Thank you for purchasing the Welch Allyn Model 679 SureTemp thermometer. The operating and maintenance instructions found in this manual should be followed to ensure years of accurate and reliable service. Please read these instructions thoroughly before using your 679 SureTemp® thermometer.
Technical service manuals may be ordered by contacting Welch Allyn Customer Service at (800) 854-2904.
The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
European Regulatory Manager, Medical Division Welch Allyn Ltd., Navan, Co. Meath
Republic of Ireland
®
© 1997 by W elch Allyn, Inc. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, without prior consent in writing from Welch Allyn. Printed in the U.S.A.
U.S. patent 5,632,555. Additional U.S. and foreign patents pending.
CONTENTS
INTRODUCTION .................................................................. 2
SAFETY & WARNINGS ......................................................... 2
P ARTS ................................................................................ 3
MOUNTING INSTRUCTIONS................................................. 4
SETUP ............................................................................... 4
CHANGING PROBES ..................................................................5
FEATURES.......................................................................... 5
DIS PLA Y IN DICAT ORS.......................................................... 6
OPERATION ........................................................................ 8
T AKING ORAL TEMPERATURES ..................................................... 8
USING MONITOR MODE ................................................................9
T AKING AXILLARY TEMPERA TURES (Under 4 years)................. 10
T AKING AXILLARY TEMPERA TURES (4 years - Adult) .................11
T AKING RECTAL TEMPERATURES ...............................................12
MAINTENANCE ................................................................. 12
CLEANING AND STERILIZA TION ......................................... 13
SPECIFICATIONS ............................................................... 13
CUST OMER SERVICE ......................................................... 13
LIMITED WARRANTY ............................................. Bac k C ove r
1

INTRODUCTION

The Welch Allyn 679 SureTemp is a portable thermistor thermometer to be used for measuring patient temperatures ranging from 84°F to 108°F (28.9°C to 42.2°C). Oral/axillary and rectal probes utilize single-use disposable probe covers which limit cross-contamination. Oral, axillary and rectal temperatures can be taken using Normal or Monitor operating modes. Axillary temperatures taken using the Normal operating mode will display temperatures in about 10 seconds for patients 3 years and under. Axillary temperatures for patients 4 years through adult are taken using the Monitor operating mode and will display temperatures in about 5 minutes. In the Normal mode, the thermometers microprocessor predicts body temperature in about 4 seconds for oral temperatures and in about 15 seconds for rectal temperatures. The Monitor mode continu­ously displays the temperature of the probe which will reflect the patients actual temperature in about 3 minutes and will continue for as long as the probe remains in place.

SAFETY & WARNINGS

DO NOT USE IN THE PRESENCE OF FLAMMABLE
ANESTHETICS.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
In order to obtain accurate and reliable temperature-taking results and ensure patient safety, it is important that this booklet be read thoroughly prior to using the instrument. If you have any technical or clinical questions concern­ing the thermometers use and/or care, please contact our Customer Service Department at (800) 854-2904.
Single-use disposable probe covers, available from Welch Allyn, will limit patient cross-contamination. The use of any other probe cover or the failure to use a probe cover may produce temperature errors and will invalidate the instruments warranty.
Oral probes (blue ejection button) are to be used for taking oral and axillary temperatures only. Rectal probes (red ejection button) are to be used for taking rectal temperatures only. The use of the wrong probe will produce temperature errors.
The thermometer case is not waterproof; do not drip fluids onto it.
Do not use this instrument for any purpose other than that specified in
this booklet. Doing so will invalidate the instruments warranty.
2
Do not use the thermometer if you notice any signs of damage to the probe or instrument. Contact our Customer Service Department for immediate assistance.
Do not ETO sterilize without using proper procedures described in the 679 SureTemp technical manual.
Figure 1
1. Probe cover box 5. Probe handle
2. Display 6. Probe storage channel
3. Mode button 7. Latching probe connector
4. Probe cover ejection button 8. Battery door