VEGA SON51K User Manual

Operating Instructions
VEGASON 51K … 53K
Level and Pressure
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1.1 Function ................................................................................. 4
1.2 Application features ............................................................. 5
1.3 Adjustment ............................................................................ 5
2 Types and versions
2.1 Type survey .......................................................................... 8
2.2 Configuration of meas. systems ....................................... 10
3 Technical data
3.1 Data ..................................................................................... 17
3.2 Approvals ........................................................................... 21
3.3 Dimensions ......................................................................... 22
Contents

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGASON 51K … 53K

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the oper­ating instructions and come with the Ex ap­proved instruments.
Contents
4 Mounting and installation
4.1 General installation instructions ........................................ 24
4.2 Measurement of liquids ..................................................... 25
4.3 Measurement of solids ...................................................... 27
4.4 Socket extensions ............................................................. 29
4.5 Flow measurement ............................................................. 30
4.6 False echoes ...................................................................... 31
4.7 Installation error .................................................................. 33
5 Electrical connection
5.1 Connection and connection cable .................................... 36
5.2 Connection of the sensor .................................................. 37
5.3 Connection of the external indicating instrument
VEGADIS 50 ....................................................................... 38
6 Set-up
6.1 Adjustment structure ......................................................... 39
6.2 Adjustment with PC ............................................................ 39
6.3 Adjustment with the adjustment module MINICOM ........ 40
6.4 Adjustment with the HART®-handheld .............................. 46
7 Diagnosis
7.1 Simulation ............................................................................ 51
7.2 Error codes ........................................................................ 51
VEGASON 51K … 53K 3

1 Product description

1.1 Function

The continuous level measurement with ultra­sonic sensors is based on the running time measurement of ultrasonic pulses.
Measuring principle
Piezoceramic high performance transducers emit focused ultrasonic pulses which are reflected by the surface of solids and liquids. The measuring electronics prepare a precise picture of the environment from the running time and the signal form of the reflected ultra­sonic pulses. The transducers work as trans­mitter and receiver. As receiver the trans­ducers are high sensitivity piezo-micro­phones.
Meas. distance
Product description
In order to be prepared for the different measuring distances and requirements, the instruments operate with emitting frequencies of 34 kHz up to 70 kHz.
As the sound velocity is subjected to a tem­perature influence, the transducer also con­tinuously detects the ambient temperature so that the level is precisely provided even in case of varying ambient temperature.
Output signals
The level proportional measuring signal is provided as 4 … 20 mA-output signal.
The parameter adjustment of the analogue 4 … 20 mA-signal is individually possible and reflects the measuring range or the adjusted adjustment range of the sensor.
Measured value indication
If desired, an indicating instrument for direct lobal survey can be mounted on to the series 50 ultrasonic sensors. The indicating instru­ment shows the precise level with the ana­logue bar-graph and with the digital figure. In addition to the indication in the sensor itself, you can have the level displayed with the external indicating instrument VEGADIS 50 up to 25 m away from the sensor.
The external measured value indication oper-
emission - reflection - receipt
The measuring electronics precisely calculate the distance between transducer and me­dium from the speed of sound and the actu­ally detected running time of the emitted sound impulse. The distance is then con­verted into a level proportional signal and provided according to the sensor parameter adjustment as precise scaled level.
4 VEGASON 51K … 53K
ates, like the integral indication, independent of the 4 … 20 mA-output signal, and indi­vidual parameter adjustment is possible.
Product description

1.2 Application features

Applications
• Level measurement of all liquids
• Level measurement of solids (only short measuring distances) such as e.g.: coal, ore, stones, stone dust, cement, gravel, crushed stones, sand, sugar, salt, cereals, flour, granules, powder, dusts, saw dust, wood chips
• Flow measurement on different flumes
• Gauge measurement, distance measure­ment, object monitoring and conveyor belt monitoring
Two-wire technology
• Supply and output signal on one two-wire line (Loop powered)
• 4 … 20 mA-output signal
Rugged and precise
• Unaffected by product features such as density, conductivity, dielectric constant …
• Suitable for aggressive substances
• Measuring ranges 0,25 m … 15 m
Adjustment choice
• With adjustment software VEGA Visual Operating (VVO) on PC
• With detachable adjustment module MINICOM
• With the HART®-Handheld
• Measured value indication integrated in the sensor
• Optional indication separate from the sen­sor

1.3 Adjustment

Each measuring distance is different, there­fore each ultrasonic sensor must be given some basic information on the application and the environment, e.g. you need to inform the sensor which level means “empty“ and which level “full“. Apart from this “empty and full adjustment“, a number of other adjust­ments can also be carried out with VEGASON ultrasonic sensors.
The adjustment and parameter adjustment of the ultrasonic sensors are carried out with
- the PC
- the detachable adjustment module MINI­COM
- the HART®--Handheld
Adjustment with PC
The set-up and adjustment of the ultrasonic sensors is generally made on the PC with the adjustment program VEGA Visual Operating (VVO) under Windows®. The program leads quickly through the ad­justment and parameter adjustment via pic­tures, graphics and process visualisations.
2
4 ...20 mA
2
Connection for each process
•G 11/2 A, 11/2" NPT
• G 2 A, 2" NPT
• Compression flange DN 100, ANSI 4"
Approvals
• CENELEC, ATEX, PTB, FM, CSA, ABS, LRS, GL, LR, FCC
VEGASON 51K … 53K 5
Adjustment with the PC on the analogue 4 … 20 mA­signal and supply line or directly on the sensor (four­wire sensor)
The PC can be connected to any individual position of the system or the signal line. It is thereby connected with the two-wire PC­interface converter VEGACONNECT 2 to the sensor or to the signal line.
The adjustment and parameter adjustment data can be saved with the adjustment soft­ware on the PC and protected with pass­words. If required, the adjustment can be transmitted quickly to other sensors.
The adjustment program recognises the sensor type
Product description
2
DCS
2
Adjustment with the PC on the 4 … 20 mA signal and supply line to the DCS or directly on the sensor (in the figure a two-wire sensor)
Adjustment with adjustment module MINICOM
With the (3,2 cm x 6,7 cm) 6-key adjustment module with display you can carry out the adjustment in clear text. The adjustment module can be plugged into the ultrasonic sensor or into the optional external indicating instrument.
Tank 1 m (d)
12.345
ESC
+
-
OK
Visualised input of a vessel linearisation curve
Detachable adjustment module MINICOM
By removing the adjustment module unau­thorised adjustments are avoided.
6 VEGASON 51K … 53K
Product description
ESC
+
-
Tank 1 m (d)
OK
12.345
2
4 ... 20 mA
ESC
+
-
Tank 1 m (d)
OK
12.345
4
Adjustment with the detachable adjustment module. The adjustment module can be plugged into the ultrasonic sensor or the external indication instrument VEGADIS 50.
Adjustment with HART®-Handheld
Series 50 sensors with 4 … 20 mA output signal can also be adjusted with the HART®­Handheld. A special DDD (Data-Device­Description) is not necessary as the sensors can be adjusted with the HART®-standard menus of the handheld.
For adjustment, just connect the HART®­Handheld in any position of the 4 … 20 mA­output signal line or insert the two communi­cation lines of the HART®-handheld into the adjustment sockets on the sensor.
2
4 ... 20 mA
2
HART®-handheld on the 4 … 20 mA-signal line
HART Communicator
HART®-handheld
VEGASON 51K … 53K 7

2 Types and versions

Types and versions
VEGASON series 50 sensors are a newly developed generation of very compact, small ultrasonic sensors. With very narrow space requirements, they are designed for short meas. distances (0 … 15 m) and for stand­ard applications such as storage tanks, gauge measurement and buffer tanks.
Due to the small housing dimensions and process connections, the compact sensors monitor your levels very price favourably. With the integral indication and the many features of the “big brothers“ of VEGASON series 80, they offer the advantages of an ultrasonic level measurement for applications where the special advantages of a non-con­tact measurement are not applicable for price reasons.
VEGASON 50 ultrasonic sensors control the two-wire technology perfectly. The supply voltage and the output signal are transmitted via a two-wire line. An analogue 4 ... 20 mA­output signal is available as output or meas­uring signal.
VEGASON 51
with 11/2“ thread
VEGASON 53
for mounting with holding strap or DN 100 compression flange (here on a standpipe)
VEGASON 52
with 2“ thread

2.1 Type survey

General features
• Application in solids and liquids
• Measuring range 0,25 … 15 m
• Ex-approved in Zone 1 (IEC) or Zone 1 (ATEX) classification EEx ia [ia] IIC T6
• Integral measured value indication
8 VEGASON 51K … 53K
Types and versions
Survey VEGASON …
51K 52K 53K
Signal output – active (4 … 20 mA) • – passive (4 … 20 mA)
Voltage supply – two-wire technology (voltage supply and
signal output via a two-wire line)
– four-wire technology (voltage supply separated
from the signal line)
Process connection
1
–G1
/2 A; 11/2“ NPT – – G 2 A; 2“ NPT – – DN 100 compression flange
Adjustment –PC • – adjustment module in the sensor • – adjustment module in external indicating instrument • – HART®-handheld
Measuring range in m – liquids 0,25 … 4 0,4 … 7 0,6 … 15 – solids 0,3 … 2 0,25…3,5 0,75 … 7
Type code
The second letter of the type designation e.g. VEGASON 5[1]… differentiates the instru­ments acc. to process connection and meas­uring range.
The letter e.g. VEGASON 51[K] characterises the output signal: K stands for an analogue 4 … 20 mA output signal (compact instrument).
VEGASON 51 K E X . X X (example)
P - 90 … 250 V AC (only in USA) N - 20 … 36 V DC, 24 V AC (only in USA) Z - supply via signal conditioning instrument (only in USA)
U - only in USA X - outside the USA
Ex approved
K - Analogue 4 … 20 mA output signal
(two-wire or four-wire technology)
V - Digital output signal (two-wire technology)
Type 51: Measuring range 0,25 … 4 m Type 52: Measuring range 0,4 … 7 m Type 53: Measuring range 0,6 … 15 m
Measuring principle (SON for ultrasonic)
VEGASON 51K … 53K 9
Types and versions

2.2 Configuration of meas. systems

Meas. systems in two-wire technology:
• 4 … 20 mA drawing without processing unit,
A meas. system consists of a sensor with a 4 … 20 mA signal output and a unit evaluat­ing or processing the level proportional cur­rent signal.
On the following pages you will see the in­strument configurations called meas. system which are shown in the following partly with a signal processing.
(bottom)
• 4 … 20 mA on active DCS,
(page 12)
• 4 … 20 mA on active DCS (Ex-area),
(page 13)
• 4 … 20 mA on passive DCS,
(page 14)
• 4 … 20 mA on indicating instrument VEGADIS 371 Ex,
(page 15)
Meas. systems in four-wire technology:
• 4 … 20 mA drawing without signal condi­tioning instrument, (page
16)
Meas. systems with VEGASON 51K … 53K
• Two-wire technology (loop powered), supply and output signal via a two-wire line
• Output signal 4 … 20 mA (passive)
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m away from the sensor)
• Adjustment with PC, HART®-handheld or adjustment module MINICOM (can be plugged into the sensor or in the external indicating instrument VEGADIS 50)
VEGADIS 50
4
2
VEGA­CONNECT 2
1)
If the resistors of the processing systems connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 Ω to 350 must be connected to the connection line during adjustment.
The digital adjustment signal would be extremely damped or short­circuited via too small input resistors of the connected processing system which means that the digital communication with the PC could no longer be ensured.
10 VEGASON 51K … 53K
4 … 20 mA
1)
HART®-hand­held
+
-
Types and versions
Measuring system with VEGASON 51K … 53K on active DCS
• Two-wire technology, supply on active DCS.
• Output signal 4 … 20 mA (passive).
• Measured value indication integrated in the sensor.
• Optional external indicating instrument (can be mounted up to 25 m away from the sensor in Ex-area).
• Adjustment with PC, HART®-handheld or adjustment module (can be plugged in the sensor or in the external indicating instrument).
VEGADIS 50
1)
4
2
VEGA­CONNECT 2
4 … 20 mA
2)
passive
2
2
2
DCS
HART®-handheld
1)
If the resistors of the processing systems connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 to 350 must be connected to the connection line during adjust­ment. The digital adjustment signal would be extremely damped or short-circuited via too small input resistors of the connected processing system which means that the digital communication with the PC could no longer be ensured.
2)
4 … 20 mA passive means that the sensor takes, depending on the level a current of 4 … 20 mA. The sensor therefore reacts electrically like a resistor (consumer) on the DCS.
VEGASON 51K … 53K 11
Types and versions
Measuring system with VEGASON 51K … 53K via separator in Ex-area on active DCS
• Two-wire technology (loop powered), supply via the signal line on the DCS; output signal 4 … 20 mA (passive)
• Separator converts the not-intrinsically safe DCS-circuit into an intrinsically safe circuit, the sensor can therefore also be used in Ex-zone 1
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m away from the sensor)
• Adjustment with PC, HART®-handheld or adjustment module MINICOM (can be plugged in the sensor or in the external indicating instrument VEGADIS 50)
Ex-area
VEGADIS 50
4
1)
If the resistors of the processing systems
Not Ex-area
2
Separator (e.g. Stahl) (see "3.2 Approvals“)
2
2
VEGA­CONNECT 2
1)
4 … 20 mA
2)
passive
2
DCS active
HART®-handheld
connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 to 350 must be connected to the connection line during adjust­ment. The digital adjustment signal would be extremely damped or short-circuited via too small input resistors of the connected processing system which means that the digital communication with the PC could no longer be ensured.
2)
4 … 20 mA passive means that the sensor takes a current of 4 … 20 mA depending on the level. The sensor, therefore, reacts electrically like a resistor (consumer) on the DCS. The DCS operates actively, i.e. as current or voltage supply.
12 VEGASON 51K … 53K
Types and versions
Measuring system with VEGASON 51K … 53K vi a separator (Smart-Trans­mitter) on passive DCS
• Two-wire technology (loop powered), intrinsically ia-supply via the signal line from the sepa­rator for operation of the sensor in Ex-Zone 1
• Output signal sensor 4 … 20 mA passive Output signal separator 4 … 20 mA active
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m away from the sensor)
• Adjustment with PC, HART®-handheld or adjustment module MINICOM (can be plugged in the sensor or in the external indicating instrument VEGADIS 50)
VEGADIS 50
4
Ex-area
Not Ex-area
-
+
Separator (e.g. VEGATRENN 149 Ex see "3.2 Approvals“)
1)
2
2
VEGA­CONNECT 2
2
4 … 20 mA
2)
(active)
DCS (passive)
HART®-handheld
3)
1)
If the resistors of the processing systems connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 to 350 must be connected to the connection line during adjust­ment. The digital adjustment signal would be extremely damped or short-circuited via too small input resistors of the connected processing system which means that the digital communication with
2)
4 … 20 mA active means that the separator delivers level dependent a current of 4 … 20 mA. The separator therefore reacts electrically against the DCS like a current source.
3)
4 … 20 mA passive means that the sensor takes, a current of 4 … 20 mA, depending on the level. The sensor therefore reacts electrically like a resistor (consumer) on the DCS.
the PC could no longer be ensured.
VEGASON 51K … 53K 13
Types and versions
Measuring system with VEGASON 51K … 53K on the indicating instrument VEGADIS 371 Ex with current and relay output
• Two-wire technology (loop powered), intrinsically safe ia-supply via the signal line from the indicating instrument VEGADIS 371 Ex for operation of the sensor in Ex-zone 1
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m away from the sensor)
• Adjustment with PC, HART®-handheld or adjustment module MINICOM (can be plugged in the sensor or in the external indicating instrument VEGADIS 50)
Ex-area
VEGADIS 50
4
Not Ex-area
2
2
VEGA­CONNECT 2
4 ... 20 mA
2
1)
(passive)
-
+
VEGADIS 371 Ex
(see "3.2 Approvals“)
HART®-handheld
Relay
0/4 … 20 mA
1)
If the resistors of the processing systems connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 to 350 must be connected to the connection line during adjust­ment. The digital adjustment signal would be extremely damped or short-circuited via too small input resistors of the connected processing system which means that the digital communication with the PC could no longer be ensured.
14 VEGASON 51K … 53K
Types and versions
Measuring system with VEGASON 51K … 53K in four-wire technology
• Four-wire technology, supply and output signal via two separate two-wire lines
• Output signal 4 … 20 mA active
• Optional external indicating instrument with analogue and digital indication (can be mounted up to 25 m away from the sensor)
• Adjustment with PC, HART®-handheld or adjustment module MINICOM (can be plugged in the sensor or in the external indicating instrument VEGADIS 50)
• Max. resistance on the signal output (load) 500
VEGADIS 50
2
-
4
2
2
2
VEGA­CONNECT 2
1)
> 200
+
4 … 20 mA
2)
(active)
HART®-handheld
1)
If the resistors of the processing systems connected to the 4 … 20 mA-signal output are less than 200 , a resistor of 250 to 350 must be connected to the connection line during adjust­ment. The digital adjustment signal would be extremely damped or short-circuited via too small input resistors of the connected processing system which means that the digital communication with the PC could no longer be ensured.
2)
4 … 20 mA active means that the sensor delivers a current of 4 … 20 mA (source) depending on the level. The sensor, therefore, reacts electrically against a processing system (e.g. indication) like a current source.
VEGASON 51K … 53K 15
Several sensors on a voltage source
250 250 250 250 250 250
Types and versions
C
R = 0…2
All sensors are dispatched with the same communication address zero. If several sen­sors are operated on one voltage source, a
resistors avoid the adjustment signals of the actually operated sensor from also being
short-circuited. communication on the signal line with the PC or with the HART®-handheld is no longer possible, as in this case all sensors are ad­dressed digitally at the same time.
With the PC and the adjustment software
VEGA Visual Operating (VVO) or with the
HART®-handheld, the sensors can be ad-
justed from the address zero to the address To avoid this, 250 … 350 resistors must be connected directly on the voltage source in the sensor supply line. The inner resist­ance of the voltage source must be very small. The digital communication signal (PC or HART®-handheld) used on the sensor line,
1 … 15. If a sensor is adjusted to the ad-
dress 1 … 15, the sensor no longer delivers
a 4 … 20 mA-signal, but is frozen to 4 mA
(power supply). Nevertheless, the sensor will
continue to provide the measured value digit-
ally. This mode is called multidrop operation. is then no longer transmitted to the other sensors. You can then communicate with the PC or HART®-handheld on the signal line of the appropriate sensor without the participa­tion of all other sensors in the communication even though all sensors react to the address zero. The low inner resistor (alternating load resistor by filter capacitor) of the voltage source short-circuits the signals before they can cross over to the other sensor lines. The
16 VEGASON 51K … 53K
Technical data

3 Technical data

3.1 Data

Power supply
Supply voltage
- four-wire sensor 230 V AC (20 … 250 V AC), 50/60 Hz
- two-wire sensor 24 V DC (16 … 36 V DC); Sicherung 0,2 A TR
Diagram of the min. supply voltage dependent on the
V
sensor current (two-wire sensor)
36 29
16 14
24 V DC (20 … 72 V DC) fuse 0,2 A TR
Range of the required and permis­sible terminal voltage
Ex-voltage limit
0
422
mA
Current consumption
- two-wire sensor max. 22,5 mA
- four-wire sensor max. 140 mA Power consumption
- two-wire sensor max. 80 mW, 0,45 VA
- four-wire sensor max. 1,2 W, 2,50 VA Load
- four-wire sensor max. 500
- two-wire sensor see diagram
Load diagram of the two-wire sensor (loop resistor of processing unit, e.g. DCS and supply line)
1000
900
800
700 680
600
500
HART®-
400
load
300
250
200
100
0
16 20 25 30 35
min. voltage limit by using the HART®-load
17 18 19 29
VEGASON 51K … 53K 17
max. voltage limit with Ex-sensors
max. voltage limit with not-Ex-sensors
V
36
Technical data
Type 51… 52… 53…
Measuring range (relating to the transducer surface)
Liquid 0,25 … 4 m 0,4 … 7 m 0,6 … 15 m Solid 0,30 … 2 m 0,5 … 3,5 m 0,75 … 7 m
Output signal
Analogue 4 … 20 mA-current signal
Adjustment
- PC with adjustment software VEGA Visual Operating
- adjustment module MINICOM
- HART®-handheld
Accuracy (typical values under reference conditions)
1)
Class accuracy (deviation in characteristics including repeatability and hysteresis acc. to the limit point method) < 0,1 % relating to max. measuring range Temperature drift (relating to 20°C) < 0,015 %/10 K of measuring range Accuracy of the 4 … 20 mA­output signal 0,025 % (D/A-converter) Resolution 1 mm
Characteristics
Min. span (between empty and full adjustment) > 20 mm (recommended > 50 mm) Ultrasonic frequency 70 kHz 55 kHz 34 kHz Intervals 1,0 s 1,0 s 1,0 s Beam angle (at -3 dB emitted power) 5,5° 5,5° 3° Response time
2)
2 … 20 s (dependent on the factory setting)
Materials
Housing PBT (Valox) Transducer, process connection thread PVDF Compression flange PP or 1.4571 Transducer diaphragm (type 53) 1.4571
Weights
Weight incl. transducer
- VEGASON 51 1,2 kg
- VEGASON 52 1,6 kg
- VEGASON 53 2,3 kg
1)
Reference conditions acc. to IEC 770, e.g.: Temperature 18°C … 30°C etc.
2)
The response time is the time the sensor requires to provide the level correctly with a quick level change (with max. 10 % deviation).
18 VEGASON 51K … 53K
Technical data
Ambient conditions
Max. vessel pressure (gauge pressure)
- VEGASON 51 and 52 3 bar
- VEGASON 53 1 bar Ambient temperature
- sensor (electronics) -20°C … +60°C
- process (transducer) -40°C … +80°C (StEx: -20°C … +75°C)
- storage and transport temperature -40°C … +80°C Protection
- sensor IP 67
- transducer, process IP 68 Protection class
- two-wire sensor II
- four-wire sensor I Overvoltage category III Self-heating at 40°C ambient temperature
- to sensor 45°C
- to transducer, process 55°C
Ex-technical data
Classification ia intrinsically safe (in conjunction with a safety
barrier or separator) Temperature class (permissible ambient temperature on the transducer when used in Ex-areas)
- T6 42°C
- T5 58°C
- T4 60°C
- T3 60°C Ex-approved in category or zone
- ATEX Zone 1 (II 2 G)
- IEC, CENELEC, PTB Zone 1 (II 2 G) Classification mark EEx ia IIC T6
Process connections
VEGASON 51 G 11/2 A, 11/2" NPT VEGASON 52 G 2 A, 2" NPT VEGASON 53 DN 100 compression flange
Connection lines
Two-wire sensors supply and voltage via one two-wire line,
load dependent on the supply voltage Four-wire sensor supply and signal separately, load of the
Cross-section area of conductor generally 2,5 mm
4 … 20 mA-signal line max. 500
Earth connection max. 4 mm
2
2
Cable entry, Pg 2 x M20 x 1,5 (cable diameter 5 … 9 mm)
VEGASON 51K … 53K 19
Technical data
CE-conformity
VEGASON series 50 ultrasonic sensors meet the protective regulations of EMVG (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMVG Emission EN 50 081 - 1: 1993
NSR EN 61 010 - 1: 1993
Susceptibility EN 50 082 - 2: 1995
Outputs and processings
Display indication
Indication - optional integrated in the sensor, scalable
Signal output
Signal output
- two-wire technology 4 … 20 mA (load dependent on supply
- four-wire technology 4 … 20 mA (load max. 500 Ω) Resolution of the 20 mA-signal 0,025 % of operating range Failure of the 20 mA-signal < 0,025 % of operating range
analogue and digital meas. value indication
- optional external meas. value indication, powered by the sensor and up to 25 m away from the sensor (also the adjustment module MINICOM can be inserted)
voltage, see load diagram page 17)
Two-wire technology: The analogue 4 … 20 mA-output signal (measuring signal) is transmitted together with the power supply via a two-wire line.
Four-wire technology: Separate power supply. The analogue 4 … 20 mA-output signal (measuring signal) is looped by a line separated from the supply voltage.
20 VEGASON 51K … 53K
Technical data

3.2 Approvals

When using ultrasonic sensors in Ex-areas or navigation, the instruments must be suitable and approved for the explosion zones and application areas. The suitability is checked by approval authorities and certified by ap­proval documents.
VEGASON 50 ultrasonic sensors are ap­proved for Ex-zone 1. Please note the at­tached approval documents when using a sensor in Ex-area.
Test and approval authorities
VEGASON ultrasonic sensors are tested and approved by the following monitoring, test and approval authorities:
- PTB
(Physikalisch Technische Bundesanstalt ­Physical Technical Test Authority)
- FM
(Factory Mutual Research)
- ABS
(American Bureau of Shipping)
- LRS
(Lloyds Register of Shipping)
- GL
(German Lloyd)
- CSA
(Canadian Standards Association)
Ex-area zone 1
Series 50 senors require special safety barri­ers and separators for operation in Ex-area zone 1. The safety barriers and separators provide intrinsically safe (ia) circuits. The followes shows a choice of instruments with which the series 50 sensors work reliably. The signal line resistance must not exceed 15 per wire.
Separators and signal conditioning in­strument:
- VEGADIS 371 Ex
- A puissance 3 PROFSI 37-24070A
- VEGAMET 614 Ex
- Apparatebau Hundsbach AH MS 271-B41EEC 010
Separators:
- VEGATRENN 149 Ex…
- Stahl 9303/15/22/11
- CEAG GHG 124 3111 C1206
Separators, safety barriers:
- Stahl 9001/01/280/110/10
- CEAG GHG 11 1 9140 V0728
- Type 9130 (VEGA)
- Stahl 9001/51/280/110/14
- MTL 787 S+
- CEAG CS 3/420-106
VEGASON 51K … 53K 21

3.3 Dimensions

VEGASON 51
Technical data
95
152
VEGASON 52
206
202
22
65
2 x M20x1,5
Thread G 11/2 A or 11/2“ NPT
Reference plane
ø39 ø60
85
95
Min.-distance to the medium
152
0,25 m
Thread G 2 A or 2" NPT
22
61
ø50 ø72
85
SW 60
Min.-distance to the medium
Reference plane
0,4 m
22 VEGASON 51K … 53K
Technical data
193
ø158
152
95
247
ø148
12
M8x10
ø12
ø12
120
0,6 m
VEGASON 53
External indicating instrument VEGADIS 50
38
5
ø
82
Mounting loop
suitable for compression flange DN 100
Min.-distance to the medium
48
10
135
118
108
85
Note:
Cable diameter of the connection cable min. 5 mm and max. 9 mm. Otherwise the seal effect of the cable entry is not ensured.
-
Pg 13,5
+
ESC
OK
Carrier rail mounting 35 x 7,5 acc. to EN 50 022 or flat screwed
Adjustment module MINICOM
Tank 1 m (d)
12.345
32,5
Adjustment module for insertion into VEGASON series 50 sensors or into the exter­nal indicating instrument VEGADIS 50
67,5
74
VEGASON 51K … 53K 23

4 Mounting and installation

Mounting and installation

4.1 General installation instructions

Measuring range
You select your instrument according to the required measuring range, as well as other criteria. The reference planes for the min. and
on instruments with flange version, the instru­ment flange. (Note the information for the reference planes in chapter "Dimensions“). The max. filling is dependent on the required min. distance of the instrument used (0,25 m to 1,4 m) and the installation place of the instrument or the transducer.
max. distance to the product or the solid is either the transducer surface (diaphragm) or,
Typ e 5 1
Min. distance
0,25 m 0,4 m
Min. distance, max. measuring range and span (example VEGASON 51, 52 and 53)
Full
1m
max. meas. range
4 m (type 51), 7 m (type 52), 15 m (type 53)
Empty
Min. distance
0,75 m
Typ e 5 3 Typ e 52
Reference plane
Span
Min. distance
Beam angle and false echoes
If possible, provide a vertical directing of the
sensor axis to the product surface and The ultrasonic impulses are focussed by the transducer. The impulses leave the trans-
avoid, if possible, struts within the 100 %-
beam, e.g. by tubes and struts. ducer in conical form similar to the beam pattern of a spotlight. The beam angle is 5,5° (VEGASON 51/52) or 3° (VEGASON 53) at
-3 dB emitted power.
Any object in this beam angle causes a false echo. Within the first few metres of the beam angle, tubes, struts or other installations cause strong false echoes. At a distance of 6 m the false echo of a strut has 9-times more amplitude than at a distance of 18 m.
24 VEGASON 51K … 53K
Mounting and installation
Meas. distance
0 m
4 m
7 m
Meas. distance
0 m
VEGASON 51/52
0,4
VEGASON 53

4.2 Measurement of liquids

Flat vessel top
On flat vessels the mounting is mainly made on a very short DIN-socket piece. Reference
5,5˚
12˚
50 %
emitted power
100 %
emitted power
0
1,21,2
0,4
emitted power
50 %
m
plane on flange versions is the instrument flange. The transducer should protrude out of the flange pipe.
Reference plane
60 mm
Min. distance
Flange version on very short DIN-socket piece
Ideal is the mounting directly on the vessel top. A round opening on the vessel is suffi­cient to fasten the VEGASON 53 sensor with a compression flange.
emitted power
100 %
Reference plane
15 m
0,4
1,2
0,4
0
m
1,2
Flange version (compression flange) on flat vessel top
VEGASON 51K … 53K 25
You can also mount the sensors with 11/2“ or 2“ thread to short socket pieces.
60 mm
Reference plane
60 mm
Mounting and installation
Reference plane
1
/2 vessel radius
Mounting on short 11/2“ or 2“ socket piece
Dished tank ends
On dished tank ends, please do not mount the instrument in the centre but approx. 1/2 vessel radius from the centre. Dished tank ends can act as paraboloidal reflectors for the ultrasonic pulses. If the transducer is placed in the "focus" of the parabolic tank end, the transducer receives amplified false echoes. The transducer should be mounted outside the "focus". Parabolic amplified ech­oes are thereby avoided.
60 mm
Reference plane
1
/2 vessel radium
Flange on dished vessel ends
Open vessles
On open vessels the use of the instruments
on a mounting lever is recommended. Mount
the low-weight sensor to a mounting lever.
Reference plane
Min. meas. distance
Open vessels
Mounting boss on dished tank ends
26 VEGASON 51K … 53K
;;;;;;;;;;;;;
Mounting and installation
Pump shaft
Narrow shafts and shaft openings (vessel openings) with very rough walls and shoul­ders aggravate an ultrasonic measurement due to strong false echoes.
Shaft and vessel openings
Narrow, very rough shaft, well and vessel openings can be overcome by a socket piece as described under "4.4 Socket exten­sions“ (left lower half of the figure).
see also “4.4 Socket extension“
Socket piece
> 100 mm
min. distance
Measuring
Shaft
You can realise very good meas. results with a meas. pipe in continous narrow shafts, see figure. The meas. pipe used must have smooth walls inside (e.g. PE-sewage pipe) and a diameter of 100 mm. This coordination works without problems as long as the inside of the meas. pipe has no build-up (cleaning). Also check the use of hydrostatic pressure transmitters or capacitive electrodes.

4.3 Measurement of solids

Flange mounting
As with liquids, the instrument can also be mounted on a short DIN-socket piece in solid vessels. The socket axis must point to the vessel outlet or should be directed vertically to the product surface and must be very short (< 100 mm).
Reference plane
range
Shaft pump
Meas. tube
Socket piece or meas. tube on the example of a shaft
VEGASON 51K … 53K 27
Shaft pump
VEGASON 53 on vessel flange
Min. distance
Swivelling holder
The accessory programme offers a swivel­ling holder (mounting loop) for mounting of the VEGASON 53. This facilitates the direct­ing of the sensor to the product surface.
Reference plane
Min. distance
VEGASON 53 on swivelling holder
Mounting and installation
Reference plane
Min. distance
VEGASON 51 or 52 on the mounting boss. The socket
axis should point to the product surface
Material cones
Large material cones are detected with sev­eral instruments which can be fastened on lifting beams.
Mounting boss
Reference plane
Max.
Transducer on lifting beam above material cone
VEGASON 52 on mounting boss
28 VEGASON 51K … 53K
Mounting and installation
;
;
;
;
;
;
;
;

4.4 Socket extensions

The ultrasonic sensor requires a min. dis­tance to the product or solid. Note this min. distance in your planning. In exceptions it is possible to reach the re­quired min. distance and hence the desired filling height with a socket piece. However, the socket piece increases the noise level of the ultrasonic signal and can interfere with the measurement. Only provide the socket ex­tension when there is no other possibility, and carry out the extension as shown in the fol­lowing figure.
Socket extension with liquids
Chamfer the socket carefully and ensure a smooth inner side of the socket. The socket should not protrude into the measured prod­uct when pollution or measured product could stick to the socket.
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;;
;;;;;;;;;;;;;;;
For measurements in products which do not cause any build-up, the socket extension, in form of a meas. pipe, can be permanently submerged into the measured product. The ultrasonic mesurement is then only carried out in the meas. pipe without being interfered with by vessel installations (see page 28 "Pump shaft“).
Type 51/52
LL
45˚
ø
Socket extension with liquids
Typ e 5 3
45˚
ø
Max. socket length dependent on the socket diameter
øL in mm in mm Type 51 Type 52 Type 53
100 200 300 300 150 300 400 400 200 500 500 250 600
Socket piece should not be submerged into the adhesive product (figure: VEGASON 53)
Choose as large a socket diameter as possi-
Socket extension with solids
For solids provide a conical socket extension with an angle of at least 15° … 20°.
ble and as small a socket length as possible. Ensure that the socket opening is burr free to minimise false echoes.
15˚
15˚
Socket extension with solids
VEGASON 51K … 53K 29
Mounting and installation

4.5 Flow measurement

The short examples on this page only give introductory information for your flow meas­urement. Planning information is given by the flume manufacturers and in special literature.
Rectangular flume
- Installation of the sensor on the upstream
side
- Note distance to the overfall edge
(3 … 4 x h
- Installation centered to the flume
- Edge opening 2 x h
- Installation perpendicular to the liquid sur-
face
- Keep min. distance relating to h
- Min. distance of the edge opening of down-
stream water 50 mm
max
90°
)
from ground
max
max
max. distance
h
max
Khafagi-Venturi flume
- Installation of the sensor on the inlet side
- Note distance to the Khafagi-Venturi flume (3 … 4 x h
- Installation perpendicular to the liquid sur­face
- Keep min. distance relating to the height of damming h
Khafagi-Venturi flume
max
max
3 … 4 x h
90°
Sensor
)
max
h
max
B
2 x h
Overfall edge
max
Flow measurement on open flumes
Overfall edge
3 … 4 x h
max
90°
5 cm
Upstream water
Downstream water
Flow measurement on open flumes
30 VEGASON 51K … 53K
Mounting and installation

4.6 False echoes

The placing of the ultrasonic senor must be chosen so as to ensure that no mountings or in-flowing filling material crosses with the ultrasonic impulses. The following examples and instructions show the most frequent measuring problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to their strong false echoes, greatly effect the measurement. The placing of shields over these flat protrusions scatter the false echoes and guarantee a definite measurement.
Correct Wrong
Vessel protrusions (slope)
Intake pipes, i.e. for the mixing of materials ­with the flat top turned towards the sensor ­should be covered with an angular shield with which the false echoes are then scat­tered.
Vessel mountings
Vessel mountings such as, for example, a ladder often cause false echoes. When pro­jecting the measuring point, please ensure the free access of the ultrasonic signals to the medium.
Correct Wrong
Ladder
Vessel mountings
Ladder
Vessel struts
Vessel struts, as with other vessel mountings, can cause strong false echoes which super­impose the useful echoes. Small shields effectively hinder a direct false echo reflec­tion. The false echoes are scattered and diffused in the area and are then filtered out as "echo noise" by the measuring electronics.
Correct Wrong
Correct Wrong
Vessel struts
Vessel protrusions (intake pipe)
VEGASON 51K … 53K 31
Shields
Mounting and installation
;;
;;
;;
Flood basin
The max. height of water to be expected determines the installation height to keep the min. distance of the transducer even with highest high water. The low water edge should be covered with a shield in the trans­ducer range.
Correct Wrong
Min. distance high water
60°
Shield
Low water
Gating out of an echo
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and note the inflowing mate­rial.
Correct
Wrong
Mounting on the vessel
If the sensor is mounted too close to the vessel wall, build-up and adhesion of the filling material to the vessel wall cause false echoes. Please position the sensor at a suffi­cient distance from the vessel wall. Please also note chapter "4.1 General installation instructions".
Correct
Build-up
Wrong
Strong product movements
Heavy turbulences in the vessel, e.g. by strong stirrers or strong chemical reactions, aggravate the measurement. A surge or bypass pipe (figure) of sufficient size always allows, provided that the product causes no build-up in the pipe, a reliable measurement even with strong turbulences in the vessel.
Correct
Wrong
Strong product movements
Inflowing material
32 VEGASON 51K … 53K
Mounting and installation
;
;
;
;

4.7 Installation error

Foam generation
Strong foam on the product can cause faulty measurements. Provide measures to avoid foam, measure in a bypass pipe or use an­other meas. principle, e.g. capacitive electro­des or hydrostatic pressure transmitters.
;;;;;;;;;;;;;
;;;;;;;;;;;;;
;;;;;;;;;;;;;
;;;;;;;;;;;;;
Foam generation
Wrong directing to the product
Weak measuring signals are caused if the sensor is not directed to the product surface. Direct the sensor axis vertically to the prod­uct surface to achieve optimum measuring results.
Strong heat fluctuations
Strong heat fluctuations, e.g. due to the sun cause measuring errors. Provide a shield.
Shield
Strong heat fluctuations
Min. distance to the medium
When the min. distance to the medium is not maintained, the instruments show wrong measured values. Mount the instruments with the required min. distance.
Correct Wrong
Sensor too close to the vessel wall
Direct sensor vertically to the product surface
VEGASON 51K … 53K 33
Mounting and installation
Sensor too close to the vessel wall
if the sensor is mounted too close to the ves­sel wall (dimension A in diagram) strong false echoes can be caused. Build-up, rivets, screws and weld joints superimpose their echoes on the product echo or useful echo. Please ensure the sufficient distance of the sensor to the vessel wall depending on the maximum measuring distance (dimension B in diagram). In case of good reflection condi­tions (liquids, no vessel installation) we rec­ommend the provision of the sensor distance according to diagram curve 1. At a max. meas. distance of, for example 10 m the distance of the transducer, according to curve 1, should be approx. 1 m. In case of solids with bad reflection conditions provide a distance to the vessel wall according to diagram curve 2. With very bad meas. con­ditions it could be necessary to increase the distance to the vessel wall or to gate out the false echoes additionally by a false echo storage and thereby adapt the sensor more precisely to the environment.
Distance of the transducer to the vessel wall
Parabolic effects on dished boiler head or basket arch vessels
Round or parabolic tank tops act like a para­bolic mirror for the signals. If the sensor is placed to the focus of such a parabolic tank top the sensor receives amplified false ech­oes. The optimum mounting is generally in the range of half the vessel radius from the cen­tre.
Correct
> 60 mm
~ 1/
2
vessel radius
Wrong
Curve 1 (liquids)
Curve 2 (solids)
Mounting on a vessel with parabolic tank top
Max. meas. distance
34 VEGASON 51K … 53K
Wrong
Mounting and installation
Socket piece too long
When mounting the sensor in a socket piece which is too long, strong false echoes are caused, aggravating the measurement. Please ensure that the transducer protrudes at least 30 mm out of the socket piece.
Wrong
Correct
> 60 mm
Reference plane
Correct and wrong length of socket piece
VEGASON 51K … 53K 35

5 Electrical connection

Electrical connection

5.1 Connection and connection cable

Safety information
Ensure that the instrument is unpressurised before you start work. Always switch off the power supply before you carry out clamping work on the ultrasonic sensors. Protect your­self and the instruments, especially when using sensors which do not operarate with low voltage.
Skilled staff
Instruments which are not operated with protective low voltage or DC voltage must only be connected by skilled staff.
Connection
A standard two or four-wire cable (sensors with separate supply) with max. 2,5 mm2 can be used as connection. Very often the "elec­tromagnetic pollution“ by electronic actuators, energy lines and transmitting stations is so considerable that the two-wire line or four­wire line should be screened.
We recommend the use of screening. This screening prevents future interferences. Only earth the cable screens on both ends (on the sensor and on the processing system) when you have determined, by measurement, that no or only lowest earth compensating cur­rents flow via the screens. Use a very low impedance earth connection (foundation, plate or mains earth).
Ex-protection
If an instrument is used in hazardous areas, the necessary regulations, conformity and type approvals for systems in Ex-areas must be noted (e.g. DIN 0165).
Intrinsically safe circuits with more than one active instrument (instrument delivering electrical energy) must not be connected. Please note the special installation conditions (DIN 0165).
Connection cable
Note that the connection cables are specified for the expected operating temperatures in your systems. The cable must have an outer diameter of 5 … 9 mm to ensure the seal effect of the cable entry. Cables for intrinsically safe circuits must be marked blue and must not be used for other circuits.
Earth conductor terminal
On VEGASON 51/52 sensors the earth con­ductor terminal is galvanically isolated. The sensors are shockproof. On the VEGASON 53 sensor the earth conductor terminal is galvanically connected to the metal trans­ducer diaphragm.
36 VEGASON 51K … 53K
Electrical connection

5.2 Connection of the sensor

After having mounted the sensor in the meas. position according to the instructions in chap­ter 4 "Mounting and installation“, loosen the closing screw on the top of the sensor. The sensor cover with the optional indication display can then be opened. Unscrew the compression screw and shift the screw over the approx. 10 cm dismantled connection cable. The compression screw of the cable entry is protected with a safety lock-in posi­tion against automatic loosening.
Supply voltage 4 … 20 mA (passive)
+
-
1)
To the indicating instrument in the sensor cover or to the external indicat­ing instrument VEGADIS 50
Now loop the cable through the cable entry into the sensor. Screw the compression screw on the cable entry and clamp the dis­mantled wires of the cable to the appropriate terminal positions.
The terminals operate without terminal screw. Press the white opening buckets with a small screwdriver and insert the copper core of the connection line into the terminal opening. Check the position of the lines in the terminal position by slightly pulling on the connection lines.
Supply voltage
+
-
4 … 20 mA (active)
+
-
2)
Terminals (max. 2,5 mm
12 C 567843
12 C 5678
Commu-
+-
nication
4...20 mA
+
-
Tank 1 m (d)
12.345
Tw o-wire technology
Display
ESC
OK
wire cross-section)
Sockets for connection of the HART®-handheld or the VEGACONNECT
pluggable adjustment module MINICOM
(loop powered)
1)
4 … 20 mA passive means that the sensor takes a current of 4 … 20 mA (consumer), depending on the level.
2
12 C 567843
12 C 567843
Commu-
(+) (-)
L1 N
Tank 1 m (d)
12.345
nication
-
+ -
4...20 mA
+
Display
ESC
OK
Four-wire technology
(separate supply)
2)
4 … 20 mA active means that the sensor provides a current of 4 … 20 mA (current source), depending on the level.
Opening buckets
VEGASON 51K … 53K 37
Electrical connection

5.3 Connection of the external indicating instrument VEGADIS 50

Loosen the 4 screws of the housing cover on VEGADIS 50. You can facilitate the connection procedure by fastening the housing cover during con­nection with one screw on the right of the housing (figure).
OUTPUT (to sensor)
3
2
1
4
5
8
6
7
DISPLAY
(in the cover of the indicating instrument)
VEGADIS 50
Adjustment module
Tank 1 m (d)
12.345
+
ESC
-
OK
Screws
Voltage supply
-
+
4 … 20 mA active
12 C 567843
Commu-
(+) (-)
L1 N
Tank 1 m (d)
12.345
nication
+-
4...20 mA
+
-
Four-wire sensor
(separate supply)
4 … 20 mA passive
+
-
to VEGADIS 50
12 C 567843
12 C 5678
Commu-
+-
Display
ESC
OK
4...20 mA
Tank 1 m (d)
12.345
nication
Display
+
ESC
-
OK
Tw o-wire sensor
(loop powered)
38 VEGASON 51K … 53K
Set-up

6 Set-up

6.1 Adjustment structure

Series 50 ultrasonic sensors can be adjusted with
- PC (adjustment program VVO)
- with the detachable adjustment module MINICOM
- with the HART®-handheld.
The adjustment must only be carried out with one adjustment medium. If, for example you try the parameter adjustment with the MINICOM and the HART®-handheld, the adjustment will not be successful.
Adjustment program VVO
With the adjustment program VVO Visual Operating System) on the PC you can adjust the ultrasonic sensors very comfort­ably. The PC communicates via the interface adapter VEGACONNECT 2 with the sensor. A digital adjustment signal is therefore superim­posed on the signal and supply line. The adjustment can be carried out directly on the sensor or on any individual position of the signal line.
Adjustment module MINICOM
With the adjustment module MINICOM you adjust in the sensor or in the external indicat­ing instrument VEGADIS 50. The adjustment module enables via the text display with 6­key field the adjustment in the same functions volume as with the adjustment program VVO.
(VEGA

6.2 Adjustment with PC

The instructions for the adjustment with the PC and the adjustment program VVO can be found in the manual "VEGA Visual Operating (VVO)" which is supplied with the latest ad­justment software. Beside the adjustment of ultrasonic sensors, the manual "VEGA Visual Operating (VVO)" also describes the setup of all other VEGA sensors with the PC.
PC on the sensor
For connection of the PC to the sensor you need the interface adapter VEGACONNECT
2. Insert VEGACONNECT 2 into the CONNECT-socket provided in the sensor.
PC on the signal line
Connect the two-wire line of VEGACONNECT 2 to the signal or supply line of the sensor. When the resistors of the systems (DCS, current source etc.,) connected to the signal/ supply line are less than 250 , a resistor of 250 … 350 must be connected to the si­gnal/supply line during adjustment. The dig­ital signals modulated to the signal line would be considerably damped via the small sys­tem resistors or "shortcircuited" which means that the communication with the PC would be interfered with.
HART®-handheld
VEGASON 50 K ultrasonic sensors, like other HART®-protocol capable instruments, can be adjusted with the HART®-handheld. A manu­facturer specific DDD (Data-Device-Descrip­tion) is not required. The ultrasonic sensors are adjusted with the HART®-standard menus. All main functions are, therefore, accessible. Some functions which are rarely used such as, for example the scaling of the A/D-con­verter for the signal output, or the adjustment with medium, are not possible or are blocked by the HART®-handheld. These functions must be carried out with the PC or the MINICOM.
VEGASON 51K … 53K 39
Set-up

6.3 Adjustment with the adjustment module MINICOM

As with the PC, you can also adjust the sen­sor with the small detachable adjustment module MINICOM. The adjustment module is thereby plugged into the sensor or into the external indicating instrument (optional).
ESC
+
-
Tank 1 m (d)
OK
12.345
2
4 ... 20 mA
ESC
+
-
Tank 1 m (d)
OK
12.345
4
For the adjustment with the adjustment mod­ule all sensor versions (adjustment options) as with the PC and the adjustment program VVO are available. The adjustment with MINICOM, however, is different.
Error codes:
E Hardware failure, electronics defect E013 No valid measured value
- Sensor in the feeding phase
- Loss of the useful echo E017 Adjustment span too small E036 Sensor program not operating
- Sensor must have a new program-
ming (service)
- Fault signal also appears during
programming
Adjustment steps
On pages 56 and 57 you can find the com­plete menu plan of the adjustment module MINICOM. Set-up the sensor in the numbered se­quence:
1. Adjustment
2. Conditioning
3. Outputs
4. Operating range
5. Meas. conditions
6. Measurement in gases only necessary
when measuring gas compositions other than air.
7. False echo memory (only required when
meas. errors occur during operation).
8. Indication of the useful and noise level Short explanations to the set-up steps 1 … 8 follow:
You carry out all adjustment steps with the 6 keys of the adjustment module. A small dis-
1. Adjustment
play shows you, apart from the measured value, a short message on the menu point or on the figure of a menu adjustment.
The information volume of the small display
Max.
100 % (1,270 m) corresp. to 1200 l
however cannot be compared with that of the adjustment program VVO, but you will soon get used to it and will be able to carry out your adjustments quickly and directly with
Span (4,58 m)
the small MINICOM.
Min.
40 VEGASON 51K … 53K
0 % (5,850 m) corresp. to 45 l
Set-up
Under the menu point "
Adjustment
“ you in­form the sensor with which meas. range it should operate.
You can carry out the adjustment without and with medium. Generally you will carry out the adjustment without medium as you can then adjust without filling cycle.
Adjustment without medium
(adjustment independent of the actual level)
Key Display
Sensor
m(d)
4.700
Para­meter
OK
OK
OK
OK
+
The distance indication flashes and you can choose "feet“ and
adjust ment
Adjust ment
w.out medium
Adjust ment in
m(d)
(Min. adjustment)
"m“.
OK
+
or
Confirm the adjustment with "OK“.
m(d)
0.0%
bei
m (d)
XX.XXX
Adjust ment in
With "+“ and "–“ you adjust the percentage value for the min. value or the lower level l (ex­ample 0,0 %).
OK
The adjusted percentage value is written in the sensor and the distance correspond­ing to the percentage value flashes.
+
or
With the "+“ or "–“-key you can co-ordinate a product dis­tance (example 5,85 m) to the previously adjusted percent­age value. When you do not know the distance, you have to sound.
OK
The adjusted product dis­tance is written in the sensor and the indication stops flash­ing.
You have thereby adjusted the lower product distance as well as the percentage filling value corresponding to the lower product distance.
100.0%
at
m (d)
XX.XXX
(Max-adjustment)
The max. adjustment (upper product dis­tance) is made in the same way (example: 100 % and 1,270 m).
Note:
The difference between the adjustment val­ues of the lower product distance and the upper product distance should be as big as possible, preferably at 0 % and 100 %. If the values are very close together, e.g. lower product distance at 40 % (3,102 m) and upper product distance at 45 % (3,331 m), the measurement will be inaccurate. A level characteristic is generated from the two points. Even with smallest deviations between actual product distance and ad­justed product distance, the slope of the characteristics will be considerably influ­enced. Thereby small failures multiply with the adjustment when the adjustment points are too close together to more considerable errors at the output of the 100 %-value or the 0 %-value.
VEGASON 51K … 53K 41
Set-up
Adjustment with medium
with medium
Max.
Min.
adjust
adjust
at %
at %
XXX.X
XXX.X
Fill the vessel, for example to 10 % and enter 10 % in the menu "
Min. adjust
“ with the "+“ and "–“ keys. Then fill the vessel, e.g. to 80 % or 100 % and enter 80 % or 100 % in the menu "
Max. adjust
“ with the "+“ and "–“ keys.
2. Conditioning
Signal condit ioning
Scal ing
0 %
100 %
Deci-
prop.
corres
corres ponds
XXXX
mal point
888.8
Signal conditioning
ponds
XXXX
Under the menu point " you choose the product distance at 0 % and at 100 % filling. Then you enter the parameter and the physical unit as well as the decimal point.
Enter in the menu window "
0 % corresponds
the value of the 0 %-filling. In the example of the adjustment with the PC and the adjust­ment software VVO this would be 45 for 45 l.
Unit
to
Mass
Kg
With the "—>“ key you change to the 100 % menu. Enter here the figure of your param­eter corresponding to a 100 %-filling. In the example 1200 for 1200 l.
• Confirm with "OK“.
If necessary choose a decimal point. How­ever note that only max. 4 digits can be dis­played. In the menu "
Prop. to
“ you choose the parameter (mass, volume, distance…) and in the menu "
Unit
“ the physical unit (kg, l,
ft3, gal, m3 …).
Linearisation:
Adjust ment
Signal condit ioning
Scal ing
Lin. curve
Linear
Integr ation time
0 s
A linear dependence between percentage value of the product distance and the per­centage value of the filling volume is preadjusted. With the menu "Lin.curve“ you can choose between linear, spherical tank and cylindrical tank. The adjustment of one's own linearisation curve is only possible with the PC and the adjustment program VVO.
3. Outputs
Under the menu "
for example the current output should be inverted or which parameter should be pro­vided by the sensor display.
Outputs
“ you determine if,
• Confirm with "OK“.
42 VEGASON 51K … 53K
Set-up
4. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range has already been ad­justed with the min./max. adjustment. Gener­ally it is useful to choose a slightly bigger (approx. 5 %) operating range than the measuring range.
Example: Set min./max. adjustment to: 0,300 … 5,850 m; operating range to approx. 0,250 … 6,000 m.
5. Meas. conditions
(see menu plan no. 5)
6. Measurement in gases
Adjustment is only necessary when the measurement is made in gases (Co2, He, etc.) deviating from air. In case of measure­ment in gases sound the distance of the sensor to the product surface and enter in the menu point "Measurement in gases“. The sensor can then consider the modified sound velocity in gases as opposed to air and pro­vide correct levels.
7. False echo storage
A false echo storage is always useful when false echo sources such as struts must be reduced. With the creating of a false echo storage, you authorise the sensor electronics to note the false echoes and save them in an internal database. The sensor electronics treat these (false) echoes differently from the useful echo and gates them out.
8. Useful and noise level
In the menu
you get important information on the signal quality of the level echo. The higher the "S-N“-value, the more reliable the measure­ment (menu plan MINICOM).
Ampl.: Means amplitude of the level echo in
S-N: Means Signal-Noise, i.e. the useful
The higher the "S-N“-value (distance of the amplitude useful level to noise level), the better the measurement: > 50 dB Measurement excellent 40 … 50 dB Measurement very good 20 … 40 dB Measurement good 10 … 20 dB Measurement satisfactory 5 … 10 dB Measurement sufficient < 5 dB Measurement poor
Example:
Ampl. = 68 dB S-N = 53 dB
68 dB – 53 dB = 15 dB
53 dB signal distance means very good reliability.
This means that the noise level is only 68 dB – 53 dB = 15 dB.
Ampl.:
XX dB
S-N:
XX
dB
dB (useful level)
level minus the level of the back­ground noise
VEGASON 51K … 53K 43
Menu plan of the adjustment module MINICOM
Set-up
Sensor
m(d)
4.700
Param eter adjust ment
Sensor optimi sation
Sensor Tag
Sensor
Meas. enviro nement
Opera­ting range
Begin
m (d)
0.50
SON 52 K
1.00
4.
When switching on, the sensor type and the software version are displayed for a few seconds.
End
6.00
m (d)
Meas. condit ions
Condit ion
liquid
5.
Fast change
No
Measur ing in tube
Meas. dist.
m (d)
4.700
Correc­tion Now!
Agitat ed sur face No
OK?
6.
Foam­ing prod. No
High dust level No
Large angle repose No
Measur ing in tube No
Viel­fach­echos No
1. 2. 3.
Adjust ment
w.out medium
Adjust ment in
m(d)
with medium
Min. adjust at %
XXX.X
0.0 %
at
m (d)
XX.XXX
Max. adjust at %
XXX.X
100.0%
at
m (d)
XX.XXX
Aus­wer­tung
Sca­ling
0 % corres ponds
XXXX
Lin. curve
Linear
100 % corres ponds
XXXX
Integr ation time
Dezi­mal point
888.8
0 s
prop. to
Mass
Unit
Kg
Out­puts
mA­output
mA­output
4-20mA
Fail­ure mode
22mA
Sensor displ.
prop. to
Di­stance
44 VEGASON 51K … 53K
Set-up
With these keys you move in the menu field to the left, right, top and bottom
ESC
7. 8.
OK?
OK?
Dis­tance
m (d)
4.700
Update
Meas. dist.
m (d)
X.XX
Update Now!
OK?
Learn­ing!
Simulation:Simulation:
Simulation:
Simulation:Simulation: One hour after the last simulation adjustment the sensor returns automati­cally to normal operating condition.
Error codes:Error codes:
Error codes:
Error codes:Error codes: E Hardware failure E013 No va li d mea sure d val ue
E017 Adju st ment span too small
%
E036 No operating sensor program
False echo memory
Create new
Meas. dist.
m (d)
X.XX
Create new Now!
Learn­ing!
Simu­lation
Simu­lation Now!
Simu­lation
XXX.X
Add'l func­tions
Ampl.:
XX dB
S-N:
XX
Delete
Delete Now!
OK?
Delet­ing!
dB
Info
Sensor Tag
Sensor
Reset to de fault
Reset Now!
OK?
Reset runs!
Sensor type
SON52 K
- Sensor in feeding phase
- Loss of the useful echo
- Sensor must have a new programming (service)
- Fault signal also appears during programming
Lan­guage
Eng­lish
Serial number
1094 0213
Meas. unit
Softw. Vers.
0.40
Dis­tance m
X,XX
Heavy dust
No
Fast change
Yes
m (d)
Softw. date
15.09. 1997
OK
max. range
m (d)
7.000
Menu points in bold print provide sensor and measured value information and cannot be modified in this position.
Light grey menu fields are only displayed if required (depending on the adjustments in other menus).
White menu points can be modi­fied with the "+“ or "–“-key and saved with the "OK“-key .
Dis­tance
m (d)
4.700
Ampl.:
XX dB
S-N:
XX
dB
VEGASON 51K … 53K 45

6.4 Adjustment with the HART®-handheld

With each HART®-handheld you can set-up the VEGASON series 50K ultrasonic sensors like all other HART®-capable sensors. A spe­cial DDD (Data Device Description) is not necessary.
Just connect the HART®-handheld to the signal line, after having connected the sensor to power supply.
+
-
Set-up
Ri < 250
Note:
If the resistor of the power supply is less than 250 Ohm, a resistor must be looped into the signal/connection line during adjustment.
250
Connection to a VEGA-signal condi­tioning instrument
When you operate a HART®-capable sensor on a VEGA-signal conditioning instrument, you have to connect the sensor via a resistor according to the following table during the HART®-adjustment, to reach together with the inner resistor of the instruments the value of 250 Ohm required for the HART®-instrument.
The digital adjustment and communication signals would be short-circuited via too small resistors of the supply current source of the processing system which means that the sensor communication could not be ensured.
+
-
Ri < 250
VEGA-signal conditioning instr. Rx
VEGAMET 513, 514, 515, 602 50 … 100 Ohm
VEGAMET 614 no additional VEGADIS 371 res istor
required
VEGAMET 601 200 … 250 Ohm
VEGASEL 643 150 … 200 Ohm
VEGAMET 513 S4, 514 S4 515 S4, VEGALOG EA-card 100 … 150 Ohm
46 VEGASON 51K … 53K
Set-up
VEGALOG VEGAMET
Rx
The most important adjustment steps
On the following pages you see a menu plan for the HART®-handheld in conjunction with VEGASON 51K … 53K sensors. The most important adjustment steps are marked in the menu plan with the letters A … E. When enter­ing parameters first of all the key " must be pressed. The adjustment is thereby saved in the handheld, but not in the sensor itself.
Generic: SENSOR PV URV
5.850 m
0.300 m
HELP DEL ESC ENTER
After having pressed "
ENTER
(here in the example for the min.-adjustment).
Generic: SENSOR
1 PV LRV 5.850 m 2 PV URV 0.300 m
HELP SEND HOME
After pressing "
SEND
“, a warning is dis­played, which informs you that you are about to modify the configuration and for that, safety reasons, you should switch your sys­tem over to manual operation.
ENTER
4.2 (5.2)
“, press "
SEND
Generic: SENSOR WARN-Loop should be removed from automatic control
ABORT OK
Press "OK“ and the adjustment will now be transferred to the sensor. After a short time you are asked to switch you system over from manual to automatic operation. Confirm with "OK“.
Generic: SENSOR NOTE- Loop may be re­turned to automatic control
OK
You see the actual carried out adjustment.
Generic: SENSOR
1 PV LRV 5.850 m 2 PV URV 0.300 m
HELP HOME
VEGASON 51K … 53K 47
HART®-menu plan VEGASON 51K… 53K
Set-up
Switch on:
Hart Communicator
Self Test
in Progress
Firmware Rev: F2.2 Module Rev: 3.6
01992-96 FRSI
after approx. 20 s
Generic: SENSOR Online(Generic)
1 Device setup 2 PV 2.281 m 3 PV AO 15.059 mA 4 PV LRV 7.000 m 5 PV URV 0.000 m
SAVE
Generic: SENSOR Device setup
1 Process variables 2 Diag/Service 3 Basic setup 4 Detailed setup 5 Review
SAVE HOME
Generic: SENSOR PV
2.281 m
HELP EXIT
Generic: SENSOR AA1
15.264 mA
1
2
3
Generic: SENSOR Process variables
1 Snsr 2.281 m 2 AI % rnge 69,140 % 3 AO 1 15.059 mA
HELP SAVE HOME
Generic: SENSOR Diag/Service
1 T est device 2 Loop test 3 Calibration 4 D/A trim
SAVE HOME
Generic: SENSOR Basic setup
1Tag 2 PV Unit 3 Range values 4 Device information 5 PV xfer fnctn
HELP SAVE HOME
6 PV Damp
Generic: SENSOR Detailed setup
1 Sensors 2 Signal condition 3 Output condition 4 Device information
SAVE HOME
Generic: SENSOR PV LRV
7.000 m
5.850 m
HE LP DE L ESC ENTER
1.1
1.4
not necessary
}
4.1 (5.1)
1.2
1.3
not necessary
linear
1.3.6
1.4.3
AA
A
AA
(max. meas. distance)
1.3.1
1.2.2
}
1.3.2
1.3.4
HELP EXIT
4
5
Generic: SENSOR PV URV
0.000 m
0.300 m
HELP DEL ESC ENTER
4.2 (5.2)
BB
B
BB
(min. meas. distance)
Important and necessary menu windows
Less important menu windows
Generic: SENSOR
1 PV LRV 7.000 m 2 PV URV 0.000 m
HELP HOME
Generic: SENSOR
1 PV LRV 7.000 m 2 PV URV 0.000 m
HELP HOME
Continue as under figure 4
48 VEGASON 51K … 53K
Set-up
Generic: SENSOR PV
2.281 m
HELP EXIT
1.1.1
Generic: SENSOR PV % rnge
70.390 %
HELP EXIT
1.1.2
Generic: SENSOR A01
15.264 mA
1.1.3
Switch off control
Generic: SENSOR Tag
EE
E
EE
SENSOR SENSOR
HELP DEL ESC ENTER
CC
C
CC
Generic: SENSOR Device information
1 Distributor 2 Model 3Dev id 4Tag 5 Date 01/01/97
HELP SAVE HOME
Generic: SENSOR Output condition
1 Analog output 2 HART output
Generic: SENSOR PV Snsr unit m
m bbl in cm
ESC ENTER
1.4.3
1.3.1
1.3.6
1.3.2
Generic: SENSOR Choose analog output level
14mA 2 20mA 3 Other 4End
ABORT ENTE R
Generic: SENSOR Tag
Sensor
Sensor
HE L P DE L ESC ENTER
6 Write protect None 7 Descriptor 8 Message 9 PV Snsr s/n
Final asmbly num Revision #'s
1.2.2
The following menu windows are not supported by the sensor. Saving of adjust­ments carried out here is not possible.
The adjustment of actual "empty“ or "full“ measured values should be carried out in the menu window 4.1 and 4.2. Enter there the meas. distance with empty vessel as zero and the meas. distance with full vessel as end.
HELP EXIT
Individual current values for test purposes (measured value simula­tion). See also menu window 1.4.3.1. on the next page
1.3.4
1.3.4.1
1.3.4.2
1.3.4.5
1.3.4.6
1.3.4.7
1.3.4.9
1.3.4.10
1.3.4.11
1.4.3.1
1.3.4.3
1.3.4.4
1.3.4.8
Sensor serial number
like menu 1.3.1
DD
D
DD
see next page
see next page
SAV E HOME
Generic: SENSOR HART output
1 Poll addr 0 2 Num req preams 0 3 Burst mode 4 Burst option
HELP SAVE HOME
VEGASON 51K … 53K 49
1.4.3.2
Generic: SENSOR Poll addr
0
0
HELP DEL ESC ENTER
Off: Sensor signals measured values only on request On: Sensor signals measured values unrequested
1.4.3.2
Multidrop: The sensor also provides te 4 … 20 mA signal, as well as a digital (HART®)-level signal :
• If the poll address is zero, the sensor takes a level dependent 4…20 mA­current and delivers a digital (HART®)­level signal.
• If the poll address is 1…15, the sensor still delivers a digital (HART®)-level signal, however the sensor current is frozen to 4 mA. The 4…20 mA-signal is then no longer available.
Continuation HART®-menu plan – VEGASON 51K…53K
DD
D
from 1.3.4
DD
Generic: SENSOR Message
..SOL
..FEN04.58M
1.3.4.8
Initialisation word
(enter at last)
Set-up
from
1.4.3.
HE L P DE L ESC ENTER
Generic: SENSOR Analogue output
1 A01 13,467mA 2 A0 Alrm typ 3 Loop test 4 D/A trim 5 Scaled D/A trim
HELP SAVE HOME
newly entered and initialisa­tion word to enquire with ENTER and SEND
Available initialisation words:Available initialisation words:
Available initialisation words:
Available initialisation words:Available initialisation words:
..SOL Meas. condition solid ..LIQ Meas. condition liquid ..FED Delete false echo memory
..FEN04.58M False echo e.g. at 4.58 m create new ..FEN48.67FT False echo e.g. at 48.67 ft create new
..FEU03.68M Extend false echo memory: ..FEU36.05FT add a new false echo to the false echo memory at 36.05 ft
1.4.3.1
Warning: Switch off control!
add a new false echo to the false echo memory at 3.68 m
Note:
After the adjustment of the initialisation word, press "ENTER“ and then "SEND“. Confirm the message to change over to manual operation with "OK“ and the message to switch over to automatic operation again with "OK“. Only then is the adjustment written in the sensor and effective.
Generic: SENSOR A0 Alrm typ High
1.4.3.1.2 Low: Not selectable r (3,6 mA)
High: In case of failure current output
Exit
Generic: SENSOR Choose analogue output level
14 mA 2 20 mA 3 Other 4End
1.4.3.1.3
Abort Enter
goes to < 21,5 mA
Generic: SENSOR Fld dev output is fixed at 4.000 mA
1.4.3.1.3.1
ABORTOK
50 VEGASON 51K … 53K
Diagnose

7 Diagnosis

7.1 Simulation

To simulate a certain filling, you can call up the function "Simulation" on the adjustment module MINICOM, in the software program VVO or in the HART® handheld.
With this function, you simulate a and hence a real sensor current vessel filling level. Please note that connected instruments, such as e.g. a PLC, react according to their adjust­ments and will probably activate alarms or system functions.

7.2 Error codes

Error codes Corrective measure
E013 No valid measured value Message is displayed during the warm-up
- Sensor in the warm-up phase phase
- Loss of the useful echo If the message remains, a false echo storage (menu "Sensor optimisation") must be carried out with the adjustment modul MINICOM or better with the PC and VVO. If the meassage is still displayed, you have to carry out a readjustment.
E017 Adjustment span too small Carry out a readjustment.
E036 Sensor software does not run Sensor requires a software (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tive Transducer defective.
VEGASON 51K … 53K 51
Make sure that the difference between min. and max. adjustment is at least 10 mm .
Message appears during a software update.
Contact our service department.
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Telefon (07836) 50-0 Fax (07836) 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
The statements on types, application, use and operating conditions of
the sensors and processing systems correspond to the actual
knowledge at the date of printing.
Technical data subject to alteration.
21758-EN-020313
Loading...