Vega VEGAPULS 61, PS60KF, PS60KL Operating Instructions Manual

Page 1
Operating Instructions
VEGAPULS 61
Probus PA
Document ID: 28444
Page 2
2
Contents
28444-EN-141215
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbols used................................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use ................................................................................................................ 5
2.3 Warning about incorrect use ............................................................................................. 5
2.4 General safety instructions ............................................................................................... 5
2.5 Safety label on the instrument .......................................................................................... 5
2.6 CE conformity ................................................................................................................... 6
2.7 Fulllment of NAMUR recommendations ......................................................................... 6
2.8 FCC/IC conformity (only for USA/Canada) ....................................................................... 6
2.9 Safety instructions for Ex areas ........................................................................................ 6
2.10 Environmental instructions ............................................................................................... 6
3 Product description
3.1 Conguration .................................................................................................................... 8
3.2 Principle of operation...................................................................................................... 10
3.3 Operation ....................................................................................................................... 11
3.4 Packaging, transport and storage ................................................................................... 11
4 Mounting
4.1 General instructions ....................................................................................................... 12
4.2 Mounting preparations, mounting strap .......................................................................... 14
4.3 Mounting instructions ..................................................................................................... 14
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 22
5.2 Connection procedure .................................................................................................... 23
5.3 Wiring plan, single chamber housing.............................................................................. 24
5.4 Wiring plan, double chamber housing ............................................................................ 25
5.5 Wiring plan, double chamber housing Ex d .................................................................... 27
5.6 Wiring plan - version IP 66/IP 68, 1 bar ........................................................................... 29
5.7 Switch-on phase............................................................................................................. 30
6 Set up with the display and adjustment module PLICSCOM
6.1 Short description ............................................................................................................ 31
6.2 Insert display and adjustment module ............................................................................ 31
6.3 Adjustment system ......................................................................................................... 32
6.4 Setup steps .................................................................................................................... 33
6.5 Menu schematic ............................................................................................................. 38
6.10 Saving the parameter adjustment data ........................................................................... 40
7 Set up with PACTware and other adjustment programs
7.1 Connect the PC .............................................................................................................. 41
7.2 Parameter adjustment with PACTware ............................................................................ 42
7.3 Parameter adjustment with PDM .................................................................................... 43
7.4 Saving the parameter adjustment data ........................................................................... 43
8 Maintenanceandfaultrectication
Page 3
3
Contents
28444-EN-141215
8.1 Maintenance .................................................................................................................. 44
8.2 Rectify faults ................................................................................................................... 44
8.3 Exchanging the electronics module ................................................................................ 45
8.4 Software update ............................................................................................................. 46
8.5 How to proceed if a repair is needed .............................................................................. 47
9 Dismount
9.1 Dismounting steps.......................................................................................................... 48
9.2 Disposal ......................................................................................................................... 48
10 Supplement
10.1 Technical data ................................................................................................................ 49
10.2 Probus PA ..................................................................................................................... 56
10.3 Dimensions .................................................................................................................... 60
Supplementary documentation Information:
Supplementary documents appropriate to the ordered version come with the delivery. You can nd them listed in chapter "Product descrip- tion".
Instructions manuals for accessories and replacement parts Tip:
To ensure reliable setup and operation of your VEGAPULS 61, we of­fer accessories and replacement parts. The corresponding documen-
tations are:
27835 - Display and adjustment module PLICSCOM
32628 - Interface adapter VEGACONNECT
43814 - Externe Anzeige VEGADIS 81
34296 - Protective cover
30176 - Electronics module VEGAPULS series 60
Editing status: 2014-12-08
Page 4
4
1 About this document
28444-EN-141215
1 About this document
1.1 Function
This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist personnel. The contents of this manual should be made available to these personnel and put into practice by them.
1.3 Symbols used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
SIL applications
This symbol indicates instructions for functional safety which must be taken into account particularly for safety-relevant applications.
List
The dot set in front indicates a list with no implied sequence.
→
Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
Page 5
5
2 For your safety
28444-EN-141215
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.
2.2 Appropriate use
VEGAPULS 61 is a sensor for continuous level measurement. You can nd detailed information about the area of application in
chapter "Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of stand­ard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specic installation standards as well as all prevailing safety regulations and accident prevention rules.
Depending on the model, the emitting frequencies of all radar sensors are either in the C or K band range. The low transmitting power lies far below the internationally permitted limit values. When the instrument is used correctly, it presents no danger to human health. It may be operated without restriction outside of closed metallic vessels.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free opera­tion of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula-
tions.
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be observed.
Page 6
6
2 For your safety
28444-EN-141215
2.6 CE conformity
The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the CE Certicate of Conformity in the download section of our homepage.
2.7 FulllmentofNAMURrecommendations
NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom­mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring transducers
NE 53 – Compatibility of eld devices and display/adjustment
components
For further information see www.namur.de.
2.8 FCC/ICconformity(onlyforUSA/Canada)
VEGAPULS sensors with all antenna versions are FCC/IC approved. Modications not expressly approved by VEGA will lead to expiry of
the operating licence according to FCC/IC. VEGAPULS 61 is in conformity with part 15 of the FCC directives and
fullls the RSS-210 regulations. Note the corresponding regulations for operation:
The instrument must not cause any interfering emissions
The device must be insensitive to interfering immissions, including those that may cause undesirable operating conditions
According to chapter "Dimensions" of this operating instructions manual, the instrument is designed for operation with an antenna with a max. amplication of 33 dB. The instrument must not be operated with antennas not listed therein or those having an amplication of more than 33 dB. The required antenna impedance is 50 Ω.
2.9 Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
2.10 Environmental instructions
Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according
to DIN EN ISO 14001.
Page 7
7
2 For your safety
28444-EN-141215
Please help us fulll this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
Page 8
8
3 Product description
28444-EN-141215
3 Product description
3.1 Conguration
The VEGAPULS 61 radar sensor is available in two electronics ver­sions:
Standard electronics type PS60KP
Electronics with increased sensitivity type PS60KK
The respective version can be determined by means of the type label on the electronics.
The electronics version inuences the CE conformity, the factory set­ting for the medium selection and vessel form, the accuracy as well as the approvals of VEGAPULS 61. The dierences are listed in the respective sections of this operating instructions manual.
The scope of delivery encompasses:
VEGAPULS 61 radar sensor
Documentation
this operating instructions manualOperating instructions manual 27835 "Display and adjustment
module PLICSCOM" (optional)
Supplementary instructions manual 31708 "Heating for display
and adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)if necessary, further certicates
The VEGAPULS 61 consists of the components:
Process tting with encapsulated antenna system
Housing with electronics, optionally available with plug connector, optionally available with connection cable
Housing cover, optionally available with display and adjustment module PLICSCOM
The components are available in dierent versions.
Versions
Scope of delivery
Constituent parts
Page 9
9
3 Product description
28444-EN-141215
3
2
1
Fig. 1: VEGAPULS 61 - threaded version with encapsulated antenna system and plastic housing
1 Housing cover with integrated PLICSCOM (optional) 2 Housing with electronics
3 Processttingwithencapsulatedantennasystem
3
4
2
1
Fig. 2: VEGAPULS 61 - threaded version with plastic horn antenna and plastic housing
1 Housing cover with integrated PLICSCOM (optional) 2 Housing with electronics
3 Processttingwithplastichornantenna
The type label contains the most important data for identication and
use of the instrument:
Instrument type
Type label
Page 10
10
3 Product description
28444-EN-141215
Article and serial number device
Article numbers, documentation
Technical data: Approvals, antenna type, process tting, process seal/temperature, signal output, voltage supply, protection, protec­tion class
With the serial number, you can access the delivery data of the instru­ment via www.vega.com, "VEGA Tools" and "serial number search". You can nd the serial number on the inside of the instrument as well as on the type label on the outside.
3.2 Principle of operation
VEGAPULS 61 is a radar sensor in K-band (emitting frequency ap­prox. 26 GHz) for continuous level measurement.
The version with encapsulated antenna system is particularly suit­able for level measurement of aggressive liquids in small vessels.
The version with plastic horn antenna is particularly suitable for ow measurement in open umes or gauge measurement in bodies of
water.
The electronics with increased sensitivity enables the use of VE­GAPULS 61 also in applications with very poor reective properties or products with low εr value.
The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reected by the product and received by the antenna as echoes. The transit time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropri­ate output signal and outputted as measured value.
Power supply via the Probus DP/PA segment coupler or VEGALOG 571 EP cards. A two-wire cable according to Probus specication serves as carrier of both power and digital data transmission for mul­tiple sensors. The instrument prole of VEGAPULS 61 corresponds to prole specication version 3.0.
The GSD (instrument master les) and bitmap les necessary for planning your Probus-DP-(PA) communication network are available from the download section on the VEGA homepage www.vega.com under "Services-Downloads-Software-Probus". There you can also nd the appropriate certicates. In a PDM environment, an EDD (Electronic Device Description) is also required to enable the full range of sensor functions (also available as a download).A CD with the appropriate les can be ordered via e-mail under info@de.vega. com or by phone from one of the VEGA agencies under the order number "DRIVER.S".
The backlight of the display and adjustment module is powered by the sensor. Prerequisite is a certain level of operating voltage.
The data for power supply are specied in chapter "Technical data". The optional heating requires its own operating voltage. You can nd
detailed information in the supplementary instructions manual "Heat-
Area of application
Functional principle
Power supply and bus communication
GSD/EDD
Page 11
11
3 Product description
28444-EN-141215
ing for display and adjustment module". This function is generally not available for approved instruments.
3.3 Operation
The instrument can be adjusted with the following adjustment media:
With the display and adjustment module
with the suitable VEGA DTM in conjunction with an adjustment software according to the FDT/DTM standard, e.g. PACTware and
PC
with the adjustment program PDM
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement ­Technical data - Ambient conditions"
Relative humidity 20 … 85 %
Packaging
Transport
Transport inspection
Storage
Storage and transport temperature
Page 12
12
4 Mounting
28444-EN-141215
4 Mounting
4.1 General instructions
Select an installation position you can easily reach for mounting and connecting as well as later retrotting of a display and adjustment module. The housing can be rotated by 330° without the use of any tools. You can also install the display and adjustment module in four dierent positions (each displaced by 90°).
Warning:
With threaded versions, the housing must not be used to screw in the instrument! Applying tightening forces on the housing can damage its internal parts.
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain o. This ap­plies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
Fig. 3: Measures against moisture ingress
The reference plane for the measuring range of the sensor depends
on the antenna version. With encapsulated antenna system, the reference plane is the seal
surface of the thread.
Installation position
Screwing in
Moisture
Measuring range
Page 13
13
4 Mounting
28444-EN-141215
1 3
4
2
100%
0%
Fig. 4: Measuring range (operating range) and max. measuring distance with encapsulated antenna system
1 full 2 empty (max. measuring distance) 3 Measuring range 4 Reference plane
With the Plastic horn antenna, the reference plane is the contact surface on the side of the focusing lens. With the version with adapter ange, the reference plane is the lower side of the ange.
Information:
If the medium reaches the antenna, buildup can form on it and cause faulty measurements later on.
The emitted radar impulses of VEGAPULS 61 are electromagnetic
waves. The polarisation plane is the direction of the electrical share. Their position is marked on the instrument.
1
Fig. 5: Position of the polarisation plane with VEGAPULS 61 with encapsulated antenna system
1 Marking hole
Polarisation plane
Page 14
14
4 Mounting
28444-EN-141215
1
Fig. 6: Position of the polarisation plane with VEGAPULS 61 with plastic horn antenna
1 Marking bars
Make sure that all parts of the instrument coming in direct contact with the process, especially the sensor element, process seal and process tting, are suitable for the existing process conditions, such as process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the nameplate.
4.2 Mounting preparations, mounting strap
The optionally available mounting strap is used to fasten the radar sensor over open vessels or umes. It is suitable for mounting on walls, ceilings or brackets. The strap is supplied unassembled and must be screwed to the sensor with the three hexagon socket screws M5 x 10 and spring washers before setup. Max. torque see chapter "Technical data". Necessary tools: Allen wrench size 4.
Screwing on is possible in two ways. Depending on the selected ver­sion, the sensors can be swivelled in the strap as follows:
Single chamber housing
180° innitely variablein three steps 0°, 90° and 180°
Double chamber housing
90° innitely variablein two steps 0° and 90°
4.3 Mounting instructions
When mounting the sensor, keep a distance of at least 200 mm (7.874 in) to the vessel wall. If the sensor is installed in the center of dished or round vessel tops, multiple echoes can arise. These can, however, be suppressed by an appropriate adjustment (see chapter "Setup").
If you cannot maintain this distance, you should carry out a false signal storage during setup. This applies particularly if buildup on the
Suitability for the process conditions
Installation position
Page 15
15
4 Mounting
28444-EN-141215
vessel wall is expected. In such cases, we recommend repeating the false signal storage at a later date with existing buildup.
> 200 mm
(7.87
")
Fig. 7: Mounting on round vessel tops 1 Reference plane
2 Vessel center or symmetry axis
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible down to the lowest point of the vessel bottom.
Fig. 8: Vessel with conical bottom
Do not mount the instruments in or above the lling stream. Make sure that you detect the product surface, not the inowing product.
Inowingmedium
Page 16
16
4 Mounting
28444-EN-141215
Fig.9:Inowingliquid
The socket piece should be dimensioned in such a way that the an­tenna end protrudes at least 10 mm (0.4 in) out of the socket.
ca. 10 mm
Fig. 10: Recommended socket mounting
If the reective properties of the medium are good, you can mount VEGAPULS 61 on sockets which are higher than the length of the antenna. You will nd recommended values for socket heights in the following illustration. The socket end should be smooth and burr-free, if possible also rounded. After installation you must carry out a false
echo storage.
Socket
Page 17
17
4 Mounting
28444-EN-141215
d
h
Fig. 11: Deviating socket dimensions
In liquids, direct the sensor as perpendicular as possible to the prod­uct surface to achieve optimum measurement results.
Fig.12:Alignmentinliquids
The mounting location of the radar sensor should be a place where no
other equipment or xtures cross the path of the microwave signals. Vessel installations such as for example, ladders, limit switches, heat-
ing spirals, struts etc. can cause false echoes that interfere with the useful echo. Make sure when planning your measuring site that the radar signals have a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal baes above the installations scatter the radar signals and prevent direct interfering reections.
Fig.13:Coverat,large-areaproleswithdeectors
Sensor orientation
Vessel installations
Page 18
18
4 Mounting
28444-EN-141215
If there are agitators in the vessel, a false signal storage should be carried out with the agitators in motion. This ensures that the interfer­ing reections from the agitators are saved with the blades in dierent
positions.
Fig. 14: Agitators
Through the action of lling, stirring and other processes in the vessel, dense foams which considerably damp the emitted signals may form on the product surface.
If foams lead to measurement errors, you should use the biggest pos­sible radar antennas and low frequency radar sensors (C-band).
As an alternative, sensors with guided microwave can be used. These are unaected by foam generation and are best suited for such ap­plications.
By using a standpipe, the inuence of vessel installations and turbu­lence can be excluded. Under these prerequisites, the measurement of products with low dielectric constant (from 1.6) is possible.
Note:
Measurement in a standpipe is not recommended for extremely adhesive products.
Surge or bypass tubes must extend all the way down to the requested min. level, as measurement is only possible within the tube.
Surge pipe
Make sure you provide the necessary upper vent hole in the surge pipe. The hole must be aligned so that it and the polarisation marking on the sensor are in the same plane (see illustration: "Pipe antenna system in a tank").
Agitators
Foam generation
Measurement in the standpipe (surge or by-
passtube)
Page 19
19
4 Mounting
28444-EN-141215
max.
min.
2
1
Fig. 15: Pipe antenna system in a tank. The vent hole in the surge pipe must be in one plane with the polarisation marking on the sensor.
1 Marking of the polarisation direction 2 Vent hole max. ø 5 mm (0.2 in)
If possible, the antenna diameter of the sensor should correspond to the inner diameter of the tube. With VEGAPULS 61 this is approx. 40 mm (1.575 in). The sensor can be used with tube diameters be­tween 40 … 80 mm (1.575 … 3.15 in).
Bypass tube
As an alternative to the surge pipe in the vessel, a tube system out­side of the vessel is possible as a bypass tube. Select during setup the function "Bypass tube".
Align the sensor in such a way that the polarisation marking on the process tting is in the same plane as the tube holes or the tube con­nection openings (see illustration: "VEGAPULS in a bypass tube").
Page 20
20
4 Mounting
28444-EN-141215
100%
0%
1
> 500 mm
Fig. 16: VEGAPULS 61 in a bypass tube. The polarisation marking on the
processttingmustbeinoneplanewiththetubeholesorthetubeconnection
openings. 1 Marking of the polarisation direction
When the sensor is mounted on a bypass tube, the distance from VE­GAPULS 61 to the upper tube connection should be approx. 500 mm (19.69 in) or more. In case of extremely rough tube inner walls, you should use an inserted tube (tube in tube) or a radar sensor with tube
antenna.
The short examples are meant to give only introductory information on the ow measurement. Planning information is available from ume manufacturers and in special literature.
Meas.wallwithrectangularume
In general, the following points must be observed:
Install the sensor on the headwater side
Installation in the centre of the ume and vertical to the liquid surface
Distance to the overfall orice
Distance of orice opening above ground
Min. distance of the orice opening to bottom water
Min. distance of the sensor to max. storage level
Flow measurement
Page 21
21
4 Mounting
28444-EN-141215
h
max
d
min
≥ 2 mm x h
max
90°
4
3 ... 4 h
max
≥ 50 mm
90°
2 3
1
Fig.17:Flowmeasurementwithrectangularume:d=min.distanceofthesen­sor; h
max.
=max.llingoftherectangularume
1 Overoworice(sideview)
2 Headwater 3 Tail water
4 Overfallorice(viewfrombottomwater)
Khafagi-Venturiume
In general, the following points must be observed:
Installation of the sensor at the input side
Installation in the centre of the ume and vertical to the liquid surface
Distance to the Venturi ume
Min. distance of the sensor to max. storage level
2
3 ... 4 x h
max
90°
h
max
d
1
B
Fig.18:FlowmeasurementwithKhafagi-Venturiume:d=Min.distancetosen­sor; h
max.
=max.llingoftheume;B=tightestconstrictionintheume
1 Position sensor
2 Venturiume
In general, the following points must be observed:
Installation of the sensor in a protected area
Installation vertical to the liquid surface
Gauge measurement
Page 22
22
5 Connecting to power supply
28444-EN-141215
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If voltage surges are expected, overvoltage arresters should be installed according to Probus specications
Tip:
We recommend VEGA overvoltage arrester B63-32.
In hazardous areas you must take note of the respective regulations, conformity and type approval certicates of the sensors and power supply units.
Power is supplied via a Probus DP/PA segment coupler or a VEGA­LOG 571 EP input card. The power supply range can dier depending
on the instrument version.
The data for power supply are specied in chapter "Technical data".
Connection is made with screened cable according to the Probus specication. Power supply and digital bus signal are carried over the same two-wire connection cable.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland. If you are using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Please make sure that your installation is carried out according to the Probus specication. In particular, make sure that the termination of the bus is done with appropriate terminating resistors.
On the instrument with cable entry ½ NPT and plastic housing there is a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No grease should be used when screwing the NPT cable gland or steel tube into the threaded insert. Standard grease can contain additives that corrode the connection between threaded insert and housing. This would inuence the stability of the connection and the tightness of the housing.
In systems with potential equalisation, connect the cable screen directly to ground potential at the power supply unit, in the connection box and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation (low impedance).
In systems without potential equalisation, connect the cable screen directly to ground potential at the power supply unit and at the sensor. In the connection box or T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another
Safety instructions
Voltage supply
Connection cable
Cablegland½NPT
Cable screening and grounding
Page 23
23
5 Connecting to power supply
28444-EN-141215
cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interference signals
remains.
The total capacitance of the cable and of all capacitors must not
exceed 10 nF in Ex applications.
Take note of the corresponding installation regulations for Ex applica-
tions. In particular, make sure that no potential equalisation currents ow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.
5.2 Connection procedure
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn­ing it to the left.
3. Loosen compression nut of the cable entry gland
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
6. Lift the opening levers of the terminals with a screwdriver (see following illustration)
7. Insert the wire ends into the open terminals according to the wir­ing plan
Fig. 19: Connection steps 6 and 7
Page 24
24
5 Connecting to power supply
28444-EN-141215
8. Press down the opening levers of the terminals, you will hear the terminal spring closing
9. Check the hold of the wires in the terminals by lightly pulling on
them
10. Connect the screen to the internal ground terminal, connect the outer ground terminal to potential equalisation
11. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable
12. Screw the housing lid back on
The electrical connection is nished.
5.3 Wiring plan, single chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-ia
version.
5
5
5
5
1 2
4
3
Fig. 20: Material versions, single chamber housing 1 Plastic
2 Aluminium 3 Stainless steel, precision casting 4 Stainless steel, electro-polished 5 Filter element for air pressure compensation of all material versions. Blind
plug with version IP 66/IP 68, 1 bar for Aluminium and stainless steel
Housing overview
Page 25
25
5 Connecting to power supply
28444-EN-141215
I²C
Display
12 5678
3
4
1
2
Fig. 21: Electronics and terminal compartment, single chamber housing 1 Plug connector for VEGACONNECT (I²C interface)
2 Spring-loaded terminals for connection of the external indication VEGADIS
61 3 Ground terminal for connection of the cable screen 4 Spring-loaded terminals for voltage supply
I2C
Display
1
12 5 678
Fig. 22: Wiring plan, single chamber housing 1 Voltage supply, signal output
5.4 Wiring plan, double chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-ia
version.
Electronics and terminal compartment
Wiring plan
Page 26
26
5 Connecting to power supply
28444-EN-141215
1 2 3
4
5
Fig. 23: Double chamber housing 1 Housing cover, connection compartment
2 Blind plug or plug M12 x 1 for VEGADIS 61 (optional) 3 Housing cover, electronics compartment 4 Filter element for air pressure compensation 5 Cable gland
1
3
2
Display
12 5678
I²C
Fig. 24: Electronics compartment, double chamber housing 1 Plug connector for VEGACONNECT (I²C interface)
2 Internal connection cable to the connection compartment 3 Terminals for VEGADIS 81
Housing overview
Electronics compartment
Page 27
27
5 Connecting to power supply
28444-EN-141215
1
2
3
Display
12
I2C
Fig. 25: Terminal compartment, double chamber housing 1 Spring-loaded terminals for voltage supply
2 Plug connector for VEGACONNECT (I²C interface) 3 Ground terminal for connection of the cable screen
I2C
1
12
Fig. 26: Wiring plan, double chamber housing 1 Voltage supply, signal output
5.5 Wiring plan, double chamber housing Ex d
Information:
Instruments in Ex d version with hardware revision …- 01 or higher as well as with national approvals such as e.g. according to FM or CSA at a later date.
Terminal compartment
Wiring plan
Page 28
28
5 Connecting to power supply
28444-EN-141215
1 2 3
4
5
Fig. 27: Double chamber housing 1 Housing cover, connection compartment
2 Blind plug or plug M12 x 1 for VEGADIS 61 (optional) 3 Housing cover, electronics compartment 4 Filter element for air pressure compensation 5 Cable gland
1
3
2
Display
12 5678
I²C
Fig. 28: Electronics compartment, double chamber housing 1 Plug connector for VEGACONNECT (I²C interface)
2 Internal connection cable to the connection compartment 3 Terminals for VEGADIS 81
Housing overview
Electronics compartment
Page 29
29
5 Connecting to power supply
28444-EN-141215
1
2
12
Fig. 29: Connection compartment, Ex-d-ia double chamber housing 1 Spring-loaded terminals for power supply and cable screen
2 Ground terminal for connection of the cable screen
1
12
Fig. 30: Wiring plan, Ex-d-ia double chamber housing 1 Voltage supply, signal output
5.6 Wiring plan - version IP 66/IP 68, 1 bar
1
2
Fig. 31: Wire assignment, connection cable 1 brown (+) and blue (-) to power supply or to the processing system
2 Shielding
Terminal compartment
Wiring plan
Wire assignment, con­nection cable
Page 30
30
5 Connecting to power supply
28444-EN-141215
5.7 Switch-on phase
After VEGAPULS 61 is connected to voltage supply or after voltage recurrence, the instrument carries out a self-check for approx. 30 seconds. The following steps are carried out:
Internal check of the electronics
Indication of the instrument type, the rmware as well as the sen-
sor TAGs (sensor designation)
Status byte goes briey to fault value
Then the current measured value will be displayed and the corre­sponding digital output signal will be outputted to the cable.
1)
Switch-on phase
1)
The values correspond to the actual measured level as well as to the settings
already carried out, e.g. default setting.
Page 31
31
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
6 Set up with the display and adjustment
module PLICSCOM
6.1 Short description
The display and adjustment module is used for measured value
display, adjustment and diagnosis. It can be mounted in the following housing versions and instruments:
All sensors of the plics® instrument family, in the single as well as
in the double chamber housing (optionally in the electronics or
connection compartment)
External display and adjustment unit VEGADIS 61
From a hardware version …- 01 or higher of PLICSCOM as well as a hardware version …- 01, 03 or higher of the corresponding sensor, an integrated backlight can be switched on via the adjustment menu. The hardware version is stated on the type label of the PLICSCOM or the sensor electronics.
Note:
You can nd detailed information on adjustment in the operating instructions manual "Display and adjustment module".
6.2 Insert display and adjustment module
The display and adjustment module can be inserted into the sensor
and removed again at any time. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the desired position
on the electronics (you can choose any one of four dierent posi­tions - each displaced by 90°)
3. Press the display and adjustment module onto the electronics
and turn it to the right until it snaps in.
4. Screw housing cover with inspection window tightly back on Disassembly is carried out in reverse order. The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
Function/Conguration
Mount/Dismount display and adjustment module
Page 32
32
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Fig. 32: Insert display and adjustment module
Note:
If you intend to retrot the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
6.3 Adjustment system
1
2
Fig. 33: Display and adjustment elements 1 LC display
2 Indication of the menu item number 3 Adjustment keys
[OK] key:Key functions
Page 33
33
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Move to the menu overviewConrm selected menuEdit parameterSave value
[->] key to select:
Menu changeSelect list entrySelect editing position
[+] key:
Change value of the parameter
[ESC] key:
– Interrupt input – Jump to next higher menu
The sensor is adjusted via the four keys of the display and adjust-
ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
6.4 Setup steps
Before starting the actual parameter adjustment of a Probus PA sensor, the address setting must rst be carried out. You will nd a detailed description in the operating instructions manual of the display and adjustment module or in the online help of PACTware or DTM.
As VEGAPULS 61 is a distance measuring instrument, the distance from the sensor to the product surface is measured. To have the real product level displayed, an allocation of the measured distance to the percentage height must be made. To carry out this adjustment, the distance is entered with full and empty vessel. If these values are not known, an adjustment with the distance values, e.g. 10 % and 90 % is also possible. Starting point for these distance specications is al­ways the seal surface of the thread or ange. With these settings, the real level is calculated. Furthermore the operating range of the sensor is limited from maximum to the required range.
The real product level during this adjustment is not important, be­cause the min./max. adjustment is always carried out without chang­ing the product level. These settings can be made ahead of time without the instrument having to be installed.
Caution:
If there is a separation of liquids with dierent dielectric constant in the vessel, e.g. from condensation, VEGAPULS 61 can detect under certain circumstances only the medium with the higher dielectric
constant.
Keep in mind that interfaces can cause faulty measurements.
Adjustment system
Address setting
Parameter adjustment
Page 34
34
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
If you want to measure the total height of both liquids reliably, please contact our service department or use an instrument specially de­signed for interface measurement.
In the main menu item "Basic adjustment", the individual submenu items should be selected one after the other and provided with the correct parameter values.
Start your parameter adjustment with the following menu items of the basic adjustment:
Proceed as follows:
1. Move from the measured value display to the main menu by
pushing [OK].
Basic adjustment Display
Diagnostics
Service
Info
2. Select the menu item "Basic adjustment" with [->] and conrm with [OK]. Now the menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
5.000m(d)
4.000 m(d)
3. Prepare the % value for editing with [OK] and set the cursor to the requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4. Enter the suitable distance value in m for the empty vessel (e.g. distance from the sensor to the vessel bottom) corresponding to the percentage value.
5. Save the settings with [OK] and move to "Max. adjustment" with [->].
Proceed as follows:
Max. adjustment
100.00 %
=
1.000m(d)
2.000 m(d)
1. Prepare the % value for editing with [OK] and set the cursor to the requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
2. Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. Keep in mind that the max. level must lie below the dead band.
3. Save the settings with [OK] and move to "Medium selection" with [->].
Carry out min. adjustment
Carry out max. adjust­ment
Page 35
35
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Each product has dierent reective properties. In addition, there are various interfering factors which have to be taken into account: agitated product surfaces and foam generation (with liquids); dust generation, material cones and echoes from the vessel wall (with solids). To adapt the sensor to these dierent conditions, you should rst select "Liquid" or "Solid".
Medium
Liquid
Information:
With VEGAPULS 61 with electronics version "Increased safety", "Solid" is preset as factory setting. However, the instrument should be used preferably in liquids. In such cases, the medium selection should be set to "Liquid" during setup.
According to the conductivity and the dielectric constant of liquids, the reection properties can dier considerably. Therefore additional options such as "Solvent", "Chem. mixture" and "Water based" are oered below the menu item Liquid.
With solids, you can also choose between "Powder/Dust", "Granular/ Pellets" or "Ballast/Pebbels".
Through this additional selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with poor reective properties, is considerably increased.
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Apart from the medium, the vessel shape can also inuence the measurement. To adapt the sensor to these measuring conditions, this menu item oers dierent options depending on whether liquid or bulk solid is selected. With "Liquids" these are "Storage tank", "Stilling tube","Open vessel" or "Stirred vessel", with "Solid", "Silo" or "Bunker".
Vessel form
Storage tank
Information:
With VEGAPULS 61 with electronics version "Increased safety", "Solid" is preset as factory setting. However, the instrument should be used preferably in liquids. In such cases, the vessel form should be set to "Storage tank" during setup.
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
A linearisation is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindri­cal or spherical tank - and the indication or output of the volume is
Medium selection
Vessel form
Linearisation curve
Page 36
36
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
required. Corresponding linearisation curves are preprogrammed for these vessels. They represent the correlation between the level per­centage and vessel volume. By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Linearisation curve
Linear
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Caution:
Note the following if the VEGAPULS 61 with corresponding approval is used as part of an overll protection system according to WHG (Water Resources Act):
If a linearisation curve is selected, the measuring signal is no longer necessarily linear to the lling height. This must be considered by the user especially when adjusting the switching point on the limit signal
transmitter.
High sockets or vessel installations, such as e. g. struts or agitators as well as buildup and weld joints on the vessel walls, cause interfering reections which can impair the measurement. A false echo storage detects and marks these false echoes, so that they are no longer taken into account for the level measurement. A false echo memory should be created with low level so that all potential interfering reec­tions can be detected.
False signal suppression
Change now?
Proceed as follows:
1. Move from the measured value display to the main menu by
pushing [OK].
2. Select the menu item "Service" with [->] and conrm with [OK]. Now the menu item "False signal suppression" is displayed.
3. Conrm "False signal suppression - Change now" with [OK] and select in the below menu "Create new". Enter the actual distance from the sensor to the product surface. All false signals in this area are detected by the sensor and saved after conrming with
[OK].
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as a false signal. The level would then no longer be detectable in this area.
False signal suppression
Page 37
37
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
The menu item "Extended setting" oers the possibility to optimise
VEGAPULS 61 for applications in which the level changes very quickly. To do this, select the function "Quick level change > 1 m/min.".
Extended setting
quick level change > 1 m/min.
Note:
Since with the function "Quick level change > 1 m/min." the genera­tion of an average value of the signal processing is considerably reduced, false reections by agitators or vessel installations can cause measured value uctuations. A false echo memory is thus recommended.
This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the display and adjustment module. A description of the function is available in the operating instructions manual "Display and adjustment module".
The following data are read out or written with this function:
Measured value presentation
Adjustment
Medium
Standpipe inner diameter
2)
Vessel form
Damping
Linearisation curve
Sensor-TAG
Displayed value
Scaling unit (Out-Scale unit)
Positions after the decimal point (scaled)
Scaling PA/Out-Scale 4 values
Unit of measurement
Language
The following safety-relevant data are not read out or written:
Sensor address
PIN
Copy sensor data
Copy sensor data?
Basic adjustment
If the "Reset" is carried out, the sensor resets the values of the follow­ing menu items to the reset values (see chart):
3)
Function Resetvalue
Sensor address 126
Extended setting/Quick level change
Copy sensor data
Reset
2)
With standpipe versions.
3)
Sensor-specic basic adjustment.
Page 38
38
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Function Resetvalue
Max. adjustment 0 m(d)
Min. adjustment Meas. range end in m(d)
4)
Medium Liquid
Vessel form not known
Damping 0 s
Linearization Linear
Channel PV lin. %
Sensor-TAG Sensor
Displayed value PA-Out
Extended settings None
Additional PA value Secondary Value 1 %
Out-Scale-Unit %
PV-Out-Scale 0.00 lin-% = 0.0 %
100.0 lin-% = 100 %
Unit of measurement m(d)
The values of the following menu items are not reset to the reset values (see chart) with "Reset":
Function Resetvalue
Sensor address No reset
Language No reset
Default setting
Like basic adjustment, but in addition, special parameters are reset to default values.
5)
Peak value indicator
The min. and max. distance values are reset to the actual value.
Additional adjustment and diagnosis options such as e.g. scaling, simulation or trend curve presentation are shown in the following menu schematic. You will nd a detailed description of these menu items in the operating instructions manual "Display and adjustment module".
6.5 Menu schematic
Information:
Depending on the version and application, the highlighted menu windows may not always be available.
Optional settings
4)
Depending on the sensor type, see chapter "Technical data".
5)
Special parameters are parameters which are set customer-specically on
the service level with the adjustment software PACTware.
Page 39
39
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Basic adjustment
Basic adjustment 1 Display
Diagnostics
Service
Info
Sensor address 1.1
126
Min. adjustment 1.2
0.00 %
=
10.000m(d)
8.000 m(d)
Max. adjustment 1.3
100.00 %
=
1.000m(d)
2.000 m(d)
Medium 1.3
not known
Vessel form 1.4
not known
Linearisation curve 1.6
Linear
Channel 1.7
PV lin. value
Damping 1.8
0 s
Sensor-TAG 1.9
Sensor
Display
Basic adjustment 2
Display
Diagnostics
Service
Info
Displayed value 2.1
PA-Out
Backlight 2.4
Switchedo
Diagnostics
Basic adjustment 3 Display
Diagnostics
Service
Info
Peak value indicator 3.1
Distance min.: 0.234 m(d) Distance max.: 5.385 m(d)
Meas. reliability 3.2
15 dB
Sensor status
OK
Curve selection 3.3
Echo curve
Echo curve 3.4
Presentation of the echo curve
Page 40
40
6 Set up with the display and adjustment module PLICSCOM
28444-EN-141215
Service
Basic adjustment 4 Display
Diagnostics
Service
Info
False signal suppression 4.1
Change now?
Extended setting 4.2
None
Quick level change
Additional PA value 4.3
Secondary Value 1
Out-Scale-Unit 4.4
Volume
hl
PV-Out-Scale 4.5
0 % = 0.0 m³ 100 % = 100 m³
Simulation 4.4
Start simulation
Reset 4.7
Select reset?
Unit of measurement 4.7
m(d)
Language 4.8
Deutsch
Copy sensor data 4.10
Copy sensor data?
PIN 4.11
Enable?
Info
Basic adjustment 5 Display
Diagnostics
Service
Info
Instrument type 5.1
Serial number
12345678
Date of manufacture 5.2
22. November 2010 Software version
3.80
Last change using PC 5.3
22. November 2010
Sensor characteristics 5.4
Display now?
6.10 Saving the parameter adjustment data
We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If VEGAPULS 61 is equipped with a display and adjustment module, the most important data can be read out of the sensor into the display and adjustment module. The procedure is described in the operating instructions manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
If it is necessary to exchange the sensor, the display and adjustment module is inserted into the replacement instrument and the data are written into the sensor under the menu item "Copy sensor data".
Page 41
41
7 Set up with PACTware and other adjustment programs
28444-EN-141215
7 Set up with PACTware and other
adjustment programs
7.1 Connect the PC
3
1
2
Fig. 34: Connection of the PC via VEGACONNECT directly to the sensor 1 USB cable to the PC
2 VEGACONNECT 3 Sensor
1
2
3
4
OPEN
TWIST
USB
LOCK
Fig. 35: Connection via VEGACONNECT externally 1 I²C bus (com.) interface on the sensor
2 I²C connection cable of VEGACONNECT 3 VEGACONNECT 4 USB cable to the PC
Necessary components:
VEGAPULS 61
PC with PACTware and suitable VEGA DTM
VEGACONNECTdirectly
on the sensor
VEGACONNECTexter­nally
Page 42
42
7 Set up with PACTware and other adjustment programs
28444-EN-141215
VEGACONNECT
Power supply unit or processing system
7.2 Parameter adjustment with PACTware
For parameter adjustment of the instrument via a Windows PC, the conguration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The latest PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
Fig. 36: Example of a DTM view
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value
Prerequisites
Standard/Full version
Page 43
43
7 Set up with PACTware and other adjustment programs
28444-EN-141215
and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega.com/downloads and "Software". The full version is avail­able on CD from the agency serving you.
7.3 Parameter adjustment with PDM
For VEGA sensors, instrument descriptions for the adjustment program PDM are available as EDD. The instrument descriptions are already implemented in the current version of PDM. For older versions of PDM, a free-of-charge download is available via Internet.
Go via www.vega.com and "Downloads" to the item "Software".
7.4 Saving the parameter adjustment data
It is recommended to document or save the parameter adjustment data. That way they are available for multiple use or service purposes.
The VEGA DTM Collection and PACTware in the licensed, profession­al version provide suitable tools for systematic project documentation and storage.
Page 44
44
8 Maintenance and fault rectication
28444-EN-141215
8 Maintenanceandfaultrectication
8.1 Maintenance
If the instrument is used properly, no special maintenance is required in normal operation.
8.2 Rectifyfaults
The operator of the system is responsible for taking suitable meas­ures to rectify faults.
VEGAPULS 61 oers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
Sensor
Process
Voltage supply
Signal processing
The rst measures to be taken are to check the output signals as well as to evaluate the error messages via the display and adjustment module. The procedure is described below. Further comprehensive diagnostics can be carried out on a PC with the software PACTware and the suitable DTM. In many cases, the causes can be determined and the faults rectied this way.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is manned 7 days a week round-the-clock. Since we oer this service worldwide, the support is only available in the English language. The service is free, only standard call charges are incurred.
The following table describes possible errors and helps to remove
them:
Error Cause Rectication
When an addi­tional instrument is connected, the segment fails.
Max. supply current of the seg­ment coupler exceeded
Measure the current consumption, re­duce size of segment
Wrong pres-
entation of the measured value in
Simatic S5
Simatic S5 cannot interpret the num-
ber format IEEE of the measured value
Insert converting component from Sie-
mens
In Simatic S7 the
measured value is always present­ed as 0
Only four bytes are consistently loaded in the PLC
Use function component SFC 14 to load 5 bytes consistently
Reactionwhenmalfunc­tions occur
Causes of malfunction
Faultrectication
24 hour service hotline
CheckingProbusPA
Page 45
45
8 Maintenance and fault rectication
28444-EN-141215
Error Cause Rectication
Measured val­ue on the display and adjustment module does not correspond to the value in the PLC
The menu item
"Display - Display value" is not set to
"PA-Out"
Check values and correct, if neces-
sary
No connection be­tween PLC and
PA network
Incorrect adjust­ment of the bus parameter and the segment coupler­dependent baud
rate
Check data and correct, if necessary
Instrument does not appear during
connection setup
Probus DP cable pole-reversed
Check cable and correct, if necessary
Incorrect termi­nation
Check termination at the beginning and end points of the bus and termi­nate, if necessary, according to the specication
Instrument not
connected to the segment, double assignment of an address
Check and correct, if necessary
In Ex applications, the regulations for the wiring of intrinsically safe circuits must be observed.
Error code Cause Rectication
E013 no measured value
available
Sensor in boot phaseSensor does not nd an echo, e.g.
due to faulty installation or wrong parameter adjustment
E017 Adjustment span too
small
– Carry out a fresh adjustment and
increase the distance between min. and max. adjustment
E036 no operable sensor
software
– Carry out a software update or send
instrument for repair
E041, E042, E043
Hardware error, elec­tronics defective
– Exchange the instrument or send it
in for repair
E113 Communication con-
ict
– Exchange the instrument or send it
in for repair
Depending on the reason for the fault and the measures taken, the steps described in chapter "Set up" may have to be carried out again.
8.3 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user. In Ex applications, only instruments and electronics modules with ap-
propriate Ex approval may be used.
Error messages via the display and adjustment module
Reactionafterfaultrecti­cation
Page 46
46
8 Maintenance and fault rectication
28444-EN-141215
If there is no electronics module available on site, one can be ordered from the VEGA agency serving you.
The new electronics module must be loaded with the settings of the
sensor. These are the options:
At the factory by VEGA
Or on site by the user
In both cases, the sensor serial number is necessary. The serial num­bers are stated on the type label of the instrument, inside the housing or on the delivery note.
Information:
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").
The electronics modules are adapted to the respective sensor and distinguish also in the signal output or power supply.
8.4 Software update
The software version of VEGAPULS 61 can be determined as follows:
on the type label of the electronics
Via the display and adjustment module
via PACTware
You can view all software histories on our website www.vega.com. Make use of this advantage and get registered for update information via e-mail.
The following components are required to update the sensor soft-
ware:
Sensor
Voltage supply
VEGACONNECT
PC with PACTware
Current sensor software as le
At "www.vega.com/downloads" go to "Software". Select under "plics sensors and instruments", "Firmware updates" the respective instru­ment series and software version. Load the zip le via the right mouse key with "Save target as" e.g. on the desktop of your PC. Move with the right mouse key to the folder and select "Extract all". Save the extracted les, for example on the desktop.
Connect the signal conditioning instrument to power supply and provide the connection from the PC to the instrument via the interface adapter. Start PACTware and go via the menu "Project" to the VEGA project assistant. Select "USB" and "Set instruments online". Activate the project assistant with "Start". The assistant establishes the con­nection automatically and opens the parameter adjustment window "Sensor # online parameter adjustment". Connect this parameter adjustment window before you carry out further steps.
Sensor serial number
Assignment
Load sensor software to PC
Prepare update
Page 47
47
8 Maintenance and fault rectication
28444-EN-141215
Select with the right mouse key the sensor in the project and go to "Additional function". Then click to "Software update". The window "Sensor # software update" opens. PACTware checks now the sensor data and displays the actual hardware and software version of the sensor. This takes approximately 60 s.
Push the button "Update software" and select the previously extracted hex le. Then the software update can be started. The additional les are installed automatically. Depending on the sensor, this procedure can last up to 1 h. Then the message appears ""Software update suc- cessfully executed".
8.5 How to proceed if a repair is needed
You can nd a repair form as well as detailed information on how to proceed at www.vega.com/downloads and "Formsandcerticates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data sheet outside on the packaging
Please contact the agency serving you to get the address for the return shipment. You can nd the agency on our home page
www.vega.com.
Load software into sen­sor
Page 48
48
9 Dismount
28444-EN-141215
9 Dismount
9.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, cor­rosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.
9.2 Disposal
The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
Page 49
49
10 Supplement
28444-EN-141215
10 Supplement
10.1 Technical data
General data
316L corresponds to 1.4404 or 1.4435, 304 corresponds to 1.4301 Materials, wetted parts with encapsulated antenna system
Ʋ Threaded tting G1½ A and 1½ NPT PVDF Ʋ Hygienic tting Adapter of 316L Ʋ Antenna PVDF Ʋ Seal thread FKM Ʋ Seal, slotted nut FKM
Materials, wetted parts with plastic horn antenna
Ʋ Horn antenna PBT-GF30 Ʋ Focussing lense PP Ʋ Adapter ange PPH Ʋ Seal, adapter ange FKM
Materials, non-wetted parts
Ʋ Compression ange PPH Ʋ Mounting strap 316L Ʋ Fixing screws, mounting strap 316L Ʋ Fixing screws, adapter ange 304 Ʋ Housing Plastic PBT (polyester), Alu die-casting, powder-coated,
316L
Ʋ Seal between housing and housing
cover
NBR (stainless steel housing), silicone (Alu/plastic hous­ing)
Ʋ Inspection window in housing cover
for PLICSCOM
Polycarbonate (UL-746-C listed)
Ʋ Ground terminal 316Ti/316L
Max. torque, mounting screws - strap on
the sensor housing
4 Nm
Weight depending on process tting and housing material
0.7 … 3.4 kg (1.543 … 7.496 lbs)
Output variable
Output signal digital output signal, format according to IEEE-754 Cycle time min. 1 s (dependent on the parameter setting) Sensor address 126 (default setting) Current value 10 mA, ±0.5 mA Damping (63 % of the input variable) 0 … 999 s, adjustable Met NAMUR recommendation NE 43 Resolution, digital > 1 mm (0.039 in)
Page 50
50
10 Supplement
28444-EN-141215
Input variable
Measured variable distance between process tting and product surface Min. distance from antenna end 50 mm (1.969 in)
6)
Recommended measuring range, encap­sulated antenna system
up to 10 m (32.81 ft)
Recommended measuring range plastic
horn antenna
up to 20 m (65.62 ft)
Referenceconditionstomeasuringaccuracy(accordingtoDINEN60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Other reference conditions
Ʋ Reector ideal reector, e.g. metal plate 2 x 2 m Ʋ False reections Biggest false signal, 20 dB smaller than the useful signal
Characteristics and performance data
Measuring frequency K-band (26 GHz technology) Meas. interval approx. 1 s Beam angle -3 dB
7)
Ʋ Encapsulated antenna system 22° Ʋ Plastic horn antenna 10°
Step response or adjustment time
8)
< 1 s (dependent on the parameter settings)
Max. level change Adjustable up to 1 m/min. (dependent on the parameter
settings)
Max. emitted HF power of the antenna system
Ʋ Pulse peak power < 2 mW Ʋ Pulse duration < 2 ns Ʋ Average power < 5 µW Ʋ Average power with 1 m distance < 200 nW/cm
2
Max. emitted HF power of the antenna system - Version with increased sensitivity
Ʋ Pulse peak power < 10 mW Ʋ Pulse duration < 2 ns Ʋ Average power < 25 µW Ʋ Average power with 1 m distance < 1 µW/cm
2
Measuring accuracy
Resolution, general < 1 mm (0.039 in)
Deviation
9)
see diagrams
6)
For products with low dielectric gure up to 50 cm (19.69 in).
7)
Corresponds to the range with 50 % of the emitted power
8)
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
9)
Incl. non-linearity, hysteresis and non-repeatability.
Page 51
51
10 Supplement
28444-EN-141215
0,5 m (1.6 ft) 10 m (32.80 ft)
10 mm (0.394 in)
5 mm (0.197 in)
- 5 mm (- 0.197 in)
- 10 mm (- 0.394 in)
0
Fig. 37: Deviation VEGAPULS 61 with encapsulated antenna system
30 mm (1.180 in)
15 mm (0.590 in)
0
-15 mm (- 0.590 in)
- 30 mm (- 1.180 in)
1,0 m (3.280 ft)
10 m (32.80 ft)
Fig. 38: Deviation VEGAPULS 61 with encapsulated antenna system and increased sensitivity
0,5 m (1.6 ft) 20 m (65.60 ft)
10 mm (0.394 in)
5 mm (0.197 in)
- 5 mm (- 0.197 in)
- 10 mm (- 0.394 in)
0
Fig. 39: Deviation VEGAPULS 61 with plastic horn antenna
30 mm (1.180 in)
15 mm (0.590 in)
0
-15 mm (- 0.590 in)
- 30 mm (- 1.180 in)
1,0 m (3.280 ft)
20 m (65.60 ft)
Fig. 40: Deviation VEGAPULS 61 with plastic horn antenna and increased sensitivity
Page 52
52
10 Supplement
28444-EN-141215
Deviation in characteristics and characteristics
10)
Average temperature coecient of the zero signal (temperature error)
0.03 %/10 K
Resolution, general > 1 mm (0.039 in) Measuring frequency K-band (26 GHz technology) Beam angle encapsulated antenna
system at -3 dB
22°
Beam angle plastic horn antenna at
-3 dB
10°
Adjustment time
11)
> 1 s (dependent on the parameter setting)
Received average emitted power reaching an object directly in front of the antenna
Ʋ Distance 1 m (39.37 in) 108 nW per cm² (108 x 10-9 W/cm²) or 696.8 nW pro in²
(696.8 x 10-9 W/in²)
Ʋ Distance 5 m (196.9 in) 4.3 nW per cm² (4.3 x 10-9 W/cm²) or 27.7 nW per in²
(27.7 x 10-9 W/in²)
Accuracy see diagram
0,5 m 10 m
10 mm
5 mm
- 5 mm
- 10 mm
Fig. 41: Accuracy VEGAPULS 61 with encapsulated antenna system
1,0 m 10 m
30 mm
15 mm
-15 mm
-30 mm
Fig. 42: Accuracy VEGAPULS 61 with encapsulated antenna system and increased sensitivity
10)
Relating to the nominal measuring range, incl. hysteresis and repeatability.
11)
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
Page 53
53
10 Supplement
28444-EN-141215
0,5 m 20 m
10 mm
5 mm
-5 mm
-10 mm
Fig. 43: Accuracy VEGAPULS 61 with plastic horn antenna
1,0 m 20 m
30 mm
15 mm
-15 mm
-30 mm
Fig. 44: Accuracy VEGAPULS 61 with plastic horn antenna and increased sensitivity
Inuenceoftheambienttemperaturetothesensorelectronics
12)
Average temperature coecient of the zero signal (temperature error)
0.03 %/10 K
Ambient conditions
Ambient, storage and transport tempera­ture
-40 … +80 °C (-40 … +176 °F)
Process conditions
Vessel pressure
Ʋ Encapsulated antenna system -100 … 300 kPa/-1 … 3 bar (-14.5 … 43.5 psi) Ʋ Plastic horn antenna -100 … 200 kPa/-1 … 2 bar (-14.5 … 29.0 psig)
Process temperature (measured on the process tting)
-40 … +80 °C (-40 … +176 °F)
Vibration resistance
13)
Ʋ Encapsulated antenna system mechanical vibrations up to 4 g in the frequency range
5 … 200 Hz
Ʋ Plastic horn antenna with compres-
sion or adapter ange
mechanical vibrations up to 2 g in the frequency range 5 … 200 Hz
12)
Relating to the nominal measuring range, in the temperature range -40 … +80 °C .
13)
Tested according to the guidelines of German Lloyd, GL directive 2.
Page 54
54
10 Supplement
28444-EN-141215
Ʋ Plastic horn antenna with mounting
strap
mechanical vibrations up to 1 g in the frequency range 5 … 200 Hz
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
14)
Ʋ Single chamber housing – 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
1 x closing cap M20 x 1.5; 1 x blind plug M20 x 1.5
or:
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
1␣x plug (depending on the version), 1␣x blind stopper
M20␣x␣1.5
Ʋ Double chamber housing – 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5; 1 x blind plug M16 x 1.5 or optionally available with 1 x plug M12 x 1 for external display and adjustment unit
or:
1 x closing cap ½ NPT, 1 x blind plug ½ NPT, 1 x blind
plug M16 x 1.5 or optionally 1 x plug M12 x 1 for exter­nal display and adjustment unit
or:
1 x plug (depending on the version), 1 x blind plug
M20 x 1.5; 1 x blind plug M16 x 1.5 or optionally avail­able with 1 x plug M12 x 1 for external display and adjustment unit
Spring-loaded terminals for wire cross-
section
< 2.5 mm² (AWG 14)
Electromechanicaldata-versionIP66/IP68(1bar)
Cable entry
Ʋ Single chamber housing 1 x IP 68 cable gland M20 x 1.5; 1 x blind plug M20 x 1.5 Ʋ Double chamber housing 1␣x IP␣68 cable gland M20␣x␣1.5; 1␣x blind stopper
M20␣x␣1.5; 1␣x blind stopper M16␣x␣1.5
Connection cable
Ʋ Wire cross-section 0.5 mm² (AWG 20) Ʋ Wire resistance < 0.036 Ω/m (0.011 Ω/ft) Ʋ Tensile strength < 1200 N (270 lbf) Ʋ Standard length 5 m (16.4 ft) Ʋ Max. length 1000 m (3280 ft) Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Colour - standard PE Black
14)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF.
Page 55
55
10 Supplement
28444-EN-141215
Ʋ Colour - standard PUR Blue Ʋ Colour - Ex-version Blue
Display and adjustment module
Voltage supply and data transmission through the sensor Indication LC display in dot matrix Adjustment elements 4 keys
Protection rating
Ʋ unassembled IP 20 Ʋ mounted into the sensor without cover IP 40
Material
Ʋ Housing ABS Ʋ Inspection window Polyester foil
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 9 … 32 V DC Ʋ Ex-ia instrument 9 … 24 V DC Ʋ Ex-d instrument 16 … 32 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Ex instrument 12 … 32 V DC Ʋ Ex-ia instrument 12 … 24 V DC Ʋ Ex-d instrument 20 … 32 V DC
Power supply by/max. number of sensors
Ʋ DP/PA segment coupler max. 32 (max. 10 with Ex) Ʋ VEGALOG 571 EP card max. 15 (max. 10 with Ex)
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67 (NEMA 4X) Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar) NEMA 6P
15)
Ʋ Aluminium and stainless housing, in-
vestment casting (optionally available)
IP 66/IP 68 (1 bar) NEMA 6P
Overvoltage category III Protection class II
Approvals
Instruments with approvals can have dierent technical specications depending on the version.
15)
A suitable cable is the prerequisite for maintaining the protection rating.
Page 56
56
10 Supplement
28444-EN-141215
For that reason the associated approval documents of these instruments have to be carefully noted. They are part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and "Instrument search" as well as via "Downloads" and "Approvals".
10.2 ProbusPA
Instrumentmasterle
The instrument master le (GSD) contains the characteristic data of the Probus PA instrument. These data are, e.g. the permissible transmission rates as well as information on diagnostics values and the format of the measured value outputted by the PA instrument.
A bitmap le is also provided for the Probus network planning tool. This le is installed automati­cally when the GSD le is integrated. The bitmap le is used for symbolic indication of the PA instru­ment in the conguration tool.
Ident number
Each Probus instrument gets an unambiguous ident number (ID number) from the Probus user organisation (PNO). This ID number is also included in the name of the GSD le. For VEGAPULS 61 the ID number is 0x0772(hex) and the GSD le PS__0772.GSD. Optionally to this manufacturer­specic GSD le, PNO provides also a general so-called prole-specic GSD le. For VEGAPULS 61 you have to use the general GSD le PA139700.GSD. If the general GSD le is used, the sensor must be set to the prole-specic ident number via the DTM software. By default, the sensor oper­ates with the manufacturer-specic ID number.
Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during opera­tion. The below block diagram below shows which data can be accessed by the PLC.
Page 57
57
10 Supplement
28444-EN-141215
m(d) % Lin%
Min-Max
adjustment
Linearization
PROFIBUS PA-output
Primary
Value
Secondary
Value 2
Secondary
Value 1
Target
Mode
Failure
mode
AlarmsScaling
Source for
scaling
t
i
Integration
Sensor
characteristics
PA-Out
TB
FB
Select additional cyclic
value
Fig. 45: VEGAPULS 61: Block diagram with AI (PA-OUT) value and additional cyclical value TB Transducer Block
FB Function Block
Module of the PA sensors
For the cyclic data trac, VEGAPULS 61 provides the following modules:
AI (PA-OUT)
PA-OUT value of the FB1 after scaling
Additional Cyclic Value
Additional cyclical value (depending on the source)
Free Place – This module must be used if a value in the data telegram of the cyclical data trac should not
be used (e.g. replacement of temperature and Additional Cyclic Value)
Max. two modules can be active. By means of the conguration software of the Probus master, you can determine the conguration of the cyclical data telegram with these modules. The procedure depends on the respective conguration software.
Note:
The modules are available in two versions:
Short for Probus master supporting only one "Identier Format" byte, e.g. Allen Bradley
Long for Probus master only supporting the "Identier Format" byte, e.g. Siemens
S7-300/400
Examplesoftelegramconguration
In the following you will see how the modules can be combined and how the appendant data tel­egram is structured.
Example 1 (standard setting) with distance value and additional cyclical value:
AI (PA-OUT)
Page 58
58
10 Supplement
28444-EN-141215
Additional Cyclic Value
Byte-No.
1234 5 6789 10
Status
Status
Status
Status
(FB1)
IEEE-754-
Floating point value
PA-OUT
(FB1)
IEEE-754-
Floating point value
Additional Cyclic
Value
Format
Value
Example 2 with distance value without additional cyclical value:
AI (PA-OUT)
Free Place
Byte-No.
1234 5
Status
Status
(FB1)
IEEE-754-
Floating point value
PA-OUT
(FB1)
Format
Value
Note:
Bytes 6-10 are not used in this example.
Data format of the output signal
Byte4
Byte3
Byte2
Byte1
Byte0
Status Value (IEEE-754)
Fig. 48: Data format of the output signal
The status byte corresponds to prole 3.0 "Probus PA Prole for Process Control Devices" coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit6 … 0 = 0).
The measured value is transferred as a 32 bit oating point number in the IEEE-754 format.
Byte n
Byte n+1
Bit
7
VZ
Exponent
Bit
6 2
Bit
5 2
Bit
4 2
Bit
3 2
Bit
2 2
Bit
1 2
Bit
0 2
Bit
7 2
Bit
6 2
Bit
5 2
Bit
4 2
Bit
3 2
Bit
2 2
Bit
1 2
Bit
0 2
Sign
Bit
Significant
7
6543
210-1
-2
-3
-4
-5
-6 -7
Byte n+2
Byte n+3
Bit
7 2
Significant
Bit
6 2
Bit
5 2
Bit
4 2
Bit
3 2
Bit
2 2
Bit
1 2
Bit
0 2
Bit
7 2
Bit
6 2
Bit
5 2
Bit
4 2
Bit
3 2
Bit
2 2
Bit
1 2
Bit
0 2
Significant
-9
-10 -11
-12
-13
-14 -15
-16
-17
-18
-19
-20
-21
-22
-23
-8
Value = (-1)
VZ
2
(Exponent - 127)
(1 + Significant)
Fig. 49: Data format of the measured value
Coding of the status byte associated with the PA output value
Status code
Description according to
Probusstandard
Possible cause
0 x 00 bad - non-specic Flash-Update active
0 x 04 bad - conguration error – Adjustment error
Conguration error with PV-Scale (PV-Span too small)Unit irregularityError in the linearization table
Page 59
59
10 Supplement
28444-EN-141215
Status code
Description according to
Probusstandard
Possible cause
0 x 0C bad - sensor failure Hardware error
– Converter error – Leakage pulse error – Trigger error
0 x 10 bad - sensor failure Measured value generation error
– Temperature measurement error
0 x 1f bad - out of service constant "Out of Service" mode switched on
0 x 44 uncertain - last unstable value Failsafe replacement value (Failsafe-Mode = "Last value" and al-
ready valid measured value since switching on)
0 x 48 uncertain substitute set – Switch on simulation
– Failsafe replacement value (Failsafe-Mode = "Fsafe value")
0 x 4c uncertain - initial value Failsafe replacement value (Failsafe-Mode = "Last valid value" and
no valid measured value since switching on)
0 x 51 uncertain - sensor; conversion
not accurate - low limited
Sensor value < lower limit
0 x 52 uncertain - sensor; conversion
not accurate - high limited
Sensor value > upper limit
0 x 80 good (non-cascade) - OK OK
0 x 84 good (non-cascade) - active
block alarm
Static revision (FB, TB) changed (10 sec. active, after the parame­ter of the static category has been written)
0 x 89 good (non-cascade) - active ad-
visory alarm - low limited
Lo-Alarm
0 x 8a good (non-cascade) - active ad-
visory alarm - high limited
Hi-Alarm
0 x 8d good (non-cascade) - active crit-
ical alarm - low limited
Lo-Lo-Alarm
0 x 8e good (non-cascade) - active crit-
ical alarm - high limited
Hi-Hi-Alarm
Page 60
60
10 Supplement
28444-EN-141215
10.3 Dimensions
HousinginprotectionIP66/IP68(0.2bar)
~ 69 mm
(2.72")
ø 77 mm
(3.03")
112 mm (4
13
/
32
")
M20x1,5/ ½ NPT
~ 69 mm
(2.72")
ø 77 mm
(3.03")
117 mm (4.61")
M20x1,5/ ½ NPT
~ 87 mm (3.43")
M16x1,5
ø 84 mm
(3.31")
120 mm (4.72")
M20x1,5/ ½ NPT
~ 116 mm (4.57")
ø 84 mm (3.31")
116 mm (4.57")
M20x1,5M20x1,5/
½ NPT
~ 59 mm
(2.32")
ø 80 mm
(3.15")
112 mm (4.41")
M20x1,5/ ½ NPT
1
3
3
2
2
Fig. 50: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Plastic housing 2 Aluminium housing 3 Aluminium double chamber housing 4 Stainless steel housing, electropolished 5 Stainless steel housing - precision casting
Page 61
61
10 Supplement
28444-EN-141215
HousinginprotectionIP66/IP68(1bar)
117 mm (4.61")
~ 103 mm
(4.06")
ø 77 mm
(3.03")
116 mm (4.57")
~ 150 mm (5.91")
ø 84 mm (3.31")
M20x1,5
2
M20x1,5M20x1,5
1
Fig.51:HousingversionswithprotectionratingIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustment modulethehousingis9␣mm/0.35␣inhigher
1 Aluminium housing 2 Stainless steel housing - precision casting
VEGAPULS61,threadedversion
SW 50 mm
(1.97")
43 mm
(1.69")
20 mm
(0.79")
78 mm (3.07")
ø 39 mm
(1.54")
G1½A / 1½ NPT
Fig. 52: VEGAPULS 61, threaded version G1½ A and 1½ NPT
Page 62
62
10 Supplement
28444-EN-141215
VEGAPULS61,hygienictting
1 2
3
1
2
3
SW 50 mm
(1.97")
SW 50 mm
(1.97")
SW 50 mm
(1.97")
121 mm (4.76")121 mm (4.76")
121 mm (4.76")
ø 64 mm (2.52") ø 91 mm (3.58")
ø 78 mm (3.07") ø 95 mm (3.74")
ø 39 mm
(1.54")
ø 68 mm
(2.68")
ø 84 mm
(3.31")
ø 39 mm
(1.54")
Fig.53:VEGAPULS61,hygienictting
1 Clamp 2" (ø64 mm), 2½" (ø77.5 mm), 3" (ø91 mm) according to DIN 32676, ISO 2852/316L 2 Slotted nut according to DIN 11851 DN 50 and DN 80 3 Tuchenhagen Varivent DN 32
Page 63
63
10 Supplement
28444-EN-141215
VEGAPULS61,versionwithmountingstrap
125 mm
(4.92")
2,5 mm
(0.10")
75 mm
(2.95")
107 mm
(4.21")
115 mm
(4.53")
300 mm
(11.81")
19 mm
(0.75")
98 mm
(3.86")
15 mm
(0.59")
8,5 mm
(0.34")
9 mm
(0.35")
85 mm
(3.35")
12 mm
(0.47")
9 mm
(0.35")
12 mm
(0.47")
170 mm
(6.69")
Fig. 54: VEGAPULS 61, version with mounting strap in 170 or 300 mm length
Page 64
64
10 Supplement
28444-EN-141215
VEGAPULS61,versionwithcompressionange
126 mm
(4.96")
10,5 mm
(0.41")
19 mm
(0.75")
ø 75 mm (2.95")
ø 115 mm (4.53")
ø 156 mm (6.14")
ø 200 mm (7.87")
ø 21 mm
(0.83")
ø 107 mm
(4.21")
Fig.55:VEGAPULS61,compressionangeDN80/3"/JIS80
Page 65
65
10 Supplement
28444-EN-141215
VEGAPULS61,versionwithadapterange
2
1
138 mm (5.43")
31 mm
(1.22")
20 mm
(0.79")
8 mm
(0.32")
ø 75 mm (2.95")
ø 98 mm (3.86")
Fig.56:VEGAPULS61,adapterangeDN100/DN150
1 Adapterange
2 Seal
Page 66
66
10 Supplement
28444-EN-141215
10.4 Industrial property rights
VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com
10.5 Trademark
All the brands as well as trade and company names used are property of their lawful proprietor/
originator.
Page 67
67
Notes
28444-EN-141215
Page 68
Printing date:
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
28444-EN-141215
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2014
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
Loading...