Please read this manual carefully, and also take
note of country-specific installation standards
(e.g. the VDE regulations in Germany) as well
as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal
work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified
VEGA personnel.
2VEGASON 54P … 56P
Note Ex area
Please note the attached safety instructions
containing important information on installation
and operation in Ex areas.
These safety instructions are part of the operating instructions and come with the Ex approved instruments.
Continuous level measurement with ultrasonic
sensors is based on the running time measurement of ultrasonic pulses.
Product description
VEGASON 54 … 56 sensors are a newly
developed generation of extremely compact
ultrasonic sensor for level measurement.
They were developed for liquids and
especially for solids and larger measuring
distances. They are the right choice for
applications in silos, bunkers and material
dumps.
Due to small housing dimensions and process fittings, the compact sensors are an
unobtrusive, and most of all, very costeffective solution for your level measurement
applications. With the integrated display and
a special sensor intelligence in conjunction
with large measuring ranges, they enable
high-precision level measurement and can
be used for applications in which the advantages of non-contact measurement could
never before be realised.
The outputs generate a digital Profibus PA
signal as output, i.e. measurement signal.
Measuring principle
High-performance piezoceramic transducers
emit focused ultrasonic pulses which are
reflected by the product surface. The measurement electronics prepares a precise image of the environment from the reflected
ultrasonic pulses. The transducers work both
as transmitter and receiver. As receiver, the
transducers are high-sensitivity piezo microphones.
Meas.
distance
emission - reflection - reception
The measurement electronics precisely calculates the distance between transducer and
medium from the speed of sound and the
measured running time of the emitted sound
impulse. The distance is then converted into
a level-proportional signal and, in conjunction
with the sensor parameter settings, made
available as a precise, calibrated level value.
Since the speed of sound is subject to temperature influence, the transducer also continuously detects the ambient temperature, so that the level is precisely measured
even in case of varying ambient temperature.
4VEGASON 54P … 56P
Product description
Output signal
The level-proportional Profibus PA measurement signal is processed and outputted
completely digitally. Digital processing of the
measurement signal ensures an accuracy
which could be never reached by an analogue measurement signal, as the digital
signal is always transmitted error-free right
up to the last decimal position. Varying
resistances or tiny leakage currents do not
influence the accuracy of digital technology.
The digital signal is always clear and unambiguous.
The digital signal, giving a faithful representation of the adjusted measuring range of the
sensor, can be modified by various parameter settings.
Measured value display
As an option, the series 50 ultrasonic sensors
can be equipped with an indicating instrument for direct, local level survey. The indicating instrument shows the precise level by
means of the analogue bar graph and the
digital number value. In addition to the indication in the sensor, you can have the level
displayed with the VEGADIS 50 external
indicating instrument at a distance of up to 25
m from the sensor. The external measured
value display operates, like the integrated
display, independently of the PA output signal
and can be modified through individual parameter settings.
1.2 Application features
Two-wire technology
• Power supply and output signal on one
two-wire cable.
• Output signal and signal processing completely digital, therefore maximum accuracy.
• Profibus profile 3 – sensor.
Applications
• Level measurement of solids, liquids and
pastes.
• Gauge measurement, distance measurement, object monitoring and conveyor belt
monitoring.
Rugged and precise
• Measurement unaffected by substance
properties such as density, conductivity,
dielectric constant…
• Suitable for corrosive substances.
• Measuring range 0.8 m … 70 m.
Adjustment options
• With adjustment software VEGA Visual
Operating (VVO) on the PC.
• With detachable adjustment module
MINICOM.
Measured value display
• Measured value display integrated in the
sensor.
• Optional display separate from sensor.
• Adjustment module MINICOM in sensor.
Connection to any process
• G 1 A, DN 50, DN 80, DN 200, DN 250
Approvals
• CENELEC, ATEX, PTB
VEGASON 54P … 56P5
Product description
1.3 Profibus output signal
PROPRO
PROcess
PROPRO
sult of a joint project of thirteen companies
and five universities. The companies Bosch,
Klöckner-Möller and Siemens played a decisive role. The specifications of the bus are
described in the protocol layers 1, 2 and 7 of
the ISO/OSI reference model and are available from the PNO (Profibus user organisation). Layers 3 … 5 have not yet been
developed as a standard, leaving Profibus
with far-reaching perspectives for the future.
Today approx. 600 companies make use of
Profibus technology and belong to the PNO.
Profibus
Specification, Profibus
Periphery and Profibus
mation.
As a process automation bus, Profibus PA
enables power supply over the bus. Up to 32
sensors can be operated on a shielded twowire cable that carries both power supply
and measurement signal. In Ex areas, up to
ten sensors can be connected from the PA
level to one two-wire cable (EEx ia).
Bus structure
The Profibus DP and PA network consists of
up to 126 master and slave participants.
Data are always exchanged from point to
point, with the data traffic being exclusively
controlled and checked by master devices.
Communication is carried out according to
the Token-Passing procedure. This means
that the master holding the Token can contact
the slaves, give instructions, enquire data
and cause the slaves to receive and transmit
data. After the work is done or after a predetermined time interval, the Token is passed
on by the master to the next master.
FIFI
BUSBUS
FIeld
BUS (PROFIBUS) is the re-
FIFI
BUSBUS
FMSFMS
FMS stands for Fieldbus Messaging
FMSFMS
DPDP
DP for Decentralised
DPDP
PP
AA
P
A for Process Auto-
PP
AA
Master-Class 1
is the actual automation system, i.e. the process control computer or the PLC that enquires and processes all measured values.
Master-Class 2
One or several Master-Class 2 can operate in
a Profibus network. As a rule, Master-Class 2
devices are engineering, adjustment or visualisation stations. The VEGA adjustment software VVO (VEGA Visual Operating) operates
as Master-Class 2 participant on the DP bus
and can work on an engineering PC, on an
adjustment PC or on the process control
computer and can access any VEGA sensor
on the PA level.
Instrument master file
A so-called GSD file is delivered with every
VEGASON Profibus sensor. This file is necessary for integrating the sensor into the bus
system. The GSD file (instrument master file)
contains, beside the sensor name and the
manufacturer, the sensor-specific communication parameters which are necessary for a
stable integration of the sensor in the bus.
Load the GSD file belonging to the sensor
into your bus configuration program. If the
GSD file is not available, it can be loaded
from the VEGA homepage:
http://www.vega.com.
Do not confuse the GSD file with the EDD
(Electronic Device Description), a file necessary for the PDM environment (this can be
also found on the VEGA homepage).
6VEGASON 54P … 56P
Product description
1.4 Adjustment
Each measuring situation is unique. For that
reason, every ultrasonic sensor needs some
basic information on the application and the
environment, e.g. which level means "empty"
and which level "full". Beside this "empty and
full adjustment", many other settings and
adjustments are possible with VEGASON
ultrasonic sensors. The output of echo
curves or the calculation of vessel linearisation curves by means of vessel dimensions are only two examples.
Profibus adjustment scheme
In the Profibus environment, there are different adjustment concepts and adjustment
tools which often differ considerably from
manufacturer to manufacturer. From the user’s point of view, a manufacturer-independent adjustment program which could be
operated directly on the Profibus DP, as well
as at any system node (e.g. the engineering
station or the process control), would be
ideal.
In the past, only the program "SIMATIC PDM",
based on the HART® adjustment scheme,
could fulfil this wish (though with the limitations common to HART®). As with HART®, the
availability of an instrument-specific database
for a comprehensive adjustment with PDM
(Process Device Managing) is a requirement.
Otherwise, only the basic instrument functions, such as adjustment, are available. In
the PDM environment, this instrument-specific database is called EDD (Electronic Device Description), in perfect analogy to the
HART® environment which also requires,
except for the VEGA HART® instruments, a
DD (Device Description) for each sensor.
We are aware of the disadvantages of the
HART® environment: for each sensor/participant, an individual DD must be loaded, which
in addition, must always be the latest and
most up-to-date DD. Special adjustment
options such as e.g., the output of an echo
curve, are available neither with HART® nor
with PDM. User-friendly adjustment is out of
the question. With VEGA’s adjustment program VVO, those restrictions belong to the
past.
The legitimate wish of many Profibus users
for a manufacturer-independent adjustment
tool without EDD has been realised in the
form of PACTware
TM 1)
. An association of a
number of process technology companies
developed PACTwareTM: a Process Automation Configuration Tool that can run different
manufacturer software tools under a standardised user interface and adjustment
scheme. Specialists call this technology Field
Device Transcription. Just as different Windows printer drivers enable operation of
completely different printers under a single
user interface, PACTware
TM
enables operation
of all field instruments under a single user
interface. Instrument-specific databases
(EDD), like those required for SIMATIC PDM,
are not necessary.
As a result of this development, four adjustment media are available for VEGA Profibus
sensors:
- adjustment with the PC and the adjustment
program VVO (VEGA Visual Operating) as
stand-alone tool, on the segment coupler
or directly on the sensor.
- adjustment with the detachable adjustment
module MINICOM in the sensor
- adjustment with the SIMATIC PDM adjust-
ment program (requires EDD instrument
databases) from the control room
- adjustment with the manufacturer-inde-
pendent user interface PACTwareTM on the
sensor, from the control room or on the
segment coupler.
VEGASON 54P … 56P7
Product description
Adjustment with the adjustment program VVO - VEGA Visual Operating
The setup and adjustment of the ultrasonic
sensors is generally done on the PC with the
adjustment program VEGA Visual Operating
(VVO) under Windows®. The program leads
quickly through adjustment and parameter
setting by means of pictures, graphics and
process visualisations.
Note:
The adjustment program VVO must be available in version 2.70 or higher.
The VEGA adjustment software VVO (VEGA
Visual Operating) operates either as a
subprogram of the host program PACTware
acc. to the FDT concept (Field Device Tool) or
as an independent adjustment program on
any PC, engineering station or process control computer.
The adjustment program recognises the sensor type
Visualised input of a vessel linearisation curve in the
adjustment program VVO
VEGA’s adjustment program VVO can access the adjustment options of VEGA sensors in their entirety and, if necessary, can
update the complete sensor software. For
communication with Profibus sensors, the
adjustment program either requires a
Profibus-Master-Class2 interface card or the
interface adapter VEGACONNECT 3.
The PC with the Profibus interface card can
be connected directly to any point on the DP
bus with the standard RS 485 Profibus cable.
In conjunction with the adapter
VEGACONNECT, the PC can be connected
directly to the sensor. VEGACONNECT communicates via a small plug directly with the
respective sensor.
The adjustment and parameter data can be
TM
saved at any time on the PC with the adjustment software and can be protected by
passwords. If necessary, the adjustments
can be transferred quickly to other sensors.
In practice, the adjustment program VVO is
often installed as a tool on an engineering
station or an operating station. VVO then
accesses all VEGA sensors directly over the
bus via the Profibus interface card (e.g. from
Softing) as Master Class 2, from the DP level
to the PA level (via segment coupler) right
down to the individual sensor.
Beside the instrument master file (GSD), with
which a sensor is logged into the Profibus
system, the majority of all Profibus sensors
requires for adjustment, beside the specific
adjustment software, also a so-called EDD
(Electronic Device Description) for each
sensor, in order to access and adjust the
sensor from the bus levels. This is not the
case with VVO. The adjustment software
VVO can communicate at any time with all
VEGA sensors without the help of a special
database. Of course, all other non-Profibus
VEGA sensors can be adjusted with the
adjustment software as well (4 … 20 mA
sensors or VBUS sensors). With VEGA sensors, it is not necessary to go looking for the
latest EDD. This is the essential requirement
for a manufacturer-independent adjustment
program, like PACTwareTM, anticipated by
many users, see following pages.
8VEGASON 54P … 56P
Product description
Adr. 21
SPS
Adr. 22
VVO
3
PA-
Adr. 23
Bus
Master-Class 1
Adr. 1
DP-Bus
Adr. 24
Adr. 10
Adr. 25 … 56
2
Profibus DP interface card as
Master-Class 2 (e.g. Softing)
3
Adr. 58
Adr. 57
Segment coupler
(max. 32 participants)
Adr. 60
Adr. 59
Adr. 26
Adr. 25
Adr. 27
Adr. 28
Adr. 29
Adjustment of the VEGASON ultrasonic sensors from process control via a Profibus interface card in the process control computer or in an additional PC. The adjustment software VEGA Visual Operating (VVO) accesses
the sensors bidirectionally via the interface (interface card).
VEGASON 54P … 56P9
Product description
Adjustment with adjustment module
MINICOM
With the small (3.2 cm x 6.7 cm) 6-key adjustment module with display, you carry out
the adjustment in clear text dialogue. The
adjustment module can be plugged into the
ultrasonic sensor or into the optional, external
indicating instrument.
Tank 1
m (d)
12.345
Detachable adjustment module MINICOM
The adjustment module can be easily removed to prevent unauthorised sensor adjustment.
ESC
+
-
Tank 1
m (d)
12.345
OK
2
Tank 1
m (d)
12.345
4
PA-Bus
ESC
+
-
OK
ESC
+
-
OK
Adjustment with the SIMA TIC PDM
adjustment program
To adjust all essential functions of the VEGA
sensor with the adjustment station SIMATIC
PDM from Siemens, a so-called EDD is required. Without this EDD, only the basic functions such as min./max. wet adjustment or
integration time can be adjusted with the
PDM adjustment program. Further important
adjustment functions, such as input of the
measuring environment or false echo storage, are not available without EDD. After
integration of the EDD files in the Simatic PDM
adjustment software, all important adjustment
functions are accessible. If it is not at hand,
the obligatory GSD (instrument master file)
as well as the EDD (Electronic Device Description) necessary for PDM can be
downloaded from the VEGA homepage
(http://www.vega.com).
Adjustment with PACTware
PACTwareTM is a manufacturer-independent
automation/configuration tool through which
access to instruments of different manufacturers (Krohne, Pepperl + Fuchs, VEGA,
WIKA-, Bürkert…) is possible. The VEGA
adjustment software VVO works as a
subprogram/menu. PACTwareTM activates the
required menu options for the sensor/instrument being accessed.
PACTwareTM looks different than VVO and is
designed with a tree structure. Operating
instructions for PACTwareTM can be found in
the PACTware
described in this operating instructions
manual.
TM
documentation. They are not
TM
max. 2.5 m
Adjustment with detachable adjustment module. The
adjustment module can be plugged into the ultrasonic
sensor or into the external indicating instrument
VEGADIS 50.
10VEGASON 54P … 56P
Product description
1.5 T ype survey
VEGASON 54 … 56 sensors are a newly
developed generation of very compact ultrasonic sensors for large measuring ranges.
For shorter measuring ranges, VEGASON
51 … 53 sensors are used.
Thanks to the small housing dimensions and
process fittings, the compact sensors are a
very cost-effective solution for your level
measurement applications. Swivelling holders
enable quick orientation of the transducers to
the liquid or solid surface. Because the sensor electronics and transducer can be separated, installation is very simple. It is possible
to separate the sensor electronics up to
300 m from the transducer. As a result, the
transducer can handle ambient temperatures
up to 150°C (VEGASON 56).
General features
• Application in solids and liquids.
• Measuring range 0.8 m … 70 m.
• Ex-approved in Zone 10 (IEC) or Zone 11
(ATEX) classification mark EEx ia [ia] IIC T6
or Zone 20/21/22, VEGASON 56 also
Zone 1.
• Integrated measured value display in the
sensor or in the external indicating instrument which can be mounted up to 25 m
away from the sensor.
VEGASON 54
Version AVersion BVersion CVersion D
VEGASON 55
Version AVersion BVersion CVersion D
VEGASON 56
Version AVersion BVersion CVersion D
VEGASON 54P … 56P11
1.6 Type code
Product description
VEGASON 54 K EX.XX X X X X X X X
K - Plastic housing PBT, M20 x 1.5 cable entry
N - Plastic housing PBT,
A - Aluminium housing, M20 x 1.5 cable entry
FEP - Version A, flange DN 200 (PP)
FEA - Version A, flange DN 200 (Aluminium)
FLP - Version A, flange 8" (PP)
FLA - Version A, flange 8" (Aluminium)
SAS - Version B, flange swivelling holder DN 50
SBS - Version B, flange swivelling holder DN 80
GAS - Version C, flange swivelling holder DN 50
GBS - Version C, flange swivelling holder DN 80
RGS - Thread G 1 A
YYY - Other process fittings
X - without display
A - with integrated display
X - without adjustment module MINICOM
B - with adjustment module MINICOM (mounted)
A - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA (four-wire)
B - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA, HART
(four-wire)
E - Power supply via signal conditioning instrument
G - Segment coupler for Profibus PA
P - 90 … 250 V AC (only in USA)
N - 20 … 36 V DC, 24 V AC (only in USA)
Z - Supply via signal conditioning instrument (only in USA)
.X - without approval
EXS.X - StEx Zone 10
K - Analogue 4 … 20 mA output signal (two-wire or
four-wire technology)
V - Digital output signal (two-wire technology)
P - Profibus
Type 54 - Measuring range 1.0 … 25 m
Type 55 - Measuring range 0.8 … 45 m
Type 56 - Measuring range 1.6 … 70 m
Measuring principle (SON for ultrasonic)
½" NPT cable entry
®
12VEGASON 54P … 56P
Mounting and installation
2 Mounting and installation
2.1 Mounting
Version A
Sensors in version A (flange version) are
supplied completely assembled and ready
for operation. Immediately after mounting on
the vessel and electrical connection, they are
ready for operation.
Version B
Sensors in version B are supplied in two
parts (transducer and sensor electronics).
First of all, mount the transducer on the vessel or above the medium. There is a four-pole
jack at the end of the transducer tube. The
respective counterpart to the jack protrudes
out of the lower side of the sensor electronics. Insert the plug of the sensor electronics
(only possible in one position) into the jack of
the transducer tube. Continue pushing the
electronics housing onto the transducer tube,
on which there is a wide and a narrow
groove.
Groove for locking the
headless screw
Mounting groove (must
no longer be visible after
mounting)
The wide groove is used for locking the
headless screws. The narrow groove is an
assembly mark. Move the electronics housing farther down over the transducer tube
until the mounting groove is no longer visible.
Fasten the housing with the headless screws
to the transducer tube. Use a 5 mm hexagon
screwdriver (or Allen wrench).
VEGASON 54P … 56P13
Mounting and installation
Ver sion C and D
The sensors in version C and D are supplied
in three parts (transducer, sensor electronics
and transducer cable). First mount the transducer (see version B). There is a four-pole
jack at the transducer tube end. A respective
counterpart to the jack is provided in the
connection cylinder of the transducer cable.
Insert the connection cylinder plug into the
jack of the transducer tube. S
Connection
cylinder
Mounting
bracket
Connection
cylinder
Transducer
cable
On the end of the transducer tube you find a
wide and a narrow groove. The wide groove
is used for locking the cylinder with the headless screws. The narrow groove is the assembly mark.
Then push the connection cylinder onto the
transducer tube (with a slight swivelling motion) until the assembly mark is no longer
visible.
When the assembly mark is covered by the
cylinder, fasten the cylinder with the two
headless screws. Use a 5 mm hexagon
screwdriver (or Allen wrench).
Now mount the sensor electronics at the
requested location. The sensor electronics is
fastened to a mounting bracket so that it can
be mounted on a plain surface or on a wall.
Make sure that the sensor electronics is
mounted in such a way that there is enough
space above the electronics to open the
cover.
Now insert the plug at the other end of the
transducer cable into the jack on the electronics housing.
Note:
Avoid bending the transducer cable too
sharply when laying it out. This is a special
cable which could otherwise be damaged.
In addition, make sure that the cable cannot
be damaged during operation. A signal with
a voltage of approx. 1 kV is transmitted
(which could be a danger in Ex areas if the
cable is damaged) via the shielded cable
cores.
Groove for locking the
headless screws
Mounting groove (must
no longer be visible after
mounting)
14VEGASON 54P … 56P
Mounting and installation
2.2 General installation instructions
Measuring range
version, the instrument flange (version A).
Please note the information on the reference
Beside other criteria, you select your instrument according to the required measuring
range. The reference plane for the min. and
planes in chapter "7.4 Dimensions". The max.
filling depends on the required min. distance
and the mounting location.
max. distance to the liquid or solids is the
transducer end, or for instruments in flange
Reference plane
min. distance
1.0 m
100 %
0 %
Span
Min. distance, max. measuring range, span and reference plane
0 %
max. meas. range
max. meas. distance 25 m (type 54), 45 m (type 55), 70 m (type 56)
min. distance 0.8 m
100 %
Span
Reference plane
min. distance 1.4 m
100 %
0 %
Span
Beam angle and false echoes
At greater distance, the energy of the ultrasonic impulses distributes over a large area,
The ultrasonic impulses are focused by the
transducers. The impulses leave the transducer in a conical form similar to the beam
pattern of a spotlight. The beam angle is 4°
thus causing weaker echoes from obstructing surfaces. The interfering signals are
therefore less critical than those at close
range.
(VEGASON 54), 5° (VEGASON 55) and 6°
(VEGASON 56) at -3 dB emitted power.
Any object, e.g. tubes or struts inside this
emission cone will cause a large false echo.
Especially within the first few meters of the
If possible, orient the sensor axis perpendicularly to the product surface and avoid
vessel installations (e.g. pipes and struts)
within the 100 % area of the emission cone.
emission cone, pipes, struts, or other installations can interfere with the measurement. At a
distance of 6 m, the false echo of a strut has
an amplitude nine times greater than at a
distance of 18 m.
VEGASON 54P … 56P15
The following illustration of the ultrasonic
beams is simplified and represents only the
main beam - a number of additional weaker
beams exists.
0 m
Meas.
distance
Mounting and installation
VEGASON 56VEGASON 54
0 m
Emitted power
50 %
100 %
4˚
Emitted power
25 m
Meas.
distance
45 m
0 m
4
0,9
2,0
5˚
10˚
242
0
8˚
m
0,9
2,0
0
VEGASON 55
50 %
Emitted power
100 %
m
Emitted power
Meas. distance
50 %
Emitted power
6˚
100 %
Emitted power
12˚
70 m
3,7
7,5
3,7
0
m
7,5
Therefore, in practical application, the transducer has to be oriented so that the lowest
possible false echo signal strength is
achieved. Only giving attention to the size of
the useful echo is usually not adequate under
difficult measuring conditions. In most cases,
a low false echo level enables the sensor to
reliably pick up the useful echo. With the
adjustment software VVO on the PC, you can
view the echo image (see manual VVO "Sensor optimisation".
16VEGASON 54P … 56P
Mounting and installation
2.3 Measurement of liquids
Flat vessel top
On flat vessels, mounting is usually done on
a very short DIN socket piece. The lower
side of the flange is the reference plane. The
transducer should protrude out of the flange
pipe.
< 100 mm
VEGASON 55 in flange version on short DIN socket
piece
< 400 mm
Reference plane
Min. distance
Type 54: 1 m
Type 55: 0.8 m
Reference plane
A mounting location directly on the vessel top
is ideal. A round opening in the vessel top is
sufficient to fasten the sensor with the flange
or version B and C with the swivelling holder.
Reference
plane
Min. meas. distance
1.8 m
Swivelling
holder
Min. meas. dis-
tance 1.4 m
Reference
plane
Flange version and swivelling holder on flat vessel
top
It is also possible to mount version C sensors
in a 1" thread.
Min. distance
1.8 m
< 60 mm
Reference plane
VEGASON 56 in flange version on short DIN socket
piece
Mounting of the transducer in a 1" thread
VEGASON 54P … 56P17
Dished tank top
On dished tank tops, please do not mount
the instrument in the centre or close to the
vessel wall, but approx. ½ vessel radius from
the centre. Dished tank ends can act as
paraboloidal reflectors. If the transducer is
placed in the focal point of the parabolic tank,
the ultrasonic sensor receives amplified false
echoes. The transducer should be mounted
outside the focal point. Parabolically amplified
echoes can be thereby avoided.
Mounting and installation
Reference plane
< 100 mm
½ vessel radius
VEGASON 54 on dished tank top; applies also to
VEGASON 55
< 400 mm
1
/2 vessel radius
VEGASON 56 on dished tank top
Reference plane
Open vessels
On open vessels, the sensors can be readily
installed on extended mounting brackets.
Mount the low-weight sensor onto such a
bracket and ensure a sufficient distance from
the vessel wall.
Reference plane
Min. meas.
Min. meas.
distance
distance
Reference plane
Open vessel
18VEGASON 54P … 56P
Mounting and installation
Pump shaft
Narrow, uneven shafts, wells and vessel
openings with very rough walls and shoulders make ultrasonic measurements extremely difficult due to strong false echoes.
This problem can be overcome by using an
extended socket piece or a complete measuring tube (see chapter "2.5 Socket extension").
see "2.5 Socket extensions"
Socket piece
Shaft pump
Measuring tube socket or measuring tube in a shaft
≥ 250 mm
Shaft pump
Measuring tube
min.
distance
Meas. range
2.4 Measurement of solids
Flange mounting
For containers with solids, as with applications for liquids, the instrument can be
mounted on a short DIN socket piece. The
transducer end should protrude out of the
socket so that it does not interfere with ultrasonic signals. The swivelling holder enables
not only correct orientation to the product
surface but also minimisation of possible
false echoes.
Different filling conditions often lead to a varying product surface orientation. This causes
the useful echo to vary in quality. For this
reason, the transducer should be mounted in
such a way that, even in an empty vessel, the
false echo intensity remains as low as possible. You can view the echo curve on the PC
with the adjustment program VVO.
Reference plane
Shaft
Min. distance
Very good measuring results can be attained
with a measuring tube in continuous narrow
shafts, see illustration. The applied measuring tube must have smooth walls inside (e.g.
PE sewage pipe) and a diameter ≥ 200 mm.
This arrangement works well as long as the
inside of the measuring tube collects no dirt
or buildup (cleaning necessary). You might
want to consider using hydrostatic pressure
transmitters or capacitive measuring probes.
The measuring tube should either be never
immersed in the medium, or always immersed, making the measurement take place
exclusively in the tube.
VEGASON 54P … 56P19
VEGASON 54C with adapter flange on a DN 200
vessel flange
Mounting and installation
Mounting boss
Reference plane
Min. distance
VEGASON 56 in 1" mounting boss.
The socket axis should be oriented to the
product surface. More suitable is the use of a
swivelling holder version (type B, C), as
described on previous pages.
Material heaps
2.5 Socket extensions
Ultrasonic sensors require a min. distance to
the liquid or solid. Take the min. distance into
account in your planning. In some situations,
it is possible to reach the required min. distance, and hence the desired filling height
with a socket extension. However, the socket
piece increases the noise level of the ultrasonic signal at the extension outlet and can
interfere with the measurement. Only use a
socket extension if all other possibilities have
to be excluded. Carry out the extension as
shown in the following illustration.
Socket extensions for liquids
Chamfer and deburr the socket carefully and
make sure that it has a smooth inner surface.
The socket should not protrude into the
measured product, in case buildup can form
on the socket through dirt or product
residues.
For non-adhesive measured products, a
socket extension in the form of a measuring
tube can be permanently submerged in the
product. The ultrasonic measurement is then
made exclusively in the measuring tube and
works very well without interference from
other vessel installations (see "Pump shaft").
Large material heaps are usually measured
with several instruments, which can be
mounted on e.g. traverse cranes. If material
cones form, we recommend orienting the
sensors to the solid surface.
Socket piece should not be immersed into adhesive
products (figure: VEGASON 54)
Transducer on traverse crane above a material heap
20VEGASON 54P … 56P
Mounting and installation
The socket diameter should be as large and
the socket length as small as possible. To
minimise false echoes, make sure that the
socket outlet is burr-free.
Typ e 5 4
L
45˚
ø
Socket extensions not immersed into the medium
Typ e 5 5
L
45˚
ø
Max. socket length in relation to socket
diameter
ø in mmL in mm
Type 54Type 55Type 56
200400– –––
250500500500
300––––600
2.6 False echoes
The mounting location of the ultrasonic sensor
must be selected such that no installations or
inflowing material are in the path of the ultrasonic impulses. The following examples and
instructions show the most frequent measuring problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to
their strong false echoes, adversely effect
the measurement. Shields above these flat
protrusions scatter the false echoes and
guarantee a reliable measurement.
CorrectIncorrect
Socket extension in solids
For solids, use a conical socket extension
with a taper of at least 15° … 20°.
Vessel protrusions (slope)
Intake pipes, e.g. for the mixing of materials,
with a flat surface directed towards the sensor should be covered with a sloping shield.
The shield will scatter the false echo.
CorrectIncorrect
15˚ 15˚
Socket extension in solids
Vessel protrusions (intake pipe)
VEGASON 54P … 56P21
Mounting and installation
Vessel installations
Vessel installations such as, for example, a
ladder, often cause false echoes. Make sure
when planning your measurement loop that
the ultrasonic signals have free access to the
measured product.
CorrectIncorrect
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can
cause strong false echoes that are superimposed over the useful echo signal. Small
shields effectively hinder a direct false echo
reflection. The false echoes are scattered
and diffused in the area and are then filtered
out as "echo noise" by the measuring electronics.
Inflowing material
Do not mount the instrument in or above the
filling stream. Ensure that you detect the
product surface and not the inflowing material.
Correct
Correct
Inflowing material
Incorrect
Incorrect
Buildup
CorrectIncorrect
If the sensor is mounted too close to the
vessel wall, residues or buildup of the measured product on the vessel wall can cause
false echoes. Position the sensor at a sufficient distance from the vessel wall. Please
Shields
also note chapter "2.2 General installation
instructions".
Struts
22VEGASON 54P … 56P
Mounting and installation
Correct
Buildup
Incorrect
Strong product movements
Strong turbulence in the vessel, e.g. caused
by powerful stirrers or intense chemical reactions, seriously interfere with the measurement. A surge or bypass tube of sufficient
size (DN 200, DN 250) always allows, provided the product causes no buildup in the
tube, a reliable measurement even with
strong turbulence in the vessel.
100 %
2.7 Installation mistakes
Foam generation
Thick foam on the product can cause incorrect measurements. Take measures to avoid
foam, carry out the measurement in a bypass
tube, or use a different measurement technology, e.g. capacitive probes or hydrostatic
pressure transmitters.
Foam generation
Wrong orientation to the product
Weak measuring signals are the result if the
sensor is not directly oriented to the product
surface. Orient the sensor axis perpendicularly to the product surface to achieve optimum measuring results.
60 %
0 %
Strong turbulence
Orient the sensor perpendicularly to the product
surface
VEGASON 54P … 56P23
Mounting and installation
Strong temperature fluctuations
Strong temperature fluctuations, e.g. caused
by the sun, cause measuring errors. Please
provide a sun shield.
Sun shield
Strong heat fluctuations
Min. distance to the medium
If the min. distance to the medium is not maintained, the instruments show wrong measured values. Mount the instrument at the
required min. distance.
Sensor too close to vessel wall
CorrectIncorrect
If the sensor is mounted too close to the
vessel wall (dimension A in diagram), strong
false echoes can be caused. Buildup, rivets,
screws or weld joints on the vessel wall superimpose their echoes on the product, i.e.
the useful echo. Please ensure a sufficient
distance from the sensor to the vessel wall
depending on the maximum measuring distance (dimension B in diagram). In case of
good reflecting conditions (liquids, no vessel
installations), we recommend determining the
sensor distance according to Diagram curve
1. At a max. measuring distance of e.g. 10 m,
the distance of the transducer (acc. to curve
1) should be approx. 1.5 m. In case of solids
with poor reflective properties, determine the
distance to the vessel wall according to Dia-gram curve 2. Under very bad measuring
conditions (rough vessel walls, struts), it
might be necessary to increase the distance
to the vessel wall, or to also filter out the false
echoes by storing them in memory, thereby
adapting the sensor more precisely to the
environment.
Distance from the
transducer to the
vessel wall
A
2 m 4 m 6 m 8 m
Curve 1 (liquids)
10 m
B
20 m
Sensor too close to the vessel wall
24VEGASON 54P … 56P
30 m
max. meas.
distance
Curve 2 (solids)
Mounting and installation
Parabolic effects of rounded or arched
vessel tops
Round or parabolic tank tops act like a parabolic mirror on the signals. If the sensor is
placed at the focal point of such a parabolic
tank top, the sensor receives amplified false
echoes. The optimum location is generally in
the area of half the vessel radius from the
centre.
Correct
< 100 mm
~ ½
vessel
radius
Incorrect
Socket piece too long
If the sensor is mounted in a socket piece
that is too long, strong false echoes are generated and measurement is made more difficult. Make sure that the transducer protrudes
at least 30 mm out of the socket piece.
Reference plane
< 100 mm
Incorrect
Correct and wrong length of socket piece
Mounting on a vessel with parabolic tank top
VEGASON 54P … 56P25
3 Electrical connection
Electrical connection
3.1 Connection
Safety information – Qualified personnel
Instruments which are not operated with
protective low voltage or DC voltage must
only be connected by qualified personnel.
This also applies to the configuration of
measuring systems planned for Ex environment.
As a rule, do all connecting work in the complete absence of line voltage. Always switch
off the power supply before you carry out
connecting work on the ultrasonic sensors.
Protect yourself and the instruments.
Note, important for connection to an EP
card
If VEGASON 54P … 56P sensors are operated on Profibus input cards (EP cards) in
the VEGALOG 571 processing system, each
card must be powered individually with an
own power supply unit. For power supply of
an EP card, use the power supply unit "ASIInterface SLA 3.100" from Puls which is available from VEGA.
By this measure, the signal circuits are galvanically separated. This avoids possible
crosstalks between the EP cards.
3.2 Connection cables and bus configuration
Note the Profibus specification. The connection cables must be specified for the expected operating temperatures in the plant
and must have an outer diameter of
6 … 12 mm, to ensure the seal effect of the
cable entry on the sensor.
For power supply and bus communication, a
two-wire cable acc. to the Profibus specification (up to max. 2.5 mm2 cross-section area
of conductor) is used. The electrical connection on the sensor is made by spring-loaded
terminals.
In a laboratory setup, a Profibus system will
also work with standard, unshielded two-wire
cable. In practice however, an automation
network and bus system can only be protected reliably against electromagnetic interference with screened cable. Acc. to the
Profibus specification (IEC 1158-2),
screened and twisted cables are prescribed.
All participants are connected in one line
(serially). At the beginning and end of the
bus segment, the bus is terminated by an
active bus termination. On the DP bus level,
most participants already have a bus termination implemented. With more than 32 participants on the DP level, a so-called repeater
must be used to open and combine another
DP level with a max. of 32 additional participants. On the PA bus branch of the segment
coupler, the PA ultrasonic sensors also work
with a maximum of 32 participants (Ex max.
10 participants).
A PA sensor can work only in conjunction with
a Profibus DP system, to which a Profibus PA
subsystem is connected. A PA Profibus participant must consume min. 10 mA supply
current.
26VEGASON 54P … 56P
Electrical connection
Cable length
Connection cables must correspond to the
Profibus specification and the FISCO model.
The sensor cable must be in conformity with
the values of the reference cable acc. to
IEC 61158-2:
0.8 mm2; R
Z
= 80 … 120 Ω; damping = 3 dB/km;
31.25kHz
C
asymmetric
The max. cable length first of all depends on
the transmission speed:
up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr
up to 94 kbit/s: 1200 m Profibus DP
up to 188 kbit/s: 1000 m Profibus DP
up to 500 kbit/s: 500 m Profibus DP
up to 1500 kbit/s: 200 m Profibus DP
up to 12000 kbit/s: 100 m Profibus DP
The distributed resistance of the cable, in
conjunction with the output voltage of the
segment coupler and the current requirement
(VEGASON 10 mA) or the voltage requirement (VEGASON 9 V) of the sensors, determines the max. length of the cable.
In a practical application of a PA bus branch,
the max. length of the cable is also determined (beside the required supply voltage
and max. current consumption of all participants on the PA bus branch) by the bus
structure and the type of segment coupler
used.
The cable length results from the sum of all
cable sections and the length of all stubs.
The length of the individual stubs must not
exceed the following lengths:
1 … 12 stubs120 m (Ex: 30 m)
13 … 18 stubs60 m (Ex: 30 m)
19 … 24 stubs30 m (Ex: 30 m)
More than 24 stubs are not permitted,
whereby each branch longer than 1.2 m is
counted as a stub. The total length of the
cable must not exceed 1900 m (in Ex version
1000 m).
= 44 Ω/km;
DCmax.
= 2 nF/km.
ofibus Pofibus P
ofibus P
ofibus Pofibus P
AA
A
AA
Ground terminal
The electronics housings of the sensors have
a protective insulation. The ground terminal in
the electronics is galvanically connected with
the metallic process connection. For sensors
with a plastic thread as process fitting, the
sensor must be grounded by connecting a
ground cable to its external ground terminal.
Screening
"Electromagnetic pollution" caused by electronic actuators, energy cables and transmitting systems has become so pervasive that
shielding for normal two-wire bus cable is
usually a necessity. According to the Profibus
specification, the screening should be made
on both ends. To avoid potential equalisation
currents, a potential equalisation system
must be provided in addition to the screening.
According to specification, we recommend
the use of twisted and screened two-wire
cable, e.g.: SINEC 6XV1 830-5AH10 (Siemens), SINEC L26XV1 830-35H10 (Siemens),
3079A (Belden).
Alternatively, when grounding at both ends in
non-Ex areas, the cable shielding can be
connected on one ground side (in the switching cabinet) via an YC-capacitor (e.g. 10 nF,
1500 V) to the ground potential. Make sure
that the ground connection has the lowest
possible resistance (foundation, plate or
mains earth).
Profibus PA in Ex environment
When used in Ex area, a PA bus with all connected instruments must be carried out in
intrinsically safe protection class "i". Four-wire
instruments requiring separate supply must
at least have an intrinsically safe PA connection. VEGA sensors for PA-Ex environment
are generally "ia” two-wire instruments.
VEGASON 54P … 56P27
Electrical connection
In the so-called Fieldbus Intrinsically Safe
Concept (FISCO), the general conditions for
an Ex safe bus configuration have been laid
down. Therein, the electrical data of the
participants and the bus cable have been
determined, so that the linking of these components always meets Ex requirements. This
makes the otherwise time-consuming Ex
calculation unnecessary. You can build your
Ex bus according to the IEC standard
1158-2.
The Ex segment coupler delivers a controlled
power supply to the PA bus. It acts as
source in the PA branch. All other components (field instruments and bus terminators)
are only consumers. A field instrument must
consume at least 10 mA. Ideally, an individual
sensor should not consume more than
10 mA, so that the number of instruments can
be as large as possible.
VEGA PA sensors, whether Ex or non Ex,
consume a constant current of 10 mA. According to the Profibus specification, this is
the minimum participant current. With VEGA
sensors it is therefore possible to connect 10
sensors (also in Ex environment) even with a
limited energy supply from the Ex segment
couplers.
Watch out for potential losses
Due to potential transfer, earthing on both
sides without a potential equalisation system
is not allowed in Ex applications. If an instrument is used in hazardous areas, the required regulations, conformity and type
approval certificates for systems in Ex areas
must be noted (e.g. DIN 0165). Please also
note the approval documents with the safety
data sheet attached to the Ex sensors.
In a Profibus system composed of Profibus
DP and Profibus PA subsystem, each participant must have a unique address. Each
participant, whether master or slave, is
accessed by means of its own address in
the bus system. The address of a participant, whether on DP or PA level, should be
assigned before connecting to the bus, because an address can be used only once. If
an address is used twice, interference in the
bus will be caused.
The address of an ultrasonic sensor can be
set in two ways:
- with the adjustment software VVO (software addressing) or
- with the DIP switch block in the sensor
(hardware addressing).
VEGA Profibus sensors are delivered with
the address set at 126 (all DIP switches to
"ON").
Remember, in a Profibus system there are
max. 126 participants possible. When the
DIP switch is set to address 126 (or higher),
the address can be adjusted with the adjustment software VVO, the adjustment module
MINICOM or another configuration tool (e.g.
PDM). However, there can be only one sensor on the bus with address 126 (delivery
status) during address assignment via software. For that reason, hardware addressing
(DIP switch) before connection to the bus is
recommended.
Hardware addressing
The DIP switches generate an address
number in the binary system. This means
that, from right to left (ascending), any switch
represents a number twice as high as the
previous switch on the right. The corresponding number in the decimal system
results from the sum of all switches set to
"ON". In the illustration, you see the decimal
number that corresponds to each individual
DIP switch.
DIP switch 8 corresponds to the number 128,
switch 1 corresponds to the number 1 and
switch 3 corresponds to the decimal number
4.
1
2
8765 4
128
64
32
Example 1
The switches 3, 5 and 7 are set to "ON". The
address is then:
DIP switch 3 to "ON" means 4
DIP switch 5 to "ON" means 16
DIP switch 7 to "ON" means 64
The sum is:
4 + 16 + 64 = Address 84
3
1
2
4
8
16
ON
1
2
8765 4
64
64 + 16 + 4 = 84
VEGASON 54P … 56P29
3
16
4
Electrical connection
Example 2
You want to set address 27.
16 + 8 + 2 + 1 = 27
You must set the DIP switches
5 = 16
4 = 8
2 = 2
1 = 1
to "ON".
Example 3
You want to set address 99
64 + 32 + 2 + 1 = 99
You must set the DIP switches
7 = 64
6 = 32
2 = 2
1 = 1
to "ON".
Software addressing
The DIP switches must be set to an address
of 126 … 255, i.e.
- either all DIP switches are set to "ON",
corresponding to address 255 (delivery
status)
OFF
1
2
64
3
ON
8765 4
32
16
1
2
3
1
2
4
8
8765 4
Addr.
- or only DIP switch 8 is set to "ON", corresponding to address 128.
128
Of course, software addressing is also possible, if the switches 7 … 2 are set to "ON"
(address 126).
The adjustment of the address with software
VVO is described in chapter "6.2 Adjustment
with VVO" under the heading "Software addressing" or in chapter "6.3 Sensor adjustment with the adjustment module MINICOM".
30VEGASON 54P … 56P
Electrical connection
3.4 Connection of the sensor
After mounting the sensor at the measurement location according to the instructions in
chapter "4 Mounting and installation", loosen
the closing screws on top of the sensor. The
sensor lid with the optional indication display
can then be opened. Unscrew the sleeve nut
and slip it over the connection cable (after
removing about 10 cm of insulation). The
sleeve nut of the cable entry has a self-locking ratchet that prevents it from opening on
its own.
Version with Aluminium housing
Power supply and
Profibus signal
+
To the indicating instrument in the
sensor lid or to the external
indicating instrument VEGADIS 50
–
M20x1.5 (diameter of the
connection cable
6…9 mm)
Now insert the cable through the cable entry
into the sensor. Screw the sleeve nut back
onto the cable entry and clamp the stripped
wires of the cable into the proper terminal
positions.
The terminals hold the wire without a screw
(spring-loaded terminals). Press the white
opening tabs with a small screwdriver and
insert the copper core of the connection
cable into the terminal opening. Check the
hold of the individual wires in the terminals by
lightly pulling on them.
Version with plastic housing
Power supply and
Profibus signal
M20x1.5 (diameter of the
connection cable
6…9 mm)
+
To the indicating display
in the sensor lid or to the
external indicating
instrument
3.5 Connection of the external indicating instrument
Loosen the four screws of the housing lid on
VEGADIS 50.
The connection procedure can be facilitated
by fixing the housing cover during connection work with one or two screws on the right
of the housing.
Adjustment
module
OUTPUT
(to the sensor)
3
2
1
4
5
8
6
7
VEGADIS 50
+
-
Tan k 1
m (d)
12.345
Electrical connection
ESC
OK
Voltage supply and
digital meas. signal
-
+
8 7654
+1 2-5 6 7 8
Addr.
Bus
Tank 1
m (d)
12.345
DISPLAY
(in the lid of the
indicating
instrument)
1
2
3
Display
ON
ESC
+
-
OK
Screws
32VEGASON 54P … 56P
Electrical connection
3.6 Bus configuration
The type of ultrasonic sensor you use depends on the process requirements and
mounting conditions, as well as on the requirements of your control, regulative, or
process control system.
VEGASON 54 … 56 Profibus sensors are
sensors for use in Profibus PA environment.
Profile 3 has been implemented in the sensors. A measuring system consists of one or
several sensors, one or several segment
couplers and one DP master computer, such
as e.g. a S7 PLC with Profibus interface or a
process control system with Profibus DPMaster-Slot. The processing unit, e.g. the
PLC, evaluates the level-proportional, digital
measuring signals in a number of evaluation
routines and puts them to use process-specifically.
On the following four pages you will see
schematic illustrations of the bus configuration.
The automation system as Master-Class 1
takes over the complete bus control. It reads
out all signals cyclically and, if necessary,
gives instructions to the participants (e.g.
sensors). Additional master systems (e.g.
visualisation systems or adjustment tools)
can also be connected to the DP bus. These
systems operate as so-called Master-Class 2
participants. Like the Master-Class 1 system,
they can read out signals, give instructions
and operate in the acyclical mode.
Each participant on the bus must have an
unambiguous address. The addressing
covers both bus levels. A Profibus DP network can have max. 126 participants, including all participants on the PA level. In
practice, each Master-Class 1 computer gets
address 1 and the Master-Class 2 computers address 10 … 20. As a rule, the slaves or
participants get the addresses 21 … 126. On
the Profibus PA network segment, a max. of
32 sensors can be connected on one PA
segment coupler.
Ex environment
In Ex environment, intrinsically safe (EEx ia)
PA sensors are used with Ex segment couplers. Generally, the number of PA sensors
on a segment coupler (Ex or non Ex) depends on the current requirement of the
sensors and on the current supplied by the
segment coupler. Segment couplers for
EEx ia environment provide 90 … 110 mA.
The number of sensors results from the sum
of:
- the basic current intake of all sensors
- plus 9 mA communication signal
- plus the leakage currents of all sensors
- plus a recommended current reserve
(approx. 10 mA)
A DP bus does not allow power supply via
signal cable, whereas the PA bus does. Both,
DP and PA, require at least one screened
two-wire cable. The DP bus can additionally
have up to 8 cores (screened), some of
which can be supply cables (see also "Installation Guides PA + DP" of the Profibus User
Organisation (PNO).
VEGASON 54P … 56P33
Profibus PA sensors on the Profibus network
1
Electrical connection
Master-Class 1
Bus terminator
3...9
Profibus PA (31,25 kBit/s)
Profibus DP
21
Profibus interface
card
RS 232
22...54
3
RS 485
10
Master-Class 2
Segment coupler
Bus terminator
2
22
23
24
53
54
VEGACONNECT 3
PA segment on segment coupler:
1 … 32 sensors on one two-wire cable
(Ex: 10 sensors)
34VEGASON 54P … 56P
Electrical connection
Profibus DP segment level
1 … 126 participants including all DP and PA participants.
Through segment couplers and PA segments, the transmission rate, also on the DP level, is determined by the
slowest coupler/participant on the Profibus DP and PA
network.
Bus terminator
3...9
M
Segment coupler
3...9
3~M
89
90
Bus terminator
2
55
56...88
2
Profibus PA
56
57
87
88
PA segment:
1 … 32 sensors on one two-wire cable
(Ex: 10 sensors)
VEGASON 54P … 56P35
Electrical connection
Profibus PA sensors with 4 … 20 mA sensors on the Profibus network
1
Master-Class 1
Bus terminator
VEGALOG
Profibus PA (31,25 kBit)
3~M
3…9
Profibus DP
Profibus interface
card
21
3
RS 485
22
10
Master-Class 2
VEGACONNECT 3
4
RS 232
1
2
3
5
11
4
1 … 15 PA sensors per two-wire cable
13
12
15
14
with independent address zone
(Ex: 10 sensors)
36VEGASON 54P … 56P
Electrical connection
Profibus DP segment level
1 … 126 participants including all DP and PA participants.
Up to 12 MBit/s transmission rate on DP level.
In the PA segments 31.25 kBit/s transmission rate.
VEGACONNECT 3
4 … 20 mA (HART )
2
4
4
4
2
2
2
Profibus PA (31,25 kBit)
3…9
23
M
3…9
24
25
VEGALOG
VBUS
Outputs
2
2
2
2
®
2
2
2
2
2
0/4…20 mA
0…10 V
2
VBUS
Bus terminator
Profibus PA:
1 … 15 sensors per two-wire cable
(Ex: 10 sensors) with independent address zone
VBUS:
1 … 15 sensors per twowire cable
Exd: also 15
Ex ia: 5 sensors
VEGASON 54P … 56P37
4 Setup
4.1 Adjustment media
Adjustment with PACTware
Setup
TM
In chapter "1.4 Adjustment" the Profibus adjustment scheme was briefly explained and
the adjustment media for VEGA Profibus
sensors were shown. All VEGA Profibus
sensors operate in profile 3 and can be adjusted with:
- the adjustment program VVO on a PC with
Profibus card
- the adjustment program PACTwareTM, under
which VVO runs as a subprogram
- the Siemens software PDM in conjunction
with an EDD (Electronic-Device-Description)
- the adjustment module MINICOM in the
sensor.
Adjustment with VV O on the PC
The adjustment program VVO enables userfriendly adjustment of VEGA Profibus PA
sensors. All functions and options relating to
the sensor adjustment are accessible. The
program runs under Windows® on a PC with
a Profibus-Master-Class 2 interface card on
Profibus DP level as Master-Class 2 tool. The
VVO program accesses the VEGA PA sensors via the DP bus, the segment coupler
and the PA bus.
The adjustment with PACTwareTM corresponds to VVO adjustment. In this case,
VVO runs as a subprogram of PACTwareTM.
The adjustment instructions can be found in
the documentation of PACTwareTM.
Adjustment with PDM
The sensors can be adjusted completely with
PDM. However, some convenient functions
and many special features, like e.g. display
of an echo curve, are not available. In addition to the PDM software, an EDD (available
upon request from VEGA) is required for
each sensor type. The adjustment instructions for PDM are described in the PDM
documentation.
Adjustment with the adjustment module
MINICOM
With the adjustment module MINICOM, you
adjust the individual sensor directly in the
sensor or in the external indicating instrument
VEGADIS 50. The adjustment module
MINICOM enables (with the 6-key adjustment
field with text display) all essential functions
of parameter setting and adjustment.
Note:
Please note that for adjustment with VVO,
version 2.70 or higher is required.
38VEGASON 54P … 56P
Setup
4.2 Adjustment with the PC
The instructions for the adjustment with the
PC and the adjustment program VVO can be
found in the manual "VEGA Visual Operating
(VVO)" which is supplied with the latest adjustment software.
Beside the adjustment of ultrasonic sensors,
the manual "VEGA Visual Operating (VVO)"
also describes the setup of all other VEGA
sensors with the PC.
4.3 Adjustment with the adjustment
module MINICOM
In addition to the PC, you can adjust the
sensor with the small, detachable adjustment
module MINICOM. The adjustment module is
plugged into the sensor or into the external
indicating instrument (optional).
ESC
+
-
Tank 1
m (d)
12.345
OK
2
Tank 1
m (d)
12.345
4
All sensor versions (adjustment options)
provided by a PC with the adjustment program VVO are also available with the adjustment module MINICOM. However, the
adjustment with MINICOM looks different.
+
-
4 ... 20 mA
ESC
OK
You can carry out all adjustment steps with
the 6 keys of the adjustment module. A small
display shows you, apart from the measured
value, a short message on the menu item or
an entered numerical value.
Although the volume of information of the
small display cannot be compared with that
of the adjustment program VVO, you will be
able to carry out your settings quickly and
efficiently with the small MINICOM.
VEGASON 54P … 56P39
Setup
Error codes:
E013 No valid measured value
- sensor in the warm-up phase
- loss of the useful echo
E017 Adjustment span too small
E036 Sensor program not operating
- sensor must be reprogrammed
(service)
- fault signal also appears during
programming
E040 Hardware failure, electronics
defective
E113 Hardware failure
Adjustment steps
On the following pages, you will find the complete menu schematic of the adjustment module MINICOM.
Set up the sensor in the numbered sequence:
1. Address
2. Measurement in gases
3. Operating range
4. Adjustment
5. Conditioning
6. Meas. conditions
7. False echo storage (only required if errors
occur during operation).
8. Indication of the useful level and noise level
9. Outputs
Below you will find short explanations to
setup steps 1 … 9.
3. Operating range
Without special adjustment, the operating
range corresponds to the measuring range.
The measuring range was already entered
with the min./max. adjustment. Generally, it is
useful to choose a slightly wider range
(approx. 5 %) for the operating range than
for the measuring range.
Example:
Min./Max. adjustment: 0.300 … 5.850 m;
adjust operating range to approx.
0.250 … 6.000 m.
4. Adjustment
Max.
Min.
Under the menu item "
the sensor of the measuring range.
100 % (1.270 m) correspond
to 1200 liters
Span (4.58 m)
0 % (5.850 m) correspond
to 45 liters
Adjustment
", you inform
1. Address
Choose a free bus address with the DIP
switch (see chapter "4.2 Sensor address").
You can carry out the adjustment with or
without medium. Generally, you will carry out
the adjustment without medium, as you can
then adjust without a filling/emptying cycle.
2. Measurement in gases
Adjustment is only necessary if the measurement is made in gases other than air (CO2,
He, etc.). When measuring in gases, sound
the distance from the sensor to the product
surface and enter this in the menu item
"
Measurement in gases
take the altered (compared to air) sonic velocity into account and output correct levels.
40VEGASON 54P … 56P
". The sensor can then
Setup
Adjustment without medium
(adjustment independent of the actual level)
KeyDisplay indication
Sensor
m(d)
4.700
Parameter
OK
OK
OK
OK
+
adjustment
Adjustment
w.o.
medium
Adjustment
in
m(d)
(min. adjustment)
The distance indication flashes
and you can choose "feet" and
"m".
OK
+–
or
Confirm the setting with "OK".
m(d)
0.0%
at
m (d)
XX.XXX
Adjustment
in
With "+" and "–" you set the
percentage value for the min.
value or the lower level (example 0.0 %).
OK
The entered percentage value
is written in the sensor and the
distance corresponding to the
percentage value begins
flashing.
+–
or
With the "+" or "–" key you can
assign a level distance (example 5.85 m) to the previously
adjusted percentage value. If
you do not know the distance,
you have to measure it by
sounding.
OK
The entered product distance
is written in the sensor and the
display stops flashing.
You thereby adjusted the lower product distance as well as the percentage filling value
corresponding to the lower product distance.
100.0%
at
m (d)
XX.XXX
(max. adjustment)
Enter the max. adjustment (upper product
distance) in the same manner (example:
100 % and 1.270 m).
Note:
The difference between the adjustment values of the lower product distance and the
upper product distance should be as big as
possible, preferably at 0 % and 100 %. If the
values are very close together, e.g. lower
product distance adjustment at 40 %
(3.102 m) and upper product distance adjustment at 45 % (3.331 m), the measurement
will be inaccurate. A characteristics curve is
generated from the two points. Even the
smallest deviations between actual product
distance and entered product distance will
considerably influence the slope of the characteristics curve. If the adjustment points are
too close together, small errors inflate to
considerably larger ones when the 100 % or
the 0 % value is outputted.
VEGASON 54P … 56P41
Setup
Adjustment with medium
with
medium
Max.
Min.
adjust at
adjust
%
at %
XXX.X
XXX.X
Fill the vessel e.g. to 10 % and enter 10 % in
the menu "
Min. adjust
" with the "+" and "–"
keys. Then fill the vessel, e.g. to 80 % or
100 % and enter 100 % in the menu "
adjust
" with the "+" and "–" keys.
Max.
5. Conditioning
Signal
conditioning
Scaling
0 %
100 %
Deci-
prop.
corres-
corresponds
XXXX
mal
point
888.8
Conditioning
ponds
XXXX
Under the menu item "
assign a product distance at 0 % and at 100
% filling. Then you enter the physical quantity
and the physical unit as well as the decimal
point.
to
Mass
Unit
Kg
", you
With the "—>" key you change to the 100 %
menu. Enter here the numerical value of your
parameter corresponding to a 100 % filling.
In the example 1200 for 1200 liters.
• Confirm with "OK".
If necessary, choose a decimal point. However, note that only max. 4 digits can be
displayed. In the menu "
prop. to
", you choose
the physical quantity (mass, volume, distance…) and in the menu "
Unit
" the physical
unit (kg, l, ft3, gal, m3 …).
Linearisation:
Adjustment
Signal
conditioning
Scaling
Lin.
curve
Linear
Integra
tion
time
0 s
A linear correlation between the percentage
value of the product distance and percentage value of the filling volume is preset. With
the menu "Lin. curve" you can choose between linear, spherical tank and cylindrical
tank. The generation of a customised
linearisation curve is only possible with the
PC and the adjustment program VVO.
6. Meas. conditions
(see menu schematic)
Enter in the menu window "
0 % corresponds
7. False echo storage
"
the numerical value of the 0 % filling. In the
example of the adjustment with the PC and
the adjustment software VVO, this would be
45 for 45 liters.
A false echo storage is always useful when
unavoidable false echo sources (e.g. struts)
must be minimised. By creating a false echo
storage, you authorise the sensor electronics
• Confirm with "OK".
to record the false echoes and save them in
an internal database. The sensor electronics
treats these (false) echoes differently from
the useful echoes and filters them out.
42VEGASON 54P … 56P
Setup
8. Signal and noise level
In the menu
Ampl.:
XX dB
S-N:
XX
dB
you get important information on the signal
quality of the product echo. The higher the
"S-N" value, the more reliable the measurement (menu schematic MINICOM).
Ampl.:means amplitude of the level echo in
dB (useful level)
S-N:means Signal-Noise, i.e. the useful
level minus the level of the background noise
The bigger the "S-N" value (difference between the amplitudes of the useful signal level
and the noise level), the better the measurement:
> 50 dBMeasurement excellent
40 … 50 dBMeasurement very good
20 … 40 dBMeasurement good
10 … 20 dBMeasurement satisfactory
5 … 10 dBMeasurement sufficient
< 5 dBMeasurement poor
Example:
Ampl. = 68 dB
S-N = 53 dB
68 dB – 53 dB = 15 dB noise level
A signal distance of 53 dB means a very high
degree of measurement reliability.
9. Outputs
Under the menu "Outputs" you determine, for
example, if the current output should be
inverted, or which unit of measurement
should be shown on the sensor display.
VEGASON 54P … 56P43
Menu schematic for the adjustment module MINICOM
Ultras
onic
Parameter
97,5
%
Sensor
optimize
SON 52
When switching on, the sensor
P
type and the software version are
displayed for a few seconds.
3.00
Configuration
1.
Sensor
addr.
(•–/ –•)
126
Meas.
enviro
nment
Sensor
Tag
Sensor
Setup
Sensor address:
• Sensor address here only adjustable, if the
DIP switch in the sensor is set to address
greater than/equal to 126.
• If the DIP switch with number 8 is set to
"On" (address 128), it is possible to adjust
address 1 … 126 here.
Meas.
unit
m (d)
Adjust
ment
w.out
medium
Operating
range
Begin
0.50
4.
m (d)
3.
End
m (d)
6.00
Meas.
condit
ions
Condit
ion
liquid
Condit
ion
Solid
6.
Fast
change
No
Fast
change
No
with
medium
Agitat
ed sur
face
No
High
dust
level
No
Foaming
prod.
No
Large
angle
repose
No
Measur
ing in
gases
No
Measur
ing in
gases
No
Signal
condit
ioning
Scaling
Multi
ple
echo
No
Multi
ple
echo
No
5.
Lin.
curve
Linear
Integr
ation
time
Meas.
dist.
mm (d)
Correc
tion
Now!
0 s
Measur
ing in
gases
2,381
OK?
2.
Sound
speed
m/s
Correc
tion
Now!
333
OK?
0.0 %
m (d)
100.0%
at
m (d)
XX.XXX
Minadjust
at %
XXX.X
Maxadjust
at %
XXX.X
0 %
corres
ponds
XXXX
100 %
corres
ponds
XXXX
Decimal
point
888.8
Prop.
to
Mass
Unit
Kg
Adjust
ment
in
m(d)
at
XX.XXX
44VEGASON 54P … 56P
Setup
With these keys you move in
the menu field to the left, right,
top and bottom
ESC
7.8.
9.
act.
dist.
m (d)
4.700
Update
Meas.
dist.
m (d)
X.XX
Update
Now!
Learning!
OK?
False
echo
memory
Create
new
Meas.
dist.
m (d)
X.XX
Update
Now!
Learning!
Outputs
OK?
Add’l
functions
Info
Ampl.:
XX dB
S-N:
XX
dB
Delete
Delete
Now!
OK?
Deleting!
Simulation:Simulation:
Simulation:
Simulation:Simulation:
Sensor
Tag
Sensor
max.
range
m (d)
7.000
One hour after the last simulation adjustment, the sensor returns automatically to normal operating mode
Simulation
Sensor
type
SON51
P
Act.
dist.
Reset
to default
Reset
Now!
OK?
Reset
ing!
m (d)
4.700
Serial
number
1094
0213
Ampl.:
S-N:
Act.
dist.
m
X,XX
Language
English
Softw.
Softw.
vers.
date
3.00
15.09.
1999
Tempe-
XX dB
rature
XX
dB
act.
temperature
18°C
Menu items in bold print provide
sensor and measured value
information and cannot be
modified in this position.
OK
Min.
temperature
14°C
Sensor
addr.
(•–/ –•)
126
Max.
temperature
59°C
PA
output
Prop.
to
distance
Failure
mode
value
Sensor
displ.
Prop.
to
distance
Simulation
Now!
Simulation
XXX.X
OK?
High
dust
level
No
%
Fast
change
No
Light grey menu fields are only
displayed if required (dependent
on the adjustments in other
menus).
White menu items can be
modified with the "+" or "–" key
and saved with the "OK" key.
VEGASON 54P … 56P45
5 PA function diagram and parameters
5.1 P A function diagram
Sensor offset (Float)
[sensor unit (m,ft ...)]
Zero adjust
Zero offset
Raw-Distance
(value, status)
[in m]
Sensor value (Float)
[sensor unit (m,ft ...)]
PA function diagram and parameters
F Time (Float)
(sec)
(not available over
PA Parameter)
Filter
Raw-Temperature
(value, status)
[in K]
Cal point hi, cal point lo (Float)
[sensor unit (m,ft ...)]
Level hi, level lo (Float)
[level unit (%,m,ft ...)] (*1)
Cal type (Unsigned8)
Lifeabgleich (min-max) und
Trockenabgleich möglich
Cal
Level hi
Level lo
Cal point hi Cal point lo
Max min memory
max
min
Simulate procent value (DS_50)
[level unit (%,m,ft...)] (*1)
(manuf. spec. parameter)
Simulation
Temperature (Float)
[temperature unit (˚C, K, ...)]
Max temperature, min temperature
(Float) [temperature unit (˚C, K, ...)]
off
on
Simulate value
Level (Float)
[level unit (%,m,ft...)] (*1)
Level offset (Float)
[level unit (%,m,ft...)] (*1)
Offset
Offset
46VEGASON 54P … 56P
PA function diagram and parameters
Simulation distance value (DS_50)
[sensor unit (m,ft ...)]
(manuf. spec. parameter)
Simulation
off
on
Simulate value
Secondary value 2 (DS_33)
[secondary value 2 unit, (m,ft ...)]
distance
Max min memory
max
min
Secondary value 1 (DS_33)
[secondary value 1 unit
(%,m,ft...)] (*1)
Max sensor value,
min sensor value (Float)
[sensor unit (m,ft ...)]
Lin type (Unsigned8),
Tab index (Unsigned8),
Tab X Y value (2 x Float)
[level unit (%,m,ft ...). (*1)
primary value unit (%,m,ft ...)]. (*1)
Tab min number (Unsigned8),
Tab max number (Unsigned8),
Tab op code (Unsigned8),
Tab status (Unsigned8),
Tab actual number (Unsigned8)
Linearization
level %
Transducer
Block
Primary value (DS_33)
[primary value unit (%,m,ft...)] (*1)
lin %
A
B
distance
C
VEGASON 54P … 56P47
lin %
level %
distance
Display source select
(Unsigned8)
Source selector
scale
PV FTime (Float)
[sec]
PA function diagram and parameters
Sensordisplay
VEGA
Hi hi limit, hi limit,
lo limit, lo lo limit,
alarm hys
(Float) [out scale unit]
Alarm checkFilter
Hi hi limit
Hi limit
Lo limit
Lo lo limit
Alarm hyst
Hi hi alarm, hi alarm,
lo alarm, lo lo alarm
(DS_39)
Alarm sum (DS_42)
lin %
A
B
C
level %
Channel (Unsigned16)
Channel
distance
Simulate Linearization value (DS_50)
[primary value unit (%,m,ft ...).(*1)
secondary value 1 unit, (%,m,ft ...).(*1)
secondary value 2 unit, (m,ft ...)
(dep. on channel)]
Simulation
off
on
Simulate value
48VEGASON 54P … 56P
PA function diagram and parameters
Fail safe value (Float)
[out scale unit]
Fail safe type (Unsigned8)
Target mode (Unsigned8)
Out (DS_33)
[out scale unit]
Fail safe
Fail safe value
Divice status
PV scale (2 x Float)
[primary value unit (%,m,ft ...).(*1)
secondary value 1 unit, (%,m,ft ...).(*1)
secondary value 2 unit, (m,ft ...)
(dep. on channel)]
PV Scale
1
0
EU 0% EU 100%
Mode
Auto
Man
Mode & Status calc.
Out scale (DS_36)
[out scale unit (included)]
Out Scale
EU 100%
EU 0%
Data value (DS_33)
[out scale unit]
Actual mode
(DS_37)
Function
Block
10
VEGASON 54P … 56P49
PA function diagram and parameters
5.2 Status bytes
Coding of the status byte on the PA output value
Status codeDescr. acc. to Profibus standard Probable causes in the VEGA sensor
block alarmfor 10 sec., after parameter was written
0x89good (non-cascade) – active- Lo-Alarm
advisory alarm – low limited
0x8agood (non-cascade) – active- Hi-Alarm
advisory alarm – high limited
0x8dgood (non-cascade) – active- Lo-Lo-Alarm
critical alarm – low limited
0x8cgood (non-cascade) – active- Hi-Hi-Alarm
critical alarm – high limited
Note:
All 256 possible status byte values can also be generated by simulation. For this reason, the Failsafe-Mode
should be set to "Wrong calculated value", so that the simulated status will not be modified by the Failsafe calculation.
- Configuration error with PV-Scale
(PV-Span too small)
- Differences in meas. unit
- Error in the linearisation table
- Converter error
- Leakage pulse error
- Trigger error
- Temperature error
"Last valid value" and already valid meas.
value since switching on)
- Failsafe substitute value (Failsafe-Mode =
"Fsafe value")
"Last valid value" and no valid meas.
value since switching on)
in static category)
50VEGASON 54P … 56P
PA function diagram and parameters
5.3 Parameter listing
The following parameter listing and function
diagram is used for setup with the automation
system if there is no adjustment software
available. The listing reflects the parameters
of the function block and the transducer
block and should only be used by experienced Profibus technicians. Please make
sure that all VEGA sensors are Profile 3 sensors. As a rule, however, you will carry out
the sensor setup more conveniently with the
VEGA adjustment software VVO or with the
adjustment module MINICOM in the sensor.
The PA parameters and the function diagram
are normally not required for setup and serve
only to deepen the knowledge of the interested user and enable him to read in the
sensor measured values into the processing
system without an adjustment tool.
Slot IndexParameterObjectRead Writ eTypeSize Reset value StoreUnit
016 Block objectPBYes NoDS-3220Cm
017 St revPBYes NoUns igned162 0Nm
018 Tag descPBYes Y esOctet St ring32 32 * ' 'Sm
019 StrategyPBYes YesUnsigned162 0Sm
0 20 Alert keyPBYes YesUnsigned81 0Sm
021 Target modePBY es Y es (*1) Unsigned81Sm
0 22 Mode blkPBYes N oDS-373(D)m
0 23 Alarm s umPBY es NoDS-428Dm
0 24 Soft ware revision PBY es NoOctetString1 6Cm
0 25 Hardware revision PBY es NoOctet String1 6Cm
0 26 Device man IDPBY es NoUns igned162Cm
0 27 Device IDPBY es NoOctetString1 6Cm
0 28 Device ser numPBYes NoOctet St ring16Cm
0 29 Di agnosisPBYes NoOct et St r ing4 -Dm
031 Diagnosis maskPBY es NoOctetSt ring4Cm
0 33 Device certification PBYes NoOctet St ring32Co
0 34 Write lockingPBY es Y esUnsigned1 62 -No
0 35 Fact ory resetPBYes Y esUnsigned1 62 -So
0 36 Decript orPBYes YesOctetS t ring32 -So
0 37 Device messagePBYes Y esOctetSt ring32 -So
Optional
manda-
tory
VEGASON 54P … 56P51
PA function diagram and parameters
Slot IndexParameterObjectRead Writ eTypeSize Reset value StoreUnit
0 38 Device install date PBYes YesOctetString16 -So
0 40 Ident number s elect PBYes YesUnsigned81 -Sm
041 HW writ e prot ect ion PBYes N oUnsigned81Do
0 49 VVOPBYes YesOctetString 32o
0 50 View object PBPBYes NoOctet String17Dm
Direct ory
10
object header
Composite list
11
directory ent ries
1 16 Block objectFB_Primary_Value Yes NoDS-3220Cm
1 17 St revFB_Primary_V alue Yes NoUns igned162 0Nm
1 18 Tag des cFB_Primary_Value Yes Y esOctetString32 32 * ' 'Sm
1 19 St rategyFB_Primary _Value Yes YesUnsigned162 0Sm
1 20 Alert keyFB_Primary_Value Yes Y esUnsigned81 0Sm
1 21 Target modeF B_Primary_V alue Yes Y esUnsigned818Sm
1 22 Mode blkFB_Primary_Value Yes NoDS-373Dm
1 23 Alarm s umFB_ Prima ry_V alu e Yes NoD S-428Dm
1 24 BatchFB_Primary_Value Yes YesDS-6710 0,0,0,0Sm
1 32 PV FTimeFB_Primary_Value Yes Y esFloat4 0Nsecm
1 33 Fsafe t ypeFB_Primary_Value Yes Y esUnsigned81 1So
1 34 Fsafe valueFB_Primary_V alue Yes Y esFloat4 -Sout scale unito
1 35 Alarm hysFB_Primary_V alue Yes Y esFloat4 0.5% of range Sout scale unitm
1 37 HI HI LimFB_Primary_Value Yes YesFloat4 max valueSout scale unitm
1 39 HI LimFB_Primary_V alue Yes YesFloat4 max valueSout scale unitm
1 41 LO LimFB_Primary_V alue Yes Y esFloat4 min valueSout scale unitm
1 43 LO LO LimFB_Primary_V alue Yes Y esFloat4 min valueSout scale unitm
1 46 HI HI AlmFB_Primary_Value Yes NoDS-391 6 0Do
1 47 HI AlmFB_ Primary_V alue Yes NoDS-3916 0Do
1 48 LO AlmFB_Prim ary_V alu e Y es NoDS-3916 0Do
1 49 LO LO AlmFB_Primary_Value Yes NoDS-391 6 0Do
1 127 Alarm sumTB_LevelYes N oDS-428Dm
1 128 Primary valueTB_LevelYes NoDS-335Dprimary value unit m
1 129 Primary value unit T B_LevelYes YesUnsigned162 %Sm
1 130 LevelTB_LevelY es NoFloat4Dlevel unitm
1 1 31Level unitTB_LevelYes Y esUnsigned1 62 %Sm
1 132 Sens or valueTB_LevelYes NoFloat4Dsens or unitm
1 133 Sens or unitTB_LevelY es Y esUnsigned1 62Sm
Secondary
1134
value 1
Secondary
1135
value 1 unit
1 136 Secondary value 2 TB_LevelYes N oDS-335D
Secondary
1137
value 2 unit
1 138 Sens or of fsetTB_LevelYes Y esFloat4 0Ssens or unitm
1 139 Cal typeTB_LevelYes Y esUnsigned81
1 140 Cal point loTB_LevelYes YesFloat4S[sensor unit]m
1 1 41Cal point hiTB_LevelYes Y esFloat4Ssensor unitm
1 142 Level loTB_LevelY es Y esFloat4 0Slevel unitm
1 143 Level hiTB_LevelY es Y esFloat4 100Slevel unitm
1 144 Level offsetTB_LevelYes Y esFloat4 0Slevel unitm
1 145 Lin typeTB_LevelYes Y esUnsigned81 0 (=linear)Sm
1 148 Sens or high limitTB_LevelYes NoFloat4Csensor unito
1 149 Sens or low limitTB_LevelYes NoFloat4Csensor unito
1 150 Max sens or valueTB_LevelYes YesFloat4
1 1 51Min sensor valueTB_LevelYes Y esFloat4
1 152 TemperatureTB_LevelYes N oFloat4Dtemperature unit o
1 153 Temperature unit TB_LevelYes Y esUnsigned1 62 °CSo
1 154 Max temperature TB_LevelYes Y esFloat4
1 155 Min temperatureTB_LevelYes Y esFloat4
1 156 Tab indexTB_LevelYes Y esUnsigned811Do
TB_LevelYes NoDS-335D
TB_LevelYes YesUnsigned162So
TB_LevelYes YesUnsigned162So
oder
Lifeabgleic h
oder
akt. Wert
oder
akt. Wert
oder
akt. Wert
oder
akt. Wert
secondary
value 1 unit
secondary
value 2 unit
Sm
Nsens or unito
Nsens or unito
Ntemperature unit o
Ntemperature unit o
Optional
manda-
tory
o
o
VEGASON 54P … 56P53
PA function diagram and parameters
Slot IndexParameterObjectRead Writ eTypeSize Reset value StoreUnit
1 157 Tab X Y valueT B_LevelYes Yes2 * Float8N
1 158 Tab min numberTB_LevelYes NoUns igned81Co
1 159 Tab max numberTB_LevelYes NoUns igned81Co
1 160 Tab op codeTB_LevelYes YesUnsigned81Do
1 1 61Tab s tatusTB_LevelYes NoUns igned81Do
1 162 Tab act ual number TB_LevelYes NoUnsigned81Do
Simulate sensor
1190
value
1 1 91Simulat e levelTB_LevelYes YesDS-506 disableNlevel unito
1 192 Füllgutt ypTB_LevelYes YesUnsigned81No
1 193 Echo qualityTB_LevelYes NoUns igned81DdBo
1 194 First echo f act orTB_LevelY es Y esUnsigned81 ausNo
216 Block objectFB_Temperature Yes NoDS-3220Cm
217 St revFB_Temperature Yes NoUns igned162 0Nm
218 Tag descFB_Temperature Yes Y esOctetString32 32 * ' 'Sm
219 S t rategyFB_Temperat ure Yes YesUnsigned162 0Sm
2 20 Alert keyFB_Temperature Yes Y esUnsigned810Sm
221 Target modeF B_Temperature Yes Y esUnsigned818Sm
TB_LevelY es Y esDS-506 disableNsens or unito
ca. 331.6
(VEGA-Wer t )
TB_LevelYes YesUnsigned81Do
TB_LevelY es Y esFloat4Nsensor unito
TB_LevelY es Y esFloat4Nsensor unito
TB_LevelYes YesUnsigned81No
TB_LevelYes YesUnsigned81No
TB_LevelYes YesUnsigned81No
TB_LevelYes YesUnsigned81No
TB_LevelYes YesFloat4N%o
TB_LevelNo Y esFloat4Nsensor unito
secondary
value 1 unit,
primary value unit
Nm/so
Optional
manda-
tory
o
54VEGASON 54P … 56P
PA function diagram and parameters
Slot IndexParameterObjectRead Writ eTypeSize Reset value StoreUnit
2 22 Mode blkFB_Temperature Yes NoDS-373Dm
2 23 Alarm sumFB_Temperat ure Yes N oDS-428Dm
2 24 B at chFB_Temperat ure Yes YesDS-6710 0,0,0,0Sm
2 26 OutFB_Temperat ure Yes Y es (*1) DS-335D
2 27 PV scaleFB_Temperature Yes Yes2 * Float8 0,100Stemperature unit m
2 28 Out scaleFB_Temperature Yes Y esDS-36110,1 00,-,-S
2 29 Lin typeFB_Temperat ure Yes Yes (*1 ) Unsigned810Sm
2 37 HI HI LimFB_Temperat ure Yes YesFloat4 max valueS
2 39 HI LimFB_Temperat ure Yes YesFloat4 max valueS
241 LO LimF B_Temperature Yes YesFloat4 min valueS
2 43 LO LO LimFB_Temperature Yes YesFloat4 min valueS
2 46 HI HI AlmFB_Temperature Yes NoDS-391 6 0Do
2 47 HI AlmFB_Temperature Yes NoDS-3916 0Do
2 48 LO AlmFB_Temperat ure Yes NoDS-3916 0Do
2 49 LO LO AlmFB_Temperat ure Yes NoDS-391 6 0Do
2 50 SimulateFB_Temperature Yes YesDS-506 disableNtemperature unit m
View object
261
FB_Temperat ure
FB_Temperature Yes NoOc t etString18Dm
temp
out s cale unit
includes temp
out s cale unit
temp
out s cale unit
temp
out s cale unit
temp
out s cale unit
temp
out s cale unit
temp
out s cale unit
temp
out s cale unit
Optional
manda-
tory
m
m
o
m
m
m
m
m
VEGASON 54P … 56P55
6 Diagnosis
Diagnosis
6.1 Simulation
To simulate a certain filling, you can call up
the function "Simulation" on the adjustment
module MINICOM or in the software program
VVO.
With this function, you simulate a real vessel
filling level. Please note that connected instruments, such as e.g. a PLC, react according
to their adjustments and will probably activate alarms or system functions.
Simulation with VVO
If you start the simulation mode with the adjustment program VVO on the PC, the simulated level is outputted until you quit the
simulation mode.
Simulation with MINICOM
If you start the simulation mode on the adjustment module MINICOM, the sensor returns to
standard operating mode after one hour.
6.2 Error codes
Error codesCorrective measure
E013 No valid measured valueMessage is displayed during the warm-up
- Sensor in the warm-up phasephase
- Loss of the useful echoIf the message remains, a false echo storage
(with the adjustment software on the PC - see
"Echo curve" under "Sensor optimisation") must
be carried out together with a modification of
mounting location and orientation to achieve the
lowest possible false echo background.
E017 Adjustment span too smallCarry out a readjustment.
E036 Sensor software does not runSensor requires a software update (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tiveTransducer defective.
56VEGASON 54P … 56P
Make sure that the difference between
min. and max. adjustment is at least 10 mm .
Message appears during a software update.
Contact our service department.
Technical data
,0
,0
,0
,0
,0
,0
,0
,0
,0
35689101315
,0
,
,
,
,0
,0
,0
,0
35689101315
g
7 Technical data
7.1 Data
Power supply
Supply voltage9 … 32 V DC
Note:Note:
Note:
Note:Note:
If VEGASON 54P … 56P sensors are operated on Profibus input cards (EP cards) in the
VEGALOG 571 processing system each card must be powered individually with own
power supply unit. For power supply of an EP card, use the power supply unit (ASIInterface SLA 3.100) from Puls which is available from VEGA.
Current consumptionmax. 81 mA, depending on the output voltage
VEGAPULS 54P, 55PVEGAPULS 56P
mA
100
90
81
80
70
60
50
40
35
30
5
0
13,5
15
10
U
= 9 V
min
25
20
Power consumption
- VEGASON 54 and 55max. 0.9 W
- VEGASON 56max. 1.1 W
Resistance of the signal cabledepending on the output voltage UO (PA voltage)
Number of sensors on a two-wire cabledepending on the output voltage UO of the
UO (PA voltage) of the segment coupler used,
see diagrams.
mA
100
90
81
80
70
60
50
40
35
= 32 V
30
V
30
U
max
5
0
13,5
15
10
U
= 9 V
min
20
of the segment coupler used
segment coupler used and the voltage losses of
the signal cable, see cable specification.
Number of sensors when being operated on an
EP card in a VEGALOG, see diagrams:
25
30
V
U
= 32 V
max
VEGAPULS 54P, 55P on EP cardVEGAPULS 56P on EP card
2000
1800
1600
1400
1200
1000,0
800
600
400
Cable length (m)
200
0,0
No. of sensors VEGAPULS 54P, 55 P
0,8 mm
2000,0
1800
1600
1400
1200,0
1000
800
0
600
0
0
400
Cable length (m)
200,0
0
No. of sensors VEGAPULS 56P
0,8 mm
VEGASON 54P … 56P57
Technical data
Measuring range
(with sensor version B … D, the transducer end is the reference plane. With sensor version A
the lower side of the flange is the reference plane)
VEGASON 541.0 … 25 m (solids 1.0 … 15 m)
VEGASON 550.8 … 45 m (solids 0.8 … 30 m)
VEGASON 56
- Version A1.8 … 70 m (solids 1.8 … 45 m)
- Version B … D1.4 … 70 m (solids 1.8 … 45 m)
Output signal
digital (Profibus)the digital output signal (meas. signal) is
modulated onto the power supply and further
processed in the PLC or process control
system.
Integration time0 … 999 seconds
Measured value display (optional)
Liquid-crystal indication
- in the sensorscalable output of meas. value as graph and
number value.
- powered externally by the sensorscalable output of meas. value as graph and
number value. Measured value display can
be mounted up to 25 m away from the sensor.
Adjustment
- adjustment software VEGA Visual Operating on Master-Class 2 PC
- adjustment module MINICOM in the sensor or in external indicating instrument (optional)
- process adjustment interface PACTwareTM (VVO runs as a subprogram)
- SIMATIC PDM in conjunction with Electronic Device Description (EDD)
- VEGASON 565 … 300 m (cable diameter 9.5 … 9.9 mm)
Cable entry for the signal and supply
cable with spring-loaded terminals
- plastic housing (PBT)2 x M20 x 1.5 (cable diameter 5 … 9 mm)
or 2 x 1/2" NPT (cable diameter
3.1 … 8.7 mm or 0.12 … 0.34 inch)
- Aluminium housing and
Ex d terminal compartment2 x 1/2" NPT (cable diameter 3.1 … 8.7 mm
or 0.12 … 0.34 inch)
60VEGASON 54P … 56P
Technical data
Materials
HousingPBT (Valox) or Aluminium
Process connection
- flange versionAlu or PP
- swivelling holder and threadgalvanized steel
Transducer
- VEGASON 54PA (1.4301 with StEx)
- VEGASON 55, 56UP
Transducer diaphragm
- VEGASON 541.4571
- VEGASON 55Alu/PE foam
- VEGASON 56Alu/PTFE non-stick coating
Transducer cable (cable cover)
- VEGASON 54, 55PUR (1.1082)
- VEGASON 56silicone (1.1083)
Weight (depending on the housing materials used)
VEGASON 54
- Version A5.6 … 10.7 kg
- Version B6.9 … 9.7 kg
- Version C7.5 … 10.5 kg
- Version D4.7 … 6.9 kg
VEGASON 55
- Version A8.0 … 13.3 kg
- Version B8.7 … 10.3 kg
- Version C9.2 … 11.1 kg
- Version D6.5 … 7.5 kg
VEGASON 56
- Version A7.3 … 11.3 kg
- Version B8.7 … 10.3 kg
- Version C9.3 … 11.1 kg
- Version D6.5 … 7.5 kg
CE conformity
VEGASON series 50 ultrasonic sensors meet the protective regulations of EMC
(89/336/EWG), NSR (73/23/EWG) and R & TTE directive (1999/5/EC).
Conformity has been judged acc. to the following standardst:
EN 300 683 - 1: 1997
EN 300 440 - 1: 1995
I-ETS 300-440
Expert opinion No. 0043052-01/SEE, Notified
The status byte corresponds to profile 3,0 "Profibus PA Profile for Process Control Devices"
coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit 6 … 0 = 0).
Measured value:
The measured value is transmitted as a 32 Bit floating point number in IEEE-754 format.
Byte nByte n+1
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
7654321076543210
VZ27262524232221202-12-22-32-42-52-62
SignExponentMantissa
Byte n+2Byte n+3
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
When using ultrasonic sensors in Ex areas or
on ships, the instruments must be suitable
and approved for the explosion zones and
applications.
The suitability is checked by the approval
authorities and is certified in approval documents.
Please note the attached approval documents when using a sensor in Ex area.
Mounting on carrier rail 35 x 7.5 acc. to EN 50 022 or
screwed to a flat surface
Note:
The cable diameter of the connection cable
should be min. 5 mm and max. 9 mm.
Otherwise, the seal effect of the cable entry
would not be ensured.
Adjustment module MINICOM
Tank 1
m (d)
12.345
67,5
VEGASON 54P … 56P63
ESC
+
-
32,5
OK
Adjustment module for insertion into
VEGASON series 50 sensors or into the external indicating instrument VEGADIS 50
74
VEGASON 54 … 56 in Version A
201
165
Plastic housing
(PBT)
10˚
101
Aluminium
housing (Al)
25
116
Technical data
215
185
VEGASON 54
VEGASON 55
397
Min. distance to
the product
Min. distance to
the product
257
1,0 m
0,8 m
90
12xø22
1)
ø190 (ø196)
ø340
ø244
ø405
75
1)
20
110 (126)
12xø26
20
128
12xø22 (12xø26)
445,8
282
Reference plane
Reference plane
20
Reference plane
VEGASON 56
423
Min. distance to
the product
1,8 m
64VEGASON 54P … 56P
ø198
2)
ø340 (ø405)
Technical data
VEGASON 54 … 56 in version B
486
4xø19
172
386
245
ø165
ø122,8
201
165
10˚
Plug connection
ø 27
> ø200
Plastic housing
(PBT)
11,5
11,5
101
90
503
4xø19
ø165
ø122,8
>ø250
435
270
215
185
65
Plug connection
798
Aluminium
housing (Al)
25
116
ø165
ø122,8
4xø19
>ø210
ø190 (ø196)
VEGASON 54
1,0 m
Reference plane
189,5
ø 244
VEGASON 55
484,5
0,8 m
ø 198
1,4 m
VEGASON 56
VEGASON 54P … 56P65
VEGASON 54 … 56 in version C
Aluminium housing (Al)
Plug
Technical data
215
25
116
185
Plug connection
ø 45
78
68
7
130
150
ø 7
85
65
170
445,8
282
10120
1,0 m
0,8 m
VEGASON 54
VEGASON 55
1,4 m
Reference plane
VEGASON 56
66VEGASON 54P … 56P
Technical data
VEGASON 54 … 56 in version D
Plug connection
3240
214
149
Plastic housing (PBT)
Plug
68
ø 45
78
233
7
3240
167,5
186
101
90
130
150
201
165
10˚
397,2
257,2
ø 7
85
65
12010
170
3240
527
462,5
Reference plane
1,0 m
0,8 m
1,4 m
VEGASON 54VEGASON 55VEGASON 56
VEGASON 54P … 56P67
VEGA Grieshaber KG
Am Hohenstein 113
D-77761 Schiltach
Phone(07836) 50-0
Fax(07836) 50-201
E-Mailinfo@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application, practical
use and operating conditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data subject to alterations
2.24 967 / 02-03-19 / EN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.