Welch Allyn Standard laryngoscope blade assemblies User Manual

Page 1
©2014 Welch Allyn, Inc. MM 721358 Ver. C
1
Standard
Directions for use
Intended
About this
Warnings and
use
The
rigid laryngoscope is a device used to examine
placement of a tracheal
document
This directions for use applies to W elch Allyn reusable standard l aryn goscope blade assemblies MacIntosh 6804X and REF 68470. Welch Allyn used
with
Welch Allyn standard laryngoscope handles 60200, 60300, 60400, 60305, and 60710.
cautions
WARNING: R each use.
WARNING: The reprocessing procedure and the equipment and materials described must be followed and cond ucted by persons trained and familiar reprocessing.
WARNING: Consult cleaning and disinfecting agent manufacturer proper preparation and use.
WARNING: Repeated procedures to assure damage has not occurred to the blade assembly.
WARNING: Only trained personnel shall use a laryngoscope for WARNING: Do not reprocess laryngoscope blade assemblies used on
Creutzfeldt-Jakob disease. These instruments should be discarded. WARNING: Discard any component that shows evidence of damage or deterioration. WARNING: Do not modify this equipment. Any modification of this
patient injury. Any modification of this equipment voids the WARNING: Personnel
personal protective equipment when handling WARNING:
WARNING: Lamps if left illuminated, could generate sufficient heat to cause burns.
CAUTION: Only use lamps specified. Failure or poor performance of this device.
eusable
Laryngoscope equipment is not suitable for use in intense
and visualize
tube.
REF:
6904X, English MacIntosh
reusable sta n d a r d
fiber optic
reprocessing may
shall follow their facility policies and procedures and
laryngoscope blade
degrade
potentially
to follow these instructions may cause damage
REF:
laryngoscope blade assemblies may be
assemblies mu s t be reprocessed
elements of the device.
a patient’s airway
6924X, and Miller REF:
with
medical device
instructions
intubation.
patients
equipment
product warranty.
contaminated
equipment.
magnetic
for their
Follow
wear
and
aid
after
inspection
with
can lead to
appropriate
fields.
CAUTION: CAUTION:
disinfectants. CAUTION: CAUTION:
CAUTION:
Only use neutral pH Only use neutral pH
Do not use other cleaners or Do not remove lamp during reprocessing. Damage to device could occur.
REF 68470 is not compatible with any sterilization process.
(6.0-8.0) (6.0-8.0)
enzymatic type cleaners. ortho-phthalahyde
disinfectants.
(OPA)
or glutaraldehyde
Page 2
s
2
Directions for use
Welch Allyn standard laryngoscope
blade assemblie
Reprocessing
instructions
Reprocessing refers to procedures for cleaning and disinfection or sterilization of devices. The standard laryngoscope blade assemblies must be reprocessed prior to first use and between each use using any of
• Cleaning
• Cleaning
• Cleaning
• Cleaning
• Cleaning
Welch Allyn has validated these instructions as being capable of preparing its laryngoscope blade for re-use. The user must ensure that the reprocessing as actually performed by the user's personnel, with the user's equipment and materials, achieves the desired result. This may require validation and routine monitoring of the user's actual process.
Cleaning instructions
CAUTION:
Point of use: 1.
Preparation decontamination:
Figure 1
the
following methods outlined in this
and cold solution high level
disinfection
document:
and gravity autoclave sterilization and pre-vacuum autoclave sterilization and Steris® Amsco® V-PRO® low temperature sterilization and
STERRAD® hydrogen peroxide gas plasma sterilization
Do not remove lamp during reprocessing. Damage to device could occur.
Separate assembly into suitable containment for
blade assembly from handle and place the blade
subsequent reprocessing. See Figure 1. Do not place the blade assembly with sharp devices.
2. Prevent the blade assembly from drying per
facility
practice (for example, use immersion, neutral pH
for
enzymatic towel).
1. Disassemble the blade assembly per Figure 2.
pre-cleaner
or foam, or cover with a
moist
2. Keep all blade components together to prevent loss.
3. Select a neutral pH enzymatic cleaner.
4.
Prepare manufacturer
cleaning solution per cleaning
instructions.
solution
Figure 2
Page 3
Directions for use
Manual cleaning process:
1. Immerse all components in a neutral pH enzymatic cleaning solution as directed by manufacturer’s
2. While components are immersed in the cleaning
Welch Allyn st andard laryngoscope
solution
instructions.
blade assemblies
solution, use a soft-bristled brush to clean all component surfaces (the blade, retaining screw, and light carrier) until all
Drying:
visible contamination is
3. Thoroughly rinse all components in one of the potable filtered water, softened water, or per cleaning solution manufacturer instructions, or per instructions to remove cleaning
After cleaning, dry all components with a clean cloth or
removed.
solution.
following:
facility
to air dry.
Maintenance, Inspection and Testing prior to disinfection or sterilization:
1. Inspect
each
component
area (per deterioration. Also, inspect the following deterioration:
hooks of laryngoscope blade
Figure
(1)
and
3)
for
damage
areas
for
light assembly
(2)
lamp (3)
threads of blade
WARNING: Discard any component that
(1)
and l oc k i n g screw (4)
shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble the blade assembly by reversing
steps
illustrated in Figure 2.
3. Attach blade assembly to a clean and disinfected handle in known working condition. Verify
Blade assembly engages and locks onto handle.
Blade assembly deploys into its locked position on
handle AND lamp
Verify that light output is satisfactory.
If the la mp fails to replace the lamp
illuminates.
light or output is low
.
that:
test
, check or
After cleaning; choose either cold solution high level disinfection method (page 4) for any blade or one of the four sterilization methods (page 4-5) except REF 68470
Figure 3
3
allow
or
1. Laryngoscope
2. Light assembly
3. Lamp
4. Locking
blade
screw
Page 4
s
4
Directions for use
Cold solution high level disinfection instructions
Preparation decontamination:
Cold solution high level disinfection:
for
1. Select a neutral pH ortho-phthalahyde glutaraldehyde
2.
Prepare manufacturer
disinfecting solutions per disinfectant
1. Disassemble the blade per Figure 2.
2. Immerse all components in disinfectant solution for the time duration specified by the disinfectant manufacturer to achieve high level
3. Thoroughly rinse all components in one of the potable water, softened water, deionized water, or high purity water (for example, RO or distilled water) or per disinfectant solution manufacturer instructions, or per facility solution.
instructions to thoroughly remove disinfection
Drying: After cleaning, dry all components with a clean cloth or
to air dry.
Assembly: Reassemble the
Packaging:
Storage: Store
End of
reprocessing
in Figure 2. Package
the clean and disinfected blade assembly per
practice for return to service.
packaged
device to remain clean, dry, and ready for service.
instructions for cold solution high level disinfection.
Sterilization instructions
CAUTION:
After maintenance, inspection and testing, select one
Gravity autoclave
Packaging: Package wrap labeled as appropriate for the steam autoclave exposure parameters below;
Gravity
package autoclave
the blade assembly per facility practice in preparation for autoclave.
the set-up and operation of autoclave equipment. Gravity autoclave settings are as follows:
Temperature: 132 C (270 F)
Exposure time: 15
Minimum dry time: 30
Storage: Store per facility practice to allow the clean, dry and ready for service.
REF 68470 is not compatible with any sterilization process.
packaging, sterilization, and storage methods:
sterilization
the blade assembly in an appropriately sized single pouch or in a
sterilization: Follow equipment manufacturer
minutes (wrapped)
minutes
Welch Allyn standard laryngoscope
disinfectant.
(OPA)
blade assemblie
or
solution
instructions.
disinfection.
following:
blade
assembly by reversing steps
illustrated
blade assembly per facility practice to
of the following four
outlined
and
facility procedures in
packaged
blade assembly to remain
allow
facility
allow
Page 5
Directions for use
Pre-vacuum autoclave
Packaging: Package wrap labeled as below;
package
Pre-vacuum
appropriate
the blade assembly per facility practice in preparation for autoclave.
autoclave sterilization: Follow equipment manufacturer and
sterilization
the blade assembly in an appropriately sized single pouch or in a
for the
pre-vacuum autoclave
procedures in the set-up and operation of autoclave equipment. settings are as
Temperature: 132 C (270 F)
Exposure time: 4
Minimum dry time: 20
follows:
minutes (wrapped)
minutes
Storage: Store per facility practice to allow the
Welch Allyn st andard laryngoscope
exposure parameters
packaged
blade assembly to remain
blade assemblies
facility
Pre-vacuum
outlined
autoclave
clean, dry and ready for service.
Steris® Amsco® V-PRO® low temperature
Packaging: Package
the blade assembly per facility practice in an appropriately sized
single pouch or in a wrap labeled as appropriate for the Steris
sterilization
®
Amsco® V-PRO
collection of low temperature sterilization equipment and exposure parameters. Steris® Amsco® V-PRO® low temperature sterilization: Follow equipment manufacturer
and facility procedures in the set-up and operation of Steris temperature sterilization equipment.
• V-PRO
• V-PRO 1 PLUS:
• V-PRO
1: Lumen cycle
Lumen cycle, non-Lumen cycle
maX: Lumen cycle, non-Lumen cycle, Flexible cycle
Storage: Store per facility practice to allow the
V-PRO® sterilization settings are as
packaged
®
Amsco® V-PRO® low
follows:
blade assembly to remain
clean, dry and ready for service.
STERRAD® hydrogen peroxide gas plasma
Packaging: Package
the blade assembly per facility practice in an appropriately sized
single pouch, tray or in a container labeled as appropriate for the
sterilization
STERRAD® hydrogen
peroxide gas plasma exposure parameters. STERRAD® hydrogen peroxide gas plasma sterilization: Follow equipment manufacturer
and facility procedures in the set-up and operation of the gas plasma sterilization equipment. sterilization systems and cycles
• STERRAD
• STERRAD
• STERRAD
• STERRAD
100S: 100S: Long cycle (outside of US only) NX: 100NX: Express,
are as
Standard
Standard,
Advanced cycles
Storage: Store per facility practice to allow the
STERRAD® hydrogen peroxide gas plasma
follows:
(short) cycle
Standard
cycles
packaged
STERRAD® hydrogen peroxide
blade assembly to remain
clean, dry and ready for service.
End of
reprocessing
instructions for sterilization.
5
®
Page 6
s
6
Directions for use
Welch Allyn standard laryngoscope
blade assemblie
Maintenance
instructions
Replace
the la m p a n d / o r light a s s em b l y
(REF:
6904X, 6924X, and 6804X)
CAUTION: Use only lamps specified to ensure proper performance. Use lamp REF 04700 for: 69041, 2, 69241, 2, 68045, 0, 1, and 68470 blade/light assemblies
Use lamp REF 04800 for: 69043, 4, 69243, 4, and 68042, 3, 4 blade/light assemblies
1. Remove the locking screw by rotating counterclockwise with a standard
screwdriver.
2. Pull light assembly away from base of laryngoscope bl a d e and slide distal end out
3. To replace lamp: Rotate lamp counterclockwise until free and insert a new lamp and tighten.
4. Position new or re-lamped light assembly the reverse order of step 2.
5. Replace locking screw and rotate clockwise until secure.
6. Verify lamp and blade engagement/operation using a known working test handle.
7. Reprocess repaired assembly as appropriate per these instructions.
Figure 2
Specifications
Electrical: For use with ISO 7376-1 (black co de) s t andard type handles only
For information about electromagnetic compatibility (EMC) see Welch Allyn website: http://www.welchallyn.com
of
blade.
32
(0 C)
104 F (40 C)
F
Operating:
Storage/Transport:
-4 F (-20 C)
120 F (49 C)
Approvals: Conf or ms to A ST M F 965 and
The
CE
mark on this product indicates that it has been tested to and
with the provisions noted within the Complies with EMC Framework of Australia
ISO-7376-1, IEC/EN
93/42/EEC
60601-1,
Medical Device
IEC/EN
60601-1-2
conforms
Directive.
Warranty:
Service Information:
For
Technical
Welch Allyn
One year
Support or to obtain information about any Welch Allyn product,
Technical
Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA www.welchallyn.com
Support: w
ww.welchallyn.com/support.
Authorized European Re presentative Add ress: Regulatory Affairs Representative Welch Allyn, Limited Navan Business Park Dublin Road Navan, County Meath, Republic
contact
of Ireland
Loading...