Welch Allyn SP-10 Service manual

Page 1
SP-I0
Cardio Menu
Service Manual
S
pi
rom
ete
r
SP-10
User Guide
Schiller.
742Welch Allyn Inc.
7420 Carroll Road
San
Diego,
CA 92121
Phone: (800) 854-2904
Fax:
(619) 621-6611
277a
lpyright
0
97
& ’ 98
by
Welch
Allyn
United Kingdom: Welch Allyn
Cubblington Road Aston Abbotts United Kingdom
Tel: 0 1296-682- 140
Fax:
01296-682-1 04
Miller
UK
HP22
www.welchallyn.com
Ltd
4ND
Canada: Welch Allyn Canada Ltd
I60
Matheson Mississauga, Ontario L4Z 1V4 Canada
Tcl:
(800)
Fax:
(905)
I
Blvd.
561-8797
890-0008
East, Unit #2
Page 2
DECLARATION
OF
CONFORMITY
Spimmeter:
Serial numbers starting with: 531.
Year
of
manufacture: 1995 Onwards
We, the undersigned, hereby declare that the medical device (class IIa) specified above conforms with the essential requirement listed in Annex
This declaration is supported by:
Certificate of approval
and
No:
-10
1
of
EC
Directive 93/42/EEC.
11425-01
45112-60-00
45112-16-00 Annex
IS0
9001 (Rev.1994) EN 46001 by
IS0
9001/08.94, EN 46001 / 12.93 by DEKRA
11,
Section 3 of the Directive 93/42/EEC
(30.04.1995)
SQS
Baar (Switzerland) 03.07.1996
Martin Spreng Research
&
Development Manager
(<
0124
Markus Butler
Quality Assurance Manager
ii
Page 3
...
111
Page 4
This equipment has been tested and found to coinply with the limits for a class A digital device, pursuant to both Part
15
of the FCC (Federal Communications Commision) Rules and the radio interference regulations of the Canadian
Deparhnent of Communications. These limits are designed to provide reasonable protection against harmful interference
is
when the equipment
frequency energy and, if not installed and used in accordance with this instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area
interference in which case the user will be required to correct the interference at his own expense.
1
Disposal Instructions and Battery Care
operated in a commercial environment. This equipment generates, uses and can radiate radio
is
likely to cause harmful
O
DO NOT DISPOSE DANGER
O
DO NOT ATTEMPT TO RECHARGE THE BATTERY - DANGER EXPLOSION
O
DO
Only dispose of the battery in official recycling centres or municipally approved areas. Alternatively used batteries can be returned to Schiller AG for disposal.
OF
NOT OPEN THE BATTERY CASING - DANGER
OF
THE BATTERY BY
EXPLOSION
FIRE
OR INCINERATOR
OF
ACID BURN
-
OF
.%c9
Power Supply
The mains connection is on the rear of the unit.
The power supply voltage is set by the factory for100-115V(nom. llOV) or 220-240V (nom. 230V) working. The setting is indicated by the indented metal strip on the fuse panel. Contact your dealer if the voltage needs to be changed.
The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. The power source is indicted on the top line of the LCD.
Changing a Mains Fuse If it
is
necessary to change a fuse, always replace with the correct rating i.e 2x200mAT for 230V, or 2x400mAT
for llOV
To change a fuse press the panel. Remove the fuse panel and replace the fuse(+ Click back the fuse panel.
.
two
Unit Disposal Instructions
Units no longer required can bereturned to Schiller AG for disposal. Alternatively dispose of the unit in municipally approved recycling centres.
retaining lugs on side of the fuse panel (situated below the mains connector on the back
L
J
iv
Page 5
PHYSICIAN’S RESPONSIBILITY
THE SP-10 SPIROMETER IS PROVIDED FOR THE EXCLUSIVE US€ OF
QUALIFIED PHYSICIANS OR PERSONNEL UNDER THEIR DIRECT SUPERVISION. THE NUMERICAL AND GRAPHICAL RESULTS FROM A
RECORDING MUST BE EXAMINED WITH RESPECT TO THE PATIENTS OVERALL CLINICAL CONDITION. THE.RECORDING PREPARATION QUALITY AND THE GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DATA ACCURACY, MUST ALSO BE TAKEN INTO ACCOUNT.
IT IS THE PHYSICIANS RESPONSIBILITY TO MAKE THE DIAGNOSIS ORTO OBTAIN EXPERT OPINION ON THE RESULTS, AND TO INSTITUTE CORRECT TREATMENT IF INDICATED.
FEDERAL LAW IN THE USA RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER
OF
A PHYSICIAN
Page 6
I
Safety
Notices
\vercqnuyrs*
Sc
hiller
F
THIS UNIT IS BF CLASSIFIED ACCORDING TO IEC 601 TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS
INSIDE. REFER SERVICING TO QUALIFIED PERSONNEL ONLY. DO NOT USE THIS UNIT IN AREAS WHERE THERE
PRESENCE OF FLAMMABLE GASES SUCH AS ANAESTHETIC AGENTS.
THIS PRODUCTIS NOT DESIGNED FOR STERILE USE.
THIS PRODUCT SWITCH THE UNITOFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS. DO NOT UNDER ANY CIRCUMSTANCES, IMMERSE THE UNITOR CABLE ASSEMBLIES IN LIOUID.
THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION IS SUSPECTOR IF THE MAINS LEAD
I
DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOTUSE E-BEAM OR GAMMA RADIATION STERILISATION.
DO NOT USE SOLVENT CLEANERS
USE ONLYACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY WELCHALLYN SCHILLER. USE OFOTHER THAN RECOMMENDED OR SUPPLIED PARTS MAY RESULTIN INJURY INACCURATE INFORMATION AND/ OR DAMAGE TO THE UNIT.
THE SP-10 COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICHAFFORDS
PROTECTION AGAINST EMISSIONS AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL
CARE MUSTBE EXERCISED WHEN THE UNITIS USED WITH HIGH FREQUENCY EQUIPMENT.
IS
NOT DESIGNED FOR OUTDOOR USE.
IS
DAMAGED OR SUSPECTED
-1.
IS
ANY DANGER
OF
EXPLOSION OR THE
OF
BEING DAMAGED.
I
vi
Page 7
SP-I0
This User’ s Guide gives instructions on how to opemte the unit and provides an overview of all the basic functions in an easy and simple to use format. The procedures are presented in a
logical, step-by step way to enable the userto quickly and easily familiarise themselves with unit opemtion. Detailed medical information is excluded from this guide except where necessary
to opemte the unit or understand the
User‘s
results.
Guide
Welc9AUyrr
Sc
hiller
9314UEEC Medical Devices:
0124 ‘Notified Body’ DEKRA AG
vii
Page 8
viii
Page 9
Chapter
Introduction and System Settings
1
Introduction
INTRODUCTION
Getting Started Location Potential Equalisation (Grounding) Switch the Unit On/Off Overview
DISPLAY
Liquid Crystal Display (LCD) Adjusting the Contrast
KEYBOARD
Alpha Numeric Keyboard General Purpose Keys Control and Function Keys
Menu Control Keys
PANELS
Right-hand Side Panel Rear Panel Mains Module
SYSTEM SElTlNGS
General Adjusting Date and 'Time User Identification Printout Settings Device Settings Device Setup Combinations
.............................................................................................
..........................................................................................
.....................................................................................................
of
the SCHILLER
..........................................................................................................
.....................................................................................................
..................................................................................
.........................................................................................................
..............................................................................................
..........................................................................................
.....................................................................................
....................................................................................................
....................................................................................
......................................................................................
.......................................................................................
and
...............................................................................
SP-10
.....................................................................
...............................................................................
..........................................................................
..............................................................................
........................................................................
............................................................................
.........................................................................
....................................................................
System
...........................................................
...........................................................
1.10
1.11 1-11
1.12
1.12
1.13
1.13
1.13
1.14
1.14
1.15
1
Settings
1.2
1.3
1.3
1.3
1.3
1.4
1.5
1.5
1.5
1.6
1.7
1.8
1.9
.
16
SCHILLER
SP-10
Spirometer . User Guide
1
.
I
Page 10
Chapter
1
lntroduction and System Settings
INTRODUCTION
The SCHILLER SP-10 is a sophisticated, compact workstation for pulmonary diagnosis with
the following characteristics:
Multiple operation possibilities, thanks to low weight and battery power.
0
LCD display. Menu guidance provided for ease The SP-10 is capable of working in four measurement modes as follows:
of
operation.
Forced Vital Capacity (FVC-Test) Vital Capacity (VC-Test) Expired or Minute Ventilation (MV-Test) Maximum Voluntary Ventilation (MVV-Test)
Additional options enable resistance testing and blood oximetry diagnosis to be carried out.
The test results are displayed on the integral LCD and recorded on the built-in thermal printer.
The open pneumotacho transducer SP-110, with disposable mouthpiece, is usedfor respiratory measurement.
As options, WELCH ALLYN offers instead disposable measurement unit integrated inside the mouthpiece
As further option in combination with resistance tests the SP-11
necessary. This sensor can also be used for normal pulmonary function tests.
of
the SP-110, either the SP-150 with
or
the
SP-20.
O/R
flow
sensor is
RS-232
connection to the SCHILLER PC-based data management program (SEMA) for
validation and archiving
The unit also has a facility for the connection
and
RS-422
interfaces are provided for data transmission and for the
of
recorded data.
of
an external video monitor.
1.2
Page 11
Introduction and System Settings
INTRODUCTION
Getting Started
Location
Do
not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid
exposure to direct sunlight or heat from other sources.
Do
not allowthe unit
cause irreparable damage.
The unit should not be placed near X-ray or diathermy units, large transformers or motors
of any kind.
to
come into contact with acidic vapours or liquids, as such contact may
WARNING:
Chapter
1
THIS APPARATUS SHOULD NOT BE OPERATED
DANGER
Potential Equalisation (Grounding)
Attach the yellow/green ground lead (supplied with the unit) between the potential equalisa­tion stud of the SP-10 (situated on the rear panel) and the hospital/building common ground.
The connection for the potential equalisation is marked with the+ symbol.
Switch the Unit OnlOff
The SCHILLER SP-10 can be operated either from the mains supply or from the built-in batteries.
To operate the unit from the mains, connect the power cable
it into the connection on the rear
TheSCHILLERSP-1Oisswitchedonoroff
connection on the rear
and the ON key is not pressed, the battery is charged
The SP-10 can remain connected
OF
EXPLOSION.
of
the unit.
withtheVOpowerswitch locatedabovethe mains
of
the unit (green lamp is lighted when on). If the unit is switched on
to
the mains supply without damage.
IN
AREAS WHERE THERE
to
the mains supply and plug
if
necessary (yellow lamp is lighted).
IS
ANY
SCHILLER
To
start
the functions the
on with the
A
self-test of the unit is automatically initiated every time the unit is switched on.
I/O
switch, the SP-10 runs on battery power (yellow lamp is lighted).
ON
key has
to
be pressed. If that is done without switching the unit
BEFORE SWITCHING THE UNIT ON, ENSURE THAT THE FLOW SENSOR PLUGGED IN.
IF
THE SENSOR
IS
NOT CONNECTED, A MESSAGE APPEARS TO CONNECT THE SENSOR. THE UNIT MUST FIRST BE SWITCHED OF5 THE SENSOR CONNECTED AND THE UNIT SWITCHED ON AGAIN.
THE FLOW SENSOR CAN ONLY BE CONNECTED WHEN THE UNIT
OFF. IF ANY KEY
IS
PRESSED WITH THE SENSOR NOT CONNECTED THE UNIT
IS
AUTOMATICALLY SWITCHES OFF.
SP-10
Spirometer - User Guide
IS
SWITCHED
1.3
Page 12
Chapter Introduction and System Settings
1
INTRODUCTION
Overview
The main operating and connection modules of the SCHILLER
Keyboard
of
the
SCHILLER
SP-10
The keyboard comprises the general function and control keys, and the alphanumeric keys for data entry.
SP-10
are as follows:
LCD Screen
Thermal Printer
Right-hand Side Panel
Left-hand Side Panel
Back Panel The back panel contains the video monitor connector and the
The LCD screen displays the pulmonary traces and certain operating and status information. Under operator control the display also gives menu options and displays operator entered data. The display is folded down when not in use.
The printer provides the hard copy of test results. The side panel contains the Spirometry connector and the
serial connectors. This panel contains the paper tray and (below the paper tray)
the program pack.
mains connector with on/off switch, voltage selector and mains
fuse.
RS
1.4
Page 13
Introduction
and
System Settings
DISPLAY
Liquid Crystal Display (LCD)
The liquid crystal display (LCD) performs both as a real-time monitor and as an alphanumeric display for menu selection, data input and the provision of important information.
Depending on the function, the displayed information will vary but the screen is normally
divided into three main parts:
Top
of
the screen
status and operating information are permanently displayed.
Chapter
1
Right side
Left side
Central Part
Lower Part of Display
of
Display
test results are displayed.
the display is divided into the following five fields from the top:
0
Date and Time
/
Function Status of CAPS key Alert field
In the central part test is displayed.
The very bottom line gives the function key for the next step.
Battery capacity in
of
the display the test curve
%
of
the selected
Adjusting the Contrast
The contrast can be adjusted by means of the LCD knob located on the right-hand side of the unit. Visibility is best when the display is fully illuminated.
SCHILLER
SP-10 Spirometer
-
User
Guide
1.5
Page 14
Chapter Introduction and System Settings
1
KEYBOARD
The keyboard is divided into the following functional sections:
alphanumeric keypad menu control keys control and function keys
test function keys
general purpose keys
1.6
Page 15
KEYBOARD
Alpha Numeric Keyboard
Chapter
Introduction and System Settings
1
The alphanumeric keyboard serves as a normal keyboard for data input and follows:
Special keys of the alphanumeric keyboard:
!QaL
I+-
(-I
ENTER
OEL
SPACE
Function
Delete character to the left of the cursor.
Confirm entry or selection.
Enter blank space into the text.
is
arranged as
It
Ir
When pressed simultaneously with a letter key it activates the uppercase letter. When pressed simultaneously with a number
key it activates an alternative symbol set.
Toggle key to set lower case or upper case letters. The Caps Lock key status is indicated on status information
field on the right-hand side of the screen.
SCHILLER
SP-10
Spirometer - User Guide
1.7
Page 16
Chapter Introduction and System Settings
1
KEYBOARD
General
The general purpose keys are:
pcTJ
.............
("p)
............
am
(a]-p-J
........................
......
Purpose
Keys
call up main menu I return to monitor mode
return to previous menu
change setting
change setting I enable command
move the cursor up
move cursor
I
disable command
I
down
to
left
I
right
1.8
Page 17
KEYBOARD
Introduction and
System
Chapter
Settings
1
Control
The control and function keys are:
p-J
(STARTJ
...........
and
Function
Keys
switch unit on
switch unit
extend I retract paper tray
enter patient data
start selected test
stop test
off
/
move paper to start position
SEND
I.
DATA
..........
STORE
mi
...........
pF-J
..........
print best test result
print all test results
select test result for printout
i
i
:
initiate data transmission
initiate data storage
initiate reception mode
enter details of medical assistant
display general settings menu
delete stored measurement
SCHlLL
,
ER
SP-
10
Spirometer - User Guide
1.9
Page 18
Chapter Introduction and System Settings
7
KEYBOARD
Menu Control Keys The menu control keys are:
display test results
IFm]
.............
FVC
E
1
I
FVC
E-l
1
(svc)
..........................
p,vv
[RESIS.)
..........................
,)
...........
inhalation
volume
Forced Vital Capacity test (Exhalation only)
Forced Vital Capacity test (Exhalation and Inhalation)
Slow
Minute Volume (volume in one minute)
Minute Volume (forced volume in one minute)
Resistance Test (Option)
/
exhalation
(I)
vs. Time
Vital Capacity test
(s)
flow
(Vs)
vs. Volume
(I)
1.10
(PRUPOST)
FALIBRATIO3
..........................
pre
/
post medication test
calibration of the unit
to
environmental conditions
Page 19
PANELS
Chapter
Introduction and System Settings
1
The SCHILLER
on
the right-hand side of the unit.
Right-hand Side Panel
SP-10
has two connector panels. One is located on the back and the other
7
RS-232C interface
1
2
RS-232C interface
3
RS-232C
interface
(Port 1) (Port 2)
(Port
3)
3
6
(not used) (not used) (for data transmission
15 2 4
(as.
to
SEMA)
4
RS-422
5
LCD
6
Connection for:
SP-110 SP-150
SP-20
SP-11
N0TE:The resistance sensor
VC, MV, MVV) tests.
7
Connection for optional
interface
contrast control
O/R
(Port 4)
for pulmonary function measurements with disposable mouthpiece
for
pulmonary function measurements for single use with disposable
f
low-sensor for pulmonary function measurements with disposable filter for resistance measurements (Option) with disposable mouthpiece
SP-lIO/R
Sp02
sensor
can also be used for pulmonary function (i.e. FVC,
SCHILLER SP-10 Spirometer - User Guide
1.11
Page 20
Chapter
1
Introduction
and
System Settings
SYSTEM
User
Identification
This function enables entry of the name of the physician, clinic or department which will then be given on each printout. The input is stored until overwritten by the user. the unit is switched off. To reach that menu, proceed as follows:
my-<
.............
The maximum number of characters is
0
SE77INGS
Confirm with
Press
ESC
.........................
User
id.
(,ENTER.
f
user
>
23.
to return to the settings menu.
(memorized
I
(confirm id
identification:
when
with
power
'ENTER')
It
is not
off)
lost
when
Printout Settings
There are five settings which may need axis for a printout of the diagnosis on the printout, including of a flow graph on the printout and the sensor that is connected and used for the tests. These settings stay stored until they are selected new.
To reach the menu proceed as follows:
Br-<
Selections in the menu:
By pressing key
10
mds
0
20
mds
By pressing key2 and setting (for explanations, check in the chapter "Value Information" under paragraph
"Diagnosis").
By pressing key 3 and setting
Print
FVC
graph, the norm values that are to be used, including
settings
1,
the time axis for the printout
to
be made before starting the test, namely the time
FVC-Graph
flow
Diagnosis
Flow-Graph
PEF
[Vmin]
of
an
FVC
graph can be set
is
given as indication on the printout
graphics
is
printed.
>
to
ON
to
ON
the diagnosis
the
of
to:
1.14
By pressing key
4
and setting
to
ON
the
PEF
values are printed.
Page 21
Chapter
1
Introduction and System Settings
I
I
SYSTEM
Device
There are seven settings which need
SETTINGS
Settings
to
be made before starting the test, namely:
the sensor that is connected and used for the tests calculation basis for norm values
norm values that are to be used race of the patient calculation units for weight and size of the patient
order of date
measurement units for temperature.
All these settings stay stored until they are selected new.
To reach the menu proceed as follows:
.
-.
...
SrnlNGS
r-7
-<
Device
Settings
>
LJnits
in
-
. ..
.
Order
of
date
Temperature
(61;
F;&;yy
Selection possibilities:
By pressing key 1, the flow sensor that is to be used can be selected. The available
settings are:
SP-110
(also used for
SP-I
1
O/R
resistance sensor)
SP-150
SP-20
CAUTION:
ENSURE THAT THE CORRECT SENSOR INACCURRATE WHEN THE INCORRECT SENSOR TYPE
By pressing key (see Chapter
0
3).
International; for all countries outside the USA and Canada
2,
IS
SELECTED. MEASURED VALUES CAN BE
IS
SET.
the diagnosis standard can be selected
SCHILL
ER
By pressing key be selected (see Chapter
SP-
10
Spirometer - User Guide
American; for USA and Canada
3,
the standards used for the various measurement calculations can
3).
1.15
Page 22
Chapter
2
Taking Tests
Taking
CONNECTING THE FLOW
MTA Identification
CALIBRATION PROCEDURE PATENT DATA PULMONARY FUNCTION TESTS
Schematic Overview of the Test Procedure General Test Procedure
Select Test Display Start Tests Display Results Print Tests
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION ONLY
Interrupt
Delete Measureme Screen Information Toggle Screen Infomation
Display Results
Insert1
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION AND INHALKTION2.14
Carry Out Test
Interrupt
Delete Measurements
Screen Information
Toggle Screen Information
Display Results
InSetl
SLOW VITAL CAPACITY (SVC) TEST
Carry
Interrupt I End Test Delete Measurements Screen Information Display Results InSelt
EXPIRED
Carry Out Test
..
............................................................
........................................................................
..
, , , ,
......................................
......................................
I
End Test
Edit Diagnosis
.........................................................................................
I
End Test
I
Edit Diagnosis
Out
Test
.........................................................................................
I
Edit Diagnosis
OR
MINUTE VENTILATION (MV) TEST
SENSOR
...............................................................................
I
ENTERING USER NAME
........
.........................................
.........................................................................
2.6
........................................................
.........................
....................................................
............................................................................
......................
....................................................................
.................
..........................................................
....................................
2.8
2.9
2.10
2.10
2.10
...................................................
..........................................................
............................................................................
...........................................................................
.............................
.....................
................................................................................
.....................................................................
.................................
.................
........................................................................................
....................................
.....................
.........................................................
..................................................................................
.............................................................................
..................................................................................
........................................................................................
.............................................................................
.........................................................................................
......................................
2.13
2.14
2.14
2.15
2.15
2.17
2.17
2.18
2.18
2.18
2.19
2.19
2.20
2.20
2.21
2.21
Tests
SCHILLER
SP-I0
Spirometer - User Guide
Delete Measurements
.......
2.1
Page 23
Chapter
Taking Tests
2
Screen Information Display
Results
I
Edit Diagnosis
Insert
MAXIMUM VOLUNTARY VENTILATION (MVV) TEST
Carry Out Test Interrupt Delete Measurements Screen Information
MAXIMUM VOLUNTARY VENTILATION (MVV) TEST
Display Results Insert
POST-MEDICATION TESTS
Select Post-Medication
Print Post-Medication Tests PRINTING TEST CURVES AND RESULTS RESISTANCE TEST OPTION
Test Procedure
Print Resistance Tests
I
End Test
/
Edit Diagnosis
..................................................................................
.........................................................................................
.............................................................................
.........................................................................................
..................................................................................
.............................................................................
..................................................................................
........................................................................................
.............................................................................
.........................................................................
...........................................................................
.....................................................................
.................................................
.......................................................................
........................................................................................
.............................................................................
................................
................................
2.22
2.23
2.23
2.24
2.24
2.24
2.25
2.25
2.26
2.26
2.26
2.27
2.27
2.27
2.28
2.29
2.29
2.30
2.2
Page 24
Chapter
Taking Tests
2
CONNECTING THE
CONNECT
OR
DISCONNECTTHE
FLOW
SENSOR/ ENTERING USER NAME
IMPORTANT:
FLOW
SENSOR ONLY WHEN THE UNIT
IS
TURNED
OFF.
Each flow sensor is supplied completely with connection cable and plug.
Insert the plug into the socket marked
EACH FLOW SENSOR CONTAINS A SENSITIVE MEASURING DEVICE AND MUST ALWAYS BE HANDLED WITH CARE. DO NOT ALLOW THE SENSOR TO BE DROPPED OR
SUBJECTED TO ANY SUDDEN BLOWS
ENSURE THAT THE CONNECTION CABLE MECHANICAL STRESS. WHENEVER DISCONNECTING THE SENSOR, HOLD THE
PLUG AND NOT THE CABLE..
"SPIRO
SENSOR' on the right-hand side of the unit.
CAUTION:
IS
NOT TO BE EXPOSED TO ANY
The Unit is turned on
If
no sensor
When this message appears, switch the unit
MTA
This function is must be entered every time the unit is used. The MTA is identified on each printout. Enter
the menu as follows:
0
is
ldentif ication
Confirm with
by
pressing
connected, the following message appears:
to
register the name of the medical assistant carrying out the recording and
&,
a
off
and connect the sensor.
Identification (stored
until
of
MTA
power
off)
I
SCHILLER
[
Press
SP-10
Spirometer - User Guide
ESC
]
to return to the main menu.
2.3
Page 25
Chapter
2
Taking Tests
CA
LIB
RA
TlON
The unit must be calibrated:
before starting the first pulmonary function test of the day
after a significant temperature change
after changing the sensor.
Press the calibration key
to
disaply the calibartion menu:
CALIBRATION
Last
calib.
BTPSFactor Calib. Factor
Temperature Measured Syringe
Vol.
Vol.
SP-110
:
:
:
: :
:
The following positions are displayed:
Headline The headline displays the selected sensor.
the sensor that is connected. If not, select the correct sensor in the Settings menu. Press the SETTINGS
'device settings' option. The sensor type is displayed. Toggle
Last Calibration BTPS Factor
with key
date of last calibration calculated BTPS
1
between SP-
(=>
1
10,
SP-
150
and SP-20.
Body Temperature, Ambient Pressure,
Saturated with water vapour) value
Calibration Factor
calculated value between measured and effective calibration
air volume
23.04.05
1.090
1.000
20.0
"C
0.00
I
3.00
I
Ensure that this is
KEY
and select the
2.4
Temperature
Measured Volume Syringe Volume
ambient temperature in
OC
(or
OF)
dependent on device setting air volume measured by the system from the calibration pump Actual (entered) air volume depending on the
size
of the calibration pump and times the air was pumped through the sensor; e.g. pumping
(the recommended volume with a
3
litre pump, 9 litres)
2
litres 4 times amounts
2
litre pump is 8 litres; with
to
8
litres
Page 26
Chapter
2
Taking Tests
CALIBRATION
1
Connect the calibration pump to the sensor. Ensure no air leaks.
Press the CALIBRATION key. The menu appears as shown opposite.
2
Enter the ambient temperature as prompted
3
4
Confirm with
5
6
7
NOTE: Make sure that the flow sensor is kept still during the pumping operation. While pumping, the unit records the actual volume being pumped through the flow sensor
and indicates it on the display.
8
9
[START)
Press
Wait 1 second.
3
Pump
Press when finished pumping.
The cusor
(depending on the size
a litre pump emptyed once
PROCEDURE
(,ENTER
...........................
litres,
is
now positioned at syringe volume. Enter the total pumped air volume
.
to
start the calibration.
or
more, air through the sensor.
of
the calibration pump and number
=
3
litres, the same pump emtgyed twice
of
times of emptyed i.e.
=
6
litre).
.
Press&
10
11
The message "Calibration completed" appears on the display
If a printout
printed with the following information:
Measured Volume in litres Effective Volume in litres Deviation in litres
Deviation in percent Temperature in Celsius
BTPS Factor
0
Calibration Factor Flow Graph
Date, Time and Software version
12
Press
The unit then calculates the calibration factor.
of
the calibaration is required, press
or
Fahrenheit
or[
ESC
1
to return
to
the main menu.
.
A calibration report is
Note: If the message 'EXCESSIVE DEVIATION! 'appears on the screen aftera calibration,
it indicates that the difference between the measured volume and the enteredvolume is
too
great
(~10%).
entered syringe volume. Reset the calibratioin to bease valuse by pressing
Check the temperature setting, the syringe volume and the
'I'
and
recalibrate.
SCHILLER
SP-10
Spirometer - User Guide
2.5
Page 27
Chapter
2
Taking Tests
PATENT
DATA
Each printout is complete with the name and other information concerning the patient. Before beginning a recording, the patient data should be entered. In order
to
call up the menu for
patient data input, proceed as follows:
The following parameters have
to
be entered:
Pat-Name Pat-No
:
:
(dd.mm.yy)
(MIF)
Pat-Name: Patient name
Pat-No:
Born:
Patient number The date of birth has to be entered in figures, separated by a
full stop as day.month.yearor month.day.year (order depending
on unit setup).
Age:
The age is calculated by the
SP-10
on the basis of the date of
birth:
-
up to 2 years: number
-
up
to
6
years: number
-
over 6 years:
Sex:
number enter
of
months;
of
years and months;
of
years
"M"
(male) or
"F"
(female).
Height: in centimetres (or inches, depending on unit setup). Weight: Race:
WIB:
CIHIBIA:
Weight in kilogrammes (or pounds, depending on unit setup).
Enter patients race (depending on unit setup):
"W"
for White,
"CI
for Caucasian,
"B
for Black
"H"
for Hispanic,
"8"
for Black or
"A"
for
Asian
=>
If this position does not appear, "NONE" was selected in the
setup (see Chapter
1)
Wrongly typed characters can be deleted before pressing the ENTER key by pressing the DELETE key. The old content
If an error is noticed after ENTER has been pressed, complete the entry
then press the
PATlENTDATAkeyandthecursor
of
the line is deleted as soon as the first character
of
all other data and
returnstothefirst linefordataentry. Press
is
entered.
ENTER until the cursor returns to the line in question and enter the data required. The existing entry is deleted as soon as new data
is
entered.
2.6
Page 28
PULMONARY FUNCTION TESTS
Chapter
Taking Tests
~
2
Schematic Overview
________--
IF---
I
of
the Test Procedure
Switch on SP-10
r
with mains switch
I
Connectsensor
I
I
........
--1
.:
L
r
Make measurements
Delete Measurement(s)
Repeat tests
with new patients
.....
,......
Clean sensor and
replace mouthpiece
Switch
with mains switch
off
Post-medication
tests
SP-10
SCHILLER
SP-10
Spirometer - User Guide
2.7
Page 29
Chapter
Taking
2
Tests
PULMONARY FUNCTION TESTS
General Test Procedure NOTE: Exact descriptions of each step is given in the appropriate section.
Select Test
The required test is selected directly with the appropriate key on the keyboard.
Display The display indicates the measurements which will be made and gives the predicted results.
to
There are three positionsfor measurement results enabling several tests
if
all the necessary information is displayed in the top part
requested setup:
PRE
FVC FEVl
I
PEF
MEAS3 MEAS2 MEASl PRED.
5.47
4.49
9.84
SPO2
-
-Yo
(
--
/mi n)
of
the screen according to the
We
be made. Check
26.APR.95 13:26:20
The following information is displayed from installed):
Test information about:
Premeditation (PRE) or Post-medication (POST) test
3
Measured values for all Predicted (calculated) values according
(PRED.)
0
Sp02 value
0
Data field containing: Date and Time Unit Information Percentage value of battery capacity Status of CAPS key (CAPS
Possible error message for Sp02 sensor
0
A
small field on the top right of the chart shows the selected sensor and the set norm
values.
The coordinates represent the graph on which the curve volume in litres being represented on thevertical axes forthe FVC and VCtests (the relative respiratory volume in litres for the horizontal axis.
(“A)
and heart rate (/min) (if Sp02 option
tests (MEAS)
=
MV
left
Key locked)
and
MVV
to
right (depending on the option that is
to
patient data and norm value settings
is
installed)
will
be drawn with the respiratory
tests) and the time in seconds on the
2.8
In the
FVC
test a second presentation possibility the vertical axis and the flow volume in litres on the horizontal axis. The top part of the screen remains
the
same.
is
in form
of
the flow speed presented on
Page 30
Chapter
Taking
2
Tests
PULMONARY
Start Tests
N0TE:The flow sensor must be held quite still and no air should be breathed into the device
for at least one second before and after the START key is pressed.
Press the green key to start the test
0
Patient carries out test
After patient has carried out the test press the
Repeat the above procedure for three or more times.
Display Results
With
............
FUNCTION
all the values can be displayed as a table.
TESTS
[
............................
STOP
1
key to register the result
Print
Tests
By pressing either
The printout contains the patient data, selected normvalues, main valuesfor that test, Sp02 percentage rate
(HR),
values for measurement (MEAS), predicted values (PRED.) and percentage rate
between measured and predicted values (%PRED.) (PRED.
@
if
option is installed (Sp02), heart rate in beats per minute
,
or
,
aprintout is made according to the selection.
=
100%).
if
option
is
installed
SCHILL
ER
SP-
10
Spirometer - User
Guide
2.9
Page 31
Chapter
2
Taking Tests
FORCED VITAL CAPACITY (FVC) TEST- EXHALATION
ONLY
Forthis test the patient must exhale as quickly as possible from the time of starting the test.
so,
Inform the patient
Note: The FVC test employs the
too
large
is
a question mark
that it is sure that he understands what is required of him.
"Back
extrapolation" method. If the extrapolated volume
(>O.
1
litres or
(?)
70%
of FVC), then a warning appears on the display and
will appear in the FEV, position for that test.
To beabletocarry outthis kindof Forcedvital Capacity (FVC)test,first pressthe key.
As
soon as the basic screen
is
displayed, check
if
all the necessary information is displayed
in the top part of the screen according to the requested setup.
Note: The flow sensor must be heldquite stillandno air shouldbe breathedinto the device
for at least one second before and after the START key is pressed.
Press the green key
to
start the test.
The message
"READY
FOR
MEASUREMENT"
appears on the display.
As
soon as the patient starts to breathe into the flow sensor, the unit begins to record the flow and the corresponding curve is represented on the display. An acoustic "beep" indicates when the break-off point for the measurement has been reached.
Five seconds later, the measurement will be automatically interrupted unless the patient carries on. Measurements will still take place should this break-off point not be reached. Once the test is completed, the measurement results are calculated and given on the display.
Interrupt / End Test
The test can be interrupted at any time by pressing
L-zy?zJ.
Delete Measurements
If one of the result columns should be deleted, proceed as follows:
1
Delete
MEAS
DELETE
:
MEAS.
!!
I..
Press the number of that column (e.g. for column
the results should be deleted,
i.e.
all three columns, press key
1,
press key
2 Delete
3
Delete
-
4
Delete
1)
that is to be deleted.
4.
MEAS
MEAS
ALL
1
2
3
meas.
If
all
2.10
Page 32
FORCED VITAL CAPACITY (FVC) TEST- EXHALATION ONLY
Screen information
Once the test iscompleted, the measurement results are calculated and given on the display. When subsequent tests are made, the result with the highest FVC
be saved and given in the right-hand column (MEAS
The result with the lowest FVC + FEV, value of the tests already made and stored will be overwritten. An asterisk
NOTE: The marked results correspond to the displayed test curve.
Toggle Screen information
("*")
indicates the measurement results of the last test made.
1).
+
FEV, value will always
Chapter
Taking Tests
2
Two different screen types can be displayed
to
achieve more information.
--
screen, press either
NOTE: The top part of the screen always remains the same.
0
With
w:
0
air volume (in litres).
The information key as long as no further test was made.
NOTE: These keys only function with the FVC test.
pmJpoJ
.._..........
or
..............
.
the chart is given as volume (in litres) in relation of time (in seconds).
the chart is given as flow speed (in litres per seconds) in relation
of
the previous screen remains stored and can be displayed with the other
To
toggle the
to
the
SCHILL
ER
SP-
10
Spirometer - User Guide
2.
I I
Page 33
bnaprer
Taking
z
Tests
FORCED VITAL CAPACITY (FVC) TEST- EXHALATION ONLY
We
PRE
I
FVC 3.06 4.04 4.70 5.47 FEVl 2.71 3.32 2.68 4.49
I
PEF 4.04 4.26 2.96 9.84
ME AS^
ME AS^
MEASI
PRED.
26.APR.95 13:26:20
SPlROVlTSP-10
I
I
95%
STOP=Stop
-10-
I I
2
measurement
12­1
0-
8-
6-
4-
-2-
4-
6-
0-
I I
t
&
....',"""
I
I
I
I
I I
4
:
6
8
10
I
I
I
12
[SI
2.12
Page 34
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION ONLY
Display Results
is
To
display results of the test, press
PRE
FVC 3.06 4.04 4.70 FEVl 2.71 3.32 2.68 4.49 PEF 4.04 4.26 2.96 9.84
ME AS^
ME AS^
MEASI
and the following information
PRED.
5.47
We
SPIROVIT SP-10
L
displayed:
26.APR.95 13:26:20
Taking Tests
I
95%
I
CAPS
Chapter
2
I
FVC FEVl FEVl IFVC FEF.2-1.2 FE F25-75% FEF7 5-85% PEF MEF75Yo M EF50Yo M EF2so/o FMFT
FlVC FlVl FI V 1 /FIVC FlVl /FVC PIF MIF50Yo
Press
lFNCTl
or
IESCI
PRED. MEASl MEAS2
1.10
to
exit.
MEAS3 YoPRED
I
Insert / Edit Diagnosis
Press the
Press
SCHILLER SP-
...._.._.._
orLJE,s,c,,J toexit.
10
Spirometer - User
key when in the results menu once more
Guide
to
insert
or
edit the diagnosis.
2.13
Page 35
Chapter
2
Taking Tests
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION AND INHALA TlON
For this test the patient must exhale and inhale as quickly and strongly as possible from the
time of starting the test.
Inform the patient
so,
that it
is
sure that he understands what is required of him.
NOTE: The FVC test employs the "Back extrapolation" method. If the extrapolated volume
is too large
a
question mark
To
beabletocarry outthis kindof
key.
As soon as the basic screen is displayed, check
in the top part of the screen according
Carry
Out Test
NOTE:
(>O.
1
litres or
(?)
10%
of FVC), then a warning appears on the display and
will appear in the FEV, position for that test.
ForcedVitalCapacity(FVC)test,first
if
all the necessary information is displayed
to
the requested setup.
pressthe
The flow sensormust be heldquite stillandno airshouldbe breathedinto the
device for at least one second before and after the START key is pressed.
Press the green
The message "READY patient starts corresponding curve is represented on the display. An acoustic "beep" indicates when the break-off point for the measurement has been reached.
to
[START
..........................
FOR
MEASUREMENT" appears on the display. As soon as the
breathe into the flow sensor, the unit begins
key
to
start the test.
to
record the flow and the
Five seconds later, the measurement will be automatically interrupted unless the patient carries on. Measurements will still take place should this break-off point not be reached. Once the
Interrupt I End Test
The test can be interrupted at any time by pressing
test
is completed, the measurement results are calculated and given on the display.
2.14
Page 36
Chapter
2
Taking Tests
I
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION AND
LA
TI0
INHA
Delete Measurements
If one of the result columns should be deleted, proceed as follows:
N
1
Delete MEAS
2
DeleteMEAS2
3
Delete MEAS
4
Delete
ALL
1
3
meas.
Press the number of that column (e.g. for column
the results should be deleted, i.e. all three columns, press key
Screen Information
Once the
test
iscompleted, the measurement results are calculated and given on the display.
1,
press key
1)
4.
When subsequent tests are made, the result with the highest FVC be saved and given in the right-hand column (MEAS
The result with the lowest FVC + FEV, value
overwritten. An asterisk
NOTE: The marked results correspond
Toggle Screen Information
("*")
indicates the measurement results of the last test made.
to
the displayed test curve.
Two different screen types can be displayed
1).
of
the tests already made and stored will be
to
achieve more information. To toggle the
screen, press either WorW.
NOTE: The top part of the screen always remains the same.
that is to be deleted.
+
FEV,
value will always
If
all
SCHILLER
With
[vol
,.__....____
the chart is given as volume (in litres) in relation of time (in seconds).
0
With
the chart
is
given as flow speed (in litres per seconds) in relation
to
air volume (in litres).
The information of the previous screen remains stored and can be displayed with the other
key as long as no further test was made.
NOTE: These keys only function with the
SP-10
Spirometer - User Guide
FVC
test.
the
2.15
Page 37
Chapter
Taking
2
Tests
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION AND
INHALATION
We
PRE
FVC 4.22 4.26
FEVI
I
PEF
ME AS^
1.64 1.83
1.73 1.87 2.21 9.84
ME AS^
MEASI
4.64
1.90 4.49
PRED.
5.47
26.APR.95 13:26:20
SPlROVlTSP-10
I
I
95%
I
CAPS
I
STOP=Stop measurement
&-J
PRE
FVC 4.22 FEV1 1.64
1
PEF
*MEAS3 MEAS2
4.26
1.83
1.73 1.87 2.21 9.84
12-
12-
10-
10-
8-
8-
6-
6-
4-
4-
2-
2-
or,
or,
4-
4-
6-
6-
-
-
-8-
-8-
""'I"""
I
MEASl PRED.
4.64 5.47
1.90 4.49
ci,
.
.
F(v)
. . .
',
.. ..
.."
j
4
We
26.APR.95 13:26:20
SPlROVlTSP-10
I
95%
I
CAPS
2.16
-1
-1
0-
0-
Page 38
FORCED VITAL CAPACITY (FVC) TEST - EXHALATION AND INHALA TlON
Display Results
RESULTS
To
display results of the test, press
PRE
FVC FEV1 1.64
PEF
ME AS^
4.22 4.26
1.73
ME AS^
1.83 1.90
1.87 2.21
MEASI
4.64 5.47
0
...........
PRED.
j j
and the following information is displayed:
We
26.APR.95 13:26:20
SPlROVlT SP-10
4.49
9.84
I
Chapter
Taking Tests
I
95%
I
CAPS
2
FVC
FlVC FlVl
FIVl/FIVC FlVl/FVC
PIF
Press
I"J.
............
PRED. M EAS 1' MEAS2 MEAS3 YoPRED
4.64
1.90
.O
41
2.02
1.35
.oo
1
3.86 4.08 4.40
2.72 1.46 1.52
70.6 35.9 35.5
58.7 34.4 36.0
3.33 1.69 2.12
1.65
1.08
1.87
1.84
1.74
i
120
1.29
1.66
1.41
0.94
1.73
1.63
1.60
1.06
1.49
ESC
.........................
5.47
4.49
83.3
8.23
4.75
1.40
9.84
8.47
5.74
2.36
---
---
---
---
---
j
toexit.
I
85 42 49 24 28 72 22 22 22 50
__-
---
__-
---
---
---
SCHILL
Insert I Edit Diagnosis
RESULTS
0
...........
Spirometer - User Guide
ER
Press the
Press
SP-
10
key when in the results menu once more
or[
ESC
.........................
1'
i
to
exit.
to
insert
or
edit the diagnosis.
2.17
Page 39
Chapter
Taking Tests
2
SLOW VITAL CAPACITY (SVC) TEST
For this test the patient should breathe normally 3 times and then inhale maximally to total lung capacity and then expire maximally.
so,
Inform the patient
To be able to carry out a Slow Vital Capacity (SVC) test, first press the key.
As soon as the basic screen is displayed, check in the top part of the screen according to the requested setup.
Carry
Out Test
NOTE:
T
device for at least
that it is sure that he understands what
if
all the necessary information is displayed
he flow sensor must be held quite still and no air should be breathed into the
one
second before and after the START key is pressed.
is
required of him.
Press the green
The message “READY FOR MEASUREMENT” appears on the display. As soon as the patient starts to breathe into the flow sensor, the unit begins to record corresponding curve is represented on the display. An acoustic “beep“ indicates when the break-off point for the measurement has been reached.
Five seconds later, the measurement will be automatically interrupted unless the patient carries on. Measurements will still take place should this break-off point not be reached. Once the test iscompleted, the measurement results are calculated and given on the display.
Interrupt I End Test
The test can be interrupted at any time by pressing
c,R.T.j
key to start the test.
[
STOP
..........................
1
.
the
flow and the
2.18
Page 40
Chapter
2
Taking Tests
SLOW VITAL CAPACITY (SVC)
Delete Measurements
If
one
of
the result columns should be deleted, proceed as follows:
MEAS.
I"'"")
Press the number of that column (e.g. for column 1, press key
the results should be deleted, i.e. all three columns, press key
Screen Information
When subsequent tests are made, the result with the highest SVC + ERVvalue will always besavedandgiven inthe right-handcolumn (MEAS value of the tests already made and stored will be overwritten. An asterisk
measurement results of the last test made.
TEST
1
Delete MEAS
2 Delete MEAS 2
3
Delete MEAS
4
Delete ALL meas.
1)
that is to be deleted.
4.
l).TheresultwiththelowestSVC+ERV
1
3
("*'I)
indicates the
NOTE: The marked results correspond to the displayed test curve.
We
PRE
svc
ERV 0.94
ME AS^
4.81
1.89 1.75 1.44
ME AS^
4.83 4.85
0.94 1.43
MEASI
PRED.
26.APR.95 13:26:20
SPlROVlTSP-10
If
all
I
95%
I
CAPS
12
10
-
-
8-
6-
SCHILLER
SP-10
Spirometer - User Guide
2.19
Page 41
Chapter
Taking Tests
2
SLOW
Display Results
To
PRE
svc
E
RV
I
RV
FVC FEV1 FEVlIFVC FEF.2-1.2 FEF25-75% FEF75-85% PEF MEF75Yo MEF5OYo MEF25% FMFT
FlVC
FlVl
FIVlIFIVC FIV11FVC PIF MIF5OYo
VITAL CAPACITY(SVC)
display results of the test, press
ME AS^
4.81 4.83
0.94 0.94 1 .a9 1.75
ME AS^
TEST
RESULTS!
0
..........
i
and the following information is displayed:
We
MEASI
4.85
1.43
1.44
PRED. MEAS1 MEASB MEASB %PRED
5.47
4.49
83.3
8.23
4.75
1.40
9.84
8.47
5.74
2.36
---
PRED.
4.26
1.83
43.1
1.67
1.65
1
.oo
2.21
1.86
1.29
1.19
1.73
4.85
1.43
1.44
1.98 2.15 1.97
1.08
1.87
1.84
1.74
1.20
1.29
4.08
1.46
35.9
34.4
1.69
1.68
4.83 4.81
0.94 0.94
1.75 1 .a9
26.APR.95 13:26:20
SPlROVlTSP-10
I
1.41
0.94
1.73
1.63
1.60
1.06
1.49
I
I
85 42 49 24 28 72
22
---
---
---
---
95%
CAPS
1
pNcl-J
Press
............
Insert / Edit Diagnosis
Press the
c"Jr
Press
. ._
...........
or(
ESC
...........................
.,.j
key when in the results menu once more
orpsc
.........................
f
toexit.
1'
j
to
exit.
to
insert or edit the diagnosis.
2.20
Page 42
Chapter
2
Taking Tests
EXPIRED
For this test the patient should breathe as normally
for at least
Inform the patient
To be able to carry out atest for Expired or Minute Ventilation (MV), first press the
key.
As
soon as the basic screen is displayed, check
in the top part of the screen according to the requested setup.
Carry Out Test
OR
20
seconds.
MINUTE VENTILATION (MV) TEST
as
possiblefor up to
so,
that it is sure that he understands what is required of him.
if
all the necessary information
60
seconds, but
L3
. . .. ....
is
displayed
NOTE: The flow sensor must be heldquite stillandno air shouldbe breathedinto the device
for at least one second before and after the START key
Press the green key to start the test.
is
pressed.
The message "READY patient starts to breath into the flow sensor, the unit begins corresponding curve is represented on the display. An acoustic "beep" indicates when the break-off point for the measurement has been reached.
Five seconds later, the measurement will be automatically interrupted unless the patient carries on. Measurements Oncethe test iscompleted, the measurement results arecalculated and given on the display.
I
End
Interrupt
The test can be interrupted at any time by pressing
Delete Measurements If one of the result columns should be deleted, proceed as follows:
Press the number of that column (e.g. for column
the results should be deleted, i.e. all three columns, press key
Test
DELETE
MEAS.
0
FOR
MEASUREMENT" appears on the display. As soon as the
to
record the flow and the
will
still take place should this break-off point not be reached.
[sToP)
...........................
1,
press key
.
1
Delete MEAS
2
Delete MEAS
3
Delete MEAS
4 Delete ALL
1)
that is
4.
1 2
3
meas.
to
be deleted. If all
SCHILLER
SP-10
Spirometer - User Guide
2.2
I
Page 43
Chapter
Taking Tests
2
EXPIRED OR MINUTE VENTILATION (MV) TEST
Screen Information
When subsequent tests are made, the result with the highest besavedandgiveninthe right-handcolurnn (MEAS value
of
the tests already made and stored will be overwritten. An asterisk
measurement results
NOTE: The marked results correspond
PRE
MV 43.46 43.80 48.27
MEAS3 MEAS 2 *MEAS 1 PRED.
19.05
2.28 2.19
10-
8-
of
the last test made.
19.97 17.99
2.60
to
the displayed test curve.
l).Theresultwiththelowest
-
6-
SVC
+
ERV
value will always
("*")
We
26.APR.95 13:26:20
SPIROVITSP-10
SVC+
ERV
indicates the
I
95%
I
f7AP.S
I
STOP=stop
measurement
2.22
Page 44
EXPIRED OR MINUTE VENTILATION (MV) TEST
Chapter
Taking Tests
2
Display
To
PRE
MV 43.46
RR
Tv
FVC FEVl FEV1 IFVC FEF.2-1.2 FEF25-75% FEF75-85% PEF
MEF75%
MEF5OYo MEF25Yo FM
FT
FlVC FlVl FIVl/FIVC FIV1 IFVC PIF
svc
E
RV
I
RV
Tv
Results
display results of the test, press
MEAS3
19.05
2.28
MEAS2 *MEAS1 PRED.
43.80 48.27
19.97 17.99
2.19 2.68
RESULTS
j
0
............:
and the following information is displayed:
We
26.APR.95 13:26:20
SPlROVlT SP-10
I
I
I
PRED. M EAS 1 MEAS2 MEAS3 YoPRED
5.47 4.64
4.49 1.90
83.3 41
~
~..
8.23
4.75 1.35
1.40 1
9.84 2.21
8.47 1.86
5.74 1.29
2.36 1.19
-__
---
---
---
---
---
---
---
---
---
.O
2.02
.oo
1.73 1.29 1.49
3.86
2.72
70.6 35.9 35.5
58.7 34.4 36.0
3.33 1.69 2.12
2.99
4.85 4.83 4.81
1.43 0.94 0.94
1.44 1.75 1.89
1.98 2.15 1.97
4.26 4.22 85
1.83 1.64 42
43.1
1.67 1.66 24
1.65 1.41 28
1.08 0.94 72
1.87 1.73 22
1.84 1.63 22
1.74 1.60 22
1.20 1.06 50
4.08 4.40
1.46 1.52
1.68 2.05
38.9 49
---
---
---
---
---
---
---
--_
---
_--
---
95%
CAPS
I
.-
SCHILLER
Press
Insert
I
Press the
Press
SP-10
Spirometer - User Guide
orl
...........
Edit
Diagnosis
RESULTS
0
or[-Tz----]
............
ESC
..........................
key when in the results menu once more to insert or edit the diagnosis.
.........................
1’
to
.:
to exit.
exit.
2.23
Page 45
Chapter Taking Tests
2
MAXIMUM VOLUNTARY VENTILATION (MVV) TEST
For this test the patient should breathe as deeply and as rapidly as possible over a period
of
6
to
12
seconds.
so,
Inform the patient
that it is sure that he understands what is required of him.
WARNING:
EXTREME CARE SHOULD BE EXERCISED WHEN
THERE
To be able
the
As soon as the basic screen is displayed, check in the top part
Carry Out Test
NOTE: The flowsensormustbe heldquite stillandno airshouldbe breathedinto the device
The message "READY patient starts corresponding curve is represented on the display. An acoustic "beep" indicates when the break-off point for the measurement has been reached.
IS
A DANGER OF HYPERVENTILATION.
to
carry out a test for Maximum Voluntary Ventilation (MVV), first press
MVV
...........................
for
Press the green
J'
key.
of
the screen according
at
least one second before and after the START key is pressed.
[
START
..........................
FOR
to
breathe into the flow sensor, the unit begins
1
MEASUREMENT" appears
if
all the necessary information is displayed
to
the requested setup.
:
key
to
start the test.
PERFORMING THIS TEST AS
on
the display. As soon as the
to
record the flow and the
Five seconds later, the measurement will be automatically interrupted unless the patient carries on. Measurements will still take place should this break-off point not be reached. Oncethe test is completed, the measurement results arecalculated and given on the display.
When subsequent tests are made, the result with the highest MVVvalue will always besaved
1).
and given in the right-hand column (MEAS
tests already made and stored will be overwritten. An asterisk
ment results of the last test made.
Interrupt / End Test
The test can be interrupted at any time by pressing
The result with the lowest MW value of the
r")
indicates the measure-
m.
2.24
Page 46
MAXIMUM VOLUNTARY VENTILATION (MVV) TEST
Delete Measurements
If one
of
the result columns should be deleted, proceed as follows:
1
Delete MEAS
2
Delete MEAS
3
Delete MEAS
4
Delete
ALL
meas.
Chapter
2
Taking Tests
1
2
3
Press the number
the results should be deleted,
Screen Information
of
that column (e.g. for column 1, press key 1) that
i.e.
all three columns, press key
4.
is
to be deleted. If all
When subsequent tests are made, the result with the highest SVC + ERV value will always besavedandgiveninthe right-handcolumn (MEASl).TheresuIt withthelowest SVC+ ERV value
of
the tests already made and stored will be overwritten. An asterisk
("*")
indicates the
measurement results of the last test made.
NOTE:
PRE
MVV
RR
Tv
10-
8-
The marked results correspond to the displayed test curve.
MEAS
3
MEAS2 *MEAS 1 PRED.
69.5
21.8
3.19
69.6
21.1
3.30
81.7
24.0
3.41
150.8
We
26.APR.95
SPlROVlTSP-10
13:26:20
I
95%
1
CAPS
SCHILL
ER
6-
SP-
10
Spirometer - User Guide
2.25
Page 47
Chapter
Taking
2
Tests
MAXIMUM VOLUNTARY VENTILATION (MVV)
Display
To
PRE
MVV 69.5 69.6 81.7 150.8
RR
FVC
FEVl
FEVlIFVC FEF.2-1.2 FEF25-75Yo FEF75-85% PEF MEF75Yo MEF50Yo MEF257o FMFT
FlVC FIV1 FIVlIFIVC
FIVlIFVC PIF MIF50%
Results
display results of the test, press
MEAS3 MEAS2 *MEAS1 PRED.
21
.8
21.1 24.0
1: I
1%:
11s
,
11s
,
:I/S
,
11s
:
,
11s.
11s
,
:
I/s
,
1s.
I:
/J0:
YO.
1
k:
5.47 4.64
4.49 1.90 1.83
83.3 41
8.23 2.02 1.67 1.66 24
4.75 1.35 1.65
1.40 1
9.84 2.21
8.47 1.86 1.84 1.63 22
5.74 1.29 1.74 1.60 22
2.36 1.19 1.20 1.06
RESULTS
El
...........I
---
---
---
---
---
---
__-
I
and the following information is displayed:
4.26 4.22 85
.O
.oo
1.73 1.29 1.49
3.86 4.08 4.40
2.72 1.46 1.52
70.6 35.9 35.5
58.7 34.4 36.0
3.33 1.69 2.12
2.99 1.68 2.05
43.1 38.9 49
1.08 0.94 72
1
TEST
We
26.APR.95 13:26:20
SPIROVIT SP-10
1.64 42
1.41 28
.a7 1.73
I
95%
I
CAPS
22
50
---
---
---
---
---
---
---
pmJ
Press
..............
Insert
I
Press the
rl,
pcrJ
Press
...........
or[
ESC
.........................
Edit Diagnosis
RESULTS
01:
key when in the results menu once more to insert
or[
ESC
.........................
)
..
to exit.
,
i’
I
to exit.
or
edit the diagnosis.
2.26
Page 48
POS
T-ME
DlCA
TlON TESTS
Chapter
2
Taking Tests
Each time the Pulmonary Function Testing program premeditation tests are top left-hand corner of the screen.
The post-medication function is selectable for the following tests:
FVC Test
SVC
Test
MV
Test
MW
Test
Select Post-Medication
In order to carry out post-medication tests for comparison, simply press
"POST" appears in the top left-hand corner as in the following example:
to
POST
FVC FEVl PEF
be carried out. This is confirmed by "PRE" which appears in the
MEAS 3 MEAS 2 MEAS 1 PRED.
is
called up, the unit assumes that
5.47
4.49
9.84
~REIPOST)
............................
and
NOTE
The post-medication tests are carried out in the same way as the premeditation tests.
Print Post-Medication Tests
The printout following post-medication tests will give the curves medication tests (the premedication curve
The measurement results are shown as the actual results, actual results as a percentage those predicted, and the percentage change (i.e. difference) from pre
results.
is
bold).
of
both pre and post-
to
post-medication
of
NOTE: The premedication results used forthe comparison printout are those with the highest
values, and not necessarily the last taken.
The diagnosis resulting from the premedication test
is
also given on this printout.
SCHILLER
SP-10
Spirometer - User Guide
2.27
Page 49
Chapter
Taking
2
Tests
PRINTING TEST
By pressing either
0
With only selected results are printed.
The patient data, the time and date and the standard used for calculation are printed out for every test but the type of curve and test results will depend on the type of test.
A
printout can be interrupted at any time by pressing
CURVES
the best result is printed.
all results are printed.
AND RESULTS
,a printout is made according to theselection:
m.
2.28
Page 50
RESISTANCE TEST OPTION
.
Resistance measurement enables the early detection of an obstruction in the respiratory
tract through measurement of the lung pressure necessary for exhalation.
Test Procedure
We
26.APR.95 13:26:20
RESISTANCE
TEST
SPlROVlTSP-10
.........................................
Taking Tests
I
95%
I
CAPS
Chapter
2
10-
8-
6-
4-
‘1
..
.
rn
.
... . ....
.
..
.
.....
.
.
........
. .
.. . .........
.....
.........
.......
.
.
..
........
.
..
.
.. .. ..
.....
...
.. ..
.....
..
.....
. .
.......
.
...
...
...
.
......
. . . . . . . . . . . . . .
...
.
STOPstop measurement
The display indicates that an average resistancevalue will be measured between 0 and
of
the expired volume of air. To the right, the actual measured resistance will be given as a pressure value in cm curve will be drawn with the expiratory flow on the vertical axis and the time in seconds on
the horizontal axis.
H,O/
litrelsecond. The coordinates represent the graph on which the
85%
SCHILLER
The patient should inhale as deeply as possible before starting the test and then exhale through the flow sensor in such a way that the curve on the display remains between the two
horizontal markers (i.e. between
the patient achieves the necessary flow of exhaled air
and that he understands what is required of him.
N0TE:At regular intervals during exhalation, the air flow will be interrupted for very short
periods. If the air flow is too strong
and the measurement will be invalid.
SP-10
Spirometer - User Guide
1
and
2.5
litres). This may require several trial runs before
so
make sure that he is aware
(>2.5
1s)
then the shutter could become blocked
of
this
2.29
Page 51
Chapter
Taking Tests
2
RESISTANCE TEST OPTION
Print Resistance Tests
gives the following:
Patent data
Norm values
Diagnosis (as indication) (if selected under "Settings")
Resistance curve as shown on the display.
Measured pressure as cmH,O / litres / second.
Where a VC test has also been carried out, the ERV and EVvalues appear together with the
resistance value
of
the expired volume at ERV.
2.30
Page 52
Chapter
Diagnosis and Norm Values
3
Diagnosis and
EXPLANATION DIAGNOSIS
Diagnosis for Countries Outside the USA Diagnosis for USA and Canada
NORM VALUES
Norm Values for Countries Outside the USA
ECCS Values Quanjer & Tammeling Comparisons Austrian Standard Values (C)sterreich)
Swedish Standards (Berglund)
Finnish Standards
Indian Equations
Norm Values for USA and Canada
Morris Norm Values
Crapo Norm Values
Knudson Norm Values
Knudson
Composite Norm Values
Polgar Norm Values
Race Influences on Norm Values
OF
MEASURED VALUES
.....................................................
.....................................................................................................
.................................................
................................................................
...............................................................................................
............................................
.............................................................................................
.........................................................
.....................................................
.....................................................................................
........................................................................................
...................................................................................
.................................................................................
.............................................................................
76
Norm Values
................................................................................
.................................................................
...........................................................
.......................................................................
.........................................................................
............................................................
Norm
3.3
3.5
3.5
3.5
3.6
3.6
3.6
3.7
3.7
3.8
3.8
3.9
3.9
3.9
3.10
3.11
3.12
3.12
3.13
3.14
Values
SCHILLER
SP-10
Spirometer . User Guide
3
.
I
Page 53
Chapter Diagnosis and Norm
3
Values
3.2
Page 54
Chapter
Diagnosis and Norm Values
3
Explanation
FVC
~ ~
FEV*S,.OI3.0
FEF
FEF25-75%
of
Measured
Values
Forced vital capacity expiration
possible exhalation after a maximal inhalation.
Forced expiratory volume lung
1
.O
or 3 seconds forced expiration.
Forced Expiratory Flow respiratory
volumes measured in litres per second. flow speed of the expired air by 25
(FW
flow speed
of
the expired air by
(FW
FEF25%
FEF,O% FEF,,,
FEF.2-1
.z
flow speed of the expired air by 25% of the forced vital capacity flow speed of the expired air by flow speed of the expired air by 75% of the forced vital capacity
averaged flow between 0.2 and 1.2 litres of the forced expired vital
capacity
PEF Peak Expiratory Flow
volume achieved by the quickest
volume in litres, measured after
flow in terms of differing lung
to
75% of the forced vital capacity
75
to
85%
of the forced vital capacity
50%
of
the forced vital capacity
0.5,
(NC)
(FVC)
(FVC)
MEF Maximum Expiratory Flow
flow
speed
of
MEF,,,
MEF50, MEF25,
MEF,,, MEF,,,
ERV
IRV
TV
svc
MV
= =
FEF,,,
FEF,,,
flow
speed of the expired air by 50%
flow speed of the expired air by 75%
Expiratory Reserve Volume possible
from the normal expiration level
lnspiratory Reserve Volume possible
from the normal inspiration level
Tidal Volume expirational
respiration
Slow
following a deep inspiration
Expired or Minute Ventilation volume
minute measured over a minimum of one minute
the expired air by 25% of the forced vital capacity
Vital Capacity lung
volume measured from a complete expiration
(NC)
of
the forced vital capacity
of
the forced vital capacity
further expiration starting
further inspiration starting
and inspirational volumes during normal
of
expired air in litres per
(FVC) (FVC)
SCHlLLER
SP-10
Spirometer - User
Guide
3.3
Page 55
Chapter
3
Diagnosis and Norm
Values
Explanation
MVV
RR
~
FMFT
FlVC
1.0
F'V FIV
l,o
/
FIVC
l,o
I
FIV
PIF
MIF
FVC
of
Measured
Values
Maximum Voluntary Ventilation maximumvolume of air which can
be moved on expiration while breathing as deeply and as rapidly as
possible Respiration Rate
-~
Forced Mid-expiratory Flow Time time difference between the
and
75%
points of the
FVC
Forced lnspiratory Vital Capacity inspiration volume achieved between a maximal expiration and a maximal inspiration
forced inspiratory air volume in litres measured in the first second forced inspiratory air volume measured in the first second as a
of
percentage
forced inspiratory vital capacity
forced inspiratory air volume measured in the first second as a percentage of forced expiratory vital capacity
Peak lnspiratory Flow maximum inspiratory flow speed in litres
second
Maximum lnspiratory Flow maximum inspiratory flow in litres
25%
/
50%
flow speed by
50
%
of the forced inspiratory vital capacity
3.4
Page 56
Diagnosis and Norm Values
Diagnosis
The diagnostic interpretation is dependent upon the country.
The factors used in the evaluation for diagnosis are automatically included in the respective language software and described in the following chapters.
Chapter
3
Diagnosis for Countries Outside the
Possible respiratory problems are diagnosed
USA
on
evaluation of the following factors:
Normal Condition Restrictive Obstructive Combined
Restrictive Ventilatory Defect is indicated
Obstructive Ventilatory Defect is indicated
If
both conditionsare satisfied, a Combined Obstructive/Restrictive Ventilatory Defect is
if
-C
predlde
if
x
100%
FEVI
SVC
x
100% is smaller than
is
smaller than
80%.
70%.
indicated.
Diagnosis for
For the
USA
and Canada
USA
and Canada, diagnosis of possible respiratory problems is based on the
ITS
interpretation standard which uses the LLN (Lower Limits of Normal) calculations. These calculations apply
to
patients between the ages of 5 and
85.
The LLN FEV1% value is
calculated as follows:
Adolescent Child
SCHILLER
19
to
25
Predicted FEV1%
yrs: Predicted FEVl
c18 vrs: Predicted FEV1%
x
0.848
x
0.826
19
to
20 yrs:
c18
yrs: Predicted FEV1% x 0.791
Using this LLN value, the diagnostic statements and their criteria are as follows:
SP-10
Spirometer - User Guide
Nonnal limits
Borderline
Obstluction Mild Obstruction Moderate Obstruction
Severe Obstruction Mild Restriction
Moderate Restriction
Severe Restriction
%FEVpredicted is ~80%
FEVl
is
80
to
100%
of
LLN
FEVl
is
60
to
80%
of
LLN
is
40
to
60%
of
of
LLN
is
is
is
LLN
60
to
80%
50
to
60%
~50%
FEVl FEVl
is
~40%
%FVCpredicted %FVCpredicted %FVCpredicted
%
x
0.805
3.5
Page 57
Chapter
3
Diagnosis
and Norm Values
Norm
The norm values used for the calculation
Values
of
predicted values are dependent upon the
country.
For Great Britain, Italy, Spain and Switzerland, the ECCS and Quanjer standards are used.
For Sweden, the Swedish (Berglund) and Quanjer standards are used. In Finland the Finnish and Quanjer standards are used. In Austria the Austrian standards are used.
0
In India the Indian norm values are used.
0
In America and Canada the norm values that are used are Knudson, Knudson76, Crapo, Morris, Composite and Polgar. The American normvalues are extended with values taken from the ITS (Intermountain Thoracic Society) recommendations.
The factors used in the evaluation for diagnosis and the specific norm values are included
in the software and are described in the following pages.
IMPORTANT:
DUE TO GREAT DIFFERENCES IN THE SIZE OF THE LUNGS OF CHILDREN, THERE ARE NO STANDARD VALUES
Norm Values
ECCS
Values
for
Countries Outside the
FOR
CHILDREN UNDER 6 YEARS OF AGE.
USA
The safety standards of the European Coal and Steel Community Standards (ECCS) are valid for adults of at least calculated on the basis
svc
FVC
FEVl
FEVlISVC MEF PEF
MEFi’5 MEF5O
of
6.103
5.757
4.301
-0.179
1.944
6.146
5.459
3.794
25
years of age. Patients between the ages of
a
25
year old. The calculation equations are as follows:
x
H
-0.028 x A -4.654
x
H
-0.026
x
A
-4.345
x
H
-0.029
x
A
-2.492
x
A
+87.21 -0.192
x
H
-0.043
x
A
+2.699
x
H
-0.043
x
A
+0.154
x
H
-0.029
x
A
-0.470
x
H
-0.031
x
A
-0.352 2.450
4.664 X H -0.024 X A -3.284
4.426
x
H
-0.026 x A -2.887
3.953 x H -0.025 x A -2.604
x
A
49.10
1.252
x
H
-0.034 x A +2.924
5.50
x
H
-0.030
x
A
3.218
x
H
-0.025
x
A
x
H
-0.025 x A +1.156
-1.106 +1.596
18
and
25
are
3.6
H: Height in meters
A:
Age
Page 58
Diagnosis
Chapter
3
and Norm Values
Norm
Values
Quanjer & Tammeling Comparisons
The Quanjer and Tammeling comparison is valid for children between the ages of 6 and
as follows:
H:
SVC = FVC FEV1 FEVIISVC
lMEF= PEF 18.2x H
MEF50 5.6 x H -4.4
Height
in
meters
1
.OO
0.84
0.84
x
H(power 2.7)
x
H(power 2.7)
-6.8
0.95 x H(power 2.7)
0.81 x H(power 2.7)
0.84
6.6 x H -5.3
4.6 x H -3.3
Austrian Standard Values (Osterreich)
The Austrian standard values are between the ages of
In(FVC) = -1.142 + 1.2591-1 + 0.004070A dW In(FEV1) =-1.178
In(PEF) ln(PEF75) In(PEF50)
ln(PEF25)=-1.576+ 1.166H +0.0219A+O.O021W FEVl %FVC
19
and
=
-0.214 + 0.921H + 0.0467A + 0.002OW
=
-0.077 + 0.770H + 0.0373A + 0.0025W
=
-0.522
=
101.99 - 1.191 H(power 2) - 3.9621n(A)
valid
for
children between
70
are
as
follows:
BOYS 5
+
+
-
17.99
years
(1.09
1.221H + 0.003841A.\IW
0.843H + 0.0300A + 0.0035W
-
1.96m)
the
ages of 7 and
1
8
and adults
17
H W
=
Height
=
weight
..
In(FEV1) = -3.877 + 3.9809dH + 0.1485dA - 1.322Fi In(PEF)
=
0.41
1
+
1.7931n(H) + 0.42511n(A) - 0.91OFi
=
ln(MEF75)
I
InlMEFSOI = 0.256 + 1.6431n(HI + 0.34811nlA) - 1.089Fi
FVC = -1 1.606 + 8.172H - 0.0339A x H + 1.2869In(A) FVC1 dPEF dPEF75 dPEF5O = 1.490 + 1.2901n(H) + 0.0125A - 0.000218A(power 2) dPEF25
FEVl %FVC
0.455 + 1.6161n(H) + 0.37381n(A) - 0.861Fi
=
-8.125 + 6.212H - 0.0300A x H + 0.97701n(A)
=
1.798 + 2.31 1 In(H) + 0.0159A - 0.000248A(power 2)
=
1.581 + 1.8541n(H) + 0.0213A - 0.000283A(power2)
=
1.314 + 0.8981n(H) - 0.0083A - 0.000026A(power 2)
=
101.99 - l.lSlH(power2) - 3.9621n(A)
..
FVCl = -6.995 + 5.174H - 0.0314A x H + 1.02511n(A)
=
1.832 + 1.8381n(H) + 0.0078A - 0.000172A(power 2)
dPEF dPEF75
dPEF50 dPEF25
FEV1 %FVC
in
meters
in
=
1.779 + 1.4211n(H) + 0.0096A - 0.000179A(power2)
=
1.561 + 1.1771n(H) + 0.0045A - 0.000140A(power2)
=
1.372 + 0.9381n(H) - 0.01 52A + 0.000036A(power 2)
=
118.993 - 3.0320H(power 2) - 6.90531n(A)
A
=
Age
in
years
Fi
=
kg
body fat index
=
H/qW
SCHILLER
SP-10
Spirometer - User Guide
3.7
Page 59
Chapter
3
Diagnosis and
Norm
Values
Norm
Values
Swedish Standards (Berglund)
The Swedish (Berglund) standard is valid for adults between the ages of follows:
FEV%
91.79 -(0.373
SVC
1.09 [ (4.81
FEV 1.01.09 [ (3.44
x
x
H)
A)
x
-(0.020
H)
-(0.033
92.1 1 -(0.261
x
A)
-2.811
x
A)
-1.001
1.09 [ (4.04
1.09 [ (2.67
x
x
H)
X
H)
A)
-(0.022
-(0.027
A: Age H: Height in meters
Finnish Standards
The Finnish standard is valid for adults from the age of
18
years
as
follows:
18
x
A)
x A)
and
-2.351
-0.541
75
years as
svc
(I)
FEVl
(I)
FVC
(I)
FEVl ISVC(%)
FEVl
I
FVC
(“A)
PEF
(11s)
svc
(I)
FEVl
(I)
I
FVC
(1)
FEVl ISVC FEVl I FVC
I
PEF
(%)
(“A)
(11s)
A: Age
H: Height in meters log: Logarithm to base
exp exp
exp exp exp exp
I
exp
I
exp
I
exp
I
exp
I
exp
I
exp
10
[(-0.00833 [(-0.00587
[(-0,00827 [(0.00246 [(0.00240 [(-0.00211
[(-0.01016 [(-0.00920 [(-0.00982
[(0.00096 [(0.00062 [(-0,00677
x
A)
+
(0.6309
x
log
A)
+
(-1.4750
x
A)
+
(0.2756
x
log
A)
+
(-1.1655
x
A)
+
(0.5860
x
log
A)
+
(-1.4468
x
A)
+
(-0.3553
x
A)
+
(-0.3104
x
A)
+
x
A)
+
x
A)
+
x
A)
+
x
A)
+
(-0.2223
x
A)
+
(-0.1586
x
A)
+
(0.1049
(0.6995 (0.4772 (0.6358
(0.4017
x
log
A)
+
(0.3095
x
log
A)
+
(0.2813
x
log
A)
+
(-0.6774
x
log
A)
+
(-1.4518
x
log
A)
+
(-1.3284
x
log
A)
+
(-1.4137
x
log
A)
+
(0.1233
x
log
A)
+
(0.0853
x
log
A)
+
(-0.7422
I
H)
+
0.90471
I
H)
+
1.09801
I
H)
+
0.94611
I
H)
+
2.19331
I
H)
+
2.15191
I
H)
+
1.32551
I
H)
+
0.77631
I
H)
+
0.92961
I
H)
+
0.83201
I
H)
+
2.15331
I
H)
+
2.09751
I
H)
+
0.96611
1
I
Page 60
Chapter
Diagnosis and Norm Values
3
Norm
Indian Equations
The Indian equations are valid for patients from the age
FVC FEVl
FVC FEV1
FEVl IFVC
svc
FEV3 FEF25-75% PEF FEF5O FEF75
MW
Values
<30
years
old:
0.055 x H + 0.019 x A -6.058
0.039
x
H
-
0.010 x A -3.266
230
years
old:
0.054 x H - 0.01
0.037
x
-0.1756~ H -0.2457xA-119.346
0.0522
x
0.0485
x
0.0173
x
0.0850 x H - 0.0187 x A -6.2083
I
0.0195
x
I
0.0088
x
I I
I
1.3056
x
8 x A -4.832
H
-
0.022 x A -2.650
H
-
0.01 14 x A -4.859
H
-0.0183
H
-
0.0407 x A -1.6108
H
-
0.0365 x A -1.7383
H
-
0.0301 x A -1.0402
H
-
0.5228
x
A
x
A
-4.138
-93.2102
of
7 years as follows:
I
0.030 x H + 0.006 x A - 2.284
1
0.025
x
H
-
0.01 1 x
A
-
A
x
x
x
1.424
-
3.755
A A
A
-
5.927
-
5.660
-
25.0208
I
10.043 x H - 0.010 x
10.032 x
10.0272
10.0113
10.7149 X H - 0.3624
H
-
0.012 x A - 2.580
-0.0334 x H - 0.2146 x A - 94.8867
0.0587
x
H
-
0.0296
0.0533 X H - 0.01 05 0,0245 X H - 0.0336 x A - 0.1399
0.0497XH-0.0018~ A-2.7154
x
H
-
0.0279 x A - 0.2704
x
H
-
0.0288 x A - 0.5012
A:
Age
H: Height in meters
Norm Values for
USA
and Canada
Morris Norm Values
The Morris equations are valid for women between
range
of
20 to
90
of 20 to
F
FEVlIFVC FEF0.2 - 1.2
FEF75 - 85
years, and for men between
90
years as follows:
0.0920
-0.31 18
0.1090
0.0470
0.0130
x
H
-0.0320 x A -1.260
X
H
-0.2422
x
H
-0.0470
x
x
H
-0.0450 x A +2.513
x
H
-0.0230 x A +1.210
X
A
A
+107.120
+2.010
58
The Morris normals are extended with the following:
56
and 72 inches tall and within the age
and
80
inches tall and within the age range
0.0890
x
H
-0.0250 x A -1.932
-0.0679
x
H
-0.1815
XA
+88.700
0.1450 x H -0.0360 x A -2.532
0.0600 x H -0.0300 x A +0.551
0.0250 x H -0.0210 x A +0.321
SCHlLL
FEV3.0 FEV3.OIFVC MVV
A: Age H: Height in inches
ER
SP-
10
Spirometer - User Guide
I
-0.1359
-0.1593
3.4040
x
H
-0.0271
x
x
H
-0.1450
x
x
H
-1.2600 x A -21.400
A A
-3.512 +I
12.090
1-0.1 123 x H -0.0257
-0.2380X H-0.1630xA+118.160
2.0500
x
H
-0.5700
x
x
A
A
-2.745
-5.500
3.9
Page 61
Chapter
Diagnosis and
3
Norm Values
Norm
Crapo
Values
Norm
Values
The Crapo equations are valid for men between 61 and 77 inches tall and within the age
range
of
18
to
89 years, and for women between
range
of
18 to 89 years as follows:
FVC 10.1247 x H -0.0216 x A -3.590 FEV1
FEV3 FEVllFVC FEF25
-
MVV Vol.
10.1524~ H -0.0214xA-4.650
10.1052xH-0.0244xA-2.190 10.0869
0.1359xH-0.0271 XA-3.512
x
H -0.1520 x A +110.490
x
H -0.0380 x A +2.133
x
H -1.2600 x A -21.400
75
-0.3302
0.0518
3.4040
The Crapo normals are extended with the following
57
and 71 inches tall and within the age
x
H -0.0255 x A -1.578
0.1 123 x H -0.0257 x A -2.745
x
H
-0.2520
xA
-0.5131
0.0391
2.0500
ITS
equations:
x
H -0.0460 x A +2.683
X
H
-0.5700
+126.580
X
A -5.500
IFEV3.0lFVC I -0.1593 x H -0.1450 x A +112.090 1-0.2380 x H -0.1630 x A +118.160
A: Age
H:
Height in inches
I
3.10
Page 62
~ ~~ ~
Diagnosis
and
Norm Values
Chapter
3
Norm
Knudson
The Knudson equations are valid for both children and adults
Values
Norm
Values
in
specific groups according
to age and height as follows:
H=44
to
61
inches,
A
=6
to
11
yrs
\
FVC FVC0.5 FEVl FEVllFVC FEF25 - 75
PEF
I
FEF5O
I
FEF75
I
MVV 14.6800
FVC FVC0.5 FEV1 FEVlIFVC FEF25 - 75 PEF
FEMO FEF75 MW
0.1 039 X H
0.0760 X H +0.0430 x A -3.050
0.0884 x H + 0.0 x A - 2.814
-0.2065
0.0859
0.1980
1
0.0960
I
0.0434 X H + 0.0
I
H
=
10.1499
0.0760
0.1318 X H + 0.0636
-0.2065
0.1369 x H + 0.0749
0.1980 x H + 0.1660
0.1379 x H + 0.1 150
0.1008
4.6800
55
+
0.0
x
H
+
0.0
x
H
+
0.0
x
H
+
0.1660
x
H
+
0.0
x
H
+
1.8 x A - 192.320
to
76
inches,
x
H
+
0.0739
x
H
+
0.0430
x
H
+
0.0
X
H
-0.0057
x
H
+
1.8000
x
A -3.376
x
A
x
A - 2.320
x
x
A - 2.545
x
A
-
A
x
x x
x
A
x x
x
x
x
+
100.439
A
-
1.01 5
=
12
A
-
A
-
A
-
+
100.439
A
-
A
-
A
-
A
-
4.242
A
-
8.061
to
25
6.887
3.050
6.1 18
6.199
8.061
6.385
192.320
yrs
H = 42
to
58
inches,
A
=
6
to
10
yrs
0.1 092
x
H
+
0.0 X A -3.749
0.0480 x H + 0.0610 X A -1.740
0.0853
x
H
+
0.0 x A - 2.758
-0.4849
x
H
+
0.6655 x A + 109.974
0.0559
x
H
+
0.0
X
A
-
0.812
0.1240 x H + 0.1570 x A - 3.920
10.0
x
H + 0.1846 x A + 0.736
10.0277 x H + 0.0
12.7600 x H + 3.4000 x A - 108.120
1
H
=
52
to
10.1057 x H + 0.0699 x A - 4.447
0.0480
x
0.0892
x
-0.4849
x
0.0709 x H + 0.1275
0.1240 x H + 0.1570 x A - 3.920
0.0732 X H + 0.11 11 X A - 2.304
2.7600 X H + 3.4000 X A - 108.12
x
A
-
0.1 66
72
inches,
A
=
H
+
0.0610 x A - 1.740
H
+
0.0694 x A - 3.762
H
+
0.6655 x A + 109.974
x
A
1 1
-
2.801
to
20
I I I
yrs
FEV3.0 FEV3.0lFVC
FVC 10.1524xH -0.0214xA-4.650 FVCO.5 FEVl
FEV3.0 FEVlIFVC
FEF25 - 75 10.1471 x H - 0.0363 x A - 4.518
FEF75-85
PEF FEF25
FEF50
FEF75 MVV
I
FEV3.0lFVC 1 -0.1 593
A:
H:
0.1359
X
H
-
-0,1593
x
H
-
I
H = 62
to
77
0.0831 xH-0.0152xA-1.914
0.1052
X
H
-
0.1359
x
H
-
0.0
x
H
-
0.1050 x A + 86.686 -0.4704 X H - 0.1896 X A + 121.678
I
0.0130~H-0.0230~A+ 1.210
0.2390 X H - 0.0350
0.0900
X
H
-
0.1 737 X H - 0.0366 x A - 5.409
0.0787
x
H
-
3.0300
X
H
-
x
H
-
Age
Height in inches
0.0271 x A - 3.512
0.1450
x
A
+
112.090
inches,
A
=
26
0.0244 x A - 2.190
0.0271 x A - 3.512
x
A
-
5.990
0.0200 x A + 2.726
0.0230 x A - 2.483
0.8160 x A - 37.900
0.1450 x A
+
1
0.1 123 x H - 0.0257 X A - 2.745
-0.2380~H-0.1630xA+118.160
I
H
=
58
to
71
inches,
A
=
21
to
-
0.809
-
1.578
+
0.321
-
4.870
120
to
120
yrs
10.1247
x
H
-
0.0216 x A - 3.590
0.0605
x
H
-
0.0185
x
A
0.0869
x
H
-
0.0255
x
A
0.1067
x
H
-
0.0257 x A - 2.745
I
(0.0531
x
H
-
0.0344 xA + 1.128
0.0250
x
H
-
0.0210
x
A
0.1240 x H - 0.0250 X A - 0.740
0.0690 x H - 0.0190 x A + 2.147
0.0681 x H - 0.0289 x A + 0.609
0.0244
x
H
-
0.0259 x A + 1.1 18
2.1400
x
H
-
0.6850
x
A
12.090 1-0.2380 x H - 0.1630 x A + 118.1 60
yrs
I
SCHILLER
SP-I0
Spirometer - User Guide
3.
I
I
Page 63
Chapter
3
Diagnosis and Norm Values
~
Norm
Knudson
The Knudson
Values
76
Norm
76
equations are valid for both males and females in specific age groups as
Values
follows:
IFEF25-75 10.1499xH +0.0xA-5.443 10.0635~ H + 0.121 x A - 0.893
PEF FEF25
FEF50 FEF75
FVC
I
FVCO.5
FEVl
.O
FEV3.0 FEVlIFVC
FEF25 - 75 PEF FEF25 FEF50 FEF75
10.1981
x
H
+
0.166 x A - 8.060
10.1778 x H + 0.147 x A - 7.054
0.1295
x
H
+
0.081
x
A
-
4.975 0.0864
0.0813
x
H
+
0.0 x A - 2.455
Age
225
yrs
0.1651
x
H
-
0.029 x A - 5.459 0.0940
I
0.0940
x
H
-
0.017
x
A
-
2.746 10.0483
0.1321
x
H
-
0.027
x
A
-
4.203
0.1600
x
H
-
0.031 x A - 5.245
-0.2210 x H - 0.140 x A + 103.64
0.1 143 x H - 0.031 x A - 1.864
0.2388
x
H
-
0.035 x A - 5.993
0.2235
x
H
-
0.035 x A - 5.618
0.1753xH-0.015xA-5.400
0.1118xH -0.012xA-4.143
10.1245 x H + 0.157 x A - 3.916
10.1118~
0.0 x H + 0.139
Age
0.0686
0.0889
-0.281 9 x H - 0.109 x A + 107.38
0.0533XH -0.024xA+ 1.171
0.1245 X H - 0.025 x A - 0.735
0.1092 X H - 0.025 x A - 0.1 32
0.0889
0.0
H
+
0.144 x A-3.365
x
H
+
0.120
x
x
A
+
220
yrs
x
H
-
0.022
x
A
x
H
-
0.01 4
x
A
x
H
-
0.021 x A - 0.794
x
H
-
0.023 x A - 1.633
x
H
-
0.013 x A - 0.444
X
H - 0.014 X A + 3.042
A
-
2.531
0.692
-
1.774
-
0.406
I
I
A:
Age
H:
Height in inches
Composite Norm
Selection of the Composite normals provides selected equations taken from other tables as
follows:
Values
FEVl
FEV3
I
svc
Knudson
Crapo
I
Knudson
(same
as
FVC)
I
3.12
Page 64
Chapter
Diagnosis and Norm Values
3
Norm
Polgar Norm Values
The Polgar equations are valid
Values
for
both children and adults in specific groups according
age as follows:
I
FVC FVC0.5 FEVl
.O
FEF25 - 75% PEF
MW
I
FVC
FEVl.0 FEF25 - 75
PEF
FEF50 FEF75
MW
I
FEF3.0 FEF0.2-1.2 FEF25 - 75%
IFEF75 - 85% I H"0.0130 -A*0.0230 +1.210
PEF FEF25
FEF50 FEF75
MW
I
Age under
H(power 3)*0.0000071 +H(power 2) '0.00057 -H*0.0123 +0.14
H*0.076 +A*0.043 -3.05 H(power 3)*0.00000087 +H*0.00035 H(power 3)*0.0000086 +H(power 2)
W0.0086 W0.1109 3.46 H*0.1109 -3.46 H*0.2219 -7.09 H*0.2219 -7.09 H"4.68 -A*l.8 -1 92.32
I
Age
H"0.1499 +A*0.0739 -6.887
H*0.1318 +A*0.0636 -6.118 H'0.1369 +A*0.0749 -6.199
H*0.1980 +A*0.1660 -8.061
H*O.
H"O.1008 -A*0.0057 -4.242 H*4.68 +A*1.8 -192.32
I
Age over26 years
H*0.1090 -A*0.0470 +2.010 H*0.1471 -A'0.0363 -4.518
H*0.2390 -A*0.0350 -5.990 Hn0.0900 -A*O.O220 +2.726
H"0.1737 -A"0.0366 -5.409 H*0.0787 -A*0.0230 -2.483 H"3.03 -A*0.816 -37.9
18
years
+0.1
18
to
25
years
1379 +A'O. 1 150 -6.385
I
Age under
H(power 3)'0.0000076 +H(power 2) *0.00048 +H*0.0112 +0.13
*0.00035 -H*0.0086 +0.1
H"2.76 -A*3.4 -108.12
I
Age
H*0.1057 +A*0.0699 -4.447
H*0.0892 +A'0.0694 -3.762 H*0.0709 +&0.1275 -2.801
H'0.1240 +A*O.1570 -3.920
-
H*0.0732 H*0.0617 +A*0.2923 -4.401 H*2.76 +A*3.4 -108.12
1
Age over20years
W0.1067 -A*0.0257 -2.745
H*0.1450 -A*0.0360 -2.532 H*0.0531 -A*0.0344 +1.128
I
H*0.0250 -A*0.0210 +0.321 H*0.1240 -A*0.0250 -0.740 H*0.0690 -A*O.0190 +2.147
H*0.0681 -A*0.0289 +0.609 H*0.0244 -A'0.0259 +1.118 H2.14 -A*0.685 -4.87
18
to
20
+A*O.
18
years
years
1 1 1 1 -2.304
to
I
I
I
I
SCHILLER
A:
Age
H:
Height in inches
The remaining values are taken from the
SP-I0
Spirometer - User Guide
ITS
equations.
3.13
Page 65
Chapter
Diagnosis
3
and
Norm Values
Norm
Race Influences
According to the setting are (according
.
Values
on
Norm
Values
of
the patients race the predicted values will differ. The differences
to
ITS
recommendations) as follows:
When in the "Settings" menu All values are calculated according When in the "Settings" menu
W (White): values are calculated according to the given formulas
B (Black):
When in the "Settings" menu
C
(Caucasian):
H
(Hispanic):
B
(Black):
85%
values are calculated according to the given formulas
valuesarecalculatedaccordingtothe
85%
NONE
was set:
to
the given formulas.
(WB)
was set:
of the given formulas
(CRVBIA)
of the given formulas
was set:
(=
(=
given formulas (=
lOOo/)
100%)
100%)
A
(Asian):
85%
of the given formulas
3.14
Page 66
Chapter
Data Storage and Transmission
4
Data Storage and Transmission
DATA STORAGE
Store Data in Memory Recall Delete Data from Memory Maximum Memory Capacity
DATA TRANSMISSION
Setups before Sending Data Line Transmission Modem Transmission Enter Phone Number for Modem Transmission Send Data Error Messages
.............................................................................................
...............................................................................
Data from Memory
.........................................................................
.........................................................................
......................................................................
...................................................................................
.....................................................................
.....................................................................................
................................................................................
...............................................................................................
.......................................................................................
........................................
4.3
4.3
4.3
4.3
4.4
4.5
4.6
4.7
4.8
4.9
4.10
4.10
SCHILLER SP-10
Spirometer . User
Guide
4
.
I
Page 67
Chapter Data Storage and Transmission
4
Page 68
Data Storage
Chapter
Data Storage and Transmission
4
Store Data
To store spirometry data in the built-in memory (e.g. for tests with PRWPOST medication),
proceed as follows:
1.
2.
in
Memory
Enter patient data (important for later identification in the memory list). Make tests.
(s,,,,1
3.
After completing the tests, press the
appears as confirmation of the saving.
NOTE: The values
4.
Return to the main menu with
Recall Data from Memory
To recall spirometry data from the memory, proceed as follows:
of
the
FVC,
SVC,
MV
and
w
DATA
'
key. The message "FILE
M
W
tests can be stored.
IS
STORED"
2.
Select file by moving the cursor with the arrow keys to the appropriate position.
3.
Confirm selection by pressing them key.
NOTE: Only the patient data and measurement values can be vie wed on the screen.
4.
The complete test can be printed as described in Chapter 3 under "Printing out Test Curves and Measurement Results".
Delete Data from Memory
To delete spirometty data from the memory, proceed as follows:
2.
Select file by moving the cursor with the arrow keys to the appropriate position.
3.
Confirmselection by pressingthek key.The message"DELETE FILE?
appears.
4.
Confirm with
Y
(yes) or retract with N (no).
(Y/N)"
SCHILL
ER
5.
Return to the main menu with
SP-
10
Spirometer - User Guide
4.3
Page 69
Chapter
4
Data Storage and Transmission
Data
Storage
Maximum
Memory
Capacity
The maximum memory capacity for the SP-10 is as
0
I
60
patients (one test only)
0
or
30
patients
0
or
30
patients
0
or
30
patients
(3 (3
(3
FVC FVC
FVC
13
SVC
I
1
MV
I
3
SVC Pre & Post Test)
I 3 SVC
I 1 MV I 1 MVV Pre & Post Test)
follows:
I 1 MVV Test)
4.4
Page 70
Data Storage and Transmission
Data Transmission
The SCHILLER SP-10 can only send spirometry data to a PC (SEMA). The units can be
connected directly or can communicate over a telephone line by modem. Data transmission is facilitated via one of the standard serial interfaces on the back panel. For shorter distances and with a modem use the longer cables the RS-422 interface (ma. 1'200 metres) is recommended.
Only the spirometry data that is currently in the memory of the SP-10 can be sent.
Direct
data sending
sending over a
In
in line mode called BLOCK. With block transmission the transmitted data is checked
for errors and corrected accordingly. Two types of transmission protocols can be selected in modem mode called CCITT
and BELL. These are standard modem protocols and are selected according
type modem connected.
telephone
spirometry mode only one type of line sending format
RS-232
to
a connected PC (SEMA) is called
line is called
Port 3 interface (ma.
MODEM.
15
metres with
LINE
(to
19'500
Baud). With
transmission and data
SEMA) is
to
be selected
to
the
Chapter
4
Other transmission settings that are applicable for both line and modemtransmission are follows:
Channel
Baud
Parity stop
ALLTHESE SETTINGS ARE LARGELY DEPENDENTON THE PCTHAT THEY MUST BE THE SAME IN BOTH THE TRANSMITTING AND RECEIVING UNITS!
Here the connector is selected for transmission. This sets the speed of transmission -the higher the Baud rate
the quicker the transmission. In good quality lines or short distance lines a high Baud rate can be set. When a transmission errors occura lower Baud rate should be selected.
This is a method of detecting errors in the transmission. This is a method that the hardware uses
and end of sections of the message.
to
determine the start
IS
CONNECTED.
lot
as
of
SCHILLER
SP-10
Spirometer
-
User
Guide
4.5
Page 71
Chapter
4
Data Storage and Transmission
Data
Line Transmission
SEMA is the SCHILLER data management system for diagnostics which allows storage, processing and archiving of pulmonary function tests on
If your
LINE
Channel
Transmission
For the data transfer to
SP-10
is connected directly to a PC with SEMA, proceed as follows:
r
m-.'
Set the following parameters for line transmission:
(2)
RS-232
SEMA
only
'-1
Select line transmission and the format with key Set to: LINE: format = BLOCKS
Use Key Select between
2
a
PC.
use
BLOCKS.
1
---
MODE
SETIING
Parity
Channel Baud
stop
to select the required channel.
RS-232
or
RS-422
RS-232/422
4
:
(normally
- - -
RS-232
9600
NO
1
RS-232).
1.
Baud
(3)
Parity
(4)
N0TE:Checkthat these settings are the same as the PC to which the
Use Key The following values are available:
2400,4800,9600, 19200,38400,76800, 153600.
Key 4 selects the parity. The available choice is either EVEN,
ODD
The length The available choice is either
3
to select the required baud rate.
or NO (normal position NO).
of
the stop bit is determined by pressing key
1, 1.5
75, 150, 300,
or 2 (normal position
SP-10
600,
1200,
5.
1).
is connected.
4.6
Page 72
Chapter
4
Data Storage and Transmission
Data
Transmission
Modem Transmission
If
your
SP-10
is connected over a modem to a PC, proceed as follows:
[srmFx;s)r<
RS-232
>
Set the following parameters for MODEM transmission:
MODEM
Select line transmission and the format with key
The following two formats are selectable:
MODEM: protocol MODEM:
protocol Select in accordance with the protocol of your telephone exchange.
Channel
(2)
Use Key
2
Select between
to select the required channel.
RS-232
Kind
of
tranmission
I
(1):
MODEM:
MODE SETTING RS-232/422
or
RS-422
(normally
-
-
-
-
protocol=
RS-422
RS-232).
9600
NO
CCln
--
1
CCITT
-
1.
BELL
Baud
Parity
(3)
(4)
Use Key 3 to select the required baud rate. The following values are available:
9600, 19200,38400,76800,153600.
75, 150, 300,
Key 4 selects the parity. The available choice is either EVEN,
ODD
or
NO
(normal position NO).
The length of the stop bit is determined by pressing key available choice is either
1, 1.5
or 2 (normal position
Selections 2 to 5 are the same as with line transmission.
N0TE:Check that these settings are the same as the device
connected.
600,
1200, 2400, 4800,
to
which the
5.
The
1).
SP-10
is
SCHILLER
SP-10
Spirometer - User Guide
4.7
Page 73
Chapter Data Storage and Transmission
4
Data
Transmission
Enter
Phone
@-<
. . . . . . .
. . . .
A
typical example
Number for Modem Transmission
To enter the Phone number of the reception unit, proceed as follows:
MODEM Phone-No.:
Phone
of
No.
Modem
a telephone number can be as follows:
>
T'=Tone:
save
PO
with
'P'=Pulse;
ENTER
','=l
sec
00.41.42.3343
Pause
53
Select Pulse Dialling Outside Line
1s
Pause
T-
International Code National Code Telephone Number
Before entering the telephone number the type of dialling must be entered. This is determined by the type of telephone exchange. dialling; older exchanges use pulse dialling.
If
it is necessary to enter a pause in the dialling sequence then enter a comma. The comma gives a one second pause; can be entered. This may be required for example, line. To give the exchange time to set up the connection, it is also a good idea to insert a pause after a national or international code.
Press
ENTER
to confirm the settings and return to the main menu.
if
a longer pause is required, two or more commas
All
modern exchanges now use tone
If
in doubt select pulse.
if
you have to wait for an outside
4.8
Page 74
Chapter
Data Storage and Transmission
4
Data
Send Data
To
NOTE: If stored data in the memory is
then be sent pressing
A
Transmission
send the spirometry data that is currently stored in the memory proceed as follows:
Check that the cable is connected to the Be sure that all the setups are done and correct.
Press the
Error
Messages
If the data can not be sent an error message will be displayed in the status field.
list of possible error messages and their possible causes together with solution possibilities
is given below:
I"'"")
,.
D.ATA
.....
key after the tests are completed.
to
w..
.
. . . . .
. . . . . .
SP-10
(RS-232
be sent, it must first recalled on the screen.
Port
3
or
RS-422).
It
can
SERIAL LINK TIME-OUT This indication appears
unit (after approximately Checkthat the remote unit is switched on and set
parameters.
Check the correct setup in both connected units
(=>
SETTINGS key). Check that the connecting cable is plugged in correctly. Check the integrity of the cable assemblies (on both units).
If
using a modem ensure that it
remote modem.
if
nosignal is receivedfromthe remote
30
seconds).
to
the correct
is
communicating with the
SCHlLL
ER
SP-
10
Spirometer - User Guide
4.9
Page 75
Chapter Data Storage and Transmission
4
Data Transmission
TX-ERROR (PARITY)
TX-ERROR (OVERRUN)
TX-ERROR (FRAMING)
DATA SET NOT READY!
This indicates a parity error.
Check that the parity is set correctly in both receiving and
transmitting units
If there is a parity error it indicates that the data may be faulty. Select other test data and re-transmit.
Carry out the RS This indicates a system timing error. Carry out the RS self test This indicates a transmission error or that the baud rate is not
the same setting in both units. Check that the baud rate is set correctly in both units
(=>
SETTINGS
Check that the connecting cable is plugged in correctly. Check the integrity
If using a modem ensure that it
remote modem. This
is
a general fault indication.
Checkthattheremote unit isswitchedon andsettothecorrect parameters.
(=>
SElTINGS
self
test procedure.
(=>
key).
of
the cable assemblies (on both units).
key).
SETTINGS
is
/
5).
communicating with the
NO DATA IN MEMORY!
Check that the connecting cable is plugged in correctly. Check the integrity of the cable assemblies (on both units).
If using a modem ensure that it is communicating with the remote modem.
A spirometry data transmission has been attempted, but no data
is
stored in the units memory.
Store a spirometry test in the memory and attempt the transmission again. your SCHILLER agent.
If
the same message appears, contact
4.10
Page 76
Chapter
Maintenance. Testing. Troubleshooting
5
Maintenance. Testing.
Trouble Shooting
REPLACING THE RECORDING PAPER CLEANING THE CASING
FLOW
SENSOR
Cleaning and Sterilising FLOW SENSOR SP-150 FLOW SENSOR SP-20 FLOW SENSOR SP-11OIR WITH RESISTANCE
Cleaning the Resistance Sensor SP-11OIR USER CHECKS
Self-Test
Technical Safety Check
USER CHECKS
Test
SP-110
.............................................................................................
of
the Unit
.............................................................................................
Output
Channels (RS-Test)
...............................................................................
.................................................................................
............................................................................
.................................................................................
...................................................................................
.................................................................................
...........................................................................
.......................................................
.............................................................
SENSOR
............................
..............................................
5.10
5.10
5.10
5.11
5.11
5.3
5.3
5.4
5.5
5.6
5.7
5.8
5.9
SCHILLER
SP-10
Spirometer . User
Guide
5
.
I
Page 77
Chapter Maintenance, Testing, Troubleshooting
5
5.2
Page 78
Ma in t e n ance, Testing, Troubleshooting
Replacing the Recording Paper
A
red strip on the bottom edge When this strip appears, there are about fifteen be replaced as soon as this indication appears. Should the paper run out completely, then an audible alarm is initiated and the message
To replace the paper, proceed as follows:
of
the recording paper indicates that it needs to be replaced.
A5
pages left. We recommendthatthe paper
"REFILL
PAPER!" appears on the display.
Chapter
5
By pressing the key
Remove any remaining paper from the paper compartment. Take a new packof recording paper and place it into the paper compartment with the
grid side upper most and with the black markings at the top back of the unit.
Pull
the first sheet out
0
Pressw again and the paper tray slides in automatically.
When printing is restarted, the warning message will be removed from the display.
NOTE:
WELCH ALLYN can onlyguarantee impeccableprinting qualityif the original
chart paper orpaper
the paper tray slides out automatically.
of
the paper tray
of
equal quality
so
that it is over the roller
is
used.
of
the paper towards the
Cleaning the Casing
The casing of the
Switch the unit off before cleaning.
SCHILLER
SP-10
should be cleaned with a soft cloth on the surface only.
Do
NOTE:
SCHILLER SP-10 Spirometer - User Guide
not, under any circumstance, immerse the apparatus into a cleaning liquid
or sterilize with hot water, steam, or air.
5.3
Page 79
Chapter Maintenance, Testing, Troubleshooting
5
Flow
Sensor
Inner
SP-110
/
1
Outer
Tube
SCHILLER
\
Rubber
Adapter
\
\
\
\
\
\
Handle
\
\
\
-\
/
/
I
rnax.
1.5
\
crn
1
5.4
Page 80
Chapter
Ma in ten an c e, Testing, Troubleshooting
5
Flow
Cleaning and Sterilising
The utmost attention must be paid
must breathe through the nozzle (inhaling as well as exhaling), it is important that the mouthpiece and filter be renewed and the measuring tube painstakingly cleaned and sterilised before the next patient uses the device.
To
Sensor
clean the flow sensor, proceed as follows:
Remove and discard the mouthpiece. Remove the rubber adapter by pulling it away from the inner tube. Remove the inner tube of the flow sensor by pushing it out of the outer tube in the
direction of the red location marks. Once the tube has been pushed half way be pulled out from the other side.
Unscrew the two halves of the inner tube and remove and discard the filter. Clean and sterilize all parts of the inner tube, the rubber adapter and the inside of the
outer tube with one of the following products or a product of equal quality:
Kohrsolin
SP-
1
10
to
thecleanliness of the flow sensor. Because the patients
it
can
-
Lysoform
The cable and handle can be wiped with soapy water.
To reassemble the flow sensor, proceed as follows:
Insert a new filter into the inner tube red location mark.
Carefully screw the two halves of the inner tube together making sure that the filter
is not displaced. Locate the end of the inner tube into the end of the outer tube and push it gently but
firmly until the shoulder of the inner tube makes contact on the outside edge of the
The
two
outer tube. Fit the rubber adapter by placing its wider end overthe end
red location mark.
Insert a new mouthpiece
The flow sensor is now ready for use.
red location marks must be together,
(ma.
so
that it sits on the inner lip of the half with the
of
the innertube with the
1.5
cm) into the end of the rubber adapter.
SCHILLER SP-10 Spirometer - User Guide
5.5
Page 81
Chapter Maintenance, Testing, Troubleshooting
5
Flow
2.
3.
Sensor
Remove disposable mouthpiece
1.
handle Discard the complete assembly.
Position new disposable mouthpiece it
in position.
NOTE:
SP-150
(2).
The disposable mouthpiece can
force is necessary to insert it
(1)
by gently but firmly pulling it away from the
(Part
No.
only
on
the handle.
2.100077)
be positioned in one direction and no
and gently but firmly click
5.6
Page 82
Chapter
Maintenance, Testing, Troubleshooting
5
Flow
1.
NOTE: This is only possible in one direction.
2.
Sensor
Slide out the combined filter/mouthpiece assembly
Unscrew the assembly.
I
S
P-20
(1).
SCHILLER
3.
Remove and discard filter
4.
Clean and disinfect the assembly after every patient.
5.
Insert a new filter
6.
Carefully screw the two halves
Make sure that the filter is not displaced.
7.
Push the combined filter/mouthpiece assembly gently but firmly into the outer tube until it makes contact on the outside edge
NOTE: The assembly can
SP-10
Spirometer - User Guide
(Part
only
(2).
No.
2.100123).
be fit
in
one direction.
of
the assembly together again.
of
the outer tube.
5.7
Page 83
Chapter Maintenance, Testing, Troubleshooting
5
Flow
I
SC
Sensor
H
I
L L
SP-IlO/R
ER
with Resistance Sensor
Resistance
\
\
\
Motor
Tube
Plug
Handle
5.8
Page 84
Chapter
Main t en ance, Testing, Tro ubles
h
oo
5
ting
Flow
Cleaning the Resistance Sensor With each resistance test, a buildup of condensation is created within the resistance sensor
of the flow sensor. This should be wiped dry after each patient.
To
-
Sensor
clean the resistance sensor, proceed as follows:
Remove and discard the mouthpiece.
Disconnect the pressure tube and motor plug from the handle. Pull the resistance sensor out of the rubber adapter. Separate the two halves of the resistance sensor, making sure that the shutter does
not fall out. Clean and sterilize all parts of the inner tube, the rubber adapter and the inside of the
outer tube with one Kohrsolin Lysoform
SP-IIO/R
of
the following products or a product
with Resistance Sensor
SP-1
1OlR
of
equal quality:
NOTE:
To reassemble the resistance sensor, proceed as follows:
The resistance sensor is now ready for use.
Before startingtests witha newpatient, the main part of the flowsensorshould also be thoroughly cleaned and sterilised as described before.
Checkthat the shutter is in place and reassemble the two halves of the sensor making sure that the two air tube apertures are on the same side. Press lightly until the two halves click together.
Insert the air tube into the rubber adapter ensuring that the pressure tube and motor plug are on the under side and pointing in the direction of the flow sensor handle (red location mark
Slide the pressure tube over the connection on the handle of the flow sensor.
Plug the motor plug into the socket on the handle of the flow sensor. Insert a new mouthpiece.
to
the left).
SCHILLER
SP-10
Spirometer - User Guide
5.9
Page 85
Chapter
Maintenance, Testing, Troubleshooting
5
User
Self-Test of the Unit
Each time the unit is switched on, a self-test is performed to check all the functions.
Technical Safety Check
At
check should include the following:
Checks
12
monthly intervals, the unit should undergo a technical safety check. The extent of this
Visual inspection of the unit and cables.
Electrical safety tests according to Functional tests according to the Service Handbook.
IEC
601
-1.
5.10
Page 86
Chapter
5
Main ten ance, Te s ting, Troubleshooting
1
User
Checks
Test Output Channels (RS-Test)
NOTE:
The tests in this menu, in particular the self test (key
T)
which has out with a special test plug, are intended for use by authorised service personnel only.
First connect the SCHILLER
To
test the transmission channels, proceed as follows:
<
RS-Test
First define the transmission channel
SP-10
a-
to an external unit that can send or receive data.
TEST
R
S
2321422
0
-Test output on channel n
I
-
Test input on channel n
of
test
Settings
with
= 3
Ek153600
OE
inputloutput
test
P:
PE
TxE
(B,
P
C!
=
End
N
P
Channel selection
+I-IPIS
T
-
Selftest
CHANNEL n
DSR
BD
000001
and
S)
before it is to be tested.
=
FE
Selectthetransmission channelthat isto betested by pressing the key N (setto either
3
or
4).
Change the baud rate with the keys
pF-JorpF-,
to
be carried
plug
0
S:
1
RxR
TxR
01
Change the parity and the stop bits with the keys P and
To
carry out an input test on the selected channel, press the key
on the selected channel, press the key
With the
With the
output
test
input test
the message
a special display appears. It shows the last three lines of the
0.
"TEXT OUTPUT"
previous display with the received text characters.
To
quit either
of
the above test sequences, press
Q.
S.
1.
To carry out an output test
appears as last line.
SCHILLER
SP-I0
Spirometer - User Guide
5.
I
I
Page 87
Chapter Maintenance, Testing, Troubleshooting
5
5’
12
Page 88
Chapter
Technical Data
6
Dimensions
Weight:
Power supply requirements:
Power consumption:
Battery: Printing process:
Paper speed: Chart paper: Liquid crystal display:
(llwlh):
Technical
320 x 264.x 74
Approx.
1 10 I 130 I 220 I 240 13
to
30
Lead-acid,2 x 12 V High-resolution thermal printhead, 8 dots per
(amplitude
mds
2.5
I5
1101
Thermoreactive, Z-folded,perforation for A5 format Backlighted liquid crystal display for monitoring and
alphanumeric information Resolution:
adjustable
4.7
kg
VA
axis),40
mm
12.5
640
VAC,50
dots per
I25
I
501
100
x
480
dots;viewing angle
I
60
Hz
mm
(time axis) @25
mds
mm
~~
Data
~
Safety standards:
Protection class:
Environmental conditions:
Keyboard:
BF
according
RL
93142lEEC EN 60601-1:1990 IEC 601-1
pr
EN 1441 :1994
IEC 51 3:1994
EN 5501 1 IIEC 801
I
according
power)
lla according
Temperature, operating: Temperature, storage:
Relative humidity:
condensing)
Splashproof keys
to
to
to
IEC
and complying with the following:
IEC,
VDE, SEV (with internal battery
RL
93142lEEC
10"
-1 15
to
40" C
0'
to
50"
C
-
85%
(non-
SCHILLER
SP-10
Spirometer - User Guide
6.
I
Page 89
Chapter
6
Technical Data
Technical
Pulmonary Functions Method of Measurement: Measurement Ranges:
Measurement Accuracy:
Flow
Impedance:
Prediction Equation:
Video Connector
Data
Pneumotachometer Flow:
0
to
+14
Vs;Volume:
0
to
+11 litres
+2%
Less than
0.5
mbar
*
s/l
at 10
Vs
Adults:ECCS / Austria / Berglund / Finnish / Indian
Morris / Crapo / Knudson / Knudson76 / Polgar
Chi1dren:Quanjer
&
Tammeling / Austria / Indian
/
Knudson / Knudson74 / Polgar
/
Connector for external monitor.
sub,
15 pole, High-density, female:
D
Resolution: Input signals:
Scanning frequency:
Pin Connections:
640 x 480 dots Horizontal sync.: TTL (positive)
Vertical sync.:TTL (negative) Video:
0
to
0.7
V
Horizontal: 31.3 kHz Vertical: 60 Hz
Pin 2 Video Pin
5
Test Pin 7 Video GND Pin
9
DCLK Pin 10 Sync GND Pin 11 DlSP Pin 13 H-SYNC Pin 14 V-SYNC
6.2
Page 90
Technical Data
Serial Interfaces
Chapter
Technical Data
6
RS-232 0124)
Protocol: Baud Rate: Byte Format:
Transfer Control: Connection Socket:
Pin Connections:
RS-422
Protocol: Baud Rate:
Interfaces (Interfaces
Interface (Interface
No.
1,2
and
3)
Asynchronous 75 to 153600 Baud
1
start bit, 8 data bits
0
or 1 parity bit
1
/
1.5 / 2 stop bits, programmable
By means
3 x D subminiature wired as DTE (Data Terminal Equipment).
Channel 1,2,3: Pin 3 TXDl Pin
2
RXDl Pin 7 RTSl Pin
8
CTSl Pin 6 DSRl Pin
4
DTRl
5
GND (ground)
Pin
No.
4)
Asynchronous
75 to 15600 Baud
(+
or
-),
programmable
of
DTR, DSR, CTS, RTS
(9
pole female),
0
(output data)
I
(input data)
0
(request for output)
I
(ready for output)
I
(transfer unit ready)
0
(SP-10 ready)
Byte Format:
Transfer Control:
Connection Socket: Pin connections:
1 start bit, 8 data bits
0
or 1 parity bit
1
/
1.5 / 2
None
1
x
D-sub, 9 pole
Pin 1 GND
2
TXC+
Pin Pin 3 TXC­Pin
4
RXC+ Pin 5 RXC Pin 6 RXD Pin 7 RXD Pin 8 TXD+
9
TXD
Pin
(+
or
stop bits, programmable
-
+
-
-
-),
programmable
SCHILLER
SP-10
Spirometer - User Guide
6.3
Page 91
Chapter
Technical Data
6
6.4
Loading...