Siemens Mammomat Novation Troubleshooting Guide

Page 1
Urdaneta
MAMMOMAT Novation
Troubleshooting Guide
System
SP
Mammomat Novation DR
08380490
Print No.: Replaces: n.a.
SPB7-250.840.01.01.02
© Siemens AG
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
English Doc. Gen. Date: 10.05
2005
Page 2

2 Revision / Disclaimer

1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 2 of 100
Medical Solutions
Page 3
Table of Contents 3
1- 0Table of Contents
1 _______ General Information______________________________________________ 5
Protective measures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Printed circuit boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measurements with the oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 _______ Error messages _________________________________________________ 6
Error messages of the master Err 0XX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Error messages of the panel Err 1XX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Error messages of the filament Err 3XX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Error messages of the AEC Err 4XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Error messages of the FFDM DR System Er 5XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Error messages of the power pack Err 6XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Error messages of the OPDIMA Err 7XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Error messages of the stand Err 8XX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3 _______ Log and error files ______________________________________________ 36
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
SW Version VA10x. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Copy MPxxxx files (detector files), dump files and syngo logs . . . . . . . . . . . . . . . . . 37
Copy Array and Brick log files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Windows Eventlog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
SW Version VA11x. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Copy MPxxxx files (detector files), dump files and syngo logs . . . . . . . . . . . . . . . . . 46
Copy Array and Brick log files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Windows Eventlog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Copy SaveLogs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4 _______ Generator checks ______________________________________________ 56
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Documents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Troubleshooting tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Signals on D750 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Tube high voltage, current and mAs values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Filament circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5 _______ Image quality __________________________________________________ 74
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Gain calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Bad pixel mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Qualify log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Image examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 3 of 100
Page 4
4 Table of Contents
6 _______ Compression Unit ______________________________________________ 94
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Documents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Check the adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Check the potentiometer R863 (position transducer) . . . . . . . . . . . . . . . . . . . . . . . . 95
Check the potentiometer R861 (compression force) . . . . . . . . . . . . . . . . . . . . . . . . . 95
Check the movements (up/down) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Check the detection of compression paddle.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
7 _______ Changes to the previous version _________________________________ 100
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 4 of 100
Medical Solutions
Page 5

General Information 5

1General Information
2-

Protective measures 0

System power supply 0

Before you begin working on equipment, it is very important that you disconnect it from the power supply at the main circuit breaker. Before removing or inserting any of the printed circuit boards, switch off the equipment.
WARNING
If the system is only switched off at the control panel, the line volt­age will still be present at the generator line connection (see wiring diagram).
¹ A life-threatening hazard of electric shock exists.
WARNING
After the system shuts down, there may still be 380 V DC present on the intermediate circuit.
¹ A life-threatening hazard of electric shock exists

Printed circuit boards 0

CAUTION
The printed circuit boards contain highly electrostatically-sensi­tive components.
If disregarded, the components could be damaged.
¹ Use ESD equipment, ground prior to making contact, and
place the components on a conductive surface.

Measurements with the oscilloscope 0

WARNING
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
Under no circumstances should the existing ground conductor in the mains cable be disconnected when operating the oscilloscope.
There is a life-threatening hazard of electric shock.
¹ For measurements in which a resulting ground loop may
affect the measuring result, use the differential amplifier (difference measurement).
Page 5 of 100
10.05 CS SD 24
Page 6

6 Error messages

2Error messages
3-

Error messages of the master Err 0XX 0

A malfunction detected by the system is indicated by " Err" followed by a three-digit error code on the kV and mAs displays of the control panel. This can be acknowledged by the operator pressing the limit button. Exceeded limits are also deleted with the limit button.
Err 004
“Communication master - AEC disturbed or AEC not ready”
Description
If the master processor (D750) can no longer activate the AEC processor via the serial interface or if any important data from the control deck is missing, the above error mes­sage is displayed.
Fault elimination
First check if PC board D750 is plugged in correctly and if the connections to the power supply and the serial interface (see wiring diagram) are functioning correctly.
If no fault can be found, PC board D750 should be replaced.
Err 005
“Flash in AEC is defective or not correctly initiated”
Description
During initialization, the AEC processor (D750) checks the data stored in its E
2
means of stored checksums. If this data is incorrect or if the E
PROM fails, the AEC pro-
2
PROM by
cessor informs the master processor via the serial interface and the above error message is displayed.
The master also generates this error if communication has not been established with the AEC during exposure release.
NOTE
In contrast, error 004 appears if communication started with the master disturbed.
Fault elimination
Reprogram E2PROM or replace E2PROM or PC board D750. Reprogram using the ser­vice PC "Backup" menu.
Err 008
“Communication master - stand disturbed or stand not ready”
Description
This error means that communication between the master processor and unit processor via the serial interface is faulty or interrupted.
Fault elimination
Check whether the power supply and the serial interface (see wiring diagram) to the unit are functioning correctly.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 6 of 100
Medical Solutions
Page 7
Error messages 7
Err 011
“Exposure aborted by user”
Description
If the exposure buttons are released prematurely, exposure is aborted immediately. On the control deck, the LIMIT LED (→0←) lights up, an acoustic signal sounds and the above error message is displayed.
Fault elimination
This error indication is purely informational and does not require further action. Should it appear frequently, however, check the contacts of the exposure release buttons. Replace the exposure release button(s), if necessary.
Err 012
“Time limit reached”
Description
In order to prevent the tube assembly from overloading, a timer runs during exposure. This is set to the time limit prior to exposure. If this timer runs down, the exposure is aborted immediately, and the above error message is displayed. Depending on the mode of operation, the time limit is calculated as follows:
AEC mode
The time limit is the power-dependent load time plus a tolerance value (approx. 400 ms).
mAs mode
The time limit is the calculated exposure time plus a certain reserve. If this period is shorter than 2 s, 2 s are assumed. If it exceeds 2 s, a reserve time of 400 ms is added.
The time limit can be read off in ms with the service PC in “normal mode” under item “AEC data”. The value shown here does not include the extra 400 ms.
Fault elimination
The tube current and the filament current must be measured again with the oscilloscope (test points MA_ACT and FIL_CURR_ACT on D750). The nominal and actual value of the tube current can be read off with the service PC in normal mode. If the actual tube current is too low, you must first ensure that the dynamic learn filament current (DLF switch) is switched on. The tube must be readjusted in any case. Should difficulties still occur, the tube current actual value acquisition must be checked (see wiring diagram).
Err 013
“mAs limit reached in AEC exposure”
Description
For safety reasons, the accumulated mAs is integrated via the timer independently of the AEC during AEC exposures. If a focus, tube assembly and kV-dependent time limit (see table) is obtained without the AEC being switched off, the exposure is aborted and the above error message is displayed.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 7 of 100
Page 8
8 Error messages
Tube assembly 25 kV 30 kV 35 kV
P40 MoW-100G F1 196 mAs 163 mAs 140 mAs
P40 MoW-100G F2 600 mAs 500 mAs 428 mAs
P40 MoW-100G F3 238 mAs 198 mAs 170 mAs
P40 MoW-100G F4 752 mAs 627 mAs 537 mAs
Fault elimination
The dose signal from the detector or the chamber to the hardware of the AEC (see wiring diagram) must be checked.
If no error can be detected, PC board D750 or D801 must be replaced.
Err 014
“kV, corrected for dose calculation, is out of range”
Description
The kV value used for glandular dose calculations is the "set kV" adjusted for the actual HVL value of the tube. This corrected kV value is outside the limits of the lookup tables used for glandular dose calculations.
Fault elimination
Reload the tube-specific HVL values using the "dose calculation configuration program".
Err 020
“PC message had bad length”
Description
The length of the message sent from the service PC to the MAMMOMAT Novation
DR
is
too long.
Fault elimination
Check that the correct version of the service program is being used.
Err 022
Receive message from external unit failed”
Description
An error occurred when the MAMMOMAT Novation
DR
was receiving data from the PC or
printer.
NOTE
May occur if the printer is switched on/off while the rest of the sys­tem is on.
Fault elimination
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 8 of 100
Medical Solutions
Page 9
Error messages 9
Check the cables and connectors to the PC or printer. Board 750 defective. PC/printer defective.
Err 024
“PC, ID camera or workstation has not sent ACK/NAK within two seconds. Three tries are made”
Description
The service PC did not answer.
Fault elimination
Check the cables and connectors to the PC. PC defective.
Err 025
“PC, ID camera or workstation sends NAK three times as response to one and the same message”
Description
The service PC did not receive information from the master. Timeout occurred and the service PC responds with NAK.
Fault elimination
1. Reset error on panel and try again.
2. Reset MAMMOMAT Novation
DR
and PC.
Err 030
“Master PLD requests interruption of tube voltage”
Description
The master PLD logic has detected a type of error that is considered fatal and must termi­nate the exposure.
Fault elimination
1. Reset error on panel and try again.
Err 040
“The radiation protection door is open”
Description
The radiation protection door switch is not active while exposure is being released.
Fault elimination
1. Check the door switch.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 9 of 100
Page 10
10 Error messages

Error messages of the panel Err 1XX 0

Err 105
“Checksum error in installation data stored in panel EEPROM”
Description
Each time the MAMMOMAT Novation
2
PROM on the control deck is checked with a shadow area also stored in the E2PROM.
E If any discrepancy is detected, the above error message is displayed.
Fault elimination
The installation data can be corrected via the service PC. If this error appears more fre­quently, exchange the deck PC board D740 and reinstall the parameters in the control panel; see Service SW Backup/Restore. The opdose parameters have to be programmed manually via the control deck. For the remaining parameters, use the service PC "Backup" menu.
Err 106
2
“E
PROM program area error”
DR
is switched on, the installation area in the
Description
When data is stored in the E
2
PROM of the control panel during programming of the opdose parameters, this is read back once more to check whether it has been stored cor­rectly. If any differences occur, the above error message is displayed.
Fault elimination
2
The E
PROM in the panel must be replaced. If the error still occurs, the whole D740 must be replaced. In both cases the parameters in the control panel have to be reinstalled. The opdose parameters have to be programmed manually via the control deck. For the remaining parameters, use the service PC, backup menu.
Err 107
“No communication panel - master”
Description
If no data transmission takes place between the deck and the master during the first 20 s after power on, the control deck switches to the so-called “standalone” mode. The key­board can be operated normally. If the exposure release button is operated in this mode, the above error message is displayed.
Fault elimination
The cause of this error can be an interruption on the serial interface to the master. For this reason, this connection must be checked carefully.
Err 121
“The version number of EEPROM and PROM are not the same”
Description
This error applies to the panel.
Fault elimination
Check EEPROM and PROM.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 10 of 100
Medical Solutions
Page 11
Error messages 11

Error messages of the filament Err 3XX 0

Err 303
“Filament current outside limits”
Description
During “Stand-by”, the master checks whether there is any filament current. If the filament current is not within 10% from nominal of a predetermined value, the above error mes­sage is displayed.
Fault elimination
Check fuse F2 (see wiring diagram).
Check filament wires.
Filament of tube assembly defective.
PC board D750 defective.
Err 304
“Filament current over limit”
Description
During “Stand-by”, the master checks whether there is any filament current. If the filament current does not reach a predetermined value or the filament current exceeds 8.5 A, the above error message is displayed.
Fault elimination
Check fuse F2 (see wiring diagram).
Check filament wires.
Filament of tube assembly defective.
PC board D750 defective.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 11 of 100
Page 12
12 Error messages

Error messages of the AEC Err 4XX 0

Err 401
“Error during configuration of PLD.”
Description
Configuration of PLD device in AEC failed.
Fault elimination
Check whether Proms are inserted properly
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
Err 402
“Self-test of PLD failed.”
Description
DR
. Wait approx. one minute. Start up the MAMMO-
Gain test of circuit board D750 failed.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 403
“Self-test timeout. Fatal error in PLD’s internal program.”
Description
Timeout occurred in the PLD device in the AEC during gain test of circuit board D750.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Turn on the MAMMO-
Err 404
“AEC CPU requests interruption of tube voltage”
Description
The processor in the AEC part of the master board requests termination of high voltage generation.
Fault elimination
Check the tube power module and the HV inverter module.
Err 405
“AEC PLD cannot perform requested operation because it is busy self- testing”
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 12 of 100
Medical Solutions
Page 13
Error messages 13
The PLD in the AEC part of the master board requests termination of high voltage gener­ation.
Fault elimination
Check the tube power module and the HV inverter module.
Err 406
“PLD cannot perform requested operation because it is busy self-testing.”
Description
The PLD in the AEC cannot perform requested operation because it is performing a self-test.
Fault elimination
Acknowledge the displayed error and wait approx. one minute.
Err 407
“Time-out error while erasing AEC flash memory”
Description
The D750 contains a memory device called “flash memory”. This memory is used for stor­ing AEC parameters and correction tables. Before writing data into the memory, it is nec­essary to erase the sector in which data is to be stored. The erasing procedure is controlled by a timer. If anything goes wrong while erasing the AEC flash memory and a timeout occurs, an error results.
Fault elimination
Switch off the MAMMOMAT Novation MAT Novation
DR
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 408
“Time-out error while writing to AEC flash memory (time-out in waiting for ok indi­cation)”
Description
The D750 contains a memory device called “flash memory”. This memory is used for stor­ing AEC parameters and correction tables. The procedure of writing this data is controlled by a timer. If anything goes wrong while writing to the AEC flash memory and a timeout occurs, an error results.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 409
“AEC flash test has locked up (time-out of 15 seconds exceeded)”
Description
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 13 of 100
Page 14
14 Error messages
During start-up of the unit, the AEC performs a number of tests, e.g. testing of the flash memory device on D750. This testing is controlled by a timer and if something is wrong, a timeout occurs.
Fault elimination
DR
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
Err 410
“Attempt to start an exposure before AEC flash memory has delivered correction values”
Description
Before an AEC exposure is released and based on the exposure settings on the panel, the AEC fetches an estimated dose from a correction table stored in the flash memory. If the exposure buttons are pressed during a reading operation from the flash memory, an error occurs.
Fault elimination
. Wait approx. one minute. Start up the MAMMO-
Acknowledge the displayed error. Wait 15 seconds. Try to perform an exposure.
Err 412
“Attempt made to start an exposure while self-testing of the PLD in AEC was in progress”
Description
Attempt made to start an exposure while a self-test of the PLD in the AEC was in progress.
Fault elimination
Acknowledge the displayed error. Wait 15 seconds. Try to perform an exposure.
Err 413
“Exposure too short, correction value could not be loaded in time”
Description
An estimated dose is calculated before an exposure is released. The minimum estimated dose is defined for a minimum object thickness. For a normal object thickness (> 5 mm), the AEC will calculate a necessary correction value to add to the estimated dose during exposure. This procedure takes a while, and in the case of particularly thin objects, the AEC may not be able to determine within the time limit whether a correction is necessary. An error results due to the small object thickness.
Fault elimination
Acknowledge the displayed error. Change the exposure parameters (e.g. lower kV).
Err 414
“Attempt to release an exposure after the PLD in AEC has failed self-test/ configu­ration”
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 14 of 100
Medical Solutions
Page 15
Error messages 15
Description
Caused by previously generated errors 401, 402, 403, 409 and not restarting
MAMMOMAT Novation
Fault elimination
Switch off the MAMMOMAT Novation
MAMMOMAT Novation
Err 416
“Estimated dose out of range”
Description
Estimated dose out of range, which can be caused by
- an extreme sensitivity setting
- an extreme sensitivity correction setting
- an extreme density correction setting.
DR
.
DR
. Wait approx. one minute. Start up the
DR
. If the error recurs, replace the D750 circuit board.
This means that the exposure will be terminated by the max. mAs limit.
Fault elimination
Acknowledge the displayed error. Make sure that the correct AEC correction tables are installed. Check settings for sensitivity, sensitivity correction and density correction.
Err 418
“Attempt to make offset compensation test while in AEC mode”
Description
During an offset compensation test, the AEC board is set to a special test mode and can­not perform an AEC exposure. Because of this, test exposures have to be performed in the mAs mode during this test.
Fault elimination
Acknowledge the displayed error. Change to mAs and follow the instructions for an offset compensation test.
Err 419
“Attempt to write to registers of PLD not available during exposure”
Description
This is an internal error that does not interfere with an exposure and is not shown on the control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it immediately.
Err 420
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 15 of 100
Page 16
16 Error messages
“Dose or time monitor overflow in PLD”
Description
This is an internal error that does not interfere with an exposure and is not shown on the control panel. The time and dose values reported by the service PC are not valid.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it immediately.
Err 421
“Wrong test mode request received by PLD”
Description
This is an internal error that does not interfere with an exposure and is not shown on the control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it immediately.
Err 422
“Attempt to read registers of PLD not available during exposure”
Description
This is an internal error that does not interfere with an exposure and is not shown on the control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it immediately.
Err 423
“Counters for estimated doses in PLD not initialized in time”
Description
If mAs mode: Internal error that does not interfere with an exposure and is not shown on the control panel. If AEC mode: Internal error that interferes with an exposure and is shown on the control panel.
Fault elimination
If mAs mode: If this error is discovered when reading the error buffer via the service PC, please report it immediately.
If AEC mode: Acknowledge the displayed error. Wait 15 seconds. Try to perform an exposure.
Err 450
“Exposure aborted by dose monitoring”
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 16 of 100
Medical Solutions
Page 17
Error messages 17
Description
An AEC exposure could not be performed due to insufficient dose rate.
Fault elimination
Acknowledge the displayed error. Change the exposure parameter settings (e.g. higher kV).
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 17 of 100
Page 18
18 Error messages

Error messages of the FFDM DR System Er 5XX 0

Err 501
“DR detector request interruption of tube voltage”
Description
The HW signal DR_DET_RDY_FOR_EXP has become inactive during exposure.
Fault elimination
Check the plugs and connections at board D750.
Err 502
“DR system request interruption of tube voltage”
Description
If the exposure buttons are released prematurely, exposure is aborted immediately.
The HW signal DR_SYST_RDY has become inactive during exposure.
Fault elimination
Check the plugs and connections at board D750.
Err 503
“Master does not sense DR signal level change within specified times”
Description
At exposure start, the HW signal DR_DET_RDY_FOR_EXP is not activated within the specified time.
Fault elimination
Check the plugs and connections at board D750.
Err 504
“Master senses an error within the DR system”
Description
The HW signal DR_DET_ERR has been activated.
Fault elimination
Check the plugs and connections at board D750.
Check the switch S1 on grid drive.
Err 506
“Master senses a tube arc during the AEC pre exposure when in DR mode”
Description
A tube arc has occurred during the AEC pre-exposure. No tube arc is allowed during the AEC pre-exposure.
Fault elimination
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 18 of 100
Medical Solutions
Page 19
Error messages 19
N/A
Err 578
“mAs exceeds maximum”
Description
The calculated AEC main exposure mAs value returned from the DR system exceeds the maximum mAs value that the MAMMOMAT Novation
Fault elimination
Check AEC functioning
Err 579
“Master is not updated with the mAs value for the main exposure in AEC mode”
Description
The mAs value for the main exposure has not been calculated/transmitted by the DR sys­tem or received by the MAMMOMAT Novation
DR
Fault elimination
DR
can handle.
within the specified time.
Check the plugs and connections at board D750.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 19 of 100
Page 20
20 Error messages

Error messages of the power pack Err 6XX 0

Err 601
“Rotation speed not reached within 3s”
Description
During starting of the rotating anode, the master counts the control pulses and compares them with a ‘tube assembly and intermediate circuit voltage’-dependent table value. If this is not reached within 3 s, err 601 is displayed.
Fault elimination
Check oscillation current (see wiring diagram). Rotating anode cable or stator defective.
Err 602
“kV
Displayed on PC: kV
- minimum tube voltage 17 kV not reached”
min
-minimum tube voltage 17kV not reached
min
Description
If 17 kV is not reached within 250 ms after the KVE signal, the above error message is displayed.
Fault elimination
Intermediate circuit voltage too low (see wiring diagram).
Tube current or power too high. See also (Tube high voltage, current and mAs
values / p. 63)
Check oscillation current (see wiring diagram).
SS relay not pulled up (see wiring diagram).
“kV
” too low.
nom
Board D750 defective.
Err 603
“kV
- tube voltage greater than 50 kV”
max
Description
During exposure, a threshold switch in the kV controller monitors whether the high voltage exceeds 50 kV. If this is the case, the exposure is aborted immediately via the KVA lead.
Fault elimination
Check the kV nominal value. Check tube current and power, see also (Tube high voltage,
current and mAs values / p. 63). Actual value acquisition defective (see wiring diagram).
Replace PC board D750.
Err 604
“WR Austast - inverter short circuit blanking signal remains”
Description
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 20 of 100
Medical Solutions
Page 21
Error messages 21
When I
(Overcurrent) in the inverter is exceeded, the control is interrupted for 200 ms.
max
The software then checks whether the blanking signal is still applied. If this is the case, the above error message is displayed.
Fault elimination
Transistor module in HV inverter defective. Check tube current and power, see also (Tube
high voltage, current and mAs values / p. 63). Check fuse (F2) . Stator short circuit. Board
D750 defective.
Err 606
“Inverter short circuit during radiation”
Description
HV-Inverter overcurrent during exposure sequence.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).Check tube current and power, see also (Tube high voltage, current and mAs values / p. 63)
Stator short circuit.
Board D750 defective.
Err 607
“Inverter short circuit during brake cycle”
Description
Actual tube voltage deviates by more than 5% from nominal value at the start of the expo­sure.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).Check tube current and power, see also (Tube high voltage, current and mAs values / p. 63). Stator short circuit.
Err 608
“Tube current is outside limits”
Description
Actual tube current deviates by more than 10% from nominal at the start of exposure.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).
Stator short circuit.
Err 611
“KVA signal disabled during radiation”
Description
During exposure, KVA becomes “H” and the inverter is thus enabled. If this enabling fails, the above error message is displayed.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 21 of 100
Page 22
22 Error messages
Fault elimination
Check KVA lead (see wiring diagram). Board D750 defective.
Err 620
“U
+15V to control inverter not present”
anst
Displayed on PC: U
+15V to control inverter not present
anst
Description
The DC bus voltage used for powering the modules of the generator assembly is too low or too high.
Fault elimination
Check the power supply and fuses.
Err 630
“PH1 pressure switch on the HV tank activated”
Description
the HV actual value does not correspond to the HV set value. The above error message is displayed.
Fault elimination
Overload of the tube assembly or of HV inverter.
Line interruption (see wiring diagram).
Board D750 defective (monitoring).
Err 631
“Open connector in the generator”
Description
At least one connector between modules of the generator assembly or between the mas­ter board and the modules of the generator assembly is not connected.
Fault elimination
Line interruption (see wiring diagram).
Board D750 defective (monitoring).
Err 632
“Bias voltage error”
Description
Active bias voltage for large focus or inactive bias voltage for small focus.
Fault elimination
Check power supplies and fuses on D750.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 22 of 100
Medical Solutions
Page 23
Error messages 23
Err 633
“The high voltage is detected during stand by”
Description
If any or both of the signals HV_ON and EN_HV_ON are active during standby (no expo­sure being performed), this error is issued. Displayed when the tube is arcing.
Fault elimination
Press the "Lim" button to clear the error message.
Check the HV inverter module.
Err 634
“Voltage outside limits for 15V supplies”
Description
This error is issued if the +15V supply voltage goes below 10.6V or if the -15V supply volt­age goes above -10.75V.
Fault elimination
Check the power supplies at the mains converter module and the fuses.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 23 of 100
Page 24
24 Error messages

Error messages of the OPDIMA Err 7XX 0

Err 777
“Exposure sequence is aborted by OPDIMA”
Description
The time between pressing acquire from the workstation and releasing the exposure from the MAMMOMAT generator has exceeded 60 seconds.
The selected exposure parameters (kV, mAs) will yield an excessively hard beam quality for the object.
The cancel button was pressed while "Expose at MAMMOMAT" or "Exposure preparation started" was displayed.
Fault elimination
Press the limit button on the MAMMOMAT generator control panel to continue.
Err 778
“No communication present between MAMMOMAT and OPDIMA”
Description
Communication problem.
Fault elimination
Press "Retry" at the workstation or press the limit button on the MAMMOMAT generator control panel to continue. Check the plugs and cables.
Err 779
“The mAs value for main exposure in OPDIMA AEC mode has not been received in time”
Description
Communication problem.
Fault elimination
Press the limit button on the MAMMOMAT generator control panel to continue.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 24 of 100
Medical Solutions
Page 25
Error messages 25

Error messages of the stand Err 8XX 0

Err 801
“Time out of AR signal”
Description
The grid has not reached its start position within 2 seconds.
Fault elimination
Board D750 defective.
Check grid functioning.
Err 802
“OKT 2 pressure switch on tube housing or beam form anode or door switch”
Description
The oil pressure in the tube has exceeded the limit.
Fault elimination
Allow the tube to cool down. Close door/switches.
Err 803
“Stand requests interruption of tube voltage”
Description
The processor on stand CPU board D801 requests termination of high voltage generation.
Fault elimination
Switch off the equipment and then switch it on again.
Err 811
“Stand not ready for exposure”
Description
Normally, the panel does not allow start of exposure if any of the following errors is present: No object table installed, no film cassette inserted, film cassette not changed after exposure, improper diaphragm mounted. Nor will it allow exposure, if the tube assembly is at a distance from the floor that could result in collision. These operator errors are indicated on the control panel (the LED at the corresponding symbol lights up). Should the control panel nevertheless allow start of exposure, this might be due to a bit error dur­ing the transmission of data between the stand and the control panel. As a precautionary measure, the stand also checks whether the conditions are fulfilled. If this is not the case, error message 811 is generated.
Fault elimination
Switch off the equipment and then switch it on again.
Err 812
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 25 of 100
Page 26
26 Error messages
“Compression protect switch not OK in stand”
Description
This signal is used to check for proper operation of the compression protect relay (Kl) on the motor board (D802). The CPU (D801) reads the signal after a short delay after releas­ing the compression pedal in order to check that the relay has opened and disabled motor operation in the compression direction. Note that the test point on D802 is protected with
4.75k.
Fault elimination, troubleshooting
Check the fuses on board D802. Normal active level 9V (test point COMP_PROT on D802). If the level is OV check 14V and Kl on the motor board or look for a short circuit on the ribbon cable between D801 and D802. If the cable is OK, D802 should be replaced.
If the voltage is 18V at the test point, the ribbon cable is open or D801 is faulty.
If the voltage is OK, replace D801.
Err 813
“Compression Motor controller, or motor, error in stand”
Description
This signal indicates an overcurrent or lack of voltage (5V or 15V) in the compression motor drive. In the case of an overcurrent, the signal appears after about 2 s and disap­pears after about 0.5 s. The signal causes D801 to stop sending pulses to the compres­sion drive and thus disables the compression motor. The red LED “comp error”, V2 lights when error (comp_ok signal low) except if 5V is missing. There is no test point.
Fault elimination, troubleshooting
If V2 is lit steadily, check fuses F3 (24V) and F2 (24VF & l5V). l5V is indicated by LED V49 and can be measured at test point l5V on D802. Test point 24VF should measure about 30V in an unloaded state. 5V is indicated by LED V7 and test point 5V on D802. If 5V and 14V (normally about 18V unloaded) are missing, check Fl on D801. Missing volt­ages can also be caused by an open connection in emergency stop switch S880.
Overcurrent is most likely caused by a faulty motor, mechanical overloading of motorized movement, short-circuited wiring of D805 or a faulty D802.
Err 815
“Compression speed too high, error suspected in stand”
Description
The CPU (D801) has detected a compression speed greater than 13 cm/s. The software causes the compression motor to reverse direction.
Fault elimination, troubleshooting
This error can be caused by play in the compression unit, defective compression mea­surement hardware, or a short-circuited power stage on the motor board. In the first two cases, replace the compression unit and recalibrate the compression values with the ser­vice PC program. In the third case, the compression motor will run at a max. speed upwards except when a compression pedal is pressed. It is not certain that error 815 will appear. Replace D802.
Err 816
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 26 of 100
Medical Solutions
Page 27
Error messages 27
“Collimator fails to reach correct position in time”
Description
Time out for filter lamella positioning during startup.
Fault elimination, troubleshooting
Check whether the collimator plates move at all.
If the collimator plates do not move:
- Check whether the CPU is sending out pulses, COLL_ST test point on D801, if not, replace D801.
- Check whether the wing board (D805) is receiving pulses, test point COLL_ST on D805; if not, check the cables and connectors between D801 and D805.
- Check voltages on the wing board.
- Check STEP_ENABLE test point on D805. It should be low (0V) when pulses are out­put to the motor.
If the collimator plates do move:
- Check whether the collimator position light switch is functioning as in the description above.
- Check whether it moves for the entire 8 s. If it does but apparently does not reach the other position in time, something may be in the way and slowing the speed, or the motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after just a short time, either the motor or the wing board (D805) is faulty. Try replacing one of them.
- Check collimator calibration (lamp and/or X-ray field). If not successful, replace collimator and/or D814.
Err 817
“Filter self-test failure”
Description
The zero position for the filter wheel is not found during start-up.
Fault elimination, troubleshooting
Switch between the three anode/filter combinations on the control panel and check if the filter disk moves.
- Check whether the filter position light switch works.
- If it only moves slowly, something may be in the way and slowing the speed, or the motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after just a short time, either the motor or the wing board (D805) is faulty. Try replacing one of them.
If not successful, replace collimator and/or D814.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 27 of 100
Page 28
28 Error messages
Err 818
“Filter positioning failure”
Description
The zero position for the filter wheel is not found after exposure.
Fault elimination
Switch between the three anode/filter combinations on the control panel and check if the filter disk moves.
- Check whether the filter position light switch works.
- If it only moves slowly, something may be in the way and slowing the speed, or the motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after a short time, either the motor or the wing board (D805) is faulty. Try replacing one of them.
If not successful, replace collimator and/or D814.
Err 819
“Drive protect switch not OK (lift and rotation), in stand”
Description
This signal is used to check for proper operation of the drive protect relay (K6) on the motor board (D802). The CPU (D801) reads the signal after a short delay after releasing the activated lift or rotation button in order to check that the relay has opened and dis­abled motor operation.
Fault elimination, troubleshooting
Normal active level 5V (test point DRIVE_PROT on D802). If the level is 0V, check 24VF and K3 on the motor board or look for a short circuit on the ribbon cable between D801 and D802. If the cable is OK, D802 should be replaced.
If the voltage is 30V at the test point, the ribbon cable is open or D801 is faulty.
If the voltage is OK, replace D801.
NOTE
There is no error for the lift & rotation drive corresponding to the compression OK error. If this fault occurs, D801 stops sending pulses to D802 and the selected motor stops. The LED V12, DRIVE_ERROR functions in the same manner as COMP_ERROR. Note that the current limit is lower when the lift motor is moving downward. If there are problems with lift or rotation operation in­dicating an overload, look at the DRIVE_ERROR LED to determine if the motor is overloaded.
Overloading is possible due to improper balancing (adjustment of the spring or improper loading of the lifting carriage) or improper adjustment of the rotation brake. Faulty motor, short-circuited wiring or faulty D802 are other possible causes.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 28 of 100
Medical Solutions
Page 29
Error messages 29
Err 820
“Write/read back failure with EEPROM in stand”
Description
If data is stored in the E to verify that it has been stored correctly. In the event of differences, the above error mes­sage is displayed.
Fault elimination
Repeat the attempt to write in the E of repeated attempts, board D801 must be replaced. All stand parameters must be rein­stalled. Use the service PC to reinstall the stand parameters from floppy or disk.
Err 821
Stand EEPROM has wrong version nbr (=old or corrupt)”
Description
The version number of the PROM stand is also stored in the E starting up the equipment, the version number stored in the E
version number of the PROM. If they do not correspond, the above error message code is displayed on the control panel. This occurs after changing the software version to indicate that a new version has been installed.
2
PROM (D801.I34) stand during adjustment, a check is performed
2
PROM. If this not does not yield any results, in spite
2
PROM (D801.I34). When
2
PROM is compared to the
Fault elimination
The version number of the new software is automatically written into the E
2
PROM. Switch
off the equipment and then switch it on again. The error should not recur.
Err 822
“Grid failed to move properly”
Description
The grid has to move properly from the start. This is supervised by the CPU. The CPU measures the time from start till the grid has reached one of the end positions and then is able to leave it. To check that the grid is functioning properly and does not jam, the time the grid requires to move between the two end positions is also measured the first time. If the time for start or run check exceeds 2 sec., this error appears.
Fault elimination, troubleshooting
Check programming of the grid fast speed (normally 80%) and the grid fast speed time (normally 500 ms).
Check the bucky board. First try another board. Make sure the grid moves freely along its entire travel path in both directions.
Check the output voltage at test point GRID_M on D802. At 80% speed and nominally 30V on 24VF, GRID_M should be 30 - 0.8*30 = 6V during fast grid movement. Check for possible short circuits if the signal remains at about 30V. This voltage goes to the bucky via board D805 and its relay Kl. Test point GRID_SP on D802 is the control signal from D801. It is a 5V, 20 kHz PWM signal. At 80% speed, the signal should be low at about 40 µs and high at about 10 µs per period. Replace D802 if the correct output is not achieved despite proper input, cabling and bucky.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 29 of 100
Page 30
30 Error messages
Err 824
“Stand motor voltage drops unnaturally”
Description
The CPU (D801) measures voltage 24VF from the motor board (D802). If this value is below about 16V, this error is reported.
Fault elimination, troubleshooting
Check that the stand has voltage (generator - X14 connected, F4 OK). The stand display should be on.
Check that the emergency stop is not activated and that its wiring is OK.
Check that LED V49, 15V, is lit. If not, check fuses F2 and F3 on D802 in the stand.
Check voltage 24VF (measure with reference to test point 0V on D802).
Check the ribbon cable between D802 and D801.
Replace D801.
Err 825
“One or more potentiometers seem to have lost contact with stand”
Description
The four potentiometers, R803-tube angle, R871-preset angle, R861-preset force and R863-thickness, have their return current through signal pot_return. This is done to assure that no potentiometer is missing or partly disconnected so that it can yield a value leading to improper stand operation. During stand configuration or after replacing a poten­tiometer or the compression unit, the correct value for the signal potentiometer check
must be read and stored in the stand’s E
2
PROM with the service PC program. This error
is reported if the measured value is more than 32 bits less than the programmed value.
Fault elimination, troubleshooting
Check that all potentiometers are functioning properly. An error can be caused by an open connection, connector or potentiometer. If all four potentiometers are functioning properly, check the programming of the potentiometer check and correct it. The need for such an adjustment may indicate a potentiometer is about to fail.
Err 826
“There is a short circuit somewhere among potentiometers in stand”
Description
The same circuitry as Err 825 is used. In this case the error indicates that the value is more than 32 bits greater than the programmed value.
Fault elimination, troubleshooting
Check the actual and programmed pot. check values with the service PC program. If the value is only slightly too high, check all potentiometers for mechanical damage or sensitiv­ity. A short circuit (pot. or wiring) should yield the max. value for pot_check (3FF).
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 30 of 100
Medical Solutions
Page 31
Error messages 31
Err 829
“Collimator motor communication error”
Description
Communication between the stand and collimator control board D814 is erroneous or dis­rupted.
Fault elimination, troubleshooting
Check cables and plugs. Replace D814.
Err 830
“Filter motor communication error”
Description
Communication between the stand and collimator control board D814 is erroneous or dis­rupted.
Fault elimination, troubleshooting
Check cables and plugs. Replace D814.
Err 831
“Paddle decoder communication error”
Description
Identification of compression paddle fails.
Fault elimination, troubleshooting
Clean the sensor at the paddle. Check whether the paddle code is working with another paddle.
Err 850
“Write/read back failure for table 1 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
R is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the stand installation menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 851
“Write/read back failure for table 2 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the calibrate compression menu, the above error message is displayed.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 31 of 100
Page 32
32 Error messages
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 852
“Write/read back failure for table 3 in stand EEPROM”
Description
Each time the MAMMOMAT Novation EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the rotation calibration menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 853
“Write/read back failure for table 4 in stand EEPROM”
DR
is switched on, the adjustment data in the stand
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the rotation calibration menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 854
“Write/read back failure for table 5 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the lift parameters menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 855
“Write/read back failure for table 6 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the best compression menu, the above error message is displayed.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 32 of 100
Medical Solutions
Page 33
Error messages 33
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 856
“Write/read back failure for table 7 in stand EEPROM”
Description
Each time the MAMMOMAT Novation EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the rotation parameters menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 857
“Write/read back failure for table 8 in stand EEPROM”
DR
is switched on, the adjustment data in the stand
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the beam limiting device menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 858
“Write/read back failure for table 9 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the rotation parameters menu, the above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 859
Write/read back failure for table 10 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has its own checksum. If a discrepancy is detected in the rotation calibration menu, the above error message is displayed.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 33 of 100
Page 34
34 Error messages
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 860
“Write/read back failure for table 11 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for small focus Mo in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 861
“Write/read back failure for table 12 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for large focus Mo in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 862
“Write/read back failure for table 13 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for small focus W in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 863
“Write/read back failure for table 14 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for large focus W in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 864
“Write/read back failure for table 15 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for the light field focus in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 34 of 100
Medical Solutions
Page 35
Error messages 35
Err 865
“Write/read back failure for table 16 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for wing differences in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 866
“Write/read back failure for table 17 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the filter in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 867
“Write/read back failure for table 18 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the optimum compression function in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 868
“Write/read back failure for table 19 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the optimum compression function in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 35 of 100
Page 36

36 Log and error files

3Log and error files
4-

General 0

Some files on Mammomat Novation system contain information about processes on the system. This information could be helpful for determination of failure, detector artifacts, communications problems, blue screen and automatic reboot of the WH AWS.The analy­sis of the files requires specific knowledge. For this reason, if support is required, the files have to be sent to USC/HSC.
If support by USC/HSC is requested, groups of files have to be sent for explicit analysis:
MPxxxx.smj, MPxxxx.map, qualify.log,MPxxxx.cfg and drapi.log: They are locat-
ed under WH AWS in folder C:\AWS\drul\data\.
Brick files: These files can be shown by using the PCMX Web software.
Savelog files: a feature on the syngo SW is implemented for extracting the files and
burning them onto CD (only as of SW VA11x).
Windows Event log.
syngo logs. They are located under WH AWS in folder C:\AWS\log\*.*
Dump files (if available). They are located under WH AWS in folder C:\winnt\
minidump\*.dmp.
brickspy.tar.gz file.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 36 of 100
Medical Solutions
Page 37
Log and error files 37

SW Version VA10x 0

These instructions describe extracting log files for WH AWS with SW VA10x.

Copy MPxxxx files (detector files), dump files and syngo logs 0

These files have to be copied to a defined folder to ensure a correct copy.
Work steps
The syngo AWS application is running.
Select Service and Local Service in the menu bar under Options
Fig. 1:
Enter the service key and press OK. The Home menu appears.
Select Utilities
Select Escape to OS
Enter start explorer under parameters.
Command: NT command Interpreter
Parameters: start explorer
Press the enter key.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 37 of 100
Page 38
38 Log and error files
Windows Explorer opens.
Fig. 2: Detector files
Under C:\AWS\drul\data, select the files Drapi.log, MPxxxx.cfg, qualify.log,
MPxxxx.map and MPxxxx.smj,(1/Fig.2/p.38)click the right mouse button and copy and paste to folder C:\AWS\Service\extract.
Select C:\winnt\minidump\*.dmp (if available), right-click, and copy and paste to fold-
er C:\AWS\Service\extract.
Under C:\AWS\log , select all files (*.*), right-click them and copy and paste them to
folder C:\AWS\Service\extract.
Close Explorer.
Close Explorer.
Select Home
Close the Service software window

Copy Array and Brick log files 0

Use the Web Brick window to display these files.
Work steps
Array files
The syngo AWS application is running.
Select Service and Local Service in the menu bar under Options
Enter the service key and press OK. The Home menu appears.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 38 of 100
Medical Solutions
Page 39
Log and error files 39
Enter http:brick under address.
Fig. 3: PXCM Web Control Interface
Select array(1/Fig.3/p.39)
Select Log File(s)
Fig. 4:
The user name and appropriate password have to be entered (Fig.4/p.39)
Login: root
Password: *****
Press OK
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 39 of 100
Page 40
40 Log and error files
Select All Log Files
Fig. 5:
Right-click syslog and select Save Target As (Fig.5/p.40)
Save the file as array_syslog in the directory C:\AWS\Service\extract.
Right-click syslog.0 and select Save Target As .
Save the file as array_syslog.0 in the directory C:\AWS\Service\extract.
Brick files
Select Brick(2/Fig.3/p.39)
Select Log File(s)
Select All Log Files
Right-click syslog and select Save Target As
Save the file as brick_syslog in the directory C:\AWS\Service\extract.
Right-click syslog.0 and select Save Target As
Save the file as brick_syslog.0 in the directory C:\AWS\Service\extract.
Close the Web Brick window.
Close the Service software,

Windows Eventlog 0

Work steps
The syngo AWS application is running.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 40 of 100
Medical Solutions
Page 41
Log and error files 41
Select Service and Local Service in the menu bar under Options
Fig. 6:
Enter the service key and press OK. The Home menu appears.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 41 of 100
Page 42
42 Log and error files
Select EventLog on the right upper edge of the menu bar.
Fig. 7: EventLog_System
Select the following parameters as shown in (Fig.7/p.42) .
Domain: System(1/Fig.7/p.42)
Severity: All(2/Fig.7/p.42)
Extended: selected(3/Fig.7/p.42)
NT-Message: selected(3/Fig.7/p.42)
Message limit: empty(4/Fig.7/p.42)
Extract to file: System.txt ((5/Fig.7/p.42)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual date)).(6/Fig.7/p.42)
Select GO.(7/Fig.7/p.42)
The log file appears.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 42 of 100
Medical Solutions
Page 43
Log and error files 43
Fig. 8: EventLog_Application
Select the following parameters as shown in (Fig.8/p.43) .
Domain: Application(1/Fig.8/p.43)
Severity: All(2/Fig.8/p.43)
Facility: Developer(3/Fig.8/p.43)
Extended: selected(4/Fig.8/p.43)
NT Message: selected(4/Fig.8/p.43)
Message limit: empty(5/Fig.8/p.43)
Extract to file:Application.txt(6/Fig.8/p.43)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date))(7/Fig.8/p.43)
Select GO.(8/Fig.8/p.43)
The log file appears.
Brickspy.tar.gz file
Use telnet to start a script.
Select Service and Local Service in the menu bar under Options
Select Utilities
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 43 of 100
Page 44
44 Log and error files
Select Escape to OS
Enter start telnet brick under parameters
Login: root (and press enter)
Password: brick (and press enter)
From the brick# command prompt, execute the following command:
brickspy all (use lowercase letters and press enter)
Wait until the script is finished and enter Exit (and press enter)
Use ftp to get the saved file.
Select Service and Local Service in the menu bar under Options
Select Utilities
Select Escape to OS
Enter start cmd under parameters
At the DOS prompt, enter cd C:\AWS\Service\extract (and press enter)
At c:\temp, enter ftp brick (and press enter)
User: ftp (and press enter)
Password: a random e-mail address, e.g.. test@siemens.com
At the ftp prompt, enter bin (and press enter)
Enter prompt (and press enter)
Enter cd tmp (and press enter)
Enter dir (and press enter)
The file brickspy.tar.gz will be displayed
Enter mget brick* (and press enter)
After the file has been transferred, enter quit to exit ftp.
The file brickspy.tar.gz can be found on WH AWS under C:\Temp.
The log files are now stored in the folder C:\AWS\Service\extract and can be copied to an external storage medium, e.g., USB stick. If no external drive is available, the log files should be saved to CD. For this procedure, do the following:
Creating a new backup package
Select Local Service/Configuration/(Backup/Restore)
Enter a new package name(e.g.:ServiceLogFiles) in the blank line below
Package(1/Fig.9/p.45)
Click add(2/Fig.9/p.45)
In the blank line “file &dir”, enter the name: C:\AWS\Service\extract(3/Fig.9/p.45)
Click add after each entry(4/Fig.9/p.45).
The new file or path selection is now visible in the files & directories list box.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 44 of 100
Medical Solutions
Page 45
Log and error files 45
Fig. 9: Backup Package
Click save(5/Fig.9/p.45).
Back up your created package to CD.
Select Local Service/Backup & Restore.
Under Command: select Backup.
Select the CD-R drive.
Select your own created backup package: ServiceLogFiles.
Click Go.
The backup process to CD will be started!
After you have completed the backup, check your CD with the stored log files.
The configured ServiceLogFiles package is stored on CD compressed as file type “*.ar”. You can use the syngo arviewer.exe to read or extract these files. For more information, contact CS HSC 24 or your local USC.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 45 of 100
Page 46
46 Log and error files

SW Version VA11x 0

These instructions describe extracting log files for WH AWS with SW VA11x.

Copy MPxxxx files (detector files), dump files and syngo logs 0

These files have to be copied to a defined folder to ensure a correct copy on the CD.
Work steps
The syngo AWS application is running.
Select Service and Local Service in the menu bar under Options
Fig. 10:
Enter the service key and press OK. The Home menu appears.
Select Utilities
Select Escape to OS
Enter start explorer under parameters
Command: NT command Interpreter
Parameters: start explorer
Press the enter key.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 46 of 100
Medical Solutions
Page 47
Log and error files 47
Windows Explorer opens.
Fig. 11: Detector files
Under C:\AWS\drul\data, select the files Drapi.log, MPxxxx.cfg, qualify.log,
MPxxxx.map and MPxxxx.smj,(1/Fig. 11 / p. 47)right-click, and copy and paste to folder C:\AWS\service\html\whaws\extract.
Select C:\winnt\minidump\*.dmp (if available), right-click, and copy and paste to fold-
er C:\AWS\Service\html\whaws\extract.
Under C:\AWS\log , select all files (*.*), right-click them and copy and paste them to
folder C:\AWS\Service\html\whaws\extract.
Close Explorer.
Select Home
Close the Service software window
The storage files will be written onto CD by burning the SaveLogs files.

Copy Array and Brick log files 0

Use the Brick Configuration feature to display these files.
Work steps
Array files
The syngo AWS application is running.
Select Service and Local Service in the menu bar under Options
Enter the service key and press OK. The Home menu appears.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 47 of 100
Page 48
48 Log and error files
Select the Acquisition System button.
Fig. 12: PXCM Web Control Interface
Select Brick Configuration(1/Fig. 18 / p. 54)
Select array(1/Fig. 12 / p. 48)
Select Log File(s)
Fig. 13:
The user name and appropriate password have to be entered(Fig. 13 / p. 48)
Login: root
Password: *****
Press OK
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 48 of 100
Medical Solutions
Page 49
Log and error files 49
Select All Log Files
Fig. 14:
Right-click syslog and select Save Target As (Fig. 14 / p. 49)
Save the file as array_syslog in the directory C:\AWS\Service\html\whaws\extract.
Right-click syslog.0 and select Save Target As .
Save the file as array_syslog.0 in the directory C:\AWS\Service\html\whaws\extract.
Brick files
Select Brick
Select Log File(s)
The user name and appropriate password have to be entered
Login: root
Password: *****
Press OK
Select All Log Files
Right-click syslog and select Save Target As
Save the file as brick_syslog in the directory C:\AWS\Service\html\whaws\extract.
Right-click syslog.0 and select Save Target As
Save the file as brick_syslog.0 in the directory C:\AWS\Service\html\whaws\extract.
Close the Web Brick window.
Select Home in the Service software
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 49 of 100
Page 50
50 Log and error files
Copied files will be written by burning the SaveLogs to CD

Windows Eventlog 0

Work steps
The syngo AWS application is running.
Select Service and Local Service in the menu bar under Options
Fig. 15:
Enter the service key and press OK. The Home menu appears.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 50 of 100
Medical Solutions
Page 51
Log and error files 51
Select EventLog on the right upper edge of the menu bar.
Fig. 16: EventLog_System
Select the following parameters as shown in (Fig. 16 / p. 51) .
Domain: System(1/Fig. 16 / p. 51)
Severity: All(2/Fig. 16 / p. 51)
Extended: selected(3/Fig. 16 / p. 51)
NT Message: selected(3/Fig. 16 / p. 51)
Message limit: empty(4/Fig. 16 / p. 51)
Extract to file: System.txt ((5/Fig. 16 / p. 51)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual date)).(6/Fig. 16 / p. 51)
Select GO.(7/Fig. 16 / p. 51)
The log file appears.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 51 of 100
Page 52
52 Log and error files
Fig. 17: EventLog_Application
Select the following parameters as shown in (Fig. 17 / p. 52) .
Domain: Application(1/Fig. 17 / p. 52)
Severity: All(2/Fig. 17 / p. 52)
Facility: Developer(3/Fig. 17 / p. 52)
Extended: selected(4/Fig. 17 / p. 52)
NT Message: selected(4/Fig. 17 / p. 52)
Message limit: empty(5/Fig. 17 / p. 52)
Extract to file: Application.txt(6/Fig. 17 / p. 52)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date))(7/Fig. 17 / p. 52)
Select GO.(8/Fig. 17 / p. 52)
The log file appears.
Under C:\AWS\Service\extract, select the files Date.txt (e.g.:mm_dd_yyyy.txt) and
Application.txt and copy and paste them to folder C:\AWS\Service\html\whaws\ex­tract
Brickspy.tar.gz file
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 52 of 100
Medical Solutions
Page 53
Log and error files 53
Use telnet to start a script.
Select Service and Local Service in the menu bar under Options
Select Utilities
Select Escape to OS
Enter start telnet brick under parameters
Login: root (and press enter)
Password: brick (and press enter)
From the brick# command prompt, execute the following command:
brickspy all (use lowercase letters and press enter)
Wait until the script is finished and enter Exit (and press enter)
Use ftp to get the saved file.
Select Service and Local Service in the menu bar under Options
Select Utilities
Select Escape to OS
Enter start cmd under parameters
At the DOS prompt, enter cd C:\AWS\Service\html\whaws\extract (and press enter)
At c:\temp, enter ftp brick (and press enter)
User: ftp (and press enter)
Password: a random e-mail address, e.g., test@siemens.com
At the ftp prompt, enter bin (and press enter)
Enter prompt (and press enter)
Enter cd tmp (and press enter)
Enter dir (and press enter)
The file brickspy.tar.gz will be displayed
Enter mget brick* (and press enter)
After the file has been transferred, enter quit to exit ftp.
The file brickspy.tar.gz can be found on WH AWS under C:\Temp.

Copy SaveLogs 0

This feature is implemented on the WH AWS starting with SW VA11x and can be used if necessary.
Work steps:
Select Service and Local Service in the menu bar under Options
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 53 of 100
Page 54
54 Log and error files
Enter the service key and press OK. The Home menu appears.
Fig. 18: Acquisition system
Select the Acquisition System button.
Click the Extract AWS Savelog button (2/Fig. 18 / p. 54).
Fig. 19:
The following mask appears: Extract Servicelog, Extract Drapi.logs, Burn files on CD,
etc.
Click Extract Service Logs (1/Fig. 19 / p. 54)
Within seconds, the extracted files are shown.
Click Extract Drapilogs (2/Fig. 19 / p. 54)
Within seconds, the extracted files are shown.
Insert an empty CD into the CD writer.
Click Burn files on CD (3/Fig. 19 / p. 54).
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 54 of 100
Medical Solutions
Page 55
Log and error files 55
The following message appears: "Insert a blank CD into the CD-Writer and press OK or
press Cancel to abort“
Press OK
After the burn procedure, the CD is ejected automatically..
If artifacts are visible in some images, these images (row data and processed data) can be stored on a second CD for explicit fact-finding in the factory.
Make a test image as described below:
Test images
Install the compression plate simulator; make sure that the collimator is open to its full
size
Install the collimator-mounted plexi phantom; make sure no other object is in the radia-
tion field
Select the "QC-RAW" technique
Take an image with 28kV and AEC
Save the image to the CD
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 55 of 100
Page 56

56 Generator checks

4Generator checks
5-

General 0

The generator section of the Mammomat Novation is of modular construction. The mod­ules were developed so that, in general, no settings need be made when they are replaced. If an error occurs in the generator (radiation generation), different signals can be measured or checked in order to delineate or to localize the error.

Documents required 0

Service Software, SPB7-250.816.03...
Wiring Diagrams, SPB7-250.844.01...
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 56 of 100
Medical Solutions
Page 57
Generator checks 57

Troubleshooting tips 0

Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 57 of 100
Page 58
58 Generator checks
Failure Check Measures to be taken if not OK
Control panel not working at switch-on
LEDs: “+5VD,
-15 V, +15 V and 28Vprim on mains input con­verter
If all LEDs are off, check if there is input voltage at the input terminals and if S700 is ON.
If there is voltage and S700 is ON, then the mains input converter is probably defective and must be replaced. Remember to switch off the input voltage before disconnecting this unit.
If only one or some of the LEDs are off, switch off S700 and disconnect the cables from X700, X701, X705 and X706. If the LEDs still are off when you switch on S700, then the mains input converter is probably defective and must be replaced. Remember to switch off the input voltage before disconnecting this unit.
If all LEDs are on when you switch on S 700, then the problem cable has to be identified. If the cable con­nected to X 700 is causing the failure, then the prob­lem is either the HV inverter or the tube power supply. If the failure remains after the cable to con­nector X720 on the tube power supply has been dis­connected, the tube power supply is probably defective and must be replaced. If the failure disap­pears after the cable to connector X720 has been disconnected, then the HV inverter is probably defec­tive and must be replaced. If the cable connected to X701 is causing the failure, then the master board is probably defective and must be replaced. If the cable connected to X706 is causing the failure, then the problem is either the AC inverter or the tube power supply. If the cable to connector X723 on the tube power supply is disconnected and the failure remains, then the AC inverter is probably defective and must be replaced.
If the failure disappears after the cable to connector X723 has been disconnected, then the tube power supply is probably defective and must be replaced.
Attention: connectors X706 and X723 and related cables are galvanic and connected to mains voltage.
Control panel not working at switch-on
LED “DC_BUS_OK” on mains input converter
If neither LED “DC_BUS_OK” or LED “DC_BUS_FAILURE” is lit, there is probably no signal from the master board to the mains input converter or from the control panel to master board.
If LED “DC_BUS_FAILURE” is lit, then the mains input converter is probably defective and must be replaced. Remember to switch off the input voltage before disconnecting this unit.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 58 of 100
Medical Solutions
Page 59
Generator checks 59
Failure Check Measures to be taken if not OK
Control panel not working at switch-on
Control panel not working at switch-on
LED “400VDC live H3”
LED “AC ON” on AC inverter
If LED H4 is lit but not “H3”, fuse F2 in the mains input converter is probably defective. Switch off the mains to the input, wait until LED “H4” is off, then remove the mains input converter and replace fuse F2. Put back the unit and connect the cables but before switching on the the mains; remove the cable from connector X712 on the AC inverter. Switch on the mains and the control panel and check if LED “H3” is lit.
If it is lit, switch off the mains to the input, wait until LED “H4” is off, then reattach the cable to X712 on the AC inverter. Switch on the mains and the control panel. If “H3” is still off, the AC inverter and fuse F2 have to be replaced. But remember to wait until “H4” is off before disconnecting cables.
If “H3” is off in the first place, the tube power supply is probably defective and and must be replaced along with fuse F2. But remember to wait until “H4” is off before disconnecting cables.
The AC inverter is probably defective and must be replaced.
If “AC ON” is OK, the failure must be located in the stand electrical devices.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 59 of 100
Page 60
60 Generator checks

Signals on D750 0

The actual values of the generator can be measured on board D750 or checked with the aid of the Mammomat Novation, Software System, SPB7-250.816.03 service program. Refer to software system Main menu-Normal mode->Actual values.
Measurement points on board D750 are shown in following picture:
Fig. 20: D750
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 60 of 100
Medical Solutions
Page 61
Generator checks 61
Test point Description Status
5VD_1 Logic supply voltage: 5 volt digital
0VD_1,2,3,4,5 Logic supply voltage: 0 volt digital
P15VA Logic supply voltage: Positive 15 volt analog
0VA_1,2,3 Logic supply voltage: 0 volt analog
N15VA Logic supply voltage: Negative 15 volt analog
5V_RS232 Logic supply voltage: 5 volt; for communication
with PXCM unit (Brick unit)
0V_RS232 Logic supply voltage: 0 volt; for communication
with PXCM unit (Brick unit)
5VI_RS232 Logic supply voltage: 5 volt; for communication
with service PC, Opdima WS and printer
0VI_RS232 Logic supply voltage: 0 volt; for communication
with service PC, Opdima WS and printer
M_WD_RESET_ Signal that resets the master CPU and causes
the entire Mammomat system to restart.
GEN_POWER_ONSignal to the generator that turns on the genera-
tor and the Mammomat system.
DC_BUS_OK Signal from the generator that indicates the
DC-bus voltage is within its allowed voltage win­dow.
HV_OFF_INT_ Indicates that the signal M_PLD_HV_OFF_ from
the PLD circuit has detected an event that will terminate or prevent high voltage generation. It causes the signal HV_ON from the master board to go inactive, which immediately prohibits high voltage generation.
REQ_HV_ON Signal from PLD and CPU that generates high
voltage and thus starts the exposure. Together with the signal GRID_O_HV_ON, it sets the sig­nal HV_ON.
HV_ON Signal set by the signals REQ_HV_ON and
GRID_O_HV_ON that are output to the genera­tor. One of two necessary signals, together with the signal EN_HV_ON, that generates high volt­age and thus starts the exposure.
EN_HV_ON Signal from the exposure release buttons at the
control panel to the generator. One of two neces­sary signals, together with the signal HV_ON, to generate high voltage and thus start the expo­sure.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 61 of 100
Page 62
62 Generator checks
Test point Description Status
GEN_POWERED_ONSignal to the PXCM unit (Brick unit) that the
power to the Mammomat system has been switched on by the operator.
PREP_DR_DET_ Signal to the PXCM unit (Brick unit). The positive
edge indicates to the Brick that the generator preparation phase to perform an exposure has started and requests that the DR detector pre­pare for image acquisition.
The negative edge of the signal comes 5 ms after the negative edge of the signal GEN_RDY_FOR_DR_EXP at normal exposure termination; if the hardware detects an error, the negative edge of the signals will coincide. If a tube arc occurs, the signal will not go inactive.
GEN_RDY_FOR_ DR_EXP
Signal to the PXCM unit (Brick unit). The genera­tor preparation phase has finished and the gen­erator is ready to generate high voltage to the tube (i.e., X-ray). The inactivation of the signal will coincide with the negative edge of the signal X_RAY_ON at normal termination of the expo­sure. If the exposure terminates abnormally, the inactivation of the signal coincides with the inac­tivation of signal PREP_DR_DET. If a tube arc occurs, the signal will not go inactive.
0 V <=> OK
4.3 V <=> Standby
0 V <=> Acquisition
4.3 V <=> Standby
0 V Acquisi­tion ON
X_RAY_ON Signal to the PXCM unit (Brick unit). The expo-
sure is in progress.
Note: The signal stays active - at the grid turning points when X-ray is actually off.
- at the first two tube arcs when X-ray is actually off.
DR_DET_RDY_F
The DR detector is acquiring image data. 5 V <=>
OR_EXP
DR_SYST_RDY The DROC of the DR system indicates that it is
ready for an exposure.
DR_DET_ERR_ The DR system indicates to the generator that it
has detected an error.
4.3 V <=> Standby
0 V <=> X-ray ON
Acquisition ON
5 V <=> OK
0 V <=>Not ready
0 V <=> OK
5 V <=>Error
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 62 of 100
Medical Solutions
Page 63
Generator checks 63

Tube high voltage, current and mAs values 0

This section describes how to check the X-Ray tube in regard to the:
High voltage, tube current, mAs value with and without AEC and with a small or large focus.
These values can be checked by measuring HV_ACT/MA_ACT on board D750 and/or by using the service PC -> SW -> Normal -> Actual values. Only one of them is necessary, but both methods are described in the following.
Preparation
Switch the system OFF (e.g., S700 on main input)
Connect the oscilloscope as follows:
Channel 1 to measurement point HV_ACT (1V=5 kV) on D750
Channel 2 to measurement point MA_ACT (1V= 40 mA) on D750.
Connect the service PC to the stand.
Switch the system ON.
Start the Service program and check in the anode menu whether the tungsten anode is
enabled.
In the Service program, select Normal->Main menu ->Normal mode->Actual values
Measurement method
The mAs value must be calculated as the product of the tube current and exposure time. The tube current and the exposure time can be obtained from the oscillograms. The dia­gram may have glitches, in which case you have to measure the exposure time t1 and t2.
Example:
Fig. 21:
In the figure, the tube current is 150 mA and the exposure time is t1+t2 in seconds.
The mAs is 150mA*(t1+t2)
NOTE
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
Be aware that with the AEC mode on, a pre-exposure of 50 ms is done.
Select the correct trigger on the scope so you get the pre-exposure and all the exposure times.
Page 63 of 100
10.05 CS SD 24
Page 64
64 Generator checks
The following oscillograms and values should be of assistance to you in troubleshooting the high-voltage circuit or radiation release.
Checks with a large focus and without AEC
Prerequisites
Deselect D and H on the control console; the Mammomat is in mAs mode without AEC.
Make sure the magnification table is removed so that a large focus is selected.
The check can be performed for each measurement of the table.
Measurement kV mAs Anode/Filter
13020Mo/Mo
23020W/Rh
3 30 100 Mo/Mo
4 30 100 W/Rh
Select the Examination tab card and load a patient for examination
Set the exposure values on the control console according to the table above.
Start an exposure.
Check and compare the values with the following diagrams.
The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
30 kV 20mAs Mo/Mo large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
40 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 170 ms
Fig. 22: 30 kV 20 mAs MoMo large
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 64 of 100
Medical Solutions
Page 65
Generator checks 65
30 kV 20mAS W/Rh large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T (1T<=>40 mA)
40 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 140 ms
Fig. 23: 30 kV 20 mAs W large
30 kV 100mAs Mo/Mo large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 950 ms
Fig. 24: 30 kV 100 mAs MoMo large
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 65 of 100
Page 66
66 Generator checks
30 kV 100mAs W/Rh large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 790 ms
Fig. 25: 30 kV 100 mAs W large
Checks with a large focus and with AEC
Prerequisites
Mount the collimator-mounted plexi phantom
Register a patient
Make sure the magnification table is removed so that a large focus is selected.
Select H on the control console; the Mammomat is now in AEC mode.
The check can be performed for each measurement of the table.
Measurement kV mAs Anode/Filter
130--Mo/Mo
230--W/Rh
Adjust the exposure values on the control console according to the table above.
Start an exposure.
Check and compare the values with the following diagrams.
The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 66 of 100
Medical Solutions
Page 67
Generator checks 67
30 kV AEC Mo/Mo large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 360 ms
Fig. 26: 30 kV AEC MoMo large
30 kV AEC W/Rh large focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 330 ms
Fig. 27: 30 kV AEC W large
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 67 of 100
Page 68
68 Generator checks
Checks with a small focus
NOTE
Prerequisites
These checks are possible only when a magnification table is available. Select the small focus. The magnification table is need­ed. This table is a option.
Make sure the magnification table is mounted so that a small focus is selected.
Set the exposure values on the control console according to the table below.
Start an exposure.
Check and compare the values with the following diagrams.
The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
Measurement kV mAs Anode/Filter
13020Mo/Mo
23020W/Rh
3 30 AEC Mo/Mo
4 30 AEC W/Rh
30 kV 20mAs Mo/Mo small focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
24 mA 0.2V/T(1T<=> 8 mA)
825 ms 100 ms/T
Service->Normal->Act. values
Act. tube current 24 mA
Act. exp. time 830 ms
Fig. 28: 30 kV 20 mAs MoMo small
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 68 of 100
Medical Solutions
Page 69
Generator checks 69
30 kV 20mAs W/Rh small focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
30 mA 0.2V/T(1T<=> 8 mA)
680 ms 100 ms/T
Service->Normal->Act. values
Act. tube current 30 mA
Act. exp. time 690 ms
Fig. 29: 30 kV 20 mAs W small
30 kV AEC Mo/Mo small focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
24 mA 0.2V/T(1T<=> 8 mA)
400 ms/T
Service->Normal->Act. values
Act. tube current 24 mA
Act. exp. time 1250 ms
Fig. 30: 30 kV AEC MoMo small
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 69 of 100
Page 70
70 Generator checks
30 kV AEC W/Rh small focus Values
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
30 mA 0.2V/T(1T<=> 8 mA)
400 ms/T
Service->Normal->Act. values
Act. tube current 30 mA
Act. exp. time 1090 ms
Fig. 31: 30 kV AEC W small
If there is an error in the high-voltage circuit, one of the following components could be defective:
X-ray tube
HV tank
HV inverter module
AC inverter module
Inverter current
This is an analog signal proportional to the output inverter current fed to the HV tank.
I-OUT 30 kV 100mAs W/Rh large focus Values
Module: HV inverter
Measurement point I_OUT
I-OUT
0.5 V/T
100 ms/T
Fig. 32: I-out, 30 kV 100 mAs W large
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 70 of 100
Medical Solutions
Page 71
Generator checks 71
If no signal can measured at I_OUT due to an error, then one of the following components could be defective:
X-ray tube
HV tank
HV inverter
Mains input converter
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 71 of 100
Page 72
72 Generator checks

Filament circuit 0

The Service SW, software system, is used for setting the filament current values.
Work steps
Connect the Service PC.
Start the SW.
Turn the system ON.
Select Main menu->Configuration->Filament.
Fig. 33: Filament
Indicates the adapted filament current of available anode materials for large and small focus. Adaptation exposures for the filament current are also made in this menu. Pressing F3 sets the exposure parameters on the control panel, and subsequently pressing Alt-F6 starts the exposure. Repeat this procedure until the tube current deviation is within ±5%. If the tube has been replaced, program the filament current value to be 200 mA less than the value stated in the test certificate as a pre-adjustment. (e.g., if tube value is 6900, pro­gram 6700.)
If a new filament current has been entered, you must save the new value (press F2) before learn filament is performed.
NOTE
The Dynamic Learn Filament may change this adjustment. But the DLF will learn only when the exposure time is longer than 60 ms.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 72 of 100
Medical Solutions
Page 73
Generator checks 73
Fig. 34: Miscellaneous, DLF switch
Activation and deactivation of the filament current adaptation. When activated, the adap­tation takes place after a total of ten exposures have been made with the actual focus, if the deviation of the average value from the nominal value is not within -15% to + 5%.
NOTE
Test points on D750
The filament current values can also be measured on board D750.
FIL_CURR_SET Analog output signal to the generator for setting the magnitude of the fil­ament current. The relation between the filament current and the analog control signal is1A/V.
FIL_CURR_ACT Analog input signal FIL_CURR_ACT, from the generator, that is propor­tional to the true filament current. The relation between the filament current and the ana­log signal is 1A/V.
Independent of the selected focus, the current value is 3.5 A (At test point D750 FIL_CURR_ACT = 3.5 V).
During preparation or exposure, the value is increased to approx. 6.6 A to 6.9 A according to selected focus.
If no filament current is present, there could be an error on one of the following compo­nents:
Filament adaptations are possible only when the detector wing is selected. "dr" is displayed on the control panel.
X-ray tube
Tube power module
AC inverter module
Cable between X-ray tube and tube power module
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 73 of 100
Page 74

74 Image quality

5Image quality
6-

General 0

The detector system is an important component of the Mammomat Novation and is key to image quality. There are 2 detector calibration procedures that should be used if image artifacts (pixel failures, bad lines, etc.) occur.
Gain calibration
Bad pixel mapping
In case of image quality problems, defined test images and files are needed for analysis:
MPXXXX.map, MPXXXX.cfg, quality.log, MPXXXX.smj
Test image: 4 cm Plexi, 28 kV, AEC, QC Raw
Calibration image
These images and files have to be supplied to USC/HSC for support.
Documents required
Startup Instructions, SPB7-250.815....
Rep. of parts, SPB7-250.841.01...
Rep. of parts, SPB7-250.841.02..
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 74 of 100
Medical Solutions
Page 75
Image quality 75
Gain calibration 0
A gain calibration should be performed first. If the image artifacts persist, then perform bad pixel mapping.

Gain calibration

Creates the offset for each pixel value.
Filename:<Serial_Number>.map
If there is no gain calibration file that matches the current array serial number, only cali-
bration scans and service scans are allowed.
Furthermore, the gain calibration procedure must be performed weekly. This procedure can be performed by the customer or the service engineer.
For SW version VA10x, see the user manual.
Fig. 35: _select_calibration
Eight exposures (accepted) have to be taken.
Work steps for SW VA11x
Select Examination->Calibration
Select 28 kV and 250 mAs.
Select the Anode/Filter combination the customer uses most frequently.
Mount the collimator-mounted plexi phantom.
Be sure that no additional objects are in the radiation field.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 75 of 100
Page 76
76 Image quality
During the calibration procedure, an image can be seen on the monitor, shown in the fol­lowing example:
Fig. 36: Normal Calibration
This image is normal during the calibration and must not be interpreted as an error.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 76 of 100
Medical Solutions
Page 77
Image quality 77

Bad pixel mapping 0

A new pixel map should be created if excessive pixel dropout occurs repeatedly on images displayed in the Examination tab card(Fig. 35 / p. 75). The procedure has to be performed by service and only when a defective line or pixel occurs.
Work steps
Select Options -> Service -> Local Service
Enter the service key and press OK.
Select Acquisition System.
Select Examination -> Calibration in the menu bar.
Select Bad Pixel Mapping.
Select 28 kV and 250 mAs.
Select the Anode/Filter combination the customer uses most frequently.
Insert the collimator-mounted plexi phantom.
One exposure has to be performed.
Be sure that no additional objects are in the radiation field.
Once accepted, the defective line(s) or pixel(s) is/are not released again.
If the line/pixel is still visible after bad pixel mapping calibration, try to lower the mAs value by 50 mAs, e.g., 200 mAs.
NOTE
NOTE
NOTE
NOTE
Bad pixel mapping cannot be repeated without restoring the old mapping files. Bad pixel mapping is a continuous procedure. Bad results remain in the mapping file. There is no undo procedure. Contact the Uptime Support Center.
The bad pixel mapping procedure cannot be performed until the detector has been powered on for at least one hour to allow it to warm up.
To go back, the original mapping files from the supplied detec­tor-specific CD must be loaded.
Bad pixels after bad pixel mapping can result from an incorrect ad­justment of the collimator setting or wrong X-ray tube settings. In case of bad pixel(s), also check the collimator adjustment and X-ray tube settings.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 77 of 100
Page 78
78 Image quality

Qualify log 0

Description
An important file of the detector system for image quality is the qualify.log. This provides information about possible pixel failures in the detector. After every bad pixel mapping, this file is overwritten with the current data of the detector. This includes possible pixel fail­ures, bad areas or defective lines.
The qualify.log can be found in folder C:\AWS\drul\data or using the Service SW, Home->Acquisition System->Bad Pixel Mapping.
Images defects in the raw, uncorrected image inside the imageable area must meet the following performance criteria:
Not to exceed a total of 2936 defective pixels exclusively or dead lines which corre-
spond to 0.02% of the total amount of pixels.
Have zero 3x3 pixel areas containing more than 8 defective pixels
Have zero 5x5 pixel areas containing more than 15 defective pixels.
Not to exceed a total number of 80 pixel clusters (2x2) where all 4 pixels are defective.
Not to exceed a total number 20 defective lines.
Have zero adjacent defective lines (e.g., two defective lines must be separated by at
the least one good line)
The following examples show qualify logs for good and defective detectors:
Fig. 37: Qualify.log
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 78 of 100
Medical Solutions
Page 79
Image quality 79
Fig. 38: Bad qualify.log
If the check is not successful (fails) and most dropouts are at the image edges, check the collimator adjustment. Refer to Collimator Adjustment, System Startup, SPB7-250.815.05...
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 79 of 100
Page 80
80 Image quality

Image examples 0

The following image examples show different effects that can occur on the Mammomat Novation. If one of these effects or artifacts can be seen, perform a gain calibration and, if required, bad pixel mapping. For grid stripes, check the mechanical grid settings. In some image examples, the details cannot be recognized quite accurately (pixel failures, grid stripes). The online version (HTML version) of this document shows higher perceptibility compared to the printed version. Should the recommended measures be unsuccessful, then contact USC/HSC. The detector may need to be replaced.
Collimator problems Description Actions
Fig. 39: Bad collimator
Shadow on the chest wall side
1. Bad line
Perform collimator adjust­ment according to Startup Instructions, SPB7-250.815... Collima­tor Adjustment chapter.
1. Perform gain calibration procedure
Perform bad pixel mapping correction.
2. Perform collimator
2. Shadow on chest wall side
Fig. 40: Bad line 1 cm
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 80 of 100
adjustment according to Startup Instructions, SPB7-250.815... Collima­tor Adjustment chapter.
Medical Solutions
Page 81
Image quality 81
Line problems Description Actions
1 cm line (1) Perform gain calibration
procedure.
If not successful, perform bad pixel mapping correc­tion.
If not successful, the detector may need to be replaced. Contact USC/HSC.
Fig. 41: Bad line 1 cm
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 81 of 100
Page 82
82 Image quality
Line problems Description Actions
1 mm line (1) Perform gain calibration
procedure.
If not successful, perform bad pixel mapping correc­tion.
If not successful, repeat bad pixel mapping with lower mAs values (e.g.: 200mAs).
If not successful, the detector may need to be replaced. Contact USC/HSC.
Fig. 42: Bad line 1 mm
Bad line (1) Perform an exposure with-
out the phantom and check whether the line(s) or bad pixels are located at the same position.
Fig. 43: Bad line into RMI156
If yes, perform the gain calibration procedure.
If not successful, perform bad pixel mapping correc­tion.
If not successful, the detector may need to be replaced. Contact USC/HSC.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 82 of 100
Medical Solutions
Page 83
Image quality 83
Quadrant failure Description Actions
Fig. 44: Quadrant failure 1
Quarter panel defect (lower right side)
Quarter panel defect (upper left side)
Restart the whole system
If not successful, the detector may need to be replaced. Contact USC/HSC.
Restart the whole system.
If not successful, the detector may need to be replaced. Contact USC/HSC.
Fig. 45: Quadrant failure 4
Fig. 46: Quadrant failure
Quarter panel defect (lower right side)
Restart the whole system.
If not successful, the detector may need to be replaced. Contact USC/HSC.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 83 of 100
Page 84
84 Image quality
Problems during calibration Description Actions
Fig. 47: Bad lines during calibration
Bad lines appear during calibration
Bad lines appear during calibration
After gain calibration, turn the whole system OFF.
Turn the system ON and repeat the gain calibration.
After gain calibration, turn the whole system OFF.
Turn the system ON and repeat the gain calibration.
Fig. 48: Bad lines during calibration
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 84 of 100
Medical Solutions
Page 85
Image quality 85
Pixel dropouts Descrip-
tion
Defective pixels or lines on image
Actions
Perform an exposure without the phantom and check whether the line(s) or bad pixels are located at the same position.
Perform gain calibra­tion procedure.
If not successful, per­form bad pixel mapping correction.
Check qualify.log file whether map file is within specification (pass).
If not, the detector may need to be replaced. Contact USC/HSC.
Fig. 49: Bad pixel 1
Fig. 50: Pixel problem
Defective pixels (1) on image
Perform an exposure without the phantom and check whether the line(s) or bad pixels are located at the same position.
Perform gain calibra­tion procedure.
If not successful, per­form bad pixel mapping correction.
Check qualify.log file whether map file is within specification (pass).
If not, the detector may need to be replaced. Contact USC/HSC.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 85 of 100
Page 86
86 Image quality
Grid lines Description Actions
Fig. 51: Raster6
Grid lines are visible on image.
Check exposure parameters: If exposure time is very short (e.g.: pre-shot), grid lines can be visible on image. In this case, select the appropriate exposure parameters.
To eliminate grid lines, check the grid adjustment according to the Startup Instructions, SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the “wheel”
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replace­ment of Parts, SPB7-250.841.02.01.02.
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 86 of 100
Medical Solutions
Page 87
Image quality 87
Grid lines Description Actions
Fig. 52: Raster8
Grid lines are visible on image.
Grid lines are visible on image.
Check exposure parameters: If exposure time is very short (e.g.: pre-shot), grid lines can be visible on image. In this case, select the appropriate exposure parameters.
To eliminate grid lines, check the grid adjustment according to the Startup Instructions, SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the “wheel”
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replace­ment of Parts, SPB7-250.841.02.01.02.
Check exposure parameters: If exposure time is very short (e.g.: pre-shot), grid lines can be visible on image. In this case, select the appropriate exposure parameters.
Fig. 53: Raster5
To eliminate grid lines, check the grid adjustment according to the Startup Instructions, SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replace­ment of Parts, SPB7-250.841.02.01.02.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 87 of 100
Page 88
88 Image quality
Bad paddle Description Actions
Fig. 54: Bad paddle
Edge of com­pression pad­dle visible on image
Replace compression paddle
Adjust the whole compression unit according to Replacement of Parts SPB7-250.841.01...
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 88 of 100
Medical Solutions
Page 89
Image quality 89
Heel effect Description Actions
Decrease of beam intensity to the stand side.
Problems dur­ing “Bad Pixel Mapping”.
Perform an exposure with the other com­pression pad­dle.
If not success­ful, check colli­mator adjustment.
If not success­ful, the X-ray tube may need to be replaced. Contact USC/HSC.
Fig. 55: Heel effect
Fig. 56:
Decrease of beam intensity to the stand side.
Perform an exposure with the other com­pression paddle
If not success­ful, check colli­mator adjustment.
If not success­ful, the X-ray tube may need to be replaced. Contact USC/HSC.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 89 of 100
Page 90
90 Image quality
Microphony Description Actions
Fig. 57: Raster7
Lines and arti­facts visible on image. Detec­tor motions dur­ing acquisition of image data
Lines and arti­facts visible on image. Detec­tor motions dur­ing acquisition of image data
Repeat image acquisition and be sure there is no motion during image data transfer.
If successful, inform the customer about handling during data acquisition.
Repeat image acquisition and be sure there is no motion during image data transfer.
If successful, inform the customer about handling during data acquisition.
Fig. 58: Microphony
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 90 of 100
Medical Solutions
Page 91
Image quality 91
Starry sky Description Actions
Fig. 59: Starry sky 1
Many spots are visible on image
Many spots are visible on image
Reboot the whole system.
If not successful, perform an exposure with the collima­tor-mounted plexi-phantom and QC-Raw.
If not successful, perform gain calibration procedure.
If not successful, perform bad pixel mapping correc­tion.
If not successful, the detec­tor may need to be replaced. Contact USC/HSC.
Reboot the whole system.
If not successful, perform an exposure with the collima­tor-mounted plexi-phantom and QC-Raw.
If not successful, perform gain calibration procedure.
Fig. 60: Starry sky 2
If not successful, perform bad pixel mapping correc­tion.
If not successful, the detec­tor may need to be replaced. Contact USC/HSC.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 91 of 100
Page 92
92 Image quality
“Scrambled” Image Description Actions
Fig. 61:
Image information “scrambled” on image.
Reboot the whole system (including Brick)
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 92 of 100
Medical Solutions
Page 93
Image quality 93
AEC/Dose problems Description Actions
Fig. 62: Bad exposure 2
Dose for exposure too low, or acquisition aborted by user, or acquisition was termi­nated before exposure was completed.
Dose for exposure too low, or acquisition aborted by user, or acquisition was termi­nated before exposure was completed.
Check exposure parame­ters. Exposure time prob­ably too short
Check generator values according to the Trouble­shooting Guide, “Genera­tor checks” chapter, (these instructions.)
Check exposure parame­ters. Exposure time prob­ably too short
Check generator values according to the Trouble­shooting Guide, “Genera­tor checks” chapter, (these instructions.)
Fig. 63: Bad exposure dose1
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 93 of 100
Page 94

94 Compression Unit

6Compression Unit
7-

General 0

These instructions describe the work steps for troubleshooting the compression unit:
Check the adjustments (see Replacement of Parts, SPB7-250.841.01..)
Check the potentiometer R863 (Position transducer of the compression unit)
Check the potentiometer R861 (compression force)
Check the movements(up/down)
Check the detection of the compression paddles

Documents required 0

Replacement of Parts, SPB7-250.841.01..
Wiring Diagrams, SPB7-250.844.01...
Service SW Stand, SPB7-250.816.03..
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 94 of 100
Medical Solutions
Page 95
Compression Unit 95

Checks 0

Check the adjustments 0

For the compression unit to function properly, it is very important that it is adjusted according to the instructions described in Replacement of Parts, SPB7-250.841.01...

Check the potentiometer R863 (position transducer) 0

Attach a compression paddle to the compression unit.
Move the compression paddle up/down.
The display for thickness changes the values during movement.
Place a defined object thickness onto the detector, e.g.: 40 mm Plexi-phantom.
Move down the compression paddle by means of the foot switch until the paddle reach-
es the object on the detector.
The value on the thickness display (D803) has to be the same as the object thickness.
If not, adjust the compression unit according to Replacement of Parts, SPB7-250.841.01.. If the potentiometer still does not work properly after the adjustment, the potentiometer might be defective and need to be replaced.

Check the potentiometer R861 (compression force) 0

Turn the potentiometer clockwise to a defined position, e.g.: 8 kp (see display on the
lower position of stand).
Place a scale with a digital display onto the detector and move down the compression
paddle until the movement stops automatically.
Read the force on the display of the scale. If a deviation >0.3 kp between the values oc-
curs, perform a force adjustment according to Replacement of Parts, SPB7-250.841.01..
If adjustment is not possible, potentiometer R861 might be defective and need to be replaced or the compression unit is defective.

Check the movements (up/down) 0

Press the compression pedals (up/down) for electrical movement of the compression
paddle. The paddle moves up/down.
If electrical movement is not possible, check the power supply:
Check 14 volt supply on board D802
If this voltage is not present, fuse F1 (5AT) on D805 or F803 might be defective or D805 is defective or motor of compression unit is defective.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 95 of 100
Page 96
96 Compression Unit

Check the detection of compression paddle. 0

One condition for an exposure release is the detection of a compression paddle. This will occur by attaching the compression paddle in the guide shaft of the compression unit. The detection of compression paddles can be checked using the Service SW Stand, SPB7-250.816.03.. Chapter Service=>Stand test.
Work steps
Connect service PC and start Service SW of stand, see Software System Stand,
SPB7-250.250.816.03..
Fig. 64: Stand test paddleID
Select Main menu=>Service=>Stand test PaddleID
Attach a compression paddle to the compression unit and start the test with the service
SW.
A paddle ID will be shown on PC display.
Repeat this procedure with each compression paddle.
A list with the corresponding PaddleID is shown below:
Compression plate Art.no Plate ID Object Table
Comp. plate simulator 10048527 00 Detector
Comp. plate 20x22H 10048520 01 Detector
Comp. plate 20x22H (Flex) 10048521 02 Detector
Comp. plate 18x24 (High Edge) 10048519 03 Detector
Comp. plate 18x24 (Low Edge) 10048516 04 Detector
Comp. plate 18x24 (Flex) 05 Detector
Comp. plate 24x30 (High Edge) 10048515 06 Detector
Comp. plate 24x30 (Low Edge) 10048518 07 Detector
Comp. plate 24x30 (Flex) 10048517 08 Detector
Comp. plate Mag.1.5 10048522 09 Mag. Table 1.5
Comp. plate Mag. 1.8 10048523 10 Mag. Table 1.8
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 96 of 100
Medical Solutions
Page 97
Compression Unit 97
Compression plate Art.no Plate ID Object Table
Comp. plate Spot 1.0 10048513 11 Detector
Comp. plate Spot 1.5 10048524 12 Mag. Table 1.5
Comp. plate Spot 1.8 10048525 13 Mag. Table 1.8
Shadow cross 14 Detector
Comp. plate Stereo 10048514 15 Stereo
Biopsy Hole Plate 10046676 04 Detector
No plate attached 99 Detector
If one of the paddles cannot be detected (ID 99), the paddle might be defective and need to be replaced.
If none of the paddles can be detected, a problem occurred on the compression unit or D805 or the ribbon cable between both:
Fig. 65: D805
Check the ribbon cable between guide shaft and D805 X831(1/Fig. 65 / p. 97). The
connector on the ribbon cable (2/Fig. 65 / p. 97) has to be plugged in so that the "nose" on the connector (1/Fig. 66 / p. 98), and the pins are visible from above; they are on the upper side.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 97 of 100
Page 98
98 Compression Unit
Fig. 66: X831
Measurement on D801
If none of the paddles can be detected, a pulse sequence can be checked at TP SMART_RX on D801:
When no paddle is attached on the compression unit, a short pulse sequence can be
measured (Fig. 67 / p. 98).
D801 TP SMART_RX
1 V/T
10 ms/T
Fig. 67: Pulse sequence without compression
paddle
When a paddle is attached on the compression unit, a longer pulse sequence can be measured than without a paddle (Fig. 68 / p. 99)
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 98 of 100
Medical Solutions
Page 99
Compression Unit 99
D801 TP SMART_RX
1V/T
10 ms/T
Fig. 68: Pulse sequence with compression plate
simulator
When no pulse sequence occurs, board D805 may be defective or there is a bad connection between the guide shaft (compression unit) and D805 (X831).
Err 831 appears when connection D801 <=> is defective or J10 (D805) is defective.
Siemens AG SPB7-250.840.01.01.02 MAMMOMAT Novation Medical Solutions
10.05 CS SD 24
Page 99 of 100
Page 100

100 Changes to the previous version

7Changes to the previous version
8-
Chapter Page Change
n.a. n.a. Initial version
MAMMOMAT Novation SPB7-250.840.01.01.02 Siemens AG
10.05 CS SD 24
Page 100 of 100
Medical Solutions
Loading...