VEGA PULSWL61 User Manual

Page 1
Operating Instructions
Radar sensor for continuous level measurement of water and wastewater
VEGAPULS WL 61
4 … 20 mA/HART - two-wire
Document ID: 38061
Page 2

Quick start

Mounting
Connect electrically
Quick start
The quick start enables a quick setup with many applications. You can nd further information in the respective chapters of the operating instructions manual.
1. Distance antenna to vessel wall > 200 mm
2. Orientation with mounting strap
> 200 mm
(7.87
")
Fig. 1: Distance antenna to vessel wall, orientation with mounting strap
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica­tions on the type label.
2. Connect the instrument according to the following illustration:
Set parameters
Parameterization example
2
1
2
Fig. 2: Wire assignment x-connected connection cable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding
For further information see chapter "Connecting to power supply".
1. Connect interface adapter
2. Start PACTware and then start the "VEGA project assistant".
3. Start the setup assistant in the DTM window and carry out the predetermined steps.
The radar sensor measures the distance from the sensor to the product surface. For indication of the real level, an allocation of the measured distance to the percentage height must be carried out.
38061-EN-121011
Page 3
Further steps
Quick start
100%
2
")
0,5 m
(19.68
")
5 m
(196.85
0%
1
Fig. 3: Parameterization example
1 Min. level = max. meas. distance 2 Max. level = min. meas. distance
For this adjustment, the distance is entered for min. and max. level. If these values are not known, an adjustment with distances, for example, of 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the ange.
1. In the menu "Additional settings", menu item "Damping" you have to adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out- put".
The quick start is then nished. For further information see chapter "Set up with PACTware".
38061-EN-121011
3
Page 4
Contents
Contents
1 About this document
1.1 Function ............................................................................. 6
1.2 Target group ....................................................................... 6
1.3 Symbolism used ................................................................. 6
2 For your safety
2.1 Authorised personnel ......................................................... 7
2.2 Appropriate use .................................................................. 7
2.3 Warning about incorrect use ............................................... 7
2.4 General safety instructions ................................................. 7
2.5 CE conformity ..................................................................... 8
2.6 NAMUR recommendations ................................................ 8
2.7 Radio license for Europe .................................................... 8
2.8 Environmental instructions ................................................. 8
3 Product description
3.1 Conguration .................................................................... 10
3.2 Principle of operation ........................................................ 11
3.3 Packaging, transport and storage ..................................... 11
3.4 Accessories and replacement parts ................................. 12
4 Mounting
4.1 General instructions ......................................................... 13
4.2 Mounting versions ............................................................ 13
4.3 Mounting preparations, mounting strap ............................ 16
4.4 Instructions for installation ................................................ 17
5 Connecting to power supply
5.1 Preparing the connection ................................................. 23
5.2 Wiring plan ....................................................................... 23
5.3 Switch-on phase ............................................................... 23
6 Set up with VEGADIS 62
6.1 Connection ....................................................................... 25
6.2 Adjust the sensor .............................................................. 25
6.3 Scale the indication .......................................................... 26
7 Setup with PACTware
7.1 Connect the PC ................................................................ 28
7.2 Parameter adjustment with PACTware .............................. 29
7.3 Saving the parameter adjustment data ............................. 31
8 Set up with other systems
8.1 DD adjustment programs ................................................. 32
8.2 Communicator 375, 475 ................................................... 32
9 Diagnosis, Asset Management and service
9.1 Maintenance .................................................................... 33
9.2 Measured value and event memory ................................. 33
9.3 Asset Management function ............................................. 34
9.4 Rectify faults ..................................................................... 37
9.5 Software update ............................................................... 41
9.6 How to proceed in case of repair ...................................... 41
38061-EN-121011
4
Page 5
10 Dismounting
10.1 Dismounting steps ............................................................ 43
10.2 Disposal ........................................................................... 43
11 Supplement
11.1 Technical data .................................................................. 44
11.2 Radio astronomy stations ................................................. 49
11.3 Dimensions ...................................................................... 49
Contents
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
38061-EN-121011
Editing status: 2012-09-27
5
Page 6

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
38061-EN-121011
6
Page 7

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGAPULS WL 61 is a sensor for continuous level measurement. You can nd detailed information on the application range in chapter
"Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula­tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.
Depending on the instrument version, the emitting frequencies are in the C or K band range. The low emitting frequencies are far below the internationally approved limit values. When used correctly, there is no danger to health.
38061-EN-121011
7
Page 8
2 For your safety

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the conformity certicate in the download section of our homepage.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom­mendations:
NE 43 – Signal level for malfunction information from measuring
transducers NE 53 – Compatibility of eld devices and indicating/adjustment
components NE 107 - Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Radio license for Europe

The instrument meets the LPR (Level Probing Radar) radio standard EN 302729-1/2. It is approved for unrestricted use inside and outside of closed vessels in countries of the EU and EFTA that have imple­mented this standard: Austria, Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, Great Britain, Ireland, Island, Italy, Liech­tenstein, Lithuania, Latvia, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Slovenia, Spain, Czech Republik and Cyprus.
Not included in the CE conrmity declaration are the countries imple­menting this radio standard at a later date: Finland and Hungary.
For operation outside of closed vessels, the following conditions must be fullled:
The installation must be carried out by trained qualied personnel
The instrument must be stationary mounted and the antenna
directed vertically downward The mounting location must be at least 4 km away from the radio
astronomy stations listed in the supplement, unless special per­mission was granted by the responsible national approval authority When installed within 4 to 40 km of one of the radio astronomy sta-
tions listed in the supplyment, the instrument must not be mounted higher than 15 m above the ground.

2.8 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro-
38061-EN-121011
8
Page 9
2 For your safety
tection. The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
38061-EN-121011
9
Page 10

3 Product description

Type label
3 Product description
3.1 Conguration
The type label contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 4: Layout of the type label (example)
1 Instrument type 2 Product code
3 Approvals
4 Power supply and signal output, electronics
5 Protection rating 6 Measuring range
7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Note to observe the instrument documentation 15 Notied authority for CE marking 16 Approval directive
16
15
14
13
12
11
Serial number
Scope of this operating instructions manual
10
With the serial number of the instrument on the type label you have access to the following data on our homepage:
Article number of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate "Measuring Accuracy" (PDF)
For this purpose, move to www.vega.com and "VEGA Tools".
This operating instructions manual applies to the following instrument versions:
Hardware from 1.0.0
Software from 4.4.0
38061-EN-121011
Page 11
3 Product description
Scope of delivery
Application area
Functional principle
Packaging
The scope of delivery encompasses:
Radar sensor
Compression ange (option)
Mounting strap with xing material (optional)
Documentation
– this operating instructions manual – Ex-specic "Safety instructions" (with Ex versions) – if necessary, further certicates

3.2 Principle of operation

The radar sensor VEGAPULS WL 61 is the ideal sensor for all appli­cations in the water and waste water industry. It is particularly suitable for level measurement in water treatment, in pump stations as well as storm water overow tanks, for ow measurement in open umes and for gauge measurement.
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The transit time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropri­ate output signal and outputted as measured value.

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport
Transport inspection
Storage
38061-EN-121011
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
11
Page 12
3 Product description
Storage and transport temperature
Interface adapter
External indicating and adjustment unit with HART protocol
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %

3.4 Accessories and replacement parts

The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, an adjustment software such as PACTware with VEGA-DTM is required.
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 62 is suitable for measured value indication and adjustment of sensors with HART protocol. It is looped into the 4 … 20 mA/HART signal cable.
You can nd further information in the operating instructions "VE- GADIS 62" (Document-ID 36469).
12
38061-EN-121011
Page 13
Suitability for the process conditions
Straining clamp

4 Mounting

4 Mounting

4.1 General instructions

Make sure that all parts of the instrument exposed to the process, in particular the active measuring component, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the type label.

4.2 Mounting versions

Most simply mount the instrument via the straining clamp. For this purpose, the connection cable is provided with a strain relief wire of Kevlar.
In order to avoid faulty measured values, make sure that the sensor does not oscillate.
Fig. 5: Mounting via a straining clamp
Mounting bracket
38061-EN-121011
For a rigid mounting, a mounting bracket with opening for thread G1½, e.g. from the VEGA product range, is recommended. The mounting of the sensor in the bracket is carried out via a G1½ counter nut of plastic. Take note of chapter "Mounting instructions" for the distance to the wall.
> 200 mm
(7.87")
13
Page 14
4 Mounting
> 200 mm
(7.87
Fig. 6: Mounting via a mounting bracket
")
Mounting strap
The optional mounting strap enables sensor mounting on e.g. a ceil­ing, wall or bracket. It is available in the following versions:
Length 300 mm for ceiling mounting
Length 170 mm for wall mounting
The instrument is normally mounted vertically on the ceiling. This ensures swivelling of the sensor up to 180° for optimum orienta-
tion.
Fig. 7: Vertical mounting on the ceiling via the mounting strap with length 300 mm
As an alternative, mounting can be carried out vertically on the wall.
38061-EN-121011
14
Page 15
4 Mounting
> 200 mm
(7.87")
Fig. 8: Vertical mounting on the wall via the mounting strap with length 170 mm
Some measuring points have only very little space between ceil­ing and water surface. In such cases, for example in closed storm overow basins, horizontal mounting of the sensor is recommended. The radar impulses must be directed via a 45° reector - for example a stainless steel sheet - to the water surface.
45°
Fig. 9: Horizontal mounting with a mounting strap of length 170 mm with reec­tor provided by the customer
Information:
With this mounting arrangement, the reference plane mentioned in chapter "Technical data" no longer applies. There is an oset that must be taken into account for the adjustment. Note the distance measured at min. level with the reector, for example 2.5 m. Enter this value as min. adjustment/measuring range begin. Determine the dierence between the min. and max. level, for example 1 m. The dis­tance for the max. adjustment results from 2.5 m - 1 m = 1.5 m. Enter this value as max. adjustment/measuring range end.
38061-EN-121011
> 50 mm
(1.97")
15
Page 16
4 Mounting
1
Fig. 10: Horizontal mounting with the optional mounting strap with integrated reector
1 Reference plane
Information:
With this combination (by default), the oset is already taken into ac­count. The lower side of the mountng seal is the reference plane.
Flange
For mounting the instrument on a socket or a manhole cover, an un­assembled combination compression ange is optionally available for DN 80 (ASME 3" or JIS 80), also as a retrotting part. As an alterna­tive, the instrument can be already supplied with a tight, x-mounted adapter ange from DN 100 (ASME 4" or JIS 100).
You can nd drawings of these mounting options in chapter "Dimen- sions".
Fig. 11: Mounting by means of an adapter ange, for example, on a manhole lid.

4.3 Mounting preparations, mounting strap

The optional mounting strap is supplied unassembled. It must be screwed to the sensor before setup with the attached screws. Max. torque, see chapter "Technical data". Required tools: Allen wrench size 4.
There are two dierent ways of screwing the strap to the sensor. Depending on the selected method, the sensor can be rotated in the strap innitely variable through 180° or in three steps 0°, 90° and 180°.
38061-EN-121011
16
Page 17
Tight installation of the plastic horn antenna
4 Mounting
Fig. 12: Rotation in the centre with ceiling mounting
Fig. 13: Adjustment of the angle of inclination in case of wall mounting

4.4 Instructions for installation

For tight installation of the version with plastic horn antenna with com­pression or adapter ange, the following conditions must be fullled:
1. Use suitable at seal, e.g. of EPDM with Shore hardness 25 or 50
2. Make sure the number of ange screws corresponds to the num­ber of ange holes
3. Tighten all screws with the torque stated in the technical data
Polarisation plane
38061-EN-121011
The emitted radar impulses of the radar sensor are electromagnetic waves. The polarisation plane is the direction of the electrical wave component. By turning the instrument in the connection ange or mounting strap, the polarisation can be used to reduce the eects of false echoes.
The position of the polarisation level is marked by marking bars on the instrument.
17
Page 18
4 Mounting
1
Fig. 14: Position of the polarisation level
1 Marking bar
Installation position
When mounting the sensor, keep a distance of at least 200 mm (7.874 in) to the vessel wall. If the sensor is installed in the center of dished or round vessel tops, multiple echoes can arise. These can, however, be suppressed by an appropriate adjustment (see chapter "Setup").
If you cannot keep this distance you should carry out a false echo storage before setup. This applies mainly if buildup on the vessel wall is expected. In this case, we recommend repeating a false echo stor­age later with existing buildup.
> 200 mm
(7.87
")
Fig. 15: Mounting of the radar sensor on round vessel tops
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible down to the lowest point of the vessel bottom.
18
38061-EN-121011
Page 19
Fig. 16: Mounting of the radar sensor on vessels with conical bottom
4 Mounting
Inowingmedium
Socket
Do not mount the instrument in or above the lling stream. Make sure that you detect the product surface, not the inowing product.
Fig. 17: Mounting of the radar sensor with inowing medium
Approximate values of the socket heights are shown in the following illustration. The socket end should be smooth and burr-free, if possi­ble also rounded. After mounting, you have to carry out a false signal memory during the parameter adjustment.
38061-EN-121011
19
Page 20
4 Mounting
Sensor orientation
Vessel installations
300 mm / 11.8
"
"
500 mm / 19.69
600 mm / 23.6
"
800 mm / 31.5
h
"
"
"
"
d
h
d
80 mm / 3.15
100 mm / 4
125 mm / 5"
150 mm / 6
Fig. 18: Deviating socket dimensions
Direct the sensor as perpendicular as possible to the product surface to achieve optimum measurement results.
Fig. 19: Orientation of the sensor
The mounting location of the radar sensor should be a place where no other equipment or xtures cross the path of the microwave signals.
Vessel installations, such as e.g. ladders, limit switches, heating spi­rals, struts, etc., can cause false echoes and impair the useful echo. Make sure when planning your measuring site that the radar sensor has a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal baes above the installations scatter the radar signals and prevent direct interfering reections.
20
38061-EN-121011
Fig. 20: Cover smooth proles with deectors
Page 21
4 Mounting
Foam generation
Flow measurement with
rectangularume
Through the action of lling, stirring and other processes in the vessel, compact foams that considerably damp the emitted signals may form on the product surface.
If foams are causing measurement errors, the biggest possible radar antennas, the electronics with increased sensitivity or low frequency radar sensors (C band) should be used.
As an alternative, sensors with guided microwave can be used. These are unaected by foam generation and are best suited for such ap­plications.
The short examples give you introductory information on the ow measurement. Detailed planning information is available from ume manufacturers and in special literature.
3 ... 4 h
90°
2 3
Fig. 21: Flow measurement with rectangular ume: d sensor (see chapter "Technical data"); h ume
1 Overow orice (side view) 2 Headwater
3 Tail water
4 Overfall orice (view from bottom water)
1
max
≥ 50 mm
min
d
h
≥ 2 mm x h
90°
max
max
4
= min. distance of the
min.
= max. lling of the rectangular
max.
In general, the following points must be observed:
Install the sensor on the headwater side
Installation in the centre of the ume and vertical to the liquid
surface Distance to the overfall orice
Distance of orice opening above ground
Min. distance of the orice opening to bottom water
Min. distance of the sensor to max. storage level
38061-EN-121011
21
Page 22
4 Mounting
Flow measurement with
KhafagiVenturiume
3 ... 4 x h
d
max
90°
h
max
2
1
Fig. 22: Flow measurement with Khafagi-Venturi ume: d = Min. distance to sen­sor; h
= max. lling of the ume; B = tightest constriction in the ume
max.
1 Position sensor
2 Venturi ume
B
In general, the following points must be observed:
Installation of the sensor at the inlet side
Installation in the centre of the ume and vertical to the liquid
surface Distance to the Venturi ume
Min. distance of the sensor to max. storage level
22
38061-EN-121011
Page 23
Safety instructions

5 Connecting to power supply

5 Connecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
Voltage supply
Connection cable
Wire assignment, con­nection cable
Power supply and current signal are carried on the same two-wire cable. The voltage supply range can dier depending on the instru­ment version.
The data for power supply are specied in chapter "Technical data". Provide a reliable separation between the supply circuit and the
mains circuits according to DIN VDE 0106 part 101. Keep in mind the following additional factors that inuence the operat-
ing voltage:
Output voltage of the power supply unit can be lower under nomi-
nal load (with a sensor current of 20.5 mA or 22 mA in case of fault message) Inuence of additional instruments in the circuit (see load values in
chapter "Technical data")
The instrument is connected with standard two-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326-1 for industrial areas, screened cable should be used.
For instruments with housing and cable gland, use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland. If you are using cable with a dierent diameter, exchange the seal or use a suitable cable gland.
We generally recommend the use of screened cable for HART multi­drop mode.

5.2 Wiring plan

1
Fig. 23: Wire assignment x-connected connection cable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding

5.3 Switch-on phase

After connecting the instrument to power supply or after a voltage recurrence, the instrument carries out a self-check for approx. 30 s:
38061-EN-121011
2
23
Page 24
5 Connecting to power supply
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC Indication of the status message "F 105 Determine measured
value" on the display or PC The output signal jumps to the set error current
As soon as a plausible measured value is found, the corresponding current is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory setting.
24
38061-EN-121011
Page 25

6 Set up with VEGADIS 62

6 Set up with VEGADIS 62

6.1 Connection

The VEGADIS 62 is an indicating and adjustment unit without external energy for looping into 4 … 20 mA/HART circuits.
The parameter adjustment of the sensor is carried out via HART com­munication. During the parameter adjustment, the VEGADIS 62 acts as a Secondary Master with respect to the sensor.
=
~
2
1
Fig. 24: Connection of VEGADIS 62 to the sensor
1 Sensor 2 VEGADIS 62
3 HART resistance > 150 Ω (necessary with low impedance power supply) 4 Voltage supply/Processing
The following adjustment volume of the connected HART sensor is available:
Min./Max. adjustment
zero/span adjustment (live adjustment)
Damping
43

6.2 Adjust the sensor

Proceed as follows for the min./max. adjustment of the sensor:
1. Press "OK" to reach the adjustment menu.
2. Select the submenu "Measurement" and conrm with "OK".
3. Move to the menu item "Unit". There the instrument unit of the sensor is displayed, for example "m".
4. Move to the menu item "MB begin", there the max. measuring distance is displayed, for example the default setting 15 m.
38061-EN-121011
25
Page 26
6 Set up with VEGADIS 62
5. Edit value via "OK" and adjust the requested value, for example, 5 m.
6. Save value with "OK", VEGAPULS WL 61 displays briey "Wait", then the value is taken over into the sensor.
7. Move to the menu item "MB end", there the min. measuring dis­tance is displayed, for example the default setting 0 m.
8. Proceed accordingly for "MB end", enter for example the value 1 m and store.
The min./max. adjustment is nished. After "[ESC]", the display shows the actually measured distance as
digital value in m and the level on the bargraph. Keep in mind that the displayed values are anticyclical. With increas-
ing distance, the 4 … 20 mA value gets smaller and vice versa.
26

6.3 Scale the indication

Proceed as follows for indication of the level as digital value in %:
1. Press "OK" to reach the adjustment menu.
2. Select the submenu "Measurement" and conrm with "OK".
3. Select the menu item "Unit"
38061-EN-121011
4. Select the unit "USER" and conrm with "OK".
Page 27
6 Set up with VEGADIS 62
Information:
The symbol for HART extinguishes, the HART communication is switched o.
5. Select the submenu "Indication" and conrm with "OK".
6. Select the menu item "Unit" and conrm with "OK".
7. Select the unit "%" and conrm with "OK".
The conversion to level indication in % is nished. The unit in the menu "Measurement" must remain at "USER".
After "[ESC]", the display shows the level as digital value in % and on the bargraph. The indication values are now synchronous.
38061-EN-121011
27
Page 28

7 Setup with PACTware

7 Setup with PACTware

7.1 Connect the PC

Via interface adapter to the signal cable
2 4
LOCK
USB
TWIST
3
OPEN
1
5
Fig. 25: Connecting the PC to the signal cable
1 Sensor
2 HART resistance 250 Ω (optional depending on processing) 3 Connection cable with 2 mm pins and terminals 4 Processing system/PLC/Voltage supply 5 Interface adapter, for example VEGACONNECT 4
Note:
With power supply units with integrated HART resistance (internal resistance approx. 250 Ω), an additional external resistance is not necessary. This applies, e.g. to the VEGA instruments VEGATRENN 149A, VEGAMET 381, VEGAMET 391. Common Ex separators are also usually equipped with a sucient current limitation resistance. In such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration).
28
38061-EN-121011
Page 29
Via interface adapter to the VEGAMET signal con­ditioning instrument
Prerequisites
7 Setup with PACTware
3
1
2
1
2
2
TWIST
LOCK
1
OPEN
4
USB
%
on
VEGAMET 381
Fig. 26: Connection of the PC to the VEGAMET signal conditioning instrument
1 Sensor
2 Connection cable with 2 mm pins 3 Signal conditioning instrument, e.g. VEGAMET 381 4 Interface adapter, for example VEGACONNECT 4

7.2 Parameter adjustment with PACTware

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
38061-EN-121011
29
Page 30
7 Setup with PACTware
Fig. 27: Example of a DTM view
Adjustment
Since a radar sensor is a distance measuring instrument, the distance from the sensor to the product surface is measured. For indication of the real level, an allocation of the measured distance to the percent­age height must be carried out. To perform the adjustment, enter the distance with full and empty vessel, see the following example:
100%
5 m
")
(196.85
0%
2
1
Fig. 28: Parameterization example
1 Min. level = max. meas. distance 2 Max. level = min. meas. distance
0,5 m
")
(19.68
38061-EN-121011
30
Page 31
7 Setup with PACTware
If these values are not known, an adjustment with the distances of for example 10 % and 90 % is possible. Starting point for these distance specications is always the seal surface of the thread or ange. By means of these settings, the real level will be calculated.
The real product level during this adjustment is not important, be­cause the min./max. adjustment is always carried out without chang­ing the product level. These settings can be made ahead of time without the instrument having to be installed.
Standard/Full version
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega. com/downloads and "Software". The full version is available on CD from the agency serving you.

7.3 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
38061-EN-121011
31
Page 32

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".

8.2 Communicator 375, 475

Device descriptions for the instrument are available as DD or EDD for parameter adjustment with the Field Communicator 375 or 475.
The les can be downloaded unter www.vega.com/downloads and "Software".
32
38061-EN-121011
Page 33

9 Diagnosis, Asset Management and service

9 Diagnosis, Asset Management and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal operation.

9.2 Measured value and event memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
Event memory
Echo curve memory
Up to 60,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switching on and o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control system with EDD.
The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup. Changes in the measure­ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via:
PC with PACTware/DTM
Control system with EDD
Indicating and adjustment module
38061-EN-121011
33
Page 34
9 Diagnosis, Asset Management and service
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD

9.3 Asset Management function

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the indicating and adjustment module, PACTware/DTM and EDD.
Status messages
The status messages are classied in the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 29: Pictograms of the status messages
1 Failure - red 2 Function check - orange 3 Out of specication - yellow 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is outputted.
This status message is always active. It cannot be deactivated by the user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
38061-EN-121011
34
Page 35
9 Diagnosis, Asset Management and service
Failure
The following table shows the error codes and text messages in the status message "Failure" and gives information on the cause and how to eliminate it. Keep in mind that some specications are only valid for four-wire instruments and the electronics of VEGAPULS WL 61 can­not be exchanged by the user.
Code
Text mes­sage
F013
no measured value avail­able
F017
Adjustment span too small
F025 Error in the
linearization table
F036
No operable software
F040 Error in the
electronics
F080 – General software error – Separate operating voltage
F105
Determine
measured value
F113
Communica-
tion error
F125 Unpermissi-
ble electronics temperature
F260
Error in the calibration
Cause Rectication
– Sensor does not detect an
echo during operation
– Antenna system contami-
nated or defective
– Adjustment not within
specication
– Index markers are not con-
tinuously rising, for examle unlogical value pairs
– Failed or interrupted soft-
ware update
– Hardware defect – Exchanging the electronics
– The instrument is still in the
start phase, the measured value could not yet be determined
– EMC interferences – Transmission error with the
external communication with 4-wire power supply
unit
– Temperature of the elec-
tronics in the non-specied
section
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Check or correct installation
and/or parameter adjust-
ment
– Clean or exchange process
component or antenna
– Change adjustment accord-
ing to the limit values (dif­ference between min. and max. ≥ 10 mm)
– Check linearization table – Delete table/Create new
– Repeat software update – Check electronics version – Exchanging the electronics – Send instrument for repair
– Send instrument for repair
briey
– Wait for the warm-up phase – Duration depending on
the version and parameter adjustment up to approxi­mately 3 min.
– Remove EMC inuences – Exchange 4-wire power
supply unit or electronics
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
– Exchanging the electronics – Send instrument for repair
38061-EN-121011
35
Page 36
9 Diagnosis, Asset Management and service
Function check
Outofspecication
Code
Text mes­sage
F261
Error in the conguration
F264 Installation/
Setup error
F265 Measurement
function dis­turbed
Cause Rectication
– Error during setup – False signal suppression
faulty
– Error when carrying out a
reset
– Adjustment not within the
vessel height/measuring range
– Max. measuring range of
the instrument not sucient
– Sensor no longer carries
out a measurement
– Operating voltage too low
– Repeat setup – Repeat reset
– Check or correct installation
and/or parameter adjust-
ment
– Use an instrument with big-
ger measuring range
– Check operating voltage – Carry out a reset – Separate operating voltage
briey
The following table shows the error codes and text messages in the status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700 Simulation ac-
tive
Cause Rectication
– A simulation is active – Finish simulation
– Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the status message "Out of specication" and provides information on causes as well as corrective measures.
Code
Text mes­sage
S600
Unpermissi­ble electronics temperature
S601
Overlling
Cause Rectication
– Temperature of the elec-
tronics in the non-specied
section
– Danger of vessel overlling – Make sure that there is no
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
further lling
– Check level in the vessel
Maintenance
36
The following table shows the error codes and text messages in the status message "Maintenance" and provides information on causes as well as corrective measures.
38061-EN-121011
Page 37
9 Diagnosis, Asset Management and service
Code
Text mes­sage
M500 Error with the
reset delivery
status
M501 Error in the
non-active linearization table
M502 Error in the
diagnosis memory
M503 Reliability too
low
M504
Error on an
device inter­face
M505 No echo avail-
able
Cause Rectication
– With the reset to delivery
status, the data could not be restored
– Hardware error EEPROM – Exchanging the electronics
– Hardware error EEPROM – Exchanging the electronics
– The echot/noise ratio is the
small for a reliable meas-
urement
– Hardware defect – Check connections
– Level echo can no longer
be detected
– Repeat reset – Load XML le with sensor
data into the sensor
– Send instrument for repair
– Send instrument for repair
– Check installation and
process conditions
– Clean the antenna – Change polarisation direc-
tion
– Use instrument with higher
sensitivity
– Exchanging the electronics – Send instrument for repair
– Clean the antenna – Use a more suitable
antenna/sensor
– Remove possible false
echoes
– Optimize sensor position
and orientation

9.4 Rectify faults

Reaction when malfunc­tions occur
Procedure for fault recti-
cation
Check the 4 … 20 mA signal
38061-EN-121011
The operator of the system is responsible for taking suitable meas­ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the indicating and
adjustment module Checking the output signal with 4 … 20 mA instruments
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
Connect a handmultimeter in the suitable measuring range according to the wiring plan. The following table describes possible errors in the current signal and helps to remove them:
37
Page 38
9 Diagnosis, Asset Management and service
Error Cause Rectication
4 … 20 mA signal not stable
4 … 20 mA signal missing
Current sig­nal greater than 22 mA or less than 3.6 mA
– Level uctua-
tions
– Electrical con-
nection faulty
– Voltage supply
missing
– Operating volt-
age too low or load resistance too high
– Electronics
module in the sensor defec­tive
– Set damping according to the
instrument via the indicating and adjustment module or PACTware/ DTM
– Check connection according to
chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan".
– Check cables for breaks; repair if
necessary
– Check, adapt if necessary
– Exchange the instrument or send it
in for repair
Treatment of measure­ment errors with liquids
The below tables show typical examples of application-related meas­urement errors with liquids. The measurement errors are dierentiated according to the following:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
Level
1
2
0
1 Real level
2 Level displayed by the sensor
time
Instructions:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
38061-EN-121011
38
Page 39
9 Diagnosis, Asset Management and service

9.4 Measurement error with constant level

Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
Level
0
– Min./max. adjustment not
correct
– Wrong linearization curve – Adapt linearization curve
time
– Installation in a bypass tube or
standpipe, hence running time error (small measurement error close to 100 %/large error close
– Adapt min./max. adjustment
– Check parameter "Application"
with respect to vessel form, adapt if necessary (bypass, standpipe, diameter)
to 0 %)
2. Measured value jumps towards 0 %
3. Measured value jumps towards 100 %
Level
– Multiple echo (vessel top,
product surface) with amplitude higher than the level echo
0
Level
time
– Due to the process, the ampli-
tude of the product echo sinks
– Check parameter "Application",
especially vessel top, product type, dished end, high dielectric gure, adapt if necessary
– Carry out false signal suppres-
sion
– A false signal suppression was
0
not carried out
time
– Amplitude or position of a
false echo has changed (e.g. condensation, buildup); false signal suppression no longer
– Determine the reason for the
changed false echo, carry out false signal suppression, e.g. with condensation
matches
9.4 Measurementerrorduringlling
Fault description Error pattern Cause Rectication
4. Measured value re­mains unchanged during lling
5. Measured value re­mains in the bottom section during lling
6. Measured value re­mains momentarily unchanged during ll­ing and then jumps to the correct level
Level
– False echoes in the close range
too big or product echo too small
0
– Strong foam or spout genera-
time
tion
– Max. adjustment not correct
Level
– Echo from the tank bottom
larger than the product echo, for example, with products with
0
Level
0
εr < 2.5 oil-based, solvents
time
– Turbulence on the product
surface, quick lling
time
– Eliminate false echoes in the
close range
– Check measurement situation:
Antenna must protrude out of the socket, installations
– Remove contamination on the
antenna
– Minimize interfering instal-
lations in the close range by changing the polarization direction
– Create a new false signal sup-
pression
– Adapt max. adjustment
– Check application parameters
Medium, Vessel height and Floor form, adapt if necessary
– Check application parameters,
change if necessary, e.g. in dosing vessel, reactor
38061-EN-121011
39
Page 40
9 Diagnosis, Asset Management and service
Fault description Error pattern Cause Rectication
7. Measured value jumps towards 0 % during lling
8. Measured value jumps towards 100 % during lling
9. Measured value jumps sporadically to 100 % during lling
10. Measured value jumps to ≥ 100 % or 0 m distance
Level
– Amplitude of a multiple echo
(vessel top - product surface) is larger than the level echo
0
time
– The level echo cannot be distin-
guished from the false echo at a false echo position (jumps to multiple echo)
Level
– Due to strong turbulence and
foam generation during lling, the amplitude of the product
0
Level
0
Level
echo sinks. Measured value
time
jumps to the false echo
– Varying condensation or con-
tamination on the antenna
time
– Level echo is no longer
detected in the close range due to foam generation or false echoes in the close range. The sensor goes into overll protec­tion mode. The max. level (0 m distance) as well as the status message "Overll protection" are outputted.
– Check parameter "Application",
especially vessel top, product type, dished end, high dielectric gure, adapt if necessary
– Remove/reduce false echo:
minimize interfering installa­tions by changing the polariza­tion direction
– Chose a more suitable installa-
tion position
– Carry out false signal suppres-
sion
– Carry out a false signal sup-
pression or increase false signal suppression with con­densation/contamination in the close range by editing
– Check measuring site: Antenna
must protrude out of the socket
– Remove contamination on the
antenna
– Use a sensor with a more suit-
able antenna

9.4 Measurement error during emptying

Fault description Error pattern Cause Rectication
11. Measured value re­mains unchanged in the close range during emptying
12. Measured value jumps towards 0 % during emptying
40
Level
– False signal larger than the
level echo
– Level echo too small
0
time
Level
– Echo from the tank bottom
larger than the product echo, for example, with products with
0
εr < 2.5 oil-based, solvents
time
– Remove false echoes in the
close range. Check: Antenna must protrude out of the socket
– Remove contamination on the
antenna
– Minimize interfering instal-
lations in the close range by changing the polarization direction
– After removing the false ech-
oes, the false signal suppres­sion must be deleted. Carry out a new false signal suppression
– Check application parameters
Medium type, Vessel height and Floor form, adapt if necessary
38061-EN-121011
Page 41
9 Diagnosis, Asset Management and service
Fault description Error pattern Cause Rectication
13. Measured value jumps sporadically to­wards 100 % during emptying
Level
0
– Varying condensation or con-
tamination on the antenna
time
– Carry out false signal suppres-
sion or increase false signal suppression in the close range by editing
– With bulk solids, use radar sen-
sor with purging air connection
Reaction after fault recti-
cation
24 hour service hotline
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside the normal working hours on seven days a week around the clock.
Since we oer this service worldwide, the support is in the English language. The service itself is free of charge, the only costs involved are the normal call charges.

9.5 Software update

The following components are required to update the sensor soft­ware:
Sensor
Voltage supply
Interface adapter VEGACONNECT 4
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".

9.6 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to proceed under www.vega.com/downloads and "Forms and certi- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
38061-EN-121011
41
Page 42
9 Diagnosis, Asset Management and service
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
42
38061-EN-121011
Page 43

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
38061-EN-121011
43
Page 44

11 Supplement

11 Supplement

11.1 Technical data

General data
Materials, wetted parts
Ʋ Adapter ange PP Ʋ Seal, adapter ange FKM (COG VI500), EPDM (COG AP310) Ʋ Antenna PBT-GF 30 Ʋ Focussing lense PP
Materials, non-wetted parts
Ʋ Compression ange PP Ʋ Mounting strap 316L Ʋ Fixing screws, mounting strap 316L Ʋ Fixing screws, adapter ange 304 Ʋ Housing plastic PBT (Polyester) Ʋ type label support on cable PE hard
Process tting, mounting thread on the housing
Ʋ Flange DIN from DN 80, ANSI from 3", JIS from DN 100 10K Ʋ Pipe thread, cylindrical (ISO 228 T1) G1½
Instrument weight, depending on pro­cess tting
Weight suspension cable 0.1 kg/m (0.07 lbs/ft) Max. torque for mounting strap on sensor
housing Max. torque ange screws
Ʋ Compression ange DN 80 5 Nm (3.689 lbf ft) Ʋ Adapter ange DN 100 7 Nm (5.163 lbf ft)
0.7 … 3.4 kg (1.543 … 7.496 lbs)
4 Nm
Input variable
Measured variable The measured variable is the distance between the pro-
44
cess tting of the sensor and the product surface. The reference plane is the lower side of the ange.
38061-EN-121011
Page 45
11 Supplement
Fig. 44: Data of the input variable
1 Reference plane 2 Measured variable, max. measuring range
3 Antenna length
4 Useful measuring range
3
2
4
1
Max. measuring range 15 m (49.21 ft)
Output variable
Output signal 4 … 20 mA/HART Fullled HART specication 7.0 Signal resolution 0.3 µA Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA Max. output current 22 mA Starting current ≤ 3.6 mA; ≤ 10 mA for 5 ms after switching on Load see load diagram under Power supply Damping (63 % of the input variable),
0 … 999 s
adjustable HART output values according to HART 7.0
1)
Ʋ PV (Primary Value) Distance to the level Ʋ SV (Secondary Value) Level as percentage value Ʋ TV (Third Value) Linearised percentage value Ʋ QV (Fourth Value) Scaled measured value
Resolution, digital < 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
1)
Default values, can be assigned individually
38061-EN-121011
45
Page 46
11 Supplement
Installation reference conditions
Ʋ Min. distance to installations > 200 mm (7.874 in) Ʋ Reector Plane plate reector Ʋ False reections Largest false echo 20 dB smaller than the useful echo
Deviation with liquids See following diagrams
10 mm (0.394 in)
2 mm (0.079 in)
- 2 mm (- 0.079 in)
- 10 mm (- 0.394 in)
Fig. 45: Deviation under reference conditions
1 Reference plane 2 Antenna edge 3 Recommended measuring range
0
0,5 m (1.6 ft)
1
2
3
Reproducibility ≤ ±1 mm
Variablesinuencingmeasurementaccuracy
SpecicationsapplytotheHARTsignalandthecurrentoutput
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
max. 10 mm
Additional deviation through strong, high
< ±50 mm frequency electromagnetic elds acc. to EN 61326
Specicationsapplyalsotothecurrentoutput
Temperature drift - Current output ±0.03 %/10 K relating to the 16 mA span max. ±0.3 % Deviation on the current output by ana-
< ±15 µA logue/digital conversion
Deviation on the current output due to
< ±150 µA strong, high frequency electromagnetic elds acc. to EN 61326
Characteristics and performance data
Measuring frequency K-band (26 GHz technology) Measuring cycle time approx. 450 ms Step response time
2)
≤ 3 s
38061-EN-121011
2)
Time span after a sudden distance change of max. 0.5 m until the output signal reaches for the rst time 90% of
the nal value (IEC 61298-2).
46
Page 47
11 Supplement
Tracking speed of the measuring window
1 m/min max.
Beam angle Emitted HF power
Ʋ Average spectral transmission power
3)
4)
10°
-34 dBm/MHz EIRP
density
Ʋ Max. spectral transmission power
+6 dBm/50 MHz EIRP
density
Ʋ Max. power density in a distance of
< 1 µW/cm²
1 m
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
For the process conditions, please also note the specications on the type label. The lower value always applies.
Vessel pressure -1 … 2 bar (-100 … 200 kPa/-14.5 … 29.0 psig) Process temperature (measured on the
-40 … +80 °C (-40 … +176 °F)
process tting) Vibration resistance
Ʋ With adapter ange 2 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ with mounting strap 1 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance) Shock resistance 100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Electromechanical data - version IP 66/IP 68 (2 bar)
Cable entry IP 68 cable gland Connection cable
Ʋ Conguration two wires, one Kevlar cable, braiding, cover Ʋ Wire cross-section 0.5 mm² (AWG 20) Ʋ Standard length 6 m (19.69 ft) Ʋ Max. length 550 m (1804 ft) Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Wire isolating and cable cover PUR Ʋ Colour - standard Black Ʋ Colour - Ex-version Blue Ʋ Fire protection classication UL94-V0
3)
Outside the specied beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)
4)
EIRP: Equivalent Isotropic Radiated Power
38061-EN-121011
47
Page 48
11 Supplement
Integrated clock
Date format Day.Month.Year Time format 12 h/24 h Time zone Ex factory CET
Measurement electronics temerature
Resolution 1 °C (1.8 °F) Accuracy ±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 9.6 … 36 V DC Ʋ Ex-ia instrument 9.6 … 30 V DC
Interpolation protection Integrated Permissible residual ripple - Non-Ex, Ex-ia instrument
Ʋ for 9.6 V< U Ʋ for 18 V< U
< 14 V ≤ 0.7 Ve (16 … 400 Hz)
N
< 36 V ≤ 1.0 Ve (16 … 400 Hz)
N
Load see diagram
1200
927
750
500
250
9,6
Fig. 46: Voltage diagram
1 HART load 2 Voltage limit Ex-ia instrument 3 Voltage limit non-Ex/Ex-d instrument
4 Operating voltage
1
12 18
16
14 2015,12224262830323436
Electrical protective measures
Protection rating IP 66/IP 68 (2 bar) Overvoltage category III Protection class III
3
2
4
V
Approvals
Instruments with approvals can have dierent technical data depending on the version. For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".
48
38061-EN-121011
Page 49
11 Supplement

11.2 Radio astronomy stations

The following table shows the geographic position of the radio astronomy stations in Europe:
Country Name of the Station Geographic Latitude Geographic Longitude
Finland Metsähovi 60°13'04'' N 24°23'37'' E
Tuorla 60°24'56'' N 24°26'31'' E
France Plateau de Bure 44°38'01'' N 05°54'26'' E
Floirac 44°50'10'' N 00°31'37'' W
Germany Eelsberg 50°31'32'' N 06°53'00'' E
Hungary Penc 47°47'22'' N 19°16'53'' E
Italy Medicina 44°31'14" N 11°38'49" E
Noto 36°52'34" N 14°59'21" E
Sardinia 39°29'50" N 09°14'40" E
Poland Krakow- Fort Skala 50°03'18" N 19°49'36" E
Russia Dmitrov 56°26'00" N 37°27'00" E
Kalyazin 57°13'22" N 37°54'01" E
Pushchino 54°49'00" N 37°40'00" E
Zelenchukskaya 43°49'53" N 41°35'32" E
Spain Yebes 40°31'27" N 03°05'22" W
Robledo 40°25'38" N 04°14'57" W
Switzerland Bleien 47°20’26" N 08°06’44" E
Sweden Onsala 57°23’45" N 11°55’35" E
UK Cambridge 52°09'59" N 00°02'20" E
Darnhall 53°09'22" N 02°32'03" W
Jodrell Bank 53°14'10" N 02°18'26" W
Knockin 52°47'24" N 02°59'45" W
Pickmere 53°17'18" N 02°26'38" W

11.3 Dimensions

The following dimensional drawings represent only an extract of the possible versions. Detailed dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
38061-EN-121011
49
Page 50
11 Supplement
VEGAPULS WL 61, basic version
(0.73")
18,5 mm
G1 1/2
ø 8 mm
ø 75 mm
(2.95")
ø 115 mm
(4.53")
(0.32")
42,5 mm
ø 72 mm
(2.84")
(1.67")
19 mm
(0.75")
(0.59")
15 mm
(11.81")
300 mm
Fig. 47: VEGAPULS WL 61, basic version
50
38061-EN-121011
Page 51
VEGAPULS WL 61, version with mounting strap
2,5 mm
(0.10")
170 mm
(6.69")
(0.34")
8,5 mm
125 mm
(4.92")
11 Supplement
(11.81")
300 mm
(3.86")
98 mm
(0.75")
19 mm
9 mm
(0.47")
12 mm
9 mm
(0.35")
(0.35")
12 mm (0.47")
(3.35")
85 mm
Fig. 48: VEGAPULS WL 61, version with mounting strap in 170 or 300 mm length
38061-EN-121011
75 mm (2.95")
107 mm
(4.21")
115 mm
(4.53")
(0.59")
15 mm
51
Page 52
11 Supplement
VEGAPULSWL61,versionwithmountingstrapandreector
60 mm (2.36")
(6.85")
174 mm
(0.69")
17,5 mm
(2.8")
71 mm
ø 9 mm (0.35")
110 mm
(4.33")
160 mm
(6.30")
Fig. 49: VEGAPULS WL 61, version with mounting strap and reector
107,5 mm
(4.23")
117,5 mm
(4.63")
392 mm (15.43")
115,5 mm (4.55")
218 mm (8.58")
45°
38061-EN-121011
52
Page 53
VEGAPULSWL61,versionwithcompressionange
(0.41")
10,5 mm
ø 21 mm
(0.83")
(4.96")
126 mm
11 Supplement
(0.75")
19 mm
ø 156 mm (6.14")
ø 200 mm (7.87")
Fig. 50: VEGAPULS WL 61, compression ange DN 80/3"/JIS80
38061-EN-121011
53
Page 54
11 Supplement
VEGAPULSWL61,versionwithadapterange
138 mm (5.43")
(1.22")
31 mm
1
(0.59")
15 mm
(0.79")
20 mm
ø 98 mm (3.86")
8 mm
(0.32")
2
Fig. 51: VEGAPULS WL 61, adapter ange DN 100/4"/JIS 100 as well as DN 150/6"/JIS 150
1 Adapter ange
2 Seal
54
38061-EN-121011
Page 55
11 Supplement

11.4 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

11.5 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/ originator.
38061-EN-121011
55
Page 56
INDEX
INDEX
A
Accessories
– External indicating and adjustment unit 12 – Interface adapter 12
Application area 11
C
Check signal 37 Connection cable 23
D
DD (Device Description) 32
E
Echo curve memory 33 EDD (Enhanced Device Description) 32 Error messages 34 Event memory 33
F
Fault rectication 37 Flow measurement
– Khafagi-Venturi ume 22 – Rectangular ume 21
Foam generation 21 Functional principle 11
H
HART resistor 28
I
Inowing medium 19 Installation
– Angle 13 – Bracket 14 – Straining clamp 13
Installation position 18
P
Packaging 11 Polarisation plane 17
R
Repair 41
S
Sensor orientation 20
Service hotline 41 Socket 19 Status messages 34 Storage 11
T
Type label 10
V
Vessel installations 20 Voltage supply 23
M
Measured value memory 33 Measurement error 38 Mounting
– Flange 16
N
NAMUR NE 107
– Failure 35 – Function check 36 – Maintenance 36 – Out of specication 36
56
38061-EN-121011
Page 57
Notes
38061-EN-121011
57
Page 58
Notes
58
38061-EN-121011
Page 59
Notes
38061-EN-121011
59
Page 60
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2012
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
38061-EN-121011
Loading...