VEGA PULS60 User Manual

Supple
mentary information
Recurring function
for VEGAPULS series 60
test
Document ID:
36553
1 Conte
nts
Conten
1 About this document
2 Prerequisites
3 Sequence of the recurring function test
4 Test protocol - Recurring function test
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Required comparative data . . . . . . . . . . . . . . . . . . . .
2.4 Required plant situation . . . . . . . . . . . . . . . . . . . . ..
3.1 Restart of the sensor . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Verication of the current output . . . . . . . . . . . . . . . .
3.3 Verication of the instrument parameters . . . . . . . . . .
3.4 Verication of echo data . . . . . . . . . . . . . . . . . . . . . .
3.5 Sensor reaction to a level change . . . . . . . . . . . . . . .
3.6 Result of the recurring function test . . . . . . . . . . . . . .
5 5 5 5
7
9 10 11 13 14
2 Recurring
36553-EN-100225
function test for VEGAPULS series 60
out this document
1 Ab
Why a function test?
Scope
1 Abou
t this document
The recurring function test serves to check the safety function and reveal possible undetected, dangerous failures
). The functional
du
capability of the measuring system has to be tested in adequate time intervals. It is the user's responsibility to select an appropriate type of testing. The time intervals are subject to the PFD
-value according to
avg
the specications in the Safety Manual (SIL).
With high demand rate, a recurring function test is not requested in IEC 61508. The functional eciency of the measuring system is
demonstrated by the frequent use of the system. In double channel architectures it is a good idea to verify the eect of the redundancy through recurring function tests at appropriate intervals.
This supplementary information applies to measuring systems con­sisting of radar sensor VEGAPULS series 60 in two-wire and four-wire 4 20 mA/HART versions:
VEGAPULS 61, 62, 63, 65, 66, 67, 68
Valid hardware and software versions:
Serial number of the
electronics
VEGAPULS 61, 62, 63, 65, 66
VEGAPULS 61, 62, 63
with increased sensitiv­ity
VEGAPULS 67, 68 > 14165303 from version 3.25 up to
> 13978716 from version 3.22 up to
> 14165303 from version 3.25 up to
Sensor software
version 3.80
version 3.80
version 3.80
Fault detection rate λ
96 %
36553-EN-100225
Recurring
function test for VEGAPULS series 60 3
With this supplementary information you can carry out a recurring
function test of VEGAPULS series 60 level sensors without dis­mounting the instrument or moving the product level to the switching point.
With this procedure, 88 % or 96 % of all dangerous undetected instrument failures
) are detected (DC
du
Proof
).
The described function test fulls the requirements according to SIL.
If you have already created a sensor documentation during setup, you
du
can check the sensor at a detection rate of 96 % of all dangerous undetected failures.
The remaining dangerous undetected failures are 11 FIT.
9
FIT = Failure In Time (1 FIT = 1 failure/10
h)
1 A bou
t this document
Make sure that also a later sensor documentation is possible. This sensor documentation of a recurring function test must be at least 6 months old.
Fault detection rate λ
88 %
If you created no sensor documentation during setup, you can check
du
the sensor only at a detection rate of 88 % of all dangerous undetected failures.
The remaining dangerous undetected failures are in this case 32 FIT.
9
FIT = Failure In Time (1 FIT = 1 failure/10
Ple
ase note the Ex-specic safety information for installation and
h)
operation in Ex areas. These safety instructions are part of the scope of delivery and come with the Ex-approved instruments.
War
ning:
The recurring function test inuences connected devices. Take note that downstream devices may be activated during the test.
Informa
tion:
Proceed according to the specied, recommended sequence of these instructions to isolate possible device failures systematically.
Informa
tion:
Document the recurring function test, for example, in the test protocol in the supplement. To facilitate the recording and for further function tests, we recommend to copy the empty test protocol before completing it.
This supplementary information manual can be downloaded from our download section.
e recurring function test cannot replace the prescribed test
Th according to WHG (Water Resources Act).
4 Recurring
36553-EN-100225
function test for VEGAPULS series 60
2 Prer
equisites
2 Prere
quisites

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Required tools

l This test instruction l PACTware l Actual VEGA DTM Collection l Device-DTM of the corresponding sensor (part of the VEGA DTM
Collection)
l Communication-DTM (part of the VEGA DTM Collection) l Interface adapter VEGACONNECT l mA-meter or PLC resp. DCS (accuracy ≤±0.2 %) l Operating instructions manual of the sensor l Safety Manual

2.3 Required comparative data

The setup data should be used for verication of the settings.
The following setup data are required:
l Sensor documentation of the setup with all parameters or a sensor
documentation created at least 6 months ago
l Documentation of all parameter changes since the setup
36553-EN-100225
Recurring
the sensor documentation of the setup or a sensor documentation
If created at least 6 months ago is not available, then the described recurring function test
96 %) cannot be carried out completely. In
du
this case, only the test at a detection rate of 88 % of all dangerous, undetected failures is possible.

2.4 Required plant situation

Caution:
e sure that there are no considerable process-relating changes in
Mak your plant during the recurring function test. This means also that the level in the vessel should not change signicantly by lling or emptying during the test. Make sure that also temperature changes, stirrers, current reactions in the vessel, etc. can cause level changes.
function test for VEGAPULS series 60 5
0
dB
1
2
2 P rereq
uisites
tion:
Informa
Document the recurring function test, for example, in the test protocol in the supplement.
The following conditions must be fullled when performing the recurring function test:
Fig. 1: Level 1 Amplitude of the useful echo above the noise level (signal/noise ratio) 2 Noise level
echo - VEGAPULS
l The level must be in the following areas:
- Min. distance to the level: Lower antenna edge +200 mm
- Level above the vessel bottom 250 mm
- Reliability at least 20 dB (amplitude of the useful logarithmic
echo above the noise level). The reliability can be veried during the test.
l The process conditions must be nearly constant (level, process
pressure, process temperature).
l The medium must be the same as during setup or the medium
must at least belong to the same product group
- Solvents/Liquid gases/Hydrocarbons/Oils (DK value < 3)
- Chemical mixtures (DK value 3 … 10)
- Water/Acids/Bases (DK value > 10 or conductive)
36553-EN-100225
6 Recurring
function test for VEGAPULS series 60
Loading...
+ 14 hidden pages