Welch Allyn Standard laryngoscope handles User Manual

©2014 Welch Allyn, Inc. MM 721357 Ver. C
1
Intended
About this
use
The laryngoscope blades which are used to examine placement of a tracheal
document
This directions for use apply to Welch Allyn reusable standard laryngoscope handles REF: 60200, 60300, 60400, 60305. Welch Allyn used MacIntosh
Warnings and
WARNING: Welch Allyn reusable standard laryngoscope handles must be reprocessed after each use.
WARNING: The reprocessing procedure and the equipment and materials described must be followed and conducted by persons trained and familiar reprocessing.
WARNING: Consult cleaning and disinfecting agent manufacturer proper preparation and use.
WARNING: Repeated inspection procedures to assure damage has not occurred to the handle.
WARNING: High level disinfection and/or sterilization are not achieved by these methods.
WARNING: Discard any component that shows evidence of damage or deterioration.
WARNING: Do not modify this equipment. Any modification of this lead to patient injury. Any modification of this equipment voids the
WARNING: Personnel shall appropriate personal protective equipment when handling equipment.
WARNING: CAUTION:
CAUTION: Do not immerse/soak handle, damage to handle may occur. CAUTION: If the device will be unused for several months or longer, remove
batteries prior to storing the device.
standard laryngoscope handle is an accessory used w i t h c o m p at i b l e r i g i d st a n d a r d
and visualize
tube.
reusable standard
with
Welch Allyn standard laryngoscope blades MacIntosh
REF:
6924X, and Miller REF: 6804X and REF 68470.
a patient’s airway
laryngoscope handles may be
REF:
and
aid
6904X, English
cautions
with
medical device
reprocessing may
follow their facility policies
Laryngoscope equipment is not suitable for use in intense
Failure
to follow these instructions may cause damage to this han d l e .
degrade
elements of the handle.
and
procedures
potentially
instructions
product warranty.
and
contaminated
magnetic
for their
Follow
equipment may
wear
the
fields
2
Directions for use
Welch Allyn standard laryngoscope handles
Reprocessing
instructions
These reprocessing instructions refer to procedures for cleaning and intermediate level disinfection. between each use using
• Cleaning
Standard
laryngoscope handles must be reprocessed prior to first use and
the
and intermediate level
following method as outlined in this
disinfection
Welch Allyn has validated the above instruction as being capable of preparin g these laryngoscope handles for re-use. The user must ensure that the reprocessing as actually performed by the user's personnel, with the user's equipment and materials, achieves the desired result. This may require validation and routine monitoring of the user's actual proce ss
NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are compatible with the autoclave methods identified which are provided for facilities who wish to autoclave after cleaning and intermediate level disinfection.
Cleaning and intermediate level disinfection instructions
Point of use: 1.
Preparation
for
decontamination:
Separate
blade assembly from handle and place handle into suitable containment for subsequent reprocessing per figure 1. Do not place handle with sharp devi ce s.
2. Prevent the handle from drying (i.e. wrap/cover in moist germicidal
wipe).
1. Select an appropriate quaternary ammonium isopropanol based germicidal cleaner labeled suitable for use on healthcare equipment and capable intermediate level disinfection. Reference EPA­registered disinfectants: http://www.epa.gov/oppad001/chemregindex.htm Outside of the U.S., please consult applicable regulatory body for equivalent quaternary ammonium isopropanol germicidal cleaner.
2. Remove batteries per figure 3.
document:
Directions for use
Cleaning
and intermediate level disinfection:
Drying:
Maintenance, Inspection and Testing
Storage: Store handle per facility practice to allow device to remain
End of
reprocessing
Welch Allyn standard laryngoscope handles
3
1. Follow the ge r m i c i da l w i p e manufacturer’s instructions to clean all exposed surfaces of the handle and end cap.
2. If necessary, brush with a dry, soft-bristled brush and re-wipe to loosen/remove excessive visible soil.
3. After all visible soil is removed, re-wipe to wet all surfaces and allow adequate contact time for disinfection as directed by the germicidal wipe manufacturer.
CAUTION: Only use quaternary ammonium
isopropanol based germicidal wipes.
Allow components to air dry.
1. Inspect
each
component
area (per f
igure
3)
for
damage
or
deterioration.
WARNING: Discard any component that
shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble batteries into handle per figure 4 with new or batteries in known good condition.
3 . Attach handle to a clean and disinfected test blade
assembly in known working condition. Verify
Blade assembly engages and locks onto handle.
Blade assembly deploys into its locked position on
handle
Light output is satisfactory.
If the la mp fails to replace the
AND lamp illuminates.
light or output is low
batteries.
that:
, check or
clean, dry, and ready for service.
instructions for intermediate level disinfection.
4
Directions for use
Autoclave
Welch Allyn standard laryngoscope handles
instructions
NOTE: The main handle and bottom cap components of handles marked “AUTOCLAVE” are compatible with the autoclave methods identified which are provided for facilities who wish to autoclave after cleaning and intermediate level disinfection.
Disassembly: Remove batteries per figure 2 and set aside.
After battery removal, select ONE of the following autoclave methods below main handle and bottom cap (only):
Gravity autoclave: Follow equipment manufacturer and set-up and operation of autoclave e quipment. Gravity autoclave settings are as follows:
Temperature: 132 C (270 F)
Exposure time: 3 minutes (unwrapped)
Minimum dry time: 1
minute
Pre-vacuum autoclave: Follow equipment manufacturer and set-up and operation of autoclave e quipment. follows:
Temperature: 132 C (270 F)
Exposure time: 3 minutes (unwrapped)
Minimum dry time: 1
minute
Pre-vacuum
each
Maintenance, Inspection and Testing
1. Inspect deterioration.
component
WARNING: Discard any component that
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble b a t t e r i e s i n t o h andle per figure 4.
3. Attach handle to a clean and disinfected test blade assembly in known working condition. Verify
Blade assembly engages and locks onto handle.
Blade assembly deploys into its locked position on
handle AND lamp
Light output is satisfactory.
If the la mp fails to replace the
illuminates.
batteries
light or output is low
Storage: Store handle per facility practice to allow device to remain
clean, dry, and ready for service.
facility
procedures in the
facility
procedures in the
autoclave settings are as
area per f
igure 3 for
damage
shows
, check or
for
that:
the
or
Directions for use
Welch Allyn standard laryngoscope handles
5
Maintenance
Replace
the
batteries
1. Unscrew bottom cap of handle per figure 5 and remove batteries.
2. Alkaline batteries are supplied with your handle for maximum performance and
3. Insert batteries and reinstall bottom cap per figure 5.
4. Reprocess repaired assembly as appropriate per these instructions
instructions
are recommended as replacements; however used.
Large handle, REF 60200 uses two “D” size
Medium handle, REF 60300 uses two “C” size
Penlight handle, REF 60400 uses two “AA” size
Stubby handle, REF 60305 uses two “AA” size
carbon-zinc
batteries may also be
Figure 5
6
Directions for use
Specifications
Electrical: For information about electromagnetic compatibility (EMC) see Welch Allyn website:
http://www.welchallyn.com
Operating:
32
(0 C)
Approvals: Con fo rm s t o A S TM F 965 and
Warranty:
One year Service Information:
For Welch Allyn
Storage/Transport:
104 F (40 C)
F
-4 F (-20 C)
120 F (49 C)
ISO-7376-1, IEC/EN
The
CE
mark on this product indicates that it has been tested to and
with the provisions noted within the Complies with EMC Framework of Australia
Technical
Support or to obtain information about any Welch Allyn product,
Technical
Support: w
ww.welchallyn.com/support.
93/42/EEC
Welch Allyn standard laryngoscope handles
60601-1,
Medical Device
IEC/EN
60601-1-2
conforms
Directive.
contact
Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com
Authorized European Repre sentative Address: Regulatory Affairs Representative Welch Allyn, Limited Navan Business Park Dublin Road Navan, County Meath, Republic
of Ireland
Loading...