VEGA PULS60 User Manual

Supple
mentary information
Recurring function
for VEGAPULS series 60
test
Document ID:
36553
1 Conte
nts
Conten
1 About this document
2 Prerequisites
3 Sequence of the recurring function test
4 Test protocol - Recurring function test
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Required comparative data . . . . . . . . . . . . . . . . . . . .
2.4 Required plant situation . . . . . . . . . . . . . . . . . . . . ..
3.1 Restart of the sensor . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Verication of the current output . . . . . . . . . . . . . . . .
3.3 Verication of the instrument parameters . . . . . . . . . .
3.4 Verication of echo data . . . . . . . . . . . . . . . . . . . . . .
3.5 Sensor reaction to a level change . . . . . . . . . . . . . . .
3.6 Result of the recurring function test . . . . . . . . . . . . . .
5 5 5 5
7
9 10 11 13 14
2 Recurring
36553-EN-100225
function test for VEGAPULS series 60
out this document
1 Ab
Why a function test?
Scope
1 Abou
t this document
The recurring function test serves to check the safety function and reveal possible undetected, dangerous failures
). The functional
du
capability of the measuring system has to be tested in adequate time intervals. It is the user's responsibility to select an appropriate type of testing. The time intervals are subject to the PFD
-value according to
avg
the specications in the Safety Manual (SIL).
With high demand rate, a recurring function test is not requested in IEC 61508. The functional eciency of the measuring system is
demonstrated by the frequent use of the system. In double channel architectures it is a good idea to verify the eect of the redundancy through recurring function tests at appropriate intervals.
This supplementary information applies to measuring systems con­sisting of radar sensor VEGAPULS series 60 in two-wire and four-wire 4 20 mA/HART versions:
VEGAPULS 61, 62, 63, 65, 66, 67, 68
Valid hardware and software versions:
Serial number of the
electronics
VEGAPULS 61, 62, 63, 65, 66
VEGAPULS 61, 62, 63
with increased sensitiv­ity
VEGAPULS 67, 68 > 14165303 from version 3.25 up to
> 13978716 from version 3.22 up to
> 14165303 from version 3.25 up to
Sensor software
version 3.80
version 3.80
version 3.80
Fault detection rate λ
96 %
36553-EN-100225
Recurring
function test for VEGAPULS series 60 3
With this supplementary information you can carry out a recurring
function test of VEGAPULS series 60 level sensors without dis­mounting the instrument or moving the product level to the switching point.
With this procedure, 88 % or 96 % of all dangerous undetected instrument failures
) are detected (DC
du
Proof
).
The described function test fulls the requirements according to SIL.
If you have already created a sensor documentation during setup, you
du
can check the sensor at a detection rate of 96 % of all dangerous undetected failures.
The remaining dangerous undetected failures are 11 FIT.
9
FIT = Failure In Time (1 FIT = 1 failure/10
h)
1 A bou
t this document
Make sure that also a later sensor documentation is possible. This sensor documentation of a recurring function test must be at least 6 months old.
Fault detection rate λ
88 %
If you created no sensor documentation during setup, you can check
du
the sensor only at a detection rate of 88 % of all dangerous undetected failures.
The remaining dangerous undetected failures are in this case 32 FIT.
9
FIT = Failure In Time (1 FIT = 1 failure/10
Ple
ase note the Ex-specic safety information for installation and
h)
operation in Ex areas. These safety instructions are part of the scope of delivery and come with the Ex-approved instruments.
War
ning:
The recurring function test inuences connected devices. Take note that downstream devices may be activated during the test.
Informa
tion:
Proceed according to the specied, recommended sequence of these instructions to isolate possible device failures systematically.
Informa
tion:
Document the recurring function test, for example, in the test protocol in the supplement. To facilitate the recording and for further function tests, we recommend to copy the empty test protocol before completing it.
This supplementary information manual can be downloaded from our download section.
e recurring function test cannot replace the prescribed test
Th according to WHG (Water Resources Act).
4 Recurring
36553-EN-100225
function test for VEGAPULS series 60
2 Prer
equisites
2 Prere
quisites

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Required tools

l This test instruction l PACTware l Actual VEGA DTM Collection l Device-DTM of the corresponding sensor (part of the VEGA DTM
Collection)
l Communication-DTM (part of the VEGA DTM Collection) l Interface adapter VEGACONNECT l mA-meter or PLC resp. DCS (accuracy ≤±0.2 %) l Operating instructions manual of the sensor l Safety Manual

2.3 Required comparative data

The setup data should be used for verication of the settings.
The following setup data are required:
l Sensor documentation of the setup with all parameters or a sensor
documentation created at least 6 months ago
l Documentation of all parameter changes since the setup
36553-EN-100225
Recurring
the sensor documentation of the setup or a sensor documentation
If created at least 6 months ago is not available, then the described recurring function test
96 %) cannot be carried out completely. In
du
this case, only the test at a detection rate of 88 % of all dangerous, undetected failures is possible.

2.4 Required plant situation

Caution:
e sure that there are no considerable process-relating changes in
Mak your plant during the recurring function test. This means also that the level in the vessel should not change signicantly by lling or emptying during the test. Make sure that also temperature changes, stirrers, current reactions in the vessel, etc. can cause level changes.
function test for VEGAPULS series 60 5
0
dB
1
2
2 P rereq
uisites
tion:
Informa
Document the recurring function test, for example, in the test protocol in the supplement.
The following conditions must be fullled when performing the recurring function test:
Fig. 1: Level 1 Amplitude of the useful echo above the noise level (signal/noise ratio) 2 Noise level
echo - VEGAPULS
l The level must be in the following areas:
- Min. distance to the level: Lower antenna edge +200 mm
- Level above the vessel bottom 250 mm
- Reliability at least 20 dB (amplitude of the useful logarithmic
echo above the noise level). The reliability can be veried during the test.
l The process conditions must be nearly constant (level, process
pressure, process temperature).
l The medium must be the same as during setup or the medium
must at least belong to the same product group
- Solvents/Liquid gases/Hydrocarbons/Oils (DK value < 3)
- Chemical mixtures (DK value 3 … 10)
- Water/Acids/Bases (DK value > 10 or conductive)
36553-EN-100225
6 Recurring
function test for VEGAPULS series 60
ence of the recurring function test
3 Sequ
Function test not suc­cessful
3 Sequen
ce of the recurring function test
Carry out the recurring function test in the following sequence:
l 3.1 Restart the sensor l 3.2 Verication of the current output l 3.3 Verication of the instrument parameters (only with sensor
documentation)
l 3.4 Verication of the echo data (only with sensor documentation) l 3.5 Sensor reaction to a level change
Informa
tion:
Document the recurring function test, for example, in the test protocol in the supplement.
If one of the test points was not terminated successfully, there is probably an undetected dangerous failure. The recurring function test has failed.
In this case, proof of functional safety can only be provided by moving the level to the switching point.

3.1 Restart of the sensor

With this test point, it is possible to check of the sensor ouputs the same value within the prescribed min. accuravy after a restart.
ning:
War
The recurring function test inuences connected devices. Take note that downstream devices may be activated during the test.
Before the restart
36553-EN-100225
Recurring
function test for VEGAPULS series 60 7
Carry out a restart of the sensor. Proceed as follows:
1 Start PACTware and the corresponding sens or DTM.
Make sure that the conditions to the plant situation are maintained. See "Required plant situation".
(range of the actual level or the reliability at least 20 dB)
2 Set the indication to "Current".
3 The level is subject to plant or process-relevant uctuations.
Control the indicated current values over an adequate period.
Make sure that a damping is probably adjusted on the sensor.
4 Note the upper and lower limit values of the measured value.
5 Measure the output current of the sensor.
Preferrably use the indication of the input current value in the processing system.
1
2
3
mA
3 Sequenc
e of the recurring function test
you do not have this possibility, connect an mA-meter according
If to the following illustration.
You require the mA-meter for the verication of the current output in the next test point. The accuracy of the mA-meter should be better than 0.2 %. Select the smallest measuring range covering 4 20 mA.
After switching on again
Fig. 2: Connection
1 Level sensor 2 Processing system 3 mA-meter
6 Switch othe power supply.
7 Switch the voltage supply on again after approx. 10 s.
If the software signals a communication error during or after switching othe power supply, you have to acknowledge it.
After connecting the sensor to power supply respectively after the voltage recurrence, the instrument carries out a self-check for approx. 30 seconds:
l Internal check of the electronics l The output signal jumps to the set error current
Then the current corresponding to the level is outputted to the cable.
1 The level is subject to plant or process-relevant uctuations.
Control the indicated current values over an adequate period.
2 Note the upper and lower limit values of the measured value.
3 Compare the actually noted current values with the previously
noted values.
The two values must correspond within the safety accuracy of 2 %
(±0.32 mA).
If the two dierential values are within the safety accuracy, then the
test of the restart was successful.
Continue with the next test point.
of the mA-meter
36553-EN-100225
8 Recurring
function test for VEGAPULS series 60
ence of the recurring function test
3 Sequ
Simulation 4 mA
Simulation 20 mA
3.2 Vericat
In this test point you simulate certain level values via the current output. With this you can test the reaction of the sensor with dierent current output values and the switching behaviour.
Warning:
The recurring function test inuences connected devices. Take note that downstream devices may be activated during the test.
1 Select in the DTM under the menu "Service" the menu item
"Simulation".
2 Select "Current" as measured variable for the simulation.
3 Activate the simulation.
4 Set the simulation value to 4 mA.
Take note that downstream devices may be activated.
5 Accept the simulation value.
Wait approx. 30 s.
The simulation is running and a corresponding current is being
outputted.
6 Note the displayed value (4 mA simulation) of the mA-meter.
The value must correspond with the simulated value within the safety accuracy of 2 % (±0.32 mA).
Continue with the simulation if the two values correspond.
1 Set the simulation value of the current simulation to 20 mA.
Take note that downstream devices may be activated.
2 Accept the simulation value.
Wait approx. 30 s.
The simulation is running and a corresponding current is being
outputted.
3 Note the displayed value (20 mA simulation) of the mA-meter.
The value must correspond with the simulated value within the safety accuracy of 2 % (±0.32 mA).
The verication of the current output was successful if the two values
correspond.
ion of the current output
36553-EN-100225
Recurring
Caution:
ate the simulation.
Deactiv
Continue with the next test point.
function test for VEGAPULS series 60 9
3 Sequenc
e of the recurring function test
Create a current sensor documentation
3.3 Vericat
For this test point, the sensor documentation of the setup or the last sensor documentation (at least 6 months old) is required. If a parameter was changed since then, you also require the protocol or the sensor documentation of this parameter change.
If this sensor documentation is not available, then the described recurring function test cannot be carried out completely. In this case, only the test with a detection rate of 88 % of all dangerous, undetected failures is possible.
In this case, you continue with the test point "Sensor reaction on level change" or create an actual sensor documentation and carry out the
function test after at least 6 months. The actual parameter adjustment must be checked for correctness in the respect.
A sensor documentation was generated directly after the setup or at least 6 months ago. For assessment of the instrument parameters, this sensor documentation of the setup, the actual sensor documentation of the last parameter change or the sensor documentation generation and checked 6 months ago must be taken into consideration.
Create now a sensor documentation with the current instrument parameters. Proceed as follows:
1 Select the function "Print" in the DTM.
2 For complete sensor documentation, you have to select all
instrument parameters (except laboratory parameters).
A multiple-page pdf documentation containing all relevant sensor data is then generated.
3 Save this documentation as pdf document and, where appropriate,
print out the documentation to be on the safe side.
4 Compare the instrument parameters of this actual sensor
documentation with the sensor documentation of the setup or the last parameter change.
Deviating parameters must be documented, justied and checked on correctness.
If the actual sensor documentation corresponds to the stored sensor documentation or if the modied parameters are correct, then the verication of the instrument parameters was successful.
Continue with the next test point.
ion of the instrument parameters
36553-EN-100225
10 Recurring
function test for VEGAPULS series 60
ence of the recurring function test
3 Sequ
Calculation of the cor­rection factor
3.4 Vericat
ion of echo data
For this test point, you require the sensor documentation of the setup or a sensor documentation created at least 6 months ago.
If none of these sensor documentations is available, the described recurring function test
96 %) cannot be carried out.
du
Only the test at a detection rate of 88 % of all dangerous undetected failures is possible. In this case, just skip this test point.
In this case, continue with the test point "Sensor reaction on level change".
Use the two pdf les of the sensor documentation again for assessment of the level echo.
Under chapter "Echo curve" you will nd a short chart containing the "Echo data". The data of this chart are relevant for the assessment.
Compare the values of the two echo data charts. The echo curve itself cannot be compared.
Proceed as follows:
1 Compare the distance [m] of the two level echoes (actual/setup).
2 Divide the distance [m] of the current level through the distance [m]
of the level echo during setup.
Round this ratio "V" mathematically to one decimal point.
3 Search in the following chart the correction value (dB) belonging to
the calculated ration "V".
4 Calculate with the correction value the corrected value of the
amplitude.
This corrected value of the amplitude [dB] of the actual level echo can be higher, but only max. 6 dB lower than the amplitude of the level echo during setup.
36553-EN-100225
Recurring
Ratio "V" Correction value (dB)
0.5 -6
0.6 -4.5
0.7 -3
0.8 -2
0.9 -1
1.0 -0
function test for VEGAPULS series 60 11
3 Sequenc
e of the recurring function test
Ratio "V" Correction
1.1 +0.8
1.2 +1.5
1.3 +2.3
1.4 +3
1.5 +3.5
1.6 +4
1.7 +4.6
1.8 +5
1.9 +5.6
2.0 +6
value (dB)
Example
During setup, the level echo had 32 dB. The distance [m] to the medium was 12.9 m.
The actual level echo has 25 dB. The distance [m] to the medium is
15.8 m.
Division of the current distance [m] by the distance [m] during setup:
15.8 m : 12.9 m = 1.224
Round mathematically to one position after the decimal point: 1.224 -->
1.2
Correction value for the ratio (1.2): + 1.5
Calculation of thecorrectedvalueoftheamplitude:25 dB + 1.5 = 26.5 dB
The result of the example (26.5 dB) is thus within the permissible tolerance range of - 6 dB (32 dB - 6 dB = 26 dB)
In this example, the test point would be successfully fullled.
The following criteria must correspond when comparing the echo data:
l If there are false echoes in front of the level echo (highest useful
echo probability), they must have a useful echo probability of 0 %.
l The amplitude [dB] (with correction factor) of the actual level echo
(highest useful echo probability) corresponds to the respective
value of the level echo from the setup (tolerance max. -6 dB). This means that the actual level echo can be higher, but only max. 6 dB lower than the level echo during setup.
12 Recurring
36553-EN-100225
function test for VEGAPULS series 60
ence of the recurring function test
3 Sequ
all the above conditions are fullled, the measurement works
If correctly and the verication of the echo data was successful.
Continue with the next test point.

3.5 Sensor reaction to a level change

With this test point, you monitor the reaction of the sensor during a level change.
1 Set the indication of the sensor DTM to "Distance".
2 Change the lling of your vessel.
It doesn't matter if you ll or empty the vessel.
The lling speed is also not relevant.
The level change must be at least 50 mm.
3 Take note of the sensor reaction.
Does the measured value [m(d)] move in the correct direction
during emptying/lling?
The displayed measured value (distance) is the distance between seal surface of the sensor and the product surface.
The measured value will decrease during lling.
The measured value will increase during emptying.
When the level value changes analogue to the level change, the measurement works correctly and the assessment of the sensor reaction was successful.
If all function tests were successful, the recurring function test is nished.
36553-EN-100225
Recurring
you used an mA-meter for the recurring function test, switch othe
If
sensor and remove the mA-meter from the sensor cable after nishing the recurring function test.
function test for VEGAPULS series 60 13
3 Sequenc
e of the recurring function test
Function test successful
Function test not suc-
cessful
3.6 Res
ult of the recurring function test
If all test points could be terminated successfully, then the recurring function test was successful.
Fullled test points
3.1 / 3.2 / 3.3 / 3.4 / 3.5 Fault detection rate λ
3.1 / 3.2 / 3.5 (sensor documentation
of the setup or a sensor documenta­tion which is at least 6 months old, is not available)
Fault detection rate λdu88 %
du
96 %
The test must be repeated in regular intervals. The time periods depend on the PFD
value according to the sepcications in the
avg
Safety Manual (SIL).
If one of the test points (3.1 / 3.2 / 3.5) could not be terminated
successfully, there is probably an undetected dangerous failure. The recurring function test has failed.
In this case, proof of functional safety can only be provided by moving the level to the switching point.
14 Recurring
36553-EN-100225
function test for VEGAPULS series 60
est protocol - Recurring function test
4 T
4 Test
protocol - Recurring function test
If you copy this protocol, please note the date of the function test, the measurement loop and the sensor serial number on each page.
Specications VEGAPULS
Tester
Measurement loop name (sensor-TAG)
Sensor type
Serial number of the sensor
Software version
Required plant situation (according to chapter 2.4) maintained
Safety-instrumented system (SIS) yes
SIL activated yes
Medium or product group
Date of the setup (sensor documentation)
Date of the last function test (if carried out)
Test 3.1 - Restart of the sensor
Measured value before switching oCurrent value min. in mA Current value max. in mA
Measured value after switching on
again
Dierence of the current values Current value min. in mA Current value max. in mA
Duration of the inspection in s
Test result Min. and max. values
Current value min. in mA Current value max. in mA
Deviation 2 %
(test point successful)
Deviation > 2 %
(test point not successful)
Test 3.2 - Verication of the current output
Lower simulation value (4 mA) Indication mA-meter in mA
Intermediate result 2 % (≤0.32 mA) Corresponds Does not correspond
Upper simulation value (20 mA) Indication mA-meter in mA
Intermediate result 2 % (≤0.32 mA) Corresponds Does not correspond
Test result - Total Corresponds Does not correspond
36553-EN-100225
Recurring
function test for VEGAPULS series 60 15
est protocol - Recurring function test
4 T
Sensor setup available
or Sensor documentation (at least 6
months old) is available
documentation of the
File name:
Parameters correspond
Parameters do not correspond -
however checked on correctness
Parameters do not correspond -
File name:
File name:
deviation not acceptable
Sensor documentation of the setup not available
Parameters checked on correct- ness and saved (recheck after 6
File name:
months necessary)
Test result All parameters correct or
Parameters not correct
checked on correctness
Tester:
Test 3.4 - Verication of the echo data
No false echoes or echoes with
Corresponds Does not correspond
useful echo probability 0 %
Corresponding amplitude In the tolerance range Not in the tolerance range
Test result Corresponds Does not correspond
Test 3.5 - Verication of the sensor reaction
Reduction of the level Distance value [m] increases Other reaction
Increase of the level Distance value [m] decreases Other reaction
Test result Corresponding sensor reaction Sensor reaction does not corre-
spond
Summary 96 % (λ
)
du
Date Signature
Test 3.1 / 3.2 / 3.3 / 3.4 / 3.5 All ve test points passed One or several test points did not
pass
Summary 88 % (λ
)
du
Date Signature
16 Recurring
function test for VEGAPULS series 60
36553-EN-100225
est protocol - Recurring function test
4 T
Summary 88 %
Test 3.1 / 3.2 / 3.5 All three test points passed One or several test points did not
)
du
pass
Date ___________________________________
Signature ___________________________________
36553-EN-100225
Recurring
function test for VEGAPULS series 60 17
est protocol - Recurring function test
4 T
18 Recurring
36553-EN-100225
function test for VEGAPULS series 60
est protocol - Recurring function test
4 T
36553-EN-100225
Recurring
function test for VEGAPULS series 60 19
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com
www.vega.com
Printing date:
statements concerning scope of delivery, application,
All
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2010
Subject to change without prior notice 36553-EN-100225
Loading...