Weinmann Oxymat 2 User manual

Instrument Description and Operating Instructions
OXYMAT2
Its own oxygen source
OXYMAT2
Fig. 1
1
15
5 6
2
8
15
7
13
3
11
1 OXYMAT2 oxygen concentrator 2 Oxygen outlet 3 Main switch 4 Running time meter 5 On indicator (green) 6 Fault indicator (red) 7 Mains lead 8 Connecting hose, 20m
9 Flow monitor 10 Oxygen dosage control knob 11 Bubbler humidifier 12 Angled hose connection 13 Extension hose for oxygen nasal cannula, 3m 14 Oxygen nasal cannula 15 Air intake slots
4
9
12
10
14
Contents
1 Instrument Description 2
1.1 Purpose 2
1.2 Function 2
1.3 Circuit diagram OXYMAT
2 Technical Data 4 3 Extent of Supply 5 4 Accessories and Spare Parts 5 5 Use 5
5.1 Setting up the instrument 5
5.2 Putting into use 6
5.3 After use 6
6 Cleaning and Disinfection 7 7 Checking the Function 8 8 Maintenance 8 9 Safety Precautions 9 10 Guarantee 10 11 Trouble-Shooting 10
2 oxygen concentrator 3
Page
1 Instrument Description
1.1 Purpose
The OXYMAT2 oxygen concentrator is used to increase the percentage of the oxygen volume in the inspired air.
Oxygen therapy is a clear indication in cardiac circulation disturbances and diseases of the respiratory organs. It is the most common prophylactic or therapeutic measure in cases of existing or impending hypoxia.
Where patients who require continuous administration of a higher oxygen content in the inspired air are treated with the OXYMAT made for emergency supply from an oxygen cylinder in the case of instrument failure (e.g. as a result of power failure).
1.2 Function
An electrically operated compressor sucks-in ambient air through dust- and fine-mesh filters and forces the slightly compressed air into containers fitted with molecular filters.
Molecular filters work on the principle of absorption of the nitrogen present in the air.
The oxygen-enriched air flows from the containers into the oxygen tank, from which it passes through a pressure controller and bacterial filter to the oxygen outlet of the OXYMAT
The rate of flow required can be set on the oxygen dosage control knob and read-off on the top of the flowmeter cone (see circuit diagram in Fig. 2).
2 oxygen concentrator, provision must always be
2 oxygen concentrator.
2
1.3 Circuit diagram
OXYMAT
2 oxygen concentrator
Fig. 2
Flow monitor with oxygen nasal cannula, bubbler humidifier, flowmeter
Oxygen outlet Bacterial filter
Pressure controller
Oxygen tank
Nozzle
Exhaust air
Ambient air
III
Non-return valves
Molecular filters I and II
Cooler
Compressor
Fine-dust filter
Dust filter
3
2 Technical Data
G
e
p
r
ü
f
t
e
r
L
ä
r
m
s
c
h
u
t
z
G
e
r
ä
u
s
c
h
a
r
m
0197
Classification acc. to EU-Guideline 93/42/EEC: II a
Dimensions: see Dimension Drawing (Fig. 3),
Weight: app. 19.5 kg Ambient temperature for operation: +10°C to +40°C for storage: –20°C to +70°C Electrical connection: 230 V ~, 50Hz Wattage and current consumption: 250 W, 1.4 A Fuse: T 1.6 AL 250V acc. IEC 127-2 Classification acc. to EN 60601-1 Type of electr. shock protection: Class II Level of electr. shock protection: Type B Noise level: 50 dB (A) acc. ISO 3743
delivery: 95 ± 2 vol.% O2 to 2 l/min flow
O
2
(values after 10 minutes 93 ± 2 vol.% O operation, at +20°C and min. 70 vol.% O 50% relative humidity and 1013 mbar) Flow pressure nasal cannula: 0,05 bar at 4 l/min flow
Flow check: 0,4 – 4 l/min,
Fault warning: visible and audible Electro magnetic compatibility (EMC) Interference suppression: EN 55011 (VDE 0875 T.11) Interference resistance: IEC 801 part 2 to part 5 Maintenance: every 5.000 running hours
345
Fig. 3
dimensions in mm
to 3 l/min flow
2
to 4 l/min flow
2
surface flowmeter ball
(Subject to changes in construction)
590
4
505
230
286
3 Extent of Supply
OXYMAT2 oxygen concentrator, complete WM 12300 Components: OXYMAT2 oxygen concentrator WM 12301 Flow monitor, complete WM 12250 Extension hose for oxygen nasal cannula, 3 m WM 5720 Oxygen nasal cannula, double outlet, with ear clips and connecting hose WM 1929 Oxygen nasal cannula, single outlet, with silicone seal and connecting hose WM 1960 Connecting hose, 20 m WM 12330
4 Accessories and Spare Parts
Bubbler humidifier with angled hose connection WM 1375 Angled hose connection WM 1305 Oxygen nasal cannula, double outlet, with ear clips and connecting hose WM 1929 Oxygen nasal cannula, single outlet, with silicone seal and connecting hose WM 1960 Connecting hose, 20 m WM 12330 Extension hose for oxygen nasal cannula, 3 m WM 5720 Oxygen flow monitor without humidifier WM 12255 Gasket (for wing nut) WM 1145/31 Gasket for bubbler humidifier WM 1145/93 Transport-caddy WM 12305
WEINMANN:HAMBURG supplies portable oxygen therapy units (e.g. OXYTRON supply outside the home.
2 Mobile) as accessories and for oxygen
5 Use
5.1 Setting up the instrument
The OXYMAT2 oxygen concentrator should be located in a fixed, not humid and not disturbing position, e.g. in a corridor. We recommend placing it on a smooth floor, not on a deep-pile carpet or rug, which might block the air circulation. It must not be placed too close to a wall (at least 15 cm away, see page 9 Fig. 9).
Fit the 20m connecting hose (do not confuse with the 3 m extension hose) at the oxygen outlet of OXYMAT2 (see Fig. 4) and at the flowmeter-inlet of the flow monitor as well (see Fig. 5).
5
Fig. 4
Fig. 5
5.2 Putting into use
If the instrument prior use has been located in
environmental conditions below +10°C, wait for warming of at least +10°C room temperature.
Check that the mains voltage agrees with that given on the
instrument (see rating plate, Fig. 6).
Push the mains lead connector into the instrument (Fig. 6)
and the mains plug into the mains socket.
Unscrew the bubbler humidifier from the flow monitor, fill up
to the mark with distilled water (Aqua Dest.) and screw on again.
Fit the 3-m extension hose (WM5720) to the oxygen
nasal cannula and to the angled socket of the bubbler humidifier (Fig. 7), ensuring that the connections are tight.
Fig. 6
Rating plate
Switch on the instrument (green "On" light comes on
permanently, red fault light comes on briefly and an audible signal sounds.
Set the flow required on the oxygen dosage control knob
(see Fig. 8).
6
Fig. 7
The instrument is now ready
for use and the oxygen nasal cannula can be put on.
Fig. 8
Top of cone
5.3 After use
Switch off the instrument by pressing the main switch. The flow setting can be left as it is. If no further therapy is to be carried out that day, empty the bubbler humidifier.
6 Cleaning and Disinfection
The oxygen nasal cannula should never be used by more than one patient. The bubbler humidifier can be autoclaved and sterilised at 134°C. We recommend cleaning the oxygen goggles, hoses, etc. under running water after each use.
The instrument casing and mains lead are kept clean simply by wiping with commercial disinfectants (observing their Instructions of Use). Do not allow water or disinfectant to penetrate into the air intake slots.
The spray humidifier, oxygen nasal cannula and hoses should be changed before use on a new patient.
For health reasons, the bubbler humidifier should be cleaned daily and the distilled water (Aqua dest.) regularly changed. In home therapy, where the bubbler humidifier is used by only one patient, it is sufficient to use a disinfectant, rinsing the parts with distilled water after disinfection and drying them.
The air intake slots of the OXYMAT should be freed from dust every week using a soft brush or the flexible tube of a vacuum cleaner. Do not probe into the air intake slots (Fig. 9) with hard objects.
2 oxygen concentrator
7
7 Checking the Function
A function check must be carried out twice a year.
Switch on the OXYMAT
Following: – the green "On" light comes on – the red fault light comes on briefly – a short audible signal sounds.
The red minute counter of the running time meter must
move round every 6 minutes.
Check that the humidifier bottle cap and humidifier/flow
monitor connections are tight, retightening by hand if necessary.
Set a flow of 2 I/min. on the oxygen dosage control knob
(Fig. 8).
To check that the dosage monitor is leak-tight, remove the
hose from the angled coupling of the bubbler humidifier and seal-off the socket.
The flow monitor system is tight if the flowmeter cone drops
down completely. Should a leak be present, change the gaskets on the humidifier.
Check the mains power failure alarm by removing the
mains plug. A continuous audible signal should sound.
Switch the instrument off at the mains switch.
2 oxygen concentrator.
8 Maintenance
In addition to the function check, we recommend that the OXYMAT2 oxygen concentrator is overhauled every 5,000 running hours (see running time meter). Once this is achieved maintenance has to be carried out by specific shilled and authorized personnel or return to the manufacturer WEINMANN:HAMBURG.
For safety reasons, a check should be made every three months to see whether the instrument has reached 5,000 running hours.
Do not despatch the hose systems and oxygen nasal cannula with the instrument.
N.B. Please store the original carton, since we recommend the instrument should be packed in for transport.
8
9 Safety Precautions
For your and your patient’s safety! To conform with EU-Guidelines (93/42/EEC), we have to point out the following:
Observe the Operating lnstructions which is a component
of the instrument.
Any use of or intervention with the instrument requires
accurate knowledge and observation of these Operating Instructions. The instrument is intended only for the purpose described.
With exchange of accessories and spare parts recom-
mended in this operating instructions by using foreign items, the function as well as the O
The OXYMAT2 oxygen concentrator produces oxygen.
Items of clothing which come into contact with oxygen­enriched air represent an increased fire risk! For this reason, smoking, naked lights and fires are strictly prohibited in the vicinity, even after therapy, until the high oxygen concentra­tion in the clothing has dissipated .
Check the mains voltage.
Keep the instrument and the hoses free from grease and oil.
Use the carrying handle to transport the instrument (Fig. 9).
The left-hand side of the instrument with the air intake slots
(top and bottom) must be at least 15 cm away from the wall. The major part the instrument must not be located close to a heat source (Fig. 9).
Air intake slots
weakly to clean.
When red alarm light
and intermittent sound comes up, switch of the instrument.
compatibility may be reduced.
2
Fig. 9
Carrying handle
9
10 Guarantee
The instrument is guaranteed for one year from the date of delivery for defects arising from material or manufacturing faults.
Defects covered by the Guarantee will be rectified under the terms of our Guarantee Conditions. Plastic, glass and rubber parts are not covered by the Guarantee. Transport to and from our works for repair is at the cost of the customer.
WEINMANN:HAMBURG offers no guarantee when the user has jeopardised the function of the instrument through non­observation of these Operating Instructions, incorrect use or handling or unauthorised interference with the instrument. In such instances, responsibility passes to the user.
Important:
The guarantee is only valid when the purchase invoice is produced.
11 Trouble-Shooting
Fault
No motor noise, operating light does not come on, power failure alarm sounds (continuous alarm).
Instrument running, but inter­mitted audible alarm and flashing fault light.
After prolonged use, instrument emits audible alarm, fault light comes on, no motor noise.
10
Cause
No mains power.
Faulty fuse.
Instrument over­heated. Air intake slots blocked or covered.
Motor not running.
Remedy
Check mains lead connection. If necessary check that socket has mains power (e.g. plug in a lamp).
Remove mains lead plug from socket and disconnect mains lead from instrument! Change fuse (T 1.6AL 250V acc. IEC 127-2) by squeezing together the two tongues of the fuse cartridge, removing and replacing the cartridge (Fig. 10).
Clean air intake slots. Instrument too close to the wall – provide minimum distance of 15 cm. Allow instrument to cool down.
Check that air intake slots are not dirty and that instrument is in correct distance from wall. Allow instrument to cool down, then restart. If alarm sounds again after a short time, instrument is faulty.
Notify manufacturer.
Fault
No flow at oxygen nasal
cannula. No
alarm given by instrument.
Reduced flow.
Cause
Hose connection interrrupted. Hose kinked, faulty gasket or leaking screw coupling.
Oxygen distributor in humidifier blocked.
Remedy
Check that there is a flow indication on the flow monitor. Check that all hose connections and screw couplings are tight and change gaskets if necessary.
Clean oxygen distributor and replace if necessary.
Should there are any faults be present which cannot be rectified, contact the manufacturer or having the instrument overhauled by the manufacturer. Do not continue to use the instrument once the red fault light has come on.
WEINMANN:HAMBURG supplies various portable oxygen therapy units suitable as emergency reserve for home use (e.g. in case of power failure). For oxygen supply outside home, WEINMANN:HAMBURG supplies the OXYTRON 2 Mobile, which enables the patient to leave his home for fairly long periods of time without having to interrupt oxygen therapy. Ask your supplier or consultant for details.
Fig. 1 0
Spare fuse
Fuse holder
11
Gottlieb Weinmann Geräte für Medizin und Arbeitsschutz GmbH+Co. P.O. Box 54 02 68 · D-22502 Hamburg Phone 40/5 47 02- 0 · Telefax 40/54 70 24 61 Teletex 40 37 53 · Telex 17 40 37 53
WM 16091b – 05/95
Paper free of chlorine bleaching-agents
Loading...