Welch Allyn Stress system Additional Protective Earth User Manual

Connecting the grounding kit to the PC
1
Stress system Additional Protective Earth (APE) grounding kit installation
102992
Material Number 714632, 80015559 Ver. B
4341 State Street Road Skaneateles Falls, NY 13153 USA www.welchallyn.com © 2013, Welch Allyn, Inc.
CAUTION If you have previously installed the resting system grounding kit (101920), you MUST replace that kit with the contents of this kit before installing any additional hardware. Failure to do so can create a fire hazard due to the wire in the old kit being a smaller gauge than the rest of the components.
1
2
3
4
1
3
2
1
3
Schiller
2
Welch Allyn Technical Support
www.welchallyn.com/about/company/locations.htm
3
Connecting the grounding kit to the treadmill
Schiller
2
CAUTION The grounding kit wire must not run under the treadmill.
1 3
Schiller
2
ler
Schil
3
Connecting the grounding kit to protective earth NOTE You must confirm that the grounding method you select is permitted by national and local codes.
3
Continuous water pipe method
This method must be a continuous copper line back to earth.
Electrical panel method
This method must be a conduit tied directly and securely to the ground electrical panel, which must be tied to the ground buss within the panel.