Welch Allyn ProXenon
Headlight and Fiberoptic Cable
Directions for Use
iiCopyright InformationWelch Allyn ProXenon Headlight and Fiberoptic Cable
Copyright 2008 Welch Allyn®
®
Welch Allyn
under various patents and patents pending.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or
improper use of the product, that may result from failure to use this product in
accordance with the instructions, cautions, warnings, or statement of intended
use published in this manual.
Welch Allyn Technical Support:
is a registered trademark of Welch Allyn. Welch Allyn is protected
USA Tel:+ 1 315 685 2993
Fax: + 1 315 685 2999
Canada 800 561 8797China + 86 216 327 9631
European Call Center + 353 46 906 7790France + 331 6009 3366
Germany + 49 747 792 7186Japan + 81 33 219 0071
Latin America + 1 305 669 9003Netherlands + 31 15 750 5000
ivContentsWelch Allyn ProXenon Headlight and Fiber
1
Introduction
Thank you for purchasing the Welch Allyn ProXenon Headlight
and Fiberoptic Cable. Follow the guidelines provided in this
manual, and you will enjoy years of dependable, trouble-free
service from your new headlight.
Intended Use
The Accessory Headlight is a passive luminaire that is illuminated
by fiberoptic light and utilized to provide supplemental light for
surgical and medical procedures.
Contraindication
This product is contraindicated for use in neonate
transillumination, ophthalmic procedures, photosensitive
patients, or patients receiving photosensitizing agents
(hematoporphyrin derivatives) within 3 months.
1
User Profile
The intended users of the ProXenon Headlight and fiber are
medical professionals who are qualified to perform surgical
procedures.
2IntroductionWelch Allyn ProXenon Headlight and Fiberoptic Cable
Symbol Descriptions
These important symbols may appear on your Welch Allyn
ProXenon Headlight and Fiberoptic Cable. Please note their
meaning.
WarningCaution
Latex FreeAttention; Read the Directions For
Caution, high
temperatures
The CE mark indicates this product has been tested to, and conforms
with, the provisions noted in both the 89/336/EEC Electromagnetic
Compatibility Directive and the 93/42/EEC Medical Device Directive.
European Contact for Regulatory Compliance:
European Regulatory Manager
Welch Allyn Ltd.
Navan Business Park
Dublin Road
Navan, County Meath
Republic of Ireland
Tel: +353 46 90 67700
Fax: +353 46 90 67755
Use Manual for all warnings,
precautions, and instructions for use.
Caution, high intensity light
Directions for Use 3
Warnings
A warning statement in this manual identifies a condition or
practice, which if not corrected or discontinued immediately,
could lead to patient injury, illness, or death.
WARNING Users of this product should read and
understand Directions For Use for the ProXenon Headlight,
Fiberoptic Cable and all equipment used with it.
WARNING Do not shine light into eyes. Eye injury may
result.
WARNING Do not use at distances closer than 10 inches
(25 cm). Tissue damage or erythema may result.
WARNING Do not remove fiberoptic cable from the
luminaire while energized. Intense light energy exiting the
tip of energized fiberoptic cables may cause burns or fires.
Cautions
A caution statement in this manual identifies a condition or
practice, which if not corrected or discontinued immediately,
could lead to equipment failure, equipment damage, or data loss.
Caution Rx Only: United States Federal Law restricts
this device to sale by or on the order of a health care
practitioner.
Caution When removed from light source, end of
fiberoptic cable may be hot. Use caution when handling.
Caution Service must be performed only by Welch Allyn
authorized repair personnel.
Caution Keep dirt, dust, and other contaminants out of
the luminaire. Damage will occur. Reinsert dust cap after
removing the fiberoptic cable.
4IntroductionWelch Allyn ProXenon Headlight and Fiberoptic Cable
Avertissements
Les avertissements de ce manuel identifient les conditions
ou pratiques qui, si elles ne sont pas corrigées ou arrêtées
immédiatement, risquent de provoquer des blessures, des
maladies ou le décès du patient.
AVERTISSEMENT Les utilisateurs de ce produit doivent
lire et comprendre le mode d’emploi de la lampe frontale,
du câble à fibres optiques ProXenon et de tout le matériel
utilisé conjointement.
AVERTISSEMENT Ne pas orienter la lumière vers les
yeux. Des lésions oculaires pourraient s’ensuivre.
AVERTISSEMENT Ne pas utiliser à des distances
inférieures à 25 cm (10 pouces). Des lésions tissulaires
ou des érythèmes pourraient s’ensuivre.
AVERTISSEMENT Ne pas retirer le câble à fibres
optiques du dispositif d’éclairage lorsqu’il est sous tension.
L’intense énergie lumineuse qui sort de l’extrémité des
câbles à fibres optiques sous tension pourrait générer des
brûlures ou des incendies.
Directions for Use 5
Précautions
Les précautions de ce manuel identifient les conditions
ou pratiques qui, si elles ne sont pas corrigées ou arrêtées
immédiatement, risquent de provoquer des pertes de données,
un endommagement ou une défaillance du matériel.
Attention Sur prescription uniquement : en vertu de la loi
fédérale des États-Unis, ce produit ne peut être vendu que
par un médecin ou sur prescription médicale.
Attention Lorsqu’elle est extraite de la source
lumineuse, l’extrémité du câble à fibres optiques peut
être chaude. Manipuler avec précaution.
Attention Seuls des techniciens Welch Allyn agréés sont
habilités à procéder à l’entretien.
Attention Protéger le dispositif d’éclairage contre les
saletés, la poussière et autres contaminants. Dans le
cas contraire, il s’en trouverait endommagé. Réinsérer le
capuchon antipoussière après avoir retiré le câble à fibres
optiques.