Welch Allyn SureSight Service Manual Rev. C

Eye Care Menu
Service Manual
French Version
Italian Version
Spanish Version
4341 State Street Road
Skaneateles Falls, NY 13153-0220
140400 Rev. B

Table of Contents

Warnings & Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Mises en garde et symboles . . . . . . . . . . . . . . . . . . . . . . . . 4
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Function and Intended Application . . . . . . . . . . . . . . . . . . 10
SureSight Display, Controls, and Features . . . . . . . . . . . . 11
First Time Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Plus or Minus Cylinder Convention . . . . . . . . . . . . . . . . . . . . . 14
Diopter Power Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Important Operating Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Helpful hints for testing pre-schoolers . . . . . . . . . . . . . . . . . . . 18
Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adult/Child Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Adult/Child Mode Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Replacement and Recycling of Battery . . . . . . . . . . . . . . . 20
Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Recycling Lithium Ion Battery . . . . . . . . . . . . . . . . . . . . . . . . . 21
Screening Mode and Software Updates . . . . . . . . . . . . . . 22
Screening Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Updating SureSight Software . . . . . . . . . . . . . . . . . . . . . . . . . 22
Hand Strap Replacement . . . . . . . . . . . . . . . . . . . . . . . . . 23
Cleaning and Sterilization of the SureSight . . . . . . . . . . . . 23
Service, Calibration, and Warranty Information . . . . . . . . . 24
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Service and Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
SureSight Rechargeable Battery . . . . . . . . . . . . . . . . . . . . . . . 25
Calibration and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Warnings & Symbols
ATTENTION: Refer to the operating instructions. This symbol ( ) is intended to advise the user of the presence of important operating or maintenance instructions in the documents accompanying the instrument.
Service or Repair to be performed by qualified, authorized personnel only. There are no user serviceable parts inside the instrument. Opening this device can expose the user to harmful invisible laser radiation.
Use only with Welch Allyn 710 series chargers.
Replacement parts and accessories - Use only approved replacement parts and accessories specified in this manual. Refer to the maintenance section of this manual for more information.
Do not sterilize the instrument or any of its components.
Use only with IEC 60601-1 approved printers, or keep printer out of patient vicinity.
Disassembly of instrument beyond the extent described in this manual will void the warranty. Refer all servicing to Welch Allyn or a Welch Allyn authorized service representative listed in the Servicing, Calibration and Warranty Information section. Opening this device can expose the user to harmful invisible laser radiation. Only the battery cover is to be opened by the user.
Battery replacement: Replace with Welch Allyn model # 72420 Lithium Ion battery only.
1
Do not attempt to disassemble or modify the battery pack. There are no user serviceable parts inside pack.
Do not attempt to directly solder to the battery pack.
Do not attempt to connect the positive and negative battery terminals to each other, nor to any other device.
“Caution” - use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
Do not operate, charge, leave or discard battery pack in any environment where it may exceed 0°C to 40°C.
Do not immerse the battery pack in water.
Do not attempt to open or pierce the battery pack.
Do not strike or throw the battery pack.
Do not use a battery pack which appears to be deformed, leaking, corroded, or is other wise irregular.
If electrolyte leaking from battery contacts your skin or eyes, rinse with running water and immediately seek medical attention.
IPXØ: Not protected against the ingress of water.
Not for use in the presence of flammable anesthetics.
2
Type BF Equipment.
Double Insulated.
ETL listed: UL2601-1, CSA C22.2 No. 601.1, IEC 60601-1, AS 3200.1. IEC 60601-1-2
Australia EMC Framework Compliance
The CE mark on this device indicates that it has been tested and conforms to the provisions noted within the 93/42/EEC Medical Device Directive.
3

Mises en garde et symboles

ATTENTION : Consulter le mode d'emploi. Ce symbole ( ) signale à l’utilisateur la présence de renseignements importants relatifs au fonctionnement ou à l'entretien de l'instrument dans les documents qui l'accompagnent.
L'entretien ou la réparation ne doivent être effectués que par un personnel compétent habilité. Cet instrument ne contient pas de pièces réparables par l'utilisateur. L'ouverture de ce dispositif risque d'exposer l'utilisateur à un rayonnement laser invisible nocif.
Utiliser uniquement avec des chargeurs Welch Allyn série 710.
Pièces et accessoires de rechange – Utiliser uniquement les pièces et accessoires de rechange autorisés spécifiés dans ce manuel. Consulter la section Entretien de ce manuel pour des informations supplémentaires.
Ne pas stériliser l'instrument ni aucun de ses composants.
Utiliser uniquement avec une imprimante conforme à la norme CEI 60601-1, ou utiliser l'imprimante à distance du patient.
Un démontage de l'instrument au-delà des recommandations de ce manuel annule la garantie. Confier toutes réparations à Welch Allyn ou à un représentant agréé par Welch Allyn listé dans la section Informations sur les réparations, l'étalonnage et la garantie. L'ouverture de ce dispositif risque d'exposer l'utilisateur à un rayonnement laser invisible nocif. Seul le couvercle de la batterie peut être ouvert par l'utilisateur.
4
Remplacement de la batterie : Remplacer uniquement par une
o
batterie au lithium Welch Allyn modèle n
72420.
Ne pas essayer de démonter ou de modifier le bloc-batterie. Il ne contient pas de pièces réparables par l'utilisateur.
Ne pas essayer de souder directement sur le bloc-batterie.
Ne pas essayer de relier les bornes positive et négative l'une à l'autre, ou à d'autres dispositifs.
Attention : Des procédures de contrôle, de réglage ou de performance autres que celles qui sont spécifiées dans cette notice peuvent entraîner une exposition dangereuse au rayonnement.
Ne pas activer, charger, laisser ou jeter le bloc-batterie à un endroit où
°
la température ambiante est inférieure à 0° C ou supérieure à 40
C.
Ne pas immerger le bloc-batterie dans l'eau.
Ne pas essayer d'ouvrir ou de percer le bloc-batterie.
Ne pas frapper ni lancer le bloc-batterie.
Ne pas utiliser un bloc-batterie qui semble fuir ou être déformé, corrodé ou anormal.
5
En cas de contact cutané ou oculaire avec une fuite d'électrolytes de la batterie, rincer à l'eau courante et consulter immédiatement un médecin.
IPXØ : Non protégé contre l’infiltration d'eau.
Ne pas utiliser en présence d'anesthésiques inflammables.
Equipement de type BF.
Double isolation.
o
Homologations ETL : UL 2601-1, CSA C22.2, N
601.1 et IEC 60601-
1,
AS 3200.1, I EC 60601-1- 2
Obedece la norma australiana EMC Framework
La marque CE appliquée sur ce dispositif indique qu'il a été testé et déclaré conforme aux dispositions de la directive 93/42 de la CEE sur les appareils médicaux.
6
+ 21 hidden pages