIT
SSR
Terminal
Ether (*A) or (*B) is selected.
AO TERMINAL BOARD
HUB TP BOARD
(*A)
OPT CONVERTER
(*B)
TERMINAL
CPU CARD
TEMP CARD
DET1DET2
I.Oven Heter
I.Oven PT100
I.Oven PT100
LSV2 Heter
LSV2 PT100
EV IF
BOARD
TEMP IF BOARD
FPD PT100 FPD PT100
LSV1 Heter
LSV1 PT100
LSV2 PT100
LSV1 PT100
FPD Heter
EPC
BOARD
P .SW
P .SW
(4) GC1000S/GC1000E/GC1000C Type X Pur ging, with T wo Pr otection
Systems
SHEETKEY LCD
LCD.KEY
BOARD
AO MOTHER BOARD
AO BOARD
AO BOARD
AO BOARD
AO BOARD
BACKBOARD
AI CARD
DET2 CARD
DET1 CARD
ETHERNET CARD
FPD AMP BOARD
MC MC
P.SW P.SW
P.M BOARD
AC/DC 24VAC/DC 24V
AO BOARD
AO BOARD
CPU CARD
TEMP CARD
DET1DET2
AC/DC 15VAC/DC 3.3V
I.Oven Heter
I.Oven PT100
I.Oven PT100
COM BOARD
COM BOARD
COM BOARD
MARSHALING.X BOARD
LSV1 Heter
LSV2 Heter
LSV2 PT100
LSV2 PT100
LSV1 PT100
EV IF BOARD
LSV1 PT100
Terminal
RELAY BOARD
Relay Relay
Terminal
POWER UNIT
SSR
Terminal
TEMP IF BOARD
FPD PT100 FPD PT100
FPD Heter
Ether (*A) or (*B) is selected.
AO TERMINAL BOARD
HUB TP BOARD
OPT CONVERTER
TERMINAL
EPC BOARD
(*A)
(*B)
P.SW
P.SW
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >
(5) GC1000E T ype Y Purging
AO MOTHER BOARD
AO BOARD
AO BOARD
AO BOARD
BACKBOARD
P.SW
AO BOARD
AO BOARD
AO BOARD
AC/DC 24V AC/DC 24V
P.SW
AC/DC 3.3V AC/DC 15V
COM BOARD
COM BOARD
COM BOARD
MARSHALING.Y BOARD
Terminal
POWER UNIT
SSR
SSR
Terminal
3-6
Ether (*A) or (*B) is selected.
AO TERMINAL BOARD
HUB TP BOARD
(*A)
OPT CONVERTER
(*B)
TERMINAL
LCDSHEETKEY
LCD.KEY
BOARD
AI CARD
DET1 CARD
DET2 CARD
FPD AMP BOARD
DET2 DET1
CPU CARD
TEMP CARD
ETHERNET CARD
MCMC
I.Oven Heter
I.Oven PT100
Name Description
CPU CARD
Controls the whole GC1000 analyzer system (controls the temperature or the isothermal oven).
TEMP CARD
Controls the temperature of the components other than the isothermal oven.
ETHERNET CARD AI CARD TCD CARD FID CARD FPD CARD BACK BOARD
RELAY BOARD P.M BOARD AI TERMINAL BOARD
Controls the TPC/IP. Controls the analog input. Controls the TCD. Controls the FID. Controls the FPD. Connects various cards.
Controls the relay inside the protection system. Monitors the inner pressure. Equipped with the connection terminal for the analog input.
I.Oven PT100
LSV2 Heter
LSV2 PT100
LSV2 PT100
LSV1 Heter
EV IF BOARD
LSV1 PT100
LSV1 PT100
TEMP IF
BOARD
FPD PT100 FPD PT100
FPD Heter
EPC BOARD
P.SW
P.SW
LCD.KEY BOARD MARSHALING.X BOARD MARSHALING.Y BOARD COM BOARD FPD AMP BOARD EV IF BOARD TEMP IF BOARD
EPC BOARD AO TERMINAL BOARD HUB TP BOARD AO MOTHER BOARD AO BOARD
Relays among the LCD, sheet key an CPU card. Relays among various boards. Relays among various boards (for Y purging). Controls various communications. An amplifier board for the FPD. For solenoid valves. Relays among the heaters except for the isothermal oven, sensors, and TEMP CARD. Controls the EPC. Equipped with the connection terminal for the analog output. A relay board for the analyzer bus. Controls the analog output. An analog output board.
T0301.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >

3.4 Internal Piping System Diagram

(1) GC1000D/GC1000T/GC1000W T ype X Purging
3-7
Isothermal oven
SPLIT1 Vent
B.F .1 Vent TCD1 Vent REF .1 Vent
SPLIT2 Vent
B.F .2 Vent TCD2 Vent REF .2 Vent
FID1 Vent
FID2 Vent
8
7
Programmed
temperature
oven
Sampling base
LSV2
LSV1
Explosionproof enclosure 2
Internal pressure detection switch
Fan
Pressure detection
FID1 FID2
Valve 2 TCD1 TCD2
Valve 6 Valv e 5 Valv e 4
8
7
Pressure detection
Atomospher pressure balancing valve
8 8
7 7
Valve 3
Valve 1
Atomospher pressure balancing valve
Fan
REF
Purge Purge
27
21
2223
24
Pressure detection
16 15 14 14 13 13 12 12 11 11 10 10 9 9 8 8 7 7 6 5 4 3 2 1
24 23 12 11 10
9 8
7 14 13
1
V
2 V2 C1
F1 A1 C2
F2 A2
Restrictor 1 Restrictor 2 Restrictor 3 Restrictor 4 Restrictor 5 Restrictor 6
P
P
P
P
P
P
Flame arrester 1 Flame arrester 2
Forced cooling valve
Cooling air
Voltex tube
27
21
8
Atomospher pressure balancing valve
Hydrogen limitter
Electromagnetic
Purge for air chamber
Isothermal oven
Programed temperature oven
7
Flow rate limitter
Cooling Forced cooling
Explosionproof enclosure 1
Purge
123456781314151622
Atomospher pressure balancing valve
P
Shut off
P
P
P
P
P
Regulator
Electric section (pressurized enclosure)
27
Carrier pressure switch
Pressure regulator
21
Electromagnetic output 1 Electromagnetic output 2 Electromagnetic output 3 Electromagnetic output 4 Electromagnetic output 5 Electromagnetic output 6 Electromagnetic output 7 Electromagnetic output 8
Sample 1 Sample vent 1 Sample 2 Sample vent 2 Carrier 1
H2(N2)-1 for combustion Air-1 for combustion Carrier 2
H2(N2)-2 for combustion Air-2 for combustion H2 Air for combustion
Manifold pressure regulator
Purge air
To sampling system Piping length: 10 m max.
To external sampling system
Samplin base
(*)
High tempreture air
) In the case that EPC is installed.
(
*
C1
F1
A1
C2
F2
A2
Flame arrester
EPC1 EPC3
EPC5 EPC2
EPC4 EPC6
Hydrogen limitter
Flow rate limitter
Shut off
P
Carrier1
H2(N2)-1 for combustion Air-1 for combustion Carrier2
H2(N2)-2 for combustion Air-2 for combustion
H2
Air for combustion
F0312.EPS
IM 11B03A03-01E
F0307.EPS
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >
(2) GC1000T Type Y Purging
Electric section (pressurized enclosure)
3-8
Isothermal oven
SPLIT1 Vent
B.F .1 Vent TCD1 Vent REF .1 Vent
SPLIT2 Vent
B.F .2 Vent TCD2 Vent REF .2 Vent
FID1 Vent
FID2 Vent
8
7
Programmed
temperature
oven
Sampling base
LSV2
LSV1
Internal pressure detection switch
Fan
Pressure detection
FID1 FID2
Valve 2 TCD1 TCD2
Valve 6 Valv e 5 Valv e 4
8
7
Pressure detection
Atomospher pressure balancing valve
8 8
7 7
Valve 3
Valve 1
Atomospher pressure balancing valve
Fan
REF
2223
24
Pressure detection
16 15 14 14 13 13 12 12 11 11 10 10 9 9 8 8 7 7 6 5 4 3 2 1
24 23 12 11 10
9 8
7 14 13
1
V
2 V2 C1
F1 A1 C2
F2 A2
Restrictor 1 Restrictor 2 Restrictor 3 Restrictor 4 Restrictor 5 Restrictor 6
P
P
P
P
P
P
Flame arrester 1 Flame arrester 2
8
Atomospher pressure balancing valve
Hydrogen limitter
Electromagnetic
Purge for air chamber
Isothermal oven
Programed temperature oven
Forced cooling valve
Cooling air
Voltex tube
Cooling Forced cooling
7
Flow rate limitter
Atomospher pressure balancing valve
Shut off
P
P
Regulator
Carrier pressure switch
Purge
Manifold pressure regulator
Pressure regulator
To sampling system Piping length: 10 m max.
Electromagnetic output 1 Electromagnetic output 2 Electromagnetic output 3 Electromagnetic output 4 Electromagnetic output 5 Electromagnetic output 6 Electromagnetic output 7 Electromagnetic output 8
Sample 1 Sample vent 1 Sample 2 Sample vent 2 Carrier 1
H2(N2)-1 for combustion Air-1 for combustion Carrier 2
H2(N2)-2 for combustion Air-2 for combustion H2 Air for combustion
To external sampling system
Samplin base
(*)
123456781314151622
P
P
P
P
Purge air
High tempreture air
) In the case that EPC is installed.
(
*
C1
F1
A1
C2
F2
A2
Flame arrester
EPC1 EPC3
EPC5 EPC2
EPC4 EPC6
Hydrogen limitter
Flow rate limitter
Shut off
P
Carrier1
H2(N2)-1 for combustion Air-1 for combustion Carrier2
H2(N2)-2 for combustion Air-2 for combustion
H2
Air for combustion
F0312.EPS
F0308.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >
(3) GC1000S/GC1000E/GC1000C Type X Pur ging, with One Protection
System
Explosionproof enclosure 1
Electric section (pressurized enclosure)
27
Isothermal oven
Internal pressure detection switch
24
REF
22
Pressure detection
Purge
27
21
3-9
SPLIT1 Vent
B.F .1 Vent TCD1 Vent REF .1 Vent
F .F.1-1 Vent F .F.1-2 Vent
TCD2 Vent REF .2 Vent
SPLIT2 Vent
B.F .2 Vent
F .F.2-1 Vent F .F.2-2 Vent
FID1 Vent
FID2 Vent
8
LSV2
7
LSV1
Programmed
temperature
oven
Sampling base
Fan
Pressure detection
FID1 FID2
Methan Converter
TCD1 TCD2
Valve 6 Valv e 5 Valv e 4
8
8
7
8
7 7
Valve 2
Valve 3
Atomospher pressure balancing valve
Valve 1
Atomospher pressure balancing valve
16 15 14 14 13 13 12 12 11 11 10 10 9 9 8 8 7 7 6 5 4 3 2 1
24
12 11 10
9 8
7 14 13
1
V
2 V2 C1
F1 A1 C2
F2 A2
Restrictor 1 Restrictor 2 Restrictor 3 Restrictor 4 Restrictor 5 Restrictor 6 Restrictor 7 Restrictor 8 Restrictor 9 Restrictor 10
P
P
P
P
P
P
Flame arrester 1 Flame arrester 2
8
Atomospher pressure balancing valve
Hydrogen limitter
Electromagnetic
Purge for air chamber
Isothermal oven
7
Atomospher pressure balancing valve
P
Flow rate
Shut off
limitter
P
P
P
Manifold pressure regulator
Purge
1234567813141522
Carrier pressure switch
21
Pressure regulator
To sampling system Piping length: 10 m max.
Electromagnetic output 1 Electromagnetic output 2 Electromagnetic output 3 Electromagnetic output 4 Electromagnetic output 5 Electromagnetic output 6 Electromagnetic output 7 Electromagnetic output 8
Sample 1 Sample vent 1 Sample 2 Sample vent 2 Carrier 1
H2(N2)-1 for combustion Air-1 for combustion Carrier 2
H2(N2)-2 for combustion Air-2 for combustion H2 Air for combustion
Purge air
To external sampling system
Samplin base
(*)
) In the case that EPC is installed.
(
*
C1
F1
A1
C2
F2
A2
Flame arrester
EPC1 EPC3
EPC5 EPC2
EPC4 EPC6
Hydrogen limitter
Flow rate limitter
Shut off
P
Carrier1
H2(N2)-1 for combustion Air-1 for combustion Carrier2
H2(N2)-2 for combustion Air-2 for combustion
H2
Air for combustion
F0312.EPS
IM 11B03A03-01E
F0309.EPS
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >
(4) GC1000S/GC1000E/GC1000C T ype X Purging, with Two Protection
Systems
3-10
Isothermal oven
SPLIT1 Vent
B.F .1 Vent TCD1 Vent REF .1 Vent
F .F.1-1 Vent F .F.1-2 Vent
TCD2 Vent REF .2 Vent
SPLIT2 Vent
B.F .2 Vent
F .F.2-1 Vent F .F.2-2 Vent
FID1 Vent
FID2 Vent
8
7
Programmed
temperature
oven
LSV2
LSV1
Explosionproof enclosure 2
Internal pressure detection switch
Fan
Pressure detection
FID1 FID2
Methan Converter
TCD1 TCD2
Valve 6 Valv e 5 Valv e 4
8
8
7
8
7 7
Valve 2
Valve 3
Atomospher pressure balancing valve
Valve 1
Atomospher pressure balancing valve
REF
Purge
27
21
22
24
Pressure detection
16 15 14 14 13 13 12 12 11 11 10 10 9 9 8 8 7 7 6 5 4 3 2 1
24
12 11 10
9 8
7 14 13
1
V
2 V2 C1
F1 A1 C2
F2 A2
Restrictor 1 Restrictor 2 Restrictor 3 Restrictor 4 Restrictor 5 Restrictor 6 Restrictor 7 Restrictor 8 Restrictor 9 Restrictor 10
P
P
P
P
P
P
Flame arrester 1 Flame arrester 2
Purge
27
21
8
Atomospher pressure balancing valve
Hydrogen limitter
Electromagnetic
Purge for air chamber
Isothermal oven
Explosionproof enclosure 1
7
Atomospher pressure balancing valve
P
Flow rate
Shut off
limitter
P
P
P
Manifold pressure regulator
Electric section (pressurized enclosure)
27
Carrier pressure switch
Purge
21
1234567813141522
Pressure regulator
To sampling system Piping length: 10 m max.
Electromagnetic output 1 Electromagnetic output 2 Electromagnetic output 3 Electromagnetic output 4 Electromagnetic output 5 Electromagnetic output 6 Electromagnetic output 7 Electromagnetic output 8
Sample 1 Sample vent 1 Sample 2 Sample vent 2 Carrier 1
H2(N2)-1 for combustion Air-1 for combustion Carrier 2
H2(N2)-2 for combustion Air-2 for combustion H2 Air for combustion
Purge air
To external sampling system
Samplin base
(*)
Sampling base
) In the case that EPC is installed.
(
*
C1
F1
A1
C2
F2
A2
Flame arrester
EPC1 EPC3
EPC5 EPC2
EPC4 EPC6
Hydrogen limitter
Flow rate limitter
Shut off
P
Carrier1
H2(N2)-1 for combustion Air-1 for combustion Carrier2
H2(N2)-2 for combustion Air-2 for combustion
H2
Air for combustion
F0312.EPS
F0310.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >
(5) GC1000E T ype Y Purging
Electric section (pressurized enclosure)
3-11
Isothermal oven
SPLIT1 Vent
B.F .1 Vent TCD1 Vent REF .1 Vent
F .F.1-1 Vent F .F.1-2 Vent
TCD2 Vent REF .2 Vent
SPLIT2 Vent
B.F .2 Vent
F .F.2-1 Vent F .F.2-2 Vent
FID1 Vent
FID2 Vent
8
7
Programmed
temperature
oven
LSV2
LSV1
Internal pressure detection switch
Fan
Pressure detection
FID1 FID2
Methan Converter
TCD1 TCD2
Valve 6 Valv e 5 Valv e 4
8
8
7
8
7 7
Valve 2
Valve 3
Atomospher pressure balancing valve
Valve 1
Atomospher pressure balancing valve
REF
22
24
Pressure detection
16 15 14 14 13 13 12 12 11 11 10 10 9 9 8 8 7 7 6 5 4 3 2 1
24
12 11 10
9 8
7 14 13
1
V
2 V2 C1
F1 A1 C2
F2 A2
Restrictor 1 Restrictor 2 Restrictor 3 Restrictor 4 Restrictor 5 Restrictor 6 Restrictor 7 Restrictor 8 Restrictor 9 Restrictor 10
P
P
P
P
P
P
Flame arrester 1 Flame arrester 2
8
Atomospher pressure balancing valve
Hydrogen limitter
Electromagnetic
Purge for air chamber
Isothermal oven
7
Atomospher pressure balancing valve
P
Flow rate
Shut off
limitter
P
P
P
Manifold pressure regulator
Purge
1234567813141522
Carrier pressure switch
Pressure regulator
To sampling system Piping length: 10 m max.
Electromagnetic output 1 Electromagnetic output 2 Electromagnetic output 3 Electromagnetic output 4 Electromagnetic output 5 Electromagnetic output 6 Electromagnetic output 7 Electromagnetic output 8
Sample 1 Sample vent 1 Sample 2 Sample vent 2 Carrier 1
H2(N2)-1 for combustion Air-1 for combustion Carrier 2
H2(N2)-2 for combustion Air-2 for combustion H2 Air for combustion
Purge air
To external sampling system
Samplin base
(*)
Sampling base
) In the case that EPC is installed.
(
*
C1
F1
A1
C2
F2
A2
Flame arrester
EPC1 EPC3
EPC5 EPC2
EPC4 EPC6
Hydrogen limitter
Flow rate limitter
Shut off
P
Carrier1
H2(N2)-1 for combustion Air-1 for combustion Carrier2
H2(N2)-2 for combustion Air-2 for combustion
H2
Air for combustion
F0312.EPS
F0311.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <3. Overview >

3.5 External Input and Output Signals

(1)Input
Item Signal level No. Description
Analog Input
Contact Input
*1) : 2-wire transmitter *2) : It means the data which is averaged by every 1 second
after filtered by a constant which the analog data is scanned by every 200msec. It can be output by DCS communication (Modbus protocol).
*3) : It means the current value which is set time in a cycle time.
Isolated
4-20mA DC 1-5V DC 4-20mA DC (with 24 or 28V DC of Power)
Specification :
5V DC, 20mA DC or more
Input
ON signal: 200 or less OFF signal: 100k or more
On operation:
NC or NO (selectable)
*1)
48Accuracy :
±0.5%FS (-10 to 50˚C)
Function :
Output of Current value and Av erage value
Function :
Alarm from outside Following command request
Stream sequence assign Stream (cont.) assign Stream (1 cycle) assign Cal (Val) assign Change of Operation mode
*2)
*3)
T0302.EPS
3-12
(2)Output
Item Signal level No. Description
Analog Output
Contact Output
Air output for stream valve
*1) : Analysis result (concentration, simulated distillation result,
base level, signal level, noise level, Deviation calculation, Liner calculation1-5, Ratio, Separation coefficient, Calorific value, Density, Compressive factor, Wobbe Index)
Isolated or No-isolated
4-20mA DC Load: 300 or less
Specification (relay) :
Voltage: 30V DC Current: 100mA DC
On operation:
NO or NC (selectable)
Air press : 350kPa
Analysis result
MAX.
36
System alarm1
8
System alarm2 Component alarm (Conc./RT) Timing signal Code signal for stream ID
MAX.
Binary code signal for 9 to 31 stream (max.)
8
*1)
(Max. 5 points)
T0303.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <3. Overview >
(3)Communication
Item Signl level No. Description
DCS communi-cation
Standard :
RS422 (4wires, Full-Duplex)
Specification :
Start bit 1, Stop bit 1, Parity 1, ASCII7 bit, Without procedure or Hand shake
Speed :
1200/2400/4800/9600/19200 bps(selectable)
For explosion protection :
RS422/RS232C converter is provided.(2 wires of power line is needed except the signal line)
The transmission type is full duplex for RS232C.
PC communi-cation
Standard :
RS422 (4wires, Full-Duplex)
Speed :
19.6/33.6kbps
For explosion protection :
RS422/RS232C converter is provided.(2 wires of power line is needed except the signal line)
The transmission type is full duplex for RS232C.
*1) : Analysis result(concentration, simulated distillation result,
base level, signal level, noise level, Deviation calculation, Liner calculation1-5, Ratio, Separation coefficient, Calorific value, Density, Compressive factor, Wobbe Index)
*2) : Operation request, (Stream sequence assign, Stream
(cont.) assign, Run command, Stop command, Pause command, Range change)
*3) : Operation Information(Chromatogram, Oven
temperature, measuring stream, Valve ON/OFF etc...)
11Transmission :
Analysis result Calibration coefficient
Alarm Reception : Operation request
GCMT (GC Maintenance Terminal) Transmission :
Analysis result
Operation information
Parameter list Reception :
Operation request
Change of Parameter list
*4)
*5)
*4)
*5)
T0304.EPS
3-13
*6)
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
Blank Page
<Toc> <Ind> <4. Outline of Software >

4. Outline of Software

4.1 Status and Operation Mode

The GC1000 Mark II has the following statuses:
Process: Normal measurement, calibration and validation
Manual: Manual operation
Lab: Measurement with automatic peak detection, like a laboratory GC
The GC1000 Mark II has the following operation modes:
Stop: Mode in which the measurement stops
Run: Mode in which the measurement is running
Pause: Mode in which the measurement pauses
Figure 4.1 shows the transitions of statuses and operation modes.
4-1
Stop
command
Stop mode
Lab
Run mode
Run
command
Figure 4.1 Status and Operation Mode Transition Diagram
Pause mode Run mode Run mode
Stop command
Change the status Change the status
Process Manual
Run command
Pause command
Pause command
Stop mode Stop mode
Run
command
Stop command
Stop
command
When the LCD panel on the GC1000 Mark II or the LCD emulator on a personal computer (PC) connected with the GC1000 Mark II is used, by changing the operation panel to the manual operation panel or lab operation panel, the status will change from Process to Manual or Lab. Likewise, by changing the manual operation panel or lab operation panel to the operation panel, the status will change from Manual or Lab to Process.
When the GCMT on a PC connected with the GC1000 Mark II is used, the status can be changed through command button operation or menu-driven operation.
The following pages describe each status and operation in detail.
Run
command
F0401.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.1.1 Process

In the Process status normal process measurements, calibration, and validation are per­formed. When power is applied, the status enters Process.
The Process status contains the following measurement statuses.
• Stream Sequence: Continuously measures streams in order specified in Stream Sequence
• Stream (continuous): Continuously measures the specified stream
• Stream (1 cycle): Measures the specified stream once
• Calibration/V alidation: Performs calibration or validation of the specified number
Figure 4.2 shows the transitions of measurement statuses.
4-2
Stream Sequence
(Same as the status)
Note: When Calibration/V alidation has been specified from Stream Sequence, the measurement status returns to Stream
Sequence after the completion of calibration/validation. When Calibration/Validation has been specified from Stream (continuous), the measurement status returns to Stream (continuous) after the completion of calibration/validation.
Figure 4.2 Measurement Status Transition Diagram
Stream (coninuous)
Cal/ValStream (1 cycle) Cal/Val
F0402.EPS
The following explain each measurement status in detail.
(1) Stream Sequence
In the Stream Sequence status, the analyzer continuously measures multiple streams in order. When power is applied, the measurement status enters Stream Sequence. Four kinds of stream sequences can be used with the GC1000 Mark II. For each steam se­quence 31 streams can be set in order. If number "0" is set as a stream number , the measurement skips the order and goes to the stream specified in the following number. Note that number "0" cannot be set as the first stream number, which needs any number between 1 and 31 to be specified.
(2) Stream (continuous)
In the Stream (continuous) status, the analyzer continuously measures the specified stream. With the GC1000 Mark II, 31 streams can be specified. In the Stream (continu­ous) measurement status, the analyzer measures the stream specified from among the 31 streams.
(3) Stream (1 cycle)
In the Stream (1 cycle) status, the GC1000 Mark II measures the specified stream once. Like Stream (continuous), the analyzer measures the stream specified from among the 31 streams.
(4) Calibration/Validation
In the Calibration/V alidation status, calibration or validation is performed using a standard sample. V alidation is to verify, using a standard sample, whether the measurement of the GC1000 Mark II is normal. This function can be used in combination with calibration: calibration is conducted when the increased error between a measurement result and a standard sample is seen during the regular validation.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >
Three ways of calibration and three ways of validation can be used with the GC1000 Mark II.
There are three methods for Calibration/V alidation: automatic, semiautomatic and manual. The following explain the actions of three methods.
•Automatic
Calibration or validation is conducted automatically at the specified time. This function is only available when automatic valves are used for streams for calibration or valida­tion.
When Calibration/V alidation is set to automatic, the analyzer starts calibration or validation at the specified starting time and subsequently conducts calibration or validation every time after the specified interval time elapses. When the operation mode is in Stop or Pause at the time of automatic calibration or validation, the calibra­tion or validation for the time is cancelled.
• Semiautomatic When the calibration number or validation number is specified, the valves for streams
for calibration or validation automatically open and the analyzer performs calibration or validation. This function is only available when automatic valves are used for the stream.
• Manual
4-3
When the calibration number or validation number is specified, the valves for all streams are closed so that the analyzer is ready to allow a standard sample to flow. The standard sample is supplied manually for displacement, making the analyzer ready to accept a command. After the displacement is conducted adequately, the analyzer performs calibration or validation by the start command.
Each measurement status has the following operation modes, with the exception of the Stream (1 cycle) and Calibration/V alidation statuses which do not have Stop and Pause.
• Stop: Mode in which the measurement stops
• Run: Mode in which the measurement is running
• Pause: Mode in which the Run mode pauses
For the transitions of the operation modes, refer to Figure 4.1. To transfer from the Run mode to the Stop mode, two commands are used: stop command
and forced stop command, which are explained below.
• Stop command
The measurement stops after an ongoing measurement has been completed.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >
The following explain each operation mode in detail.
(1) Stop
In the Stop mode the measurement stops. When power is applied, the operation mode enters Stop. Settings should be changed in this Stop mode.
CAUTION
If settings are changed in an operation mode other than the Stop mode, the analyzer may not accept the change and run under the previous setting.
When the operation mode is transferred to the Stop mode from others, the on-off statuses of various valves and stream valves remains unchanged.
(2) Run
In the Run mode, the measurement runs. If the Auto Start Set is set to Enabled, the GC1000 Mark II automatically enters the Run mode after the time set for the "Auto start time" elapses after power is applied. When the operation mode is transferred from the Stop mode to the Run mode, the measurement starts after the Warming up time for dis­placement in the sample stream elapses. For the Warming up time, see Section 4.3, “Method”. If a level 1 alarm occurs during the Run mode, the mode changes to the Stop mode after an ongoing measurement of the stream has been completed.
4-4
(3) Pause
In the Pause mode, the measurement pauses at the specified Pause Time. For the Pause Time, see Section 4.3, “Method.”
When the operation mode is transferred from the Pause mode to the Run mode, the measurement resumes at the Pause Time.
When the operation mode is transferred from the Stop mode to the Pause mode, the on-off statuses of various valves including stream valves become those at the Pause Time of the stream to be measured next.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.1.2 Manual

In the Manual status, various operations and measurements can be performed manually . The operations available in the Manual mode are:
• On/off of various valves
• On/off of stream valves
• On/off of several heaters
• On/off of several types of detector (only On for FID and FPD)
When the status is transferred from Process to Manual, all the various valves and stream valves turn Off.
The Manual mode contains the following operation modes.
• Stop: Mode in which measurement stops
• Run: Mode in which measurement runs
For the transitions of the operation modes, refer to Figure 4.1. The following explain each operation mode in detail.
(1) Stop
In the Stop mode the manual measurement stops.
4-5

4.1.3 L a b

(2) Run
In the Run mode the manual measurement runs. During the Run mode, the GC1000 Mark II detects peaks automatically until the mode turns to Stop. When the mode turns to Stop, the concentration in area percentage is calculated from the area of peaks detected. The calculated concentration values are displayed on the Laboratory Analysis Results display panel. For five seconds after the operation mode turns from Run to Stop, the GC1000 Mark II does not accept the run command in order to process analysis calculations and others.
In the Lab status the lab measurement runs. The GC1000 Mark II performs continuous measurements for times specified in the lab settings.
In the Run mode in Process, peaks are detected according to the on/off of the set gate, while in Lab, peaks are detected automatically like a laboratory GC. This mode can be used for determining gate timing when a column system is changed.
The Lab mode contains the following operation modes.
• Stop: Mode in which measurement stops
• Run: Mode in which measurement runs
For the transitions of the operation modes, refer to Figure 4.1. The following explain each operation mode in detail.
(1) Stop
In the Stop mode the lab measurement stops.
(2) Run
In the Run mode the lab measurement runs. In the Run mode, GC1000 Mark II performs measurements according to the lab settings. If a level 1 alarm occurs during the Run mode, the mode changes to the Stop mode after an ongoing lab measurement has been completed.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.2 Stream

In the GC1000 Mark II, 31 streams can be used. Parameters needed to be specified for streams are:
• Stream valve number
Any one stream valve number from among the numbers of 1 to 31 should be specified for a stream.
If the stream valves are automatic, the stream valve of the specified valve turns on/off. If the stream valves are manual, this setting does not have any effect on the operation.
The same stream valve number can be specified for different streams. When multiple measurements are performed in one stream, the same stream valve number should be specified.
A standard sample stream should be specified as one of the streams.
• Method number
Any number from 1 to 4 should be specified for a method number.
There are three types for the stream:
• Measurement stream
4-6
Used for measuring process samples.
• Validation stream
Comparing measured results before and after calibration, or checking if the GC1000 Mark II is operating properly can be performed using a standard sample. The validation stream is used for standard sample measurement in such case.
• Calibration stream
Used for measuring a standard sample at calibration. The calibration stream may be used in combination with the validation stream.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.3 Method

Activation timing for various valves and parameters with regard to the temperature of the isothermal/programmed-temperature ovens should be specified for Method. Four Methods can be used with the GC1000 Mark II. Parameters needed to be specified are described below. For setting procedures, refer to the LCD Panel Operation Manual, IM 11B03A03-05 E.
• Analysis cycle
Time from start (0 second) to stop of analysis
• Warming up time Time required for displacement in sample streams. An appropriate displacement time
should be determined considering the diameter and length of the sample stream tubing, sample flow rate and delay time.
A positive value should be specified for the W arming up time. But displacement in sample streams is implemented before the start of analysis (0 second) so that the Warming up time actually acts as a negative value.
• Peak detection stop time
Time to stop peak detection compulsorily, which is usually specified to around five seconds earlier than the pause time. The time must be the analysis cycle time minus one second, or less.
4-7
In case of the GC1000D/GC1000T/GC1000W, the heater should be turned of f at the time of peak detection stop time.
• Pause time
Time when the GC1000 Mark II can be stopped without adversely affecting the column system. Usually the pause time should be set after peaks of all components have been eluted (after the peak detection stop time) for the analysis cycle time minus 3 seconds, or less.
• DO on time/DO off time
These can be used for notifying a supervisory computer or system of analysis end timing and other information by using contacts. They can also be used as a reading signal when analog outputs are employed.
• Stream valve on time/stream valve off time
Time to turn on/off stream valves. As stream valves usually turns on from the Warming up time, the stream valve on time should be set as a negative value. The values to be set should be the Warming up time minus one second, or more, and the analysis cycle time minus two seconds, or less.
• V1 to V8 on/off time
Time to turn on/off valve 1 to valve 8 - sample valves, back flush valves, column switching valves, and atmosphere balance valves. On/off can be set for three times each. The time to be set should be the Warming up time minus one second, or more, and the Analysis Cycle Time minus two seconds, or less.
• Initial control temperature
Displayed on the GC1000D/GC1000T/GC1000W that uses a temperature program. The temperature of the programmed-temperature oven at the start of analysis should be speci­fied.
• Initial temperature time
Displayed on the GC1000D/GC1000T/GC1000W that uses a temperature program. The time for holding the initial control temperature should be specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >
• The first temperature gradient/first control temperature/fir st temperature time (only for the GC1000D/GC1000T/GC1000W)
• The second temperature gradient/second control temperature/second temperature time (only for the GC1000D/GC1000T/GC1000W)
• The thir d temperature gradient/third contr ol temperature /thir d temperature time (only for the GC1000D/GC1000T/GC1000W)
• Initial control pressure Displayed when EPC is installed that uses a pressure program. The pressure at the
start of analysis should be specified.
• Initial pressure time Displayed when EPC is installed that uses a pressure program. The time for holding
the initial control pressure should be specified.
• First pressure gradient/first control pressure/fir st control time (only when EPC is installed)
• Second pressure gradient/second control pressure/second control time (only when EPC is installed)
• Thir d pressure gradient/third contr ol pressure/thir d control time (onl y when EPC is installed)
4-8
In the GC1000D/GC1000T/GC1000W, three stages of temperature program can be speci­fied. Figure 4.3 shows the operations for temperature programs.
Temperature
Third control
temperature
Second control
temperature
First control
temperature
Initial control
temperature
Figure 4.3 Operation for Temperature Programs
First
temperature
gradient
Initial
temperature
time
Second
temperature
gradient
First
temperature
time
Third
temperature
gradient
Second
temperature
time
Analysis cycle
End of peak detection
time
Heater off
Third
temperature
time
Heater on
Analysis endAnalysis start
Time
F0403.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >
In the case that EPC is installed, three stages of pressure program can be specified. Figure 4.4 shows the operations for EPC pressure programs.
4-9
Pressure
Third control
pressure
Second control
pressure
First control
pressure
Initial control
pressure
Figure 4.4 Operation for EPC Pressure Programs
First pressure gradient
Initial
control
time
Second
pressure
gradient
First
control
time
Analysis cycle
Third
pressure
gradient
Second
control
time
End of peak detection
time
Third
control
time
Analysis endAnalysis start
Time
F0404.EPS

4.4 Description of Actions

Actions under the conditions below - stream sequence in Table 4.1, calibration in Table 4.2, and validation in Table 4.3 - are described as examples.
T able 4.1
First stream
number
Stream Sequnce 1 Stream 1 Stream 2 Stream 3 0 Stream Sequnce 2 Stream 4 0 Stream 5 Stream 6
T able 4.2
Calibration
stream number
Calibration 1 Stream 7 2 Stream 8
T able 4.3
Validation
stream number
Validation 1 Stream 9 2
Second stream
number
Measuring
times
Measuring
times
T0403.EPS
Third stream
number
Validation
stream number
T0402.EPS
Fourth stream
number
T0401.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.4.1 Actions of Stream Sequence

The same action is performed when the Stream Sequence is specified at any timing during analysis.
(1) An action when the operation mode is transferred from Stop to Run while the mea-
surement status is in Stream Sequence 1, is shown in Figure 4.5.
Run command
S1 S2 S3 S1 S2 S3
W2 W3 W1 W2 W3 W1W1
Time
Stream Sequence 1
S1 to S3: Analysis cycles for streams 1 to 3 W1 to W3: Warming up time for streams 1 to 3
Figure 4.5 Action of Stream Sequence (1)
(2) An action when Stream Sequence 2 is specified while the measurement status is in
Stream Sequence 1 and the operation mode is in Run, is shown in Figure 4.6.
4-10
F0405.EPS
Selet the Stream Sequence 2
S1 S2 S5
W4
Time
Stream Sequence 1 Stream Sequence 2
S1 to S6: Analysis cycles for streams 1 to 6 W1 to W6: Warming up time for streams 1 to 6
S4 S6 S4
W5 W6 W4 W5W2
F0406.EPS
Figure 4.6 Action of Stream Sequence (2)
(3) An action when Stream Sequence 1 is specified while the measurement status is in
Stream (continuous) 2 and the operation mode is in Run, is shown in Figure 4.7.
Select the Stream Sequence 1
S2 S2 S2
W1
Time
Stream (Cont) 2 Stream Sequence 1
S1 S3 S1
W2 W3 W1 W2W2
S1 to S3: Analysis cycles for streams 1 to 3 W1 to W3: Warming up time for streams 1 to 3
F0407.EPS
Figure 4.7 Action of Stream Sequence (3)
(4) When Stream Sequence 1 is specified while the measurement status is in Stream
Sequence 2 to 4 or in Stream (continuous) 1 to 31 and the operation mode is in Stop or in Pause, the measurement status is immediately changed to Stream Sequence 1 and the operation mode to Stop. The same action is performed when Stream Se­quence 2 to 4 is specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.4.2 Actions of Stream (Continuous)

Actions are different according to the timing when Stream (continuous) is specified. (1) An action when the operation mode is transferred from Stop to Run while the mea-
surement status is in Stream (continuous) 2 and the operation mode is in Stop, is shown in Figure 4.8.
Run command
S2 S2 S2 S2 S2 S2
W2 W2 W2 W2 W2 W2W2
Time
Stream (Cont) 2
S2: Analysis cycles for stream 2 W2: Warming up time for stream 2
Figure 4.8 Action of Stream (Continuous) (1)
(2) An action when Stream (continuous) 3 is specified before the following warming up
time starts while the measurement status is in Stream (continuous) 2 and the opera­tion mode is in Run (i.e., Stream 2 is being measured), is shown in Figure 4.9.
4-11
F0408.EPS
Select the Stream (cont) 3
S2S2 S2 S3 S3 S3 S3
W2 W3 W3 W3 W3 W3W2
Time
Stram (Cont) 2
S2, S3: Analysis cycles for streams 2 and 3 W2, W3: Warming up time for streams 2 and 3
Stream (Cont) 3
F0409.EPS
Figure 4.9 Action of Stream (Continuous) (2)
(3) An action when Stream (continuous) 3 is specified after the following warming up time
has started while the measurement status is in Stream (continuous) 2 and the opera­tion mode is in Run (i.e., Stream 2 is being measured), is shown in Figure 4.10.
Select the Stream (cont) 3
S2S2 S2 S2 S3 S3 S3
W2 W2 W3 W3 W3 W3W2
Time
Stram (Cont) 2
Stram (Cont) 3
S2, S3: Analysis cycles for streams 2 and 3 W2, W3: Warming up time for streams 2 and 3
Figure 4.10 Action of Stream (Continuous) (3)
IM 11B03A03-01E
F0410.EPS
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >
(4) An action when Stream (continuous) 4 is specified while the measurement status is in
Stream Sequence 1 and the operation mode is in Run, is shown in figure 4.1 1.
Select the Stream (cont) 4
4-12
S1 S2 S4
W4
Time
Stream Sequence 1 Stream (Cont) 4
S1 to S4: Analysis cycles for streams 1 to 4 W1 to W4: Warming up time for streams 1 to 4
S4 S4 S4
W4 W4 W4 W4W2
F0411.EPS
Figure 4.11 Action of Stream (Continuous) (4)
(5) When stream (continuous) 1 is specified while the measurement status is in Stream
Sequence 1 to 4 or in Stream (continuous) 2 to 31 and the operation mode is in Stop or in Pause, the measurement status is immediately changed to Stream (continuous) 1 and the operation mode to Stop. The same action is performed when Stream (continuous) 2 to 31 is specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.4.3 Actions of Stream (1 cycle)

Stream (1 cycle) 1 to 31 is allowed to be specified only when the measurement status is in Stream Sequence 1 to 4. The same action is performed when the Stream (1 cycle) is specified at any timing during analysis.
(1) When Stream (1 cycle) 1 to 31 is specified while the measurement status is in Stream
Sequence 1 to 4 and the operation mode is in Stop or in Pause, the specified stream is measured once and then the measurement status turns to Stream Sequence 1 and the operation mode returns to Stop.
(2) An action when Stream (1 cycle) 4 is specified while the measurement status is in
Stream Sequence 1 and the operation mode is in Run, is shown in Figure 4.12.
Select the Stream (1 cycle) 4
4-13
S1 S2 S3
W4
Time
Stream Sequence 1 Stream Sequence 1Stream (1 cycle) 4
S1 to S4: Analysis cycles for streams 1 to 4 W1 to W4: Warming up time for streams 1 to 4
Figure 4.12 Action of Stream (1 cycle)
S4 S1 S2
W3 W1 W2 W3W2
F0412.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.4.4 Actions of Calibration

Calibration 1 to 3 is allowed to be specified only when the measurement status is in Stream Sequence 1 to 4 or in Stream (continuous) 1 to 4. Actions are dif ferent according to the specified calibration (validation) method.
• Automatic Calibration
The same action is performed when the starting time comes at any timing during analysis. (1) An action when the starting time of Automatic Calibration 1 comes while the measure-
ment status is in Stream Sequence 1 and the operation mode is in Run, is shown in Figure 4.13.
Starting time of Automatic Calibration 1
4-14
S1 S2 S7
W7
Time
Stream Sequence 1 Stream Sequence 1Calibration 1
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8
S7 S8 S3
W7 W8 W3 W1W2
F0413.EPS
Figure 4.13 Action of Automatic Calibration (1)
(2) An action when the starting time of Automatic Calibration 1 comes while the measure-
ment status is in Stream (continuous) 1 and the operation mode is in Run, is shown in Figure 4.14.
Starting time of Automatic Calibration 1
S1 S1 S7
W7
Time
Stream (Cont) 1 Stream (Cont) 1Calibration 1
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8
Figure 4.14 Action of Automatic Calibration (2)
S7 S8 S1
W7 W8 W1 W1W1
F0414.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >
• Semiautomatic Calibration
The same action is performed when the calibration is specified at any timing during analy­sis.
(1) An action when Calibration 1 is specified while the measurement status is in Stream
Sequence 1 and the operation mode is in Run, is shown in Figure 4.15.
Select the Calibration 1
4-15
S1 S2 S7
W7
Time
Stream Sequence 1 Stream Sequence 1Calibration 1
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8
S7 S8 S3
W7 W8 W3 W1W2
F0415.EPS
Figure 4.15 Action of Semiautomatic Calibration (1)
(2) An action when calibration 1 is specified while the measurement status is in Stream
(continuous) 1 and the operation mode is in Run, is shown in Figure 4.16.
Select the Calibration 1
S1 S1 S7
W7
Time
Stream (Cont) 1 Stream (Cont) 1Calibration 1
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8
Figure 4.16 Action of Semiautomatic Calibration (2)
S7 S8 S1
W7 W8 W1 W1W1
F0416.EPS
(3) When Calibration 1 is specified while the measurement status is in Stream Sequence
1 to 4 or in Stream (continuous) 1 to 31 and the operation mode is in Stop or in Pause, the measurement status is immediately changed to Calibration 1 and then Calibration 1 is performed. After Calibration 1 has been completed, the previous measurement status is retrieved and the operation mode turns to Stop. The same action is per­formed when Calibration 1 or 2 is specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >
• Manual Calibration
The same action is performed when the calibration is specified at any timing during analy­sis.
(1) An action when Calibration 1 is specified while the measurement status is in Stream
Sequence 1 and the operation mode is in Run, is shown in Figure 4.17. After Calibra­tion 1 has been completed, the measurement status returns to Stream Sequence 1 and the operation mode turns to Stop.
Finishing the calibrationSelect the Calibration 1
Starting the calibration
4-16
S1 S7
TE TE
Time
Stream Sequence 1 Stream Sequence 1Calibration 1
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8 TE: Displacement enabled (repeat of 20-second cycle)
S7 S8
W7 W8W7
F0417.EPS
Figure 4.17 Action of Manual Calibration (1)
(2) An action when Calibration 1 is specified while the measurement status is in Stream
Sequence (continuous) 1 and the operation mode is in Run, is shown in Figure 4.18. After Calibration 1 has been completed, the measurement status returns to Stream (continuous) 1 and the operation mode turns to Stop.
Finishing the calibrationSelect the Calibration 1
Starting the calibration
S1 S7
TE TE
Time
Stream (Cont) 1 Stream (Cont) 1Calibration 1
S7 S8
W7 W8W7
S1 to S8: Analysis cycles for streams 1 to 8 W1 to W8: Warming up time for streams 1 to 8 TE: Displacement enabled (repeat of 20-second cycle)
F0418.EPS
Figure 4.18 Action of Manual Calibration (2)
(3) When calibration 1 is specified while the measurement status is in Stream Sequence
1 to 4 or in Stream (continuous) 1 to 31 and the operation mode is in Stop or in Pause, displacement become available immediately . The following actions are the same as those described in (1) or (2). Refer to the action after the start of calibration in Figure
4.17 or Figure 4.18. The same action is performed when Calibration 2 or 3 is speci­fied.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.4.5 Actions of Validation

V alidation 1 to 3 is allowed to be specified only when the measurement status is in Stream Sequence 1 to 4 or in Stream (continuous) 1 to 31. Actions are dif ferent according to settings. The same action is performed when the starting time comes at any timing during analysis.
• Automatic V alidation
(1) An action when the starting time of Automatic Validation 1 comes while the measure-
ment status is in Stream Sequence 1 and the operation mode is in Run, is shown in Figure 4.19.
Starting time of Automatic Validation 1
4-17
S1 S2 S9
W9
Time
Stream Sequence 1 Stream Sequence 1Validation 1
S1 to S9: Analysis cycles for streams 1 to 9 W1 to W9: Warming up time for streams 1 to 9
S9 S3 S1
W9 W3 W1 W2W2
F0419.EPS
Figure 4.19 Action of Automatic Validation (1)
(2) An action when the starting time of Automatic Validation 1 comes while the measure-
ment status is in Stream (continuous) 1 and the operation mode is in Run, is shown in Figure 4.20.
Starting time of Automatic Validation 1
S1 S1 S9
W9
Time
Stream (Cont) 1 Stream (Cont) 1Validation 1
S1, S9: Analysis cycles for streams 1 and 9 W1, W9: Warming up time for streams 1 and 9
Figure 4.20 Action of Automatic Validation (2)
S9 S1 S1
W9 W1 W1 W1W1
F0420.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >
• Semiautomatic Validation
The same action is performed when the validation is specified at any timing during analy­sis.
(1) An action when V alidation 1 is specified while the measurement status is in Stream
Sequence 1 and the operation mode is in Run, is shown in Figure 4.21.
Select the Validation 1
4-18
S1 S2 S9
W9
Time
Stream Sequence 1 Stream Sequence 1Validation 1
S1 to 9: Analysis cycles for streams 1 to 9 W1 to W9: Warming up time for streams 1 to 9
S9 S3 S1
W9 W3 W1 W2W2
F0421.EPS
Figure 4.21 Action of Semiautomatic Validation (1)
(2) An action when V alidation 1 is specified while the measurement status is in Stream
(continuous) 1 and the operation mode is in Run, is shown in Figure 4.22.
Select the Validation 1
S1 S1 S9
W9
Time
Stream (Cont) 1 Stream (Cont) 1Validation 1
S1, 9: Analysis cycles for streams 1 and 9 W1, W9: Warming up time for streams 1 and 9
Figure 4.22 Action of Semiautomatic Validation (2)
S9 S1 S1
W9 W1 W1 W1W1
F0422.EPS
(3) When V alidation 1 is specified while the measurement status is in Stream Sequence 1
to 4 or in Stream (continuous) 1 to 31 and the operation mode is in Stop or in Pause, the measurement status is immediately changed to V alidation 1 and then Validation 1 is performed. After Validation 1 has been completed, the previous measurement status is retrieved and the operation mode turns to Stop. The same action is per­formed when V alidation 2 or 3 is specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >
• Manual Validation
The same action is performed when the validation is specified at any timing during analy­sis.
(1) An actin when V alidation 1 is specified while the measurement status is in Stream
Sequence 1 and the operation mode is in Run, is shown in Figure 4.23. After Valida­tion 1 has been completed, the measurement status returns to Stream Sequence 1 and the operation mode turns to Stop.
Finishing the validationSelect the Validation 1
Starting the validation
4-19
S1 S9
TE TE
Time
Stream Sequence 1 Stream Sequence 1Validation 1
S1 to S9: Analysis cycles for streams 1 to 9 W1 to W9: Warming up time for streams 1 to 9 TE: Displacement enabled (repeat of 20-second cycle)
S9
W9W9
F0423.EPS
Figure 4.23 Action of Manual Validation (1)
(2) An action when V alidation 1 is specified while the measurement status is in Stream
(continuous) 1 and the operation mode is in Run, is shown in Figure 4.24. After V alidation 1 has been completed, the measurement status returns to stream (continu­ous) 1 and the operation mode turns to Stop.
Starting the validation
S1 S9
TE TE
Time
Stream (Cont) 1 Stream (Cont) 1Validation 1
S9
W9W9
Finishing the validationSelect the Validation 1
S1, 9: Analysis cycles for streams 1 and 9 W1, W9: Warming up time for streams 1 and 9 TE: Displacement enabled (repeat of 20-second cycle)
F0424.EPS
Figure 4.24 Action of Manual Validation (2)
(3) When V alidation 1 is specified while the measurement status is in Stream Sequence 1
to 4 or in Stream (continuous) 1 to 31 and the operation mode is in Stop or in Pause, displacement become available immediately . The following actions are the same as those described in (1) or (2). Refer to the action after the start of validation in Figure
4.23 or Figure 4.24. The same action is performed when V alidation 2 or 3 is specified.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.5 Computation and Processing

The following are processed using chromatograms obtained as detection signals. For the setting procedure, see the LCD Panel Operation Manual, IM 1 1B03A03-05E.
Peak processing
Deviation processing
Additional processing
Signal processing
Each processing is outlined below. For details, see the Technical Guide, IM 11B03A03-03E.

4.5.1 Peak Processing

This processing detects peaks from a detector signal and determines the concentrations of components based on the area or height of each peak. The retention times, peak half-width and so on are also determined from the detected peaks.

4.5.2 Deviation processing

This processing determines the value obtained by subtracting the sum of the concentra­tions of components computed by peak processing from the total sum of set concentra­tions.
4-20

4.5.3 Additional processing

There are the following three types of processing in the additional processing:
Linear computation processing This processing determines the total sum of the concentrations of each component
computed in the peak processing multiplied by individual coefficients.
Ratio computation processing This processing determines the ratio of concentrations of two components obtained
by the peak processing.
Separation capability processing This processing determines the separation capability using retention time and peak
half-widths of two adjacent components determined by the peak processing.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <4. Outline of Software >

4.5.4 Signal Processing

There are three types of processing in the signal processing:
Base level processing The detector signal at a time set for base level processing within the analysis cycle
avoiding component signals is used as the base level value. If the base level process­ing is not set, the detector signal immediately after the start of analysis is employed as the base level value.
Signal level processing The differences between the detector signals at each time set for signal level process-
ing within the analysis cycle and the base level value, are used as the signal level values. If more than one base level processing is set, the base level value immedi­ately before a base level processing is used for signal level processing.
Noise level processing Among the preceding 20 data for detector signals at a time set for noise level process-
ing within the analysis cycle, the maximum and minimum values are used as noise level values.
4-21
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <4. Outline of Software >

4.6 Alarm Processing

There are four alarm levels. For details of alarms, see the Alarm Message Manual, IM 1 1B03A03-06E.
Level 1 Alarms for system or hardware failures. When an alarm is generated, the alarm
condition is maintained until it is reset. If a level-1 alarm is generated in the Run or Lab mode, the operation mode changes to
Stop at the end of the measurement being executed.
Level 2 Alarms for measuring conditions failure. When an alarm is generated, the alarm condition is maintained until it is reset.
Component Alarms when measurement results such as concentrations and retention times be-
come out of each specified range. If an alarm is generated, the alarm condition is maintained until it is reset.
Level 3
4-22
Alarms for minor failures or information other than level-1 and level-2 alarms. The alarm condition is not maintained.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >
5-1

5. Actions of External Input and Output Signals

5.1 Analog Hold Output

Up to 36 outputs are available as option. Actions of analog hold output are different between when the actual stream is specified and
when "99" is specified for the stream number in setting analog hold output. Actions of contact output are different whether a contact output is used or not as a stream identifica­tion signal. The following explains actions of analog output as well as contact output. If "99" is specified for the stream number, contact output must be used as a stream identifica­tion signal. For actions of contact output other than stream identification signals, refer to Section 5.2, “Contact Output.”
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >

5.1.1 When Actual Stream is Specified for Stream Number (without Stream Identification Signal)

An action when analog hold output is specified as shown in Table 5.1, is illustrated in Figure
5.1.
Table 5.1 Setting of Analog Output
Output stream #1
CH.1 CH.2 CH.3 CH.4 CH.5 CH.6 CH.7 CH.8
Output peak #1
1 1 1 1 2 2 2 2
1 2 3 4 1 2 3 4
Output stream #2
None None None None None None None None
Output peak #2
None None None None None None None None
T0501.EPS
5-2
Stream 1
End of Analysis
*1 *1 *1 *1
CH.1
CH.2
Analog Hold Output
CH.3
CH.4
CH.5
CH.6
CH.7
CH.8
End of Analysis: Time when the final peak detection is completed or peak detection stop time. *1: One to two seconds (depending on analysis specifications)
Figure 5.1 Action of Analog Output
Stream 2
End of Analysis
Stream 1
End of Analysis
Stream 2
End of Analysis
F0501.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >

5.1.2 When Actual Stream is Specified for Stream Number (with Stream Identification Signal)

An action when analog hold output is specified as shown in Table 5.2, is illustrated in Figure
5.2.
Table 5.2 Setting of Analog Output
Output stream #1
CH.1 CH.2 CH.3 CH.4
Output peak #1
1 1 1 1
1 2 3 4
Output stream #2
2 2 2 2
Output peak #2
1 2 3 4
T0502.EPS
5-3
Stream 1
End of Analysis
Analog Hold Output Contact Output
CH.1
CH.2
CH.3
CH.4
CH.3
(Read-in signal)
CH.4
(Read-in signal)
CH.5
(Read-in signal)
End of Analysis: Time when the final peak detection is completed or peak detection stop time. *1: One to two seconds (depending on analysis specifications)
*1 2s4s 2s4s 2s4s 2s4s*1 *1 *1
Stream 2
End of Analysis
Figure 5.2 Actions of Analog Output and Contact Output
Stream 1
End of Analysis
Stream 2
End of Analysis
F0502.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >

5.1.3 When "99" is Specified for Stream Number (with Stream Identification Signal)

An action when analog hold output is specified as shown in Table 5.4 under the peak setting conditions of each stream as shown in Table 5.3, is illustrated in Figure 5.3.
Table 5.3 Conditions of Peaks for Streams
Stream 1
Peak 1 Peak 2 Peak 3 Peak 4
Exist Exist Exist Exist
Table 5.4 Setting of Analog Output
Output stream #1
CH.1 CH.2 CH.3 CH.4
99 99 99 99
Stream 2
Exist
None
Exist
None
Output peak #1
Stream 3
Exist
Exist None Exist
T0503.EPS
1 2 3 4
Output stream #2
None None None None
Output peak #2
None None None None
T0504.EPS
5-4
Stream 1
End of Analysis
Analog Hold Output Contact Output
CH.1
CH.2
CH.3
CH.4
CH.3
(Read-in signal)
CH.4
(Read-in signal)
CH.5
(Read-in signal)
End of Analysis: Time when the final peak detection is completed or peak detection stop time. *1: One to two seconds (depending on analysis specifications)
*1 2s4s 2s4s 2s4s 2s4s*1 *1 *1
Stream 2
End of Analysis
Figure 5.3 Actions of Analog Output and Contact Output
Stream 1
End of Analysis
Stream 2
End of Analysis
F0503.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >

5.2 Contact Output

The following five types can be specified for up to eight contact outputs as standard.
• Stream Sequence
• Stream
• Operation Mode
• Alarm
• Timing
Actions when each type is specified, are described below.
CAUTION
If settings are changed in the DO Setup panel while the contact output is On, the contact output may remain unchanged in On until the power is turned off and on. In changing settings, make sure that the contact output is Off.
5-5

5.2.1 Stream Sequence

When the stream sequence with the number specified in the DO Setup panel runs, the contact output turns On.
Actions when Stream Sequence 1 is specified for Contact Output 1, are described below. (1) When the operation mode is changed from Stop to Run:
When the Warming up time for the first stream of Stream Sequence 1 starts, the contact output turns On.
(2) When the operation mode is changed from Run to Stop:
The contact output is kept On until the measurement status is changed next.
(3) When the operation mode is changed from Run to Pause:
The contact output is kept On.
(4) When Stream Sequence is changed from 1 to 2:
When the Warming up time for the first stream of Stream Sequence 2 starts, the contact output turns Off.
(5) When Stream Sequence is changed from 2 to 1:
When the Warming up time for the first stream of Stream Sequence 1 starts, the contact output turns On.
(6) When Stream Sequence 1 is changed to other measurement status:
When the Warming up time for the first stream of the other measurement status starts, the contact output turns Off.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >

5.2.2 Stream

When the stream with the number specified in the DO Setup panel runs, the contact output turns On.
Actions when Stream 1 is specified for Contact Output 1, are described below. (1) When the operation mode is changed from Stop to Run:
When the Warming up time of Stream 1 starts, the contact output turns On.
(2) When the operation mode is changed from Run to Stop:
The contact output is kept On until the measurement status is changed next.
(3) When the operation mode is changed from Run to Pause:
The contact output is kept On.
(4) When Stream (continuous) is changed from 1 to 2:
When the analysis of Stream 2 starts, the contact output turns Off.
(5) When Stream (continuous) is changed from 2 to 1:
When the analysis of Stream 1 starts, the contact output turns On.
(6) When Stream (continuous) 1 is changed to other measurement status:
When the Warming up time for the first stream of the other measurement status starts, the contact output turns Off.
5-6

5.2.3 Op eration Mode

When the operation mode specified in the DO Setup panel is enabled, the contact output turns On. If Run is specified, the start of the Warming up time triggers the contact output to turn On.

5.2.4 Alarm

If an alarm specified in the DO Setup panel is generated, the contact output turns On. When the alarm is cleared, the contact output turns Off.

5.2.5 Timing

The contact output turns on and off according to the timing specified in the Method Setup panel.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >

5.3 Contact Input

The following six types can be specified for up to eight contact inputs as standard.
• Stream Sequence command
• Stream (continuous) command
• Stream (1 cycle) command
• Calibration (V alidation)
• Operation Mode
• Alarm processing
CAUTION
Contact input requires closed, pulse input. For input, closed, pulse input for at least 2 seconds must be performed. Closed, pulse input for less than 2 seconds may result in invalid function. Status input must not be performed.
5-7
For actions other than alarm processing, refer to Section 4.4, “Description of Actions.” When alarm processing is specified for contact input, the alarm specified in the DI Setup
panel is generated.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >
5-8

5.4 Communication Input and Output

5.4.1 GC6 T ype Output Data Format (Fixed to 45 Characters)

(1)Analysis value data (component concentration)
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
DS
a: Stream number (01 to 24) b: Component number (01 to 24) c: Component concentration (without zero suppression) d: Unit e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL) f: Retention time (seconds) g: Retention time error alarm Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs)
,,
bacd e fg
PPM A: CHL
%
,,,
A: CLL
T
(2)Analysis value data (component ratio operation, linear polynormial operation)
A: RT
Dummy
CR: Carriage return LF: Line feed
space
F0504.EPS
C R
L F
1 2345678 9 10 11 12 13 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
DS
a: Stream number (01 to 24) b: Component number (01 to 24) c: Operation value (without zero suppression) d: Unit
,,
bac
(3)Calibration factor
1 234567 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
CC
a: Standard sample (STD1=1, STD2=2) b: Component number (01 to 24) c: Calibration factor (without zero suppression) d: Calibration factor error (only if an error occurs)
,,
bacd
(4)Error alarm
• Memory pattern check error • Watchdog timer
• Calibration out of range • Detector 1 (or 2) calibration error
• Isothermal oven temperature error • Power off
• Pressure switch 1 (or 2) OFF • External contact inputs 1 to 8
• Communication error • FID1 (or 2) extinguished
• Calibration repeatability error
14
PPM
%
d
COE
,
Dummy space
Dummy space
CRL
F0505.EPS
CRL
F0506.EPS
F
F
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
AS / :
a: Month, b: Day, c: Hour, d: Minute e: Error alarm types
,,
bacde
Note: For 3 characters, the 18th character is a dummy space.
Dummy space
MEM: Memory pattern check error WDT: Watchdog timer AD1: Detector 1 calibration error AD2: Detector 2 calibration error TMPH: Isothermal oven temperature error CAR1: Pressure switch 1 OFF CAR2: Pressure switch 2 OFF
EXT1: External contact
input 1
EXT2: External contact
input 2
EXT8: External contact
input 8 FLM1: FID1 extinguished FLM2: FID2 extinguished RPT: Calibration repeatability error CAL: Calibration out of range POF: Power off NSD: Communication error
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
8 items in total
(GC1000 - GCCU)
CRL
F0507.EPS
F
<Toc> <Ind> <5. Actions of External Input and Output Signals >

5.4.2 GC6 T ype Input Data Format

(a) Stream change command (d) Start/stop command
1 23456
SC
a: Stream number
CRL
a
F
¥ Start command
1 234
CRL
BE
F
5-9
(b) Calibration command
1 2345
CRL
CA
a: Standard sample (STD1=1, STD2=2)
F
a
¥ Stop command
1 234
CRL
FI
F
(c) Range change command
1 2345678 9 10 11 12
RA
a b c
a: Stream number b: Component number c: Component list number
CRL
F
F0508.EPS
Non-protocol output signal transmission timing
Send Data (SD)
Analysis data
Analysis completion point
Calibration factor
+9V
+9V
(Transmission
completion)
During calibration
Error alarm
When an alarm occurs (after transmission completion if the component concentration value or calibration factor is being transmitted)
Request to Send (RS)
Terminal Equipment Ready (ER)
(Always ON)
Processor power ON
+9V
+9V
F0509.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >
I/O instruction format and communication control procedure for handshaking
5-10
I/O instruction format
E: Data receiving
request
F: Data transmission
completion
Output instruction Input instruction
T: Input command
prompt
B: Answerback
(command executed)
W: Answerback
(command not executed)
If the command is in an incorrect format, answerback is in the format shown at right.)
A: Data transmission
request
R: Data
retransmission request
CRL
#E
CRL
#F
CRL
#T
CRL
#B
C
#W
RLF
##W
CRL
#A
CRL
#R
Normal communication
F
F
F
F
F
F
Supervisory computer GCCU
Retransmission
Supervisory computer GCCU
Input command
Supervisory computer GCCU
Error handling
Supervisory computer GCCU
Communication control procedure
A AA
E
Output upon analysis completion or occurrence of an error alarm (after transmission completion if the component concentration value or calibration factor is being transmitted)
A RR
E Data 1
Note: If retransmission is requested, data can be transmitted any number of times. If the
total communication time exceeds one analysis cycle, the data is changed into new data. (The limiter should be considered for the supervisory computer side.)
Command
1
5 seconds
T B or W B or W
Output after completion of data transmission or every 8 to 10 sec periodically
Note: If the sequence has stopped, input commands can be received
continuously until the start signal is received.
(No response from the computer)
E E
3
seconds3seconds
Re-prompted if there is no command. If there is no command upon re-prompting, it is assumed that there is no command.
E
Data 1 Data 1
Command
2
E
Up to 6 commands including those not executed
FData 1 Data n
5 seconds
If there is no command during this period, it is assumed that there is no command.
3.2 seconds (for reference)
E E
(No response from GCCU)
The same command is repeated.
F0510.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >
5-11

5.4.3 GC8/GC1000 T ype Output Data Format (Fixed to 45 Characters)

(1)Analysis value data (concentation)
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
DS
a: Stream number (01 to 24) b: Component number (01 to 24) c: Concentration (without zero suppression) d: Unit e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL) f: Retention time (seconds) g: Retention time error alarm h: Analyzer number 1 to 9 (Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs)
,,
bacd e fgh
PPM A: CHL
%
,,,
A: CLL
T
A: RT
CR: Carriage return, LF: Line feed
(2)Analysis value data (component ratio operation, linear polynormial operation)
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
DS1
2,,
b eac
a: Stream number (01 to 24) b: Component number (01 to 24) c: Operation value (without zero suppression) d: Unit e:Analyzer number 1 to 9
PPM
%
d
Dummy space
C
L
R
F
F0511.EPS
CRL
F0512.EPS
F
(3)Calibration factor
1 234567 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
CC
a:
d:
,,
b eacd
Standard sample (STD1=1, STD2=2, STD3=3) Sensitivity error (if an error occurs)
COE
,
b:
Component number (01 to 24)
e:
Analyzer number 1 to 9
Dummy space
c:
Calibration factor (without zero suppression)
(4)Error alarm
• Memory pattern check error • Watchdog timer
• Calibration out of range • Detector 1 (or 2) calibration error
• Isothermal oven temperature error • Power off
• Pressure switch 1 (or 2) OFF • External contact inputs 1 to 8
• Communication error • FID1 (or 2) extinguished
• Calibration repeatability error
1 2345678 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44434241 45
AS /:
a: Month, b: Day, c: Hour, d: Minute e: Error alarm types f: Analyzer No.1 to 9
,,
bacde f
Note: For 3 characters, the 18th character is a dummy space.
MEM: Memory pattern check error WDT: Watchdog timer AD1: Detector 1 calibration error AD2: Detector 2 calibration error TMPH: Isothermal oven temperature error CAR1: Pressure switch 1 OFF CAR2: Pressure switch 2 OFF
Dummy space
EXT1: External contact input 1 EXT2: External contact input 2
EXT8: External contact input 8 FLM1: FID1 extinguished FLM2: FID2 extinguished RPT: Calibration repeatability error CAL: Calibration out of range POF: Power off NSD: Communication error
CRL
F0513.EPS
CRL
F
8 items in total
(GC1000 - GCCU)
F0514.EPS
F
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >

5.4.4 GC8/GC1000 T ype Input Data Format

5-12
(a) Stream change command
1 2345678
SM
aab
1 2345678910
SM
CRL
b
F
CRL
F
a: Stream number b: Analyzer number (1 to 6) or (001 to 255)
(c) Range change command
1 2345678 9 10 11 12
RM
a b c
1 2345678 9 10 11 12
RE
a b c
a: Stream number b: Component number c: Component list number d: Analyzer number (1 to 6) or (001 to 255)
(b) Calibration command
1 234567
CM
1 23456789
CE
CRL
a b
a b
F
CRL
F
a: Standard sample
(STD1=1, STD2=2, STD3=3)
b: Analyzer number (1 to 6) or (001 to 255)
13 14
CRL
F
d
15 1613 14
CRL
F
d
(d) Start/stop command
¥ Start command
1 2
345
CRL
BM
1 2 3 4 567
BE
F
a
a
a: Analyzer number (1 to 6) or (001 to 255)
CRL
F
¥ Stop command
1 245
3
CRL
FM
1 2 3 467
FE
F
a
5
CRL
a
a: Analyzer number (1 to 6) or (001 to 255)
F
F0515.EPS
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >
Non-protocol output signal transmission timing
Send Data (SD)
Analysis data
AUTO mode analysis
completion point
+9V (RS232C)
(Transmission completion)
5-13
I/O instruction format
E: Data receiving request
F: Data transmission
completion
Output instruction Input instruction
T: Input command prompt
B: Answerback
(command executed)
W: Answerback
(command not executed)
If the command is in an incorrect format, answerback is in the format shown at right.)
A: Data transmission
request
R: Data retransmission
request
Calibration factor
During AUTO mode calibration
Error alarm
When an alarm occurs (after transmission completion if the component concentration value or calibration factor is being transmitted)
+9V (RS232C)
(Transmission completion)
+9V (RS232C)
Request to Send (RS)
Terminal Equipment Ready (ER)
(Always ON)
Processor power ON
+9V (RS232C)
F0516.EPS
I/O instruction format and communication control procedure for handshaking
Communication control procedure
t (0.4 to 0.5 second)
A AA
Output upon analysis completion or occurrence of an error alarm (after transmission completion if the component concentration value or calibration factor is being transmitted)
t
A RR
total communication time exceeds one analysis cycle, the data is changed into new data. (The limiter should be considered for the supervisory computer side.)
t t
Command
1
5 seconds
Re-prompted if there is no command. If there is no command upon re-prompting, it is assumed that there is no command.
the start signal is received.
(No response from the computer)
E
3
seconds
seconds
E
3
t t
Data 1 Data 1
Command
2
Up to 6 commands including those not executed
CRL
#E
CRL
#F
CRL
#T
CRL
#B
C
#W
RLF
##W
CRL
#A
CRL
#R
Normal communication
F
F
F
F
F
F
Supervisory computer GCCU
Retransmission
Supervisory computer GCCU
Input command
Supervisory computer GCCU
Error handling
Supervisory computer GCCU
E
E Data 1
Note: If retransmission is requested, data can be transmitted any number of times. If the
T B or W B or W
Output after completion of data transmission or every 8 to 10 sec periodically
Note: If the sequence has stopped, input commands can be received continuously until
E E
t
FData 1 Data n
5 seconds
If there is no command during this period, it is assumed that there is no command.
3.2 seconds (for reference)
E E
(No response from GCCU)
The same command is repeated.
Note: If #ACRLF or #RCRLF is not received in 20 seconds or more after the data has been sent, the transmission of the
analysis data of that cycle will be aborted.
IM 11B03A03-01E 3rd Edition : Aug. 23. 2006-00
F0517.EPS
<T oc> <Ind> <5. Actions of External Input and Output Signals >

5.4.5 MODBUS Communication Data Specification

(1) Coil (Command Contact)
(1) Run command
Commands the start of continuous analysis to the GC1000 Mark II.
(2) Stop command
Commands the stop of continuous analysis to the GC1000 Mark II.
(3) Time setting request
Requests the GC1000 Mark II to set the device clock to values in addresses 40001-
40004. Before this request, time should be set by a holding register.
(4) Stream Sequence setting command
Commands the GC1000 Mark II to perform the specified stream sequence. This command is rejected if the measurement status is in Calibration/V alidation, including wait status, or in Stream (1 cycle), including wait status.
(5) Calibration/Validation command
Commands the GC1000 Mark II to perform the specified calibration or validation. Corresponding numbers to the Calibration/V alidation numbers to be specified are as follows:
5-14
Calibration 1 to 3: 1 to 3 V alidation 1 to 3: 4 to 6
This command is rejected if the measurement status is in Stream (1 cycle), including wait status, or the calibration method is set to Automatic/Manual.
(6) Stream (continuous) setting command
Commands the GC1000 Mark II to perform a continuous measurement of the speci­fied stream. This command is rejected if the measurement status is in Calibration/ V alidation, including wait status, or in Stream (1 cycle), including wait status.
Note 1: Coils are automatically reset from 1 to 0 when the slave accepts the message. Note 2: Pattern requests are queued when received but may not activate immediately. Note 3: Input relays should be monitored to determine when commands actually activate. Note 4: When the master sends multiple commands to the slave simultaneously, the slave executes it in the order received
regardless of the content of the commands.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >
(2) Input Relay (Status Contact)
(1) Analyzer normal
The GC1000 Mark II is normal. A 1 is displayed if there is no active level 1 (critical failure) or level 2 (minor failure) alarm.
(2) Analyzer error
The GC1000 Mark II is faulty . A 1 is displayed if there is at least one active level 1 (critical failure) alarm.
(3) Alarm status change
A 1 is displayed when a new alarm occurs on the GC1000 Mark II. After this status is read and when at least one alarm status is read, this bit is automatically reset to 0.
(4) Measuring (Run mode)
The GC1000 Mark II is analyzing. A 1 is displayed when it is in Run mode. Otherwise, 0 is displayed.
(5) Stop (Stop mode)
The GC1000 Mark II is not analyzing. A 1 is displayed when it is in Stop mode. Other­wise, 0 is displayed.
(6) Maintenance (other modes)
5-15
If the GC1000 Mark II is in Manual, Lab, or Pause mode, a 1 is displayed. Otherwise, 0 is displayed.
(7) Data update
New analysis data is now available from the GC1000 Mark II. Data update is auto­matically reset to 0 once data update is read and at least one analysis value is read.
(8) Data valid
Data is valid for a particular peak on the GC1000 Mark II. For analysis peaks, the analysis value must be between the upper and lower limits and the retention time is at or below the upper limit. For operation peaks, the data to be used for operation is valid. A 1 is displayed when data is valid.
(9) Alarm status
The alarm status of the GC1000 Mark II is displayed for each alarm number. If an alarm occurs, 1 is displayed. Otherwise, 0 is displayed. The alarm number is 1 to 99 for level 1 alarms of the GC1000 Mark II, and 101 to 199 for level 2 alarms of the GC1000 Mark II.
(10) Calibration factor update
The GC1000 Mark II has new calibration factors for a particular stream. The address is reset to 0 after the calibration factor update is read and when at least one calibration factor is read.
(1 1) Executing the Stream Sequence
The corresponding stream sequence on the GC1000 Mark II is being executed.
(12) Measurement status command failure
If the command is not executed after any of the Stream (continuous) setting com­mand, Calibration/V alidation command, Stream Sequence setting command has been sent by the coil, 1 is displayed. When the command is executed after the next com­mand has been sent, this bit is reset.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >
(3) Holding Register (Set Data)
(1) Time setting value
This is a set of four registers used by the DCS to set the device clock. When the time setting request coil (address 00003 for GC1000 Mark II and GCIU, 00001 for the analyzer server) is activated, these entries are used for the year, month/day, hour and minute/second. Since all devices on the highway synchronize to the analyzer server clock on a regular basis, the analyzer server clock should be regularly set.
Example : September 25, 1996, 15:23:10
5-16
Year
Month Day
Hour
Minute Second
Figure 5.4 Time Register Configuration
07CC (hexadecimal)
0919 (hexadecimal)
000F (hexadecimal)
170A (hexadecimal)
F0518.EPS
(2) Range setting command
This address allows the range of the component indicated by the specified stream number and peak number to be changed on the GC1000 Mark II.
(3) Analysis value
These addresses display the same data as the analysis value in the input resister, however values cannot be written to theses addresses. Only these addresses support real number (floating point) form.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Toc> <Ind> <5. Actions of External Input and Output Signals >
(4) Input Register (Measured Data)
(1) Stream number
The currently active stream number on the GC1000 Mark II is displayed. In Stop or Manual mode, the stream number is 0. In Lab mode, the stream number is 32.
(2) Starting peak number
The starting peak number assigned to each stream on the GC1000 Mark II is dis­played. The maximum number of analysis values, including peaks of all streams, is
255. The number is 0 if no peak is assigned.
(3) Peak number
Peak number assigned to each stream on the GC1000 Mark II is displayed. See the Figure 5.5 below.
Example: the peak number of Stream 2, Peak 3 is 7.
5-17
Starting peak number = 1 Starting peak number = 5
Starting peak number = 15
Starting peak number = 25
255
Figure 5.5 Example of Peak Allocation
Stream 1
Stream 2
Stream 3
Number of peaks = 4 Number of peaks = 10
Number of peaks = 10
F0519.EPS
(4) Sampling time
This register contains the latest sampling time for each stream on the GC1000 Mark II. Hour and minutes are stored.
Example : 15:23
Hour Minute
0F17 (hexadecimal)
F0520.EPS
(5) Analysis value
This register contains each analysis value. The value is represented by a fraction to the full scale or by a real number (floating point format). The full scale is set in ad­vance for each analysis value and the scaling factor is user selectable as either 9999 or 65535. The real number format conforms to the IEEE standard and requires two registers per peak.
The fraction format is calculated as follows: (Analysis value x Scaling factor)/Full scale value For example, if the analysis result is 5 ppm and the range is 0-20 ppm, the value read
using a scaling factor of 9999 is
5
x 9999 = 2499 (09C3)
20
F0521.EPS
For the real number format, units such as % and ppm are considered. For example, the real number value converted from 1.5 is directly transmitted as 1.5%. The GC1000 Mark II updates analysis values at the end of each cycle. The GCIU updates the average values according to the userset scan interval.
(6) Retention time
This register contains the retention time for each measured stream of the GC1000 Mark II. The unit is in seconds.
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<T oc> <Ind> <5. Actions of External Input and Output Signals >
(7) Calibration factor
This register contains the calibration factor for each calibrated stream of the GC1000 Mark II. Since the factors are in the range of 0.000 to 9.999, each value is multiplied by 1000 and displayed as integers, i.e., 0000 to 9999.
Tab le 5.5 Address Table (GC1000)
Name
Run command Stop command Time setting request Stream sequence select
command Calibration (validation)
command Stream (Cont) select
command Analyzer normal Analyzer error
Analyzer status change
Address
00001 00002 00003
0001P
0002P
001TT
10001
10002
10003
Recieved the message by the master, the slave reset. Same as above Same as above
Same as above P: Stream Sequence number (1 to 4)
Same as above P: 1 to 3 (Cal. 1 to 3), 4 to 6 (Val. 1 to 3)
TT: Stream number (01 to 31)
Reset when the alarm status is read after alarm status change is read.
Description
5-18
Measuring Stop Maintenance Stream
sequence Stream
(Cont) Calibration (Validation)
Stream sequence
Data update
Calibration factor update
Data valid Alarm status
Time setting request
Analysis value
Range select Stream number Starting peak number Peak number Sampling time
Analysis value
Retention time Calibration factor
10004 10005 10006
1001P
10021
10022
10023
101TT
102TT
11CCC 12AAA
40001 to 40004
41DDD
4NNQQ
30001 301TT 302TT
303TT
31CCC/-
31DDD
32CCC 33CCC
P: Stream Sequence number (1 to 4)
Reset when the analysis value is read after data update is read. (TT: Stream number)
Reset when the calibration factor is read after calibration factor update is read. (TT: Stream number)
CCC: Component number AAA: Alarm number (001 to 249) Year (40001), month/day (40002), hour (40003),
minute/second (40004) Read only (Write disabled)
DDD(real format)5CCC*2-1(CCC: Component number) NN: Stream number +40, QQ: Peak number
Starting peak number of each stream (TT: Stream number) Peak number of each stream (TT: Stream number) Hour/Minute (TT: Stream number) Analysis value (CCC: component number)
DDD(real format): CCC*2-1(CCC: Peak number) Seconds
Analysis value (CCC: component number)
T0505.EPS
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
<Int> <T oc>
Revision Record
Manual No. : IM 11B03A03-01E
Title : GC1000 Mark II Process Gas Chromatograph Overview
Oct. 2001/1st Edition
Newly published
July. 2002/2nd Edition Introduction modified below;
Added “(5) When touching LCD Panel switches” to “WARNING”; Corrected a part of “CENELEC (ATEX directive) certification”; Corrected a part of “Installation site and environment”; Added “Explanation f or type Y purging” to “WARNING of explosionpr oof enc losure” of FM
and CSA certification; Corrected title of Sec. 3.3 (2); Corrected title of Sec. 3.4 (2);
Aug. 2006/3rd Edition
Added the items for EPC and ethernet network (Page 2-2, 3-1, 3-2, 3-4 to 3-11, 4-8, 4-9)
i
IM 11B03A03-01E
3rd Edition : Aug. 23. 2006-00
Blank Page
Loading...