For many decades WEINMANN has developed,
produced and marketed appliances for oxygen
therapy, inhalation therapy and emergency
medicine.
In 1957 WEINMANN put the first diaphragm
pressure reducer on the market.
Pressure reducers are used in medical oxygen
inhalation. They reduce the pressure delivered by
oxygen bottles from 200 bar to the specified
operating pressure.
The flowmeter is used during the inhalation of medical oxygen via the central gas supply system of a
hospital, for example.
Pressure reducers are a functional component of a
wide range of appliance combinations such as
stationary oxygen appliances and portable oxygen devices. They are therefore an important
component in the treatment of chronic respiratory
diseases and/or disorders of the cardiovascular
system.
The aim of these Service and Repair Instructions is to
familiarize you, a
the function, technology, servicing and repair of
pressure reducers and flowmeters. In conjunction
with a training course that you have already been
trained and competent expert,
with
given by WEINMANN, you now belong to that
group of “trained and competent experts”, which
means you can give your customers proper instruction, rectify faults yourself and perform the prescribed functional check and any repairs in
accordance with these Service and Repair Instructions.
In the event of any warranty claim the device must be
sent to WEINMANN
So that we can process any warranty claims or
requests for favorable treatment, please send us
the customer’s purchase slip (invoice) with the
appliance.
Repairs and/or repair work may only be performed by WEINMANN or by trained and competent experts.
You are responsible for any repairs you carry out
yourself and the relevant guarantees!
Only genuine WEINMANN replacement parts
used for repairs.
Please remember:
Your customers put their trust in you and rely on
your efficiency, just as you rely on WEINMANN.
.
may be
Note:
You will find the following information in the operating instructions for the devices:
Multistep pressure reducer Fast II High Flow
WM 31899
8 Flow gage
Indexed pressure reducer FAST
WM 30600; 30800; 30850
1 Contents gage
6 Index head
Oxygen outlet
(to consumer)
1 Contents gage
15 Connection nut
2 Connector (oxygen
bottle)
4 Body
22 Cover
Oxygen outlet
(to consumer)
9 Adjuster knob
Variable pressure reducer FINE
WM 30500; 30700; 30750
4
Overview
Page 5
1 Contents gage
15 Connection nut
2 Connector (oxygen
bottle)
4 Body
22 Cover
52 Body stage 2
Oxygen outlet
(to consumer)
Two-stage pressure reducer
OXYTRON WM 30150
2 Connector
22 Cover
3 Rating plate
6 Index head
13 Service plate
20 Plate for changing
filter screw
14 Sealing ring
Oxygen outlet
(to consumer)
Indexed flowmeter
OXYWAY Click WM 31030
12 Filter screw
10 Appliance number
11 Safety valve
Rear view of pressure reducer
Overview
5
Page 6
1.1 Explanation of Numbers
Type: 30800
The type row indicates the basic design of the
pressure reducer, e.g. “30800” for indexed
pressure reducers (Fast I).
Order number:
In the order number the first three numbers are a
short indication of the type, and the last two digits
indicate the specific version (e.g. the connection
variants).
Appliance number:
The first two digits of the appliance number show
the year of manufacture. The next five digits are the
consecutive appliance serial number within each
type. The appliance number is stamped on the
back of the pressure reducer.
1.2 Information about conformity with standards
Up to serial number 0849999, OXYWAY
pressure reducers meet standard EN 738-1.
From serial number 0850000, OXYWAY
pressure reducers meet standard EN ISO 105241:2006.
In the course of adapting the products to standard
EN ISO 10524-1:2006, it was not possible to
design all components to be downward-compatible. Parts which are not downward-compatible are
available for both device statuses separately and
are marked accordingly in the replacement parts
lists (see “8. Replacement Parts” on page 64) of
these instructions.
•Fit replacement parts which are not downward-compatible only
pressure reducer serial numbers quoted.
•Do not convert pressure reducers up to
SN 0849999 “to suit the new standard”.
in accordance
with the
Caution!
Malfunction as a result of unclear device status.
If replacement parts which are not downward-compatible are used in the wrong pressure reducers,
connected devices will malfunction.
6
Overview
Page 7
2. Function
2.1 General
During operation, the oxygen coming from the
oxygen bottle passes through the bottle valve and
the connector 2 into the body 4 of the pressure
reducer. You can read off the bottle pressure (delivery pressure) at the contents gage 1.
WEINMANN piston pressure reducers are of
modular design. This makes it possible to keep the
number of replacement parts to a minimum.
A filter screw with integrated sinter filter (see explanation on right) traps any impurities. A further sinter
filter absorbs pressure surges in the oxygen flow.
The filter screw 12 is located at the entrance to the
connector 2. The second sinter filter is fitted between
the connector 2 and the body 4.
In the reduction stage the bottle pressure is reduced
to the required nominal outlet pressure, depending
on type. The principle used, with crater drill hole
and spring-loaded piston, ensures especially constant pressure in spite of variations in flow.
All pressure reducers are protected against excess
pressure by a safety valve 11.
Explanations:
Sinter filter: Small spherical bronze particles are
pressed together under heat, creating a filter with a
particularly fine pore structure (50 - 75 µm).
Flow: Outlet volume; expressed in liters per minute.
A fixed orifice in oxygen outlet 7 creates a constant
flow.
The pressure reducers in this series differ in the
length and position of the connector 2 and the
oxygen outlet 7.
2
7
Function
7
Page 8
2.3 Variable pressure reducer FINE (WM 30500; 30700; 30750)
The piston is acted upon by a second spring which
can be adjusted via a spindle system and thereby
creates different outlet pressures. In conjunction
with the fixed orifice in the oxygen outlet 7 you can
therefore make continuously variable adjustments
to the flow. You can read off the flow setting at the
flow gage 8.
8
2
The pressure reducers in this series differ in the
length and position of the connector 2.
7
2.4 Indexed pressure reducer FAST (WM 30600; 30800; 30850)
As in the fixed pressure reducers, this creates a
constant operating pressure (outlet pressure).
The index head 6 contains an orifice disc with
various orifice diameters which are aligned by an
index (click-in) system with the oxygen outlet 35.
This enables you to set various flow rates.
The pressure reducers in this series differ in the
length and position of the connector 2. A pressure
reducer of type WM 30850 has two oxygen
outlets.
7
35
2
6
2.5 Multistep pressure reducer Fast II High Flow (WM 31899)
As in the Fast multistep pressure reducers, there is an
orifice disk with a variety of orifice diameters in adjuster 6 which is brought into line with oxygen outlet 35
by means of a locating system.
35
In addition to the flow outlet, there is a pressure outlet
for medical devices with an oxygen requirement in excess of 90 l/min. The pressure outlet and the flow outlet may not be in operation simultaneously.
7
6
8
Function
Page 9
2.6 Two-stage pressure reducer OXYTRON (WM 30150)
The two-stage pressure reducer has a first stage in
the form of a fixed pressure reducer WM 30100,
and a second reduction stage 52 to permit optimal
adjustment of pressure and flow to suit the
OXYTRON oxygen device.
In the second stage the pressure of 4.5 bar from
the first stage is similarly reduced by a system of
crater drill hole and spring-loaded pistion to
1.6 bar.
A fixed orifice in the oxygen outlet 7 generates a
constant flow of 12 l/min.
2.7 Indexed flowmeter OXYWAY Click
During operation, the oxygen supplied by the central
gas system at a pressure of 4.5 bar passes through the
inlet connector 2 into the body 22 of the OXYWAY
Click.
The dial head 6 contains an orifice disk with orifices of different diameters which are lined up with
the oxygen outlet 35 by means of a click dial system. You can set the following flow rates: 1, 2, 3,
4, 5, 6, 9, 12, 15 l/min..
52
7
235
6
22
Function
9
Page 10
2.8 Oxygen outlets pressure reducer OXYWAY
Oxyway pressure reducers are available with different oxygen outlets: the pressure outlet and the
flow outlet are different(see also “10. Technical
Data” on page 79).
Flow outlets are intended for the direct supply of
the patient, e.g. via nasal cannula, mask or
Flow outlet
(Fix I, Lateral outlet Fix I)
9/16 UNF thread
Pressure and flow outlet
(Fast II)
(Fine I up to III)
9/16 UNF thread
Pressure and flow outlet
(Fast II High Flow)
intermediate humidifier. Flow outlets have a 9/16
UNF thread.
Pressure outlets are intended to supply ventilators
such as the Medumat, modules or distributor bars.
Pressure outlets have a G 3/8" thread.
Flow outlet
Flow outlet
(Fast I and III)
9/16 UNF thread
Pressure outlet
(Fix III, Fix III left)
G 3/8"
thread
9/16 UNF
thread
Flow outlet
(Oxytron)
18 x 11 thread
G 3/8"
thread
G 3/8" thread
9/16 UNF
thread
Flow outlet
(Oxyway Click)
9/16 UNF thread
10
Function
Page 11
3. Final Check
3.1 General
Test conditions:
•Ambient temperature: 15 °C
•Ambient pressure: 1013 hPa
•Connection to oxygen supply. For flow tests, a
constant pressure* of 100 bar or 200 bar is
required.
* 100 bar up to SN 0849999
200 bar from SN 0850000
All the tolerances listed in these instructions (see
“10. Technical Data” on page 79) relate to these
conditions. Note that if different ambient conditions prevail, deviating measuring results may be
obtained.
For the check you will need:
•shutoff valve (max. 6 bar) after oxygen outlet
•test pressure gage 0 - 10 bar
•test pressure gage 0 - 2.5 bar
and three flow meters:
•0 - 50 ml/min
•0 - 20 l/min
•0 - 220 l/min.
If the final check reveals faults, the device must not
be used until the defect has been rectified. For possible causes of the defect and how to remedy
them, see Chapter “5. Troubleshooting” on
page 18.
A complete check of the OXYWAY Fix, Fast and
Fine pressure reducer includes:
•Visual inspection for mechanical damage,
•“3.3.1 Testing system for leaks” on page 12,
•“3.3.2 Testing safety valve for leaks” on
page 13,
•“3.3.3 Checking “0” position of contents
gage” on page 13,
•“3.3.4 Checking “0” position of flow gage on
pressure reducer FINE” on page 13
•“3.3.5 Checking for leaks with valve closed
on pressure reducer FINE” on page 14,
•“3.3.6 Checking adjustable flow rate on pressure reducer FINE” on page 14,
•“3.3.7 Checking adjustable flow rate on pressure reducer FAST” on page 14.
•“3.3.8 Checking constant flow rate in pressure reducers FIX, FAST and
FAST II high Flow” on page 15
•“3.3.9 Checking static outlet pressure p4 for
all pressure reducers” on page 15.
A complete check of the OXYWAY Click flowmeter includes:
•“3.4 Checking the adjustable output (flow)
on the OXYWAY Click” on page 15.
We recommend that you always keep in stock:
•Replacement seal set WM 1148,
•Filter screw WM 30905.
3.2 Intervals
After any servicing or repair work
•Perform a final check.
Final Check
11
Page 12
3.3 Performing check
3.3.1 Testing system for leaks
1. Close system after pressure reducer, e.g. by
closing a valve after the pressure reducer.
2. Check that all screwed joints and hoses are
firmly seated. If necessary tighten joints by
hand.
3. Slowly open valve of oxygen bottle. Contents
gage 1 of pressure reducer now shows bottle
pressure, e.g. 200 bar.
4. To detect leaks, wet screwed connections with
a soap-and-water solution. Use Lanosan
for this purpose.
Warning!
Risk of explosion from soap penetration.
Leaks can cause soap to penetrate the system
and combine with the oxygen to form an explosive mixture.
•Open the O2 valve before wetting parts of the
system with a soap-and-water solution. The
system is then pressurized. No soap can penetrate the system.
•Remove the soap-and-water solution without
leaving any residues after the test is complete.
5. Close bottle valve again.
6. Observe needle of contents gage 1 for about
1 minute. If needle stays in same position, the
system is gas-tight. If the needle falls steadily,
there is a leak.
®
med
1
Eliminating leaks
We recommend that you keep a stock of replacement seals for the connections.
1. Prepare solution of soapy water, using perfume-free soap.
2. Wet all screwed joints with solution. If bubbles
form, this indicates a leak.
3. Release pressure in system:
– Close oxygen bottle valve.
– Open valve after pressure reducer until
contents gage 1 reads “0”.
12Final Check
4. Replace faulty seal responsible for leak (see
“6.2 Replacing sealing ring in connector” on
page 20).
Important!
Screwed joints of oxygen line must only be hand
tightened.
5. Check for leaks again.
6. If leak cannot be eliminated, unit must be
repaired.
Page 13
3.3.2 Testing safety valve for leaks
1. Test must be performed at maximum flow. In
the Fine and Fast series this is set accordingly.
2. Slowly open oxygen bottle valve.
3. Test with finger whether oxygen is escaping
from safety valve 11.
If oxygen is escaping, have pressure reducer
repaired by manufacturer (WEINMANN).
Alternatively, you can also wet the safety valve
with soap-and-water solution to find leaks. Use
Lanosan® med.
Warning!
Risk of explosion from soap penetration.
Leaks can cause soap to penetrate the system
and combine with the oxygen to form an explosive mixture.
•Open the O2 valve before wetting parts of the
system with a soap-and-water solution. The
system is then pressurized. No soap can
penetrate the system.
•Remove the soap-and-water solution without
leaving any residues after the test is complete.
11
3.3.3 Checking “0” position of contents gage
1. Slowly open oxygen bottle valve.
Contents gage
1 now shows bottle pressure.
2. To release pressure in system:
– Close oxygen bottle valve
– Switch on connected appliance
– Open shutoff valve after oxygen outlet.
3. Needle of contents gage
“0”. Wait until needle stops moving. Now
switch appliance off.
4. Check whether needle points exactly to “0”.
If not, fit replacement contents gage (see “6.5
Replacing gage” on page 27).
3.3.4 Checking “0” position of flow gage on pressure reducer FINE
1. Slowly open oxygen bottle valve.
2. Set flow rate (e.g. 7 l/min) at adjuster knob
You can read the set value at the flow gage 8.
3. To release pressure in system:
– Close oxygen bottle valve
– Open shutoff valve after oxygen outlet
– Switch on connected appliance.
4. Needle of flow gage 8 now falls toward “0”.
Wait until needle stops moving. Now switch
appliance off.
9.
1
8
1 now falls toward
9
5. Check whether needle points exactly to “0”.
If not, fit replacement gage (see “6.5 Replacing gage” on page 27).
Final Check13
Page 14
3.3.5 Checking for leaks with valve closed on pressure reducer FINE
You need a flow meter with a range of 0 – 50 ml/min.
Perform measurement at an inlet pressure of 200 bar.
1. Slowly open oxygen bottle valve.
Contents gage 1 now shows bottle pressure.
2. Check whether flow gage 8 shows a flow rate
of “0”.
If not, use adjuster knob 9 to set flow rate to
“0”.
3. Connect flow meter (0 - 50 ml/min) to oxygen
outlet 7.
4. Wait until flow meter shows constant reading.
5. Read off value.
If value is greater than 30 ml/min, fit replacement piston (see “6.7 Replacing piston in pressure reducer FINE (WM 30500, 30700,
30750)” on page 30).
3.3.6 Checking adjustable flow rate on pressure reducer FINE
You need a flow meter with a range of 0 – 20 l/min.
Perform measurement at an inlet pressure of 100 bar ± 10 bar.
1. Slowly open oxygen bottle valve.
Contents gage 1 now shows bottle pressure.
2. Turn adjuster knob 9 to the left as far as it will
go, until flow gage 8 shows a flow rate of “0”.
3. Connect flow meter to oxygen outlet 7.
4. Set any desired flow rate at adjuster knob 9.
Note: Checking the flow of the second oxygen outlet is described in a later section (3.3.8, page 15).
5. Check on flow meter whether actual value
agrees with set value.
Note tolerances of pressure reducer:
1 to 5 l/min: ± 0.5 liters,
6 to 15 l/min: ± 10 %.
6. If any value does not agree, have pressure reducer repaired by manufacturer
(WEINMANN).
3.3.7 Checking adjustable flow rate on pressure reducer FAST
You need a flow meter with a range of 0 – 20 l/min.
Perform measurement at an inlet pressure of 100 bar ± 10 bar.
1. Slowly open oxygen bottle valve.
Contents gage 1 now shows bottle pressure.
2. Set flow of “0” at index head 6.
3. Connect flow meter to oxygen outlet 7.
4. Set any desired flow rate at index head 6.
Note: Checking the flow of the second oxygen outlet is described in a later section (3.3.8, page 15).
5. Check on flow meter whether actual value
agrees with set value.
Note tolerances of pressure reducer:
1 to 5 l/min: ± 0.5 liters,
6 to 15 l/min: ± 10 %.
6. If any value does not agree, have pressure
reducer repaired by manufacturer
(WEINMANN).
14Final Check
Page 15
3.3.8 Checking constant flow rate in pressure reducers FIX, FAST and
FAST II high Flow
– Contents gage 1 now shows bottle pressure
– Flow meter shows flow rate.
4. Check flow rate reading against set value.
Pressure reducers must maintain the following
flow rates:
WM 30050, 30100: 4 l/min ± 0.5 l/min
WM 30150:12 l/min ± 1.2 l/min
WM (30200), 30300,
30350:120 l/min ± 15 l/min
WM 30850:90 l/min ± 10 l/min
WM 31899:min. 160 l/min
5. In the event of discrepancies, have pressure reducer repaired by manufacturer
(WEINMANN).
6. There is a second oxygen outlet on pressure
reducer WM 30200. This is a free pressure
outlet.
3.3.9 Checking static outlet pressure p4 for all pressure reducers
1. Connect appropriate test gage to oxygen
outlet 7.
2. Check closing pressures p4 (see “10. Technical Data” on page 79). They are valid for a
delivery pressure of p
* = 100 bar or
1
200 bar and must be reached within 1 minute
at most. After this they must not show any further increase.
* 100 bar up to SN 0849999
200 bar from SN 0850000
3.4 Checking the adjustable output (flow) on the OXYWAY Click
You need a flow meter with a range of 0 – 20 l/min.
Perform the measurement at an inlet pressure of 4.5 bar ±0.2 bar.
1. Set the flow "0" at index head 6.
2. Plug the flowmeter into a suitable compressed
gas supply (e.g. a central gas supply system).
3. Connect the flow meter to the oxygen outlet.
4. Set any flow value at index head 6.
5. Check on the flow meter whether the set value
corresponds to the actual value.
Please observe the tolerances of the flowmeter in
the process.
Rising test:
1 to 5 l/min: ±0.5 liter,
6 to 15 l/min: ±10 %.
Falling test:
6 ±10 % and 0/min: ±0.15 l/min
Final Check15
Page 16
4. Servicing
4.1 General
We recommend that all maintenance work, such
as inspections and repairs, be carried out by the
manufacturer (WEINMANN) or by expert technical personnel trained by WEINMANN.
When carrying our repairs, be sure to observe the
directions in these Service and Repair Instructions.
4.2 Intervals and scope for OXYWAY and OXYTRON pressure
reducers
In cases of dirt accumulation, or at least every 2 years:
1. Fit replacement filter screw 12 (see “6.3 Re-
placing filter screw” on page 21).
Every 4 years
1. Fit replacement filter screw 12 (see “6.3 Re-
placing filter screw” on page 21);
2. Completely replace piston 16.
– For pressure reducers WM 30050,
30100, 30200, 30300 and 30350:
see Section 6.6, page 28,
– For pressure reducers WM 30500,
30700, 30300 and 30750:
see Section 6.7, page 30,
– For pressure reducers WM 30600;
30800, 30850:
see Section 6.8, page 33;
2. Perform a functional check (see “6. Functional
Check“ on page 12 of the Operating Instructions”).
– For pressure reducers WM 31899;
30800, 30850:
see Section 6.9, page 36;
3. Fit replacement sinter filter 19 (see “6.12 Replacing Connector and Sinter Filter” on
page 50);
4. Fit replacement sealing ring 14 (see “6.2 Replacing sealing ring in connector” on
page 20)
5. Perform a final check (see “3. Final Check” on
page 11). If you find any faults, rectify them.
6. Replace service plate 13 with a new one with
the new data from “8. Replacement Parts” on
page 64.
4.3 Intervals and scope for OXYWAY Click flowmeter
Check metering accuracy by having a check measurement performed every 4 years.
We recommend that repair work be performed
only by the manufacturer, WEINMANN, or by
trained qualified experts expressly authorized by
WEINMANN.
16Servicing
Page 17
4.4 Storage
If the pressure reducer is to remain unused for long periods, we recommend the following procedure:
1. Clean pressure reducer (see “Hygienic
preparation” in the Operating Instructions).
2. Store pressure reducer in a dry place.
4.5 Disposal
Do not dispose of the unit with domestic
waste. To dispose of the unit properly,
please contact a licensed, certified
electronic scrap disposal merchant. This
address is available from your Environment Officer or from your local authority.
Important:
Be sure to observe the servicing intervals for stored appliances as well. Otherwise the appliance is no longer
fit for use on removal from storage.
Servicing17
Page 18
5. Troubleshooting
5.1 Pressure reducer OXYWAY and OXYTRON
FaultCauseRemedy
Replace sealing ring
(Section 6.2, page 20).
Replace connector
(Section 6.12, page 50).
Replace sealing ring
(Section 6.2, page 20).
Replace piston
(Section 6.6, page 28 or
Section 6.7, page 30 or
Section 6.8, page 33).
Have pressure reducer repaired by
manufacturer (WEINMANN).
Replace pressure gage
(Section 6.5, page 27).
Replace oxygen outlet
(Section 6.4, page 21)
Replace index disc (Section 6.9,
page 36)
Replace index disc (Section 6.9,
page 36)
Leak at connector
(bottle connection).
Leak at oxygen outlet
(appliance connection).
Leak at relief valve.
Pressure gage does not read
“0” at zero pressure.
Outlet flow rate too low.
Damaged sealing ring.
Damaged seat for connector sealing
ring.
Damaged sealing ring.
Unacceptable pressure rise in
pressure reducer (delayed rise
effect).
Relief valve not gas-tight.
Crater in body faulty.
Pressure gage faulty.
Oxygen outlet faulty.
Index disk faulty.
Index disk incorrectly adjusted.
Outlet flow rate of OXYTRON
WM 30150 outside
tolerance of
12 l/min ± 1.2 l/min
Outlet pressure of OXYTRON
WM 30150 outside
tolerance of
1.5 bar ± 0.08 bar
Variable pressure reducer:
leak at oxygen outlet 7
> 30 ml/min
(with valve closed and
200 bar bottle pressure).
Second stage faulty
Unacceptable pressure rise in
pressure reducer.
Replace complete piston
(Section 6.16, page 56)
Replace piston
(Section 6.6, page 28 or
Section 6.7, page 30 or
Section 6.8, page 33)
or
have pressure reducer repaired by
manufacturer (WEINMANN).
18Troubleshooting
Page 19
5.2 Flowmeter OXYWAY Click
FaultCauseRemedy
Unusually high oxygen
consumption
Inadequate O2 supply during
inhalation
Flowmeter not working
Leak in systemFind and eliminate leak
Leak in systemFind and eliminate leak
Defect or internal soiling in
flowmeter
Flowmeter defectiveHave flowmeter repaired
Oxygen source defective or emptyCheck oxygen source
Have flowmeter repaired
Troubleshooting19
Page 20
6. Repair Information and Repair Instructions for
pressure reducer OXYWAY and OXYTRON
6.1 General
•Observe the safety information on page 4 of the
Operating Instructions for OXYWAY.
•Caution: Risk of explosion!
To prevent explosion risks, make sure your
hands, tools and workplace are absolutely
free from oil and grease during the repair
work.
For this reason you should always wash your
hands before starting work.
•Any operations on the appliance presuppose
detailed knowledge and observance of the
Operating Instructions and the Service and
Repair Instructions.
•You should only carry out repairs described in
these Service and Repair Instructions. Otherwise proper functioning of OXYWAY cannot
be guaranteed.
•Be sure to perform a final check after every
•If you replace components or individual parts,
•If there is a fault in the body or the relief valve,
•Certain parts must be screwed up using a
•Note:
6.2 Replacing sealing ring in connector
repair (see “3. Final Check” on page 11).
always use genuine WEINMANN parts only.
have OXYWAY repaired by the manufacturer
(WEINMANN).
torque wrench. To ensure compliance with the
specified torque, check your torque wrench
regularly for torque compliance (control of inspection, measuring and test equipment).
The item numbers in the following text are the
same as the item numbers in the Replacement
Parts List on page 64 and the overview on
page 4.
Tools required:
•From tool set WM 15366:
– Watchmaker’s screwdriver WM 3004130041 (cleaned with methylated spirit).
1. Using a cleaned screwdriver, ease sealing
ring 14 carefully out of its groove.
Be careful not to damage the groove in any way.
2. If necessary, clean the groove with a dry cloth
or a cloth moistened with clean water.
3. Carefully press a new sealing ring into the
groove. Do not use any tools for this.
4. Perform a final check (see “3. Final Check” on
page 11).
14
20Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 21
6.3 Replacing filter screw
Tools required:
•From tool set WM 15366:
– Allen wrench 4 mm WM 30042 (cleaned with methylated spirit).
1. If necessary, clean out the hexagonal socket of
the filter screw 12 with a small screwdriver.
2. Loosen the filter screw
12, but do not unscrew
it completely yet.
3. So that no dirt can fall into the pressure reducer,
hold it with the connector pointing down. In this
position, remove the filter screw
12.
4. Screw a new filter screw 12 into the connector
stub and tighten.
5. Perform a final check. Enter the change of filter
screw with date in the service record.
6.4 Replacing oxygen outlet
12
6.4.1 Replacing G 3/8 pressure outlet
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035,
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit);
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
•Vise;
•Sealing ring
•Loctite 245 WM 14920 (never use other Loctite products).
•Torque wrench 25 Nm ± 1 Nm,
Note:
The oxygen outlet must always be replaced as a complete unit, because the flow rate cannot be guaranteed if
individual parts are replaced.
1. Clamp pressure reducer firmly in counter tool.
To do this, perform the following steps:
– Using a cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Be careful not to damage the groove at all.
– Clamp bottom part of counter tool firmly in
a vise.
– Place the top part of the counter tool on the
bottom part of the counter tool.
– Screw the connection nut onto the bottom
part of the counter tool.
– Tighten the connection nut using the top part
of the counter tool.
14 (WM 1145);
7
2
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON21
Page 22
2. Using cleaned screwdriver bit, unscrew defective oxygen outlet 7 by turning to left.
3. Fit new oxygen outlet 7:
– Remove any adhesive residues from screw
threads. Use a brass wire brush for the external thread and if necessary a tap G 1/
8 for the internal thread.
Caution:
When cleaning threads, no dirt must enter
body of pressure reducer.
– Take a new oxygen outlet .
Caution:
The oxygen outlet must be clean and free
from oil and grease.
– Apply a little Loctite 245 to screw thread
with the exception of the first two turns.
The first two turns must be kept free of Loctite
to ensure that no Loctite enters the body of the
pressure reducer.
– Screw oxygen outlet 7 by hand into pres-
sure reducer.
– Tighten oxygen outlet 7 with 9 mm screw-
driver bit and a torque wrench
(25 Nm ± 1 Nm).
7
2
7
4. Release pressure reducer from counter tool.
5. If necessary, clean groove in connector
2 with
dry cloth or cloth moistened with clean water.
6. Now press
new sealing ring 14 carefully into
groove. Do not use any tools.
7. Perform a final check
(see “3. Final Check” on page 11).
22Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 23
2. Screw the UNF counter tool completely onto
the flow outlet of the pressure reducer.
3. Clamp the pressure reducer in a vise so that
only counter tool WM 14224 is actually in the
vise.
4. To clamp the defective flow outlet firmly in the
counter tool, turn locknut WM 14223 clockwise with a 22 mm open-end wrench.
5. To release the pressure reducer from the flow
outlet, turn the pressure reducer anticlockwise
by hand.
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON23
Page 24
6. Take the pressure reducer off the flow outlet.
7. To release the defective flow outlet in the
counter tool, turn locknut WM 14223 anticlockwise with the 22 mm open-end wrench.
8. Unscrew the defective flow outlet from the locknut.
Assembling new flow outlet
Caution:
When cleaning threads, no dirt must enter body
of pressure reducer.
1. Take a new oxygen outlet.
Caution:
The oxygen outlet
from oil and grease.
35 must be clean and free
2. Screw the new flow outlet fully into the counter
tool.
3. Wet about 2 threads of the new flow outlet
with Loctite 245 (WM 14920).
The first two turns must be kept free of Loctite to
ensure that no Loctite enters the body of the pressure reducer.
4. Clamp the counter tool in a vise so that only
counter tool WM 14224 is actually in the
vise.
5. To clamp the new flow outlet firmly in the
counter tool, turn the counter tool clockwise
with the 22 mm open-end wrench.
35
24Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 25
6. Screw the pressure reducer onto the new flow
outlet hand-tight.
7. Take the pressure reducer with the counter tool
screwed onto it out of the vise.
8. Clamp counter tool WM 30035 in the vise.
9. Screw the pressure reducer and oxygen inlet
hand-tight to clamped counter tool WM
30035.
10. Put a torque wrench with open-end wrench insert SW 22 on counter tool WM 14224.
11. Tighten counter tool WM 14224 and thus the
new flow outlet on the pressure reducer
(torque: 25 Nm ±1 Nm).
12. Release the pressure reducer from counter tool
WM 30035.
13. Take counter tool WM 30035 out of the vise.
14. Clamp the pressure reducer in the vise so that
only counter tool WM 14224 is actually in the
vise.
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON25
Page 26
15. To release locknut WM 14223 from the flow
outlet, turn it anticlockwise with the 22 mm
open-end wrench.
16. Unscrew the pressure reducer and flow outlet
from locknut WM 14223.
17. Take the counter tool out of the vise.
18. Perform a final check
(see “3. Final Check” on page 11).
26Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 27
6.5 Replacing gage
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035,
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
•Vise;
•Sealing ring 14 (WM 1145);
•Packing washer 0.5 mm 17 (WM 1145/35);
•Packing washer 0.8 mm
•Torque wrench 23 Nm – 1 Nm,
•Torque wrench 28 Nm + 2 Nm,
•Open-end wrench SW 12 or SW 14;
1. Clamp pressure reducer firmly in counter tool.
To do this, perform the following steps:
– Using a
cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Be careful not to damage the groove at all.
– Clamp bottom part of counter tool firmly in
a vise.
– Place the top part of the counter tool on the
bottom part of the counter tool.
– Screw the connection nut onto the bottom
part of the counter tool.
– Tighten the connection nut using the top part
of the counter tool.
18 (WM 1145/36).
2
2. Loosen defective gage 1 or 8 with an openend wrench, then unscrew it by hand.
3. Carefully remove two packing washers 17 and
18 from threaded hole.
Take care not to damage sealing surface in body
of pressure reducer.
Ensure when taking out the packing washers
that no particles of dirt get into the bore of the
inside of the pressure reducer housing.
To guarantee the function of the pressure reducer, we recommend checking the inside of
the pressure reducer housing for any contamination before the final check. To do so, proceed in accordance with the relevant section
“Replacing piston on pressure reducer” (without changing the piston).
If you cannot remove the packing washers, do
not insert an additional packing washer
, but
send the pressure reducer to WEINMANN for
repair.
1 or 81718
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON27
Page 28
4. Of the two new packing washers, first insert
only the thicker 0.8 mm packing washer 18 in
the threaded hole.
5. Place open-end wrench insert in torque
wrench (23 Nm – 1 Nm).
6. Screw gage into pressure reducer by hand, then
tighten firmly with torque wrench.
7. Check position of gage. If not correct,
– unscrew gage again,
– insert second packing washer 17 (0.5 or
0.8 mm) in threaded hole and
– tighten gage with torque wrench
(23 Nm – 1 Nm).
8. If position of gage in relation to body is still not
correct, turn gage to correct position using a
maximum torque of 28 Nm + 2 Nm.
9. Release pressure reducer from counter tool.
10. If necessary, clean groove in connector 2 with
dry cloth or cloth moistened with clean water.
17 (Do not use unless position of fitted gage is
incorrect. See step 7.)
11. Carefully press new sealing ring 14 into
groove. Do not use any tools.
12. Perform final check
(see “3. Final Check” on page 11).
1. Turn adjuster knob 9 to left until it meets the
stop.
2. Now remove cap 30 from adjuster knob by
sliding knife blade under cap and levering
cap off.
3. Hold adjuster knob tightly and unscrew nut 32
with a socket wrench.
4. Pull adjuster knob 9 off pressure reducer.
Take care not to lose driving plate 31 on inside of
adjuster knob.
309
32931
30Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 31
5. Gently clamp cover clamping device in a vise.
6. Insert pressure reducer in cover clamping device.
7. Now clamp cover clamping device firmly in
vise. This presses the two halves of the cover
clamping device together and holds the pressure reducer in position.
8. Place counter crown over body. “A” on
counter crown must face toward connector 2.
9. Turn body to left to loosen it, but do not unscrew
completely yet
.
10. Undo vise and remove pressure reducer.
11. Carefully unscrew cover by hand.
Caution!
The piston spring is under strong compression.
Cover and spring could fly out at you.
12. Remove any adjustment washers inserted by
the factory. Be sure to note their positions.
13. Pull faulty piston 16 out of cover.
14. Remove small spring from piston.
15. Blow out body and cover with oxygen.
Caution!
Make sure the oxygen you use for blowing out is
directed away from you in order to prevent any
accumulation of oxygen in your clothing.
Counter crown
2
2116WasherBody
Small spring
Cover
When reassembling as described below, make
sure that all parts are clean and free from oil and
grease
.
16. Put the small spring in the piston.
17. Push the piston carefully into the cover.
Important!
Take care not to damage the O-rings.
18. Place body 4 in cover fitting device.
19. If adjustment washers were fitted, place them
in body taking care to ensure correct
positions.
20. Insert spring 21 in body.
21. Carefully place cover on spring.
Small spring16Cover
O-rings
Washer
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON31
Page 32
22. Use cover fitting device to press cover down
onto body.
23. Maintaining compression, screw cover by
hand onto body (two to three turns is sufficient).
24. Remove pressure reducer from cover fitting
device.
25. Tighten cover by hand.
26. Clamp pressure reducer in cover clamping
device.
27. Place counter crown over body 4. “A” on
counter crown must face connector 2.
28. Screw body firmly to cover using a torque
wrench (8 Nm + 2 Nm).
29. Undo vise and remove pressure reducer.
Cover
21
4
Counter crown
2
30. Push driver plate
31 onto square part of
spindle 28.
31. Now push adjuster knob onto spindle
9 so that
driver plate 31 engages in recess in adjuster
knob.
32. Screw adjuster knob on firmly with new nut 32.
33. Push cap 30 onto adjuster knob.
34. Perform final check
(see “3. Final Check” on page 11).
32931
30
28
32Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 33
6.8 Replacing piston in pressure reducer FAST
(WM 30600, 30800, 30850)
Tools and material required:
•From tool set WM 15366:
– Cover clamping device WM 30036,
– Cover fitting device WM 30037,
– Counter crown WM 30038,
– Screwdriver 3 mm WM 30043.;
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit),
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit),
•Vise;
•Torque wrench 8 Nm + 2 Nm,
•Open-end wrench insert SW 12;
•Flat-bladed knife;
•Oxygen lubricant WM 14934;
•plus for repairing pressure reducer WM 30850:
– Counter tool with clamping handle WM 30035 from tool set WM 15366,
•Torque wrench 25 Nm ± 1 Nm,
•Thread tap G 1/8,
•Brass wire brush,
•Sealing ring 14 (WM 1145);
•Loctite 245 WM 14920 (Never use a different Loctite product).
1. Turn index head 6 to position “0”.
2. Remove cap 51 from index head by sliding
knife blade under cap and levering off.
3. Hold index head tightly and unscrew the two
countersunk screws 52.
Take care not to change position of index head.
Otherwise you cannot be sure that the actual
flow values after reassembly will agree with
those shown on the index head.
4. Lever off loose part of index head 6.
5. Pull long part 50 of index head off body.
rear
oxygen
outlet
front
oxygen outlet
index head
front
oxygen
outlet
516
52650
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON33
Page 34
6. Gently clamp cover clamping device in a vise.
7. Slide the pressure reducer into the cover
clamping device so that the front oxygen outlet
fits into the opening of the cover clamping
device.
Please note!
Slide the pressure reducer into the cover clamping device until you clamp on to the cover only.
8. Now clamp cover clamping device firmly in
vise. This presses two halves of clamping
device together and holds pressure reducer in
position.
9. Place counter crown over body. “A” on
counter crown must face connector 2.
10. Now turn body to left to loosen it, but do not
completely unscrew it yet
.
11. Undo vise and remove pressure reducer.
12. Carefully unscrew cover by hand.
Caution!
The piston spring is under strong compression.
Cover and spring could fly out at you.
13. Remove spring 21 and any adjustment washers from body. Be sure to note their positions.
14. Pull faulty piston 16 out of cover.
15. Blow out body and cover with oxygen.
Counter crown
2
2116
Cover
Body
Caution!
When blowing out with oxygen, take care that
the oxygen is directed away from you to prevent
any accumulation of oxygen in your clothing.
Cover assembly should not be loosened (dismantled) any farther, as this would cause changes in
the relevant flow rates and the alignment of the
two oxygen outlets in relation to each other.
When reassembling as described below, make
sure all parts are clean and free from oil and grease.
16. Carefully insert piston in cover.
Important!
Take care not to damage O-rings.
O-rings
16
Cover
Index heac
Cover
Front
oxygen outlet
index head
34Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 35
17. Place body in cover fitting device.
18. If adjustment washers were fitted, place them
in body taking care to ensure correct positions.
19. Insert spring 21 in cover.
20. Carefully place cover onto body.
21. Use cover fitting device to press cover down
onto body.
22. Maintaining pressure, screw cover onto body
by hand (two to three turns are sufficient).
23. Remove pressure reducer from cover fitting
device.
24. Screw on cover hand tight.
25. Clamp pressure reducer in cover clamping
device.
26. Place counter crown over body. “A” on
counter crown must face connector 2.
27. Screw body firmly to cover using a torque
wrench (8 Nm + 2 Nm).
Caution!
The two oxygen outlets must be aligned.
Washer
Cover
21
4
Counter
crown
2
28. Undo vise and remove pressure reducer.
29. Push long part 50 of index head onto body.
30. Place index head 6 on pressure reducer.
Take care to fit it in correct position. Triangular
marker on long part of index head must then
point to “0”.
31. Screw on index head with two countersunk
screws 52.
32. Push cap 51 onto index head.
33. Perform final check
(see “3. Final Check” on page 11).
Front
oxygen outlet
Cover
Index head
52
51650
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON35
Page 36
6.9 Replacing piston in pressure reducer Fast II High Flow
(WM 31899)
Tools and material required:
•From tool set WM 15366:
– Cover clamping device WM 30036,
– Cover fitting device WM 30037,
– Counter crown WM 30038,
– Screwdriver 3 mm WM 30043.;
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit),
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit),
•Vise;
•Torque wrench 8 Nm + 2 Nm,
•Open-end wrench insert SW 12;
•Flat-bladed knife;
•Oxygen lubricant WM 14934;
•plus for repairing pressure reducer WM 31899:
– Counter tool with clamping handle WM 30035 from tool set WM 15366,
•Torque wrench 25 Nm ± 1 Nm,
•Thread tap G 1/8,
•Brass wire brush,
•Sealing ring 14 (WM 1145);
•Loctite 245 WM 14920 (Never use a different Loctite product).
1. Turn index head 6 to position “0”.
2. Remove cap 51 from index head by sliding
knife blade under cap and levering off.
3. Hold index head tightly and unscrew the two
countersunk screws 52.
Take care not to change position of index head.
Otherwise you cannot be sure that the actual
flow values after reassembly will agree with
those shown on the index head.
4. Lever off loose part of index head 6.
5. Pull long part 50 of index head off body.
rear
oxygen
outlet
front
oxygen outlet
index head
front
oxygen
outlet
516
52650
36Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 37
6. Gently clamp cover clamping device in a vise.
7. Slide the pressure reducer into the cover
clamping device so that the front oxygen outlet
fits into the opening of the cover clamping device.
Please note!
Slide the pressure reducer into the cover clamping device until you clamp on to the cover only.
8. Now clamp cover clamping device firmly in
vise. This presses two halves of clamping
device together and holds pressure reducer in
position.
9. Place counter crown over body. “A” on
counter crown must face connector 2.
10. Now turn body to left to loosen it, but do not
completely unscrew it yet
.
11. Undo vise and remove pressure reducer.
12. Carefully unscrew cover by hand.
Caution!
The piston spring is under strong compression.
Cover and spring could fly out at you.
13. Remove spring 21 and any adjustment washers from body. Be sure to note their positions.
14. Pull faulty piston 16 out of cover.
15. Blow out body and cover with oxygen.
Caution!
When blowing out with oxygen, take care that
the oxygen is directed away from you to prevent
any accumulation of oxygen in your clothing.
Caution!
Cover assembly should not be loosened (dismantled) any farther, as this would cause changes in
the relevant flow rates and the alignment of the
two oxygen outlets in relation to each other.
2
Body
Counter crown
2116
Cover
When reassembling as described below, make
sure all parts are
clean and free from oil and grease.
16. Carefully insert piston in cover.
Important!
Take care not to damage O-rings.
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON37
O-rings
16
Cover
Index heac
Cover
Front
oxygen outlet
index head
Page 38
17. Place body in cover fitting device.
18. If adjustment washers were fitted, place them
in body taking care to ensure correct positions.
19. Insert spring 21 in body.
20. Carefully place cover on spring.
21. Use cover fitting device to press cover down
onto body.
22. Maintaining pressure, screw cover onto body
by hand (two to three turns are sufficient).
23. Remove pressure reducer from cover fitting
device.
24. Screw on cover hand tight.
25. Clamp pressure reducer in cover clamping
device.
26. Place counter crown over body. “A” on
counter crown must face connector 2.
27. Screw body firmly to cover using a torque
wrench (8 Nm + 2 Nm).
Washer
Cover
21
4
Counter
crown
2
28. Undo vise and remove pressure reducer.
29. Push long part 50 of index head onto body.
30. Place index head 6 on pressure reducer.
Take care to fit it in correct position. Triangular
marker on long part of index head must then
point to “0”.
31. Screw on index head with two countersunk
screws 52.
32. Push cap 51 onto index head.
33. Perform final check
(see “3. Final Check” on page 11).
front
oxygen outlet
index headcap
52650
38Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
G 3/8 WM 22827 (from 15348 or WM 15349),
– Fitting aid,
– Allen (internal hexagon) key, 2 mm,
– Screwdriver 3 mm WM 30043;
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit),
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit),
•Vise;
•Torque wrench 8 Nm + 2 Nm;
•Open-end wrench insert SW 12;
•Flat-bladed knife;
•Oxygen lubricant WM 14934;
•plus for repairing pressure reducer WM 30850:
– Counter tool with clamping handle WM 30035 from tool set WM 15366,
•Torque wrench 25 Nm ± 1 Nm,
•Thread tap G 1/8,
•Brass wire brush,
•Sealing ring 14 (WM 1145);
•Loctite 245 WM 14920 (Never use a different Loctite product).
6.10.1 Dismantling adjuster knob
1. Set index head 6 to “0”.
2. Remove cap 51 from index head by sliding
knife blade under cap and levering off.
Rear
oxygen
outlet
Front
oxygen
outlet
516
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON39
Page 40
3. Hold index head firmly and unscrew the two
countersunk screws 52.
Take care not to change position of index head.
Otherwise you cannot be sure that the actual
flow values after reassembly will agree with
those shown on the index head.
Front
oxygen outlet
4. Lift off loose part of index head 6.
5. Pull long part 50 of index head off body.
6.10.2 Removing cover with index head from body
1. Gently clamp cover clamping device in a vise.
2. Insert pressure reducer in cover clamping
device so that front oxygen outlet fits into
opening in cover clamping device.
Caution!
Push pressure reducer into cover clamping device
only just far enough to clamp the cover itself.
3. Now clamp cover clamping device firmly in
vise. This presses the two halves of the cover
clamping device together and holds the
pressure reducer in position.
4. Place counter crown over body. “A” on
counter crown must face towards connector 2.
5. Turn body to left to loosen it, but do not unscrew
completely yet
.
6. Open vise and remove pressure reducer.
7. Carefully unscrew cover from index head.
Caution!
The piston spring is under strong compression.
Cover and spring could fly out at you.
8. Remove spring 21 and any adjustment
washers from body. Be sure to note their
positions.
A
2
Index head
2116
52650
Counter crown
Front oxygen
outlet
Body
40Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Cover
22
Page 41
6.10.3 Removing index head from cover
Caution!
Under the index disk are balls and springs which
are under compression. Perform the following
steps carefully, otherwise the springs and balls
could fly out at you.
1. Loosen grub screw 45 of index disk 43.
2. Carefully remove index disc 43, slide ring 44,
springs 42 and balls 41 from index head.
3. Push the locator for the orifice disk completely
out of the index head.
4. Clamp the cover about 20 mm down in the
cover clamping device.
5. Screw the special counter tool onto the oxygen
outlet and use the special pliers to remove the
union nut of the index head from the cover.
45
44
42
37
Cover
clamping
device
WM 30047
Special pliers
WM 30046
45
43
41
22
43
22
Counter tool
6.10.4 Assembling index head and cover
1. Take a new index head or cover, as
necessary.
2. Screw cover hand tight onto body.
3. Clamp pressure reducer in cover clamping
device.
4. Place counter crown over body. “A” on
counter crown must face towards connector.
5. Screw body onto cover using a torque wrench
(8 Nm +2 Nm).
22
A
Counter
crown
2
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON41
Page 42
6.10.5 Aligning oxygen outlet
1. Screw the empty index head onto the bottom
part of the housing.
2. Schrauben Sie die Anschlussmutter auf das
Konterwerkzeug.
3. Befestigen Sie das Konterwerkzeug an dem
Sauerstoffausgang.
4. Use the rubber-jawed pliers to slacken the
screwed connection on the index head.
5. Now use the special counter tool to turn the oxygen outlet into its final position. It should now
no longer be possible to turn the index head.
6. Both oxygen outlets are in line.
7. Release the special counter tools.
8. Unscrew the connection nut from the counter
tool.
9. Clamp the pressure reducer in the cover
clamping device.
10. Clamp the counter crown over the housing.
The A on the counter crown must point towards
the connection bolt.
11. Now release the housing by turning it
anticlockwise.
Counter tool
A
2
Special pliers
WM 30046
37
22
Counter
crown
6.10.6 Inserting orifice disk and fitting index disk
1. Place orifice holder 38 complete with orifice
disk 39 and slide disk 40 onto fitting aid and
push cover with index head onto it as far as
possible.
2. Remove pre-assembled cover from fitting aid.
40
39
38
42Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Cover
clamping
device
Page 43
3. Align hole on inside of cover with hole in notch
pin.
4. Put the preassembled housing back on the
fitting aid. There must be no change in the
position of orifice holder in relation to cover.
hole on inside
of cover
hole in orifice disk
hole in notch pin
of orifice holder
5. Place slide disc 44 over projecting spindle of
orifice holder.
6. Insert springs 42 in corresponding holes and
place balls 41 on springs 42.
7. Place index disk 43 on balls 41 so that thread
of fixing screw is opposite oxygen outlet 35.
8. Use fitting aid to press index disk onto index
head. Secure index head with fixing screw.
9. Take assembled unit from fitting aid and check
position of holes in cover and notch pin.
44
45
37
36
43
41
42
35
22
40
39
38
6.10.7 Fitting index head with cover on body
1. Blow out body and cover with oxygen.
Caution!
When blowing out with oxygen, take care that
the oxygen is directed away from you to prevent
any accumulation of oxygen in your clothing.
When reassembling as described below, make
absolutely sure that all parts are clean and free from
oil and grease
2. Push the piston carefully into the cover.
Important
Take care not to damage O-rings.
3. Place body in cover fitting device.
4. If any adjustment washers were present, place
them in the body. Take care to fit in correct
position.
5. Insert spring
.
21 into cover.
O-rings
16
Cover
Index head
Cover
Front oxygen
outlet
Index head
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON43
Page 44
6. Carefully place cover onto body.
7. Now use cover fitting device to press cover
down onto body.
8. Maintaining the pressure, screw cover onto
body by hand (two to three turns is sufficient).
9. Remove pressure reducer from cover fitting
device.
10. Screw cover on firmly by hand.
11. Clamp pressure reducer in cover clamping
device.
12. Place counter crown over body. “A” on
counter crown must face towards connector 2.
13. Screw body onto cover using a torque wrench
(8 Nm + 2 Nm).
Caution!
The two oxygen outlets must be aligned.
14. Undo vise and remove pressure reducer.
Cover
21
4
Counter crown
2
6.10.8 Fitting adjuster knob
1. Push long part 50 of index head onto body.
2. Place index head
Take care to fit in correct position. Triangular
marker on long part of index head must then
point to “0”.
3. Screw on index head with two countersunk
screws 52.
4. Push cap
51 onto index head.
5. Perform final check.
(see “3. Final Check” on page 11).
6 on pressure reducer.
Front
oxygen outlet
Cover
Index head
52
51650
44Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 45
6.11 Replacing index head/cover on Fast II High Flow pressure
reducer
•From tool set WM 15366:
– Cover clamping device WM 30036,
– Cover fitting device, WM 30037,
– Counter crown WM 30038,
– Special pliers for WM 30046,
– Counter tool UNF (from WM 14223 and WM 14224),
– Fitting aid,
– Allen (internal hexagon) key, 2 mm,
– Screwdriver 3 mm WM 30043;
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit),
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit),
•Vise;
•Torque wrench 8 Nm + 2 Nm;
•Open-end wrench insert SW 12;
•Flat-bladed knife;
•Oxygen lubricant WM 14934;
•plus for repairing pressure reducer WM 30850:
– Counter tool with clamping handle WM 30035 from tool set WM 15366,
•Torque wrench 25 Nm ± 1 Nm,
•Thread tap G 1/8,
•Brass wire brush,
•Sealing ring 14 (WM 1145);
•Loctite 245 WM 14920 (Never use a different Loctite product).
6.11.1 Dismantling adjuster knob
1. Set index head 6 to “0”.
2. Remove cap 51 from index head by sliding
knife blade under cap and levering off.
rear
oxygen
outlet
front
oxygen
outlet
516
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON45
Page 46
3. Hold index head firmly and unscrew the two
countersunk screws 52.
Take care not to change position of index head.
Otherwise you cannot be sure that the actual
flow values after reassembly will agree with
those shown on the index head.
front
oxygen outlet
4. Lift off loose part of index head 6.
5. Pull long part 50 of index head off body.
6.11.2 Removing cover with index head from body
1. Gently clamp cover clamping device in a vise.
2. Insert pressure reducer in cover clamping
device so that front oxygen outlet fits into
opening in cover clamping device.
Caution!
Push pressure reducer into cover clamping device
only just far enough to clamp the cover itself.
3. Now clamp cover clamping device firmly in
vise. This presses the two halves of the cover
clamping device together and holds the
pressure reducer in position.
4. Place counter crown over body 22. “A” on
counter crown must face towards connector 2.
5. Turn body 22 to left to loosen it, but do not
unscrew completely yet
.
6. Undo vise and remove pressure reducer.
7. Carefully unscrew cover from index head.
Caution!
The piston spring is under strong compression.
Cover and spring could fly out at you.
8. Remove spring 21 and any adjustment
washers from body. Be sure to note their
positions.
A
2
Body
52650
index headcap
Counter crown
22
2116
Cover
6.11.3 Removing index head from cover
Caution!
Under the index disk are balls and springs which
are under compression. Perform the following
steps carefully, otherwise the springs and balls
could fly out at you.
46Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
45
44
42
37
43
41
22
Page 47
1. Loosen grub screw 45 of index disk 43.
2. Carefully remove index disc 43, slide ring 44,
springs 42 and balls 41 from index head.
3. Push orifice disk holder completely out of index
head.
4. Clamp cover about 20 mm deep in cover
clamping device.
Cover
clamping
device
WM 30047
45
43
22
5. Screw counter tool onto oxygen outlet and use
special pliers to remove union nut of index
Special pliers
WM 30046
head from cover.
6.11.4 Assembling and aligning index head and cover
1. Take a new index head or cover, as
necessary.
2. Screw cover hand tight onto body.
3. Clamp pressure reducer in cover clamping
device.
4. Place counter crown over body. “A” on
counter crown must face towards connector.
5. Screw body onto cover using a torque wrench
(8 Nm +2 Nm).
Special pliers
WM 30046
Counter tool UNF
37
22
Counter tool UNF
37
22
6. Place index head on cover so that oxygen
outlet is about 45 degrees before its end
position. Tighten union nut of index head by
hand.
7. Screw union nut of index head against cover
until you meet appreciable resistance.
8. Screw counter tool onto oxygen outlet.
9. Now use counter tool to turn oxygen outlet to
its end position. It should no longer be
possible to turn the index head out of position.
10. Remove counter tool from oxygen outlet.
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON47
Page 48
6.11.5 Inserting orifice disk and fitting index disk
1. Place orifice holder 38 complete with orifice
disk and notch pin onto fitting aid and push
cover with index head onto it as far as
possible.
2. Remove pre-assembled cover from fitting aid.
orifice disk
38
Cover
clamping
device
3. Align hole on inside of cover with hole in notch
pin.
4. Put the preassembled housing back on the
fitting aid. There must be no change in the
position of orifice holder in relation to cover.
5. Place slide disc 44 over projecting spindle of
orifice holder.
6. Insert springs 42 in corresponding holes and
place balls 41 on springs 42.
7. Place index disk
43 on balls 41 so that thread
of fixing screw is opposite oxygen outlet 35.
8. Use fitting aid to press index disk onto index
head. Secure index head with fixing screw.
9. Take assembled unit from fitting aid and check
position of holes in cover and notch pin.
hole on inside
of cover
hole in orifice disk
hole in notch pin
of orifice holder
45
44
37
36
43
41
42
35
22
48Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
40
39
38
Page 49
6.11.6 Fitting index head with cover on body
1. Blow out body and cover with oxygen.
Caution!
When blowing out with oxygen, take care that
the oxygen is directed away from you to prevent
any accumulation of oxygen in your clothing.
When reassembling as described below, make
absolutely sure that all parts are clean and free from
oil and grease
.
2. Carefully insert piston in cover.
Important
Take care not to damage O-rings.
3. Place body 4 in cover fitting device.
4. If any adjustment washers were present, place
them in the body. Take care to fit in correct
position.
5. Insert spring 21 in body.
6. Carefully place cover on spring.
7. Now use cover fitting device to press cover
down onto body.
8. Maintaining the pressure, screw cover onto
body by hand (two to three turns is sufficient).
9. Remove pressure reducer from cover fitting
device.
10. Screw cover on firmly by hand.
O-rings
Disk
16
Cover
Index head
Cover
Front oxygen
outlet
Index head
11. Clamp pressure reducer in cover clamping
device.
12. Place counter crown over body 4. “A” on
counter crown must face towards connector 2.
13. Screw body onto cover using a torque wrench
(8 Nm + 2 Nm).
Caution!
The two oxygen outlets must be aligned.
14. Undo vise and remove pressure reducer.
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON49
Cover
21
4
Counter crown
2
22
Page 50
6.11.7 Fitting adjuster knob
1. Push long part 50 of index head onto body.
2. Place index head 6 on pressure reducer.
Take care to fit in correct position. Triangular
marker on long part of index head must then
point to “0”.
Front
oxygen outlet
3. Screw on index head with two countersunk
screws 52.
4. Push cap 51 onto index head.
5. Perform final check.
Cover
Index head
(see “3. Final Check” on page 11).
6.12 Replacing Connector and Sinter Filter
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035,
– Counter crown WM 30038,
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
•Vise;
•Torque wrench 55 Nm ± 1 Nm,
•Open-end wrench insert SW 12;
•Oxygen lubricant WM 14934;
•Sealing ring 14 (WM 1145);
•Sinter filter 19 (WM 6619);
•Filter screw 12 (WM 30905).
52
51650
1. Clamp pressure reducer firmly in counter tool.
Proceed as follows:
– Using cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Take care not to damage groove in any way.
– Clamp bottom part of counter tool in a vise.
– Place the top part of the counter tool on the
bottom part of the counter tool.
– Screw connection nut 2 onto the bottom
part of the counter tool.
– Tighten connection nut 2 using the top part
of the counter tool.
50Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
2
15
Page 51
2. Place counter crown over body 4. “A” on
counter crown must face connector 2.
3. Place torque wrench (with open-end wrench
insert) lengthwise on square end of counter
crown.
4. Loosen body by unscrewing to left, then remove completely by hand.
5. Remove connector 2 from counter tool.
6. Check connection nut 15 and new
connector 2 for defects (e.g. thread, chrome
plating, sealing surface).
Do not refit faulty parts.
7. If necessary clean groove in connector 2 with
dry cloth or cloth moistened with clean water.
8. Fasten connector 2 firmly in counter tool with
connection nut 15.
9. Blow out connector with oxygen.
Caution!
When blowing out, make sure oxygen is directed
away from you to prevent any accumulation of
oxygen in your clothing.
10. Insert a new sinter filter 19 in connector.
Do not on any account use Loctite or other thread
sealants.
19
11. Screw body onto connector by hand.
12. Place counter crown over body 4. “A” on
counter crown must face connector 2.
13. Using torque wrench (55 Nm ± 1 Nm),
tighten body securely on connector.
14. Release pressure reducer from counter tool.
15. Insert new filter screw 12 in connector (see
“6.3 Replacing filter screw” on page 21).
16. Carefully press new sealing ring 14 into
cleaned groove. Do not use tools.
17. Perform final check
(see “3. Final Check” on page 11).
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON51
Page 52
6.13 Changing connector and sinter filter in OXYTRON pressure
reducer
Tools and material required:
•From tool set WM 15366:
– Bottom part of counter tool WM 30035,
– Counter crown WM 30038,
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
– Single-end wrench SW 19 WM 30040 (cleaned with methylated spirit).
•Vise;
•Torque wrench 55 Nm ± 1 Nm,
•Torque wrench 10 Nm + 2 Nm,
•Open-end wrench insert SW 12;
•Open-end wrench SW 30;
•Oxygen lubricant WM 14934;
•Sealing ring
•Sinter filter
•Filter screw 12 (WM 30905).
1. Clamp pressure reducer firmly in counter tool.
Proceed as follows:
– Using cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Take care not to damage groove in any way.
– Clamp bottom part of counter tool firmly in
a vise.
– Insert connector
– Use open-end wrench SW 30 to tighten
connection nut
14 (WM 1145);
19 (WM 6619);
52
2
15
2 in stub of bottom part.
15.
2. Using the cleaned single-end wrench SW 19,
turn second stage 52 to left and unscrew it (see
“6.14 Replacing second stage of OXYTRON
pressure reducer (WM 30150)” on
page 54).
3. Place counter crown over body 4. “A” on
counter crown must face connector 2.
4. Place torque wrench (with open-end wrench
insert) lengthwise on square end of counter
crown.
5. Loosen body by turning to left, then remove by
hand.
6. Release connector 2 from counter tool.
7. Check connection nut 15 and new
connector 2 for defects (e.g. thread, chrome
plating, sealing surface).
Do not refit faulty parts.
8. If necessary clean groove in connector 2 with
dry cloth or cloth moistened with clean water.
52Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 53
9. Using open-end wrench SW 30, fix new
connector 2 firmly in bottom part of counter
tool with connection nut 15.
10. Blow out connector with oxygen.
Caution!
When blowing out connector, make sure that
oxygen is directed away from you to prevent any
accumulation of oxygen in your clothing.
11. Insert a new sinter filter 19 in connector.
Do not on any account use Loctite or other thread
sealants.
12. Screw body onto connector by hand.
13. Place counter crown over body 4. “A” on
counter crown must face connector 2.
14. Use torque wrench (55 Nm ± 1 Nm) to screw
body firmly onto connector.
15. Fit new second stage:
– Remove adhesive residues from threads.
Use a brass wire brush for the external
thread and, if necessary, a thread tap
G 1/8 for the internal thread.
Caution:
When cleaning threads, make sure no dirt
enters body.
– Pick up second stage 52.
Caution:
Second stage 52 must be clean and free from oil
and grease.
– Apply a little Loctite 245 to thread, except
for the first two turns.
First two turns are left free so no Loctite can get
into the body.
– Screw second stage 52 into the pressure
reducer by hand.
– Use screwdriver bit and torque wrench
(10 Nm + 2 Nm) to tighten second
stage 52.
16. Release pressure reducer from counter tool.
17. Insert a new filter screw 12 in connector (see “6.3
Replacing filter screw” on page 21).
18. Carefully press new sealing ring 14 into groove. Do
not use tools.
19
2
15
52
19. Perform final check
(see “3. Final Check” on page 11).
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON53
Page 54
6.14 Replacing second stage of OXYTRON pressure reducer
(WM 30150)
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035;
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit);
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
– Single-end wrench SW 19 WM 30040.
•Vise;
•Torque wrench 10 Nm + 2 Nm
•Open-end wrench SW 30
•Sealing ring
•Loctite 245 WM 14920 (never use a different Loctite product).
Note:
Position of set screw in outlet from second stage must not be changed, as this would change the flow rate.
1. Clamp pressure reducer firmly in counter tool.
Proceed as follows:
– Using cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Take care not to damage groove in any way.
– Clamp bottom part of counter tool firmly in
a vise.
– Insert connector 2 in stub of bottom part.
– Use open-end wrench SW 30 to screw
connection nut 15 tight.
14 (WM 1145);
52
2
15
2. Using cleaned single-end wrench, turn faulty
second stage 52 to left and unscrew.
3. Fit new second stage:
– Remove adhesive residues from screw
threads. Use screw tap G 1/8 for internal
thread.
Caution:
When cleaning screw threads, do not let any
dirt enter the body.
– Pick up a new second stage 52.
Caution:
Second stage 52 must be clean and free from
oil and grease.
– Apply a little Loctite 245 to screw thread,
except for first two turns.
First two turns must be left free of Loctite so no
Loctite can get inside body.
52
54Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 55
– Screw second stage 52 into pressure
reducer by hand.
– Using screwdriver bit 9 mm and torque
wrench (10 Nm + 2 Nm), screw second
stage 52 up tight.
4. Release pressure reducer from counter tool.
5. If necessary, clean groove in connector 2 with
dry cloth or cloth moistened with clean water.
6. Carefully press new sealing ring 14 into
groove. Do not use tools.
7. Perform final check
(see “3. Final Check” on page 11).
52
6.15 Replacing first-stage piston in OXYTRON pressure reducer
(WM 30150)
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035,
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit);
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
– Single-end wrench SW 19 WM 30040 (cleaned with methylated spirit).
•Vise;
•Torque wrench 25 Nm ± 1 Nm
•Open-end wrench SW 30
•Sealing ring 14 (WM 1145);
•Loctite 245 WM 14920 (Never use a different Loctite product).
Note:
Position of set screw in second-stage outlet must not be altered, as this would change the flow rate.
1. Unscrew second stage (see “6.14 Replacing
second stage of OXYTRON pressure reducer
(WM 30150)” on page 54).
2. Replace piston (see “6.6 Replacing piston in
pressure reducer FIX (WM 30050, 30100,
30200, 30300, 30350)” on page 28).
3. Screw second stage on again (see “6.14 Replacing second stage of OXYTRON
pressure reducer (WM 30150)” on
page 54).
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON55
Page 56
6.16 Replacing second-stage piston in OXYTRON pressure reducer
(WM 30150)
Tools and material required:
•From tool set WM 15366:
– Counter tool with clamping handle WM 30035,
– Screwdriver bit 9 mm WM 30039 (cleaned with methylated spirit);
– Watchmaker’s screwdriver WM 30041 (cleaned with methylated spirit);
– Single-end wrench SW 19 WM 30040 (cleaned with methylated spirit).
•Vise;
•Oxygen lubricant WM 14934.
•Torque wrench 10 Nm + 2 Nm
•Sealing ring
•Loctite 245 WM 14920 (Never use a different Loctite product).
Note:
Position of set screw in second-stage outlet must not be altered, as this would change the flow rate.
1. Clamp pressure reducer firmly in counter tool.
Proceed as follows:
– Using cleaned screwdriver, carefully ease
sealing ring 14 out of groove in connector.
Take care not to damage groove in any way.
– Clamp bottom part of counter tool in a vise.
– Screw up connection nut by hand.
14 (WM 1145);
2. Open cover 54 of second stage by holding
body 53 with special open-end wrench
SW 19 and loosening cover 54 with 9 mm
screwdriver bit.
3. Slowly unscrew cover 54 by hand.
Note:
Piston spring is under compression!
4. Remove spring 55.
5. Remove faulty piston 56 from cover.
Note:
When reassembling as described below, make sure
all parts are clean and absolutely free from grease.
Sliding surface of O-ring on inside of cover 54 must
not be scored or scratched.
53
55
53 54
54
56
56Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON
Page 57
6. Apply a little oxygen lubricant to O-ring and
carefully insert new piston 56 in cover.
Caution:
Do this very carefully. New O-rings must not be
damaged during insertion.
7. Place spring 55 on piston.
8. Screw cover 54 onto body 53 by hand.
9. Hold body with special open-end wrench
SW 19. Then use 9 mm screwdriver bit and
torque wrench (10 Nm + 2) to tighten cover.
10. Release pressure reducer from counter tool
with open-end wrench SW 30.
11. If necessary clean groove in connector 2 with
dry cloth or cloth moistened with clean water.
12. Carefully press new sealing ring 14 into
groove. Do not use tools.
13. Perform final check
(see “3. Final Check” on page 11).
Repair Information and Repair Instructions for pressure reducer OXYWAY and OXYTRON57
Page 58
7. Repair Information and Repair Instructions for
flowmeter OXYWAY Click
7.1 General
•Observe the safety information on page 4 of the
Operating Instructions for OXYWAY Click.
•Caution:
Spontaneous explosive reactions can occur if
compressed oxygen comes into contact with
flammable substances (grease, oil, alcohol etc.)!
To prevent explosion risks, make sure your
hands, tools and workplace are absolutely
free from oil and grease during the repair
work.
For this reason you should always wash your
hands before starting work.
•Any operations on the appliance presuppose
detailed knowledge and observance of the
Operating Instructions and the Service and
Repair Instructions.
7.2 Replacing index head / cover
•You should only carry out repairs described in
these Service and Repair Instructions. Otherwise proper functioning of OXYWAY Click
cannot be guaranteed.
•Be sure to perform a final check after every
repair (see “3. Final Check” on page 11).
•If you replace components or individual parts,
always use genuine WEINMANN parts only.
•Certain parts must be screwed up using a
torque wrench. To ensure compliance with the
specified torque, check your torque wrench
regularly for torque compliance (control of inspection, measuring and test equipment).
•Note:
The item numbers in the following text are the
same as the item numbers in the Replacement
Parts List on page 64 and the overview on
page 4.
•From tool set WM 15366:
– Cover clamping device WM 30036,
– Counter tool:
G 3/8 WM 22827 (from WM 15348 or WM 15349),
UNF (from WM 14223 or WM 14224)
– Fitting aid,
– Allen (internal hexagon) key, 2 mm,
– Screwdriver 3 mm WM 30043;
•Vise;
•Torque wrench 16 Nm + 2 Nm;
•Flat-bladed knife;
•Oxygen lubricant WM 14934;
58Repair Information and Repair Instructions for flowmeter OXYWAY Click
Page 59
7.2.1 Dismantling adjuster knob
1. Set index head 6 to “0”.
2. Remove cap 51 from index head by sliding
knife blade under cap and levering off.
3. Hold index head firmly and unscrew the two
countersunk screws 52.
Take care not to change position of index head.
Otherwise you cannot be sure that the actual
flow values after reassembly will agree with
those shown on the index head.
Oxygen outlet
4. Lift off loose part of index head 6.
5. Pull long part 50 of index head off body.
7.2.2 Removing cover with index head from body
1. Gently clamp cover clamping device in a vise.
2. Now clamp cover clamping device firmly in
vise. This presses the two halves of the cover
clamping device together and holds the
flowmeter in position.
Oxygen
outlet
Index head
516
52650
Counter tool
22
3. Now loosen body (do not unscrew completely
yet!
) by turning it anti-clockwise.
4. Undo vise and remove flowmeter.
5. Now slowly unscrew the body from the index
head.
Repair Information and Repair Instructions for flowmeter OXYWAY Click59
Oxygen outlet
Cover
Page 60
7.2.3 Removing index head
Caution!
Under the index disk are balls and springs which
are under compression. Perform the following
steps carefully, otherwise the springs and balls
could fly out at you.
1. Loosen grub screw 45 of index disk 43.
2. Carefully remove index disc 43, slide ring 44,
springs 42 and balls 41 from index head.
3. Push orifice disk holder completely out of index
head.
4. Clamp cover about 20 mm deep in cover
clamping device.
5. Screw counter tool onto oxygen outlet and use
special pliers to remove union nut of index
head from cover.
Cover
clamping
device
WM 30047
45
44
42
37
43
41
36
22
43
22
Counter tool
22
7.2.4 Assembling index head and cover
1. Take a new index head or cover, as
necessary.
2. Tighten up the body hand-tight on the cover for
the central gas supply system.
3. Clamp flowmeter in cover clamping device.
4. Screw the body into the central gas supply
cover using a torque wrench (8 Nm +2 Nm).
5. Place index head on cover so that oxygen
outlet is about 45 degrees before its end
position. Tighten union nut of index head by
hand.
Counter tool
22
60Repair Information and Repair Instructions for flowmeter OXYWAY Click
Page 61
6. Screw union nut of index head against cover
until you meet appreciable resistance.
7. Screw special counter tool onto oxygen outlet.
8. Now use special counter tool to turn oxygen
outlet to its end position. It should no longer be
possible to turn the index head out of position.
9. Remove special counter tool from oxygen
outlet.
7.2.5 Inserting orifice disk and fitting index disk
1. Use cover clamping device to remove index
head with cover from body.
2. Place orifice holder complete with orifice disk
and notch pin onto fitting aid and push cover
with index head onto it as far as possible.
3. Remove pre-assembled cover from fitting aid.
Counter tool
Special pliers
22
38
4. Align hole on inside of cover with hole in notch
pin.
5. Place pre-assembled cover on fitting aid
again. There must be no change in the
position of orifice holder in relation to cover.
6. Place slide ring over projecting spindle of
orifice holder.
7. Insert springs in corresponding holes and
place balls on springs.
Repair Information and Repair Instructions for flowmeter OXYWAY Click61
Cover
clamping
device
hole on inside
of cover
hole in orifice disk
hole in notch pin
of orifice holder
Page 62
8. Place index disk on balls so that thread of
fixing screw is opposite oxygen outlet.
9. Use fitting aid to press index disk onto index
head. Secure index head with fixing screw.
10. Take assembled unit from fitting aid and check
position of holes in cover and to notch pin.
44
45
37
36
43
41
42
35
22
40
39
38
7.2.6 Assembling index head and housing
1. Clamp the cover clamping device gently in a
vise.
2. Now clamp the cover clamping device in the
vise. This presses the two halves of the clamping device firmly together and holds the cover
of the flowmeter in position.
3. Screw the housing on the central gas supply
system cover tight.
7.2.7 Fitting adjuster knob
1. Push long part 50 of index head onto body.
2. Place index head
Take care to fit in correct position. Triangular
marker on long part of index head must then
point to “0”.
3. Screw on index head with two countersunk
screws 52.
4. Push cap
51 onto index head.
5. Perform final check.
(see “3. Final Check” on page 11).
6 on pressure reducer.
Special pliers
Oxygen outlet
Cover
Counter tool
22
52650
51
Index head
62Repair Information and Repair Instructions for flowmeter OXYWAY Click
Page 63
7.3 Changing filter screw
Tools required:
•From tool set WM 15366:
– Allen wrench 4 mm WM 30042 (cleaned with methylated spirit).
1. Release the index head and the body of the
flowmeter from the central gas supply system
cover as described in “7.2 Replacing index
head / cover” on page 58.
The filter screw is located on the inside of the central gas supply system cover (see drawing).
2. If necessary, clean the hexagon socket of filter
screw 12 using a small screwdriver.
3. Loosen filter screw
yet.
4. So that no dirt can fall into the flowmeter, hold
the
central gas supply system cover with the con-
nector pointing down. In this position, unscrew
filter screw 12 completely.
5. Screw a new filter screw 12 firmly into the cen-
tral gas supply system cover
12, but do not unscrew it
6. Refit the index head and cover as described in
“7.2 Replacing index head / cover” on
page 58.
7. Perform a final check. Enter the filter screw
change and date in the service record.
.
central gas
supply system
cover
12
7.4 Replace central gas supply system plug connection
Tools required:
•Cover clamping device WM 30036,
•Open-end wrench SW 17
•Loctite WM 14901
1. Undo the index head as described in “7.2 Replacing index head / cover” on page 58.
2. Clamp the flowmeter in the cover clamp so
that only the cover is clamped.
3. Release the old plug connection and take it off
the body.
4. Remove the old seal and put in a new one (unless it is a Swedish AGA connector).
5. Wet the thread (leave first two turns) of the
plug connector with a little Loctite 601.
6. Insert the new plug connector in the body and
tighten it up with an open-ended wrench.
Ensure that the plug connector is correctly lined up.
Two of the six sides (see drawing) must be flush with
the O
outlet. This ensures that the O2 outlet of the
2
flowmeter points vertically downwards when it is
connected to the central gas supply system.
7. Perform a final check (see “3. Final Check” on
page 11).
Repair Information and Repair Instructions for flowmeter OXYWAY Click63
Page 64
8. Replacement Parts
Note:
The item numbers in the following tables are the same as the numbers in the text of these Service and Repair
Instructions.
8.1 Information about conformity with standards
Up to serial number 0849999, OXYWAY
pressure reducers meet standard EN 738-1.
From serial number 0850000, OXYWAY
pressure reducers meet standard EN ISO 105241:2006.
In the course of adapting the products to standard
EN ISO 10524-1:2006, it was not possible to
design all components to be downward-compatible. Parts which are not downward-compatible are
available for both device statuses and are marked
accordingly in the replacement parts lists (see “8.
Ersatzteile” on page 69) of these instructions.
Caution!
Malfunction as a result of unclear device status.
If replacement parts which are not downward-compatible are used in the wrong pressure reducers,
connected devices will malfunction.
•Fit replacement parts which are not downward-compatible only in accordance with the
pressure reducer serial numbers quoted.
•Do not convert pressure reducers up to
SN 0849999 “to suit the new standard”.
64Replacement Parts
Page 65
8.2 Replacement parts list for all pressure reducers
1
17
18
13 Plate, service
19
15
14
12
20
Plate, filter screw change
16
Item No.DescriptionOrder No.
Maintenance kitWM 15458
1Contents gage 0 – 315 barWM 30914
12Filter screwWM 30905
13Plate, service, 4 yearsWM 75341
14Sealing ring, O-ring 12.3 x 2.4WM1145
Connection nut for all connectors except WM 30150 (DE) (see
15
“8.10 Additional replacement parts 2nd stage OXYTRON
WM 30150” on page 73 and “8.11 Connections for pressure
WM 1144/16
reducer” on page 74)
16Piston, completeWM 30910
17Sealing washer 4 x 8 x 0.5 (for gage)WM 1145/35
18Sealing washer 4 x 8 x 0.8 (for gage)WM 1145/36
19Sinter filterWM6619
20Plate, filter screw changeWM0495
Gage capWM 30960
Safety tagWM 16211
Set of replacement sealsWM1148
Oxygen lubricantWM 14934
The instructions for use for the devices are available on the Internet at www.weinmann.de or contact
WEINMANN.
Replacement Parts65
Page 66
8.3 Additional replacement parts for WM 30050
7
23
2
24
2122
Item No.DescriptionOrder No.
2Connectorsee chapter 8.11
Thread connection, complete (oxygen outlet)
7
21Spring for pistonWM 30924
22Cover (body cover)WM 30922
23Screw plug G 1/8WM1519
24CapWM 30993
– Threaded connection up to SN 0849999
– Threaded connection from SN 0850000
WM 30927
WM 30002
66Replacement Parts
Page 67
8.4 Additional replacement parts for WM 30100, 30150,
30200, 30300
7
Item No.DescriptionOrder No.
2Connectorsee chapter 8.11
Threaded connection, complete (oxygen outlet)
– WM 30100 up to SN 0849999
7
21Spring for pistonWM 30924
22Cover (body cover)WM 30922
– WM 30100 from SN 0850000
– WM 30150
– WM 30200, 30300
2
21 22
WM 30926
WM 30005
WM 30904
WM 30936
Replacement Parts67
Page 68
8.5 Additional replacement parts for WM 30500, 30700, 30750
26 27 28
21
25
7
8
30
2
22
29
32
9
31
Item No.DescriptionItem No.
2Connectorsee chapter 8.11
Threaded connection, complete (oxygen outlet) up to SN 0849999:
– Threaded connection, complete 3 – 15 l/min for WM 30500
– Threaded connection, complete 1.5 – 8 l/min for WM 30700
– Threaded connection, complete 0.5 – 3 l/min für WM 30750
7
Threaded connection, complete (oxygen outlet) from SN 0850000:
– Threaded connection, complete 3 – 15 l/min for WM 30500
– Threaded connection, complete 1.5 – 8 l/min for WM 30700
– Threaded connection, complete 0.5 – 3 l/min für WM 30750
WM 30946
WM 30966
WM 30969
WM 31119
WM 31120
WM 31121
Flow gage for devices up to SN 0849999:
– Flow gage 3 – 15 l/min for WM 30500
– Flow gage 1.5 – 8 l/min for WM 30700
– Flow gage 0.5 – 3 l/min for WM 30750
8
Flow gage for devices from SN 0850000:
– Flow gage 3 – 15 l/min for WM 30500
– Flow gage 1.5 – 8 l/min for WM 30700
– Flow gage 0.5 – 3 l/min for WM 30750
9Rotary knob (adjuster knob)WM 30949
21Spring for pistonWM 30954
22Cover (body cover)WM 30952
25Washer 18 x 27.2 x 2WM 30944
26Spring for lifting pinWM 30959
WM 30961
WM 30962
WM 30963
WM 31112
WM 31113
WM 31114
– Lifting pin (up to SN 0849999)
27
– Lifting pin, long (from SN 0850000)
– Spindle (up to SN 0849999)
28
– Spindle (from SN 0850000)
29O-ring 11.91 x 2.62WM 1145/96
30Cap for adjuster knob 9, printedWM 30950
31Driver plateWM3812
32Hexagonal nut DIN 985 – M5 – 8 ST–GalZnWM 51446
68Replacement Parts
WM 30956
WM 31102
WM 30957
WM 31103
Page 69
8.6 Additional replacement parts for WM 30800, 30850
Threaded connection, complete (oxygen outlet):
– Threaded connection up to SN 0849999
– Threaded connection from SN 0850000
– Union nut
– Index head
– Orifice disk holder, complete
– Orifice disk 0-15 l/min
– Slide disk
– Ball
– Spring
– Index washer
– Slide disk 6 x 12 x 0.5
– Set screw DIN EN ISO 4029 M4X8-V2A with Tuflok
– Threaded connection, complete (oxygen outlet)
– Union nut
– Index head
– Orifice disk holder, complete
– Orifice disk 0-15 l/min
– Slide disk
– Ball
– Spring
– Index washer
– Slide disk 6 x 12 x 0.5
– Set screw DIN EN ISO 4029 M4X8-V2A with Tuflok
– Threaded connection from SN 0850000
– Union nut
– Index head
– Orifice disk holder, complete
– Orifice disk 0-6 l/min
– Slide disk
– Ball
– Spring
– Index washer
– Set screw DIN EN ISO 4029 M4X8-V2A with Tuflok
– Threaded connection from SN 0850000
– Union nut
– Index head (long part with triangular marker)
– Orifice disk holder, complete
– Orifice disk 0-15 l/min
– Slide disk
– Ball
– Spring
– Index washer
– Slide disk 6 x 12 x 0.5
– Set screw DIN EN ISO 4029 M4X8-V2A with Tuflok
Connection bolt, complete, 30 mm, G 3/4 for WM 30051, WM 30101,
WM 30301, WM 30351, WM 30501, WM 30701, WM 30751
Connection bolt, complete, 50 mm, G 3/4WM 1141/18
Connection bolt, complete, 80 mm, G 3/4WM 1141/42
Connection bolt, complete, 100 mm, G 3/4WM 1141
Connection bolt, complete, 140 mm, G 3/4WM 1141/92
Connection nut for connection bolt G 3/4WM 1144/16
Seal for G 3/4 connection boltWM 1145
Designation for G 5/8 bolt
Connection bolt, complete, 30 mm, G 5/8, 60° BullnoseWM 1141/66
Connection bolt, complete, 50 mm, G 5/8, 60° BullnoseWM 1141/55
Connection bolt, complete, 100 mm, G 5/8, 60° BullnoseWM 1141/56
Connection bolt, complete, 80 mm, G 5/8, 60° BullnoseWM 1141/45
Connection bolt, complete, 30 mm, G 5/8, 72° BullnoseWM 1141/53
Connection bolt, complete, 50 mm, G 5/8, 72° BullnoseWM 1141/52
Connection bolt, complete, 80 mm, G 5/8, 72° BullnoseWM 1141/44
Connection bolt, complete, 100 mm, G 5/8, 72° BullnoseWM 1141/59
Connection bolt, complete, 140 mm, G 5/8, 72° BullnoseWM 1141/28
Connection bolt, complete, 50 mm, G 5/8, 90° BullnoseWM 1141/98
Connection bolt, complete, 80 mm, G 5/8, 90° BullnoseWM 1141/79
Connection bolt, complete, 100 mm, G 5/8, 90° BullnoseWM 1141/75
Connection bolt, complete, 140 mm, G 5/8, 90° BullnoseWM 1141/84
Connection bolt, complete, 30 mm, G 5/8, 90° BullnoseWM 1141/82
Order No.
WM 1141/16
Order No.
Seal for G 5/8 connection boltWM 1145/4
Connection nut for connection bolt G 5/8 for BullnoseWM 1144/52
Connection nut for connection bolt G 5/8 for BullnoseWM 1144/54
Designation for NGO bolt
Connection bolt, complete, 40 mm, NGOWM 1141/77
Connection bolt, complete, 50 mm, NGOWM 1141/24
Connection bolt, complete, 80 mm, NGOWM 1141/43
Connection bolt, complete, 100 mm, NGOWM 1141/2
Connection nut for connection bolt NGOWM 1144/27
Seal for NGO connection boltWM 1145/50
Designation for W21.8/21.7 bolt
Connection bolt, complete, 30 mm, W21,8/21,7WM 1141/71
Connection bolt, complete, 50 mm, W21,8/21,7WM 1141/24
Connection bolt, complete, 80 mm, W21,8/21,7WM 1141/43
Connection bolt, complete, 100 mm, W21,8/21,7WM 1141/2
Order No.
Order No.
74Replacement Parts
Page 75
Connection nut for connection bolt W21,8/21,7WM 1144/2
Seal for W21,8/21,7 connection bolt WM 1145/2
Designation for bolt PIN index
Order No.
Connection bolt, PIN IndexWM 1142/86
Connection bolt, chrome-plated for 30 mm PIN indexWM 1142/88
Connection bolt, chrome-plated for 50 mm PIN indexWM 1142/87
Sealing ring 15,2-6,0-2,0WM 1145/86
PIN index hoop, chrome-platedWM 2841
PIN index pinWM 2842
T-screw, chrome-platedWM 2843
Replacement Parts75
Page 76
9. Tools, Inspection/Measuring/Test Equipment
Following is a list of all tools and inspection, measuring and test equipment mentioned in these Service and
Repair Instructions.
For details of the tools and equipment needed in the individual case, see the relevant section.
Special tools are obtainable through the manufacturer (WEINMANN).
9.1 General tools and resources
•Socket wrench 8 mm,
for slackening and tightening hexagonal
nut 32 in adjustment knob
(for WM 30500, 30700, 30750);
•Open-end wrench SW 12 or SW 14,
for slackening gage;
•Open-ended wrench, WAF 17, to undo the
plug connector for the central gas supply system (OXYWAY Click flowmeter);
•Special wrench SW 20 WM 22391 for tightening and slackening UNF 9/16 counter tool
•Open-end wrench SW 30,
for tightening connection nut of pressure
reducer OXYTRON (WM 30171).
9.2 Special tools
The following tools and inspection, measuring and test equipment are obtainable through the manufacturer
(WEINMANN):
WM 30035
comprising
– Bottom part
– Top part with clamping handle with
G 3/4 socket
– The following sockets are available
on special request:
G 5/8 Holland;
G 5/8 England;
G 21.8;
G NGO.
– Cover clamping device, for removing body
cover 22WM 30036
– Cover fitting device, for tightening body
cover 22WM 30037
– Counter crown, for unscrewing pressure
reducer bodyWM 30038
76Tools, Inspection/Measuring/Test Equipment
– Screwdriver bit 9 mm, for
unscrewing and screwing up
oxygen outlet 7WM 30039
– Single-end wrench SW 19, for 2nd stage
OXYTRON (WM 30150)WM 30040
– Watchmaker’s screwdriver, for
changing sealing ring 14WM 30041
– Allen key 4 mm, for
filter screw 12WM 30042
– Screwdriver 3 mm for slackening/
tightening screw 52WM 30043
– Adapter for
torque wrenchWM 30044
Page 77
Additional tools for pressure reducer FAST:
•Special gripping pliersWM 30046
•Clamping device,
for assembling coverWM 30047
•Loctite 601WM 14901
Oxygen lubricant,WM 14934
for O-rings of piston and
spindle;
Loctite 245WM 14920
for securing oxygen outlet
and second stage of WM 30150).
(BAM - tested to 15 bar, never
use a different Loctite product)
9.3 Inspection, measuring and test equipment
•Pressure gage 0 to 10 bar, class 1.0,
oxygen version,
for measuring outlet pressure at
oxygen outlet ;
•Pressure gage 0 to 2.5 bar, class 1.0,
oxygen version,
for measuring outlet pressure at
oxygen outlet ;
•Flow measuring tube 0 to 50 ml/min,
class 2.5, oxygen version,
for measuring leakages at oxygen
outlet with flow setting “0”
(for WM 30500, 30700, 30750);
•Flow measuring tube 0 to 20 l/min,
class 1.0, oxygen version,
for measuring
– outlet flow of 0 to 15 l/min
(for WM 30600, 30800, 30850),
– constant flow of 12 l/min
(for WM 30150);
•Flow measuring tube 0 to 220 l/min,
class 1.0, oxygen version,
for measuring
– constant flow of 90 l/min
(for WM 30850),
– constant flow of 110 l/min
(for WM 30200, 30300, 30350).
The following items of inspection, measuring and test equipment are obtainable from the firms listed below:
•Torque wrench, fixed setting of
– 55 Nm ± 1 Nm, Hazet No. 6291-1CT
for fitting connector
– 28 Nm + 2 Nm, Hazet No. 6391-35
for fitting gages
– 25 Nm ± 1 Nm, Hazet No. 6110-1CT
for fitting oxygen outlet (pressure outlet);
– 23 Nm – 1 Nm, Hazet No. 6391-35
for fitting gages 1 and 8;
– 16 Nm ± 2 Nm, for fitting oxygen outlet
(flow outlet);
– 10 Nm + 2 Nm, Hazet No. 6110-1CT
for screwing on second stage 52 of
pressure reducer WM 30150.