Welch Allyn Propaq CS User manual

Propaq
®
CS
Directions for Use
Models 242, 244, 246
Software Version 3.5X
Copyright © 2004 by Welch Allyn. Welch Allyn
are registered trademarks of Welch Allyn. Welch Allyn is protected under various patents and patents pending. Masimo SET
are registered trademarks and Signal Extraction Technology
®
DURASENSOR LIFEPAK
®
, C-LOCK
®
6s
are registered trademarks of Medtronic Physio-Control Corporation.
®
and OXISENSOR
®
is a registered trademark of Welch Allyn. Propaq
is a trademark of Masimo Corporation. NELLCOR
®
are registered trademarks of Nellcor Puritan Bennett, Inc. LIFEPAK
®
, Propaq
®
CS, and Acuity
and
®
,
®
®
5
and
Copyright Notice: Software in this Product is Copyright© 2004 by Welch Allyn or its vendors. All rights are reserved. The
software is protected by United States of America copyright laws and international treaty provisions applicable all over the world. Under such laws, you are licensed to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded, but the software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn or its vendors. Welch Allyn will make available specifications necessary for interoperability of this software on request; however, users should be aware that use of Welch Allyn hardware and software with devices or software not sold by Welch Allyn or its authorized dealers and affiliates may lead to erroneous results and consequent danger in patient care, and may also void Welch Allyn's warranty.
This device complies with Part 15 of the FCC rules and the Canadian ICES-003. Operation is subject to the following conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including any interference that may cause undesired operation.
Disclaimers: Welch Allyn cautions the reader of this manual:
• This manual may be wholly or partially subject to change without notice.
• All rights are reserved. No one is permitted to reproduce or duplicate, in any form, the whole or part of this manual without permission from Welch Allyn.
•Welch Allyn will not be responsible for any injury to the user or other person(s) that may result from accidents during operation of the Propaq monitor.
•Welch Allyn assumes no responsibility for usage not in accordance with this manual that results in illegal or improper use of the Propaq monitor.
• No implied license: Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
Welch Allyn Technical Support:
USA 1-800-
Latin America (+1) 305-669-9591
European Call Center (+353) 469-067-790
United Kingdom 0-207-365-6780
535-6663
France (+33) 1-60-09-33-66
Australia (+61) 2-9638-3000
Germany (+49) 7477-927-173 Singapore (+65) 6291-0882
Canada 1-800-561-8797
South Africa (+27) 11-777-7509
Japan (+81) 3-5212-7391
China (+86) 21-6327-9631
For information concerning this document or any Welch Allyn product, contact:
Welch Allyn
Customer Service 8500 SW Creekside Place Beaverton, Oregon 97008-7107 USA
Within USA, toll free:
Phone: (800) 289-2500 Phone Technical Services: (800)
535-6663
WorldWide:
Phone: (503) 530-7500 Fax: (503) 526-4200 Fax Technical Services: (503) 526-4970
Internet: http://www.welchallyn.com E-mail Technical Services: solutions@monitoring.welchallyn.com E-mail Marketing Dept.: marketing@monitoring.welchallyn.com
Welch Allyn European Customer Service
IPA Business Park, Dublin Road Navan, County Meath, Ireland Phone: 353-46-67700 Fax: 353-46-27128
Welch Allyn U.K. Ltd.
Cublington Road, Aston Abbotts Buckinhamshire HP22 4ND, England Phone: 44-1296-682140 Fax: 44-1296-682104
Welch Allyn Italia
Via Napo Torriani, 29, 20124 Milan, Italy Phone: 39-02-6699-291 Fax: 39-02-6671-3599
Welch Allyn GmbH: Germany
Postfach 31, Zollerstrasse 2-4 72417 Jungingen, Germany Phone: 49-7477-92-710 Fax: 49-7477-92-7190
Welch Allyn: France
814 Rue Charles de Gaulle 77100 Mareuil les Meaux, France Phone: 01-6009-3366 Fax: 01-6009-6797
Welch Allyn: Pacific
P. O. Box 39-293 Howick, Auckland, New Zealand Phone: 64-9-532-9524 Fax: 64-9-532-9526
Welch Allyn: Asia
Room1002, 10/F Tung Sun Comm. Centre 194-200 Lockhart Road, Wanchai, H.K. Phone: (852) 9016-7812 Fax: (852) 2535-5650
Welch Allyn: Latin America
MD International, 11300 NW 41st Street Miami, FL 33172 USA Phone: (305) 669-9003 Fax: (305) 669-8951
Reorder Kit No: 810-1441-XX
Manual Part No: 810-1097-03 Rev. A 8/04 Printed in USA
®
0123

Contents

1 – General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Intended Use, Symbols, and Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Controls and Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Learn Propaq CS Operation with In-Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2 – Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Prepare the Propaq CS Monitor for a New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Set Patient Alarms and Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Change the Current Patient Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Change Powerup Patient Mode or Store Customized Settings . . . . . . . . . . . . . . . . . . 21
3
3 – Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Perform ECG/RESP Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Use the Propaq CS Monitor With Pacemaker Patients . . . . . . . . . . . . . . . . . . . . . . . . 28
Perform Invasive Blood Pressure (IBP) Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Take a Non-Invasive Blood Pressure (NIBP) Reading . . . . . . . . . . . . . . . . . . . . . . . . . 32
Perform Temperature Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Perform SpO Perform Mainstream CO Perform Sidestream CO Set Up the CO
4 – Alarms & Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Respond to Patient Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Customize Alarm Limits Based on Patient’s Current Vital Signs . . . . . . . . . . . . . . . . . 50
Alarm Holdoffs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Connect Nurse Call Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Respond to An Equipment Alert. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Troubleshooting Equipment Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5 – Printing & Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Print Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Display or Print Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Set Printer Options and Automatic Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Print OxyCRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
2
Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
2
Display and Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2
Propaq CS Directions for Use
4
6 – Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Understanding the Propaq CS, Acuity, and the FlexNet™ Network . . . . . . . . . . . . . . 67
Hardwired (Serial) Connection to Acuity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Wireless Connection to Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 – Defibrillator Synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Synchronous Cardioversion with LIFEPAK 5 Defibrillator . . . . . . . . . . . . . . . . . . . . . . 79
Synchronous Cardioversion with LIFEPAK 6s Defibrillator . . . . . . . . . . . . . . . . . . . . . 82
8 – Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Connect the AC Power Adapter to Recharge the Battery . . . . . . . . . . . . . . . . . . . . . . 85
Replace Monitor Input Power Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Install Printer Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Inspect and Clean the Monitor and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Service Interval Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Change the Wireless Propaq CS Network Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
9 – Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Set the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Change the Date Format, ECG Filter, and Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
10 – Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Welch Allyn
1 – General Information

Intended Use, Symbols, and Safety Information

Intended Use
The Propaq CS monitor is intended to be used by skilled clinicians for multiparameter vital signs monitoring of neonatal, pediatric, and adult patients in health care facility bedside applications. It is also intended for intra-facility and ambulance transport.
The ECG channel is intended for five-lead or three-lead ECG monitoring.
The Respiration (RESP) channel is intended to detect the rate or absence of respiratory effort, deriving the signal by measuring the ac impedance between selected terminals of ECG electrodes.
The Invasive Pressure (IBP) channel is intended for measuring arterial, venous, and intracranial pressures (and umbilical artery and vein pressures for neonates) using invasive transducers.
5
The Noninvasive Blood Pressure (NIBP) channel is intended for indirectly measuring arterial pressures using an inflatable cuff. If ECG is also monitored, the Propaq CS Smartcuf™ software algorithm automatically synchronizes the NIBP measurement process to the occurrences of the R-wave, increasing accuracy in cases of extreme artifact and diminished pulses. The operator may disable or enable the Smartcuf algorithm in the NIBP Menu.
The Temperature (TEMP) channel is intended to measure temperature using an attachable probe.
The Pulse Oximetry (SpO hemoglobin at a peripheral measurement site.
The Capnography (CO End-tidal CO
This guide was written for clinicians. Although this guide may describe some monitoring techniques, Welch Allyn expects that you are a trained clinician who knows how to take and interpret a patient’s vital signs. This monitor has been designed as a quality monitor; however, inherent limitations require that good clinical judgment always prevails.
(ETCO
2
) channel is intended to noninvasively measure oxygen saturation of arteriolar
2
) channel is intended to noninvasively measure the following vital signs or events:
2
), Inspired CO
2
(INCO
2
), Breath Rate, and Apnea.
2
Symbols
Warning
WARNING statements in this manual identify conditions or practices that could result in personal injury.
Caution
CAUTION statements in this manual identify conditions or practices that could result in damage to the equipment or other property.
Note
NOTE statements provide additional important information.
Propaq CS Directions for Use
6
The following symbols may appear on the Propaq CS monitor or accessories. They are defined by the International Electrotechnical Commission, IEC 878 and IEC 417A.
Off (Standby)
Tw o way communication port
On Input port
For continued fire protection, use only the
Output port
specified fuse
Direct current Alternating current
Direct current or alternating current
Separate batteries from other disposables for
1
recycling.
Caution: Refer to Directions For Use and accompanying documentation
Battery charging when green indicator
IPX1
Enclosure Protection Drip proof: Classification IPX1 per IEC Publication 529
Temperature sensor input
illuminated
Patient connections are Type CF, isolated for direct cardiac application, and protected against defibrillation
Patient connections are Type BF, and
Tr ansformer meets requirements of a short-circuit-proof safety-isolating power transformer
1
For indoor use only (on power adapter only)
protected against defibrillation
Protected during defibrillation Stacking limit by number
n
This way up Temperature limits
Fragile
n%
Keep away from rain Altitude limit
Humidity limit
n
Patient connections are Type B Signifies the device has met all essential
requirements of European Medical Device Directive 93/42/EEC for a Class 1 product
The CE Mark and Notified Body Registration Number signify the device has met all
0123
essential requirements of European Medical Device Directive 93/42/EEC
Urgent alarm notification (output to Nurse Call system)
NRTL/C
Evaluated to CSA 601-1 and UL2601-1
C
The Canadian Standards Association has evaluated this device according to CSA 601-1 and Underwriters Laboratory Standard UL 2601-1
1
.
This device has been tested and certified by the Canadian Standards Association International to comply with applicable U.S.
US
and Canadian medical safety standards.
Apply the NIBP cuff as shown.
NIBP cuff sizes: Thigh Large adult Adult
Small adult Child Infant
1. This symbol is on the Universal Power Adapter.
2
Single-use only (not reusable).
1
Welch Allyn
Chapter 1 – General Information 7

General Warnings and Cautions

Familiarize yourself with all warnings and cautions before using the Propaq CS monitor. In addition to the following, other warnings and cautions appear throughout this manual.
Warning
systems safety standards such as IEC 601-1-1. Within certain governmental jurisdictions, all interconnected accessory equipment must be labeled by an approved testing laboratory. After interconnection with accessory equipment, risk (leakage) current and grounding requirements must be maintained.
manual and all accessory Directions For Use before operating the monitor.
it on again. This clears the prior patient’s trend values, alarm limit settings, and NIBP cuff inflation target.
limits, maximum cuff inflation pressure, and internal algorithm settings.
and humidity ranges.
their shelf or mount. Lift the monitor only by its handle; do not lift it by any attached cables.
not operate the apparatus from ac power with a damaged power adapter cord or plug.
electrode wires must be inspected, properly maintained, and in proper working order to allow the equipment to function properly and protect patient safety.
entanglement or strangulation.
circuit so that the return paths cannot be made through monitoring electrodes and probes.
conductive parts in contact with the patient. Avoid contact with any accessories connected to the monitor’s left side panel.
other patient-applied parts should not contact other conductive parts, including earth ground, at any time.
combination with air, oxygen-enriched environments, or nitrous oxide; explosion can result.
interference with ECG monitor operation. This interference may distort the displayed or recorded ECG signal, thereby preventing accurate rhythm analysis. Avoid operating this device near equipment of this type.
hospital-grade receptacle. Do not under any circumstances attempt to remove the grounding conductor from the power plug of the power adapter. Do not plug the power adapter into an extension cord. If there is any doubt about the integrity of the protective earth ground of the receptacle for the power adapter, do not plug in the power adapter; operate the monitor only on battery power. Contact your biomedical engineering department for assistance in identifying the proper power receptacle and making appropriate power connections.
Safe interconnection between the Propaq CS monitor and other devices must comply with applicable medical
This monitor is to be operated by qualified personnel only. The operator of this monitor should read this entire
Before you use a Propaq CS monitor on a new patient, always turn off the monitor for a few seconds, then turn
Always check the patient mode when monitoring a new patient. The patient mode determines default alarm
The monitor may not meet its performance specifications if stored or used outside the specified temperature
Place the Propaq monitor and accessories in locations where they cannot harm the patient if they fall from
Do not connect more than one patient to a monitor. Do not connect more than one monitor to a patient.
Inspect the power adapter cord periodically for fraying or other damage, and replace the adapter as needed. Do
Make frequent electrical and visual checks on cables, sensors, and electrode wires. All cables, sensors, and
As with all medical equipment, carefully route the patient cabling to reduce the possibility of patient
Avoid electrosurgery burns at monitoring sites by ensuring proper connection of the electrosurgery return
During defibrillation, keep the discharge paddles away from ECG and other electrodes, as well as other
To ensure patient safety, the conductive parts of the ECG electrodes (including associated connectors) and
Do not operate this product in the presence of flammable anesthetics or other flammable substance in
Electronic equipment that emits strong electromagnetic or radio frequency signals can cause electrical
When using a power adapter with this monitor, be sure to connect the power adapter to a three-wire, grounded,
Propaq CS Directions for Use
8
Warning
To help protect against electrical shock due to leakage current, use only monitor ac power adapters
recommended in the Welch Allyn Products and Accessories booklet (P/N 810-0409-XX).
For best product performance and measurement accuracy, use only accessories supplied by Welch Allyn or recommended in the Welch Allyn Products and Accessories booklet. Use accessories according to your facility’s standards and the manufacturer’s recommendations. Always refer to the manufacturer’s Directions for Use.
If a product has been dropped or severely abused, send it to a qualified service person to confirm proper operation and acceptable risk (leakage) current values.
Some or all NIBP safety functions are disabled in the NIBP TEST screen in the Service Menu. Do not attempt to conduct NIBP TEST when the cuff is attached to a patient.
When using the motion tolerant pulse oximetry channel, a very sudden and substantial change in pulse rate can result in erroneous pulse rate readings. Be sure to validate the patient data and patient condition before intervention or change in patient care.
Do not use the Propaq CS monitor in a Magnetic Resonance Imaging (MRI) suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in patient injury and monitor damage.
Impedance pneumography and CO high-frequency jet ventilation or high-frequency oscillatory ventilation.
This monitor should only be repaired by qualified service personnel. The operator should not attempt to open the monitor case or perform any maintenance on the monitor except for procedures explicitly described in this manual that can be performed by operators such as inspection and cleaning.
monitoring may not operate properly when used in conjunction with
2
Caution
Do not autoclave the Propaq CS monitor. Autoclave accessories only if the manufacturer's instructions clearly approve it. Many accessories can be severely damaged by autoclaving.
Federal USA law restricts this device to sale, distribution, or use by or on the order of a licensed medical practitioner.
It is possible for the monitor to detect a problem that prevents the monitor from operating properly. If this occurs, the monitor displays an error message and error number. Report such errors to Welch Allyn.
The Propaq CS monitor should be serviced only by a Welch Allyn service technician while under warranty. The Propaq CS Service Manual (P/N 810-1101-XX) is available from Welch Allyn to assist the biomedical engineer during post-warranty period service.
Welch Allyn

Controls and Connectors

Nurse Call
Option
Chapter 1 – General Information 9
Yellow antenna cap indicates Wireless Propaq CS
Alarm Light
Alarm(s) Off Light
Suspend/ Resume Alarm
Suspends alarm tone for 90 seconds or resumes alarm tone (if active)
SpO
Option
Mainstream CO2 Option
Printer Option
Sidestream
CO2 Option
2
Starts or stops NIBP reading
Returns the display to the Main Menu
Printer Buttons
Start/ Stop NIBP
Home
Touch-Screen Controls
The front panel touch-screen provides five softkeys along the bottom and three icon-labeled keys along the right side. An Acuity connected to an Acuity system. These keys allow control of all monitoring and setup functions.
NET OFF
key is displayed in the upper left corner if the monitor is
Note
touch-screen to misinterpret the command and respond to the wrong key.
Avoid pressing more than one touch-screen key at a time. Touching more than one key area at a time can cause the
Caution
touch-screen. Use your finger to press the touch-screen keys.
Propaq CS Directions for Use
Do not touch the screen with a sharp object such as a pen or pencil. Sharp objects can damage the
10
System Control and Connectors (Right Side Panel)
On/Off Switch
Input Fuse
3A
2AG
Power Input Connector
12-28V, 3A
Patient Connectors (Left Side Panel)
Model 242
ECG NIBP Temperature (two channels)*
ECG / EKG RESP
!
NIBP PSNI
T1
T2
Model 244
ECG Invasiv Pressure (one channel) NIBP Temperature (two channels)*
NIBP PSNI
DEFIB SYNCHROMONITOR
!
!
Defib Sync Connector
Real-time ECG Output Connector
EKG x 1000
!
!
Speaker
Connector for Acuity or Modem-Propaq
Battery Charging Light
Model 246
ECG Invasive Pressure (two channels) NIBP Temperature (two channels)*
ECG / EKG RESPINV. BP
P1
NIBP
T1
!
T2
PSNI
ECG / EKG RESPINV. BP
P1
T1
!
T2
Propaq CS Model 242
ECG / EKG RESP
P
!
S
COMPATIBLE
N
I
HP
NIBP
HP Model 242*
*The HP (Hewlett-Packard) side panels provide only one temperature connector.
Propaq CS Model 244
INV. BP
COMPATIBLE
ECG / EKG RESP
P1
P
!
S N
I
HP
HP Model 244*
NIBP
INV. BP
P2
Propaq CS Model 246
INV. BP
COMPATIBLE
HP
INV. BP
ECG / EKG RESP
P1
!
P S N
I
P2
HP Model 246*
NIBP
Welch Allyn
Option Connectors
!
CO
Chapter 1 – General Information 11
CO
2
2
!
Mainstream CO2 Connector Sidestream CO2 Connector
Cable: see Welch Allyn
Products and
Accessories Guide
(810-0409-XX)
Masimo SpO2 Connector
(motion tolerant)
Nurse Call Connector
Masimo SpO2 Connector
(motion tolerant)
SpO
2
!
NELLCOR SENSORS/ CAPTEURS
Cable: see Welch Allyn
Products and
Accessories Guide
(810-0409-XX)
Propaq CS Directions for Use
Nellcor SpO2 Connector
(newer style, motion tolerant)
Nellcor SpO2 Connector
(older style, without motion
tolerance)
12

Display

You can select up to four waveforms to be shown on the Propaq CS monitor. When only one waveform is selected, a trend window automatically appears beneath the waveform.
NET OFF
to disconnect
monitor from
Acuity
network.
Patient name
entered at
Acuity Central
Station.
When
selected,
trends are
displayed
here.
STATSCALE
automatically adjusts
all waveform scales for
optimum viewing.
Patient mode Status messages can appear here.Time of day
ADULT
NET OFF
ON NETWORK
II
mV
1
cm
MON
V
mV
1
cm
MON
130
P
100
1
70
30
P
15
2
0
STATSCALE
ALARMS
17:05:10
ALLEN KIM
PRINT FAULT
ACUITY
TRENDS
HR
BPM
112
1
P
(M)
mmHg
( 96)
P2
(M)
( 15)
mmHg
NIBP
(M)
( 85)
mmHg
TEMP
°F
MCO2
38
mmHg
SpO2
%
MORE
S
121
D
79
S D
S
122
D
58
100.4
T1
98.6
T2 T
BR
Br/m
97
25
9
1.8
12
While changing monitor settings, a status window may appear below the waveform:
Heart Rate Source: HR indicates ECG; PR indicates blood pressure or SpO
Heart Rate in beats per minute.
Bells indicate alarm limit status
All numeric values are continuously displayed and updated.
If the monitor detects a vital sign outside the measurable range, it displays
- - - (below the range) or + + + (above the range).
2
.
ECG1
waveform
is always
displayed
if active.
Status
window
ADULT
NET OFF
ON NETWORK
17:05:10
ALLEN KIM
II
mV
1
cm
MON
ECG/RESP
HR/PR TONE
PACER DISPLAY
ECG BANDWIDTH
RESP LEAD
: OFF
: ON
: MONITOR (0.5-40Hz)
: Ld2 (RA-LL)
RESP MONITORING : ON
NEXT
CHANGE
PRINT FAULT
HR
BPM
112
1
P
( 96)
P2
( 15)
NIBP
( 85)
TEMP 100.4
°F
MCO2
mmHg
(M)
(M)
(M)
38
mmHg
mmHg
mmHg
S
121
D
79
S D
S
122
D
58
T1 T2 T
BR
Br/m
25
9
98.6
1.8
12
SpO2
%
97
PREVIOUS
MENU
Systolic, diastolic, and mean pressures are uniquely labeled and displayed.
ETCO2 concentration
Breath Rate from CO
2
Oxygen saturation is a percentage value.
Welch Allyn
Chapter 1 – General Information 13

Menus

Menus for some patient vital signs are displayed only if the option is included in your Propaq CS monitor.
Main Menu
MAIN MENU
NIBP
ECG/
RESP
INVASIVE
PRESSURE
SpO2/
CO2
SETUP
SpO2 CO2
CO2 MENU 1
RANGE
mm/s MORE
CO2 MENU 2A (NO CO2 SENSORS)
GAS
COMP
RESPONSE
CO2 MENU 2B (MAINSTREAM CO2 ACTIVE)
GAS
COMP
RESPONSE SOURCE
CO2 MENU 2C (SIDESTREAM CO2 ACTIVE)
GAS
COMP
RESPONSE SOURCE
SpO2 MENU 1
SIZE MORE
SpO2 MENU 2
RESPONSE C-LOCK PREVIOUS
INVASIVE PRESSURE MENU 1
RANGE RESCALE ZERO P1 MORE
ZERO P2
PREVIOUS
MENU
FLOW
RATE
PREVIOUS
MENU
MENU
PREVIOUS
MENU
PREVIOUS
MENU
PREVIOUS
MENU
PREVIOUS
MENU
ECG/RESP MENU 1
ECG1
SIZE
ECG/RESP MENU 2
NEXT PREVIOUS
NIBP MENU
START AU TO/
STOP
MANUAL
Propaq CS Directions for Use
CANCEL
INVASIVE PRESSURE MENU 2
LABELP1FORMAT
ECG1 LEAD
CHANGE
INTERVAL
P1
ECG2
SIZE
TURBOCUF
LABEL
P2
RESP
SIZE
SMARTCUF
CANCEL
FORMAT
P2
MORE
MENU
PREVIOUS
MENU
14
Setup Menus
NIBP ECG/
RESP
SETUP MENU 1
STATSCALE
TREND MENU
PRINT NEXT
ACUITY MENU (HARDWIRED CONNECTION)
ALARMS MENU
STAT SET SUSPEND ALL
RESUME
INVASIVE
PRESSURE
WAVE
WAVE SEL
SELECT
ACUITY
NET OFF
ALARMS
SpO2/
CO2
TREND
SNAPSHOT PREVIOUS
LIMITS PREVIOUS
SETUP
MOREALARMS TRENDS
PRINT
OXYCRG
1
MENU
MENU
SETUP MENU 2
NEXT CHANGE WAVE
SELECT
PRINTER
MORE
ALARM LIMITS MENU
NEXT
PARAMETER
1
Acuity Menu is displayed if the monitor is connected to Acuity. See page 67 for details about Acuity hardwired and wireless communication options.
2
ON/OFF key is not displayed for HR/PR alarm limits if the HR/PR ALARM LIMITS setting is set to CANNOT TURN OFF.
3
Service menu tests are only for use by authorized service personnel and are only available in the Adult patient mode. RADIO key is only displayed for Wireless Propaq CS.
4
Patient Mode menu is accessed when CHANGE is pressed for PATIENT MODE.
5
Radio menu is displayed for Wireless Propaq CS.
SETTING
2
UP DOWNNEXT
ON/
OFF
TIME/DAY MENU
NEXT UP DOWN ENTER
SERVICE MENU 1
TEST
SERVICE MENU 2
TEMP TEST
SERVICE MENU 3
3
IBP
TEST
3
PIXEL
TEST
3
CAL
SpO2 TEST
NET
TEST
SYSTEM RADIO
CO2
TEST
PCMCIA
TEST
SERVICE
MORENIBP
MORE
MORESETTINGS SCREEN
SETTINGS MENU
NEXT CHANGE
PREVIOUS
MENU
PRINTER SETUP MENU
NEXT
CHANGE
CHANGE
PRINT
TREND
PREVIOUS
MENU
WAVE SELECT MENU
NEXT INSERVICE PREVIOUS
PATIENT MODE MENU
NEONATAL PEDIATRIC SETUP
ON/
OFF
4
ADULT
NO
INSERVICE
PREVIOUS
MENU
MENU
MODE SETUP MENU
NEXT POWERUP* USE
NOW
SAVE
PREVIOUS
MENU
RADIO MENU5
NETWORK
STATUS
NETWORK MONITOR
CHANGE
NET NAME
PREVIOUS
MENU
Welch Allyn
Chapter 1 – General Information 15

Learn Propaq CS Operation with In-Service Mode

You can practice using the Propaq CS monitor with the in-service mode of operation.
The in-service mode cannot be activated while you are monitoring a patient. During in-service mode, the monitor display and all printouts include the message SIMULATING or SIMULATED DATA.
To practice with your Propaq CS monitor:
1. Disconnect all patient cables connected to the monitor. You can leave the NIBP cuff connected to the monitor so you can take NIBP measurements.
2. If you have been monitoring a patient, turn off the monitor and turn it back on.
If your monitor is programmed so that the NIBP Automatic Mode is selected at powerup, select one of the Factory Patient Modes as the powerup patient mode (see page 21). Then turn the monitor off and turn it back on. (The in-service mode is not available if the NIBP Automatic Mode has been selected.)
3. From the Main Menu, press SETUP, WAVE SELECT, INSERVICE.
The Propaq CS monitor has two sets of simulated patient information. To change between the sets, from the Main Menu press SETUP, WAVE SELECT, and INSERVICE again.
While in the in-service mode, you can press any monitor keys (except the AUTO/MANUAL key in the NIBP Menu) to change a function setting. For example, you can change ECG and RESP waveform sizes, set alarm limits, or set custom settings.
You can also apply the NIBP cuff to yourself and take NIBP measurements.
4. To exit the in-service mode, turn off the monitor.
If you changed the powerup patient mode in step 2, be sure to restore the appropriate powerup patient mode according to your local protocol.
Note
or the NIBP Automatic Mode has been selected. If the monitor is in in-service mode and you connect a sensor (except for an NIBP cuff) or press the NIBP AUTO/MANUAL key, the monitor will turn off power to exit the in-service mode, and then turn on in the normal operating mode.
The in-service mode is not available if the monitor detects that a sensor has been connected (except for an NIBP cuff)
The pacemaker signal indicators are not displayed in the in-service mode.
Propaq CS Directions for Use
16
Welch Allyn
2 – Setup

Prepare the Propaq CS Monitor for a New Patient

Warning
17
again. This clears the prior patient’s trend values, alarm limit settings, and NIBP cuff inflation target.
Before you use a Propaq CS monitor on a new patient, always turn it off for a few seconds, then turn it on
1. Press the gray recessed MONITOR button on the right side of the monitor to turn the monitor off (if it is on). Press it again to turn the monitor on. The monitor displays the powerup screen for about 10 seconds, then displays the Main Menu. The monitor is in the powerup patient mode with the associated settings.
2. Confirm that the monitor emits a tone. If the monitor has SpO2, listen for two tones and confirm both speakers are working.
Note
SETUP, MORE, MORE). If the battery voltage is 7.4V or less or a low battery message is displayed, connect the
monitor to an ac power adapter to recharge the battery (see page 22). Connecting the adapter does not interrupt patient monitoring.
Check the battery voltage level on the powerup screen (or check it on the Time/Day window: Home,
3. Confirm the monitor is in the correct patient mode according to the patient’s age. If the patient mode is not correct, from the Main Menu press SETUP, MORE, CHANGE to access the Patient Mode window:
PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:
NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT:
> 9 YEARS
BATTERY: 8.2 VOLTS
Home
NEONATAL
PEDIATRIC
ADULT
SETUP
PREVIOUS
MENU
4. Based on the patient’s age, press NEONATAL, PEDIATRIC, or ADULT. When the confirmation window appears, press YES to confirm your selection.
Whenever you change the patient mode, the alarm limit settings, maximum NIBP cuff inflation pressures, and internal computations are automatically changed to the defaults for that patient mode. See page 21 for information about preset Factory patient modes or programmable Custom patient modes.
Note
originally-programmed CO2 alarm limits for the new mode. Check the CO2 alarm limits.
Propaq CS Directions for Use
If you change the patient mode, the CO2 alarm limits in the new mode might vary slightly from the
18
5. To select which vital sign waveforms will display, from the Main Menu press
WAVE SELECT.
Use NEXT and ON/OFF to turn on the desired waveforms in the Wave Select
SETUP, MORE,
window..
SETUP
ECG
1 : ON
ECG
2 : ON
PA
: ON
: ON
2
P
NEXT
ON/ OFF
You can turn on all waveforms, but only the first four waveforms selected as ON in the Wave Select window are displayed. You cannot turn off the ECG1 waveform.
WAVE SELECT
2 : ON
CO
RESP : ON
2 : ON
SpO
NIBP : ON
INSERVICE
PREVIOUS
MENU
6. To set the HR/PR source, display sweep speed, tone volumes, and display brightness, from the Main Menu press SETUP, MORE to access Setup Menu 2. Use NEXT and CHANGE to select settings.
SETUP
HR/PR RR/BR
CURRENT SOURCE : ??? SELECTED SOURCE : ECG SWEEP (mm/s)
ALARM TONE HR/PR TONE
PATIENT MODE
BRIGHTNESS
:
: :
:
:
25.0
LOW
OFF
ADULT
NORMAL
???
---
6.25
The ? ? ? is displayed until a source is active.
NEXT
CURRENT SOURCE When the selected HR/PR source is no longer available, the current
CHANGE
WAVE
SELECT
PRINTER
MORE
source is the active source with highest priority. The RR/BR source cannot be manually selected. It is always CO2 if CO2 is active. Otherwise, it is ECG/RESP.
SELECTED SOURCE The user-selected HR/PR source is displayed along with the HR/PR
source currently used by the monitor.
SWEEP (mm/s) Selectable sweep speeds for HR/PR: 12.5, 25, 50 mm/sec.
Selectable sweep speeds for RR/BR: 3.13, 6.25, 12.5 mm/sec.
ALARM TONE Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.
HR/PR TONE Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.
BRIGHTNESS Sets the display brightness to NORMAL or LOW.
Warning
At the highest volume alarm level, the sound pressure level does not exceed safe limits (OSHA HSM 73-1101,
1972). However, additional precautions may be required in patients under treatment with ototoxic medications.
Welch Allyn
Chapter 2 – Setup 19
P2
25
25
S

Set Patient Alarms and Alarm Limits

1. From the Main Menu, press SETUP, ALARMS to access the Alarms Status Menu:
S
(M)
ALARM STATUS
APNEA :
HR/PR :
P1 :
P2 :
2 :
CO
STAT SET
SUSPEND
RR/BR :
SpO
2 :
NIBP :
TEMP :
ALL
ALARMS
( 15)
NIBP
( 85)
TEMP
°F
MCO2
mmHg
SpO2
%
LIMITS
All alarm limits are on.
At least one alarm limit is off.
mmHg
(M)
mmHg
T1 T2 T
38
PREVIOUS
MENU
D
S
122
D
58
100.4
98.6
1.8
BR
12
Br/m
97
9
Bells also appear in numerics windows.
(No bell) All alarm limits are off or the vital
sign is not being monitored.
2. Press LIMITS to display the Alarms Limits window:
(M)
(M)
38
mmHg
mmHg
T1 T2 T
ON/ OFF
D
S
122
D
58
99.8
98.6
1.2
BR
12
Br/m
97
9
Arrow indicates
there are more
parameters that
are not displayed.
Press NEXT
PARAMETER to
scroll down. After
the selection
reaches the last
parameter, it
returns to the top
(HR/PR).
ALARM LIMITS
UPPER LOWER
HR/PR BPM 120 50
*
30 20 100 OFF OFF 220 110 120
100.0
100.0
5.0
UP
RR/BR APNEA DELAY SpO
2
ETCO2 INCO2 NIBP S NIBP D NIBP (M) T1
T2 T
NEXT
PARAMETER
An asterisk indicates this alarm limit was violated during monitoring.
Red asterisk = alarm is occurring now.
Yellow asterisk = alarm has occurred since the last time this window was entered. The asterisk is removed when you exit this menu. The asterisk reappears if the limit is violated again.
Br/M
mmHg mmHg mmHg mmHg mmHg
NEXT
SETTING
sec %
°F °F °F
5
90
OFF
75 35 50
95.0
95.0
0.0
DOWN
( 15)
NIBP
( 85)
TEMP
°F
MCO2
mmHg
SpO2
%
3. Press NEXT PARAMETER to highlight the parameter you want to change, then press
NEXT SETTING to highlight the limit you want to change.
4. Press UP, DOWN, or ON/OFF to change the limits.
The apnea alarm cannot be turned off at any time.
5. After setting the desired limits, press Home to return to the Main Menu.
Propaq CS Directions for Use
Home
20

Change the Current Patient Mode

1. To change the current patient mode, from the Main Menu press SETUP, MORE, CHANGE to access the Patient Mode window:
PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:
NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT:
NEONATAL
> 9 YEARS
PEDIATRIC
BATTERY: 8.2 VOLTS
ADULT
SETUP
PREVIOUS
MENU
2. Based on the patient’s age, press NEONATAL, PEDIATRIC, or ADULT. When the confirmation window appears, press YES to confirm your selection.
Whenever you change the patient mode, the alarm limit settings, maximum NIBP cuff inflation pressures, and internal computations are automatically changed to the defaults for that patient mode. See page 21 for information about preset Factory patient modes or programmable Custom patient modes.
Note
If you change the patient mode, the CO2 alarm limits in the new mode might vary slightly from the
originally-programmed CO2 alarm limits for the new mode. Check the CO2 alarm limits.
Welch Allyn
Chapter 2 – Setup 21
P2
25

Change Powerup Patient Mode or Store Customized Settings

The Propaq CS monitor has standard, preset, default powerup settings and alarm limits for each patient mode: Adult, Pediatric, and Neonatal. These are “Factory Patient Mode” settings (listed on page 95).
You can also choose to customize and store programmable powerup settings and alarm limits for each patient mode. These are “Custom Patient Mode” settings.
The instructions below describe how to change the powerup patient mode and how to select and store new Custom Patient Mode settings.
Note
When you change patient modes, you also change the alarm limits associated with the new patient mode.
Change the Powerup Patient Mode
1. From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the Mode Setup window.
S
(M)
mmHg
(M)
mmHg
T1 T2 T
38
PREVIOUS
MENU
D
S
122
D
58
100.4
98.6
1.8
BR
12
Br/m
97
9
Asterisk indicates which patient mode is
currently
selected for
powerup.
MODE SETUP
FACTORY:
CUSTOM
* = PATIENT MODE ON POWERUP
NEXT
*ADULT PED
:
ADULT PED
POWERUP* USE
BATTERY: 9.2 VOLTS
NOW
NEO
NEO
( 15)
NIBP
( 85)
TEMP
°F
MCO2
mmHg
SpO2
%
SAVE
2. Press NEXT to highlight the desired Factory or Custom powerup mode, then press POWERUP* and YES. The new powerup selection is marked by the asterisk.
Changing the powerup mode does not affect the patient mode currently used.
Propaq CS Directions for Use
22

Customize Patient Mode Settings

1. From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the Mode Setup window.
2. The patient mode you want to reprogram (ADULT, PED, or NEO) must be currently selected.
To make sure it is currently selected, press NEXT as needed to highlight the desired mode, then press USE NOW and YES.
3. Press Home to exit the Mode Setup window, then use other menus and keys to set the
monitor settings and alarm limits as desired.
A convenient way to access settings and alarm limits for all functions without connecting cables is to select the in-service mode (disconnect all patient cables, turn the monitor power off and then on, then press
SETUP, WAVE SELECT, INSERVICE from the Main Menu).
Home
Warning
If any alarms are set to OFF and you select SAVE to store settings for a Custom patient mode, those alarms will be OFF when the monitor powers up in that Custom patient mode or that Custom patient mode is selected. Consider carefully before setting Custom patient mode powerup alarms to OFF.
4. Re-enter the Mode Setup window, press NEXT as needed to highlight the desired Custom mode, then press SAVE and YES.
If the in-service mode is used, turn off the monitor to exit the in-service mode.
Welch Allyn
3 – Monitoring

Perform ECG/RESP Monitoring

Warning
23
pneumography can be used to detect central apnea. However, apnea episodes with continued respiratory effort, such as obstructive apnea and mixed apnea, may go undetected. Also, artifact due to patient motion, apnea mattress shaking, or electrocautery use may cause apnea episodes to go undetected. Always monitor and set alarms for SpO2 when using impedance pneumography to monitor respiratory function.
ECG R-wave detection. Therefore, always select the ECG lead with the most prominent QRS complex when monitoring respiration via impedance pneumography.
RESP measurement frequencies may interfere with one another.
is not recommended for use on paced patients.
possible.
failure to detect apnea, or a failure to display a Lead Fail message.
may cause the monitor to display inaccurate heart rates.
per minute, it may display incorrectly low heart rates, due to intermittent picking of R-waves.
electrosurgical unit, can induce heat into electrodes and cables which can cause burns on the patient. Reading errors and damage to equipment may also result. This hazard can be reduced by (1) avoiding the use of small ECG electrodes, (2) selecting ECG electrode attachment points remote from the surgical site and from the electrosurgical return electrode, (3) using electrosurgical return electrodes with the largest practical contact area, and (4) assuring proper application of the electrosurgical return electrode to the patient.
settings.
SpO
outlet or make contact with other hazardous electrical potentials including earth ground. To prevent damage during defibrillation, don’t use ECG cables without 1 k series resistors.
again. This clears the prior patient’s trend values and alarm limit settings.
Impedance pneumography detects respiratory effort via changes in chest volume; therefore, impedance
The Propaq CS monitor automatically rejects cardiovascular artifact. This function is dependent upon accurate
Don’t place the Propaq CS monitor with RESP in close proximity to another respiration monitor because the
Because pacemaker pulses in some instances may be falsely counted as breaths, impedance pneumography
Motion artifact can cause incorrect breath rate or heart rate readings. Minimize patient motion whenever
If a disconnected lead is in too close proximity to other electrical devices, it may cause false heart rate, a
The Propaq CS monitor does not provide arrhythmia analysis. Therefore, arrhythmias are not analyzed and
The Propaq CS monitor will show + + + for HR numerics between 301-350 beats per minute. Above 350 beats
High-intensity radio frequency (RF) energy from external sources, such as an improperly connected
Verify patient mode. Incorrect patient mode may result in inaccurate heart rates and inappropriate alarm
To help prevent injury, use the provided garment clips to route the ECG cables away from the patient’s head.
Use of ECG cables with loose or faulty detachable lead wires may cause erratic behavior of the ECG waveform,
(C-LOCK), and NIBP (Smartcuf) due to intermittent ECG lead wire connections.
2
Use only ECG safety cables that are designed so that they cannot accidently be plugged into an ac mains
Before you use a Propaq CS monitor on a new patient, always turn it off for a few seconds, then turn it on
Propaq CS Directions for Use
24
Caution
To protect the Propaq CS monitor from damage during defibrillation, for accurate ECG information, and for protection against noise and other interference, use only ECG electrodes and cables specified or supplied by Welch Allyn (these cables have the required current-limiting resistors). Follow recommended application procedures.
• Impedance pneumography (RESP) is not recommended for use with high frequency ventilation.
• Since RESP is derived from the same leads as the ECG channel, the Propaq CS monitor determines which signals are cardiovascular artifact and which signals are a result of respiratory effort. If the breath rate is within five percent of the heart rate or a multiple or sub-multiple of the heart rate, the monitor may ignore breaths and trigger an apnea alarm.
• When monitoring RESP it is highly recommended that you use SpO2 monitoring as a backup monitoring method.
• The Propaq CS monitor counts as “breaths” respiratory efforts that are larger than two times background cardiovascular artifact.
• Even though the Propaq CS monitor contains fully isolated patient-connected circuitry, it has not been specially designed for direct application on a patient’s heart.
• Use only with accessories provided or recommended in the Welch Allyn Products and Accessories booklet.
• Severe artifact and interference (such as defibrillation interference) can cause the waveform to move off the display for a few seconds before it is restored.
Prepare for ECG/RESP Monitoring
1. Inspect the ECG cable and replace it if it shows signs of wear, breakage, or fraying.
2. Select the appropriate patient mode. To change patient modes, from the Main Menu press
SETUP, MORE, CHANGE, then the desired patient mode (NEONATAL, PEDIATRIC, or ADULT) and then YES.
3. Select electrode sites on the patient.
Choose flat areas; avoid fatty areas and major muscles.
4. Shave or clip hair from electrode sites, thoroughly clean skin, and lightly rub dry.
You may use soap and water, isopropyl alcohol or special skin preparation pads. To avoid allergic reactions to electrodes, refer to the electrode manufacturer’s directions.
5. If you are using pre-gelled electrodes, make sure the electrode date is not expired and the gel is intact and not dried out. For best results, use only silver/silver chloride electrode.
If you are using non-gelled electrodes, apply a 1/4 to 1/2 inch mound of gel over the electrode contact area.
For best product performance and measurement accuracy, do not use stainless steel needle electrodes, squeeze bulb electrodes, or electrodes with dissimilar metals. Do not use electrodes from more than one manufacturer on the same patient.
Welch Allyn
Chapter 3 – Monitoring 25
6. Attach lead wires to the electrodes before applying them to the patient. Apply the electrodes to the patient as shown.
RA
LA
G
+
If using MCL1,
Standard 3-lead Configuration
RA
V1
5-lead Configuration
RL
LL
LA
LL
select LEAD II on the monitor, and use all three electrodes.
V6
One of six V lead electrode placement sites for the C lead.
7. Plug the ECG cable into the ECG connector on the monitor's left side panel.
Marriott Configuration MCL 1
RA
Neonate 3-lead Configuration
(G)=LA (-)=RA (+)=LL
LA
LL
8. Support the ECG cable so it does not stress the electrode wires, ECG cable connectors, or electrodes.
9. If an electrosurgical unit will be used, place the ECG cable and electrode wires as far as possible from the surgical site and from the electrosurgical return electrode and its cables. This minimizes interference.
Although the ECG channel contains electrosurgical interference suppression (ESIS) circuitry, noise artifact may be displayed on the ECG trace while an electrosurgical device is in use. Choose electrode placement to minimize interference.
10. Look for an ECG waveform and heart rate on the monitor. Depending on how your monitor
is programmed, a beep tone may occur with each detected QRS.
If there is no waveform, check the electrodes, wires, cable, and the monitor for a possible misconnection or lead fault.
Propaq CS Directions for Use
26
P1
121
S
mV
11. To set up the ECG/RESP display, from the Main Menu press
ECG/RESP to display the first
ECG/RESP menu:
ECG1 lead
selected
ECG1 size
selected
ECG2 lead
The ECG2 lead
is always V,
except when the
ECG1 lead is V
(in that case the
ECG2 lead is II).
RESP waveform
is displayed (with
RESP option)
RESP lead
selected
RESP size
selected
NET OFF
II
MON
V
MON
2
Ld
RESP
mV
1
cm
mV
1
cm
ECG1
SIZE
ADULT ON NETWORK
2X
17:05:10
ALLEN KIM
ECG
1
LEAD
PRINT FAULT
ECG2 SIZE
RESP SIZE
HR
BPM
112
1
P
(M)
( 96)
P2
(M)
( 15)
NIBP
(M)
( 85)
TEMP
°F
MCO2
38
mmHg
SpO2
%
12. Press buttons as desired to adjust the display:
ECG1 SIZE Selects the ECG1 waveform size: 4, 2, 1, 0.5, or 0.2 mV/cm.
ECG1 LEAD Selects the ECG1 lead: I, II, III, aVR, aVL, aVF, or V.
mmHg
mmHg
mmHg
T1 T2 T
MORE
S
121
D
79
S
25
D
S
122
D
58
100.4
98.6
1.8
BR
12
Br/m
97
9
BR indicates respiration rate is derived from active CO2
channel. RR indicates CO
is not monitored and respiration rate is derived from ECG/RESP.
2
Selections aVR, aVL, aVF, and V are only available with a 5-lead ECG cable.
The ECG2 lead is always V, except when the ECG1 lead is V (in that case the ECG2 lead is II).
ECG2 SIZE Selects the ECG2 waveform size: 4, 2, 1, 0.5, or 0.2 mV/cm.
RESP SIZE Selects the RESP waveform size: 1x, 2x, 4x, 8x, or 16x.
The QRS detector sensitivity threshold is not affected by changing the ECG display size. Likewise, the RESP breath detector threshold is not affected by changing the RESP display size.
13. Press MORE to display the second ECG/RESP menu and status window:
ECG bandwidth selected:
MON = Monitor
EXT = Extended
1
cm
MON
ECG/RESP
HR/PR TONE
PACER DISPLAY
ECG BANDWIDTH
RESP LEAD
: OFF
: ON
: MONITOR (0.5-40Hz)
: Ld2 (RA-LL)
RESP MONITORING : ON
NEXT
CHANGE
(M)
mmHg
( 96)
P2
(M)
( 15)
mmHg
NIBP
(M)
( 85)
mmHg
TEMP 100.4
°F
MCO2
38
mmHg
SpO2
%
PREVIOUS
D
79
S D
S
122
D
58
T1 T2 T
BR
Br/m
97
MENU
25
9
98.6
1.8
12
Welch Allyn
Chapter 3 – Monitoring 27
14. Press NEXT and CHANGE as desired to adjust the display.
HR/PR TONE Sets heart tone loudness to LOW, MEDIUM, HIGH, or OFF.
If SpO2 is monitored, tone pitch varies with the SpO2 value.
PA CER DISPLAY Turns on and off the pacer indicator in the ECG waveform.
If the patient has a pacemaker, you may want to turn on the pacer indicator (see page 28).
ECG BANDWIDTH Selects the bandwidth for displayed and printed data.
MONITOR is 0.5-40 Hz (Adult mode) or 0.5-120 Hz (Pediatric and
Neonatal mode).
Monitor Mode filters out extraneous noise and artifact to provide a more stable display.
EXTENDED is 0.05-40 Hz (Adult mode) or 0.05-120 Hz (Pediatric and
Neonatal mode).
Extended Mode is a higher-resolution setting that allow more detailed analysis.
Always use Extended Mode when observing ST segment morphology on the display or printer. Although Monitor Mode is useful to minimize baseline wander due to artifact, ST segments can be distorted in Monitor Mode. This can potentially cause underestimation of ST elevation and overestimation of ST depression. Although the monitor does not have automated ST segment monitoring, ST segments may be accurately displayed and printed in Extended Mode.
RESP LEAD Selects the RESP lead: Ld1 (RA-LA) or Ld2 (RA-LL). RESP lead
selection is independent of ECG lead selection.
Choose the RESP lead that gives you the best signal. If neither signal is adequate, experiment with nonstandard electrode placement such as placing the RA and LA electrodes on the respective mid-axillary lines just above the level of the nipples.
RESP MONITORING Turns RESP on or off.
15. Set alarms according to your facility’s standards.
Use the ECG Filter to Display a Better Waveform
If the ECG waveform appears unclear or distorted, make sure the monitor ECG filter is properly set to reduce interference from your facility’s ac power frequency. To check the filter:
1. Press SETUP, MORE, MORE, SERVICE, YES to access the Service Menu.
2. Press MORE, MORE, SETTINGS to display the Settings Menu.
3. If the FILTER setting does not match your ac power frequency (60 or 50 Hz), press NEXT to
highlight FILTER, then press CHANGE to change settings.
Contact a qualified service person if you have questions.
Propaq CS Directions for Use
28

Use the Propaq CS Monitor With Pacemaker Patients

Warning
Pacemaker signals can differ from one pacemaker to the next. The Association for Advancement of Medical Instrumentation (AAMI) cautions that “in some devices, rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. All pacemaker patients should be kept under close or constant observation.”
The presence of much pacer-like noise can cause the displayed heart rate to be erratic even though the ECG trace may look undistorted with the pacer indicator off. To help avoid this noise problem, use fresh ECG electrodes and make sure the ECG cable lead wires make good connections.
If the patient being monitored has a pacemaker, the Propaq CS monitor detects and can indicate the occurrence of pacemaker signals.
1. To access the monitor pacer indicator, from the Main Menu press ECG/RESP, MORE, and then
NEXT to select the PACER DISPLAY.
2. Press CHANGE to set the PACER DISPLAY either ON or OFF.
When ON, the monitor displays (and prints on printouts) vertical dashed lines to indicate each time a pacemaker signal is detected. (If the pacemaker signal is strong enough, the monitor also displays it as a waveform “spike.”)
Pacemaker signal indicators
NET OFF
II
MON
mV
1
cm
ADULT ON NETWORK
17:05:10
ALLEN KIM
PRINT FAULT
HR
BPM
80
1
P
(M)
mmHg
( 96)
P2
S
121
D
79
S
25
When OFF, the vertical lines are not displayed (or printed), but the pacemaker signal waveform spike is still displayed if strong enough.
Note
Pacemaker pulses are not counted as heartbeats as defined by the Pacer Pulse Rejection specifications (see page 98).
Noise on the ECG signal may be detected as pacer signals, causing the pacer indicator to appear on the display. If you don't need to indicate pacemaker signals, turn off the pacemaker indicator for a better ECG waveform display.
Welch Allyn
Chapter 3 – Monitoring 29
II
BPM
112

Perform Invasive Blood Pressure (IBP) Monitoring

Warning
connected to its cable shield. Using a conductive transducer case with such a shield connection risks high-frequency burns at the ECG electrodes if the transducer case becomes earth grounded.
functions, partial disconnection will not be detected, nor will the use of some incompatible transducers. The user must exercise reasonable measures to ensure that approved transducers are used and that pressure transducers are connected properly.
again. This clears the prior patient’s trend values and alarm limit settings.
recommended in the Welch Allyn Products and Accessories booklet. Use accessories according to your facility’s standards and the manufacturer’s recommendations. Always refer to the manufacturer’s Directions for Use. Do not use light-sensitive disposable transducers.
If electrocautery is used, always avoid using any transducer with a conductive (metal) case that is electrically
Although complete disconnections of invasive pressure transducers will be detected by the normal alarm
Before you use a Propaq CS monitor on a new patient, always turn it off for a few seconds, then turn it on
For best product performance and measurement accuracy, use only accessories supplied by Welch Allyn or
1. Inspect the transducer cable and transducer dome for wear, breakage, or fraying. Replace any worn or broken accessory.
2. Set up the transducer according to your hospital's procedures. Always refer to the transducer manufacturer’s Directions for Use. If the transducer is a disposable unit with separate cable, connect the transducer to the transducer cable.
3. Plug the transducer (or transducer cable) into an invasive pressure connector on the monitor left side panel.
MON
MON
P
1
mV
1
cm
V
mV
1
cm
RANGE
RESCALE
ZERO P1
ZERO P
P
1
NIBP
( 85)
TEMP
°F
MCO2
mmHg
SpO2
%
2
ZEROED
(M)
38
NOT
mmHg
T1 T2 T
MORE
S
122
D
58
100.4
98.6
1.8
BR
12
Br/m
97
The message NOT ZEROED immediately appears in the IBP numerics window for the invasive pressure channel being used.
4. To zero the transducer, open the transducer's stopcock to atmospheric air. Wait a few seconds for the transducer to settle.
Before zeroing, make sure the transducer cable is properly connected to the monitor and the transducer is open to atmospheric air and positioned at the same level as the patient’s heart. The monitor will not zero the transducer if the pressure waveform is pulsatile, there is too much signal noise, or the transducer’s offset is too great.
Propaq CS Directions for Use
30
25
S
5. If the ZERO menu is not displayed, from the Main Menu press
P1
(or ZERO P2). The word ZEROING appears in the numerics window during zeroing.
INVASIVE PRESSURE, then ZERO
If you want to cancel the zeroing process, press CANCEL.
6. Wait for a brief tone to sound and the word ZEROED to appear in the blood pressure numerics window.
7. Close the transducer's stopcock. The monitor displays the pressure scale and numerics.
8. If the transducer will not zero, the monitor displays the words ZERO REJECTED in the
numerics window. Press CANCEL and try zeroing again beginning at step 4. The monitor does not display numerics or scales until an acceptable zero reference is established.
You can rezero an IBP transducer at any time after you again open the transducer stopcock to atmospheric air. If the transducer has already produced pressure readings, rezeroing provides a new zero reference for the monitor.
If the zero value is not accepted, the monitor continues to use the previous zero reference and displays numerics and waveforms based on that value.
If the transducer still does not zero, try another transducer or another cable.
Warning
If you press ZERO after an invasive pressure channel has been successfully zeroed and the channel is currently monitoring a pressure waveform, the message ZERO REJECTED will display in the IBP numerics window. This message continues to display in place of the valid invasive pressure numerics until you press Home, INVASIVE PRESSURE, and then CANCEL in the IBP menu. If an IBP alarm occurs while ZERO REJECTED is displayed in place of IBP numerics, the IBP numerics will not flash to indicate invasive pressure is in alarm.
9. To set up the IBP display, from the Main Menu press INVASIVE PRESSSURE to display the first IBP menu:
(M)
D
mmHg
S
(M)
D
mmHg
T1 T2 T
38
MORE
PREVIOUS
MENU
122 58
100.4
98.6
1.8
BR
12
Br/m
97
9
First IBP menu
Second IBP menu
IBP labels are
user-selectable.
V
MON
P
1
P
2
mV
1
cm
130
100
70
30
15
0
RANGE
LABEL
P
( 15)
NIBP
( 85)
TEMP
°F
MCO2
mmHg
SpO2
%
RESCALE
1
FORMAT
P
1
ZERO P1
LABEL
P
2
ZERO P2
FORMAT
2
P
Welch Allyn
Loading...
+ 92 hidden pages