Wavetek 75 Instruction Manual

INSTRUCTION MANUAL
Model
75
Arbitrary
Waveform Generator
0
1986
This document contains information
proprietary to Wavetek and is provided solely for instrument operation and maintenance. The information in this document may not be duplicated in any manner with­out the prior approval in writing from Wavetek.
WAVETEIQ
WAVETEK SAN DIEGO, INC.
9045
Balboa Ave.. San Diego, CA
92123
P.
0.
Box
85265.
San Diego. CA
921 38
Tel
61 912792200
TWX
91 013352007
Manual Revision
2/89
Manual Part Number:
1300-00-0366
Scans
by
ArtekMedia
O
2006
Scans
BY
Artek
Media
Artek
Media
1042
Plummer
Cir.SW
Rochester,
MN
55902
www
.artekmedia
.co
m
"High
resolution
scans
of
obsolete
technical
manuals"
if
your
looking
foraquality
scanned
technical
manual
in
PDF
format please
visit
our
WEB
site
at
www
.artekmedia
.co
m
or
drop
us
an
email
at
manuals
@
artekme
dia
.co
m
and
we
will
be
happy
to
email
you
a
current
list
of the
manuals
we
have
available
.
If
you
don't
see
the
manualyou
need
on
the
list
drop
us
a
line
anyway
we
may
stillbeable
to
point
you
to
other
sources.If
you
have an
existing
manual
you
would
like
scanned
please
write
for
details,
This
can
often
be
done
very
reasonablyinconsideration
for
adding
your
manual
to
our
library
.
Typically
the
scans
in
our
manuals
are
done
as
follows
;
1)
Typed
text
pages
are
typically
scanned
in
black
and
white
at
300
dpi
.
2)
Photo
pages
are
typically
scanned
in
gray
scale
mode
at
600
dpi
3)
Schematic
diagram
pages
are
typically
scanned
in
black
and
white
at
600
dpi
unless
the
original
manual
had
colored
high
lighting
(asisthe
case
for
some
70's
vintage
Tektronix
manuals)
.
if
you
purchased
this
manual
from
us
(typically
through
our
Ebay
name
of
ArtekMedia)
thank
you
very
much
.Ifyou
received
this
from
a
well-meaning
"friend"
for
free
we
would
appreciate
your
treating
this
much
like
you
would
"share
ware".By
that
we
mean
a
donation
of at
least
$5-10
per
manual
is
appreciated
in
recognition
of
the time
(a
manualcan
take as
much
as40hours
to
reproduce,
book,
link
etc
.),
energy
and
qualityofeffort
that
went
into
preserving
this
manual
.
Donations
via
PayPal
goto:
manmanuals@artekmedia
.co
m
or
can
be
mailedtous
the
address
above
.
Dave&Lynn
Henderson
Artek
Media
CONTENTS
SECTION
1
GENERAL DESCRIPTION
1.1 MODEL 75
.................................................
1-1
........................................
1.2 WAVEFORM EDITING 1-1
.......................................
1.2.1 Standard Waveforms 1-1
......................................
1.2.2 Rubber Band Editing 1-1
............................
1.2.3 Digital Amplitude and Digital Offset 1-1
1.2.4 Point Edit
................................................
1-1
......................................
1.3 STANDARD FEATURES 1-1
........................
1.3.1 81 92 x 4095 Point Waveform Resolution 1-1
................................
1.3.2 Crystal Controlled Frequency 1-1
..........................................
1.3.3 Operating Modes 1-1
..................................
1.3.4 Internal Waveform Counter 1-1
1.3.5 Hold Function
.............................................
1-1
...........................................
1.3.6 Return Function 1-2
.........................................
1.3.7 Stored Waveforms 1-2
...............................
1.3.8 Non-Volatile Instrument Set-Up 1-2
...........................................
1.3.9 Series Operation 1-2
.........................................
1.3.1 0 Parallel Operation 1-2
...................................
1.3.1 1 Communications Options 1-2
...........................................
1.4 SPECIFICATIONS 1-2
.......................................
1.4.1 Standard Waveforms 1-2
..........................................
1.4.2 Operational Mode 1-2
.......................................
1.4.3 Waveform Resolution 1-2
..............................................
1.4.4 Sample Rate 1-2
............................................
1.4.5 Output Signals 1-3
......................................
1.4.6 Output Characteristics 1-3
...................................................
1.4.7 Inputs 1-3
1.4.8 Trigger Sources
...........................................
1-3
..............................
1.4.9 Communication Ports (Optional) 1-4
................................................
1.4.10 General 1-4
1.4.11 Options
.................................................
1-4
SECTION
2
INITIAL PREPARATION
....................................
2.1 UNPACKING INSPECTION 2-1
..................................
2.2 ELECTRICAL INSTALLATION 2-1
.........................................
2.2.1 Power Connection 2-1
...................................
2.2.2 Verifying the Line Voltage 2-1
..................................
2.2.3 Fuse and Voltage Selection 2-2
.......................................
2.2.4 Signal Connections 2-3
......................................
2.2.5 Interface Connections 2-3
.............................................
2.2.6 GPlB Address 2-4
......................
2.2.7
RS-232
Baud Rate and DTElDCE Selection 2-4
..............................
2.2.8 Connecting the RS-232 Interface 2-5
..............................
2.3 INITIAL CHECKOUT PROCEDURE 2-7
Scans
by
Artekhledia
O
2006
CONTENTS (Continued)
SECTION
3
OPERATION
...
......
3.1 INTRODUCTION
.........
...........
3-1
....
3.2 CONTROLS, CONNECTORS. AND INDICATORS
..........
3-1
.................................
3.3 OPERATION
......
3-7
.......
3.4 INITIAL SETUP
......
................... 3-7
................................
3.5 EDITING WAVEFORMS 3-7
...........................
3.6 THUMBTACKS AND CURSOR 3-8
...........................
3.7 RUBBER BAND EDITING 3-8
3.8 INSERTING STANDARD WAVEFORMS
................ 3-8
3.9 DIGITAL AMPLITUDE
...............................
.
.
3-8
......
.........
3.1 0 NORMALIZING
......
. 3-8
.....
3.11 DIGITAL OFFSET
.............................
3-8
........................................
3.1 2 POINT EDITING 3-8
.............................
3.1 3 START AND STOP ADDRESSES 3-11
............................
3.1 4 WAVEFORM PARAMETERS 3-11
............
3.1 4.1 Waveform Frequency, Period, and Sample Freq 3-11
3.1 4.2 External Reference
................................
3-12
3.1 4.3 Waveform Amplitude
..........................
...
3-12
............................
3.1 4.4 Waveform Offset
...
3-12
................................
3.1 5 OPERATIONAL MODES 3-12
3.1 5.1 Continuous Mode
...............................
. .
3-12
......................................
3.1 5.2 Triggered Mode 3-12
3.15.3 Gated Mode
.............................
........ 3-12
.........................................
3.1 5.4 Toggled Mode 3-12
.....
3.15.5 Burst Mode
.......................
. .
3-1 3
....
3.16 SETTING AND CLEARING BREAK POINTS
...........
3-13
.............................
3.1 7 SYNC ADDRESS AND PHASE 3-13
......
3.1 8 STORllVG AND RECALLING WAVEFORMS
.......
3-13
3.19 CHAINED OPERATION
...............................
3-13
................................
3.20 WAVEFORM COUNTER 3-13
..................................
3.21 WAVEFORM HOLD 3-13
....................................
3.22 RETURN TO START 3-13
................................
3.23 LOW BATTERY INDICATOR 3-14
3.24 SERIES OPERATION
...............................
. 3-14
3.25 PARALLEL OPERATION
....................
...
3-14
.....................
3.26 GPlB CONTROLS (Option 001 Only) 3-15
...........................
3.26.1 GPlB Address DIP Switches 3-15
.........................................
3.26.2 Setup Menu Key 3-15
.........................................
3.26.3 BLIFFR Soft Key 3-15
........................................
3.26.4 ADRS Soft Key 3-15
.....................................
3.26.5 LOCAL Soft Key 3-15
.........................................
3.26.6 Control Knob 3-15
.................
3.27 GENERAL PLIRPOSE INTERFACE BUS (GPIB) 3-16
.................................
3.27.1 Bus Line Definitions 3-17
...................................
3.27.2 Handshake Sequence 3-18
..........................................
3.27.3 Commands 3-18
........................................
3.27.4 Listen Address 3-18
............................................
3.27.5 Talk Address 3-18
....................
.....
...
3.27.6 Secondary Address
..
..
3-18
....................................
3.27.7 Universal Commands 3-18
Scans
by
ArtekMedia
O
2006
CONTENTS (Continued)
3.27.8 Addressed Commands
..................................
3-19
3.27.9 Data Transfer
.......................................
3-19
3.28 MODEL 75
GPlB PROGRAMMIIVG
.......................
3-19
3.28.1 General Programming Information
..................
3-19
3.28.2 Characters
.....................................
. . 3-19
3.28.3 Action and Parameter Commands
...............
...
3-20
3.28.4 Action Commands
.....................
.......
3-20
.....
3.28.5 Programming Parameters
......................
3-20
3.28.6 Sending Waveforms to the Model 75
.......
...........
3-20
3.28.7 High Speed Binary Waveform Transfer
.........
3-21
...
3.28.8 Reading Waveforms From the Model 75
...
3-21
..............................
3.28.9 Programming Reference Guide 3-21
...........
3.29 ERRORS - GPlB REPORTING
...........
3-26
............................................
3.29.1 Class 1 Errors 3-26
.........................................
3.29.2 Class 2 Errors 3-27
......................................
3.29.3 Class 3 Errors 3-27
3.30 RS-2326 CONTROLS
.....
. .
.
.
3-27
....
3.30.1 Setup Menu Key 3-27
3.30.2 BUFFR Soft Key
. . .
. 3-28
....................................
3.30.3 LOCAL Soft Key 3-28
...
3.30.4 HAND Soft Key
.
. . .
3-28
..................................
3.30.5 Control Knob 3-28
.........................................
3.30.6 Baud Rate Switch 3-28
...
......
3.31 GENERAL RS-2326 INFORMATION
.....
3-28
......
3.31.1 Characters
......
...
.
. 3-29
.........
3.31.2 RS-2326 Terminators ..... 3-29
............................
3.31.3 Action and Parameter Commands 3-30
.....
3.31.4 Action Commands
. .
.......
.
. 3-30
...
....
3.31.5 Programming Parameters
...........
3-30
.........................
3.31.6 Sending Waveforms to the Model 75 3-31
3.31.7 Reading Waveforms from the Model 75
.....................
3-31
3.31.8 Programming Reference Guide
...........................
3-31
...
3.32 ERRORS - RS-232-C
REPORTING
....................
3-31
3.32.1 Class 1 Errors
........................................
3-31
....
3.32.2 Class 2 Errors
....
3-32
3.32.3 Class 3 Errors
. .....
.....................
3-32
SECTION
4
CIRCUIT DESCRIPTION
4 1 INTRODUCTION 4-1 4 2 DETAILED CIRCUIT DESCRIPTION 4-1
4 2 1
S~gnal Path 4-1
4 2 2
T~m~ng And Tr~gger~ng 4-9 4 2 3 Instrument Control 4-1 2 4 2 4
Commun~cat~on Optlons 4-1 8 4 2 5 Power Supply 4-1 9
Scans
by
ArtekMedia
O
2006
CONTENTS (Continued)
SECTION
5
CALIBRATION
SCOPE
................................
5-1
...........
..........................
FACTORY REPAIR
....
..... 5-1
REQUIRED TEST EQUIPMENT
... .................. 5-1
CALIBRATION
.
.
......
........
..............
5-1
AMPLITUDE
..................... ...........
5-3
OFFSET
....................................
......
5-4
FREQUENCY
..................................
....
5-5
OUTPUT WAVEFORM
RISE AND FALL TIME
.................
5-6
................
SUM IN
........................
5-7
SECTION
6
TROUBLESHOOTING
6.1 FACTORY REPAIR
................................... 6-1
6.2 BEFORE YOU START
..................................
6-1
6.2.1 Isolating a Problem
......................................
6-1
6.2.2 Troubleshooting Dynamic Digital Circuits
....................
6-3
6.3 TROUBLESHOOTING GUIDES
............................... 6-3
6.3.1 Power Supply
...........................................
6-3
6.3.2 Microprocessor and Main Memory
.........................
6-16
6.4 TROUBLESHOOTING
l NDlVlDUAL COMPONENTS
.................
6-20
6.4.1 Transistor
..........................................
3-20
6.4.2 Diode
............................................
6-20
6.4.3 Operational Amplifier
...............................
6-23
6.4.4 Field Effect Transistor (FET)
..............................
6-20
................................
6.4.5 Capacitor
....
6-20
6.4.6 Logic Levels
........................................
6-20
SECTION
7
PARTS LIST AND SCHEMATICS
7.1 DRAWINGS
.......................................
7-1
.............................................
7.2 ERRATA 7-1
7.3 ORDERING PARTS
.................................
7-1
APPENDIX
APPENDIX A AMERICAN STANDARD CODE FOR INFORMATION INTERCHANGE
(ASCII)
.................................................. A-1
APPENDIX B MODEL 75 QUICK REFERENCE GUIDE
.............................
B-1
Scans
by
ArtekMedia
O
2006
viii
Model
75
Arbitrary Waveform Generator
Scans
by
ArtekMedia
O
2006
GENERAL DESCRIPTION
1.1 MODEL 75
The Model 75 is a versatile, yet easy to use arbitrary waveform generator. It contains advanced waveform editing features that allow the user to easily create and edit complex waveforms.
1.2 WAVEFORM EDITING
The editing featuresare based on a pair of points called "thumbtacks". These thumbtacks are placed at each end of the portion of the waveform to be edited, then the
appropriate editing feature is invoked. The thumbtacks will appear as bright dots on an oscilloscope using Z-axis modulation.
1.2.1 Standard Waveforms
Standard waveforms are inserted between the thumb­tacks. Astandard waveform can be the entire waveform
or only part of it. The standard waveforms are sine, cosine, inverse sine, haversine (inverse cosine), triangle, square, ramp up, ramp down, and dc. These standard waveforms can then be edited using the other editing features if desired.
1.2.2 Rubber Band Editing
Rubber band editing is analogous tostretching a rubber band between two thumbtacks. A cursor is placed
between the thumbtacks which can be moved in either theXor they direction.The waveform is then stretched from each thumbtack to the cursor in real time.
1.2.3 Digital Amplitude and Digital Offset
The sectionof the waveform that is between the thumb­tacks can beadjusted in amplitude and offset in real time without affecting the rest of the waveform.
1.2.4 Point Edit
Individual points of the waveform can also be adjusted in real time.
1.3 STANDARD FEATURES
1.3.1 81 92 x 4095 Point Waveform Resolution
The waveforms have a vertical resolution of 4095 points and a horizontal resolution adjustable from 2 to 8192 points. The waveform memory is broken up into four blocks of 2048 horizontal points. This can be used as one waveform of up to 8192 points or as four different waveforms (one active and three stored) of up to 2048 points each.
1.3.2 Crystal Controlled Frequency
The waveform clock (sample frequency) is crystal con-
trolled and adjustable from 20
mHz to 2 MHz (500 ns to
50s), allowing a waveform periodof 1 ps to 11 3.8 hours. The frequency is adjustable as either waveform fre­quency, waveform period, or sample frequency. An
external
TTL
level clock of up to 2 MHz can also be used
as the waveform clock (sample freq).
1.3.3 Operating Modes
The Model 75 can operate in continuous, triggered, gated, toggled, or burst modes. All modes except con­tinuous are initiated (triggered) by either a front panel
pushbutton or an external signal. In these modes the waveform stops at break points which can be inserted anywhere in the waveform (by default one breakpoint will be at the start address), then waits for another trigger.
In burst mode, a predetermined number of waveforms (from
1
to 1,048,575) is generated before stopping.
1.3.4 Internal Waveform Counter
In gated, toggled, and burst modes an internal counter can be used to display the number of complete waveforms since the last trigger. This is particularly useful for monitoring a burst of waveforms or counting the number of waveforms for fatigue testing.
1.3.5 Hold Function
A front panel pushbutton or external signal will cause the waveform to stop at the current output level and
remain there until the hold is released.
Scans
by
ArtekMedia
O
2006
1.3.6 Return Function
A front panel pushbutton or external signal will cause the output to ramp back to the start value. The rate of this ramp is adjustable from
0 (instant return tostart) to
100 ms per vertical point.
1.3.7 Stored Waveforms
Three additional waveforms may be stored (with their parameter settings) in waveform memory. These may later be recalled into, or exchanged with the active waveform. This feature is only available with horizontal resolutions of 2048 points or less.
1.3.8 Nonvolatile Instrument Setup
The complete instrument setup is stored in battery backed up memory so that it will not have to be reset after the power has been turned off.
1.3.9 Series Operation
Several Model 75's can be connected in series, each one producing a burst of waveforms, with their outputs
summed together (using the "sum in" connector). This can be used to produce long waveforms or unusual sequences such as 50 "normal" waveforms and then 1 "abnormal" waveform.
1.3.1 0 Parallel Operation
Several Model 75's can be connected in parallel to create phase coherent arbitrary waveforms. The phase relation­ship is completely adjustable with a resolution equal to the horizontal resolution of the waveform. This feature is useful in three axis mechanical testing and multi-phase power applications.
1.3.1 1 Communications Options
Either an IEEE-488or RS232Ccommunication interface is available to transfer waveform data and control the Model 75 from a computer.
1.4 SPECIFICATIONS
1.4.1 Standard Waveforms
DC, square, triangle, up-ramp, down-ramp, sine, cosine, inverse sine, and haversine waveforms are provided.
Each of these waveforms can be stored into the waveform memory between thumbtacks automatically, and then edited with the waveform editing functions if desired.
1.4.2 Operational Modes
Continuous Output runs repetitively from the programmed start address to the programmed stop address.
Triggered Output quiescent until triggered by external signal or front panel pushbutton, then runs until another break point is encountered. Break points can be inserted anywhere in the waveform with the default break point at the beginning of the waveform.
Gated Same as triggered mode except that the output is con­tinuous for the duration of the gate signal and until the first break point encountered after the gate signal ends.
Burst
Same as in the triggered mode except that a program-
mable number of waveforms is counted before the
waveforms stop. The maximum burst is 1,048,575 waveforms.
Toggled The output is quiescent until triggered. It then runs con­tinuously until another trigger stops the waveform at the next break point.
1.4.3 Waveform Resolution
Normal Waveform 2048 points horizontal
x
4095 points vertical.
Chained Waveform 81 92 points horizontal
x
4095 points vertical. Stored
waveforms are not available with chained operation.
1.4.4 Sample Rate
Range
0.02 Hz to 2.0 MHz (50
s to 500 ns)
Waveform Period and Frequency The actual waveform period is the number of horizontal points (from start to stop address) times the sample time. This can range from 1 microsecond for a square wave to 11 3.8 hours for a waveform using all 81 92 points.
Frequency Resolution 4 digits
Frequency Accuracy
+
0.01 5
'10
Repeatability (24 Hr)
+
25 ppm
Temperature Coefficient
<5 ppml°C
Scans
by
Artekhledia
O
2006
1.4.5
Output Signals Z-Axis Out
Amplitude
This signal is a pulse, adjustable in amplitude from 2 to
14 volts, that is used to modulate the Z-axis of an
Range:
+
0.005 to 5 Vp into 509 (0.01 to 10 VP-P into
oscilloscope when using the waveform editingfeatures.
509). + 0.01 to 1 OVp into an open circuit (0.02 to 20Vp-p
The logic polarity is also selectable.
into an open circuit). Resolution: 3 digits. Accuracy:
+
1.8% of setting and + 2.5 mVfor amplitude
1.4.7
Inputs
+
offset 20.5V.
k
1.8% of setting and ? 0.25 mV for 0.5V > amplitude
sum
In (509)
+
offset 250 mV.
The signal at this input is added to the arbitrary waveform
+
1.8% of setting and + 25 pV for 50 mV > amplitude
to produce an output (at waveform out)of the sum of the
+
offset.
two. A peak input of
+
5V w~ll produce an output equal
Repeatability (24 Hr):
k
0.1
O/O
of setting.
to the amplitude setting. Input is fuse protected.
Tem~erature Coefficient: 61 50 ppml°C of setting.
Accuracy is
4% (2% if internal fuse is shorted).
Offset
Trig In (TTL)
lm~edance is 100 KR. Sianal is used to control the
Range:
+
5~ to - 5V into 50R
(-
10V to + 1 OV into an
ge"erator in the trigger, gaie, toggle, and burst modes.
open circuit). Absolute amplitude
PIUS
offset may not
Slope polarity is selectable,
exceed + 5V into 50R
(+
10V into an open circuit).
Svnc In
(TTL)
Resolution: 3 digits.
lkpedanbe is 100 KR. A high input causes the waveform
Accuracy:
2% of setting and + 15 mV for amplitude
+
to restart at the start address. Primarily used for syn-
offset
20.5V.
chronizing several arbitrary waveform generators in
2% of setting and k 1.5 mVfor 0.5V>amplitude + off-
parallel operation.
set 250
mV.
Ref In (TTL)
2% of setting and + 0.1 5 mV for 50 mV > amplitude
+
Impedance is 100 KR. This signal input is used for the
offset.
sample clock when the external ref is selected.
Repeatability (24 Hr):
+
5 mV and + 0.1
O/O
of setting.
Temperature Coefficient:
6(+
1 mV + 150 ppm)l°C of
setting.
1.4.6
Outputs
ARB Out
(50R)
Source of primary waveform output.
Maximum
amplitude is
+
5 Vp into 50R. Output is fuse protected.
Sync Out
(TTL) Will drive up to 10 LSTTL loads.This is a pulse 1 sample time wide that is usually used to sync an oscilloscope or another ARB generator. Pulse position is adjustable to any point in the waveform.
Burst Done Out (TTL) Will drive up to 10 LSTrL loads. ATTL low indicates that the waveform is running. TTL high indicates that the instrument is waiting for a trigger in triggered, gated, toggle, or burst modes. Used primarily to start another ARB generator in series operation.
Ref Out (TTL) Will drive up to 10 LSTTL loads. This pulse train at the sample frequency is used primarily as a reference source for other ARB generators in series operation or in parallel operation.
Hold In Impedance is 100 KR. ATTL low or switch closure at this input causes the generator to stop at the current output level and remain there until released by a TTL high or
switch opening.
Return In Impedance is 100 KR.
ATTLlowor switch closureat this
input causes the output level to ramp back to the start
value.
'The ramp rate is front panel selectable from 0 to
100 ms per point.
1.4.8
Trigger Sources
Manual Use the Trigger pushbutton switch on front panel to operate internal trigger circuitry. The slope polarity can be set via the display menus.
External Use the communication
ports(GPIB or RS-232C)or the
Trig In connector on front panel. Trig In connector fre­quency input can be up to 1 MHz. The slope polaritycan be set via the display menus.
Scans
by
ArtekMedia O 2006
1.4.9 Waveform Characteristics
Horizontal Resolution Up to 81 92 points.
Vertical Resolution 4095 bits
(
-
2047 to + 2047).
RiselFall Time
<350 ns.
1.4.10 Communication Ports (Optional)
IEEE-488 Port (Option 001)
Type:
IEEE-488-1978 compatible, non-isolated, double
buffered. Address: 0-30, internal DIP switch selectable or front
panel selectable. Subsets:
SHl, AHl, T6, TEO, L4, SRl, RLI, PPO, DCl,
CO. El.
RS-232-C Serial Port (Option 002)
Communication is full duplex
(bi-directional) with
CTSIDTR or XONIXOFF handshaking. Connector: Rear panel mounted DB-25 (female) with
DCE or DTE configuration. Data Format: 8 bits, no parity, one stop bit Data Rate: 14 steps internal switch selectable (50, 75,
1
10,134.5,150,200,300,600,1200,1800,2400,4800,
9600, 19,200).
1.4.1 1 General Environmental
Temperature Range: + 23°C + 5°C for specified opera­tion, operates
0 to + 50°C, - 20 to + 75°C for storage.
WarmupTime: 20 minutes for specified(guaranteed for 1 year) operation.
Vibration: 5 to 55 Hz with maximum of
29 at 55 Hz.
Shock:
309, 11 ms half sine.
Altitude: Sea level to 10,000 ft. for operation. Sea level to 40,000 ft. for storage.
Relative Humidity:
95% at 25°C and at sea level
(non-condensing).
Dimensions: 21 cm (8.27 in.) wide,
10 cm (3.94 in.) high,
32.3 cm (1 2.72 in.) deep.
Weight: 3.55 kg (7.8 Ib) net, 4.68 kg (1 0.3 Ib) shipping.
Power: 90 to 128,180 to 256 rms; 48 to 66 Hz; 1 phase;
<25 VA.
1.4.1 2 Options
001: IEEE-488 (GPIB) Port
For instrument control or entering waveform data.
002:
RS-232-C Serial Port
For instrument control or entering waveform data.
NOTE
Options
001
and
002
are mutually exclusive.
Scans
by
ArtekMedia
O
2006
INITIAL PREPARATION
2.1 UNPACKING INSPECTION
After carefully unpacking the instrument, visually inspect all external parts for possibledamage toconnectors, sur­face areas, etc. If damage is discovered, filea claim with the carrier who transported the unit. The shipping con­tainer and packing material should be saved in case reshipment is required.
2.2 ELECTRICAL INSTALLATION
2.2.1 Power Connection
WARNING
To preclude injury
ordeath due toshock, the
third wire earth ground must be continuous to the facility power outlet. Before connect-
ing to the facility power outlet, examine extension cords, autotransformers, etc., between the instrument and the facility power outlet for acontinuous earth ground
path.The earth ground path can be identified at the plug on the instrument power cord; of three ter­minals, the earth ground terminal is the non­matching shape, usually cylindrical.
CAUTION
To prevent damage to the instrument, check for proper match of line and instrument voltage and proper fuse type and rating.
Line Voltage
Refer to table 2-1 to determine the proper selection of the line voltage connector. Unless otherwise specified at the time of purchase, this instrument was shipped from the factory with the power transformer connected for operation on a
1 1
5
Vac line supply and with a 318 amp
fuse. If the unit is shipped for
11
5
Vac operation, there
will be no markings or tags on the unit. If the unit is
shipped for 220
Vac operation, there will be a 220 Vac
tag on the rear panel of the unit and a 3/16 amp fuse
installed.
2.2.2 Verifying the Line Voltage
To verify the line voltage (or change the fuse), the operator must first remove the top and bottom covers.
Remove the top and bottom using the following steps and figure 2-1.
1. Remove two (2) screws holding top and bottom covers to rear panel.
2. Slide both covers (together as a unit) to the rear and remove from the chassis assembly.
Figure 2-1. Top and Bottom Cover Removal
After the covers have been removed, the line voltage can be checked by viewing the voltage label through the
inspection hole as shown in figure 2-2.
INSPECTION SLOT
GUARD PLATE
'
REAR PANEL
Figure 2-2. Line Voltage Inspection Hole
Scans
by
ArtekMedia
O
2006
2.2.3 Fuse and Voltage Selection
If
the line voltage is not correct according to table 2-1,
perform the following steps and refer to figure 2-3 for steps 1 and 2, and figure 2-4 for steps 3 thru 6
tochange
the line voltage and fuse.
1. Remove the two screws holding guard plate to the rear panel.
2. Remove the guard plate from the rear panel.
GUARD
PLATE
SCREWS
/
VOLTAGE
SELECTOR
CONNECTOR
F
U
AC
PRIMARY
-
VOLTAGE
SELECTOR
Figure 2-3. Guard Plate Removal
Figure 2-4. Fuse and Voltage Selection
3. Remove the voltage selector connector from the ac
primary board. Rotate the connector until the cor­rect voltage selector indicator is on top.
4. Reinstall the voltage selector connector.
5. Remove the fuse and install new fuse as called out in table
2-1.
Connector Voltage
Position Range
90
to 128 Vac 318 amp
Table 2-1. VoltagelFuse Selection
220V
WARNING
Because lethal voltages are exposed, do not apply ac power to the unit until the guard plate is attached to the unit.
Reassembly
Refer to figure 2-5 for step 1 and figure 2-6 for steps 2 thru 5.
1. Insert the guard plate over the ac primary board and voltage selector connector and align with screw holes in the rear panel. Secure with two screws.
180
to
256 Vac
BOARD
311 6 amp
Figure 2-5. Rear Chassis Assembly
Scans
by
ArtekMedia
63
2006
Figure
2-6.
Top and Bottom Cover Installation
2. Turn the instrument upside down, position the bottom cover over the guard shield and rear panel.
Slide the bottom cover forward approximately two
(2) inches while engaging the
topcover shield slides (see figure 2-6, detail A) and the outside slide rails (see figure 2-6, detail B).
3.
Turn the instrument right side up. Install the topcover using the same procedure as in step
2.
4. Align the rear of both the topand bottom cover with each each other so that the cover interlocks are properly mated. Once mated, hold the covers firmly together and slide the chassis assembly into top and
bottom covers.
CAUTION
When sliding on the bottom cover, avoid
moving the ribbon cables located
nearthe
power transformer.
5. Secure covers to unit using two screws as shown in figure 2-1
.
2.2.4
Signal Connections
Use RG58U 509 coaxial cables equipped with BNC con­nectors to distribute signals when connecting this instru­ment to associated equipment.
NOTE
Signal ground may be floated up to
_t
42V
with respect to chassisground. Bea ware that all signal grounds are common and must all be floated together.
2.2.5
Interface Connections
The GPlB (Option 001) 110 rear panel connector is an Amphenol 57-10240 or equivalent and connects to a GPlB bus cable connector (available from Wavetek in
1 and 2 meter lengths).
NOTE
The terms
"GP1B"and "IEEE-488"are used
synonymously in this manual.
Scans
by
Artekhledia
O
2006
2.2.6 GPlB Address
For Option 001 instruments on the General Purpose
Interface Bus (GPIB), ensure that the instrument GPlB address is correct. The GPlB default address can be changed by resetting the internal DIP switch (for access,
remove the top cover, see paragraphs 2.2.2 and 2.2.3) or changed temporarily by pressing the front panel ADRS soft key and rotating the Control knob when the GPlB address is displayed.
The switch sections are labeled from 1 thru 5 and their OPEN position noted (OPEN
=
Binary "0" in table 2-2).
To verify the address, press ADRS soft key on the front panel.
Table 2-2. GPlB Address Code
Device / Listen I Talk
1
1
2
3
4 5
Listen
1
Talk
2.2.7 The RS-2324 Connectors
The RS-232C 110 rear panel pin connections and signal
names are given in table
2-3. The panel connector is a DB-25 female connector (see figure 2-7) and will con­nect to a standard
RS-232C cable (interface type
E).
The data format is an 8-bit, no parity, one stop bit. The baud rate is selectable from those listed in table 2-5 using dip switch
S1 (figure 2-8).
Table 2-3. RS-232-C Data InlOut
40
4
1
42
(Remaining Pins Not Connected)
43
'Transmit Data 'Receive Data 3~equest to Send 4~lear to Send 'Data Set Ready
'Receive Line Signal Detect
7~ata Terminal Ready
Pin
2
3
4
5
6
7 8
20
-
/
RS-232-C Connector
Name
TXD' RXD~
RTS3 CTS~ DSR5 SIG GRD RLSD'
DTR~
Direction (DCE Configuration)
IN
OUT
N IC OUT OUT
-
OUT
IN
Z-AXIS
out
I60011
1
Ref Ref
Hold
In
Out
In
In
ITTLI ITTL) lTTLl ITTLI
Direction (DTE Configuration)
OUT
IN
OUT
IN IN
-
IN
OUT
nnnnnnnnn
Figure 2-7. RS-232-C Connector Pinouts
NOTE:
Address
31
is
not allowed
Scans
by
ArtekMedia
O
2006
The following is a description of pin functions. Actual
installation is covered in paragraph 2.2.8. In the DCE configuration, pins 2 and 3 are data input and
output respectively and must be used. Pin 5 may be used
to indicate that the Model 75 is ready to receive data.
Pin 20 tells the Model 75 that the connected device is ready to receive data, but may be connected to pin 6 of
the Model 75 if the other device is always ready. Pin 7
is signal ground and must be used. Pins 6 and 8 will be
positive (asserted) when the Model 75 is on.
In the DTE configuration, pins 2 and 3 are data output and input respectively and must be used. Pin 4 will be positive whenever the Model 75 is On. Pin 5 tells the
Model 75 that the connected device is ready to receive data.
If
the other device is always ready, pins 4 and 5 of the Model 75 may be connected together. Pins 6 and 8 are properly terminated inputs but have no function. Pin
20
will be positive when the Model 75 is ready to
receive data. Pin 7 is signal ground and must be used.
1. Determine if the other device is a DCE device or a DTE device. If this is not explicitly spelled out you will need to look at the pin out for its connector. If
pin 2 is the data output, the device is a DTE (most
computers). If pin 3 is the data output, the device
is a DCE (most modems). Configure the Model 75
as the opposite type of device by connecting the
internal optioncable to theappropriate
header(see
figure 2-8).
In most cases a straight through cable can be used
to connect the Model 75 with the other device. The
Model 75 only requires the connection of pins 2, 3, 5,7, and 20 but the other device may require more. Examples of typical cable assemblies are shown in
figure
2-9. However, there are many interpretations
of the
RS-232Cstandard and either of the following procedures (a or b) should be used to ensure that the control and handshake signals are properly connected.
The handshaking method is front panel selectable as
a. If the Model 75 is configured as a DCE then;
CTSIDTR (hardware) or XONIXOFF (see paragraph 3-32 for more information).
(1) Pin 7 (signal ground) is always connected
straight through.
2.2.8 Connecting the RS-232-C Interface
Before connecting the Model 75 using the RS-232C
interface, the correct configuration (DCE or DTE) must be selected and the internal ribbon cable connected to either the DCE or DTE location (see figure 2-8) on the RS-232-C option board. To determine which configura­tion is correct, you will need the manual for the device to which you are connecting the Model 75. The follow­ing assumes that the other device has a DB-25, RS-232-C connector.
DCE
HEADER
Figure 2-8. Baud Rate Switch and DTElDCE
Connections
(2) Pins 2 and 3 are connected straight through
if
DCEIDTE configuration is correct.
(3) Pin 5 of the Model 75 will need to be con-
nected to an input of the other device that will enable and disable the character transmission from that device. If the other device does not support this, XONIXOFF handshaking will be required and pin 5 can be left disconnected.
(4) Pin 20 of the Model 75 should be connected
to an output pin on the other device that
indicates when it is ready to accept data. If the other device does not have such a pin but is not always ready, it can send
XOFFs
and XONs to indicate readiness. When
using this method, or if the other device is
always ready
toaccept data, this pin can be connected to an always positive pin on the other
deviceor topin 60r 8on the Model 75.
(5) Pins 6 and 8 are always positive when the
power is on and can be connected to any pins on the other device that require being
held positive.
b. If the Model 75 is configured as a DTE then;
(1) Pin 7 (signal ground) is always connected
straight through.
(2) Pins 2 and 3 are connected straight through
if
DCEIDTE configuration is correct.
Scans
by
ArtekMedia
O
2006
(3) Pin 20 of the Model 75 will need to be con-
nected to an input of the other device that will enable (positive) and disable (negative) the character transmission from that device. If the other device does not support this,
XONIXOFF handshaking will be
required and pin
20 can be left dis-
connected.
(4)
Pin 5 of the Model 75 should be connected to an output pin on the other device that
indicates whether or not it is ready to accept data. If the other devicedoes not have such a pin but is not always ready, it can send XOFFs and XONs to indicate readiness. When using this method, or if the other device is always ready to accept data, this
IBM-PC
APPLE
II
WITH SUPER SERIAL CARD
IBM-PC IN COMMUNICATIONS MODE
MODEL 75 MODEL 75
(DCE) APPLE II
DB-25
FEMALE
DB-25 DB-25
MALE MALE
DB-25
MALE
WAVETEK MODEL
6000
APPLE MACINTOSH
6000
MODEL 75 MODEL 75
(DCE) MACINTOSH (DTE)
D 6-25 MALE
D 6-25 DE-9
MALE MALE
Figure
2-9.
Typical Examples
of
Serial Cables
DB-25 MALE
2-6
Scans
by
Artekhledia
O
2006
pin can be connected to an always positive pin on the other device or to pin 4 on the Model 75.
(5) Pin 4 is always positive when the power is
on and can be connected to any pin on the other device that requires being held positive.
(6) Pins 6and 8are properly terminated inputs
but have no function.
NOTE
Anypins not mentioned in the previous list are
not connected in the Model
75and therefore can be connected to anything if the connec­tions are already present in the cable.
3. The handshaking method is front panel selectable as
CTSIDTR (hardware) or XONIXOFF (see para-
graph 3.31 for more information).
4. Set the baud
rateon the Model 75 to the highest rate that is available on both the Model 75 and the other device by setting the dip switches (figure 2-8) according to table
2-4.
5. Configure the other device for the same baud rate as in step 4, 1 stop bit, 8 data bits, and no parity.
Table 2-4. RS-232-C DIP Switch Settings
Baud Rate
SW1-4 SW1-3 SW1-2 SW1-1
Closed Open
Open Closed Open
Open Closed
Open Closed Open Open Closed Open Closed Open Open Closed
Closed Open
Closed Closed
Closed Open Open Closed
Open
1
Open I Closed
Closed
1
(3;:;
I
Closed
Open Open
Closed Closed
Closed
Open Closed Closed Open
*This is the factory set position
NOTE
1.
SW1-5 should always be OPEN
2.
Factory connector configuration is DCE
Closed Closed Closed Open Open Open Open
Open Open Closed Closed
2.3 INITIAL CHECKOUT PROCEDURE
The Model 75 ARB waveform generator
will
supply a variety of waveforms at the ARB Out connector that can be modified at will using the front panel controls. In order toverify the operation of the controls and the functions1 modeslvalues that they control, the generator must first be set up to produce an output and then the effect of the controls on the output and the display indications noted. While this procedure verifies functional operation of this
instrument, it does not verify the
calibration.Thevalues shown are typical values and should only be used as a guide. Required tools and test equipment are shown in table
2-5.
Closed Open Open Open
Table 2-5. Test Equipment and Tools
Instrument Comments
Oscilloscope
*
50R Feedthrough BNC Tee
BNC Coax Cable
Signal Source*
Frequency Counter
Dual channel. 20 MHz rnin band­width with external trigger and Z­axis capability.
2% accuracy, 2W
1 male to 2 female
RG58U, Q 3 ft. length (4 each)
1 MHz capability with square wave (1 0 Vp-p into
50R) and TTL
output. Capable of both frequency and
period measurements.
*Check the oscilloscope manual to determine the
maximum allowable Z-axis input voltage. Do not exceed that voltage.
*A second Model 75 can be used as the signal source.
For the TTL signal requirements, use the Ref Out connector on the rear panel and use the SAMPL soft key to set the frequency. For the 1 kHz, 10 Vp-p signal, use the STOP soft key to set the stop address to 1999, use the FREQ soft key to set the frequency, and use the FUNC and EXEC soft keys to select a square wave. See section 3 for more detail.
To verify the overall operation of the Model 75 ARB waveform generator, follow the procedure in table
2-6. Refer to the referenced figures and tables for appropriate test equipment connections and settings. Refer to figure 3-1 and paragraph 3.2 for control and connector locations.
Scans
by
ArtekMedia
O
2006'
Table
2-6.
Checkout Procedure
Step
1
ControllSwitch
Power
PositionlOperation
Press On
ObservationlComments
Display indicates "WAVETEK MODEL 75".
I I
Adjust the LCD viewing control for best contrast.
--
-
-
This step verifies the "Reset" functions of the Model 75.
Press three times
Press once Press once
2
Display indicates "RESET Ver
x.xx".
Display indicates "RESET Ver x.xx
X".
Setup F2 (Reset) F3 (Exec)
This step verifies the "Z-Axis" functions of the Model 75.
Setup
F2 (ZLOGIC) Control knob
Control knob
F1 (ZLEVL)
Control knob
Control knob
Press once
Press once Rotate until display
indicates "ZLOGIC POS".
Rotate until display indicates "ZLOGIC
NEG".
Press once
Rotate CW until display indicates
"ZLEVEL
14V".
Rotate CCW to step display through the selectable settings.
Connect the Model 75 as shown in figure 2-10. Set
equipment controls as indicated in table
2-7.
Display indicates "ZLOGIC NEG" or "ZLOGIC POS". Oscilloscope will indicate a dc level of
0
Vdc + 1V
with positive-going pulses.
Oscilloscope will indicate a dc level of
2,
4,
6,
8, 10,
12, or 14 Vdc with negative-going pulses.
Oscilloscope will indicate a dc level of
+
14V
+_
1V
with negative-going pulses.
Oscilloscope will indicate the selected voltage
+_
1V with negative-going pulses. Leave Setting at "ZLEVEL 2V".
This step sets the "Z-Axis Out" voltage level and polarity.
Control knob Rotate CW or CCW
to set acceptable contrast between trace and thumb-
tacks. Adjust oscillo­scope intensity if
required.
Connect the Model 75 as shown in figure 2-1 1. Set equipment controls as indicated in table
2-8.
If thumbtack intensity is too bright at + 2V, terminate the Z-axis cable with a 50Q load. If the thumbtacks are dimmed instead of intensified, set ZLOGIC to POS.
Scans
by
ArtekMedia
O
2006
Table 2-6. Checkout Procedure (Continued)
This step verifies the "Waveform Memory'' functions of the Model 75.
Store F2 (CHAIN)
Control knob
ObservationlComments
F3 (EXEC) Edit F1 (FUNC) Control knob
PositionlOperation Step
F3 (EXEC)
ControllSwitch
Edit F1 (START)
Control knob
Control knob
F2 (STOP)
Control knob
Control knob
Store
F2 (CHAIN)
Control knob
F3 (EXEC) Edit F1 (FUNC)
F3 (EXEC)
Press two times Press once Rotate until display
indicates "CHAIN 8K".
Press once Press two times Press once Rotate until display
indicates "FUNC SINE"
Press once
Press two times Press once Rotate CW
Rotate CCW
Press once Rotate CCW
Rotate CW
Press two times Press once Rotate until display
indicates "CHAIN 2K".
Press once Press two times Press once
Press once
Reset the oscilloscope horizontal sweep to 1
msIDIV.
Oscilloscope will indicate slightly more than one cycle of the sine wave at approximately 10 Vp-p.
Display will indicate
">ST = 0000 SP = 8191
".
Oscilloscope waveform start point will move from left to right until waveform disappears. Display will now
indicate
">ST = 81 91 SP = 81 91
".
Oscilloscope waveform start point will move from right to left until waveform is complete. Display will now
indicate
">ST = 0000 SP = 81 91
".
Display will indicate "ST = 0000 >SP = 81 91.
Oscilloscope waveform stop point will move from right to left until the waveform disappears. The display will
now indicate "ST
=
0000 >SP = 0000".
Oscilloscope waveform stop point will move from left to right until the waveform is complete. The display
will
now indicate "ST = 0000 >SP = 81 91
".
Display indicates "CHAIN 8K"
Display indicates "FUNC SINE".
Set the oscilloscope sweep to
.5
mslDIV.
Display will indicate "FUNC SINE X". Oscilloscope will display less than 3 cycles of a sine wave of approx­imately 10 Vp-p.
Scans
by
ArtekMedia
O
2006
Table
2-6.
Checkout Procedure (Continued)
This step verifies the "Amplitude Control" functions of the Model
75.
Step
This step verifies the "Offset" functions of the Model
75.
ControllSwitch
Display will indicate "AMPL 5.000 Vp".
Oscilloscope amplitude decreases smoothly to minimum of approximately
10 mVp-p. Display will now
indicate "AMPL 5.00
mVp".
6
This step verifies the "Frequency" functions of the Model
75.
PositionlOperation
7
ObservationlComments
Parameter F1 (AMPL)
Control knob
Press two times Press once Rotate CCW
F2 (OFFSET)
Control knob
Control knob
Control knob
F1 (AMPL)
Control knob
F3 (SAMPL)
Connect a coax cable from the Ref Out connector on the Model 75 rear panel to the frequency counter input. Set frequency counter to frequency mode.
Press once
Control knob
Display will indicate "SAMPL 1.0000 MHz'' and counter will indicate approximately 1 MHz.
Press once
Rotate CW
Rotate CCW
Rotate CW until display indicates "OFFSET 0.0
mV". Press once Rotate CW until
display indicates
"AMPL 5.000 Vp".
Rotate CCW until
display indicates
"SAMPL 500.0 KHz".
Counter will indicate approximately 500 kHz.
Display will indicate "OFFSET 0.0
mV".
Reset oscilloscope vertical gain to 2VIDIV.
Oscilloscope will increase smoothly to approximately
5.1 OV. The display will now indicate
"OFFSET5.1 OV".
Oscilloscope level decreases smoothly to
approxi-
mately - 5.10V. Display will now indicate "OFFSET
-
5.10V".
Resolution near zero is much greater so that the level will change slower as zero is approached and passed.
Scans
by
Artekhledia
O
2006
Table
2-6.
Checkout Procedure (Continued)
This step verifies the "Hold" functions of the Model
75.
-
9
1
Hold
ObservationlComments
Counter will indicate approximately 300 kHz.
Counter will indicate approximately 100 kHz.
Counter will indicate approximately 10 kHz.
Counter will indicate approximately 1 kHz.
Counter will indicate approximately 100 Hz.
Set frequency counter to measure period.
Counter will indicate approximately 100 ms.
Counter will indicate approximately 1
sec.
Counter will indicate approximately 10 sec.
Oscilloscope will indicate a slow sine wave of
10Vp-p.
Press once
PositionlOperation
Rotate CCW until
display indicates
"SAMPL 300.0
KHz". Rotate CCW until
display indicates "SAMPL 1 00.00
KHZ". Rotate CCW until
display indicates
"SAMPL 10.000 KHz".
Rotate CCW until display indicates "SAMPL 1.0000 KHz".
Rotate CCW until display indicates "SAMPL 100.00 HZ".
Rotate CCW until display indicates
"SAMPL 10.000 HZ".
Rotate CCW until display indicates
"SAMPL 1.0000 HZ".
Rotate CCW until display indicates
"SAMPL 100.00
mHZ". Rotate CW until
display indicates
"SAMPL 1.0000
KHz".
Step
8
(Cont)
Display indicates
"H0LD:ON". Oscilloscope indicates
that waveform stops at instantaneous dc level.
ControllSwitch
Control knob
Control knob
Control knob
Control knob
Control knob
Control knob
Control knob
Control knob
Control knob
Scans
by
ArtekMedia
O
2006
Table
2-6.
Checkout Procedure (Continued)
9
(Cont)
Hold
Step
Press once
ControllSwitch
PositionlOperation
Apply a 7TL low or switch closure input connector.
ObservationlComments
Remove the 7TL low
or switch closure
from the rear panel connector.
Oscilloscope waveform starts again. Connect a coax cable from the
7TL source output to
the Hold In connector on the Model
75
rear panel.
Oscilloscope indicates that waveform stops at instantaneous dc level.
Oscilloscope indicates that waveform starts again.
Return
Hold
This step verifies the "Return" functions of the Model
75.
Press once
Press once
Apply a
7TL low or switch closure input to the rear panel connector.
Oscilloscope indicates that waveform stops at instan­taneous dc level and moves slowly to approximately
0
Vdc. Display indicates "H0LD:ON
Y
=
0000".
OsciIloscope waveform starts again. Display indicates
"HOLD:OFFM.
Connect a coax cable from the TrL source output to the Return In connector on the Model
75
rear panel.
Oscilloscope indicates that waveform stops at instan­taneous dc level and moves slowly to approximately
0
Vdc. Display indicates "H0LD:ON
Y
=
0000".
Remove the 7TL low or switch closure input from the rear panel connector.
I
Hold press once Oscilloscope indicates that waveform starts again.
This step verifies the "Mode" functions of the Model
75.
I
Parameter
F1
(MODE)
Control knob
Trigger
Press two times
Press once
Rotate CW until dis-
play indicates
"MODE
TRIGGERED".
Press once
Oscilloscope trace level approximately
0
Vdc. Set
oscilloscope sweep to
.5
seclDIV.
Oscilloscope waveform runs for one cycle, then stops at
0
Vdc.
Scans
by
ArtekMedia
O
2006
Table
2-6.
Checkout Procedure (Continued)
This step verifies the "Burst Done Out" signal of the Model 75.
0
bservationlComments
Oscilloscope waveform runs continuously. Oscilloscope waveform stops at
0 Vdc.
Oscilloscope waveform runs continuously. Oscilloscope waveform stops at
0 Vdc.
Oscilloscope waveform runs for 3 cycles, then stops at
0 Vdc.
Display will indicate "COLINTER
=
3".
This step verifies the "Trig In" functions of the Model 75.
PositionlOperation
Rotate CW until
display indicates
"MODE GATED". Press and hold Release Rotate CW until
display indicates
"MODE TOGGLED". Press once Press once Press once
Rotate CCW until display indicates
"BURST CT
3". Press once Rotate CW until
display indicates
"MODE BURST".
Press once
Press once
Step
11
(Cont)
12
Connect the Model 75 as shown in figure 2-12. Set equipment controls as indicated in table
2-9.
ControllSwitch
Control knob
Trigger Trigger Control knob
Trigger Trigger
F3 (BLIRST)
Control knob
F1 (MODE)
Control knob
Trigger
Counter
Scans
by
Artekhledia
O
2006
Trigger Press Once
Connect a cable from Burst Done Out connector to oscilloscope CH 2 vertical input. Reset oscilloscope vertical input to CHOPPED with
CHI and CH2 both
"On" and at
2VIDIV.
Oscilloscope Burst Done Out signal goes to OVdc while
waveform is running and then returns to
+
5V when
waveform stops.
Table
2-6.
Checkout Procedure (Continued)
This step verifies the "Trigger Slope" function of the Model
75.
7
Parameter
1
press two times
Step
13
(Cont)
PositionlOperation
Press once Rotate CW until
display indicates
"MODE
TRIGGERED". Press once Press once Rotate CW until
display indicates "FREQ
900.00
HZ".
ControllSwitch
F1 (MODE)
Control knob
Parameter F1 (FREQ)
Control knob
1
Control knob
)
Rotate CW until
ObservationlComments
Oscilloscope displays one cycle of sine wave whose start is coincident with rising edge of the trigger signal.
I
F2(TRSLP)
Oscilloscope displays one cycle of sine wave whose
Press once
This step verifies the "Sync In" functions of the Model
75
display indicates "TRIG SLOPE NEG".
start is coincident with falling edge of the trigger signal.
Scans
by
ArtekMedia O 2006
Move the cable from the Trig In connector to the Sync
In connector.
Oscilloscope waveform runs continuously when the Sync In signal is low. Waveform goes to
0
Vdc when
Sync In is high.
15
This step verifies the "Ref In" functions of the Model
75.
F1 (MODE)
Control knob
16
Press once Rotate CCW until
display indicates "MODE
CONTINUOUS".
Press two times
Parameter
Connect the Model
75
as shown in figure 2-13. Set
equipment controls as indicated in table 2-1
0.
Table 2-6. Checkout Procedure (Continued)
F3 (REF) Control knob
TTL
Signal
source
ObservationlComments
Press once
Rotate CW until
PositionlOperation
Step
display indicates
"REF EXTERNAL".
ControllSwitch
Vary the frequency
up and down.
Oscilloscope waveform frequency varies with the TTL frequency.
Control knob
Edit F1 (FLINC)
Control knob
This step verifies the "Sum In" functions of the Model
75.
Rotate CW until
display indicates
"REF INTERNAL" Press two times Press once
Rotate CW until display indicates
"FUIVC DC".
1
F3 (EXEC)
I
Press once
Connect the Model
75
as shown in figure 2-14. Set
equipment controls as indicated in table 2-1 1.
Oscilloscope
(CHI) waveform is the same frequency
and approximately the same amplitude as the
(CH2)
waveform.
Z-AXIS
OUT
I
MODEL
75
I
OSCILLOSCOPE
Figure 2-10. Equipment Interconnection #1
Table 2-7. Test Equipment Control Setup#l
Oscilloscope
CHI Vert Input
5
VIDIV
Horiz Sweep
.5
mslDlV
Trigger Source
CHI Trigger Mode Auto Remaining Controls As Required
Scans
by
ArtekMedia
O
2006
Figure 2-1 1.
Equipment lnterconnection
#2
Table 2-8. Test Equipment Control Setup #2 Table 2-9. Equipment Control Setup #3
Oscilloscope
CHI Vert Input 2 VlDlV
Horiz Sweep
.5
mslDlV
Trigger Source
Ext
(
-
10 if available)
Trigger Mode
Auto
Trigger Slope
Positive
Trigger Coupling
AC
Remaining Controls As Required
TTL SIGNAL SOURCE MODEL
75
OSCILLOSCOPE
50Q
00000000
..@ @ @ @
@
0
TRIG
ARB
IN
OUT
TTL Signal Source
Frequency 200 Hz Function
TTL
Figure 2-12. Equipment lnterconnection #3
Oscilloscope
CHI Vert Input
2
VIDIV
CH2 Vert Input
2
VIDIV
Vert Mode Alt Horiz Sweep
.5
ms1DIV
Trigger Source CH2
Trigger Mode Auto Trigger Slope Positive
Trigger Coupling AC
Remaining Controls As Required
2-1
6
Scans
by
ArtekMedia
O
2006
Figure 2-1
3.
Equipment lnterconnection #4
TTL SIGNAL SOURCE
MODEL 75 OSCILLOSCOPE
SIG
OUT
Q
Table 2-10. Test Equipment Control Setup #4 Table 2-1 1. Test Equipment Control Setup #5
Frequency 1 .OO
KHz
Function Square
Wave
Amplitude 1 0 Vp-p
REF
IN
-
TTL Signal Source
Frequency 1 MHz Function
lTL
Signal Source
CHI Vert Input 2 VIDIV
SYNC
OUT
Oscilloscope
CHI Vert Input 2 VIDIV Trigger Source Ext (-10 if
avail)
Horiz Sweep
.5
ms1DIV
Trigger Mode
Auto
Trigger Slope
Positive Trigger Coupling AC Remaining Controls As Required
Oscilloscope
CH2 Vert Input 2 VIDIV
'rn-mo'
0
0
.
-
-
...
.
OOOOOOOO0
@
L@
@
69
@
..q
ARB
OUT
Vert Mode Alt
Horiz Sweep
.2 ms1DIV Trigger Source CH2 Trigger Mode Auto Trigger Slope Positive Trigger Coupling AC Remaining Controls As Required
50Q
cl
CH2
,;
-
0
Q
SIGNAL SOURCE MODEL 75 OSCILLOSCOPE
CH2
0
SUM
ARB
IN
OUT
Figure 2-14. Equipment lnterconnection #5
Scans
by
ArtekMedia
O
2006
r
WAMTCK:
1
San
Dleao
CA USA
-
I
MODEL
75
I
BEFOREOPENING UNPLUGPOWEO CORD DANGEROUS
VOLTAGES
INSIDE SEE MANUAL FOO SAFETY
IY9TFtllCTIOII
Z
AXIS
out
IEEE-488
(600nI
Ref Ref
Hold
Return
In
Out
In In
(TTL)
(TTLI (TTL)
(TTL)
Figure
3-1.
Controls, Connectors, and Indicators
Scans
by
ArtekMedia
O
2006
3.1 INTRODUC'I'ION
This section describes the operation of the Model 75.
Paragraph 3.2 describes the controls and connectors
of the instrument. The following paragraphs describe
how to use the various features and modes of the Model
75.
3.2 CONTROLS, CONNECTORS, AND INDICATORS
The front panel controls are organized into four menu
selection keys, three soft keys, three real time keys, the Counter key, and the Control knob. To change any instru­ment setting, the appropriate menu key is pressed
repeatedly until thedesiredsoft key labelappears in the bottom line of the display (see Appendix B for menu organization). Pressing the associated soft key will cause
the top line of the display to indicate the setting to be
changed. Rotating the Control knob will change the
setting. When the setting being changed has the
possibility of irreversibly changing a waveform, an
execute label (EXEC) will appear for the F3 soft key and
must be pressed before the change takes place. The
Trigger, Hold, and Return keys control the waveform in
real time and are not directly associated with the Con­trol knob. The Counter key causes the display to indicate a waveform count as will be discussed below. The front panel controls and connectors are shown in figure 3-1 and keyed (bold numbers) to the following descriptions.
W-TEK
Arbitrar, Waveform Generator
FFtEl;!
F;ER
;!;RplFL
1 Liquid Crystal Display
(See figure 3-2). The lower
row of this display will provide labels for the soft keys that are located directly below the display. Each time one of the menu keys is pressed the labels will be updated to match the menu selection (see Appen­dix B). The upper row of the display will respond to the soft keys and control knob by indicating the setting or parameter value corresponding to the last soft key that was pressed.
2
LCD Viewing Control.
This control allows the user to adjust the intensity of the display to match the viewing angle of the instrument.
3 Edit Key.
Each time the edit key is pressed the lower line of the display will change to indicate three dif­ferent waveform editing options. There are a total of five different sets of three editing options. Any of these can be invoked by pressing the soft key
[Fl
(21), F2 (1
9),
or F3 (1 7)] below the corresponding
label on the display.
NOTE
Refer to appendix B for a summary of soft key options.
'
WAVETEK
Arbitrary Waveform Generator
L
FREQ SELECTED FROM PARAMETER MENU
ADJX
SELECTED FROM EDIT MENU
Figure 3-2. Typical Display Indications
3-1
Scans
by
ArtekMedia
O
2006
Loading...
+ 114 hidden pages