Testing equipment required
Preparing for final check
Entering the device data
Testing for leaks and pressure reading
Self-test when device is switched on
Functional check on controls
(button check)
Functional test and alarms
Battery power
Test pressure sensors
Functional check on frequency setting
Functional check
at 4.5 bar input pressure and
10 mbar back pressure
Checking O2 concentration
Functional check on pressure limit
Functional check on relief valve
without patient valve
Checking the type plate data
Check on external condition
Documentation
gas connection
Opening the device
Closing the device
Replacing button cell
Changing the speaker
Changing the board
Replacing the pneumatic block
Replacing the 3/2-way magnetic valve 29
Calibrating the potentiometer
(after changing pneumatic block
including potentiometer)
Changing upper part of housing
Changing the fascia film
Changing lower part of housing
For decades WEINMANN has been developing,
manufacturing and marketing devices for emergency medical care, oxygen therapy and inhalation
therapy.
In 1972 WEINMANN put the first MEDUMAT
emergency respirator on the market.
MEDUMAT emergency respirators are automatic
respirators. They are used for controlled respiration
in emergency medical care, e.g. in cases of acute
respiratory disorders, and also secondary obstructions.
The new generation of devices, developed specifically to meet users’ requirements and put on the
market in 1997, offers users and patients increased security. An intelligent alarm system monitors the patient’s breathing and informs the user
about any problems that occur. These devices thus
offer even greater security and reliability during
respiration.
The aim of these servicing and repair instructions is
to familiarise you, as an expert in the field, with the
function, technology, servicing and repair of the
MEDUMAT respirator. Thanks to training which you
have already received from WEINMANN, you now
count as “trained expert personnel” and can therefore give your customers appropriate instructions,
remedy problems on your own and perform the
functional checks prescribed in the operating instructions and any repairs required in accordance with
these Service and Repair Instructions.
In the event of a warranty claim, send the MEDUMAT
to WEINMANN.
To enable us to process ex gratia requests or warranty claims, please enclose the customer’s proof
of purchase (invoice) with the device.
Repairs or servicing work may be performed only
by WEINMANN or by trained specialist staff.
You are responsible for repairs carried out yourself
and for their warranty!
Use only original WEINMANN spares for repairs.
Please bear in mind:
Your customer trusts you and relies on your expert
capability, just as you rely on WEINMANN.
Note:
For the following information, please consult the Operating Instructions for MEDUMAT:
•Safety Information
•Fitting accessories
•Operating MEDUMAT emergency respirators
•Hygienic preparation
•Functional check
Introduction3
Page 4
1. Overview
Control panel MEDUMAT Easy
1 Mask/tube ventilation
switch with indicator
LEDs
2 Respiratory pressure
indicator (bar graph)
60
mbar
50
40
30
20
10
0
11
9
10
7
11
16
5
30
3
MV (l/min)
Stenosis
Disconnection
< 2,7 bar O
12
10
12
Freq.(min
3 Alarm panel
MEDUMAT
Easy
2
14
15
13
14
16
Demandflow
-1
)
4 Alarm confirmation
5 Colour code
6 Adjuster knob, ventilation parameters
7 Index position
8 LED Demandflow
9 ON/OFF switch
Connections MEDUMAT Easy
r
a
b
10 Pressurized gas
connection
15 Relief outlet valve
m
60
50
40
30
20
11
10
10
9
0
10
11
7
16
30
5
eq.(
r
F
3
l/min
(
V
M
T
A
M
U
D
E
M
sy
a
E
is
s
o
n
e
t
S
n
io
t
c
e
n
n
o
c
s
i
D
O
2
r
a
b
,7
2
<
12
14
12
15
13
16
14
-
d
n
a
m
e
D
w
lo
f
-1
)
in
m
)
60
50
40
30
20
10
0
T
A
M
U
D
E
M
sy
a
E
is
s
o
n
e
t
S
n
io
t
c
e
n
n
r
a
o
c
is
mb
D
O
2
r
a
b
7
,
2
<
12
1
1
4
1
12
10
9
15
3
10
1
6
1
14
1
1
7
-
d
n
a
m
e
D
w
o
l
16
f
30
5
1
-
)
n
i
m
(
.
eq
Fr
3
)
n
i
m
/
l
(
V
M
11 Battery compartment
12 Speaker
13 Pressure gauge hose
connection
14 Ventilation hose connection
4Overview
Page 5
1.1 Special symbols on the ventilator
Symbols on MEDUMAT Easy
Typ :
T
A
M
EDU
M
y
as
E
s
i
s
o
n
e
t
S
n
o
i
t
c
e
n
n
r
o
c
ba
s
i
m
D
O
2
r
a
60
b
7
,
2
<
50
40
30
20
12
11
10
14
12
10
9
15
0
10
13
16
14
11
7
-
d
n
a
m
e
D
w
o
l
16
f
30
5
1
-
)
n
i
m
(
.
q
e
r
F
3
)
n
i
m
/
l
(
V
M
Inlet 2,7 - 6 bar O2.
T
A
UM
ED
M
Easy
e
t
u
a
h
n
o
i
s
s
e
r
e
s
P
s
a
b
n
Ty
o
p
i
s
:
s
e
r
P
mbar
O
2
r
0
a
6
b
,7
2
<
0
5
0
4
0
3
0
2
2
1
1
1
0
4
1
1
2
1
0
1
9
5
1
0
0
3
1
1
6
1
4
1
1
1
7
6
1
30
5
3
SN
MEDUMAT Easy device information plate
Serial number of device
Year of manufacture
Do not dispose of device in domestic waste.
Safety check and servicing label
Servicing label: indicates when the next service is due.
Safety check label: (in Germany only) marks when the next safety check as per §6 of the German
law relating to users of medical devices is required.
Overview5
Page 6
2. Description
2.1 Uses
MEDUMAT Easy is an automatic oxygen respiration device (short-term ventilator) with additional inhalation
facility.
You can use MEDUMAT Easy:
•to revive patients at the site of the emergency;
•for longer periods in more protracted
emergencies, e.g. fires;
•for short-term O2 inhalation using a respiration
mask.
You can use MEDUMAT Easy while transporting
patients:
•between the various rooms and departments
of a hospital;
•between the hospital and other premises;
2.2 Ventilation function
MEDUMAT Easy operates within a pressure range
of 2.7 to 6 bar and at a flow rate of not less than
70 l/min O2. It has a built-in power supply.
It uses high-pressure, medicinal-grade oxygen. An
external pressure reducer brings this down to the
required operating pressure. The oxygen supply is
fed in at input valve 10.
The ventilation settings are continuously variable.
These settings (frequency and volume per minute
are coupled) and the inspiration/expiration ratio
of 1:1.67 are regulated by internal electronic
control mechanisms.
•in emergencies;
•when transport over considerable distances is
planned.
MEDUMAT Easy:
•is designed to provide controlled ventilation to
persons of 10 kg body weight or more;
•is used to treat respiratory arrest;
•can be preset to parameters that ensure evenly
balanced ventilation, provided that the
selected maximum ventilation pressure P
max
is
not exceeded;
•permits breathing-controlled oxygen inhalation
in Demand mode.
MEDUMAT
Easy
Stenosis
r
a
b
m
Disconnection
2
60
10
< 2,7 bar O
50
40
30
20
12
11
10
14
12
10
9
15
0
10
13
16
14
11
7
Demand-
w
16
flo
30
5
-1
)
q.(min
Fre
3
MV (l/min)
The gas for inspiration flows along the hose and
through the patient valve and either the mask or
tube into the patient’s airways. The patient valve is
fitted with a lip membrane that enables expired
gas to be conducted away through the expiration
tube.
6Description
Page 7
You can check the course of ventilation on the
10
respiration pressure indicator 2.
2.3 Demandflow function
mbar
60
50
2
40
30
20
10
9
0
The Demandflow setting switches the
MEDUMAT Easy to breathing-controlled O2
inhalation. Such inhalation must be carried out
with the respiration mask. A small inspiration
(trigger) pulse causes oxygen to continue flowing
until slight overpressure interrupts the flow.
Expiration then takes place via the patient valve as
in ventilation.
2.4 Patient valve
The gas for inspiration is channelled into the
patient’s airways through the patient valve.
The valve is designed to enable spontaneous
breathing in the event of failure of the
MEDUMAT Easy.
0
Demandflow setting
12
11
9
10
10
7
11
16
5
30
3
Freq.(min
MV (l/min)
12
14
15
13
14
16
Demandflow
-1
)
38
37
39
403435 36 41
2.5 Audio response
The device has an audio response facility that can
be switched on for user guidance, especially for
users who have little practice.
If audio guidance is not wanted, it can be
switched off by pressing a combination of keys
(see “4.10 Audio response for user guidance” in
the operating instructions).
Description7
Page 8
3. Final check
After every repair and every service, the device
must be subjected to the following final check in
accordance with Test Instructions WM 28001
and the test record.
Note:
For the final check on the MEDUMAT Easy you
must connect the ventilation hose and the patient
valve to the device.
MEDUMAT Easy must not be used if the final
check reveals defects or deviations from the
specified parameters.
We recommend that you always hold reserve
stocks of the following items:
•replacement washers for the connections;
•lip membrane for the patient valve.
•membrane for spontaneous breathing tube;
•membrane for exhalation tube;
•O-ring 1145/118.
3.1 Testing equipment required
•Volume flow meter, Type RT 200 (Timeter), Type EKU VIP – ventilator, PF 300 Imtmedical or comparable
test device
•Adjustable orifice, e.g. ball valve, internal diameter ≥ 10 mm
•Test set for functional checks WM 15323
•Oxygen concentration meter, 0 – 100% ± 1%,
e.g. Type Oxycontrol WM 13550
•Set, hose with syringe WM 15359
•Pressure gauge 0 - 6.3 bar, class 1.6
•Pressure gauge 0 -100 mbar, class 1.6
•Set, supply test Medumat / Modules WM 15440
3.2 Preparing for final check
1. Connect MEDUMAT Easy to the 4.5 - 6 bar pressure supply.
2. Connect ventilation hose and pressure measurement tube to MEDUMAT Easy.
If audio response is enabled, you will hear the sentence “Open oxygen cylinder” before the self-test starts.
1. Apply 4.5 bar to the input.
2. Switch on MEDUMAT Easy at button 9.
Requirement: The self-test starts, the ventilation mode LEDs light up once each one after the other, the alarm
LEDs flash, the pressure indicator LEDs are run through 3 times, a signal tone sounds and an audio
response sentence is spoken.
3.6 Functional check on controls (button check)
1. Open the pressure supply.
2. Switch on the device.
3. Switch from mask ventilation to tube ventilation.
4. Press the alarm acknowledgement button.
Final check9
Page 10
3.7 Functional test and alarms
3.7.1 Test Stenosis alarm
•Switch to tube ventilation () and close patient valve outlet.
•Operate device at Freq. setting = 30 min-1 and MV = 3 l/min.
Requirement: The Stenosis alarm must be triggered after two ventilation cycles. If audio response is
enabled, the ventilator announces “Check airways and minute volume”.
3.7.2 Checking alarm confirmation
•Immediately after the first alarm tone sounds, press Alarm confirmation button 4.
Requirement: The alarm tone must be suppressed immediately.
3.7.3 Test Disconnection alarm
•Open patient valve outlet.
Requirement: The Disconnection alarm must be triggered after two ventilation cycles. If audio response
is enabled, the ventilator announces “Check ventilation system and settings”.
3.7.4 Test Pressure alarm
•Shut off pressurised gas connection of MEDUMAT (2.7 - 6.0 bar).
Requirement: The Pressure alarm must be triggered. If audio response is enabled, the ventilator
announces “Check pressure hose system and gas supply”.
3.7.5 Checking Demand mode LED
•Open the pressure supply.
Requirement: Demand mode LED must come on (flickering).
3.7.6 Check voice output
3.8 Battery power
3.8.1 3.0 V battery (on board)
Since the 3.0 V lithium cell is difficult to access, its
charge status can be checked both via the
interface and in a relevant menu.
To do so, hold down the alarm confirmation button
while switching the device on. The device is then
in battery check mode for 3 seconds, before
switching to normal operating mode.
During this 3-second period the “Power supply”
alarm in the alarm panel lights up and the voltage
measured for the lithium cell is shown on the
pressure bar graph.
The adjacent table shows how the voltage
readings correspond to the bar graph values.
Bar graph value (mbar)Battery voltage (V)
603.21
553.19
503.17
453.15
403.13
353.11
303.09
253.07
203.05
153.03
103.01
52.99
02.97
10Final check
Page 11
Requirement: Voltage measured for 3.0 V battery is in the range 3.0 to 3.2 V; the alarm for this cell is
triggered at 2.7 V.
3.8.2 3.6 V battery (battery compartment)
Remove the 3.6 V battery from the battery compartment and measure the voltage with a digital multimeter.
Requirement: Voltage measured for 3.6 V battery is in the range 3.4 to 3.7 V.
3.9 Test pressure sensors
3.9.1 Input pressure sensor
•Shut off pressurised gas connection of MEDUMAT (2.7 - 6.0 bar).
Requirement: The Pressure alarm must be triggered. If audio response is enabled, the ventilator
announces “Check pressure hose system and gas supply”.
3.9.2 Ventilation pressure sensor
•Connect ventilation hose to test bag.
•Run device with settings: Freq. = 10 min-1 and MV = 11 l/min.
•Switch device to mask ventilation.
Requirement: The pressure limit must respond at 20 ± 5 mbar.
•Switch device to tube ventilation.
Requirement: The pressure limit must respond at 45 ± 5 mbar.
3.9.3 Demand flow sensor
1. Supply device with pressure.
2. Switch on MEDUMAT Easy at button 9.
3. Connect ventilation tube and patient valve to test bag.
4. Set MEDUMAT Easy to demand flow mode.
5. Green LED in demand mode button illuminates.
6. Squeeze the test bag.
Requirement: Releasing the test bag triggers a short ventilation.
Final check11
Page 12
3.10 Functional check on frequency setting
T
A
M
U
D
E
M
y
s
a
E
Stenosis
tion
c
r
a
b
Disconne
m
O
2
r
60
,7 ba
2
<
50
40
30
20
12
11
10
14
12
0
1
9
5
1
0
0
3
1
1
6
1
14
11
7
16
0
3
5
3
Patient valve with
hose system
Volume flow meter
Orifice 10 mbar
open
•Connect ventilation hose to 10 mbar orifice and to volume flow meter.
•Run device with settings: Freq. = 14 min-1 and MV = 16 l/min.
Requirement: The frequency must be 14 ± 2 min
-1
.
•Run device with settings: Freq. = 10 min-1 and MV = 11 l/min.
Requirement: The frequency must be 10 ± 2 min
-1
.
•Run device with settings: Freq. = 30 min-1 and MV = 3 l/min.
Requirement: The frequency must be 30 ± 2 min
-1
.
3.11 Functional check breath volume at 4.5 bar input pressure and
10 mbar back pressure
T
A
M
U
D
E
M
y
s
a
E
is
s
o
n
te
n
S
io
t
c
e
n
n
o
c
is
bar
D
m
O
2
r
0
a
6
b
7
,
2
<
0
5
0
4
0
3
0
2
2
1
1
1
0
4
1
1
2
1
0
1
9
5
1
0
0
3
1
1
6
1
4
1
1
1
7
6
1
0
3
5
3
Patient valve with
hose system
Volume flow meter
Orifice at 10 mbar
setting
•Run device with settings: Freq. = 14 min-1 and MV = 16 l/min.
Requirement: Breath volume must be 1140 ± 170 ml.
•Run device with settings: Freq. = 10 min-1 and MV = 11 l/min.
Requirement: Breath volume must be 1100 ± 170 ml.
•Run device with settings: Freq. = 30 min-1 and MV = 3 l/min.
Requirement: Breath volume must be 100 ± 20 ml.
12Final check
Page 13
3.12 Checking O2 concentration
•Run device with settings: Freq. = 10 min-1, MV = 11 l/min and 100% O2.
Requirement: The O
concentration must be > 98%.
2
3.13 Functional check on pressure limit
•Connect ventilation hose to test bag.
•Run device with settings: Freq. = 10 min-1 and MV = 11 l/min.
•Switch device to mask ventilation.
Requirement: The pressure limit must respond at 20 ± 5 mbar.
•Switch device to tube ventilation.
Requirement: The pressure limit must respond at 45 ± 5 mbar.
3.14 Functional check on relief valve without patient valve
•Run device with settings: Freq. = 11min-1 and MV = 7 l/min.
•Switch ventilation hose to test bag.
Requirement: The test bag is fully inflated during the inspiration stroke. The ventilator is then heard to
release pressure.
3.15 Checking the type plate data
•Check type plate data against drawing.
Requirement: The type plate data must be correctly entered in accordance with the drawing.
Final check13
Page 14
3.16 Check on external condition
•Check external condition.
Requirements: The outside of the device is not scratched and there are no flaws.
The connection thread is undamaged and screws easily.
The adjuster knob is secured by self-locking against inadvertent changes.
Elbow outlet turns easily.
3.17 Documentation
•Document items 3.3 to 3.16, and also test date and tester number, in the test record.
14Final check
Page 15
4. Servicing
Note:
Remember to perform a final check after every repair.
We recommend that maintenance work such as inspections and repairs be performed only by the
manufacturer, i.e WEINMANN, or by qualified technicians expressly authorized by WEINMANN.
4.1 Intervals and scope
Every 2 years:
Every 2 years the device (incl. patient valve and hose system) must undergo servicing and be subjected to a
safety check as specified below.
You can also have servicing and the safety check performed by WEINMANN.
Be sure to check the following items:
•Check equipment for completeness;
•Visual check:
– Mechanical damage
– Labelling of controls
– Damage to all external hoses;
•Renew parts subject to wear / parts requiring
compulsory replacement (see “7.2 Service
sets” on page 37);
•Final check in accordance with Test Instructions/
Test Record STK WM 28001 (see “3. Final check”
on page 8 and see “11. Repair and service
records” on page 43).
Every 4 years:
•Servicing of the fittings in the oxygen supply system (e.g. pressure reducer) either by the manufacturer or
by a specialist expressly authorized by the manufacturer.
Every 10 years:
•Repeat testing of conventional steel or aluminium oxygen cylinders by the responsible testing
organisation. The repeat testing date is stamped on the shoulder of the cylinder.
Servicing15
Page 16
4.2 Batteries
MEDUMAT Easy is equipped with two batteries.
The main battery 18 must always be changed. The
button cell 17 (CR2430) only has to be replaced
every 4 years.
The button cell 17 supplies auxiliary power to the
electronic system if the main battery 18 fails. This
makes it possible to set off an alarm even if the
main battery suddenly fails. The device then
switches to exhalation.
In principle, the battery capacity is designed so
that under normal conditions of use, it will not need
changing in the 2 years between services. The
main battery 18 is to be renewed during the 2yearly service, the button cell 17 only every 4
years.
17
60
MEDUMAT
m
b
a
r
Easy
S
te
n
os
50
is
D
isco
n
n
ec
t
io
n
<
2
,7
ba
40
r
O
2
30
20
10
11
12
0
9
10
10
12
14
7
11
15
13
14
16
16
5
30
D
em
an
d
-
3
flow
Freq
.(
m
-
1
in
)
M
V
(
l/m
i
n
)
42
18
When changing batteries, special precautions
must be taken to protect the electronic circuits (see
“6.5 Replacing button cell” on page 24).
17: Button cell for back-up power
18: Lithium battery 3.6 V for main power
supply MEDUMAT Easy
When changing the main battery 18, you must
change the O-ring 42 on the battery compartment
cover as well.
4.3 Storage
If you do not intend to use MEDUMAT Easy for a long period, we recommend the following storage
precautions:
1. Clean and disinfect the device (see “5.
Hygienic Preparation” in the operating
instructions).
2. Store MEDUMAT Easy in a dry place.
Important
Remember that devices still require servicing at the
specified intervals even when in storage, otherwise
they are not allowed to be used when removed
from storage.
16Servicing
Page 17
5. Troubleshooting
FaultCauseRemedy
A battery is exhausted
Fuse defectiveChange board (6.7, page 26)
MEDUMAT Easy does not
start up when switched on
MEDUMAT Easy will not
switch off
MEDUMAT Easy is
functioning, but without any
displays
MV too high
MV not correct
Ribbon cable X100 of fascia film is
faulty or not connected
On/Off button 9 defectiveChange fascia film (6.12, page 32)
Board defectiveChange board (6.7, page 26)
Magnetic valve defective
Operating error
On/Off button 9 defectiveChange fascia film (6.12, page 32)
Pressure gauge hose on
MEDUMAT Easy or on patient valve
has slipped off
Kink in pressure gauge hose
Pressure gauge hose within device is
kinked or has slipped off
Measured without 10 mbar back
pressure
Measuring device not calibratedCalibrate measuring device
Input pressure > 6 bar
Patient valve not in order
Potentiometer wrongly adjusted
Adjuster knob out of adjustment
Pneumatic block leaking
Change main battery 18 (4.2,
page 16), check button cell 17
(3.8.1, page 10) and replace if
necessary (4.2, page 16)
Check plug-in connection and cable
(6.7, page 26)
If necessary, replace upper part of
housing (6.11, page 31)
Change magnetic valve (6.9,
page 29)
Keep On/Off button 9 pressed for
at least 2 seconds
Check pressure gauge hose
Set to 10 mbar back pressure
Reduce system setting to less than
6 bar
Check membranes and O-ring,
replace if necessary (Chapter 6.7 of
operating instructions)
Readjust potentiometer (6.10,
page 30)
Readjust adjuster knob (6.10,
page 30)
Replace pneumatic block (6.8,
page 27)
Troubleshooting17
Page 18
FaultCauseRemedy
+–
Incorrect setting selected on device
Make correct setting (Chapter 6.5 of
operating instructions)
Check membranes and O-ring,
Patient valve not in order
replace if necessary (Chapter 6.7 of
operating instructions)
Pressure limit (P
max
) not in
Patient valve or test bag not correctly
connected
MV not in orderSee fault “MV not in order”
Check hose connections and test
bag
order
Hose connections in device not in
order
Check hoses, replace if necessary
(6.8, page 27)
Pressure sensor on board is faultyChange board (6.7, page 26)
Adjuster knob for ventilation
defective
Pressure measurement connection
blocked
Replace adjuster knob (6.3,
page 21 and 6.4, page 22)
Replace (6.8, page 27)
LED’s do not light upChange board (6.7, page 26)
Alarms (visual + acoustic) not
in order
Incorrect indication (Stenosis/
Disconnection)
Check settings, check hose
connection to patient valve (Chapter
6.7 of operating instructions)
No (visual + acoustic) alarmBoard defectiveChange board (6.7, page 26)
Alarm confirmation pressed?Wait for between 30 – 120 s
page 16), check button cell 17
(3.8.1, page 10) and replace if
necessary (4.2, page 16)
Pressure inlet leaking
Hose system in device is
leaking
Pressure sensor on board is
leaking
Angled connector in device is loose
or defective
Check (6.8, page 27)
Check hoses, replace if necessary
(6.8, page 27)
Change board (6.7, page 26)
Mask/tube switch 1 defectiveChange fascia film (6.12, page 32)
Pneumatic block leaking
Potentiometer wrongly adjusted
Replace pneumatic block (6.8,
page 27)
Adjust potentiometer (6.10,
page 30)
Frequencies not in order
Replace pneumatic block (6.8,
page 27)
Test bag is not filled
sufficiently during functional
check, Disconnection alarm
Potentiometer defective
Ventilation parameters incorrectly set Correct ventilation parameters
Patient valve not working properlyCheck lip membrane
Pressure measurement tube not fitted Fit pressure measurement tube
18Troubleshooting
Page 19
FaultCauseRemedy
No Stenosis alarm when
patient valve closed during
functional check (see
“Checking the breath volume”
in the operating instructions)
Patient valve not working properlyCheck lip membrane
Troubleshooting19
Page 20
6. Repair information and instructions
6.1 General
Always perform repairs to MEDUMAT Easy at an ESD-protected workplace.
•Observe the safety information in the Operating
Instructions for MEDUMAT Easy.
•All operations on this device require detailed
knowledge and observation of the Operating
Instructions and the Service and Repair
Instructions.
•Do not carry out any repairs that are not
described in these Service and Repair
Instructions. This is the only way to guarantee
trouble-free functioning of MEDUMAT Easy.
•Make sure that your hands and workplace are
clean during the repair work.
•Be sure to carry out a final check after every
repair (see “3. Final check” on page 8).
•If you replace components or individual parts,
use only genuine WEINMANN parts.
•When ordering the lower part of the
housing 24, please specify device type, year
of manufacture and device number.
•Note:
The item numbers quoted in the following text
are identical to the item numbers in the spare
parts list on Page 35 and the overview on
Page 4.
6.2 Changing the filter in the pressurised gas connection
Tools and equipment required
•Flat-head screwdriver
•Tweezers.
1. Unscrew the slot-head screw from the
pressurised gas connection 10.
2. Use tweezers to remove filter set 57.
3. Carefully insert a new filter set 57 in the
pressurised gas connection.
4. Screw the slot-head screw firmly into the
pressurised gas connection.
57
20Repair information and instructions
Page 21
6.3 Opening the device
12
11
Tools and equipment required
•ESD-protected workplace,
•Phillips screwdriver, size 1,
•Tubular hexagon box spanner 10 mm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the battery compartment with a coin
and remove the battery.
60
M
E
D
U
M
m
b
A
a
r
T
E
a
s
y
S
te
no
50
s
is
D
is
c
o
n
n
e
c
tio
n
<
2
,7
b
40
a
r
O
2
30
20
10
11
12
0
9
10
10
12
14
7
11
15
13
14
16
16
5
30
D
e
m
a
n
d
-
3
flo
F
r
w
e
q
.(
m
1
in
)
M
V
(
l/m
in
)
2. Lift off the lid 32.
3. Hold the adjuster knob 33 still with the special
tool and loosen the nut with a tubular hexagon
socket spanner (10 mm).
Do not unscrew the nut completely – only loosen
it. Otherwise the knob will split into its separate
parts.
4. Pull off the adjuster knob 33.
5. Place the device on a non-slip surface and
unscrew the 4 screws 43 from the rear of the
device.
6. Place the device on its side and carefully pull
the two parts of the housing apart until the
locating pin of the adjuster knob is pulled
completely out of the upper part of the
housing 20.
7. Remove the speaker 12 from the lower part of
the housing 24 and disconnect it.
10
0
7
5
14
12
10
9
15
10
13
16
14
11
16
Demand-
flow
30
-1
)
q.(min
Fre
3
33
32
MV (l/min)
43
24
43
60
M
E
D
U
M
m
b
A
a
r
T
E
a
s
y
Stenosis
50
Disconnection
< 2,7 bar O
40
2
30
20
10
11
12
0
9
10
10
12
14
7
11
15
13
14
16
16
5
30
Demand-
3
flow
Freq.(
min
1
)
MV (
l/m
in)
24
12
20
Repair information and instructions21
Page 22
8. Disconnect the plug of wiring harness 51 from
the board.
Note:
This is best done with flat-nose pliers. Be very
careful not to damage the board with the
pliers.
9. Unscrew the blue sensor tube 53 from the
pneumatic block.
Note:
Do NOT pull sensor tube 53 off the pressure
sensor on the board, as this will damage the
sensor.
10. Pull the ends of the Y-shaped sensor tube 54 off
the sensors on the board.
11. Flap the two parts of the housing apart and
place them with the outside downwards.
12. Unscrew screw 47 with the serrated washer 48
from the earth connection to the pneumatic
block.
13. Unscrew the two screws 46 with the spring
washers 45 and remove the battery contacts
from the battery compartment.
The two halves of the housing are now
separated.
Earth connection
53515447
46
45
48
6.4 Closing the device
Tools and equipment required
•ESD-protected workplace,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Tubular hexagon box spanner 10 mm,
•Special tool WM 22829 from special tool set WM 15349,
•Allen key SW 2.5.
1. Place the two parts of the housing side by side
with the outside facing downwards.
2. Place the battery contacts in the battery
compartment and screw them in place with the
two screws 46 and the spring washers 45.
46
45
22Repair information and instructions
Page 23
3. Push the ends of the Y-shaped sensor tube 54
12
11
onto the sensors on the board.
4. Screw the blue sensor tube 53 onto the
pneumatic block.
5. Connect the plug of wiring harness 51 to the
board.
6. Attach the earth connection to the pneumatic
block using screw 47 and serrated washer 48.
Earth connection
7. Connect up the speaker 12 and insert it in the
recess in the lower part of the housing 24. The
centering lug of the speaker must be located in
the corresponding groove in the lower part of
the housing.
8. Push the adjuster knob locating pin through the
hole in the board and put the two halves
together.
9. Insert the battery in the battery compartment
and close the lid with a coin. Make sure the
battery is correctly connected.
10. Now screw up the lower part of the housing
with the 4 screws 43 using a torque of
50 ±5 Ncm.
53515447
60
M
E
D
U
M
m
b
A
a
r
T
E
a
s
y
Stenosis
50
Disconnection
< 2,7 bar O
40
2
30
20
10
11
12
0
9
10
10
12
14
7
11
15
13
14
16
16
5
30
Demand-
3
flow
Freq.(
min
1
)
MV (
l/m
in)
48
24
Centering
lug
1220
60
M
E
D
U
M
m
b
A
a
r
T
E
a
s
y
S
te
n
o
50
s
is
D
is
c
o
n
n
e
c
tio
n
<
2
,7
b
40
a
r
O
2
30
20
10
11
12
0
9
10
10
12
14
7
11
15
13
14
16
16
5
30
D
e
m
a
n
d
-
3
flo
F
r
w
e
q
.(m
1
in
)
M
V
(
l/m
in
)
11. Secure the adjuster knob 33:
– Push the adjuster knob onto the spindle as
far as it will go, then pull it out a fraction.
– Hold the knob firm with the special tool and
screw it tight with a torque of
200 ±10 Ncm.
43
24
43
10
0
7
5
14
12
10
9
15
10
13
16
14
11
16
30
Fre
3
q.(min
and-
Dem
flow
-1
)
33
32
MV (l/min)
Repair information and instructions23
Page 24
12. Check the alignment of the adjuster knob.
When the knob is turned fully to the left, the
white line must point to MV: 3 l/min.
If it does not, slacken the nut and realign the
adjuster knob.
13. Place the cap 32 on the adjuster knob 33.
14. Perform the final check (see “3. Final check” on
page 8).
6.5 Replacing button cell
Tools and equipment required
•ESD-protected workplace,
•Phillips screwdriver, size 1,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Release the ribbon cable from its clamp X100:
To do so, lift up the top part of the clamp. Then
you can pull out the cable.
3. Unscrew the two screws 49 and remove the
board 21 from the upper part of the housing.
4. To remove the button cell 17, use a match or
similar to ease it slightly out of the holder and
pull the button cell out sideways with the other
hand.
Caution!
Do not use sharp or pointed objects for this
purpose, as this could damage the board.
5. Insert a new button cell 17.
Make sure it is installed the right way round.
X100
Push here
49
21
49
17
24Repair information and instructions
Page 25
6. Insert the board 21 in the upper part of the
housing.
Make sure that the ribbon cable is not under the
board or jammed between it and the housing.
49
21
7. Attach the board 21 with the two screws 49.
8. Place the ribbon cable in the clamp X100 on
the board and then press down the top part of
the clamp.
X100
49
9. Close the device (see “6.4 Closing the
device” on page 22). Use a new main
battery 18.
10. Perform a final check (see “3. Final check” on
page 8).
Please remember that old batteries must not be disposed of in household waste. Always take old batteries
to a local collection point.
6.6 Changing the speaker
Tools and equipment required
•ESD-protected workplace,
•Phillips screwdriver, size 1,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349.
1. Open the device (see „6.3 Opening the
device“ on page 21, steps 1. to 6.).
2. Remove the speaker 12 from the lower part of
the housing 24 and disconnect the speaker
cable plugs. Put the old speaker on one side.
3. Take the new speaker 12 and connect the
speaker cable plugs.
4. Insert the speaker 12 in the recess in the lower
60
50
40
30
20
10
0
part of the housing 24. The centering lug of the
speaker must be located in the corresponding
groove in the lower part of the housing.
5. Close the device (see „6.4 Closing the
device“ on page 22, steps 8. to 13.).
6. Perform the final check (see “3. Final check” on
page 8).
M
E
D
U
M
m
b
A
a
r
T
E
a
s
y
Stenosis
Disconnection
< 2,7 bar O
2
24
11
12
9
10
10
12
14
7
11
15
13
14
16
16
5
30
Demand-
3
flow
Freq.(
min
1
)
MV (
l/m
in)
Centering
lug
1220
Repair information and instructions25
Page 26
6.7 Changing the board
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 1,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Release the ribbon cable from its clamp X100:
To do so, lift up the top part of the clamp. Then
you can pull out the cable.
3. Unscrew the two screws 49 and remove the
board 21 from the upper part of the housing.
4. Place the new board 21 on the spacers.
Make sure that the ribbon cable is not under the
board or jammed between it and the housing.
X100
49
21
49
5. Attach the board 21 with the two screws 49.
6. Place the ribbon cable in the clamp X100 on
the board and then press down the top part of
the clamp.
7. The screw of the blue sensor tube 53 is secured
with a nut for transport. Remove the nut before
screwing the tube to the pneumatic block
8. Close the device (see “6.4 Closing the
device” on page 22).
9. Perform a final check (see “3. Final check” on
page 8).
49
21
53
X100
49
26Repair information and instructions
Page 27
6.8 Replacing the pneumatic block
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 2,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Press the catch and unplug the magnetic valve
control cable.
3. Unplug the potentiometer cable.
Potentiometer
cable
4. Remove the clip 27 from the pneumatic
block 22.
5. Pull the elbow connector 26 off the pneumatic
block 22.
6. Turn the lower part of the housing over, hold it
and the pneumatic block tight and unscrew
both screws 44 together with the spring
washers 45.
7. Turn the lower part of the housing over again.
8. Press the catch towards the elbow connector
and pull the pneumatic block 22 off the
tube 56.
When you remove the pneumatic block from
the lower part of the housing, the O-ring 28
will fall out of the connection. Make sure that
the O-ring 31 does not fall out of the relief
outlet valve.
56
Catch
22
Magnetic valve
control cable
22
27
26
45
44
2831
Repair information and instructions27
Page 28
9. When inserting the new or replacement
pneumatic block 22, insert the tube 56 in the
elbow connector. Make sure that the cables
are routed below the pneumatic block to the
right-hand side.
10. Turn the lower part of the housing over, hold it
and the pneumatic block tight and screw up
both screws 44 with their spring washers 45.
56
45
44
22
Cable
11. Lubricate the O-rings of the elbow connector
26 with a little O
lubricant (WM 14298).
2
Make sure that the compensating hole remains
free.
12. Insert the O-ring 28 in the groove in the outlet
from the pneumatic block.
13. Insert the elbow connector 26 in the pneumatic
block 22.
14. Insert the clip 27 in the pneumatic block 22.
15. Plug in the potentiometer cable.
16. Plug in the magnetic valve control cable.
17. Close the device (see “6.4 Closing the
device” on page 22).
18. Perform a final check (see “3. Final check” on
page 8).
22
27
28
26
26
Compensating
hole
Potentiometer
cable
Magnetic valve
control cable
28Repair information and instructions
Page 29
6.9 Replacing the 3/2-way magnetic valve
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 2,
•Phillips screwdriver, size 0,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Remove the pneumatic block (see “6.8
Replacing the pneumatic block” on page 27).
3. Place the pneumatic block on the side with the
type plate.
4. Unscrew the two securing screws and remove
the 3/2-way magnetic valve 23.
5. Insert the new 3/2-way magnetic valve 23 in
the correct position.
Make sure that the seal for the 3/2-way magnetic
valve is seated in the correct position.
6. Fasten the 3/2-way magnetic valve with the
screws supplied with it.
7. Reinstall the pneumatic block (see “6.8
Replacing the pneumatic block” on page 27).
8. Close the device (see “6.4 Closing the
device” on page 22).
9. Perform a final check (see “3. Final check” on
page 8).
23
Repair information and instructions29
Page 30
6.10 Calibrating the potentiometer
(after changing pneumatic block including potentiometer)
When the pneumatic block including potentiometer is changed, the potentiometer must be recalibrated for
the board.
1. Make sure the device is switched off.
2. Press the On/Off switch 9 and immediately
hold down the alarm confirmation button 4
and the mask/tube ventilation switch 1. Then
release the button. After a brief interval, the 30
mbar LED on the bar graph indicator lights up.
3. To set the first calibration value, turn the
adjuster knob 6 fully to the left to the value
MV 3, Freq. 30.
4. Press the mask/tube switch 1. The device now
makes a plausibility check.
– If the value is not correct, all alarm LEDs light
up until the correct value is present at the
adjuster knob 6 and has been confirmed
with the mask/tube ventilation switch 1. Or
you can cancel the potentiometer
calibration process by pressing the On/Off
switch 9.
– If the value is correct, this is indicated by the
0 mbar LED on the bar graph lighting up.
You can now set the next value.
5. Turn the adjuster knob 6 fully to the right to the
white Demandflow zone.
6. Press the mask/tube switch 1. The device
accepts this value The LED on the bar graph
lights up, and you can move on to setting the
next value.
7. Turn the adjuster knob 6 back beyond the
index point 7 to the ventilation mode zone,
then to the right again until it reaches the index
point (zone MV 16, Freq. 14 ).
8. Press the mask/tube switch 1 again. The
calibration values are stored and the device
exits the calibration mode.
Note:
60
mbar
50
40
30
20
10
0
19
11
9
10
7
11
16
5
30
3
MV (l/min)
Stenosis
Disconnection
< 2,7 bar O
12
10
Freq.(min
MEDUMAT
Easy
2
14
12
15
13
14
16
Demandflow
-1
)
4
6
7
Until the last step the calibration mode can be
cancelled at any time by pressing the On/Off
switch 9, without storing the new values. If an
invalid value is detected during calibration (all
alarm LEDs light up), no value is stored either.
30Repair information and instructions
Page 31
6.11 Changing upper part of housing
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 2,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Remove the board (see „6.7 Changing the
board“ on page 26, steps 2. and 3.).
3. Remove the grommet 52 from the upper part of
the housing 20.
You have now removed all the parts. You can start
reassembling.
4. Insert the grommet 52 in the new upper part of
the housing 20.
5. Refit the board (see „6.7 Changing the board“
on page 26, steps 4. to 6.).
6. Close the device (see “6.4 Closing the
device” on page 22).
7. Perform a final check (see “3. Final check” on
page 8).
20
52
Repair information and instructions31
Page 32
6.12 Changing the fascia film
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 2,
•Tubular hexagon box spanner 8 mm,
•Tubular hexagon box spanner 10 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special tool WM 22829 from special tool set WM 15349,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Remove the board (see „6.7 Changing the
board“ on page 26, steps 2. and 3.).
3. Remove the grommet 52 from the upper part of
the housing 20.
4. From inside the housing, insert the tubular box
spanner through the hole for the grommet 52
and press the fascia film 19 outwards until you
can grasp it on one side. Then completely
remove the fascia film from the upper part of
the housing.
5. Use 70% isopropanol to remove all traces of
adhesive from the upper part of the housing.
Then wait until the isopropanol has completely
evaporated from the housing surface.
You have now removed all the parts. You can start
reassembling.
6. Stick the new fascia film 19 to the upper part
of the housing:
– First position the fascia film on the upper
part of the housing on the side where the
ribbon cable is attached.
– Run the ribbon cable through the slit in the
upper part of the housing.
– Then affix the entire fascia film, taking care
to avoid bubbles.
19
20
19
20
52
6
0
5
0
4
0
3
0
2
0
1
0
0
MEDUMAT
Easy
m
b
a
r
S
t
e
n
o
s
is
D
is
c
o
n
n
e
c
t
io
n
<
2
,7
b
a
r
O
2
1
1
1
2
9
1
0
1
1
2
4
1
0
7
1
5
1
1
1
3
1
4
1
6
1
6
5
3
0
D
e
m
a
n
d
-
flow
3
F
re
q
.(m
-1
i
n
)
M
V
(l/min
)
Ribbon
cable
32Repair information and instructions
Page 33
7. Refit the grommet 52 in the upper part of the
housing 20.
8. Refit the board (see „6.7 Changing the board“
on page 26, steps 4. to 6.).
9. Close the device (see “6.4 Closing the
device” on page 22).
20
52
10. Perform a final check (see “3. Final check” on
page 8).
6.13 Changing lower part of housing
Tools and equipment required:
•ESD-protected workplace,
•Phillips screwdriver, size 2,
•Open-ended spanner, 13 mm,
•Open-ended spanner, 22 mm,
•Torque spanner 50 ±5 Ncm,
•Torque spanner 200 ±10 Ncm,
•Special counter tool G 3/8 (WM 22827) and special spanner 17 mm (WM 22828) from special tool
set WM 15349,
•Vice with jaw protectors,
•Flat-nose pliers.
1. Open the device (see “6.3 Opening the
device” on page 21).
2. Remove the pneumatic block (see „6.8
Replacing the pneumatic block“ on page 27,
steps 2. to 8.).
3. Undo the screws 50 and remove the wiring
harness 51.
4. Use a pointed object to push the blanking plug
59 of the interface connection out of the lower
part of the housing 24 from inside.
5. Remove the O-ring 31 from the lower part of
the housing 24.
50
51
24
59
24
31
Repair information and instructions33
Page 34
6. Screw the special counter tool onto the
pressurised gas connection 10.
7. Clamp the special counter tool in a vice.
10
8. Use a 22-mm open-ended spanner to screw
the nut of the special counter tool firmly again
the pressurised gas connection.
9. Screw the elbow connector 25 off with a
13-mm open-ended spanner.
10. Use the 17-mm special spanner to slacken the
nut 30 and unscrew it completely.
11. Lift out the plate 29.
12. Remove the lower part of the housing 24.
13. Take the new lower part 24 and place it on the
pressurised gas connection 10.
14. Slide the plate 29 on the inside of the housing
onto the connection.
15. Tighten the 17-mm nut 30 on the inside of the
connection.
16. Fasten the elbow connector 25 to the
connection.
17. Use the 22-mm open-ended spanner to undo
the nut of the special counter tool.
18. Open the vice.
29
30
10
24
25
19. Screw the special counter tool off the
pressurised gas connection 10.
20. Using screws 50, screw the interface cable of
the wiring harness 51 into the new lower part
of the housing 24.
21. Push the blanking plug 59 into the interface
connection from outside.
22. Refit the O-ring 31 in the lower part of the
housing 24.
23. Reinstall the pneumatic block (see „6.8
Replacing the pneumatic block“ on page 27,
steps 9. to 16.).
24. Close the device (see “6.4 Closing the
device” on page 22).
25. Perform the final check (see “3. Final check” on
page 8).
50
51
31
59
24
34Repair information and instructions
Page 35
7. Spare parts
7.1 Spare parts list
Note:
The item numbers in the following table are identical to the numbers in the text of these Service and Repair
Instructions and in the Operating Instructions.
Item No.DescriptionOrder No.
10Pressurised gas connection (threaded), preassembledWM 22685
12
16Ventilation hoseWM 22647
17
18
19Fascia filmWM 28009
20
Speaker
– Seal
– Button cell CR 2430
– Battery 3.6 volt
Upper part of housing with film up to device no. 2279*
comprising:
– Upper part of housing with film
– Label for languages
– Label for patient connection
Upper part of housing with film from device no. 2280*
comprising:
– Upper part of housing with film
– Rating plate
– Label for patient connection
Lower part of housing up to device no. 2279*
comprising:
– Lower part of housing
– Rating plate
– Label for O2 input
– Label for excess pressure valve
24
25Elbow connector 4/6WM 22552
26
– Label for battery position
Lower part of housing from device no. 2280*
comprising:
– Lower part of housing
– Label for O2 input
– Label for languages
– Label for excess pressure valve
– Label for battery position
The input pressure at p is max. 6bar. This is dynamically reduced by V1 to 2.5 bar. This is the input pressure
at V2, V3 and V4.
Inspiration
An electrical impulse to V2 opens V3 and closes V4. Oxygen flows through the ventilation hose to the patient
valve. If the ventilation pressure in the patient valve reaches >100 mbar, the relief valve V5 will open.
Expiration
A fresh electrical impulse closes V2. The relief valve V4 opens and vents the ventilation hose. The patient
breathes out through the patient valve.
Demandflow
An inspiration impulse (trigger) at V2 opens valves V3 and V4.
Electronic system
The microprocessor-controlled electronic system sets the ventilation parameters and monitors ventilation, and
also O2 supply and power supply. If necessary, a visual and acoustic alarm is given. The ventilator has an
audio response facility that can be switched on for user guidance.
Patient valve
During inspiration the respiratory gas flows to the patient.
During expiration the expiration pressure switches the valve so
that the patient can breathe out.
Technical data41
Page 42
10. Technical Changes
Technical ChangesFrom Device No.Date
New position of device information plates228027.04.07
42Technical Changes
Page 43
11. Repair and service records
DateSignature
Service performed in accordance
with MEDUMAT service
instructions
Company
_______________________________
Measures / Comments
Company
DateSignature
_______________________________
Company
DateSignature
_______________________________
Company
DateSignature
_______________________________
22525 Hamburg
Device master dataService and repair work carried out in accordance with service instructions
Manufacturer: WEINMANN GmbH +
Co.
Device type:MEDUMAT Easy
Order No.: ________________________
Date of manufacture: ________________
Safety check - 2 years _______________
Safety check - 4 years _______________
Repair and service records43
Safety check - 6 years _______________
Safety check - 8 years _______________
Safety check - 10 years _______________
Page 44
Test record: Safety check in accordance with Test Instructions WM 28001
Device: MEDUMAT EasyWM No.: 28000Device No.: .......………........Date of manufacture: …...................
1. Testing equipment
• Pressure gauge 0 – 6.3 bar, class 1.6 (test pressure 6 ± 0.15 bar, operating pressure 4.5 ± 0.15 bar)
• O2 concentration meter
• Volume flow meter (e.g. RT 200, ViP, PF-300); adjustable orifice 10 mbar; test set WM 15323
• Digital multimeter
2. Preparations for test
• Connect MEDUMAT to test equipment.
• Set language “English”
• Set MEDUMAT to f = 10 min-1, MV = 11 l/min and “tube” (p
= 45 mbar).
max
3. Enter device data
• Entry of above device data MeasurementOK not OK
4. Leak test at 6 bar
• Pressure drop inlet side ≤ 0.2 bar/min
• Pressure drop pressure measurement zone < 2.0 mbar/min
5. Self-test when device is switched on
• 4 red and 1 green LED (i.e. one each for “Tube” and “Mask”) light up
alternately, alarm tone and audio response sound
6. Functional check, control elements
• Key check
7. Functional check, display elements
• Check on alarm LEDs
• Check on Demand mode LED
• Check on Ventilation pressure indicator LEDs
• Check on audio response
8. Check batteries
• 3.1 V ± 0.1 V
• 3.5 V + 0.2 / -0.1 V
9. Check pressure sensors
• Input pressure sensor
• Ventilation pressure sensor
• Demandflow sensor
10.Check frequency setting
• Frequency 14 ± 2 min
• Frequency 10 ± 2 min
• Frequency 30 ± 2 min
-1
-1
-1
11.Check breath volume at 4.5 bar supply pressure and 10 mbar back pressure
• f = 14 min-1, MV = 16 l/min:BV = 1140 ± 170 ml
• f = 10 min-1, MV = 11 l/min:BV = 1100 ± 170 ml
• f = 30 min-1, MV = 3 l/min:BV = 100 ± 20 ml
12.Check O2 concentration at f = 10 min-1 and MV = 11 l/min
• O2 concentration > 98 vol.%
13.Functional check on pressure limit at f = 10 min-1 and MV = 11 l/min
• Pressure limit “Mask” responds at 20 ± 5 mbar
• Pressure limit “Tube” responds at 45 ± 5 mbar
14.Functional check on relief valve
• Test bag is fully inflated, then ventilator vents audibly
15.Check on equipment and accessories
Presentyesno
• Ventilation hose and patient valve undamaged and operational
• Test set for functional check operational
• Pressure reducer operational
• O2 cylinder within test deadline, valve operational
• Carrying platform complete and operational
• Medical products book
• Operating instructions
16.Check on external condition
• Connection thread undamaged, adjuster knob operational, index function
working, swivelling elbow outlet moves easily
Service performed: yes no Final check performed:_______________________________