Yokogawa FieldMate User Manual

User’s Manual
R2.05
R1.01
AXF Verication Tool
IM 01R01A11-01EN
IM 01R01A11-01EN
1st Edition
AXF Verication Tool
IM 01R01A11-01EN 1st Edition

CONTENTS

1 INTRODUCTION ........................................................................................ 1-1
1.1 About This Manual ............................................................................................1-2
1.2 Safety and Modication Precautions .............................................................1-2
1.3 Trademarks ........................................................................................................1-3
1.4 Software License Agreement ..........................................................................1-3
2 OUTILINE ................................................................................................... 2-1
2.1 Conguration .................................................................................................... 2-2
2.2 Verication Item ................................................................................................2-3
2.3 Operation Procedure ........................................................................................2-4
Toc-1
3 PREPARATION .......................................................................................... 3-1
3.1 Package and Installation ..................................................................................3-2
3.2 AXF Write Protect Cancellation .......................................................................3-2
3.2.1 Before Verication ..............................................................................3-2
3.2.2 During Verication ..............................................................................3-3
4 START-UP FROM FieldMate .................................................................... 4-1
4.1 From “Segment Viewer”
(AXF is reading in “online” mode in FieldMate) ............................................4-2
4.2 From “Device Navigator” .................................................................................4-4
4.3 From “Verication Data” in “Device Maintenance Info” ............................... 4-6
4.4 From “History” ..................................................................................................4-8
5 LAUNCHING .............................................................................................. 5-1
5.1 Creating New Verication Data ........................................................................5-2
5.2 Loading Existing Verication Data .................................................................5-4
5.3 After Unexpected Termination.........................................................................5-5
6 OPERATION .............................................................................................. 6-1
6.1 Main Screen ....................................................................................................... 6-2
6.1.1 Load ...................................................................................................6-3
6.1.2 Save ...................................................................................................6-4
6.1.3 Print out ..............................................................................................6-8
6.1.4 Menu bar ............................................................................................6-9
1st Edition’ Aug. 2012 All Rights Reserved, Copyright © 2012, Yokogawa Electric Corporation
IM 01R01A11-01EN
Toc-2
6.2 Standard VF ..................................................................................................... 6-11
6.2.1 “Circuit” check and “Device Status” check
(Verication Tool is in “online” mode) ...............................................6-12
6.2.2 Physical Appearance .......................................................................6-21
6.2.3 Standard VF Result ..........................................................................6-22
6.3 Enhanced VF ...................................................................................................6-23
6.3.1 Current Output ................................................................................6-25
6.3.2 Pulse Output ....................................................................................6-28
6.3.3 Converter .........................................................................................6-35
6.3.4 Insulation Resistance .......................................................................6-38
6.4 Result ...............................................................................................................6-39
7 VARIATION OF VERIFICATION DATA ..................................................... 7-1
7.1 Installation Data ................................................................................................7-1
7.2 Locked Data .......................................................................................................7-2
7.3 Others .................................................................................................................7-2
8 TERMINATION ........................................................................................... 8-1
8.1 Normal Termination ..........................................................................................8-1
8.1.1 When the verication data has not been changed after saving:........8-1
8.1.2 When the verication data has been changed after saving:..............8-1
8.2 Unexpected Termination ..................................................................................8-2
9 COMMUNICATION ERROR ...................................................................... 9-1
9.1 Error Message ...................................................................................................9-1
9.2 Device Address Setting ....................................................................................9-3
9.2.1 Multidrop Mode Canceling ................................................................ 9-3
9.2.2 Multidrop Mode Setting ......................................................................9-3
Revision Information ..............................................................................................1
IM 01R01A11-01EN
<1 INTRODUCTION>

1 INTRODUCTION

This User’s Manual gives instructions on “AXF Verication Tool”.
This software is used to set up magnetic owmeter AXF; therefore, it is indispensable for users to read, understand and follow the instructions on all the following user’s manual before actually starting the operation.
Table 1.1.1 User’s Manuals for magnetic owmeter ADMAG AXF
Title IM No.
AXF Magnetic Flowmeter, Integral Flowmeter/ Remote Flowtube [Hardware Edition]
AXFA11G Magnetic Flowmeter, Remote Converter [Hardware Edition/Software Edition]
AXFA14G/C Magnetic Flowmeter Remote Converter [Hardware Edition/Software Edition]
AXF Magnetic Flowmeter Integral Flowmeter [Software Edition]
Table 1.1.2 User’s Manuals for external instruments
Title IM No.
Models AM012 Magnetic Flowmeter Calibrator FY2-XJAM012-40-2E
CA150 HANDY CAL IM CA150E
IM 01E20D01-01E
IM 01E20C01-01E
IM 01E20C02-01E
1-1
Refer to FieldMate User’s Manual when using AXF Verication Tool.
Table 1.1.3 User’s Manuals for FieldMate
Title IM No.
FieldMate Versatile Device Management Wizard
FieldMate Operational Precaution IM 01R01A01-91E
FieldMate Wizard Getting Started
IM 01R01A01-01E
IM 01R01A04-01E
IM 01R01A11-01EN
<1 INTRODUCTION>

1.1 About This Manual

• This manual should be provided to the end user.
• Before using the AXF Verication Tool, read this manual thoroughly to comprehend its con­tents.
• The contents of this manual may be changed without prior notice.
• All rights are reserved. No part of this manual may be reproduced in any form without Yokogawa’s written permission.
• Yokogawa makes no warranty of any kind with regard to this material, including, but not limited to, implied warranties of merchantability and suitability for a particular purpose.
• All reasonable effort has been made to ensure the accuracy of the contents of this manual. However, if any errors or omissions are found, please inform Yokogawa.
• Yokogawa assumes no responsibilities for this product except as stated in the warranty.
• Please note that this user’s manual may not be revised for any changes in specications, construction changes or operating part changes that are not considered to affect function or performance.
• If the customer or any third party is harmed by the use of this product, Yokogawa assumes no responsibility for any such harm owing to any defects in the product which were not pre­dictable, or for any indirect damages.
1-2
• This manual describes the operation of the AXF Verication Tool with FieldMate Advance. For detail installation and operation of FieldMate and additional functions, please refer to the FieldMate User’s Manual.
1.2 Safety and Modication Precautions
• The following general safety precautions must be observed during all phases of opera­tion, service, and repair of this instrument. Failure to comply with these precautions or with specic WARNINGS given elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Yokogawa assumes no liability for the customer’s failure to comply with these requirements. If this instrument is used in a manner not specied in this manual, the protection provided by this instrument may be impaired.
• Yokogawa will not be liable for malfunctions or damage resulting from any modication made to this instrument by the customer.
• The following safety symbols are used in this user’s manual and on the instrument.
WARNING
A WARNING sign denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury or death of personnel
CAUTION
A CAUTION sign denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product.
IM 01R01A11-01EN
<1 INTRODUCTION>
IMPORTANT
An IMPORTANT sign denotes that attention is required to avoid damage to the instrument or system failure.
NOTE
A NOTE sign denotes essential information for understanding operations and features.

1.3 Trademarks

All the brand names or product names of Yokogawa Electric used in this document are either trademarks or registered trademarks of Yokogawa Electric Corporation.
All the brand names or product names of other companies mentioned in this document are either trademarks or registered trademarks of their respective holders.
1-3

1.4 Software License Agreement

Refer to IM 01R01A01-01E Part A.
IM 01R01A11-01EN
<2 OUTILINE>

2 OUTILINE

AXF Verication Tool is a PC application software running in conjunction with FieldMate R2.05.00 or later.
Without having to remove the AXF magnetic owmeter from process line, its condition can be veried by checking several items and this tool provides a certicate that the unit is operating in accordance with YOKOGAWA standards. It is essentially an indication that the AXF is operating to its standards and is useful in supporting plant maintenance practices. It differs from actual-ow calibration at a manufacturing plant.
This tool stores verication data (*1) in a database (*2) in an organized manner, and can be used to print a Verication Report that has not only the individual verication items (*3) result but also the overall status of “passed” or “failed”. It is thus useful for device maintenance management.
*1: Verication data means the data saved by the user. *2: Database means a collection of verication data saved in Device Maintenance Information. *3: For the verication items, refer to 2.2.
2-1
IM 01R01A11-01EN
<2 OUTILINE>
2.1 Conguration
The following equipment is required to use Verication Tool ver. 1.01.00;
• YOKOGAWA magnetic owmeter AXF with HART Communication (AXF Output Signal and Communication sufx code: -E).
• PC (*1) and Printer
• Magnetic Flowmeter Calibrator AM012 (Only for Enhanced VF (*2), Optional)
• Handy Calibrator CA150 or equivalent (Only for Enhanced VF (*2), Optional)
• Insulation Checker (Only for Enhanced VF (*2), Optional)
*1: Regarding the PC environment, refer to IM 01R01A01-01E Part B,C. *2: VF stands for “Verication”.
- Circuit Check
- Device Status Check
- Physical Appearance Check
USB FieldMate Modem
2-2
Magnetic Flowmeter
AXF
MA012
Calibrator
(Optional)
- Converter Check
Figure 2.1.1 Congurations
Output Signal 4-20mA or pulse
CA150
Handy Calibrator
(Multi-meter)
(Optional)
- Current Output Check
- Pulse Output Check
Printer
PC
Insulation checker
(Optional)
- Insulation Resistance Check
Magnetic
Flowmeter
AXF
F020301.ai
IM 01R01A11-01EN
<2 OUTILINE>
2.2 Verication Item
This Verication Tool has 2 modes, which can be run separately.
1. Standard Verication (Standard VF)
• This mode veries the condition of the AXF without difculty by checking its internal param­eters. No external instruments are required.
• Some items are run only when FieldMate is in “online (*1)” mode.
2. Enhanced Verication (Enhanced VF) [Optional]
• This mode provides a more reliable verication by performing a detailed overview equivalent to SERVICE.
• This mode is optional. If only this mode is completed, “Overall Status” is left blank.
• This mode veries AXF by using external instruments, such as Magnetic Flowmeter Calibra­tor AM012, Handy Calibrator CA150, and so on.
• Even when FieldMate is in “ofine (*2)” mode, this mode can be run; in this case, some set­tings must be made manually.
There are only 2 possible combinations:
1. Standard VF
2-3
2. Standard VF + Enhanced VF
Table 2.2.1 Verication Items
Mode Verication Item Description
Standard VF Circuit (*3)
Magnetic Circuit Checks the Magnetic Circuit.
Excitation Circuit Checks the Excitation Circuit.
Calculation Circuit Checks the Calculation Circuit.
Device Status (*3)
Alarm Check Checks occurring Alarms.
Alarm History Checks historical Alarms.
Physical Appearance
Flowtube Checks the Flowtube appearance.
Converter Checks the Converter appearance.
Enhanced VF Current Output (*4) Checks the Current Output.
Pulse Output (*4) Checks the Pulse Output.
Converter (*4) Checks the Converter Accuracy.
Insulation Resistance
Coil
Signal (Electrode)
Checks the Excitation Coil. (Remote Flowtube Only)
Checks the Insulation Resistance. (Remote Flowtube Only)
*1: Online: The status in which the software communicates with a device. *2: Ofine: The status in which the software does not communicate with a device. The communication status between FieldMate and AXF can be conrmed on the FieldMate window. *3: These items are checked only when Verication Tool is in “online” mode. *4: When Verication Tool is in “ofine” mode, manual setting is necessary.
IM 01R01A11-01EN
<2 OUTILINE>

2.3 Operation Procedure

Figure 2.3.1 summarizes operation procedure.
To complete “Standard VF”, start-up Verication Tool from “Segment Viewer”. See 4.1.
1 to 3 PREPARATION
4 START-UP
2-4
Online Offline
IMPOSSIBLE
4.1 Segment Viewer 4.2 Device Navigator
No. New data Yes
5.3 General Information
Online
6.2 Standard VF (Mandatory)
FieldMate and AXF
5 LAUNCHING
Use the existing data?
6.1 Main Screen
Verification Tool and AXF
6.3 Enhanced VF
4.3 Verification Data
in “Device Maintenance Info”
Verification Tool and AXF: OfflineVerification Tool and AXF: Online
IMPOSSIBLE
(Optional)
4.4 History
5.2 Load
Offline
Verification
Operation
No Yes
Figure 2.3.1 Operation Procedure
6.4 Result
Save the verification data ?
6.1.3 Print out
8 TERMINATION
6.1.2 Save
F020301.ai
IM 01R01A11-01EN
<3 PREPARATION>

3 PREPARATION

3-1
WARNING
• Before running the Verication Tool, loop should be set to manual mode in the host system.
• The following applications may cause inadequate results:
- Signicantly-low ow rate
- Slurry liquid
- Stray current
• When “unexpected issue (*1)” occurs while running Verication Tool, follow “8.2 Unexpect­ed Termination” and “9 Communication Error”
*1: “Unexpected issue” means disconnection between AXF and PC:
• Physical disconnection between AXF and PC
• Forced shutdown of PC
• Unforeseen AXF power-off
IM 01R01A11-01EN
<3 PREPARATION>

3.1 Package and Installation

Regarding the packaging and installation procedure, refer to IM 01R01A01-01E Part B.

3.2 AXF Write Protect Cancellation

3.2.1 Before Verication
1) AXF parameters are changed during the verication process. Cancel the write protect switch.
Hardware switch off
Set the "Write Protect Switch" to "Enable" before this verification. Return the switch to the original position after this verification is complete.
3-2
Switch 1
Switch 2
Enable Protect
(A) AXFA11
2 Burnout setting switch
1 Write protect setting switch
F030201-A.ai
Hardware switch off
Set the "Write Protect Switch" to "Enable" before this verification. Return the switch to the original position after this verification is complete.
Switch 1
Switch 2
Enable Protect
(B) AXFA14 or AXF Integral Flowmeter
Figure 3.2.1 Write Protect Cancellation
2 Burnout setting switch
Write protect setting switch
1
F030201-B.ai
2) After the verication is complete, return the switch to the original position.
IM 01R01A11-01EN
<3 PREPARATION>
3.2.2 During Verication
1) If the AXF write protect function is set to “Protect” during the verication, Figure 3.2.2 will be displayed.
Figure 3.2.2 “Write Protect Cancellation” Dialog
3-3
1
2
3
4
F030202.ai
Table 3.2.1 “Write Protect Cancellation” Dialog
Item
No.
1 ‘If hardware….’
2 Enable Write
3 OK Runs the software protection cancellation.
4 Cancel Cancels ‘Write Protect cancellation’
Item Name Description
Shows the gure of hardware switch. See Figure 3.2.1. Click the text.
Cancels the software protection of the device. Enter the password.
2) Set the “Write Protect Switch” to “Enable” . Then, enter the password in (2) and click “OK (3)” in Figure 3.2.2.
3) Return the switch to the original position after this verication is complete.
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>

4 START-UP FROM FieldMate

Verication Tool can be started by the following:
■ From “Segment Viewer” (FieldMate is online to AXF), see 4.1.
■ From “Device Navigator”, see 4.2.
■ From “Verication Data” in “Device Maintenance Info”, see 4.3.
■ From “History”, see 4.4.
Select the start-up procedure depending on the items to be veried and connection status be-
tween FieldMate and AXF.
NOTE
• To run Verication Tool in “online” mode, start the Verication Tool from “Segment Viewer”.
• To complete Standard VF, start Verication Tool from “Segment Viewer”.
4-1
Online
IMPOSSIBLE
4.1 Segment Viewer
6.2 Standard VF (Mandatory)
Figure 4.1.1 “Start-up” Procedure
FieldMate and AXF
4.2 Device Navigator
IMPOSSIBLE
6.3 Enhanced VF (Optional)
6.4 Result
Offline
4.3 Verification Data
in “Device Maintenance Info”
Verification Tool and AXF: OfflineVerification Tool and AXF: Online
4.4 History
F040101.ai
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
4-2

4.1 From “Segment Viewer” (AXF is reading in “online” mode in FieldMate)

NOTE
Set the device address to 0. Verication Tool can be started only when the device address is set to 0. For details on how to set the device address, refer to 9.2.
From “Segment Viewer”, “Verication Tool” can be started by either method (A) or (B).
(A) By right-clicking
Right-click on the target (AXF) icon displayed at “Segment Viewer”, and select “Verication Tool”.
Figure 4.1.2 Start-up from “Segment Viewer” by Right-clicking
F040102.ai
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
(B) By selecting menu “Action”
Select the target (AXF) icon displayed at “Segment Viewer”, and in the “FieldMate” menu bar, select “Action - Verication Tool”.
4-3
F040103.ai
Figure 4.1.3 Start-up from “Segment Viewer” by Selecting the “Action” Menu
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>

4.2 From “Device Navigator”

NOTE
When Verication Tool is started from “Device Navigator”, it will be in “ofine” mode even if the AXF is in “online” mode in FieldMate.
From “Device Navigator”, “Verication Tool” can be started by either method (A) or (B).
(A) By right-clicking
Right-click on the target (AXF) icon displayed at “Device Navigator”, and select “Verication Tool”.
4-4
F040201.ai
Figure 4.2.1 Start-up from “Device Navigator” by Right-clicking
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
(B) By selecting menu “Action”
Select the target (AXF) icon displayed at “Device Navigator”, and in the “FieldMate” menu bar, select “Action - Verication Tool”.
4-5
F040202.ai
Figure 4.2.2 Start-up from “Device Navigator” by Selecting the “Action” Menu
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
4.3 From “Verication Data” in “Device Maintenance Info”
NOTE
When Verication Tool is started from “Device Maintenance Info”, it will be in “ofine” mode even if the AXF is in “online” mode in FieldMate.
From the “Verication Data” tab in “Device Maintenance Info”, “Verication Tool” can be started by either method (A) or (B). Select the target (AXF) icon and click the “Open Device Maintenance Info” icon. The “Device Maintenance Info” screen appeares.
4-6
Figure 4.3.1 “Device Maintenance Info” Icon
(A) By right-clicking
Select the “Verication Data” tab on the “Device Maintenance Info” screen. Right-click on the verication data listed on the “Verication Data” tab under “Device Maintenance Info, and select “View”. Verication Tool is started up and the selected verication data is loaded.
Figure 4.3.2 Start-up from “Verication Data” in “Device Maintenance Info” by Right-clicking
F040301.ai
F040302.ai
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
(B) By selecting menu “Action”
In the “Device Maintenance Info” menu bar, select “Action”-“View”. Verication Tool is started up and the selected verication data is loaded.
Figure 4.3.3 Start-up from “Verication Data” in “Device Maintenance Info” by Selecting the “Action”
Menu
4-7
F040303.ai
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>

4.4 From “History”

NOTE
When Verication Tool is started from “History”, it will be in “ofine” mode even if the AXF is in “online” mode in FieldMate.
From “History”, “Verication Tool” can be started by either method (A) or (B).
(A) By right-clicking
Right-click on the target device (AXF) listed in “History”, and select “Verication Tool”.
4-8
Figure 4.4.1 Start-up from “History” by Right-clicking
F040401.ai
IM 01R01A11-01EN
<4 START-UP FROM FieldMate>
(B) By selecting menu “Action”
Select the target (AXF) listed in “Device Maintenance Info”, and in the “FieldMate” menu bar, select “Action”-“Verication Tool”.
F040402.ai
Figure 4.4.2 Start-up from “History” by Selecting the “Action” Menu
4-9
IM 01R01A11-01EN
<5 LAUNCHING>

5 LAUNCHING

When Verication Tool is launched, Figure 5.1.1 is displayed automatically.
F050101.ai
Figure 5.1.1 Splash Screen
Select the verication data to be used for the verication:
1. Creating new verication data: see 5.1.
5-1
2. Loading existing verication data: see 5.2.
3. After unexpected termination: see 5.3
IM 01R01A11-01EN
<5 LAUNCHING>
5.1 Creating New Verication Data
1) After the “Splash Screen”,
1)-1 When there is no existing verication data in the database:
The splash screen is closed automatically, and then Figure 5.1.3 is displayed.
1)-2 When there is existing verication data in the database:
1)-2-1 The splash screen is closed automatically, and then Figure 5.1.2 is displayed.
1
2
F050102.ai
Figure 5.1.2 “Conrmation” Dialog
5-2
Table 5.1.1 “Conrmation” Dialog
Item
No.
1 Load
2 Cancel
Item Name Description
Loads the verication data from the database. See 5.2. To load the verication data, click “Load”.
Creates new verication data and shows Figure 5.1.3. To create new verication data, click “Cancel”.
1)-2-2 Click “Cancel (2)”.
If “Load (1)” is selected by accident, Figure 5.2.1 is displayed. Click “Cancel (3)” in
Figure 5.2.1.
IM 01R01A11-01EN
<5 LAUNCHING>
2) “General Information” is displayed. See Figure 5.1.3.
Enter each eld and click “Next (8)”. Figure 6.1.1 is displayed automatically.
1
2
3
4
5
6
7
5-3
8
F050103.ai
Figure 5.1.3 “General Information” Dialog
Table 5.1.2 “General Information” Dialog
Item
No.
1 User Name (*)
2 Location. (*) Enter the user location.
3 Veried by (*) Enter the name of the person to operate the verication.
4 Tag (Converter)
5 Tag (Flowtube) Enter the tag number of the owtube.
6 Serial No. (Converter) Enter the serial number of the converter.
7 Serial No. (Flowtube) Enter the serial number of the owtube.
8 Next
*: Mandatory eld
Item Name Description
Enter the user name. The user name of FieldMate is displayed by default. This eld can be edited.
Enter the tag number of the converter. The value of HART parameter “Tag” is displayed by default. This eld can be edited. Even if a different value is input from the HART parameter, the HART parameter is not overwritten.
Moves to “Main Screen”. See Chapter 6. Is enabled only after the mandatory elds have been lled in. Click “Next” to move to “Main Screen”.
IM 01R01A11-01EN
<5 LAUNCHING>
5.2 Loading Existing Verication Data
NOTE
In terms of keeping the integrity of the data, loaded verication data cannot be copied, nor can it be saved under a different name. Loaded verication data can only be overwritten by new saved data, except for “Locked data” which cannot be modied.
1) After the “Splash screen” is closed automatically, Figure 5.1.2 is displayed.
2) Click “Load (1)” in Figure 5.1.2. Figure 5.2.1 is displayed.
3) Select the preferred data from “Data list (1)” and click “OK (2)” in Figure 5.2.1.
5-4
1
2
3
F050201.ai
Figure 5.2.1 “Select Verication Data” Dialog
Table 5.2.1 “Select Verication Data” Dialog
Item
No.
1 Data list
2 OK
3 Cancel
2) After loading, Figure 5.2.2 is displayed automatically. Click “OK”. “Main Screen” is dis- played. See Chapter 6.
Item Name Description
Shows the verication data in the database. For the meaning of the icon in the “Data Status”, see Chapter 7.
Loads the verication data. To use the verication data in the database, select the preferred one among “Data list” and click “OK”. Figure 5.2.2 is displayed.
Creates new verication data and shows Figure 5.1.3. See 5.1. To create the new verication data, click “Cancel”.
F050202.ai
Figure 5.2.2 “Verication Data Loaded” Dialog
IM 01R01A11-01EN
<5 LAUNCHING>

5.3 After Unexpected Termination

In case of launching after an unexpected termination of the program, see 8.2.
5-5
IM 01R01A11-01EN
<6 OPERATION>

6 OPERATION

This chapter describes how to use Verication Tool.
For details of the verication procedure, see 6.2 and 6.3.
NOTE
• Be sure to run Standard VF.
• “Overall Status” in “Result” is carried out only when “Standard VF” is complete. When “Standard VF” is not complete, “Overall Status” is left blank.
• “Result sheet” can be printed even when “Standard VF” is not complete.
6-1
IM 01R01A11-01EN
<6 OPERATION>

6.1 Main Screen

This screen performs the following actions:
■ Loads the verication data. See 6.1.1.
■ Saves the verication data. See 6.1.2.
■ Prints the result sheet. See 6.1.3.
■ Displays the “Standard” VF screen. See 6.2.
■ Displays the“Enhanced” VF screen. See 6.3.
■ Displays the “Result” screen. See 6.4.
4
1 2 3
6-2
Figure 6.1.1 Main Screen
Table 6.1.1 Main Screen
Item
No.
1 Load
2 Save
3 Print
4 Menu bar Shows the “File” and “Help” menus. See 6.1.4.
5 Standard Selects Standard VF mode. See 6.2.
6 Enhanced Selects Enhanced VF mode. See 6.3.
7 Result Displays the result sheet image. See 6.4.
Item Name Description
Loads the verication data from the database. See 6.1.1. To load the verication data, click “Load”.
Saves the current data to the database. See 6.1.2. To save the current data, click “Save”.
Prints out the result. See 6.1.3. To print the verication result, click “Print”.
7
6
5
F060101.ai
IM 01R01A11-01EN
<6 OPERATION>

6.1.1 Load

1) Click the “Load” icon.
F060102.ai
Figure 6.1.2 “Load” Icon
2) After clicking the “Load” icon, Figure 6.1.3 is displayed. Select the preferred data from “Data list (1)”, then click “OK (2)”.
6-3
1
Figure 6.1.3 “Select Veri cation Data” Dialog
Table 6.1.2 “Select Veri cation Data” Dialog
Item
No.
1 Data list
2 OK
3 Cancel
Item Name Description
Shows the veri cation data in the database. Select the preferred data.
Performs data loading. To load the veri cation data from the database, click “OK”.
Cancels loading the veri cation data and returns to “Main Screen”. See
6.1.
4) After loading, Figure 6.1.4 is displayed automatically. Click “OK”.
2
3
F060103.ai
F060104.ai
Figure 6.1.4 “Veri cation Data Loaded” Dialog
IM 01R01A11-01EN
<6 OPERATION>

6.1.2 Save

By clicking the “Save” icon, the veri cation data is saved to the database.
After clicking “Save”, the screen changes as follows:
• When “No Flow + Flow (*1)” is not complete, see (A).
• When “No Flow + Flow” is complete, see (B).
*1: “No Flow + Flow”;
“Circuit” check and “Device Status” check for both “No Flow” and “Flow” conditions. Refer to 6.2.1.
F060105.ai
Figure 6.1.5 “Save” Icon
6-4
(A) When “No Flow + Flow” is not complete:
1) After clicking the “Save” icon, Figure 6.1.6 is displayed.
1 2 3
Figure 6.1.6 “Saving Data When “No Flow + Flow” Uncompleted” Dialog
Table 6.1.3 “Saving Data When “No Flow + Flow” Uncompleted” Dialog
Item
No.
1 Reason
2 Saved as Installation Data.
3 Saved as Locked Data.
4 OK Saves the data.
5 Cancel Cancels saving the data and returns to “Main Screen”. See 6.1.
Item Name Description
Shows the reason or comment. It can be used as a marker when loading veri cation data is selected. Refer to Figure 6.1.3. Enter preferred text. (Optional  eld)
Saves the data as “Installation Data”. See 7.1. This  eld is disabled until “No Flow + Flow” is complete.
Saves the data as “Locked Data”. See 7.2. To save the data as read-only, check the box. (Always valid)
4
5
F060106.ai
IM 01R01A11-01EN
<6 OPERATION>
2) Save the data as “Read-only” or “Editable”.
2)-1 To save the verication data as read-only: Click “OK” after checking the “Saved as Locked Data” box in Figure 6.1.6. Figure 6.1.7 is displayed. Click “Yes” in Figure 6.1.7.
1
2
F060107.ai
Figure 6.1.7 “Conrm Locked Data” Dialog
Table 6.1.4 “Conrm Locked Data” Dialog
6-5
Item
No.
1 Yes Saves verication data to the database as read-only.
2 No Cancels saving the data and returns to Figure 6.1.6.
Item Name Description
2)-2 To save the verication data as editable: Click “OK” without checking the “Saved as Locked Data” box in Figure 6.1.6. No dialogs appear.
3) Figure 6.1.8 is displayed automatically. Click “OK”.
F060108.ai
Figure 6.1.8 “Saving Completed” Dialog
IM 01R01A11-01EN
<6 OPERATION>
(B) When “No Flow + Flow” is complete:
1) After clicking the “Save” icon, Figure 6.1.9 is displayed.
1 2 3
4
5
F060109.ai
Figure 6.1.9 “Saving Data When “No Flow + Flow” Completed” Dialog
Table 6.1.5 “Saving Data When “No Flow + Flow” Completed” Dialog
6-6
Item
No.
1 Reason
Saved as Installation Data.
2
(*1)
Saved as Locked Data.
3
(*1)
4 OK Saves the data.
5 Cancel Cancels saving the data and shows “Main Screen”. See 6.1.
*1: These items can be selected at the same time.
Item Name Description
Shows the reason or comment. It can be used as a marker when the load­ing verication data is selected. Refer to Figure 6.1.3. Enter preferred text. (Optional eld)
Saves the data as installation data. See 7.1. To save the data as installation data, check the box.
Saves the data as locked data. See 7.2. To save the data as locked data, check the box.
2) Select the preferred one accordingly.
2)-1 To save the verication data as “Installation Data”: Click “OK” after checking the “Saved as Installation Data” box. If there is no existing installation data in the database, no dialogs appear. If there is existing installation data in the database, Figure 6.1.10 is displayed. Click “Yes” in Figure 6.1.10.
1
Figure 6.1.10 “Conrm Replacing Data” Dialog
2
F060110.ai
IM 01R01A11-01EN
<6 OPERATION>
Table 6.1.6 “Conrm Replacing Data” Dialog
6-7
Item
No.
1 Yes Replace the data.
2 No Cancels replacing the data and returns to Figure 6.1.9.
Item Name Description
2)-2 To save the verication data as read-only: Click “OK” after checking the “Saved as Locked Data” box in Figure 6.1.9. Figure 6.1.11 is displayed. Click “Yes” in Figure 6.1.11.
1
2
F060111.ai
Figure 6.1.11 “Conrm Locked Data” Dialog
Table 6.1.7 “Conrm Locked Data” Dialog
Item
No.
1 Yes Saves the data as locked data.
2 No Cancels saving the data and returns to Figure 6.1.9.
Item Name Description
2)-3 To save the verication data as editable and non-installation data: Click “OK” without checking any boxes in Figure 6.1.9. No dialogs appear.
3) Figure 6.1.12 is displayed automatically. Click “OK”.
F060112.ai
Figure 6.1.12 “Saving Completed” Dialog
IM 01R01A11-01EN
<6 OPERATION>

6.1.3 Print out

To print the result sheet, click the “Print” icon. Figure 6.1.14 is displayed.
F060113.ai
Figure 6.1.13 “Print” Icon
6-8
F060114.ai
Figure 6.1.14 Example of “Print” Dialog
The printed image is the same as the “Result” tab in “Main Screen”. See 6.4.
The recommended paper size is A4 or letter size.
• If the paper size is smaller than the recommended one, the printed size is reduced to  t the paper.
• If the paper size is bigger than the recommended one, the printed size will be the recom­mended one.
IM 01R01A11-01EN
<6 OPERATION>

6.1.4 Menu bar

There are “File” and “Help” in the menu bar.
(A) File
Select the preferred menu.
1
2
3
4
F060115.ai
Figure 6.1.15 Menu bar
Table 6.1.8 Menu bar
6-9
Item
No.
Load from
1
Database
2 Save to Database Same as “Save” in Table 6.1.1. See 6.1.2.
3 Print Same as “Print” in Table 6.1.1. See 6.1.3.
4 Exit Terminates the Verication Tool. See 8.1.
Item Name Description
Same as “Load” in Table 6.1.1. See 6.1.1.
IM 01R01A11-01EN
<6 OPERATION>
(B) Help
Upon selecting “About AXF Verication Tool…” in the “Help” menu, Figure 6.1.17 is displayed.
1
F060116.ai
Figure 6.1.16 Help Menu
Table 6.1.9 Help Menu
6-10
Item
No.
About AXF Verication
1
Tool…
Figure 6.1.17 Example of Version Information
Item Name Description
Shows the version of Verication Tool. See Figure 6.1.17.
F060117.ai
IM 01R01A11-01EN
<6 OPERATION>

6.2 Standard VF

■ This mode veries the condition of the AXF without difculty, by checking its internal param­eters. No external instruments are requied.
■ This mode is mandatory.
■ “Circuit” check and “Device Status” check are run only when Verication Tool is in “online” mode.
■ Each item can be run separately.
6-11
1
2
Figure 6.2.1 “Standard VF” Screen
Table 6.2.1 “Standard VF” Screen
Item
No.
1 Circuit Checks the status of device circuits. See 6.2.1.
2 Device Status Checks the alarm in AXF parameters. See 6.2.1.
3 Physical Appearance Checks the physical appearance. See 6.2.2.
Item Name Description
3
F060201.ai
IM 01R01A11-01EN
<6 OPERATION>
6.2.1 “Circuit” check and “Device Status” check (Verication Tool is in “online” mode)
CAUTION
• “Circuit” check and “Device Status” check temporarily change the parameters of AXF. In case of unexpected termination, follow 8.2.
• The AXF owtube should be completely lled with uid, otherwise the results may be inadequate.
• These items are run only when Verication Tool is in “online” mode.
• These items are veried for 2 types. Each condition check can be run separately.
(A) “No Flow” condition:
Veries the condition of the AXF by checking;
- Circuits including the excitation circuit and the magnetic circuit inside the owtube when the uid velocity is completely zero.
- The present and past history of alarms.
6-12
(B) “Flow” condition:
Veries the condition of the AXF by checking;
- Circuits including the excitation circuit and the operation circuit which calculates the ow velocity when the uid velocity is sufcient.
- The present and past history of alarms.
• There are 3 combinations:
(1) No Flow
(2) Flow
(3) Flow + No Flow
• The check items of the “Circuit” check depend on the case as follows:
• Nominal size of the unit.
• With or without the optional code ELC (DC Noise Cut Circuit).
• Test condition type; “Flow” or “No Flow”.
For details, see Table 6.2.2.
IM 01R01A11-01EN
<6 OPERATION>
1
6-13
2
3
4
5
6
7
Figure 6.2.2 “Circuit and Device Status” Screen
Table 6.2.2 “Circuit and Device Status” Screen
(1) Nominal size is 400mm or smaller, without optional code ELC
Item
No.
1
2 Test Condition
3
4
5
6
7
Item Name Description
No Flow Flow
Circuit
Device Status
Magnetic Circuit
Excitation Circuit
Calculation Circuit
Alarm Check
Alarm History
Shows test condition. Select the test condition
Shows operated test condition. “No Flow” or ” Flow” or ” No Flow + Flow” See Figure 6.2.13.
Shows the result of the Magnetic Circuit check. “Magnetic Circuit” means check­ing the magnetic circuit inside the owtube.
Shows the result of the Excitation Circuit check. “Excitation Circuit” means check­ing the excitation circuit by the excitation current measurement.
Shows the result of the Calculation Circuit check. “Calculation Circuit” means check­ing the operation circuit which calculates the ow velocity.
Shows the result of the Alarm Check. “Alarm Check” means checking the occurrence of alarms.
Shows the result of Alarm History Check. “Alarm History Check” means checking the history of alarms that occurred.
F060202.ai
Check is [possible/ not available]
No Flow Flow
Possible
Possible Possible Possible
Not Avail-
able
[Shows :N/A]
Possible Possible Possible
Possible Possible Possible
Not Avail-
able
[Shows: N/A]
Possible Possible
No Flow +
Flow
Possible
IM 01R01A11-01EN
<6 OPERATION>
(2) Nominal size is over 400mm, without optional code ELC (DC Noise Cut Circuit)
6-14
Item
No.
1
2 Test Condition Same as Table 6.2.2 (1) “No.2”
3
4
5
6
7
(3) With optional code ELC (DC Noise Cut Circuit)
Item
No.
1
2 Test Condition Same as Table 6.2.2 (1) “No.2”
3
4
5
6
7
Item Name Description
No Flow Flow
Magnetic Circuit
Circuit
Device Status
No Flow Flow
Circuit
Device Status
Excitation Circuit
Calculation Circuit
Alarm Check
Alarm History
Item Name Description
Magnetic Circuit
Excitation Circuit
Calculation Circuit
Alarm Check
Alarm History
Same as Table 6.2.2 (1) “No.1”
Same as Table 6.2.2 (1) “No.3” Possible
Same as Table 6.2.2 (1) “No.4” Possible Possible Possible
Same as Table 6.2.2 (1) “No.5”
Same as Table 6.2.2 (1) “No.6” Possible Possible Possible
Same as Table 6.2.2 (1) “No.7” Possible Possible Possible
Same as Table 6.2.2 (1) “No.1”
Same as Table 6.2.2 (1) “No.3”
Same as Table 6.2.2 (1) “No.4” Possible Possible Possible
Same as Table 6.2.2 (1) “No.5”
Same as Table 6.2.2 (1) “No.6” Possible Possible Possible
Same as Table 6.2.2 (1) “No.7” Possible Possible Possible
Check is [possible/ not available]
No Flow Flow
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
Check is [possible/ not available]
No Flow Flow
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
No Flow +
Flow
Possible
Not Avail-
able
[Shows :N/A]
No Flow +
Flow
Not Avail-
able
[Shows :N/A]
Not Avail-
able
[Shows :N/A]
IM 01R01A11-01EN
<6 OPERATION>
(A) No Flow (The uid velocity is completely zero)
1) Fill the owtube with uid completely. Be sure that the uid velocity is completely zero.
1
2
3
6-15
F060203.ai
Figure 6.2.3 “Flow/ No Flow” Screen
Table 6.2.3 “Flow/ No Flow” Screen
Item
No.
1 No Flow Selects the “No Flow” test condition.
2 Start Verication (*1)
3 Cancel Stops running “Circuit” check and “Device Status” check.
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed. Follow 3.2.2.
Item Name Description
Runs “Circuit” check and “Device Status” check. These checks are performed automatically. The Progress Bar is displayed at the bottom left.
2) Select “No Flow (1)” and click “Start Verication (2)”.
3) Before running “Circuit” check and “Device Status” check, Figure 6.2.4 is displayed.
IM 01R01A11-01EN
<6 OPERATION>
1
2
F060204.ai
Figure 6.2.4 “No Flow Conrmation” Dialog
Table 6.2.4 “No Flow Conrmation” Dialog
6-16
Item
No.
1 OK Runs “Circuit” check and “Device Status” check.
2 Cancel
Item Name Description
Stops running “Circuit” check and “Device Status” check, and displays “Main Screen”.
4) Click “OK (1)“ after conrming that the owtube is lled with uid and the uid velocity is completely zero. “Circuit” check and “Device Status” check are run. During these checks, the status bar and remaining time are displayed. See Figure 6.2.5. Regarding the execution time, see Table 6.2.5.
F060205.ai
Figure 6.2.5 “During ”Circuit” Check and “Device Status” Check” Screen
Table 6.2.5 Execution Time [No Flow]
No. Optional code Execution time
1 Without ELC 630 s
2 With ELC (DC Noise Cut Circuit) 120 s
IM 01R01A11-01EN
<6 OPERATION>
5) After the “Circuit” check and “Device Status” check are completed for the “No Flow” condi­tion, Figure 6.2.6 is displayed automatically. Click “OK”.
Figure 6.2.6 “No Flow Verication Completion” Dialog
6) “Result” is entered automatically. See Figure 6.2.13.
6-17
F060206.ai
IM 01R01A11-01EN
<6 OPERATION>
(B) Flow (The uid velocity is sufcient)
1) Fill the owtube with uid. Be sure that the uid velocity is sufcient.
1
2
3
6-18
F060207.ai
Figure 6.2.7 “Flow/ No Flow” Screen
Table 6.2.6 “Flow/ No Flow” Screen
Item
No.
1 Flow Selects the “Flow” test condition.
2 Start Verication (*1)
3 Cancel Stops running “Circuit” check and “Device Status” check.
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed. Follow 3.2.2.
Item Name Description
Runs “Circuit” check and “Device Status” check. Those checks are performed automatically. The progress bar is displayed at the bottom left.
2) Select “Flow (1)” and click “Start Verication (2)”.
3) Before running “Circuit” check and “Device Status” check, Figure 6.2.8 is displayed auto- matically.
IM 01R01A11-01EN
<6 OPERATION>
1
2
F060208.ai
Figure 6.2.8 “Flow Conrmation” Dialog
Table 6.2.7 “Flow Conrmation” Dialog
6-19
Item
No.
1 OK Runs “Circuit” check and “Device Status” check.
2 Cancel
Item Name Description
Stops running “Circuit” check and “Device Status” check, and displays “Main Screen”.
4) Click “OK (1)” after conrming that the owtube is lled with uid and the uid velocity is suf­cient. “Circuit” check and “Device Status” check are run. During these checks, the status bar and remaining time are displayed. See Figure 6.2.9. Regarding the execution time, see Table 6.2.8.
F060209.ai
Figure 6.2.9 “During ”Circuit” Check and “Device Status” Check” Screen
Table 6.2.8 Execution time [Flow]
No. Optional code Execution time
1 Without ELC 630 s
2 With ELC (DC Noise Cut Circuit) 120 s
IM 01R01A11-01EN
<6 OPERATION>
5) After the “Circuit” check and the “Device Status” check are completed for the “Flow” condi­tion, Figure 6.2.10 is displayed automatically. Click “OK”.
Figure 6.2.10 “Flow Verication Completed” Dialog
6) “Result” is entered automatically. See Figure 6.2.13.
6-20
F060210.ai
IM 01R01A11-01EN
<6 OPERATION>

6.2.2 Physical Appearance

1
2
3
F060211.ai
Figure 6.2.11 “Physical Appearance” Screen
Table 6.2.9 “Physical Appearance” Screen
6-21
Item
No.
1 Note Opens another window as shown in Figure 6.2.12. Click “Note”.
2 Flowtube Selects the result of the Flowtube Appearance Check.
3 Converter Selects the result of the Converter Appearance Check.
Item Name Description
Physical Appearance
A physical observation of the Flowmeter is required, check the following items:
Flowtube
Check the Flowtube and the Cables connected to the Flowtube for: * Deformation, Damage, Corrosion, Cracking, Signs of Abrasion and any Leaks.
Converter
Check the Converter and the Cables connected to the Converter for: * Corrosion, Damage, moisture in the housing, corroded terminals.
Observations are objective.
Figure 6.2.12 Example of Physical Appearance “Note”
1) Check the physical appearance of “Flowtube” and “Converter”. For details, see Figure 6.2.12.
F060212.ai
2) Select an applicable result manually in Table 6.2.10.
Table 6.2.10 Physical Appearance
Result Description
Passed Is selected when the result is good and there is no problem.
Failed Is selected when the result is bad.
N/A Is selected when the appearance check is not possible.
3) “Total Result” is displayed automatically. See Figure 6.2.13.
IM 01R01A11-01EN
<6 OPERATION>

6.2.3 Standard VF Result

6-22
Figure 6.2.13 Example of “Result of Standard VF”
F060213.ai
IM 01R01A11-01EN
<6 OPERATION>

6.3 Enhanced VF

• This mode provides more reliable verication by performed inspections equivalent to SERV­ICE.
• This mode is optional.
• This mode veries AXF by using external instruments, such as Magnetic Flowmeter Calibra­tor AM012, Handy Calibrator CA150, and so on.
• Even when FieldMate is in “ofine” mode, this mode can be run, but some manual setting will have to be performed.
NOTE
When Verication Tool is in “online” mode:
If Verication Tool turns to “ofine” mode due to wiring disconnections, reconnect FieldMate and AXF. Verication Tool will restore to “online” mode again.
6-23
Figure 6.3.1 “Enhanced VF” Screen
1
2
3
4
5
F060301.ai
IM 01R01A11-01EN
<6 OPERATION>
Table 6.3.1 “Enhanced VF” Screen
6-24
Item
No.
1 Current Output
2 Pulse Output
3 Output Selection
4 Converter
5 Insulation Resistance
Item Name Description
Checks the current output. See 6.3.1.
Checks the pulse output. See 6.3.2.
Outputs the current or pulse in accordance with the check items. See 6.3.1 or 6.3.2
Checks the accuracy of the converter by using AM012. See 6.3.3.
Checks the insulation resistance. See 6.3.4.
IM 01R01A11-01EN
<6 OPERATION>

6.3.1 Current Output

This item checks the current output from the converter.
1 2
F060302.ai
Figure 6.3.2 “Current Output” Screen
Table 6.3.2 “Current Output” Screen
6-25
3
4
Item
No.
1 Connection diagram Shows the connection diagram on another screen. See Figure 6.3.3.
2 Measured (mA) Enter the measured value manually.
Output (*1)
3
in Output Selection
Exit
4
in Output Selection
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed after clicking “Output”. Follow 3.2.2.
Item Name Description
Outputs the current. By clicking “Output” after selecting each setting value, the preferred cur­rent is transferred to AXF. (Only when Veri cation Tool is in “online” mode)
Stops the current output and enables other operations. (Only when Veri cation Tool is in “online” mode)
IM 01R01A11-01EN
<6 OPERATION>
Current Output
Use a Yokogawa CA150 Handy Calibrator or similar.
NOTE
Set the Digital MultiMeter (CA150 or similar) to measure the current in “mA”.
6-26
I+ I–
CUR OUT STATUS OUT ALARM OUT SIGNAL
POWER SUPPLY EXCITER
P–P+
PULSE OUT STATUS IN
SBBASACAL–AL+COMSO2+SO1+
COMSI2+SI1+EX2EX1L/+N/–
250 to 600
Digital
Multi Meter
F060303-A.ai
(A) AXFA11
Current Output
Use a Yokogawa CA150 Handy Calibrator or similar.
NOTE
Set the Digital MultiMeter (CA150 or similar) to measure the current in “mA”.
250 to 600
Digital
Multi Meter
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.3 Current Output Connection Diagram
F060303-B.ai
IM 01R01A11-01EN
<6 OPERATION>
1) Wire AXF and CA150 Handy Calibrator or similar as shown in Figure 6.3.3.
2) Set a current value to output.
2)-1 When Veri cation Tool is in “online” mode: Select 4 mA, 12 mA, or 20 mA in “Output Selection” and click “Output”.
F060304.ai
Figure 6.3.4 “Output Selection” in “Current Output”
2)-2 When Veri cation Tool is in “of ine” mode: Set the AXF HART parameter “Loop test” to the preferred value.
Table 6.3.3 Parameter Setting of “Current Output”
6-27
HART Parameter (DD)
Loop test Device Setup → Diag/Service → Output Test→ Loop test -
Parameter of
AXF Display
Setting Value
• 4 mA
• 20 mA
• Other (12 mA)
3) Measure the current output value by using CA150 Handy Calibrator or similar for each set­ting value.
4) Enter the measured current value in each “Mesured(mA)” manually. “Error(%)” is calculated automatically.
F060305.ai
Figure 6.3.5 “Current Output” Result
5) End processing
5)-1 When Veri cation Tool is in “online” mode: Click “Exit” in “Output Selection”. “Result” is displayed automatically after  lling in all  elds.
5)-2 When Veri cation Tool is in “of ine” mode: “Result” is displayed after  lling in all  elds.
IM 01R01A11-01EN
<6 OPERATION>

6.3.2 Pulse Output

This item checks the pulse output from the converter.
Check the Suf x Code of AXF converter or AXF integral  owmeter.
■ Without optional code EM, see “(A) Pulse Output (without optional code EM)”.
■ With optional code EM, see “(B) Active Pulse Output (with optional code EM)”.
1 2 3
Figure 6.3.6 “Pulse Output” Screen
Table 6.3.4 “Pulse Output” Screen
6-28
4
5
F060306.ai
Item
No.
1 Connection diagram
2 Setting Value Selects and shows the setting pulse value.
3 Measured (pps) Shows the measured value. Enter the measured value manually.
Output (*1)
4
in Output Selection
Exit
5
In Output Selection
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed after clicking “Output”. Follow 3.2.2.
Item Name Description
Shows the connection diagram on another screen. The diagram is any of those in Figure 6.3.9 or Figure 6.3.13 depending on the case.
Outputs the pulse. By clicking “Output” after selecting “Pulse”, the pulse with the value of “Setting value (2)” is transferred to AXF. (Only when Veri cation Tool is in “online” mode)
Stops pulse output and enables other operations. (Only when Veri cation Tool is in “online” mode)
IM 01R01A11-01EN
<6 OPERATION>
(A) Pulse Output (without optional code EM)
1) Wire AXF and CA150 Handy Calibrator or similar as shown in Figure 6.3.9.
2) Select “100 pps” in “Setting Value”. See Figure 6.3.7.
F060307.ai
Figure 6.3.7 “Setting Value” in “Pulse Output”
3) Output
3)-1 When Veri cation Tool is in “online” mode: Select “Pulse” in “Output Selection” and click “Output”. See Figure 6.3.8
6-29
F060308.ai
Figure 6.3.8 “Output Selection” in “Pulse Output”
3)-2 When Veri cation Tool is in “of ine” mode:
3)-2-1 Record the original AXF parameters, and then set them as shown below.
Table 6.3.5 Parameter Settings of “Pulse Output”
HART Parameter
Pulse Scale Device Setup → Easy Setup → Pulse Scale B33/E11 100
Pulse Width Device Setup → Pulse Set → Pulse Width E12 50% Duty
Pulse Unit Device Setup → Easy Setup → Pulse Scale B32/E10 Pulse/s
Test DO (*1) Device Setup → Diag/Service → Output Test→ Test DO N30 Pulse
DO Function(*1)
*1: Only for AXFA14 or AXF Integral Flowmeter
Device Setup → Con guration → Status Function → DO Function
Parameter of
AXF Display
F20 Pulse Output
Setting Value
3)-2-2 Output pulse by setting AXF HART parameter “Loop test” to “20 mA”.
Table 6.3.6 Parameter Setting of “Pulse Output”
HART Parameter (DD)
Loop test Device Setup → Diag/Service → Output Test→ Loop test - 20 mA
Parameter of
AXF Display
Setting Value
IM 01R01A11-01EN
<6 OPERATION>
NOTE
■ Measure for more than 60 seconds.
■ If using a YOKOGAWA CA150, set the mode to “Contact Input” and the range to “CPM”.
4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or equivalent to simplify the measurement of pulses. If such a device is unavailable, an appro­priate load (30V DC, 0.2A max) is required and use the con guration shown in Figure 6.3.9.
30VDC , 0.2A. max
6-30
CUR OUT STATUS OUT ALARM OUT SIGNAL
POWER SUPPLY EXCITER
(A) AXFA11
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
P–P+
PULSE OUT STATUS IN
30VDC , 0.2A. max
SBBASACAL–AL+COMSO2+SO1+I+ I–
COMSI2+SI1+EX2EX1L/+N/–
External
power supply
Load
Electronic Counter
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
F060309-A.ai
Electronic Counter
Load
External
power supply
F060309-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.9 Pulse Output Diagram (without optional code EM)
5) Enter the measured value manually.
6) “Result” is displayed automatically.
Figure 6.3.10 “Pulse Output” Result
F060310.ai
IM 01R01A11-01EN
<6 OPERATION>
7) End processing
7)-1 When Verication Tool is in “online” mode: Click “Exit” in “Output Selection”.
7)-2 When Verication Tool is in “ofine” mode: Set the AXF parameters back to their original values manually.
6-31
IM 01R01A11-01EN
<6 OPERATION>
(B) Active Pulse Output (with optional code EM)
1) Wire AXF and CA150 Handy Calibrator or similar as shown in Figure 6.3.13.
2) Select “2 pps” in “Setting Value”. See Figure 6.3.11.
F060311.ai
Figure 6.3.11 “Setting Value” in “Active Pulse Output”
3) Output
3)-1 When Veri cation Tool is in “online” mode: Select “Pulse” in “Output Selection” and click “Output”. See Figure 6.3.12.
6-32
F060312.ai
Figure 6.3.12 “Output Selection” in “Active Pulse Output”
3)-2 When Veri cation Tool is in “of ine” mode:
3)-2-1 Record the original AXF parameters, and then set them as shown below.
Table 6.3.7 Parameter Settings of “Active Pulse Output”
HART Parameter (DD)
Pulse Scale Device Setup → Easy Setup → Pulse Scale B33/E11 2
Pulse Width Device Setup → Pulse Set → Pulse Width E12 100ms
Pulse Unit Device Setup → Easy Setup → Pulse Scale B32/E10 Pulse/s
Test DO (*1) Device Setup → Diag/Service → Output Test→ Test DO N30 Pulse
DO Function(*1)
(*1) only for AXFA14 or AXF Integral Flowmeter
Device Setup → Con guration → Status Function → DO Function
Parameter of
AXF Display
F20 Pulse Output
Setting Value
3)-2-2 Output pulse by setting AXF HART parameter “Loop test” to “20 mA”.
Table 6.3.8 Parameter Setting of “Active Pulse Output”
HART Parameter (DD)
Loop test Device Setup → Diag/Service → Output Test→ Loop test - 20 mA
Parameter of
AXF Display
Setting Value
IM 01R01A11-01EN
<6 OPERATION>
NOTE
• Measure for more than 60 seconds.
• If using a YOKOGAWA CA150, set the mode to “Frequency (FREQ) and Pulse” and the range to “CPM”.
• An appropriate load (24 VDC ± 20%, 150 mA max) is required.
4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or equivalent to simplify the measurement of pulses. If such a device is unavailable, use the con guration shown in Figure 6.3.13.
Output voltage: 24 V DC ±20% Current: 150 mA max
CUR OUT STATUS OUT ALARM OUT SIGNAL
POWER SUPPLY EXCITER
P–P+
PULSE OUT STATUS IN
SBBASACAL–AL+COMSO2+SO1+I+ I–
COMSI2+SI1+EX2EX1L/+N/–
6-33
Load
Electronic Counter
Dashed box : Yokogawa CA150 Handy Calibrator or similar
(A) AXFA11
Load
Dashed box : Yokogawa CA150 Handy Calibrator or similar
Output voltage: 24 V DC ±20% Current: 150 mA max
Electronic Counter
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.13 Active Pulse Output Diagram (with optional code EM)
5) Enter the measured value manually.
6) “Result” is displayed automatically.
F060313-A.ai
F060313-B.ai
Figure 6.3.14 “Active Pulse Output” Result
F060314.ai
IM 01R01A11-01EN
<6 OPERATION>
7) End processing
7)-1 When Verication Tool is in “online” mode: Click “Exit” in “Output Selection”.
7)-2 When Verication Tool is in “ofine” mode: Set the AXF parameters back to their original values manually.
6-34
IM 01R01A11-01EN
<6 OPERATION>

6.3.3 Converter

6-35
WARNING
Follow User’s Manual FY2-XJAM012-40-2E for correct operation and safe handling.
This item checks the accuracy of the AXF converter with Magnetic Flowmeter Calibrator AM012.
1 2 3
4
5
F060315.ai
Figure 6.3.15 “Converter” Screen
Table 6.3.9 Converter
Item
No.
1 Connection diagram
2 Span
3 Read Span
4 Measured (m/s)
5 Read
Item Name Description
Shows the connection diagram on another screen. The diagram is (A) or (B) in Figure 6.3.16 depending on the case.
Shows the Span of the converter. This eld can be edited.
Reads the Span of the converter. Press “Read Span”. (Only when Verication Tool is online to AXF)
Shows the measured value. This eld can be edited.
Reads the measured value. Press “Read”. (Only when Verication Tool is online to AXF)
AXFA11
(White)
F060316-A.ai
SB
BASAC
(Black)
COMSI2+SI1+
AM012
AM012
CUR OUT STATUS OUT ALARM OUT SIGNAL
POWER SUPPLY EXCITER
(Shield) (White) (Black)
AL–AL+COMSO2+SO1+I+ I–
(Shield)
EX2EX1
L/+N/–
PULSE OUT STATUS IN
P–P+
(A) AXFA11 (B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.16 “Converter” Connection Diagram
Junction Box Terminal
(Shield)
(Shield)
(White)
(Black)
(Black)
N/
L/
+
+
SA
(White)
+
SB
+
F060316-B.ai
IM 01R01A11-01EN
<6 OPERATION>
1) For details of the calibration procedures, follow User’s Manual FY2-XJAM012-40-2E.
Record the following original parameters and set the AXF parameters as shown below
manually. (These settings are described in FY2-XJAM012-40-2E.)
Table 6.3.10 AXF parameter setting of Converter Check
6-36
HART Parameter (DD)
Select Flow Tube (*1)
Nominal Size
Nominal Size
MF Set
Low MF
High MF
Low MF (EDF) (*2)
High MF (EDF) (*2)
*1: Only for AXFA11 *2: Only for optional code HF1 or HF2 (Enhanced Dual Frequency Excitation) *3: Only for optional code HF1 or HF2, select “Enhanced DF”.
Device setup → Detailed Setup → Basic Setup →Se­lect Flow Tube
Device setup → Detailed Setup → Basic Setup →Nominal Size
Device setup → Detailed Setup → Basic Setup →Nominal Size unit
Device setup → Detailed Setup → Basic Setup→MF Set
Device setup → Detailed Setup → Basic Setup→MF Set → Standard DF → Low MF
Device setup → Detailed Setup → Basic Setup→MF Set → Standard DF → High MF
Device setup → Detailed Setup → Basic Setup→MF Set → Enhanced DF → Low MF (EDF)
Device setup → Detailed Setup → Basic Setup→MF Set → Enhanced DF → High MF (EDF)
2) Enter the span on the “Converter” screen.
Parameter of
AXF Display
C30 Calibrator
C32
C31 mm
C20
(Measure Mode)
C21 1.0
C22 1.0
C23 1.0
C24 1.0
nominal size in
Enhanced DF (*3)
Setting Value
Enter owtube
“mm” unit
Standard DF
or
2)-1 When Verication Tool is in “online” mode: Click “Read Span (3)” in Figure 6.3.15. The value of parameter “Velocity Check” is entered automatically. If the AXF parameters are different from those in Table 6.3.10, the following error message is displayed. Set the AXF parameters to those in Table 6.3.10.
F060317.ai
Figure 6.3.17 Example of “Parameter Error” Dialog
IM 01R01A11-01EN
<6 OPERATION>
2)-2 When Verication Tool is in “ofine” mode: Read the value of parameter “Velocity Check” manually. Refer to Table 6.3.11. And enter the value in “cell (2)” in Figure 6.3.15 manually.
Table 6.3.11 AXF Parameter Settings of Converter Check
6-37
HART Parameter (DD)
Velocity Check Device setup → Detailed Setup → Basic Setup →Velocity Check C44
Parameter of
AXF Display
3) Measure the output accuracy at the preferred setting (0%, 25%, 50%, 75%, or 100%). After setting the value, let the AM012 stabilize at least for 1 minute.
For operation procedure, follow User’s Manual FY2-XJAM012-40-2E.
4) Enter the measured value.
4)-1 When Verication Tool is in “online” mode: Click “Read (5)” in Figure 6.3.15. The value of instantaneous ow rate is entered automati- cally.
4)-2 When Verication Tool is in “ofine” mode: Read the value of parameter “PV” or “PV %rnge” manually. Refer to Table 6.3.12, and convert the value in “m/s” unit.
Enter the calculated value in “cell (4)” in Figure 6.3.15 manually.
Table 6.3.12 AXF Measured Value of “Converter” Check
Parameter of
AXF Display
A20
A10
PV
PV %rnge
HART Parameter (DD)
■ Process Variables → PV or
■ Device Status → Process Variable → PV
■ Process Variables → PV %rnge or
■ Device Status → Process Variable → PV %rnge
5) “Error (%)” is calculated automatically.
6) “Result” is displayed automatically after lling in all elds.
Figure 6.3.18 “Converter” Result
7) Set the AXF parameters back to their original values manually.
F060318.ai
IM 01R01A11-01EN
<6 OPERATION>
F060319-A.ai

6.3.4 Insulation Resistance

6-38
WARNING
Follow User’s Manual IM 01E20D01-01E for correct operational and safe handling.
Where the insulation resistance check is not possible, select “N/A”
1) Check the insulation resistance in the terminal box in accordance with the tables below. Fol­low User’s Manual IM 01E20D01-01E.
(A) Coil
Table 6.3.13 Coil Insulation Resistance Check
Test Terminals Test Voltage
Between terminals EX1 and C 500 V DC (Use an insulation tester or equivalent.)
(B) Signal (Electrode)
Table 6.3.14 Electrode Insulation Resistance Check
Test Terminals Test Voltage
- Between terminals A and C
- Between terminals B and C
500 V DC (Use an insulation tester or equivalent.)
2) Enter the measured value manually. Where the insulation resistance check is not possible, select “N/A”. Where the measured value is over-range, select “Open Loop (O.L.)”.
3) “Result” is displayed automatically.
(A) Example 1
(B) Example 2
Figure 6.3.19 “Insulation Resistance” Result
F060319-B.ai
IM 01R01A11-01EN
<6 OPERATION>

6.4 Result

The results of Standard VF and Enhanced VF are available in the “Result” tab. Uncompleted check items are displayed as blank.
6-39
Figure 6.4.1 “Result” Screen
F060401.ai
IM 01R01A11-01EN
<6 OPERATION>
The result in full view is shown Figure 6.4.2.
If the user needs a hard copy of the report, refer to 6.1.3.
6-40
1
Figure 6.4.2 “Result” in Full View
2
3
4
5
6
F060402.ai
IM 01R01A11-01EN
<6 OPERATION>
Table 6.4.1 Result
6-41
Item
No.
Tag (Converter) Tag (Flowtube)
1
Serial No. (Converter) Serial No. (Flowtube)
2 Overall Status
3 Comments
4 Veried by
5 Signature
6 Legend symbols
Item Name Description
Shows AXF device information. The contents entered in General Informa­tion are displayed by default. This elds can be edited in the “Result” tab.
Shows the result of the whole verication.
• Passed This owmeter was veried to be functioning within +/- □% of the original factory calibration.
• Failed This owmeter did not pass the verication test.
Shows user comment. This item can be edited. Enter any comment (e.g. user’s document No.).
Shows the name of the person who performed the verication. The name entered in General Information is displayed by default. This eld can be edited in the “Result” tab.
Shows the signature of the person who performed the verication. The person should sign this eld on the printout.
Shows the meaning of symbols in “Result” elds.
● [----] : Inconclusive
● [ ] : Not performed
IM 01R01A11-01EN
<7 VARIATION OF VERIFICATION DATA>
7-1

7 VARIATION OF VERIFICATION DATA

In terms of keeping its own data integrity, veri cation data cannot be copied, nor can it be saved as other name. Veri cation data except for “Locked data” can be saved only by newly saving or over-writing.
There are 3 types of veri cation data:
1. Installation data
2. Locked data
3. Others
1
2
Figure 7.1.1 Variation of veri cation data
Table 7.1.1 Variation of veri cation data
Item
No.
1 Installation data
2 Locked data
3 Others See 7.3.
Item Name Icon Description

7.1 Installation Data

“Installation data” functions as a reference. Each device can have only one “Installation data”. When the veri cation data is saved by checking the “Save as installation data” box in Figure
6.1.9, the data becomes the “Installation data”.
Only when “No Flow + Flow (*1)” is complete, veri cation data can be saved as an “Installation data”. See 6.1.2.
3
F070101.ai
See 7.1.
See 7.1 and 7.2.
Save the  rst veri cation data after the delivery as “Installation data”. (Recommended)
When the user wants to edit “Installation data” later:
Save the veri cation data by checking only the “Saved as installation data” box in Figure
6.1.9. The icon in Data Status is “
When it is necessary to save installation data as read-only:
Save the veri cation data by checking the “Saved as installation data” box and the “Save as locked data box” in Figure 6.1.9. The icon in Data Status is “
*1: “No Flow + Flow”:
“Circuit” check and “Device Status” check for both “No Flow” and “Flow” conditions. Refer to 6.2.1.
“.
”.
IM 01R01A11-01EN
<7 VARIATION OF VERIFICATION DATA>

7.2 Locked Data

“Locked data” is saved as read-only. The icon in Data Status is “ ”.
After verication data is saved as “Locked data”, the data:
• can be loaded.
• can be printed out.
• cannot be over-written.
When verication is complete, save the verication data as “Locked data” to prevent unexpected over-writing or falsifying, before printing. (Recommended)

7.3 Others

Verication data which is neither “Installation data” nor “Locked data” can be edited. When veri­cation is to be continued later, do not save as either “Installation data” or “Locked data”.
7-2
IM 01R01A11-01EN
<8 TERMINATION>
8-1

8 TERMINATION

This chapter describes “Normal Termination” and “Unexpected Termination”.

8.1 Normal Termination

Select “File”-“Exit” or click the “Close” button at the upper-right corner of “Main Screen”.
F080101.ai
Figure 8.1.1 Normal Termination
8.1.1 When the verication data has not been changed after saving:
This tool is terminated without any dialogs appearing.
8.1.2 When the verication data has been changed after saving:
1) Figure 8.1.2 is displayed.
1
2
F080102.ai
Figure 8.1.2 “Conrmation” Dialog
Table 8.1.1 “Conrmation” Dialog
Item
No.
1 Yes Saves the verication data by over-writing. See 6.1.2.
2 No Cancels saving the data.
Item Name Description
2) Select whether the latest verication data is saved to the database or not.
2)-1 To save the latest verication data: Select “Yes (1)”. See 6.1.2.
2)-2 When the latest verication data is not saved: Select “No (2)”. No dialogs appear and the data is not saved.
3) This Verication Tool is terminated automatically.
IM 01R01A11-01EN
<8 TERMINATION>

8.2 Unexpected Termination

If an unexpected termination occurs, follow this procedure.
1) When an unexpected termination occurs, be sure to re-connect FieldMate and AXF, and reboot Verication Tool. If the AXF parameters have been changed from the original setting, Figure 8.2.1 “Error Recovery” Dialog is displayed after reboot.
1
2
8-2
F080201.ai
Figure 8.2.1 “Error Recovery” Dialog
Table 8.2.1 “Error Recovery” Dialog
Item
No.
1 Error contents
2 OK
Item Name Description
Shows “Last verication error date” and “Parameter items which will be recovered”.
Resets the AXF internal parameters to their original values. Click “OK (2)”.
2) To reset the AXF parameters to their original values, click “OK (2)”. Figure 8.2.2 is displayed and then the AXF parameters are recovered.
Figure 8.2.2 “Recovering” Screen
F080202.ai
IM 01R01A11-01EN
<9 COMMUNICATION ERROR>

9 COMMUNICATION ERROR

9.1 Error Message

This chapter shows error dialogs. See Table 9.1.1 for details.
After termination, follow 8.2.
F090101.ai
Figure 9.1.1 Error Dialog 1
9-1
Figure 9.1.2 Error Dialog 2
Figure 9.1.3 Error Dialog 3
Figure 9.1.4 Error Dialog 4
F090102.ai
F090103.ai
F090104.ai
IM 01R01A11-01EN
<9 COMMUNICATION ERROR>
F090105.ai
Figure 9.1.5 Error Dialog 5
Table 9.1.1 Communication Error Dialog
9-2
Figure
No.
9.1.1
9.1.2
9.1.3
9.1.4
9.1.5
Message When Description
Setting Error. Set the device address to 0, and then restart this applica­tion.
Error occurred during verica­tion. Application has terminated. Check connections and then restart this application. Press OK to continue.
Device parameters have been changed. Check connections and then restart this application. Press OK to continue.
Connection Error. Check connections. Press OK to continue.
Connection Error. Check connections and then restart this application. Press OK to continue.
■ When starting Verica­tion Tool from “Segment Viewer” with status where FieldMate is online to AXF whose device address is not “0”.
■ During running “Circuit” check or “Device Sta­tus” check.
■ During setting pulse output.
■ During running “Cur­rent Output” or “Pulse Output”.
■ During preparing “Cir­cuit” check or “Device Status” check after click­ing “Start Verication” in Standard VF.
Upon clicking;
■ “Output” in Output Se­lection
■ “Exit” in Output Selec­tion
■ “Read Span” in Con­verter
■ “Yes” in Figure 8.2.1.
■ During launching “Veri­cation Tool”
■ During recovery of AXF parameters after unexpected termination (Figure 8.2.2)
■ If the user wants to run Veri­cation Tool with status where Veri­cation Tool is in “online” mode: Follow 9.2.
■ If Verication Tool need not be in “online” mode: Restart Verication Tool from “Device Navigator”, “Device Info”, or “History”.
After this dialog, Figure 8.1.2 is dis- played and the tool is terminated. As AXF parameters might have changed, follow 8.2.
After this dialog, Figure 8.1.2 is dis- played and the tool is terminated. As AXF parameters might have changed, follow 8.2.
After this dialog, the tool returns to the previous step.
After this dialog, the tool is termi­nated. As AXF parameters might have changed, follow 8.2.
IM 01R01A11-01EN
<9COMMUNICATIONERROR>
9.2 DeviceAddressSetting
9-3
WARNING
This operation must be carried out by the trained personnel having knowledge of multidrop loop. For the details of the parameter setting procedure, refer to IM 01E20C01-01E or IM 01E20C02­01E. After using Verication Tool, set the AXF to the original multidrop loop.
9.2.1 MultidropModeCanceling
USB FieldMate Modem
Device A
Device Address: 1
Device B
Device Address: 2
Device C
Device Address: 3
Device X
Device Address: 15
-
+
-
+
-
+
-
+
Load: 250 Ω
Device Address: 1
F090201-A.ai
Device A
Set to “0” .
-
+
Load: 250 Ω
USB FieldMate Modem
Multidrop Loop side
Load: 250 Ω
F090201-B.ai
(A)Multidroploop (B)AftercancelingMultidroploop
Figure9.2.1Conguration
1) Disconnect the target device (AXF) from the multidrop loop physically.
2) Connect the AXF to FieldMate one to one. A load of 250Ω is required. See Figure9.2.1(B).
3) Record the following original AXF parameter, and then set the parameter as shown.
Table9.2.1ParameterofCancellingMultidropmode
HARTParameter(DTM) SettingValue
Poll Addr Conguration → HART → Poll addr 0
4) Start Verication Tool.
9.2.2 MultidropModeSetting
After using Verication Tool, set the AXF back to its original status.
1) After Verication Tool termination, set the AXF parameter back to its original value manually.
Table9.2.2ParameterSettingofMultidropmode
HARTParameter(DTM) SettingValue
Poll Addr Conguration → HART → Poll addr Original Value
2) Connect the AXF back to the original conguration.
IM 01R01A11-01EN

Revision Information

AXF Verication Tool
Manual No.: IM 01R01A11-01E
Aug. 2012 / 1st Edition
IM 01R01A11-01EN
Loading...