Tanita BD-815U User Manual

Page 1
Professional Baby Scale
INSTRUCTION MANUAL
BD-815U
GB
Thank you for purchasing the TANITA digital baby scale. For optimum performance and safety, please read this manual carefully, and keep it handy for future reference.
Page 2
Contents
Before Use
Safety Notes
GB GB
Caution Symbols
WarningCautionMaintenanceInstructions for Measurement
<Usage Conditions> <Storage Conditions> <Power Source> <To measure correctly> <Explanation of terminology>
Baby Basket Attachment & Removal Part Names & Accessories Names and Functions of Display Panel and Operation Keys
How to Use
Power supply How to Set Up Correctly How to Use the Scale
•When tare is not specified
•Tare weighing function
•Canceling the tare function
Memory Function for Measured Values
•Memory Function
•Memory Value “Recall” Method
•Display Weight Gain
Various Setting Methods
•Display Hold Time
•Auto Power Off Time
Output Data Format
•••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••
••••••••••••••••
•••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••
•••••
10 11
12 13
22
4 4
4 5 5 6
6 6 6 6 7
8
14 16 17
18 19 19
20 21
When necessary
Troubleshooting Specifications
•••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••
26 27
Page 3
Safety Notes
Caution Symbols
Thank you for purchasing this precision crafted Tanita product. For optimum per­formance and safety, please familiarize yourself with the Caution Symbols below.
GB GB
These symbols are designed to alert the user to potential hazards when using this equipment. Ignoring these Caution Symbols may result in serious injury, or damage to the product. Please be sure to review before proceeding with the INSTRUCTION MANUAL.
This symbol indicates the possibility of serious
Warning
injury if the product is mishandled or instruc­tions are ignored.
Caution
• Please make sure you place the scale on a level and stable surface. If the equipment is used when the scale is unstable because not all feet are on the surface, there may be a risk of stumbling or inaccurate measurement. Never jump on the scale, there may be a risk of stumbling and malfunction of the equip­ment.
• For the BD-815U; Ensure you use the original AC Adapter (MODEL: AD-0920M). Using an AC Adapter other than the original one may cause malfunction. Do not insert or remove the plug by the cable.
• Do not place a finger into the gap or hole of this equipment. You may get injured.
This symbol indicates the possibility of physi-
Caution
Warning
• Inserting and Removing the AC adapter To reduce the risk of electric shock or product damage, never insert or remove the AC adapter with wet hands.
• Do not under any circumstances dismantle or alter the device, as this could result in electric shock or injury as well as adversely affect the preci­sion of measurements.
• To prevent fire hazard Use only a correctly wired (120V AC) outlet, and do not use a multiple outlet extension cable.
• Do not sit on the edge of the scale, jump on it, or bounce on the baby bas­ket, as any resultant fall or slip may cause serious injury.
• The baby basket is not designed to accommodate individuals weighing more than 33lb(15kg).
• As a safety precaution, infants must not be left unattended on the baby basket.
cal injury or equipment damage if instructions are ignored.
This symbol indicates general precautions that should be taken when using this product.
Maintenance
This is a precision manufactured and accurately calibrated product. Please observe the following instructions.
• Never disassemble or adjust the equipment, as this may cause malfunctions. The equip­ment should only be inspected in accordance with the regulations in each country.
• When not in use for a long time, unplug the AC Adapter from the wall socket.
• Do not wipe the equipment with corrosive chemicals (benzine, cleaner, etc.). Please use a neutral detergent to clean the equipment.
• Avoid subjecting the equipment to excessive shocks or vibrations.
• When disposing of this unit, please do so in accordance with the prevailing regulations in each country.
• If an unauthorised person attempts to disassemble or repair any of the part, the warran­ty will become invalid. When the unit malfunctions, please consult your nearest Tanita sales office or agent.
• Do not wash any part of the scale other than the basket with water.
Do not lean the product against the wall where people and objects pass by.
Someone may touch the product and it may fall down
Page 4
Safety Notes (continued)
Instructions for Measurement
GB GB
Please observe the following instructions for accurate measurement.
• Measurement may not be possible on a surface that is strongly vibrating. In such cases transfer the equipment to a stable surface with little or no vibration.
• In the case to move to a place with a temperature difference of 36˚F (20˚C) or more, leave it before use for 2 hours or more after moving.
• While taking measurements, please keep any persons using transmitters, such as a cellular phone, away from the unit to avoid causing margin errors.
<Usage Conditions>
Temperature Range for Use : 41˚F - 95˚F (5˚C - 35˚C) Relative Humidity : 30% - 80% (without condensation)
<Storage Conditions>
Temperature Range of Environment : 14˚F - 122˚F (-10˚C - +50˚C) Range of Relative Humidity : 10% - 90% (without condensation) To avoid malfunctions, avoid storing the equipment where there is direct sunlight, significant temperature changes, the risk of dampness, a large amount of dust, in the vicinity of fires, or where there is the risk of receiving vibrations or shocks.
• Capacity
This shows the maximum weight that can be measured by the scale.
[Example]
In the case of Maximum capacity 33lb (15kg), it can measure up to 33lb (15kg). * One exceeding the weighing capacity cannot be measured.
•d
This shows the amount of one scale mark. In the case of a digital display, this shows the inter­val of the numeric value to change.
[Example]
In the case of Minimum Graduation 0.1oz (0 - 12lb) 0.2oz (12 - 33lb) or 2g (0 - 6kg) 5g (6 - 15kg) the display increases and decreases.
• Extend Minimum Graduation
Enables changing precision of measurements.
• Tare
The container, towel, blanket, etc., used when measuring weight is called “tare”. To measure the net-weight after subtracting the weight of the tare is called "tare weighing."
Explanation of terminology
<AC Adapter>
Model
Input
Output
AD-0920M
120 VAC 60Hz
9 VDC 200mA
To measure correctly
Do not measure at a place where the temperature fluctuates greatly.
Errors may be observed in the measured value. In the case to move to a place with a temperature difference of 36˚F (20˚C) or more, leave it before use for 2 hours or more after moving.
• Do not use cellular phones or microwave therapy equipment near this equipment. It may not measure correctly due to equipment malfunction.
Page 5
Baby Basket Attachment & Removal
A.Attachment
The baby basket is not attached to the main unit when unpacked.
GB GB
1
Pull out the lever.
Line
2
3
Align the far edge of the basket with the stickered line.
Slide the basket over the plat­form until it stops.
B.Removal
The baby basket can be removed and washed.
1
Pull out the lever.
While gently pushing down the
2
3
lever, slide out the basket in the displayed direction.
It is ready to remove the basket once its far edge is aligned with the stickered line.
4
Make sure the basket is firmly secured and then fix it to the platform by pushing in the lever.
Caution
Make sure the basket can't drop.
Caution
Once the basket is removed, take care not to insert a finger or drop anything into the gap.
Page 6
Part Names & Accessories
Names and Functions of Display Panel and Operation Keys
Front
Baby basket (for infant)
Display Panel
GB GB
Lever
Measurement is stabilized.
Tare is specified.
Display is on hold.
* The pointers (black triangles) indicate active items.
* The backlight is on only when the AC adapter is used.
* Thesmark appears in the digital display next to each of the
following operations, indicating which are in effect.
Display panel
Operation keys
Refer to page 11
Back
Battery cover
Level gauge
Adjusters (4)
Weight is memorized.
Operation Keys
MEMORY
* Memory of
current weight.
* Sets display
hold time.
Weight gain is displayed for 5 seconds after pressing the Weight Gain key.
* Available key make a “beep” sound when pressed.
WEIGHT GAIN
* Indicate the weight dif-
ference between memo­ry weight and current weight.
* Set auto power off time.
RECALL
Indicates the memory weight when it has memory weight.
WEIGHT LOCK OFF
Sets the weight lock off.
Memorized value is recalled and displayed.
g/lb•oz
AC adapter inlet
RS-232C connector
Battery housing
Selects grams or pounds and ounces for weighing unit.
Confirmation of accessories
Instruction manual (this book) AC adapter (9V)
AA-size batteries(4)
ON/TARE
Turns the power on or sets the tare.
ZERO
Scale resets to zero.
OFF
Turns power off.
10 11
EXTEND
Provides greater precision.
Page 7
Power Supply
How to Set Up Correctly
Using batteries
(1) Turn the scale upside-down,
GB GB
and turn the finger screw on the battery cover counter­clockwise (as indicated) to release the screw.
•When the batteries start to run low, appears on the display panel as a warning.
All (four) batteries should be immediately replaced with new ones at the same time. Note) Do not use a combination of alkaline and manganese batteries, as such an arrangement may result in equipment failure. Note) Rechargeable batteries are NOT recom­mended.
•A set of standard batteries should last about one year (assuming four readings a day).
(2) Lift off the battery cover. (3) Insert new batteries.
* Insert the batteries
such that the negative terminals of the batter­ies are pushed up against the coils within the battery housing.
Image
Use the scale in a firm, flat and stable location.
• Do not position it anywhere that is subject to direct sunlight, near heating equipment, or directly in the path of an air conditioner outlet.
• Do not use it anywhere that is subject to significant temperature change.
• Do not position it anywhere that is either damp or subject to high humidity.
• After positioning the scale, pick up the baby basket from both sides to confirm that the basket is not easily removed before use.
• Ensure that the lever is pushed in.
Level gauge Check
•Place the scale on a flat surface to ensure
accurate measurement.
•Turn the four adjusters such that the air bub-
ble in the level guide is positioned exactly in the center. Confirm that all four adjusters are in contact with the surface.
Level gauge
<View from above>
Red
Using the AC adapter
(1) Insert the AC adapter jack into the AC
adapter inlet on the right side of the main unit.
(2) Plug the AC adapter into the power outlet.
Caution
This equipment must be used with the included AC adapter.
Air bubble
Adjusters (4)
1312
Page 8
How to Use the Scale
取り扱い上の注意
General measurement
GB
Press the key to turn on
1
the power. After flashing the display.
is displayed.
< Example of the lb • oz display >
Press the key to select the unit.
2
The display alternates between the two units by pressing the key.
* Unit selection can be made valid only when “0” is displayed.
st
lb • oz unit g unit
Note
•Press the key to cancel measurements or turn off the display.
•The extended display function cannot be used if “Hold Time” is canceled, or if the “Recall” or “Weight Gain” is displayed.
After measurement, remove the
2
baby from the baby basket.
If the display hold time has been set, the measurement remains on display for the set time. The display can be reset to
at any time by pressing the
key.
Extended Display Function
When the key is pressed while the
display is on hold, the weight is displayed
0.05oz (1g) units for as long as the key remains pressed.
* The extended display function cannot be used if
“Hold Time” is canceled, or if the “Recall” or “Weight Gain” is displayed.
GB
Keep pressing
14
When tare is not specified
Place the baby on the baby basket.
Once the scale stabilizes and the weight is con-
1
firmed, stabilized (") and weight lock (t) will appear. (Beep sound)
Measurement instructions
•A warning sound beeps if a baby is placed
on the basket before is dis-
played. If it happens, remove the baby from the bas­ket.
By releasing the key simultaneously
return to the previous display.
Release
15
Page 9
How to Use the Scale (continued)
Tare weighing function
After turning on the power, the bath towel, diaper, blanket etc. is
GB GB
deducted by pressing key. Then, the baby’s actual (i.e. net)
weight (excluding the weight of the towel, diaper, etc.) can be meas­ured.
Place the bath towel or blanket
1
on the scale.
The weight of the bath towel is displayed.
End of measurement.
4
Remove the baby from the baby basket.
When Hold Time is set, the measurement is retained for the set time, after which the
display resets to .
Canceling the tare function
Press the key.
2
(Beep sound)
is displayed, and the pointer
beside NET is shown. (Beep sound)
Place the baby on the baby bas-
3
ket.
The net weight of the baby (excluding the weight of the bath towel, etc.) is dis­played. (Beep sound)
Remove the tare item(s) from
1
the basket.
Press the key with noth-
2
ing on the basket.
(Beep sound)
The tare function is canceled and
is displayed.
Caution
•Tare weighing is possible only within the maximum tare weight 4lb (2kg).
1716
Page 10
Memory Function for Measured Values
Memory Function
When the key is pressed while the display is on hold, the dis-
GB GB
played value is memorized.
Press the key.
1
(Beep sound)
* When the previous reading is memorized, the
“Memory” pointer is shown.
Pressing the key overwrites the existing record with the currently dis-
played value as the new memorized value while the display is on hold.
< Display is on hold >
Memory cancellation
Press the key while
1
This function is used to check the value memorized previously.
Press the key.
1
(Beep sound)
The value memorized previously is shown for 5 seconds.
* This key is valid only when this scale has a mem-
orized weight.
“Memorize” the weight before breastfeeding, and then take another measurement after breastfeeding. The increase is calculated automati­cally, by pressing weight gain.
Place the baby on the baby basket.
(Beep sound)
1
Memory Value “Recall” Method
5 seconds
Display Weight Gain
is displayed.
The memorized value is deleted, and the “Memory” pointer disappears.
Press the key.
2
“The currently displayed value – memorized value” is displayed as the weight gain (breastfeeding amount) for 5 seconds.
* If the new reading (after breastfeeding) is less than the memorized value (before breast-
feeding), a minus sign (–) is displayed.
(Beep sound)
Caution
•When checking the weight gain for another baby, be sure to memorize his/her weight before measuring the weight gain. Each baby’s weight must be record­ed before and after breastfeeding for accurate measurement of the breastfeed­ing amount.
1918
Page 11
Various Setting Methods
Three types of setting, namely “Display Hold Time” and “Auto Power Off Time” can be used on this scale.
GB GB
Press key.
1
Press key while the screen is
(Beep sound)
2
flashing.
(Beep sound)
The “Auto Power Off Time” setting automatically turns off the power when the scale is unused for the specified length of time.
Press key.
3
The auto power off time is shown (the default is 15 minutes), and changed by, pressing this key.
* The auto power off times can be set are 15 minutes, 30 minutes, 45 minutes, 60 minutes, or
-- (None).
--(None) means that the auto power off function is deactivated.
Display Hold Time
You can set certain seconds to keep the measurement on the display after remov­ing the baby.
Press key.
3
The hold time is shown (the default is 10 seconds), and the display changed by pressing this key.
* The hold times that can be set are 10 seconds, 20 seconds, 30 seconds, or -- (None).
-- (None) means the hold function is deactivated.
(Beep sound)
4
t
To complete to set the auto power
off time, press again.
Auto Power Off Time
(Beep sound)
ss
t
t
20
t
To complete to set the hold time,
4
press again.
(Beep sound)
ss
s
(Beep sound) To save and complete that setting.
21
Page 12
Output Data Format
Explanation for RS-232C-compliant output
This explanation is for the interface to output the readings from the
GB GB
Digital Baby-Scale BD-815U to external equipment (e.g. PC) as a RS­232C-compliant signal.
Grams (g) mode (transmission data)
Transmission data is output immediately after measurement regard­less of the status of the reception side (e.g. PC). Therefore, the recep­tion side must always be ready to receive data before any measure­ments are taken.
Specifications
Communication standard EIA RS-232C
Communication method Asynchronous
Signal speed 9,600bps
Data bit length 8bit
Parity NONE
Stop bit 1bit
Terminator CR+LF
Flow control None
Item Header Output data (ASCII code)
Control data Control data Fixed at 1
Model No.
Body weight XXXXX 3~5 bytes variable length (unit: g)
123456789101112131415161718192021222324252627
{{
28 29 30 31 32
,,6611,,00
A. When the display is held
Fixed at 16
{0
~0
"XXXXXX" 8-byte fixed length (“BD-815”)
MO
Wg
˜˜
Details
33,,ggWW,,""551188--DDBB"",,OOMM,,11,,00
LFLFCRCR227799
Signal line name and connection method
Terminal no. Signal name
1 2 3 4 5 6 7 8 9
TXD
GND
Item Header Output data (ASCII code)
Weight gain
Memorized weight XXXXX 3~5 bytes variable length (unit: g)
12345678910111213141516
Item Header Output data (ASCII code)
Memorized weight XXXXX 3~5 bytes variable length (unit: g)
1234567 8 9
B. When the Weight Gain key is pressed
XXXXXX 1~6 bytes variable length (unit: g)
Gg Rg
LFLFCRCR668822,,ggRR,,002211,,55ggGG
C. When the Recall key is pressed
Rg
Details
Details
LFLFCRCR88668833,,ggRR
Connection example
BD-815U PC, etc.
Note
•Commas (,) are used to separate data.
•Terminator (end of data) is CR (ASCII code: 0DH), LF (ASCII code: 0AH).
2322
Page 13
Output Data Format (continued)
Pounds/ounces (lb•oz) mode (transmission data)
Transmission data is output immediately after measurement regard-
GB GB
less of the status at the reception side (e.g. PC). Therefore, the recep­tion side must always be ready to receive data before any measure­ments are taken.
A. When the display is held
Item Header Output data (ASCII code)
Control data Fixed at 16 Control data Fixed at 1
Model No. "XXXXXX" 8-byte fixed length (“BD-815”)
Body weight XX.XX.X 5~7 bytes variable length (unit: lb•oz)
123456789101112131415161718192021222324252627
{{
28 29 30 31 32 33
{0
~0
MO
Wo
˜˜
,,6611,,00
Details
99,,ooWW,,""551188--DDBB"",,OOMM,,11,,00
LFLFCRCR33..33..
B. When the Weight Gain key is pressed
Item Header Output data (ASCII code)
Weight gain
Memorized weight XX.XX.X 5~7 bytes variable length (unit: lb•oz)
1234567891011121314151617181920
XXX.XX.X 5~8 bytes variable length (unit: lb•oz)
Go Ro
Details
LFLFCRCR77..5511..88,,ooRR,,55..33..00,,ooGG
C. When the Recall key is pressed
Item Header Output data (ASCII code)
Memorized weight
1234567891011
XX.XX.X 5~7 bytes variable length (unit: lb•oz)
Ro
LFLFCRCR77..5511..88,,ooRR
Details
Note
•Commas (,) are used to separate data.
•Terminator (end of data) is CR (ASCII code: 0DH), LF (ASCII code: 0AH).
24
25
Page 14
Troubleshooting
Specifications
If you are concerned that the scale may not be functioning correctly, please check the following points before requesting repairs.
GB GB
Check pointsProblem
Nothing is displayed
when the key is
pressed.
is displayed.
• Have the batteries been removed, or has the AC adapter been unplugged?
• Are the batteries running low?
s
• Is anything stuck to the terminals of the batter­ies?
• The batteries are running out. Please replace
s
them immediately or use the AC adapter.
Page
P.12
P.12
Extended calibration (uncertified)
After the power is turned on, is displayed, and an
• Is any pressure being applied to the baby basket
s
when the power is turned on?
• Has basket been removed?
P.14
Model
Capacity 33 lb (15000g)
d (division)
Maximum tare 4lb (2000g)
Display panel
External interface
Power supply
Regular voltage
0.1oz(0-12lb) 0.2oz(12-33lb) 2g (0-6000g) 5g (6000-15000g)
0.05oz (1g)
Backlit LCD (Backlight is on only when the AC adapter is used)
RS-232C (D-sub 9-pin female connector), unidirectional output
Input: AC 120 V (60 Hz) Output: DC 9 V / 200mA
DC 6 V (using the batteries) DC 9 V (using the AC adapter)
BD-815U
alarm sounds.
0.12 W or less (using the batteries)
0.35 W or less (using the AC adapter)
20mA (using the batteries) 38mA (using the AC adapter)
is shown during measure­ment.
• The maximum measurable weight capacity for
s
the scale (33lb / 15 kg) is exceeded.
Power consumption
P.7
Consumption current
Operating temperature 41˚F - 95˚F (5˚C - 35˚C)
Storage temperature 14˚F - 122˚F (-10˚C - 50˚C)
Overall dimensions
Main unit mass Approx. 10.1lb(4.6kg)
Main material ABS resin
Made in Japan
D16.1 W23.6 H6.2in (D409 W600 H157mm)