Supplementary documents appropriate to the ordered version come
with the delivery. You can nd them listed in chapter "Product descrip-tion".
Instructions manuals for accessories and replacement parts
Tip:
To ensure reliable setup and operation of your VEGASON 62, we oer
accessories and replacement parts. The corresponding documenta-
tions are:
27835 - Display and adjustment module PLICSCOM
•
32628 - Interface adapter VEGACONNECT
•
27720 - External indication VEGADIS 61
•
34296 - Protective cover
•
30176 - Electronics module VEGASON series 60
•
Editing status: 2013-04-11
3
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of batteries and accumulators.
28785-EN-130417
4
VEGASON 62 • Probus PA
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGASON 62 is a sensor for continuous level measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment.
28785-EN-130417
VEGASON 62 • Probus PA
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety
instructions in this operating instructions manual, the country-specic
installation standards as well as all prevailing safety regulations and
accident prevention rules.
The instrument must only be operated in a technically awless and
reliable condition. The operator is responsible for trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.
2.6 CE conformity
This device fullls the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a conrmation of
5
2 For your safety
successful testing. You can nd the CE conformity declaration in the
download area of "www.vega.com".
2.7 FulllmentofNAMURrecommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recommendations:
NE 21 – Electromagnetic compatibility of equipment
•
NE 43 – Signal level for malfunction information from measuring
•
transducers
NE 53 – Compatibility of eld devices and display/adjustment
•
components
For further information see www.namur.de.
2.8 Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
2.9 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental protection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
Chapter "Packaging, transport and storage"
•
Chapter "Disposal"
•
28785-EN-130417
6
VEGASON 62 • Probus PA
Scope of delivery
3 Product description
3 Product description
3.1 Conguration
The scope of delivery encompasses:
VEGASON 62 ultrasonic sensor
•
Documentation
•
– this operating instructions manual
– Operating instructions manual 27835 "Display and adjustment
module PLICSCOM" (optional)
– Supplementary instructions manual 31708 "Heating for display
and adjustment module" (optional)
– Supplementary instructions manual "Plug connector" (optional)
– Ex-specic "Safety instructions" (with Ex versions)
– if necessary, further certicates
Constituent parts
Application area
The VEGASON 62 consists of the components:
Process tting with transducer
•
Housing with electronics, optionally available with plug connector
•
Housing cover, optionally available with display and adjustment
•
module PLICSCOM
The components are available in dierent versions.
1
2
3
Fig. 1: VEGASON 62 - threaded version with plastic housing
1 Housing cover with integrated PLICSCOM (optional)
2 Housing with electronics
3 Processttingwithtransducer
3.2 Principle of operation
VEGASON 62 is an ultrasonic sensor for continuous level measurement. It is suitable for liquids and solids in virtually all industries,
particularly in the water and waste water industry.
Functional principle
28785-EN-130417
VEGASON 62 • Probus PA
The transducer of the ultrasonic sensor transmits short ultrasonic
pulses to the measured product. These pulses are reected by
7
3 Product description
product surface and received back by the transducer as echoes.
The running time of the ultrasonic pulses from emission to reception
is proportional to the distance and hence the level. The determined
level is converted into an appropriate output signal and outputted as
measured value.
Power supply and bus
communication
GSD/EDD
Power supply via the Probus DP/PA segment coupler or VEGALOG
571 EP cards. A two-wire cable according to Probus specication
serves as carrier of both power and digital data transmission for
multiple sensors. The instrument prole of VEGASON 62 corresponds
to prole specication version 3.0.
The backlight of the display and adjustment module is powered by the
sensor. Prerequisite is a certain level of operating voltage.
The data for power supply are specied in chapter "Technical data".
The optional heating requires its own operating voltage. You can nd
details in the supplementary instructions manual "Heating for display and adjustment module".
This function is generally not available for approved instruments.
The GSD (instrument master les) and bitmap les necessary for
planning your Probus-DP-(PA) communication network are available
from the download section on the VEGA homepage www.vega.com
under "Services-Downloads-Software-Probus". There you can
also nd the appropriate certicates. In a PDM environment, an EDD
(Electronic Device Description) is also required to enable the full
range of sensor functions (also available as a download).A CD with
the appropriate les can be ordered via e-mail under info@de.vega.
com or by phone from one of the VEGA agencies under the order
number "DRIVER.S".
3.3 Operation
The instrument can be adjusted with the following adjustment media:
With the display and adjustment module
•
with the suitable VEGA DTM in conjunction with an adjustment
•
software according to the FDT/DTM standard, e.g. PACTware and
PC
with the adjustment program PDM
•
Packaging
8
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
VEGASON 62 • Probus PA
28785-EN-130417
3 Product description
Transport
Transport inspection
Storage
Storage and transport
temperature
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
Not in the open
•
Dry and dust free
•
Not exposed to corrosive media
•
Protected against solar radiation
•
Avoiding mechanical shock and vibration
•
Storage and transport temperature see chapter "Supplement -
Make sure that all parts of the instrument coming in direct contact
with the process, especially the sensor element, process seal and
process tting, are suitable for the existing process conditions, such
as process pressure, process temperature as well as the chemical
properties of the medium.
You can nd the specications in chapter "Technical data" and on the
nameplate.
Installation position
Moisture
Measuring range
Select an installation position you can easily reach for mounting and
connecting as well as later retrotting of a display and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the display and adjustment module in four
dierent positions (each displaced by 90°).
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain o. This applies mainly to outdoor mounting as well as installation in areas where
high humidity is expected (e.g. through cleaning processes) or on
cooled or heated vessels.
Fig. 2: Measures against moisture penetration
The reference plane for the measuring range is the lower edge of the
transducer.
Make sure that a minimum distance from the reference plane - the
so-called dead band, in which measurement is not possible - is
maintained. The exact value of the dead band is stated in chapter
"Technical data".
28785-EN-130417
10
VEGASON 62 • Probus PA
4 Mounting
2
1
Fig. 3: Minimum distance to the max. level
1 Deadband
2 Reference plane
Information:
If the medium reaches the transducer, buildup can form on it and
cause faulty measurements later on.
132
100%
Pressure/Vacuum
Screwing in
28785-EN-130417
VEGASON 62 • Probus PA
0%
Fig. 4: Measuring range (operating range) and max. measuring distance
1 full
2 empty (max. measuring distance)
3 Measuring range
Gauge pressure in the vessel does not inuence VEGASON 62. Low
pressure or vacuum does, however, damp the ultrasonic pulses. This
inuences the measuring result, particularly if the level is very low.
With pressures under -0.2 bar (-20 kPa) you should use a dierent
measuring principle, e.g. radar or guided microwave.
4.2 Mounting instructions
Screw VEGASON 62 into the mounting socket with an appropriate
spanner applied to the hexagon of the process tting. Max. torque see
chapter "Technical data".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
11
4 Mounting
Installation position
When mounting the sensor, keep a distance of at least 200 mm
(7.874 in) to the vessel wall. If the sensor is installed in the center of
dished or round vessel tops, multiple echoes can arise. These can,
however, be suppressed by an appropriate adjustment (see chapter
"Setup").
2
1
> 200 mm
Fig. 5: Mounting on round vessel tops
1 Reference plane
2 Vessel center or symmetry axis
If you cannot keep this distance you should carry out a false echo
storage before setup. This applies mainly if buildup on the vessel wall
is expected. In this case, we recommend repeating a false echo storage later with existing buildup.
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
down to the lowest point of the vessel bottom.
Socket
12
Fig.6:Vesselwithconicalbottom
28785-EN-130417
Socket pieces should be dimensioned so that the lower end of the
transducer protrudes at least 10 mm (0.394 in) out of the socket.
VEGASON 62 • Probus PA
4 Mounting
ca. 10 mm
Fig. 7: Recommended socket mounting
If the reective properties of the medium are good, you can mount
VEGASON 62 on sockets which are higher than the length of the
transducer. You will nd recommended values for socket heights in the
following illustration. The socket end should be smooth and burr-free,
if possible also rounded. Carry out a false echo storage.
Sensor orientation
28785-EN-130417
VEGASON 62 • Probus PA
h
d
Fig. 8: Deviating socket dimensions
d
100 mm/4"
150 mm/6"
200 mm/8"
h
300 mm/12"
400 mm/16"
500 mm/20"
Align the sensor in liquids as vertical as possible to the product surface to achieve an optimum measurement result.
Fig. 9: Alignment in liquids
To reduce the min. distance to the medium, you can also mount VEGASON 62 with a beam deector. By doing this, it is possible to ll the
vessel nearly to maximum. Such an arrangement is suitable primarily
for open vessels such as e.g. overow basins.
13
4 Mounting
~400x400
45°
~200
Fig.10:Beamdeector
Vessel installations
Agitators
The ultrasonic sensor should be installed at a location where no
installations cross the ultrasonic beam.
Vessel installations such as for example, ladders, limit switches, heat-
ing spirals, struts etc. can cause false echoes that interfere with the
useful echo. Make sure when planning your measuring site that the
ultrasonic signals have a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be
carried out during setup.
If large vessel installations such as struts or supports cause false
echoes, these can be attenuated through supplementary measures.
Small, inclined sheet metal or plastic baes above the installations
scatter the ultrasonic signals and avoid direct false echoes.
α
α
Fig.11:Coversmoothproleswithdeectors
α
α
α
α
If there are agitators in the vessel, a false signal memory should be
carried out with the agitators in motion. This ensures that the interfering reections from the agitators are saved with the blades in dierent
positions.
14
28785-EN-130417
VEGASON 62 • Probus PA
Fig. 12: Agitators
4 Mounting
Inowingmedium
Foam
Air turbulences
Do not mount the instruments in or above the lling stream. Make sure
that you detect the product surface, not the inowing product.
Fig.13:Inowingliquid
Through the action of lling, stirring and other processes in the vessel,
dense foams which considerably damp the emitted signals may form
on the product surface.
If foams are causing measurement errors, the sensor should be used
in a standpipe or, alternatively, the more suitable guided radar sensors (TDR) should be used.
Guided wave radar is unaected by foam generation and is particularly suitable for such applications.
If there are strong air currents in the vessel, e.g. due to strong winds
in outdoor installations or air turbulence, e.g. by cyclone extraction
you should mount VEGASON 62 in a standpipe or use a dierent
measuring principle, e.g. radar or guided radar (TDR).
28785-EN-130417
VEGASON 62 • Probus PA
15
4 Mounting
Standpipe measurement
By using a standpipe (surge or bypass tube), the inuence of vessel
installations, foam generation and turbulence is excluded.
Standpipes must extend all the way down to the requested min. level,
as measurement is only possible within the tube.
max.
1
min.
Fig. 14: Standpipe in the tank
1 Vent hole: ø 5 … 10 mm (0.197 … 0.394 in)
VEGASON 62 can be used from tube diameters of 50 mm (1.969 in).
Avoid large gaps and thick welding joints when connecting the tubes.
Generally carry out a false echo storage.
Measurement in a standpipe is not recommended for extremely
adhesive products.
16
28785-EN-130417
VEGASON 62 • Probus PA
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.