Weinmann Oxymat 3 Service manual

Servicing and
repair instructions
OXYMAT 3
Oxygen Concentrator
Contents
Introduction
1.
Overview
2.
Functional description
3.
Servicing
3.1
3.2
3.3
4.
Hygienic Preparation
4.1
4.2
5.
Functional check
5.1
5.2
5.3
6.
Troubleshooting
6.1
6.2
6.3
6.4
. . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Implementation Disposal
No change of patient Change of patient or reuse
General Intervals Implementation
5.3.1
5.3.2
5.3.3
5.3.4
5.3.5
5.3.6
5.3.7
5.3.8
5.3.9
5.3.10
5.3.11
5.3.12
5.3.13
5.3.14
Visual and acoustic signals from
6.1.1
6.1.2
6.1.3 Remedies
O2 concentration outside
tolerance range . . . . . . . . . . . . . . . . .
System pressure fault
. . . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . 10
Checking the housing Checking the coarse dust filter Checking the power cord Checking the fuse Checking the main switch and LED displays Checking the dosage monitor for leaks Checking the flowmeter for leaks Checking the operating hours meter Checking the O2 concentration Checking the mains failure alarm Checking the indirect cator (yellow LED) with appliance
closed
Checking the indirect status indicator (yellow LED) with the appliance System pressure test Checking the O2 system for leaks
OXYMAT 3 . . . . . . . . . . . . . . . .
Alarms Signal patterns Deactivating the signal functions
. . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . . 9
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . 9
. . . . . . . . . . 9
. . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . 10
status indi
open
. . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . 22
. . . . . . . . . . . . . 27
10
-
19
19
27
3
4
5
8
9
7.
Repair information and repair instructions
7.1
General
7.2
Replacing the castors
7.3
Opening the appliance
7.4
Closing the
7.5
Replacing the compressor
(up to appliance no. 13350) . . . . . . . .
7.6
Replacing the compressor (from appliance no. 13351 to 14999) .
7.7
Replacing the compressor
(from appliance no. 15000) . . . . . . . .
7.8
Replacing the operating hours meter
7.9
Replacing the circuit board
7.10
Replacing the fan
7.11
Replace adsorption containers
(Mac valves) . . . . . . . . . . . . . . . . . . .
7.12
Replace adsorption container
(plastic magnetic valve block) . . . . . . . .
7.13
Replace adsorption containers
(aluminum magnetic valve block) . . . . . .
7.14
Replace cover
7.15
Replacing the pressure reducer
7.16
Replacing the frame
(up to appliance no. 19999) . . . . . . . .
7.17
Replacing the frame
(from appliance no. 20000) . . . . . . . .
7.18
Replacing magnetic valves with valve block
(up to appliance no. 34999) . . . . . . . .
7.19
Replace magnetic valve block
8.
Spare parts
8.1 List of spare parts . . . . . . . . . . . . . . . 69
8.2 Maintenance set. . . . . . . . . . . . . . . . 72
9. Tools, testing equipment, disinfectants . . . . . 73
9.1 General tools and test equipment. . . . . 73
9.2 Special tools and test equipment . . . . . 73
9.3 Disinfectants . . . . . . . . . . . . . . . . . . 74
10. Technical Data. . . . . . . . . . . . . . . . . . . . . . . 75
10.1 Arrangement of the PCBs . . . . . . . . . . 76
11. Technical Changes . . . . . . . . . . . . . . . . . . . . 77
12. Repair and inspection log . . . . . . . . . . . . . . 78
13. Test report. . . . . . . . . . . . . . . . . . . . . . . . . . 79
. . . . . . . . . . . . . . . . . . . . . 28
. . . . . . . . . . . . . 28
. . . . . . . . . . . 29
appliance
. . . . . . . . . . . . . . . . . 53
. . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 30
. . . . . . . . . 40
. . . . . . . . . . . . . . . 42
. .
28
30
33
36
. . . 39
43
46
49
. . . . . . 54
57
59
60
. . . . . . . 66
69
© Copyright Weinmann GmbH & Co. KG. The content and presentation are copyright protected and may only be used by authorised Weinmann Service Partners in the course of their service operations. The content must not be reproduced or passed on to third parties. The complete documents must be returned on termination of the cooperation with Weinmann.
2

Introduction

For decades, Weinmann has developed, manu­factured and distributed equipment for oxygen ther­apy, inhalation therapy, sleep apnea therapy and emergency medicine.
In 1982, Weinmann introduced the first oxygen concentrator to the market.
Long-term oxygen therapy with an oxygen concen­trator now has a firm place in the treatment of chronic obstructive diseases of the respiratory tracts. Many studies have proven that long-term oxygen therapy can bring about a tangible im­provement in life expectancy and quality of life. The OXYMAT 3 oxygen concentrator from Wein­mann is part of this therapy concept.
The aim of these service and repair instructions is to familiarise you, as a knowledgeable expert, with the OXYMAT 3 with regard to function, techno­logy, servicing and repairs. In conjunction with the training you have already received from Wein­mann, you are now a "trained, qualified expert" and are able to instruct your clients correctly, rectify faults yourself, and carry out the functional checks
described in the instructions, as well as conduct any repairs which may be necessary, as outlined in these service and repair instructions.
In the event of a guarantee claim, OXYMAT 3 should be returned to Weinmann.
In order to enable us to process any guarantee or goodwill claims, please return the consumer's proof of purchase (invoice) together with the appli­ance.
Repair and maintenance work must be carried out only by Weinmann or by knowledgeable expert staff.
You are responsible for any repairs performed by yourself and for guaranteeing them!
Only original Weinmann spare parts should be used
for repair purposes.
Please remember: Your customer trusts you and relies on your exper­tise, just as you rely on Weinmann.
Note:
The following information can be found in the description and operating instructions for OXYMAT 3:
Safety instructions
Installation
Operation
Hygienic preparation
Warranty
3

1. Overview

Front
(with dosage monitor)
12 Nasal cannula
11 Connecting tube
(3 m) for nasal cannula
1 Handle for lifting or
moving
2 On/Off knob
3 Warning notices
4 Operating hours meter
5 Oxygen outlet
10 Power cord
Dosage monitor
9
Back
up to appliance no. 19999 from appliance no. 20000
13 Coarse dust filter
14 Power cord socket
15 Fuse holder
Front
(with Flowmeter)
16 Flowmeter
17 Connecting tube
(10 m)
Indicators
19 Oxygen concentration status
indicator (yellow)
20 Power indicator (green)
6 Connecting nipple
with union nut (fitted)
7 Connecting tube
(20 m)
8 Castors with brake
18 Fault indicator (red)
O
O/I
2
Oxygen appliance!
Increased
Dosage monitor
26 Connection,
3 m hose to nasal cannula
25 Humidifier
glass of bubble humidifier
4 Overview
21 Dosage unit (flowmeter)
0
,
5
0
,
4
0
,
3
0
,
2
0
,
1
5
,
0
24 Flow control knob
22 Ball, indicates
flow
23 Connection, long
tube to OXYMAT 3 (20 m)
Flowmeter
16 Flowmeter
22 Ball, indicates
flow
28 Connection,
OXYMAT 3
29 Connection for
tubes (3 m or 10 m)
27 Flow control knob

2. Functional description

OXYMAT 3 is a time-controlled appliance, i.e. the changeover time remains constant; only the pressure is monitored.
Layout
The appliance essentially consists of three compo­nents:
the compressor,
the O2 system,
the electronic/pneumatic control.
During operation, the compressor draws in ambi­ent air via a coarse dust filter and a suction filter, and passes the compressed air (approx. 1.7 bar overpressure) into containers with molecular sieves where the actual oxygen enrichment takes place.
The functional principle of molecular sieves is based on their property of absorbing the nitrogen contained in the air. The oxygen contained in the air is then collected in the oxygen tank (further de­tails can be found in the chapter “ The O2 enrich­ment process” on page 6).
The outlet pressure is regulated by a pressure­reducer set to 0.6 bar.
The required quantity (flow) of oxygen is set on the oxygen dosage control knob 24. Where a flow­meter is used, the flow is controlled via the flow control knob 27. The flow is read at the top edge of the flowmeter ball 22.
Every time it is switched on, the appliance control system performs an internal self-test of the entire O2 system. This test takes about a minute. If the test de­tects an error in the O2 system that results in a drop in the oxygen concentration, it activates the yellow status indicator 19.
After the self-test, OXYMAT 3 switches to normal operation. During normal operation the sequence control constantly monitors the electrical and pneu­matic circuits of the oxygen system.
Dosage monitor
Pressure-reducer
Oxygen tank
Container with molecular sieve
Noise suppressor Cooler
Compressor
Suction filter
Coarse dust filter
5,0
4,0
3,0
2,0
1,0 0,5
Ambient air
In order to allow the process to run continuously, the containers operate alternately, i.e. before one container is saturated with nitrogen, the system switches to the second container, which then continues with oxygen enrichment. During this period, the first container is regenerated by a brief reduction in pressure, and at the same time is flushed by a partial flow of enriched oxygen. This procedure is implemented cyclically at 7­second intervals, thereby ensuring continuous oxygen enrichment of the inspiratory air.
Functional description 5
The level of oxygen concentration depends on the set flow and is available within a few minutes.
The waste air is expelled to the rear via a noise suppressor on the underside of the appliance.
The electronic control with microprocessor consists of one circuit board. It is housed separately from the oxygen system on the front panel.
The O2 enrichment process
The compressor draws the ambient air into the housing via a coarse dust filter. Part of this air is used to cool the compressor assembly. The other part enters the compressor via the suction filter, where it is compressed. A cooler reduces the tem­perature of the air. The compessed and cooled air is then passed to the O2 system to enrich the oxygen.
Oxygen concentration
Flow [ l/min ]
Ambient air
The adsorption containers (ADS containers) are alternately filled with compressed air in fixed cycles, then regenerated by expanding the compressed air and flushing with oxygen.
At the end of each enrichment, a pressure of around 1.7 bar will have built up in the currently active container.
After 7 seconds, the magnetic valves are alter­nately triggered by the electronic control and switch between the two ADS containers.
Of the valves in the two ADS containers, one is open and the other closed during operation. Once the changeover time has been reached, pressure compensation is implemented between the two ADS containers. During pressure compen­sation, both valves are open for approximately 1 second.
The actual process of oxygen enrichment takes place in the ADS containers. A molecular sieve adsorbs the nitrogen from the air, thereby increas­ing the proportion of oxygen by volume.
As soon as the pressure in the ADS container has risen sufficiently, air enriched with oxygen flows into the oxygen tank at the other end of this con­tainer via the non-return valve.
Noise suppressor
Magnetic valves
Non-return valves
Suction filter
Compressor
Cooler
ADS container
Oxygen tank
Coarse dust filter
ADS container
Flushing nozzle
Throttle
Shutoff valve
outlet
2
O
Pressure sensor
Pressure reducer
6 Functional description
Part of the oxygen flows via the flushing nozzle into the other ADS container for flushing.
During the oxygen enrichment process, the area where adsorption of the nitrogen occurs will shift from the beginning of the ADS container towards the end. Before reaching the end of the container, the pressure between the two ADS containers is compensated via the pressure compensating valve. Parallel to this, the pressure escapes via the noise suppressor and the other ADS container is opened. The same procedure then takes place there.
Via pressure compensation, the remaining oxygen passes to the other ADS container, compression energy is saved, and blow-off noise is reduced.
The nitrogen adsorbed in the first ADS container is desorbed as a result of the pressure drop when the valve is opened and is expelled via the noise suppressor.
Displays and alarms
In order to attain adequate desorption of the mo­lecular sieve, part of the oxygen extracted in one ADS container is continuously passed to the other ADS container via the flushing nozzle.
The oxygen stored in the oxygen tank at an over­pressure of up to 1.7 bar is reduced to 0.6 bar via the pressure-reducer.
The pressure sensor monitors this system pressure. If the pressure falls below 0.5 bar or rises above
2.5 bar, the system activates visual and acoustic alerts.
At an overpressure of around 0.6 bar, the oxygen flows towards the outlet of the appliance.
The prescribed flow is set on the dosage monitor or on the flowmeter. It then flows to the patient via the nasal cannula.
If a leak that is likely to cause a drop in the oxy-
gen concentration occurs after the internal self-
test, this situation is indicated immediately by a visual warning signal (yellow lamp).
If the compressor output falls off, a visual alarm signal is displayed (red lamp flashes) and an acoustic warning (intermittent tone) sounds.
Further information on displays and alarms can be found in Chapter “6.1 Visual and acoustic signals from OXYMAT 3” on page 19“.
The green lamp indicates that the system is op­erational.
In the event of a power failure, an intermittent warning signal will sound and the red lamp will flash. The green lamp is not illuminated.
Functional description 7

3. Servicing

3.1 General

Important! Enter the operating hours of OXYMAT 3 and the oxygen concentration in your service record (see page 78).
The number of operating hours can be read from the operating hours meter 4 on the front of the device.
As a precautionary measure, we recommend that the OXYMAT 3 be serviced once a year or every 5,000 operating hours.
After each servicing, a functional check should be carried out (see "5. Functional check" on page 10).
Note: Dark discoloration of the intake filter 30 does not mean that its flow characteristics are impaired, provided its operating hours do not exceed 5,000 h; an exception applies if the device is op­erated in a very dusty environment.

3.2 Implementation

Servicing the OXYMAT 3 involves changing the coarse dust filter and the suction filter and performing a final check (see "5. Functional check" on page 10).
The filters are located on the back of the appliance.
1. Detach the coarse dust filter 13 from the service flap 38.
13
2. Release the lock on the service flap 38 and remove the latter from the appliance.
When removing the suction filter, please proceed with caution, since the connector and tube may be torn off if the suction filter is handled carelessly.
3. You will find a black clip in the recess to the left of the suction filter. Use your fingers or an object to press against this clip to prevent it from being pulled out as well.
4. Pull the suction filter 30 off the connector 71
without twisting. The suction filter must not be
twisted, as the connector could become detached from the tube.
5. Connect the new filter 30 to the connector.
6. Clean the service flap 38 and insert it.
7. Clip a new coarse dust filter 13 into the service flap 38.
38
71
30
8 Servicing

3.3 Disposal

Do not dispose of the unit with domestic waste. For proper disposal of the device and its compo­nents, please contact a certified waste disposal site for electronic goods. Ask your Environmental Officer or local council for the address. Appliance packaging (cardboard and inserts) may be dis­posed of in the paper recycling bin.

4. Hygienic Preparation

4.1 No change of patient

Proceed as described in the operating instructions for the OXYMAT 3.

4.2 Change of patient or reuse

Dispose of extension tubes and nasal cannu­las, and replace with new parts.
Dispose of coarse dust filter and suction filter, and replace with new parts.
For the bubble humidifier, please observe the separate operating instructions.
Wipe housing and power cord with TERRALIN disinfectant.
Clean inside of housing with vacuum cleaner, clean any particularly dirty parts with house­hold cleaner (to open and close the appli­ance, see sections 7.3 and 7.4 on page 29 and page 30).
Hygienic Preparation 9

5. Functional check

5.1 General

Should you discover any faults or deviations from the set values during your functional check, OXYMAT 3 must not be used again un­til the faults have been rectified.
The possible causes underlying the fault and how to rectify them are outlined in Chapter “6. Troubleshooting” on page 19.
Please use the following equipment to carry out the functional check:
Oxygen measuring device,
Filled oxygen cylinder with adjustable pressure reducer (e.g. HIT,WM 33100),
Respiratory tubes WM 5726 and WM 5772.
For environmental reasons a capacitor back­up unit has been installed instead of a battery to serve the power failure alarm function. As soon as OXYMAT 3 is operational, the ca­pacitor is recharged. Charging takes approxi­mately 5 minutes.
If the device has been out of service for some time, an alarm cannot be triggered until the de­vice has been operational for at least 5 minutes.

5.2 Intervals

A final check must be performed on the appliance:
after every service;
after every repair.

5.3 Implementation

5.3.1 Checking the housing

Check the housing to determine its general condition.
Dirt can be wiped off with a damp cloth. The all-purpose cleaner WM 14937 is also very suitable.
If you want to use a disinfectant, we recom­mend TERRALIN.
10 Functional check
If the housing is defective, replace it (see "7.3 Opening the appliance" on page 29) and „7.4 Closing the appliance“ auf Seite 30.
If the frame is defective, please replace it (see "7.16 Replacing the frame (up to appliance no. 19999)" on page 57).

5.3.2 Checking the coarse dust filter

Check the coarse dust filter 13 to determine its general condition:
1. Unclip the coarse dust filter from the back of OXYMAT 3.
2. If the filter is dirty, clean it. As a general rule, the filter should be freed from dust once a week (either by tapping or vacuuming).
3. If the filter is defective, replace it.
13

5.3.3 Checking the power cord

1. Check the power cord 10.
Make sure that
– the insulation is undamaged, – the cable is undamaged, – none of the contacts are wobbly.
2. If any of these requirements are not satisfied, replace the power cord 10.

5.3.4 Checking the fuse

The fuse holder 15 can be found on the rear of the appliance beneath the power socket 14.
1. Pull the power cord 10 from the socket and from the device.
2. Press the side tabs on the fuse holder 15 to­gether and pull out the fuse holder.
3. Check that the correct fuses are installed. There should always be two fuses installed: A mains fuse and a spare fuse.
The fuses 39 should carry the following label: DIN EN 60127-T 2.5 A H 250 V.
4. If necessary, replace the defective or missing fuses.
5. Push the fuse holder 15 into the housing until it locks home.
15

5.3.5 Checking the main switch and LED displays

1. Connect the power cord 10 to OXYMAT 3 and to a wall socket.
2. Connect the dosage monitor or the flowmeter to the oxygen outlet 5.
Functional check 11
3. Switch on the OXYMAT 3 at the On/Off knob 2. The following LED displays will ap­pear:
– The power indicator (green LED) 20 lights
up.
– The fault indicator (red LED) 18 and status
indicator (yellow LED) 19 light up briefly.
Important
The program then makes an internal check for leaks:
– Pressure is built up within 20 seconds. The
green operating indicator 20 flashes light/ dark.
– Then the compressor is switched off and the
existing pressure is monitored for about 20 seconds. If there is a fault and the pressure drops, the yellow status indicator 19 lights up.
– Finally the system is vented so that the com-
pressor can start up again without any back-pressure.

5.3.6 Checking the dosage monitor for leaks

19
20
O/I
Oxygen appliance!
Increased fire risk!
O/ I
2
O
2
18
1. Check the screw connections between the hu­midifier glass/cap and humidifier/dosage monitor to verify that they are firmly seated. If necessary, re-tighten the connections by hand.
2. Set a flow of 4 l/min on the flow control knob 24 for oxygen dosage.
3. To check the dosage monitor 9 for leaks,
– bend the 3 m tube 11 – or pull the 3 m tube 11 from the connection
of the bubble humidifier 25 and hold the
connection 26 closed. If the ball 22 sinks all the way to the bottom, there are no leaks in the dosage monitor sys­tem.
4. In the event of leaks, change the sealing rings on the humidifer.

5.3.7 Checking the flowmeter for leaks

1. Set a flow of 4 l/min on the knob 27.
2. To check the flowmeter 16 for leaks:
– Bend the connection tube 17 – or pull the tube 17 from the connection of
the flowmeter and hold the connection 29
closed. If the ball 22 sinks all the way to the bottom, there are no leaks in the flowmeter.
3. In the event of leaks, change the sealing rings on the flowmeter.
26
25
0
,
5
0
,
4
0
,
3
0
,
2
0
,
1
5
,
0
22
9
24
16
22
27
29
17
12 Functional check

5.3.8 Checking the operating hours meter

Check that the red minute counter on the operating hours meter 4 is incremented every 6 minutes.

5.3.9 Checking the O2 concentration

Leave the OXYMAT 3 to run for at least 10 minutes before starting the measurement.
Before you check the oxygen concentration on the OXYMAT 3, you must first calibrate the oxygen measuring device (e.g. OXYcontrol WM 13550).
For this purpose, you will need the respiratory tube WM 5726 and an oxygen cylinder with an ad­justable pressure reducer (e.g. WM 33100).
Caution ! Please observe the instructions for handling oxygen!
Important!
Before calibrating the oxygen sensor for the first time, the sensor must be stored in ambient air for at least 15 minutes (refer to instructions for use of the sensor).
Observe instructions for use!
4
000000
How to calibrate the oxygen sensor:
1. Connect the oxygen measuring device to the pressure reducer of the oxygen cylinder using the respiratory tube WM 5726.
2. Slowly open the cylinder valve. One revolution is sufficient!
3. Set a flow of 4 l/min at the pressure reducer.
4. Leave the oxygen measuring device (e.g. OXYcontrol) flowing for approximately 2 minutes,
until the displayed oxygen level remains steady.
5. Set a value of "100" using the calibration knob.
6. Close the cylinder valve again.
The oxygen measuring device is now calibrated and you can check the oxygen concentration on OXYMAT 3.
Please remember that OXYMAT 3 must have already been operational for at least 10 minutes.
When operating with a dosage monitor:
Important! It is vital that you empty the humidifier glass 25 of the bubble humidifier; otherwise, the oxygen sensor of the oxygen measuring device will be damaged.
7. Connect the oxygen measuring device:
– To do so, remove connecting tube 11 from
connection 26 of the dosage monitor and use breathing tube WM 5772 to make a connection between connection 26 and the oxygen measuring device;
or
– Unscrew the bubble humidifer (sterile water
system) from the dosage monitor and attach the connecting tube to the OXYcontrol at this point with the union nut.
8. Using the flow control knob 24, set a flow of 4 l/min.
11
26
24
Functional check 13
9. After waiting for a short while, read the oxy­gen concentration from the oxygen measuring device.
It should be between 90 and 95 % by volume.
10. Re-connect the connecting tube 11.
11. If the O2 concentration is outside the tolerance range, proceed as described in “6.3 O2 con­centration outside tolerance range” on page 27.
When operating with a flowmeter:
7. Connect the oxygen measuring device:
– To do this, remove connecting tube 17 from
connection 29 of the flowmeter and use breathing tube WM 5772 to make a con­nection between connection 29 and the oxygen measuring device;
or
– Unscrew the union nut from the flowmeter
and attach the connecting tube to the OXYcontrol at this point with the union nut.
8. Using the flow control knob 27 set a flow of 4 l/min.
9. After waiting for a short while, read the oxy­gen concentration from the oxygen measuring device.
27
29
17
It should be between 90 and 95 % by volume.
10. Re-connect the connecting tube 17.
11. If the O2 concentration is outside the tolerance range, proceed as described in “6.3 O2 con­centration outside tolerance range” on page 27.

5.3.10 Checking the mains failure alarm

Note: As power is supplied to the alarm function via a capacitor, the OXYMAT 3 should first be operated for 5 minutes to allow the capacitor to charge.
1. Pull the power cord from the socket or from the device.
2. Check whether the following displays and alarms occur:
Acoustic: The device emits a continuous
tone.
Optical: The red fault indicator 18 lights up;
the green power indicator 20 is off.
3. Re-make the connection to the mains.
Now:
– The acoustic alarm should be silenced. – The red fault indicator 18 should go out.
O
20 18
O/I
Oxygen appliance!
Increased fire risk!
– The green power indicator 20 should be
on.
– The appliance should start up.
O/ I
2
14 Functional check
4. Switch off the appliance at the On/Off knob 2.

5.3.11 Checking the indirect status indicator (yellow LED) with appliance closed

The purpose of this check is to establish whether the status indicator that warns if a leak that is likely to cause a drop in the oxygen concentration is working properly. This is done by an internal program that creates an artificial leak which results in the alarm being triggered.
1. Switch on the OXYMAT 3 at the On/Off switch 2. An internal leak test will run for about 1 minute.
19
The green operating indicator 20 flashes while this is in progress.
20
2. With the appliance running, set a flow of 4 l/min at the top of the ball.
To start the test program, proceed as follows:
3. Remove power cord 10 from the wall socket or the appliance.
4. Wait for the acoustic power failure alarm to sound (two groups of five tones – – – – –).
5. Switch off the OXYMAT 3 at the On/Off switch 2. Do not connect the power cord again.
6. Switch OXYMAT 3 on again at On/Off switch 2.
11. Check that the yellow status indicator 19 is on while this is happening.
If the status indicator does not light up, either: – you have not set the flow to exactly 4 l/min or – the status indicator is faulty.
12. Switch off the OXYMAT 3 at the On/Off switch 2, then switch it on again.
The test program is now deactivated and the OXYMAT 3 is in normal operating mode.
7. Wait for the acoustic power failure alarm to sound again (two groups of five tones – – – – –).
8. Switch off the OXYMAT 3 at the On/Off switch 2.
The yellow status indicator 19 must now light up. If it does not, repeat the procedure.
9. Connect the power cord again.
10. Switch on the OXYMAT 3 at the On/Off switch 2.
The following LED indicators appear: – green operating indicator 20 lights up – red fault indicator 18 lights up briefly – yellow status indicator 19 lights up briefly.
O/I
Oxygen appliance!
Increased fire risk!
O/ I
2
O
2
18
The program then runs an internal leak check routine: – pressure is built up within 20 s. The green
operating indicator 20 flashes.
– the pressure is checked for about 20 s with
the compressor switched off.
An artificial leak is created in the OXYMAT 3, and this results in the oxygen concentration falling be­low 82 %.
Functional check 15

5.3.12 Checking the indirect status indicator (yellow LED) with the appliance open

Note: Not necessary as part of a normal functional check or service.
The purpose of this check is to establish whether the status indicator that warns if a leak that is likely to cause a drop in the oxygen concentration is working properly. This is done by an internal program that creates an artificial leak which results in the alarm being triggered. If the appliance is open (e.g. for repair work) you can start this test program with switch S5 on the printed circuit board.
1. Switch on the OXYMAT 3 at the On/Off switch 2. An internal leak test will run for about 1 minute.
19
The green operating indicator 20 flashes while this is in progress.
20
2. With the appliance running, set a flow of 4 l/min at the top of the ball.
To start the test program, proceed as follows:
3. Switch off the OXYMAT 3 at the On/Off switch 2.
4. Open coding switch S5 on the printed circuit board.
5. Switch on the OXYMAT 3 at the On/Off switch 2.
The following LED indicators appear:
– green operating indicator 20 lights up – red fault indicator 18 lights up briefly
O
O/I
– yellow status indicator 19 lights up briefly.
The program then runs an internal leak check routine: – pressure is built up within 20 s. The green
operating indicator 20 flashes.
– the pressure is checked for about 20 s with
the compressor switched off.
O/I
Oxygen appliance!
Increased fire risk!
O/I
2
O/ I
2
O
2
18
S5
An artificial leak is created in the OXYMAT 3, and this results in the oxygen concentration falling be­low 82 %.
6. Check that the yellow status indicator 19 is on while this is happening.
If the status indicator does not light up, either: – you have not set the flow to exactly 4 l/min or – the status indicator is faulty.
7. Switch off the OXYMAT 3 at the On/Off switch 2.
8. Close coding switch S5 on the printed circuit board.
The test program is now deactivated and the OXYMAT 3 is in normal operating mode.
9. Close the appliance (see "7.4 Closing the appliance" on page 30).
16 Functional check

5.3.13 System pressure test

General
The system pressure should only be tested if there is a fault in the appliance.
Fitting the test pressure gauge
1. Connect the test pressure gauge inlet directly to the oxygen outlet 5.
If your pressure gauge also has an outlet, close it (e.g. with your thumb).
2. Open the appliance housing (see "7.3 Open­ing the appliance" on page 29).
3. Open the strap system 42 without pulling the strap out of its buckle.
4. Carefully remove foam section 44.
5. To prevent the ADS containers from slipping out, lay the strap system 42 over the remaining foam sections and gently tighten it.
44
I / O
2
O
/I
O
O
2
0000 0
0
5
To make fitting easier, the test pressure gauge is not fitted directly to the inlet of the O2 tank, but before the 6 mm diameter tube 56 which leads directly to the O2 tank.
6. At the T-connector 74, push back the sleeve of the elbow connector and remove the tube 56.
7. Push the tube of the test pressure gauge firmly into the elbow connector.
8. Push the tube 56 into the free inlet of the test pressure gauge.
Pressure test
Oxygen outlet
1. Plug the power cord 10 into the appliance and insert the mains plug in a wall socket.
2. Switch the OXYMAT 3 on.
3. After 10 minutes, attach the pressure gauge to the oxygen outlet 5.
4. Read off the pressure.
The reading must be in the range
0.6 bar ± 0.1 bar (at sea level).
5. Rectify any discrepancies you find (see "6.2 Remedies" on page 22).
42
56
74
6. Switch the appliance off, then remove the mains plug from the wall socket and the power cord from the appliance.
7. Remove the pressure gauge.
8. Perform a functional check (see "5. Functional check" on page 10).
Functional check 17
O2 tank inlet
9. Attach the dosage unit 21 or the flowmeter 16.
10. Plug the power cord 10 into the appliance and insert the mains plug in a wall socket.
11. Switch the OXYMAT 3 on.
Caution! Risk of injury or death! Beware of live or moving components.
12. Set the flow to 4 l/min.
13. After 10 minutes, read off the pressure on the pressure gauge.
The reading must be in the following range:
At the O2 tank inlet:
min. 1.2 + 0.3/- 0.2 bar to max. 1.65 bar + 0.3/- 0.2 bar
at 4 l/min (at sea level).
Rectify any discrepancies you find(siehe „6.2 Remedies“ auf Seite 22).

5.3.14 Checking the O2 system for leaks

14. Switch the appliance off, then remove the mains plug from the wall socket and the power cord from the appliance.
15. Remove the dosage unit 21 or flowmeter 16 from the appliance.
16. Remove the pressure gauge.
17. Attach the tube 56.
18. Slacken the strap system.
19. Carefully insert the foam section 44 and tight­en the straps again.
20. Close the appliance (see "7.4 Closing the appliance" on page 30).
21. Perform a functional check (see "5. Functional check" on page 10).
General
A check for leaks should only be made if there is a problem with the appliance.
Method
Soap all connections between the compressor exit, the pressure equalization system (tubes, exit) and the shutoff valve after the pressure reducer.
18 Functional check

6. Troubleshooting

6.1 Visual and acoustic signals from OXYMAT 3

6.1.1 Alarms

The following appliance states trigger an alarm:
1. Power failure
2. Overheating
3. Memory error
4. Under-pressure
5. Over-pressure
6. Implausible pressure measurement
7. Leak in O2 system
1. Power failure
The energy for a power failure alarm is supplied by a capacitor back-up unit. If the capacitor back-up unit is fully discharged, the power failure alarm is not available for the first minute after switching on the OXYMAT 3 until the back-up unit has charged up again.
2. Overheating
A temperature sensor is mounted on the printed cir­cuit board. If the temperature rises above 60 °C (± 2.5 °C), an alarm is given and at the same time the compressor is switched off. The compressor re­mains switched off as long as the temperature alarm is activated. The cause may for example be a malfunction of the fan.
3. Memory error
Every time the appliance is switched on, the EEPROM is first read out. Among other things, this contains the operating cycles for the valves and the calibration data for the temperature sensor. A plau­sibility check is performed on the data read out.
If a memory error occurs:
– default values (basic settings) are read from
the program memory and used to start the
control program; – the memory error alarm is activated; – after every alarm cycle a fresh attempt is
made to read the EEPROM; – after a successful read attempt the alarm is
terminated.
4. Under-pressure
This alarm is activated if the pressure sensor at the inlet to the O2 system measures a pressure lower than the required operating pressure.
5. Over-pressure
This alarm is activated if the pressure sensor at the inlet to the O2 system measures a pressure higher than the required operating pressure.
6. Implausible pressure measurement
Since the OXYMAT 3 has a time control, there must be a defined pressure pattern over time. This pressure pattern is measured by the pressure sensor and analyzed by the appliance software. If necessary an alarm is activated.
7. Leak in O2 system
Every time the appliance is switched on, the con­trol system runs a leak test on the entire O2 system. If a leak is found, the yellow status indicator 19 is switched on, as sooner or later this leak in the O2 system may result in the oxygen concentration at the outlet falling below 82 %.
During the leak test the green power indicator 20 flashes.After the leak test the OXYMAT 3 switches to normal operating mode.
If the leak test is not successfully completed within one minute, the control system stops the test cycle and switches on the yellow status indicator 19 to indicate an operating fault. The cause of the leak must now be identified.
No further test for leaks is performed during normal operation.
Troubleshooting 19

6.1.2 Signal patterns

The alarm signal patterns are in accordance with EN 475 (see Table of Signal Patterns). To make it possible to distinguish the alarm sources, this standard assigns specific signal patterns to individual error situations (see Table of Error Situations).
Since it is possible that two or more errors may occur at the same time, the alarms are weighted by the software and the appropriate alarm pattern is output. For example: if under-pressure and overheating occur simultaneously, the alarm pattern for overheating is output (see Table of Alarm Situations).
Table of Alarm Situations:
Appliance Status
Power failure
Overheating
Implausible pressure reading
Memory error
Under-pressure
Over-pressure
O2 status
Release after
x minutes
1 minute after cold start
Immediately after switching on
1 minute after end of leak test
Immediately after switching on
1 minute after end of leak test
1 minute after end of leak test
Test immediately after switching on
Pattern of alarm
signal indicators
Buzzer: A Red LED: A Yellow LED: off Green LED: off
Buzzer: A Red LED: A Yellow LED: off Green LED: on
Buzzer: A Red LED: C Yellow LED: off Green LED: on
Buzzer: B Red LED: B Yellow LED: off Green LED: on
Buzzer: C Red LED: C Yellow LED: off Green LED: on
Buzzer: C Red LED: D Yellow LED: off Green LED: on
Buzzer: off Red LED: off Yellow LED: on Green LED: on
Classification of
Weighting
1 High priority
1 High priority
2 High priority
3 High priority
4 High priority
5 High priority
Is indicated independently by the yellow LED
signal patterns
under EN 475
Low priority
Note: The green power indicator 20 is not an alarm signal indicator. Its status is only given here for completeness’ sake.
20 Troubleshooting
The signal patterns for the buzzer consist of 2 identical pulse sequences, repeated every 10 seconds. Each pulse sequence consists of 5 individual pulses (see Table of Signal Patterns).
Table of Signal Patterns:
Buzzer pattern
for one pulse
sequence
Pulse duration / ms
(= buzzer on)
Pulse interval/ ms
(= buzzer off)
Pulse interval
between 3rd and
4th pulse / ms
Graphic representation
A 150 100 500 — — — — —
B 150 10 320 — — — — —
C 150 0 300 ——— ——
Flashing pattern
for
red LED
Flashing
frequency / Hz
On time / ms Off time / ms Graphic representation
A 1.43 200 500 — —
B 1.43 400 300 —— —— C 2.5 80 320 – – D 2.5 240 160 — —

6.1.3 Deactivating the signal functions

The signal functions can be deactivated by means of the switches S2, S3, S4 and S5 on the printed circuit board. To see where the switches are located on the PCB, consult the illustrations in Section “10.1 Arrange­ment of the PCBs” on page 76.
Coding switch Open Closed
S2
S3
Power supply for power failure alarm deactivated.
Buzzer deactivated. Buzzer activated.
Alarm suppression during warming-up
S4
phase deactivated (i.e. alarm function activated).
S5
Tip:
Activates program for checking status indicator
Power supply for power failure alarm activated.
Alarm suppression during warming-up phase activated (i.e. alarm function deactivated).
Normal operation
For fault-finding purposes it can be useful to open coding switch S4. This enables the alarms immediately on completion of the leak test.
Troubleshooting 21

6.2 Remedies

Fault Cause Localisation Remedy
No motor noises, power indicator does not light up, mains failure alarm sounds.
No power.
Faulty fuse.
Check mains lead for secure connection. If applicable, check whether power is available at the mains (e.g. with a lamp).
Replace fuse 39 (see "5.3.4 Checking the fuse" on page 11). A spare fuse can be found in the fuse box.
Power indicator fails to illuminate, no acoustic alarm.
No flow at nasal cannula, although there is flow at the appliance outlet. In such cases, the appliance does not emit an alarm.
After running for a long period, appliance emits intermittent acoustic warnings and red fault indicator flashes; no flow.
After running for a long period, appliance emits intermittent acoustic warnings and red fault indicator flashes.
Power cord damaged. Replace power cord.
Capacitor for mains failure alarm is completely discharged.
Tube connections interrupted.
Oxygen distributor in bubble humidifier blocked.
Motor compressor not rotating.
Crank mechanism stiff, motor circuit-breaker of compressor is activated.
Appliance too hot, coarse dust filter blocked.
Leak at compressor outlet port elbow connector
Suction filter blocked.
Overtemperature of 60 °C ± 2.5 °C.
Disconnect tube from inlet of dosage monitor; check whether flow is escaping from the tube.
Loosen union nut of bubble humidifier; the ball should then rise.
Leave the appliance to run for 5 minutes.
Seal all connection and sealing points from the appliance outlet to the inlet of the dosage monitor.
Clean or replace the oxygen distributor.
Capacitor incorrectly connected or not connected at all. Connect the capacitor correctly (7.5, page 30).
Check the voltage supply at the motor. Motor defective: Replace the compressor (7.5, page 30).
Compressor defective: Replace compressor (7.5, page 30).
Clean or replace coarse dust filter. (3.2, page 8).
Replace parts (7.5, page 30)
Replace suction filter. (3.2, page 8).
Check fan (7.10, page 42).
22 Troubleshooting
The following faults are based on a defective circuit. They occur with the compressor and fan running.
Caution ! Always pull out the mains plug before working on electrical components.
Remember that all coding switches on the circuit board must be closed during normal operation.
Fault Cause Localisation Remedy
After switching on, first of all
After running for a long period, appliance emits intermittent acoustic warnings and red fault indicator flashes.
Memory failure.
the EEPROM is read. In the event of a memory failure, an alarm is issued and it is then re-read. If this readout attempt is successful, the fault will be cleared. If not, the circuit board should be replaced. (7.9, page 40).
Blow-off noises on compressor (overpressure valve activated).
Connectors X10 and/ or X11 have worked loose.
Inadequate voltage supply to magnetic valves.
No voltage supply to magnetic valves.
Fuse F1 tripped.
Green LED is illuminated and an acoustic pressure alarm sounds.
See above.
See above.
Green LED not illuminated, mains failure alarm.
Re-make the connection (7.9, page 40).
Ensure that the mains voltage is at least 207 V.
Replace fuse F2 (7.9, page 40).
Replace fuse F1 (7.9, page 40).
Overpressure alarm (acoustic alarm, red LED 18 flashing
2.5 times per second).
No overpressure
alarm, even though overpressure exists.
Underpressure alarm (acoustic alarm, red LED 18 flashes briefly 2.5 times per second).
Thermal release in transformer tripped.
Overpressure exists.
Valves of O2 system jamming
Circuit board defective.
Warm-up phase not yet complete.
Pressure measurement tube faulty.
Circuit board defective.
Underpressure exists.
Circuit board defective.
Pressure measurement tube has come loose.
Green LED not illuminated, mains failure alarm.
Check pneumatic system for blockages.
Perform a system pressure test (5.3.13, page 17)
Each time the device is switched on, the overpressure alarm is blocked for one minute.
Visual check on tube for: properly seated connection; not blocked.
Check pneumatic system for leaks.
Perform a system pressure test (5.3.13, page 17)
Visual check for proper seating of pressure measurement tube.
Replace the circuit board (7.9, page 40).
Rectify blockage (7.9, page 40).
(see " Blow-off noises on compressor (overpressure valve activated)." on page 23)
Replace the circuit board (7.9, page 40).
Wait for one minute.
Replace pressure measure­ment tube if necessary.
Replace the circuit board (7.9, page 40).
Remedy the leak.
Replace circuit board
Establish trouble-free pneumatic connection.
Troubleshooting 23
Fault Cause Localisation Remedy
Each time the appliance is
No underpressure alarm.
Warm-up phase not yet complete.
Circuit board defective.
Miniature fuse in non­heating appliance connector has tripped.
switched on, the overpressure alarm is blocked for one minute.
Compressor and fan not running and green LED is extinguished.
Wait for one minute.
Replace the circuit board (7.9, page 40).
Check fuse 39 and replace if necessary (5.3.4, page 11).
Internal connection has worked loose at non­heating appliance
See above.
Check connection and re­make if necessary.
connector.
Mains failure alarm
Connector X1 has worked loose.
Cable harness defective.
Fuse F1 has tripped.
See above.
See above.
Compressor and fan not running and green LED is extinguished.
Check connection and re­make if necessary.
Replace the cable harness (item no. 86 in the spare parts list).
Check fuse F1 and replace if necessary (7.9, page 40) (10.1, page 76).
Device fails to emit correct mains failure alarm.
Thermal release in trans­former has tripped.
Circuit board defective.
Coding switch S3 is open.
Coding switch S2 is open.
Voltage supply for mains failure alarm discharged.
Mains failure alarm is
shorter than 1 minute.
See above.
Green LED illuminated, compressor and fan running. Valves are switching.
Green LED is extinguished. Buzzer off. Red LED is flashing.
Green LED is extinguished. Buzzer off. Red LED is extinguished.
See above.
Operate the appliance for 30 minutes. If the alarm duration is then shorter than 1 minute, the circuit board is defective.
Replace the circuit board (7.9, page 40).
Replace the circuit board (7.9, page 40).
Close coding switch S3 (7.9, page 40).
Close coding switch S2 (7.9, page 40).
Operate the device for one minute to recharge the voltage supply.
Replace the circuit board (7.9, page 40).
24 Troubleshooting
Fault Cause Localisation Remedy
Leak in O2 system
Switch S5 for checking status indicator was not closed again (5.3.12, page 16)
Examine pneumatic system for leaks.
Switch off OXYMAT 3. Close switch S5. Switch on OXYMAT 3.
Eliminate leaks.
Close switch S5.
During a power failure
O2 alarm
alarm the OXYMAT 3 was switched off and on twice, thereby starting a status indicator check (5.3.11, page 15)
Check plug-in connection X13.
Connect OXYMAT 3 to mains. Then switch OXYMAT 3 off and on again.
Ensure trouble-free plug-in connection.
No voltage at shutoff valve.
Check fuse F3.
Change fuse F3 (7.9, page 40), (10.1, page 76).
Shutoff valve 83 faulty.
Pressure measurement tube 63 blocked.
Flow starts immediately after switching on.
Self-test is stopped after 60 s without stopping compressor.
Change shutoff valve 83.
Re-route tube 63, replace if necessary.
No O2 alarm, even though concentration is < 82 % by volume.
Compressor not running
Temperature alarm, fan not running.
Pressure sensor B1 faulty.
Leak occurred since last switching on.
Pressure sensor faulty.
Compressor overpressure valves activate after 30 s.
Switch OXYMAT 3 off then on again to initiate self-test.
Check indirect status indicator (5.3.11, page 15) (5.3.12, page 16).
Change board (7.9, page 40).
Eliminate leak.
If yellow LED does not light up, change board (7.9, page 40).
OXYMAT 3 was oper­ated for long time at relative humidity in
Weigh ADS container (7.11, 7.12 and 7.13).
Change ADS container (7.11, 7.12 and 7.13).
excess of 75 %.
Fuse F3 defective. Green LED is on. Replace fuse F3.
Suction filter blocked.
Plug-in connection X2 faulty.
Check plug-in connection
Clean or replace suction filter (3.2, page 8).
Replace wiring harness if necessary.
Allow OXYMAT 3 to cool
Overheating alarm.
Appliance control system switches off compressor if overheating alarm given.
down. After the end of the overheating alarm the con­trol system switches the OXYMAT 3 on again.
Board faulty.
Change board (7.9, page 40).
Check connection at fan. If
Fan failure.
No draught can be felt at the coarse dust filter, compressor is running.
necessary, replace cable harness (item no. 86 in spare parts list) or fan (7.10, page 42).
Troubleshooting 25
Fault Cause Localisation Remedy
Comply with maximum
Temperature alarm although fan is running.
Appliance was operated at an ambient temperature of > 41 °C, or was switched off briefly and back on.
Accumulated heat warmed the temperature sensor after switching off.
ambient temperature of 40 °C. After operating for a short time, the accumulated heat is eliminated by the fan and the temperature alarm must stop.
Ambient temperature is < 40 °C, fan and compressor working correctly.
Visual inspection to verify correct seating of the connector.
Install a replacement meter as a test. If it moves forward by
00000.1 after 6 minutes, the old meter was defective.
If the above measures have failed to solve the problem.
Replace the circuit board (7.9, page 40).
Re-make the connection (7.9, page 40).
Replace the operating hours meter (7.8, page 39).
Replace the circuit board (7.9, page 40).
Operating hours meter no longer counting.
Circuit board defective.
Connector at X9 has worked loose.
Meter is defective.
Circuit board defective.
26 Troubleshooting
Loading...
+ 58 hidden pages