Connecting the grounding kit to the PC
1
Stress system Additional
Protective Earth (APE)
grounding kit installation
102992
Material Number 714632, 80015559 Ver. B
4341 State Street Road
Skaneateles Falls, NY 13153 USA
www.welchallyn.com
© 2013, Welch Allyn, Inc.
CAUTION If you have previously
installed the resting system grounding kit
(101920), you MUST replace that kit with
the contents of this kit before installing
any additional hardware. Failure to do so
can create a fire hazard due to the wire
in the old kit being a smaller gauge than
the rest of the components.
1
2
3
4
1
3
2
1
3
Schiller
2
Welch Allyn Technical Support
www.welchallyn.com/about/company/locations.htm
3
Connecting the grounding kit to the treadmill
2
CAUTION The grounding kit wire must not run under the treadmill.
1 3
Schiller
2
ler
Schil
3
Connecting the grounding kit to protective earth NOTE You must confirm that the grounding method you select is permitted by national and local codes.
3
Continuous water pipe method
This method must be a continuous
copper line back to earth.
Electrical panel method
This method must be a conduit tied
directly and securely to the ground
electrical panel, which must be tied
to the ground buss within the panel.
1 3