Welch Allyn SP-10 User manual

..
Cardio Menu
Operator Manual
.
SCHILLER
SERVICE
SP-I
WDBOOK
Spirometry
7
Unit
2
July
I998
Sp
SP=l
irom
etry
Unit
.
Service
CH-6340
Phone:
Fax:
Home page:
Hundbook
SCHILLER
Altgasse
Baar, Switzerland
41
A141
http://www.schiller-ag.ch
68
41
760 87 87
761
AG
08
80
July1
998
Article Number:
i
2.
540
OII
Issue Issue
1:
2
March July
1998,
1996
Update
SP-1
Article
and
revision
Service
Number
to
incorporate
SERVICE
Handbook
2.540
011
latest
hardware
SCHlUER
SP-I
HANDBOOK
and
software.
Spirornetry
Issue 2 July
Unit
1998
f
Associated
Guide
and
SCHILLER SCHILLER SP-1 USER GUIDE SCHILLER SP-1 USER GUIDE
Documents
to
SCHILLER Interpretation
Measurement Rogram
SP-1
USER GUIDE - English / German
v/
E
/
D / F
-
French
/
Italian
-
Spanish
I
Portuguese
93/42/EEC Medical Devices: 0124
'Notified Body'
DEKRA
AG
Article Article Article Article
No. No. No.
No.
2.510
2.510
2.510
2.510
179
196
197
103
0
Copyright
1998
by
SCHILLER
AG
11
SCHILLER SERVICE
HANDBOOK
SP-I
Spiromem
Issue
Unit
2
July
1998
DECLARATION
Diagnostic
Serial numbers Year of manufacture:
System
starting
:
with:
OF'
CONFORMITY
Spirovit
SP-1
040.
1997 Onwards
We, the undersigned, hereby declare that the medical device (class above conforms with the essential requirement listed
in
Annex 1 of
93/42/EEC.
This
declaration
Certificate
of
is
approval
supported
No.:
11425-01 45112-60-01 45112-16-01
by:
IS0
9001
(REV.
1994)
IS0
9001/08.94EN46001/
Annex
II,
Section
3
of
EN
the
directive
46001
12.93
IIa)
specified
EC
Directive
by
SQS
byDEKRAmd
93/42"EEC
Baar
(Switzerland) Dated 30.06.1998
J.J.
Schmid
&
Research
Development Manager
(€01
.
24
Markus
Butler
Quality Assurance Manager
...
111
SCHILLER
SERVICE
SP-1
HANDBOOK
Spirometry
Issue
2
July
Unit
1998
Where
Asia Pacific
to
Obtain
America
Austria
Canada
France
Service
1
WELCH
92121,
Telephone:
Fax:
SCHILLER Asia Pacific, Petaling Jaya, Malaysia
Telephone: Fax:
I
I
SCHlLLER Handelsgesellschaft mbH Medizintechnik, , Kampmiillerweg
24,
Telephone: Fax:
I
I
SCHILLER France
F
Telephone: Fax:
ALLYN
USA
800-535-6663
315-685-3361
(603) 777 5336 (603) 777
A4044
Linz,
0732 70 9 90 0732 7157
-94350
Villiers s/Marne, France
01 49 41 24 40
01
SCHILLER Inc.,
1OA
Jalan
5744
Ausma
OOO
SA.,
Zac des Luats
49 41 24 49
,7420 Carroll
SS
3/33,
58,
Road, San Diego, CA
Taman University,
route de Champigny,
47300
1
Germany
(EU-Bevollmachtigter))
India
IMY
Switzerland
All
Other
Countries
SCHILLER Medizintechnik GmbH, Rudolf Diesel
D
-85521
i
Telephone:
1
F~:
~
Hotline:
SCHILLER Healthcare India Pvt. Ltd. Kalpatarn Motwane Lane, Ground Floor, Fon Mumbai
Telephone:
Fax:
ESPAOTE SPA (SCHILLER)) Via di Caciolle Telephone:
Fax: SCHILLER Reomed AG, Riedstrasse
Telephone: Fax:
SCHILLER AG, Altgasse Telephone:
Fax:
1
Home page:
Ottobrunn, Germany
089 629 981 089 609 089 629 981 36
+91 22 263 4381 +9122 263 4384
+39
055
+39
055
01 741 02 09 01 740 37 10
41
41 760 87 87
41
41 761 08 80
http://www.schiller-agxh
0
50
9
422 91 422 920 8
68,
CH-6340
0
14, CH-8953
Baar,
Su.
14,
Chambers,
400
023
15,1-50125
Dietikon, Switzerland
Switzerland
Nanik
Firenze, Italy
I
iv
-
&.
-.
SCHILLER
SERVICE
SP-1
HANDBOOK

Warranty

Disclaimer
The Information given
as
regarding the contents changes any person of such revision
Trademarks
SCHILLERandSP-1 of their
Copyright Notice
0
Copyright
transcribe, store in a retrieval system
electronic, mechanical, magnetic, optical, chemical,
publication without express written permission
Spiromev Unit
issue
2
July
1998
in
this
guide
has
been carefully checked for reliability; however no guarantee is
to
the correcmess
in
the
specification of the product described within at any
owners.
1998
of
the
contents and SCHILLERmakes no representations
of
this
manual. We
or
change.
areregisteredtrademarksofSCHlLLERAG.All
by
SCHILLER
AG.
reserve
All
rights reserved.
or
translate into any language, in any
of
the
right
to
You
manual
SCHILLER
AG.
or
revise
this
document and make
time
without obligation
mdemarksare the property
may
not reproduce, transmit,
fonn
or
by any means,
or
otherwise,
any
part
warranties
to
notify
of
this
Terms
of
Warranty
The SCHILLER SP- 1 is warranted against defects in material and manufacture for one year or Any unqualified persons attempt
In case The manufacturer
apparatus
'
persons authorized by
'
instructions.
THERE
WARRANTES
MERCHANTABILITY ORFITNESS FOR APARTICULARPURPOSE
THE
(as
from date of purchase). Excluded from
as
a result of improper handling. The warranty entitles free replacement of the defective part.
liability
assembly operations, extensions, readjustments, modifications, or repairs
the
PRODUCT
for subsequent damage
to
make repairs.
of
a
defect send the apparatus to your dealer or directly
can
only be held responsible for the safety, reliability, and
if:
him,
and
unit
and approved attached equipment is
ARE
NO
EXPRESS
HEFEINABOVE
OR
OR
PARTS
THEREOF.
is
IMPLIED
SET
FORTH. SCHTLLER
this
guarantee is damage
excluded. The warranty
to
the manufacturer.
used
in
accordance with
WARRANTES WHICH
MAKES
is
void
EXTEND
NO
the
duration
caused
by
an
accident
if
unauthorized or
performance
are
carried out by
the
manufacturers
BEYOND
WARRANTY OF
WITH
RESPECTTO
of
the
THE
of
V
-.
.
..
.
c
...
.-
SCHILLER
SERVICE
.
._
.
_.
...
.
i
.
Safety
Notices
TO
PREVENTELECTRICSHOCKDONOTDISASSEMBLETHEUNIT.NOSERVICEABLE
PARTS INSIDE. DO NOT USE
THE
OR
PRESENCE
REFER
THIS
UNIT
SERVICING TO
IN AREAS
OF
FLAMMABLE
QUALIFIED
WHERE
THERE
GASES SUCH
PERSONNEL ONLY.
IS
ANY
AS
ANAESTHETIC AGENTS.
HANDBOOK
DANGER
SP- I Spirometry
OF
EXPLOSION
Issue
2
July
Unit
I998
IFTHEDISPL.AY IS DAMAGED, ALEAKAGEOFFLUIDMAY OCCUR.DONOTINHALE
THE
VAPOUR FROM THIS
CONTACT
IS
MADE, CLEAN CONTAMINATED AREA
FLUID
AND
AVOID CONTACT
WITH
MOUTH
IMMEDIATELY
AND
WITH
SKIN.
FRESH
IF
WATER. THIS PRODUCT IS NOT DESIGNED FOR
SWITCH
DONOT,LJNDERANY
IN
DO NOT OPERATE MAINS
THE
UNIT
OFF
BEFORE CLEANING
CIRCUMSTANCES,JMMERSETHEUNITORCABLEASSEMBLIES
LIQUID.
THE
UNIT
IF
THE
LEAD IS DAMAGED OR SUSPECTED
STERILE
USE.
AND
DISCONNECT FROM
THE
EARTH CONNECTION IS SUSPECT OR
OF
BEING DAMAGED.
MAINS.
IF
THE
I
DO
NOT USE HIGH TEMPERAW STERILISATION PROCESSES (SUCH AS
AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA RADIATION STERDLISATION.
DO
NOT
USE
SOLVENT
USE ONLY ACCESSORIES SCHILLER AG. USE RESULT
THE
AFFORDS
IN
INJURY INACCURATE INFORMATION
SP-1 COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH
PROTECTION AGAINST EMISSIONS
HOWEVERSPECIALCAREMUSTBEEXERCISEDWHENTHESP-1
CLEANERS
AND
OF
OTHER
OTHER PARTS RECOMMENDED OR SUPPLIED BY
THAN
RECOMMENDED
AM)
OR SUPPLIED PARTS MAY
AND/
OR DAMAGE TO
THE
UNIT.
ELECTRICAL INTERFERENCE.
ISUSED WITHHIGH
FREQUENCY EQUIPMEN?:
IT
MUST BE ENSURED THAT NEITHER THE PATIENT NOR
(INCLUDING
PERSONS OR CONDUCTING OBJECTS WHEN
ENSURE
THE
NEUI'RAL
NON-LMEDICAL DEVICES
ELECTRODE) COME
EVEN
IF
THESE
ARE
CONNECI'ED TO
INTO
THAT BOTH UNITS ARE SECURELY CONNECTED TO
ARE
THE
ELECTRODES
CONTACT
WITH
EARTHED).
THE
Rs-232 INTERFACE
THE
SAME
OTHER
EARTH
POTENTIAL.
WHEN
NONMDICAL DEVICES,
OPERATING
THE
UNIT ON BA?TERY
THE
RS-232
AND
INTERFACE
MUST BE FULLY ISOLATED.
SIMULTANEOUSLY USING
vi
SCHILLER
SERVICE
HANDBOOK
SP-1
Spirometry
c
Unit
Issue
2
July
1998
Safety
BEFORE USING
UNIT
SCHLLER REPRESENTATIVE.
THE GUIDELINES FOR PATIENT ELECI'RODE PLACEMENT
OVERVIEW
~
THE
PHYSICIANS ORPERSONNEL UNDERTHEIRDIRECTSUPERVISION. THENUMERICAL
AND
RECORDING
CLINICAL CONDITION.
GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT
ACCURACY, MUST
IT IS THE PHYSICIANS RESPONSIBILITY
EXPERT OPINION ON
INDICATED.
Notices
THE
UNIT,
ENSURE
FUNCTIONS
SP-1 ECG UNIT IS PROVIDED FOR
GRAPHICAL RESULTS
ONLY.
MUST
AND
THEY
BE
ALSO
THE
THE
SAFETY
ARE
NOT A
AND
EXAMINED
THE
RECORDING PREPARATION QUALITY AND
BE TAKEN
RESULTS,
THAT
AN
INTRODUCTION REGARDING THE
PRECAUTIONS
SUBSTl"E
THE
ANY
INTERPRETATION DERIVED
WITH
RESPECT
INTO
ACCOUNT.
TO
MAKE
AND
TO
INSTITUTE
HAS
FOR MEDICAL EXPERTISE.
EXCLUSIVE USE
TO
THE
THE
DIAGNOSIS OR TO OBTAIN
CORRECT TREATMENT
BEEN PROVIDED BY A
ARE
PROVIDED AS
OF
QUALIFIED
PATlENTS
OVERALL
FROM
ON
A
THE
THE
REPORT DATA
IF
vii
-
SCHILLER SP-1
SERVICE
HANDBOOK
Spirometr?,
2
July
Unit
1998
~
._..
_-
What’s
in
THE
SERVICE
MODULE
TION
NECESSARY
AND
REPLACE A FAULTY
EDGEOFTHESP-1
WITH
STANDARD
The
book
Chapter
The purpose
to give a basic introduction operation and function of each key briefly explained. The information in background to SCHILLER
Chapter
This chapter provides a functional overview by functional block diagrams.
Chapter
2
3
this
Book
PHILOSOPHY
LEVEL.
is
1
-
Operating Elements
SP-1
-
Functional Overview
-
Fault Diagnosis
THE
TO
BUTDOESREQUIRETHATTHESERVICE~G~RISFAMlLIAR
WORKSHOP PRACTICES.
divided
into the following
of
this chapter is
the
operating functions
Operating Manual.
PURPOSE OFTHIS
ENABLE
to
to
the
FOR
ALL
THE
MODULE.
chapters:
provide
an
SP-1.
This chapter gives details of the operator controls with the
only.
SCHILLER
BOOK
SERVICE
THIS
BOOK
easy reference for all the
Complete opting information is provided in the
of
the
SP-1
UMTS
IS
TO
ENGINEER
ASSUMES
The functional description
IS FAULT
PROVIDE
TO
EFFlCIENTLY LOCATE
NO
main
ALL
DETAILED
operator functions and
this
FINDING
THE
INFORMA-
KNOWL-
chapter provides
is
supported
TO
a
!-
-\
7’
3
1
This chapter provides a guide logical sequence of the text where needed.
Chapter
and after replacement. Functional checks that must be
also
Chapter
test procedures
Chapter
This Chapter provides the part numbers and reordering information for
Also
adjustment or fault finding procedures.
Chapter
4
-
This
chapter gives attributes of the unit briefly described. The physical description showing the internal location of all modules. Removal and Replacement instructions for replaceable modules are also provided in of
tools,
jumper settings, adjustments,
provided.
5
-
Adjustments
This chapter provides
6
-
Spare Parts
included
7
-
Technical
fault
Module Removal
an
all
that
can
in
this chapter
to
locate a fault
finding algorithms and procedures. Illustrations are provided to support
and
overview of the physical construction of the
and
adjustments and settings.
be
performed to check the functioning
are
details of my
to
module level. The diagnostics are presented in a
Replacement
SP-1
with the main physical
is
supported by illustrations
this
chapter. Each procedure
settings or
special
carried
Also
detailed in
special
test
requirements
out after replacing a new module are
equipment or
is
autonomous with details
that
are
this
chapter are basic functional
of
the unit.
all
replaceable modules.
special
tools
required before
Data
all
required for
The full technical specification of the
Chapter
This Chapter explains diagrams.
8
-
Glossary
all
the acronyms and signal titles used in
SP-1
is
given in
...
Vlll
this
chapter.
this
book and in the
SP-1
circuit
,
..
SCHILUR
SERVICE
.-.
.
What’s in this
SP-I
HANDBOOK
Index
Spiromrry
Issue
Book
Unit
2
July
I998
Circuit
The
circuit
for
information
Diagrams & Board
diagrams
and component
only.
The philosophy ment.
Our
aim
information and the latest technological developments. We reserve the right improvements at any time.
Your
suggestions and comments are welcome. Please contact
the
SCHILLER
Fax
41
41
Layouts
layouts
are
provided
We
Need
Your
of
SCHILLER
is
to provide the
to
revise
this
document and make changes
Technical Documentation Department:
761
03
34
for
all
boards.
Help
is
one
of
continuous improve-
user
with the most up-to-date
These details are provided
or
.
. . .
.
-
__
.
.
__
.
.
.
-
.
.
.
-
List
qf
Symbols
6
A
Mains
Supply
Battery Operation
FllSe
LCD Contrast
Potential
Attention - General Warning
Equalisation
(common
SCHILLER
SERWCE
Ground)
Sign
-
see accompanying documentation
HANDBOOK
SP-I
Spirometry
Issue
2
July
Unit
I998
Type
BF
equipment - safe for external applications
Manufacturers Take-back
0
-
Mains Power
I
-
Mains Power
ONlOFF
Off
-
mains supply disconnected from unit
On
-
mains
guarantee
supply
for used batteries
connected
to
unit
X
SCHILLER
SERVICE
HANDBOOK
SP-1
Spirometry
Issue
Unit
2
July
1998
Chapter
Operating Elements
I
Procedure Overview
Introduction Location
Keyboard
Main
Menu Overview
Setting Screen Contrast
System
Pulmonary
Obtaining
Flow Sensors
Setup Menu
Sending
Function Test Procedure
a
Printout
and
Contents
Entering
Operating
Data
Chapter
Elem
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.10
1.13
1.15
1.16
en
1
ts
Page
1.1
Chapter Operaring Elements
I
SCHILLER
SERVICE
HANDBOOK
SP-I
Spirometry Unit
Issue
2
July
I998
Procedure
.
.
.,
--
/
General Settings
1
Date Time
1
Device Norm values
1
Contrast
1
Print User
1
Memory
etc.
L--
Overview
for
every
Insert new filterkcreen
Insert new sensor
-7
I
I
I
I
I
I
Enter Patient Data
newpafient
(SP-20)
I
Repeat tests
or
I
(3
more times)
1
Make Measurements
P
1
I
Post-medication
tests
1
Switch
ON
2
Make
Settings preference
3
Insert new screen/filter
4
Calibrateunit
5
Enter patient
6
Select test
7
PressSTART
8
Patient Makes test
9
RessSTOP
10
Repeat
11
Display
12
PrintResult
13
Send
result
as
many
result
data
if
required
times
if
required (remembered
or
sensor
as
required
(Best,
or
best
two
when
+
unit switched
last printed)
ofl
Page
1.2
SCHILLER
SERVICE
HANDBOOK
SP-I
Spirometry
Issue
Unit
2
July
1998
Chapter
Operming
I
Elements
-

Introduction

TEST INDICATORS POWER
The SCHILLER SPIROVIT
SP-1
PAPER REEASE
INMCATORS
is
a stand-alone pulmonary function
CATCH
SYSTEM
INDICATORS
MSPUY
WINS
AN0
RS-232
tb
POTENTIAL
CONNECTOR
features
EauALlSAnONCONNECTORS
sting unit four modes for the measurement of inspiratory andexpiratory values. The unit incorporates aliquid crystal display clear and comprehensive result analysis is provided for msmission of
are
initiated
test being carried out and function
sensor disconnected, paper and hygienic open pneumotacho sensor. Two sensors are available. The
to
instantly show the test results.
test
results to
via
an
easy to use keypad with dedicated test keys. Test indicators show the current
LEDs
tray
empty and paperjam. Measurements are
An
integral high quality thermal printer provides
in
anumber of different formats and
a
PC
with the
provide visual indication
SEMA
program.
for
an
RS-232
All
tests
functions such
made
with a lightweight
SP-20
open pneumotacho
interface
and settings
as
power,
sensor incorporates a disposable filter and is easily dismantled for cleaning and sterilisation. The
SP-150 sensor
the sensor after each use. The SP-1
has
a disposable, single patient sensor/fdter assembly eliminating the need to clean
has
the following features:
Direct function keys for measuring the following:
FVC
Forced Vital Capacity
SVC
Slow
Vital
Capacity
MVV
Maximum Voluntary Ventilation
MV
Expired or Minute Ventilation
Low weight and compact dimensions Built-in rechargeable battery for mains-independent use Selectable printing formats with integrated quality thermal printer
9
Patient memory for up Choice
of
6
languages for display and printing
to
100
measurements (option)
Page
1.3
ic
Chapter
I
Operating Elements

Location

As
with
all
The
unit
The
unit
The
unit
transformers unit
and
Potential Equalisation
In certain
equipment. Connect the earth potential'equalisation hospital common frame. Note same common ground. The part number of
Power Supply
SERVICE
medical
should
units,
the following general
not
be
stored
or
rules
are
to
be
not&
operated in a wet, humid or dusty environment. may not come into contact with acidic steams or liquids. should not
the
circumstances
be
set up near radiology or diathermic systems, or near Iarge
or
electrical motors.
mains
supply
intederence
Ensure
is
maintained
can
that
a
minimum
be
caused by externai electrical equipment or
distance of 1 meter between the
connection
ground
or,
if
a
that
all
other electrical equipment
common
ground is
not
present,
in
close
proximity
the
potential equalisation cable is
to
should
SCHILLER
HANDBOOK
at the back
of
the
SP-1
SP-1
Issue
to
a metal framework e.g. the
also
be
connected to the
2.310
005.
Spirometry
2
July
radio
the
bed
Unit
1998
The unit can
connection
connected to the battery capacity the
mains
15
hours
A
fully
remain connected to the
Switching
The
SPIROVlT
be
operated from the
is
on the
mains
is
supply
by
to
be
fully
rear
of the unit. The
supply. A battery indicator lamp confirms
limited, the indicator flashes.
means of the supplied power cable. A totally discharged battery needs less than
recharged
(60%
mains
in less
or
from
mains
than
To
charged battery lasts approximately 7 hours of normal use
mains
supply without any danger of damage toeither the battery or the
On
and
Off
SP-1
is
switched on and off with the
the built-in rechargeable battery. The mains
indicator lamp
is
always lit when the unit
battery
operation. When the
recharge the battery, connect the apparatus
3
hours,
90%
in 6 hours).
(SP-1R
6
hours). The unit can
ON
and
OFF
keys.
is
to
unit.
Page
1.4
..
.
:-
..
..
SCHILLEU
SERVICE
SP-I
HANDBOOK

Keyboard

Indicates the being carried
Spirometv
test
out
Unit
Issue
2
July
1998
currently (Commence each measurement with
Enter Lung Function Test Menu as designated each measurement with STOP) unit
L
START;
Chapter
Operaring Elements
end Switch Switch
off
unit on
I
Move Display
Stop Measurement
Display Menu options
Cursor
Start
Print Measurement
or
1
Measurement
Stop Printout
for
Enter Patlent Data
the indirect functions
Display Result
Change key to second function
Delete a character/delete last measurment
or
Escape
of
the test
1
Calibibrate unit
Enter menu option displayed
Pre and Post medication test
Battery Operation
Paper jam
or
confirm entry
/
Mains Connected'
Sensor not connected
or
no paper in paper tray
Post test being carried out
Page
IS
Chapter
Operating
I
Elements

Main Menu Overview

SCHILLER
SERVICE
SP-I
HANDBOOK
Issue
Spirometry Unit
2
July
1998
1
Main
Menu
Memory
MTA Ident
Send
Setup
When a menu
to
move the cursor
enter field
use these keys
RIGHT
:/
. . . .
is
.
....
..c
is
displayed use these keys
UP
displayed e.g.
to
move the
Use
this
used the same
The
SP-1
has
a memory for up to
memory menu item
Display a previously recorded patient
.
To
.
Enter the name (medical technical Assistant) is given on printouts. The users name entered here, needs updating when a new user takes over. See below.
When selected interface (for example the
delete a patient record
RS-232
protocol settings must
to:
of
the person carrying out the
is
remembered when the unit
all
recordings are
to
the
100
sent
SEMA
be
measurements. Select the
test.
The
MTA
is
switched
automatically
PC
storage program).
defined
in
the
off;
over the
Note
setup
menu before
it
name only
RS-232
that
transmission.
See System Setup Menu following
I
WWN.
When
h4TA
Ident
cursor
LEn
key
to
display the main menu
as
the
ENTER
an
I
key
to
confii
When a menu select highlighted menu option. When
is displayed e.g.
field
to
confm
as
above. When
entry and
is
displayed - press this key
entry and
an
enter field
to
exit field
h4TA
to
exit field
is
Ideat, use
displayed,
an
this
this
-.
to
enter
key
I
key
is
Page
1.6
_-
-
..-
SCHILLER
SERVICE
SP-I
Spirometry Unit
HANDBOOK
Issue
2
July
I998
Setting Screen Contrast
Setting
Change the contrast
Press the
and
Entering
Printed below certain keys on the keyboard key once to enter the first the third character.
The character is displayed on the screen and move
Use
Press the
When an enuy field requires anumericai input e.g. date of selected. When an entry field requires a letter input e.g. patient name, the letter keys are automatically selected. conversely a letter input required number key enables
the
Contrast
of
the screen
ALT
key and the
MENU
DOWN
data
in
a field
character,
to
the
next character position.
the left and right cursor keys at any the to select a position
ENTER
key or the MENU key to
If
a numerical input is required for example in the patient name, or
in
this.
Therefore when entering
Operaring Elements
and
to
Entering
your preference
key together then change the contrast with the cursor keys
are
letters.
twice
to enter the second character, and three times to enter
as
entered.
the date
Press
confum
of
birth,
dam
Data
as
follows:
In
a
data
field (e.g.
the right cursor key to confm the entry
to
entry and exit the field.
birth,
the number keys are automatically
pressing the
ALT
MTA
Ident) press the
change that character.
key before the character/
UP
Chapter
I
ALT
+
key = number input in character field
ALT
+
key = character input in number field
Page
1.7
L
Chapter
Operating Elements
I
SCHULER
SERVICE
SP-I
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
System
Press this option
key
Setup
to
Menu
key
or
to
change setting.
program
to
select highlighted menu
Press
the
settings and
exit
menu
i
.=-
\\
MENU
This menu option defmes switched off) until overwritten. The settings are
User Ident
Measurement
all
the general
Use
this
operates
This
is how the
key toggle the options
Row Row=lOop
Vol
settings.
to
enter the name of the consultant
the
unit
FVC
gmph is displayed and printed. With the
=
f(v0l)
=f(t)
All
enaies remain stored (even when the unit is
as
follows:
or
department
to
set a flow display of one of the following:
that
ENTER
Page
1.8
SCHILLXR
SERVICE
SP-I
Spirometry Unit
HANDBOOK Issue 2 July
I998
System Setup Menu (cont.)
This
Print
defines the format and content of the printout when the
key
is
pressed. The settings
are
as
follows:
\
Chapter
Operating Elements
I
PRINT
Device
STANDARD
NORMALS
UNITS
DATE
RACE
Graph
Time
Axis
Flow
graph
Volume
Diagnosis
Graph
PEF
Select
ALL
or
BEST
(measurements)
This
defines
the
time
axis
1-s
Set
to
ON
or
or
2Ws
OFF
-
gaph printed or not printed
Print the volume graph - select
Diagnosis
printed
or
not printed -select
Peak Expiratory flow calculation - select litred min or litredsecond
Set
to
'International' for
all
countries outside the USA and Canada or
to 'American' for USA and Canada
The
diagnosis
value standards are given at the end
is
based on the standard defrned here. The Normal
of
this
section- 'Diagnosis
Norm Value Tables'
cm
/
kg
PF,
Set
to
:
cm
/
kg PC,
Set the order of the date to month
y
ear-month-day
inch / Ib. /OF
-
day - year,
NONE (no race); recommended for Europe,
W/B
(White/ Black) for
C/H/B/A (Caucasian
all
countries outside the
/
Hispanic / Black / Asian) for USA and Canada
of
the
graph
YES
day
-
month - year or
USA
and Canada
or
-
set
NO
YES
to
or NO
and
The standards and (Diagnosis and
of
effect
page)
race
Norm
Values are detailed at the end
Norm
Values).
on
results is detailed in the Patient data (see following
LANGUAGE This defines the language on the printout and menu smcture. Set to
Clock
RS-232
English, German, French, Set the time and date
The
Baud rate and parity setting (Noloddeven) can be stored.
Italian,
The following table shows typical setup combinations.
The
dividual
setting
Spanish, or Portuguese
iohl
IblF
of
for
race,
this section
and
the
Page
I.9
Chapter Operating
I
Elements
SCHILLER
SERVICE
SP-I
HANDBOOK
Spirometry Unit
Issue
2
July
I998
Pulmonary
The
1.
2.
The relevant keyboard lamp display
The coordinates represent the in
litres
Function
test
method for
Calibrate
At
If
When
Press the
as
shown
being represented on the vertical
the
FVC, SVC,
the
unit
THE
the beginning
the sensor
is
a
siNicant
function
FVC
SVC
MVV
MV
in
this
example for
Test Procedure
MVV
and
MV
UNIT
of
changed
ternmature or
test key
lights
graph
MUST
each
day
and the corresponding coordinate presentation appears on the
FVC
on
which
BE
mssure
the curve
axis
and the time in seconds
test
is
the same.
Proceed
CALIBRATED
difference
will
be
drawn
occurs
with
on
as
follows:
the respiratory volume the horizontal
axis.
:-
I
.”;
3
Page
1.10
3.
Press
the START key
c
SCHILLER SP-I Spirometry
SERVICE
Pulmonary
HANDBOOK
Issue
Function
The
flow
sensor must be
least one second before
The message volume counter graph for sensor, the display. The break-off point for the expiration measurement the Stop key
4.
Press the
5.
Repeat steps 3 and 4 until three measurements have
Note:
"READY
the
unit
is
STOP
Ifthe Start
measurements are stored.
Unit
2
July
1998
Test
held
quite
and
after the
FOR
MEASUREMENT'
the
value.
begins to
pressed).
record
the expiratory
key on completion
key
is again pressed (a founh
Procedure
still
and
no
air
should
START
key
is
pressed
appears
As
soon
as
the
patient
flow.
The corresponding curve
of
the
test.
or
more times), only the best three
(cont.)
be
breathed into the device
on
the display together with a
starts
to
breathe into the flow
is
reached automatically (or
been
taken.
Chapter
Operating Elements
for
at
is
represented on
I
When American deviation measurements
Following a series the
last
test
printed graphical dong with The table of the best measurements can
Definition
In accordance with the defined
The measurement table (on the printout and on the display) gives the highest individual measure­ments (e.g. last recording made. This is true for all parameters except the folIowing which are the highest
'
as
FEFo.2.
EEF,%.75%
FEF,,%.,S%
standards
(as
stipulated by the
is
not greater than
of
made (which may
of
Best
the highest value from the calculation:
Best
=
FVC + FEVl
FVC,
FEV1,
FVC
+
1.2
are stipulated,
patient measurements the best two results are recorded and stored along
or
may
a
table of
ATS
Spirometry Standard
SVC,
MVV
FEVl
value:
the
message
American
2OOml
Thorasic Society) between
or
5%.
not
be
one of the
the
best individual measurements (from
also
be displayed on
etc.) from
the
"ATS
criteriamet" appears on the printout
the
best and second best
best
three). These three results can then be
these
the
screen.
(11
November
two
best recordings
19941,
the
best recordings
as
defined above, and the
three
based
if
with
results).
are
soleIy on
FEFZ5
'
FEF,,
FEF,,%
%?%
and
when loop graph selected:
ERV
IRV
*Tv
the
Page
I.11
Chapter
I
Operating Elements
_-.
.
-..
L
c
SCHILLER
SERVICE
HANDBOOK
SP-I
Spirometry
Issue
2
July
Unit
1998
i'
F
,
_.
I-
Pulmonary Function
Displaying
Press
.
the
measurement table is displayed for the
the
Results
RESULT
Key
Test
in Tabular
after
a
measurement,
Procedure (cont.)
Form
or
series
of
measurements have been
test
mode
(as
indicated by the test lamp)
as
follows:
made.
A
FVC
svc
..
;
MV
The display gives the measured values and the predicted values (calculated from the entered patient The predicted values
The result displayed is in accordance with
parameter,
from the
When subsequent measurement column.
It
can
happen that
Measurement
three
tests
the
bestlVC value
2.
can
only
tests recorded.
are
made,
Press
the
be
calculated when the patient
ATS
recommendations and is the best resultfor
the
result with the highest
MENU
Key to return
is
taken from Measurement 1 and the best FEVl value is taken from
data
has
been entered.
value
will
always
to
the graphic display.
be
saved and given in the
data).
each individual
Page
1.12
SCHILLER
SERVICE
SP-I
Spiromtry
HANDBOOKIssue
Unit
2
July

Obtaining a Printout

1998
Chapter
Operaring
I
Elements
Press
The following
The patient
is
data
given:
after
taking
all
measurements.
The selected norm values The
date
of
last calibration
The number of tests stored. If more than
three
tests
are
made the figure given in brackets indicates the total number of tests wries out The number of post tests stored, and the total number of post test made are given in
opposite, a The
value The diagnostic statement The
FVC
and
is
within
MTA
identification
total
of
5
tests were canied out and a total of 6 post tests.
FEVl variation
2oOml
or
in
5%
the message
(if
enabled
the
same manner after the slash. In
ml
between the best and second best measurements.
'ATS
criteria
met'
is
printed.
in
Settings menu)
the
example given
The user identification The date and The software version and the
bottom
The second section
the test results presented values
the graph of the best test or the curves for
16).
other curves and
time
of
the printout
installed
of
this
page.
of
the printout comprises:
as
a
table
as
a
percentage of the predicted.
data
dependent on test made
options
giving the
all
(M
=
memory) are also indicated
best
values, the predicted values and
three
tests depending on selection
at
(
see
the
the
If
this
best
page
The Graphs of the measurements are given on the subsequent page(s). Note and the
ERV
values
are
only given
if
Inspiration
is
taken with the
FVC
test and if Loop is selected
in for measurement (see system setup menu).
that
the
FIVC
values
Page
1.13
c
Chapter
I
Operating Elements
Pulmonary
_.
..
..
-.
-.
..~
SERVICE
Function Test Procedure (cont.)
. . . . .
SCHILLER
HANDBOOK
SP-1
.
..
. . .
Spirometry Unit
Issue
2
July
1998
._.
-.
T.
L
crl
::.
.
j
Forced
For Note:
To
Vital
this
Capacity (FVC)
test
the
patient must exhale
Test
as
quickly
as
possible
The FVC test employs the "Back extrapolation" method.
(>0.15
lim
or
5%
of
FVO,
is too large
carry
out
the
test
for Forced
Vital
Capacity
then a waming appears
0,
from
press
key
the time of
If
"FVC"
coordinate presentation appears on the display The patient must exhale
understands what
be
immediately followed by a
is
required
as
quickly
as
of
him.
If inspimtory
maximum
possible &om the time
measurements
of
starring
are
required, the exhaiation can
inhalation. The inspimion results will
printout.
Before initiating a printout previously. Press the
Slow
Vital
PRINT
Capacity (SVC)
key to
obtain
Test
ensure that the
if
printout
as
seuings
defined.
of
the
FVC
Test,
The patient should breathe normally 3 times and then inhale maximally then exhale maximally. Make
MVV
Test
The patient should breathe make
sure
that he understands what is required of him.
sure
as
deeply and
that
the patient understands
as
rapidly
as
what
possible over a
period
starring
the test.
the extrapolated volume
on
the display
and the corresponding
the test
so
be
sure
that
be
given on the
are
correct
as
detailed
to
total lung capacity and
is
required
of 6 to
of
him.
12
seconds
he
so
,
WARNING
EXTREMECARESHOULDBEEXERCISEDWHENPERFORMINGTHIS~TAS
IS
A
DANGER
OF
HYPERVENTILATION.
ENSURE
THAT
THE
PATIENT
DOWN.
MV
Test
The patient should breathe
that
Make sure
the patient understands what is required
as
normally
as
possible for up
to
of
60
seconds,
him.
but
for at
least
THERE
IS
SITTING
20
seconds.
Page
1.14
SCHILLER
SERVICE
HANDBOOK
SP-I
Spiromrry
Issue 2 July
Unit-
I998
Chapter
Operating Elements
I
Pulmonary Function
...
.
:.
7
Post-Medication
In
order
to
cany
Lamp
lights
The post-medication measurements stored). The printout following post-medication pre and post-medication shown
as
the best results (pre/post), results
and the percentage change (i.e. difference) between pre and post-medication results.
Tests
out
post-medication
tests
tests
Test
are carried
Procedure
tests
for
comparison,
out
in
the
(the premedication curve is bold). The measurement results are
as
a percentage
(cont.)
press
same way
of
the
PEWPOST
as
the premedication tests (three
tests
will give the curves
those predicted,
key and
(both
pre and post)
"POST"
of
both
The
diagnosis resulting
from
the premedicarion
Sending
Measurements can be sent cation parameters,
side. See Chapter
Connect the Carry out the measurement(s)
Press the
When the transmission is complete a message
has
been completed.
PC
and on the unit. Check the cable connection. Ensure
are
the
same
3
PC
MENU
in
both the
to
the
ie
baud
rate, parity etc. are
(RS-232
Check) forfurtherdetails.
to
the
RS-232
key and select
If
an error message
PC
and the
SEMA
connector on the right hand side
as
descriid previously.
SEND
is
unit
fest
is
also
given
on
data
management system. Make sure
set
the Same on the transmitting
To
send ameasurement proceed
is
displayed showing
displayed e.g.
(SETUP
'serial
that
menu).
link
time-out' check settings in the
the
Baud rate and the parity settings
this
printout.
that
the communi-
and
the receiving
as
of
the
unit.
that
successful transmission
follows:
Page
1.15
Chuprer
I
Operating
Elements
SCHILLER
SERVICE
HANDBOOKIssue
SP-1
Spirometry
2
July
Unit
1998
Flow
Sensors
Flow
Sensor
1.
Remove disposable mouthpiece
(2).
2.
Discard
3.
Position new disposable mouthpiece position.
NOTE:
SP-150
the complete assembly.
The disposable mouthpiece can only necessary to insert it on the handle.
(1)
by gently but firmly pulling it away from the handle
(Part
No.
2.100077)
be
positioned
and gently but fdy click it
in
one direction and
no
force
in
is
d'
.
1
-.
Flow
Sensor
I
1.
Slide out
SP-20
the
combined filter/mouthpiece assembly
II
(1).
This
is
only possible
in
one
direction.
2.
Unscrew
3.
Remove and
4.
Clean
5.
Insert a new filter
6.
Carefully screw
not
the
assembly.
discard
Nter
and
disinfect the assembly after
(Part
the
two
displaced
(2).
No.
2.100123).
halves
every
patient.
of
the assembly together again.
Make
sure
that
thefilter
is
Page
1.16
7.
Push
makes
direction.
the combined filtedmouthpiece assembly gently but fmly
contact on the outside edge
of
the
outer
tube.
The assembly
into
the outer tube until it
can
only go in
one
.-
SCHILLER
SERVICE
HANDBOOK
SP-I Spiromtly
Introduction
SPll-
Power Supply CPU and Processing Circuits Memory Them1 Print Head Controller Paper Mark Power
Stepper Motor Controller
Issue
1
On
Unit
2
July
Reset
1998
Chapter
Functional
Contents
Chapter
Functional Overview
2
2
Overview
2.2
2.4
2.4
2.4
2.4
2.5
2.5
2.5
2.5
Page
2.
I
Chapter
Functional
2
Overview

Introduction

SCHILLER
SERVICE
SP-I
HANDBOOK
Spiromtv
Issue
2
July
Unit
I998
..
This
Chapter
to
enable
modules.
provides a functional overview
the
service
engineer
A
functional block diagram supports the
to
identify
processing
of
the
SP-1
electronics. The
paths in
text.
order
aim
of
this
overview is
to
help iden@ possible faulty
Page
2.2
SCHILLZR
SERVICE
HANDBOOK
SP-I
Spirometry
Issue
Unit
2
July
1998
Chapter
Functional
2
Overview
I
i
*
?
w
Page
2.3
Chapter
Functionul
2
Oveniew
\
SERVICE
c
SCHILLLB
HANDBOOK
SP-1
Issue
Spirometry
2
July
Unit
1998
SPll-
Power
The
1
Supply
mains
approximately voltage generator
When
An
ON/OFF
directly from the keyboard
overvoltage
undervoltage The mains
dc supply (+US) and activates signal
connected (i.e. the
A
Battery low signal compensates only
at
Note:
CPU
and
supply
is
full
15V
and
to
produce
mains
is
not connected,
control logic switches
on
either the
is
detected on
LED
is
lit
directly when mains
unit
(BLOW)
for
voltage drop when the printer stepper motor
1O.W.
The battery voltage is
off
(WFF)
Processing
wave
28 V
depending on
+UD
+UD
and
5.2V
+US
is
running on
is
rectified
or
also
to
produce
(13.5V).
is
the
battq
+UD
to
+m
an
unregulated
mains
voltage.
This
chges the battery when
voltage.
three
voltage
regulators.
voltage
then held on from the CPU (signal
24V
supplies directly switches
(indicating over current) the
is
connected. The same circuit
+BAIT
battery
generated when
monitored directly
when
power).
battexy
the
by
the
unit
is directly switched off.
unit is switched on and
voltage
(+US)
is
active and the Blow signal is active
the CPU which switches
when the voltage falls below approxhately
Circuits
dc
supply
(+W
of
between
is
used
by a switched
mains
is
connected.
The unit is switched on
WE).
unit
off.
also
monitors the switched
falls
Detection
Similarly when an
mains
to
11.3V.
A
the
9.4V.
is not
circuit
unit
of
-\
Overall control and coordination
of
the SP-1
Memory
Program Memory
An
EPROM contains the unit software. The EPROM
Static
The battery backup for
Serial
The serial EEPROM
RAM
RAM
memory stores
EEPROM
Memory
data
retention when the unit is switched off.
(U12)
the
data
and comprises
stores the unit
is
base
by
68331CPU.
has
two
128
settings.
128kByte of memory.
kbyte
RAM
chips. The static
RAM
has
-.
.,
Page
2.4
Loading...
+ 86 hidden pages