VEGA PULS66 User Manual

Operating Instructions
Radar sensor for continuous level measurement of liquids
VEGAPULS 66 standpipe ver­sion
Foundation Fieldbus
Document ID: 36526
Mounting
Connect electrically
Quick start
The quick start enables a quick setup with many applications. You can nd further information in the respective chapters of the operating instructions manual.
1. In case of turbulences or strong product movement in the vessel, long standpipe antennas should be fastened to the vessel wall.
2. The measurement is only possible inside the tube, the standpipe antenna must hence reach up to the requested min. level
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica­tions on the type label.
2. Connect the instrument according to the following illustration
2
0
1
1
0
Bus
( )
(-)
1
+
2
5
678
1
Fig. 1: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("1" = mode for simulation release) 4 For external indicating and adjustment unit 5 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
3
4
5
Set parameters
Further steps
2
The instrument is preset by default with the suitable parameters for "Application", "Min. adjustment" and "Max. adjustment".
For adaptation of the parameter "Medium" you have to proceed as follows:
1. Go via the indicating and adjustment module to the menu "Setup".
2. In the menu item "Medium"you select the medium of your applica­tion, for example "Solvent".
36526-EN-121011
1. In the menu "Additional settings", menu item "Damping" you have to adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out- put".
VEGAPULS 66 standpipe version • Foundation Fieldbus
The quick start is then nished. For further information see chapter "Parameter adjustment".
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
3
Contents
Contents
1 About this document
1.1 Function ............................................................................. 6
1.2 Target group ....................................................................... 6
1.3 Symbolism used ................................................................. 6
2 For your safety
2.1 Authorised personnel ......................................................... 7
2.2 Appropriate use .................................................................. 7
2.3 Warning about incorrect use ............................................... 7
2.4 General safety instructions ................................................. 7
2.5 CE conformity ..................................................................... 8
2.6 NAMUR recommendations ................................................ 8
2.7 Radio license for Europe .................................................... 8
2.8 Radio license for USA/Canada ........................................... 8
2.9 Environmental instructions ................................................. 9
3 Product description
3.1 Conguration .................................................................... 10
3.2 Principle of operation ........................................................ 11
3.3 Packaging, transport and storage ..................................... 11
3.4 Accessories and replacement parts ................................. 12
4 Mounting
4.1 General instructions ......................................................... 14
4.2 Instructions for installation ................................................ 14
5 Connecting to the bus system
5.1 Preparing the connection ................................................. 15
5.2 Connecting ....................................................................... 16
5.3 Wiring plan, single chamber housing ................................ 17
5.4 Wiring plan, double chamber housing .............................. 17
5.5 Wiring plan, double chamber housing Ex d ...................... 18
5.6 Wiring plan - version IP 66/IP 68, 1 bar ............................. 19
5.7 Switch-on phase ............................................................... 19
6 Set up with the indicating and adjustment module
6.1 Insert indicating and adjustment module .......................... 21
6.2 Adjustment system ........................................................... 22
6.3 Parameter adjustment ...................................................... 23
6.4 Saving the parameter adjustment data ............................. 31
7 Setup with PACTware
7.1 Connect the PC ................................................................ 32
7.2 Parameter adjustment with PACTware .............................. 32
7.3 Saving the parameter adjustment data ............................. 33
8 Set up with other systems
8.1 DD adjustment programs ................................................. 34
8.2 Communicator 375, 475 ................................................... 34
9 Diagnosis, Asset Management and service
9.1 Maintenance .................................................................... 35
9.2 Measured value and event memory ................................. 35
36526-EN-121011
4
VEGAPULS 66 standpipe version • Foundation Fieldbus
9.3 Asset Management function ............................................. 36
9.4 Rectify faults ..................................................................... 39
9.5 Exchanging the electronics module .................................. 41
9.6 Software update ............................................................... 42
9.7 How to proceed in case of repair ...................................... 42
10 Dismounting
10.1 Dismounting steps ............................................................ 44
10.2 Disposal ........................................................................... 44
11 Supplement
11.1 Technical data .................................................................. 45
11.2 Supplementary information Foundation Fieldbus ............. 51
11.3 Dimensions ...................................................................... 59
Contents
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
Editing status: 2012-09-27
5

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
36526-EN-121011
6
VEGAPULS 66 standpipe version • Foundation Fieldbus

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGAPULS 66 is a sensor for continuous level measurement. You can nd detailed information on the application range in chapter
"Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula­tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.
Depending on the instrument version, the emitting frequencies are in the C or K band range. The low emitting frequencies are far below the internationally approved limit values. When used correctly, there is no danger to health.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
7
2 For your safety

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the conformity certicate in the download section of our homepage.
Only with Ex-d-ia version
The instrument is designed for use in an industrial environment. Nevertheless, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a dierent environment, its electromagnetic compatibility with other devices must be ensured by suitable measures.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom-
mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring
transducers
NE 53 – Compatibility of eld devices and indicating/adjustment
components
NE 107 - Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Radio license for Europe

The instrument is approved according to EN 302372-1/2 (2006-04) for use in closed vessels.

2.8 Radio license for USA/Canada

The instrument is in conformity with part 15 of the FCC regulations. Take note of the following two regulations:
The instrument must not cause any interfering emissions
The device must be insensitive to interfering immissions, including
those that may cause undesirable operating conditions
Modications not expressly approved by the manufacturer will lead to expiry of the operating licence according to FCC/IC.
The instrument is in conformity with RSS-210 of the IC regulations. The instrument may only be used in closed vessels made of metal,
concrete, or bre-reinforced plastic.
8
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
2 For your safety

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
9

3 Product description

Type label
3 Product description
3.1 Conguration
The type label contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 2: Layout of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Measuring range 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Note to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directive
16
15
14
13
12
11
Serial number
Scope of this operating instructions manual
10
With the serial number of the instrument on the type label you have access to the following data on our homepage:
Article number of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate "Measuring Accuracy" (PDF)
For this purpose, move to www.vega.com and "VEGA Tools".
This operating instructions manual applies to the following instrument
versions:
Hardware from 2.1.1
Software from 4.5.0
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
3 Product description
Versions
Scope of delivery
Application area
The instrument is available in two dierent electronics versions. Each version can be identied via the product code on the type label as well as on the electronics.
Standard electronics type PS60FFC.-
Electronics with increased sensitivity type PS60FFS.-
The scope of delivery encompasses:
Radar sensor
Documentation
this operating instructions manualTest certicate Measurement accuracy, depending on the
instrument VEGAPULS 66 (optional)
Operating instructions manual "Indicating and adjustment mod-
ule PLICSCOM" (optional)
– Supplementary instructions "GSM/GPRS radio module"
(optional)
Supplementary instructions manual "Heating for indicating and
adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)if necessary, further certicates

3.2 Principle of operation

The instrument is suitable for the measurement of liquids under dif­cult and extreme process conditions. Application possibilities can be found in the chemical industry, in environmental and recycling technology as well as in the petrochemical industry.
The version with standpipe antenna is particularly suitable for measurement of solvents and liquid gases, for vessels with foam generation or for measurement of products with low dielectric values (DK > 1.6).
Measurement in a standpipe is not recommended for extremely adhesive products.
The instrument can be used with products with an ε actually achievable value depends on the measuring conditions, the antenna system, the standpipe or bypass.
value ≥1.8. The
r
Functional principle
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The transit time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropri­ate output signal and outputted as measured value.

3.3 Packaging, transport and storage

Packaging
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
11
3 Product description
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport
Transport inspection
Storage
Storage and transport temperature
Indicating and adjust­ment module
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %

3.4 Accessories and replacement parts

The indicating and adjustment module PLICSCOM is used for meas­ured value indication, adjustment and diagnosis. It can be inserted into the sensor and removed at any time.
You can nd further information in the operating instructions "Indicat- ing and adjustment module PLICSCOM" (Document-ID 27835).
Interface adapter
External indicating and adjustment unit
12
The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, an adjustment software such as PACTware with VEGA-DTM is required.
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 61 is an external indicating and adjustment unit for sensors with single chamber housing and Ex-d double chamber housing.
It is suitable for measured value indication and adjustment of plics sensors and is connected to the sensor with a four-wire standard cable up to 50 m long.
You can nd further information in the operating instructions "VE- GADIS 61" (Document-ID 27720).
VEGAPULS 66 standpipe version • Foundation Fieldbus
®
36526-EN-121011
3 Product description
External radio unit
Flanges
Protective cover
Electronics module
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for transmission of measured values and for remote parameter adjust­ment of plics® sensors. The adjustment is carried out via PACTware/ DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions "PLICSMOBILE T61" (Document-ID 36849).
Screwed anges are available in dierent versions according to the following standards: DIN 2501, EN 1092-1, ANSI B 16.5, JIS B 2210­1984, GOST 12821-80.
You can nd additional information in the supplementary instructions manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
The protective cover protects the sensor housing against soiling and intense heat from solar radiation.
You will nd additional information in the supplementary instructions manual "Protective cover" (Document-ID 34296).
The electronics module VEGAPULS series 60 is a replacement part for radar sensors of VEGAPULS series 60. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions "Elec- tronics module VEGAPULS series 60" (Document-ID 36801).
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
13

4 Mounting

Screwing in
4 Mounting

4.1 General instructions

With instruments with threaded process tting, suitable tools must be applied for tightening the hexagon.
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
Moisture
Suitability for the process conditions
Installation
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain o. This ap­plies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
Make sure that all parts of the instrument exposed to the process, in particular the active measuring component, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the type label.

4.2 Instructions for installation

By using the standpipe version, the inuence of turbulence and ves­sel installations, such as e.g. heating spirals or agitators, is excluded. If turbulence or vigorous product movement occurs in the vessel, long standpipe antennas should be fastened to the vessel wall.
The standpipe antenna must extend all the way down to the request­ed min. level, as measurement is only possible within the tube. If a good mixing of the product is important, you should use a version with perforated surge pipe.
14
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
Safety instructions

5 Connecting to the bus system

5 Connecting to the bus system

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
Voltage supply
Connection cable
Cable gland ½ NPT
Cable screening and grounding
The instrument requires a operating voltage of 9 … 32 V DC. Operat­ing voltage and the digital bus signal are carried on the same two-wire connection cable. Power is supplied via the H1 power supply.
Connection is carried out with screened cable according to Fieldbus specication.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland. If you are using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Make sure that the entire installation is carried out according to the Fieldbus specication. In particular, make sure that the bus is termi­nated with suitable terminating resistors.
With plastic housing, the NPT cable gland or the Conduit steel tube must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data"
Make sure that the cable screening and ground is executed accord­ing to the Fielbus specication. If electromagnetic interference is expected which is above the test values of EN 61326-1 for industrial areas, we recommend to connect the cable screen on both ends to ground potential.
With systems with potential equalisation, connect the cable screen directly to ground potential at the power supply unit, in the connection box and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation (low impedance).
In systems without potential equalisation with cable screening on both sides, connect the cable screen directly to ground potential at the power supply unit and at the sensor. In the connection box or T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). Low-frequency potential equali­sation currents are thus suppressed, but the protective eect against high frequency interference signals remains.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
15
5 Connecting to the bus system
Connection technology

5.2 Connecting

The voltage supply and signal output are connected via the spring­loaded terminals in the housing.
The connection to the indicating and adjustment module or to the interface adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the elec­tronics. To do this, lift the terminal block with a small screwdriver and pull it out. When inserting the terminal block again, you should hear it snap in.
Connection procedure
Proceed as follows:
1. Unscrew the housing cover
2. If an indicating and adjustment module is installed, remove it by turning it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Fig. 3: Connection steps 5 and 6
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with cable end sleeves are inserted directly into the terminal openings. In case of exible cores without end sleeves, press the terminal head with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the terminal closes again.
You can nd further information to the max. wire cross-section under "Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
36526-EN-121011
16
VEGAPULS 66 standpipe version • Foundation Fieldbus
5 Connecting to the bus system
8. Connect the screen to the internal ground terminal, connect the outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable
10. Place probably existing indicating and adjustment module back
on
11. Screw the housing cover back on
The electrical connection is hence nished.

5.3 Wiring plan, single chamber housing

Electronics and connec­tion compartment
Electronics compartment
2
3
0
1
1
0
Bus
( )
(-)
1
+
2
5
678
4
5
1
Fig. 4: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("1" = mode for simulation release) 4 For external indicating and adjustment unit 5 Ground terminal for connection of the cable screen

5.4 Wiring plan, double chamber housing

2
3
0
1
1
0
Bus
( )
(-)
+
1
2
5
678
11
Fig. 5: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("on" = simulation mode)
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
17
5 Connecting to the bus system
Connection compartment
Radio module PLICSMO­BILE integrated in the connection compartment
2
Bus
(-)
( )
+
1
2
3
1
Fig. 6: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For indicating and adjustment module or interface adapter 3 Ground terminal for connection of the cable screen
Status
SIM-Card
Test
Bus
( )
+
1
2
USB
(-)
1
Fig. 7: CConnection of the voltage supply of the radio module
1 Voltage supply
You can nd detailed information for connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".
Electronics compartment
18

5.5 Wiring plan, double chamber housing Ex d

2
3
0
1
1
0
Bus
( )
(-)
+
1
2
Fig. 8: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the indicating and adjustment module or interface adapter 3 Simulation switch ("on" = simulation mode)
VEGAPULS 66 standpipe version • Foundation Fieldbus
5
678
11
36526-EN-121011
Connection compartment
5 Connecting to the bus system
Bus
( )
(-)
+
1
2
2
1
Fig. 9: Connection compartment, double chamber housing Ex d
1 Voltage supply, signal output 2 Ground terminal for connection of the cable screen
Plug M12 x 1 for external indicating and adjust­ment unit
Wire assignment, con­nection cable
34
1
Fig. 10: Top view of the plug connector
1 Pin 1 2 Pin 2 3 Pin 3 4 Pin 4
Contact pin Colour connection ca-
ble in the sensor
Pin 1 Brown 5
Pin 2 White 6
Pin 3 Blue 7
Pin 4 Black 8
2
Terminal, electronics module

5.6 Wiring plan - version IP 66/IP 68, 1 bar

1
2
Fig.11:Wireassignmentx-connectedconnectioncable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding

5.7 Switch-on phase

After VEGAPULS 66 is connected to the bus system, the instrument carries out a self-test for approx. 30 seconds. The following steps are
carried out:
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
19
5 Connecting to the bus system
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC Indication of the status message "F 105 Determine measured
value" on the display or PC Status byte goes briey to fault value
As soon as a plausible measured value is found, it is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory settings.
20
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus

6 Set up with the indicating and adjustment module

6 Set up with the indicating and adjustment
module

6.1 Insert indicating and adjustment module

The indicating and adjustment module can be inserted into the sen­sor and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the indicating and adjustment module in the requested position onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order. The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Fig. 12: Insertion of the indicating and adjustment module with single chamber housing into the electronics compartment
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
21
6 Set up with the indicating and adjustment module
1 2
Fig. 13: Insertion of the indicating and adjustment module with double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with an indicating and adjust­ment module for continuous measured value indication, a higher cover with an inspection glass is required.
Key functions
22

6.2 Adjustment system

Fig. 14: Indicating and adjustment elements
1 LC display 2 Adjustment keys
[OK] key:
Move to the menu overviewConrm selected menu
VEGAPULS 66 standpipe version • Foundation Fieldbus
1
2
36526-EN-121011
6 Set up with the indicating and adjustment module
– Edit parameter – Save value
[->] key:
Presentation change measured valueSelect list entrySelect editing position
[+] key:
– Change value of the parameter
[ESC] key:
– interrupt input – Jump to next higher menu
Adjustment system
Main menu
The sensor is adjusted via the four keys of the indicating and adjust­ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.

6.3 Parameter adjustment

Through the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g., for measurement loop name, medium, applica­tion, vessel, adjustment, signal output
Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g. on the instrument status, pointer, reli-
ability, simulation, echo curve Further settings: Instrument unit, false signal suppression, linearisa-
tion curve, reset, date/time, reset, copy function Info: Instrument name, hardware and software version, calibration
date, instrument features
Information:
In this operating instructions manual, the instrument-specic param­eters in the menu sections "Setup", "Diagnosis" and "Additional set- tings" are described. The general parameters in these menu section are described in the operating instructions manual "Indicating and adjustment module".
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
23
6 Set up with the indicating and adjustment module
You can nd in the operating instructions manual "Indicating and ad- justment module" also the description of the menu sections "Display" and "Info".
In the main menu point "Setup", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the measurement. The proce­dure is described in the following.
Setup
Setup/Medium
Each medium has dierent reection properties. With liquids, further interfering factors are uctuation product surface and foam genera­tion. With bulk solids, these are dust generation, material cone and additional echoes from the vessel wall.
To adapt the sensor to these dierent measuring conditions, the selection "Liquid" or "Bulk solid" should be made in this menu item.
Through this selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with poor reec­tive properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup/Adjustment
24
Since a radar sensor is a distance measuring instrument, the distance from the sensor to the product surface is measured. For indication of the real level, an allocation of the measured distance to the percent­age height must be carried out.
To perform the adjustment, enter the distance with full and empty ves­sel. For instruments with standpipe antenna, these values are already preset by default, see the following example:
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
6 Set up with the indicating and adjustment module
")
0,5 m
(19.68
100%
2
")
5 m
(196.9
0%
1
Fig. 15: Parameter adjustment example min./max. adjustment
1 Min. level = max. meas. distance 2 Max. level = min. meas. distance
If these values are not known, an adjustment with the distances of for example 10 % and 90 % is possible. Starting point for these distance specications is always the seal surface of the thread or ange. By means of these settings, the real level will be calculated.
The real product level during this adjustment is not important, be­cause the min./max. adjustment is always carried out without chang­ing the product level. These settings can be made ahead of time without the instrument having to be installed.
Setup/Min. adjustment
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
25
6 Set up with the indicating and adjustment module
4. Enter the suitable distance value in m for the empty vessel (e.g. distance from the sensor to the vessel bottom) corresponding to the percentage value.
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
Setup/Max. adjustment
Diagnosis/Peak value
Proceed as follows:
1. Select with [->] the menu item max. adjustment and conrm with [OK].
2. Prepare the percentage value for editing with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4. Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. Keep in mind that the max. level must lie below the min. distance to the antenna edge.
5. Save settings with [OK]
The respective min. and max. measured value is saved in the sensor. The values are displayed in the menu item "Peak values".
Diagnosis/Measurement reliability
26
When non-contact level sensors are used, the measurement can be inuenced by the respective process conditions. In this menu item, the measurement reliability of the level echo is displayed as dB value. The measurement reliability equals signal strength minus noise. The higher the value, the more reliable the measurement. With a function­ing measurement, the values are > 10 dB.
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
6 Set up with the indicating and adjustment module
Diagnoses/Curve indica­tion
The "Echo curve" shows the signal strength of the echoes over the measuring range in dB. The signal strength enables an evaluation of the quality of the measurement.
The "False signal suppression" displays the saved false echoes (see menu "Additional settings") of the empty vessel with signal strength in "dB" over the measuring range.
A comparison of echo curve and false signal suppression allows a more accurate conclusion on measurement reliability. The selected curve is continuously updated. With the [OK] key, a submenu with zoom functions is opened:
"X-Zoom": Zoom function for the meas. distance
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "dB"
"Unzoom": Reset the presentation to the nominal measuring range
with single magnication
Diagnostics/Echo curve memory
Additional adjustments/ False signal suppression
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
With the function "Echo curve memory" the echo curve can be saved at the time of setup. This is generally recommended; for using the As­set Management functions it is absolutely necessary. If possible, the curve should be saved with a low level in the vessel.
With the adjustment software PACTware and the PC, the high resolu­tion echo curve can be displayed and used to recognize signal changes over the operating time. In addition, the echo curve of the setup can be also displayed in the echo curve window and compared with the actual echo curve.
The following circumstances cause interfering reections and can inuence the measurement:
High sockets
27
6 Set up with the indicating and adjustment module
Vessel installations such as struts
Agitators
Buildup or welded joints on vessel walls
Note:
A false signal suppression detects, marks and saves these false sig­nals so that they are no longer taken into account for level measure­ment.
This should be done with the low level so that all potential interfering reections can be detected.
Proceed as follows:
1. Select the menu item "Additional settins" with [->] and conrm with [OK]. With [->] you have to select the menu item "False signal suppression" and conrm with [OK].
2. Conrm again with [OK].
3. Conrm again with [OK].
28
4. Conrm again with [OK] and enter the actual distance from the sensor to the product surface.
5. All interfering signals in this section are detected by the sensor and stored after conrming with [OK].
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as false signal. The lling level would then no longer be detectable in this area.
If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression":
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
6 Set up with the indicating and adjustment module
The menu item "Delete" is used to completely delete an already cre­ated false signal suppression. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel.
The menu item "Extend" is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with a too high level and not all false signals could be de­tected. When selecting "Extend", the distance to the product surface of the created false signal suppression is displayed. This value can now be changed and the false signal suppression can be extended to this range.
Additional adjustments/ Linearization curve
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindri­cal or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [ESC] and [->] key.
Caution:
Note the following if the instrument with corresponding approval is used as part of an overll protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer linearly proportional to the level. This must be taken into consideration by the user, particularly when setting the switching point on the level switch.
Additional adjustments/ Probe length
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
With radar sensors with standpipe antenna, the sensor length is already preset in this menu item. When shortening the standpipe antenna afterwards, this value must be appropriately corrected.
29
6 Set up with the indicating and adjustment module
Additional adjustments/ Reset
When a reset is carried out, all settings (with only a few exceptions) are reset. The exceptions are: PIN, language and lighting.
The following reset functions are available: Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specic settings. A created false signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special and laboratory parameters to the default values of the respective instru­ment. A created false signal suppression, user programmable lineari­zation curve as well as the measured value memory will be deleted.
Setup: Resetting of the parameter settings to the default values of the respective instrument in the menu item Setup. Order-related settings remain but are not taken over into the current parameters. User­generated false signal suppression, user-programmed linearization curve, measured value memory as well as event memory remain untouched. The linearization is set to linear.
False signal suppression: Deleting a previously created false signal suppression. The false signal suppression created in the factory remains active.
Peak values distance: Resetting the measured min. and max. dis­tances to the actual measured value.
Select the requested reset function [->] and conrm with [OK]. The following table shows the default values of VEGAPULS 66:
Menu section Menu item Default value
Setup Measurement
Display Language Like order
Additional adjust-
ments
loop name
Min. adjustment Length of the standpipe Ex factory
Damping 0.0 s
Displayed value Distance
Display unit m(d)
Scaling 0.00 %, 0 l
Distance unit m
Temperature unit °C
Unit SV2 m
Probe length Length of the standpipe Ex factory
Linearisation curve
Sensor
100.00 %, 100 l
Linear
36526-EN-121011
30
VEGAPULS 66 standpipe version • Foundation Fieldbus
6 Set up with the indicating and adjustment module

6.4 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If the instrument is equipped with an indicating and adjustment module, the data in the sensor can be saved in the indicating and adjustment module. The procedure is described in the operating instructions manual "Indicating and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
The following data or settings for adjustment of the indicating and adjustment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Sensor-specic
units, temperature unit and linearization" The values of the user programmable linearisation curve
The function can also be used to transfer settings from one instru­ment to another instrument of the same type. If it is necessary to exchange a sensor, the indicating and adjustment module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy sensor data".
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
31

7 Setup with PACTware

7 Setup with PACTware

7.1 Connect the PC

Via the interface adapter directly on the sensor
Prerequisites
2
1
3
Fig. 16: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 4 3 Sensor
Information:
The interface adapter VEGACONNECT 3 is not suitable for connec­tion to the sensor.

7.2 Parameter adjustment with PACTware

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
36526-EN-121011
32
VEGAPULS 66 standpipe version • Foundation Fieldbus
Fig. 17: Example of a DTM view
7 Setup with PACTware
Standard/Full version
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega. com/downloads and "Software". The full version is available on CD from the agency serving you.

7.3 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
33

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".

8.2 Communicator 375, 475

Device descriptions for the instrument are available as DD or EDD for parameter adjustment with the Field Communicator 375 or 475.
The les can be downloaded at www.vega.com/downloads under "Software".
34
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus

9 Diagnosis, Asset Management and service

9 Diagnosis, Asset Management and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal operation.

9.2 Measured value and event memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
Event memory
Echo curve memory
Up to 60,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switching on and o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control system with EDD.
The echo curves are stored with date and time and the corresponding echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup. Changes in the measure­ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via:
PC with PACTware/DTM
Control system with EDD
Indicating and adjustment module
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
35
9 Diagnosis, Asset Management and service
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD

9.3 Asset Management function

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the indicating and adjustment module, PACTware/DTM and EDD.
Status messages
The status messages are classied in the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 18: Pictograms of the status messages
1 Failure - red 2 Function check - orange
3 Outofspecication-yellow
4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is outputted.
This status message is always active. It cannot be deactivated by the user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
36526-EN-121011
36
VEGAPULS 66 standpipe version • Foundation Fieldbus
9 Diagnosis, Asset Management and service
Failure (failure)
The following table shows the codes and text messages of the status message "Failure" and provides information on causes as well as corrective measures.
Code
Text mes­sage
F013
no meas-
ured value available
F017 Adjustment
span too
small
F025
Error in the
lineariza­tion table
F036
No op-
erable software
F040
Error in the
electronics
F080 – General software error – Separate operating
F105
Determine measured
value
F113 Communi-
cation error
F125
Unper-
missible electronics
tempera­ture
Cause Rectication PA DevS-
– Sensor does not
detect an echo during
operation
Antenna system con-
taminated or defective
– Adjustment not within
specication
– Index markers are not
continuously rising, for examle unlogical value
pairs
Failed or interrupted
software update
Hardware defect Exchanging the elec-
The instrument is still
in the start phase, the measured value could not yet be determined
Error in the internal
instrument communi­cation
Temperature of the
electronics in the non­specied section
– Check or correct
installation and/or parameter adjustment
Clean or exchange
process component or antenna
Change adjustment
according to the limit values (dierence between min. and max. ≥ 10 mm)
– Check linearization
table
– Delete table/Create
new
Repeat software
update
Check electronics
version
Exchanging the elec-
tronics
Send instrument for
repair
tronics
Send instrument for
repair
voltage briey
Wait for the warm-up
phase
Duration depending
on the version and
parameter adjustment up to approximately 3 min.
– Separate operating
voltage briey
Send instrument for
repair
Check ambient tem-
perature
Isolate electronicsUse instrument with
higher temperature range
pec
Diagnosis
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 7
Bit 8
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
37
9 Diagnosis, Asset Management and service
Function check
Code
Text mes­sage
F260
Error in the
calibration
F261
Error in the congura­tion
F264 Installa-
tion/Setup
error
F265 Meas-
urement
function disturbed
Cause Rectication PA DevS-
Error in the calibra-
tion carried out in the
factory
Error in the EEPROM
Error during setupFalse signal suppres-
sion faulty
Error when carrying
out a reset
Adjustment not within
the vessel height/ measuring range
– Max. measuring range
of the instrument not sucient
Sensor no longer
carries out a measure­ment
Operating voltage
too low
Exchanging the elec-
tronics
Send instrument for
repair
Repeat setupRepeat reset
Check or correct
installation and/or parameter adjustment
– Use an instrument
with bigger measuring range
– Check operating
voltage
Carry out a resetSeparate operating
voltage briey
pec
Diagnosis
Bit 9
Bit 10
Bit 11
Bit 12
The following table shows the error codes and text messages in the status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700 Simulation ac-
tive
Cause Rectication
A simulation is active Finish simulation
Wait for the automatic end
after 60 mins.
Outofspecication
38
The following table shows the error codes and text messages in the status message "Outofspecication" and provides information on causes as well as corrective measures.
Code
Text mes­sage
S600
Unpermissi-
ble electronics
temperature
S601
Overlling
Cause Rectication
– Temperature of the elec-
tronics in the non-specied
section
Danger of vessel overlling Make sure that there is no
Check ambient temperatureIsolate electronicsUse instrument with higher
temperature range
further lling
– Check level in the vessel
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
9 Diagnosis, Asset Management and service
Maintenance
The following table shows the error codes and text messages in the status message "Maintenance" and provides information on causes as well as corrective measures.
Code
Text mes­sage
M500 Error with the
reset delivery
status
M501
Error in the non-active
linearization table
M502
Error in the
diagnosis memory
M503 Reliability too
low
M504
Error on an device inter-
face
M505 No echo avail-
able
Cause Rectication
– With the reset to delivery
status, the data could not be restored
Hardware error EEPROM Exchanging the electronics
Hardware error EEPROM Exchanging the electronics
The echot/noise ratio is the
small for a reliable meas­urement
Hardware defect Check connections
Level echo can no longer
be detected
Repeat resetLoad XML le with sensor
data into the sensor
Send instrument for repair
Send instrument for repair
Check installation and
process conditions
Clean the antennaChange polarisation direc-
tion
Use instrument with higher
sensitivity
Exchanging the electronicsSend instrument for repair
Clean the antennaUse a more suitable
antenna/sensor
Remove possible false
echoes
Optimize sensor position
and orientation

9.4 Rectify faults

Reaction when malfunc­tions occur
Procedure for fault recti-
cation
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
The operator of the system is responsible for taking suitable meas­ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the indicating and
adjustment module Checking the output signal with 4 … 20 mA instruments
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
39
9 Diagnosis, Asset Management and service
Treatment of measure­ment errors with stand­pipe versions
The below tables show typical examples of application-relevant meas­urement errors with standpipe versions. The measurement errors are dierentiated according to:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
Level
1
2
0
Fig. 19: The broken line 1 shows the real level, the continuous line 2 shows the level displayed by the sensor
time
Instructions:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high

9.4 Measurement error with constant level

Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
2. Measured value jumps towards 100 %
Level
0
Level
0
Min./max. adjustment not
correct
Wrong linearization curve Adapt linearization curve
time
– Installation in a bypass tube or
standpipe, hence running time error (small measurement error close to 100 %/large error close to 0 %)
– Due to the process, the ampli-
tude of the product echo sinks
A false signal suppression was
not carried out
time
– Amplitude or position of a
false echo has changed (e.g. condensation, buildup); false signal suppression no longer
matches
Adapt min./max. adjustment
Check parameter "Application"
with respect to vessel form, adapt if necessary (bypass, standpipe, diameter)
Carry out false signal suppres-
sion
Determine the reason for the
changed false echo, carry out false signal suppression, e.g. with condensation
36526-EN-121011
40
VEGAPULS 66 standpipe version • Foundation Fieldbus
9 Diagnosis, Asset Management and service
9.4 Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value re­mains in the area of the bottom during lling
4. Measured value jumps to ≥ 100 % or
0 m distance
Level
– Echo from the tank bottom
larger than the product echo, for example, with products with
0
Level
εr < 2.5 oil-based, solvents
time
– Level echo is no longer
detected in the close range due to foam generation or false echoes in the close range. The sensor goes into overll protec­tion mode. The max. level (0 m distance) as well as the status message "Overll protection"
– Check application parameters
Medium, Vessel height and Floor form, adapt if necessary
– Check measuring site: Antenna
must protrude out of the socket
Remove contamination on the
antenna
Use a sensor with a more suit-
able antenna
are outputted.

9.4 Measurement error during emptying

Fault description Error pattern Cause Rectication
5. Measured value re­mains unchanged in the close range during emptying
6. Measured value jumps towards 0 % during emptying
Level
False echo larger than the level
echo
Level echo too small
0
time
Level
– Echo from the tank bottom
larger than the product echo, for example, with products with
0
εr < 2.5 oil-based, solvents
time
– Remove false echoes in the
close range. Check: Antenna must protrude out of the socket
Remove contamination on the
antenna
Minimize interfering instal-
lations in the close range by changing the polarization
direction
After removing the false ech-
oes, the false signal suppres­sion must be deleted. Carry out a new false signal suppression
– Check application parameters
Medium type, Vessel height and Floor form, adapt if necessary
Reaction after fault recti-
cation
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
24 hour service hotline
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside the normal working hours on seven days a week around the clock.
Since we oer this service worldwide, the support is in the English language. The service itself is free of charge, the only costs involved are the normal call charges.

9.5 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
41
9 Diagnosis, Asset Management and service
In Ex applications only one instrument and one electronics module with respective Ex approval may be used.
If there is no electronics module available on site, the electronics module can be ordered through the agency serving you. The electron­ics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").

9.6 Software update

The following components are required to update the sensor soft­ware:
Sensor
Voltage supply
Interface adapter VEGACONNECT 4
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".
42

9.7 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Formsandcerti- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
9 Diagnosis, Asset Management and service
Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
43

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
44
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus

11 Supplement

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435 Materials, wetted parts
Ʋ Process tting 316L, Hastelloy C22, Monell Alloy Ʋ Process seal On site Ʋ Antenna 316L, Hastelloy C22 Ʋ Antenna impedance cone PTFE, PP, PEEK, ceramic (99.7 % Al Ʋ seal, antenna system FKM (A+P GLT FPM 70.16-06), FFKM (Kalrez 6375),
Standpipe 316L, Hastelloy C22
Materials, non-wetted parts
Ʋ Plastic housing plastic PBT (Polyester) Ʋ Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated - basis:
Ʋ Stainless steel housing 316L Ʋ Seal between housing and housing
cover
Ʋ Inspection window in housing cover
(optional)
Ʋ Ground terminal 316L
Ohmic contact Between ground terminal, process tting and standpipe
Process tting Flanges DIN from DN 50, ANSI from 2" Weight
Ʋ Instrument (depending on housing,
process tting and antenna)
Ʋ Standpipe 3.2 kg/m (2.2 lbs/ft) Length standpipe max. 5.85 m (19.19 ft) Torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing max. 10 Nm (7.376 lbf ft)
Ʋ Aluminium/Stainless steel housing max. 50 Nm (36.88 lbf ft)
silicone FEP coated (A+P FEP-O-SEAL), EPDM (A+P 75.5/KW75F), PTFE, graphite (99.9 %) and
ceramic
Polyester
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housng, electropolished)
Polycarbonate
approx. 6.3 … 136 kg (13.89 … 299.8 lbs)
2O3
)
Input variable
Measured variable The measured quantity is the distance between process
tting of the sensor and product surface. The reference plane is the seal surface on the hexagon or the lower side of the ange.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
45
11 Supplement
1
3
2
4
Fig. 26: Data of the input variable
1 Reference plane 2 Measured variable, max. measuring range 3 Antenna length 4 Useful measuring range
Max. measuring range 35 m (114.83 ft) Useful measuring range According to the respective standpipe length. Min. distance 0 mm (0 in)
Output variable
Output
Ʋ Signal digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer according to IEC 61158-2
Damping (63 % of the input variable) 0 … 999 s, adjustable Channel Numbers
Ʋ Channel 1 Process value
Ʋ Channel 8 Electronics temperature
Ʋ Channel 9 Count rate Transmission rate 31.25 Kbit/s
Current value
Ʋ Non-Ex and Ex ia instrument 10 mA, ±0.5 mA
Ʋ Ex d instruments 16 mA, ±0.5 mA
Resolution, digital > 1 mm (0.039 in)
1)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F)
Ʋ Relative humidity 45 … 75 %
Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
1)
Standpipe supply Ex factory up to 5.85 m (19.19 ft)
46
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
Deviation with liquids See following diagrams
10 mm (0.394 in)
11 Supplement
5 mm (0.197 in)
- 5 mm (- 0.197 in)
- 10 mm (- 0.394 in)
Fig. 27: Deviation under reference conditions
1 Reference plane 2 Distance vent hole from reference plane 47 mm (4.85 in) 3 Standpipe end
0
0,5 m
(1.6 ft)
2
1 3
Reproducibility ≤ ±1 mm
Variablesinuencingmeasurementaccuracy
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
max. 10 mm
Additional deviation through strong, high
< ±50 mm
frequency electromagnetic elds acc. to
EN 61326
Inuenceofthesuperimposedgasandpressuretotheaccuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by high pressure. This eect depends on the superimposed gas or vapour and is especially large at low temperatures.
The following table shows the resulting deviation for some typical gases and vapours. The specied values refer to the distance. Positive values mean that the measured distance is too large, negative values that the measured distance is too small.
Gas phase Temperature Pressure
1 bar (14.5 psig)
Air 20 °C/68 °F 0.00 % 0.22 % 1.2 % 2.4 % 4.9 %
200 °C/392 °F -0.01 % 0.13 % 0.74 % 1.5 % 3.0 %
400 °C/752 °F -0.02 % 0.08 % 0.52 % 1.1 % 2.1 %
Hydrogen 20 °C/68 °F -0.01 % 0.10 % 0.61 % 1.2 % 2.5 %
200 °C/392 °F -0.02 % 0.05 % 0.37 % 0.76 % 1.6 %
400 °C/752 °F -0.02 % 0.03 % 0.25 % 0.53 % 1.1 %
10 bar (145 psig)
50 bar (725 psig)
100 bar (1450 psig)
200 bar (2900 psig)
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
47
11 Supplement
Gas phase Temperature Pressure
Steam (satu­rated steam)
1 bar (14.5 psig)
100 °C/212 °F 0.26 % - - - -
180 °C/356 °F 0.17 % 2.1 % - - -
264 °C/507 °F 0.12 % 1.44 % 9.2 % - -
366 °C/691 °F 0.07 % 1.01 % 5.7 % 13.2 % 76.0 %
10 bar (145 psig)
50 bar (725 psig)
Characteristics and performance data
Measuring frequency C-band (6 GHz technology) Measuring cycle time approx. 350 ms
Step response time
Tracking speed of the measuring window
2)
≤ 3 s 1 m/min
max. Emitted HF power (depending on the parameter adjustment)
Ʋ Average spectral transmission power
-31 dBm/MHz EIRP
3)
density
Ʋ Max. spectral transmission power
+24 dBm/50 MHz EIRP
density
Ʋ Specic absorption rate (SAR) 0.47 mW/kg
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
100 bar (1450 psig)
200 bar (2900 psig)
Process conditions
For the process conditions, please also note the specications on the type label. The lower value always applies.
Seal Antenna impedance
FKM (A+P GLT FPM
70.16-06)
FFKM (Kalrez 6375) PFFE -20 … +150 °C (-4 … +302 °F)
Silicone FEP coated (A+P FEP-O-SEAL)
EPDM (A+P 75.5/KW/75F) PFFE -40 … +150 °C (-40 … +302 °F)
Graphite Ceramic -60 … +250 °C (-76 … +482 °F)
cone
PFFE -40 … +150 °C (-40 … +302 °F)
PFFE -40 … +150 °C (-40 … +302 °F)
Ceramic -60 … +300 °C (-76 … +572 °F)
Ceramic -60 … +400 °C (-76 … +752 °F)
Process temperature (measured on the process
tting)
Vessel pressure relating to the antenna cone
Ʋ PTFE -100 … 4000 kPa/-1 … 40 bar (-14.5 … 580 psi) with
PN 40
2)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications until the output signal
has taken for the rst time 90 % of the nal value (IEC 61298-2).
3)
EIRP: Equivalent Isotropic Radiated Power
48
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement
Ʋ Ceramic -100 … 16000 kPa/-1 … 160 bar (-14.5 … 2321 psi) with
Vessel pressure relating to the ange nominal stage
Vibration resistance 1 g at 5 … 200 Hz according to EN60068-2-6 (vibration
Shock resistance 100 g, 6 ms according to EN 60068-2-27 (mechanical
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland M20 x 1.5 (cable: ø 5 … 9 mm)
Ʋ Cable entry ½ NPT
Ʋ Blind stopper M20 x 1.5; ½ NPT
Ʋ Closing cap M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit M12 x 1, according to ISO 4400, Harting HAN, 7/8" FF
Ʋ Indication circuit M12 x 1
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
tion cable
Ʋ Cable entry ½ NPT
Ʋ Blind stopper M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section 0.5 mm² (AWG 20)
Ʋ Wire resistance < 0.036 Ω/m
Ʋ Tensile strength < 1200 N (270 lbf)
Ʋ Standard length 5 m (16.4 ft)
Ʋ Max. length 180 m (590.6 ft)
Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F)
Ʋ Diameter approx. 8 mm (0.315 in)
Ʋ Colour - Non-Ex version Blue
Ʋ Colour - Ex-version Blue
PN 160
see supplementary instructions manual "Flanges ac­cording to DIN-EN-ASME-JIS"
with resonance)
shock)
M20 x 1.5 (cable: ø 5 … 9 mm)
Indicating and adjustment module
Voltage supply and data transmission through the sensor Indication LC display in dot matrix Measured value indication
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
49
11 Supplement
Ʋ Number of digits 5
Ʋ Size of digits W x H = 7 x 13 mm
Adjustment elements 4 keys Protection rating
Ʋ unassembled IP 20
Ʋ mounted into the sensor without cover IP 40
Materials
Ʋ Housing ABS
Ʋ Inspection window Polyester foil
Integrated clock
Date format Day.Month.Year Time format 12 h/24 h Time zone Ex factory CET
Measurement electronics temerature
Resolution 1 °C (1.8 °F) Accuracy ±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 9 … 32 V DC
Ʋ Non-Ex instrument 9 … 32 V DC
Ʋ EEx-ia instrument - Power supply
9 … 17.5 V DC
FISCO model
Ʋ EEx-ia instrument - Power supply
9 … 24 V DC
ENTITY model
Ʋ EEx-d instrument 14 … 32 V DC Power supply by/max. number of sensors
Ʋ H1 power supply max. 32 (max. 10 with Ex)
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67
Ʋ Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar)
4)
housing - investment casting, stain­less steel housing - electro-polished
Ʋ Aluminium and stainless housing, in-
IP 66/IP 68 (1 bar)
vestment casting (optionally available) Overvoltage category III Protection class III
Approvals
Instruments with approvals can have dierent technical data depending on the version.
4)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
50
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement
For that reason the associated approval documents of these instruments must be carefully noted. They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".

11.2 Supplementary information Foundation Fieldbus

The following table gives you an overview of the instrument versions and the corresponding device descriptions, the electrical characteristics of the bus system as well as the applied function blocks.
Revisions Data DD-Revision Rev_01
CFF-File 010101.c
Device Revision 0101.o
0101.sym
C-Revision xx xx 01
Device software revision > 4.4.0
ITK (Interoperability Test Kit) Number 5.0.2
Electricial Characteristics Physicial Layer Type Low-power signaling, bus-pow-
Input Impedance > 3000 Ohms between
Unbalanced Capacitance < 250 pF to ground from either
Output Amplitude 0.8 V P-P
Electrical Connection 2 Wire
Polarity Insensitive Ye s
Max. Current Load 10 mA
Device minimum operating voltage 9 V
Transmitter Function Blocks Resource Block (RB) 1
Transducer Block (TB) 1
Standard Block (AI) 3
Execution Time 30 mS
Advanced Function Blocks Discret Input (DI) Yes
PID Control Ye s
Output Splitter (OS) Yes
Signal Characterizer (SC) Yes
Integrator Yes
Input Selector (IS) Ye s
Arithmetic (AR) Yes
Diagnostics Standard Yes
Advanced Yes
Performance No
Function Blocks Instantiable No
ered, FISCO I.S.
7.8 KHz - 39 KHz
input terminal
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
51
11 Supplement
General Information LAS (Link Active Scheduler) Ye s
Master Capable Ye s
Number of VCRs (Virtual Communication Re­lationships)
24

11.2 Function blocks

Transducer Block (TB)
The Transducer Block "Analog Input (AI)" takes the original measured value (Secondary Value 2), carries out the min./max. adjustment (Secondary Value 1), carries out a linearization (Primary Value) and makes the values on its output available for further function blocks.
min/max
adjustment
Secondary
Value 2
m(d)
Secondary
Value 1
Sensor_Value
Fig. 28: Schematic presentation Transducer Block (TB)
Function block Analog Input (AI)
The function block "Analog Input (AI)" takes the original measured value selected by a Channel Number and makes it available to additional function blocks on its output.
Linearization
% Lin%
Primary
Value
TB
Fig. 29: Schematic presentation function block Analog Input (AI)
Function block Discret Input (DI)
The function block "Discret Input (DI)" takes the original measured value selected by a Channel Number and makes it available to additional function blocks on its output.
52
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement
Fig. 30: Schematic presentation function block Discret Input (DI)
Function block PID Control
The function block "PID Control " is a key component for various tasks in the process automation and is used universally. PID blocks can be cascaded if this is necessary or requested due to dier­ent time constants with the primary and secondary process measurement.
Fig. 31: Schematic presentation function block PID Control
Function block Output Splitter
The function block "Output Splitter" generates two control outputs out of one input. Each output is a linear image of a part of the input. A retrograde calculation function is realised by using the linear imaging function inversely. A cascading of several Output Splitters is supported by an integrated decision table for the combinability of inputs and outputs.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
53
11 Supplement
Fig. 32: Schematic presentation function block Output Splitter
Function block Signal Characterizer
The function block "Signal Characterizer" has two channels the outputs of which are not in linear relation with the respective input. The non-linear relation is dened by a look-up table with individu­ally selectable x/y-pairs. The respective input signal is imaged on the corresponding output, hence this function block can be used in a control loop or signal path. Optionally the function axis can be exchanged in channel 2 so that the block can be also used in a reverse control loop.
Fig. 33: Schematic presentation function block Signal Characterizer
Function block Integrator
The function block "Integrator" integrates a continuous input signal over the time and sums the results of an impulse input block. It is used as a totalizer up to a reset or as a subtotalizer up to a reference point at which the integrated and accumulated value is compared with the default values. When these default values are reached, digital output signals will be outputted. The integration func­tion is carried out upwardly starting with zero and downwards with a default value. Two ow values are also available so that the net ow volume can be calculated and integrated. This can be used for calculation of volume and mass changes in the vessel or for optimisation of ow controls.
54
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement
Fig. 34: Schematic presentation function block Integrator
Function block Input Selector
The function block "Input Selector" oers selection possibilities for up to four inputs and generates an output signal according to the selection criteria. Typical input signals are AI blocks. Selection possibilities are maximum, minimum, mean value, average value and rst useful signal. Through parameter combination, the block can be used as rotary switch or as preselection switch for the rst useful value. Switch information can be received by other input blocks or the user. Mean value selection is also supported.
Fig. 35: Schematic presentation function block Input Selector
Function block Arithmetic
The function block "Arithmetic" allows the simple integration of usual metrological calculation functions. The user can select the requested algortihm according to the name without known the formula.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
55
11 Supplement
The following algorithms are available:
Flow compensation, linear
Flow compensation, square root
Flow compensation, approximate
BTU ow
Traditional Multiply Divide
Average
Traditional Summer
Fourth order polynomial
Simple HTG compensated level
Fourth order Polynomial Based on PV
Fig. 36: Schematic presentation function block Arithmetic

11.2 Parameter list

The following table gives you an overview of the parameters used.
FF desciptor Description Unit
PRIMARY_VALUE PRIMARY_VALUE (Linearized value). This is the process val-
PRIMARY_VALUE_UNIT Selected unit code for "PRIMARY_VALUE"
SECONDARY_VALUE_1 This is the measured value after min/max adjustment with the
SECONDARY_VALUE_1_
UNIT
SECONDARY_VALUE_2 This is the distance value ("sensor_value") with the status of the
FILL_HEIGHT_VALUE Filling height. The unit is dened in "FILL_HEIGHT_VALUE_
FILL_HEIGHT_VALUE_UNIT Filling height unit
CONST_VALUE Constant value
56
ue after min/max adjustment and Linearization with the status of the transducer block. The unit is dened in "PRIMARY_VAL- UE_UNIT"
status of the transducer block. The unit is dened in "SECOND- ARY_VALUE_1_UNIT"
Selected unit code for "SECONDARY_VALUE_1"
transducer block. The unit is dened in "SECONDARY_VAL-
UE_2_UNIT"
UNIT"
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement
FF desciptor Description Unit
SECONDARY_VALUE_1_ TYPE
SECONDARY_VALUE_2_ TYPE
FILL_HEIGHT_VALUE_Type Filling height value type
DIAGNOSIS AITB Diagnosis
DIAG_MASK_1
DIAG_OUT_1
DIAG_MASK_2
DIAG_OUT_2
DEVICE_IDENTIFICATION Manufacturer ID, device type, bus type ID, measurement princi-
DEVICE_NAME Device name
IS-SPARE_ELECTRONICS Device name
DEVICE_VERSION_INFO Hard- and software version for system, function and error
CALIBRATION_DATE Day, month and year
FIRMWARE_VERSION_ASCII Software version
HW_VERSION_ASCII Hardware version
ADJUSTMENT_DATA Min./max.-adjustment physical, percent and oset
FIRMWARE_VERSION_MAIN Firmware versions major, minor, revision and build
PHYSICAL_VALUES Distance, distance unit, distance status, level and status
DEVICE_UNITS Distance and temperature units of the instrument
APPLICATION_CONFIG Medium type, media, application type, vessel bottom, vessel
LINEARIZATION_TYPE_SEL Type of linearization
SIMULATION_PHYSCAL
INTEGRATION_DATA Physical oset and integration time
DEVICE_CONFIG_PULS_ RADAR
ADJUSTMENT_LIMITS_MIN Min. range min.-/max.- values physical, percent, oset
ADJUSTMENT_LIMITS_MAX Max. range min.-/max.- values physical, percent, oset %
FALSE_SIGNAL_COMMAND %
FALSE_SIGNAL_CMD_CRE­ATE_EXTEND
FALSE_SIGNAL_CMD_DE­LET_REGION
FALSE_SIGNAL_CMD_STATE Busy, last command, errorcode
FALSE_SIGNAL_CMD_CON­FIGURATION1
Secondary value 1 type
Secondary value 2 type
ple, serial number, DTM ID, device revision
height
Electronics variant, probe type, max. measuring range, anten­na extension length, adjustment propagation antenna extension lprapproval conguration
Amplitude safety of the 0 % curve, safety of the false signal sup­pression, position of the 0 % and 100 % curve in near and far range
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
57
11 Supplement
FF desciptor Description Unit
FALSE_SIGNAL_CMD_CON­FIGURATION2
ECP_CURVE_AVARAGING_ CONFIG
LEVEL_ECHO_MEASURE­MENT
ECHO_CURVE_STATUS
PACKET_COUNT
GU_ID_END
ECHO_CURVE_READ Echo curve data
ECHO_EVALUATOR Echo parameters, rst large echo, amplitude threshold rst
ECHO_DECIDER Echo selection criteria, fault signal on loss of echo, delay on
DISPLAY_SETTINGS Indication value, menu language, lightning
SIL_MODE
EDENVELOPE_CURVE_FIL­TER
EDDETECTION_CURVE_FIL­TER
EDECHO_COMBINATION Parameters for echo combination, function combine echoes,
LIN_TABLE_A … LIN_ TABLE_Q
ELECTRONICS_INFORMA-
TION
APPLICATION_CONFIG_ SERVICE
LEVEL_ECHO_INFO Level echo ID, amplitude, measurement safety
DEVICE_STATUS Device status
FALSE_SIGNAL_LIMITS False signal distance min./max.
USER_PEAK_ELEC_TEMP Min.-/max.- values of electronics temperature, date
USER_MIN_MAX_PHYSI­CAL_VALUE
RESET_PEAK_PHYSICAL_
VALUE
RESET_LINEARIZATION_ CURVE
DEVICE_STATUS_ASCII Device status
ECHO_CURVE_PLICSCOM_ REQUEST
Gradient of the manual sectors, safety at the end of false echo memory and depending on the import range gating out the false signals
Averaging factor on increasing and decreasing amplitude
Function measured value lter
large echo
fault signal on loss of echo
Parameters of envelope curve lter, activation of smooth raw value curve
Parameters of the detection lter, oset threshold value curve
amplitude dierence of combined echoes, position dierence of combined echoes
32 couples of percentage and lin. percentage values
Electronics version
Limitation measuring range begin, safety of measuring range
end
Min.-/max.- distance values, date
Parameters as curve selection and resolution
36526-EN-121011
58
VEGAPULS 66 standpipe version • Foundation Fieldbus
11 Supplement
FF desciptor Description Unit
ECHO_CURVE_PLICSCOM_
Parameters as start and end
LIMITS
APPROVAL_WHG Sensor acc. to WHG
DEVICE_STATE_CONFIG Function check, maintenance required, out of specication
ELECTRONIC_TEMPERA-
Electronics temperature
TURE
RESET_PEAK_ELECTRON­IC_TEMP
FOCUS_RANGE_CONFIG Width focusing range, time for opening the focusing range, min.
measurement reliability in and outside the focusing range
NOISE_DETECTION_INFO Increase of the system noise
NOISE_DETECTION_CON-
System noise treatment
FIG
ECHO_MEM_SAVE_CURVE_ TYPE
ECHO_MEM_STATE Busy, curve type, error code

11.3 Dimensions

The following dimensional drawings represent only an extract of all possible versions. Detailed dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.03")
~ 84 mm (3.31")
M16x1,5
ø 79 mm
(3.31")
M20x1,5/ ½ NPT
112 mm (4.41")
1
M20x1,5/ ½ NPT
2
112 mm (4.41")
Fig. 37: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
59
11 Supplement
Aluminium housing
~ 116 mm (4.57")
ø 86 mm (3.39")
~ 87 mm (3.43")
M16x1,5
ø 86 mm
(3.39")
M20x1,5/ ½ NPT
M20x1,5
116 mm (4.57")
M20x1,5/ ½ NPT
21
120 mm (4.72")
Fig. 38: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
Aluminium housing in protection rating IP 66/IP 68 (1 bar)
~ 150 mm (5.91")
ø 84 mm (3.31")
116 mm (4.57")
M20x1,5M20x1,5
1
Fig. 39: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
~ 105 mm (4.13")
M20x1,5/ ½ NPT
ø 84 mm
(3.31")
2
M16x1,5
120 mm (4.72")
60
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
Stainless steel housing
~ 59 mm
(2.32")
ø 80 mm
(3.15")
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 87 mm (3.43")
M16x1,5
ø 86 mm
(3.39")
11 Supplement
M20x1,5/ ½ NPT
112 mm (4.41")
M20x1,5/ ½ NPT
117 mm (4.61")
M20x1,5/ ½ NPT
321
120 mm (4.72")
Fig. 40: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
Stainless steel housing in protection rating IP 66/IP 68, 1 bar
~ 93 mm
(3.66")
M20x1,5/ ½ NPT
ø 80 mm
(3.15")
112 mm (4.41")
~ 105 mm (4.13")
M20x1,5/ ½ NPT
ø 84 mm
(3.31")
21
M16x1,5
120 mm (4.72")
Fig. 41: Housing versions in protection IP 66/IP 68 (1 bar) - with integrated indicating and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 2 Double chamber version, precision casting
~ 103 mm
(4.06")
M20x1,5
ø 77 mm
(3.03")
3
117 mm (4.61")
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
61
11 Supplement
VEGAPULS 66, standpipe version
60 mm
ø 8 mm
(1.85")
L
47 mm
(0.32")
ø 56 mm
(2.21")
ø 52 mm
(2.05")
(2.36")
2 3 1
(7.09")
180 mm
(12.48")
317 mm
Fig. 42: VEGAPULS 66, standpipe version
5)
1 Standard version 2 For process temperatures up to 250 °C (482 °F) 3 For process temperatures up to 400 °C (752 °F)
5)
The plant operator must provide a suitable support for the standpipe depending on the length and the process
conditions.
62
VEGAPULS 66 standpipe version • Foundation Fieldbus
36526-EN-121011
11 Supplement

11.4 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

11.5 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/ originator.
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
63
INDEX
INDEX
A
Accessories
External indicating and adjustment unit 12External radio unit 13Flanges 13Indicating and adjustment module 12Interface adapter 12 – Protective cover 13
Adjustment 24
Max. adjustment 26Min. adjustment 25
Adjustment system 23
C
Cable entry 15 Connection procedure 16 Connection technology 16 Curve indication
– Echo curve 27 – False signal suppression 27
D
DD (Device Description) 34
E
Echo curve memory 35 Echo curve of the setup 27 EDD (Enhanced Device Description) 34 Error messages 36 Event memory 35
F
False signal suppression 27 Fault rectication 39 FF parameter 56 Functional principle 11 Function blocks
Analog Input (AI) 52 – Arithmetic 55 – Discret Input (AI) 52Input Selector 55Integrator 54Output Splitter 53PID Control 53Signal Characterizer 54Transducer Block (TB) 52
L
Linearisation curve 29
M
Main menu 23 Meas. reliability 26 Measured value memory 35 Measurement error 40 Medium
– Liquid Bulk solid 24
Moisture 14
N
NAMUR NE 107
Failure 37Function check 38Maintenance 39Out of specication 38
O
Overll protection according to WHG 29
P
Packaging 11 Peak value 26
R
Reection properties 24 Repair 42 Replacement parts
– Electronics module 13
S
Service hotline 41 Status messages 36 Storage 12
T
Type label 10
36526-EN-121011
K
Key function 22
64
VEGAPULS 66 standpipe version • Foundation Fieldbus
Notes
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
65
Notes
66
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
Notes
36526-EN-121011
VEGAPULS 66 standpipe version • Foundation Fieldbus
67
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2012
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
36526-EN-121011
Loading...