VEGA SON54K User Manual

Page 1
Operating Instructions
VEGASON 54K … 56K
Page 2
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description .................................................................. 4
1.2 Application features ............................................................. 5
1.3 Adjustment ............................................................................ 5
2 Types and v ersions ................................................................... 7
2.1 Type survey .......................................................................... 7
2.2 Type code ............................................................................. 8
2.3 Approvals ............................................................................. 8
2.4 Configuration of measuring systems ................................. 8
3 Technic al data .......................................................................... 12
3.1 Data ..................................................................................... 12
3.2 Dimensions ......................................................................... 17
Contents

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules.
For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGASON 54K … 56K

Note Ex area

Please note the approval documents (yellow binder), and especially the included safety data sheet.
Page 3
Contents
4 Mo unting and in stalla tion ..................................................... 22
4.1 Mounting ............................................................................. 22
4.2 General installation instructions ........................................ 24
4.3 Measurement of liquids ..................................................... 26
4.4 Measurement of solids ...................................................... 28
4.5 Socket extensions ............................................................. 29
4.6 False echoes ...................................................................... 30
4.7 Incorrect mounting ............................................................. 32
5 Electrical connection .............................................................. 35
5.1 Connection and connection cable .................................... 35
5.2 Connection of the sensor .................................................. 36
5.3 Connection of the external indicating instrument
VEGADIS 50 ....................................................................... 37
6 Set-up ........................................................................................ 38
6.1 Adjustment methods .......................................................... 38
6.2 Adjustment with the PC ...................................................... 38
6.3 Adjustment with adjustment module MINICOM ............... 40
7 Diagnosis .................................................................................. 46
7.1 Simulation ............................................................................ 46
7.2 Error codes ........................................................................ 46
VEGASON 54K … 56K 3
Page 4

1 Product description

1.1 Function

Continuous level measurement with ultrasonic sensors is based on the running time meas­urement of ultrasonic pulses.
VEGASON series 54 … 56 sensors are a newly developed generation of extremely compact ultrasonic sensors for level meas­urement. They were developed for liquids and especially for solids and larger measur­ing ranges. Preferred areas of application are silos, bunkers or material heaps.
Due to the small housing dimensions and process fittings, the compact sensors are a very reasonable solution for your level meas­urement applications. With the integrated display and a special sensor intelligence, in conjunction with large measuring ranges, they can be used for applications in which the advantages of a non-contact measure­ment could never before be realized.
The instruments produce an analogue 4 … 20 mA output signal.
Measuring principle
Piezoceramic high performance transducers emit focused ultrasonic pulses which are reflected by the product surface. The meas­urement electronics prepares a precise im­age of the environment from the reflected ultrasonic pulses. The transducers work both as transmitter and receiver. As receiver, the transducers are high-sensitivity piezo micro­phones.
Product description
Meas. dis­tance
emission - reflection - reception
The measurement electronics precisely cal­culates the distance between transducer and medium from the speed of sound and the measured running time of the emitted sound pulses. The distance is then converted into a level-proportional signal and made available acc. to sensor settings as a precise, cali­brated level value.
Since the speed of sound is subject to tem­perature influence, the transducer also con­tinuously detects the ambient temperature, so that the level is precisely measured even if the ambient temperature varies.
Output signal
The level-proportional measurement signal is made available as 4 … 20 mA output signal. The analogue 4 … 20 mA signal is freely programmable and reflects the measuring range, i.e. the adjusted operating range, of the sensor.
4 VEGASON 54K … 56K
Page 5
Product description
Display of measured values
As an option, the series 50 ultrasonic sensors can be equipped with an indicating instru­ment for direct, local level survey. The indi­cating instrument shows the precise level by means of the analogue bar graph and the digital number value. In addition to the indica­tion in the sensor, you can have the level displayed with the VEGADIS 50 external indicating instrument at a distance of up to 25 m from the sensor. The external display of measured values operates, like the inte­grated display, independently of the 4 … 20 mA output signal, can be modified through individual parameter settings, and is powered by the sensor.

1.2 Application features

Voltage supply
• 24 V DC four-wire technology (supply and signal separate)
Applications
• Level measurement of solids, liquids and pastes.
• Gauge measurement, distance measure­ment, object monitoring and conveyor belt monitoring
Display of measured value
• Display module integrated in sensor
• Optional display module separated from sensor
• Adjustment module MINICOM in the sen­sor.
Process fittings
• G 1 A, DN 50, DN 80, DN 200, DN 250
Approvals
• CENELEC, ATEX, PTB

1.3 Adjustment

Each measuring situation is unique. For that reason, every ultrasonic sensor needs some basic information on the application and the environment, e.g. a false echo storage is important for a reliable measurement. Beside this, many other settings and adjustments are possible on VEGASON ultrasonic sen­sors.
The adjustment and parameter setting of ultrasonic sensors are carried out with
- the PC
- the detachable adjustment module MINICOM.
Signal output
• 4 … 20 mA
Rugged and precise
• Measurement unaffected by substance properties such as density, conductivity, dielectric constant…
• Suitable for corrosive substances
• Measuring range 0.8 m … 70 m
Means of adjustment
• With adjustment software VEGA Visual Operating (VVO) on the PC
• With detachable adjustment module MINICOM
• With the HART® handheld
VEGASON 54K … 56K 5
Adjustment with PC
The setup and adjustment of the ultrasonic sensors is generally done on the PC with the adjustment software PACT dows®. The program leads quickly through the adjustment and parameter setting by means of pictures, graphics and process visualisations.
The PC can be connected to the sensor or to any individual position of the system or the signal cable. It is connected by means of the two-wire PC interface converter VEGACONNECT 2 to the sensor or the signal cable. The adjustment and parameter data can be saved with the adjustment software on the PC and can be protected by pass­words. On request, the adjustments can be quickly transferred to other sensors.
ware
TM
under Win-
Page 6
-+
2
4 … 20 mA
2
Adjustment with the PC on the analogue 4 … 20 mA signal and supply cable or directly on the sensor (four-wire sensor)
Adjustment with adjustment module MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad­justment module with display, the adjustment can be carried out in clear text dialogue. The adjustment module can be plugged into the ultrasonic sensor or into the optional, external indicating instrument.
Tank 1 m (d)
12.345
ESC
+
-
OK
Product description
ESC
+
-
Tank 1 m (d)
12.345
OK
2
Tank 1 m (d)
12.345
+
-
4 ... 20 mA
ESC
OK
4
Adjustment with detachable adjustment module. The adjustment module can be plugged into the ultrasonic sensor or the external indicating instrument VEGADIS
50.
Adjustment with HART® handheld
Beside the PC and the adjustment module MINICOM, it is also possible to adjust the sensors with the HART® handheld. A special DDD (Data Device Description) is not neces­sary, so that the sensors can be adjusted with the HART® standard menus of the handheld.
To adjust the sensor, just connect the HART handheld to the 4 … 20 mA output signal cable or insert the two communication cables of the HART® handheld into the adjustment jacks on the sensor.
®
20 ... 72 V DC
2
4 ... 20 mA
2
Detachable adjustment module MINICOM
HART® handheld
Unauthorised sensor adjustments can be prevented by removing the adjustment mod­ule.
6 VEGASON 54K … 56K
HART® handheld on the 4 … 20 mA signal cable
Page 7
Types and versions

2 Types and versions

2.1 Type survey

VEGASON series 54 … 56 sensors are a newly developed generation of extremely compact ultrasonic sensors for large measur­ing ranges. VEGASON 51 … 53 are used for shorter measuring ranges.
Due to the small housing dimensions and process fittings, the compact sensors do your level monitoring inconspicuously, and above all, at reasonable cost. Swivelling holders allow a quick alignment of the trans­ducer to the product and solid surface. Mounting is simplified by the option of sepa­rating the sensor electronics from the trans­ducer. The sensor electronics can be mounted at a distance of 300 m from the transducer. It is then possible to mount the transducer in environments with an ambient temperature up to 150°C (VEGASON 56).
Common features
• Application to solids and liquids
• Measuring range 0.8 m … 70 m
• Ex approved in Zone 10 (IEC) and Zone 11 (ATEX) classification EEx ia [ia] IIC T6 or Zone 20/21/22, VEGASON 56 also Zone 1.
• Display module integrated in the sensor or in the external indicating instrument sepa­rated up to 25 m from the sensor
VEGASON 54
Version A Version B Version C Version D
VEGASON 55
Version A Version B Version C Version D
VEGASON 56
Version A Version B Version C Version D
VEGASON 54K … 56K 7
Page 8

2.2 Type code

Types and versions
VEGASON 54 P EX.XX X X X X X X X
K - Plastic housing PBT, M20 x 1.5 cable entry N - Plastic housing PBT, ½ NPT cable entry A - Aluminium housing, M20 x 1.5 cable entry
FEP - Version A, flange DN 200 (PP) FEA - Version A, flange DN 200 (Aluminium) FLP - Version A, flange 8“ (PP) FLA - Version A, flange 8“ (Aluminium) SAS - Version B, flange swivelling holder DN 50 SBS - Version B, flange swivelling holder DN 80 GAS - Version C, flange swivelling holder DN 50 GBS - Version C, flange swivelling holder DN 80 RGS - Thread G 1 A
X - without display A - with integrated display module
X - without adjustment module MINICOM B - with adjustment module MINICOM (mounted)
A - Four-wire 20…72 V DC; 20…253 V AC; 4…20 mA B - Four-wire 20…72 V DC; 20…253 V AC; 4…20 mA, HART G - Profibus PA
.X - without approval EXS.X ATEX II 1/2D IP68 TStEx Zone 10
K - Analogue 4 … 20 mA output signal (two-wire or
four-wire technology)
P - Profibus
Type 54 - Measuring range 1.0 … 25 m Type 55 - Measuring range 0.8 … 45 m Type 56 - Measuring range 1.6 … 70 m
Meas. principle (SON for ultrasonics)
®

2.3 Approvals

2.4 Configuration of measuring sys­tems
When using ultrasonic sensors in Ex areas or in marine applications, the instruments must be suitable and approved for the explosion zones and application areas. The suitability is tested by approval authori­ties and is certified in approval documents.
VEGASON 50 ultrasonic sensors are ap­proved for Ex zone 1, 10, 11, 21 and 22. Please note the attached approval docu­ments when using a sensor in Ex area.
8 VEGASON 54K … 56K
A measuring system consists of a sensor with a 4 … 20 mA signal output and a module that evaluates and further processes the level-proportional current signal.
On the following pages you will find instru­ment configurations designated as measur­ing systems, several of which are shown with signal conditioning units.
Page 9
Types and versions
Measuring system consisting of VEGASON 54K … 56K with any 4 … 20 mA processing unit
• Four-wire technology, supply and output signal via two separate two-wire cables
• Output signal 4 … 20 mA active
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m from the sensor)
• Adjustment with PC, HART® handheld or adjustment module MINICOM (pluggable in the sensor or in the external VEGADIS 50 indicating instrument)
• Max. resistance on the signal output (load) 500
VEGADIS 50
3
4
1)
2
200
2
2
VEGA­CONNECT 2
-+
4 … 20 mA
2)
(active)
HART® handheld
1)
If the resistance of the processing systems con­nected to the 4 … 20 mA signal output is less than 200 , a resistor must be connected to the con­nection cable during adjustment, to reach an ad­justment load of 200 to 350 Ω.
2)
4 … 20 mA active means, that the sensor delivers a level-dependent current of 4 … 20 mA (source). With respect to the processing system (e.g. dis­play), the sensor acts electrically as a current
source. The digital adjustment signal would otherwise be extremely damped or short circuited because of in­sufficient resistance of the connected processing system. Communication with the PC would no longer be ensured.
VEGASON 54K … 56K 9
Page 10
Types and versions
Measuring system consisting of VEGASON 54K … 56K with PLC
• Output signal 4 … 20 mA active
• Display module integrated into the sensor
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m from the sensor in Ex area)
• Adjustment with PC, HART® handheld or adjustment module MINICOM (pluggable in the sensor or in the external VEGADIS 50 indicating instrument)
VEGADIS 50
3
-+
1)
4
2
VEGA­CONNECT 2
4 … 20 mA
2)
active
2
2
2
PLC
3)
passive
HART® handheld
1)
If the resistance of the processing systems con­nected to the 4 … 20 mA signal output is less than 200 , a resistor must be connected to the con­nection cable during adjustment, to reach an ad­justment load of 200 to 350 Ω. The digital adjustment signal would otherwise be extremely damped or short circuited because of in­sufficient resistance of the connected processing system. Communication with the PC would no
2)
4 … 20 mA active means, that the sensor delivers a level-dependent current of 4 … 20 mA (source). With respect to the processing system (e.g. dis­play), the sensor acts electrically as a current source.
3)
4 … 20 mA passive means that the PLC consumes a level-dependent current of 4 … 20 mA. The PLC acts electrically as a resistor (consumer).
longer be ensured.
10 VEGASON 54K … 56K
Page 11
Types and versions
Measuring system consisting of VEGASON 54K … 56K and VEGADIS 371 Ex indicating instrument with current and relay output
• Output signal 4 … 20 mA active
• Display module integrated into the sensor
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m from the sensor)
• Adjustment with PC, HART® handheld or adjustment module MINICOM (pluggable in the sensor or in the external VEGADIS 50 indicating instrument)
StEx area
VEGADIS 50
4
Non Ex area
2
VEGA­CONNECT 2
3
2
4 … 20 mA
(active)
-+-+
3
2
VEGADIS 371 Ex
(see "3.2 Approvals“)
Relay 1 Relay 2 Relay 3
0/4 ... 20 mA
HART® handheld
VEGASON 54K … 56K 11
Page 12
Technical data

3 Technical data

3.1 Data

Power supply
Power supply 230 V AC (20 … 253 V AC), 50/60 Hz
Current consumption max. 160 mA Power consumption max. 2 W, 3.5 VA
Measuring range
(the transducer end is the reference plane for sensor version B … D. The lower edge of the flange is the reference plane for sensor version A)
VEGASON 54 1.0 … 25 m (solid: 1.0 … 15 m) VEGASON 55 0.8 … 45 m VEGASON 56
- version A 1.8 … 70 m
- version B … D 1.4 … 70 m
Output signal
Signal output 4 … 20 mA current signal in four-wire technology Load 500 Integration time 0 … 999 seconds (adjustable) Fault signal current signal unchanged 20.5 mA, 22 mA
24 V DC (20 … 72 V DC) fuse 0.315 A TR
(adjustable)
Four-wire technology: Separate power supply. The analogue 4 … 20 mA output signal (measuring signal) is car­ried on a cable separate from the power supply.
Display module (optional)
Liquid crystal display
- in the sensor scalable measured value output as graph and
- powered externally by the sensor scalable measured value output as graph and
Adjustment
- PC with adjustment software VEGA Visual Operating
- Adjustment module MINICOM
- HART® handheld
12 VEGASON 54K … 56K
numerical value
numerical value. The display module can be mounted up to 25 m away from the sensor.
Page 13
Technical data
Accuracy
1)
(typical values under reference conditions, all statements relate to the nominal measuring range)
Characteristics linear Deviation in characteristics including
linearity, reproducibility and hysteresis (determined by
limit point methods) < 0.1 % Linearity better than 0.05 % Average temperature coefficient of the
zero signal 0.06 %/10 K Resolution in general max. 1 mm Resolution of the output signal 0.01 % or 1 mm
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring range)
Min. measuring range
(between empty and full adjustment) > 20 mm (recommended > 50 mm) Ultrasonic frequency (at 20°C)
- VEGASON 54 30 kHz
- VEGASON 55 18 kHz
- VEGASON 56 10 kHz Measuring intervals
- VEGASON 54 1.0 s
- VEGASON 55 1.5 s
- VEGASON 56 2.0 s Beam angle at -3 dB emitted power
- VEGASON 54
- VEGASON 55
- VEGASON 56 6° Influence of the process temperature 1.8 %/10 K, however is compensated by an
integral, dynamic temperature detection system in the transducer
Influence of the process pressure negligible within the approved sensor pressure
Adjustment time
2)
range
- VEGASON 54 > 2 s (depending on parameter adjustment)
- VEGASON 55, 56 > 4 s (depending on parameter adjustment)
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770; temperature 15°C … 35°C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar.
2)
The adjustment time is the time, required by the sensor, to output the correct level after a quick level change (with max. 10 % deviation).
VEGASON 54K … 56K 13
Page 14
Ambient conditions
Ambient temperature (housing) -20°C … +60°C Process temperature (transducer)
- VEGASON 54, 55 -40°C … +80°C (StEx: -20°C … +75°C)
- VEGASON 56 -40°C … +150°C
- storage and transport temperature -40°C … +80°C
Vessel pressure max. (gauge pressure)
- VEGASON 54 Version A 0.5 bar (flange version) Version B … C 0.5 bar Version D 0.5 bar
- VEGASON 55 Version A 0.5 bar (flange version) Version B … C 0.5 bar Version D 0.5 bar
- VEGASON 56 Version A 0.5 bar (flange version) Version B … C 0.5 bar Version D 2 bar
Protection
- sensor IP 67
- transducer, process IP 68
Protection class
- two-wire sensor II
- four-wire sensor I Overvoltage category III Self-heating at 40°C ambient temperature
- sensor 45°C
- transducer, process 55°C
Technical data
Ex technical data
Classification m (encapsulation of the transducer) Temperature class (permissible ambient temperature around the transducer when used in Ex areas)
- T6 42°C
- T5 58°C
- T4 60°C
- T3 60°C Ex approved in category or zone
- VEGASON 56
ATEX Zone 1 (II 2G) IEC, CENELEC, PTB Zone 1
- VEGASON 54 … 56
ATEX Zone 21/22 (II 2D/3D) IEC, CENELEC, DMT Zone 10/11
14 VEGASON 54K … 56K
Page 15
Technical data
Process fittings
VEGASON 54 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200 VEGASON 55 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 250 VEGASON 56 G 1 A, 1-11.5 NPT, DN 50, DN 80, DN 200
Note: Generally an adapter flange is necessary for G 1 A, 1-11.5 NPT, DN 50 and DN 80.
Connection cables
Four-wire sensors supply and signal in two separate cables Electrical connection spring-loaded terminals, wire cross-section
generally 2.5 mm
- option screw connection Ground connection max. 4 mm
2
2
Transducer cable
- VEGASON 54, 55 5 … 300 m (cable diameter 7.2 … 7.6 mm)
- VEGASON 56 5 … 300 m (cable diameter 9.5 … 9.9 mm) Cable entry for signal and supply cable with spring-loaded terminals
- plastic housing (PBT) 2 x M20 x 1.5 (cable diameter 5 … 9 mm) or 2 x 1/2“ NPT (cable diameter
3.1 … 8.7 mm or 0.12 … 0.34 inch)
- Aluminium housing and
Ex d terminal compartment 2 x 1/2“ NPT (cable diameter 3.1 … 8.7 mm
or 0.12 … 0.34 inch)
Materials
Housing PBT (Valox) or aluminium Process fitting
- flange version Alu or PP
- swivelling holder and thread galvanised steel
Transducer
- VEGASON 54 PA (1.4301 at StEx)
- VEGASON 55, 56 UP
Transducer diaphragm
- VEGASON 54 1.4571
- VEGASON 55 Alu/PE foam
- VEGASON 56 Alu/PTFE nonadhesive coating
Transducer cable (cable cover)
- VEGASON 54, 55 PUR (1.1082)
- VEGASON 56 silicone (1.1083)
VEGASON 54K … 56K 15
Page 16
Technical data
Weight (depending on the housing materials used)
VEGASON 54
- Version A 5.6 … 10.7 kg
- Version B 6.9 … 9.7 kg
- Version C 7.5 … 10.5 kg
- Version D 4.7 … 6.9 kg
VEGASON 55
- Version A 8.0 … 13.3 kg
- Version B 8.7 … 10.3 kg
- Version C 9.2 … 11.1 kg
- Version D 6.5 … 7.5 kg
VEGASON 56
- Version A 7.3 … 11.3 kg
- Version B 8.7 … 10.3 kg
- Version C 9.3 … 11.1 kg
- Version D 6.5 … 7.5 kg
CE conformity
VEGASON series 50 ultrasonic sensors meet the protective regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1993
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
EN 61 326 - 1: 1997/A1:1998
16 VEGASON 54K … 56K
Page 17
Technical data

3.2 Dimensions

VEGASON 54 … 56 in version A
201
165
Plastic housing (PBT)
10˚
101
Aluminium housing (Al)
25
116
215
185
VEGASON 54
VEGASON 55
397
Min. distance to the product
Min. distance to the product
257
1,0 m
0,8 m
90
12xø22
1)
ø190 (ø196)
ø340
ø244 ø405
75
1)
20
110 (126)
12xø26
20
128
12xø22 (12xø26)
445,8
282
Reference plane
Reference plane
20
Reference plane
VEGASON 56
423
Min. distance to the product
1,8 m
VEGASON 54K … 56K 17
ø198
2)
ø340 (ø405)
Page 18
VEGASON 54 … 56 in version B
Technical data
6
8
4
4xø19
2
7
1
386
245
ø165
ø122,8
201
165
0
1
Plug connection
ø
2
> ø200
˚
7
11,5
11,5
Plastic housing (PBT)
101
90
ø122,8
4xø19
3
0
5
ø165
>ø250
435
270
215
185
65
Plug connection
8
9
7
Aluminium housing (Al)
25
116
ø165
ø122,8
4xø19
>ø210
,5
9
8
ø
1
9
0
VEGASON 54
1
9
6
)
m
,0
1
Reference plane
1
ø
2
4
4
m
,8
0
VEGASON 55
,5
4
8
4
ø
1
9
8
1,4 m
VEGASON 56
Note:
The swivelling holder is shown in DN 50.
18 VEGASON 54K … 56K
Page 19
Technical data
VEGASON 54 … 56 in version C
Aluminium housing (Al)
Plug
215
25
116
185
m
,0
1
Plug connection
ø 45
78
68
7
130 150
ø 7
85
65
170
445,8
282
10120
m
,8
0
VEGASON 54
VEGASON 55
m
,4
1
Reference plane
VEGASON 56
VEGASON 54K … 56K 19
Page 20
VEGASON 54 … 56 in version D
Technical data
Plug connection
3240
214
149
Plastic housing (PBT)
Plug
68
ø 45
78
233
7
3240
167,5
186
101
130 150
201
165
10˚
397,2
90
ø 7
85
65
120 10
170
257,2
3240
527
462,5
Reference plane
1,0 m
0,8 m
1,4 m
VEGASON 54 VEGASON 55 VEGASON 56
20 VEGASON 54K … 56K
Page 21
Technical data
External indicating instrument VEGADIS 50
38
ø5
48
10
Mounting on carrier rail 35 x 7.5 acc. to EN 50 022 or flat screwed
Pg 13,5
135
118
108
Adjustment module MINICOM
Tank 1 m (d)
12.345
67,5
ESC
+
-
74
32,5
OK
82
85
Note:
The diameter of the connection cable must be 5 … 9 mm. Otherwise the seal effect of the cable entry will not be ensured.
Adjustment module for insertion into VEGASON series 50 sensor or into the exter­nal indicating instrument VEGADIS 50
VEGASON 54K … 56K 21
Page 22

4 Mounting and installation

4.1 Mounting

Mounting and installation
Version A
Sensors in version A (flange version) are supplied completely mounted and ready for operation. Immediately after mounting on the vessel and electrical connection, they are ready for operation.
Version B
The sensors in version B are supplied in two parts (transducer and sensor electronics). First of all, mount the transducer on the ves­sel or above the medium. There is a four-pole jack at the end of the transducer tube. The respective counterpart to the jack protrudes out of the lower side of the sensor electronics housing. Insert the plug of the sensor elec­tronics (only possible in one position), into the jack of the transducer tube. Continue push­ing the electronics housing onto the trans­ducer tube, on which there is a wide and a narrow groove.
Groove for locking the headless screw
Mounting groove (must no longer be visible after mounting)
The wide groove is used for locking the headless screws. The narrow groove is a marking for mounting. Move the electronics housing farther down over the transducer tube until the mounting groove is no longer visible. Fasten the housing with the headless screws to the transducer tube. Use a 5 mm hexagon screwdriver (or Allen wrench).
22 VEGASON 54K … 56K
Page 23
Mounting and installation
Version C and D
The sensors in version C and D are supplied in three parts (transducer, sensor electronics and transducer cable). First mount the trans­ducer (see version B). There is a four-pole jack at the transducer tube end. A respective counterpart to the jack is provided in the connection cylinder of the transducer cable. Insert the connection cylinder plug into the jack of the transducer tube.
Connection cylinder
Mounting bracket
Connection cylinder
Transducer cable
On the end of the transducer tube you find a wide and a narrow groove. The wide groove is used for locking the cylinder with the head­less screws. The narrow groove is the mounting mark.
Then push the connection cylinder onto the transducer tube (with a slight swivelling mo­tion) until the mounting mark is no longer visible.
When the mounting mark is covered by the cylinder, fasten the cylinder with the two headless screws. Use a 5 mm hexagon screwdriver (or Allen wrench).
Now mount the sensor electronics in the requested location. The sensor electronics is fastened to a mounting bracket, so that it can be mounted to a plane surface or to the wall. Make sure that the sensor housing is mounted in such a way that there is enough space above the housing to open the cover.
Now insert the plug at the other end of the transducer cable into the jack on the elec­tronics housing.
Note:
Avoid bending the transducer cable too sharply when laying it out. This is special cable which could otherwise be damaged.
In addition, make sure that the cable is not damaged during operation. A signal with a voltage of approx. 1 kV is transmitted via the screened wires of the cable . If the cable is damaged, it could create a danger in Ex areas.
Groove for locking the headless screws
Mounting groove (must no longer be visible after mounting)
VEGASON 54K … 56K 23
Page 24

4.2 General installation instructions

Mounting and installation
Measuring range
ducer end, or with instruments in flange ver­sion, the instrument flange (version A).
Beside other criteria, you select your instru­ment according to the required measuring range. The reference planes for the min. and max.
Please note the information on the reference planes in chapter "3.2 Dimensions“. The max. filling depends on the required min. distance and the mounting location.
distance to the liquid or solid is the trans-
Reference plane
min. meas. distance 1.0 m
100 %
0 %
Span
Min. distance, max. measuring range, span and reference plane
0 %
max. meas. range
max. meas. distance 25 m (type 54), 45 m (type 55), 70 m (type 56)
min. meas. distance 0.8 m
100 %
Span
Reference plane
min. meas. distance 1.4 m
100 %
0 %
Span
Beam angle and false echoes
At greater distances, the energy of the ultra­sonic impulses distributes over a large area,
The ultrasonic impulses are focused by the transducer. The impulses leave the trans­ducer in conical form similar to the beam pattern of a spotlight. The beam angle is 4°
thus causing weaker echoes from obstruct­ing surfaces. The interfering signals are therefore less critical than those at close range.
(VEGASON 54), 5° (VEGASON 55) and 6° (VEGASON 56) at -3 dB emitted power. Any object inside this emission cone will cause a false echo. Especially within the first few meters of the emission cone, pipes,
If possible, orient the sensor axis perpen­dicularly to the product surface and avoid vessel installations (e.g. pipes and struts) within the 100 % area of the emission cone.
struts, or other installations can interfere with the measurement. At a distance of 6 m, the false echo of a strut has an amplitude nine times greater than at a distance of 18 m.
The following illustration of the ultrasonic beams is simplified and represents only the main beam. In reality, there are quite a number of ancillary beams.
24 VEGASON 54K … 56K
Page 25
Mounting and installation
100 %
50 %
70 m
0 m
7,5
3,7
7,5
3,7
0
m
12˚
0 m
Meas. distance
Emitted power
50 %
100 %
VEGASON 56VEGASON 54
Emitted power
25 m
Meas. distance
45 m
0 m
4
0,9
2,0
10˚
242
0
0,9
2,0
0
VEGASON 55
50 %
100 %
m
Emitted power
Emitted power
m
Meas. distance
Emitted power
Emitted power
Therefore in practical application, the trans­ducer has to be oriented so that lowest pos­sible false echo signal strength is achieved. Only giving attention to the large useful echo is therefore not always sufficient. In most cases, a low false echo level enables the sensor to reliably pick up the useful echo. With the adjustment software VVO on the PC you can have a look at the echo image, see chapter "6.2 Adjustment with the PC“ under the subheading "Sensor optimisation/Echo curve“.
VEGASON 54K … 56K 25
Page 26
Mounting and installation

4.3 Measurement of liquids

Flat vessel top
On flat vessels, mounting is usually done on a very short DIN socket piece. Reference plane is the lower edge of the flange. The transducer should protrude out of the flange tube.
< 100 mm
VEGASON 55 in flange version on short DIN socket piece
< 400 mm
Reference plane
Min. distance Type 54: 1 m Type 55: 0.8 m
Reference plane
A mounting location directly on the vessel top is ideal. A round opening in the vessel top is sufficient to fasten the sensor with the flange, or version B and C with swivelling holder.
Reference plane
Min. meas. distance
1.8 m
Swivelling holder
Min. meas. dis-
tance
1.4 m
Reference plane
Flange version and swivelling holder on flat vessel top
It is also possible to mount the sensors in version C in a 1“ thread.
Min. distance
1.8 m
< 60 mm
VEGASON 56 in flange version on short DIN socket piece
Reference plane
Mounting of the transducer in 1“ thread
26 VEGASON 54K … 56K
Page 27
Mounting and installation
Dished tank ceiling
On dished tank ceilings, please do not mount the instrument in the centre, but approx. 1/ vessel radius from the centre. Dished tank ceilings can act as paraboloidal reflectors. If the transducer is placed at the focal point of the parabolic ceiling, the transducer receives amplified false echoes. The transducer should be mounted outside the focal point. Amplified echoes caused by parabolic sur­faces are thereby avoided.
2
Reference plane
< 100 mm
1
/2 vessel radius
VEGASON 54 on dished vessel ceiling; the state­ments are also valid for VEGASON 55
< 400 mm
1
/2 vessel radius
VEGASON 56 on dished tank ceiling
Reference plane
Open vessels
On open vessels, using the instrument on an extended mounting bracket is a practical solution. Mount the low-weight sensor onto such a bracket and ensure a sufficient dis­tance to the vessel wall.
Reference plane
Min. meas.
Min. meas. distance
distance
Reference plane
Open vessels
VEGASON 54K … 56K 27
Page 28
Mounting and installation
Pump shaft
Narrow shafts and shaft openings (vessel openings) with very rough walls and shoul­ders make an ultrasonic measurement ex­tremely difficult due to strong false echoes. This problem can be overcome by using an extended socket piece or a complete meas­uring tube (see chapter "4.5 Socket exten­sion“).
see also „4.5 Socket extension“
Socket piece
250 mm
min. distance

4.4 Measurement of solids

Flange mounting
As with applications for liquids, the instrument can be mounted on a short DIN socket con­nection on vessels for solids. However the socket must be short enough to allow the transducer end to protrude from it and to not get in the way of the ultrasonic signals. The swivelling holder enables not only the align­ment to the product surface but also the minimisation of false echoes.
In the course of filling and emptying, the product surface often takes on different shapes. This can change the quality of the useful echo. The alignment of the transducer should thus ensure that false echoes are always at a minimum, even when the con­tainer is empty. You can have a look at the echoes on the PC with adjustment program VVO (see chapter "6 Setup/Adjustment with the PC/Sensor optimisation/Echo curve“).
Meas. range
Shaft pump
Example of a socket extension or measuring tube in a shaft
Measuring tube
Shaft
Very good measuring results can be attained
Shaft pump
Reference plane
Min. distance
with a measuring tube in continuous narrow shafts, see figure. The applied measuring tube must have smooth walls inside (e.g. PE sewage pipe) and a diameter ≥ 200 mm. This arrangement works well as long as the inside of the measuring tube collects no dirt or buildup (cleaning necessary). You might want to instead consider using hydrostatic pressure transmitters or capacitive measur­ing probes. The measuring tube must either be never immersed, or always immersed in the medium (measurement carried out exclu­sively within the tube).
28 VEGASON 54K … 56K
VEGASON 54C with an adapter flange on a DN 200 vessel flange
Page 29
Mounting and installation
Mounting boss
Reference plane
Min. distance
VEGASON 56 in 1“ mounting boss
The socket axis should point to the product surface. The use of a swivelling holder would be even better (type B, C), as described previously.
Material heaps

4.5 Socket extensions

The ultrasonic sensors require a min. dis­tance to the product or solid. Take the min. distance into account in your planning. In some situations, it is possible to reach the required min. distance, and hence the de­sired filling height, with a socket extension. However, the socket extension increases the noise level of the ultrasonic signal at the ex­tension outlet and can interfere with the measurement. Only use a socket extension if all other possibilities have to be excluded. Carry out the extension as shown in the fol­lowing illustration.
Socket extensions in liquids
Chamfer and deburr the socket carefully and make sure it has a smooth inner surface. The socket should not protrude into the meas­ured product, in case buildup can form on the socket through impurities or product residues.
With nonadhesive measured products, a socket extension in the form of a measuring tube can be permanently submerged in the product. The ultrasonic measurement is then carried out exclusively in the measuring tube and works very well without interference from other vessel installations (see "Pump shaft“).
Large material heaps are detected with sev­eral instruments, which can be mounted on e.g. traverse cranes. For this type of applica­tion, it is best to orient the sensor toward the solid surface.
Socket piece should not be immersed into adhesive products (figure: VEGASON 54)
Transducer on traverse crane above material heaps
VEGASON 54K … 56K 29
Page 30
Mounting and installation
The socket diameter should be as large and the socket length as small as possible. Make sure that the socket outlet is burr-free to minimise false echoes.
Typ e 5 4
L
˚
5
4
ø
Socket extensions that do not protrude into the measured product
Typ e 5 5
L
45˚
ø
Max. socket length in relation to socket diam­eter
ø in mm L in mm
Type 54 Type 55 Type 56
200 400 – – –– 250 500 500 500 300 –– –– 600

4.6 False echoes

The mounting location of the ultrasonic sensor must be selected such that no installations or inflowing material are in the path of the ultra­sonic impulses. The following examples and instructions show the most frequent measur­ing problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to their strong false echoes, adversely effect the measurement. Shields above these flat protrusions scatter the false echoes and guarantee a reliable measurement.
Correct Wrong
Socket extensions for solids
For solids, use a conical socket extension with a taper of at least 15° … 20°.
Vessel protrusions (slope)
Intake pipes, e.g. for the mixing of materials ­with a flat surface directed towards the sen­sor - should be covered with a sloping shield. This shield will scatter false echoes.
Correct Wrong
15˚ 15˚
Socket extension in solids
Vessel protrusions (intake pipe)
30 VEGASON 54K … 56K
Page 31
Mounting and installation
Vessel installations
Vessel installations such as, for example, a ladder, often cause false echoes. Make sure when planning your measurement loop that the ultrasonic signals have free access to the measured product.
Correct Wrong
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can cause strong false echoes that are superim­posed on the useful echo signals. Small shields effectively hinder a direct false echo reflection. These false echoes are scattered and diffused in the area and are then filtered out as "echo noise“ by the measuring elec­tronics.
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and not the inflowing mate­rial.
Correct
Correct
Inflowing material
Wrong
Wrong
Buildup
Correct Wrong
If the sensor is mounted too close to the vessel wall, and clumps of the product on the vessel wall can cause false echoes. Position the sensor at a sufficient distance from the vessel wall. Please also note chapter "4.2
Shields
Struts
VEGASON 54K … 56K 31
General installation instructions“.
Page 32
Mounting and installation
Correct
Buildup
Wrong
Strong product movements
Strong turbulences in the vessel, e.g. by power ful stirrers or strong chemical reac­tions, seriously interfere with the measure­ment. A surge or bypass tube of sufficient size (DN 200, DN 250) always allows, pro­vided the product causes no buildup in the tube, a reliable measurement even with strong turbulence in the vessel.
100 %

4.7 Incorrect mounting

Foam generation
Thick foam on the product can cause incor­rect measurements. Take measures to avoid foam, carry out the measurement in a bypass tube, or use a different measuring technol­ogy, e.g. capacitive measuring probes or hydrostatic pressure transmitters.
Foam generation
Wrong orientation to the product
Weak measuring signals are the result if the sensor is not directly pointed at the product surface. Orient the sensor axis perpendicu­larly to the product surface to achieve opti­mum measuring results.
60 %
0 %
Strong turbulences
Orient the sensor perpendicularly to the product surface
32 VEGASON 54K … 56K
Page 33
Mounting and installation
Strong heat fluctuations
Strong heat fluctuations, e.g. due to the sun, cause measuring errors. Please provide a sun shield.
Shield
Strong heat fluctuations
Min. distance to the medium
If the min. distance to the medium is not main­tained, the instruments show wrong meas­ured values. Mount the instrument at the required min. distance.
Correct Wrong
Sensor too close to the vessel wall
If the sensor is mounted too close to the vessel wall (dimension A in diagram), strong false echoes can be caused. Buildup, rivets, screws or weld joints on the vessel wall su­perimpose their echoes on the product or useful echo. Please ensure the sufficient distance of the sensor to the vessel wall, depending on the maximum measuring dis­tance (dimension B in diagram). In case of good reflection conditions (liquids, no vessel installations), we recommend determining the sensor distance according to Diagram curve
1. At a max. meas. distance of e.g. 10 m, the distance of the transducer (according to curve 1) should be approx. 1.5 m. In case of solids with bad reflection properties, deter­mine the distance to the vessel wall accord­ing to Diagram curve 2. Under very bad measuring conditions (rough vessel walls, struts), it might be necessary to increase the distance to the vessel wall, or to also filter out the false echoes by storing them in memory, thereby adapting the sensor more precisely to the environment.
Distance of the transducer to the vessel wall
A
2 m 4 m 6 m 8 m
Curve 1 (liquids)
10 m
Sensor too close to the vessel wall
VEGASON 54K … 56K 33
B
20 m
30 m
max. meas. distance
Curve 2 (solids)
Page 34
Mounting and installation
Parabolic effects of rounded or arched vessel tops
Round or parabolic tank tops act like a para­bolic mirror for the signals. If the sensor is placed at the focal point of such a parabolic tank top, the sensor receives amplified false echoes. The optimum location is generally in the area of half the vessel radius from the centre.
Correct
< 100 mm
~ 1/
2
vessel radius
Wrong
Socket piece too long
If the sensor is mounted in a socket exten­sion that is too long, strong false echoes are caused, and measurement is hindered. Make sure that the transducer protrudes at least 30 mm out of the socket piece.
Reference plane
< 100 mm
Wrong
Correct and wrong length of socket piece
Mounting on a vessel with parabolic tank top
34 VEGASON 54K … 56K
Page 35
Electrical connection

5 Electrical connection

5.1 Connection and connection cable

Safety information
As a rule, do the work in the complete ab­sence of line voltage. Always switch off the power supply before you carry out connect­ing work on the ultrasonic sensors. Protect yourself and the instruments, especially when using sensors which do not operate with low voltage.
Qualified personnel
Instruments which are not operated with protective low voltage or DC voltage must only be connected by qualified personnel.
Connection
A standard two or four-wire cable (sensors with separate supply) with max. 2.5 mm2 can be used for connection. Very often the "elec­tromagnetic pollution“ from electronic actua­tors, energy cables and transmitting stations is so considerable that the two-wire cable or the four-wire cable should be shielded.
We recommend the use of a screened cable. Screening is also a good preventative meas­ure against future sources of interference. However, you must make sure that no ground potential currents flow through the sensor cable shields. Ground potential cur­rents can be avoided by potential equalisa­tion cables. When grounding at both ends, it is possible to connect the cable shield on one side (e.g. in the switching cabinet) via an Y­capacitor1) to the ground potential. Use a low­resistance ground connection (foundation, plate or mains earth).
Note!
In Ex applications, grounding on both ends is not allowed due to potential losses.
Ex protection
If an instrument is used in hazardous areas, the respective regulations, conformity certifi­cates and type approvals for systems in Ex areas must be noted (e.g. DIN 0165).
Please note the approval documents with the safety data sheet attached to the Ex sensors.
Connection cable
Please make sure that the connection cables are specified for the expected operating tem­peratures in your systems. The cable must have an outer diameter of 5 … 9 mm or 3.6 … 8.6 mm, to ensure the seal effect of the cable entry.
Protective conductor terminal
The electronics housing of the sensors has a protective insulation. The protective conduc­tor terminal and the earth terminal in the elec­tronics housing are galvanically connected with the metallic transducer diaphragm. On sensors in version B, the protective con­ductor terminal is galvanically connected to the transducer diaphragm via the transducer tube when the sensor is completely mounted. On version C and D, the connection is made via the cable screen of the transducer cable and the transducer tube.
1)
max. 10 nF, e.g. voltage resistance 1500 V, ce­ramic
VEGASON 54K … 56K 35
Page 36
ESC
OK
Electrical connection

5.2 Connection of the sensor

After mounting the sensor at the measure­ment location according to the instructions in chapter "4 Mounting and installation“, loosen the closing screw on top of the sensor. The sensor lid with the optional indication display can then be opened. Unscrew the sleeve nut and slip it over the connection cable (after removing about 10 cm of insulation). The sleeve nut of the cable entry has a self-lock­ing ratchet that prevents it from opening on its own.
Version with aluminium housing
-
To the indicating instrument in the sensor cover or to the external indicat­ing instrument VEGADIS 50
1)
Voltage supply
Voltage supply
4 … 20 mA active
+
-
+
Now insert the cable through the cable entry into the sensor. Screw the sleeve nut back onto the cable entry and clamp the stripped wires of the cable into the proper terminal positions.
The terminals hold the wire without a screw. Press the white opening levers with a small screwdriver and insert the copper core of the connection cable into the terminal opening. Check the hold of the individual wires in the terminals by lightly pulling on them.
Version with plastic housing
Voltage supply
Cable entry M20 x 1.5
4 … 20 mA active
-
+
+
-
To the indicating instru­ment in the cover or to the external indicating instrument VEGADIS 50
1)
M20 x 1.5
12 C 567843
Sockets for connection of VEGACONNECT 2 (communi­cation sockets)
1)
4 … 20 mA active means, that the sensor provides a level dependent current of 4 … 20 mA (current source).
12 C 5 6 7 843
(+) (-) L1 N
Commu­nication+-4...20mA
-
+
Display
ESC
OK
Terminals (max. 2.5 mm wire cross-section)
Sockets for connec­tion of the HART handheld or VEGA­CONNECT 2
pluggable adjustment module MINICOM
2
12 C 567843
12 C 567843
Commu-
(+) (-) L1 N
®
Tank 1 m (d)
12.345
nication
-
+ -
4...20 mA
+
Display
ESC
OK
Opening tabs
36 VEGASON 54K … 56K
Page 37
Electrical connection

5.3 Connection of the external indicating instrument VEGADIS 50

Loosen the four screws of the housing cover on VEGADIS 50. The connection procedure can be facilitated by fixing the housing cover during connec­tion work with one or two screws on the right of the housing (figure).
OUTPUT (to the sensor)
2
1
3
5
7
4
8
6
Adjustment module
VEGADIS 50
+
-
Tank 1 m (d)
12.345
ESC
OK
Voltage supply
+
12 C 567843
2
(+) (-)
L1 N
Tank 1 m (d)
12.345
-
4 … 20 mA active
C567843
Commu-
+
nication
4...20 mA
-
DISPLAY
(in the cover of the indicating instrument)
­Display
ESC
+
OK
Screws
Four-wire sensor
(separate supply)
VEGASON 54K … 56K 37
Page 38

6 Set-up

Set-up

6.1 Adjustment methods

Series 50 ultrasonic sensors can be adjusted with
- PC (adjustment program VVO)
- with detachable adjustment module MINICOM
- with HART® handheld.
The adjustment must only be carried out with one adjustment device. If, for example you try the parameter adjustment with the MINICOM and the HART® handheld, the ad­justment will not work.
PC
With the adjustment program VVO Visual Operating System) on the PC you can adjust the ultrasonic sensors very easily. The PC communicates with the sensor via the interface adapter VEGACONNECT 2. During the process, a digital adjustment signal is superimposed on the signal and supply cable. The adjustment can be carried out directly on the sensor or at any desired loca­tion along the signal cable.
Adjustment module MINICOM
With the adjustment module MINICOM, you adjust in the sensor or in the external indicat­ing instrument VEGADIS 50. With a dialogue text display and 6 keys, the module offers the same adjustment functionality as the adjustment software VVO.
HART® handheld
VEGASON 50K ultrasonic sensors, like other HART® protocol-compatible instruments, can be adjusted with the HART® handheld. A manufacturer-specific DDD (Data-Device­Description) is not required. The ultrasonic sensors are adjusted with the HART® stand­ard menus. All main functions are therefore accessible. Functions that are rarely used, such as for example, the scaling of the A/D converter for the signal output, or the adjustment with me­dium, are not possible or are blocked with the HART® handheld. These functions must be carried out with the PC or the MINICOM.
(VEGA

6.2 Adjustment with the PC

The adjustment with the PC and the adjust­ment program VVO is no longer outlined in this operating instructions manual but in the „VEGA Visual Operating (VVO)“ manual. Apart from the adjustment of the sensor, will find further instructions on the functionality of the adjustment media with the PC in the VEGA Visual Operating manual.
PC on the sensor
To connect the PC to the sensor, the interface converter VEGACONNECT 3 is required. Make sure that the pins of VEGACONNECT 3 are completely inserted into the sensor sock­ets, as new pins have a slightly increased resistance to insertion. The pins should be inserted up to a depth of approx. 13 mm to 15 mm.
PC on the signal cable
Connect the two-wire cable of VEGACONNECT 3 to the signal cable of the sensor. If the resistance of the systems (PLC, current source etc.) connected to the signal cable is less than 250 , a supplementary resistor must be connected to the supply cable during adjustment, in order to get a signal cable load of 250 ... 350 . Reason: The digital signals superimposed on the signal cable would be considerably damped or even „short circuited“ due to insufficient system resistance.
When using a sensor in conjunction with a VEGA signal conditioning instrument, use a communication resistor according to the following schedule:
VEGA signal conditioning instr. Rx
VEGADIS 371 no additional
resistor necessary
VEGASEL 643 150 … 200 Ohm
VEGAMET 513 S1, 514 S2 515 S1, S2, VEGALOG EA card 100 … 150 Ohm
38 VEGASON 54K … 56K
Page 39
Set-up
250
PLC
Ri > 250
PLC
Ri < 250
Rx
VEGAMET/VEGALOG
VEGA signal conditioning instr. Rx
VEGADIS 371 no additional
resistor necessary
VEGASEL 643 150 … 200 Ohm
VEGAMET 513 S1, 514 S2 515 S1, S2, VEGALOG EA card 100 … 150 Ohm
VEGASON 54K … 56K 39
Page 40
Set-up
6.3 Adjustment with adjustment
module MINICOM
You can also adjust the sensor with the small, detachable adjustment module MINICOM. The adjustment module is plugged into the sensor or into the external indicating instru­ment (optional).
ESC
+
-
Tank 1 m (d)
12.345
OK
2
Tank 1 m (d)
12.345
4
When adjusting with the adjustment module, all sensor versions (adjustment options), as with the PC and the adjustment program VVO, are available. The adjustment with MINICOM, however, is different.
+
-
4 ... 20 mA
ESC
OK
Error codes:
E013 No valid measured value
- Sensor in the warm-up phase
- Loss of the useful echo E017 Adjustment span too small E036 Sensor program not operating
- Sensor must be programmed
(service)
- Fault signal also appears during
programming
E040 Hardware failure, electronics
defective
Adjustment steps
On pages 64 and 65 you can find the com­plete menu diagram of the adjustment mod­ule MINICOM. Set up the sensor in the numbered se­quence:
1. Measurement in gases
2. Operating range
3. Adjustment
4. Conditioning
5. Meas. conditions
6. False echo storage (only required when
errors occur during operation)
7. Indication of the useful and noise level
8. Outputs Short explanations to the setup steps 1 … 8 follow.
You carry out all adjustment steps with the 6
1. Measurement in gases
keys of the adjustment module. A small dis­play shows you, apart from the measured value, a short message on the menu item or the value of a menu adjustment.
Adjustment is only necessary when the measurement is done in gases (Co2, He, etc.) deviating from air. In case of measurement in gases, sound the distance of the sensor to
The volume of information of the small display, however, cannot be compared with that of the adjustment program VVO, but you will soon get used to it and be able to carry out your adjustments quickly and directly with the
the product surface and enter it in the menu item "Measurement in gases“. The sensor can then take the modified sound velocity in gases into account (as opposed to air) and output correct levels.
small MINICOM.
40 VEGASON 54K … 56K
Page 41
Set-up
2. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range has already been ad­justed with the min./max. adjustment. Gener­ally it is useful to set the operatign range slightly larger (approx. 5 %) than the measur­ing range.
Example: Min./max. adjustment: 0.300 … 5.850 m; adjust operating range to approx.
0.250 … 6.000 m.
Max.
Min.
100 % (1.270 m) correspond to 1200 liters
Span (4.58 m)
0 % (5.850 m) corresponds to 45 liters
Adjustment without medium
(adjustment independent of the actual level)
Key adjustment Display indication
Sensor
m(d)
4.700
Para-
OK
OK
OK
OK
+
meter
Adjust­ment
w.o medium
Ad­just­ment in
m(d)
(min. adjustment)
The distance indication flashes and you can choose "feet“ and "m“.
3. Adjustment
OK
Confirm the adjustment with
"OK“. Under the menu item " the sensor which measuring range it should operate with.
Adjustment
“ you inform
Ad­just­ment in
m(d)
0.0%
at
m (d)
XX.XXX
You can carry out the adjustment with or without medium. Generally you will carry out the adjustment without medium, as you can then adjust without a filling cycle.
+
or
With"+“ and "–“ you adjust the
percentage value for the min.
value or the lower level (exam-
ple 0.0 %).
OK
The adjusted percentage
value is written in the sensor
and the distance corres-
ponding to the percentage
value flashes.
VEGASON 54K … 56K 41
Page 42
Set-up - Adjustment with the adjustment module MINICOM
+
or
With the "+“ or "–“ key you can assign a level distance (ex­ample 5.85 m) to the previ­ously adjusted percentage value. If you do not know the distance, you have to do a sounding.
OK
The adjusted product dis­tance is written in the sensor and the display stops flash­ing.
You thereby adjusted the lower product dis­tance as well as the percentage filling value corresponding to the lower product distance.
100.0%
at
m (d)
XX.XXX
(max. adjustment)
The max. adjustment (upper product dis­tance) is made in the same way (example: 100 % and 1.270 m).
Note:
The difference between the adjustment val­ues of the lower product distance and the upper product distance should be as big as possible, preferably at 0 % and 100 %. If the values are very close together, e.g. lower product distance at 40 % (3,102 m) and upper product distance at 45 % (3.331 m), the measurement will be inaccurate. A char­acteristics curve is generated from the two points. Even the smallest deviations between actual product distance and entered product distance will considerably influence the slope of the characteristic curve. When the adjust­ment points are too close together, small errors multiply to considerably larger errors at the output of the 0 % or the 100 % value.
Adjustment with medium
with medium
Max.
Min.
adjust
adjust
at %
at %
XXX.X
XXX.X
Fill the vessel e.g. to 10 % and enter 10 % in the menu "
Min. adjust
“ with the "+“ and "–“ keys. Then fill the vessel, e.g. to 80 % or 100 % and enter 100 % in the menu "
adjust
“ with the "+“ and "–“ keys.
Max.
4. Conditioning
Signal condit ioning
Scal ing
0 %
100 %
corres
corres
ponds
ponds
XXXX
XXXX
Under the menu item " choose the product distance at 0 % and at 100 % filling. Then you enter the parameter and the physical unit as well as the decimal point.
Enter in the menu window " the figure of the 0 % filling. In the example of the adjustment with the PC and the adjust­ment software VVO this would be 45 for 45 liters.
prop.
Deci-
to
mal point
888.8
Conditioning
0 % corresponds
Mass
Unit
Kg
“ you
• Confirm with "OK“.
42 VEGASON 54K … 56K
Page 43
Set-up - Adjustment with the adjustment module MINICOM
With the "—>“ key you change to the 100 % menu. Enter here the value of your parameter corresponding to a 100 % filling. In the exam­ple 1200 for 1200 liters.
• Confirm with "OK“.
If necessary, choose a decimal point. How­ever, note that only max. 4 digits can be displayed. In the menu "
prop. to
“ you choose the parameter (mass, volume, distance…) and in the menu "
Unit
“ the physical unit (kg, l,
ft3, gal, m3 …).
Linearisation:
Adjust ment
Signal condit ioning
Scal ing
Lin. curve
Linear
Integr ation time
0 s
A linear correlation between the percentage value of the product distance and percent­age value of the filling volume is preadjusted. With the menu "Lin. curve“ you can choose between linear, spherical tank and cylindrical tank. The creation of a customised linearisation curve is only possible with the PC and the adjustment program VVO.
5. Meas. conditions
(see menu diagram page 64)
7. Useful level and noise level
In the menu
you get important information on the signal quality of the product echo. The higher the "S-N“ value, the more reliable the measure­ment (menu diagram MINICOM).
Ampl.: means amplitude of the level echo in
S-N: means Signal-Noise, i.e. the useful
The bigger the "S-N“ value (difference be­tween the amplitude useful level and the noise level), the better the measurement: > 50 dB Measurement excellent 40 … 50 dB Measurement very good 20 … 40 dB Measurement good 10 … 20 dB Measurement satisfactory 5 … 10 dB Measurement sufficient < 5 dB Measurement poor
Example:
Ampl. = 68 dB S-N = 53 dB
68 dB – 53 dB = 15 dB noise level
53 dB signal level difference indicates very high measurement reliability.
Ampl.:
XX dB
S-N:
XX
dB
dB (useful level)
level minus the level of the back­ground noise
8. Outputs
6. False echo storage
Under the menu "Outputs“ you determine, if
for example, the current output should be A false echo storage is always useful when unavoidable false echo sources (e.g. struts)
inverted, or which parameter should be
shown on the sensor display. must be minimised. By creating a false echo memory, you authorise the sensor electronics to note the false echoes and save them in an internal database. The sensor electronics treats these (false) echoes differently from the useful echoes and filters them out.
VEGASON 54K … 56K 43
Page 44
Set-up - Adjustment with the adjustment module MINICOM
Menu schematic for the adjustment module MINICOM
Sensor
m(d)
4.700
PPara­meter
Sensor opti­mise
Meas. enviro nment
Opera­ting range
Begin
m (d)
0.50
2.
End
SON 51K
2.00
m (d)
6.00
When switching on, the sensor type and the software version are displayed for a few seconds.
Confi­gura­tion
Sensor Tag
Sensor
5.
Meas. condit ions
Condit ions
Liquid
Condit ion
Solid
Fast change
No
Fast change
No
Agitat ed sur face No
High dust level No
Sensor addr.
Foam­ing prod. No
Large angle repose No
Multidrop operation (HART® sensor ad­dress):
• Sensor address zero: The sensor outputs beside the 4…20 mA signal also a digital (HART®) level signal.
• Sensor address 1…15: the sensor delivers only a digital (HART®) level signal. The sensor current is frozen to 4 mA (power supply).
Meas. unit
0
m (d)
Measur ing in gas
Meas. dist.
Measur ing in gas No
Measur ing in gas No
Mul­tiple echoes No
Mul­tiple echoes No
mm (d)
2,381
Correc­tion Now!
OK?
1.
Sound speed
m/s
333
Correc­tion Now!
OK?
3. 4. 8.
Adjust ment
w.out medium
Adjust ment in
m(d)
with medium
Min. adjust at %
XXX.X
0.0 %
at
m (d)
XX.XXX
Max. adjust at %
100.0%
at
m (d)
XX.XXX
XXX.X
Signal condit ioning
Sca­ing
0 % corres ponds
XXXX
Lin. curve
Linear
100 % corres ponds
XXXX
Integr ation time
Dezi­mal point
888.8
Out­puts
Curr. out-
0 s
Prop.
Unit
to
Mass
Kg
put
Curr. out­put
4-20mA
Prop. to
Dis­tance
Error mode
22mA
Sensor dis­play
44 VEGASON 54K … 56K
Page 45
Set-up - Adjustment with the adjustment module MINICOM
With these keys you move in the menu field to the left, right, top and bottom
ESC
6. 7.
False echo memory
Create new
Meas. dist.
m (d)
X.XX
Meas. dist.
learn­ing
Simu­lation
Simu­lation Now!
Simu­lation
XXX.X
OK?
OK?
%
Dis­tance
m (d)
4.700
Update
Meas. dist.
m (d)
X.XX
Update Now!
OK?
learn­ing
Simulation:Simulation:
Simulation:
Simulation:Simulation: One hour after the last simulation adjust­ment, the sensor returns automatically to normal operating mode.
Error codes: E013 No v alid meas ured valu e
- Sensor in the warm-up phase
- Loss of the useful echo E017 Adj us tmen t span too small E036 Sensor program not operating
- Sensor must be programmed
(service)
- Fault signal also appears during
programming
E040 Hardware failure, electronics
defective
Delete
Delete Now!
delet­ing
Ampl.: S-N:
OK?
XX dB
XX
dB
max. range
m (d)
7.000
Add´l func­tions
Info
Sensor Tag
Sensor
Dis­tance
m (d)
4.700
Sensor type
PULS54 K
Ampl.: S-N:
act. dist. m
X,XX
High dust level No
Fast change
Yes
Basic reset
Reset Now!
OK?
Reset­ing
XX dB
XX
Serial number
1094 0213
dB
Lan­guage
Eng­lish
Softw.
Softw.
vers.
date
2.00
15.09. 1999
Cur-
Tempe-
rent
rature
mA
8.565
Actual
Min.
temp.
temp.
18°C
Menu items in bolt print provide sensor and measured value information and cannot be modified in this position.
Light grey menu fields are only displayed if required (dependent on the adjustments in other menus).
White menu items can be modified with the "+“ or "–“ key and saved with the "OK“ key .
OK
4°C
Sensor addr.
0
Max. temp.
59°C
VEGASON 54K … 56K 45
Page 46

7 Diagnosis

Diagnosis

7.1 Simulation

To simulate a certain filling, you can call up the function "Simulation“ on the adjustment module MINICOM, in the software program VVO or on the signal conditioning instrument.
You simulate a vessel filling and thereby a certain sensor current. Please note that con­nected instruments, such as e.g. a PLC react according to their adjustments and will prob­ably activate alarms or system functions.
Simulation with VVO or PC
If you start the simulation mode with the ad­justment program VVO on the PC, the simu­lated level is outputted until you quit the simulation mode.
Simulation with HART® handheld
If you call up the simulation mode with the HART® handheld, the sensor only returns to operating mode after quitting the simulation mode.
Simulation with MINICOM
If you start the simulation mode on the adjust­ment module MINICOM, the sensor returns to standard operating mode after one hour.

7.2 Error codes

Error codes Rectifying measure
E013 No valid measured value Message is displayed during warm-up phase
- Sensor in the warm-up phase
- Loss of the useful echo If the message remains, an echo treatment and false echo reduction must be carried out in the menu "Sensor optimisation“.
E017 Adjustment span too small Carry out a readjustment.
E036 Sensor software does not run Sensor must have a new software (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tive Transducer defective.
46 VEGASON 54K … 56K
Make sure that the difference between min. and max. adjustment is at least 10 mm.
Message appears during a software update.
Contact our service department.
Page 47
Notes
VEGASON 54K … 56K 47
Page 48
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone (07836) 50-0 Fax (07836) 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application, use and operating conditions of the sensors and processing systems corre­spond to the information available at the time of printing.
Technical data subject to alterations
24106-EN-030723
Loading...