VEGA SON51V User Manual

Operating Instruction
VEGASON 51 V … 53 V
Level and Pressure
Contents
Safety information ........................................................................... 2
1 Product description
1.1 Function ................................................................................... 4
1.2 Application feature .................................................................. 5
1.3 Adjustment ..............................................................................6
2 Types and versions
2.1 Survey ..................................................................................... 9
2.2 Configuration of measuring systems .................................. 12
3 Technical data
3.1 Data ....................................................................................... 17
3.2 Approvals ............................................................................. 20
3.3 Dimensions ............................................................................. 21
Contents

Safety information

The described module must only be installed and operated as described in this operating instruction. Please note that other action can cause damage for which VEGA does not take responsibility.
2 VEGASON 51 V … 53 V
Contents
4 Mounting and installation
4.1 General installation instructions ........................................... 23
4.2 Measurement of liquids ........................................................ 2 5
4.3 Measurement of solids ......................................................... 28
4.4 Socket extensions ................................................................ 30
4.5 Flow measurement ............................................................... 31
4.6 False echoes ......................................................................... 32
4.7 Installation error .................................................................... 35
5 Electrical connection
5.1 Connection and connection cable ....................................... 38
5.2 Connection of the sensor ..................................................... 3 9
5.3 Connection of an external indicating instrument
VEGADIS 50 ......................................................................... 40
6 Set-up
6.1 Adjustment structure ............................................................ 41
6.2 Adjustment with PC on VEGAMET ...................................... 42
6.3 Adjustment with VEGAMET or MINICOM ........................... 59
6.4 Adjustment with PC on VEGALOG ...................................... 72
VEGASON 51 V … 53 V 3

1 Product description

1.1 Function

The continuous level measurement with ultra­sonic sensors is based on the running time measurement of ultrasonic pulses.
Measuring principle:
Piezoceramic high-performance transducers emit focussed ultrasonic pulses reflected from the surface of solids and liquids. The meas. signal prepares a precise picture of the meas. environment out of the running time and signal shape of the reflected ultrasonic pulses. The transducers operate as emitter and receiver. As receiver the transducers are high sensitivity piezomicrophones.
Meas. distance
Product description - Function
Out of the physical sound velocity and the detected actual running time of the emitted sound impulses, the meas. electronics calcu­lates precisely the distance between trans­ducer and product. The distance is con­verted into a level proportional meas. signal and provided acc. to the sensor parameter adjustment as exact scaled level. The instru­ments operate with emitting frequencies from 34 kHz to 70 kHz to be prepared for the different distances and requirements.
As the sound velocity is subjected to a tem­perature influence, the transducer detects continuously the ambient temperature so that the level is provided precisely even with varying ambient temperature.
emission - reflection - receipt
4 VEGASON 51 V … 53 V
Product description - Application feature
Output signal
The level proportional meas. signal is either processed in the ultrasonic instrument itself or in the connected signal conditioning instru­ment and provided as output signal. In digital technology varying line resistances or smallest leakage currents have no influ­ence to the accuracy. The digital signal is always clear.
Individual parameter adjustment of the digital signal is possible and the signal always re­flects the adjusted meas. range of the sen­sor.
Measured value indication
An indicating instrument for direct, local level survey is integrated in the series 50 ultra­sonic sensors. The indicating instrument shows with the analogue bargraph immedi­ately the level and with the digital figure value precisely the level. Additionally to the indica­tion in the sensor, the level can be indicated with the external indicating instrument VEGADIS 50 separated up to 25 m from the sensor. The external measured value indication like the integral indication works completely inde­pendent from the output signal and individual parameter adjustment is possible.

1.2 Application feature

Applications
• Level measurement of liquids
• Level measurement of solids (only short meas. distances) such as e.g.: coal, ore, stones, stone dust, cement, gravel, crushed stones, sand, sugar, salt, cereals, flour, granules, powder, dusts, saw dust, wood chips
• Flow measurement on various flumes
• Gauge measurement, distance meas­urement, object monitoring and con­veyor belt monitoring
Two-wire technology
• Supply and output signal on one two­wire line (loop powered)
• Output signal and signal processing completely digital
Rugged and precise
• Unaffected by product features such as density, conductivity and dielectric constant figure …
• Suitable for aggressive substances
• Meas. ranges 0,25m…15m
• Precise due to digital measured value processing and transmission
Adjustment choice
• With adjustment software VEGA Visual Operating (VVO) on PC
• With detachable adjustment module MINICOM
• With signal conditioning instrument VEGAMET
• Measured value indication integrated in the sensor
• Optionally indication separated from the sensor
Connection for each process
•G 11/2 A, 11/2" NPT
• G 2 A, 2" NPT
• Compression flange DN 100, ANSI 4"
Approvals
• CENELEC, FM, CSA, ABS, LRS, GL, LR
VEGASON 51 V … 53 V 5

1.3 Adjustment

Each measuring distance is different, hence each ultrasonic sensor must be given some basic information on the application and the environment. E.g. you inform the sensor which level means "empty" and which level "full". Beside this "Empty and full adjustment" also a number of other adjustments can be carried out with VEGASON ultrasonic sensors.
The adjustment and parameter adjustment of the ultrasonic sensors are hence carried out with:
- the PC
- the signal conditioning instrument VEGAMET 514 V or VEGAMET 514 V
- the detachable adjustment module MINICOM
Product description - Adjustment
Adjustment with PC
The set-up and adjustment of the ultrasonic sensors is generally made on the PC with the adjustment program VVO (VEGA Visual Op­erating) under Windows®. The program leads quickly through the ad­justment and parameter adjustment via pic­tures, graphics and process visualizations.
2
2
Adjustment with PC on the digital signal and supply line between the sensors and the signal conditioning instrument VEGAMET or on the sensor itself (figure: VEGASON 51 V)
The PC can be connected to any individual position of the system or the signal line. It is hence connected with the two-wire PC-inter­face converter VEGACONNECT 2 to the sen­sor, the signal line or to the signal condition­ing instrument, but also with a standard cable (RS232) directly to the processing system VEGALOG.
The adjustment and parameter adjustment data can be saved with the adjustment soft­ware on the PC and protected with pass­words. If required the adjustments can be transferred quickly to other sensors.
6 VEGASON 51 V … 53 V
Product description - Adjustment
2
……
The adjustment program recognizes the sensor type and the place of the connection
Visualized input of a vessel linearization curve
2
1…15
VEGALOG
VEGALOG
571 CPU
571 EA
Adjustment with PC on the processing system VEGALOG with standard cable RS 232 (on the processing system up to 15 sensors can be operated on one two-wire line)
2
……
1 … 15
VEGALOG
VEGALOG
571 CPU
571 EA
2
Adjustment on the digital signal and supply line to the processing system VEGALOG 571 or directly on the sensor
Adjustment with PC on signal conditioning instrument VEGAMET 515 V (two sensors) or 514 V (one sensor)
VEGASON 51 V … 53 V 7
Product description - Adjustment
Adjustment with signal conditioning in­strument VEGAMET
The ultrasonic sensors with digital output signal can be operated beside the PC also with the signal conditioning instrument VEGAMET.
%
100
+
-
OK
ESC
CONNECT
2
1
on
514 Ex
6-key adjustment field on the instrument front of a signal conditioning instrument VEGAMET
For adjustment the digital signal conditioning instruments VEGAMET 514 V and 515 V are provided with a 6-key adjustment field with display. Here the parameter adjustment in clear text can be made. The adjustment structure corresponds to the adjustment on the adjustment module MINICOM.
Adjustment with adjustment module MINICOM
The adjustment with the small (3,2 cm x 6,7 cm) 6-key adjustment module with display can be compared with that of the signal conditioning instrument. It is possible to carry out some sensor relevant adjust­ments directly on the meas. loop which can naturally also be made with the signal condi­tioning instrument.
+
ESC
-
Tank 1 m (d)
12.345
Detachable adjustment module MINICOM
The adjustment module can be plugged into and removed out of the ultrasonic sensor or the optional external indicating instrument.
ESC
+
-
Tank 1 m (d)
OK
12.345
2
4 … 20 mA
ESC
+
-
Tank 1 m (d)
OK
12.345
4
OK
%
100
+
-
OK
ESC
CONNECT 2
1
on
514 Ex
Adjustment with the detachable adjustment module on the ultrasonic sensor or on the external indicating instrument VEGADIS 10.
8 VEGASON 51 V … 53 V
Types and versions

2 Types and versions

2.1 Survey

VEGASON series 50 sensors are a newly developed generation of very compact, small ultrasonic sensors. With very narrow space requirements, they are designed for short meas. distances (0 ... 15 m) and for standard applications such as storage tanks, gauge measurement and buffer tanks.
Due to the small housing dimensions and process connections the compact sensors monitor very price favourable your levels. With the integral indication and the many features of the "big brothers" of VEGASON series 80, they open the advantages of an ultrasonic measurement for applications where the special advantages of a non-con­tact measurement were not applicable due to price reasons.
VEGASON 50 ultrasonic sensors dominate the two-wire technology perfectly. The supply voltage and the digital output signal are transmitted via one two-wire line to the con­nected signal conditioning instrument VEGAMET or the processing system VEGALOG.
VEGASON 51
with 11/2" thread
VEGASON 52
with 2" thread
VEGASON 53
DN 100 compression flange (here on a standpipe) or for mounting with holding strap
VEGASON 51 V … 53 V 9
Types and versions - Survey
Survey of the features
• Application in solids and liquids
• Meas. range 0,25 … 15 m
Signal output
digital meas. signal
Voltage supply
– two-wire technology (voltage
supply and signal output via one two-wire line)
Process connection
– G11/2 A; 11/2" NPT – – G 2 ; 2" NPT – – DN 100 compression flange
Adjustment
– with PC and adjustment software • – with adjustmet module in sensor • – with adjustmet module in external
indicating instrument • – with signal conditioning instrument VEGAMET
• Ex-approved in IEC or ATEX classification EEx ia [ia] IIC T 6
• Integral indication of measured values
VEGASON …
51 V 52 V 53 V
Meas. range in m
– liquids 0,25 … 4 0,4 … 7 0,6 … 15 – solids 0,3 … 2 0,25…3,5 0,75 … 7
10 VEGASON 51 V … 53 V
Types and versions - Survey
Type code
The second figure of the type designation e.g. VEGASON 5[1] differentiates the instru­ments acc. to process connection and meas. range
VEGASON 51 V E X X . X X X X X (example)
G - Process connection G 11/2 A N - Process connection 11/2 NPT X - Process connection DN 100 PN (without compr. flange) A - Process connection DN 100 PN (PPH compr. flange) B - Process connection DN 100 PN (1.4571 compr. flange) C - 1.4301 mounting loop
X - without indication A - with integral indication
X - without MINICOM adjustment module B - with MINICOM adjustment module (pluggable)
The figure e.g. VEGASON 51[V] character­izes the output signal: V stands for a digital output signal.
A - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA B - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA; HART C -Two-wire (loop powered); 4 … 20 mA D -Two-wire (loop powered); 4 … 20 mA; HART
®
®
E - Supply via signal conditioning instrument P - 90 … 250 V AC (only in USA) N - 20 … 36 V DC, 24 V AC (only in USA) Z - Supply via signal conditioning instrument (only in USA)
U - in USA X - outside USA
.X - without approval EX.X - Ex approved CENELEC EEx ia IIC T6 EXS.X - StEx Zone 10
K - Analogue 0 … 20 mA output signal (two-wire or four-wire technolgoy) V - Digital output signal (two-wire technology)
Type 51: Meas. range 0,25 … 4 m Type 52: Meas. range 0,4 … 7 m Type 53: Meas. range 0,6 … 15 m
Meas. principle (SON for ultrasonics)
VEGASON 51 V … 53 V 11
Types and versions - Configuration of measuring systems
2.2 Configuration of measuring sys­tems
A measuring system consists of a sensor and a processing unit. The processing unit (signal conditioning instrument VEGAMET or processing system VEGALOG) processes the level proportional digital meas. signals and evaluates these in a number of process­ing routines. The levels can then be provided as current, voltage or switching signals and further proc­essed, e.g. in a weir control. On the following pages you see the different instrument configurations called measuring systems which are shown in the following partly with a signal processing.
Meas. systems in two-wire technology:
• 2 sensors on one two-wire line
(page 13)
• 2 sensors in Ex on one two-wire line
(page 14)
• 15 sensors on one two-wire line
(page 15)
• 3 sensors in Ex on one two-wire line
(page 16)
Ex
Series 50 sensors require for operation in Ex­areas the Ex-separator VEGATRENN 548 V Ex, providing intrinsically safe Ex-circuits.
Note to page 14 and 16:
2)
Sensor lines should be looped in screened cables. It is suitable to earth the cable screens on both ends. However it must be noted that no earth compensation currents flow via the screens.
Earth compensation currents can be avoided with earthing on both ends by connecting the cable screen on one earth end (e.g. in the switching cabinet) via a capacitor (e.g. 1 µF, 100 V) to the earth potential.
Sensor lines leading to the same separator card, can be looped together in one screened multiple wire cable. Sensor lines leading to other separator cards, must be looped in separate, screened cables.
12 VEGASON 51 V … 53 V
Types and versions - Configuration of measuring systems
1 … 2 sensors on signal conditioning instrument VEGAMET 515 V
• Two-wire technology, supply from signal conditioning instrument. Output signals and voltage supply via one two-wire line.
• Digital output signal, two sensors on one line.
• Measured value indication in sensor and in signal conditioning instrument.
• Optionally external indicating instrument. (can be mounted up to 25 m separated from the sensor in Ex-area).
• Adjustment with PC, signal conditioning instrument or adjustment module (can be plugged in sensor or in external indicating instrument).
• Max. resistance of the signal line 15 per wire or 1000 m cable length
VEGADIS 50
Screened line in case of electro­magnetic interferences
4
2
2
VEGADIS 50
Current outputs Voltage outputs Relays Digital connection Fault signals
4
2
VEGACONNECT 2
VEGAMET
515V
Signal conditioning instru­ment VEGAMET 515 V in housing type 505
1) Sensor lines should be looped in screened cables. It is suitable to earth the cable screens on both ends. However it must be noted that no earth compensation currents
Processings see also product information "Signal conditioning instruments series 500"
flow via the screens. Earth compensation currents can be avoided with earthing on both ends by connecting the cable screen on one earth end (e.g. in the switching cabinet) via a capacitor (e.g. 1 µF, 100 V) to the earth potential.
VEGASON 51 V … 53 V 13
Types and versions - Configuration of measuring systems
1 … 2 sensors in Ex-area via separator VEGA TRENN 548 V Ex on signal condi­tioning instrument VEGAMET 515 V
• Two-wire technology, supply from the signal conditioning instrument. Output signals and voltage supply via one two-wire line (optionally with VEGASON 53 four-wire technology)
• VEGASON 51 … 52: CENELEC EEx ia IIC T6 VEGASON 53: StEx Zone 10
• Digital output signal, two sensors on one line.
• Measured value indication in the sensor and in the signal conditioning in strument.
• Optionally external indicating instrument (can be mounted up to 25 m separated from the sensor in Ex-area).
• Adjustment with PC, the signal conditioning instrument or the adjustment module (can be plugged into the sensor or the external indicating instru ment).
• Max. resistance of the signal line 15 per wire or 1000 m cable length note the permissible values for capacitance and inductance at the connection cable (see approval certificate of the separators)
VEGADIS 50
4
VEGADIS 50
Ex-area
Not Ex-area
Screened line in case of electro­magnetic interferences
2
Current outputs
2
Voltage outputs Relays Digital wiring Fault signals
4
2
VEGACONNECT 2
VEGAMET
VEGATRENN
515V
547
Signal conditioning instrument VEGAMET 515 V with Ex-separator VEGA TRENN 548 V Ex in housing type 506
Processings see also product information "Signal conditioning instruments series 500"
14 VEGASON 51 V … 53 V
Types and versions - Configuration of measuring systems
15 sensors via one two-wire line on the processing system VEGALOG 571
• Two-wire technology, power supply and digital output signals via one two­wire line from the processing system VEGALOG 571.
• 15 sensors on one two-wire line.
• Measured value indication integrated in the sensor.
• Optionally external indicating instrument (can be mounted up to 25 m separated from the sensor in Ex-area).
• Adjustment with PC or adjustment module (can be plugged into the sensor or external indicating instrument).
• Max. resistance of the signal line 15 per wire or 1000 m cable length.
4
VEGADIS 50
4
4
Screened line in case of electromagnetic interferences
2
2
2
2
2
2
VEGASON 51 … 53 (15 sensors per two-wire line can be individually grouped)
1)
2
CPU
VEGALOG
VEGALOG
571 CPU
571 EV
Processing system VEGALOG 571 with input cards in 19"-rack. 15 sensors on one module card and two-wire line
Processings see also product information
VEGA­CONNECT 2
Interface cable RS 232
"Signal conditioning instruments series 500"
1) Sensor lines should be looped in screened cables. It is suitable to earth the cable screens on both ends. However it must be noted that no earth compensation currents flow via the screens. Earth compensation currents can be avoided with earthing on both ends by connecting the cable screen on one earth end (e.g. in the switching cabinet) via a capacitor (e.g. 1 µF, 100 V) to the earth potential.
Current outputs Voltage outputs Relays Digital wiring Fault signals Connection to all Bus­systems Transistor outputs
VEGASON 51 V … 53 V 15
Types and versions - Configuration of measuring systems
3 sensors per two-wire line via separator VEGA TRENN 548 V Ex on the processing system VEGALOG 571
• Two-wire technology, power supply and digital output signals via one two­wire line from the separator.
• Three sensors on one two-wire line.
• Measured value indication integrated in the sensor.
• Optionally external indicating instrument (can be mounted up to 25 m separated from the sensor in Ex-area).
• Adjustment with PC or adjustment module (can be plugged in the sensor or external indicating instrument).
• Max. resistance of the signal line 7,5 per wire or 1000 m cable length, note the permissible values for capacitance and inductance of the connec­tion cable (see also approval certificates of the separators).
VEGADIS 50
2
Ex-area
Not Ex-area
Screened line in case of electromagnetic interferences
2
2
2
22
2
2)
(see page 12)
2 2 2
2 2 2
2 2 2
Processing system
CPU
VEGALOG 571 (19" module card)
VEGALOG
VEGALOG
VEGATRENN
571 CPU
VEGATRENN
571 EV
548
VEGACONNECT 2
VEGATRENN
VEGALOG
VEGATRENN
VEGATRENN
571 EV
548
548
548
548
Separator VEGATRENN 548V Ex (max. 9 sensors per card)
Input card of VEGALOG 571 (max. 15 sensors per card)
Current and voltage outputs Digital wiring Fault signals Transistor outputs Connection to all Bus-systems Relays
Interface cable RS 232
2
VEGASON 51 … 53 3 sensors per two-wire line, individual
2
grouping
16 VEGASON 51 V … 53 V
Technical data

3 Technical data

3.1 Data

Type 51… 52… 53…
Power supply
Supply voltage from signal conditioning instrument VEGAMET or
Current consumption max 22,5 mA Power consumption max. 80 mW 0,45 VA
Meas. range (relating to the transducer diaphragm)
Liquid 0,25 … 4 m 0,4 … 7 m 0,6 … 15 m Solid 0,30 … 2 m 0,5 … 3,5 m 0,75 … 7 m
Output signal
Digital meas. signal (VBUS)
Adjustment
- PC with adjustment software software VEGA Visual Operating
- adjustment module MINICOM
processing system VEGALOG 571 (max. 36 V DC) fuse 0,5 A (slow-blow)
Accuracy (type, values under reference conditions)
1)
Linearity error after adjustment < 0,1 % (relating to max. meas. range) Average temperature error of sensor electronics < 0,03 %/10 K of meas. range Resolution 1 mm
Characteristics
Min. span (between empty and full adjustment) > 10 mm (recommended > 50 mm) Ultrasonic frequency 70 kHz 55 kHz 38 kHz Meas. intervals 1,0 s 1,0 s 0,6 s Beam angle 5,5° 5,5° 3° (at -3 dB emitted power)
1) Reference conditions acc. to IEC 770, e.g.: Temperature: 18 ... 30°C etc.
VEGASON 51 V … 53 V 17
Technical data
Ambient conditions
max. vessel pressure (gauge pressure)
- VEGASON 51 and 52 3 bar
- VEGASON 53 2,5 bar
Ambient temperature
- sensor (electronics) -20°C … +60°C
- process (transducer) -20°C … +80°C (StEx: -20°C … +75°C)
- storage and transport temperature -40°C … +80°C
Protection
- sensor IP 67
- transducer, process IP 68
Protection class II Overvoltage category III
Ex-technical data (note approval documents)
Classification ia (intrinsically safe in conjunction with a safety
barrier or separator) Temperature class (permissible ambient temperature on transducer when used in Ex-areas)
- T6 45°C
- T5 60°C
- T4 60°C
- T3 60°C
Ex-approved in category or zone
- EC-type approval Zone 1 (II 2 G); Type 53: Zone 21 (II 2 D)
- conformity certificate Zone 1; Type 53: Zone 10
Classification EEx ia IIC T6
Materials
Housing PBT (Valox) Transducer, process thread PVDF Compression flange PP or 1.4571 Transducer diaphragm (type 53) 1.4571
Weights (incl. transducer)
- VEGASON 51 1,2 kg
- VEGASON 52 1,6 kg
- VEGASON 53 2,3 kg
18 VEGASON 51 V … 53 V
Technical data
Process connections
VEGASON 51 G 11/2 A, 11/2" NPT VEGASON 52 G 2 A, 2" NPT VEGASON 53 DN 100 compression flange
Connection lines
Two-wire sensors
- supply and voltage via one two-wire line line resistance max. 15 per wire or 1000 m cable length
Cross-section area of conductor generally 2,5 mm Earth connection max. 4 mm
2
2
Cable entry 2 x M20 x 1,5 (cable diameter 5 … 9 mm)
CE-Conformity
VEGASON series 50 ultrasonic sensors meet the protective regulations of EMVG (89/336/ EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following stand­ards: EMVG Emission EN 50 081 - 1: 1993
Susceptibility EN 50 082 - 1: 1995
NSR EN 61 010 - 1: 1993
Outputs and processinds
Display indication
Indication optionally integral, scalable analogue and digital
Signal output
Signal output digital output signal in two-wire technology
Two-wire technololy: The digital output signal (meas. signal) is modulated to the power supply and processed in the signal conditioning instrument or in the processing system.
VEGASON 51 V … 53 V 19
measured value indication optionally external measured value indication powered by the sensor which can be mounted up to 25 m from sensor
(VBUS)
Technical data - Approvals

3.2 Approvals

When using ultrasonic sensors in Ex-areas the instruments must be suitable and ap­proved for these explosion zones and appli­cations. For the use on ships special type approval certificates are available. In hazardous areas (due to gases, vapours, fog and dusts) only appropriately approved instruments must be used. The suitability is checked by the approval authorities and is certified in approval docu­ments.
VEGASON 51 V Ex and VEGASON 52 V Ex sensors must be powered for the use in Ex­zone from one intrinsically safe circuit. This is ensured by the separator VEGATRENN 548 V Ex. The separator provides intrinsically (ia) cir­cuits. The resistance of the signal line must not exceed 15 per wire.
VEGASON 51 V Ex and VEGASON 52 V Ex sensors are approved for Ex-zone 1. VEGASON 53 V ExS is approved for Ex­zone 10 or zone 21. As proof for the Ex-approval appropriate confirmity certificates and type approval certificates of the approval authorities are attached to the sensors.
Test and approval authorities
VEGASON ultrasonic sensors are tested and approved by the following monitoring, test and approval authorities:
- PTB (Physikalisch Technische Bundesanstalt ­Physical Technical Approval Authority)
- FM (Factory Mutual Research)
- ABS (American Bureau of Shipping)
- LRS (Lloyds Register of Shipping)
- GL (German Lloyd)
- CSA (Canadian Standards Association)
Please note the attached approval docu­ments when you want to use a sensor in Ex­areas.
20 VEGASON 51 V … 53 V
Technical data - Dimensions

3.3 Dimensions

VEGASON 51
VEGASON 52
2 x M20x1,5
Thread G 11/2 A or 11/2" NPT
Reference pane
Min.-distance to the product
Thread G 2 A or 2" NPT
SW 60
Reference pane
Min.-distance to the product
VEGASON 51 V … 53 V 21
Technical data - Dimensions
VEGASON 53
ø12
Mounting loop
95
247
193
120 12
M8x10
ø148 ø158
External indicating instrument VEGADIS 50
38
ø5
82
ø12
152
suitable for compression flange DN 100
Min.-distance to the product
0,6 m
118
108
135
Pg 13,5
M20 x 1,5
Mounting on carrier rail 35 x 7,5 acc. to EN 50 022 or flat screwed
85
Note:
The cable diameter of the connection cable must be 5 … 9 mm. The seal effect of the cable entry is otherwise not ensured.
22 VEGASON 51 V … 53 V
Mounting and installation - General installation instructions

4 Mounting and installation

4.1 General installation instructions

Measuring range
You select your instrument beside other crite­ria according to the required measuring range. The reference panes for the min. and max. distance to the product or the solid is either the transducer surface (diaphragm) or on instruments with flange version the instru-
ment flange. Note the information for the refer­ence panes in chapter "Dimensions". The max. filling is dependent on the required min. distance of the instrument used (0,25 m to 0,75 m) and the installation place of the in­strument or the transducer.
Type 51
Min. distance
0,25 m 0,4 m
4 m (type 51)
full
1m
15 m (type 53)
empty
Min. distance
0,75 m
Type 53 Type 52
Reference pane
Meas. range
Min. distance, max. distance and meas. range (example VEGASON 51, 52 and 53)
Min. distance
7 m (type 52)
VEGASON 51 V … 53 V 23
Mounting and installation - General installation instructions
Beam angle and false echoes
The ultrasonic impulses are focussed by the transducer. The impulses leave the trans­ducer in conical form similar to the beam pattern of a spotlight. The beam angle is 5,5° (type 51/52) or 3° (type 53) at -3 dB emitted power.
Any object in this beam angle causes a false echo. Within the first few metres of the beam angle, tubes, struts or other installations cause strong false echoes. In a distance of 2 m the false echo of a strut has 9-times more amplitude than at a distance of 6 m.
If possible, provide a vertical directing of the sensor axis to the product surface and avoid if possible, struts within the 100%-beam angle, e.g. by tubes and struts.
Meas. distance
0 m
4 m
7 m
Meas. distance
0 m
VEGASON 51/52
0,4
0
VEGASON 53
0,4
5,5°
12°
50 %
Emitted power
100 %
Emitted power
1,21,2
m
50 %
Emitted power
Emitted power
100 %
15 m
0,4
1,2
0,4
0
m
1,2
24 VEGASON 51 V … 53 V
Mounting and installation - Measurement of liquids

4.2 Measurement of liquids

Flat vessel tops
On flat vessels the mounting is mainly made on a very short DIN-socket piece. Reference pane on flange versions is the instrument flange. The transducer should protrude out of the flange pipe.
Reference pane
You can also mount the sensors with 11/2" or 2" thread to short socket pieces.
< 60 mm
Reference pane
< 60 mm
Flange version on very short DIN-socket piece
Min. distance
Very suitable is the mounting directly on the vessel top. A round opening on the vessel is sufficient to fasten VEGASON 53 sensor with a compression flange.
Reference pane
Mounting on short 11/2" or 2" socket piece
Flange version (compression flange) on flat vessel top
VEGASON 51 V … 53 V 25
Mounting and installation - Measurement of liquids
Dished tank ends
On dished tank ends please do not mount the instrument in the centre but approx. 1/2 vessel radius from the centre. Dished tank ends can act for the ultrasonic pulses as paraboloidal refelctors. If the transducer is placed in the "focus" of the parabolic tank end, the transducer receives amplified false echoes. Hence the transducer should be mounted outside the "focus". Parabolic ampli­fied false echoes are hence avoided.
Reference pane
< 60 mm
1
/2 vessel radius
Open vessels
On open vessels the use of the instruments on a mounting lever is recommended. Mount the low-weight sensor to a mounting lever.
Reference pane
Min. meas. distance
Min. meas. distance
Flange on dished vessel ends
Open vessels
< 60 mm
Reference pane
1
/2 vessel radius
Mounting boss on a dished tank end
26 VEGASON 51 V … 53 V
,,
Mounting and installation - Measurement of liquids
Pump shaft
Narrow shafts and shaft openings (vessel openings) with very rough walls and shoul­ders aggravate an ultrasonic measurement due to strong false echoes.
Shaft and vessel openings
Narrow, very rough shaft, well and vessel openings can be overcome by a socket piece as described under "4.4 Socket exten­sions" (left lower half of the figure).
see "4.4 Socket extension"
100 mm
,,,,,,,,,,,
min. distance
Socket piece
Shaft
In completely narrow shafts you can realize very good meas. results with a meas. pipe, see figure. The meas. pipe used must have smooth walls inside (e.g. PE-sewage pipe) and at least a diameter of 100 mm. This coordination works without problems as long as the inside of the meas. pipe has no build-up (cleaning). Also check the use of hydrostatic pressure transmitters or capaci­tive electrodes.
Meas. range
Shaft pump
Meas. tube
Shaft pump
Socket piece or meas. tube on the example of a shaft
VEGASON 51 V … 53 V 27
Mounting and installation - Measurement of solids

4.3 Measurement of solids

Flange mounting
As with liquids, also in solid vessels the in­strument can be mounted on a short DIN­socket piece. The socket axis must point to the vessel outlet or should be directed verti­cally to the product surface and must be very short (< 100 mm).
Reference pane
Min. distance
Swivelling holder
The accessory program offers a swivelling holder (mounting loop) for mounting of VEGASON 53. This facilitates the directing of the sensor to the product surface.
Reference pane
Min. distance
VEGASON 53 on vessel flange
28 VEGASON 51 V … 53 V
VEGASON 53 on swivelling holder
,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,,
Mounting and installation - Measurement of solids
Mounting boss
Reference pane
Min. distance
VEGASON 51 or 52 on the mounting boss. The socket axis should point to the product surface
Material cones
Large material cones are detected with sev­eral instruments which you can fasten on lifting beams.
T ransducer on lifting beam above material cone
VEGASON 52 on mounting boss
VEGASON 51 V … 53 V 29
,
,
,
,
,
,
,
,
,

4.4 Socket extensions

The ultrasonic sensor requires a min. dis­tance to the product or solid. Note this min. distance in your planning. In exceptions it is possible to reach the re­quired min. distance and hence the desired filling height with a socket piece. However the socket piece increases the noise level of the ultrasonic signal and can interfere the measurement. Only provide the socket ex­tension when there is no other possibility and carry out the extension as shown in the fol­lowing figure.
Socket extensions with liquids
Chamfer the socket carefully and ensure a smooth inner side of the socket. The socket should not protrude into the measured prod­uct when pollution or measured product could stick to the socket.
Mounting and installation - Socket extensions
Type 51/52
L
Socket extension with liquids
Type 53
L
45˚45˚
Max. socket length dependent on the socket diameter
ø L in mm in mm Type 51 Type 52 Type 53
100 200 300 300 150 300 400 400 200 500 500 250 600
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
,,,,,,,,,,,,,,,
Socket extension with solids
For solids provide a conical socket extension with an angle of at least 15° … 20°.
Socket piece should not penetrate into adhesive products (VEGASON 53)
Choose a possibly large socket diameter and a possibly small socket length, ensure that the socket opening is burr free to mini­mize false echoes.
15°
15°
For measurements in products which do not cause any build-up, the socket extension in form of a meas. pipe can be submerged permanently into the measured product. The
Socket extension with solids
ultrasonic measurement is then only carried out in the meas. pipe without being interfered by vessel installations (see page 28 "Pump shaft").
30 VEGASON 51 V … 53 V
Mounting and installation - Flow measurement

4.5 Flow measurement

The short examples on this page should give just introductory information for your flow measurement. Planning information is given by the flume manufacturers and in special literature.
Rectangular flume
- Installation of the sensor on the upstream side
- Note distance to the overfall edge (3…4xh
- Installation centered to the flume
- Edge opening 2xh
- Installation perpendicular to the liquid surface
- Keep min. distance relating to h
- Min. distance of the edge opening of downstream water 50 mm
max
)
from ground
max
max
Khafagi-Venturi flume
- Installation of the sensor on the inlet side
- Note distance to the Khafagi-Venturi flume (3…4xh
- Installation perpendicular to the liquid surface
- Keep min. distance relating to the height of damming h
max
max
3 … 4 x h
90°
)
max
h
max
90°
Overfall edge
< min. distance
h
max
2 x h
max
Khafagi-Venturi flume
Sensor
B
Flow measurement on open flumes
Overfall edge
3 … 4 x h
max
90°
5 cm
Upstream water
Downstream water
Flow measurement on open flumes
VEGASON 51 V … 53 V 31

4.6 False echoes

The installation place of the ultrasonic sensor must be selected such that no installations or inflowing material cross the ultrasonic im­pulses. The following examples and instruc­tions show frequent meas. problems and how they can be avoided.
Shoulders
Vessel forms with flat shoulder can aggravate the measurement due to the strong false echoes. Shields above these flat shoulders spread the false echoes and ensure reliable measurement.
Correct Wrong
Shoulders
Inlets, e.g. for material mixing with flat upper side pointing to the sensor, should be cov­ered by an angle shield. The false echo is hence gated out.
Correct Wrong
Mounting and installation - False echoes
Vessel installations
Vessel installations such as e.g. a ladder often cause false echoes. Note when plan­ning your measurement loop that the ultra­sonic signals should reach the product with­out problems.
Correct Wrong
Ladder
Vessel installations
Ladder
Struts
Struts such as vessel installations can cause strong false echoes which can overlay the useful echo. Small shields avoid a direct false echo reflection. The false echoes are diffused and filtered out by the meas. electronics as "echo noise".
Correct Wrong
Shields
Shoulder (inlet)
32 VEGASON 51 V … 53 V
Struts
,,
,,
,
Mounting and installation - False echoes
,
Flood basin
The max. high water to be expected deter­mines the installation height to keep the min. distance of the transducer even with highest high water.
The low water edge should be covered in the transducer range with a shield.
Correct Wrong
Min. distance high water
60°
Shield
,,,,,
Low water
,,,,,,
Gating out of an echo
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and note the inflowing mate­rial.
Correct
Wrong
Inflowing liquid
Build-up
If the sensor is mounted too close to the vessel wall, build-up on the vessel walls causes false echoes. Position the sensor in a sufficient distance to the vessel wall. Also note chapter "4.1 General installation instruc­tions".
Correct
Wrong
Correct
Wrong
Build-up
Inflowing liquid
VEGASON 51 V … 53 V 33
Strong product movements
Heavy turbulences in the vessel, e.g. by strong stirrers or strong chemical reactions, aggravate the measurement. A surge or bypass pipe (figure) of sufficient size allows, provided that the product causes no build­up in the pipe, always a reliable measure­ment even with strong turbulences in the vessel.
Mounting and installation - False echoes
Strong product movements
34 VEGASON 51 V … 53 V
,
,
,
,
Mounting and installation - Installation error

4.7 Installation error

Foam generation
Strong foam on the product can cause faulty measurements. Provide measures to avoid foam, measure in a bypass pipe or use an­other meas. principle, e.g. capacitive elec­trodes of hydrostatic pressure transmitters.
,,,,,,,,,,,,,
,,,,,,,,,,,,,
,,,,,,,,,,,,,
,,,,,,,,,,,,,
Strong heat fluctuations
Strong heat fluctuations, e.g. due to the sun, cause meas. errors. Provide a shield.
Shield
Foam generation
Wrong directing to the product
If the sensor is not directed to the product surface, weak measuring signals are caused. Direct the sensor axis vertically to the prod­uct surface to reach optimum meas. results.
Direct sensor vertically to the product surface
Strong heat fluctuations
Min. distance to the product
When the min. distance to the product is not maintained, the instruments show wrong measured values (see page 23). Mount the instruments with the required min. distance.
VEGASON 51 V … 53 V 35
Mounting and installation - Installation error
Sensor too close to the vessel wall
Correct Wrong
Sensor too close to the vessel wall
If the sensor is mounted too close to the vessel wall (dimension A in diagram) strong false echoes can be caused. Build-up, rivets, screws and weld joints superimpose their echoes to the product echo or useful echo. Hence note a sufficient distance of the sensor to the vessel wall dependent on the max. meas. distance (dimension B in diagram). In case of good reflection conditions (liquids, no vessel installation) we recommend to provide the sensor distance according to diagram curve 1. At a max. meas. distance of e.g. 10 m the distance of the transducer accord­ing to curve 1 would be approx. 1 m.
In case of solids with bad reflection condi­tions provide a distance to the vessel wall according to diagram curve 2. With very bad meas. conditions it can be necessary to increase the distance to the vessel wall or to gate out the false echoes additionally by a false echo storage and hence to adapt the sensor more precisely to the environment.
Distance of the transducer to the vessel wall
A
1 m 2 m 3 m 4 m 5 m
Curve 1 (liquids)
5 m
B
Max. meas. distance
10 m
15 m
Curve 2 (solids)
36 VEGASON 51 V … 53 V
Mounting and installation - Installation error
Socket piece too long
When mounting the sensor in a too long socket piece, strong false echoes are caused, aggravating the measurement. Note that the transducer protrudes at least 30 mm out of the socket piece.
Wrong
Correct
< 60 mm
Reference pane
Correct and wrong length of socket piece
Parabolic effects on dished boiler head or basket arch vessels
Round or paraboilic tank tops act for the signals like a paraboilic mirror. If the sensor is placed to the focus of such a paraboilic tank top the sensor receives amplified false ech­oes. The optimum mounting is generally in the range of the half vessel radius from the cen­tre.
Correct
< 60 mm
~ 1/
2
vessel radius
Wrong
Wrong
Mounting on a vessel with parabolic tank top
VEGASON 51 V … 53 V 37
5 Electrical connection

5.1 Connection and connection cable

Safety information
Ensure that the instrument is unpressurized before you start work. Always switch off the power supply before you carry out clamping work on the ultrasonic sensors. Protect your­self and the instruments.
Electrical connection
Ex-protection
If an instrument is used in hazardous areas the appropriate regulations, conformity certifi­cates and type approvals of the sensors and separators or safety barriers must be noted (e.g. DIN VDE 0165).
Skilled staff
Instruments which are operated in Ex-areas must only be connected by skilled staff. The skilled staff must know the installation condi­tions and note the attached type approval and conformity certificates.
Connection
A standard two-wire cable with max. 2,5 mm can be used for connection. Very often the "electromagnetic pollution" by electronic ac­tuators, energy lines and transmitting stations is so considerable that the two-wire line should be screened.
We recommend to use a screening. This screening prevents against future interfer­ences. It is suitable to earth the cable screens on both ends. However it must be noted that no earth compensation currents flow via the screens.
Earth compensation currents can be avoided with earthing on both ends by connecting the cable screen on one earth end (e.g. in the switching cabinet) via a capacitor (e.g. 1 µF, 100 V) to the earth potential. Use a very low impedance earth connection (foundation, plate or mains earth).
Sensors which are used in Ex-areas must only be operated on intrinsically safe circuits. The permissible electrical values are stated in the conformity certificate or the type approval.
Connection cable
Note that the connection cable must be speci-
2
fied for the expected operating temperatures in your systems. The cable must have an outer diameter of 5 … 9 mm otherwise the seal effect of the cable entry will not be ensured.
Cables for intrinsically safe circuits must be marked blue and must not be used for other circuits.
Earth conductor terminal
On VEGASON 51/52 sensors the earth con­ductor terminal is galvanically isolated. On VEGASON 53 sensor the earth conductor terminal is galvanically connected to the metal transducer diaphragm.
Note: In Ex-applications the cable screen of the connection cable must only be earthed at one end.
38 VEGASON 51 V … 53 V
Electrical connection - Connection of the sensor

5.2 Connection of the sensor

After having mounted the sensor in the meas. position according to the instructions in chap­ter "4 Mounting and installation", loosen the closing screw on the top of the sensor. The sensor cover with the optional indication display can then be opened. Unscrew the compression screw and shift the screw over the approx. 10 cm dismantled connection cable. The compression screw of the cable entry is protected with a safety lock-in posi­tion against automatic loosening.
Supply voltage and digital meas. signal
+
-
Now loop the cable through the cable entry into the sensor. Screw the compression screw to the cable entry and clamp the dis­mantled wires of the cable to the appropriate terminal positions. The terminals operate without terminal screw. Press the white opening buckets with a small screwdriver and insert the copper core of the connection line into the terminal opening. Check the position of the lines in the terminal position by slightly pulling on the connection lines.
T o the indicating instrument in the sensor cover or to the external indicat­ing instrument VEGADIS 50
Earth terminal The earth terminal must be connected to system earth
Spring-loaded terminal
12 C 567843
12 C 5678
Commu-
+-
Tank 1 m (d)
12.345
nication
-
+
4...20 mA
Two-wire technology
Display
ESC
OK
Terminals (max. 2,5 mm wire cross-section)
Sockets for connection of the interface adapter VEGACONNECT
Pluggable adjustment module MINICOM
2
VEGASON 51 V … 53 V 39
Electrical connection - Connection of the indicating instrument

5.3 Connection of an external indicating instrument VEGADIS 50

Loosen the 4 screws of the housing cover on VEGADIS 50. You can facilitate the connection procedure by fastening the housing cover during con­nection with two screws on the right of the housing (figure).
Terminal strip VEGADIS 50
3
2
1
4
5
to DISPLAY in cover of VEGADIS 50
8
6
7
Adjustment module
-
Tank 1 m (d)
12.345
321 5678
-
+
OUTPUT
+
DISPLAY
VEGADIS 50
ESC
OK
16.85
Voltage supply
and
digital meas. signal
+
-
12 C 567843
-
+
Tank 1 m (d)
12.345
Sensor terminal box
(open)
+
ESC
-
OK
Screws
40 VEGASON 51 V … 53 V
Set-up - Adjustment structure

6 Set-up

6.1 Adjustment structure

Series 50 ultrasonic sensors can be adjusted with
- PC (adjustment program VVO)
- detachable adjustment module MINICOM
- signal conditioning instrument VEGAMET.
Signal conditioning instrument VEGAMET
The signal conditioning instrument VEGAMET allows with the 6-key adjustment field with text display the parameter adjustment in the same functional spectrum than the adjust­ment program VVO on the PC.
The adjustment must only be carried out with one adjustment medium.
Adjustment program VVO
With the adjustment program VVO (VEGA Visual Operating) on the PC you adjust the ultrasonic sensors very comfortably. The PC communicates via the interface adapter VEGACONNECT 2 or the standard RS232 interface cable digitally with the sensor. The adjustment can be carried out hence directly on the sensor, at any position of the signal line or on the processing system VEGAMET/ VEGALOG.
Note: The adjustment with the PC via the interface adapter VEGACONNECT 2 directly on the sensor or on the signal line only allows the "
Parameter adjustment
the adjustment with the PC on VEGAMET or VEGALOG on the following pages. Adjust­ments for " direct connection of VEGACONNECT 2 to the sensor or the signal line. Therefore this is described separately in this operating in­struction.
Configuration
" and corresponds to
" are not possible with
Adjustment module MINICOM
With the adjustment module MINICOM you adjust in the sensor or in the external indicat­ing instrument VEGADIS 50. The adjustment module ensures with the 6-key adjustment field with text display the parameter adjust­ment in the same functional spectrum like the adjustment program VVO or the 6-key ad­justment field with text display on the signal conditioning instrument VEGAMET.
The configuration adjustments however are only possible with the adjustment program VVO or the 6-key adjustment field with text display on the signal conditioning instrument VEGAMET.
Independent whether you carry out the set­up of a meas. system (consisting of sensor and signal conditioning instrument VEGAMET or sensor and processing system VEGALOG) with the adjustment software VVO, with the signal conditioning instrument or with the adjustment module MINICOM, the way is always the same:
- Configure first a meas. system in the menu "
Configuration
- carry out the parameter adjustment of the sensors in the menu "
" and
Instrument data
".
VEGASON 51 V … 53 V 41
Configuration means to coordinate or deter ­mine once. In the menu Configuration the signal conditioning instrument is informed about the general configuration:
- which kind of sensor is connected (radar, ultrasonic, process pressure…),
- what kind of parameter or application is concerned (level, gauge, distance…)
- to which input the sensor is connected
- which outputs (current, voltage, relay, fault signals, indication…) should be coordi­nated to which input in which way (in­verted, threshold value controlled…).
After these adjustments (configuration) had been carried out, the meas. system goes to operating condition and the signal condition­ing instrument will display a measured value. Now the sensor parameter adjustment (ad­justment, unit, linearisation curve, sensor adaption…) can be carried out.
Parameter adjustment means to enter values. Parameters are entered in the signal condi­tioning instrument as well as in the connected sensors. For example:
- Min. and max. adjustment
- Meas. range limits
- Physical unit, decimal point
- Linearisation curves
- Integration time
- Meas. environment (solid, liquid, foam generation, operating range…)
- False echo memory
- Inversion of the measured value etc.
Now all required adjustments for a precisely adapted sensor are carried out for a reliable measurement.
Set-up - Adjustment with PC on VEGAMET

6.2 Adjustment with PC on VEGAMET

The PC with the adjustment program VVO (VEGA Visual Operating) can be connected:
- to the sensor
- to the signal line
- to the signal conditioning instrument
VEGAMET 514V/515V
- to the processing system VEGALOG 571
For connection of a PC to a signal condition­ing instrument you require the interface adapter VEGACONNECT 2. The PC commu­nicates via the interface adapter VEGACONNECT 2 with the signal condition­ing instrument and the sensor. Hence a digital adjustment signal is superim­posed to the signal and supply line between sensor and signal conditioning instrument. In chapter "2.2 Configuration of measuring sys­tems" the connection of the PC in different coordinations is shown.
Before starting the set-up:
Take some time to carry out the set-up step­by-step with the PC and soon you will no more require the following pages.
In any case you are requested to enter or enquire something, this is marked in the following with a dot, such as e.g.:
• Choose …
• Start …
• Click to …
Now start to
• connect the standard plug of
VEGACONNECT 2 (9-pole) with the inter­face COM 1 or COM 2 of your PC.
42 VEGASON 51 V … 53 V
Set-up - Adjustment with PC on VEGAMET
• Plug the two small pin plugs of VEGACONNECT 2 into the CONNECT­socket of the signal conditioning instru­ment.
• Now switch on the voltage supply of the signal conditioning instrument.
After approx. 1…2 minutes (selfcheck) the meas. system is generally ready for opera­tion and indicates measured values.
• Now start the adjustment software VVO on your PC.
• Choose in the entrance screen with the arrow keys or the mouse the item "
ning
" and click to "OK".
In the next window you are asked for the user identification.
Plan-
The adjustment program, just called VVO, gets in contact with connected coordination/
sensor...... and indicates after a few seconds
if and with which coordination/sensor a con­nection exists. The preadjusted identification can be modi­fied later in the menu "User access".
Note:
If you do not get a sensor connection, please check:
- is the sensor fed with power supply (min. 20 V)?
- do you use an older VEGACONNECT instead of the new VEGACONNECT 2 ?
If VVO (adjustment software) gets in contact with the signal conditioning instrument for the first time, you are asked if the data should be transferred from the signal conditioning in­strument to the PC.
• Click to "
In the following menu window "
choice
• Enter under the name "
• Also enter under the password
VEGASON 51 V … 53 V 43
VEGA
".
"
VEGA
".
or keep the suggested file name.
Yes
"
DISBUS-group
" you can give a name to the database
Set-up - Adjustment with PC on VEGAMET
Configuration
Create new measurement loop
• Click to "OK", and you are in the main menu window.
Before starting the configuration:
The signal conditioning instruments are preconfigured dependent on the sensor type you have ordered with the signal conditioning instrument.
Generally you will use a preconfigured signal conditioning instrument. Hence you normally do not have to make any adjustments in the menu " ning on this page and in this case you can directly choose the menu "Parameter adjust­ment" (on page 46).
Configuration
" begin-
• Choose the menu "
ment loop/Create new
menu window
loop - Application"
• Choose the parameter "
ment
" (gauge or distance) and the sensor
type ("
pulse echo
• Click to "
Continue
Configuration/Measure-
" and you are in the
"Create new measurement
Level measure-
" for ultrasonic).
"
Nevertheless the most important steps in the menu "
Configuration
exception your signal conditioning instruments are not preconfigured, then start with the subchapter " continue with the adjustments in the subchapter "
Parameter adjustment
44 VEGASON 51 V … 53 V
" will be shown. If as an
Configuration
" on this page and
" on page 46.
• Choose "Standard level measurement" and "
no options
• Click to " "New application - select meas. loop" opens.
"
Continue
", and the menu window
Set-up - Adjustment with PC on VEGAMET
• Choose one of the two inputs of the signal conditioning instrument VEGAMET (VEGAMET 514 V has just one sensor input) and click to "OK"
After a few seconds the menu window "Cre­ate new measurement loop - Sensor configu­ration" opens.
• Confirm with "OK".
• Click in the menu window "Sensor coordi­nation" again to "OK".
You are again in the menu window "Create new measurement loop - Sensor configura­tion"
• Click in the menu window "Create new measurement loop - Sensor configuration" to "
Sensor coordination
"
Sensor coordination
• Click to "
• Click to "
• Choose the serial number of the sensor which you want to coordinate e.g. to input 1.
Sensor search
Input
"
". The menu window
" opens.
"
• Click to "
• Click in the menu window
Continue
"…
"Create new measurement loop - Measurement loop designation
" to "
Level
".
VEGASON 51 V … 53 V 45
• Enter in the menu window "Create new measurement loop - Measurement loop designation" a measurement loop number and a designation.
Set-up - Adjustment with PC on VEGAMET
Parameter adjustment
In the menu "
justment
adjustments.
Instrument data/Parameter ad-
" you carry out all important sensor
In this menu window you can choose with which output signals your level should be provided, e.g. as current, voltage, relay signal etc.
• Confirm with "OK", and the previous menu window appears again.
• Click to " adjustments are transmitted.
Configuration information
In the menu "
tem
Quit
", and wait a moment until the
Configuration/Measuring sys-
" you can check the configuration.
Adjustment
• Choose the menu "
Instrument data/Parameter adjustment
and then the sensor for which you want to carry out the parameter adjustment.
When you have configured just one sensor on the signal conditioning instrument, natu­rally only one sensor will be offered as choice.
",
In the heading of the menu window which is going to open, you now see the previously entered measurement loop name and the designation.
• First choose "
• Click in the menu window "Adjustment" to "
Min / Max - Adjustment
46 VEGASON 51 V … 53 V
Adjustment
"
"
Set-up - Adjustment with PC on VEGAMET
You can carry out the min./max. adjustment "
with medium
definite level) or considering the definite level, also with empty vessel).
Generally the adjustment is carried out with­out medium whereby during adjustment you are completely independent from the actual vessel filling. If you want to carry out the adjustment with medium, you have to carry out the min. ad­justment with emptied (also partly emptied) vessel and the max. adjustment with filled (also partly filled) vessel. It is hence comfortable and quick to carry out the adjustment without medium as shown in the example.
" (adjustment by means of a
"without medium"
(without
In the example the 0 % filling corresponds to a product distance of 3,400 m and the 100 % filling to a product distance of 0,500 m.
Note:
The sensor can only detect levels within the defined operating range. For level detection outside the operating range, the operating range must be corrected appropriately in the menu "S
ment
• Confirm your adjustments with "OK" and
• Click in the menu window "Adjustment" to
You are now again in the menu window "
strument data parameter adjustment
ensor optimisation/Meas. environ-
" on page 52/53.
you are now again in the menu window "
Adjustment
"
Quit
".
".
".
In-
• Choose "
• Choose in the following window if you want
• Enter a distance for the upper and lower
VEGASON 51 V … 53 V 47
no (adjustment without medium)
to make the adjustment in meters (m) or in
feet
(ft).
level and the appropriate filling degree in %.
"
Set-up - Adjustment with PC on VEGAMET
Hence the sensor electronics has two char­acteristics points (MIN and MAX), out of which a linear proportionality between prod­uct distance and the percentage filling of a vessel is generated. However the characteristics points must not be at 0 % and 100 % but the difference should be as large as possible (e.g. at 20 % and at 80 %). The min. product distance of the characteristics point for min./max. adjust­ment should be 50 mm. When the character­istics points are too close together, the possi­ble meas. error increases. It is hence suitable to carry out the adjust­ment at 0 % and at 100 %.
In the menu "
justment/Conditioning/Linearisation
adjust later another linear dependence be­tween product distance and the percentage filling degree (see later subpoint linearisation).
Conditioning
• Click in the menu window "
parameter adjustment
The menu window "
Instrument data/Parameter ad-
" you can
Instrument data
" to "
Conditioning
Conditioning
" opens.
".
• Click to "
In the menu " 0 % and 100 %-values of the parameter and their unit. Hence you inform the sensor, e.g. that at 0 % filling there are 45,5 litres and at 100 % filling 1200 litres in the vessel. The sensor indication then displays with empty vessel 45,5 litres (0 %) and at full vessel 1200 litres (100 %).
Scaling
Scaling
".
" you enter the actual
48 VEGASON 51 V … 53 V
Set-up - Adjustment with PC on VEGAMET
As parameter you can choose (figures), and coordinate an appropriate meas. unit (e.g. l, hl). The sensor indication displays then the figure in the chosen parameter and unit.
• Save the adjustments in the menu "
The adjustments are now transmitted to the sensor and you are again in the menu win­dow "
Linearisation
The relation of level and volume is described with so called linearisation curves. If in your vessel there is another than a linear dependence between level ("percentage value" of the level) and the volume ("linearised" value of the volume), choose the menu "
ment/Conditioning
volume, mass, height
with "OK".
Conditioning
".
Instrument data/Parameter adjust-
"
"dimensionless
and
distance"
Scaling
• Click in the menu window " the menu point "
"
In the menu window " that "
Linear
the dependence between the percentage value of the filling volume and the value of the level is linear. Beside the two given linearisation curves "
Cylindrical tank
can also adjust six "
curves
".
• To enter an own vessel geometry or a user programmable filling curve, click to "
Linearisation
" is preadjusted. This means that
" and "
user programmable
programmble curve
• Click to "
Edit
".
Conditioning
".
Linearisation
Spherical tank
".
" you see
" to
" you
User
VEGASON 51 V … 53 V 49
Set-up - Adjustment with PC on VEGAMET
First of all a linear relation (a straight line) is displayed.
In the field "
Transfer measured value
" the actual level in percent of the adjusted meas. range (meas. window) is displayed. You have ad­justed the meas. range during the min./max. adjustment to 0,500 … 3,400 m.
Max.
Min.
100 % (0,500 m) correspond to 1200 litres
Meas. range
0 % (3,400 m) correspond to
45,5 litres
The user programmable linearisation curve is generated out of linearisation points, the so called value pairs. A value pair consists of "
Linearised”
and "
(percentage value of the filling)
Percentage value
" (percentage value of the level relating to the meas. range). If the linearisation points or value pairs of your vessel are not known to you, you must gauge the vessel by litres.
Linearisation point 0 is at 0 % filling (
age value [%]
), which corresponds to an
percent-
actual distance to the product surface of 3,400 m in the example (empty vessel) . The volume value is hence 45,5 litres (rest filling of the vessel).
Linearisation point 1 is at a filling level of 20 % (20 % of the meas. distance of 0,500 m … 3,400 m). In our example there are 100 litres at 20 % filling in the vessel. Linearisation point 2 is at a filling level of 40 %. At this filling level there are 250 litres in the vessel. Linearisation point 3 is at a filling level of 80 %, where 1000 litres are in the vessel. Linearisation point 4 is at a filling level of 100 % (level distance 0,500 m), where 1200 litres are in the vessel.
Gauging by litres
In the characteristics of the following figure you see five linearisation points (0, 1, 2, 3, and 4) or value pairs. Between the linearisation points there is always a linear interpolation. Click to "
Indication in scaled values
", to have
the adjusted meas. unit indicated on the y-
You can enter max. 32 linearisation points (value pairs) per linearisation curve.
• Quit the menu with "OK".
• Confirm the message with "OK", and your individual linearisation curve will be saved in the sensor.
axis (left bottom part of the menu window).
50 VEGASON 51 V … 53 V
Set-up - Adjustment with PC on VEGAMET
Back to the menu window " can enter a measured value integration with the menu point " ful for fluctuating product surfaces to avoid that the measured value indication and output changes permanently. As a standard feature an integration time of 0 seconds is adjusted.
• Quit the menu with "OK" and you are now again in the menu window "Instrument data parameter adjustment".
• Quit the menu window with "OK".
Outputs
• Choose "
justment
• Choose in the menu window "
Integration time
Instrument data parameter ad-
".
data parameter adjustment
"
Outputs
".
Conditioning
". This is use-
Instrument
" the menu point
" you
• Click to "
• If you have carried out adjustments in this
• If you want to keep the adjustments without
You are again in the menu window "
• Click in the menu window "
Current output
mode of the 0/4…20 mA output signal.
menu window, click to " saving, click to "
Quit
" to adjust the signal
Save
".
".
Outputs
Outputs
menu point " choose "
Sensor display
Display of measured value
".
" to the
", and
".
• Choose under "Sensor No." "A" and under "Parameter" "Scaled".
You are in the menu window "
VEGASON 51 V … 53 V 51
Outputs
".
• Choose " cation should display e.g. litres. In the example a level of 45 … 1200 litres would be displayed.
• Enter the meas. distance in metres which you have entered in the min./max. adjust­ment and the appropriate litre information corresponding to the min. and the max. value.
scaled
", "
Volume
" when the indi-
Set-up - Adjustment with PC on VEGAMET
Sensor optimisation
In the menu " carry out special optimizing adjustments of the sensors and can optimize e.g. by means of the echo curve the installation position of the sensor.
Meas. environment
• Choose the menu "
Instrument data/Parameter adjustment
• Choose in the menu window "
data parameter adjustment
"
Sensor optimisation
A
"
• First click to "
Sensor optimisation
Instrument
" the menu point
", and click to "
Meas. environment
" you can
"
Sensor
".
• Click to
• Click in the window "
• Click in the menu window "Display of meas-
You are now again in the menu window "
puts
• Click to
You are again in the menu window "
ment data parameter adjustment
52 VEGASON 51 V … 53 V
"
Save
"
"Quit"
ured value” to "Quit".
"
"
Quit
"
Outputs
" in the menu window
"
Sensor display
".
" to
Out-
Instru-
Set-up - Adjustment with PC on VEGAMET
The window "
With the menu point " can define the measuring range of the sensor deviating from the " As a standard feature the measuring range normally corresponds to the min./max. ad­justment.
Generally it will be suitable to choose the measuring range approx. 5 % larger than the range determined by the min./max. adjust­ment. In the example the min. adjustment was at 0,500 m and the max. adjustment at 3,400 m. In the example you would adjust the measur­ing range to 0,4 … 3,5 m.
Meas. environment
Measuring range
Min/Max-Adjustment
" opens
" you
".
• Click in the menu window "
ment
" to "
• In the menu window "
• Confirm with "OK". After some seconds of saving during which the adjustments are saved in the sensor, you are again in the window "
• Click in the menu window "
You are again in the menu window "
optimisation
Echo curve
With the menu point " look at the course and the strength of the detected ultrasonic echo.
Measuring condition
you click to the options corresponding to your application.
ment
" to "
Quit
".
".
Meas. environ-
Measuring conditions
".
Meas. environment
Meas. environ-
Sensor
Echo curve
" you can
"
".
If due to vessel installations, strong false echoes have to be expected, a correction of the installation position (if possible) can help to localize and reduce the size of the false echo by monitoring the echo curve.
• Save the adjustments with "OK"
VEGASON 51 V … 53 V 53
Set-up - Adjustment with PC on VEGAMET
In the following figure you see the echo curve before correction of the installation angle (directing to the product surface) with a false echo which has nearly the same size than the product echo and which is caused by a strut.
In the next figure you see the echo curve after optimum directing of the sensor to the product surface (sensor axis vertically to the product surface).
With the menu point " the menu window " can cause the sensor to mark false echoes. The sensor electronics then saves the false echoes in an internal database and treats the false echoes at a poorer grade than the use­ful echo.
• Click in the menu window "
optimisation
storage
".
• Click in the menu window "
age
" to "
Learn false echoes
window "Learn false echoes" opens.
False echo storage
Sensor optimisation
Sensor
" to the menu point "
False echo stor-
" in
" you
False echo
". The small
You see that the false echo caused by the strut is here reduced by approx. 20 dB against the useful echo and hence can no more influence the measurement.
• Quit the menu window "Echo curve" with "
Quit
".
54 VEGASON 51 V … 53 V
• Enter here the checked product distance and click to "
Hence you cause the sensor to mark all ech­oes in front of the product echo als false echoes. This avoids that the sensor detects erroneously a false echo as level echo.
Create new
".
Set-up - Adjustment with PC on VEGAMET
• Click to "
The echo curve and the false echo marking is displayed.
• Quit the menu with " You are again in the menu window "
optimisation
With the menu point " options of the menu " basic adjustment.
• Quit the menu window "Sensor
You are then in the menu window "
data parameter adjustment
Show echo curve
".
optimisation" with " "
Quit
" the menu window "
sensor optimisation
"
Quit
".
Sensor
Reset
" you reset all
Sensor optimisation
Quit
" and with
Sensor choice for
"
Instrument
".
" to
• Click to "
Application
"
VEGAMET
your measuring system in the information windows.
", to get detailled information on
", to "
Input No. A
" and
• Click in the window "Instrument data pa­rameter adjustment" to "Meas. Loop Data".
VEGASON 51 V … 53 V 55
• Close the information windows.
• Quit the menu "
• Click in the menu window "Instrument data parameter adjustment" to "
You are now again in the main menu window.
Change COM-interface
In the menu "
nication
" you can adjust the interface param­eters of your PC or change the used COM­Port.
Meas. Loop Data
Quit
".
".
Configuration/Program/Commu-
Set-up - Adjustment with PC on VEGAMET
Display of measured value
• Click in the main menu window to the menu "
Display/Display of measured value
choose the measurement loop (in the ex­ample only one is available).
", and
• Choose in the line "Indication value" "
Scaled
played e.g. in litres and volume percent as
56 VEGASON 51 V … 53 V
well as the actual signal current in the 0/4 ... 20 mA signal line.
" and the measured value is dis-
Set-up - Adjustment with PC on VEGAMET
Simulation
• Click to the menu " and choose the measurement loop (in the example only one is available).
Diagnostics/Simulation
"
• Click in the turquoise window cutout to "
Start
".
The menu window " which is similar to the previous menu window opens. In this menu window however you can simulate the filling of the vessel or the signal current and the indication to an individual value (measured value simulation). First of all the actual measured value and the signal current are displayed.
VEGASON 51 V … 53 V 57
Simulation of outputs
",
The grey scrollbar in the turquoise window cutout becomes active. With this scollbar you can modify the measured value in the range of -10 % … 110 % and hence simulate the filling or emptying of your vessel. In the field with the figures above the scroll­bar you can enter an individual %-value for the filling degree.
Backup
• Click to "
Services/Backup/Sensors
Set-up - Adjustment with PC on VEGAMET
"
Saved sensor data can be transferred later to other sensors. If you have e.g. a system with several of the same storage vessels and identical sensors, it is sufficient to configure one sensor and then transfer to the other sensors.
In the menu window " the serial number is displayed. You can save the sensor individually or in groups with all adjustments in a directory of your choice on your PC. In addition you can add a short note to each backup.
You also save the adjustments of the signal conditioning instrument VEGAMET:
Backup
" the sensor with
• Choose the menu "
figuration/Sensors
In this menu window you choose in the yellow window cutout a sensor (or a signal condi­tioning instrument) which should be overwritten by the adjustments of another sensor.
Choose in the field " serial number of the sensor of which you want to transfer the adjustments. With "
store to
" you transfer these sensor adjust­ments to the sensor you have chosen in the yellow window cutout.
Services/Restore con-
"
Selection of backups
" the
Re-
58 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM

6.3 Adjustment with VEGAMET or MINICOM

Beside the PC you can also adjust the ultra­sonic sensors VEGASON 51 V … VEGASON 53 V also
- with signal conditioning instrument
- or with the small, detachable adjustment module MINICOM.
All adjustment options are available for the adjustment with signal conditioning instru- ment VEGAMET like with PC. The adjustment just differs in the demonstration however not in the function.
With the adjustment module MINICOM all sensor relevant adjustments (adjustment, meas. range, meas. conditionins, sensor­indication-scaling or false echo memory) are available. However not possible are the adjustment steps relating to the configuration of the sig­nal conditioning instrument VEGAMET or the processing system VEGALOG and their signal processing (configuraton of inputs and outputs, linearisation curves, simulation…).
The adjustment of VEGAMET and MINICOM is identical. They are adjusted with 6 keys. A small display shows you beside the meas­ured value, in short words a message on the menu point or on a figure of a menu adjust­ment. The information quantity of the small display however cannot be compared with the ad­justment program VVO, however in conjunc­tion with the following menu plan of MINICOM and VEGAMET 515 V you will have no problems with the adjustment. Some users however will find the work with the small adjustment field quicker and more direct to make than with the PC. The menu plan to VEGAMET 514 V is pro­vided in the operating instruction of VEGAMET 514 V.
Indication and adjustment surfaces
Signal conditioning in­strument VEGAMET
Display,
Branch, i.e. jump to the lower menu with [OK]
%
100
+
ESC
OK
CONNECT
2
1
on
515 V
Interrupt the adjust­ment or jump to the upper menu
Analogue LED-indication (0 … 100 %)
VEGASON 51 V … 53 V 59
indication of
- measured value
- menu point
- parameter
- value
Acc. to the menu point, change value or choose out of a list
Choose menu window or shift flashing cursor
Save the adjusted value or change to the lower menu
Adjustment module MINICOM
Adjustment structure
Set-up - Adjustment with VEGAMET or MINICOM
Measured value indi-
TAG1
OK
36.9
TAG2
%
TAG3
%
%
cation
Main menu
Param. TAG1
OK
ESC
ESC
ESC
Adjust-
OK
ment
OK
Param. TAG2
ioning
Scal­ing
Param. TAG3
OutputCondit
Lin. curve
The most important adjustment steps
On page 62 you see the complete menu plan of the signal conditioning instrument VEGAMET 515 V as well as the adjustment module MINICOM. The menu plan of the signal conditioning instrument VEGAMET 514V is nearly identical (see operating in­struction to signal conditioning instruments VEGAMET 514 V).
Set-up the sensor in the following numbered sequence. The appropriate numbers are stated in the menu plans on page 64 … 71.
The items in brackets can be carried out or adjusted beside the signal conditioning in­strument VEGAMET also with the adjustment module MINICOM.
1. Configuration of the input (generally al- ready preconfigured)
Menu plan page 66
MST1­2
Simu­lation
Integra­tion time
10
Confi­gura­tion
In­puts
Menu window Parameter Value
Selection of the menu window in the horizontal menu stage as well as choice of fixed parameters
Meas. loop
Out­puts
2.) Operating range
Menu plan page 70 (via I on page 66)
3.)
Measuring conditions
Menu plan page 70 (via I on page 66)
4. Min./Max. adjustment
Menu plan page 64
5.) Conditioning/Scaling VEGAMET
Menu plan page 64
6. Outputs
Menu plan page 64
7.) False echo storage (only required if meas. errors occur during operation)
Menu plan page 70
8.) Indication of the useful and noise level
Menu plan page 71
9. Conditioning/Scaling
of the sensor display
Menu plan page 70
Add’l func­tions
Pass­word off
60 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM
Following short explanations to the set-up steps 1 … 9.
1. Configuration of the input
First of all you have to inform the signal condi­tioning instrument (only VEGAMET 515 V) to which input (input 1 or input 2) the sensor is connected.
2.) Operating range
Without special adjustment, the operating range corresponds to the meas. range. The meas. range has already been adjusted with the min./max. adjustment. Generally it is use­ful to choose the operating range slightly bigger (approx. 5 %) than the meas. range. Example: Min./max. adjustment: 0,500 … 3,400 m; adjust operating range to approx. 0,400 … 3,500 m.
Adjustment without medium
Key adjustment Display indication
OK
Max.
Min.
100 %
(0,500 m product distance) correspond to1200 litres
Meas. range 0,5…3,4 m
0 % (3,400 m product distance) correspond to 45 litres
Sensor
m(d)
4.700
Param. adjust ment
Adjust
OK
ment
3.) Measuring conditions
see menu plan figure 3 on page 70.
4.) Adjustment
Under the menu point " the sensor in which meas. range it should operate.
Adjustment
" you inform
OK
OK
+
The adjustment can be carried out with and without medium. Generally you will carry out the adjustment without medium as you can adjust without filling cycle.
VEGASON 51 V … 53 V 61
Now you can enter with the "+" and "–" key the distance, which your sensor has to the product at 0 % filling (example: 5,850 m)
w.out medium
Adjust ment in
(Min. adjustment)
m(d)
0.0 %
at
m (d)
XX.XXX
The distance-indication flashes slowly
Set-up - Adjustment with VEGAMET or MINICOM
Adjustment with medium
with medium
Max.
Min.
set
set
at %
at %
XXX.X
XXX.X
Fill the vessel e.g. to 10 % and enter in the menu "
Min. set
" with the "+" and "–" keys 10 %. Then fill the vessel e.g. to 80 % or 100 % and enter in the menu "
Max. set
" with the
"+" and "–" keys 80 % or 100 %.
If you do not know the distance, you have to sound.
The indication stops flashing
OK
and the adjustment will be saved.
Hence you have adjusted the min. value.
100.0%
at
m (d)
XX.XXX
(Max. adjustment)
The max. adjustment is made like the min. adjustment with "+", "–" or "OK" (example: 1,270 m).
5.) Conditioning / Scaling VEGAMET
Under the menu point " you enter the figure which corresponds to a 0 % and 100 % filling and choose beside the position of the decimal point a physical unit, e.g. distance.
Condit ioning
Sca­ling
0 %
100 %
corres
corres
ponds
ponds
XXXX
Enter in the menu window " the value of the 0 %-filling (in the example of the adjustment with PC these were 45 litres). Confirm with "OK". With the key you change to the 100 % menu. Enter the value of your parameter corresponding to a 100 %-filling. In the examble this had been 1200 for 1200 litres. Confirm with "OK".
If necessary choose a decimal point. How­ever note that only max. 4 digits can be dis­played. In the menu "
prop. to
eter (mass, volume, distance…) and in the menu "
Unit
" the physical unit (kg, l, ft3, gal, m
…).
Conditioning/Scaling
prop.
Deci-
to
mal point
XXXX
888.8
Mass
0 % corresponds
" you choose the param-
Unit
Kg
"
"
3
62 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM
Linearisation:
Ab­gleich
Aus­wer­tung
Skalie rung
Lin. kurve
Linear
Integr ations zeit
0 s
A linear dependence between the percent­age value of the level distance and the per­centage value of the filling volume is preadjusted. You can choose in the menu "Lin.curve" be­tween linear, cylindrical tank and spherical tank. The adjustment of an own linearisation curve is only possible with the PC and the adjustment program VVO (see page 48).
6. Outputs
Under the menu " e.g. the current output should be inverted or which parameter should be provided by the sensor indication.
Outputs
" you determine if
8.) Useful and noise level
In the menu
you get an important information via the sig­nal quality of the product echo. The higher the amount out of " "
S-N
Ampl.: Means amplitude of the level echo S-N: Means Signal-Noise, i.e. the level of
The larger the distance of the amplitude (Ampl.) to the noise level (S-N), the better is your measurement: > 18 dB Meeasurement very good 18 … 13 dB Measurement good 13 … 8 dB Measurement satisfactory 8 … 5 dB Measurement sufficient < 5 dB Measurement very bad
Example:
Ampl. = 68 dB S-N = 53 dB
Ampl.:
XX dB
S-N:
XX
dB
Ampl
." minus
", the more reliable is the measurement.
in dB (useful level) the background noise (noise level)
68 dB – 53 dB = 15 dB
With 15 dB signal distance there is a high reliability.
7.) False echo storage
A false echo storage is always useful when false echo sources such as e.g. struts must be reduced. With the creation of a false echo storage you cause the sensor electronics to note the false echoes and to save them in an internal database. The sensor electronics treats these (false) echoes differently than the useful echo and gates them out.
VEGASON 51 V … 53 V 63
9. Conditioning / Scaling of the sensor display
The menu point "Conditioning" in the MINICOM-menu only relates to the SENSOR-display and will be overwritten by the adjustments in the signal condi tioning instrument.
Menu plan page 70
Set-up - Adjustment with VEGAMET or MINICOM
MST1-3 xx,x
xx,x
xx,x
TAG ­No. 1
%
Param.
TAG­No. 1
xx,x
TAG ­No. 2
%
xx,x
Param.
TAG­No. 2
AC
Adjust ment
like TAG-No.1
Signal condit ioning
TAG ­No. 3
%
xx,x
Param.
TAG­No. 3
Out­puts
mA out­puts
mA output no.1
Prop. to
Per­cent
6.
Unit
0,0%
Volt. out­puts
Volt. output no.1
Prop. to
Per­cent
mA output no.2
VEGAMET 515 V - Menu plan
Confi­gura­tion
B
Volt. output no.3
Volts at 0%
0,000
mA at 100%
20,000
see page 68-69
Fail­ure mode
0mA
see page 66-67
Unit
0,0%
mA Range
0% +100%
Volt. output no.2
Volt­Range
0% +100%
mA output no.3
mA output
4/20mA
mA at 0%
4,000
Volt­age output custom
Add´l func­tions
Volts at 100%
10,000
mA Limita tion
on
Fail ure mode
0V
Volt­limita tion
on
4.
with medium
Min-Ab gleich bei % 0,0
5.
Sca­ling
0% corres ponds 0
Max-Ab gleich bei % 100,0
100% corres ponds 1000
w.out medium
Adjust ment
bar
Deci­mal­point
888,8
Offset correc tion
Sensor unpres sur´d OK ?
Offset corr. Now! OK ?
Prop. to
unde­fined
0% at
bar
0,000
Unit
– –
100% at
bar
1,000
Lin. curve
linear
Integr ation time 0
Densi­ty in kg/ dm3
1,000
Meas. value limita tion
nega­tive Values Yes
Fail. mode
Stan­dard
64 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM
Bolt print are the sensor or measured value information which cannot be modified in this position.
The menu points in white letters can be modified with the or key abd confirmed with the key.
Light grey menu fields are only displayed if necessary (dependent on the adjustments in the other menus).
MET­Dis­play
Prop.
Unit
to
Per­cent
0,0%
PC/PCS out­puts
OK
DIS out­puts
DIS­outp. 1
Prop. to
Per­cent
Unit
0,0%
Simu­lation
Simu­lation Now! OK?
Simu­lation %
XX,X
Relay out­puts
Dis­tance m
X,XX
Fast change
Yes
High dust level
No
With these keys you move in the menu field to the left, right, top and bottom.
Spe­cial funct.
Real value corr.% 0
DIS­outp.7
DIS­outp.2
Reset
Level
Reset
OK? Reset
Now! OK?
Densi­ty corr.% 0
ESC
OK
Manual correc tion
Offset correc tion
Offset correc tion OK?
Correc tion Now! OK ?
Real value corr.
Real value corr. OK?
Correc tion % 0,0
Correc tion Now! OK ?
Deviat ion period
Scan time s 1
DC/PCS outp.7
No. of scans
10
Fail­ure mode off
Add´l func tions
Switch ing delay
ton
s
1
t off
s
1
Relay out­put 1
Prop. to
Per­cent
Relay out­put 2
like relay output 1
Unit
0,0%
Mode
Over­fill protec
PC/DCS outp.1
Monit. on
Low& High
Prop. to
Per­cent
Low %
0,0
Unit
0,0%
High %
100,0
PC/DCS outp.2
Deviat ion % 1,0
VEGASON 51 V … 53 V 65
Set-up - Adjustment with VEGAMET or MINICOM
MST1-3 xx,x
xx,x
xx,x
TAG ­No. 1
%
A
Config Inputs
xx,x
Param.
TAG­No. 1
see page 64-65
TAG ­No. 2
%
xx,x
Param.
TAG­No. 2
Config meas. loop
Single measurement
TAG 1
Level
TAG ­No. 3
%
xx,x
Param.
TAG­No. 3
Combi application
Appli­cation
level
1)
1)
The parameters in these menu points can be only modified when a "Reset to combi application" was carried out before.
TAG 2
Level
like TAG 1
Sensor type
Hydro­static
1)
Mode
press­urized vessel
1) 1)
VEGAMET 515 V - Menu plan
B
Confi­gura­tion
Option
no option
Sensor coordi nation
Loca­tion A
Input no.1
Add´l func­tions
C
see page 68-69
TAG 1
Level
TAG-ID
TAG ­No. 1
Loca­tion B
Input no.2
Fault signal ?
on
Loca­tion C
Input no.3
TAG 2
Top press.
like TAG 1
T are
Input from
unde­fined
Loca­tion D
unde­fined
TAG 3
Total press.
Loca­tion E
unde­fined
Moni­toring
Input from
unde­fined
Appli-
Sensor
1.
cation
type
Hydro-
level
static
2)
2) 2)
The parameters in these menu points can only be modified when a "Reset to single measurement" was carried out before.
Edit ser.no
0000 0000
Input no. 1
Input from
local MET
2)
Serial no.
xxxx
xxxx
Mode
stan­dard
Input no.
unde­fined
Option
no option
2)
Sensor coordi nation
Loca­tion A
Input no.1
Sensor charac terist ics
Min. meas. range
0,00
Loca­tion D
unde­fined
Max. meas. range
1,00
Loca­tion E
unde­fined
2.
TAG-ID
TAG ­No. 1
Sensor opti­mize
Fault signal ?
on
Tare
Input from
unde­fined
I
continue in the MINICOM-menu
on page 70
Moni­toring
Input from
unde­fined
66 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM
Config out­puts
Config curr. output
Dis­tance m
X,XX
Fast change
Yes
High dust level
No
Config volt. output
Bolt print are the sensor or measured value information which cannot be modified in this position.
The menu points in white letters can be modified with the or key abd confirmed with the key.
Light grey menu fields are only displayed if necessary (dependent on the adjustments in the other menus).
Config relay output
Operat
Fail
ing
safe
relay
relay
Relay
Stan­dard
Rel. 1
Rel. 1 to
Stan-
TAG -
dard
No. 1
OK
Input no.
unde­fined
Config PC/DCS output
PC/DCS Meas. values
Rel. 2 to
TAG ­No. 2
With these keys you move in the menu field to the left, right, top and bottom.
PC/DCS Relay status
off
DCS 1 to
TAG ­No. 1
Rel. 2
Input no.
Stan-
unde-
dard
fined
PC/DCS Input status
off
DCS 7 to
----
Config VEGA­DIS
DIS 1 to
TAG ­No. 1
ESC
OK
DIS 7 to
----
V no2
mA no1 to
TAG ­No. 1
Input no. 2
like input no. 1
mA no2 to
TAG ­No. 2
V no1 to
TAG ­No. 1
mA no3 to
TAG ­No. 3
to
TAG ­No. 2
V no3 to
TAG ­No. 3
Input no. 4
Input from
local MET
Input no. 5
like input no. 4
Chan­nel no. K1
Input no.
unde­fined
Autom. sensor search
Sensor search
OK ? Sensor
search Now! OK ?
VEGASON 51 V … 53 V 67
Set-up - Adjustment with VEGAMET or MINICOM
MST1-3 xx,x
xx,x
xx,x
A
TAG ­No. 1
%
xx,x
Param.
TAG­No. 1
see page 64 -65
TAG ­No. 2
%
xx,x
Param.
TAG­No. 2
Pass­word
off
TAG ­No. 3
%
xx,x
Param.
TAG­No. 3
Edit lin. curves
Info
Input info
VEGAMET 515 V - Menu plan
VEGA­MET Info
Type MET515
V
Confi­gura tion
B
see page 66-67
Pro­gram­info
Pro­gram info
xxxxxx
Serial number
xxxx
xxxx
Instr. addr.
1
Meas. loop info
TAG 1
TAG ­No. 1
Min set at %
0,0
Softw. Vers.
01.13 97
C
Add´l func­tions
Min set at
bar
0,000
TAG 2
TAG ­No. 2
Softw. Date
47/97
Lang­uage
Eng­lish
like TAG 1
Max set at %
100,0
Param. Vers.
5
TAG 3
TAG ­No. 3
Max set at
bar
1,000
Para­meter
Level
Curve no.1
Add Lin.­point
x % 0,0 y V% 0,0
Input no.1
II
continue in the
MINICOM-menu
on page 71
Edit curve no.1
x 0 %
0,0
y 0 V% 0,0
Input no.2
like input 1
Curve no.2
x 1 %
100,0
y 1 V% 100,0
Lin. point numbers (0...32)
Curve no.3
like LIN-curve 1
x 2 %
100,0
y 2 V% 100,0
Input no.4
Sensor­type
Vibra­tion
x32 % y32 V%
100,0
100,0
Input from
local MET
Chan­nel no.
K1
Input no.5
like input 4
Input no.
unde­fined
Delete Lin.­point
x 0,0 y 0,0
delete
Delete now ?
Actual switch status
68 VEGASON 51 V … 53 V
Set-up - Adjustment with VEGAMET or MINICOM
Reset VEGA­MET
Reset confi­gura­tion
TAG 1 to De­fault
Reset
OK?
Reset Now!
OK?
Dis-
tance
m
X,XX
Fast change
Yes
High dust level
No
Ser­vice
only accessible with service password
TAG 2 to De­fault
Reset
OK? Reset
Now! OK?
Bolt print are the sensor or measured value information which cannot be modified in this position.
The menu points in white letters can be modified with the or key abd confirmed with the key.
Light grey menu fields are only displayed if necessary (dependent on the adjustments in the other menus).
Reset sensor charac
Not available with VEGASON 51 V … 53 V
to linked appli­cat.
Delete all TAG´s?
Reset
OK? Reset
Now! OK?
to single meas.
Reset TAG 1
Delete TAG 1?
Reset
OK? Reset
Now! OK?
Reset TAG 2
Delete TAG 2?
Reset
OK? Reset
Now! OK?
OK
With these keys you move in the menu field to the left, right, top and bottom. .
Reset lin. curves
Reset curve no.1
Reset
OK?
Reset Now!
OK?
Reset curve no.2
Reset
OK?
Reset Now!
OK?
ESC
OK
Reset curve no.3
Reset
OK? Reset
Now! OK?
Reset all curves
Delete all curves ?
Reset
OK?
Reset Now!
OK?
VEGASON 51 V … 53 V 69
MINICOM - Menu plan
Set-up - Adjustment with VEGAMET or MINICOM
Sensor
m(d)
4,700
Para­meter
Sensor opti­mise
Sensor Tag
Sensor
2.
Sensor opti­mise
Meas. Enviro nment
Opera­ting range
Begin
m (d)
0.50
SON 51 V
1.00
When switching on the power supply, the sensor type and the software version are displayed for a few seconds.
I
f
rom the menu of the signal conditioning instrument VEGAMET on page 66
Meas. Condiat ions
End
m (d)
6.00
Measur ing in gas
Meas. Dist.
m (d)
4.700
Correc tion Now! OK?
7.
False echo memory
Create new
Meas. Dist.
m (d)
X.XX
echo learn Now! OK ?
Update
Meas. Dist.
m (d)
X.XX
Update Now!
OK ?
Delete
Delete Now!
OK ?
delet­ing
3.
Condit ion
Liquid
Fast change
No
agitat ed sur face
No
Foam­ing prod.
No
learn­ing
High dust level
No
learn­ing
Large angle repose
No
9.
Displ. adjust ment
0.0 %
at
m (d)
XX.XXX
100.0%
at
m (d)
XX.XXX
Signal Condit ioning
Sca­ling
0 % corres ponds
XXXX
100 % corres ponds
XXXX
Deci­mal­point
888.8
Prop. to
Volume
Unit
Kg
70 VEGASON 51 V … 53 V
Out­puts
Sensor Dis­play
Prop. to
Scaled
Set-up - Adjustment with VEGAMET or MINICOM
Basic Reset
Reset Now!
OK ?
Reset­ing
8.
Dis­tance
m (d)
4.700
Ampl.:
XX dB
S-N:
XX
II
from the menu of the signal conditioning instrument VEGAMET on page 68
Dis­tance m
X,XX
Fast change
Basic Reset
Reset Now!
OK ?
Reset­ing
Yes
High dust level
No
Lan­guage
Eng­lish
Add´l func­tions
Info
Pass­word
dB
Input no. 1
Off
Bolt print are the sensor or measured value information which cannot be modified in this position.
The menu points in white letters can be modified with the or key abd confirmed with the key.
Light grey menu fields are only displayed if necessary (dependent on the adjustments in the other menus).
Meas. Unit
m (d)
OK
ESC
OK
With these keys you move in the menu field to the left, right, top and bottom.
Sensor Tag
Sensor
Sensor type
PULS 52 V
Serial number
1094 0218
Softw. Vers.
1.00
Softw. Date
10.09. 1997
max. range
m (d)
7.000
Dis­tance
m (d)
4.700
Ampl.:
XX dB
S-N:
XX
dB
VEGASON 51 V … 53 V 71

6.4 Adjustment with PC on VEGALOG

For connection of the PC to the processing system VEGALOG you require a standard interface RS232 DTE-DTE (Data Terminal Equipment) interface cable. With this cable you connect the PC with the processing system.
DTE DTE
DCD 1 1 DCD RxD 2 2 RxD TxD 3 3 TxD DTR 4 4 DTR GND 5 5 GND
--- 6 6 ---
--- 7 7 ---
--- 8 8 ---
--- 9 9 ---
Via the signal and supply lines between sen­sor and processing system also adjustment signals are transmitted digitally beside the measured values. The adjustment program VVO can then communicate with the processing system and all connected sen­sors. In chapter "2.2 Configuration of measur­ing systems" the connection of the PC to the various meas. systems is shown.
Set-up - Adjustment with PC on VEGALOG
The first set-up steps with the PC in conjunc­tion with the processing system VEGALOG correspond to the adjustment in chapter "
Adjustment with VEGAMET or MINICOM
6.2
".
• Connect the standard output of your PC with the standard RS232-interlink cable (9­pole) with the processing system VEGALOG.
• Now switch on the power supply of the processing system.
After approx. 1…2 minutes (selfcheck) the meas. system consisting out of processing system and sensors is normally ready for operation and the sensors indicate measured values.
• Now start the adjustment software VVO on your PC.
Before starting with the set-up:
The many pictures, adjustment steps and menus may be confusing but just carry out the set-up with the PC step-by-step and you will soon no more require the following pages.
When you want to enter or enquire something this is marked in the following with a dot, e.g.
• Choose …
• Start …
• Click to …
72 VEGASON 51 V … 53 V
• Choose in the initial screen with the arrow keys or the mouse the point "
Planning
" and
click to "OK".
In the next window you are asked for the identification.
Set-up - Adjustment with PC on VEGALOG
When you connect VVO to a VEGALOG where data have been already saved, a message is displayed whether you want to transfer the saved data from the PC to VEGALOG or whether the data of VEGALOG should be transferred to the PC.
• Enter under name "
• Also enter under password
VEGA
".
"
VEGA
".
The adjustment program, called in the follow­ing VVO, contacts the connected Meas. system/sensor …… and shows after a few seconds if and with which Meas. system/ Sensors a connection exists.
Note:
If you do not get a sensor connection, check the following:
- is the processing system powered with supply voltage?
- do you erroneously use a wrong RS232­cable instead of the VEGA RS-232-connec­tion cable?
If VVO (adjustment software) contacts the processing system for the first time, you are asked if the data should be transferred from the processing system to the PC.
• Click to "
Yes
"
• Click to "OK", and you are in the main menu window.
The preadjusted identification can be modi­fied later in the menu "
User access
" (see page 43 right column).
Configuration/Program/
VEGASON 51 V … 53 V 73
Configuration
The adjustment of an ultrasonic sensor on a VEGALOG with the PC generally correponds to the adjustment of an ultrasonic sensor on the signal conditioning instrument, like in chapter "
VEGAMET
The difference is just in the higher configura­tion requirement of VEGALOG. A number of possible sensor inputs and signal outputs as well as various processing routines must be coordinated. First of all it must be defined which sensor (input) should be processed where and how.
Configuration Info
• Choose the menu "
6.2 Adjustment with PC on
".
Configuration/Measur-
ing system
"
Set-up - Adjustment with PC on VEGALOG
• Click in the card info window of your EV­card (input card VBUS) to "
Measuring system
You get a survey of the sensors connected to the card.
You are in the menu "
system
", where a picture of the connected
VEGALOG is shown on the PC.
• Click to an individual card number and you get in the window a "
74 VEGASON 51 V … 53 V
Configuration/Measuring
Card-Info
".
Configuration/
".
Set-up - Adjustment with PC on VEGALOG
Create new measurement loop
• Choose the menu "
ment loop/New
Configuration/Measure-
".
• Choose "
• Click to "
• Choose "Standard level measurement" and
• Click to " After a few seconds the menu window "
Level measurement
Echo
"
Continue
"
no options
"
Continue
" and "
Pulse-
"
".
Cre­ate new measurement loop - Sensor configu­ration
" opens.
You are in the menu window "
measurement loop - Application
parameter (level, gauge or distance) and the sensor type (Pulse-Echo for ultrasonic).
VEGASON 51 V … 53 V 75
Create new
". Choose the
• Click to "
Sensor coordination
"
Set-up - Adjustment with PC on VEGALOG
The small menu window "
Sensor coordination
opens.
• Choose the serial number of the sensor you want to coordinate and confirm with "OK".
"
The menu window "Create new measurement loop - Measurement loop designation" opens.
Enter: – a meas. loop number – a measurement loop description – and coordinate to your sensor one or sev-
• Click in the menu window "
Create new
measurement loop - Sensor configuration
to "
Continue
"
• Click in the menu window "Create new measurement loop - Measurement loop designation" to "
Level
"
eral output signals.
"
• E.g. configure a current output by clicking to "
Current output
".
In the menu window "Configure current out­put" you choose in your VEGALOG a current output card and coordinate one or several sensor outputs to the sensor.
76 VEGASON 51 V … 53 V
Set-up - Adjustment with PC on VEGALOG
• Click to "Quit" and you are again in the initial menu.
You have carried out the special additional configuration adjustments in conjunction with a VEGALOG. Congratulation!
• Confirm your adjustment with “OK” and you are again in the window “
Create new meas­urement loop - Measurement loop designa­tion”
.
• Confirm your adjustments with "OK".
In the menu "
tem
Now you have to carry out the parameter adjustment. The parameter adjustment is nearly identical with the parameter adjustment in chapter
"6.2 Adjustment with PC on VEGAMET"
Configuration/Measuring sys-
"
you can see that a sensor is configured:
.
VEGASON 51 V … 53 V 77
Parameter adjustment
In the menu "
justment
adjustments.
Adjustment
• Choose the menu "
Instrument data/Parameter adjustment
and then the sensor for which you want to carry out the parameter adjustment.
In the heading of the opening menu window you now see the previously adjusted meas­urement loop name and the measurement loop description. When you have only configured one sensor on the signal conditioning instrument, natu­rally only one sensor is available.
Instrument data/Parameter ad-
" you carry out all important sensor
Set-up - Adjustment with PC on VEGALOG
",
You can carry out the min./max.-adjustment
"with medium"
definite level) or considering the definite level, i.e. also with empty vessel).
Generally you will carry out the adjustment without medium, so you are completely inde­pendent of the actual vessel filling. When you want to carry out the adjustment with medium, you have to carry out the min. adjustment with emptied (also partly emp­tied) vessel and the max. adjustment with filled (also partly filled) vessel. It is hence comfortable and quick to carry out the adjustment without medium like in the example.
(adjustment by means of the
"without medium"
(without
• Choose "
• Choose in the following window if you want
• Enter a distance for the upper and lower
In the example the 0 % filling is at a level
• First choose "
• Click in the menu window "Adjustment" to "
Min / Max - Adjustment
78 VEGASON 51 V … 53 V
Adjustment
"
"
distance of 3,400 m and the 100 % filling at a level distance of 0,500 m.
no (adjustment without medium)
to carry out the adjustment in metres (m) or in
Feet
(ft).
level and the appropriate filling degree in %.
"
Set-up - Adjustment with PC on VEGALOG
When the characteristics points are too close together, the possible meas. error increases. It is hence suitable when you carry out the adjustment at 0 % and at 100 %.
• Confirm the adjustments with "OK", and you are again in the menu window "
• Click in the menu window "Adjustment" to "
Quit
".
You are now again in the menu window "
strument data parameter adjustment
Hence the sensor electronics has two char­acteristics points, one at MIN and one at MAX, out of which a linear proportionality between level distance and the percentage filling of a vessel is generated. The characteristics points however must not be at 0 % and 100 % however the difference should be as large as possible (e.g. at 20 % and at 80 %). The min. distance between the characteristics points for the min./max. ad­justment should be 50 mm.
Adjustment
In-
".
In the menu "
justment/Conditioning/Linearisation
enter later if necessary another linear de­pendence between the level distance and the percentage filling degree (see later subpoint linearisation).
Conditioning
".
• Click in the menu window "
parameter adjustment
The menu window "
• Click to " In the menu "
and 100 %-values of the parameter and their unit. Hence you inform the sensor, e.g. that at 0 % filling there are still 45,5 litres and at 100 % filling1200 litres in the vessel. The sensor indication shows in case of empty vessel 45,5 litres (0 %) and in case of full ves­sel 1200 litres (100 %).
Instrument data/Parameter ad-
" you can
Instrument data
" to "
Conditioning
Scaling
Scaling
Conditioning
".
" you enter the actual 0 %
" opens.
".
VEGASON 51 V … 53 V 79
Set-up - Adjustment with PC on VEGALOG
Linearisation
The relation of level and filling volume is de­scribed with so called linearisation curves. If in your vessel there is another than a linear dependence between the level ("
value
" of the level) and the filling volume
("
Linearised
"), choose the menu "
data/Parameter adjustment/Conditioning
percentage
Instrument
"
• Click in the menu window "
As parameter you can choose (figures), and coordinate to the parameter the appro­priate unit (e.g. litres, hl). The sensor indica­tion then displays the figure in the selected parameter and unit.
• Save the adjustments in the menu "
The adjustments are now transferred to the sensor and you are again in the menu win­dow "
• Click in the menu window "Conditioning" to
• Click in the menu window "Instrument data
80 VEGASON 51 V … 53 V
volume, mass, height
with "OK".
Conditioning
"Quit" parameter adjustment" to "Quit"
".
"dimensionless
and
distance"
Scaling
the menu point "
In the menu window " that "
Linear
the dependence between the percentage value of the filling volume and the value of the level is linear. Beside the two given linearisation curves
"
"
cylindrical tank
also enter six "
Linearisation
" is preadjusted. This means that
" and "
user programmable curves
Conditioning
".
Linearisation
spherical tank
" to
" you see
" you can
".
Set-up - Adjustment with PC on VEGALOG
• Click to "
• Click at the bottom left side to "
First of all a linear relation (straight line) is displayed. In the field " level in percent of the adjusted meas. range (meas. window) is displayed.
The user programmable curve is generated with linearisation points, the so called value pairs. One value pair consists of " (percentage value of the filling) and "
age value
relating to the meas. range). If the linearisation points or value pairs of your vessel are not known to you, you have to gauge the vessel by litres.
User programmable curve
then to "
Edit
".
" and
Show
scaled values
displayed on the y-axis (bottom left in the menu window).
", to have the adjusted unit
Transfer measured value
" the actual
Linearised
Percent-
" (percentage value of the level
Gauging by litres
In the characteristics of the following figure you see five linearisation points or value pairs (0, 1, 2, 3, and 4). There is always a linear interpolation between the linearisation points.
Linearisation point 0 is at 0 % filling (
age value [%]
actual distance to the product surface of 3,400 m in the example (empty vessel). The volume value is 45,5 litres (rest filling of the vessel).
Linearisation point 1 is at a level of 20 % (20 % of the meas. distance of 0,500 m … 3,400 m). At 20 % filling there are in our example 100 litres in the vessel. Linearisation point 2 is at a level of 40 %. At this level there are 250 litres in the vessel. Linearisation point 3 is at a level of 80 %, where 1000 litres are in the vessel. Linearisation point 4 is at a level of 100 % (product distance 0,500 m), where 1200 litres
"
are in the vessel. You can enter max. 32 linearisation points
(value pairs) per linearisation curve.
), which corresponds to an
percent-
VEGASON 51 V … 53 V 81
Set-up - Adjustment with PC on VEGALOG
• Quit the menu with "OK".
• Confirm the message with "OK", and your individual linearisation curve will be saved in the sensor.
Again in the menu window " can enter with the menu point "
time
" a measured value integration. This is useful for fluctuating product surfaces to avoid a permanently changing measured value indication and output. As a standard feature an integration time of 0 seconds is adjusted.
• Quit the menu with "OK".
• Quit the menu window "Instrument data
parameter adjustment" with "OK".
Outputs
• Choose in the main menu window "
ment data/Parameter adjustment
the menu point "
Outputs
Conditioning
Integration
Instru-
".
" you
", and then
• In the menu window "Outputs" click to "
rent output
the 0/4…20 mA output signal.
• If you have made adjustments in the menu window "
Save
• If you want to keep the adjustments un­changed, click to "
You are now again in the menu window "
Outputs
measured value
cation or probably additional indicating in­struments connected to VEGALOG (see also page 51 right column).
", to adjust the signal mode of
"Current output
".
". With the menu point "
", click to
Quit
".
" you scale the sensor indi-
Display of
Cur-
• Click in the menu window " "Quit" and you are in the menu window "
Instrument data parameter adjustment
• Click in the menu window "Instrument data parameter adjustment" again to "
82 VEGASON 51 V … 53 V
Outputs
Quit
" to
"
".
Set-up - Adjustment with PC on VEGALOG
Sensor optimisation
In the menu " carry out special optimising adjustments on the sensors and optimise e.g. by means of the echos curve the installation position of the sensor.
Meas. environment
• Choose the menu "
Instrument data/Parameter adjustment
• Choose in the menu window "
data parameter adjustment
"
Sensor optimisation
A
"
• First click to "
Sensor optimisation
" the menu point
", and click to "
Meas. environment
" you can
"
Instrument
Sensor
".
The window "
Meas. environment
" opens.
With the menu point " can define the operating range of the sensor deviating from the " As a standard feature the operating range corresponds otherwise to the Min-/Max­Adjustment.
Generally it is useful to choose the operating range approx. 5 % larger than the meas. range.
VEGASON 51 V … 53 V 83
In the example the meas. range has been adjusted by the min./max. adjustment to 0,500 ... 3,400 m (min. adjustment 0,500 m, max. adjustment 3,400 m).
You would then adjust the operating range to 0,400 … 3,500 m.
• Save the adjustments with "OK"
• Click in the menu window "
ment
" to "
Measuring conditions
Operating range
" you
Min/Max-Adjustment
Meas. environ-
".
".
Set-up - Adjustment with PC on VEGALOG
• In the menu window " you click to the options corresponding to your application.
• Confirm with "OK".
After some seconds of saving you are again in the window "
• Click in the menu window "
ment
" to "
You are now again in the menu window "
Meas. environment
Quit
".
sor optimisation
Measuring conditions
".
Meas. environ-
Sen-
".
Echo curve
"
With the menu point " have a look to the course and the strength of the detected ultrasonic echo.
If you have to expect strong false echoes due to vessel installations, a correction of the installation position (if possible) can help by monitoring the echo curve, to localize and reduce the size of the false echoes.
Echo curve
" you can
Proceed like in the chapter "Echo curve" on page 53.
84 VEGASON 51 V … 53 V
Set-up - Adjustment with PC on VEGALOG
Meas. loop data
• Click in the window "
rameter adjustment
• Click to "
Application
"
Input No. A
tailed information on your measuring sys­tem in the information windows.
Instrument data pa-
" to "
Meas. loop data
",
" and "
VEGLOG
", to get de-
• Close the information windows.
• Quit the menu "
• Click in the menu window "
".
parameter adjustment
You are now again in the main menu window.
Change COM-interface
In the menu "
nication
" you can adjust the interface para­meters of your PC or change the used COM­port.
Meas. loop data
Instrument data
" to "
Quit
Configuration/Program/Commu-
".
".
Display of measured value
See page 56
VEGASON 51 V … 53 V 85
Simulation
See page 57
Backup
• Click to "
Services/Backup/Signal conditioning
instruments
"
Set-up - Adjustment with PC on VEGALOG
Saved sensor data can be later transferred to other sensors.
E.g. if you have a system with several of the same storage tanks and identical sensors, it is sufficient to configure one sensor, to save the adjustments and transfer them to the other sensors.
This is naturally also valid for the processing system VEGALOG:
• Choose the menu "
Services/Restore con-
figuration/Signal conditioning instruments
"
In the menu window " with the serial number is displayed. You can save the data of VEGALOG individually or in groups with all adjustments in a directory of your choice on the PC. You can also add short backup notes.
In the same way you save the adjustments of the sensors:
Backup
" the VEGALOG
In this menu window you choose in the yellow window cutout a VEGALOG (or a sensor), which should be overwritten with the adjust­ments of another VEGALOG.
Choose in the field " serial number of the VEGALOG, of which you want to transfer the adjustments. With "
store to
" you transfer these VEGALOG-ad­justments to the VEGALOG, which you have selected in the yellow window cutout.
Selection of backups
Re-
" the
86 VEGASON 51 V … 53 V
Notes
VEGASON 51 V … 53 V 87
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail vega@vega-g.de
ISO 9001
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the actual knowledge at the date of printing.
Technical data subject to alteration
2.21 763 / March ‘98
Loading...