Welch Allyn SP-10 Operator Manual

Page 1
.
Cardio Menu
Operator Manual
.
..
..
.
SCHILLER
HANDBOOK
SP-I0
Spirometry
Unit
Issue
2
July
1998
SP-10
Spirometry
Unit
July1
998
SCHILLER
Altgasse
CH-6340 Baar, Switzerland
41
Phone:
Fax:
Home page:
41 760 87 87
i414176108
http://www.schiller-ag.ch
i
68
AG
80
Article
Number:
2.
540
006
Page 2
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
Issue
1:
Issue
2:
Associated
Guide
to
SCHlLLER Interpretation
and Measurement SCHILLER SCHILLER SCHILLER SCHILLER
SP-10
May 1996 July 1998, Update and
Documents
Program
SP-I0
USER
SP-I0
USER GUIDE
SP-10
USER GUIDE - French
SP-10
USER GUIDE -USA
E / D I F
GUIDE - English
Service
Article
Number
revision
-
G~~III~II
Handbook
2.540
to
incorporate
006
latest hardware and
Article Article Article Article Article
software.
No. 2. No. 2.
No.
2.510
No.
2.
No.
2.
510
510
510
510
179
130 129
131
277
0
Copyright 1998 by SCHJLLER
AG
93/42/EEC 'Notified
Body'
Medical Devices:
DEKRA
AG
0124
Page 3
-_
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometv
Issue
Unit
2
July
I998
DECLARATION
Diagnostic
Serial numbers Year
of
System
starting
manufacture:
:
with:
OF
CONFORMITY
Spirovit
SP-10
040. 1997
Onwards
We, the undersigned, hereby declare that the medical device (class above conforms with the essential requirement listed in Annex
93/42/
E
EC
.
This
declaration is supported by:
Certificate
of
approval
No.:
11425-01 45112-60-01 45112-16-01
IS0
9001
(REV.
1994)
IS0
9001/08.94EN466001/ 12.93
Annex
II,
Section 3 of
EN
46001
the
directive
1
of
IIa)
specified
EC
Directive
by
SQS
by
DEKFUmd
93/42fEEC
(€01
24
Baar (Switzerland) Dated 30.06.1998
J.J.
Schmid Markus Butler
Research
&
Development Manager Quality Assurance Manager
...
111
Page 4
Where
to
Obtain
Service
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
1
I
I
J
1-
America
Asia Pacific
Austria
I
WELCH ALLYN SCHILLER Inc., 92121,
USA Telephone: Fax:
SCHILLER Asia Pacific, Petaling Jaya, Malaysia
Telephone:
Fax:
SCHILLER Handelsgesellschaft 24,
A444 Linz, Austria
Telephone:
Fax:
800-535-6663 3 15-685-3361
(603) 777 5336 (603) 777 5744
0732 70 9 90 0732 7157
Canada
SCHILLER France S.A., Zac des Luats
F
-94350
France
Germany
(EU-Bevollmiichtigter)
Telephone:
F2.X:
SCHILLER Medizintechnik GrnbH, Rudolf Diesel D Telephone:
Fax:
Hotlie:
Villiers sMarne, France
01 49 41 24 40
0149412449
-85521
Ottobmnn, Germany
089 629 981 089 609 089 629 981 36
10A Jalan
OOO
0
50
9
0
,7420
Carroll Road,
SS
3/33,
Taman University,
mbH
Medizintechnik, , Kampmiillerweg
58,
route de Champigny,
Str.
San
14,
Diego, CA
47300
India
IdY
Switzerland
All
Other
!
Countries
SCHILLER Healthcare India Pvt. Ltd. Kalpatarn Chambers, Nanik Motwane Lane, Ground Floor, Fort, Mumbai Telephone:
Fax:
ESPAOTE SPA (SCHILLER)) Via di Caciolle Telephone:
Fax:
SCHILLER Reomed AG,
Telephone:
Fax:
SCHILLER Telephone: Fa:
I
Home page:
+91 22 263 4381 +91 22 263 4384
+39 055 422 91 +39 055 422 920 8
Riedstrasse
01 741
02
09
01 740 37 10
AG,
Altgasse
41
41
http://www.schiller-ag.ch
68,
CH-6340 Baar, Switzerland 41 760 87 87 41 761 08 80
iv
400 023
15,1-50125
14, CH-8953
Firenze, Italy
Dietikon, Switzerland
I
Page 5
SCHILLER SERVICE
SP-
I0
HANDBOOK
Warranty
Disclaimer
Spirometry Unit
Issue 2 July
1998
The Information in this guide given
as
to the correcmess regarding the contents of this changes any person
in
the specification
of
such revision
has
been carefully checked for reliability; however no guarantee
of
the contents and SCHILLERmakes no representations
manual.
of
the product described within at any time without obligation
or
change.
We
reserve the right to revise
this
document and make
or
warranties
Trademarks
SCHILLER and property
of
Copyright
0
Copyright transcribe, store in electronic, mechanical, magnetic, optical, chemical,
publication without express written permission
Terms
of
Warranty
The SCHILLER SP-10 is warranted against defects in material and manufacture for the duration of one year accident or part. Any liability for subsequent damage unqualified persons attempt
in
case
OC
SP-10
are registered trademarks
their owners.
of
SCHILLER
AG.
All
trademarks are the
Notice
I998
by SCHILLER
a
retrieval system
(as
from date of purchase). Excluded from this guarantee
as
aresult of improper handling. The warranty entitles free replacement of the defective
a
defect, send the apparatus to your dealer or directly
AG. All
or
to
make repairs.
rights
reserved.
tranrfate into any language, in any
of
SCHILLER
is
excluded. The warranty is void if unauthorized or
You
manual
may
not reproduce, transmit,
form
or
by
or otherwise, any part
AG.
is
damage caused by
to
the manufacturer.
any
is
to
notify
means,
of
this
an
The manufacturer apparatus
O
persons authorized by him, and
O the
instructions.
THERE
WARRANTIES
MERCHANTABILITY THE
if:
assembly operations, extensions, readjusbnents, modifications,
unit
ARE
PRODUCT OR PARTS THEREOF.
can
only be held responsible for the safety, reliability, and performance of the
and approved attached equipment is
NO
EXPRESS
HEREINABOVE SET
OR
OR
F?TNESS
&PLIED
FORTH.
FOR
WARRANTES WHICHEXEND
A
PARTICULAR PURPOSE
V
used
in
SCHlLLER
or
repairs are carried
accordance with the manufacturers
BEYOND
MAKES
NO
WARRANTY
WITH
RESPECT TO
out
THE
by
OF
Page 6
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spiroinetry
Issue
2
July
Unit
I998
Safety
TO PREVENTELECTRIC PARTS INSIDE.
DO
OR
IFTHEDISPLAY IS DAMAGED, ALEAKAGEOFFLUIDMAY OCCUR.DONOTWHALE
THE VAPOUR FROM THIS
CONTACT IS MADE, CLEAN CONTAMINATED AREA IMMEDIATELY WITH FRESH
WATER. THIS PRODUCT IS NOT DESIGNED FOR STERILE USE. SWITCH THE UNIT DONOT, UNDER
IN LIQUID. DO NOT OPERATE
MAINS DO NOT USE HIGH TEMPERATURE STEFULISATION PROCESSES (SUCH AS
AUTOCLAVWG). DO NOT USE E-BEAM OR GAMMA RADIATION STERILISATION. DO NOT USE SOLVENT CLEANERS
Notices
REFER
NOT USE
THE PRESENCE
THIS
ANY
LEAD
IS DAMAGED OR SUSPECTED
SHOCKDO
SERVICING TO QUALIFIED PERSONNEL ONLY.
UNIT
IN
AREAS
OF
FLAMMABLE
FLUlD
OFF
BEFORE CLEANING
CIRCUMSTANCES, IMMERSETHE UNITORCABLE ASSEMBLIES
THE
UNIT
NOTDISASSEMBLETHEUNIT. NO SERVICEABLE
IF
WHERE
AND
THE
THERE
GASES SUCH AS ANAESTHETIC AGENTS.
AVOID CONTACT
EARTH CONNECTION IS SUSPECT OR
IS
ANY
DANGER
WlTH
AND
DISCONNECT FROM
OF
BEING DAMAGED.
OF
MOUTH
EXPLOSION
AND
THE
SKIN.
MAINS.
IF
IF
THE
,I
/
,
AND
USE ONLY ACCESSORIES SCHILLER AG. USE RESULT IN INJURY INACCURATE INFORMATION
THE SP-10 COMPLIES WJTH
AFFORDS
HOWEVER SPECIAL CARE MUST BE EXERCISED HIGH FREOUENCY EOUIPMENT.
IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN
WHEN NON-MEDICAL DEVICES
ENSURE
POTENTIAL.
PROTECTION AGAINST EMISSIONS
THAT BOTH UNITS
OF
OTHER
OTHER PARTS RECOMMENDED OR SUPPLIED BY
THAN
RECOMMENDED OR SUPPLIED PARTS MAY
AND/
OR DAMAGE TO THE
EMC
REGULATIONS FOR MEDICAL PRODUCTS WHICH
AND
ELECTRICAL INTERFERENCE.
WHEN
IF
THESE
ARE
CONNECTED TO
ARE
SECURELY CONNECTED TO THE
THE
ARE
EARTHED).
THE
SP-10 IS USED WITH
RS-232 INTERFACE
SAME
UNIT.
EARTH
vi
Page 7
SCHILLER SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
-_
Unit
2
Julv
I998
Safety
Notices
U"
NONMEDICAL DEVICES, BEFORE USING THE
UMT SCHILLER REPRESENTATFE.
THE
OVERVIEW
THE
PHYSICIANS OR PERSONNEL AND RECORDING MUST BE CLINICAL CONDITION. THE RECORDING PREPARATION QUALITY GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DATA
ACCURACY, MUST ALSO BE TAKEN IT IS
EXPERT OPINION ON INDICATED.
OPERATING
FUNCTIONS
GUIDELINES FOR PATIENT ELECTRODE PLACEMENT
ONLY. THEY
SP-IOSPIROMETRYUNITIS
GRAPHICAL RESULTS
THk
PHYSICIANS RESPONSIBILITY TO
THE
UNIT,
AND
THE
EXAMINED
THE
UNIT ON BATTERY
THE
RS-232 INTERFACE
ENSURE
SAFEI'Y
ARE
NOT
THAT
PRECAUTIONS
A
SUBSTITUTE
AND
SIMULTANEOUSLY USING
MUST
BE FULLY ISOLATED.
AN
INTRODUCTION REGARDING
HAS
FOR
MEDICAL EXPERTISE.
PROVIDEDFORTHEEXCLUSIVEUSE
UNDER
RESULTS,
THEIR DIRECT SUPERVISION. THE NUMERICAL
AND
ANY
INTERPRETATION DERIVED
WITH RESPECT
INTO
ACCOUNT.
MAKE
AND
TO
INSTITUTE
TO
THE PATIENTS OVERALL
THE
DIAGNOSIS OR TO OBTAIN
CORRECT TREATMENT
THE
BEEN PROVIDED BY A
ARE
PROVIDED AS ON
OFQUALIFIED
FROM
AND
THE
IF
A
vii
Page 8
-_
SCHILLER
SERVICE
HANDBOOK
SP-I0
Issue
Spirometry
2
July
Unit
1998
I
What's
THE SERVICE PHILOSOPHY FOR MODULE
TION
AND EDGE
WITH
The book
Chapter
The purpose of to give the operation and function of each key briefly explained. The information in
a
background to the
SCHILLER SP-10
Chapter
This chapter provides a functional overview of the SP-10 The functional description is supported by functional block diagrams.
Chapter
in
this
Book
ALL
SCHILLER
LEVEL. THE
NECESSARY
REPLACE
OFTHE
STANDARD
a
2
SP-I0
is
divided into
1
-
Operating Elements
this
basic inbroduction to
-
Functional Description
PURPOSE
TO
ENABLE
A
FAULTY MODULE.
BUT
DOES
REQUIRE
WORKSHOP
the
chapter is
operating
Operating
PR.4CI'ICE.S.
following
to
provide
the
SP-10.
functions
Manual.
OF
THIS
BOOK
THE
SERVICE ENGINEER
THIS
BOOK
THATTHE SERVICEENGINEERIS
chapters:
an
easy reference for
This
chapter gives details of the operator controls with
only.
Complete operating information is provided in the
3 - Fault Diagnosis and Check Procedures
UNITS
IS
TO
PROVIDE
ASSUMES
all
IS
FAULT FINDING TO
ALL
THE
INFORMA-
TO
EFFICIENTLY LOCATE
NO
DETAILED
FAMILIAR
the
main
operator functions and
this
chapter provides
-i
KNOWL-
This chapter provides a guide to locate a fault logical sequence of fault finding algorithms and procedures. Illustrations are provided the text where needed.
Chapter
replaceable modules are also provided of and after replacement. Functional checks that must
also
Chapter
This Chapter provides the part numbers and reordering information for
Also
adjustment or fault finding procedures.
Chapter
The full technical specification
Chapter
This Chapter explains diagrams.
4
-
Physical Overview & Module Replacement
This chapter gives attributes of the unit briefly described. The physical description is supported by illustrations showing the internal location of all modules. Removal and Replacement instructions for
tools,
jumper settings, adjustments, and settings or special requirements that
provided.
5
-
included in this chapter
6
-
7
-
an
overview of the physical construction of the
Spare Parts
are
Technical Data
Glossary
all
the acronyms and signal titles used in this book and
details of any
of
the
to
module level. The diagnostics are presented in
SP-10
with the
in
this
chapter. Each procedure is autonomous with details
are
required before
be
carried
out after replacing a new module
all
replaceable modules.
special
test
SP-
10
is given in
equipment or
this
chapter.
special
in
tools
the SP-10 circuit
to
main
required for
a
support
physical
all
are
viii
Page 9
SCHILLER
SERWCE
HANDBOOK
SP-10
Spirometry
Issue
Unit
2
July
I998
What’s in this
Circuit
Diagrams
The circuit for information only.
diagrams
Book
&
Board
and component layouts
The philosophy of
ment
Our
information and reserve the right to revise improvements at any time.
Your
suggestions and comments are welcome. Please contact
the
SCMLLER
Layouts
are
provided for
We
Need
Your
SCHILLER
aim
is
to provide the
the
latest
Technical Documentation Department:
is
one of continuous improve-
user
technological developments. We
this
document and make changes
all
Help
with the most up-to-date
boards.
These
details
or
are provided
Fax
41 41
761
03
34
ix
Page 10
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
X
Page 11
SCHILLER
SERVICE
SP-I
0
Spirometry
HANDBOOK
Issue
Unit
2
July
I998
c_
Chapter
Operating
I
Elements
Chapter
I.
Operating Elem en
1
ts
Contents
introduction
~Tv
SCHILLER SPIROVITSP-I0
The
Keyboard
Alpha
Numeric Keyboard General Purpose Keys Control and Function Keys Test Function Keys
Connector
panels
Side Panel Back Panel
Mains
Panel
1.2
I.2
1.4
1.4
1.5
1.6
1.7
1.8
1.8
1.9
1.10
Page
1.1
Page 12
Chapter Operaring
I
Elements
-
SCHILLER
SERVICE
SP-10
HANDBOOK
Spirometry
Issue
2
July
-_
Unit
1998
Introduction
/
The
SCHILLER
The
SPIROVIT
capable of working in four measurement
0
Forced
0
Vital
capacity
0
Expired or Minute Ventilation
0
Maximum In addition options enable resistance testing, and blood oximetry diagnosis The
test
results are displayed on the integral
Theopenpneumotacho transducers
for respiratory measurement.
Three
RS-232
The unit The
(designated
also
SP-10
SP-10
is
a
Vital
Capacity
Voluntary
interfaces and one
has
can
also
SEMA),
ventilation
facilities for connection
be
connected to the
for the validation and archiving
sophisticated compact work-station for pulmonary diagnosis. It
SP-20/SP-1 lO/SP-150,
RS-422
SPIROVIT
modes
as
follows:
LCD
and recorded on the built-in thermal printer.
interface are provided for
of
an external video monitor.
SCHILLER
SP-10
withdisposablemouthpiece,
PC
based
of
recorded
data
Iransmissiodreception.
data
management program
data.
to
be
carried out.
are
is
used
Page 1.2
Page 13
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
July
I998
Chapter
Operating Elements
I
Introduction
The main operating and connection modules of the
Keyboard
LCDScreen
Thermalprinter The printer provides hard copy of test results. Right-hand Side Panel
Left-hand Side Panel
RearPanel
(cont.)
SPIROVIT
SP-10 are
as
follows:
The keyboard comprises the general function and control
data
keys, and the alpha-numeric keys for The LCD screen displays the pulmonary traces and
operating and
the
display
entered
side
The RS
serial
is installed, the disk drive
This
panel contains the paper
status
information. Under operator control
also
gives menu options
data
The display is folded
down
panel contains the Spirometry connector and the
COM~C~~IS.
When the floppy disk storage option
is
installed on
tray
entry.
certain
and
displays operator
when not
this
side
in
use.
panel.
and (below the paper
my) the program pack. The
rear
panel
the
and
mains
mains
contains
connector
fuse.
the video monitor connector and
with
odoff switch, voltage selector
Page
1.3
Page 14
Chapter
Operaing Elements
I
SCHILLER
SERVICE
SP-I0
HANDBOOK
Issue
Spirometry
2
July
Unit
I998
The
Keyboard
Alpha
Numeric
The
alphanumeric
Keyboard
keyboard
serves
as
a
normal
keyboard
for
data
input.
Page
1.4
Page 15
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
July
1998
%_
I
Chapter
1
Operating Elements
The
Keyboard
General
Purpose
The general purpose
(cont.)
Keys
keys
to
the
right
of
the alpha-numeric keypad,
Call
up
main
menu /return
to
Monitor mode
Return
to
previous menu
are
as
follows:
The direction keys are
used
to
move the cursor in order
Change setting / disable
command
Change setting / enable
to
make menu selections
command
Page
1.5
Page 16
Chapter Operating Elements
I
SCHILLER
SERVICE
SP-I0
HANDBOOK
Issue
Spirometry
2
July
Unit
I998
The
Keyboard
Control
and
The
Control
function key
(cont.)
Function
and
function
are
as
follows:
Keys
keys
are
situated
the
Switches
Switches the unit
Extend Paper Tray Retract Paper Tray
unit
above
on
off
I
the
alpha
numeric
keyboard.
The
control
Print the best test result
all
Print
Select test result for print
Initiate data transmission
Initiate data storage
test results
and
Note:
Entry
Start selected test
Stop printout paper to start position
Slight
functional differences may
of
patient data
/
move
exist
in different modes
Initiate receive mode
of
general
the
settings
Enter details Medical Assistant
Display menu
Delete stored
measunnent
-.
Page
1.6
Page 17
SCHILLER SERVICE
HANDBOOK
SP-I0
Spiromtv
Issue
Unit
2
Jury
I998
Operating
Chapter
Eletnents
I
The
Keyboard
Test
Function
The function keys
and
their functions:
(cont.)
Keys
are
arranged
in
a
group
in
Display test reults
Flow
the middle
(against Volume)
of
keyboard. The
following
is
a
summary
of
the keys
Volume (against
Forced Vital Capactity test (Exhalation only)
Forced Vital Capactity test (Exhalation and Inhalation)
Vital Capacity
Minute Volume (volume in one minute)
Minute Volume (forced volume in
Reistance test (Option)
Test pre and post (medication)
Spo2
test (Option)
time)
one
minute)
Calibration
of
the
unit for the environmental conditions
Page
I.
7
Page 18
Chapter Operaring Elements
I
c..
SP-I0
SCHILLER
SERVICE
HANDBOOK
Spirometry
Issue 2 July
-
Unit
1998
Connector
Side
Panel
panels
The following are located
1. 1
X
RS-232C interface (Port
2. 1 X RS-232C interface
3. 1 X
RS-232C
4.
1
X
RS-422 interface (Port
5.
LCD contrast control
6.
Connection for SP-20, SP-110 or SP-150 flow sensor for lung function measurements (with or without resistance test facility)
7.
SpO, connection
interface (Port
on
the side panel:
1)
(Port
2)
3)
4)
Page
1.8
Page 19
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry Unit
Issue
2
July
I998
Chapter
Operating Elements
1
Connector
Back
Panel
panels
(cont.)
0
The
following connectors are located on
1.
Identification Plate
2.
MainsModule
3.
Connector
4.
Potential equalisation (ground) connection
5.
Local area network Connector (Option)
for
external video
0
the
back panel:
Page
1.9
Page 20
Chapter
Operating
I
Elements
Connector panels (cont.)
SP-10
SCHILLER
Spirometry Unit
SERVICE HANDBOOK Issue
2
July
1998
~
-3
I
.I
.
Mains
Panel
The following are located on the
1.
Mains
fuse
(200
mA
slow) and
2.
Mains
connection
0
mains
voltage
panel:
selector
3.
Voltage
4.
Power on
indicator
I
off
switch
(mains)
Page
1.10
Page 21
SCHILLER SP-I0 Spirometry Unit SERVICE HANDBOOK Issue
2
July
1998
Chapter
Functional Description
2
SP-10 Basic Block
Brief Description Overall
Power Supply
Processor
Interface Board
Functional
Power Supply Processor and Memory Printer Control
Graphics Control
Miscellaneous InputdOutputs
OdOg
Control and Low Battery Voltage Monitoring
+5V and
Battery Controller
CPU Program Pack Memory
Communications Controller Thermal Print Head Controller Programmable Printer Timer Graphics Controller
Video Power
Thermal Printer Control Paper Carriage Control
kI2V
Board
-
LCD Converter
On
Reset
and
Description
Switching Regulators
Chapter
Functional Description
Contents
Interconnection Diagram
and
Beeper
2.2
2.3
2.5
2.5
2.5
2.5
2.5
2.6
2.8
2.8
2.8
2.9
2.11
2.11
2.11
2.11
2.12
2.12
2.13
2.13
2.13
2.13
2.14
2.14
2.15
2
Page
2.1
Page 22
Chapter
2
Functional Description
i
SP-10
,MAINS
El-
SCHILLER SP-I
SERVICE
HANOBOOK
0
Spirometry Unit
Issue
2
July
-7
1
i
1998
SP-I
lorn
Page
2.2
&
SP-10
Basic Block
and
Interconnection
Diagram
RS-232
RS-422
VDEO
(a
Page 23
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue 2 July
Unit
1998
Chapter
Functional Description
2
Brief
Description
The
SP-10
comprises the following boards and major functional
Power Supply
Processor Board
areas:
The power supply rectifies the mains input
all
the required voltages for the rechargeable the power for the printer for about
The Processor Board and contains the unitRAMmemory.
Video
conversion circuit for the LCD Program Pack containing the unit software
Print control, keyboard circuits
The
Processor
following connectors:
Serial transmission for
12V
batteries. connected in series, provide
and
45
minutes,
signal generation circuits and the videoLCD
RS-232
data
evaluation and storage.
when mains is not present.
is
themaster control for
main
CPU, system timing circuits and the
This
boardalsocontains the following:
Board communicates externally over the
and
to
other units or
SP-10
enable
timing,
RS-422
to
electronics.
the
SP-10
and graphics control
interfaces for data
an
external computer
and
generates
to
operate
all
peripherals
Two
Interface Board
LCD/Keyboard Buffer Board
Video connector for connection of an external monochrome signal reflects the current LCD display.
Spirometry connector for connection of a spirometer
(SP110).
The Interface board contains the following:
Paper mark detection circuit Thermal print head
Stepper motor control circuit
Control circuit for the paper tray dc motor
This board buffers the
the high voltage circuit for the LCD.
VGA
standard monitor. The digital video
data
buffers
LCD
and keyboard
data
and contains
Page
2.3
Page 24
x
SNlVW
tllOLI-dS
Page 25
c_
SCHILLER
SERVICE
SP-I0
Spirometry Unit
HANDBOOK
Issue
.
..
-
2
July
1998
Functional
Chapter
Description
2
,.
.
'.
..
.
.,
...
Overall
Power Supply
Mains
The voltage selection switch connects the supply to
transformer. The ac voltage +27V. required by the
The power is controlled by the following conditions:
0
0
0
Two 12V batteries, connected in series, directly provide the power for provide power for the switching regulators when battery
Processor
The SP-10 is controlled by a high speed, CMOS bit address bus. The clock speed is 16.7 MHz. The processor works gate array which comprises measured unit software comprises 256kSytes
Functional
is applied
This
On signal from the keyboard Off signal from the Microprocessor (Off key pressed on keyboard, 5V supply below 4.75V) Overcurrent condition on the +12V
is
to
voltage
SP-I0
odoff
monitored by a battery charge monitor circuit.
and
Memory
is
256
kByte of static
data
and other variables and an EEPROM holds the operator selectable parameters. The
Description
the power supply viaamains filter,
is
then rectified to produce a resultant dc voltage
is
applied
electronics.
control circuit switches the supply voltage
custom programmed
to
regulator
RAM
circuits
rail
to
perform specific
and 2MByte of dynamic
of
EPROM
odoff
the
relevant winding
which generate
mains
is
not connected. The charge state
68OOO
processor with a 16-bit data bus and
contained on the program pack.
switch andavoltage selection switch.
of
the mains isolation
of
approximately
all
of
the dc power supplies
to
the regulators. The power odoff
the
thermal printer, and
of
the
23-
in
conjunction with adedicated
tasks.
RAM.
The working
The dynamic
RAM
RAM
memory
holds the
The keyboard
Keyboard Interface circuit When a key board for interpretation
Printer
Control
The Thermal Print Head The print head controller serialises the patallel The Timer circuit controls out.
The printer motor speed
is
a matrix style circuit which is periodically scanned by the processor via the
is
pressed the key information is clocked to the processor
by
the master CPU.
is
controlled by the Print Head Controller circuit and the Timer circuit.
data
written by the CPU into a
how
long current is applied to the head, and thus the intensity
is
controlled by the master CPU
via
the Timer circuit.
16-bit FIFO
of
the print-
register.
Graphics Control
The display The graphics processor to
an
/
data
LCD signal
is
generated in the graphics processor circuit under the control of the master CPU.
outputs
a VGA standard serial video signal. The video signal is converted
by
the Video - LCD circuit.
Page
2.5
Page 26
Chapter
Functional Description
2
-_
SCHILLER
SP-I0
HANDBOOK
Spirometry Unit
Issue
2
July
1998
I
Overall
Miscellaneous
Spirometry
The
clockedintoashiftregkterin
on the
Video
The connector reflects
Serial
The
sion signal level compatibility with required
Functional
Description
Inputs/Outputs
serial
input
signal
from
the spirometer
theinterfacecircuitand, whenenabledthe
data
bus
for
interpretation by
the
Connector
video
connector outputs the
the
Same
serial
video signal generated by
data
as
that displayed on the LCD.
Ports
Communications Controller contains
and
reception
or
of
data
The Communications Driver circuit contains interface circuits
RS-232.
when longer cable lengths
The
are
(cont.)
is
connected to the interface circuit. The serial
datais
present din parallel
master
CPU.
the
USARTS
RS-422
used.
which generate
connector
is
graphics
serial
used when high
processor.
signals
for
data
The video
the
transmis-
to rates
data
is
ensure
are
'/
I
4
Page
2.6
Page 27
SCHILLER SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
July
I998
Chapter
Functional Description
2
Page
2.7
Page 28
...
Chapter
2
Functional
Power
Mains
filter incorporates switch-on. After a few seconds the resistor supplied. The voltage switch ensures
is
The transformed ac input approximately generate the
The switching regulator comprises a pulse widthregulator chip
FET(Ql).TheFETactsasaswitch
D4
The pulse width set by by the feedback on the current sense
VRI).
An
exceeds The overvoltage protection circuit monitors the output voltage viaQ3 which
mains
switching regulator is cut
Description
Supply
is applied
used for
and output filter
The frequency
overvoltage protection circuit cuts the voltage
+35V.
is
connected
to
the line filter via the onfoff switch and
a
tempexature variable resistor
warms,
that
the correct
the
mains voltage applied.
is
full-wave rectified by a bridge rectifier
+22V.
The rectified voltage
+uB
supply
.
tocontrolthecurrentflow inloadingcoilT1 andthus, viadiode
((289,
C13 etch regulate
U1.
is determined by
(via
is
between
The overvoltage protection circuit is formed by Q30 and associated components.
to
the
40kHz
unit.
When
by
switching off Q5.
is
the
an
Q1 power
and
5001rHz.
an
overvoltage
SCHIUR
SERVICE
two
500
mA,
to
ensure that
the resistance drops and normal power is
primary
used by a switching regulator and filter circuit
the
Winding
Wl)
voltage of
oscillator circuit
FET),
+UB.
(C8
and the voltage feedback from
to
the power supply
is
detected,
the
HANDBOOK
slow-blow fuses. The line
current
of
the isolation transformer
to
produce a dc voltage of
which controls a power
and
C9)
if
+UB
is
dc input voltage
SP-I0
Spirometry Unit
Issue
is
limited
at
initial
MOS
which is controlled
+UJ3
(via
output voltage
switched on when
to
the
UB
2
to
July
I998
A
short
circuit protection circuit formed by undervoltage which indicates high current and detected
A
too low (in effect ensuring
circuit uses uses the same tapping.
On/Off
The supply for the of the odoff control circuit is Q1 and then latched by the the processor unit is switched off.
The voltage voltage drops below overcurrent or overvoltage
Q5
is
switched off.
low current detector circuit
that
a
tapping from T1
Control and
+5
Low
and
+12V
OFF
goes
low, Q11 is switched off, the power to the switching regulators
of
the battery is monitored by comparator
18.8V
(battery voltage low)
(+12V)
is
incorporated to monitor the output current and ensure that it
the output voltage
to
monitor the current. The overcurrent detector ( ZCUDE of
Battery Voltage Monitoring
regulators
1.
It
is
initially switched on by the
signal (logic
1)
is detected, Qll is switched off via
+5V and k12V Switching Regulators
The
5V
and
12V
step down switching regulatorsboth function
given here is for the 12V regulator.
Q4,
C2
and associated components, monitors for
a
possible short circuit. If an undervoltage is
does
not exceed
is
controlled by an onfoff control circuit. The switch
from
the
Microprocessor. When the
U6C
Q11
is switched off. In the same manner
+27V).
The low current detector
ON
signal from the keyboard,
(monitoring the
+US
powerrail).
OFF
signal from
is
cut, and the
U7B.
in
the samemanner. The description
is
U1)
If
if
not
the
an
Page
2.8
Page 29
-..
c_
SCHILLER SERVICE
1
-..
Power
SP-I0
Spirometry
HANDBOOK
Supply
PowerFETsQ9 andQ10formapush-pull circuitwhichswitchesthe supply toloadingcoilT?,.The
power
FET
control circuit is formed by U5A
feedback
FETs
Secondary windings of T2 provide voltages of 13V and 25V. These
regulators to produce the regulated -12V
distribution
is
formed
(via
is
I
Unit
Issue
2
July
1998
(cont.)
R47)
to
control the switching oscillator
by
U4.
given on the following table.
IJlJlJl
Functional Description
which
monitors the
W5B,
and
-18V supplies respectively. The power supply
ac
feedback (via C24) and
C17). A gate
are
booster
used by linear voltage
for
the power
Chapter
dc
2
Battery
Controller
The battery controller circuit measures the ChargeJdischarge current CPU to calculate the capacity of the battery. The battery controller circuit
LED
to give a visual indication when the battery is charging and when the
power.
The negative side
an
gives amplifiers U15C and U15A form a current rectification
which is fed to a current frequency convener Circuit. The current/frequency converter provides frequency output that of the current/frequency converter clocks discharge current and chargeldischarge time of the battery, and thus battery.
The frequency
mately 31.3Hz. The counter counts up when the banery is discharging and
count is converted charge state is
The counter is preset when the battery is which acts Via Q23, the
(via
indication to the counter of the current direction ie charging or discharging. Operational
OV.
Q24)
of
the battery is connected
is
directly proportional to the battery charge or discharge current. The output
of
the currenrffrequency converter at a charge or discharge rate of 1A is approxi-
to
an
analog
value by
of
the battery. When the battery is
as
acurrent sense and
PRESET
to
light
signal goes low and the counter is preset to
the battery
goes
LED
when the battery is charging.
to
acurrent sense circuit.
and
an
up/down counter which is used
down
D/A
converter U13 for
fully
charged ie counter
fully
charged. This
high when the charging rate of the battery is less than IOmA.
is
of
the battery
also
Operational
measurement circuit the output
to
the
actual capacity
when the battery
use
by
the CPU in determining the
all
’1’s
carried out by a comparator U14C
all
-1’s. This circuit is
to
enable the
controls the battery
unit
is using battery
amplifier U15B
indicate the charge/
is
charging. The
the DAC output
also
of
used
of
a
the
Page
2.9
Page 30
1
RS-232
PORT
RS-232
PORT 2
3
AS-232
PORT
PORT
RS-422
VIDEO
CONNECTOR
Diagram
Block
Functional
-
Processor
1
SP-
Page 31
SCHILLER SP-I0 Spirometry Unit
SERVICE
HANDBOOK Issue
2
Julv
I998
Chapter
2
Functional Description
.
--
I
..
.
Processor
Board
CPU
Overall
tray caniage is the
of
Program Pack
The Program Pack comprises two EPROM chips containing the SP-10 software. The EPROMs each have 256 or 5
by
Memory
Dynamic RAM
control and coordination
68000
the gate
To
decode
motor).
and it works
array
is by the CPU
circuit
the address bus (A17 to
are
of
all
circuits
(U46)
in
conjunction with a
as
follows:
A23)
situated on the Microprocessor Board. The CPU
and generate the chip select (CS) signals for the
memory and the peripherals.
To
generate interrupt priority
signals
(PLO, IPL1, IPL2) for
highest priority interrupt is ’7’ the lowest -1’. To generate the row address strobe
multipiex control signals for the
12
Byte of memory which is addressed by Address lines A0 to
(RAS),
Dynamic
RAM.
the EPROM CS signal.
and
all
peripherals (with the exception
dedicated
gate array IC (U47). The main functions
the
main
processor - the
of
column address strobe (CAS) and address
A20
and enabled
the paper
used
The Dynamic generated
by
the gate
RAM
comprises four 1 Mbyte
array
(U47). The
control protocol at approximately
Printer pixel pattern Printer text data
Static RAM and
The
two
128k
EEPROM
RAM
chips (U25 & U26) are addressed viaaddress lines A0 to A19; they are enabled
by the CSSRAM signal. The static
not
lost when the
The static
unit
is
RAM
stores the following:
switched off.
Test Data Patient
data
The
SP-
10
EEPROM
64k
EEPROM chip (U27). It shares
memory stores some
RAM
modules. The
RAM
data
is
transparently refreshed
ZkHz.
The Dynarmc
RAMS
use the back-up power supply
user
defined parameter settings and comprises a single
an
enable signal (CSRTCC) with the real time clock
RAM
stores the following
RAM
control signals are
via
so
that the stored
CAS before
data:
data
(U30).
RAS
is
Page
2.1
I
Page 32
Chapter
Functional
Processor
2
Description
Board
Communications Controller
The Communications Conaoller carries Out the following functions:
Controls the
serial
(cont.)
ports
SCHILLER
SERVICE
SP-I0
Spirometry
HANDBOOK Issue
2
July
Unit
1998
1
d
I
Provides synchronization
The communications controller comprises
ROM
to
control the
Controller and the
CPU,
serial
interfaces and
is
via a dual
of
the
peripheral
devices
a
single chip processor
Data
port
RAM
(U13)
YO
with
ports.
which
the CPU
(U14)
with
Data
between the Communications
acts
as
a
data
the CSCC signal.
Temperaturesignalfromthethermalprintheadisinputto
the
ADconverterofthecommunications
controller processor.
Communication over the controllers monitor
"I7
RS
ports
two
RS
ports
PORT
1
present
1
PORT2-notusedat
is via
as
follows:
-
not
Dual
Communication Interface Controllers. Each of the
used
at
"I6
PORT
TRANSMISSION
]
PORT4-
The enable signals for the communication controllers are generated by U14 converter
(LJlS).
built in
buffer. It
3
-
DATA
via
a
RAM
is
selected by
binary
to
octal
and
Thermal Print Head Controller
Page
2.12
The Thermal chips and fust out
10 input,
in the required format for the printer.
so
that each pulse is repeated eight times.
TPD
is
MHz
temperature control for the print head
programmable timer), which is
Print
latch
buffers.
(FIFO)
memories
8
output
the print head
serial
data
Head Controller circuit comprises programmable generic array logic
GALS
clock. The output
Data
to
be
printed
is
presented
(v60
and
U61)
in
two
are programmed to serialise
To
ensure asuperior quality print the print head
This
serial
data
signal, TPSO and
to
the printer
is
set by the
strobed
by signal TGATE (6kHz).
on
the
data
bytes (lower
dara
from the
is carried out by
TPS
1
are printer strobes, and
is
synchronized
TPDUR
signal (thermal print duration from the
-
DO
FIFO
bus and
strobed
to
D7,
upperD8
FlFOs
memories and send
memory U59.
with
the 1.9kHz clock. The
(GAL)
through Fist in,
to
D15). The
data
is
'cycled'
TP
clock
is
the
it
4
Page 33
SCHILLER SP-10
SERVICE
HANDBOOK
Spirometry
Issue
Unit
2
July
c-
1998
Functional
Chapter
Description
2
L
Processor
Board
Programmable Printer Timer
The Timer circuit comprises U28 and encoded on the
are as
follows:
MCLK
ISYS
TPDUR
ALBEEP,
The timers are addressed of the master clock input, is 5
data
KBBEJZP
Graphics Controller
A
VGA standard signal
31.8
kHz
and a vertical sync of associated components. The video controller also generates (HSYNC, VSYNC etc).
(cont.
bus. The
The duration CPU, from the resistance ambient temperature
Audio signals
by
address line A1 and
(480
1
and
Beeper
U29
which decode the timing instructions from the CPU
Timers
are
accessed by the CSEE signal. The timing signals produced
The printer motor clock. The faster the speed faster the paper feed.
A
general purpose timing clock of
(determines sampling
that
rate,
the print head is
of
of
the print head.
for
the
alarm
A2
to select the required timing. The frequency
MHz.
rows
with
640
pixelskow
60
Hz),
is generated
by
2
Wz,
used
printing
with
rate
etc.)
heated.
This is calculated
the print head and the measured
and keyboard beeper.
a horizontal sync of approximately
a graphics processor chip
all
the necessary protocol signals
of
the motor the
as
system interrupt
(U39)
by
the
and
The parallel which together convert the 16-bit parallel output from the video controller to the required serial video output. Two
picture freeze when required. The video signal to the VideoLCD Converter is gated with signal
SCINV
Certain control and enable
Video
-
The videoLCD converter chip (U74) generates LCD and converts the lower control and memory for the converter chip
Power
The Power on reset circuit is formed by master reset of the CPU. This circuit has three functions
Provides a delay on to give the
Ensures that the system is inoperative and in Reset mode when the the Program Pack
Disables the unit
data
output of the video controller
RAM
chips
(U31
and
which inverts the screen image (changes the LCD
signals
from the
LCD
Converter
serial
video signal into the correct format for display. Four upper and four
data
signals
are output to the LCD along
is
provided (U75).
On
Reset
U54
initial
switch on
20ms
reset time required by the
is
not present
if
the +5V
rail
drops below 4.75V.
is
fed
U32)
form a latch circuit for the pixel memory and allow
video
controller
all
and associated components. The circuit controls the
to
ensure that the power supply is
68000
to
the
timing
processor.
two
8-bit shift registers
to
white on black image), when set.
are
further decoded by
and strobe signals required by the
with
two
clock signals. A working
as
follows:
SYSEN
(U33
fully
stabilized and
indicates that
GAL
and
U34)
U38.
Page
2.13
Page 34
-_
Chapter
Functional
InteMace
2
Description
Board
Thermal Printer Control
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometn,
Issue
2
July
Unit
I998
*.
The Thermal hnter stepper motor from the processor gives a paper speed of incorporates a current sense circuit is
active and the motor stopped. Pull-up resistors on the output lines protect against spikes from the
stepper motor. The printer
and
TpS1)
Because a stable supply to the
The paper mark detector
operational
optocoupler PM output
of
is
the enable signal for the controller and the MCLK signal sets the speed;
2.5
data,
the strobe and the clock signals
are
2H2,
the clock (TPCLCK)
the high power peaks
thermal
is
amplifier
is
is
(IJ4)
to set the
0.4V
to
0.6V
logic 0 when a paper mark
is
controlled by stepper motor controller
mm/s,
2.09kHz
drawn
print head (via
an opto sense circuit. The output from the detector is fed
PM
for
white
gives a paper speed
If
the current
by the
signal when a paper mark is detected. The input from the
paper and
is
drawn
are
buffered by
is
4MHZ
WH.
the
controller).
3.5V
detected.
is greater than
and the latch frequency
reservoir
to
4.OV
of
U5.
The strobe frequencies (TPSO
capacitors on the
for black paper ie paper mark. The
25
U1.
The MON signal
mm/s.
The driver circuit
100
mA
the sense signal
(TPL)
+24V
is
209Hz
2kHz.
rail ensure
to
-,
an
Page
2.14
Page 35
SCHILLER
,
SERVICE
SP-I0
Spirontetty
HANDBOOK
Issue
Unit
2
July
1998
Functional
Chapter
Description
2
3
Interface Board
(cont.)
Paper Carriage Control
REPUCE
PAPER
-0-
4-
FROM
CPU
I
I
RUN.
STOP
GlCT
I
CURRENT
SENSE
PAPER
MOTOR
TRAY
The paper tray carriage motor control circuit works independently from the power supply which extended, the power supply
Paper
Pressing the
signals
+UB
by the The
by U6C,
U7A
Tray
REPLACE
to
a preset circuit state. Signal
supply. cUop
5Vod
supply.
5Vod
also
U6D
and
U7B.
polarity motor sense circuit and when is fed back via a delay circuit, incorporated
motor
to
is
initially activated.
is
enabled only when the paper
is
inactive.
Not
Extended
acts
PAPER
is
regulated by
as
the
‘start
key switches the power supply on and presets some internal
EJCT
switches transistor
U4C
to generate the
signal’ to operate the motor via a delay and timing circuit formed
and associated components. This sets the
The flip-flops enable the motor confrol chip
(GO
IN,
GO
OUT
signals). The motor controller incorporates
the
paper tray
is
fully
extended the motor controller
to
reset the flip-flops and stop the motor. The delay circuit
my
is
in
use; when the paper tray
47
on
to
generate +Uop from the
+5V
supply (5Vod)
ENABLE
U8
and
INIT
and set the direction of the
sets
prevent the motor being stopped prematurely during the power surge when the
CPU
and
has
its
is
and
47
is latched on
signals
to
flip-flops
24V
dual
an
‘H’bridge current
the sense line which
(C34)
own
not
is
Page
2.15
Page 36
-..
Chapter Functional Description
Interfuce
2
Board
(cont.)
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry Unit
Issue
2
July
I998
-
,f
L
-
%>9
Paper
When the paper tray is extended, pressing the and activates the signal to the motor controller. current sense circuit
When the tray the +Uop
The flip-flop control circuit monitors the printer motor motor via the REPLACE
Tray
has
rail
and switches
run
PAPER
Extended
EJECT
signal.
stops
returned, the
signal
(paper
key
is
pressed when either the printer motor, or paper tray motors are active.
REPLACE
This
sets the clock input
In
the Same
the motor when the paper tray
DIRIN
off
the power supply.
tray
manna
signal
is
active which switches
motor active).
as
when the paper
This
PAPER
to
has
via
ensures that the circuit
key switches
the flip-flops and activates the
tray
fully
returned.
Q9
the
MON
signal, and the paper tray
off
is being extended, the
on, and
Q7
is
transistor
off.
This
disabled
GO
if
Q8
IN
cuts
i
the
Page
2.16
Page 37
SCHILLER
SERVICE
HANDBOOK
SP-10
Spiromtry Unit
Issue 2 July
1998
Fault Diagnosis
and
Check Procedures
Chupter 3
Chapter
Fau
and
Contents
Introduction 3.2 Test Equipment 3.3
Fauif
Diagnosis
&
Warnings
SP-I0
SP-10 Initial Fault Diagnosis Chart (Sheet
SP-I0
SP-I0 Fault Diagnosis Chart (Sheet SP-I0 Fault Diagnosis Chart (Sheet C - General Problems) 3.10 SP-I0 Fault Diagnosis SP-I0 Fault Diagnosis Chart (Sheet
SP-I0
Check
General Confidence Check and Calibration Procedure 3.14 Power-Supply 3.17
+
UB
Power Rail Measurement 3.19 Printer 3.20 Printer Resistance 3.20 Print Head Alignment 3.21
RS Interface Testing 3.22
RS Self Test with Test Plug
RS-232 TransmissionlReception Test 3.26 Interface Board Adjustments 3.29
Paper Mark Adjustment
Service Screen
Unit Software 3.33
Replacing the
Cautions
Initial Fault Diagnosis Chart (Sheet
Fault Diagnosis Chart (Sheet A - Power Problems)
Chart
(Sheet D - Printer Problems)
Fault Diagnosis Chart (Sheet
and
Adjustment Procedures 3.14
Adjustment 3.19
and
Printer Resistance Setting
Flow
Sensor
Filter
Check
I)
3)
B
-
Power Problems)
G
-
Spirometry Problems) 3.12
H
-
Communication (RS) Problems) 3.13
It
Diagnosis
Procedures
3.4
3.5
3.6
3.7
3.5
3.9
3.11
3.22
3.30
3.31
3.34
3
Page 3.1
Page 38
Chapter
Fault
3
Diagnosis
and
Check Procedures
Introduction
SCHILLER SP-I0 Spirometry
SERVICE
HANDBOOK
Issue 2 July
Unit
I998
The purpose
identify afault test equipment
An
initial
When this is followed it will indicate
of
this
chapter
on
an
is
kept
fault
diagnosis char&
SP-10
to a
is
to
provide fault-fmding procedures
to
a
specific module. The fault-finding procedures are designed
minimum.
is
provided detailing
the
module where the fault lies or specify a further
all
the
finding procedure. When more than one module is stated, likely
to
contain the
checked in
If
the
settings
the
initial fault-finding chart
and
check the software. If When
a
key operation ormenu selection replacement, the key sequence required parenthesis
is
fault.
If
the fault
is
not found on
order given.
does
not indicate
parameters that have been entered (for
this
is correct suspect external connections.
is
required when carrying
is
given
the
actual charact&) printed on
the
the
first module,
the
area
the
particular
in
parenthesis
key. When a key sequence
be followed in the order given.
that
will quickly and efficiently
general fault indications
the
first module given
the
other modules should
where
the
fault exists, re-check
task
that
fails).
If
these are correct,
out
the functional checks after
‘e’.
The chmcter(s) given
is
provided it must
that
may occur.
is
the
so
that
fault-
one most
be
all
the
in
Page
3.2
Page 39
SCHILLER SP-I0 Spiromtry Unit SERVICE HANDBOOK Issue
2
July
I998
Fault
Diagnosis
Chapter
and Check Procedures
3
Test
Equipment
Thefollowingproprietaryanddedicatedtestequipmentistheminimumthatisrequired
and
carry
out
any
board
The
list
of
proprietary equipment
test equipment
can
be
obtained
checks and
adjustments,
is
not comprehensive. Recommendations
from
the
Schiller
on
the
SP-10.
Service Department
of
suitable proprietary
Dedicated Test Equipment
0
RS
Test Cable Assembly required for the
RS
self
test
Proprietary Test Equipment/Tools
0
Calibration Pump (2litles or more)*
0
oscilloscope
0
Digital Multimeter*
0
Stabilised
0
Standard tool general tools*
0
SolderingIron
0
VGAMonitor
Power Supply
kit
with
a selection of cross-bladed and flat-bladed screwdrivers, Pliers and
to faultfind
*
Indicates Essential Equipment for basic
Special
The following table
or
Tools
/
Equipment
lists
all the special tools and equipment that may be required when fault finding
carrying out certain maintenance procedures
I
RS TEST CABLE ASSEMBLY 3-WAY CONNECTORS
2MM HEXAGONAL KEY FOR THE REMOVAL
/REPLACEMENT
HEXAGONAL CROSS-BLADED SCREWDRIVER USED FOR THE REMOVAL
AND
OF
THE PROGRAM PACK
REPLACEMENT
SP-10
OF
THE PRINT
fault finding
on
the SP-10.
HEAD
1
2.310042
STANDARD
AVAILABLE ON
1
THEMARKET
NUMBER4.950074. MANUFAO
NUMBER
367TX
TOOL
SCHILLER
BY
WERA,
-
WEFA
210180
I
Page
3.3
Page 40
Chapter
Fault
3
Diagnosis
Fault
-_
and
Check
Procedures
Diagnosis
-_
SCHILLER SP-I0 Spirometry Unit
SERVICE
HANDBOOK
Issue
2
July
1998
i
i
‘1
/
Use the fault finding charts and procedures on the following pages to indicate a faulty module.
When apossible When applicable. The Removal and Replacement instructions set-up
In
most cases the
a
module
or
check
fault
finding
charts
faulty
module is indicated by the fault fmding
bas
been
replaced
procedures
specific test
required,
are given in
should indicate the most likely
charts.
the module must
parameters
Chapter
and setting-up
for
all
replaceable modules, along with any
4.
of
faulty
area.
be
the module may
area
or
replaced.
be
Page
3.4
Page 41
--.
SCHILLER
SERVICE
SP-I0
Spirometry
HANDBOOK
Issue 2 July
Unit
I998
Warnings & Cautions
r
MAINSPOMRISPRF.SENTWHENTHEUNITCOVERISREM0VED.CERTAINCHECKS
AND
ADJUSTMENTS
REMOVED
BEWARE
THE
STATIC PRECAUTIONS. ALWAYS PLACE
MATWHENCARRYINGOUTANY~~~PROCEDURES.PERSO~MUST
AND
THAT
SP-10
CONTAINS STATIC SENSITIVE
BEEARTI-IEDWHENHANDLINGANYBOARDSORC0MPONENTS.ALWAYSUSEAN
ANTI-STATIC BAG
CAN
MAINS
CONNECTED.
POTENTIALLY
WHEN
WARNINGS
ONLY
BE
CARRIED
WHEN
LETHAL
VOLTAGES
CAUTIONS
THE
TRANSPORTING
Fault Diagnosis and Check Procedures
OUT
WITH
THE
TOP ASSEMBLY
CARRYING OUTTHESE PROCEDURES
ARE
PRESENT.
CMOS
COMPONENTS; OBSERVE
UNIT
ON AN
BOARDS
EARTHED
OR COMPONENTS
ANTI-STATIC
ANTI-
Chapter
3
CARE MUST BE TAKEN WHEN USE FORCE. NEVER
STRAIN
THE
REMOVING
CABLE ASSEMBLIES.
AND
REPLACING CONNECTORS. NEVER
Page
3.5
Page 42
Chapter Fault
Diagnosis
3
THE
and
UNIT
AND SWITCH
Check Procedures
SCHILLER
SP-10
Spirometry
SERVICE HANDBOOK Issue
2
July
Unit
1998
A
IS THE GREEN
\/
I
DISCONNECT THE
MAINS
SUPPLY AND
CHECK THE
FUSE
IM
THE
MAINS
MAINS
I
I
REMOVETHETOP
ASSEMBLY AND CHECK
FUSES
SI
1
AND
SI
4
/
YES
1
GO
TO
NO
1
I
2
INVESITGATETHE
REASON
FOR
RUPTUREDFUSE
ON
THE
I
FAULT DIAGNO-
SIS (SHEET A
PROBLEMS)
-
POWER
Page
3.6
PRESS THE SP-LO
<ON> KEY
DOES
THE LCD
\/
V
SHEET
TWO
NO
SP-10
GO
TO FAULT DIAGNO-
GENERAL PROBLEMS)
I
Initial Fault
1
Diagnosis
REASON FOR THE
RUPTURED FUSE
CHECK THE BATTER-
Note:
The removal instructions
all
and
Chapter
Chart
(Sheet
REPLACE THE POWER
SUPPLY BOARD
and
replacement
and
the location
boards,
cable assemblies
connectors are given
4.
1)
of
in
Page 43
SCHILLER SERVICE
HANDBOOK
SP-I0
Spirometry
Issue 2 July
Unit
1998
Fault
Diagnosis
Chapter
and
Check Procedures
3
a
MESSAGE DIS-
SENSOR TO THE SP-I
CONNECTOR ON THE SP-LO.
PRESS THE PATIENT DATA
KEY AND ENTER SOME
PATIENT DETAILS. PRESS THE FVC E KEY FOLLOWED BY START. BLOW INTOTHE
FLOW SENSOR.
FROM
SHEET
ONE
TYES
SOR BOARD AND POWER
SUPPLY(P1). CHECK
CABLE ASSEMBLY
BETWEEN POWER
SUPPLY AND K/B/LCD
BUFFER. REPLACE
I0
CHECK THE LCD
CONTRAST CONTROL
ON THE RIGHT-HAND
PANEL OF THE
CONNECT AN EXTER-
NAL VGA MONITOR TO
THE
vmeo
ON THE SIDE PANEL
UNIT.
CONNECTOR
I
PROCESSOR BOARD
AND PROGRAM PACK
REPLACE THE
GO TO FAULT DIAGNO-
SIS CHART (SHEET B
POWER PROBLEMS). IF
THE POWER SUPPLY IS
I
OK CHECK THE PRINTER
MOTOR.
IYES
A
GO TO FAULT DIAGNO
GO TO FAULT DIAGNO
SIS
CHART (SHEET D
SIS
CHART (SHEET D
I
GO
TO FAULT DIAGNO-
SIS
CHART (SHEET G
SPIROMETRY PROB-
GO
TO FAULT DIAGNO-
SIS CHART (SHEET H
FUNCIION? PROBLEMS)
COMMUNICATION
-
-
-
-
-
1
I
I
I
I
I
A
I
INTERFACE BOARD TO
lYES
SP-10
Initial
REPLACE
me
SENSOR AND REPEAT
mST.
THE
STILL PRESENT
REPLACE THE
PROCESSOR BOARD
CHECKCABLE
ASSEMBLY BEIWEEN
POWER SUPPLY AND
INTERFACEBOARD
(no).
ASSEMBLIES FROM
PRINTER. REPLACE
1. PROCESSOR BOARD
2.
3.
THERMAL PRINTER
FLOW
IF FAULT
CHECK CABLE
UP
BOARD
Fault
Diagnosis
I
1
I
CHECK ALL PARAM-
ENSURE THAT PARAM-
mERS ARE
ANCE. CONTACT SCHILLER CANNOTBE FOUND
Chart
(Sheet
IN
IF
2)
FAULT
TOLER-
NO
ENSURE THAT THE
BAlTERIES ARE FULLY
CHARGED. REMOVE
THE PAPER
.
WORMDRNE JAMMED. CHECKTHE DC MOTOR FOR OPW
CIRCUIT. CHANGE THE
JXTERFACE BOARD
TRAY
IS
NOT
Page
3.7
AND
ENSURE THAT THE
MECHANICAL
Page 44
Chapter Fault
Dkwnosis
3
and
Check
Procedures
SCHILLER
SERVICE
HANDBOOK
SP-10
Issue
Spirometry
2
July
Unit
1998
,
MAINS
FUSE
RUFTURE
ii
REMOVE
THE TOP ASSEM-
BLY.
VISUALLY
ALL CABLE ASSEMBLIES
AND
CONNECTORS. ENSURE
THAT
ALL CONNECTORS ARE SECURE. LOOK IN THE AREA OF THE TRANSFORMER AND
MAINS
mpm
BURNING OR SCORCHING.
FOR
msPEcr
SIGNS
OF
NO
RECTIFY THE
THE PROBLEM.
REPLACE FAULTY
COMPO"T(S).
CAUSE
OF
FLYING
LEADS
THE
FROM
MAINS
THE MAINS TRANS-
FORMER TO THE WWER SUPPLY.
REPLACE
A
I
POWER SUPPLY.
SUPPLY
DISCONNECTTHE
INTERNAL CABLES
THE MAINS ASSEMBLY.
REPLACE THE MAINS
FUSE AND SWlTCH
hums
TRANSFORMER.
REPLACE THE MAINS
TRANSFORMER
TO
ON.
Note:
MAINS ASSEMBLY.
REPLACE
I
The removal instructions
all
boards,
and
connectors are given
Chapter
THE
ASSEMBLY
and
and
cable assemblies
4.
MAINS
replacement
the
location
in
of
Page
3.8
SP-10
Fault
Diagnosis
Chart
(Sheet
A - Power
Problems)
Page 45
,
SCHIUER
SP-10
Spiromtry
SERVICE HANDBOOK Issue
REMOVE THE TOP
ASSEMBLY. VISUALLY
mspecr
ALL
ASSEMBLIES
CONNEffORS. ENSURE
THAT
TORS ARE SECURE. CHECK THE CABLE
ASSEMBLY
POWER SUPPLY
.
TOPASSEMBLY.
POWER SUPPLY.
CABLE
AND
ALL CONNEC-
FROM
mE
TO
Unit
2
THE
July
I998
Fault
Diagnosis
Chapter
and
Check Procedures
3
F.3
REPLACE
ASSEMBLY BETWEEN
BUFFER BOARD (BASE ASSEMBLY). REPLACE
THE LCD/WB BUFFER
BOARD.
mE
CABLE
THE
POWER SUPPLY
AND THE LCDIK/B
ASSEMBLY
TO
REPLACE
TOP
me
DISCONNECT THE MAINS.
REMOVE THE CABLES
NO
THE MAINS ASSEMBLY. SWITCH ON THE MAINS
AND MEASURE FOR AC
.MAINS
AND
3.
REPLACE THE
TRANSFORMER
NO
REPLACE THE POWER
ACROSS
AND PINS 2 AND
pms
MAWS
SUPPLY
TO
1
4
REPLACE THE MAINS
ASSEMBLY
Note:
The instructions
all
and
Chapter
removal
and
and
boards,
cable assemblies
connectors
are given
4.
replacement
the location
in
of
SP-10
Fault Diagnosis Chart (Sheet
B
-
Power
Problems)
Page
3.9
Page 46
Chapter
Fault
Diagnosis
3
and
Check
Procedures
ENSURE THATTHE PROGRAM
PACK
IS
INSERTED. REMOVE THE
TOP ASSEMBLY. VISUALLY
INSPECT ALL CABLE ASSEMBLIES
AND CONNECTORS.
ALL CONNECTORS ARE SECURE.
P,"sURE
'THAT
SCHILLER
SERVICE
HANDBOOK
SP-10
Spirometry Unit
Issue
2
July
1998
l-
I
I
CHECK ALL POWER
RAILS AS DETAILED IN
'POWER CHECKS
PARAGRAPH
SUPPLIES OK?
CHECK THE LCD
CONTRAST CONTROL.
CONNECT AN
EXTERNAL VGA
TO
I
AND
THE
me
POWER
MONlTOR
VIDEO CONNECTUR
MONlTOR DIS- SOR BOARD AND
CHECK THE CABLE ASSEMBLY BE-
'
WEEN
THE
LCD/K/B BUFFER BOARD (BOTTOM
ASSEMBLY
THE CABLE ASSEMBLY BETWEEN THE
PROCESSOR
SUPPLY (Pl). REPLACE THE LCD/K/B
BUFFEJt
REPLACE THE VIDEOLCD CONVERTER
POWER SUPPLY AND THE
TO
TOP ASSEMBLY). CHECK
BOARD
BOARD. REPLACE THE LCD.
(PROCESSOR BOARD)
REPLACE THE POWER
SUPPLY
REPLACE THE PROCES-
PROGRAM PACK
Note:
The removal
instructions
all
boards,
and
connectors are given
Chapter
and
replacement
and
the-location
cable assemblies
4.
of
in
Page
3.
I0
SP-10
Fault Diagnosis Chart (Sheet
C - General
Problems)
Page 47
SCHILLER SERVICE
HANDBOOK
SP-I0
Spirometry Unit
Issue
2
July
I998
CORRECT PRIEmR
IS
RESISTANCE
MEMORY AND THATTHE
PRINT HEAD AMBm
TEMPERATURE
CORRECT AS DmAILED
THIS CHAPTER.
STORED IN
IS
Chapter
Fault Diagnosis and
IN
Check Procedures
3
NO
QUAW. FRESH PAPER
IS
INSTALLED INTHE
TRAY.
PAPER
IF
THE PAPER MARK
DETECTOR AND
CURRENT LIMlTER
THRESHOLDS AS
DETAILED IN
CHANGE
NECESSARY.
STOP ATTHE CHAPTER. REPLACE
STOP ATTHE CHAPTER. REPLACE
THIS
RESISTANCE
AMBIENTTEMPERA-
NRE. REPLACE
1.
THERMAL PRINT
HEAD
2.
PROCESSOR BOARD
CARRY OUTTHE
CARRY OUTTHE
PRINTER CHECK
PRINTER CHECK
DETAILED
DETAILED
THE THERMAL PRINT
THE THERMAL PRINT HEAD IF NECESSARY
HEAD IF NECESSARY
ENTERTHE
CORRECT PRINER
RESISTANCE
DETAILED IN THIS
CHAPTER.
IN
THIS
IN
THIS
SP-10
DOES THE
PROBLEM STILL
EXIST?
REPLACE:
1.
PROCESSOR BOARD
2.
INTERFACE BOARD
3.
THERMAL PRINTER
Fault
Diagnosis
Chart (Sheet
D
Note:
-
Printer
The removal insmctions
all
boards,
and
connectors are given
Chapter
and
and
cable assemblies
4.
Problems)
replacement
the
location
in
of
Page
3. I I
Page 48
..
Chapter
Fault
Diagnosis
3
and
Check
Procedures
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
1998
CHECK THE
FLOW SENSOR. ENSURE
I
THATlTIS NOTBLOCKED
AND THATm
REPLACE
AS
I
FILTER
IN
IS
CLEAN.
IF
NECESSARY
DETAILED
CHECK ALL SPIROM-
ETRY
ENSURE
CONNECTIONS ARE
CABLE
ENSURETHATTHE
CALIBRATION
CORRECT AND THAT
ALL SE?TINGS ARE
IN
CHAPTER.
CABLING.
THAT
SECURE.
I
OK?
CORRECT.
THIS
ALL
IS
THE
1
I
NO
I
REPLACE THE FLOW
SENSOR.
Page
3.12
SP-10
WSSIBLE OPERATOR
ERROR.
Fault Diagnosis Chart (Sheet
CHECK FLOW SENSOR.
CHANGE:
1.
PROGRAM PACK
2.
PROCESSOR BOARD
Note:
The removal instructions
all
boards,
cable assemblies
and
connectors are given
Chapter
G
-
Spirometry Problems)
4.
and
replacement
and
the
location
in
of
Page 49
SCHILLER SP-I0 Spirometry Unit SERVICE
HANDBOOK
Issue
2
July
232).
SETlTNGS(
BJT) AND THE
ARE THE
PRESS <SETnNGS>
232).
SETlTNGS( LINE OR MODEM,
RECORD OR BLOCK) ARE
CORRECT FOR THE APPLICA-
I998
CHECKTHAT ALL
PARITY. STOP
BAUD
IN
RATE
BOTH
<4>
SAME
DEVICES.
CHECK THAT ALL
(RS-
Fault
Diagnosis
Chapter
and
Check Procedures
-.
3
YES
SELF TEST PROCEDURE
DETAILED
IN
TMS
I
A
PASSED?
TEST
POSSIBLE OPERATOR
ERROR.
CHECK ALL
ASSEMBLIES. REPLACE
1
IF
NECESSARY
CABLE
ENSURE COMPATIBILITY
BETWEEN THE COMMU-
r
1.
2.
CONNECTOR ASSEMBLY
TO
NICATING DEVICES
CHECK THE
12V SUPPLY VOLTAGES
AS DETAILED
CHANGE THE PROCES-
3.
Note:
+12V
The
IN
5.
removal
PARAGRAPH
SOR BOARD
CHANGE THE RS
AND
instructions
all
boards, cable assemblies
and
connectors
Chapter
1
-
and
replacement
and the location
are
4.
given
of
in
SP-10
Fault Diagnosis
Chart
(Sheet
H
-
Communication
(RS)
Problems)
Page
3.13
Page 50
.L.
....
Chapter
Fault
Check
3
Diagnosis
and
and
Check
Procedures
Adjustment Procedures
General Confidence Check and Caiibration Procedure
The procedure detailed here is a general confidence check in the board has been replaced. It
be
present after the
10
proceed
1.
Ensure that the voltage selector
mains
2.
Connect the flow sensor Connect a caliiration pump to the sensor.
3.
Switch the unit on by pressing the
and that for
test screen disappears, check that
are
4.
Press <SETTINGS> complete. Ensure that the correct date and time is displayed on the screen.
5.
Press the <CALIBRATION> key. Enter the ambient temperature and press Press <START> and empty the calibration pump*. Press ‘Effective VoT, enter
as
follows:
to
the
unit.
a
few seconds
shown in the
is
not a full functional test but is intended
unit
has
been reassembled.
Switch on the
SP-110
data
box
<1>
the
To
carry
on
the back panel
mains
and
ensure
to
the connector on the right hand panel
<ON>
key on
the
test screen is displayed at the bottom of the screen. When the
the
welcome display
in the right hand corner of
and enter the time and date details. Press <FNCT> when
calibrated volume of the pump, in litres.
SCHILLER
SERVICE
unit
out the general confidence check
is
set
for the required voltage and connect
that
the green mains
the
Keyboard. Ensure that the LCD lights
is
given and that the
the
LCD.
HANDBOOK
after an internal module or
to
identify any faults that may
LED
SP-10 Spirornetry
of
lights.
of
the unit.
date
and time
<ENTER>.
<ENTER>
and at the prompt
Issue
the
SP-
2
July
Unit
1998
SNote:
To
obtain a more accurate calibration, it is recommended that the calibration pump
is emptied more
6.
Repeat step 5 and ensure calibration gives the following
0 0
0
0
0
0
7.
Ensure that the print-out
can
be
MeasuredVol
Effective Vol
Deviation
Temperature BTPS Factor
Calib. Factor
than
that
obtained
data:
:
: :
:
is
once
to
achieve a total volume
the measured volume is correct. If required a print-out
by
pressing
: :
:
even and of good quality.
the
<PRINT
The calculated volume using the existing calibration. The actual volume received by the
calibration pump volume) entered manually by the user. The difference between the measured volume and the
actual volume The ambient temperature entered by the user The Body Temperature Pressure Saturated factor
compensates for humidity, and allreadingsapartfrom
to
the ATPS (Ambient Temperature Pressure Saturated))
The calibration factor is individually set for the attached transducer and used by
the
SP-10
is
applied
in
volume calculations
of
more than 6 litres.
SEL>
key. The print-out obtained
SP-10
the
difference in inhaled and exhaled
to
all
subsequentreadings (that is,
the
calibration procedure
is
asoftware calculated factor which
of
(that
(as
opposed
is
the
the
Page
3.14
Page 51
SCHIU.FR SERVICE
HANDBOOK
SP-I0
Spirometry Unit
Issue 2 July
1998
Fault Diagnosis
Chapter
and
Check Procedures
3
Page
3.IS
Page 52
Chapter
Fault
3
Diagnosis
Check
8.
and
Check Procedures
and
Adjustment
Note:
If the print-out quality printer resistance resisrance setting see Paragraph 'Self Test Screens and Offset Adjustments'. If problems
may
Press the <PATIENT DATA> key and systematically press
are
experienced with paper
be
necessary to adjust the factory pre-sets on the Interface board.
Procedures
is
poor and the printer
data
may
have been lost from
(cont.)
has
memory.
mark
detection or setting
SCHILLER SP-I0
SERVICE
not been changed, the stored
the
HANDBOOK
To
check the stored
the
printer speed it
keys:
-_
Spirometry
Issue
2
July
Unit
I998
I
I
~
-1
<QWERTY
Ensure that the entered
to
exit the patient
9.
Select
a
test
key (eg manually blow into the flow sensor. Ensure screen and that after six seconds the
10.
Press the
and
11.
Press
Note: A full description of the programmable parameters and settings, and the procedure
12.
If a module Paragraph supply for
<PRINT
of good
quality.
<SElTINGS>
change them,
has
(if
not already carried out). Switch the unit off and leave connected
24
hours
data
data
display.
&WE>)
ALL>
key and ensure that the settings menu is displayed
is
provided
been replaced cany out the checks and tests detailed in the relative
to
fully charge the battery.
UIOb,
is correctly presented on the screen
and press <START>. With
OR
<PRINT BEST> key and ensure that the print-out
in
<ENTER>,
that
unit
prompts for the second xeading.
the SP-10 Operating Manual.
<1234567890>
the
the pulmonary
calibration pump, or
as
entered.
graph
is
Press
<FNCT>
traced on the
is
as
follows:
to
the
accurate
to
mains
Page
3.16
Page 53
SCHILLER SP-I0 SERVICE
-
HANDBOOK Issue 2 July
Check
Spirometry
and
Adjustment Procedures
Unit
1998
(cont.)
Fault Diagnosis
and
Check Procedures
Chapter
3
Power Supply
1.
Connect
LED
light
2.
Connect an
lights and that
3.
Press the spirograph
4.
Press
measurement
5.
Press any key and ensure
6.
Check the
procedure detailed
7.
Disconnect the mains supply and ensure that the yellow
mains
to
the
lights. (Ensure
is
extinguished when
SP-110
flow sensor
the
welcome screen
FVC
E
key followed by
is
plotted.
the
<PRINT
+12V
battery
SEL.>
is
given.
supplies and communications interface by wrying out the
in
LED
lights.
unit
and switch on.
that
when the battery
mains
key and select
Ensm
that
that
the buzzer beeps.
this chapter.
Ensure
is
fully charged
is still connected).
to
the
SP-10.
Switch the unit on and ensure that
is
displayed.
START.
the print-out
Blow into the sensor and ensure that the
Meas
1.
is
that the
Ensure
of
good
mains
mains
LED
lights and the Battery
-
maximum
that a print-out of the
quality and
LED
is extinguished and
15
hours - the battery
is
complete.
RS
the
last
test
that
LCD
the
are
If these checks
checks fail, the individual power supplies must
Power
Rail
The power supply voltages be assembled Supply is COMeCted achieved.
Test measuring points are provided on the Power Individual adjustment is not possible for any
supplies are generated and the batteries are charged).
satishctory it indicates that
Adjustment Procedure
can
be
checked
to
check the power supplies.
to
the Processor board however, a more accurate measurement may
in
If
of
all
the power supply
be
checked.
isolation: it is not a requirement that the unit has
the power supplies are checked when the Power
S
upply board for measurement of
the voltage rails except
mils
are correct.
+UB
If
any of the
all
power rails.
(from which
all
other
to
be
Page
3.1
7
Page 54
c..
-c
Chapter
Fault
3
Diagnosis
Check
and
Check
and
Adjustment
PI7
-
GND BAYONET
CONNECTOR
Procedures
Procedures
P12
-
TO
LCDKEYBOARD BUFFER
(IN
TOP
ASSEMBLY)
SERVICE
(cont.)
SCHILLER SP-I0
HANDBOOK
Issue
Spirometry
2
July
Unit
1998
.
p20
-
TO
INTERFACE
BOARD
P2.5,
€26
BATTERY
CONNECTORS
+UB VOLTAGE ADJUSTMENT
Power
Supply
VRI
Board
MK
p27
MAWS
TRANSFORMER
8-6
-
FROM
Page
3.
I8
Page 55
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
Julv
I998
Fault
Diagnosis
and
Check Procedures
Chapter
3
Check
LETHALVOLTAGESAREPRESENT~CHECKINGANDADJUSTINGTHEPOWER
SUPPLY.
TAKINGMEASUREMENTS.
ON
A
ON
1.
2.
Note:
+UB
and
Adjustment
DO
NOT
LET
Procedures
CONCENTRATION LAPSE. EXERCISE
Signal Simulation
latch circuit exists on the power supply.
signal must
Connect mains to the unit and switch the
Using a suitable
supply will now be correct voltage.
be
connected
lead,
To switch the power supply off, connect
connect
latched
to
GND. To
GND
on and
all
Adjustment
(cont.)
WARNING
GREAT
If
the Power Supply
simulate
to Plug P12 pin
generated power supplies can be checked for the
mains
the
on.
GND
board
is
checked in isolation the
ON
signal, proceed
25,
and hold for 2 seconds. The power
to
PI
pin
50.
as
follows:
CARE
WHEN
To
check and adjust
1.
Connect mains power to the unit and switch on. Latch the power supply on, above.
2.
Measure the voltage at connector
k2OOUlV.
Note:
The test point where measurement is taken voltage is present ensure that the
voltage before the On/Off switch measure the voltage on fuse
+UB
supply proceed
as
€25
pin
follows:
1.
Adjust
odoff
as
VR1
to
achieve a voltage of
is
situated after the WOff switch. If no
switch
Q11,
is switched on.
To
SI2.
detailed
+27V
check the
Power Rail Measurement
No
adjustment is possible for any
relevantconnectorpinandmeasure withaDVM.Ifany ofthe valuesareoutoftolerance firstensure
that
+UB
voltage
is
correct.
of
the power
If
this is correct the power supply must be changed.
rails.
To
check the power
rail
voltages go on the
Page
3.19
Page 56
Chapter
Fault
Diagnosis
c
3
and
Check Procedures
SCHILLER
SERVICE
HANDBOOK
SP-I0
Issue
Spirometry
2
July
Unit
1998
-_
Check
Printer
To
The printer check is entered from the settings menu; press the keypad,
A
lines are even
the print-out for evenness of print.
and
Adjustment Procedures
check the printer and to ensure that every pixel
to
display the settings menu and press
print-out of a series of diagonal lines
and
unintempted. Any faulty print-head pixels will
will
be
(cont.)
is
operational,
<p>
to commence the printer
given. Carefully examine the print-out and ensure that
a
built-in printer
~SETI”GS>
-
FAULTY PIXEL
be
seen
key
as
a
horizontal white line. Examine
test is provided.
on
the left
test
of
the
all
the
If a faulty pixel
is
detected the printer must be replaced. If
the
darker at the top than the bottom), it indicates that the printer alignment out
is
too
faint or
too
dark,
check that the correct resistance
is
Paragraph ‘Printer Resistance‘ below.
Printer Resistance
Check
that
the correct resistance of the printer
Note: The printer resistance will
1.
Display the settings menu by pressing the
be
When in the settings menu press<> (Test).
2.
R
TPH is
the
resistance
shown).
value
value
To
enter a new resistance press
is
highlighted and the correct resistance
has
been entered, press
accept a resistance range
of
the printer (See Paragraph ‘Service Screen’ for
<ENTER>
of
between
is
entered in
the
found on a label attached to the printer.
<SETTINGS>
</>
(slash key).
can
to
save the setting. Note
lo00
and
1800
key on
The
be
entered. When the correct resistance
Ohms.
print-out
is
uneven (for example
is
not correct.
entered in memory
SP-10 memory
the
left
of
as
the keypad.
the
first figure of
that
the
the SP-10 will only
If
the print-
as
detailed
follows:
display that
resistance
in
is
Page
3.20
Page 57
SCHILLER
SERVICE
HANDBOOK Issue
SP-I0
Spirometry
Unit
2
Julv
I998
Fault Diagnosis
Chapter
and
Check Procedures
3
Check
and
Print Head Alignment
Printer alignment screws
so
out.
Adjustment Procedtlres
is
achieved with the three print head alignment (tensioning) screws. Adjust the
that they
are
slightly indented
below
the frame. Adjust to achieve the best quality print-
I
P32
I
GUARD
(cont.)
p29
PRINT
DATA CABLE
HEAD
PRINTHEAD
TENSIONING
SCREWS
(X3)
VRI
P4 P23PRINTHEAD p24-TO PRINTHEAD
(TEST
POINT
1)
DATA CABLE PHOTO RETAINING
SENSOR ROD
Page
3.21
Page 58
Chapter
Fault
RS
3
Diagllosis
Interfue
RS
-
and
Interface
-_
Check Procedures
Testing
Error
messages
SCHIL.LER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
,'
The following errormessages
are
displayed,
SERIAL
TRANSMISSION or
DATA SET
carry
LINK
out the checks suggested.
TIME-OUT
ERROR
NOT
READY
are
associated with
This
remote unit (after approximately
0
0
0
0
0
This is a general fault indication
0
0 Check that the connecting cable is correctly plugged in 0 Check the integrity
the
RS
Interface.
indication appears
Check that
correct parameters Check the correct set-up in both units (<STORAGE>
key)
Check that the connecting cable
Check the integrity
Units)
If
using a modem ensure that it
the remote modem
Check that the remote unit
correct parameters
units)
the
remote unit is switched on and set
If
any
of
these
emr
if
no signal
of
the cable assemblies
of
the
is
received from the
30
seconds).
is
correctly plugged in
is
communicating with
is
switched on and set
cable assemblies (at both
messages
to
the
(at
both
to
the
NO
DATA
IN
MEMORY
RS
Seif
Test
with
Test
Plug
The RS interface self test ensures that the
functioning.
1.
Connect
2.
Switch
Select option RS-test (menu option
the
the
To
SP-10
carry
out the
RS
test cable to
on. Press
RS
self test proceed
RS
the
0
If
using amodem, ensure that it
the remote modem
A
data
transmission stored in the memory message appears, change the Processor board.
ports
4E"GS)
5)
and
RS
communication circuit on the Processor board is
as
1.2,
and 3
key
to
display the RS-test options:
has
been attempted, but no
units'
memory.
attempt the transmission again.
follows:
as
indicated on the RS plugs.
to
display the settings menu
is
communicating with
Store
a
data
is
measurement in the
If
the same
.
Page
3.22
Page 59
~
..
.
. ..
..
..
.
.~. . .
. . .
.. . . . .
..
. .
.
..
~
..
SCHILLER SERVICE
RS
SP-I0
Spirometry Unit
HANDBOOK
Issue
2
July
Interface Testing (cont.)
3.
Press
<T>
to
commence the
1998
RS
self
test with test plug.
Fault Diagnosis
and
Check Procedures
Chapter
3
The self test appears on
Test
If
this
assembly
Test
When this test passes
starts
the
Failure
test
fails
is
correct, a fault
Pass
and a pass (audio sequence)
LCD.
An
audible indication
is
or
given
fail
(TRANSMISSION
that
the test is complete.
ERROR) message
ensure that the cable assembly is wired and connected correctly. If the test cable
is
indicated on the Processor Board
it
indicates that the
SP-10
is
functioning correctly and
or
Power Supply.
that
the fault lies in
the remote equipment, the transmission line or that the protocol of the two communicating devices
are
incompatible.
Check that Reset
all
settings ( Line
settings compatible in
Check the integrity
all
settings
all
settings
(Parity,
Stop bit) and the Baud
to
ensure compatibility between the communicating devices. Check that
or
Modem, Record
both
the
of
all
cable connections. Are
or
Block) are correct for the application. Are the
uansmitting
and receiving devices?
rate
are the same
all
cables correctly connected
in
both devices.
to
the programmed
port?
Page
3.23
Page 60
Chapter
Fault
RS
3
Diagnosis
Interface
and
Check Procedures
Testing (cont.)
RS
Test
Cable
PDRT
Art,
3
No.
2.310
SCHILLER
SERVICE
042
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
I3
6
I
2
7
8
RTSl
CTS 1 (Input - Clear
DSRl
DTRl
Out ( Data Output)
In
(
Input Data)
(
Output
-
Request
(Input
-
Receive
(
Output
-
AT-60
Ground
to
to
Send)
Unit
Ready)
I
Send)
Ready)
Page
3.24
Page 61
SCHILLER SERVICE HANDBOOK Issue
RS
SP-10
Spirometry Unit
Interfme
Test
Input / Output
The
test
Use
these
the cable assemblies, connectors etc. between the communicating
Testing
transmission
test
options
2
July
1998
(cont.)
/
test
reception options enables a test sequence
to
ensure that the receiving and transmitting units
Fault
Diagnosis
to
be
generated and received.
can
communicate and that
units
are
good.
and
Check Procedures
Chapter
3
The test message generated is a string of
the
test
transmission/ reception option unit are displayed on the indicating that
To
perform these tests, you will need a second
to
be
connected
Windows 95TM. The following equipment is needed:
A
personal computer
An
the 094 and
1.
Start
Hyperterminal.
a
RS-232
SP-10
cable assembly,
with the COM
an
adapter
from Windows 95 desktop. Click on
..................................
IAT&T
;.""
LCD
in
test transmission is in progress.
to
an active terminal. for example the
(PC)
with Windows 95TM Hyperterminal installed.
port
DB
9 / DB
Maill
-...
..""....""-.
:
all
ASCII
is
selected the string
the
receiving unit.
Art.No. 2.310
of
the
PC.
24,
Art.
No.
AT-1
04
characters - ABCD
In
SP-
10
159, for connecting the
This assembly consists of a cable,
2.100
START.
....
12
M....
abcd... etc. When
of
characters sent by the transmiaing
the transmitting unit a message is displayed
unit, or the
552.
SP-10
RS-232 (#3)
Hyperterminal,
Select Programs / Accessories
which
RS-232 (#3)
interface has
is
available under
interface on
Art
No.
2.310
/
AT-4
2.
Double-click OK.
CompuSe
on
Hypertrm(.exe). Enter the name
...
Hypertrm
MCI Mail
SP-10
for the new connection and click on
Page
3.25
Page 62
Chapter
Fault
Diagnosis
3
and Check Procedures
SCHILLER
SERVICE
SP-I0
HANDBOOK
Issue
Spiromtry
2
Julv
1998
Unit
RS-232
3.
4.
Transmission/Reception
Set
the
direct connection to the proper
Enter the
port
settings
19200,8,
N,
1,
COM-port
Hardware
Test
and click
and
confirm with
,
.f
OK.
OK..
Page
3.26
5.
6.
7.
You
are now
Go
to
rab
Click on
in
the
SP-10
Hyperterminal window.
Settings and set Terminal keys,
ASCII
Setup.
Auto
Click
detect and
on File and select Properties.
500
buffer lines.
,
Page 63
SCHILLER SERVICE
SP-10
Spirometv
HANDBOOK
Issue
Unit
2
July
1998
Fault
Diagnosis
Chapter
and Check Procedures
3
RS-232
8.
You
Transmission/Reception
Check the "Wrap lines that exceed terminal width". Confirm with
back
are
boxes
to the open
now
ready
for "Send line ends with
Hyperterminal
to
start
the
window.
input and output
line
tests.
Test
feeds",
(cont.)
"Echo
typed
OK.
Click once
characters locally'' and
more
on
OK
to
get
Test Output
On the
SP-10,
Baud rate to
3
with key
Press
The Hyperterminal window on
a>.
key
SP-10
19200,
<O>
now
press the
key
<p>
(Output)
sends
<SETTINGS>
to
set parity Nand key
Test
output on channel n' on the
a
string
of
alphanumerical charactm, which
the
PC.
key and select
<S>
<5>
RS-Test.
to
set the stopbit to
SP-10.
Use
the
+/-
1.
are
echoed in the
keys
to
set the
Set Channel No.
Page
3.27
Page 64
Chapter
Fault
Diagnosis
3
and
Check
Procedures
SCHILLER
SERVICE
HANDBOOK
SP-10
Issue
Spirometry
2
July
Unit
1998
RS-232
Test Input
of
Transmission/Reception Test (cont.)
345671119:
Stop
In
the RS-232
On
;~>?~~CDEPGHI.JKLUNOEQR~T-
the test by pressing key cQ> (Quit) 'End
the
RS-232
the
test
PC,
test
window
type
any
window, press
is
the
erased to
characters and verify
key
be
able
of
<I>
(Input)
to
display incoming messages.
that
they
test
inputloutput'.
Test
input on channel n'. The
are
echoed on the
SP-10.
upper
part
3
45
678
9:
;
<=>?
This
i.3
an
Stop
the test by pressing key
Error Codes
The error codes shown
DSR BD
FE
OE
PE TXE RxR
Data set ready Break detected Frame error Overrun error
Parity error Transmitter empty Receiver ready
@ABCDEFGHIJ'KLMNOPCIRZT
RZ-232
test
-
<Q>.
at
the
bottom
Leave the RS-232 test window by pressing
of
the display have the following interpretations:
ESC.
Page
3.28
TxR
Transmitter ready
Page 65
SCHILLER
SERVICE
SP-10 SpirometQ' Unit
WDBOOK
Issue 2 July
1998
Interface Board Adjustments
Chapter
3
Fault Diagnosis and Check Procedures
One adjustment is possible
the
paper
mark
detection circuit
used
for
calculation
Paper
When
displayed. Before window
Jammed
the detection circuit does not recognise the paper
carrying
is
clean.
on
the Interface board - Paper
threshold
of
print-out commencement position, paper detection, paper present etc.
Indication
the paper
mark
P32
GUARD DATA CABLE
Mark
(VRl).
This
pre-set
is
used
to
adjust
so
that
the paper
mark
mark,
is
recognised. The paper
a paper
jam
error message
mark
may
adjustment procedure. ensure that the paper detector
P29
PRINT
HEAD
P24
-
TO
PHOTO
SENSOR
is
be
P22-TOTP
MOTOR
P4
(TEST
POINT
Interface
1)
Board
MK
TO
PAPER
CARRIAGE
8-5
,,!TO
Page
3.29
Page 66
Chapter
Fault Diagnosis
3
and
Check Procedures
SCHILLER SP-10 Spirometry
SERVICE
HANDBOOK
Issue
2
July
Unit
1998
Interface
Paper
Mark
Set the paper
1.
Connect a digital
2.
Switch the
NB
Board
Adjustments
Adjustment
mark
threshold
voltmeter across the
unit
on and adjust
>3.5v
0.4
to0.5V
IF
THE
DETECTION
ADTUSTMENT,
as
follows:
VR1
to
when the
there
when
DOES
ENSURE
THAT
(cont.)
P4
connector
obtain
a reading
black
paper
is
no
paper
NOT
RECOGMSE
THE
(test
as
mark
mark
PAPER
point
1).
follows:
is
under
the
paper
mark
detector
under the detector ie white
THE
PAPER
MARK
MARK
WINDOW
AFTER
IS
CLEAN.
paper
-1
*
-I
\
Page
3.30
Page 67
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
July
I998
Fault Diagnosis
Chapter
and
Check Procedures
3
I
,
..
.-.
i
1
Service Screen
The service screen provides readings print head resistance
To
enter the service screen press
T
(for test)
The following display
This screen gives the following
R
TPH
and
setting
is
Printer Resistance
of
some
of
the
intend
option.
the
shown:
settings key
<SEITINGS>
data
and adjustments:
to
display the Settings menu then press
<T>
Setting
offsets
and also contains the thermal
the
letter
The value indicated here is the resistance resistance to ensure a
good
quality
of
the thermal print head and must
be
set to the correct
print-out. The print head resistance is found on a label on the
printer. To set the resistance
now highlighted and the new resistance value
loo0
Ohms
and less
user
is
prompted to enter the value again. When the resistance
T
TPH
This displays the thermal print head temperature. and the
of
the thermal print head press
than
1800
Ohms.
If
a resistance value outside
d>.
The
first
figure of the resistance value is
can
be
entered. The entered value must be more than
of
these limits
is
thermal
set press
print head temperature in volts
<ENTER>.
is
entered, the
(received from the transducer).
B
TEMP
and
AIN6
This displays the battery temperature, and the battery temperature in volts (received from the transducer).
BV
and
AINO
BV
indicates the voltage level of
generated by the power factor controller from the
of
+US
voltage level
diagram)
and
shown
when working from
is calculated from the
on this screen
the
mains,
indication (signal BA'ITV) when working from the
mains
+US
(that
as
AINO.
and
between
is
4ooomV+100
is
the battery voltage, or the primary dc voltage
mains
(and
used
for battery charging)). The actual
BATW
The value
+24V
signal (sheet
of
BV
and
+21V
5
of
the power supply circuit
is
approximately
when using battery
26.8V power.
(nom
The
+27V)
AINO
when working from the battery and 440OmV+100
Page
3.3I
Page 68
..
.
Chapter
Fault
3
Diagnosis
and
Check Procedures
Service Screen
BCHARGE,
AINl
gives a voltage value indicating the loading control level of the battery (from the BAnZC). The voltage range 25OOmV
Because every battery varies, it knows when capacity of the discharged
new battery. The BCHARGE indication shows the remaining
signal
Note:
indicates
to
installed
a
minimum of
(AINl
indication).
The battery capacity displayed on the capacity remaining capacity displayed
as
that
and
Printer Resistance Setting (cont.)
AINl
and
B
CAPACITY
is
GmV
to
that
the battery is fully discharged.
the
processor initially calculates the capacity of the battery
generate the battery
battery. When anew battery
two
of
the battery and
shown
for BCHARGE.
times,
low
so
is
that
SCHILLER
SERVICE
25OOmV
signal
calculated
in
(OV
indicates
-
the B CAPACITY indication gives the calculated
is
installed,
the processor
battery
capacity and is calculated
LCD
information
from
the information
that
itmust
can
accurately assess the capacity of the
B
CAPACITY and
box
will
HANDBOOK
the battery
be
fully charged and fully
box
is the actual remaining
AINl .
not necessarily
SP-I0
Issue
PSU
-signal
is
fully charged,
so
from
BATI'L,C
The
be
the same
Spirometry
2
July
that
Unit
I998
Page
3.32
Page 69
-_
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue
Unit
2
Julv
1998
Fault
Diagnosis
and
Check
Chapter
Procedures
3
Unit
Sohare
To
check the software version of the
installed,
The software version and the
press
the setiings key followed by the letter
installed
unit
and to display
options
all
x:
are
displayed on the
the sysrem~ options
LCD
as
follows:
that
have been
The top line gives the equipment name and the overall system software version of
the uniC that is the software version of the EEPROMs in the followed by a combination of characters to indicate the type of software (or basic options) that is installed.
The second line gives the software version of the communications controller on the Processor Board
Below the software version derails is the options line. This details the options that
are
installed. These are
0
RES
The number below, and to the right of the Copyright notice, is a counter number of unauthorized attempts that have been made to incorporate extra options to the unit.
Important
upgrades
The number immediately below the unauthorized attempts counter is the system
serial
number.
The options on the bottom of the screen are the software updates that incorporated into the unit when the correct code is entered. these software options the code must and
<->
keys and the
Schiller
AG
-
COMM:
-
Lung Resistance
-
only
9
attempts are allowed and
are possible.
<ENTER>
through
your
as
follows:
local
0.1
be
entered, the option selected with
key pressed. The code
dealer.
if
this
figure
To
can
Program
exceeds
Pack. This is
of
9,
no more
can
incorporate any of
the
be
purchased from
the
be
<+>
Page
3.33
Page 70
Chaprer
Fault
Y
Replacing the
3
Diagnosis
and
Check Procedures
Flow
Sensor Filter
SCHILLER SP-I0
SERVICE HANDBOOK
Spirometry
Issue
2
July
Unit
1998
!
4
I
Spirometer
The
1.
Remove and discard the mouthpiece.
2.
Remove the rubber adapter by pulling
3.
Remove the inner tube of the flow sensor by pushing it out of
of
SP-20 / SP-110 / SP-150
filter is the only serviceable item on
the red location
marks.
Once
the
the
flow meter.
it
away from the inner tube.
tube
has
To
been pushed
replace
half
the other side.
4.
Unscrew the two halves of the inner tube and remove and discard the filter.
5.
Clean and sterilize
all
parrs of the inner tube, the rubber adapter and the inside of the outer
tube with one of the following products:
0
IncidinGG
0
Amocid
0
Lysoformin
0
Alhydex
6.
The cable and handle
can
be
wiped
with
soapy water (do not
the
fdter
proceed
the
outer tube in the direction
way
it
can
be
dip
into liquid!).
as
pulled
follows:
out
from
Page
3.34
Page 71
.
..
-4
.
...
~ -.
.
-.
I
.I
--...
1
-*
.
..
SCHILLER SERVICE HANDBOOK Issue
SP-I0
Spirometry Unit
2
Julv
1998
Replacing the Flow Sensor Filter (cont.)
7.
Insert
a new
fdter
into
location
8.
Carefully screw the
mark.
displaced.
9.
Locate the end
until
the
10.
Fit
the rubber adapter by placing its
location
11.
Insert
a new mouthpiece
of
shoulder
mark
the inner tube
of
the inner tube
two
halves
the inner
(max.
so
that
it
sits
on the
her
of
the inner
into
tube
makes conract on the outside edge
1.5cm)
tube
the end
of
wider
end over the end
into
the end
together
the outer
making
tube
of
the rubber adapter.
of
Fault
lip
Diagnosis
of
the
sure
that
half
the
and
with
filter
and push it gently but
of
the outer tube.
the inner
tube
with the red
Chapter
Check Procedures
the red
is
not
fumly
3
Page
3.35
Page 72
Chapter
Fault
Diagnosis
3
Lllld
Check Procedures
SCHILLER
SERVICE
SP-10
HANDBOOK
Spirometry
Issue
2
July
Unit
1998
~
Page
3.36
Page 73
SCHILLER SERVICE
SP-10
Spiromerry Unit
HANDBOOK
Issue 2 July
I998
Chapter
Physical Overview & Module Replacement
4
Physical Overview
Module Replacement
Contents
Physical Descr@ion
Base Assembly
Top Assembly
Removal Safety Notices Opening
Power Supply board
Leadaid
Interface
Paper
and
Rephcement
the
case
Top Assembly Removal Top Assembly Replacement Functional Check After Assembling tk SP-I0
Power Supply Board Removal
Power Supply Board Replacement Checks and Tests after Power Supply Board Replacement Checking the Power Supplies
On
Signal Simulation
+
UB Adjustment
Power Rail Measurement
Battkries
Battery Removal Battery Replacement
Checks and Tests After Battery Replacement
Board
Interface Board Removal Interjiie Board Replacement
Checks and Tests After Interface Board Replacement
Tray
Assembly
Paper Tray Removal
Paper Tray Replacement Checks and Tests after Paper Tray Assembly Replacement
Chapter
&
4.3
4.5
4.8
4.11
4.12
4.13
4.13
4.13
4.14
4.16
4.17
4.17
4.17
4.18
4.18
4.19
4.19
4.20
4.20
4.21
4.21
4.22
4.22
4.22
4.22
4.24
4.24
4.25
4.25
Page
4
4.
I
Page 74
._
-..
.
Chapter
Physical Overview & Module Replacement
4
SCHILLER SP-I0 Spirometry Unit
SERVICE
HANDBOOK
c
Issue 2 July
1998
Microprocessor
Board
Microprocessor Board Removal Microprocessor Board Replacement
Checks
and
Tests After Microprocessor Board Replacement
Program
Pack
(and Software Modules)
Program Pack
(and
program EPROM ) Removal
Program Pack Replacement
Checks and Tests After
Mains
assembly
Program
Mains Assembly Removal Mains Assembly Replacement
Checks
and Tests afier Replacement
LCD / Keyboard Buffer
MK
8-4
LCD / Keyboard Assembly LCD/Keyboard Buffer Removal L CD/Ke yboard Buffer Replacement Checks after Replacement
Mains transformer
Mains Transformer Removal Mains Transformer Replacement
Check and Tests after Replacement
LCD
LCD Removal LCD Replacement Checks after LCD Replacement
Thermal Printer and
Print
Head
Thermal Printer Removal Print Head Removal Print Head Replacement Interjime Board with Printer Thermal Printer Replacement Checks, Tests
and
Adjustments afrer Printer Replacement
Keyboard
Keyboard Removal
Keyboard Replacement
Checks after Replacement
Replacement
4.26
4.26
4.27
4.28
4.29
4.29
4.30
4.30
4.31
4.31
4.31
4.32
4.33
4.34
4.35
4.35
4.35
4.36
4.36
4.36
4.36
4.37
4.38
4.39
4.39
4.40
4.40
4.40
4.40
4.41
4.42
4.42
4.43
4.43
4.44
4.44
Page
4.2
Page 75
SCHILLER
SERVICE
HANDBOOK
SP-10
Spirometry Unit
Issue
2
July
I998
Physical
Chapter
Overview & Module Replacement
4
Physical
Introduction
The
TheBaseAssembly
TheTopAssembly
The Base Assembly contains most Assembly together and achieved with a ribbon cable assembly.
Because
Description
SP-10
comprises two
houses
are
secured with captive
of
the plastic construction, threadedmetalinsem are used
main
assemblies
of
the
the keyboard and the
mws.
as
follows:
SP-
10
electronics, and
LCD
display. The
Electrical connection between the two assemblies
all
external connectors. The Top
two
assemblies
for
all
are
moulded to fit
screw fixngs in the unit.
is
Page
4.3
Page 76
-.
Chapter Phvsical Overview
4
Physical
PRINTER
MOTOR
L.
&
Module
ReDlacemnt
Description
p32
GROUND)
i
I
BATTERY
P29 - TO
PRINTER
(DATA)
(cont.)
GROUNDNG KEYBOARD
PONT BUFFER
I
I
I I
I\
SCHILLER SP-I0 Spirometry
SERVICE HANDBOOK
P1 -TO
MICROPROCESSOR MKXUX"X
I
PI2
LCDl
I
-TO
P2
BA'ITERY
-
RS
Issue
INTER
I
2
July
:FACE
TO
3ESSOR
Unit
I998
_-
I
.-
THERMAL
PRINTER
Page
4.1
p24-To
PAGE
MARK
PHOTO (UNDER)
SENSOR
P23 -TO
PRINTER iMOTOR INTERFACE
(DATA) BOARD
P22
I
TOPAPER
I
-
TO
TP
1
,
P20
POWER - TO ~25, ~26
1
SUPPLY
!
(UNDER)
TERY CONNEC-
i
BUZZER P20 - TO P27 - FROM MAINS
Base Assembly
-
TORS
BAT-
TRANSFORMER
Page 77
i
SCHILLER SERVICE HANDBOOKIssue
SP-I
0
Spirometry Unit
2
July
1998
Physical Description (cont.)
Base
Assembly
The base assembly comprises a moulded plastic tray onto which is mounted the following:
0
Mains
Isolation Transformer
0
Two Lead-Acid Baaeries
0 Power Supply PCB
0
Interface Board
0
Microprocessor Board
0
Program
0 ThedPrinter
0
Paper Tray Assembly with Carriage Motor
Pack
MK
MK
MK
8-5
8-11
8-6
MK
8-1
Chapter
4
Physical Overview & Module Replacement
TheMicroprocessorBoardissecuredtomouldedsupportson
is mountedon fourspacers above the Microprocessor board and the Interface board and the thermal
printer
are
mounted on the metal paper tray framework. Electrical connection between
is achieved with ribbon cable assemblies The
Mains
Connector and Switch assembly is secured to the Back Panel
The
Mains
Transformer the Power Supply board, the Transformer, the Batteries, and the Mains Connector and Switch assembly,
The Program Pack is inserted via a cut-out in the side panel under the paper tray. It slides and connects in position.
Cut-outsin the
110)
mounted on dedicated mounting plates.
is
achieved by flying leads.
backandsidepanelsareprovidedforexternalco~ectors.TheVideoandSpiro
connectors are all mounted directly on the Microprocessor board.
is
secured to the tray with a centre screw. Electrical connection between
directly
with the Processor Board. One captive screw secures the
or
with flying leads.
thebottomoftray.ThePowerSupply
all
boards
with
spring loaded clips.
is
mounted
Program
Pack
(SP-
All
other connectors
are
in
Page
4.5
Page 78
..._
C.
Chapter
4
Physical Overview & Module Replacement
SCHILLER
SERVICE
HANDBOOK
SP-I0 Spirometry
Issue
2
July
1998
Unit
'_
Physical
P3
CONNECTOR
(MOUNTED ON
Description
-
VIDEO
BOARD
WORM
FOR PAPER CARRIAGE
MOTOR
(cont.)
CONNECTOR
AND
ASSEMBLY SUPPLY
DW
MAINS
SWITCH
P1- TO POWER
GROUND
~
i
P2 - TO RS
BOARD
Page
4.6
,
I
P.4PER TRAY
FRAMEWORK
Base
Assembly
without
SUPPORT
BRACKET
Interface
1
PAPER
CARRIAGE
MOTOR
and
Power
Supply
SP-100
CONNECTOR
board.
Page 79
-
SCHILLER
SERVICE
SP-20
HANDBOOK
Physical
The metal carriage
Spirometry Unit
Issue
2
July
Description
paper
tray
framework
motor
tums
a
worn drive
2998
(cont.)
allows
to
slide
the
paper
the
tray
paper
to
tray
be
in
Chapter
Physical Overview & Module Replacement
withdrawn. The
or
out.
dual
rotation
paper
4
Page
4.7
Page 80
-_
Chapter
Physical Overview & Module Replacement
4
SCHILLER SP-10 Spirometry
SERVICE HANDBOOK Issue 2 July
Unit
1998
Physical
Description
PI3
-
TO
LCDBUFEER
BOARD
(cont.)
KEYBOARD
Page
4.8
TO CONTRAST
CONTROL
TO
Top
I
LCD
Assembly
TO
LCD
CONTRAST
CONTROL
Page 81
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spiromv Unit
Issue
2
July
I998
Physical Overview
&
Module
Chapter
Replacement
4
Physical
Top
assembly
The Top Assembly is manufactured from moulded plastic and comprises hinged
Secured
of
these
achieved with ribbon cable assemblies
Description
LCD
assembly.
in
the
main
housing
boards
are
secured with double sided
(cont.)
is
the
LCDkeyboard Buffer
or
with
MK8-4
tape.
Electrical connection
flying
leads.
and
a
the
Keyboard
between
Main
Housing
MK
8-3.
the
modules is
with
Both
a
Page
4.9
Page 82
c_
Chapter
4
Physical Overview & Module Replacement
M
BASE
ASSEMBLY ACCESS
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
I I
M
TOP
ASSEMBLY
ACCESS
Y-5
Page
4.10
Note:
Access
Access
to
the Program Pack is gained via the left-hand Side Panel under the paper
to
the mains
fuse
and the voltage selector
is
from
the back panel
(Mains
tray
Module).
Page 83
SCHXLER
SERVICE
SP-I0
Spirometry
HANDBOOK
Issue
Unit
2
July
1998
Removal and Replacement
Introduction
This
part provides
spared at service level. The instructions given in
containing the following:
The pre-requisites
Tools
and equipment that
the functional checks and adjusrments. Removal Procedures Replacement Procedures
the
procedures
that
must
to
carry
be
fulfilled before removal
are
required
Physical Overview & Module Replacement
out the removal and replacement of
this
chapter
are
autonomous, with
of
the module
to
remove and replace the module and
all
modules
each
to
carry out
that
are
module
Chapter
4
Checks and Tests that must
Any
adjustments, jumper settings,
a procedure, are detailed in the relevant step. In-text
diagrams
points
and adjustment potentiometers.
Specific Warnings and cautions could cause personal injury. Cautions indicate
If
a key operation character (or character string) given in parenthesis key sequence is provided it must
time and date screen
This requires that the settings key on the keypad is pressed, followed by the number 1 key.
support the text where required and provide location details of connectors, test
or
menu selection
is
as
be
are
follows:
carried out after replacement.
special
checks or functional procedures that are required during
given where applicable. Warnings indicate potential danger that
areas
that could
is
required. the key sequence required is given in parenthesis
is
the actual character that is printed on the key. When a
be
followed
in
the order given. For example the key sequence to display the
<SETTINGS>
<1>
cause
damage
to
the equipment
.
'e'.
The
Page
4.11
Page 84
Chapter
Physical Overview
4
&
Module
Replacement
.--
SCHILLER SP-10 Spirometry Unit
SERVICE
HANDBOOK
Issue
2
July
I998
Sizfety
Notices
WARNINGS
BEFORE COMMENCING THATTHEMAINS
IS
REMOVED.
CERTAINCHECKS ANDADJUSTMENTSCANONLY BECARFUEDOUTWITHTHETOP ASSEMBLY
THESE PROCEDURES BEWARE THAT FO-Y LETHAL VOLTAGES
PRESENT.
THE SP-10 CONTAINS STATIC SENSITIVE STATIC PRECAUTIONS:
0
WHENCARRYING0UTANYMAI”ANCEPROCEDURES
UNIT ON
REMOVED
AN
EARTHED
ANY
REMOVAL
POWERSUPPLYISS~CHEDOFFANDTHATTHEMAINSCABLE
AND
WITH MAINS
ORREPLACEMENT PROCEDURES ENSURE
CONNECTED.
WHEN
CARRYING
OUT
ARE
CAUTIONS
CMOS
COMPONENTS; OBSERVE
ALWAYS PLACETHE
ANTI-STATIC MAT.
ANTI-
0
PERSONAL MUST BE EARTHED WHEN HANDLING ANY BOARDS OR
COMPONENTS
ALWAYS USE
0
COMPONENTS
THE UNIT
ALWAYS PLACE
OUT
TAKE CARE NOT TO PLACE WHEN NOTCRIMPED CAI3LE ASSEMBLY.
CARE
USE FORCE.
THE
ORDER GIVEN.
IS
MA”ANCE PROCEDURES
REMOVING
iWST BE TAKEN WHEN
PROCEDURAL STEPS GIVEN FOR EACH MODULE MUST BE FOLLOWED
AN
ANTI-STATIC BAG WHEN TRANSPORTING BOARDS OR
SUSCEPTIBLE TO ABRASION DAMAGE. TO PREVENT SCRATCHING,
THE
UNTT
ON A
SOFT,
NON-ABRASIVE CLOTH WHEN CARRYING
ANY
STRAIN ON
THE TOP ASSEMBLY.
ORTWISTEDANDTHATTHETOPASSEMBLY
REiMOVING
NEVER
STRA.tN
THE CABLE ASSEMBLIES.
THE
CONNECTING RIBBON CABLE
ENSURE
AND
THAT THE CABLE ASSEMBLY
IS
NOTPLACEDONTHE
REPLACING CONNECTORS. NEVER
IN
IS
THE
Page
4.12
Page 85
SCHILLER SERVICE
HANDBOOK
SP-I0
Spirometry Unit
Issue
2
July
I998
Chapter
Physical Overview & Module Replacement
4
Opening
the
Pre-requisi tes
The unit must
any maintenance and
28
Mains
supply is required
Tools
Posi-drive screwdriver
Test Equipment
The following test equipment is required
Flow Sensor Spirometer SP-110 Calibration pump RS Test Cable (supplied by Schiller)
Top Assembly Removal
The Top Assembly is mounted on the six screws and washers; access
the Top Assembly, proceed
1. Remove cable is removed.
case
be
placed on an anti-static
is
carried
out
on
degrees.
to
carry
to
the screws
as
follows:
all
cable assemblies connected to the side and rear panels. Ensure that the mains
mat
and anti-static precautions observed when
the SP-10. The
out the
unit
functional test after assembly.
to
carry
out the functional test after unit assembly
Base
Assembly and
is
gained from the underside of
room
temm
is
secured to the Base Assembly with
should
be
the
unit To remove
between
18
2.
Taking care up-side-down and rest on a soft cloth.
3.
Unscrew the
of the unit, indicated by
4.
Grasping the top and bottom of the unit to ensure
carefully
5.
Gently lift the Top Assembly sufficiently to gain access disconnect the connector from the
6.
Gently
Top
Assembly Replacement
To replace the Top Assembly proceed
1.
Check that all boards and components are fmly secured. Check for loose screws. Ensure
that
no screws or foreign
2,
Inspect all the internal cable assemblies and ensure that they that no visible damage or caught
3.
Ensure that
to
secure the hinged
six
retaining
rem
the
an
unit
to the standing position.
LCD
screws
and washers situated in the extreme comers and edges
arrow.
Power
lift
the Top Assembly away hm the
as
bodies
are
loose
can
be
seen. Ensure that no cable assemblies are strained, crushed
all
connectors
are
firmly home.
module
follows:
so
that
it
cannot swing open,
that
the two assemblies cannot
to
the interconnecting cables, and
Supply Board and the ground connector.
Base
Assembly and place
in
the bottom of the
case.
are
on
in good condition and
turn
the unit
part,
a soft cloth.
Page
4.13
Page 86
Chapter
4
Physical Overview
Opening
the
4.
Position the Top Assembly adjacent to the cable, plug in the interconnecting cable from the Top Assembly
(PIZ).
Note:
5.
Carefully position the Top Assembly on the
6.
Grasping down and replace the the unit. Return the unit
&
Module
Reconnect the ground cable
It
case
may
be
the
two
Replacement
(cont.)
necessary
assemblies
to
to
six
securing screws and washers in the extreme comers and edges of
to
the upright position.
tilt
SCHILLER
SERVICE
Base
Assembly and without
from
the keyboard.
the
Top Assembly for the cable assemblies to reach.
Base
Assembly.
ensure
that
they cannot part, carefully turn the unit up-side-
HANDBOOK
straining
to
the
Power
SP-I0
the ribbon
Supply
Spirometry Unit
Issue 2 July
I998
,
Functional
The procedure detailed here board confidence check after
1.
2.
3.
4.
5.
6.
Check
has
been replaced. It
Ensure that the voltage selector on the back panel
connect
mains
that
<SETTINGS>
emptied more than once to achieve a total volume of more than
Switch the unit on by pressing the and the test screen disappears, check that time are shown in
Press
complete. Ensure Connect the
Connect
Press the <CALIBRATION> key. Enter the ambient temperature and press Press <START> and empty the calibration pump*. 'Effective Vol', enter
+Note:To obtain
Repeat step 5 and ensure that the measured volume
After Assembling the
is
a
general confidence check in the unit after an internal module or
is
not a full functional test but is intended
a
module has been replaced.
to
the unit. Switch on the mains and ensure that the green mains
<ON>
for a few seconds
the
that
flow
sensor
a
calibration pump to the sensor.
a
more accurate calibration,
the
test screen
data
box in the right hand corner
<I> and enter the time and
the correct date and time is displayed on
SP-110
the
to
the connector on the right hand panel of the unit.
calibrated volume of the pump,
SP-10
To
carry
out the check proceed
is
set for the required voltage and
key on
the
Keyboard. Ensure that the LCD lights
is
displayed at the bottom of
the
welcome display is given and that the date
of
the
date
details. Press mCT> when
Press
<ENTER>
in
litres.
it
is
recommended that the calibration pump is
is
correct.
to
the
LCD.
the
screen.
and at the prompt
6
provide a general
as
follows:
LED
lights.
screen. When
and
<ENTER>.
litres.
Page
4.14
7.
Press the <PATIENT DATA> key and systematically press
<QWERTYUIOP>,
Ensure that the entered
to
exit
the
patient
8.
Select a test key (eg cFCV manually blow into the flow sensor. Ensure screen and that after six seconds the
data
data
is
display.
Ej)
COITK~~Y
and press
&NTEb,
presented on the screen
<START>.
that
the pulmonary
unit
prompts for the second reading.
the
keys:
4234567890s
as
entered. Press <FNCT>
With the calibration pump, or
graph
is rraced on the
Page 87
SCHILLER
SERVICE
WDBOOK
SP-10
Spirometry Unit
Issue
2
Julv
1998
Physical Overview
Chapter
d?
Module Replacement
4
Opening the
9.
Press the and
of
good
Note:
If
printer resistance resistance setting experienced with paper mark detection necessary to adjust
10.
If
a
module
Paragraph
supply
for
case
<PRINT
the print-out
(if
24
(cont.)
ALL>
quality.
quality
has
been replaced carry out
not
already
hours
to
fully
or
<PRINT
data
see
paragraph ‘Service
the
carried out).
charge the battery.
BEST>
is
poor and the printer
may
have been
factory pre-sets
Switch
key and ensure that the print-out is accurate
has
not been changed, the stored
lost
from
memory.
Screen’
or
setting the printer
on
the Interface
the
checks and tests detailed in the relative
the unit
off
To
check the stored
in
chapter
and leave connected
board.
3.
If
problems are
speed
it
may
to
be
the
mains
Page
4.15
Page 88
Chapter
4
Physical Overview
&
Module
Replacement
-_
SCHILLER
SERVICE
SP-I0 Spirometry Unit
HANDBOOK
Issue
2
July
I
I998
Power
Supply
P17
board
-
GND BAYONET
CONNECTOR
P1-
TO MICROPROCESSOR
PI2 - TO
LCD/KEYBOARD
(IN
TOP
ASSEMBLY)
I
BUFFER
I
I
P20 - TO
INTERFACE
BOARD
P25,
P26
BATTERY
CONNECTORS
+UB VOLTAGE ADJUSTMENT
Power
Supply
VRl
Board
MK
F'27
MAINS
TRANSFORMER
8-6
-
FROM
Page
4.16
Page 89
SCHILLER
SERVICE
HANDBOOK
SP-I0
Spirometry
Issue 2 July
Unit
I998
-_
Physical Overview & Module Replacement
c
Chapter
4
Power
Supply
The Power Supply board
board
Pre-requisites
The Warnings and Cautions The Top Assembly must
Tools
and Equipment
Cross-Bladed screwdriver Small flat-bladed screwdriver
Digital Voltmeter
In addition up to 30V,
if
the Power Supply boardis checked in isolation, aPower Supply capable of supplying
is
required.
Parts
Power Supply
MK
Power Supply Board Removal
(cont.)
is
positioned on four
at
be
removed and
8-6.
Part number
the beginning
all
as
detailed
spacers
above the Processor board.
of
the Chapter must
external cable assemblies disconnected
in
Chapter
5.
be
observed.
WARNING
ENSURE
To remove the Power Supply board proceed
1. Disconnect the Interface board, Keyboard Buffer
2.
Unscrew the four screws securing the
THAT
the
THE
flying
MK
MAINS
leads
P1
8-4
CABLE
to
the
to
the Microprocessor and the GND guard P17. (P12
will already
IS
DISCONNECTED
as
follows:
mains transformer
be
removed).
board
to
the spacers and remove the board
BEFORE
and
to
the batteries. Disconnect
Power Supply Board Replacement
To
replace the Power Supply proceed
1. Position the board on the spacers and secure with the four securing screws.
2.
Connect the Mains Transformer leads and the battery leads. Connect P1,
(GMD).
as
follows:
Checks and Tests after Power Supply Board Replacement
After assembly, connect the is lit. Switch the unit on and ensure the
<SETTINGS>
Check that the battery charge monitor circuit is functioning
key followed by key
mains
supply to the unit and ensure
that
the LCD is lit and that meaningful
45..
Ensure that the printer functions.
that
the green mains supply
as
follows:
COMMENCING
to
the LCD/
P20
and P17
data
is displayed. Press
P20
LED
to
Page
4.17
Page 90
Chapter
Physical
4
Overview
&
Module
Replacement
SCHILLER
SERVICE
SP-I0
HANDBOOK
Spirometry Unit
Issue
2
July
1998
Power
Supply
1.
Switch the unit
hours. Check that the battery light
when the battery is fully charged
2.
Disconnect the mains and printer). Check that
0 0 0
Note:
board
the battery the battery the unit switches Control of the Battery
(BA'ITLC giving the expected indications, the fault connector). The signal
The nominal battery capacity at which the Battery Low is active
25%.
The nominal voltage at which
switches off, is
Checking the Power Supplies
The power supply voltages can
Power Supply board the unit assembled and the Top Assembly positioned by the side. isolation a latch circuit must the
<ON>
key).
(cont.)
off
but leave the
LED
lights when
LED
flashes
off
)
to calculate the charge state
21V.
is
easily accessible and it
be
mains
flashes
(after
run
the
unit
mnning
when
the
when the
LED
be
manually set to switch on the power supply ( equivalent to pressing
battery
is
from
to
control the battery
checked without the SP-10 being assembled. However the
connected. Leave the battery
when the battery
a
maximum
on battery power for
on battery power
battery
capacity is low
capacity
the CPU which
of
the battery.
may
LED
the
Battery
is
usually easier to check the power supplies with
is
time of
15
60
falls
below a preset value
uses
lie on the Processor board (or P1
is
LEDB.
down
If
to
charge for
charging and extinguishes
hours).
minutes (without using the
the counter signal
If
the Battery
signal is active and
the board is to
(LED
LED
flashing) is
be
10
is
not
the
unit
checked
in
Test points are provided on Individual adjustment is not possible for
supplies are generated and the batteries are charged). Note that mains connected.
ON
Signal Simulation
If
the
Power Supply board is checked
the power supply on. To simulate the
Note:
The ON simulation signal is only required when be used to switch the power supply on, that is when P12
disconnected.
the
Power Supply board for measurement
any
of
the voltage
in
isolation the
ON
signal, proceed
ON
rails
except
+UB
line must
as
follows:
the
ON
of
all
power
rails.
+UB
(from which
can
only be adjusted with
be
connected to
key on the keypad, cannot
to
the Top Assembly is
GND
all
to
other
latch
Page
4.18
Page 91
SCHILLER
SERVICE
WDBOOK
SP-I0
Spirometry Unit
Issue
2
July
I998
Chapter
Physical Overview & Module Replacement
4
Power
1.
2.
Note:
Supply
Connect power
0
0
Using a suitable power supply will now the correct voltage.
board
to
the Power Supply
If the unit charged batteries connected
If
the unit the Power Supply negative
To switch the power supply off, connect
is
is
to
pin 2 of plug
lead,
(cont.)
board
as
follows:
assembled,
not assembled, connect a power supply to the banery connectors
connect
be
use
either
to
the battery connectors
board.
Connect the positive to
P26.
Set the Power Supply to 24V.
P25
PIN
1
GND
to
Plug P12 pin
latched on and
mains
(via the
mains
pin
1
of plug
P26
PIN
2
25,
and hold for 2 seconds. The
all
generated power supplies
OV
to
P1
pin
50.
transformer) or two fully
P25,
and the
can
be
checked for
on
+UB
Adjustment
This adjustment sets the voltage supplied from the switching regulator on the mains power factor controller circuit. To adjust this voltage the unit must
be
assembled and mains connected.
WARNING
LETHALVOLTAGES SUPPLY. DO NOT LET CONCENTRATION TAKING MEASUREMENTS.
To check and adjust
1.
Connect mains power to the unit and switch on by pressing the <ON> key keyboard.
2.
Measure the voltage at
Power
Rail
Measurement
No
adjustment is possible for any of the power
relevant test point (component layout given at the end
any
of
the values are
power supply must
AREPRESENTWHENCHECKINGANDADJUSTINGTHEPOWER
+UB
supply proceed
P25
pin
1.
out
of
tolerance fvst ensure that
be
changed.
as
Adjust
LAPSE.
follows:
VRl
rails.
EXERCISE GREAT
to achieve a voltage of
To
check the power
of
this book) and measure with a
+uB
voltage is correct.
rail
CARE
on
the
+27V
SOOmV
voltages go on the
If
this is correct the
WHEN
DVM.
If
Page
4.1
9
Page 92
Chupter
Physical
4
Overview & Module
Replacement
SCXILLER
SP-I0
HANDBOOK
Spirometry
Issue
2
July
Unit
I998
Lead-acid Batteries
Two batteries are installed
Pre-requisites
The
Warnings
The
Top
Parts
The part numbers
and Cautions at
Assembly must
of
all
replaceable items are given
in
the
SP-10.
be
removed and
They
the
beginning
BATTERY
I
are
selfcontained with integral leads andconnectors.
of the Chapter must be observed.
all
external cable assemblies disconnected.
in
Chapter
2
5.
BATTERY
I
I
1
I
i
Page
Battery
THE
To
1.
2.
4.20
Removai
WARNING
MAINS
remove the battery proceed
Ensure
assemblies connected Disconnect the battery connectors from the Power Supply board and remove the batteries.
SUPPLY
that the unit is switched off and that
MUST
BE
DISCONNECTED
as
follows:
to
the front and rear panels.
DURING THIS
the
mains
PROCEDURE
is disconnected. Remove
all
cable
Page 93
SCHILLER
SERVICE
HANDBOOK
SP-I0 Spiromtv Unit
Issue
2
July
1998
’-_
Chapter
Physical Overview & Module Replacement
4
Lead-acid Batteries
Battery Replacement
To
replace the battery position the battery and connect the leads to the battery connector on the
Power Supply.
Red
Note:
Checks
and
To
ensure that the
replacement, proceed
1.
Connect mains to the unit and ensure indicators, next to the display,
2.
Ensure
So
that the capacity
3. run the unit on battery power
Repeat registered by the processor. The calculated value pressing:
lead is the + battery
Black lead is the - battery
Tests After Battery Replacement
battery
as
that
the Battery indicator is extinguished after
of
this
process two times.
(cont.)
terminal
terminal
and
the
battery recharging circuit
follows:
that
the green
are
lit.
Charge
the newly installed battery
until
the battery
This
ensures
that
<SETTINGS>
are
functioning after battery
mains
and the yellow battery
the
battery
for approximately
15
hours (battery fully charged).
is
calculated correctly by the processor,
is
fully discharged and the unit switches
the comct
of
c’b
capacity
the battery capacity
15
hours.
of the battery
can
be
seen by
off.
is
The
B
CAPACITY
for further details.
4.
Program all static settings which will have
(S)
indication displays the calculated battery capacity.
been
lost when the battery
See
was
disconnected.
Chapter 3
Page
4.21
Page 94
c
-.
..
,
Chapter
4
Physical Overview & Module Replacement
Interface Board
Pre-requisites
The Warnings and Cautions
The Top Assembly must
Tools
Cross-bladed screwdriver Small flat-bladed screwdriver
Parts
Interface
Interface
To remove the Interface Board disconnect board
Board
MK
Board
to
the Paper Tray Assembly, and
Removal
be
8-5.
Part number
at
removed and
the
beginning
lift
SCHILLER SP-I0 Spirometry Unit
SERVICE
of
the
Chapter
all
extemal able assemblies disconnected.
as
detailed
all
in
Chapter
COM~C~O~~.
must
5.
unscrew the four screws securing the
HANDBOOK
be
observed.
the board away from the Paper Tray Assembly.
Issue 2 July
1998
.I
.1
.
i
Interface
Board
Replacement
The replace the Bus Interface position the board on the Paper four captive screws. Reconnect
Checks and
Tests
After Interface Board Replacement
all
connectors to the
Thermal
The paper mark detection offset must be checked. To set the paper
follows:
1.
Connect a digital voltmeter across the P4 connector (test point
2.
Switch the unit on and adjust
>
3.5v
0.4
to0.5V
IF
THE
NB
DETECTION DOES
ADJUSTMENT,
3.
Carry out the printer check to ensure that the
VR1
to obtain a reading
when the black paper when there
is
NOT
ENSURE
THAT
mark
no paper
mark
RECOGNISE
THE
PAPER
board
is
as
functioning correctly . Enter the
following key sequence:
4E"GS>
<P>
Tray
Assembly and secure using the
Printer and the Power Supply.
mark
threshold proceed
1).
follows:
is
under the paper mark detector
under the detector ie white paper
THE
PAPER
MARK
MARK
WINDOW
AFTER
IS
CLEAN.
as
Page
4.22
Page 95
SCHILLER
SERVICE
HANDBOOK
SP-10 Spirometry Unit
Issue
2
July
I998
Chapter
Physical Overview & Module Replacement
4
Interface
A
print-out of a
all
close pixels
Note: Unevenness of print or faulty individual pixels indicates a problem with the printer or
Board
the
lines
diagonal
will
printer alignment Interface board could
the
(cont.)
series
are
even
lines and
be
seen
Processor
of diagonal lines
and
unintermpted.
that
no blocks
as
a
horizontal white line. Examine the
If
a
complete
be
faulty
board.
P32 P29 PRINT
GUARD DATA CABLE SENSOR
will
be
given. Examine the print-out and ensure
Ensure
of
piain paper
block
is
left
or the fault
that rhe complete paper
are
present.
unprinted
lies
with the printer
HEAD
Any
print-out
for evenness of print.
it
additionally indicates that the
data
P24 - TO PHOTO
is
covered with
faulty print-head
control
circuits
that
on
/
P22-TOTP
MOTOR
(TEST
P4 POINT
1)
TO PAPER
CARRIAGE
I
Page
4.23
Page 96
Chapter
Physical Overview & Module Replacement
4
SCHILLER
SERVICE
SP-10 Spirometry
HANOBOOK
Issue
2
July
Unit
1998
Paper
The Paper Tray Assembly motor,
is
he-requisites
Tools
Parts
5.
ENSURE
Tray
Assembly
comprises
the
thennal
mounted
The
Warnings
The Top Assembly must
Cross-Bladed Screwdriver Fiat-Bladed Screwdriver
The
part
THAT
printer
on
four spacers.
and Cautions
number for the Paper Tray Assembly, complete with the
THE
MAINS
and
the
at
be
removed and
CABLE
a
metal framework onto
interface
the beginning
board.
It
of
all
external cable assemblies disconnected.
WARNING
IS
DISCONNECTED
which
is
positioned above
the Chapter
must
BEFORE
is
mounted
the
Processor board and
be
observed.
DC
motor
COMMENCING
the
paper my and
is
given
in
Chapter
Paper
Tray
Removal
To remove
1.
Remove
2.
Carefully retaining the paper my assembly.
3.
Hold
4.
Taking care not away from the unit.
the
Paper Tray Assembly proceed
P20
on the Power Supply.
turn
the unit upside down and rest it
the
two parts together and turn them right side up. Watch for screws falling out
to
damage
as
follows:
on
a
soft cloth. Remove the four screws
the
printer motor gently ease the complete Paper Tray Assembly
!
Page
4.24
Page 97
SCHILLER SP-I0 Spirometry Unit SERVICE
...
1
..
Paper Tray Assembly
HANDBOOK
Issue 2 July
1998
(cont.)
Physical Overview & Module Replacement
Chapter
4
Paper Tray Replacement
To
replace
1.
2.
Hold
3.
Turn
4.
Replace the Top Assembly
the
Paper Tray Assembly
Gently
ease
the assembly
the
two
parts
firmy
the Paper Tray Assembly
right side up and replace
proceed
in
position on the four spacers
together and
to
the
Base
p20
as
detailed earlier.
as
follow:
turn
upside
Assembly.
on the Power Supply.
down.
above
the Processor board.
Secure
the
four
screws retaining
Checks
and
Tests after Paper Tray Assembly Replacement
After
assembly connect <REPLACE PAPER> key and ensure that the paper entering enter the following key sequence:
mains
to the unit and ensure
<SEl"GS>
A
series of
diagonal
line
will
be
printed. Ensure that the print-out
that
the green power
tray
retracts smoothly. Check the printer by
<p>
is
even.
LED
lights. Press the
Page
4.25
Page 98
-_
Chapter Physical Overview & Module Replacement
Microprocessor
4
Board
he-requisites
The Warnings and Cautions at the
beginning
of
the
SERVICE
Chapter must
SCHILLER SP-I0 Spirometry
HANDBOOK
be
observed.
Issue
2
July
Unit
1998
~
*.
,?
The Top Assembly must The Power Supply board and the paper tray assembly must
the Processor board.
be
removed and
all
external cable assemblies disconnected.
be
removed
before
removing
Tools
*
Cross-Bladed screwdriver
Parts
Microprocessor Board
Microprocessor
The Microprocessor board
Microprocessor Board proceed
Board
MK
8-1. Part number
as
detailed
in
Chapter
5.
Removal
is
secured to the Base Assembly with screws. To remove the
as
follows:
CAUTION
THE MICROPROCESSOR BOARD CONTAINS CMOS COMPONENTS. ANTI-STATIC PRECAUTIONS MUST BE OBSERVED AT
1.
Remove the Top Assembly.
2.
Remove the Power Supply board.
ALL
TIMES.
.,
I
3.
Remove the Paper Tray Assembly.
4.
Remove the Program Pack.
5.
Disconnect the following connectors from the Microprocessor board:
P4
to
the
ERGO
connector on the back panel
P2
to
the
RS
connectors
P11 to the Beeper Connector
on
the Back Panel
GND
6.
Unscrew and remove the ten Microprocessor board securing screws and remove the Microprocessor Board.
Note:
The Program Pack is mounted in slides on the Microprocessor directly with it.
board
and connects
Page
4.26
Page 99
._..
.
SCHILLER
SERVICE
SP-10
Spiromet?
HANDBOOK
Issue 2 July
Microprocessor
Unit
1998
Board
VIDEO
CONNEmOR
(cont.)
Physical Overview
DC
P1
CONNECTORS
Chapter 4
&
Module Replacement
Microprocessor
To
replace the
1.
Position the Microprocessor screws.
2.
Reconnect the following connectors
PAPER
TRAY
SECURING
POINT
PROGRAM
Board
P4
p2
P11
GND
Replacement
Microprocessor
to the ERGO connector on the back panel to the
RS
to
the Beeper Connector
PACK
board
proceed
board
connectors on the Back Panel
as
in
the
Base
to
the Microprocessor
P11
TO
BEEPER
SP-110
CONNECTOR
follows:
Assembly and secure with the ten securing
board:
Page 4.27
Page 100
.
-!-
~
-
..
..
~.
.
~ . ~
.. . .~..
...
.
.
~.
...
.
.
L
.I
..
/.I
...,.
..
_I.
.-
..
..,
Chapter
4
Physical Overview & Module Replacement
Microprocessor
3.
Slide in,
4.
Replace the Paper Tray Assembly.
5.
Replace the Power Supply
6.
Replace the Top Assembly.
,
Checks
and
When the unit is reassembled after Microprocessor detailed in the beginning of this Chapter must
and
secure the
Board
(cont.)
Program
Pack
as
detailed below.
board.
Tests After Microprocessor
be
Board
board
carried out
Replacement
replacement, the functional check
to
ensure the integrity of the assembled
SCHILLER SP-I0 Spirometry
SERVICE
HANDBOOK
Issue
2
unit.
The capacity of the battery
and
store the correct battery capacity proceed
1.
Fully charge the battery by leaving the unit connected to the
2.
Run the
unit
on battery power until the battery
is
calculated
and
stored by the processor for reference. To recalculate
as
follows:
mains
for
15
hours.
is
fully discharged
and
the
unit
switches off.
July
Unit
I998
I
3.
Repeat this process two times.
The value
of
the battery capacity calculated (and stored) by the processor,
<SE?TINGS>
<’I5
The B CAPACITY indication displays the calculated battery capacity. details.
all
Program
static settings.
can
be seen by pressing:
See
Chapter 3 for further
Page
4.28
Loading...