VEGA PULSSR68 User Manual

Operating Instructions
Radar sensor for continuous level measurement of bulk solids
VEGAPULS SR 68
Probus PA
Document ID: 38296

Quick start

Mounting
Electrical connection
Quick start
The quick start procedure enables a quick setup with many applica­tions. You can nd further information in the respective chapters of the operating instructions manual.
1. Distance from the vessel wall > 200 mm, the antenna should protrude > 10 mm into the vessel
200 mm
(7.87
")
Fig. 1: Distance of the antenna to the vessel wall/vessel ceiling
2. Note min. socket diameter depending on the socket length
3. Note the instructions for sealing
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica­tions on the type label.
2. Connect the instrument according to the following illustration
00
0
1
1
9
9
2
2
1
8
8
3
3
1
7
7
4
4
6
6
5
5
0
Bus
ca. 10 mm
2
3
4
Set parameters
2
+
( )
(-)
1
2
5
678
5
1
Fig. 2: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Selection switch for bus address 4 For external display and adjustment unit 5 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
1. Go via the display and adjustment module to the menu "Setup".
2. Select in the menu item "Medium" the medium of your applica­tion, for example "Powder/Dust".
38296-EN-130624
Quick start
3. Select in the menu item "Application" the vessel, the application and the vessel form, for example, silo.
4. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment".
Parameterization example
F
ther steps
ur
The radar sensor as a distance measuring instrument measures the
distance from the sensor to the product surface. For indication of the real level, an allocation of the measured distance to the percentage height must be carried out.
")
1 m
(39.37
100%
2
")
35 m
(1378
0%
1
Fig. 3: Parameterization example 1 Min. level max. meas. distance
2 Max. level min. meas. distance
For this adjustment, the distance is entered for min. and max. level. If these values are not known, an adjustment with distances, for example, of 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the thread or ange.
1. In the menu "Additional settings", menu item "Damping" you can adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out- put".
The quick start is then nished. For further information see chapter "Parameter adjustment".
38296-EN-130624
3
Contents
Contents
1 About this document
1.1 Function ........................................................................................................................... 6
1.2 Target group ..................................................................................................................... 6
1.3 Symbolism used ............................................................................................................... 6
2 For your safety
2.1 Authorised personnel ....................................................................................................... 7
2.2 Appropriate use ................................................................................................................ 7
2.3 Warning about incorrect use ............................................................................................. 7
2.4 General safety instructions ............................................................................................... 7
2.5 CE conformity ................................................................................................................... 8
2.6 NAMUR recommendations .............................................................................................. 8
2.7 Radio license for Europe .................................................................................................. 8
2.8 Radio license for USA/Canada ......................................................................................... 8
2.9 Environmental instructions ............................................................................................... 9
3 Product description
3.1 Conguration .................................................................................................................. 10
3.2 Principle of operation...................................................................................................... 11
3.3 Packaging, transport and storage ................................................................................... 11
3.4 Accessories and replacement parts ............................................................................... 12
4 Mounting
4.1 General instructions ....................................................................................................... 14
4.2 Mounting preparations - Horn antenna ........................................................................... 14
4.3 Mounting preparations - Parabolic antenna .................................................................... 15
4.4 Mounting instructions ..................................................................................................... 16
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 30
5.2 Connecting ..................................................................................................................... 31
5.3 Wiring plan, single chamber housing.............................................................................. 32
5.4 Wiring plan, double chamber housing ............................................................................ 33
5.5 Wiring plan, double chamber housing Ex d .................................................................... 34
5.6 Wiring plan - version IP 66/IP 68, 1 bar ........................................................................... 36
5.7 Switch-on phase............................................................................................................. 36
6 Set up with the display and adjustment module
6.1 Insert display and adjustment module ............................................................................ 37
6.2 Adjustment system ......................................................................................................... 38
6.3 Parameter adjustment .................................................................................................... 39
6.4 Saving the parameter adjustment data ........................................................................... 52
7 Setup with PACTware
7.1 Connect the PC .............................................................................................................. 54
7.2 Parameter adjustment .................................................................................................... 54
7.3 Saving the parameter adjustment data ........................................................................... 55
8 Set up with other systems
8.1 DD adjustment programs ............................................................................................... 56
9 Diagnosis, Asset Management and service
9.1 Maintenance .................................................................................................................. 57
38296-EN-130624
4
9.2 Measured value and event memory ............................................................................... 57
9.3 Asset Management function ........................................................................................... 58
9.4 Rectify faults ................................................................................................................... 61
9.5 Exchanging the electronics module ................................................................................ 64
9.6 Software update ............................................................................................................. 64
9.7 How to proceed in case of repair .................................................................................... 65
10 Dismounting
10.1 Dismounting steps.......................................................................................................... 66
10.2 Disposal ......................................................................................................................... 66
11 Supplement
11.1 Technical data ................................................................................................................ 67
11.2 Communication Probus PA ........................................................................................... 73
11.3 Dimensions .................................................................................................................... 77
Contents
38296-EN-130624
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
Editing status: 2013-05-16
5

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained specialist personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
38296-EN-130624
6

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGAPULS SR 68 is a sensor for continuous level measurement. You can nd detailed information on the application range in chapter
"Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.
Depending on the instrument version, the emitting frequencies are in the C or K band range. The low emitting frequencies are far below the internationally approved limit values. When used correctly, there is no danger to health.
38296-EN-130624
7
2 For your safety

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the
product.
You can nd the CE Certicate of Conformity in the download section of our homepage.
Electromagnetic compatibility
Instruments with plastic housing as well as in four-wire or Ex-d-ia version are designed for use in an industrial environment. Neverthe­less, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a dierent environment, the electromagnetic compatibility to other instruments must be ensured by suitable measures.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom-
mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring
transducers
NE 53 – Compatibility of eld devices and display/adjustment
components
NE 107 – Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Radio license for Europe

The instrument is approved according to EN 302372-1/2 (2006-04) for use in closed vessels.

2.8 Radio license for USA/Canada

The instrument is in conformity with part 15 of the FCC regulations. Take note of the following two regulations:
The instrument must not cause any interfering emissions
The device must be insensitive to interfering immissions, including
those that may cause undesirable operating conditions
Modications not expressly approved by the manufacturer will lead to expiry of the operating licence according to FCC/IC.
The instrument is in conformity with RSS-210 of the IC regulations. The instrument may only be used in closed vessels made of metal,
concrete, or bre-reinforced plastic.
8
38296-EN-130624
2 For your safety

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
38296-EN-130624
9

3 Product description

Type plate
3 Product description
3.1 Conguration
The nameplate contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 4: Layout of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Measuring range 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Reminder to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directive
16
15
14
13
12
11
Serial number
10
With the serial number of the instrument on the type label you can nd the following data on our homepage:
Article number of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate pressure transmitters (PDF)
Go to www.vega.com, "VEGA Tools" and "Serial number search".
As an alternative, you can nd the data via your Smartphone:
Download the smartphone app "VEGA Tools" from the "Apple App
Store " or the "Google Play Store" Scan the Data Matrix code on the type label of the instrument or
Enter the serial number manually in the app
38296-EN-130624
3 Product description
Scope of this operating instructions manual
Scope of deliver
Application area
Functional principle
y
This operating instructions manual applies to the following instrument versions:
Hardware from 2.1.1
Software from 4.5.0
The scope of delivery encompasses:
Radar sensor
Documentation
this operating instructions manualTest certicate Measurement accuracy, depending on the
instrument VEGAPULS SR 68 (optional)
Operating instructions manual "DIsplay and adjustment module
PLICSCOM" (optional)
– Supplementary instructions "GSM/GPRS radio module"
(optional)
Supplementary instructions manual "Heating for display and
adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)if necessary, further certicates

3.2 Principle of operation

The VEGAPULS SR 68 is a radar sensor for continuous measurement of bulk solids also under dicult process conditions. It is particularly suitable for level measurement in high silos and large bunkers.
The instrument is also suitable for applications in liquids.
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The transit time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropri­ate output signal and outputted as measured value.
Packaging
Transport
38296-EN-130624

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
11
3 Product description
Transport inspection
Stor
age
age and transport
Stor temperature
PLICSCOM
VEGACONNECT
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %

3.4 Accessories and replacement parts

The display and adjustment module PLICSCOM is used for measured value indication, adjustment and diagnosis. It can be inserted into the sensor or the external display and adjustment unit and removed at any time.
You can nd further information in the operating instructions "Display and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, the adjustment software PACTware with VEGA-DTM is required.
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 81
PLICSMOBILE T61
Protective cap
12
The VEGADIS 81 is an external display and adjustment unit for VEGA
®
plics
sensors.
For sensors with double chamber housing the interface adapter "DISADAPT" is also required for VEGADIS 81.
You can nd further information in the operating instructions "VE- GADIS 81" (Document-ID 43814).
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for transmission of measured values and for remote parameter adjust­ment of plics DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions "PLICSMOBILE T61" (Document-ID 37700).
The protective cover protects the sensor housing against soiling and intense heat from solar radiation.
®
sensors. The adjustment is carried out via PACTware/
38296-EN-130624
3 Product description
You will nd additional information in the supplementary instructions manual "Protective cover" (Document-ID 34296).
Flanges
Electr
onics module
Antenna impedance cone
Screwed anges are available in dierent versions according to the following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5, JIS B 2210-1984, GOST 12821-80.
You can nd additional information in the supplementary instructions manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
The electronics module VEGAPULS series 60 is a replacement part for radar sensors of VEGAPULS series 60. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions "Elec- tronics module VEGAPULS series 60" (Document-ID 36801).
The antenna impedance cone is used for optimum transmission of microwaves and for sealing against the process.
You nd further information in the operating instructions "Antenna impedance cone VEGAPULS 62 and 68" (Document-ID 31381).
38296-EN-130624
13

4 Mounting

Screwing in
4 Mounting

4.1 General instructions

On instruments with process tting thread, the hexagon must be tight­ened with a suitable screwdriver. Wrench size see chapter "Dimen- sions".
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
Protection against mois­ture
Suitability for the process conditions
Protect your instrument further through the following measures
against moisture penetration:
Use the recommended cable (see chapter "Connecting to power
supply") Tighten the cable gland
Loop the connection cable downward in front of the cable gland
This applies particularly to:
Outdoor mounting
Installations in areas where high humidity is expected (e.g. through
cleaning processes) Installations on cooled or heated vessels
Make sure that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
Active measuring component
Process tting
Process seal
Process conditions are particularly:
Process pressure
Process temperature
Chemical properties of the medium
Abrasion and mechanical inuences
You can nd the specications of the process conditions in chapter "Technical data" as well as on the nameplate.

4.2 Mounting preparations - Horn antenna

The instrument is also available in versions with an antenna whose diameter is larger than the process tting (thread, ange). In such cases the antenna must be disconnected from the process tting before mounting. Proceed as follows:
1. Loosen the hexagon socket screws (3) on the antenna socket with an Allen wrench (size 3)
2. Remove the antenna (4)
Note:
The plastic cone may not be pulled out of the antenna socket.
38296-EN-130624
14
4 Mounting
3. Insert the antenna from below into the vessel socket and secure it against falling o
4. Retighten the antenna with hexagon screws to the antenna socket; torque max. 2.5 Nm (1.8 lbf ft)
Note:
The radar sensor with rinsing air connection or antenna extension is provided with a notch on the antenna socket. This notch must be aligned with the marking on the process tting (the marking species the position of the polarisation of the radar signal).
1
2
3
4
Fig. 5: Dismounting of the horn antenna
1 Marking for the polarisation 2 Marking at the antenna socket 3 Hexagon screws on the antenna socket 4 Antenna
38296-EN-130624

4.3 Mounting preparations - Parabolic antenna

The instrument is also available in versions where the antenna has a diameter larger than the process tting (thread, ange). With such versions the antenna must be disconnected from the process tting before mounting. Proceed as follows:
1. Clamp VEGAPULS SR 68 with the ange, e.g. in a bench vice
2. Hold the connection piece (3) with a wrench on the at surfaces (width across ats 22 mm)
3. Loosen counter nut (2) completely with a wrench (width across ats 36 mm) in the direction of the antenna
4. Loosen compression nut (1) completely with a wrench (width across ats 41 mm) in the direction of the antenna
5. Remove the parabolic antenna (4) axially
6. Mount sensor ange on the adapter ange and fasten it tightly
7. Check if the O-ring seal is present on the connection piece and
make sure it is not damaged.
15
4 Mounting
Note:
A damaged O-ring seal must be replaced: FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375)
8. Remount the parabolic antenna (4)
9. Fasten compression nut (1) with a wrench (width across ats 41), torque max. 50 Nm
10. Fasten counter nut (2) with a wrench (width across ats 36), torque max. 40 Nm
Note:
On the version with rinsing air connection, make sure that the holes in the antenna and in the process tting coincide. This ensures a suf­cient air ow (the air is led through the holes to the feed system. A rinsing of the whole parabolic antenna is not intended).
1
2
3
4
Fig. 6: Dismounting, parabolic antenna
1 Connection piece 2 Compression nut 3 Locknut 4 Parabolic antenna
Polarisation
16

4.4 Mounting instructions

The emitted radar impulses of the radar sensor are electromagnetic waves. The polarisation is the direction of the electrical wave compo­nent. By turning the instrument in the connection ange or mounting boss, the polarisation can be used to reduce the eects of false
echoes.
The position of the polarisation is marked on the process tting of the
instrument.
38296-EN-130624
4 Mounting
Installation position
1
Fig. 7: Position of the polarisation
1 Marking with screwed version
2 Markingwithangeversion
2
Mount the sensor at least 200 mm (7.874 in) away from the vessel wall.
200 mm
")
(7.87
Fig. 8: Mounting of the radar sensor on the vessel top
If you cannot maintain this distance, you should carry out a false signal storage during setup. This applies particularly if buildup on the vessel wall is expected. In such a case, it is recommended to repeat the false signal storage at a later date with existing buildup.
Inowingmedium
38296-EN-130624
Mounting should not be too close to the inowing material as the microwave signal will be interferred. The optimum mounting position is on the opposite of the lling. To avoid strong pollution, the distance to the lter or dust extraction must be as big as possible.
17
4 Mounting
Fig.9:Mountingoftheradarsensorwithinowingmedium
With bulk solids silos with lateral pneumatic lling, mounting should not be in the lling stream as the microwave signal will be interferred. The optimum mounting position is next to the lling. To avoid strong pollution, the distance to the lter or dust extraction must be as big as possible.
18
38296-EN-130624
4 Mounting
Socket
38296-EN-130624
Fig.10:Mountingoftheradarsensorwithinowingmedium
The socket piece should be dimensioned in such a way that the antenna end protrudes slightly out of the socket.
19
4 Mounting
ca. 10 mm
Fig. 11: Recommended socket mounting with horn antenna
20
(0.39")
> 10 mm
Fig. 12: Recommended socket mounting with parabolic antenna
When using a swivelling holder, keep in mind that the distance between antenna and socket gets smaller as the inclination of the sensor increases. Additional false reections may be generated which can inuence the measuring result at close range.
38296-EN-130624
Fig. 13: Distance between antenna and socket with horn antenna
4 Mounting
38296-EN-130624
Fig. 14: Distance between antenna and socket with parabolic antenna
If the medium has good reective properties, VEGAPULS SR 68 with horn antenna can also be mounted on a longer socket piece. Recommended values for socket heights are specied in the following illustration. You must carry out a false echo storage afterwards.
max.
h
d
Fig. 15: Deviating socket dimensions
21
4 Mounting
Orientation
Socket diameter d Socket length h Recommended anten-
40 mm 100 mm 40 mm
50 mm 150 mm 48 mm
80 mm 250 mm 75 mm
100 mm 500 mm 95 mm
150 mm 800 mm 95 mm
Socket diameter d Socket length h Recommended anten-
1½"
2"
3"
4"
6"
3.9 in 1½"
5.9 in 2"
9.8 in 3"
19.7 in 4"
31.5 in 4"
na diameter
na diameter
Tip:
The instrument is also optionally available with an antenna extension. The antenna length can be selected (either ex works or later) to allow the antenna to protrude slightly out of the end of the mounting socket. Due to the antenna extension however, disturbing reections are generated in the close-up range. This can lead to an increase in the required minimum distance, especially with poorly reecting media such as plastic powder. In practice, a cleanly constructed mounting socket, if necessary with rounded edges, introduces fewer disturbing inuences than an antenna extension.
To measure as much of the vessel volume as possible, the sensor should be aligned so that the measuring beam reaches the lowest level in the vessel. In a cylindrical silo with conical outlet mounting is carried out on a socket. It should be positioned on one third up to the half of the vessel radius.
22
38296-EN-130624
4 Mounting
r
1
1
rr...
3
2
/
/
Fig. 16: Orientation
If mounting in the centre of the silo is not possible, the sensor can be directed to the vessel center by using the optional swivelling holder. The following description shows a simple way to determine the required angle of inclination.
38296-EN-130624
23
4 Mounting
α
d
a
Fig. 17: Proposal for installation after orientation VEGAPULS SR 68
The angle of inclination depends on the vessel dimensions. It can be easily checked with a suitable level or water leve on the sensor.
The following chart species the distance "a" between installation position and vessel centre dependent on the measuring distance for inclination angles of 2° … 10°.
Distance d (m)
2 0.1 0.1 0.2 0.3 0.4
4 0.1 0.3 0.4 0.6 0.7
6 0.2 0.4 0.6 0.8 1.1
8 0.3 0.6 0.8 1.1 1.4
10 0.3 0.7 1.1 1.4 1.8
15 0.5 1.0 1.6 2.1 2.6
20 0.7 1.4 2.1 2.8 3.5
25 0.9 1.7 2.6 3.5 4.4
30 1.0 2.1 3.2 4.2 5.3
10°
Example: In a vessel 20 m high, the installation position of the sensor is 1.4 m
from the vessel centre. The necessary angle of inclination of 4° can be read out from this
chart.
38296-EN-130624
24
4 Mounting
Proceed as follows to adjust the angle of inclination with the swivelling holder:
1. Loosen terminal screw of the swivelling holder with a fork spanner
SW 13
2. Direct the sensor, check angle of inclination
Information:
The max. angle of inclination of the swivelling holder is approx. 15°
3. Tighten the terminal screw, torque max. 20 Nm.
Information:
The hexagon screws must not be loosened.
Vessel installations
Agitators
The mounting location of the radar sensor should be a place where no other equipment or xtures cross the path of the microwave signals.
Vessel installations, such as e.g. ladders, limit switches, heating spi­rals, struts, etc., can cause false echoes and impair the useful echo. Make sure when planning your measuring site that the radar sensor has a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be
carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal baes above the installations scatter the radar signals and prevent direct interfering reections.
Fig.18:Coversmoothproleswithdeectors
If there are agitators in the vessel, a false signal suppression should be carried out with the agitators in motion. This ensures that the interfering reections from the agitators are saved with the blades in dierent positions.
38296-EN-130624
25
4 Mounting
Fig. 19: Agitators
Material heaps
Mounting in the vessel insulation
Large material heaps are detected with several sensors, which can be mounted on e.g. traverse cranes. For this type of application, it is best to orient the sensor toward the solid surface. A mutual infuence of the sensors is not possible.
Fig. 20: Radar sensors on traverse crane
Information:
Keep in mind that for these applications, the sensors are designed for relatively slow level changes. When using VEGAPULS SR 68 on a movable bracket, the max. measuring rate must be observed (see chapter "Technical data").
Instruments for a temperature range up to 250 °C or up to 450 °C have a distance piece between process tting and electronics hous­ing. Ths distance piece is used for thermal decoupling of the electron-
ics against high process temperatures.
Information:
The distance piece must only be incorporated up to max. 50 mm in the vessel isolation. Only then, a reliable temperature decoupling is
guaranteed.
38296-EN-130624
26
(1.97")
max. 50 mm
Fig. 21: Mounting of the instrument on insulated vessels.
1 Electronics housing 2 Distance piece 3 Vessel insulation
4 Mounting
1 2
3
Installation in subsurface enclosures
Mounting in multiple chamber silo
38296-EN-130624
For level measurements in concrete silos, the sensors are often mounted in protective boxes. These boxes can be for example metal­lic, closed subsurface enclosures.
1 2
Fig. 22: Mounting of the instrument in an subsurface enclosure
1 Subsurface enclosure 2 Concrete bottom
For this application, the minimal amounts of stray radiation from the sensor can be reected and strengthened by the walls of the subsur­face enclosures. In the case of sensors with plastic housings, this can lead to coupling disturbances. This can be avoided by using a sensor with aluminium or stainless steel housing.
The silo walls of multiple chamber silos are often made of prole walls, such as e.g. prole sheeting, to ensure the required stability. If the radar sensor is mounted very close to a heavily structured vessel wall, considerable false reections can be generated. Hence the sen-
27
4 Mounting
sor should be mounted at a large distance from the separating wall. The optimal mounting position is on the outer wall of the silo with the sensor directed towards the emptying aperture in the silo center.
Fig. 23: Mounting of VEGAPULS SR 68 in multiple chamber silos
Dust layers
28
Fig. 24: Orientation of VEGAPULS SR 68 for emptying in the silo center
To avoid strong buildup and dust in the antenna system, the sensor should not be mounted directly at the dust extraction of the vessel.
38296-EN-130624
4 Mounting
In case of extreme dust deposits in the antenna system, VEGAPULS SR 68 is available with a rinsing air connection. The air is distributed via channels in the antenna system and keeps it largely free of dust.
Fig. 25: Purging air connection with horn antenna
38296-EN-130624
Fig. 26: Purging air connection with parabolic antenna
The practice has shown that a pressure of approx. 0.2 … 1 bar provides a sucient air ow (see diagram in chapter "Technical data", "Purging air connection".
29

5 Connecting to power supply

Safety instructions
5 Connecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
V
oltage supply
Connection cable
Cable gland ½ NPT
Cable screening and grounding
The voltage supply is provided by a Probus DP /PA segment coupler. The voltage supply range can dier depending on the instrument
version. You can nd the data for voltage supply in chapter "Technical data".
Connection is made with screened cable according to the Probus specication. Power supply and digital bus signal are carried over the same two-wire connection cable.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland. If you are using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Please make sure that your installation is carried out according to the Probus specication. In particular, make sure that the termination of the bus is done with appropriate terminating resistors.
You can nd detailed information of the cable specication, installa­tion and topology in the " line" on www.probus.com.
With plastic housing, the NPT cable gland or the Conduit steel tube must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data".
Make sure that the cable screening and ground is executed accord­ing to the Fielbus specication. If electromagnetic interference is expected which is above the test values of EN 61326-1 for industrial areas, we recommend to connect the cable screen on both ends to ground potential.
With systems with potential equalisation, connect the cable screen directly to ground potential at the power supply unit, in the connection box and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation (low impedance).
In systems without potential equalisation with cable screening on both sides, connect the cable screen directly to ground potential at the power supply unit and at the sensor. In the connection box or T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a
ProbusPA-UserandInstallationGuide-
38296-EN-130624
30
Connection technology
5 Connecting to power supply
ceramic capacitor (e.g. 1 nF, 1500 V). Low-frequency potential equali­sation currents are thus suppressed, but the protective eect against high frequency interference signals remains.

5.2 Connecting

The voltage supply and signal output are connected via the spring­loaded terminals in the housing.
The connection to the display and adjustment module or to the inter­face adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the electronics. To do this, lift the terminal block with a small screwdriver and pull it out. When reinserting the terminal block, you should hear it
snap in.
Connection procedure
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn­ing it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Fig. 27: Connection steps 5 and 6 - Single chamber housing
38296-EN-130624
31
5 Connecting to power supply
Fig. 28: Connection steps 5 and 6 - Double chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are insert­ed directly into the terminal openings. In case of exible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the terminal closes again.
You can nd further information on the max. wire cross-section under "Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
The electrical connection is hence nished.
32

5.3 Wiring plan, single chamber housing

The following illustration applies to the non-Ex as well as to the Ex-ia version.
38296-EN-130624
2
2
Electronics and connec­tion compartment
Electronics compartment
5 Connecting to power supply
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
+
( )
1
3
7
7
4
4
6
6
5
5
(-)
2
5
678
4
5
1
Fig. 29: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Selection switch for bus address 4 For external display and adjustment unit 5 Ground terminal for connection of the cable screen

5.4 Wiring plan, double chamber housing

The following illustrations apply to the non-Ex as well as to the Ex-ia version.
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
3
7
7
4
4
6
6
5
5
38296-EN-130624
( )
+
(-)
1
2
5
678
11
Fig. 30: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the display and adjustment module or interface adapter 3 Selection switch for bus address
Information:
The connection of an external display and adjustment unit is not pos­sible with this double chamber housing.
33
5 Connecting to power supply
2
2
Connection compartment
Connection compartment
- Radio module PLICS­MOBILE
Bus
(-)
( )
+
1
2
3
1
Fig. 31: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Ground terminal for connection of the cable screen
Status
SIM-Card
Test
Bus
( )
+
(-)
1
2
USB
1
Fig. 32: Connection compartment radio module PLICSMOBILE
1 Voltage supply
You can nd detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".
Electronics compartment
34

5.5 Wiring plan, double chamber housing Ex d

4...20mA
+
( )
(-)
1
2
Fig. 33: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 For display and adjustment module or interface adapter 3 Internal connection to the plug connector for external display and adjust-
ment unit (optional)
8
567
31
38296-EN-130624
2
Connection compartment
DIS-ADAPT with plug connector M12 x 1 for external display and adjustment unit
5 Connecting to power supply
Note:
If an instrument with Ex-d-ia approval is used, HART multidrop op­eration is not possible.
Bus
(-)
( )
+
1
2
3
1
Fig. 34: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Ground terminal for connection of the cable screen
1
2
3
38296-EN-130624
Fig. 35: View to the electronics compartment
1 DIS-ADAPT 2 Internal plug connection 3 Plug connector M12 x 1
1
Fig. 36: Top view of the plug connector
1 Pin 1 2 Pin 2 3 Pin 3 4 Pin 4
34
2
35
5 Connecting to power supply
Wire assignment, con­nection cable
Contact pin Colour connection ca-
ble in the sensor
Pin 1 Brown 5
Pin 2 White 6
Pin 3 Blue 7
Pin 4 Black 8
Terminal, electronics module

5.6 Wiring plan - version IP 66/IP 68, 1 bar

1
2
Fig.37:Wireassignmentx-connectedconnectioncable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding

5.7 Switch-on phase

After VEGAPULS SR 68 is connected to the bus system, the instru­ment carries out a self-test for approx. 30 seconds. The following
steps are carried out:
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC Indication of the status message "F 105 Determine measured
value" on the display or PC Status byte goes briey to fault value
As soon as a plausible measured value is found, it is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory settings.
36
38296-EN-130624

6 Set up with the display and adjustment module

6 Set up with the display and adjustment
module

6.1 Insert display and adjustment module

The display and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be se­lected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the requested posi­tion onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order. The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
38296-EN-130624
Fig. 38: Insertion of the display and adjustment module with single chamber housing into the electronics compartment
37
6 Set up with the display and adjustment module
Fig. 39: Insertion of the display and adjustment module into the double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
1 2
Key functions
38

6.2 Adjustment system

Fig. 40: Display and adjustment elements
1 LC display 2 Adjustment keys
[OK] key:
Move to the menu overview
1
2
38296-EN-130624
6 Set up with the display and adjustment module
Conrm selected menu – Edit parameter – Save value
[->] key:
Presentation, change measured valueSelect list entrySelect editing position
[+] key:
Change value of the parameter
[ESC] key:
– Interrupt input – Jump to next higher menu
Adjustment system
Main menu
The sensor is adjusted via the four keys of the display and adjust­ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.

6.3 Parameter adjustment

Through the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g., for measurement loop name, medium, applica­tion, vessel, adjustment, signal output
Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g. on instrument status, pointer, measure-
ment reliability, simulation, echo curve Further settings: Instrument unit, false signal suppression, linearisa-
tion curve, reset, date/time, reset, copy function Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Information:
In this operating instructions manual, the instrument-specic param­eters in the menu sections "Setup", "Diagnosis" and "Additional set- tings" are described. The general parameters in these menu section are described in the operating instructions manual "Indicating and adjustment module".
38296-EN-130624
39
6 Set up with the display and adjustment module
You can nd in the operating instructions manual "Display and adjust- ment module" also the description of the menu sections "Display" and "Info".
In the main menu point "Setup", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the measurement. The proce­dure is described in the following.
Setup
Setup - Medium
Each medium has dierent reection properties. With liquids, further interfering factors are uctuation product surface and foam genera­tion. With bulk solids, these are dust generation, material cone and additional echoes from the vessel wall.
To adapt the sensor to these dierent measuring conditions, the selection "Liquid" or "Bulk solid" should be made in this menu item.
Through this selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with poor reec­tive properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup - Application
40
In addition to the medium, also the application, i.e. the measuring site, can inuence the measurement.
With this menu item, the sensor can be adapted to the applications. The adjustment possibilities depend on the selection "Liquid" or "
Bulk
solid" under "Medium".
The following options are available when "Liquid" is selected:
38296-EN-130624
6 Set up with the display and adjustment module
The selection "Standpipe" opens a new window in which the inner diameter of the applied standpipe is entered.
The following features form the basis of the applications:
Storage tank:
Setup: large-volumed, upright cylindrical, spherical
Product speed: slow lling and emptying
Process/measurement conditions:
– Condensation – Smooth product surfaceMax. requirement to the measurement accuracy
Storage tanke with product circulation:
Setup: large-volumed, upright cylindrical, spherical
Product speed: slow lling and emptying
Vessel: small laterally mounted or large top mounted stirrer
Process/measurement conditions:
Relatively smooth product surfaceMax. requirement to the measurement accuracy – Condensation – Slight foam generationOverlling possible
38296-EN-130624
Storage tank on ships (Cargo Tank):
Product speed: slow lling and emptying
Vessel:
Installations in the bottom section (bracers, heating spirals)High sockets 200 … 500 mm, also with large diameters
Process/measurement conditions:
Condensation, buildup by movementMax. requirement on measurement accuracy from 95 %
Stirrer vessel (reactor):
Setup: all vessel sizes possible
Product speed:
Fast to slow lling possibleVessel is very often lled and emptied
Vessel:
Socket availableLarge agitator blades of metalVortex breakers, heating spirals
Process/measurement conditions:
Condensation, buildup by movement
41
6 Set up with the display and adjustment module
– Strong spout generation – Very agitated surface, foam generation
Dosing vessel:
Setup: all vessel sizes possible
Product speed:
Fast lling and emptyingVessel is very often lled and emptied
Vessel: narrow installation situation
Process/measurement conditions:
Condensation, buildup on the antenna – Foam generation
Standpipe:
Product speed: very fast lling and emptying
Vessel:
Vent holeJoins like anges, weld jointsShifting of the running time in the tube
Process/measurement conditions:
– Condensation – Buildup
Bypass:
Product speed:
Fast up to slow lling with short up to long bypass tube possibleOften the level is hold via a control facility
Vessel:
Lateral outlets and inletsJoins like anges, weld jointsShifting of the running time in the tube
Process/measurement conditions:
– Condensation – BuildupSeparation of oil and water possibleOverlling into the antenna possible
42
Plastic tank:
Vessel:
Measurement x mounted or integratedMeasurement depending on the application through the vessel
top
With empty vessel, the measurement can be carried out
through the bottom
Process/measurement conditions:
Condensation on the plastic ceilingIn outside facilities water and snow on the vessel top possible
Transportable plastic tank:
Vessel:
Material and thickness dierentMeasurement through the vessel top
Process/measurement conditions:
38296-EN-130624
6 Set up with the display and adjustment module
– Measured value jump with vessel change
Open water (gauge measurement):
Gauge rate of change: slow gauge change
Process/measurement conditions:
Distance sensor to water surface to bigExtreme damping of output signal due to wave generationIce and condensation on the antenna possibleSpiders and insect nestle in the antennasFloating material and animals sporadically on the water surface
Openume(owmeasurement):
Gauge rate of change: slow gauge change
Process/measurement conditions:
Ice and condensation on the antenna possibleSpiders and insect nestle in the antennasSmooth water surfaceExact measurement result requiredDistance to the water surface normally relatively high
Rain water overfall (weir):
Gauge rate of change: slow gauge change
Process/measurement conditions:
Ice and condensation on the antenna possibleSpiders and insect nestle in the antennasTurbulent water surfaceSensor ooding possible
38296-EN-130624
Demonstration:
Adjustment for all applications which are not typically level meas-
urement
Sensor accepts all measured value changes within the measuring
range immediately Typical applications:
– Instrument demonstration – Object recognition/monitoring (additional settings required)
Caution:
If a separation of liquids with dierent dielectric gure occurs in the vessel, for example through condensation, the radar sensor can detect under certain circumstances only the medium with the higher dielectric gure. Keep in mind that layer interfaces can cause faulty
measurements.
If you want to measure the total height of both liquids reliably, please contact our service department or use an instrument specially de­signed for interface measurement.
The following options are available when "Bulk solid" is selected:
43
6 Set up with the display and adjustment module
The following features form the basis of the applications:
Silo (slim and high):
Vessel of metal: weld joints
Process/measurement conditions:
Filling too close to the sensorSystem noise with completely empty silo increasedAutomatic false signal suppression with partly lled vessel
Bunker (large-volumed):
Vessel of concrete or metal:
Structured vessel wallsInstallations present
Process/measurement conditions:
– Large distance to the medium – Large angles of repose
Bunkerwithfastlling:
Vessel of concrete or metal, also multiple chamber silo:
Structured vessel wallsInstallations present
Process/measurement conditions:
Measured value jumps, e.g. by truck loading – Large distance to the medium – Large angles of repose
Heap:
Sensor mounting on movable conveyor belts
Detection of the heap prole
Height detection during lling
Process/measurement conditions:
Measured value jumps, e.g. by the prole of the heap or trav-
erses
Large angles of reposeMeasurement near the lling stream
44
Crusher:
Vessel: installations, wear and protective facilities available
Process/measurement conditions:
Measured value jumps, e.g. by truck loading – Fast reaction time – Large distance to the medium
Demonstration:
Adjustment for all applications which are not typically level meas-
urement
Sensor accepts all measured value changes within the measuring
range immediately Typical applications:
– Instrument demonstration – Object recognition/monitoring (additional settings required)
38296-EN-130624
6 Set up with the display and adjustment module
Through this selection, the sensor is adapted optimally to the applica­tion or the location and measurement reliability under the various basic conditions is increased considerably.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup - Vessel height, measuring range
Setup -
Vessel form
With this selection, the operating range of the sensor is adapted to the vessel height and the reliability with dierent frame conditions is increased considerably.
Independent from this, the min. adjustment must be carried out.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Also the vessel form can inuence the measurement apart from the medium and the application. To adapt the sensor to these measure­ment conditions, this menu item oers you dierent options for vessel bottom and ceiling in case of certain applications.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup - Adjustment
38296-EN-130624
Since the radar sensor is a distance measuring instrument, the distance from the sensor to the product surface is measured. For indication of the real level, an allocation of the measured distance to the percentage height must be carried out.
To perform the adjustment, enter the distance with full and empty ves­sel, see the following example:
45
6 Set up with the display and adjustment module
Fig. 41: Parameter adjustment example min./max. adjustment
1 Min. level = max. meas. distance 2 Max. level = min. meas. distance
If these values are not known, an adjustment with the distances of for example 10 % and 90 % is possible. Starting point for these distance specications is always the seal surface of the thread or ange. By means of these settings, the real level will be calculated.
The real product level during this adjustment is not important, be­cause the min./max. adjustment is always carried out without chang­ing the product level. These settings can be made ahead of time without the instrument having to be installed.
1 m
")
(39.37
35 m
100%
")
(1378
0%
2
1
Setup - Min. adjustment
46
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
38296-EN-130624
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
6 Set up with the display and adjustment module
4. Enter the suitable distance value in m for the empty vessel (e.g. distance from the sensor to the vessel bottom) corresponding to the percentage value.
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
Setup - Max. adjustment
Diagnosis/Peak value
Proceed as follows:
1. Select with [->] the menu item max. adjustment and conrm with
[OK].
2. Prepare the percentage value for editing with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4. Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. Keep in mind that the max. level must lie below the min. distance to the antenna edge.
5. Save settings with [OK]
The respective min. and max. measured value is saved in the sensor. The values are displayed in the menu item "Peak values".
Diagnosis - Measurement reliability
38296-EN-130624
When non-contact level sensors are used, the measurement can be inuenced by the respective process conditions. In this menu item, the measurement reliability of the level echo is displayed as dB value. The measurement reliability equals signal strength minus noise. The higher the value, the more reliable the measurement. With a function­ing measurement, the values are > 10 dB.
47
6 Set up with the display and adjustment module
Diagnoses - Curve indica­tion
The "Echo curve" shows the signal strength of the echoes over the measuring range in dB. The signal strength enables an evaluation of the quality of the measurement.
The "False signal suppression" displays the saved false echoes (see menu "Additional settings") of the empty vessel with signal strength in "dB" over the measuring range.
A comparison of echo curve and false signal suppression allows a more accurate conclusion on measurement reliability. The selected curve is continuously updated. With the [OK] key, a submenu with zoom functions is opened:
"X-Zoom": Zoom function for the meas. distance
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "dB"
"Unzoom": Reset the presentation to the nominal measuring range
with single magnication
Diagnostics - Echo cur memory
48
With the function "Echo curve memory" the echo curve can be saved
v
e
at the time of setup. This is generally recommended; for using the As­set Management functions it is absolutely necessary. If possible, the curve should be saved with a low level in the vessel.
With the adjustment software PACTware and the PC, the high resolu­tion echo curve can be displayed and used to recognize signal changes over the operating time. In addition, the echo curve of the setup can be also displayed in the echo curve window and compared with the actual echo curve.
38296-EN-130624
6 Set up with the display and adjustment module
Additional adjustments ­False signal suppr
ession
The following circumstances cause interfering reections and can inuence the measurement:
High sockets
Vessel installations such as struts
Agitators
Buildup or welded joints on vessel walls
Note:
A false signal suppression detects, marks and saves these false sig­nals so that they are no longer taken into account for level measure-
ment.
This should be done with the low level so that all potential interfering reections can be detected.
Proceed as follows:
1. Select the menu item "Additional settins" with [->] and conrm with [OK]. With [->] you have to select the menu item "False signal suppression" and conrm with [OK].
2. Conrm again with [OK].
3. Conrm again with [OK].
38296-EN-130624
4. Conrm again with [OK] and enter the actual distance from the sensor to the product surface.
5. All interfering signals in this section are detected by the sensor and stored after conrming with [OK].
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as a false echo. The lling level would then no longer be detectable in this area.
49
6 Set up with the display and adjustment module
If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression":
The menu item "Delete" is used to completely delete an already cre­ated false signal suppression. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel.
The menu item "Extend" is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with a too high level and not all false signals could be de­tected. When selecting "Extend", the distance to the product surface of the created false signal suppression is displayed. This value can now be changed and the false signal suppression can be extended to
this range.
Additional adjustments ­Linearization cur
ve
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindri­cal or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [ESC] and [->] key.
Caution:
Note the following if the instrument with corresponding approval is used as part of an overll protection system according to WHG:
If a linearization curve is selected, the measuring signal is no longer necessarily linear to the lling height. This must be considered by the user especially when adjusting the switching point on the limit signal
transmitter.
38296-EN-130624
50
6 Set up with the display and adjustment module
Additional adjustments
- Reset
When a reset is carried out, all settings (with only a few exceptions) are reset. The exceptions are: PIN, language, lighting, SIL and HART
mode.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of
shipment from the factory incl. the order-specic settings. A cre­ated false signal suppression, user-programmable linearization curve, the measured value memory, echo curve memory as well as event memory will be deleted. Basic settings: Resetting of the parameter settings incl. special
parameters to the default values of the respective instrument. Any created false signal suppression, user programmable linearization curve, measured value memory as well as event memory will be deleted. Setup: Resetting of the parameter settings to the default values of
the respective instrument . Order-related settings remain but are not taken over into the current parameters. User-generated false signal suppression, user-programmed linearization curve, meas­ured value memory, echo curve memory as well as event memory remain untouched. The linearization is set to linear. False signal suppression: Deleting a previously created false
signal suppression. The false signal suppression created in the factory remains active. Peak values measured value: Resetting of the measured min.
and max. distances to the actual measured value.
Select the requested reset function [->] and conrm with [OK]. The following table shows the default values of VEGAPULS SR 68:
38296-EN-130624
51
6 Set up with the display and adjustment module
Menu Menu item Default value
Setup Measurement
Display Language Like order
Additional adjust-
ments
loop name
Medium Liquid/Water
Application Storage tank
Vessel form Vessel bottom, dished boiler end
Vesell height/
Measuring range
Min. adjustment Recommended measuring range, see
Max. adjustment
AI FB1 Channnel PV (lin. perc.)
AI FB1 scaling
unit
AI FB1 scaling
AI FB1 damping 0 s
Block adjustment Released
Displayed value
Backlight Switched o
Distance unit m
Temperature unit °C
Unit SV2 m
Probe length Length of the standpipe Ex factory
Linearization curve
Sensor address 126
Sensor
Bulk solids/Crushed stones, gravel
Silo
Vessel top, dished boiler end
Recommended measuring range, see "Technical data" in the supplement
"Technical data" in the supplement
0,000 m(d)
Height
%
0.00 lin %, 0.00 %
100.00 lin %, 100.00 %
SV 1
Linear
52

6.4 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module, the data in the sensor can be saved in the display and adjustment module. The procedure is described in the operating instructions manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
38296-EN-130624
6 Set up with the display and adjustment module
The following data or settings for adjustment of the display and ad­justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Sensor-specic
units, temperature unit and linearization" The values of the user programmable linearization curve
The function can also be used to transfer settings from one instru­ment to another instrument of the same type. If it is necessary to exchange a sensor, the display and adjustment module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy sensor data".
38296-EN-130624
53

7 Setup with PACTware

2
Via the interface adapter directly on the sensor
erequisites
Pr
7 Setup with PACTware

7.1 Connect the PC

1
3
Fig. 42: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 3 Sensor

7.2 Parameter adjustment

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications
according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
38296-EN-130624
54
Fig. 43: Example of a DTM view
7 Setup with PACTware
Standard/Full version
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega.com/downloads and "Software". The full version is avail­able on CD from the agency serving you.

7.3 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
38296-EN-130624
55

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™
and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".
56
38296-EN-130624

9 Diagnosis, Asset Management and service

9 Diagnosis, Asset Management and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal
operation.
In some applications, buildup on the antenna system can inuence the measuring result. Depending on the sensor and application, make arrangements to avoid strong pollution of the antenna system. If nec­essary, clean the antenna system in certain intervals.

9.2 Measured value and event memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
E
ent memory
v
Echo curve memory
Up to 60,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switch-on and switch-o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control system with EDD.
The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for
the measurement conditions during setup. Changes in the measure-
ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via:
38296-EN-130624
57
9 Diagnosis, Asset Management and service
PC with PACTware/DTM
Control system with EDD
Display and adjustment module
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD

9.3 Asset Management function

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the display and adjustment module, PACTware/DTM and EDD.
Status messages
The status messages are divided into the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 44: Pictographs of the status messages
1 Failure - red
2 Outofspecication-yellow
3 Function check - orange 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is
outputted.
This status message is always active. It cannot be deactivated by the
user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
38296-EN-130624
58
9 Diagnosis, Asset Management and service
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Failure
The following table shows the error codes in the status message "Failure" and gives information on the reason and rectication. Keep in mind that some information is only valid with four-wire instruments.
Code
Text mes­sage
F013
no measured
value avail­able
F017
Adjustment span too small
F025
Error in the
linearization table
F036
No operable software
F040
Error in the
electronics
F080
F105
Determine measured
value
F113
Communica­tion error
F125
Unpermissi-
ble electronics
temperature
Cause Rectication
– Sensor does not detect an
echo during operation
Antenna system contami-
nated or defective
– Adjustment not within
specication
– Index markers are not con-
tinuously rising, for examle unlogical value pairs
– Failed or interrupted soft-
ware update
Hardware defect Exchanging the electronics
General software error Separate operating voltage
The instrument is still in the
start phase, the measured value could not yet be
determined
EMC interferencesTransmission error with the
external communication with 4-wire power supply
unit
Temperature of the elec-
tronics in the non-specied
section
Check or correct installation
and/or parameter adjust­ment
Clean or exchange process
component or antenna
Change adjustment accord-
ing to the limit values (dif­ference between min. and max. ≥ 10 mm)
Check linearization tableDelete table/Create new
Repeat software updateCheck electronics versionExchanging the electronicsSend instrument for repair
Send instrument for repair
briey
– Wait for the end of the
switch-on phase
– Duration depending on
the version and parameter adjustment up to approxi­mately 3 min.
Remove EMC inuencesExchange 4-wire power
supply unit or electronics
Check ambient temperatureIsolate electronicsUse instrument with higher
temperature range
38296-EN-130624
59
9 Diagnosis, Asset Management and service
Function check
Code
Text mes­sage
F260
Error in the
calibration
F261
Error in the
conguration
F264
Installation/
Setup error
F265
Measurement
function dis­turbed
Cause Rectication
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Error during setup – False signal suppression
faulty
Error when carrying out a
reset
Adjustment not within the
vessel height/measuring
range
Max. measuring range of
the instrument not sucient
Sensor no longer carries
out a measurement
Operating voltage too low
Exchanging the electronicsSend instrument for repair
– Repeat setup – Repeat reset
Check or correct installation
and/or parameter adjust­ment
Use an instrument with big-
ger measuring range
Check operating voltageCarry out a resetSeparate operating voltage
briey
The following table shows the error codes and text messages in the status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700
Simulation ac­tive
Cause Rectication
A simulation is active Finish simulation
Wait for the automatic end
after 60 mins.
Outofspecication
Maintenance
60
The following table shows the error codes and text messages in the status message "Outofspecication" and provides information on causes as well as corrective measures.
Code
Text mes­sage
S600
Unpermissi-
ble electronics
temperature
S601
Overlling
Cause Rectication
– Temperature of the elec-
tronics in the non-specied
section
Danger of vessel overlling – Make sure that there is no
Check ambient temperatureIsolate electronicsUse instrument with higher
temperature range
further lling
– Check level in the vessel
The following table shows the error codes and text messages in the status message "Maintenance" and provides information on causes as well as corrective measures.
38296-EN-130624
9 Diagnosis, Asset Management and service
Code
Text mes­sage
M500
Error with the reset delivery
status
M501
Error in the
non-active linearization table
M502
Error in the diagnosis
memory
M503
Reliability too low
M504
Error on an
device inter­face
M505
No echo avail­able
Cause Rectication
– With the reset to delivery
status, the data could not be restored
Hardware error EEPROM Exchanging the electronics
Hardware error EEPROM Exchanging the electronics
The echot/noise ratio is the
small for a reliable meas­urement
Hardware defect – Check connections
Level echo can no longer
be detected
– Repeat reset – Load XML le with sensor
data into the sensor
Send instrument for repair
Send instrument for repair
Check installation and
process conditions
Clean the antennaChange polarisation direc-
tion
Use instrument with higher
sensitivity
Exchanging the electronicsSend instrument for repair
Clean the antennaUse a more suitable
antenna/sensor
Remove possible false
echoes
Optimize sensor position
and orientation
Reaction when malfunc­tions occur
Procedur
e for fault recti-
cation
Treatment of measure­ment errors with bulk solids
38296-EN-130624

9.4 Rectify faults

The operator of the system is responsible for taking suitable meas­ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the display and
adjustment module Checking the output signal
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
The below tables show typical examples for application-relevant measurement errors with bulk solids. There are two measurement
errors:
61
9 Diagnosis, Asset Management and service
time
0
time
0
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
Level
1
2
0
1 Real level 2 Level displayed by the sensor
Notes:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
2. Measured value jumps towards 100 %
Level
Level
Min./max. adjustment not
correct
Wrong linearization curve Adapt linearization curve
Due to the process, the ampli-
tude of the product echo sinks
A false signal suppression was
not carried out
Amplitude or position of a
false echo has changed (e.g. condensation, buildup); false signal suppression no longer
matches
Adapt min./max. adjustment
Carry out a false signal sup-
pression
Determine the reason for the
changed false echoes, carry out false signal suppression, e.g. with condensation
time
38296-EN-130624
62
9 Diagnosis, Asset Management and service
time
0
time
0
time
0
time
0
Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value jumps towards 0 % during lling
4. Measured val­ue uctuates around 10 … 20 %
5. Measured value jumps sporadically to 100 % during lling
Level
– Amplitude of a multiple echo
(vessel top - product surface) is larger than the level echo
– The level echo cannot be distin-
guished from the false echo at a false echo position (jumps to multiple echo)
– Transverse reection from an
extraction funnel, amplitude of the transverse reection larger than the level echo
Level
– Various echoes from an uneven
product surface, e.g. a material
cone
Reections from the product
surface via the vessel wall (deection)
Level
– Changing condensation or con-
tamination on the antenna
Check parameter "Application",
especially vessel top, product type, dished end, high dielectric gure, adapt if necessary
– Remove/reduce false echo:
minimize interfering installa­tions by changing the polariza-
tion direction
Chose a more suitable installa-
tion position
– Direct sensor to the opposite
funnel wall, avoid crossing with the lling stream
– Check parameter "Medium
type" and adapt, if necessary
Optimize installation position
and sensor orientation
Select a more suitable installa-
tion position, optimize sensor orientation, e.g. with a swivel­ling holder
– Carry out a false signal sup-
pression or increase false signal suppression with con­densation/contamination in the close range by editing
– With bulk solids use radar sen-
sor with purging air connection or exible antenna cover
38296-EN-130624
Measurement error during emptying
Fault description Error pattern Cause Rectication
6. Measured value re­mains unchanged in
the close range during emptying
Level
False echo larger than the level
echo
Level echo too small
Remove false echoes in the
close range. Check: Antenna must protrude out of the socket
Remove contamination on the
antenna
Minimize interfering instal-
lations in the close range by changing the polarization
direction
After removing the false ech-
oes, the false signal suppres­sion must be deleted. Carry out a new false signal suppression
63
9 Diagnosis, Asset Management and service
time
0
time
Fault description Error pattern Cause Rectication
7. Measured value jumps sporadically to­wards 100 % during emptying
8. Measured val­ue uctuates around 10 … 20 %
Level
Level
0
– Changing condensation or con-
tamination on the antenna
Various echoes from an uneven
product surface, e.g. an extrac­tion funnel
– Reections from the product
surface via the vessel wall (deection)
– Carry out false signal suppres-
sion or increase false signal suppression in the close range by editing
– With bulk solids use radar sen-
sor with purging air connection or exible antenna cover
– Check parameter "Medium
type" and adapt, if necessary
Optimize installation position
and sensor orientation
Reaction after fault recti-
cation
24 hour service hotline
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no.
+49 1805 858550.
The hotline is also available outside normal working hours, seven days a week around the clock.
Since we oer this service worldwide, the support is provided in English. The service itself is free of charge, the only costs involved are the normal call charges.

9.5 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user. In Ex applications only one instrument and one electronics module
with respective Ex approval may be used. If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron­ics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").
38296-EN-130624

9.6 Software update

The following components are required to update the sensor soft­ware:
64
9 Diagnosis, Asset Management and service
Sensor
Voltage supply
Interface adapter VEGACONNECT
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
You can nd information about the installation in the download le.
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".

9.7 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Formsandcerti- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
38296-EN-130624
65

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, cor­rosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
66
38296-EN-130624

11 Supplement

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
Ʋ Process tting 316L Ʋ Process seal On site (instruments with thread: Klingersil C-4400 is
Ʋ Antenna 316L, 316L electropolished, 316L Safecoat-coated Ʋ Antenna impedance cone PTFE TFM 1600 oder PTFE INOFLON M290, PEEK Ʋ seal, antenna system FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375), FFKM
Materials, non-wetted parts
Ʋ Plastic housing plastic PBT (Polyester) Ʋ Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated - basis:
Ʋ Stainless steel housing 316L Ʋ Seal between housing and housing
cover
Ʋ Inspection window in housing cover
(optional)
Ʋ Ground terminal 316L
Ohmic contact Between ground terminal, process tting and antenna
Process ttings
Ʋ Pipe thread, cylindrical (ISO 228 T1) G1½ A according to DIN 3852-A Ʋ American pipe thread, tapered 1½ NPT, 2 NPT Ʋ Flanges DIN from DN 25, ANSI from 1"
Weights
Ʋ Instrument (depending on housing,
process tting and antenna)
Ʋ Antenna extension 1.6 kg/m (1.157 lbs/ft) Length antenna extension max. 5.85 m (19.19 ft) Torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing max. 10 Nm (7.376 lbf ft)
Ʋ Aluminium/Stainless steel housing max. 50 Nm (36.88 lbf ft)
attached)
(Kalrez 2035), FFKM (Kalrez 6230 - FDA)
Polyester
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housng, electropolished)
Polycarbonate
approx. 2 … 17.2 kg (4.409 … 37.92 lbs)
Input variable
Measured variable The measured quantity is the distance between process
tting of the sensor and product surface. The reference plane is the seal surface on the hexagon or the lower side of the ange.
38296-EN-130624
67
11 Supplement
3
4
Fig. 54: Data of the input variable
1 Reference plane 2 Measured variable, max. measuring range 3 Antenna length 4 Useful measuring range
Max. measuring range 30 m (98.43 ft)
Recommended meas. range depending on the antenna diameter
Ʋ ø 40 mm (1.575 in) up to 15 m (49.21 ft)
Ʋ ø 48 mm (1.89 in) up to 20 m (65.62 ft)
Ʋ ø 75 mm (2.953 in), 95 mm (3.74 in) up to 30 m (98.43 ft)
Ʋ Parabolic antenna up to 30 m (98.43 ft)
1
2
Output variable
Output signal digital output signal, format according to IEEE-754 Sensor address 126 (default setting) Damping (63 % of the input variable) 0 … 999 s, adjustable Probus PA prole 3.02 Number of FBs with AI (function blocks
with analogue input) Default values
Ʋ 1. FB PV
Ʋ 2. FB SV 1
Ʋ 3. FB SV 2
Current value
68
3
38296-EN-130624
11 Supplement
Ʋ Non-Ex and Ex ia instrument 10 mA, ±0.5 mA
Ʋ Ex-d instruments 16 mA, ±0.5 mA
Resolution, digital < 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F)
Ʋ Relative humidity 45 … 75 %
Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Installation reference conditions
Ʋ Min. distance to installations > 200 mm (7.874 in)
Ʋ Reector Plane plate reector
Ʋ False reections Biggest false signal, 20 dB smaller than the useful signal
Deviation with liquids See following diagrams
10 mm (0.394 in)
2 mm (0.079 in)
- 2 mm (- 0.079 in)
- 10 mm (- 0.394 in)
0
1,0 m (3.28 ft)
1 2 3
Fig. 55: Deviation under reference conditions
1 Reference plane 2 Antenna edge 3 Recommended measuring range
Repeatability ≤ ±1 mm Deviation with bulk solids The values depend to a great extent on the application.
Deviation under EMC inuence
Variablesinuencingmeasurementaccuracy
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
Additional deviation through electromag­netic interference acc. to EN 61326
Characteristics and performance data
Measuring frequency
38296-EN-130624
Binding specications are thus not possible. ≤ ±30 mm
max. 10 mm < ±50 mm
K-band (26 GHz technology)
69
11 Supplement
Measuring cycle time approx. 700 ms
Step response time
Beam angle
1)
2)
≤ 3 s
Ʋ Horn antenna ø 40 mm (1.575 in) 20°
Ʋ Horn antenna ø 48 mm (1.89 in) 15°
Ʋ Horn antenna ø 75 mm (2.953 in) 10°
Ʋ Horn antenna ø 95 mm (3.74 in)
Ʋ Parabolic antenna
Emitted HF power (depending on the parameter adjustment)
Ʋ Average spectral transmission power
-14 dBm/MHz EIRP
3)
density
Ʋ Max. spectral transmission power
+43 dBm/50 MHz EIRP
density
Ʋ Max. power density in a distance of
< 1 µW/cm²
1 m
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
For the process conditions, please also note the specications on the type label. The lowest value always applies.
Seal Antenna impedance
FKM (SHS FPM 70C3 GLT)
FFKM (Kalrez 6375) PFFE -20 … +130 °C (-4 … +266 °F)
FFKM (Kalrez 2035) PFFE -15 … +130 °C (5 … +266 °F)
FFKM (Kalrez 6230) PFFE -15 … +130 °C (5 … +266 °F)
cone
PTFE
PEEK -20 … +250 °C (-4 … +482 °F)
PEEK -15 … +210 °C (5 … +410 °F)
PEEK -15 … +250 °C (5 … +482 °F)
Process temperature (measured on the process
tting)
-40 … +130 °C (-40 … +266 °F)
Vessel pressure - horn antenna
Ʋ Antenna impedance cone PTFE -1 … 40 bar/-100 … 4000 kPa (-14.5 … 580 psi)
Ʋ Antenna impedance cone PEEK -1 … 100 bar/-100 … 10000 kPa (-14.5 … 1450 psi)
Vessel pressure with swivelling holder -1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig) Vessel pressure relating to the ange
nominal stage
see supplementary instructions manual "Flanges ac- cording to DIN-EN-ASME-JIS"
Vibration resistance
1)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
2)
Outside the specied beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)
3)
EIRP: Equivalent Isotropic Radiated Power
38296-EN-130624
70
11 Supplement
Ʋ Horn antenna 4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Parabolic antenna 1 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Shock resistance
Ʋ Horn antenna 100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Ʋ Parabolic antenna 25 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Data on rinsing air connection
Pressure max. 6 bar (87.02 psig) Air quantity without reux valve, with
Ʋ 0.2 bar (2.9 psig) 2.0 m3/h
Ʋ 0.5 bar (7.25 psig) 3.5 m3/h
Ʋ 1 bar (14.5 psig) 4.5 m3/h
Ʋ 1.5 bar (21.76 psig) 5 m3/h
Ʋ 2 bar (29.0 psig) 5.5 m3/h
Thread
G⅛ A
Closure with
Ʋ Non-Ex Dust protection cover of PE
Ʋ Ex Threaded plug of 316Ti
Reux valve - unmounted (as option with non-Ex version, included in the scope of delivery with Ex version)
Ʋ Material 316Ti
Ʋ Seal FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375)
Ʋ for tube diameter 6 mm
Ʋ opening pressure 0.5 bar (7.25 psig)
Ʋ Nominal pressure stage PN 250
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland M20 x 1.5 (cable: ø 5 … 9 mm)
Ʋ Cable entry ½ NPT
Ʋ Blind plug M20 x 1.5; ½ NPT
Ʋ Closing cap M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF
Ʋ Indication circuit M12 x 1 plug Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
38296-EN-130624
71
11 Supplement
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
M20 x 1.5 (cable: ø 5 … 9 mm)
tion cable
Ʋ Cable entry ½ NPT
Ʋ Blind plug M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section 0.5 mm² (AWG 20)
Ʋ Wire resistance < 0.036 Ω/m
Ʋ Tensile strength < 1200 N (270 lbf)
Ʋ Standard length 5 m (16.4 ft)
Ʋ Max. length 180 m (590.6 ft)
Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F)
Ʋ Diameter approx. 8 mm (0.315 in)
Ʋ Colour - Non-Ex version Black
Ʋ Colour - Ex-version Blue
Display and adjustment module
Display element Display with backlight Measured value indication
Ʋ Number of digits 5
Ʋ Size of digits W x H = 7 x 13 mm
Adjustment elements 4 keys
Protection rating
Ʋ unassembled IP 20
Ʋ mounted into the housing without
IP 40
cover
Materials
Ʋ Housing ABS
Ʋ Inspection window Polyester foil
Integrated clock
Date format Day.Month.Year Time format
12 h/24 h
Time zone Ex factory CET
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy ±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 9 … 32 V DC
72
38296-EN-130624
11 Supplement
Ʋ EEx-ia instrument - Power supply
9 … 17.5 V DC
FISCO model
Ʋ EEx-ia instrument - Power supply
9 … 24 V DC
ENTITY model
Ʋ EEx-d instrument 14 … 32 V DC Number of sensors per DP/PA segment coupler max.
Ʋ Non-Ex 32
Ʋ Ex 10
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67
Ʋ Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar)
4)
housing - investment casting, stain­less steel housing - electro-polished
Ʋ Aluminium and stainless housing, in-
IP 66/IP 68 (1 bar)
vestment casting (optionally available) Overvoltage category III Protection class III
Approvals
Instruments with approvals can have dierent technical data depending on the version. For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".
11.2 CommunicationProbusPA
Instrumentmasterle
The instrument master le (GSD) contains the characteristic data of the Probus PA instrument. These data are, e.g. the permissible transmission rates as well as information on diagnostics values and the format of the measured value outputted by the PA instrument.
A bitmap le is also provided for the Probus network planning tool. This le is installed automati­cally when the GSD le is integrated. The bitmap le is used for symbolic indication of the PA instru­ment in the conguration tool.
ID number
Each Probus instrument gets an unambiguous ident number (ID number) from the Probus user organisation (PNO). This ID number is also included in the name of the GSD le. Optionally to this manufacturer-specic GSD le, PNO provides also a general so-called prole-specic GSD le. If the general GSD le is used, the sensor must be set to the prole-specic ident number via the DTM software. By default, the sensor operates with the manufacturer-specic ID number. When using the instruments on a segment coupler SK-2 or SK-3, no special GSD les are required.
The following table shows the instrument ID and the GSD le names for VEGAPULS radar sensors.
4)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
38296-EN-130624
73
11 Supplement
Device name Instrument ID GSDlename
VEGA Instrument class in
prole3.02
VEGAPULS WL 61 0x0CDB 0x9702 PS61WL0CDB.GSD PA139702.GSD
VEGAPULS 61 0x0BFC 0x9702 PS610BFC.GSD PA139702.GSD
VEGAPULS 62 0x0BFD 0x9702 PS620BFD.GSD PA139702.GSD
VEGAPULS 63 0x0BFE 0x9702 PS630BFE.GSD PA139702.GSD
VEGAPULS 65 0x0BFF 0x9702 PS650BFF.GSD PA139702.GSD
VEGAPULS 66 0x0C00 0x9702 PS660C00.GSD PA139702.GSD
VEGAPULS 67 0x0C01 0x9702 PS670C01.GSD PA139702.GSD
VEGAPULS SR 68 0x0CDC 0x9702 PS68R0CDC.GSD PA139702.GSD
VEGAPULS 68 0x0C02 0x9702 PS680C02.GSD PA139702.GSD
VEGA Prole-specic
Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during opera­tion. The below block diagram below shows which data can be accessed by the PLC.
Fig. 56: VEGAPULS SR 68: Block diagram with AI FB 1 … AI FB 3 OUT values
TB Transducer Block FB 1 … FB 3 Function Block
74
38296-EN-130624
11 Supplement
Module of the PA sensors
For the cyclic data trac, VEGAPULS SR 68 provides the following modules:
AI FB1 (OUT)
Out value of the AI FB1 after scaling
AI FB2 (OUT)
Out value of the AI FB2 after scaling
AI FB3 (OUT)
Out value of the AI FB3 after scaling
Free Place
This module must be used if a value in the data telegram of the cyclical data trac should not
be used (e.g. replacement of temperature and Additional Cyclic Value)
A maximum of three modules can be active. By means of the conguration software of the Probus master you can determine the conguration of the cyclical data telegram with these modules. The procedure depends on the respective conguration software.
Note:
The modules are available in two versions:
Short for Probus master supporting only one "Identier Format" byte, e.g. Allen
Bradley Long for Probus master only supporting the "Identier Format" byte, e.g. Siemens
S7-300/400
Examplesoftelegramconguration
In the following you will see how the modules can be combined and how the appendant data tel­egram is structured.
Example 1
AI FB1 (OUT)
AI FB2 (OUT)
AI FB3 (OUT)
Byte-
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
No.
For-
IEEE-754-Floating point
mat
Value
Example 2
AI FB1 (OUT)
Free Place
Free Place
Byte-No. 1 2 3 4 5
Format IEEE-754-Floating point value Status
Value
value
AI FB1 (OUT) AI
Note:
Bytes 6-15 are not used in this example.
Sta-
IEEE-754-Floating point
tus
FB1
Sta-
value
AI FB2 (OUT) AI
AI FB1 (OUT) AI FB1
IEEE-754-Floating point
tus
FB2
value
AI FB3 (OUT) AI
Sta­tus
FB3
38296-EN-130624
75
11 Supplement
Sign
-23
Data format of the output signal
Byte4
Byte3
Byte2
Byte1
Byte0
Status Value (IEEE-754)
Fig. 57: Data format of the output signal
The status byte corresponds to prole 3.02 "Probus PA Prole for Process Control Devices" coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit6 … 0 = 0).
The measured value is transferred as a 32 bit oating point number in the IEEE-754 format.
Byte n
Bit
6
7
2
Bit
Bit
4
5
6543
2
2
Exponent
VZ
Bit
Bit
7
VZ
Bit
Value = (-1)
Fig. 58: Data format of the measured value
Bit
Bit
3
2
1
2
2
2
(Exponent - 127)
2
Bit
Bit
Bit
0
7
210-1
2
2
(1 + Significant)
Bit
6
5
-2-3-4-5-6 -7
2
2
Byte n+1 Bit
Bit
Bit
4
3
2
2
2
2
Significant
Bit
Byte n+2
Bit
Bit
7
0
1
2
2
2
Bit
-9
Bit
5
-10 -11
2
Bit
4
3
2
2
Bit
6
-8
2
Bit
Bit
Bit
Bit
0
2
1
-12
2
7
-13
-14 -15
2
2
2
Significant
Byte n+3
Bit
Bit
6
-16
2
Bit
Bit
5
-17
2
Bit
Bit
4
-19
-18
2
Bit
3
2
1
0
-20
-22
-21
2
2
2
2
Significant
Coding of the status byte associated with the PA output value
You can nd further information for the coding of the status byte in the Device Description 3.02 on www.probus.com.
Status code Description according to
Probusstandard
0 x 00 bad - non-specic Flash-Update active
0 x 04 bad - conguration error – Adjustment error
0 x 0C bad - sensor failure – Hardware error
0 x 10 bad - sensor failure – Measured value generation error
0 x 1f bad - out of service con-
stant
0 x 44 uncertain - last unstable
value
0 x 48 uncertain substitute set – Switch on simulation
0 x 4c uncertain - initial value Failsafe replacement value (Failsafe-Mode = "Last valid
Possible cause
– Conguration error with PV-Scale (PV-Span too
small)
Unit irregularityError in the linearization table
Converter errorLeakage pulse error – Trigger error
– Temperature measurement error
"Out of Service" mode switched on
Failsafe replacement value (Failsafe-Mode = "Last val­ue" and already valid measured value since switching on)
– Failsafe replacement value (Failsafe-Mode = "Fsafe
value")
value" and no valid measured value since switching on)
38296-EN-130624
76
11 Supplement
Status code Description according to
Probusstandard
0 x 51 uncertain - sensor; con-
version not accurate - low limited
0 x 52 uncertain - sensor; con-
version not accurate - high limited
0 x 80 good (non-cascade) - OK OK
0 x 84 good (non-cascade) - ac-
tive block alarm
0 x 89 good (non-cascade) - ac-
tive advisory alarm - low limited
0 x 8a good (non-cascade) - ac-
tive advisory alarm - high limited
0 x 8d good (non-cascade) - ac-
tive critical alarm - low limited
0 x 8e good (non-cascade) - ac-
tive critical alarm - high limited
Possible cause
Sensor value < lower limit
Sensor value > upper limit
Static revision (FB, TB) changed (10 sec. active, after the parameter of the static category has been written)
Lo-Alarm
Hi-Alarm
Lo-Lo-Alarm
Hi-Hi-Alarm

11.3 Dimensions

The following dimensional drawings represent only an extract of all possible versions. Detailed dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 84 mm
(3.31")
ø 79 mm
(3.11")
M16x1,5
(4.41")
112 mm
(4.41")
112 mm
M20x1,5/ ½ NPT
M20x1,5/
1
Fig. 59: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
½ NPT
2
38296-EN-130624
77
11 Supplement
Aluminium housing
~ 116 mm
(4.57")
ø 86 mm
(3.39")
(4.57")
116 mm
~ 87 mm
(3.43")
M16x1,5
ø 86 mm
(3.39")
(4.72")
120 mm
M20x1,5/
M20x1,5
½ NPT
M20x1,5/ ½ NPT
Fig. 60: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
21
Aluminium housing in protection rating IP 66/IP 68 (1 bar)
~ 150 mm
(5.91")
ø 84 mm
(3.31")
(4.57")
116 mm
M20x1,5
M20x1,5
1
Fig.61:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version 2 Double chamber version
~ 105 mm
(4.13")
M20x1,5/ ½ NPT
ø 84 mm
(3.31")
2
M16x1,5
(4.72")
120 mm
78
38296-EN-130624
Stainless steel housing
~ 59 mm
(2.32")
ø 80 mm
(3.15")
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 87 mm
(3.43")
M16x1,5
11 Supplement
ø 86 mm
(3.39")
(4.41")
112 mm
(4.61")
117 mm
120 mm
M20x1,5/
½ NPT M20x1,5/ ½ NPT
Fig. 62: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
M20x1,5/ ½ NPT
321
Stainless steel housing in protection rating IP 66/IP 68, 1 bar
(4.41")
112 mm
~ 105 mm (4.13")
M20x1,5/ ½ NPT
ø 86 mm
(3.39")
2
M16x1,5
(4.72")
120 mm
~ 103 mm
(4.06")
M20x1,5
ø 79 mm
(3.11")
3
117 mm
~ 93 mm
(3.66")
M20x1,5/ ½ NPT
ø 80 mm
(3.15")
1
Fig.63:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
(4.72")
(4.61")
38296-EN-130624
79
11 Supplement
VEGAPULS SR 68, horn antenna in threaded version
38 mm (1.50")
SW 46 mm
(1.81")
G1½A / 1½ NPT
x
Fig. 64: VEGAPULS SR 68, horn antenna in threaded version
1 Standard 2 With temperature adapter up to 250 °C
22 mm (0.87")
100 ø40
120 ø48
216 ø75
430 ø95
3.94" ø1.58" ø1.89"
4.72"
8.50"
ø2.95"
16.93"
ø3.74"
x y
x y
mm
1½"
y
2"
3"
4"
inch
1½"
2"
3"
4"
2 1
144 mm (5.67")
22 mm (0.87")
y
x
80
38296-EN-130624
VEGAPULSSR68,hornantennainangeversion
(2.36")
60 mm
11 Supplement
21
y
x
mm
1½"
2"
3"
4"
100 ø40
120 ø48
216 ø75
430 ø95
xy
inch
1½"
2"
3"
4"
xy
3.94" ø1.58" ø1.89"
4.72"
8.50"
ø2.95"
16.93"
ø3.74"
Fig.65:VEGAPULSSR68,hornantennainangeversion
1 Standard 2 With temperature adapter up to 250 °C
165 mm (6.5")
38296-EN-130624
81
11 Supplement
VEGAPULS SR 68, horn antenna and swivelling holder
1
40 mm
(1.58")
20 mm
(0.79")
100 ø 40
120 ø 48
216 ø 75
430 ø 95
3.94" ø 1.58" ø 1.89"
4.72" ø 2.95"
8.50"
ø 3.74"
16.93"
x y
x y
mm
y
1½"
2"
3"
x
4"
inch
1½"
2"
3"
4"
146 mm (5.75")
x
max. 15°
(0.59")
2
20 mm
(0.79")
y
Fig. 66: VEGAPULS SR 68, horn antenna and swivelling holder
1 Standard 2 With temperature adapter up to 250 °C
82
38296-EN-130624
VEGAPULS SR 68, parabolic antenna and swivelling holder
11 Supplement
max. 15°
(0.59")
1
40 mm
(1.58")
ø 243 mm (9.57")
Fig. 67: VEGAPULS SR 68, parabolic antenna and swivelling holder
1 Standard 2 With temperature adapter up to 200 °C
130 mm (5.12")
2
120 mm (4.72")
38296-EN-130624
83
11 Supplement

11.4 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

11.5 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/
originator.
84
38296-EN-130624
INDEX
INDEX
A
Accessories
Display and adjustment module 12 – External display and adjustment unit 12 – External radio unit 12 – Flanges 13 – Interface adapter 12 – Protective cap 12
Adjustment
Max. adjustment 47Min. adjustment 46
Adjustment system 39 Agitator 25
Application
Bulk solid 43 – Liquid 40 – Standpipe 41
C
Cable entry 30 Connection procedure 31
Connection technology 31 Curve indication
Echo curve 48 – False signal suppression 48
Cyclical data trac 74
D
Data format, output signal 76 Deviation 61
E
Echo curve memory 57 Echo curve of the setup 48 EDD (Enhanced Device Description) 56 Electronics compartment with Ex-d double chamber housing 34
Error messages 58 Event memory 57
F
False signal suppression 49 Fault rectication 61 Functional principle 11
K
Key function 38
L
Linearization curve 50
M
Main menu 39
Meas. reliability 47 Measured value memory 57
Medium
Liquid Bulk solid 40
Moisture 14
N
NAMUR NE 107
Failure 59 – Function check 60 – Maintenance 60 – Out of specication 60
O
Overll protection according to WHG 50
P
Packaging 11
PA modules 75 Peak value 47 Polarisation 16
R
Reection properties 40 Repair 65
Replacement parts
– Antenna impedance cone 13 – Electronics module 13
S
Service hotline 64 Socket 19 Status bytes PA output value 76 Status messages 58 Storage 12 Subsurface enclosure 27
G
GSD le 73
I
Instrument master le 73
38296-EN-130624
T
Telegram conguration 75 Type plate 10
85
INDEX
V
Vessel form 45 Vessel height 45 Vessel installations 25 Vessel insulation 26
86
38296-EN-130624
Notes
38296-EN-130624
87
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2013
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
38296-EN-130624
Loading...