VEGA SON54P User Manual

Operating Instructions
VEGASON 54P … 56P (Profibus PA)
Level and Pressure
PROF I
PROCESS FIELD BUS
BUS
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function ................................................................................. 4
1.2 Application features ............................................................. 5
1.3 Profibus output signal .......................................................... 6
1.4 Adjustment ............................................................................ 7
1.5 Type survey ........................................................................ 11
1.6 Type code ........................................................................... 12
2 Mounting and installation
2.1 Mounting ............................................................................. 13
2.2 General installation instructions ........................................ 15
2.3 Measurement of liquids ..................................................... 17
2.4 Measurement of solids ...................................................... 19
2.5 Socket extensions ............................................................. 20
2.6 False echoes ...................................................................... 21
2.7 Installation mistakes ........................................................... 23
Contents

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGASON 54P … 56P

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the oper­ating instructions and come with the Ex ap­proved instruments.
Contents
3 Electrical connection
3.1 Connection .......................................................................... 26
3.2 Connection cables and bus configuration ....................... 26
3.3 Sensor address ................................................................. 29
3.4 Connection of the sensor .................................................. 31
3.5 Connection of the external indicating instrument ............ 32
3.6 Bus configuration ............................................................... 33
4 Setup
4.1 Adjustment media .............................................................. 38
4.2 Adjustment with the PC ...................................................... 39
4.3 Adjustment with the adjustment module MINICOM ........ 39
5 PA function diagram and parameters
5.1 PA function diagram........................................................... 46
5.2 Status bytes ....................................................................... 50
5.3 Parameter listing ................................................................ 51
6 Diagnosis
6.1 Simulation ............................................................................ 56
6.2 Error codes ........................................................................ 56
7 Technical data
7.1 Data ..................................................................................... 57
7.2 Data format of the output signal ........................................ 62
7.3 Approvals ........................................................................... 63
7.4 Dimensions ......................................................................... 63
VEGASON 54P … 56P 3

1 Product description

1.1 Function

Continuous level measurement with ultrasonic sensors is based on the running time meas­urement of ultrasonic pulses.
Product description
VEGASON 54 … 56 sensors are a newly developed generation of extremely compact ultrasonic sensor for level measurement. They were developed for liquids and especially for solids and larger measuring distances. They are the right choice for applications in silos, bunkers and material dumps.
Due to small housing dimensions and proc­ess fittings, the compact sensors are an unobtrusive, and most of all, very cost­effective solution for your level measurement applications. With the integrated display and a special sensor intelligence in conjunction with large measuring ranges, they enable high-precision level measurement and can be used for applications in which the advan­tages of non-contact measurement could never before be realised.
The outputs generate a digital Profibus PA signal as output, i.e. measurement signal.
Measuring principle
High-performance piezoceramic transducers emit focused ultrasonic pulses which are reflected by the product surface. The meas­urement electronics prepares a precise im­age of the environment from the reflected ultrasonic pulses. The transducers work both as transmitter and receiver. As receiver, the transducers are high-sensitivity piezo micro­phones.
Meas. distance
emission - reflection - reception
The measurement electronics precisely cal­culates the distance between transducer and medium from the speed of sound and the measured running time of the emitted sound impulse. The distance is then converted into a level-proportional signal and, in conjunction with the sensor parameter settings, made available as a precise, calibrated level value.
Since the speed of sound is subject to tem­perature influence, the transducer also con­tinuously detects the ambient tempera­ture, so that the level is precisely measured even in case of varying ambient temperature.
4 VEGASON 54P … 56P
Product description
Output signal
The level-proportional Profibus PA measure­ment signal is processed and outputted completely digitally. Digital processing of the measurement signal ensures an accuracy which could be never reached by an ana­logue measurement signal, as the digital signal is always transmitted error-free right up to the last decimal position. Varying resistances or tiny leakage currents do not influence the accuracy of digital technology. The digital signal is always clear and unam­biguous.
The digital signal, giving a faithful representa­tion of the adjusted measuring range of the sensor, can be modified by various param­eter settings.
Measured value display
As an option, the series 50 ultrasonic sensors can be equipped with an indicating instru­ment for direct, local level survey. The indi­cating instrument shows the precise level by means of the analogue bar graph and the digital number value. In addition to the indica­tion in the sensor, you can have the level displayed with the VEGADIS 50 external indicating instrument at a distance of up to 25 m from the sensor. The external measured value display operates, like the integrated display, independently of the PA output signal and can be modified through individual pa­rameter settings.

1.2 Application features

Two-wire technology
• Power supply and output signal on one two-wire cable.
• Output signal and signal processing com­pletely digital, therefore maximum accu­racy.
• Profibus profile 3 – sensor.
Applications
• Level measurement of solids, liquids and pastes.
• Gauge measurement, distance measure­ment, object monitoring and conveyor belt monitoring.
Rugged and precise
• Measurement unaffected by substance properties such as density, conductivity, dielectric constant…
• Suitable for corrosive substances.
• Measuring range 0.8 m … 70 m.
Adjustment options
• With adjustment software VEGA Visual Operating (VVO) on the PC.
• With detachable adjustment module MINICOM.
Measured value display
• Measured value display integrated in the sensor.
• Optional display separate from sensor.
• Adjustment module MINICOM in sensor.
Connection to any process
• G 1 A, DN 50, DN 80, DN 200, DN 250
Approvals
• CENELEC, ATEX, PTB
VEGASON 54P … 56P 5
Product description

1.3 Profibus output signal

PROPRO
PROcess
PROPRO sult of a joint project of thirteen companies and five universities. The companies Bosch, Klöckner-Möller and Siemens played a deci­sive role. The specifications of the bus are described in the protocol layers 1, 2 and 7 of the ISO/OSI reference model and are avail­able from the PNO (Profibus user organisa­tion). Layers 3 … 5 have not yet been developed as a standard, leaving Profibus with far-reaching perspectives for the future.
Today approx. 600 companies make use of Profibus technology and belong to the PNO. Profibus Specification, Profibus Periphery and Profibus mation.
As a process automation bus, Profibus PA enables power supply over the bus. Up to 32 sensors can be operated on a shielded two­wire cable that carries both power supply and measurement signal. In Ex areas, up to ten sensors can be connected from the PA level to one two-wire cable (EEx ia).
Bus structure
The Profibus DP and PA network consists of up to 126 master and slave participants. Data are always exchanged from point to point, with the data traffic being exclusively controlled and checked by master devices. Communication is carried out according to the Token-Passing procedure. This means that the master holding the Token can contact the slaves, give instructions, enquire data and cause the slaves to receive and transmit data. After the work is done or after a prede­termined time interval, the Token is passed on by the master to the next master.
FIFI
BUSBUS
FIeld
BUS (PROFIBUS) is the re-
FIFI
BUSBUS
FMSFMS
FMS stands for Fieldbus Messaging
FMSFMS
DPDP
DP for Decentralised
DPDP
PP
AA
P
A for Process Auto-
PP
AA
Master-Class 1
is the actual automation system, i.e. the proc­ess control computer or the PLC that en­quires and processes all measured values.
Master-Class 2
One or several Master-Class 2 can operate in a Profibus network. As a rule, Master-Class 2 devices are engineering, adjustment or visu­alisation stations. The VEGA adjustment soft­ware VVO (VEGA Visual Operating) operates as Master-Class 2 participant on the DP bus and can work on an engineering PC, on an adjustment PC or on the process control computer and can access any VEGA sensor on the PA level.
Instrument master file
A so-called GSD file is delivered with every VEGASON Profibus sensor. This file is nec­essary for integrating the sensor into the bus system. The GSD file (instrument master file) contains, beside the sensor name and the manufacturer, the sensor-specific communi­cation parameters which are necessary for a stable integration of the sensor in the bus.
Load the GSD file belonging to the sensor into your bus configuration program. If the GSD file is not available, it can be loaded from the VEGA homepage: http://www.vega.com.
Do not confuse the GSD file with the EDD (Electronic Device Description), a file neces­sary for the PDM environment (this can be also found on the VEGA homepage).
6 VEGASON 54P … 56P
Product description

1.4 Adjustment

Each measuring situation is unique. For that reason, every ultrasonic sensor needs some basic information on the application and the environment, e.g. which level means "empty" and which level "full". Beside this "empty and full adjustment", many other settings and adjustments are possible with VEGASON ultrasonic sensors. The output of echo curves or the calculation of vessel lineari­sation curves by means of vessel dimen­sions are only two examples.
Profibus adjustment scheme
In the Profibus environment, there are differ­ent adjustment concepts and adjustment tools which often differ considerably from manufacturer to manufacturer. From the us­er’s point of view, a manufacturer-independ­ent adjustment program which could be operated directly on the Profibus DP, as well as at any system node (e.g. the engineering station or the process control), would be ideal.
In the past, only the program "SIMATIC PDM", based on the HART® adjustment scheme, could fulfil this wish (though with the limita­tions common to HART®). As with HART®, the availability of an instrument-specific database for a comprehensive adjustment with PDM (Process Device Managing) is a requirement. Otherwise, only the basic instrument func­tions, such as adjustment, are available. In the PDM environment, this instrument-spe­cific database is called EDD (Electronic De­vice Description), in perfect analogy to the HART® environment which also requires, except for the VEGA HART® instruments, a DD (Device Description) for each sensor.
We are aware of the disadvantages of the HART® environment: for each sensor/partici­pant, an individual DD must be loaded, which in addition, must always be the latest and most up-to-date DD. Special adjustment options such as e.g., the output of an echo curve, are available neither with HART® nor with PDM. User-friendly adjustment is out of the question. With VEGA’s adjustment pro­gram VVO, those restrictions belong to the past.
The legitimate wish of many Profibus users for a manufacturer-independent adjustment tool without EDD has been realised in the form of PACTware
TM 1)
. An association of a number of process technology companies developed PACTwareTM: a Process Automa­tion Configuration Tool that can run different manufacturer software tools under a stand­ardised user interface and adjustment scheme. Specialists call this technology Field Device Transcription. Just as different Win­dows printer drivers enable operation of completely different printers under a single user interface, PACTware
TM
enables operation of all field instruments under a single user interface. Instrument-specific databases (EDD), like those required for SIMATIC PDM, are not necessary.
As a result of this development, four adjust­ment media are available for VEGA Profibus sensors:
- adjustment with the PC and the adjustment
program VVO (VEGA Visual Operating) as stand-alone tool, on the segment coupler or directly on the sensor.
- adjustment with the detachable adjustment
module MINICOM in the sensor
- adjustment with the SIMATIC PDM adjust-
ment program (requires EDD instrument databases) from the control room
- adjustment with the manufacturer-inde-
pendent user interface PACTwareTM on the sensor, from the control room or on the segment coupler.
VEGASON 54P … 56P 7
Product description
Adjustment with the adjustment pro­gram VVO - VEGA Visual Operating
The setup and adjustment of the ultrasonic sensors is generally done on the PC with the adjustment program VEGA Visual Operating (VVO) under Windows®. The program leads quickly through adjustment and parameter setting by means of pictures, graphics and process visualisations.
Note:
The adjustment program VVO must be avail­able in version 2.70 or higher.
The VEGA adjustment software VVO (VEGA Visual Operating) operates either as a subprogram of the host program PACTware acc. to the FDT concept (Field Device Tool) or as an independent adjustment program on any PC, engineering station or process con­trol computer.
The adjustment program recognises the sensor type
Visualised input of a vessel linearisation curve in the adjustment program VVO
VEGA’s adjustment program VVO can ac­cess the adjustment options of VEGA sen­sors in their entirety and, if necessary, can update the complete sensor software. For communication with Profibus sensors, the adjustment program either requires a Profibus-Master-Class2 interface card or the interface adapter VEGACONNECT 3. The PC with the Profibus interface card can be connected directly to any point on the DP bus with the standard RS 485 Profibus cable. In conjunction with the adapter VEGACONNECT, the PC can be connected directly to the sensor. VEGACONNECT com­municates via a small plug directly with the respective sensor. The adjustment and parameter data can be
TM
saved at any time on the PC with the adjust­ment software and can be protected by passwords. If necessary, the adjustments can be transferred quickly to other sensors. In practice, the adjustment program VVO is often installed as a tool on an engineering station or an operating station. VVO then accesses all VEGA sensors directly over the bus via the Profibus interface card (e.g. from Softing) as Master Class 2, from the DP level to the PA level (via segment coupler) right down to the individual sensor. Beside the instrument master file (GSD), with which a sensor is logged into the Profibus system, the majority of all Profibus sensors requires for adjustment, beside the specific adjustment software, also a so-called EDD (Electronic Device Description) for each sensor, in order to access and adjust the sensor from the bus levels. This is not the case with VVO. The adjustment software VVO can communicate at any time with all VEGA sensors without the help of a special database. Of course, all other non-Profibus VEGA sensors can be adjusted with the adjustment software as well (4 … 20 mA sensors or VBUS sensors). With VEGA sen­sors, it is not necessary to go looking for the latest EDD. This is the essential requirement for a manufacturer-independent adjustment program, like PACTwareTM, anticipated by many users, see following pages.
8 VEGASON 54P … 56P
Product description
Adr. 21
SPS
Adr. 22
VVO
3
PA-
Adr. 23
Bus
Master-Class 1
Adr. 1
DP-Bus
Adr. 24
Adr. 10
Adr. 25 … 56
2
Profibus DP interface card as Master-Class 2 (e.g. Softing)
3
Adr. 58
Adr. 57
Segment coupler
(max. 32 participants)
Adr. 60
Adr. 59
Adr. 26
Adr. 25
Adr. 27
Adr. 28
Adr. 29
Adjustment of the VEGASON ultrasonic sensors from process control via a Profibus interface card in the proc­ess control computer or in an additional PC. The adjustment software VEGA Visual Operating (VVO) accesses the sensors bidirectionally via the interface (interface card).
VEGASON 54P … 56P 9
Product description
Adjustment with adjustment module MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad­justment module with display, you carry out the adjustment in clear text dialogue. The adjustment module can be plugged into the ultrasonic sensor or into the optional, external indicating instrument.
Tank 1 m (d)
12.345
Detachable adjustment module MINICOM
The adjustment module can be easily re­moved to prevent unauthorised sensor ad­justment.
ESC
+
-
Tank 1 m (d)
12.345
OK
2
Tank 1 m (d)
12.345
4
PA-Bus
ESC
+
-
OK
ESC
+
-
OK
Adjustment with the SIMA TIC PDM adjustment program
To adjust all essential functions of the VEGA sensor with the adjustment station SIMATIC PDM from Siemens, a so-called EDD is re­quired. Without this EDD, only the basic func­tions such as min./max. wet adjustment or integration time can be adjusted with the PDM adjustment program. Further important adjustment functions, such as input of the measuring environment or false echo stor­age, are not available without EDD. After integration of the EDD files in the Simatic PDM adjustment software, all important adjustment functions are accessible. If it is not at hand, the obligatory GSD (instrument master file) as well as the EDD (Electronic Device De­scription) necessary for PDM can be downloaded from the VEGA homepage (http://www.vega.com).
Adjustment with PACTware
PACTwareTM is a manufacturer-independent automation/configuration tool through which access to instruments of different manufac­turers (Krohne, Pepperl + Fuchs, VEGA, WIKA-, Bürkert…) is possible. The VEGA adjustment software VVO works as a subprogram/menu. PACTwareTM activates the required menu options for the sensor/instru­ment being accessed.
PACTwareTM looks different than VVO and is designed with a tree structure. Operating instructions for PACTwareTM can be found in the PACTware described in this operating instructions manual.
TM
documentation. They are not
TM
max. 2.5 m
Adjustment with detachable adjustment module. The adjustment module can be plugged into the ultrasonic sensor or into the external indicating instrument VEGADIS 50.
10 VEGASON 54P … 56P
Product description

1.5 T ype survey

VEGASON 54 … 56 sensors are a newly developed generation of very compact ultra­sonic sensors for large measuring ranges. For shorter measuring ranges, VEGASON 51 … 53 sensors are used.
Thanks to the small housing dimensions and process fittings, the compact sensors are a very cost-effective solution for your level measurement applications. Swivelling holders enable quick orientation of the transducers to the liquid or solid surface. Because the sen­sor electronics and transducer can be sepa­rated, installation is very simple. It is possible to separate the sensor electronics up to 300 m from the transducer. As a result, the transducer can handle ambient temperatures up to 150°C (VEGASON 56).
General features
• Application in solids and liquids.
• Measuring range 0.8 m … 70 m.
• Ex-approved in Zone 10 (IEC) or Zone 11 (ATEX) classification mark EEx ia [ia] IIC T6 or Zone 20/21/22, VEGASON 56 also Zone 1.
• Integrated measured value display in the sensor or in the external indicating instru­ment which can be mounted up to 25 m away from the sensor.
VEGASON 54
Version A Version B Version C Version D
VEGASON 55
Version A Version B Version C Version D
VEGASON 56
Version A Version B Version C Version D
VEGASON 54P … 56P 11

1.6 Type code

Product description
VEGASON 54 K EX.XX X X X X X X X
K - Plastic housing PBT, M20 x 1.5 cable entry N - Plastic housing PBT, A - Aluminium housing, M20 x 1.5 cable entry
FEP - Version A, flange DN 200 (PP) FEA - Version A, flange DN 200 (Aluminium) FLP - Version A, flange 8" (PP) FLA - Version A, flange 8" (Aluminium) SAS - Version B, flange swivelling holder DN 50 SBS - Version B, flange swivelling holder DN 80 GAS - Version C, flange swivelling holder DN 50 GBS - Version C, flange swivelling holder DN 80 RGS - Thread G 1 A YYY - Other process fittings
X - without display A - with integrated display
X - without adjustment module MINICOM B - with adjustment module MINICOM (mounted)
A - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA (four-wire) B - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA, HART
(four-wire) E - Power supply via signal conditioning instrument G - Segment coupler for Profibus PA P - 90 … 250 V AC (only in USA) N - 20 … 36 V DC, 24 V AC (only in USA) Z - Supply via signal conditioning instrument (only in USA)
.X - without approval EXS.X - StEx Zone 10
K - Analogue 4 … 20 mA output signal (two-wire or
four-wire technology) V - Digital output signal (two-wire technology) P - Profibus
Type 54 - Measuring range 1.0 … 25 m Type 55 - Measuring range 0.8 … 45 m Type 56 - Measuring range 1.6 … 70 m
Measuring principle (SON for ultrasonic)
½" NPT cable entry
®
12 VEGASON 54P … 56P
Mounting and installation

2 Mounting and installation

2.1 Mounting

Version A
Sensors in version A (flange version) are supplied completely assembled and ready for operation. Immediately after mounting on the vessel and electrical connection, they are ready for operation.
Version B
Sensors in version B are supplied in two parts (transducer and sensor electronics). First of all, mount the transducer on the ves­sel or above the medium. There is a four-pole jack at the end of the transducer tube. The respective counterpart to the jack protrudes out of the lower side of the sensor electron­ics. Insert the plug of the sensor electronics (only possible in one position) into the jack of the transducer tube. Continue pushing the electronics housing onto the transducer tube, on which there is a wide and a narrow groove.
Groove for locking the headless screw
Mounting groove (must no longer be visible after mounting)
The wide groove is used for locking the headless screws. The narrow groove is an assembly mark. Move the electronics hous­ing farther down over the transducer tube until the mounting groove is no longer visible. Fasten the housing with the headless screws to the transducer tube. Use a 5 mm hexagon screwdriver (or Allen wrench).
VEGASON 54P … 56P 13
Mounting and installation
Ver sion C and D
The sensors in version C and D are supplied in three parts (transducer, sensor electronics and transducer cable). First mount the trans­ducer (see version B). There is a four-pole jack at the transducer tube end. A respective counterpart to the jack is provided in the connection cylinder of the transducer cable. Insert the connection cylinder plug into the jack of the transducer tube. S
Connection cylinder
Mounting bracket
Connection cylinder
Transducer cable
On the end of the transducer tube you find a wide and a narrow groove. The wide groove is used for locking the cylinder with the head­less screws. The narrow groove is the as­sembly mark.
Then push the connection cylinder onto the transducer tube (with a slight swivelling mo­tion) until the assembly mark is no longer visible.
When the assembly mark is covered by the cylinder, fasten the cylinder with the two headless screws. Use a 5 mm hexagon screwdriver (or Allen wrench).
Now mount the sensor electronics at the requested location. The sensor electronics is fastened to a mounting bracket so that it can be mounted on a plain surface or on a wall. Make sure that the sensor electronics is mounted in such a way that there is enough space above the electronics to open the cover.
Now insert the plug at the other end of the transducer cable into the jack on the elec­tronics housing.
Note:
Avoid bending the transducer cable too sharply when laying it out. This is a special cable which could otherwise be damaged.
In addition, make sure that the cable cannot be damaged during operation. A signal with a voltage of approx. 1 kV is transmitted (which could be a danger in Ex areas if the cable is damaged) via the shielded cable cores.
Groove for locking the headless screws
Mounting groove (must no longer be visible after mounting)
14 VEGASON 54P … 56P
Mounting and installation

2.2 General installation instructions

Measuring range
version, the instrument flange (version A). Please note the information on the reference
Beside other criteria, you select your instru­ment according to the required measuring range. The reference plane for the min. and
planes in chapter "7.4 Dimensions". The max. filling depends on the required min. distance and the mounting location.
max. distance to the liquid or solids is the transducer end, or for instruments in flange
Reference plane
min. distance
1.0 m
100 %
0 %
Span
Min. distance, max. measuring range, span and reference plane
0 %
max. meas. range
max. meas. distance 25 m (type 54), 45 m (type 55), 70 m (type 56)
min. distance 0.8 m
100 %
Span
Reference plane
min. distance 1.4 m
100 %
0 %
Span
Beam angle and false echoes
At greater distance, the energy of the ultra­sonic impulses distributes over a large area,
The ultrasonic impulses are focused by the transducers. The impulses leave the trans­ducer in a conical form similar to the beam pattern of a spotlight. The beam angle is 4°
thus causing weaker echoes from obstruct­ing surfaces. The interfering signals are therefore less critical than those at close range.
(VEGASON 54), 5° (VEGASON 55) and 6° (VEGASON 56) at -3 dB emitted power. Any object, e.g. tubes or struts inside this emission cone will cause a large false echo. Especially within the first few meters of the
If possible, orient the sensor axis perpen­dicularly to the product surface and avoid vessel installations (e.g. pipes and struts) within the 100 % area of the emission cone.
emission cone, pipes, struts, or other installa­tions can interfere with the measurement. At a distance of 6 m, the false echo of a strut has an amplitude nine times greater than at a distance of 18 m.
VEGASON 54P … 56P 15
The following illustration of the ultrasonic beams is simplified and represents only the main beam - a number of additional weaker beams exists.
0 m
Meas. distance
Mounting and installation
VEGASON 56VEGASON 54
0 m
Emitted power
50 %
100 %
Emitted power
25 m
Meas. distance
45 m
0 m
4
0,9
2,0
10˚
242
0
m
0,9
2,0
0
VEGASON 55
50 %
Emitted power
100 %
m
Emitted power
Meas. distance
50 %
Emitted power
100 %
Emitted power
12˚
70 m
3,7
7,5
3,7
0
m
7,5
Therefore, in practical application, the trans­ducer has to be oriented so that the lowest possible false echo signal strength is achieved. Only giving attention to the size of the useful echo is usually not adequate under difficult measuring conditions. In most cases, a low false echo level enables the sensor to reliably pick up the useful echo. With the adjustment software VVO on the PC, you can view the echo image (see manual VVO "Sen­sor optimisation".
16 VEGASON 54P … 56P
Mounting and installation

2.3 Measurement of liquids

Flat vessel top
On flat vessels, mounting is usually done on a very short DIN socket piece. The lower side of the flange is the reference plane. The transducer should protrude out of the flange pipe.
< 100 mm
VEGASON 55 in flange version on short DIN socket piece
< 400 mm
Reference plane
Min. distance Type 54: 1 m Type 55: 0.8 m
Reference plane
A mounting location directly on the vessel top is ideal. A round opening in the vessel top is sufficient to fasten the sensor with the flange or version B and C with the swivelling holder.
Reference plane
Min. meas. distance
1.8 m
Swivelling holder
Min. meas. dis-
tance 1.4 m
Reference plane
Flange version and swivelling holder on flat vessel top
It is also possible to mount version C sensors in a 1" thread.
Min. distance
1.8 m
< 60 mm
Reference plane
VEGASON 56 in flange version on short DIN socket piece
Mounting of the transducer in a 1" thread
VEGASON 54P … 56P 17
Dished tank top
On dished tank tops, please do not mount the instrument in the centre or close to the vessel wall, but approx. ½ vessel radius from the centre. Dished tank ends can act as paraboloidal reflectors. If the transducer is placed in the focal point of the parabolic tank, the ultrasonic sensor receives amplified false echoes. The transducer should be mounted outside the focal point. Parabolically amplified echoes can be thereby avoided.
Mounting and installation
Reference plane
< 100 mm
½ vessel radius
VEGASON 54 on dished tank top; applies also to VEGASON 55
< 400 mm
1
/2 vessel radius
VEGASON 56 on dished tank top
Reference plane
Open vessels
On open vessels, the sensors can be readily installed on extended mounting brackets. Mount the low-weight sensor onto such a bracket and ensure a sufficient distance from the vessel wall.
Reference plane
Min. meas.
Min. meas. distance
distance
Reference plane
Open vessel
18 VEGASON 54P … 56P
Mounting and installation
Pump shaft
Narrow, uneven shafts, wells and vessel openings with very rough walls and shoul­ders make ultrasonic measurements ex­tremely difficult due to strong false echoes. This problem can be overcome by using an extended socket piece or a complete meas­uring tube (see chapter "2.5 Socket exten­sion").
see "2.5 Socket extensions"
Socket piece
Shaft pump
Measuring tube socket or measuring tube in a shaft
250 mm
Shaft pump
Measuring tube
min. distance
Meas. range

2.4 Measurement of solids

Flange mounting
For containers with solids, as with applica­tions for liquids, the instrument can be mounted on a short DIN socket piece. The transducer end should protrude out of the socket so that it does not interfere with ultra­sonic signals. The swivelling holder enables not only correct orientation to the product surface but also minimisation of possible false echoes.
Different filling conditions often lead to a vary­ing product surface orientation. This causes the useful echo to vary in quality. For this reason, the transducer should be mounted in such a way that, even in an empty vessel, the false echo intensity remains as low as possi­ble. You can view the echo curve on the PC with the adjustment program VVO.
Reference plane
Shaft
Min. distance
Very good measuring results can be attained with a measuring tube in continuous narrow shafts, see illustration. The applied measur­ing tube must have smooth walls inside (e.g. PE sewage pipe) and a diameter 200 mm. This arrangement works well as long as the inside of the measuring tube collects no dirt or buildup (cleaning necessary). You might want to consider using hydrostatic pressure transmitters or capacitive measuring probes. The measuring tube should either be never immersed in the medium, or always im­mersed, making the measurement take place exclusively in the tube.
VEGASON 54P … 56P 19
VEGASON 54C with adapter flange on a DN 200 vessel flange
Mounting and installation
Mounting boss
Reference plane
Min. distance
VEGASON 56 in 1" mounting boss.
The socket axis should be oriented to the product surface. More suitable is the use of a swivelling holder version (type B, C), as described on previous pages.
Material heaps

2.5 Socket extensions

Ultrasonic sensors require a min. distance to the liquid or solid. Take the min. distance into account in your planning. In some situations, it is possible to reach the required min. dis­tance, and hence the desired filling height with a socket extension. However, the socket piece increases the noise level of the ultra­sonic signal at the extension outlet and can interfere with the measurement. Only use a socket extension if all other possibilities have to be excluded. Carry out the extension as shown in the following illustration.
Socket extensions for liquids
Chamfer and deburr the socket carefully and make sure that it has a smooth inner surface. The socket should not protrude into the measured product, in case buildup can form on the socket through dirt or product residues.
For non-adhesive measured products, a socket extension in the form of a measuring tube can be permanently submerged in the product. The ultrasonic measurement is then made exclusively in the measuring tube and works very well without interference from other vessel installations (see "Pump shaft").
Large material heaps are usually measured with several instruments, which can be mounted on e.g. traverse cranes. If material cones form, we recommend orienting the sensors to the solid surface.
Socket piece should not be immersed into adhesive products (figure: VEGASON 54)
Transducer on traverse crane above a material heap
20 VEGASON 54P … 56P
Mounting and installation
The socket diameter should be as large and the socket length as small as possible. To minimise false echoes, make sure that the socket outlet is burr-free.
Typ e 5 4
L
45˚
ø
Socket extensions not immersed into the medium
Typ e 5 5
L
45˚
ø
Max. socket length in relation to socket diameter
ø in mm L in mm
Type 54 Type 55 Type 56
200 400 – – –– 250 500 500 500 300 –– –– 600

2.6 False echoes

The mounting location of the ultrasonic sensor must be selected such that no installations or inflowing material are in the path of the ultra­sonic impulses. The following examples and instructions show the most frequent measur­ing problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to their strong false echoes, adversely effect the measurement. Shields above these flat protrusions scatter the false echoes and guarantee a reliable measurement.
Correct Incorrect
Socket extension in solids
For solids, use a conical socket extension with a taper of at least 15° … 20°.
Vessel protrusions (slope)
Intake pipes, e.g. for the mixing of materials, with a flat surface directed towards the sen­sor should be covered with a sloping shield. The shield will scatter the false echo.
Correct Incorrect
15˚ 15˚
Socket extension in solids
Vessel protrusions (intake pipe)
VEGASON 54P … 56P 21
Mounting and installation
Vessel installations
Vessel installations such as, for example, a ladder, often cause false echoes. Make sure when planning your measurement loop that the ultrasonic signals have free access to the measured product.
Correct Incorrect
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can cause strong false echoes that are superim­posed over the useful echo signal. Small shields effectively hinder a direct false echo reflection. The false echoes are scattered and diffused in the area and are then filtered out as "echo noise" by the measuring elec­tronics.
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and not the inflowing mate­rial.
Correct
Correct
Inflowing material
Incorrect
Incorrect
Buildup
Correct Incorrect
If the sensor is mounted too close to the vessel wall, residues or buildup of the meas­ured product on the vessel wall can cause false echoes. Position the sensor at a suffi­cient distance from the vessel wall. Please
Shields
also note chapter "2.2 General installation instructions".
Struts
22 VEGASON 54P … 56P
Mounting and installation
Correct
Buildup
Incorrect
Strong product movements
Strong turbulence in the vessel, e.g. caused by powerful stirrers or intense chemical reac­tions, seriously interfere with the measure­ment. A surge or bypass tube of sufficient size (DN 200, DN 250) always allows, pro­vided the product causes no buildup in the tube, a reliable measurement even with strong turbulence in the vessel.
100 %

2.7 Installation mistakes

Foam generation
Thick foam on the product can cause incor­rect measurements. Take measures to avoid foam, carry out the measurement in a bypass tube, or use a different measurement technol­ogy, e.g. capacitive probes or hydrostatic pressure transmitters.
Foam generation
Wrong orientation to the product
Weak measuring signals are the result if the sensor is not directly oriented to the product surface. Orient the sensor axis perpendicu­larly to the product surface to achieve opti­mum measuring results.
60 %
0 %
Strong turbulence
Orient the sensor perpendicularly to the product surface
VEGASON 54P … 56P 23
Mounting and installation
Strong temperature fluctuations
Strong temperature fluctuations, e.g. caused by the sun, cause measuring errors. Please provide a sun shield.
Sun shield
Strong heat fluctuations
Min. distance to the medium
If the min. distance to the medium is not main­tained, the instruments show wrong meas­ured values. Mount the instrument at the required min. distance.
Sensor too close to vessel wall
Correct Incorrect
If the sensor is mounted too close to the vessel wall (dimension A in diagram), strong false echoes can be caused. Buildup, rivets, screws or weld joints on the vessel wall su­perimpose their echoes on the product, i.e. the useful echo. Please ensure a sufficient distance from the sensor to the vessel wall depending on the maximum measuring dis­tance (dimension B in diagram). In case of good reflecting conditions (liquids, no vessel installations), we recommend determining the sensor distance according to Diagram curve
1. At a max. measuring distance of e.g. 10 m, the distance of the transducer (acc. to curve
1) should be approx. 1.5 m. In case of solids with poor reflective properties, determine the distance to the vessel wall according to Dia- gram curve 2. Under very bad measuring conditions (rough vessel walls, struts), it might be necessary to increase the distance to the vessel wall, or to also filter out the false echoes by storing them in memory, thereby adapting the sensor more precisely to the environment.
Distance from the transducer to the vessel wall
A
2 m 4 m 6 m 8 m
Curve 1 (liquids)
10 m
B
20 m
Sensor too close to the vessel wall
24 VEGASON 54P … 56P
30 m
max. meas. distance
Curve 2 (solids)
Mounting and installation
Parabolic effects of rounded or arched vessel tops
Round or parabolic tank tops act like a para­bolic mirror on the signals. If the sensor is placed at the focal point of such a parabolic tank top, the sensor receives amplified false echoes. The optimum location is generally in the area of half the vessel radius from the centre.
Correct
< 100 mm
~ ½ vessel radius
Incorrect
Socket piece too long
If the sensor is mounted in a socket piece that is too long, strong false echoes are gen­erated and measurement is made more diffi­cult. Make sure that the transducer protrudes at least 30 mm out of the socket piece.
Reference plane
< 100 mm
Incorrect
Correct and wrong length of socket piece
Mounting on a vessel with parabolic tank top
VEGASON 54P … 56P 25

3 Electrical connection

Electrical connection

3.1 Connection

Safety information – Qualified person­nel
Instruments which are not operated with protective low voltage or DC voltage must only be connected by qualified personnel. This also applies to the configuration of measuring systems planned for Ex environ­ment.
As a rule, do all connecting work in the com­plete absence of line voltage. Always switch off the power supply before you carry out connecting work on the ultrasonic sensors. Protect yourself and the instruments.
Note, important for connection to an EP card
If VEGASON 54P … 56P sensors are oper­ated on Profibus input cards (EP cards) in the VEGALOG 571 processing system, each card must be powered individually with an own power supply unit. For power supply of an EP card, use the power supply unit "ASI­Interface SLA 3.100" from Puls which is avail­able from VEGA.
By this measure, the signal circuits are gal­vanically separated. This avoids possible crosstalks between the EP cards.
3.2 Connection cables and bus con­figuration
Note the Profibus specification. The connec­tion cables must be specified for the ex­pected operating temperatures in the plant and must have an outer diameter of 6 … 12 mm, to ensure the seal effect of the cable entry on the sensor.
For power supply and bus communication, a two-wire cable acc. to the Profibus specifica­tion (up to max. 2.5 mm2 cross-section area of conductor) is used. The electrical connec­tion on the sensor is made by spring-loaded terminals.
In a laboratory setup, a Profibus system will also work with standard, unshielded two-wire cable. In practice however, an automation network and bus system can only be pro­tected reliably against electromagnetic inter­ference with screened cable. Acc. to the Profibus specification (IEC 1158-2), screened and twisted cables are prescribed.
All participants are connected in one line (serially). At the beginning and end of the bus segment, the bus is terminated by an active bus termination. On the DP bus level, most participants already have a bus termi­nation implemented. With more than 32 par­ticipants on the DP level, a so-called repeater must be used to open and combine another DP level with a max. of 32 additional partici­pants. On the PA bus branch of the segment coupler, the PA ultrasonic sensors also work with a maximum of 32 participants (Ex max. 10 participants).
A PA sensor can work only in conjunction with a Profibus DP system, to which a Profibus PA subsystem is connected. A PA Profibus par­ticipant must consume min. 10 mA supply current.
26 VEGASON 54P … 56P
Electrical connection
Cable length
Connection cables must correspond to the Profibus specification and the FISCO model. The sensor cable must be in conformity with the values of the reference cable acc. to IEC 61158-2:
0.8 mm2; R Z
= 80 … 120 ; damping = 3 dB/km;
31.25kHz
C
asymmetric
The max. cable length first of all depends on the transmission speed: up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr up to 94 kbit/s: 1200 m Profibus DP up to 188 kbit/s: 1000 m Profibus DP up to 500 kbit/s: 500 m Profibus DP up to 1500 kbit/s: 200 m Profibus DP up to 12000 kbit/s: 100 m Profibus DP
The distributed resistance of the cable, in conjunction with the output voltage of the segment coupler and the current requirement (VEGASON 10 mA) or the voltage require­ment (VEGASON 9 V) of the sensors, deter­mines the max. length of the cable.
In a practical application of a PA bus branch, the max. length of the cable is also deter­mined (beside the required supply voltage and max. current consumption of all partici­pants on the PA bus branch) by the bus structure and the type of segment coupler used.
The cable length results from the sum of all cable sections and the length of all stubs. The length of the individual stubs must not exceed the following lengths: 1 … 12 stubs 120 m (Ex: 30 m) 13 … 18 stubs 60 m (Ex: 30 m) 19 … 24 stubs 30 m (Ex: 30 m)
More than 24 stubs are not permitted, whereby each branch longer than 1.2 m is counted as a stub. The total length of the cable must not exceed 1900 m (in Ex version 1000 m).
= 44 Ω/km;
DCmax.
= 2 nF/km.
ofibus Pofibus P
ofibus P
ofibus Pofibus P
AA
A
AA
Ground terminal
The electronics housings of the sensors have a protective insulation. The ground terminal in the electronics is galvanically connected with the metallic process connection. For sensors with a plastic thread as process fitting, the sensor must be grounded by connecting a ground cable to its external ground terminal.
Screening
"Electromagnetic pollution" caused by elec­tronic actuators, energy cables and transmit­ting systems has become so pervasive that shielding for normal two-wire bus cable is usually a necessity. According to the Profibus specification, the screening should be made on both ends. To avoid potential equalisation currents, a potential equalisation system must be provided in addition to the screen­ing.
According to specification, we recommend the use of twisted and screened two-wire cable, e.g.: SINEC 6XV1 830-5AH10 (Sie­mens), SINEC L26XV1 830-35H10 (Siemens), 3079A (Belden).
Alternatively, when grounding at both ends in non-Ex areas, the cable shielding can be connected on one ground side (in the switch­ing cabinet) via an YC-capacitor (e.g. 10 nF, 1500 V) to the ground potential. Make sure that the ground connection has the lowest possible resistance (foundation, plate or mains earth).
Profibus PA in Ex environment
When used in Ex area, a PA bus with all con­nected instruments must be carried out in intrinsically safe protection class "i". Four-wire instruments requiring separate supply must at least have an intrinsically safe PA connec­tion. VEGA sensors for PA-Ex environment are generally "ia” two-wire instruments.
VEGASON 54P … 56P 27
Electrical connection
In the so-called Fieldbus Intrinsically Safe Concept (FISCO), the general conditions for an Ex safe bus configuration have been laid down. Therein, the electrical data of the participants and the bus cable have been determined, so that the linking of these com­ponents always meets Ex requirements. This makes the otherwise time-consuming Ex calculation unnecessary. You can build your Ex bus according to the IEC standard 1158-2.
The Ex segment coupler delivers a controlled power supply to the PA bus. It acts as source in the PA branch. All other compo­nents (field instruments and bus terminators) are only consumers. A field instrument must consume at least 10 mA. Ideally, an individual sensor should not consume more than 10 mA, so that the number of instruments can be as large as possible.
VEGA PA sensors, whether Ex or non Ex, consume a constant current of 10 mA. Ac­cording to the Profibus specification, this is the minimum participant current. With VEGA sensors it is therefore possible to connect 10 sensors (also in Ex environment) even with a limited energy supply from the Ex segment couplers.
Watch out for potential losses
Due to potential transfer, earthing on both sides without a potential equalisation system is not allowed in Ex applications. If an instru­ment is used in hazardous areas, the re­quired regulations, conformity and type approval certificates for systems in Ex areas must be noted (e.g. DIN 0165). Please also note the approval documents with the safety data sheet attached to the Ex sensors.
Electrical data of the cables
R
DC
No. of A in Z cores mm
2
31.25kHz
C in Damping Screen nF/km
SINEC 6XV1 44 Ω/km 2 0.75 100 Ω < 90 < 3 dB/km Cu braiding 830-SEH10 +/- 20 39 kHz (Siemens)
SINEC L26XV1 44 Ω/km 2 0.75 100 Ω < 90 < 3 dB/km Cu braiding 830-SFH10 +/- 20 39 kHz (Siemens)
3079A 105 /km 2 0.32 150 29.5 < 3 dB/km Foil (Belden) 39 kHz
28 VEGASON 54P … 56P
Electrical connection

3.3 Sensor address

In a Profibus system composed of Profibus DP and Profibus PA subsystem, each partici­pant must have a unique address. Each participant, whether master or slave, is accessed by means of its own address in the bus system. The address of a partici­pant, whether on DP or PA level, should be assigned before connecting to the bus, be­cause an address can be used only once. If an address is used twice, interference in the bus will be caused.
The address of an ultrasonic sensor can be set in two ways:
- with the adjustment software VVO (soft­ware addressing) or
- with the DIP switch block in the sensor (hardware addressing).
VEGA Profibus sensors are delivered with the address set at 126 (all DIP switches to "ON"). Remember, in a Profibus system there are max. 126 participants possible. When the DIP switch is set to address 126 (or higher), the address can be adjusted with the adjust­ment software VVO, the adjustment module MINICOM or another configuration tool (e.g. PDM). However, there can be only one sen­sor on the bus with address 126 (delivery status) during address assignment via soft­ware. For that reason, hardware addressing (DIP switch) before connection to the bus is recommended.
Hardware addressing
The DIP switches generate an address number in the binary system. This means that, from right to left (ascending), any switch represents a number twice as high as the previous switch on the right. The corre­sponding number in the decimal system results from the sum of all switches set to "ON". In the illustration, you see the decimal number that corresponds to each individual DIP switch.
DIP switch 8 corresponds to the number 128, switch 1 corresponds to the number 1 and switch 3 corresponds to the decimal number
4.
1
2
8765 4
128
64
32
Example 1
The switches 3, 5 and 7 are set to "ON". The address is then: DIP switch 3 to "ON" means 4 DIP switch 5 to "ON" means 16 DIP switch 7 to "ON" means 64
The sum is: 4 + 16 + 64 = Address 84
3
1
2
4
8
16
ON
1
2
8765 4
64
64 + 16 + 4 = 84
VEGASON 54P … 56P 29
3
16
4
Electrical connection
Example 2
You want to set address 27. 16 + 8 + 2 + 1 = 27
You must set the DIP switches 5 = 16 4 = 8 2 = 2 1 = 1 to "ON".
Example 3
You want to set address 99 64 + 32 + 2 + 1 = 99
You must set the DIP switches 7 = 64 6 = 32 2 = 2 1 = 1 to "ON".
Software addressing
The DIP switches must be set to an address of 126 … 255, i.e.
- either all DIP switches are set to "ON", corresponding to address 255 (delivery status)
OFF
1
2
64
3
ON
8765 4
32
16
1
2
3
1
2
4
8
8765 4 Addr.
- or only DIP switch 8 is set to "ON", corre­sponding to address 128.
128
Of course, software addressing is also pos­sible, if the switches 7 … 2 are set to "ON" (address 126).
The adjustment of the address with software VVO is described in chapter "6.2 Adjustment with VVO" under the heading "Software ad­dressing" or in chapter "6.3 Sensor adjust­ment with the adjustment module MINICOM".
30 VEGASON 54P … 56P
Electrical connection

3.4 Connection of the sensor

After mounting the sensor at the measure­ment location according to the instructions in chapter "4 Mounting and installation", loosen the closing screws on top of the sensor. The sensor lid with the optional indication display can then be opened. Unscrew the sleeve nut and slip it over the connection cable (after removing about 10 cm of insulation). The sleeve nut of the cable entry has a self-lock­ing ratchet that prevents it from opening on its own.
Version with Aluminium housing
Power supply and Profibus signal
+
To the indicating instrument in the sensor lid or to the external indicating instrument VEGADIS 50
M20x1.5 (diam­eter of the connection cable 6…9 mm)
Now insert the cable through the cable entry into the sensor. Screw the sleeve nut back onto the cable entry and clamp the stripped wires of the cable into the proper terminal positions.
The terminals hold the wire without a screw (spring-loaded terminals). Press the white opening tabs with a small screwdriver and insert the copper core of the connection cable into the terminal opening. Check the hold of the individual wires in the terminals by lightly pulling on them.
Version with plastic housing
Power supply and Profibus signal
M20x1.5 (diam­eter of the connection cable 6…9 mm)
+
To the indicating display in the sensor lid or to the external indicating instrument
Spring-loaded terminals (max. 2.5 mm2 wire cross-section)
1
2
3
87654
+1 2- 5 6 7 8
Addr.
Bus
Display
ON
ESC
-
+
OK
1
2
3
8 765 4
Opening tabs
Pluggable adjustment module MINICOM
+1 2- 5 6 7 8
Addr.
Bus
Tank 1 m (d)
12.345
Display
ON
ESC
+
-
OK
VEGASON 54P … 56P 31

3.5 Connection of the external indicating instrument

Loosen the four screws of the housing lid on VEGADIS 50. The connection procedure can be facilitated by fixing the housing cover during connec­tion work with one or two screws on the right of the housing.
Adjustment module
OUTPUT (to the sensor)
3
2
1
4
5
8
6
7
VEGADIS 50
+
-
Tan k 1 m (d)
12.345
Electrical connection
ESC
OK
Voltage supply and digital meas. signal
-
+
8 7654
+1 2- 5 6 7 8
Addr.
Bus
Tank 1 m (d)
12.345
DISPLAY (in the lid of the indicating instrument)
1
2
3
Display
ON
ESC
+
-
OK
Screws
32 VEGASON 54P … 56P
Electrical connection

3.6 Bus configuration

The type of ultrasonic sensor you use de­pends on the process requirements and mounting conditions, as well as on the re­quirements of your control, regulative, or process control system.
VEGASON 54 … 56 Profibus sensors are sensors for use in Profibus PA environment. Profile 3 has been implemented in the sen­sors. A measuring system consists of one or several sensors, one or several segment couplers and one DP master computer, such as e.g. a S7 PLC with Profibus interface or a process control system with Profibus DP­Master-Slot. The processing unit, e.g. the PLC, evaluates the level-proportional, digital measuring signals in a number of evaluation routines and puts them to use process-spe­cifically.
On the following four pages you will see schematic illustrations of the bus configura­tion.
The automation system as Master-Class 1 takes over the complete bus control. It reads out all signals cyclically and, if necessary, gives instructions to the participants (e.g. sensors). Additional master systems (e.g. visualisation systems or adjustment tools) can also be connected to the DP bus. These systems operate as so-called Master-Class 2 participants. Like the Master-Class 1 system, they can read out signals, give instructions and operate in the acyclical mode.
Each participant on the bus must have an unambiguous address. The addressing covers both bus levels. A Profibus DP net­work can have max. 126 participants, includ­ing all participants on the PA level. In practice, each Master-Class 1 computer gets address 1 and the Master-Class 2 comput­ers address 10 … 20. As a rule, the slaves or participants get the addresses 21 … 126. On the Profibus PA network segment, a max. of 32 sensors can be connected on one PA segment coupler.
Ex environment
In Ex environment, intrinsically safe (EEx ia) PA sensors are used with Ex segment cou­plers. Generally, the number of PA sensors on a segment coupler (Ex or non Ex) de­pends on the current requirement of the sensors and on the current supplied by the segment coupler. Segment couplers for EEx ia environment provide 90 … 110 mA. The number of sensors results from the sum of:
- the basic current intake of all sensors
- plus 9 mA communication signal
- plus the leakage currents of all sensors
- plus a recommended current reserve (approx. 10 mA)
A DP bus does not allow power supply via signal cable, whereas the PA bus does. Both, DP and PA, require at least one screened two-wire cable. The DP bus can additionally have up to 8 cores (screened), some of which can be supply cables (see also "Instal­lation Guides PA + DP" of the Profibus User Organisation (PNO).
VEGASON 54P … 56P 33
Profibus PA sensors on the Profibus network
1
Electrical connection
Master-Class 1
Bus terminator
3...9
Profibus PA (31,25 kBit/s)
Profibus DP
21
Profibus interface card
RS 232
22...54
3
RS 485
10
Master-Class 2
Segment coupler
Bus terminator
2
22
23
24
53
54
VEGACONNECT 3
PA segment on segment coupler: 1 … 32 sensors on one two-wire cable (Ex: 10 sensors)
34 VEGASON 54P … 56P
Electrical connection
Profibus DP segment level 1 … 126 participants including all DP and PA participants. Through segment couplers and PA segments, the trans­mission rate, also on the DP level, is determined by the slowest coupler/participant on the Profibus DP and PA network.
Bus terminator
3...9
M
Segment coupler
3...9
3~M
89
90
Bus terminator
2
55
56...88
2
Profibus PA
56
57
87
88
PA segment: 1 … 32 sensors on one two-wire cable (Ex: 10 sensors)
VEGASON 54P … 56P 35
Electrical connection
Profibus PA sensors with 4 … 20 mA sensors on the Profibus network
1
Master-Class 1
Bus terminator
VEGALOG
Profibus PA (31,25 kBit)
3~M
3…9
Profibus DP
Profibus interface card
21
3
RS 485
22
10
Master-Class 2
VEGACONNECT 3
4
RS 232
1
2
3
5
11
4
1 … 15 PA sensors per two-wire cable
13
12
15
14
with independent address zone (Ex: 10 sensors)
36 VEGASON 54P … 56P
Electrical connection
Profibus DP segment level 1 … 126 participants including all DP and PA participants. Up to 12 MBit/s transmission rate on DP level. In the PA segments 31.25 kBit/s transmission rate.
VEGACONNECT 3
4 … 20 mA (HART )
2
4
4
4
2
2
2
Profibus PA (31,25 kBit)
3…9
23
M
3…9
24
25
VEGALOG
VBUS
Outputs
2
2
2
2
®
2
2
2
2
2
0/4…20 mA
0…10 V
2
VBUS
Bus termi­nator
Profibus PA: 1 … 15 sensors per two-wire cable (Ex: 10 sensors) with independent address zone
VBUS: 1 … 15 sensors per two­wire cable Exd: also 15 Ex ia: 5 sensors
VEGASON 54P … 56P 37

4 Setup

4.1 Adjustment media

Adjustment with PACTware
Setup
TM
In chapter "1.4 Adjustment" the Profibus ad­justment scheme was briefly explained and the adjustment media for VEGA Profibus sensors were shown. All VEGA Profibus sensors operate in profile 3 and can be ad­justed with:
- the adjustment program VVO on a PC with Profibus card
- the adjustment program PACTwareTM, under which VVO runs as a subprogram
- the Siemens software PDM in conjunction with an EDD (Electronic-Device-Descrip­tion)
- the adjustment module MINICOM in the sensor.
Adjustment with VV O on the PC
The adjustment program VVO enables user­friendly adjustment of VEGA Profibus PA sensors. All functions and options relating to the sensor adjustment are accessible. The program runs under Windows® on a PC with a Profibus-Master-Class 2 interface card on Profibus DP level as Master-Class 2 tool. The VVO program accesses the VEGA PA sen­sors via the DP bus, the segment coupler and the PA bus.
The adjustment with PACTwareTM corre­sponds to VVO adjustment. In this case, VVO runs as a subprogram of PACTwareTM. The adjustment instructions can be found in the documentation of PACTwareTM.
Adjustment with PDM
The sensors can be adjusted completely with PDM. However, some convenient functions and many special features, like e.g. display of an echo curve, are not available. In addi­tion to the PDM software, an EDD (available upon request from VEGA) is required for each sensor type. The adjustment instruc­tions for PDM are described in the PDM documentation.
Adjustment with the adjustment module MINICOM
With the adjustment module MINICOM, you adjust the individual sensor directly in the sensor or in the external indicating instrument VEGADIS 50. The adjustment module MINICOM enables (with the 6-key adjustment field with text display) all essential functions of parameter setting and adjustment.
Note:
Please note that for adjustment with VVO, version 2.70 or higher is required.
38 VEGASON 54P … 56P
Setup

4.2 Adjustment with the PC

The instructions for the adjustment with the PC and the adjustment program VVO can be found in the manual "VEGA Visual Operating (VVO)" which is supplied with the latest ad­justment software.
Beside the adjustment of ultrasonic sensors, the manual "VEGA Visual Operating (VVO)" also describes the setup of all other VEGA sensors with the PC.

4.3 Adjustment with the adjustment module MINICOM

In addition to the PC, you can adjust the sensor with the small, detachable adjustment module MINICOM. The adjustment module is plugged into the sensor or into the external indicating instrument (optional).
ESC
+
-
Tank 1 m (d)
12.345
OK
2
Tank 1 m (d)
12.345
4
All sensor versions (adjustment options) provided by a PC with the adjustment pro­gram VVO are also available with the adjust­ment module MINICOM. However, the adjustment with MINICOM looks different.
+
-
4 ... 20 mA
ESC
OK
You can carry out all adjustment steps with the 6 keys of the adjustment module. A small display shows you, apart from the measured value, a short message on the menu item or an entered numerical value.
Although the volume of information of the small display cannot be compared with that of the adjustment program VVO, you will be able to carry out your settings quickly and efficiently with the small MINICOM.
VEGASON 54P … 56P 39
Setup
Error codes:
E013 No valid measured value
- sensor in the warm-up phase
- loss of the useful echo E017 Adjustment span too small E036 Sensor program not operating
- sensor must be reprogrammed
(service)
- fault signal also appears during
programming
E040 Hardware failure, electronics
defective
E113 Hardware failure
Adjustment steps
On the following pages, you will find the com­plete menu schematic of the adjustment mod­ule MINICOM. Set up the sensor in the numbered se­quence:
1. Address
2. Measurement in gases
3. Operating range
4. Adjustment
5. Conditioning
6. Meas. conditions
7. False echo storage (only required if errors
occur during operation).
8. Indication of the useful level and noise level
9. Outputs Below you will find short explanations to setup steps 1 … 9.
3. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range was already entered with the min./max. adjustment. Generally, it is useful to choose a slightly wider range (approx. 5 %) for the operating range than for the measuring range.
Example: Min./Max. adjustment: 0.300 … 5.850 m; adjust operating range to approx.
0.250 … 6.000 m.
4. Adjustment
Max.
Min.
Under the menu item " the sensor of the measuring range.
100 % (1.270 m) correspond to 1200 liters
Span (4.58 m)
0 % (5.850 m) correspond to 45 liters
Adjustment
", you inform
1. Address
Choose a free bus address with the DIP switch (see chapter "4.2 Sensor address").
You can carry out the adjustment with or without medium. Generally, you will carry out the adjustment without medium, as you can then adjust without a filling/emptying cycle.
2. Measurement in gases
Adjustment is only necessary if the measure­ment is made in gases other than air (CO2, He, etc.). When measuring in gases, sound the distance from the sensor to the product surface and enter this in the menu item "
Measurement in gases
take the altered (compared to air) sonic ve­locity into account and output correct levels.
40 VEGASON 54P … 56P
". The sensor can then
Setup
Adjustment without medium
(adjustment independent of the actual level)
Key Display indication
Sensor
m(d)
4.700
Para­meter
OK
OK
OK
OK
+
adjust­ment
Adjust­ment
w.o. medium
Ad­just­ment in
m(d)
(min. adjustment)
The distance indication flashes and you can choose "feet" and "m".
OK
+
or
Confirm the setting with "OK".
m(d)
0.0%
at
m (d)
XX.XXX
Ad­just­ment in
With "+" and "–" you set the percentage value for the min. value or the lower level (exam­ple 0.0 %).
OK
The entered percentage value is written in the sensor and the distance corresponding to the percentage value begins flashing.
+
or
With the "+" or "–" key you can assign a level distance (exam­ple 5.85 m) to the previously adjusted percentage value. If you do not know the distance, you have to measure it by sounding.
OK
The entered product distance is written in the sensor and the display stops flashing.
You thereby adjusted the lower product dis­tance as well as the percentage filling value corresponding to the lower product distance.
100.0%
at
m (d)
XX.XXX
(max. adjustment)
Enter the max. adjustment (upper product distance) in the same manner (example: 100 % and 1.270 m).
Note:
The difference between the adjustment val­ues of the lower product distance and the upper product distance should be as big as possible, preferably at 0 % and 100 %. If the values are very close together, e.g. lower product distance adjustment at 40 % (3.102 m) and upper product distance ad­justment at 45 % (3.331 m), the measurement will be inaccurate. A characteristics curve is generated from the two points. Even the smallest deviations between actual product distance and entered product distance will considerably influence the slope of the char­acteristics curve. If the adjustment points are too close together, small errors inflate to considerably larger ones when the 100 % or the 0 % value is outputted.
VEGASON 54P … 56P 41
Setup
Adjustment with medium
with medium
Max.
Min.
adjust at
adjust
%
at %
XXX.X
XXX.X
Fill the vessel e.g. to 10 % and enter 10 % in the menu "
Min. adjust
" with the "+" and "–" keys. Then fill the vessel, e.g. to 80 % or 100 % and enter 100 % in the menu "
adjust
" with the "+" and "–" keys.
Max.
5. Conditioning
Signal condit­ioning
Scal­ing
0 %
100 %
Deci-
prop.
corres-
corres­ponds
XXXX
mal point
888.8
Conditioning
ponds
XXXX
Under the menu item " assign a product distance at 0 % and at 100 % filling. Then you enter the physical quantity and the physical unit as well as the decimal point.
to
Mass
Unit
Kg
", you
With the "—>" key you change to the 100 % menu. Enter here the numerical value of your parameter corresponding to a 100 % filling. In the example 1200 for 1200 liters.
• Confirm with "OK".
If necessary, choose a decimal point. How­ever, note that only max. 4 digits can be displayed. In the menu "
prop. to
", you choose the physical quantity (mass, volume, dis­tance…) and in the menu "
Unit
" the physical
unit (kg, l, ft3, gal, m3 …).
Linearisation:
Adjust­ment
Signal condit­ioning
Scal­ing
Lin. curve
Linear
Integra tion time
0 s
A linear correlation between the percentage value of the product distance and percent­age value of the filling volume is preset. With the menu "Lin. curve" you can choose be­tween linear, spherical tank and cylindrical tank. The generation of a customised linearisation curve is only possible with the PC and the adjustment program VVO.
6. Meas. conditions
(see menu schematic)
Enter in the menu window "
0 % corresponds
7. False echo storage
" the numerical value of the 0 % filling. In the example of the adjustment with the PC and the adjustment software VVO, this would be 45 for 45 liters.
A false echo storage is always useful when unavoidable false echo sources (e.g. struts) must be minimised. By creating a false echo storage, you authorise the sensor electronics
• Confirm with "OK".
to record the false echoes and save them in an internal database. The sensor electronics treats these (false) echoes differently from the useful echoes and filters them out.
42 VEGASON 54P … 56P
Setup
8. Signal and noise level
In the menu
Ampl.:
XX dB
S-N:
XX
dB
you get important information on the signal quality of the product echo. The higher the "S-N" value, the more reliable the measure­ment (menu schematic MINICOM).
Ampl.: means amplitude of the level echo in
dB (useful level)
S-N: means Signal-Noise, i.e. the useful
level minus the level of the back­ground noise
The bigger the "S-N" value (difference be­tween the amplitudes of the useful signal level and the noise level), the better the measure­ment: > 50 dB Measurement excellent 40 … 50 dB Measurement very good 20 … 40 dB Measurement good 10 … 20 dB Measurement satisfactory 5 … 10 dB Measurement sufficient < 5 dB Measurement poor
Example:
Ampl. = 68 dB S-N = 53 dB
68 dB – 53 dB = 15 dB noise level
A signal distance of 53 dB means a very high degree of measurement reliability.
9. Outputs
Under the menu "Outputs" you determine, for example, if the current output should be inverted, or which unit of measurement should be shown on the sensor display.
VEGASON 54P … 56P 43
Menu schematic for the adjustment module MINICOM
Ultras onic
Para­meter
97,5
%
Sensor opti­mize
SON 52
When switching on, the sensor
P
type and the software version are displayed for a few seconds.
3.00
Confi­gura­tion
1.
Sensor addr.
(•–/ –•)
126
Meas. enviro nment
Sensor Tag
Sensor
Setup
Sensor address:
• Sensor address here only adjustable, if the DIP switch in the sensor is set to address greater than/equal to 126.
• If the DIP switch with number 8 is set to "On" (address 128), it is possible to adjust address 1 … 126 here.
Meas. unit
m (d)
Adjust ment
w.out medium
Opera­ting range
Begin
0.50
4.
m (d)
3.
End
m (d)
6.00
Meas. condit ions
Condit ion
liquid
Condit ion
Solid
6.
Fast change
No
Fast change
No
with medium
Agitat ed sur face No
High dust level No
Foam­ing prod. No
Large angle repose No
Measur ing in gases No
Measur ing in gases No
Signal condit ioning
Sca­ling
Multi ple echo No
Multi ple echo No
5.
Lin. curve
Linear
Integr ation time
Meas. dist.
mm (d)
Correc tion Now!
0 s
Measur ing in gases
2,381
OK?
2.
Sound speed
m/s
Correc tion Now!
333
OK?
0.0 %
m (d)
100.0%
at
m (d)
XX.XXX
Min­adjust at %
XXX.X
Max­adjust at %
XXX.X
0 % corres ponds
XXXX
100 % corres ponds
XXXX
Deci­mal point
888.8
Prop. to
Mass
Unit
Kg
Adjust ment in
m(d)
at
XX.XXX
44 VEGASON 54P … 56P
Setup
With these keys you move in the menu field to the left, right, top and bottom
ESC
7. 8.
9.
act. dist.
m (d)
4.700
Update
Meas. dist.
m (d)
X.XX
Update Now!
Learn­ing!
OK?
False echo memory
Create new
Meas. dist.
m (d)
X.XX
Update Now!
Learn­ing!
Out­puts
OK?
Add’l func­tions
Info
Ampl.:
XX dB
S-N:
XX
dB
Delete
Delete Now!
OK?
Delet­ing!
Simulation:Simulation:
Simulation:
Simulation:Simulation:
Sensor Tag
Sensor
max. range
m (d)
7.000
One hour after the last simulation ad­justment, the sensor returns automati­cally to normal operating mode
Simu­lation
Sensor type
SON51 P
Act. dist.
Reset to de­fault
Reset Now!
OK?
Reset ing!
m (d)
4.700
Serial number
1094 0213
Ampl.: S-N:
Act. dist. m
X,XX
Lan­guage
Eng­lish
Softw.
Softw.
vers.
date
3.00
15.09. 1999
Tempe-
XX dB
rature
XX
dB
act. tempe­rature
18°C
Menu items in bold print provide sensor and measured value information and cannot be modified in this position.
OK
Min. tempe­rature
14°C
Sensor addr.
(•–/ –•)
126
Max. tempe­rature
59°C
PA output
Prop. to
di­stance
Failure mode
value
Sensor displ.
Prop. to
di­stance
Simu­lation Now!
Simu­lation
XXX.X
OK?
High dust level No
%
Fast change
No
Light grey menu fields are only displayed if required (dependent on the adjustments in other menus).
White menu items can be modified with the "+" or "–" key and saved with the "OK" key.
VEGASON 54P … 56P 45

5 PA function diagram and parameters

5.1 P A function diagram

Sensor offset (Float) [sensor unit (m,ft ...)]
Zero adjust
Zero offset
Raw-Distance (value, status) [in m]
Sensor value (Float) [sensor unit (m,ft ...)]
PA function diagram and parameters
F Time (Float) (sec) (not available over PA Parameter)
Filter
Raw-Temperature (value, status) [in K]
Cal point hi, cal point lo (Float) [sensor unit (m,ft ...)]
Level hi, level lo (Float) [level unit (%,m,ft ...)] (*1)
Cal type (Unsigned8) Lifeabgleich (min-max) und Trockenabgleich möglich
Cal
Level hi
Level lo
Cal point hi Cal point lo
Max min memory
max min
Simulate procent value (DS_50) [level unit (%,m,ft...)] (*1) (manuf. spec. parameter)
Simulation
Temperature (Float) [temperature unit (˚C, K, ...)]
Max temperature, min temperature (Float) [temperature unit (˚C, K, ...)]
off
on
Simulate value
Level (Float) [level unit (%,m,ft...)] (*1)
Level offset (Float) [level unit (%,m,ft...)] (*1)
Offset
Offset
46 VEGASON 54P … 56P
PA function diagram and parameters
Simulation distance value (DS_50) [sensor unit (m,ft ...)] (manuf. spec. parameter)
Simulation
off
on
Simulate value
Secondary value 2 (DS_33) [secondary value 2 unit, (m,ft ...)]
distance
Max min memory
max min
Secondary value 1 (DS_33) [secondary value 1 unit (%,m,ft...)] (*1)
Max sensor value, min sensor value (Float) [sensor unit (m,ft ...)]
Lin type (Unsigned8), Tab index (Unsigned8), Tab X Y value (2 x Float) [level unit (%,m,ft ...). (*1) primary value unit (%,m,ft ...)]. (*1) Tab min number (Unsigned8), Tab max number (Unsigned8), Tab op code (Unsigned8), Tab status (Unsigned8), Tab actual number (Unsigned8)
Linearization
level %
Transducer
Block
Primary value (DS_33) [primary value unit (%,m,ft...)] (*1)
lin %
A
B
distance
C
VEGASON 54P … 56P 47
lin % level % distance
Display source select (Unsigned8)
Source selector
scale
PV FTime (Float) [sec]
PA function diagram and parameters
Sensordisplay
VEGA
Hi hi limit, hi limit, lo limit, lo lo limit, alarm hys (Float) [out scale unit]
Alarm checkFilter
Hi hi limit Hi limit
Lo limit Lo lo limit
Alarm hyst
Hi hi alarm, hi alarm, lo alarm, lo lo alarm (DS_39) Alarm sum (DS_42)
lin %
A
B
C
level %
Channel (Unsigned16)
Channel
distance
Simulate Linearization value (DS_50) [primary value unit (%,m,ft ...).(*1) secondary value 1 unit, (%,m,ft ...).(*1) secondary value 2 unit, (m,ft ...) (dep. on channel)]
Simulation
off
on
Simulate value
48 VEGASON 54P … 56P
PA function diagram and parameters
Fail safe value (Float) [out scale unit]
Fail safe type (Unsigned8)
Target mode (Unsigned8)
Out (DS_33) [out scale unit]
Fail safe
Fail safe value
Divice status
PV scale (2 x Float) [primary value unit (%,m,ft ...).(*1) secondary value 1 unit, (%,m,ft ...).(*1) secondary value 2 unit, (m,ft ...) (dep. on channel)]
PV Scale
1
0
EU 0% EU 100%
Mode
Auto
Man
Mode & Status calc.
Out scale (DS_36) [out scale unit (included)]
Out Scale
EU 100%
EU 0%
Data value (DS_33) [out scale unit]
Actual mode (DS_37)
Function
Block
10
VEGASON 54P … 56P 49
PA function diagram and parameters

5.2 Status bytes

Coding of the status byte on the PA output value
Status code Descr. acc. to Profibus standard Probable causes in the VEGA sensor
0x00 bad – non-specific - Flash-Update active 0x04 bad – configuration error - Adjustment error
0x0c bad – device failure - Hardware error
0x10 bad – sensor failure - Meas. value error
0x1f bad – out of service - constant - "Out of Service" mode switched on 0x44 uncertain – last unusable value - Failsafe useable value (Failsafe-Mode =
0x48 uncertain – substitute set - Simulation switched on
0x4c uncertain – initial value - Failsafe substitute value (Failsafe-Mode =
0x51 uncertain – sensor conversion - Sensor value < lower limit
not accurate – low limited
0x52 uncertain – sensor conversion - Sensor value > upper limit
not accurate – high limited
0x80 good (non-cascade) - ok - ok
0x84 good (non-cascade) – active - Static revision (FB, TB) changed (active
block alarm for 10 sec., after parameter was written
0x89 good (non-cascade) – active - Lo-Alarm
advisory alarm – low limited
0x8a good (non-cascade) – active - Hi-Alarm
advisory alarm – high limited
0x8d good (non-cascade) – active - Lo-Lo-Alarm
critical alarm – low limited
0x8c good (non-cascade) – active - Hi-Hi-Alarm
critical alarm – high limited
Note: All 256 possible status byte values can also be generated by simulation. For this reason, the Failsafe-Mode should be set to "Wrong calculated value", so that the simulated status will not be modified by the Failsafe cal­culation.
- Configuration error with PV-Scale (PV-Span too small)
- Differences in meas. unit
- Error in the linearisation table
- Converter error
- Leakage pulse error
- Trigger error
- Temperature error
"Last valid value" and already valid meas. value since switching on)
- Failsafe substitute value (Failsafe-Mode = "Fsafe value")
"Last valid value" and no valid meas. value since switching on)
in static category)
50 VEGASON 54P … 56P
PA function diagram and parameters

5.3 Parameter listing

The following parameter listing and function diagram is used for setup with the automation system if there is no adjustment software available. The listing reflects the parameters of the function block and the transducer block and should only be used by experi­enced Profibus technicians. Please make sure that all VEGA sensors are Profile 3 sen­sors. As a rule, however, you will carry out the sensor setup more conveniently with the VEGA adjustment software VVO or with the adjustment module MINICOM in the sensor.
The PA parameters and the function diagram are normally not required for setup and serve only to deepen the knowledge of the inter­ested user and enable him to read in the sensor measured values into the processing system without an adjustment tool.
Slot Index Parameter Object Read Writ e Type Size Reset value Store Unit
0 16 Block object PB Yes No DS-32 20 C m 0 17 St rev PB Yes No Uns igned16 2 0 N m 0 18 Tag desc PB Yes Y es Octet St ring 32 32 * ' ' S m 0 19 Strategy PB Yes Yes Unsigned16 2 0 S m 0 20 Alert key PB Yes Yes Unsigned8 1 0 S m
0 21 Target mode PB Y es Y es (*1) Unsigned8 1 S m 0 22 Mode blk PB Yes N o DS-37 3 (D) m
0 23 Alarm s um PB Y es No DS-42 8 D m 0 24 Soft ware revision PB Y es No OctetString 1 6 C m 0 25 Hardware revision PB Y es No Octet String 1 6 C m 0 26 Device man ID PB Y es No Uns igned16 2 C m 0 27 Device ID PB Y es No OctetString 1 6 C m 0 28 Device ser num PB Yes No Octet St ring 16 C m 0 29 Di agnosis PB Yes No Oct et St r ing 4 - D m 0 31 Diagnosis mask PB Y es No OctetSt ring 4 C m
0 33 Device certification PB Yes No Octet St ring 32 C o 0 34 Write locking PB Y es Y es Unsigned1 6 2 - N o
0 35 Fact ory reset PB Yes Y es Unsigned1 6 2 - S o 0 36 Decript or PB Yes Yes OctetS t ring 32 - S o 0 37 Device message PB Yes Y es OctetSt ring 32 - S o
Optional
manda-
tory
VEGASON 54P … 56P 51
PA function diagram and parameters
Slot Index Parameter Object Read Writ e Type Size Reset value Store Unit
0 38 Device install date PB Yes Yes OctetString 16 - S o 0 40 Ident number s elect PB Yes Yes Unsigned8 1 - S m
0 41 HW writ e prot ect ion PB Yes N o Unsigned8 1 D o 0 49 VVO PB Yes Yes OctetString 32 o
0 50 View object PB PB Yes No Octet String 17 D m
Direct ory
10
object header Composite list
11
directory ent ries 1 16 Block object FB_Primary_Value Yes No DS-32 20 C m 1 17 St rev FB_Primary_V alue Yes No Uns igned16 2 0 N m 1 18 Tag des c FB_Primary_Value Yes Y es OctetString 32 32 * ' ' S m 1 19 St rategy FB_Primary _Value Yes Yes Unsigned16 2 0 S m 1 20 Alert key FB_Primary_Value Yes Y es Unsigned8 1 0 S m 1 21 Target mode F B_Primary_V alue Yes Y es Unsigned8 18 S m 1 22 Mode blk FB_Primary_Value Yes No DS-37 3 D m 1 23 Alarm s um FB_ Prima ry_V alu e Yes No D S-42 8 D m 1 24 Batch FB_Primary_Value Yes Yes DS-67 10 0,0,0,0 S m
1 26 Out FB_Primary_Value Yes Yes (*1) DS-33 5 D out scale unit m
1 27 PV scale FB_Primary_Value Yes Y es 2 * Float 8 0,100 S
1 28 Out scale FB_Primary_V alue Yes Y es DS-36 1 1 0,100,-,- S
1 29 Lin type FB_Primary_Value Yes Yes (*1 ) Unsigned8 1 0 S m 1 30 Channel FB_Primary_Value Yes Yes Unsigned16 2 - S m
1 32 PV FTime FB_Primary_Value Yes Y es Float 4 0 N sec m 1 33 Fsafe t ype FB_Primary_Value Yes Y es Unsigned8 1 1 S o 1 34 Fsafe value FB_Primary_V alue Yes Y es Float 4 - S out scale unit o 1 35 Alarm hys FB_Primary_V alue Yes Y es Float 4 0.5% of range S out scale unit m 1 37 HI HI Lim FB_Primary_Value Yes Yes Float 4 max value S out scale unit m 1 39 HI Lim FB_Primary_V alue Yes Yes Float 4 max value S out scale unit m 1 41 LO Lim FB_Primary_V alue Yes Y es Float 4 min value S out scale unit m 1 43 LO LO Lim FB_Primary_V alue Yes Y es Float 4 min value S out scale unit m 1 46 HI HI Alm FB_Primary_Value Yes No DS-39 1 6 0 D o 1 47 HI Alm FB_ Primary_V alue Yes No DS-39 16 0 D o 1 48 LO Alm FB_Prim ary_V alu e Y es No DS-39 16 0 D o 1 49 LO LO Alm FB_Primary_Value Yes No DS-39 1 6 0 D o
1 50 Simulate FB_Pr imary_Value Yes Yes DS-50 6 disable N
View object 161
FB_Primary_Value
Dis p lay s o urce 1180
selec t
DM Yes No
DM Yes No
FB_Primary_Value Yes No Octet St ring 18 D m
TB Yes Yes Unsigned8 1 ? N o
Arra y of Unsigned16 Arra y of Unsigned16
12 C m
24 C m
prim, s ec1 , sec2 value unit (dep. on channel) includes out s cale unit
prim, s e c1 , s ec2 value unit (dep. on channel)
Optional
manda-
tory
m
m
m
52 VEGASON 54P … 56P
PA function diagram and parameters
Slot Index Parameter Object Read Writ e Type Size Reset value Store Unit
1 120 Block object TB_Level Yes No DS-32 20 C m 1 1 21St rev TB_Level Yes N o Unsigned16 2 0 N m 1 122 Tag desc TB_Level Y es Y es OctetString 32 32 * ' ' S m 1 123 S t rategy TB_Level Yes Yes Unsigned16 2 0 S m 1 124 Alert key TB_Level Yes Y es Unsigned8 1 0 S m
1 125 Target mode T B_Level Yes Yes (*1) Unsigned8 1 S m 1 126 Mode blk TB_Level Yes No DS-37 3 (D) m
1 127 Alarm sum TB_Level Yes N o DS-42 8 D m 1 128 Primary value TB_Level Yes No DS-33 5 D primary value unit m 1 129 Primary value unit T B_Level Yes Yes Unsigned16 2 % S m 1 130 Level TB_Level Y es No Float 4 D level unit m 1 1 31Level unit TB_Level Yes Y es Unsigned1 6 2 % S m 1 132 Sens or value TB_Level Yes No Float 4 D sens or unit m 1 133 Sens or unit TB_Level Y es Y es Unsigned1 6 2 S m
Secondary
1134
value 1 Secondary
1135
value 1 unit
1 136 Secondary value 2 TB_Level Yes N o DS-33 5 D
Secondary
1137
value 2 unit
1 138 Sens or of fset TB_Level Yes Y es Float 4 0 S sens or unit m 1 139 Cal type TB_Level Yes Y es Unsigned8 1 1 140 Cal point lo TB_Level Yes Yes Float 4 S [sensor unit] m
1 1 41Cal point hi TB_Level Yes Y es Float 4 S sensor unit m 1 142 Level lo TB_Level Y es Y es Float 4 0 S level unit m 1 143 Level hi TB_Level Y es Y es Float 4 100 S level unit m 1 144 Level offset TB_Level Yes Y es Float 4 0 S level unit m 1 145 Lin type TB_Level Yes Y es Unsigned8 1 0 (=linear) S m 1 148 Sens or high limit TB_Level Yes No Float 4 C sensor unit o 1 149 Sens or low limit TB_Level Yes No Float 4 C sensor unit o
1 150 Max sens or value TB_Level Yes Yes Float 4
1 1 51Min sensor value TB_Level Yes Y es Float 4 1 152 Temperature TB_Level Yes N o Float 4 D temperature unit o
1 153 Temperature unit TB_Level Yes Y es Unsigned1 6 2 °C S o 1 154 Max temperature TB_Level Yes Y es Float 4
1 155 Min temperature TB_Level Yes Y es Float 4 1 156 Tab index TB_Level Yes Y es Unsigned8 11 D o
TB_Level Yes No DS-33 5 D
TB_Level Yes Yes Unsigned16 2 S o
TB_Level Yes Yes Unsigned16 2 S o
oder Lifeabgleic h
oder akt. Wert oder akt. Wert
oder akt. Wert oder akt. Wert
secondary value 1 unit
secondary value 2 unit
Sm
N sens or unit o
N sens or unit o
N temperature unit o
N temperature unit o
Optional
manda-
tory
o
o
VEGASON 54P … 56P 53
PA function diagram and parameters
Slot Index Parameter Object Read Writ e Type Size Reset value Store Unit
1 157 Tab X Y value T B_Level Yes Yes 2 * Float 8 N
1 158 Tab min number TB_Level Yes No Uns igned8 1 C o 1 159 Tab max number TB_Level Yes No Uns igned8 1 C o 1 160 Tab op code TB_Level Yes Yes Unsigned8 1 D o 1 1 61Tab s tatus TB_Level Yes No Uns igned8 1 D o 1 162 Tab act ual number TB_Level Yes No Unsigned8 1 D o
Simulate sensor 1190
value 1 1 91Simulat e level TB_Level Yes Yes DS-50 6 disable N level unit o 1 192 Füllgutt yp TB_Level Yes Yes Unsigned8 1 N o 1 193 Echo quality TB_Level Yes No Uns igned8 1 D dB o 1 194 First echo f act or TB_Level Y es Y es Unsigned8 1 aus N o
1 195 Sound velocity TB_Level Yes Yes Float 4 1 196 Lerndistanz TB_Level Yes Y es Float 4 D sensor unit o
Störechospeicher ­1197
Akt ion
Arbeitsberei ch 1198
Ende
Arbeitsberei ch 1199
Anfang
schnelle 1 200
Meßwertänderung
unruhige 1 201
Füllgutoberfläche 1 202 Schaumbil dung TB_Level Yes Yes Unsigned8 1 N o
star ke 1 203
Staubentwicklung
Bodenref lexion; 1 204
großer
Schüttwinkel
Korrekturfaktor 1210
Ausbr.geschw.
Korrekturfakt or der
Ausbr.geschw. per 1211
Distanzeingabe
berec hnen
2 16 Block object FB_Temperature Yes No DS-32 20 C m 2 17 St rev FB_Temperature Yes No Uns igned16 2 0 N m 2 18 Tag desc FB_Temperature Yes Y es OctetString 32 32 * ' ' S m 2 19 S t rategy FB_Temperat ure Yes Yes Unsigned16 2 0 S m 2 20 Alert key FB_Temperature Yes Y es Unsigned8 10 S m 2 21 Target mode F B_Temperature Yes Y es Unsigned8 18 S m
TB_Level Y es Y es DS-50 6 disable N sens or unit o
ca. 331.6 (VEGA-Wer t )
TB_Level Yes Yes Unsigned8 1 D o
TB_Level Y es Y es Float 4 N sensor unit o
TB_Level Y es Y es Float 4 N sensor unit o
TB_Level Yes Yes Unsigned8 1 N o
TB_Level Yes Yes Unsigned8 1 N o
TB_Level Yes Yes Unsigned8 1 N o
TB_Level Yes Yes Unsigned8 1 N o
TB_Level Yes Yes Float 4 N % o
TB_Level No Y es Float 4 N sensor unit o
secondary value 1 unit, primary value unit
Nm/s o
Optional
manda-
tory
o
54 VEGASON 54P … 56P
PA function diagram and parameters
Slot Index Parameter Object Read Writ e Type Size Reset value Store Unit
2 22 Mode blk FB_Temperature Yes No DS-37 3 D m 2 23 Alarm sum FB_Temperat ure Yes N o DS-42 8 D m 2 24 B at ch FB_Temperat ure Yes Yes DS-67 10 0,0,0,0 S m
2 26 Out FB_Temperat ure Yes Y es (*1) DS-33 5 D 2 27 PV scale FB_Temperature Yes Yes 2 * Float 8 0,100 S temperature unit m 2 28 Out scale FB_Temperature Yes Y es DS-36 110,1 00,-,- S
2 29 Lin type FB_Temperat ure Yes Yes (*1 ) Unsigned8 10 S m
2 30 Channel FB_Temperature Yes Yes (*1 ) Unsigned16 2 - S m 2 32 PV FT ime FB_Temperature Yes Yes Float 4 0 N sec m
2 33 Fsafe t ype FB_Temperature Yes Yes Unsigned8 1 1 S o 2 34 Fsafe value FB_Temperat ure Yes Yes Float 4 - S
2 35 Alarm hys FB_Temperat ure Yes Yes Float 4 1K S
2 37 HI HI Lim FB_Temperat ure Yes Yes Float 4 max value S
2 39 HI Lim FB_Temperat ure Yes Yes Float 4 max value S
2 41 LO Lim F B_Temperature Yes Yes Float 4 min value S
2 43 LO LO Lim FB_Temperature Yes Yes Float 4 min value S 2 46 HI HI Alm FB_Temperature Yes No DS-39 1 6 0 D o
2 47 HI Alm FB_Temperature Yes No DS-39 16 0 D o 2 48 LO Alm FB_Temperat ure Yes No DS-39 16 0 D o 2 49 LO LO Alm FB_Temperat ure Yes No DS-39 1 6 0 D o 2 50 Simulate FB_Temperature Yes Yes DS-50 6 disable N temperature unit m
View object
261
FB_Temperat ure
FB_Temperature Yes No Oc t etString 18 D m
temp out s cale unit
includes temp out s cale unit
temp out s cale unit temp out s cale unit temp out s cale unit temp out s cale unit temp out s cale unit temp out s cale unit
Optional
manda-
tory
m
m
o
m
m
m
m
m
VEGASON 54P … 56P 55

6 Diagnosis

Diagnosis

6.1 Simulation

To simulate a certain filling, you can call up the function "Simulation" on the adjustment module MINICOM or in the software program VVO.
With this function, you simulate a real vessel filling level. Please note that connected instru­ments, such as e.g. a PLC, react according to their adjustments and will probably acti­vate alarms or system functions.
Simulation with VVO
If you start the simulation mode with the ad­justment program VVO on the PC, the simu­lated level is outputted until you quit the simulation mode.
Simulation with MINICOM
If you start the simulation mode on the adjust­ment module MINICOM, the sensor returns to standard operating mode after one hour.

6.2 Error codes

Error codes Corrective measure
E013 No valid measured value Message is displayed during the warm-up
- Sensor in the warm-up phase phase
- Loss of the useful echo If the message remains, a false echo storage (with the adjustment software on the PC - see "Echo curve" under "Sensor optimisation") must be carried out together with a modification of mounting location and orientation to achieve the lowest possible false echo background.
E017 Adjustment span too small Carry out a readjustment.
E036 Sensor software does not run Sensor requires a software update (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tive Transducer defective.
56 VEGASON 54P … 56P
Make sure that the difference between min. and max. adjustment is at least 10 mm .
Message appears during a software update.
Contact our service department.
Technical data
,0
,0
,0
,0
,0
,0
,0
,0
,0
35689101315
,0
,
,
,
,0
,0
,0
,0
35689101315
g

7 Technical data

7.1 Data

Power supply
Supply voltage 9 … 32 V DC
Note:Note:
Note:
Note:Note: If VEGASON 54P … 56P sensors are operated on Profibus input cards (EP cards) in the VEGALOG 571 processing system each card must be powered individually with own power supply unit. For power supply of an EP card, use the power supply unit (ASI­Interface SLA 3.100) from Puls which is available from VEGA.
Current consumption max. 81 mA, depending on the output voltage
VEGAPULS 54P, 55P VEGAPULS 56P
mA
100
90
81
80
70
60
50 40
35
30
5
0
13,5
15
10
U
= 9 V
min
25
20
Power consumption
- VEGASON 54 and 55 max. 0.9 W
- VEGASON 56 max. 1.1 W
Resistance of the signal cable depending on the output voltage UO (PA voltage)
Number of sensors on a two-wire cable depending on the output voltage UO of the
UO (PA voltage) of the segment coupler used, see diagrams.
mA
100
90
81
80
70
60
50 40
35
= 32 V
30
V
30
U
max
5
0
13,5
15
10
U
= 9 V
min
20
of the segment coupler used
segment coupler used and the voltage losses of the signal cable, see cable specification. Number of sensors when being operated on an EP card in a VEGALOG, see diagrams:
25
30
V
U
= 32 V
max
VEGAPULS 54P, 55P on EP card VEGAPULS 56P on EP card
2000
1800
1600
1400
1200
1000,0
800
600
400
Cable length (m)
200
0,0
No. of sensors VEGAPULS 54P, 55 P
0,8 mm
2000,0
1800
1600
1400
1200,0
1000
800
0
600
0
0
400
Cable length (m)
200,0
0
No. of sensors VEGAPULS 56P
0,8 mm
VEGASON 54P … 56P 57
Technical data
Measuring range
(with sensor version B … D, the transducer end is the reference plane. With sensor version A the lower side of the flange is the reference plane)
VEGASON 54 1.0 … 25 m (solids 1.0 … 15 m) VEGASON 55 0.8 … 45 m (solids 0.8 … 30 m) VEGASON 56
- Version A 1.8 … 70 m (solids 1.8 … 45 m)
- Version B … D 1.4 … 70 m (solids 1.8 … 45 m)
Output signal
digital (Profibus) the digital output signal (meas. signal) is
modulated onto the power supply and further processed in the PLC or process control system.
Integration time 0 … 999 seconds
Measured value display (optional)
Liquid-crystal indication
- in the sensor scalable output of meas. value as graph and number value.
- powered externally by the sensor scalable output of meas. value as graph and number value. Measured value display can be mounted up to 25 m away from the sensor.
Adjustment
- adjustment software VEGA Visual Operating on Master-Class 2 PC
- adjustment module MINICOM in the sensor or in external indicating instrument (optional)
- process adjustment interface PACTwareTM (VVO runs as a subprogram)
- SIMATIC PDM in conjunction with Electronic Device Description (EDD)
Process fitting
VEGASON 54 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200 VEGASON 55 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 250 VEGASON 56 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200
Note: For G 1 A, 1-11.5 NPT, DN 50 and DN 80, an adapter flange is normally required.
Accuracy
1)
(typical values under reference conditions, all statements relate to the nominal measuring range)
Characteristics linear Deviation in characteristics incl.
linearity, reproducibility and hysteresis (determined acc. to the
limit point method) < 0.1 % Linearity better than 0.05 % Average temperature coefficient of the
zero signal 0.06 %/10 K General resolution max. 1 mm Resolution of the output signal 0.01 % or 1 mm
58 VEGASON 54P … 56P
Technical data
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring
range)
Min. span
(between empty and full adjustment) > 20 mm (recommended > 50 mm)
Ultrasonic frequency (at 20°C)
- VEGASON 54 30 kHz
- VEGASON 55 18 kHz
- VEGASON 56 10 kHz
Intervals
- VEGASON 54 1.0 s
- VEGASON 55 1.5 s
- VEGASON 56 2.0 s
Beam angle at -3 dB emitted power
- VEGASON 54
- VEGASON 55
- VEGASON 56
Influence of the process temperature negligible, is compensated by a dynamic
temperature detection system integrated in the transducer. Without temperature compensation,
the ambient temperature error will be 0.18 %/K Influence of the process pressure negligible within the approved sensor pressures Adjustment time
2)
- VEGASON 54 > 2 s (depending on parameter setting)
- VEGASON 55, 56 > 4 s (depending on parameter setting)
Ex technical data
Classification m (encapsulation of the transducer) Temperature class (permissible ambient
temperature on the transducer when used in Ex areas)
- T6 42°C
- T5 58°C
- T4 60°C
- T3 60°C
Ex approved in category or Zone
- VEGASON 56 ATEX Zone 1 (II 2G) IEC, CENELEC, PTB Zone 1
- VEGASON 54 … 56 ATEX Zone 21/22 (II 2D/3D) IEC, CENELEC, DMT Zone 10/11
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770; temperature 15°C … 35°C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar.
2)
The adjustment time is the time required by the sensor to output a correct level (with max. 10 % deviation) after a sudden level change.
VEGASON 54P … 56P 59
Ambient conditions
Ambient temperature (housing) -20°C … +60°C Process temperature (transducer)
- VEGASON 54, 55 -40°C … +80°C (StEx: -20°C … +75°C)
- VEGASON 56 -40°C … +150°C
- storage and transport temperature -40°C … +80°C
Vessel pressure max. (gauge pressure)
- VEGASON 54 Version A 1.5 bar (flange version) Version B … C 0.5 bar Version D 3 bar
- VEGASON 55 Version A 1.5 bar (flange version) Version B … C 0.5 bar Version D 3 bar
- VEGASON 56 Version A 3 bar (flange version) Version B … C 0.5 bar Version D 3 bar
Protection
- sensor IP 67
- transducer, process IP 68
Protection class
- two-wire sensor II
- four-wire sensor I Overvoltage category III Self-heating
at 40°C ambient temperature
- sensor 45°C
- transducer, process 55°C
Technical data
Connection cables
Power supply power supply and signal via one two-wire cable
from the segment coupler
Electrical connection spring-loaded terminals, cross-section area of
conductor generally 2.5 mm
- option screwed plug connection Ground connection max. 4 mm
2
2
Transducer cable
- VEGASON 54, 55 5 … 300 m (cable diameter 7.2 … 7.6 mm)
- VEGASON 56 5 … 300 m (cable diameter 9.5 … 9.9 mm) Cable entry for the signal and supply
cable with spring-loaded terminals
- plastic housing (PBT) 2 x M20 x 1.5 (cable diameter 5 … 9 mm)
or 2 x 1/2" NPT (cable diameter
3.1 … 8.7 mm or 0.12 … 0.34 inch)
- Aluminium housing and
Ex d terminal compartment 2 x 1/2" NPT (cable diameter 3.1 … 8.7 mm
or 0.12 … 0.34 inch)
60 VEGASON 54P … 56P
Technical data
Materials
Housing PBT (Valox) or Aluminium Process connection
- flange version Alu or PP
- swivelling holder and thread galvanized steel
Transducer
- VEGASON 54 PA (1.4301 with StEx)
- VEGASON 55, 56 UP
Transducer diaphragm
- VEGASON 54 1.4571
- VEGASON 55 Alu/PE foam
- VEGASON 56 Alu/PTFE non-stick coating
Transducer cable (cable cover)
- VEGASON 54, 55 PUR (1.1082)
- VEGASON 56 silicone (1.1083)
Weight (depending on the housing materials used)
VEGASON 54
- Version A 5.6 … 10.7 kg
- Version B 6.9 … 9.7 kg
- Version C 7.5 … 10.5 kg
- Version D 4.7 … 6.9 kg
VEGASON 55
- Version A 8.0 … 13.3 kg
- Version B 8.7 … 10.3 kg
- Version C 9.2 … 11.1 kg
- Version D 6.5 … 7.5 kg
VEGASON 56
- Version A 7.3 … 11.3 kg
- Version B 8.7 … 10.3 kg
- Version C 9.3 … 11.1 kg
- Version D 6.5 … 7.5 kg
CE conformity
VEGASON series 50 ultrasonic sensors meet the protective regulations of EMC (89/336/EWG), NSR (73/23/EWG) and R & TTE directive (1999/5/EC). Conformity has been judged acc. to the following standardst:
EN 300 683 - 1: 1997 EN 300 440 - 1: 1995 I-ETS 300-440 Expert opinion No. 0043052-01/SEE, Notified
Body No. 0499 EMC Emission/Susceptibility EN 61 326: 1997/A1: 1998 ATEX EN 50 020: 1994
EN 50 018: 1994
EN 50 014: 1997 NSR EN 61 010 - 1: 1993
VEGASON 54P … 56P 61
Technical data

7.2 Data format of the output signal

Byte4 Byte3 Byte2 Byte1 Byte0
Status Measured value (IEEE-754 Format, see below)
Status byte:
The status byte corresponds to profile 3,0 "Profibus PA Profile for Process Control Devices" coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit 6 … 0 = 0).
Measured value:
The measured value is transmitted as a 32 Bit floating point number in IEEE-754 format.
Byte n Byte n+1
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
7654321076543210
VZ 27262524232221202-12-22-32-42-52-62
Sign Exponent Mantissa
Byte n+2 Byte n+3
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
7654321076543210
2-82-92
-102-112-122-132-142-152-162-172-182-192-202-212-222-23
-7
Mantissa Mantissa
Formula: Measured value = (-1)VZ • 2
Examples: 41 70 00 00 (hex) = 0100 0001 0111 0000 0000 0000 0000 0000 (bin)
Meas. value = (-1)0 • 2
(Exponent - 127)
(130 - 127)
• (1 + 2-1 + 2-2 + 2-3)
• (1 + Mantissa)
= 1 • 23 • (1 + 0.5 + 0.25 + 0.125) = 1 • 8 • 1.875 = 15.0
62 VEGASON 54P … 56P
Technical data

7.3 Approvals

When using ultrasonic sensors in Ex areas or on ships, the instruments must be suitable and approved for the explosion zones and applications. The suitability is checked by the approval authorities and is certified in approval docu­ments.
Please note the attached approval docu­ments when using a sensor in Ex area.

7.4 Dimensions

External indicating instrument VEGADIS 50
38
ø5
48
10
135
118
108
82
Test and approval authorities
- PTB
(Physikalisch Technische Bundesanstalt ­Physical Technical Approval Authority)
- FM
(Factory Mutual Research)
- ABS
(American Bureau of Shipping)
- LRS
(Lloyds Register of Shipping)
- GL
(German Lloyd)
- CSA
(Canadian Standards Association)
85
Pg 13,5
Mounting on carrier rail 35 x 7.5 acc. to EN 50 022 or screwed to a flat surface
Note:
The cable diameter of the connection cable should be min. 5 mm and max. 9 mm. Otherwise, the seal effect of the cable entry would not be ensured.
Adjustment module MINICOM
Tank 1 m (d)
12.345
67,5
VEGASON 54P … 56P 63
ESC
+
-
32,5
OK
Adjustment module for insertion into VEGASON series 50 sensors or into the exter­nal indicating instrument VEGADIS 50
74
VEGASON 54 … 56 in Version A
201
165
Plastic housing (PBT)
10˚
101
Aluminium housing (Al)
25
116
Technical data
215
185
VEGASON 54
VEGASON 55
397
Min. distance to the product
Min. distance to the product
257
1,0 m
0,8 m
90
12xø22
1)
ø190 (ø196)
ø340
ø244 ø405
75
1)
20
110 (126)
12xø26
20
128
12xø22 (12xø26)
445,8
282
Reference plane
Reference plane
20
Reference plane
VEGASON 56
423
Min. distance to the product
1,8 m
64 VEGASON 54P … 56P
ø198
2)
ø340 (ø405)
Technical data
VEGASON 54 … 56 in version B
486
4xø19
172
386
245
ø165
ø122,8
201
165
10˚
Plug connection
ø 27
> ø200
Plastic housing (PBT)
11,5
11,5
101
90
503
4xø19
ø165
ø122,8
>ø250
435
270
215
185
65
Plug connection
798
Aluminium housing (Al)
25
116
ø165
ø122,8
4xø19
>ø210
ø190 (ø196)
VEGASON 54
1,0 m
Reference plane
189,5
ø 244
VEGASON 55
484,5
0,8 m
ø 198
1,4 m
VEGASON 56
VEGASON 54P … 56P 65
VEGASON 54 … 56 in version C
Aluminium housing (Al)
Plug
Technical data
215
25
116
185
Plug connection
ø 45
78
68
7
130 150
ø 7
85
65
170
445,8
282
10120
1,0 m
0,8 m
VEGASON 54
VEGASON 55
1,4 m
Reference plane
VEGASON 56
66 VEGASON 54P … 56P
Technical data
VEGASON 54 … 56 in version D
Plug connection
3240
214
149
Plastic housing (PBT)
Plug
68
ø 45
78
233
7
3240
167,5
186
101
90
130 150
201
165
10˚
397,2
257,2
ø 7
85
65
120 10
170
3240
527
462,5
Reference plane
1,0 m
0,8 m
1,4 m
VEGASON 54 VEGASON 55 VEGASON 56
VEGASON 54P … 56P 67
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (07836) 50-0 Fax (07836) 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing sys­tems correspond to the latest information at the time of printing.
Technical data subject to alterations
2.24 967 / 02-03-19 / EN
Loading...