VEGA PULS67 User Manual

Operating Instructions
Radar sensor for continuous level measurement of bulk solids
VEGAPULS 67
HART - integrated accumulator
Document ID: 40795

Quick start

Mounting
Connect electrically
Quick start
The quick start enables a quick setup with many applications. You can nd further information in the respective chapters of the operating instructions manual.
1. Distance from the vessel wall > 200 mm
200 mm
(7.87
")
Fig. 1: Distance of the antenna to the vessel wall/socket conguration
2. Keep the socket short, round o the socket end For further information see chapter "Mounting".
1. Insert the battery charger into the socket of the power supply compartment and load the integrated accumulator before setting up the instrument.
2. Switch on the instrument by means of the button outside on the housing.
1
Set parameters
2
Fig. 2: Switch on/o button outside on the housing
1 On/O button
For further information see chapter "Connecting to power supply".
1. Go via the indicating and adjustment module to the menu "Setup".
2. Select in the menu item "Medium" the medium of your applica­tion, for example "Gravel/Pebble".
3. Select in the menu item "Application" the vessel, the application and the vessel form, for example, silo.
40795-EN-121011
Quick start
4. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment".
Parameterization example
Further steps
The radar sensor measures the distance from the sensor to the product surface. For indication of the real level, an allocation of the measured distance to the percentage height must be carried out.
")
0,5 m
(19.68
100%
2
")
10 m
(393.7
0%
1
Fig. 3: Parameterization example
1 Min. level max. meas. distance 2 Max. level min. meas. distance
For this adjustment, the distance is entered with full and virtually empty vessel. If these values are not known, an adjustment with distances, for example, of 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the thread or ange.
1. In the menu "Additional settings", menu item "Damping" you have to adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out- put".
The quick start is then nished. For further information see chapter "Parameter adjustment".
40795-EN-121011
3
Contents
Contents
1 About this document
1.1 Function ............................................................................. 6
1.2 Target group ....................................................................... 6
1.3 Symbolism used ................................................................. 6
2 For your safety
2.1 Authorised personnel ......................................................... 7
2.2 Appropriate use .................................................................. 7
2.3 Warning about incorrect use ............................................... 7
2.4 General safety instructions ................................................. 7
2.5 CE conformity ..................................................................... 8
2.6 NAMUR recommendations ................................................ 8
2.7 Radio license for Europe .................................................... 8
2.8 Radio license for USA/Canada ........................................... 8
2.9 Environmental instructions ................................................. 8
3 Product description
3.1 Conguration .................................................................... 10
3.2 Principle of operation ........................................................ 11
3.3 Packaging, transport and storage ..................................... 11
3.4 Accessories and replacement parts ................................. 12
4 Mounting
4.1 General instructions ......................................................... 13
4.2 Collar or adapter ange .................................................... 13
4.3 Mounting strap ................................................................. 14
4.4 Instructions for installation ................................................ 15
5 Connecting to power supply
5.1 Connection of the battery charger .................................... 24
5.2 Wiring plan ....................................................................... 24
5.3 Switch-on phase ............................................................... 25
6 Set up with the indicating and adjustment module
6.1 Insert indicating and adjustment module .......................... 27
6.2 Adjustment system ........................................................... 28
6.3 Parameter adjustment ...................................................... 29
6.4 Saving the parameter adjustment data ............................. 42
7 Setup with PACTware
7.1 Connect the PC ................................................................ 44
7.2 Parameter adjustment with PACTware .............................. 45
7.3 Saving the parameter adjustment data ............................. 46
8 Set up with other systems
8.1 DD adjustment programs ................................................. 47
8.2 Communicator 375, 475 ................................................... 47
9 Diagnosis, Asset Management and service
9.1 Maintenance .................................................................... 48
9.2 Measured value and event memory ................................. 48
9.3 Asset Management function ............................................. 49
9.4 Rectify faults ..................................................................... 52
40795-EN-121011
4
9.5 Exchanging the electronics module .................................. 55
9.6 Software update ............................................................... 56
9.7 How to proceed in case of repair ...................................... 56
10 Dismounting
10.1 Dismounting steps ............................................................ 57
10.2 Disposal ........................................................................... 57
11 Supplement
11.1 Technical data .................................................................. 58
11.2 Dimensions ...................................................................... 63
Contents
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
40795-EN-121011
Editing status: 2012-09-27
5

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
40795-EN-121011
6

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

The VEGAPULS 67 is a sensor for continuous level measurement. Due to the integrated accumulator the instrument is particularly suitable as a portable measuring system or test sensor for special applications.
You can nd detailed information on the application range in chapter "Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula­tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.
Depending on the instrument version, the emitting frequencies are in the C or K band range. The low emitting frequencies are far below the internationally approved limit values. When used correctly, there is no danger to health.
40795-EN-121011
7
2 For your safety

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the conformity certicate in the download section of our homepage.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom­mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring
transducers NE 53 – Compatibility of eld devices and indicating/adjustment
components NE 107 - Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Radio license for Europe

The instrument is approved according to EN 302372-1/2 (2006-04) for use in closed vessels.

2.8 Radio license for USA/Canada

The instrument is in conformity with part 15 of the FCC regulations. Take note of the following two regulations:
The instrument must not cause any interfering emissions
The device must be insensitive to interfering immissions, including
those that may cause undesirable operating conditions
Modications not expressly approved by the manufacturer will lead to expiry of the operating licence according to FCC/IC.
The instrument is in conformity with RSS-210 of the IC regulations. The instrument may only be used in closed vessels made of metal,
concrete, or bre-reinforced plastic.

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
8
40795-EN-121011
Chapter "Disposal"
2 For your safety
40795-EN-121011
9

3 Product description

Type label
3 Product description
3.1 Conguration
The type label contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 4: Layout of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Measuring range 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Note to observe the instrument documentation 15 Notied authority for CE marking 16 Approval directive
16
15
14
13
12
11
Serial number
Scope of this operating instructions manual
10
With the serial number of the instrument on the type label you have access to the following data on our homepage:
Article number of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate "Measuring Accuracy" (PDF)
For this purpose, move to www.vega.com and "VEGA Tools".
This operating instructions manual applies to the following instrument versions:
Hardware from 2.1.0
Software from 4.5.1
40795-EN-121011
3 Product description
Scope of delivery
Application area
Functional principle
Voltage supply
The scope of delivery encompasses:
Radar sensor with integrated accumulator
Battery charger
Documentation
– this operating instructions manual – Test certicate measuring accuracy (optional) – Operating instructions manual "Indicating and adjustment
module" (optional)
– Ex-specic "Safety instructions" (with Ex versions) – if necessary, further certicates

3.2 Principle of operation

The VEGAPULS 67 is a radar sensor for continuous level measure­ment of bulk solids under simple process conditions. It is suitable for smaller silos and vessels.
The instrument is also suitable for applications in liquids.
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The transit time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropri­ate output signal and outputted as measured value.
The instrument is powered via an integrated accumulator. This allows operation of the instrument independently of the mains or a control system. To achieve a longer battery lifetime, the instrument is preset to HART multidrop. This puts a load of only 4 mA current on the ac­cumulator.

3.3 Packaging, transport and storage

Packaging
Transport
Transport inspection
Storage
40795-EN-121011
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
11
3 Product description
Storage and transport temperature
Indicating and adjust­ment module
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %

3.4 Accessories and replacement parts

The indicating and adjustment module PLICSCOM is used for meas­ured value indication, adjustment and diagnosis. It can be inserted into the sensor and removed at any time.
You can nd further information in the operating instructions "Indicat- ing and adjustment module PLICSCOM" (Document-ID 27835).
Interface adapter
Protective cover
Electronics module
Accumulator insert
The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, an adjustment software such as PACTware with VEGA-DTM is required.
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
The protective cover protects the sensor housing against soiling and intense heat from solar radiation.
You will nd additional information in the supplementary instructions manual "Protective cover" (Document-ID 34296).
The electronics module VEGAPULS series 60 is a replacement part for radar sensors of VEGAPULS series 60. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions "Elec- tronics module VEGAPULS series 60" (Document-ID 36801).
The accumulator insert VEGAPULS Serie 60 is an replacement part for radar sensors of VEGAPULS series 60 with integrated accumula­tor.
You can nd further information in the operating instructions "Accumu- lator insert VEGAPULS series 60" (Document-ID 41033).
40795-EN-121011
12
Screwing in

4 Mounting

4 Mounting

4.1 General instructions

With instruments with threaded process tting, suitable tools must be applied for tightening the hexagon.
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
Moisture
Suitability for the process conditions
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain o. This ap­plies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
Make sure that all parts of the instrument exposed to the process, in particular the active measuring component, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the type label.
4.2 Collaroradapterange
For mounting the instrument on a socket, a combi compression ange for DN 80 (ASME 3" or JIS 80) is also available for retro tting. Option­ally, the instrument can be also equipped with an adapter ange from DN 100 (ASME 4" or JIS 100).
With the housing versions plastic, aluminium single chamber and stainless steel, the collar ange can be placed directly over the hous­ing. With the aluminium double chamber housing, retroactive mount­ing in this way is not possible - the mounting type must be specied with the order.
You can nd drawings of these mounting options in chapter "Dimen- sions".
40795-EN-121011
13
4 Mounting
Mounting strap
Fig. 5: Flange mounting of the radar sensor

4.3 Mounting strap

The mounting strap enables simple mounting on the vessel wall or silo top. It is suitable for wall, ceiling or boom mounting. Especially in open vessels this is a very easy and eective way to align the sensor to the bulk solid surface.
14
Fig. 6: Radar sensor with mounting strap
The strap is supplied unassembled and must be screwed to the sensor before setup with three hexagon screws M5 x 10 and spring washers. Max. torque, see chapter "Technical data". Required tools: Allen wrench size 4.
There are two ways to screw the strap onto the sensor. Depending on the selected version, the sensors can be swivelled in the strap as follows:
Single chamber housing
– Angle of inclination 180°, innitely variable
40795-EN-121011
Tight installation of the plastic horn antenna
4 Mounting
– Angle of inclination in three steps 0°, 90° and 180°
Double chamber housing
– Angle of inclination 90°, innitely variable – Angle of inclination in two steps 0° and 90°
Fig. 7: Adjustment of the angle of inclination
Fig. 8: Turning by fastening in the centre

4.4 Instructions for installation

For tight installation of the version with plastic horn antenna with com­pression or adapter ange, the following conditions must be fullled:
1. Use suitable at seal, e.g. of EPDM with Shore hardness 25 or 50
2. Make sure the number of ange screws corresponds to the num­ber of ange holes
3. Tighten all screws with the torque stated in the technical data
Polarisation plane
40795-EN-121011
The emitted radar impulses of the radar sensor are electromagnetic waves. The polarisation plane is the direction of the electrical wave component. By turning the instrument in the connection ange or mounting boss, the polarisation can be used to reduce the eects of false echoes.
The position of the polarisation level is marked by markings on the instrument.
Fig. 9: Position of the polarisation level
1 Marking bar
1
15
4 Mounting
Installation position
Inowingmedium
Mount the instrument at least 200 mm (7.874 in) away from the vessel wall.
200 mm
")
(7.87
Fig. 10: Mounting of the radar sensor on the vessel top
If you cannot keep this distance you should carry out a false echo storage before setup. This applies mainly if buildup on the vessel wall is expected. In this case, we recommend repeating a false echo stor­age later with existing buildup.
Filling from the top: Mounting should not be too close to the inow­ing material as the microwave signal will be interferred. The optimum mounting position is on the opposite of the lling. To avoid strong pollution, the distance to the lter or dust extraction must be as big as possible.
16
Fig. 11: Mounting of the radar sensor with inowing medium with lling from the
top
Laterallling: With bulk solids silos with lateral pneumatic lling, mounting should not be in the lling stream as the microwave signal will be interferred. The optimum mounting position is next to the lling. To avoid strong pollution, the distance to the lter or dust extraction must be as big as possible.
40795-EN-121011
4 Mounting
Fig. 12: Mounting of the radar sensor with inowing medium with lateral lling
Socket with plastic horn antenna
A corresponding collar ange for DN 80 (ASME 3" or JIS 80) as well as a suitable adapter ange are available for mounting VEGAPULS
67.
With the housing versions plastic, aluminium single chamber and stainless steel, the collar ange can be placed directly over the hous­ing. With the aluminium double chamber housing, retroactive mount­ing in this way is not possible - the mounting type must be specied with the order.
Information:
The socket should be as short as possible and the socket end rounded. Thus, false reections by the vessel socket are reduced.
Fig. 13: Recommended socket mounting
40795-EN-121011
17
4 Mounting
If the medium has good reective properties, VEGAPULS 67 can also be mounted on a longer socket piece. Recommended values for socket heights are specied in the following illustration. You must carry out a false echo storage afterwards.
h
d
The below charts specify the max. socket piece length h depending on the diameter d.
Socket diameter d Socket length h
80 mm 300 mm
100 mm 400 mm
150 mm 500 mm
Socket diameter d Socket length h
3" 11.8 in
4" 15.8 in
6" 19.7 in
Tip:
In new facilities it is useful to incline the vessel socket in the direction of the outlet. False reections from the vessel wall are thus reduced and measurement all the way down to the bottom of the conical outlet is possible.
18
40795-EN-121011
Fig. 15: Alignment in silos
4 Mounting
Orientation
To detect nearly the complete vessel volume the sensor should be directed in such a way that the measuring beam reaches the lowest level in the vessel, i.e. the outlet.
Fig. 16: Mounting in open vessel
40795-EN-121011
19
4 Mounting
Fig. 17: Mounting on the vessel ceiling
Caution:
When mounting on the vessel ceiling mark sure that the sensor does not overll. Depending on the measured product, this can cause a mechanical damage.
In a cylindrical silo with conical outlet, the mounting iis carried out on a socket. It should be positioned on one third up to the half of the vessel radius.
r
1
1
rr...
2
3
/
/
Material heaps
20
Fig. 18: Mounting position and orientation
Large material heaps are best measured with several instruments, which can be mounted on e.g. traverse cranes. For this type of ap­plication it is advantageous to orient the sensor perpendicular to the bulk solid surface.
The sensors do not inuence each other.
40795-EN-121011
4 Mounting
Information:
Keep in mind that for these applications, the sensors are designed for relatively slow level changes. If the sensor is used on a movable boom, the max. measuring rate must be observed (see chapter "Technical data").
Fig. 19: Radar sensors on traverse crane
Vessel installations
Agitators
The mounting location of the radar sensor should be a place where no other equipment or xtures cross the path of the microwave signals.
Vessel installations, such as e.g. ladders, limit switches, heating spi­rals, struts, etc., can cause false echoes and impair the useful echo. Make sure when planning your measuring site that the radar sensor has a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal baes above the installations scatter the radar signals and prevent direct interfering reections.
Fig. 20: Cover smooth proles with deectors
If there are agitators in the vessel, a false signal suppression should be carried out with the agitators in motion. This ensures that the interfering reections from the agitators are saved with the blades in dierent positions.
40795-EN-121011
21
4 Mounting
Fig. 21: Agitators
Installation in subsurface enclosures
Mounting in multiple chamber silo
For level measurements in concrete silos, the sensors are often mounted in protective boxes. These boxes can be for example metal­lic, closed subsurface enclosures.
Minimal amounts of stray radiation from the sensor can be reected and strengthened by the walls of the subsurface enclosures. In the case of sensors with plastic housings, this can lead to coupling disturbances. This is avoided by using sensor housings of aluminium or stainless steel as well as electromagnetic absorber material in subsurface enclosures.
1 2
Fig. 22: Mounting of the instrument in an subsurface enclosure
1 Subsurface enclosure 2 Concrete bottom
The silo walls of multiple chamber silos are often made of prole walls, like e.g. prole sheeting, to ensure the required stability. If the radar sensor is mounted very close to a heavily structured vessel wall, substantial interfering reections can be generated. Hence the sensor should be mounted at a large distance from the separating wall.
The optimum mounting position is on the outer wall of the silo with the sensor oriented towards the outlet in the silo center.
40795-EN-121011
22
Fig. 23: Installation and orientation in multiple chamber silos
4 Mounting
Fig. 24: Installation and orientation in multiple chamber silos
Air rinsing
Air rinsing is useful for avoiding buildup, particularly when there is strong condensation. Since VEGAPULS 67 has no direct rinsing air connection, a separate rinsing air connection must be provided in the mounting socket. By inclining this connection towards the top, a particularly eective cleaning of the antenna cover is achieved.
Fig. 25: Purging air connection
40795-EN-121011
23

5 Connecting to power supply

5 Connecting to power supply

5.1 Connection of the battery charger

We recommended charging the integrated accumulator completely before setting up the instrument. You can nd the charging time in chapter "Technical data".
The battery charger is plugged into a socket in the supply room, see chapter "Wiring plan".
The LEDs in the supply room show the charging process and condi­tion of the accumulator, see chapter "Wiring plan".
Cable screening and grounding
Overview
Electronics compartment
If screened cable is necessary, connect the cable screen on both ends to ground potential. In the sensor, the screen must be connected directly to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisa­tion (low impedance).
If potential equalisation currents are expected, the connection on the processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interfer­ence signals remains.

5.2 Wiring plan

1
Fig. 26: Position of the power supply and electronics compartment
1 Supply room (accumulator) 2 Electronics compartment
4...20mA
2
2
24
( )
+
(-)
1
2
567
8
1
Fig. 27: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the indicating and adjustment module
40795-EN-121011
Supply room
5 Connecting to power supply
( )
+
1
1,2 OFF
3 ON/OFF 1h 4 ON/OFF
0
1
9
2
8
3
5..9 auto cycle
7
4
6
5
(-)
2
modes
Charge
Accu
3
4
5
2
1
Fig. 28: Supply room
1 Internal connection of the socket for the battery charger 2 Mode switch 3 LED green, charging process 4 LED yellow, charging status 5 Socket for the battery charger
The mode switch enables the selection of the following modes:
0 = sensor o, LEDs show the accumulator status
1, 2 = sensor o, LEDs o
3 = sensor on for 1 hour after pressing a key (delivery status)
4 = sensor permanently on, switching on/o via button
5 = sensor is switched on every 30 min. for 3 min.
6 = sensor is switched on every hour for 3 min.
7 = sensor is switched on every 6 hours for 3 min.
8 = sensor is switched on every 12 hours for 3 min.
9 = sensor is switched on every 24 hours for 3 min.
The green LED characterizes the charging process:
LED ashes = Accumulator is charging
LED lights = accumulator is full, battery charger should be
unplugged (accumulator life time)
After pressing the key or changing the mode the yellow LED shows the accumulator status for approximately 10 s as follows:
LED lights = accumulator is full
LED ashes = accumulator should be charged
LED o = accumulator is empty

5.3 Switch-on phase

The instrument is switched on and o by means of a button outside on the housing.
40795-EN-121011
25
5 Connecting to power supply
1
Fig. 29: Switch on/o button outside on the housing
1 On/O button
After switching on, the instrument carries out a self-check for approxi­mately 30 s:
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC Indication of the status message "F 105 Determine measured
value" on the display or PC
As soon as a plausible measured value is found, this value will be displayed. The value corresponds to the actual level as well as the settings already carried out, e.g. factory setting.
26
40795-EN-121011

6 Set up with the indicating and adjustment module

6 Set up with the indicating and adjustment
module

6.1 Insert indicating and adjustment module

The indicating and adjustment module can be inserted into the sen­sor and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the indicating and adjustment module in the requested position onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order. The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Fig. 30: Insertion of the indicating and adjustment module with single chamber housing into the electronics compartment
40795-EN-121011
27
6 Set up with the indicating and adjustment module
1 2
Fig. 31: Insertion of the indicating and adjustment module with double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with an indicating and adjust­ment module for continuous measured value indication, a higher cover with an inspection glass is required.
Key functions
28

6.2 Adjustment system

Fig. 32: Indicating and adjustment elements
1 LC display 2 Adjustment keys
[OK] key:
– Move to the menu overview – Conrm selected menu
1
2
40795-EN-121011
6 Set up with the indicating and adjustment module
– Edit parameter – Save value
[->] key:
– Presentation change measured value – Select list entry – Select editing position
[+] key:
– Change value of the parameter
[ESC] key:
– interrupt input – Jump to next higher menu
Adjustment system
Main menu
The sensor is adjusted via the four keys of the indicating and adjust­ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.

6.3 Parameter adjustment

Through the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g., for measurement loop name, medium, applica­tion, vessel, adjustment, signal output
Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g. on the instrument status, pointer, reli-
ability, simulation, echo curve Further settings: Instrument unit, false signal suppression, linearisa-
tion curve, reset, date/time, reset, copy function Info: Instrument name, hardware and software version, calibration
date, instrument features
Information:
In this operating instructions manual, the instrument-specic param­eters in the menu sections "Setup", "Diagnosis" and "Additional set- tings" are described. The general parameters in these menu section are described in the operating instructions manual "Indicating and adjustment module".
40795-EN-121011
29
6 Set up with the indicating and adjustment module
You can nd in the operating instructions manual "Indicating and ad- justment module" also the description of the menu sections "Display" and "Info".
In the main menu point "Setup", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the measurement. The proce­dure is described in the following.
Setup/Medium
Setup/Application
Each medium has dierent reection properties. With liquids, further interfering factors are uctuation product surface and foam genera­tion. With bulk solids, these are dust generation, material cone and additional echoes from the vessel wall.
To adapt the sensor to these dierent measuring conditions, the selection "Liquid" or "Bulk solid" should be made in this menu item.
Through this selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with poor reec­tive properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
In addition to the medium, also the application, i.e. the measuring site, can inuence the measurement.
With this menu item, the sensor can be adapted to the applications. The adjustment possibilities depend on the selection "Liquid" or "Bulk solid" under "Medium".
30
The following options are available when "Liquid" is selected:
40795-EN-121011
The selection "Standpipe" opens a new window in which the inner diameter of the applied standpipe is entered.
The following features form the basis of the applications:
Storage tank:
Setup: large-volumed, upright cylindrical, spherical
Product speed: slow lling and emptying
Process/measurement conditions:
– Condensation – Smooth product surface – Max. requirement to the measurement accuracy
Storage tanke with product circulation:
Setup: large-volumed, upright cylindrical, spherical
Product speed: slow lling and emptying
Vessel: small laterally mounted or large top mounted stirrer
Process/measurement conditions:
– Relatively smooth product surface – Max. requirement to the measurement accuracy – Condensation – Slight foam generation – Overlling possible
Storage tank on ships (Cargo Tank):
Product speed: slow lling and emptying
Vessel:
– Installations in the bottom section (bracers, heating spirals) – High sockets 200 … 500 mm, also with large diameters
Process/measurement conditions:
– Condensation, buildup by movement – Max. requirement on measurement accuracy from 95 %
Stirrer vessel (reactor):
Setup: all vessel sizes possible
Product speed:
– Fast to slow lling possible – Vessel is very often lled and emptied
Vessel:
– Socket available – Large agitator blades of metal – Vortex breakers, heating spirals
Process/measurement conditions:
– Condensation, buildup by movement – Strong spout generation – Very agitated surface, foam generation
Dosing vessel:
Setup: all vessel sizes possible
Product speed:
40795-EN-121011
6 Set up with the indicating and adjustment module
31
6 Set up with the indicating and adjustment module
– Fast lling and emptying – Vessel is very often lled and emptied
Vessel: narrow installation situation
Process/measurement conditions:
– Condensation, buildup on the antenna – Foam generation
Standpipe:
Product speed: very fast lling and emptying
Vessel:
– Vent hole – Joins like anges, weld joints – Shifting of the running time in the tube
Process/measurement conditions:
– Condensation – Buildup
Bypass:
Product speed:
– Fast up to slow lling with short up to long bypass tube possible – Often the level is hold via a control facility
Vessel:
– Lateral outlets and inlets – Joins like anges, weld joints – Shifting of the running time in the tube
Process/measurement conditions:
– Condensation – Buildup – Separation of oil and water possible – Overlling into the antenna possible
Plastic tank:
Vessel:
– Measurement x mounted or integrated – Measurement depending on the application through the vessel
top
– With empty vessel, the measurement can be carried out
through the bottom
Process/measurement conditions:
– Condensation on the plastic ceiling – In outside facilities water and snow on the vessel top possible
Transportable plastic tank:
Vessel:
– Material and thickness dierent – Measurement through the vessel top
Process/measurement conditions:
– Measured value jump with vessel change
Open water (gauge measurement):
Gauge rate of change: slow gauge change
Process/measurement conditions:
– Distance sensor to water surface to big
40795-EN-121011
32
6 Set up with the indicating and adjustment module
– Extreme damping of output signal due to wave generation – Ice and condensation on the antenna possible – Spiders and insect nestle in the antennas – Floating material and animals sporadically on the water surface
Openume(owmeasurement):
Gauge rate of change: slow gauge change
Process/measurement conditions:
– Ice and condensation on the antenna possible – Spiders and insect nestle in the antennas – Smooth water surface – Exact measurement result required – Distance to the water surface normally relatively high
Rain water overfall (weir):
Gauge rate of change: slow gauge change
Process/measurement conditions:
– Ice and condensation on the antenna possible – Spiders and insect nestle in the antennas – Turbulent water surface – Sensor ooding possible
Demonstration:
Adjustment for all applications which are not typically level meas-
urement
Sensor accepts all measured value changes within the measuring
range immediately Typical applications:
– Instrument demonstration – Object recognition/monitoring (additional settings required)
Caution:
If a separation of liquids with dierent dielectric gure occurs in the vessel, for example through condensation, the radar sensor can detect under certain circumstances only the medium with the higher dielectric gure. Keep in mind that layer interfaces can cause faulty measurements.
If you want to measure the total height of both liquids reliably, please contact our service department or use an instrument specially de­signed for interface measurement.
The following options are available when "Bulk solid" is selected:
The following features form the basis of the applications:
Silo (slim and high):
Vessel of metal: weld joints
Process/measurement conditions:
– Filling too close to the sensor – System noise with completely empty silo increased
40795-EN-121011
33
6 Set up with the indicating and adjustment module
– Automatic false signal suppression with partly lled vessel
Bunker (large-volumed):
Vessel of concrete or metal:
– Structured vessel walls – Installations present
Process/measurement conditions:
– Large distance to the medium – Large angles of repose
Bunkerwithfastlling:
Vessel of concrete or metal, also multiple chamber silo:
– Structured vessel walls – Installations present
Process/measurement conditions:
– Measured value jumps, e.g. by truck loading – Large distance to the medium – Large angles of repose
Heap:
Sensor mounting on movable conveyor belts
Detection of the heap prole
Height detection during lling
Process/measurement conditions:
– Measured value jumps, e.g. by the prole of the heap or trav-
erses
– Large angles of repose – Measurement near the lling stream
Crusher:
Vessel: installations, wear and protective facilities available
Process/measurement conditions:
– Measured value jumps, e.g. by truck loading – Fast reaction time – Large distance to the medium
Demonstration:
Adjustment for all applications which are not typically level meas-
urement
Sensor accepts all measured value changes within the measuring
range immediately Typical applications:
– Instrument demonstration – Object recognition/monitoring (additional settings required)
Through this selection, the sensor is adapted optimally to the applica­tion or the location and measurement reliability under the various basic conditions is increased considerably.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
40795-EN-121011
34
6 Set up with the indicating and adjustment module
Setup/Vessel height, measuring range
Setup/Vessel form
With this selection, the operating range of the sensor is adapted to the vessel height and the reliability with dierent frame conditions is increased considerably.
Independent from this, the min. adjustment must be carried out.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Also the vessel form can inuence the measurement apart from the medium and the application. To adapt the sensor to these measure­ment conditions, this menu item oers you dierent options for vessel bottom and ceiling in case of certain applications.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup/Adjustment
40795-EN-121011
Since a radar sensor is a distance measuring instrument, the distance from the sensor to the product surface is measured. For indication of the real level, an allocation of the measured distance to the percent­age height must be carried out.
To perform the adjustment, enter the distance with full and empty ves­sel, see the following example:
35
6 Set up with the indicating and adjustment module
Fig. 33: Parameter adjustment example min./max. adjustment
1 Min. level = max. meas. distance 2 Max. level = min. meas. distance
If these values are not known, an adjustment with the distances of for example 10 % and 90 % is possible. Starting point for these distance specications is always the seal surface of the thread or ange. By means of these settings, the real level will be calculated.
The real product level during this adjustment is not important, be­cause the min./max. adjustment is always carried out without chang­ing the product level. These settings can be made ahead of time without the instrument having to be installed.
0,5 m
")
(19.68
10 m
100%
")
(393.7
0%
2
1
Setup/Min. adjustment
36
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
40795-EN-121011
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
6 Set up with the indicating and adjustment module
4. Enter the suitable distance value in m for the empty vessel (e.g. distance from the sensor to the vessel bottom) corresponding to the percentage value.
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
Setup/Max. adjustment
Diagnosis/Peak value
Proceed as follows:
1. Select with [->] the menu item max. adjustment and conrm with [OK].
2. Prepare the percentage value for editing with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4. Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. Keep in mind that the max. level must lie below the min. distance to the antenna edge.
5. Save settings with [OK]
The respective min. and max. measured value is saved in the sensor. The values are displayed in the menu item "Peak values".
Diagnosis/Measurement reliability
40795-EN-121011
When non-contact level sensors are used, the measurement can be inuenced by the respective process conditions. In this menu item, the measurement reliability of the level echo is displayed as dB value. The measurement reliability equals signal strength minus noise. The higher the value, the more reliable the measurement. With a function­ing measurement, the values are > 10 dB.
37
6 Set up with the indicating and adjustment module
Diagnoses/Curve indica­tion
The "Echo curve" shows the signal strength of the echoes over the measuring range in dB. The signal strength enables an evaluation of the quality of the measurement.
The "False signal suppression" displays the saved false echoes (see menu "Additional settings") of the empty vessel with signal strength in "dB" over the measuring range.
A comparison of echo curve and false signal suppression allows a more accurate conclusion on measurement reliability. The selected curve is continuously updated. With the [OK] key, a submenu with zoom functions is opened:
"X-Zoom": Zoom function for the meas. distance
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "dB"
"Unzoom": Reset the presentation to the nominal measuring range
with single magnication
Diagnostics/Echo curve memory
38
With the function "Echo curve memory" the echo curve can be saved at the time of setup. This is generally recommended; for using the As­set Management functions it is absolutely necessary. If possible, the curve should be saved with a low level in the vessel.
With the adjustment software PACTware and the PC, the high resolu­tion echo curve can be displayed and used to recognize signal changes over the operating time. In addition, the echo curve of the setup can be also displayed in the echo curve window and compared with the actual echo curve.
40795-EN-121011
6 Set up with the indicating and adjustment module
Additional adjustments/ False signal suppression
The following circumstances cause interfering reections and can inuence the measurement:
High sockets
Vessel installations such as struts
Agitators
Buildup or welded joints on vessel walls
Note:
A false signal suppression detects, marks and saves these false sig­nals so that they are no longer taken into account for level measure­ment.
This should be done with the low level so that all potential interfering reections can be detected.
Proceed as follows:
1. Select the menu item "Additional settins" with [->] and conrm with [OK]. With [->] you have to select the menu item "False signal suppression" and conrm with [OK].
2. Conrm again with [OK].
3. Conrm again with [OK].
4. Conrm again with [OK] and enter the actual distance from the sensor to the product surface.
5. All interfering signals in this section are detected by the sensor and stored after conrming with [OK].
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as false signal. The lling level would then no longer be detectable in this area.
40795-EN-121011
39
6 Set up with the indicating and adjustment module
If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression":
The menu item "Delete" is used to completely delete an already cre­ated false signal suppression. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel.
The menu item "Extend" is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with a too high level and not all false signals could be de­tected. When selecting "Extend", the distance to the product surface of the created false signal suppression is displayed. This value can now be changed and the false signal suppression can be extended to this range.
Additional adjustments/ Linearization curve
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindri­cal or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [ESC] and [->] key.
Caution:
Note the following if the instrument with corresponding approval is used as part of an overll protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer linearly proportional to the level. This must be taken into consideration by the user, particularly when setting the switching point on the level switch.
40795-EN-121011
40
6 Set up with the indicating and adjustment module
Additional adjustments
- Reset
When a reset is carried out, all settings (with only a few exceptions) are reset. The exceptions are: PIN, language, lighting, SIL and HART mode.
The following reset functions are available: Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specic settings. A created false signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special and laboratory parameters to the default values of the respective instru­ment. A created false signal suppression, user programmable lineari­zation curve as well as the measured value memory will be deleted.
Setup: Resetting of the parameter settings to the default values of the respective instrument in the menu item Setup. User-generated false signal suppression, user-programmed linearization curve, measured value memory as well as event memory remain untouched. The linearization is set to linear.
False signal suppression: Deleting a previously created false signal suppression. The false signal suppression created in the factory remains active.
Peak values distance: Resetting the measured min. and max. dis­tances to the actual measured value.
The following table shows the default values of the instrument. De­pending on the instrument version, not all menu items are available or dierently assigned:
40795-EN-121011
41
6 Set up with the indicating and adjustment module
Menu Menu item Default value
Setup Measurement
Display Language Like order
Additional adjust-
ments
loop name
Medium Dependent on the sensor:
Applications Demonstration
Vessel form Vessel bottom, dished boiler end
Vesell height/ Measuring range
Min. adjustment Recommended measuring range, see
Max. adjustment 0,000 m(d)
Damping 0.0 s
Current output mode
Current output Min./Max.
Lock operation Released
Displayed value Distance
Display unit m
Scaling size Volume
Scaling 0.00 lin %, 0 l
Backlight Switched o
Distance unit m
Temperature unit °C
Probe length Length of the standpipe Ex factory
Linearisation
curve
HART mode Multidrop
Sensor
Liquid/Water Bulk solids/Crushed stones, gravel
Vessel top, dished boiler end
Recommended measuring range, see "Technical data" in the supplement
"Technical data" in the supplement
4 … 20 mA, < 3.6 mA
Min. current 3.8 mA, max. current
20.5 mA
l
100.00 lin %, 100 l
Linear
Address 1
42

6.4 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If the instrument is equipped with an indicating and adjustment module, the data in the sensor can be saved in the indicating and adjustment module. The procedure is described in the operating
40795-EN-121011
6 Set up with the indicating and adjustment module
instructions manual "Indicating and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
The following data or settings for adjustment of the indicating and adjustment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Sensor-specic
units, temperature unit and linearization" The values of the user programmable linearisation curve
The function can also be used to transfer settings from one instru­ment to another instrument of the same type. If it is necessary to exchange a sensor, the indicating and adjustment module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy sensor data".
40795-EN-121011
43

7 Setup with PACTware

7 Setup with PACTware

7.1 Connect the PC

Via the interface adapter directly on the sensor
Via the interface adapter and HART
2
1
3
Fig. 34: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 4 3 Sensor
Information:
The interface adapter VEGACONNECT 3 is not suitable for connec­tion to the sensor.
2 4
1
LOCK
USB
TWIST
3
OPEN
44
Fig. 35: Connecting the PC via HART to the signal cable
1 Sensor 2 HART resistance 250 Ω (optional depending on processing) 3 Connection cable with 2 mm pins and terminals 4 Processing system/PLC/Voltage supply 5 Interface adapter, for example VEGACONNECT 4
Note:
With power supply units with integrated HART resistance (internal resistance approx. 250 Ω), an additional external resistance is not necessary. This applies, e.g. to the VEGA instruments VEGATRENN 149A, VEGAMET 381, VEGAMET 391. Common Ex separators are also usually equipped with a sucient current limitation resistance. In
40795-EN-121011
Prerequisites
7 Setup with PACTware
such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration).

7.2 Parameter adjustment with PACTware

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
Fig. 36: Example of a DTM view
Standard/Full version
40795-EN-121011
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value
45
7 Setup with PACTware
and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega. com/downloads and "Software". The full version is available on CD from the agency serving you.

7.3 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
46
40795-EN-121011

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".

8.2 Communicator 375, 475

Device descriptions for the instrument are available as DD or EDD for parameter adjustment with the Field Communicator 375 or 475.
The les can be downloaded at www.vega.com/downloads under "Software".
40795-EN-121011
47

9 Diagnosis, Asset Management and service

9 Diagnosis, Asset Management and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal operation.
In some applications, buildup on the antenna system can inuence the measuring result. Depending on the sensor and application, make arrangements to avoid strong pollution of the antenna system. If nec­essary, clean the antenna system in certain intervals.

9.2 Measured value and event memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
Event memory
Echo curve memory
Up to 60,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switching on and o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control system with EDD.
The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup. Changes in the measure­ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via:
40795-EN-121011
48
9 Diagnosis, Asset Management and service
PC with PACTware/DTM
Control system with EDD
Indicating and adjustment module
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD

9.3 Asset Management function

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the indicating and adjustment module, PACTware/DTM and EDD.
Status messages
40795-EN-121011
The status messages are classied in the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
Fig. 37: Pictograms of the status messages
1 Failure - red 2 Function check - orange 3 Out of specication - yellow 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is outputted.
This status message is always active. It cannot be deactivated by the user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
41 2 3
49
9 Diagnosis, Asset Management and service
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Failure
The following table shows the error codes in the status message "Failure" and gives information on the reason and rectication. Keep in mind that some information is only valid with four-wire instruments.
Code
Text mes­sage
F013 no measured
value avail­able
F017 Adjustment
span too small
F025
Error in the linearization table
F036
No operable software
F040
Error in the electronics
F080 – General software error – Separate operating voltage
F105
Determine
measured value
F113 Communica-
tion error
F125 Unpermissi-
ble electronics temperature
F260
Error in the calibration
Cause Rectication
– Sensor does not detect an
echo during operation
– Antenna system contami-
nated or defective
– Adjustment not within
specication
– Index markers are not con-
tinuously rising, for examle unlogical value pairs
– Failed or interrupted soft-
ware update
– Hardware defect – Exchanging the electronics
– The instrument is still in the
start phase, the measured value could not yet be determined
– EMC interferences – Transmission error with the
external communication with 4-wire power supply
unit
– Temperature of the elec-
tronics in the non-specied
section
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Check or correct installation
and/or parameter adjust-
ment
– Clean or exchange process
component or antenna
– Change adjustment accord-
ing to the limit values (dif­ference between min. and max. ≥ 10 mm)
– Check linearization table – Delete table/Create new
– Repeat software update – Check electronics version – Exchanging the electronics – Send instrument for repair
– Send instrument for repair
briey
– Wait for the warm-up phase – Duration depending on
the version and parameter adjustment up to approxi­mately 3 min.
– Remove EMC inuences – Exchange 4-wire power
supply unit or electronics
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
– Exchanging the electronics – Send instrument for repair
40795-EN-121011
50
9 Diagnosis, Asset Management and service
Function check
Outofspecication
Code
Text mes­sage
F261 Error in the
conguration
F264
Installation/ Setup error
F265
Measurement function dis­turbed
Cause Rectication
– Error during setup – False signal suppression
faulty
– Error when carrying out a
reset
– Adjustment not within the
vessel height/measuring range
– Max. measuring range of
the instrument not sucient
– Sensor no longer carries
out a measurement
– Operating voltage too low
– Repeat setup – Repeat reset
– Check or correct installation
and/or parameter adjust-
ment
– Use an instrument with big-
ger measuring range
– Check operating voltage – Carry out a reset – Separate operating voltage
briey
The following table shows the error codes and text messages in the status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700 Simulation ac-
tive
Cause Rectication
– A simulation is active – Finish simulation
– Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the status message "Out of specication" and provides information on causes as well as corrective measures.
Code
Text mes­sage
S600
Unpermissi­ble electronics temperature
S601 Overlling
Cause Rectication
– Temperature of the elec-
tronics in the non-specied
section
– Danger of vessel overlling – Make sure that there is no
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
further lling
– Check level in the vessel
Maintenance
The following table shows the error codes and text messages in the status message "Maintenance" and provides information on causes as well as corrective measures.
40795-EN-121011
51
9 Diagnosis, Asset Management and service
Code
Text mes­sage
M500 Error with the
reset delivery
status
M501 Error in the
non-active linearization table
M502 Error in the
diagnosis memory
M503 Reliability too
low
M504
Error on an
device inter­face
M505 No echo avail-
able
Cause Rectication
– With the reset to delivery
status, the data could not be restored
– Hardware error EEPROM – Exchanging the electronics
– Hardware error EEPROM – Exchanging the electronics
– The echot/noise ratio is the
small for a reliable meas-
urement
– Hardware defect – Check connections
– Level echo can no longer
be detected
– Repeat reset – Load XML le with sensor
data into the sensor
– Send instrument for repair
– Send instrument for repair
– Check installation and
process conditions
– Clean the antenna – Change polarisation direc-
tion
– Use instrument with higher
sensitivity
– Exchanging the electronics – Send instrument for repair
– Clean the antenna – Use a more suitable
antenna/sensor
– Remove possible false
echoes
– Optimize sensor position
and orientation
Reaction when malfunc­tions occur
Procedure for fault recti-
cation
Check the 4 … 20 mA signal
52

9.4 Rectify faults

The operator of the system is responsible for taking suitable meas­ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the indicating and
adjustment module Checking the output signal with 4 … 20 mA instruments
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
Connect a handmultimeter in the suitable measuring range according to the wiring plan. The following table describes possible errors in the current signal and helps to remove them:
40795-EN-121011
9 Diagnosis, Asset Management and service
Error Cause Rectication
4 … 20 mA signal not stable
4 … 20 mA signal missing
Current sig­nal greater than 22 mA or less than 3.6 mA
– Level uctua-
tions
– Electrical con-
nection faulty
– Voltage supply
missing
– Operating volt-
age too low or load resistance too high
– Electronics
module in the sensor defec­tive
– Set damping according to the
instrument via the indicating and adjustment module or PACTware/ DTM
– Check connection according to
chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan".
– Check cables for breaks; repair if
necessary
– Check, adapt if necessary
– Exchange the instrument or send it
in for repair
Treatment of measure­ment errors with bulk solids
The below tables show typical examples for application-relevant measurement errors with bulk solids. There are two measurement
errors:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
Level
1
2
0
1 Real level
2 Level displayed by the sensor
time
Instructions:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
40795-EN-121011
53
9 Diagnosis, Asset Management and service

9.4 Measurement error with constant level

Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
2. Measured value jumps towards 100 %
Level
0
Level
– Min./max. adjustment not
correct
– Wrong linearization curve – Adapt linearization curve
time
– Due to the process, the ampli-
tude of the product echo sinks
– Adapt min./max. adjustment
– Carry out false signal suppres-
sion
– A false signal suppression was
0
not carried out
time
– Amplitude or position of a
false echo has changed (e.g. condensation, buildup); false signal suppression no longer
– Determine the reason for the
changed false echo, carry out false signal suppression, e.g. with condensation
matches
9.4 Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3rd measured val­ue jumps towards 0 % during lling
4th measured val­ue uctuates around 10 … 20 %
5. measured value jumps sporadically to 100 % during lling
Level
– Amplitude of a multiple echo
(vessel top - product surface) is larger than the level echo
0
time
– The level echo cannot be distin-
guished from the false echo at a false echo position (jumps to multiple echo)
– Transverse reection from an
extraction funnel, amplitude of the transverse reection larger than the level echo
Level
– Various echoes from an uneven
product surface, e.g. a material
cone
0
time
– Reections from the product
surface via the vessel wall (deection)
Level
0
– Varying condensation or con-
tamination on the antenna
time
– Check parameter "Application",
especially vessel top, product type, dished end, high dielectric gure, adapt if necessary
– Remove/reduce false echo:
minimize interfering installa­tions by changing the polariza­tion direction
– Chose a more suitable installa-
tion position
– Direct sensor to the opposite
funnel wall, avoid crossing with the lling stream
– Check application parameter
"Medium type" and adapt, if necessary
– Optimize installation position
and sensor orientation
– Select a more suitable installa-
tion position, optimize sensor orientation, e.g. with a swivel­ling holder
– Carry out a false signal sup-
pression or increase false signal suppression with con­densation/contamination in the close range by editing
– With bulk solids use radar sen-
sor with purging air connection or exible antenna cover
40795-EN-121011
54
9 Diagnosis, Asset Management and service

9.4 Measurement error during emptying

Fault description Error pattern Cause Rectication
6th measured value re­mains unchanged in the close range during emptying
7th measured value jumps sporadically to­wards 100 % during emptying
8th measured val­ue uctuates around 10 … 20 %
Level
– False echo larger than the level
echo
– Level echo too small
0
time
Level
0
Level
– Varying condensation or con-
tamination on the antenna
time
– Various echoes from an uneven
product surface, e.g. an extrac­tion funnel
0
time
– Reections from the product
surface via the vessel wall (deection)
– Remove false echoes in the
close range. Check: Antenna must protrude out of the socket
– Remove contamination on the
antenna
– Minimize interfering instal-
lations in the close range by changing the polarization direction
– After removing the false ech-
oes, the false signal suppres­sion must be deleted. Carry out a new false signal suppression
– Carry out false signal suppres-
sion or increase false signal suppression in the close range by editing
– With bulk solids use radar sen-
sor with purging air connection or exible antenna cover
– Check application parameter
"Medium type" and adapt, if necessary
– Optimize installation position
and sensor orientation
Reaction after fault recti-
cation
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
24 hour service hotline
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside the normal working hours on seven days a week around the clock.
Since we oer this service worldwide, the support is in the English language. The service itself is free of charge, the only costs involved are the normal call charges.

9.5 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user. In Ex applications only one instrument and one electronics module
with respective Ex approval may be used. If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron­ics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
40795-EN-121011
55
9 Diagnosis, Asset Management and service
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").

9.6 Software update

The following components are required to update the sensor soft­ware:
Sensor
Voltage supply
Interface adapter VEGACONNECT 4
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".
56

9.7 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to proceed under www.vega.com/downloads and "Forms and certi- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
40795-EN-121011

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.
Materials: see chapter "Technical data" For disposal or recycling within the European Union proceed accord-
ing to the "Electronics recycling" and "Battery/Accumulator recycling" below. Outside the European Union you should take note of the valid national regulations.
Electronics recycling
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Therefore pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Battery/accumulator recycling
Note:
The disposal is subject to the directive 2006/66/EG on batteries and accumulators.
Batteries and accumulators contain some environmentally harm­ful but also some valuable raw materials that can be recycled. For that reason batteries and accumulators must not be disposed of in household waste.
All users are legally obligated to bring batteries to a suitable collection point, e.g. public collection points. You can also return the batteries and accumulators to us for correct disposal. Due to the very strict transport regulations for lithium-based batteries/accumulators, this is normally not a good idea because shipment is very expensive.
Proceed as follows to dismount the accumulator:
Unscrew the cover of the supply room
Loosen the plug connector
Loosen the xing screws
Pull out the complete insert by means of the plastic strap
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
40795-EN-121011
57

11 Supplement

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435 Materials, wetted parts
Ʋ Adapter ange PP Ʋ Seal, adapter ange FKM (COG VI500), EPDM (COG AP310) Ʋ Antenna PBT-GF 30 Ʋ Focussing lense PP
Materials, non-wetted parts
Ʋ Compression ange PP Ʋ Mounting strap 316L Ʋ Fixing screws, mounting strap 316L Ʋ Fixing screws, adapter ange 304 Ʋ Plastic housing plastic PBT (Polyester) Ʋ Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated - basis:
Ʋ Stainless steel housing 316L Ʋ Seal between housing and housing
cover
Ʋ Inspection window in housing cover
(optional)
Ʋ Ground terminal 316L
Process ttings
Ʋ Flanges DIN from DN 80, ANSI from 3", JIS from DN 100 10K
Weight depending on process tting and housing material
Max. torque, mounting screws - strap on the sensor housing
Max. torque ange screws
Ʋ Compression ange DN 80 5 Nm (3.689 lbf ft) Ʋ Adapter ange DN 100 7 Nm (5.163 lbf ft)
Max. torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing 10 Nm (7.376 lbf ft) Ʋ Aluminium/Stainless steel housing 50 Nm (36.88 lbf ft)
Input variable
Measured variable The measured variable is the distance between the pro-
Polyester
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housng, electropolished)
Polycarbonate
0.7 … 3.4 kg (1.543 … 7.496 lbs)
4 Nm
cess tting of the sensor and the product surface. The reference plane is the seal surface of the ange.
40795-EN-121011
58
11 Supplement
Fig. 47: Data of the input variable
1 Reference plane 2 Measured variable, max. measuring range
3 Antenna length
4 Useful measuring range
3
2
4
1
Max. measuring range 15 m (49.21 ft) Recommended measuring range up to 15 m (49.21 ft)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Installation reference conditions
Ʋ Min. distance to installations > 200 mm (7.874 in) Ʋ Reector Plane plate reector Ʋ False reections Largest false echo 20 dB smaller than the useful echo
Deviation with liquids See following diagrams
40795-EN-121011
59
11 Supplement
10 mm (0.394 in)
2 mm (0.079 in)
- 2 mm (- 0.079 in)
- 10 mm (- 0.394 in)
Fig. 48: Deviation under reference conditions
1 Reference plane 2 Antenna edge 3 Recommended measuring range
0
1,0 m
(3.280 ft)
1
2 3
Reproducibility ≤ ±1 mm Deviation with bulk solids The values depend to a great extent on the application.
Binding specications are thus not possible.
Deviation under EMC inuence ≤ ±30 mm
Variablesinuencingmeasurementaccuracy
SpecicationsapplytotheHARTsignalandthecurrentoutput
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
max. 10 mm
Additional deviation through strong, high
< ±50 mm frequency electromagnetic elds acc. to EN 61326
Specicationsapplyalsotothecurrentoutput
Temperature drift - Current output ±0.03 %/10 K relating to the 16 mA span max. ±0.3 % Deviation on the current output by ana-
< ±15 µA logue/digital conversion
Deviation on the current output due to
< ±150 µA strong, high frequency electromagnetic elds acc. to EN 61326
Characteristics and performance data
Measuring frequency K-band (26 GHz technology) Measuring cycle time approx. 700 ms Step response time Tracking speed of the measuring window
1)
≤ 3 s
1 m/min max.
Beam angle Emitted HF power (depending on the parameter adjustment)
1)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
2)
Outside the specied beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)
3)
EIRP: Equivalent Isotropic Radiated Power
60
2)
10°
3)
40795-EN-121011
11 Supplement
Ʋ Average spectral transmission power
-14 dBm/MHz EIRP
density
Ʋ Max. spectral transmission power
+43 dBm/50 MHz EIRP
density
Ʋ Max. power density in a distance of
< 1 µW/cm²
1 m
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
Process temperature (measured on the
-40 … +80 °C (-40 … +176 °F)
process tting) Vessel pressure -1 … 2 bar (-100 … 200 kPa/-14.5 … 29.0 psig) Vessel pressure - version with adapter
-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)
ange from DN 150 PP Vibration resistance
Ʋ Plastic horn antenna with compres-
sion or adapter ange
Ʋ Plastic horn antenna with mounting
strap
4)
mechanical vibrations up to 2 g in the frequency range
5 … 200 Hz
mechanical vibrations up to 1 g in the frequency range
5 … 200 Hz
Indicating and adjustment module
Voltage supply and data transmission through the sensor Indication LC display in dot matrix Measured value indication
Ʋ Number of digits 5 Ʋ Size of digits W x H = 7 x 13 mm
Adjustment elements 4 keys Protection rating
Ʋ unassembled IP 20 Ʋ mounted into the sensor without cover IP 40
Materials
Ʋ Housing ABS Ʋ Inspection window Polyester foil
Indicating and adjustment elements
Indicating elements
Ʋ Green LED in the supply room Indication of the charging process Ʋ Yellow LED in the supply room Indication of the charging condition
Adjustment elements
Ʋ Rotary switch in the supply room Selection of the mode Ʋ Button outside on the housing Switching on and o
4)
Tested according to the guidelines of German Lloyd, GL directive 2.
40795-EN-121011
61
11 Supplement
Integrated clock
Date format Day.Month.Year Time format 12 h/24 h Time zone Ex factory CET
Measurement electronics temerature
Resolution 1 °C (1.8 °F) Accuracy ±1 °C (1.8 °F)
External battery charger
Mains voltage 100 … 240 V AC Output voltage 24 V DC Output current (shortcircuit proof) max. 500 mA Load current limitation 70 mA DC plug (inside plus, outside minus) 2.1 mm
Integrated accumulator
Type Lithium ions Voltage 14.8 V Accumulator capacity 4.7 Wh Charging period from 0 % to 100 %
4 h approx.
Operating time after 10 min. charging of
> 3 h 0 %
Operating time in mode 4 (sensor per-
> 60 h manently on) with completely charged accumulator
Temperature range
Ʋ Charge accumulator 0 … +45° C (+32 … +167 °F) Ʋ Accumulator operation -20 … +60° C (-4 … +140 °F)
Temperature derating accumulator capacity
Ʋ +25° C (+77 °F) 100 % Ʋ -10° C (+14 °F) 50 %
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67 Ʋ Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar)
5)
housing - investment casting, stain-
less steel housing - electro-polished Overvoltage category III Protection class III
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
62
40795-EN-121011
11 Supplement
Approvals
Instruments with approvals can have dierent technical data depending on the version. For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".

11.2 Dimensions

The following dimensional drawings represent only an extract of all possible versions. Detailed dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
Housing
~ 87 mm (3.43")
M16x1,5
ø 84 mm
(3.31")
120 mm (4.72")
M20x1,5/ ½ NPT
Fig. 49: Dimensions housing - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1
40795-EN-121011
63
11 Supplement
VEGAPULS 67, version with mounting strap
1.4301
PBT-GF30
2,5 mm
(0.10")
8,5 mm
(0.34")
125 mm
(4.92")
(0.75")
19 mm
(6.69")
170 mm
(11.81")
300 mm
(3.86")
98 mm
PP
9 mm
(0.47")
12 mm
9 mm
(0.35")
(0.35")
12 mm (0.47")
85 mm
Fig. 50: VEGAPULS 67, with mounting strap
64
(3.35")
75 mm (2.95")
107 mm
(4.21")
115 mm
(4.53")
(0.59")
15 mm
40795-EN-121011
VEGAPULS67,versionwithcompressionange
ø 107 mm
(0.41")
10,5 mm
(4.21")
ø 21 mm
(0.83")
(4.96")
126 mm
11 Supplement
(0.75")
19 mm
ø 75 mm (2.95")
ø 115 mm (4.53")
ø 156 mm (6.14")
ø 200 mm (7.87")
Fig. 51: VEGAPULS 67, compression ange suitable for DN 80 PN 16/ASME 3" 150lbs/JIS80 10K
VEGAPULS67,versionwithadapterange
1
138 mm (5.43")
(1.22")
31 mm
(0.79")
20 mm
ø 75 mm (2.95")
ø 98 mm (3.86")
8 mm
(0.32")
2
Fig. 52: VEGAPULS 67, adapter ange
1 Adapter ange
2 Seal
40795-EN-121011
65
11 Supplement

11.3 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

11.4 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/ originator.
66
40795-EN-121011
INDEX
INDEX
A
Accessories
– Indicating and adjustment module 12 – Interface adapter 12 – Protective cover 12
Adjustment
– Max. adjustment 37 – Min. adjustment 36
Adjustment system 29 Agitator 21 Applications
– Bulk solid 33 – Liquid 30 – Standpipe 31
C
Check signal 52 Curve indication
– Echo curve 38 – False signal suppression 38
D
DD (Device Description) 47 Default values 41
E
Echo curve memory 48 Echo curve of the setup 38 EDD (Enhanced Device Description) 47 Electronics compartment 24 Error messages 49 Event memory 48
F
False signal suppression 39 Fault rectication 52 Functional principle 11
G
Grounding 24
H
HART resistor 44
K
Key function 28
L
Linearisation curve 40
M
Main menu 29 Meas. reliability 37 Measured value memory 48 Measurement error 53 Medium
– Liquid Bulk solid 30
Moisture 13
N
NAMUR NE 107
– Failure 50 – Function check 51 – Maintenance 51 – Out of specication 51
O
Overll protection according to WHG 40
P
Packaging 11 Peak value 37 Potential equalisation 24
R
Reection properties 30 Repair 56 Replacement parts
– Accumulator insert 12 – Electronics module 12
Reset 41
S
Service hotline 55 Shielding 24 Socket 17 Status messages 49 Storage 11 Subsurface enclosure 22
T
Type label 10
V
Vessel form 35 Vessel height 35 Vessel installations 21 Voltage supply 11
40795-EN-121011
67
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2012
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
40795-EN-121011
Loading...