HM70 Hand-Held Humidity and Temperature Meter for Spot-Checking Applications
Features/Benefits
Features/Benefits
Multilingual user interface
Multilingual user interface
Shows measurement trends
Shows measurement trends
graphically
graphically
Proven Vaisala HUMICAP® Sensor
Proven Vaisala HUMICAP® Sensor
technology
technology
3 probes, temperature measurement
3 probe alternatives, temperature
ranges between -70 and +180 °C
measurement ranges between -70
Multiprobe operation; dewpoint and
and +180 °C CO2 probes can also be connected
Multiprobe operation; dewpoint and
Displays various humidity
CO2 probes can also be connected parameters
2 probes can be connected
Sensor preheat and chemical purge
simultaneously
options for demanding conditions
Displays various humidity
Data can be logged, and transferred
parameters to a PC via MI70 Link software
Sensor preheat and chemical purge
Rugged and reliable - even in
options for demanding conditions extreme conditions
Data can be logged, and transferred
NIST traceable (certicate included)
e Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM70 is a high­performance, portable humidity reference that is exciting to use.
to a PC via MI70 Link software
NIST traceable (certicate included)
e Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM70 is a user-friendly meter for demanding spot-checking humidity measurements. It is ideal for eld checking and calibration of Vaisala's xed humidity instruments.
Three probes to choose from
e HMP75 is a general purpose probe for humidity measurements between
-20 and +60 °C. e HMP76 is a long, stainless steel probe for humidity measurements between -50 and +120 °C. It is especially suitable for spot-checking in ducts. e HMP77 is a small probe at the end of a 5-meter cable. e probe is ideal for dicult-to­reach areas and for on-site calibration of Vaisala's process transmitters. e measurement range is for temperatures up to +180 °C.
One or two probes can be connected simultaneously. In addition, the HM70 supports the use of Vaisala's dewpoint, carbon dioxide and moisture in oil probes, allowing measurements in multiparameter applications.
Easy-to-use user interface and three-variable display
e HM70 features a multilingual, menu-based user interface, and a clear graphical LCD display. Overall three measurement parameters can be displayed and logged into the meter's memory at the same time. An analog output option is also available.
Proven HUMICAP® technology
e HM70 incorporates the latest generation of the Vaisala HUMICAP® Sensor. Its proven reliable and high performance humidity measurement ability is combined with better than ever long-term stability. Additionally, a sensor that has been developed to cope with chemical interference is available to provide accuracy that lasts in demanding conditions.
Stable performance in demanding applications
e chemical purge option maintains measurement accuracy in environments with high concentrations of chemicals. e sensor preheat option reduces measurement delays by keeping the
sensor dry when the probe is inserted into hot and humid processes.
Calibration
e meter is calibrated at the factory against traceable standards, and delivered with a calibration certicate. e HM70 is also available with accredited calibration. e HM70 probes can be sent to a Vaisala Service Center for a traceable recalibration.
MI70 Link
e optional MI70 Link Windows® software and the USB connection cable form a practical tool for transferring logged data and real time measurement data from the HM70 to a PC.
e graphic display shows the stabilization of measurements and process trends.
38
39
Vaisala Instruments Catalog 2009
Ref. B210768EN rev. A
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
HM70
Technical Data
HUMIDITY
DEWPOINT
HMP75, HMP76 and HMP77 probes Measured variables
Relative humidity
Measurement range 0 ... 100 %RH Accuracy (including nonlinearity, hysteresis and repeatability)
at +15 ... +25 °C (+59 ... +77 °F) ±1 %RH (0 ... 90 %RH)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C / +68 °F)
(Dened as ±2 standard deviation limits.)
Response time (90%) at +20 °C (+68 °F) in still air
(with standard plastic grid) 17 s
HMP75
(with standard sintered bronze lter) 60 s
HMP76
(with standard plastic grid and stainless steel netting) 50 s
HMP77
Sensor HUMICAP® 180R
HUMICAP® 180RC (chemical purge, sensor preheat)
Typical long-term stability better than 1 %RH / year
Temperature
Measurement range
HMP75 -20 ... +60 °C (-4 ... +140 °F) HMP76 -50 ... +120 °C (-58 ... +248 F) short time -50 ... +180 °C (-58 ... +356 °F)
HMP77 -70 ... +180 °C (-94 ... +356 °F) Accuracy at +20 °C (+68 °F) ±0.2 °C (±0.36 °F) Accuracy over temperature range (see graph)
Temperature sensor Pt100 IEC751 1/3 class B
±1.7 %RH (90 ... 100 %RH)
±0.6 %RH (0 ... 40 %RH)
±1.0 %RH (40 ... 97 %RH)
MI70 measurement indicator Indicator general
Menu languages English, Chinese, French, Spanish,
Display LCD with backlight, graphic trend display of
Max. no of probes 2 Power supply Rechargeable NiMH battery pack with AC-adapter
Analog output 0 ... 1 VDC Output resolution 0.6 mV PC interface MI70 Link software with USB or serial port cable Data logging capacity 2700 points Alarm audible alarm function Operating temperature range -10 ... +40 °C (+14 ... +104 °F) Operating humidity range non-condensing Housing classication IP54 Weight 400 g Battery operation time
Continuous use 48 h typical at +20 °C (+68 F) Data logging use up to a month, depending on logging interval
Electromagnetic compatibility Complies with EMC standard
German, Russian, Japanese, Swedish, Finnish
any parameter, character height up to 16 mm
or 4xAA-size alkalines, type IEC LR6
EN61326-1, Portable Equipment
MI70 indicator + probe = HM70
Accessories
Carrying cases
for MI70 and HMP75/77 probe MI70CASE for MI70 and HMP75/76 probe MI70CASE2
Transmitter connection cables for
HMP230 Series 27159ZZ HMT330 & HMT100 Series 211339 HM60/70/140 Series HMA6070
HM20/30/130 Series HMA2030 MI70 Link software with USB cable 219687 MI70 Link software with serial port cable MI70LINK Analog output cable 27168ZZ Sensor protection
HMP75
Plastic PC grid (HMP75 standard) 6221 Membrane lter 10159HM Sintered bronze lter DRW212987SP HMP76/77 Plastic PPS grid DRW010276SP
Sintered stainless steel lter Sintered bronze lter (HMP76 standard) DRW212987SP PPS grid with SS netting (HMP77 standard) DRW010281SP
HM47280SP
Calculated variables (typical ranges)
Dewpoint temperature -20 ... +100 °C (-4 ... +212 °F) Mixing ratio 0 ... 600 g/kg dry air Absolute humidity 0 ... 600 g/m Wet bulb temperature 0 ... 100 °C (+32 ... +212 °F) Enthalpy 0 ... 160 kJ/g Water vapor pressure 0 ... 1000 hPa
Dimensions
Dimensions in mm (inches)
3
HMP75
HMP76
HMP77
e accuracies of these calculated variables are limited by the accuracies of the measured variables on which they are based, namely RH and T. e RH and T accuracies are stated above under Measured Variables.
Probe general
Operating temperature range for electronics -40 ... +60 °C
Housing classication IP65 (NEMA 4) Housing material ABS/PC blend Probe material Stainless steel (AISI316L) Cable length between probe and indicator 1.9 m Weight
HMP75 250 g HMP76 350 g HMP77 500 g
(-40 ... +140 °F)
HUMICAP® is a registered trademark of Vaisala. Specications subject to change without prior notice. © Vaisala Oyj