The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 10.05
2005
Page 2
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 2 of 100
Medical Solutions
Page 3
Table of Contents3
1-0Table of Contents
1 _______ General Information______________________________________________ 5
7 _______ Changes to the previous version _________________________________ 100
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 4 of 100
Medical Solutions
Page 5
General Information5
1General Information
2-
Protective measures0
System power supply0
Before you begin working on equipment, it is very important that you disconnect it from
the power supply at the main circuit breaker. Before removing or inserting any of the
printed circuit boards, switch off the equipment.
WARNING
If the system is only switched off at the control panel, the line voltage will still be present at the generator line connection (see wiring
diagram).
¹ A life-threatening hazard of electric shock exists.
WARNING
After the system shuts down, there may still be 380 V DC present
on the intermediate circuit.
¹ A life-threatening hazard of electric shock exists
Printed circuit boards0
CAUTION
The printed circuit boards contain highly electrostatically-sensitive components.
If disregarded, the components could be damaged.
¹ Use ESD equipment, ground prior to making contact, and
place the components on a conductive surface.
Measurements with the oscilloscope0
WARNING
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
Under no circumstances should the existing ground conductor in
the mains cable be disconnected when operating the oscilloscope.
There is a life-threatening hazard of electric shock.
¹ For measurements in which a resulting ground loop may
affect the measuring result, use the differential amplifier
(difference measurement).
Page 5 of 100
10.05CS SD 24
Page 6
6Error messages
2Error messages
3-
Error messages of the master Err 0XX0
A malfunction detected by the system is indicated by " Err" followed by a three-digit error
code on the kV and mAs displays of the control panel. This can be acknowledged by the
operator pressing the limit button. Exceeded limits are also deleted with the limit button.
Err 004
“Communication master - AEC disturbed or AEC not ready”
Description
If the master processor (D750) can no longer activate the AEC processor via the serial
interface or if any important data from the control deck is missing, the above error message is displayed.
Fault elimination
First check if PC board D750 is plugged in correctly and if the connections to the power
supply and the serial interface (see wiring diagram) are functioning correctly.
If no fault can be found, PC board D750 should be replaced.
Err 005
“Flash in AEC is defective or not correctly initiated”
Description
During initialization, the AEC processor (D750) checks the data stored in its E
2
means of stored checksums. If this data is incorrect or if the E
PROM fails, the AEC pro-
2
PROM by
cessor informs the master processor via the serial interface and the above error message
is displayed.
The master also generates this error if communication has not been established with the
AEC during exposure release.
NOTE
In contrast, error 004 appears if communication started with the
master disturbed.
Fault elimination
Reprogram E2PROM or replace E2PROM or PC board D750. Reprogram using the service PC "Backup" menu.
Err 008
“Communication master - stand disturbed or stand not ready”
Description
This error means that communication between the master processor and unit processor
via the serial interface is faulty or interrupted.
Fault elimination
Check whether the power supply and the serial interface (see wiring diagram) to the unit
are functioning correctly.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 6 of 100
Medical Solutions
Page 7
Error messages7
Err 011
“Exposure aborted by user”
Description
If the exposure buttons are released prematurely, exposure is aborted immediately. On
the control deck, the LIMIT LED (→0←) lights up, an acoustic signal sounds and the
above error message is displayed.
Fault elimination
This error indication is purely informational and does not require further action. Should it
appear frequently, however, check the contacts of the exposure release buttons. Replace
the exposure release button(s), if necessary.
Err 012
“Time limit reached”
Description
In order to prevent the tube assembly from overloading, a timer runs during exposure.
This is set to the time limit prior to exposure. If this timer runs down, the exposure is
aborted immediately, and the above error message is displayed. Depending on the mode
of operation, the time limit is calculated as follows:
AEC mode
The time limit is the power-dependent load time plus a tolerance value (approx. 400 ms).
mAs mode
The time limit is the calculated exposure time plus a certain reserve. If this period is
shorter than 2 s, 2 s are assumed. If it exceeds 2 s, a reserve time of 400 ms is added.
The time limit can be read off in ms with the service PC in “normal mode” under item “AEC
data”. The value shown here does not include the extra 400 ms.
Fault elimination
The tube current and the filament current must be measured again with the oscilloscope
(test points MA_ACT and FIL_CURR_ACT on D750). The nominal and actual value of the
tube current can be read off with the service PC in normal mode. If the actual tube current
is too low, you must first ensure that the dynamic learn filament current (DLF switch) is
switched on. The tube must be readjusted in any case. Should difficulties still occur, the
tube current actual value acquisition must be checked (see wiring diagram).
Err 013
“mAs limit reached in AEC exposure”
Description
For safety reasons, the accumulated mAs is integrated via the timer independently of the
AEC during AEC exposures. If a focus, tube assembly and kV-dependent time limit (see
table) is obtained without the AEC being switched off, the exposure is aborted and the
above error message is displayed.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 7 of 100
Page 8
8Error messages
Tube assembly25 kV30 kV35 kV
P40 MoW-100G F1196 mAs163 mAs140 mAs
P40 MoW-100G F2600 mAs500 mAs428 mAs
P40 MoW-100G F3238 mAs198 mAs170 mAs
P40 MoW-100G F4752 mAs627 mAs537 mAs
Fault elimination
The dose signal from the detector or the chamber to the hardware of the AEC (see wiring
diagram) must be checked.
If no error can be detected, PC board D750 or D801 must be replaced.
Err 014
“kV, corrected for dose calculation, is out of range”
Description
The kV value used for glandular dose calculations is the "set kV" adjusted for the actual
HVL value of the tube. This corrected kV value is outside the limits of the lookup tables
used for glandular dose calculations.
Fault elimination
Reload the tube-specific HVL values using the "dose calculation configuration program".
Err 020
“PC message had bad length”
Description
The length of the message sent from the service PC to the MAMMOMAT Novation
DR
is
too long.
Fault elimination
Check that the correct version of the service program is being used.
Err 022
“Receive message from external unit failed”
Description
An error occurred when the MAMMOMAT Novation
DR
was receiving data from the PC or
printer.
NOTE
May occur if the printer is switched on/off while the rest of the system is on.
Fault elimination
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 8 of 100
Medical Solutions
Page 9
Error messages9
Check the cables and connectors to the PC or printer. Board 750 defective. PC/printer
defective.
Err 024
“PC, ID camera or workstation has not sent ACK/NAK within two seconds. Three
tries are made”
Description
The service PC did not answer.
Fault elimination
Check the cables and connectors to the PC. PC defective.
Err 025
“PC, ID camera or workstation sends NAK three times as response to one and the
same message”
Description
The service PC did not receive information from the master. Timeout occurred and the
service PC responds with NAK.
Fault elimination
1. Reset error on panel and try again.
2. Reset MAMMOMAT Novation
DR
and PC.
Err 030
“Master PLD requests interruption of tube voltage”
Description
The master PLD logic has detected a type of error that is considered fatal and must terminate the exposure.
Fault elimination
1. Reset error on panel and try again.
Err 040
“The radiation protection door is open”
Description
The radiation protection door switch is not active while exposure is being released.
Fault elimination
1. Check the door switch.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 9 of 100
Page 10
10Error messages
Error messages of the panel Err 1XX0
Err 105
“Checksum error in installation data stored in panel EEPROM”
Description
Each time the MAMMOMAT Novation
2
PROM on the control deck is checked with a shadow area also stored in the E2PROM.
E
If any discrepancy is detected, the above error message is displayed.
Fault elimination
The installation data can be corrected via the service PC. If this error appears more frequently, exchange the deck PC board D740 and reinstall the parameters in the control
panel; see Service SW Backup/Restore. The opdose parameters have to be programmed
manually via the control deck. For the remaining parameters, use the service PC
"Backup" menu.
Err 106
2
“E
PROM program area error”
DR
is switched on, the installation area in the
Description
When data is stored in the E
2
PROM of the control panel during programming of the
opdose parameters, this is read back once more to check whether it has been stored correctly. If any differences occur, the above error message is displayed.
Fault elimination
2
The E
PROM in the panel must be replaced. If the error still occurs, the whole D740 must
be replaced. In both cases the parameters in the control panel have to be reinstalled. The
opdose parameters have to be programmed manually via the control deck. For the
remaining parameters, use the service PC, backup menu.
Err 107
“No communication panel - master”
Description
If no data transmission takes place between the deck and the master during the first 20 s
after power on, the control deck switches to the so-called “standalone” mode. The keyboard can be operated normally. If the exposure release button is operated in this mode,
the above error message is displayed.
Fault elimination
The cause of this error can be an interruption on the serial interface to the master. For this
reason, this connection must be checked carefully.
Err 121
“The version number of EEPROM and PROM are not the same”
Description
This error applies to the panel.
Fault elimination
Check EEPROM and PROM.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 10 of 100
Medical Solutions
Page 11
Error messages11
Error messages of the filament Err 3XX0
Err 303
“Filament current outside limits”
Description
During “Stand-by”, the master checks whether there is any filament current. If the filament
current is not within 10% from nominal of a predetermined value, the above error message is displayed.
Fault elimination
Check fuse F2 (see wiring diagram).
Check filament wires.
Filament of tube assembly defective.
PC board D750 defective.
Err 304
“Filament current over limit”
Description
During “Stand-by”, the master checks whether there is any filament current. If the filament
current does not reach a predetermined value or the filament current exceeds 8.5 A, the
above error message is displayed.
Fault elimination
Check fuse F2 (see wiring diagram).
Check filament wires.
Filament of tube assembly defective.
PC board D750 defective.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 11 of 100
Page 12
12Error messages
Error messages of the AEC Err 4XX0
Err 401
“Error during configuration of PLD.”
Description
Configuration of PLD device in AEC failed.
Fault elimination
Check whether Proms are inserted properly
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
Err 402
“Self-test of PLD failed.”
Description
DR
. Wait approx. one minute. Start up the MAMMO-
Gain test of circuit board D750 failed.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 403
“Self-test timeout. Fatal error in PLD’s internal program.”
Description
Timeout occurred in the PLD device in the AEC during gain test of circuit board D750.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Turn on the MAMMO-
Err 404
“AEC CPU requests interruption of tube voltage”
Description
The processor in the AEC part of the master board requests termination of high voltage
generation.
Fault elimination
Check the tube power module and the HV inverter module.
Err 405
“AEC PLD cannot perform requested operation because it is busy self- testing”
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 12 of 100
Medical Solutions
Page 13
Error messages13
The PLD in the AEC part of the master board requests termination of high voltage generation.
Fault elimination
Check the tube power module and the HV inverter module.
Err 406
“PLD cannot perform requested operation because it is busy self-testing.”
Description
The PLD in the AEC cannot perform requested operation because it is performing a
self-test.
Fault elimination
Acknowledge the displayed error and wait approx. one minute.
Err 407
“Time-out error while erasing AEC flash memory”
Description
The D750 contains a memory device called “flash memory”. This memory is used for storing AEC parameters and correction tables. Before writing data into the memory, it is necessary to erase the sector in which data is to be stored. The erasing procedure is
controlled by a timer. If anything goes wrong while erasing the AEC flash memory and a
timeout occurs, an error results.
Fault elimination
Switch off the MAMMOMAT Novation
MAT Novation
DR
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 408
“Time-out error while writing to AEC flash memory (time-out in waiting for ok indication)”
Description
The D750 contains a memory device called “flash memory”. This memory is used for storing AEC parameters and correction tables. The procedure of writing this data is controlled
by a timer. If anything goes wrong while writing to the AEC flash memory and a timeout
occurs, an error results.
Fault elimination
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
DR
. Wait approx. one minute. Start up the MAMMO-
Err 409
“AEC flash test has locked up (time-out of 15 seconds exceeded)”
Description
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 13 of 100
Page 14
14Error messages
During start-up of the unit, the AEC performs a number of tests, e.g. testing of the flash
memory device on D750. This testing is controlled by a timer and if something is wrong, a
timeout occurs.
Fault elimination
DR
Switch off the MAMMOMAT Novation
DR
MAT Novation
. If the error recurs, replace the D750 circuit board.
Err 410
“Attempt to start an exposure before AEC flash memory has delivered correction
values”
Description
Before an AEC exposure is released and based on the exposure settings on the panel,
the AEC fetches an estimated dose from a correction table stored in the flash memory. If
the exposure buttons are pressed during a reading operation from the flash memory, an
error occurs.
Fault elimination
. Wait approx. one minute. Start up the MAMMO-
Acknowledge the displayed error. Wait 15 seconds. Try to perform an exposure.
Err 412
“Attempt made to start an exposure while self-testing of the PLD in AEC was in
progress”
Description
Attempt made to start an exposure while a self-test of the PLD in the AEC was in
progress.
Fault elimination
Acknowledge the displayed error. Wait 15 seconds. Try to perform an exposure.
Err 413
“Exposure too short, correction value could not be loaded in time”
Description
An estimated dose is calculated before an exposure is released. The minimum estimated
dose is defined for a minimum object thickness. For a normal object thickness (> 5 mm),
the AEC will calculate a necessary correction value to add to the estimated dose during
exposure. This procedure takes a while, and in the case of particularly thin objects, the
AEC may not be able to determine within the time limit whether a correction is necessary.
An error results due to the small object thickness.
Fault elimination
Acknowledge the displayed error. Change the exposure parameters (e.g. lower kV).
Err 414
“Attempt to release an exposure after the PLD in AEC has failed self-test/ configuration”
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 14 of 100
Medical Solutions
Page 15
Error messages15
Description
Caused by previously generated errors 401, 402, 403, 409 and not restarting
MAMMOMAT Novation
Fault elimination
Switch off the MAMMOMAT Novation
MAMMOMAT Novation
Err 416
“Estimated dose out of range”
Description
Estimated dose out of range, which can be caused by
- an extreme sensitivity setting
- an extreme sensitivity correction setting
- an extreme density correction setting.
DR
.
DR
. Wait approx. one minute. Start up the
DR
. If the error recurs, replace the D750 circuit board.
This means that the exposure will be terminated by the max. mAs limit.
Fault elimination
Acknowledge the displayed error. Make sure that the correct AEC correction tables are
installed. Check settings for sensitivity, sensitivity correction and density correction.
Err 418
“Attempt to make offset compensation test while in AEC mode”
Description
During an offset compensation test, the AEC board is set to a special test mode and cannot perform an AEC exposure. Because of this, test exposures have to be performed in
the mAs mode during this test.
Fault elimination
Acknowledge the displayed error. Change to mAs and follow the instructions for an offset
compensation test.
Err 419
“Attempt to write to registers of PLD not available during exposure”
Description
This is an internal error that does not interfere with an exposure and is not shown on the
control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it
immediately.
Err 420
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 15 of 100
Page 16
16Error messages
“Dose or time monitor overflow in PLD”
Description
This is an internal error that does not interfere with an exposure and is not shown on the
control panel. The time and dose values reported by the service PC are not valid.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it
immediately.
Err 421
“Wrong test mode request received by PLD”
Description
This is an internal error that does not interfere with an exposure and is not shown on the
control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it
immediately.
Err 422
“Attempt to read registers of PLD not available during exposure”
Description
This is an internal error that does not interfere with an exposure and is not shown on the
control panel.
Fault elimination
If this error is discovered when reading the error buffer via the service PC, please report it
immediately.
Err 423
“Counters for estimated doses in PLD not initialized in time”
Description
If mAs mode: Internal error that does not interfere with an exposure and is not shown on
the control panel. If AEC mode: Internal error that interferes with an exposure and is
shown on the control panel.
Fault elimination
If mAs mode: If this error is discovered when reading the error buffer via the service PC,
please report it immediately.
If AEC mode: Acknowledge the displayed error. Wait 15 seconds. Try to perform an
exposure.
Err 450
“Exposure aborted by dose monitoring”
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 16 of 100
Medical Solutions
Page 17
Error messages17
Description
An AEC exposure could not be performed due to insufficient dose rate.
Fault elimination
Acknowledge the displayed error. Change the exposure parameter settings (e.g. higher
kV).
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 17 of 100
Page 18
18Error messages
Error messages of the FFDM DR System Er 5XX0
Err 501
“DR detector request interruption of tube voltage”
Description
The HW signal DR_DET_RDY_FOR_EXP has become inactive during exposure.
Fault elimination
Check the plugs and connections at board D750.
Err 502
“DR system request interruption of tube voltage”
Description
If the exposure buttons are released prematurely, exposure is aborted immediately.
The HW signal DR_SYST_RDY has become inactive during exposure.
Fault elimination
Check the plugs and connections at board D750.
Err 503
“Master does not sense DR signal level change within specified times”
Description
At exposure start, the HW signal DR_DET_RDY_FOR_EXP is not activated within the
specified time.
Fault elimination
Check the plugs and connections at board D750.
Err 504
“Master senses an error within the DR system”
Description
The HW signal DR_DET_ERR has been activated.
Fault elimination
Check the plugs and connections at board D750.
Check the switch S1 on grid drive.
Err 506
“Master senses a tube arc during the AEC pre exposure when in DR mode”
Description
A tube arc has occurred during the AEC pre-exposure. No tube arc is allowed during the
AEC pre-exposure.
Fault elimination
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 18 of 100
Medical Solutions
Page 19
Error messages19
N/A
Err 578
“mAs exceeds maximum”
Description
The calculated AEC main exposure mAs value returned from the DR system exceeds the
maximum mAs value that the MAMMOMAT Novation
Fault elimination
Check AEC functioning
Err 579
“Master is not updated with the mAs value for the main exposure in AEC mode”
Description
The mAs value for the main exposure has not been calculated/transmitted by the DR system or received by the MAMMOMAT Novation
DR
Fault elimination
DR
can handle.
within the specified time.
Check the plugs and connections at board D750.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 19 of 100
Page 20
20Error messages
Error messages of the power pack Err 6XX0
Err 601
“Rotation speed not reached within 3s”
Description
During starting of the rotating anode, the master counts the control pulses and compares
them with a ‘tube assembly and intermediate circuit voltage’-dependent table value. If this
is not reached within 3 s, err 601 is displayed.
Fault elimination
Check oscillation current (see wiring diagram). Rotating anode cable or stator defective.
Err 602
“kV
Displayed on PC:kV
- minimum tube voltage 17 kV not reached”
min
-minimum tube voltage 17kV not reached
min
Description
If 17 kV is not reached within 250 ms after the KVE signal, the above error message is
displayed.
Fault elimination
Intermediate circuit voltage too low (see wiring diagram).
Tube current or power too high. See also (Tube high voltage, current and mAs
values / p. 63)
Check oscillation current (see wiring diagram).
SS relay not pulled up (see wiring diagram).
“kV
” too low.
nom
Board D750 defective.
Err 603
“kV
- tube voltage greater than 50 kV”
max
Description
During exposure, a threshold switch in the kV controller monitors whether the high voltage
exceeds 50 kV. If this is the case, the exposure is aborted immediately via the KVA lead.
Fault elimination
Check the kV nominal value. Check tube current and power, see also (Tube high voltage,
current and mAs values / p. 63). Actual value acquisition defective (see wiring diagram).
Replace PC board D750.
Err 604
“WR Austast - inverter short circuit blanking signal remains”
Description
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 20 of 100
Medical Solutions
Page 21
Error messages21
When I
(Overcurrent) in the inverter is exceeded, the control is interrupted for 200 ms.
max
The software then checks whether the blanking signal is still applied. If this is the case,
the above error message is displayed.
Fault elimination
Transistor module in HV inverter defective. Check tube current and power, see also (Tube
high voltage, current and mAs values / p. 63). Check fuse (F2) . Stator short circuit. Board
D750 defective.
Err 606
“Inverter short circuit during radiation”
Description
HV-Inverter overcurrent during exposure sequence.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).Check tube current and
power, see also (Tube high voltage, current and mAs values / p. 63)
Stator short circuit.
Board D750 defective.
Err 607
“Inverter short circuit during brake cycle”
Description
Actual tube voltage deviates by more than 5% from nominal value at the start of the exposure.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).Check tube current and
power, see also (Tube high voltage, current and mAs values / p. 63). Stator short circuit.
Err 608
“Tube current is outside limits”
Description
Actual tube current deviates by more than 10% from nominal at the start of exposure.
Fault elimination
Transistor module in HV inverter defective. Check fuse (F2).
Stator short circuit.
Err 611
“KVA signal disabled during radiation”
Description
During exposure, KVA becomes “H” and the inverter is thus enabled. If this enabling fails,
the above error message is displayed.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 21 of 100
Page 22
22Error messages
Fault elimination
Check KVA lead (see wiring diagram). Board D750 defective.
Err 620
“U
+15V to control inverter not present”
anst
Displayed on PC:U
+15V to control inverter not present
anst
Description
The DC bus voltage used for powering the modules of the generator assembly is too low
or too high.
Fault elimination
Check the power supply and fuses.
Err 630
“PH1 pressure switch on the HV tank activated”
Description
the HV actual value does not correspond to the HV set value. The above error message is
displayed.
Fault elimination
Overload of the tube assembly or of HV inverter.
Line interruption (see wiring diagram).
Board D750 defective (monitoring).
Err 631
“Open connector in the generator”
Description
At least one connector between modules of the generator assembly or between the master board and the modules of the generator assembly is not connected.
Fault elimination
Line interruption (see wiring diagram).
Board D750 defective (monitoring).
Err 632
“Bias voltage error”
Description
Active bias voltage for large focus or inactive bias voltage for small focus.
Fault elimination
Check power supplies and fuses on D750.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 22 of 100
Medical Solutions
Page 23
Error messages23
Err 633
“The high voltage is detected during stand by”
Description
If any or both of the signals HV_ON and EN_HV_ON are active during standby (no exposure being performed), this error is issued. Displayed when the tube is arcing.
Fault elimination
Press the "Lim" button to clear the error message.
Check the HV inverter module.
Err 634
“Voltage outside limits for 15V supplies”
Description
This error is issued if the +15V supply voltage goes below 10.6V or if the -15V supply voltage goes above -10.75V.
Fault elimination
Check the power supplies at the mains converter module and the fuses.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 23 of 100
Page 24
24Error messages
Error messages of the OPDIMA Err 7XX0
Err 777
“Exposure sequence is aborted by OPDIMA”
Description
The time between pressing acquire from the workstation and releasing the exposure from
the MAMMOMAT generator has exceeded 60 seconds.
The selected exposure parameters (kV, mAs) will yield an excessively hard beam quality
for the object.
The cancel button was pressed while "Expose at MAMMOMAT" or "Exposure preparation
started" was displayed.
Fault elimination
Press the limit button on the MAMMOMAT generator control panel to continue.
Err 778
“No communication present between MAMMOMAT and OPDIMA”
Description
Communication problem.
Fault elimination
Press "Retry" at the workstation or press the limit button on the MAMMOMAT generator
control panel to continue. Check the plugs and cables.
Err 779
“The mAs value for main exposure in OPDIMA AEC mode has not been received in
time”
Description
Communication problem.
Fault elimination
Press the limit button on the MAMMOMAT generator control panel to continue.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 24 of 100
Medical Solutions
Page 25
Error messages25
Error messages of the stand Err 8XX0
Err 801
“Time out of AR signal”
Description
The grid has not reached its start position within 2 seconds.
Fault elimination
Board D750 defective.
Check grid functioning.
Err 802
“OKT 2 pressure switch on tube housing or beam form anode or door switch”
Description
The oil pressure in the tube has exceeded the limit.
Fault elimination
Allow the tube to cool down. Close door/switches.
Err 803
“Stand requests interruption of tube voltage”
Description
The processor on stand CPU board D801 requests termination of high voltage generation.
Fault elimination
Switch off the equipment and then switch it on again.
Err 811
“Stand not ready for exposure”
Description
Normally, the panel does not allow start of exposure if any of the following errors is
present: No object table installed, no film cassette inserted, film cassette not changed
after exposure, improper diaphragm mounted. Nor will it allow exposure, if the tube
assembly is at a distance from the floor that could result in collision. These operator errors
are indicated on the control panel (the LED at the corresponding symbol lights up). Should
the control panel nevertheless allow start of exposure, this might be due to a bit error during the transmission of data between the stand and the control panel. As a precautionary
measure, the stand also checks whether the conditions are fulfilled. If this is not the case,
error message 811 is generated.
Fault elimination
Switch off the equipment and then switch it on again.
Err 812
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 25 of 100
Page 26
26Error messages
“Compression protect switch not OK in stand”
Description
This signal is used to check for proper operation of the compression protect relay (Kl) on
the motor board (D802). The CPU (D801) reads the signal after a short delay after releasing the compression pedal in order to check that the relay has opened and disabled motor
operation in the compression direction. Note that the test point on D802 is protected with
4.75k.
Fault elimination, troubleshooting
Check the fuses on board D802. Normal active level 9V (test point COMP_PROT on
D802). If the level is OV check 14V and Kl on the motor board or look for a short circuit on
the ribbon cable between D801 and D802. If the cable is OK, D802 should be replaced.
If the voltage is 18V at the test point, the ribbon cable is open or D801 is faulty.
If the voltage is OK, replace D801.
Err 813
“Compression Motor controller, or motor, error in stand”
Description
This signal indicates an overcurrent or lack of voltage (5V or 15V) in the compression
motor drive. In the case of an overcurrent, the signal appears after about 2 s and disappears after about 0.5 s. The signal causes D801 to stop sending pulses to the compression drive and thus disables the compression motor. The red LED “comp error”, V2 lights
when error (comp_ok signal low) except if 5V is missing. There is no test point.
Fault elimination, troubleshooting
If V2 is lit steadily, check fuses F3 (24V) and F2 (24VF & l5V). l5V is indicated by LED
V49 and can be measured at test point l5V on D802. Test point 24VF should measure
about 30V in an unloaded state. 5V is indicated by LED V7 and test point 5V on D802. If
5V and 14V (normally about 18V unloaded) are missing, check Fl on D801. Missing voltages can also be caused by an open connection in emergency stop switch S880.
Overcurrent is most likely caused by a faulty motor, mechanical overloading of motorized
movement, short-circuited wiring of D805 or a faulty D802.
Err 815
“Compression speed too high, error suspected in stand”
Description
The CPU (D801) has detected a compression speed greater than 13 cm/s. The software
causes the compression motor to reverse direction.
Fault elimination, troubleshooting
This error can be caused by play in the compression unit, defective compression measurement hardware, or a short-circuited power stage on the motor board. In the first two
cases, replace the compression unit and recalibrate the compression values with the service PC program. In the third case, the compression motor will run at a max. speed
upwards except when a compression pedal is pressed. It is not certain that error 815 will
appear. Replace D802.
Err 816
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 26 of 100
Medical Solutions
Page 27
Error messages27
“Collimator fails to reach correct position in time”
Description
Time out for filter lamella positioning during startup.
Fault elimination, troubleshooting
Check whether the collimator plates move at all.
If the collimator plates do not move:
- Check whether the CPU is sending out pulses, COLL_ST test point on D801, if not,
replace D801.
- Check whether the wing board (D805) is receiving pulses, test point COLL_ST on
D805; if not, check the cables and connectors between D801 and D805.
- Check voltages on the wing board.
- Check STEP_ENABLE test point on D805. It should be low (0V) when pulses are output to the motor.
If the collimator plates do move:
- Check whether the collimator position light switch is functioning as in the description
above.
- Check whether it moves for the entire 8 s. If it does but apparently does not reach the
other position in time, something may be in the way and slowing the speed, or the
motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after just a short time, either the motor or the wing board (D805) is
faulty. Try replacing one of them.
- Check collimator calibration (lamp and/or X-ray field).
If not successful, replace collimator and/or D814.
Err 817
“Filter self-test failure”
Description
The zero position for the filter wheel is not found during start-up.
Fault elimination, troubleshooting
Switch between the three anode/filter combinations on the control panel and check if the
filter disk moves.
- Check whether the filter position light switch works.
- If it only moves slowly, something may be in the way and slowing the speed, or the
motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after just a short time, either the motor or the wing board (D805) is
faulty. Try replacing one of them.
If not successful, replace collimator and/or D814.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 27 of 100
Page 28
28Error messages
Err 818
“Filter positioning failure”
Description
The zero position for the filter wheel is not found after exposure.
Fault elimination
Switch between the three anode/filter combinations on the control panel and check if the
filter disk moves.
- Check whether the filter position light switch works.
- If it only moves slowly, something may be in the way and slowing the speed, or the
motor may be faulty. In the latter case, try replacing the motor.
- If the motor stops after a short time, either the motor or the wing board (D805) is
faulty. Try replacing one of them.
If not successful, replace collimator and/or D814.
Err 819
“Drive protect switch not OK (lift and rotation), in stand”
Description
This signal is used to check for proper operation of the drive protect relay (K6) on the
motor board (D802). The CPU (D801) reads the signal after a short delay after releasing
the activated lift or rotation button in order to check that the relay has opened and disabled motor operation.
Fault elimination, troubleshooting
Normal active level 5V (test point DRIVE_PROT on D802). If the level is 0V, check 24VF
and K3 on the motor board or look for a short circuit on the ribbon cable between D801
and D802. If the cable is OK, D802 should be replaced.
If the voltage is 30V at the test point, the ribbon cable is open or D801 is faulty.
If the voltage is OK, replace D801.
NOTE
There is no error for the lift & rotation drive corresponding to the
compression OK error. If this fault occurs, D801 stops sending
pulses to D802 and the selected motor stops. The LED V12,
DRIVE_ERROR functions in the same manner as COMP_ERROR.
Note that the current limit is lower when the lift motor is moving
downward. If there are problems with lift or rotation operation indicating an overload, look at the DRIVE_ERROR LED to determine
if the motor is overloaded.
Overloading is possible due to improper balancing (adjustment of the spring or improper
loading of the lifting carriage) or improper adjustment of the rotation brake. Faulty motor,
short-circuited wiring or faulty D802 are other possible causes.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 28 of 100
Medical Solutions
Page 29
Error messages29
Err 820
“Write/read back failure with EEPROM in stand”
Description
If data is stored in the E
to verify that it has been stored correctly. In the event of differences, the above error message is displayed.
Fault elimination
Repeat the attempt to write in the E
of repeated attempts, board D801 must be replaced. All stand parameters must be reinstalled. Use the service PC to reinstall the stand parameters from floppy or disk.
Err 821
“Stand EEPROM has wrong version nbr (=old or corrupt)”
Description
The version number of the PROM stand is also stored in the E
starting up the equipment, the version number stored in the E
version number of the PROM. If they do not correspond, the above error message code is
displayed on the control panel. This occurs after changing the software version to indicate
that a new version has been installed.
2
PROM (D801.I34) stand during adjustment, a check is performed
2
PROM. If this not does not yield any results, in spite
2
PROM (D801.I34). When
2
PROM is compared to the
Fault elimination
The version number of the new software is automatically written into the E
2
PROM. Switch
off the equipment and then switch it on again. The error should not recur.
Err 822
“Grid failed to move properly”
Description
The grid has to move properly from the start. This is supervised by the CPU. The CPU
measures the time from start till the grid has reached one of the end positions and then is
able to leave it. To check that the grid is functioning properly and does not jam, the time
the grid requires to move between the two end positions is also measured the first time. If
the time for start or run check exceeds 2 sec., this error appears.
Fault elimination, troubleshooting
Check programming of the grid fast speed (normally 80%) and the grid fast speed time
(normally 500 ms).
Check the bucky board. First try another board. Make sure the grid moves freely along its
entire travel path in both directions.
Check the output voltage at test point GRID_M on D802. At 80% speed and nominally
30V on 24VF, GRID_M should be 30 - 0.8*30 = 6V during fast grid movement. Check for
possible short circuits if the signal remains at about 30V. This voltage goes to the bucky
via board D805 and its relay Kl. Test point GRID_SP on D802 is the control signal from
D801. It is a 5V, 20 kHz PWM signal. At 80% speed, the signal should be low at about 40
µs and high at about 10 µs per period. Replace D802 if the correct output is not achieved
despite proper input, cabling and bucky.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 29 of 100
Page 30
30Error messages
Err 824
“Stand motor voltage drops unnaturally”
Description
The CPU (D801) measures voltage 24VF from the motor board (D802). If this value is
below about 16V, this error is reported.
Fault elimination, troubleshooting
Check that the stand has voltage (generator - X14 connected, F4 OK). The stand display
should be on.
Check that the emergency stop is not activated and that its wiring is OK.
Check that LED V49, 15V, is lit. If not, check fuses F2 and F3 on D802 in the stand.
Check voltage 24VF (measure with reference to test point 0V on D802).
Check the ribbon cable between D802 and D801.
Replace D801.
Err 825
“One or more potentiometers seem to have lost contact with stand”
Description
The four potentiometers, R803-tube angle, R871-preset angle, R861-preset force and
R863-thickness, have their return current through signal pot_return. This is done to
assure that no potentiometer is missing or partly disconnected so that it can yield a value
leading to improper stand operation. During stand configuration or after replacing a potentiometer or the compression unit, the correct value for the signal potentiometer check
must be read and stored in the stand’s E
2
PROM with the service PC program. This error
is reported if the measured value is more than 32 bits less than the programmed value.
Fault elimination, troubleshooting
Check that all potentiometers are functioning properly. An error can be caused by an open
connection, connector or potentiometer. If all four potentiometers are functioning properly,
check the programming of the potentiometer check and correct it. The need for such an
adjustment may indicate a potentiometer is about to fail.
Err 826
“There is a short circuit somewhere among potentiometers in stand”
Description
The same circuitry as Err 825 is used. In this case the error indicates that the value is
more than 32 bits greater than the programmed value.
Fault elimination, troubleshooting
Check the actual and programmed pot. check values with the service PC program. If the
value is only slightly too high, check all potentiometers for mechanical damage or sensitivity. A short circuit (pot. or wiring) should yield the max. value for pot_check (3FF).
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 30 of 100
Medical Solutions
Page 31
Error messages31
Err 829
“Collimator motor communication error”
Description
Communication between the stand and collimator control board D814 is erroneous or disrupted.
Fault elimination, troubleshooting
Check cables and plugs. Replace D814.
Err 830
“Filter motor communication error”
Description
Communication between the stand and collimator control board D814 is erroneous or disrupted.
Fault elimination, troubleshooting
Check cables and plugs. Replace D814.
Err 831
“Paddle decoder communication error”
Description
Identification of compression paddle fails.
Fault elimination, troubleshooting
Clean the sensor at the paddle. Check whether the paddle code is working with another
paddle.
Err 850
“Write/read back failure for table 1 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
R is switched on, the adjustment data in the
stand EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config”
has its own checksum. If a discrepancy is detected in the stand installation menu, the
above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 851
“Write/read back failure for table 2 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the calibrate compression menu, the
above error message is displayed.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 31 of 100
Page 32
32Error messages
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 852
“Write/read back failure for table 3 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the rotation calibration menu, the above
error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 853
“Write/read back failure for table 4 in stand EEPROM”
DR
is switched on, the adjustment data in the stand
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the rotation calibration menu, the above
error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 854
“Write/read back failure for table 5 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the lift parameters menu, the above
error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 855
“Write/read back failure for table 6 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the best compression menu, the above
error message is displayed.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 32 of 100
Medical Solutions
Page 33
Error messages33
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 856
“Write/read back failure for table 7 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the rotation parameters menu, the
above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 857
“Write/read back failure for table 8 in stand EEPROM”
DR
is switched on, the adjustment data in the stand
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the beam limiting device menu, the
above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 858
“Write/read back failure for table 9 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the rotation parameters menu, the
above error message is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 859
“Write/read back failure for table 10 in stand EEPROM”
Description
Each time the MAMMOMAT Novation
DR
is switched on, the adjustment data in the stand
EEPROM (D801.I34) is checked with a checksum. Each menu under “Stand config” has
its own checksum. If a discrepancy is detected in the rotation calibration menu, the above
error message is displayed.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 33 of 100
Page 34
34Error messages
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 860
“Write/read back failure for table 11 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for small focus Mo in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 861
“Write/read back failure for table 12 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for large focus Mo in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 862
“Write/read back failure for table 13 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for small focus W in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 863
“Write/read back failure for table 14 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for large focus W in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 864
“Write/read back failure for table 15 in stand EEPROM”
Description
Incorrect checksum for collimator calibration parameters for the light field focus in
EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 34 of 100
Medical Solutions
Page 35
Error messages35
Err 865
“Write/read back failure for table 16 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for wing differences in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 866
“Write/read back failure for table 17 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the filter in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 867
“Write/read back failure for table 18 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the optimum compression function
in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Err 868
“Write/read back failure for table 19 in stand EEPROM”
Description
Incorrect checksum for the calibration parameters for the optimum compression function
in EEPROM.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 35 of 100
Page 36
36Log and error files
3Log and error files
4-
General0
Some files on Mammomat Novation system contain information about processes on the
system. This information could be helpful for determination of failure, detector artifacts,
communications problems, blue screen and automatic reboot of the WH AWS.The analysis of the files requires specific knowledge. For this reason, if support is required, the files
have to be sent to USC/HSC.
If support by USC/HSC is requested, groups of files have to be sent for explicit analysis:
•MPxxxx.smj, MPxxxx.map, qualify.log,MPxxxx.cfg and drapi.log: They are locat-
ed under WH AWS in folder C:\AWS\drul\data\.
•Brick files: These files can be shown by using the PCMX Web software.
•Savelog files: a feature on the syngo SW is implemented for extracting the files and
burning them onto CD (only as of SW VA11x).
•Windows Event log.
•syngo logs. They are located under WH AWS in folder C:\AWS\log\*.*
•Dump files (if available). They are located under WH AWS in folder C:\winnt\
minidump\*.dmp.
•brickspy.tar.gz file.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 36 of 100
Medical Solutions
Page 37
Log and error files37
SW Version VA10x0
These instructions describe extracting log files for WH AWS with SW VA10x.
Copy MPxxxx files (detector files), dump files and syngo logs0
These files have to be copied to a defined folder to ensure a correct copy.
Work steps
•The syngo AWS application is running.
•Select Service and Local Service in the menu bar under Options
Fig. 1:
•Enter the service key and press OK. The Home menu appears.
•Select Utilities
•Select Escape to OS
•Enter start explorer under parameters.
Command: NT command Interpreter
Parameters: start explorer
•Press the enter key.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 37 of 100
Page 38
38Log and error files
•Windows Explorer opens.
Fig. 2:Detector files
•Under C:\AWS\drul\data, select the files Drapi.log, MPxxxx.cfg, qualify.log,
MPxxxx.map and MPxxxx.smj,(1/Fig.2/p.38)click the right mouse button and copy and paste to folder C:\AWS\Service\extract.
•Select C:\winnt\minidump\*.dmp (if available), right-click, and copy and paste to fold-
er C:\AWS\Service\extract.
•Under C:\AWS\log , select all files (*.*), right-click them and copy and paste them to
folder C:\AWS\Service\extract.
•Close Explorer.
•Close Explorer.
•Select Home
•Close the Service software window
Copy Array and Brick log files0
Use the Web Brick window to display these files.
Work steps
Array files
•The syngo AWS application is running.
•Select Service and Local Service in the menu bar under Options
•Enter the service key and press OK. The Home menu appears.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 38 of 100
Medical Solutions
Page 39
Log and error files39
•Enter http:brick under address.
Fig. 3:PXCM Web Control Interface
•Select array(1/Fig.3/p.39)
•Select Log File(s)
Fig. 4:
•The user name and appropriate password have to be entered (Fig.4/p.39)
Login: root
Password: *****
•Press OK
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 39 of 100
Page 40
40Log and error files
•Select All Log Files
Fig. 5:
•Right-click syslog and select Save Target As (Fig.5/p.40)
•Save the file as array_syslog in the directory C:\AWS\Service\extract.
•Right-click syslog.0 and select Save Target As .
•Save the file as array_syslog.0 in the directory C:\AWS\Service\extract.
•Brick files
•Select Brick(2/Fig.3/p.39)
•Select Log File(s)
•Select All Log Files
•Right-click syslog and select Save Target As
•Save the file as brick_syslog in the directory C:\AWS\Service\extract.
•Right-click syslog.0 and select Save Target As
•Save the file as brick_syslog.0 in the directory C:\AWS\Service\extract.
•Close the Web Brick window.
•Close the Service software,
Windows Eventlog0
Work steps
•The syngo AWS application is running.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 40 of 100
Medical Solutions
Page 41
Log and error files41
•Select Service and Local Service in the menu bar under Options
Fig. 6:
•Enter the service key and press OK. The Home menu appears.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 41 of 100
Page 42
42Log and error files
•Select EventLog on the right upper edge of the menu bar.
Fig. 7:EventLog_System
•Select the following parameters as shown in (Fig.7/p.42) .
Domain: System(1/Fig.7/p.42)
Severity: All(2/Fig.7/p.42)
Extended: selected(3/Fig.7/p.42)
NT-Message: selected(3/Fig.7/p.42)
Message limit: empty(4/Fig.7/p.42)
Extract to file: System.txt ((5/Fig.7/p.42)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date)).(6/Fig.7/p.42)
Select GO.(7/Fig.7/p.42)
•The log file appears.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 42 of 100
Medical Solutions
Page 43
Log and error files43
Fig. 8:EventLog_Application
•Select the following parameters as shown in (Fig.8/p.43) .
Domain: Application(1/Fig.8/p.43)
Severity: All(2/Fig.8/p.43)
Facility: Developer(3/Fig.8/p.43)
Extended: selected(4/Fig.8/p.43)
NT Message: selected(4/Fig.8/p.43)
Message limit: empty(5/Fig.8/p.43)
Extract to file:Application.txt(6/Fig.8/p.43)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date))(7/Fig.8/p.43)
Select GO.(8/Fig.8/p.43)
•The log file appears.
Brickspy.tar.gzfile
Use telnet to start a script.
•Select Service and Local Service in the menu bar under Options
•Select Utilities
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 43 of 100
Page 44
44Log and error files
•Select Escape to OS
•Enter start telnet brick under parameters
Login: root (and press enter)
Password: brick (and press enter)
•From the brick# command prompt, execute the following command:
brickspy all (use lowercase letters and press enter)
•Wait until the script is finished and enter Exit (and press enter)
Use ftp to get the saved file.
•Select Service and Local Service in the menu bar under Options
•Select Utilities
•Select Escape to OS
•Enter start cmd under parameters
•At the DOS prompt, enter cd C:\AWS\Service\extract (and press enter)
•At c:\temp, enter ftp brick (and press enter)
•User: ftp (and press enter)
•Password: a random e-mail address, e.g.. test@siemens.com
•At the ftp prompt, enter bin (and press enter)
•Enter prompt (and press enter)
•Enter cd tmp (and press enter)
•Enter dir (and press enter)
The file brickspy.tar.gz will be displayed
•Enter mget brick* (and press enter)
•After the file has been transferred, enter quit to exit ftp.
•The file brickspy.tar.gz can be found on WH AWS under C:\Temp.
The log files are now stored in the folder C:\AWS\Service\extract and can be copied to
an external storage medium, e.g., USB stick. If no external drive is available, the log files
should be saved to CD. For this procedure, do the following:
Creating a new backup package
•Select Local Service/Configuration/(Backup/Restore)
•Enter a new package name(e.g.:ServiceLogFiles) in the blank line below
Package(1/Fig.9/p.45)
•Click add(2/Fig.9/p.45)
•In the blank line “file &dir”, enter the name: C:\AWS\Service\extract(3/Fig.9/p.45)
•Click add after each entry(4/Fig.9/p.45).
The new file or path selection is now visible in the files & directories list box.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 44 of 100
Medical Solutions
Page 45
Log and error files45
Fig. 9:Backup Package
•Click save(5/Fig.9/p.45).
Back up your created package to CD.
•Select Local Service/Backup & Restore.
•Under Command: select Backup.
•Select the CD-R drive.
•Select your own created backup package: ServiceLogFiles.
•Click Go.
The backup process to CD will be started!
•After you have completed the backup, check your CD with the stored log files.
The configured ServiceLogFiles package is stored on CD compressed as file type “*.ar”.
You can use the syngo arviewer.exe to read or extract these files. For more information,
contact CS HSC 24 or your local USC.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 45 of 100
Page 46
46Log and error files
SW Version VA11x0
These instructions describe extracting log files for WH AWS with SW VA11x.
Copy MPxxxx files (detector files), dump files and syngo logs0
These files have to be copied to a defined folder to ensure a correct copy on the CD.
Work steps
•The syngo AWS application is running.
•Select Service and Local Service in the menu bar under Options
Fig. 10:
•Enter the service key and press OK. The Home menu appears.
•Select Utilities
•Select Escape to OS
•Enter start explorer under parameters
Command: NT command Interpreter
Parameters: start explorer
•Press the enter key.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 46 of 100
Medical Solutions
Page 47
Log and error files47
•Windows Explorer opens.
Fig. 11: Detector files
•Under C:\AWS\drul\data, select the files Drapi.log, MPxxxx.cfg, qualify.log,
MPxxxx.map and MPxxxx.smj,(1/Fig. 11 / p. 47)right-click, and copy and paste to folder C:\AWS\service\html\whaws\extract.
•Select C:\winnt\minidump\*.dmp (if available), right-click, and copy and paste to fold-
er C:\AWS\Service\html\whaws\extract.
•Under C:\AWS\log , select all files (*.*), right-click them and copy and paste them to
folder C:\AWS\Service\html\whaws\extract.
•Close Explorer.
•Select Home
•Close the Service software window
The storage files will be written onto CD by burning the SaveLogs files.
Copy Array and Brick log files0
Use the Brick Configuration feature to display these files.
Work steps
Array files
•The syngo AWS application is running.
•Select Service and Local Service in the menu bar under Options
•Enter the service key and press OK. The Home menu appears.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 47 of 100
Page 48
48Log and error files
•Select the Acquisition System button.
Fig. 12: PXCM Web Control Interface
•Select Brick Configuration(1/Fig. 18 / p. 54)
•Select array(1/Fig. 12 / p. 48)
•Select Log File(s)
Fig. 13:
•The user name and appropriate password have to be entered(Fig. 13 / p. 48)
Login: root
Password: *****
•Press OK
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 48 of 100
Medical Solutions
Page 49
Log and error files49
•Select All Log Files
Fig. 14:
•Right-click syslog and select Save Target As (Fig. 14 / p. 49)
•Save the file as array_syslog in the directory C:\AWS\Service\html\whaws\extract.
•Right-click syslog.0 and select Save Target As .
•Save the file as array_syslog.0 in the directory C:\AWS\Service\html\whaws\extract.
•Brick files
•Select Brick
•Select Log File(s)
•The user name and appropriate password have to be entered
Login: root
Password: *****
•Press OK
•Select All Log Files
•Right-click syslog and select Save Target As
•Save the file as brick_syslog in the directory C:\AWS\Service\html\whaws\extract.
•Right-click syslog.0 and select Save Target As
•Save the file as brick_syslog.0 in the directory C:\AWS\Service\html\whaws\extract.
•Close the Web Brick window.
•Select Home in the Service software
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 49 of 100
Page 50
50Log and error files
•Copied files will be written by burning the SaveLogs to CD
Windows Eventlog0
Work steps
•The syngo AWS application is running.
•Select Service and Local Service in the menu bar under Options
Fig. 15:
•Enter the service key and press OK. The Home menu appears.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 50 of 100
Medical Solutions
Page 51
Log and error files51
•Select EventLog on the right upper edge of the menu bar.
Fig. 16: EventLog_System
•Select the following parameters as shown in (Fig. 16 / p. 51) .
Domain: System(1/Fig. 16 / p. 51)
Severity: All(2/Fig. 16 / p. 51)
Extended: selected(3/Fig. 16 / p. 51)
NT Message: selected(3/Fig. 16 / p. 51)
Message limit: empty(4/Fig. 16 / p. 51)
Extract to file: System.txt ((5/Fig. 16 / p. 51)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date)).(6/Fig. 16 / p. 51)
Select GO.(7/Fig. 16 / p. 51)
•The log file appears.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 51 of 100
Page 52
52Log and error files
Fig. 17: EventLog_Application
•Select the following parameters as shown in (Fig. 17 / p. 52) .
Domain: Application(1/Fig. 17 / p. 52)
Severity: All(2/Fig. 17 / p. 52)
Facility: Developer(3/Fig. 17 / p. 52)
Extended: selected(4/Fig. 17 / p. 52)
NT Message: selected(4/Fig. 17 / p. 52)
Message limit: empty(5/Fig. 17 / p. 52)
Extract to file: Application.txt(6/Fig. 17 / p. 52)
Time range: last month (e.g.: from dd_mm_yyyy to dd_mm_yyyy (actual
date))(7/Fig. 17 / p. 52)
Select GO.(8/Fig. 17 / p. 52)
•The log file appears.
•Under C:\AWS\Service\extract, select the files Date.txt (e.g.:mm_dd_yyyy.txt) and
Application.txt and copy and paste them to folder C:\AWS\Service\html\whaws\extract
Brickspy.tar.gz file
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 52 of 100
Medical Solutions
Page 53
Log and error files53
Use telnet to start a script.
•Select Service and Local Service in the menu bar under Options
•Select Utilities
•Select Escape to OS
•Enter start telnet brick under parameters
Login: root (and press enter)
Password: brick (and press enter)
•From the brick# command prompt, execute the following command:
brickspy all (use lowercase letters and press enter)
•Wait until the script is finished and enter Exit (and press enter)
Use ftp to get the saved file.
•Select Service and Local Service in the menu bar under Options
•Select Utilities
•Select Escape to OS
•Enter start cmd under parameters
•At the DOS prompt, enter cd C:\AWS\Service\html\whaws\extract (and press enter)
•At c:\temp, enter ftp brick (and press enter)
•User: ftp (and press enter)
•Password: a random e-mail address, e.g., test@siemens.com
•At the ftp prompt, enter bin (and press enter)
•Enter prompt (and press enter)
•Enter cd tmp (and press enter)
•Enter dir (and press enter)
The file brickspy.tar.gz will be displayed
•Enter mget brick* (and press enter)
•After the file has been transferred, enter quit to exit ftp.
•The file brickspy.tar.gz can be found on WH AWS under C:\Temp.
Copy SaveLogs0
This feature is implemented on the WH AWS starting with SW VA11x and can be used if
necessary.
Work steps:
•Select Service and Local Service in the menu bar under Options
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 53 of 100
Page 54
54Log and error files
•Enter the service key and press OK. The Home menu appears.
Fig. 18: Acquisition system
•Select the Acquisition System button.
•Click the Extract AWS Savelog button (2/Fig. 18 / p. 54).
Fig. 19:
•The following mask appears: Extract Servicelog, Extract Drapi.logs, Burn files on CD,
etc.
•Click Extract Service Logs(1/Fig. 19 / p. 54)
•Within seconds, the extracted files are shown.
•Click Extract Drapilogs(2/Fig. 19 / p. 54)
•Within seconds, the extracted files are shown.
•Insert an empty CD into the CD writer.
•Click Burn files on CD(3/Fig. 19 / p. 54).
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 54 of 100
Medical Solutions
Page 55
Log and error files55
•The following message appears: "Insert a blank CD into the CD-Writer and press OK or
press Cancel to abort“
•Press OK
•After the burn procedure, the CD is ejected automatically..
If artifacts are visible in some images, these images (row data and processed data) can
be stored on a second CD for explicit fact-finding in the factory.
Make a test image as described below:
Test images
•Install the compression plate simulator; make sure that the collimator is open to its full
size
•Install the collimator-mounted plexi phantom; make sure no other object is in the radia-
tion field
•Select the "QC-RAW" technique
•Take an image with 28kV and AEC
•Save the image to the CD
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 55 of 100
Page 56
56Generator checks
4Generator checks
5-
General0
The generator section of the Mammomat Novation is of modular construction. The modules were developed so that, in general, no settings need be made when they are
replaced. If an error occurs in the generator (radiation generation), different signals can
be measured or checked in order to delineate or to localize the error.
Documents required0
Service Software, SPB7-250.816.03...
Wiring Diagrams, SPB7-250.844.01...
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 56 of 100
Medical Solutions
Page 57
Generator checks57
Troubleshooting tips0
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 57 of 100
Page 58
58Generator checks
FailureCheckMeasures to be taken if not OK
Control panel
not working
at switch-on
LEDs: “+5VD,
-15 V, +15 V and
28Vprim on
mains input converter
If all LEDs are off, check if there is input voltage at
the input terminals and if S700 is ON.
If there is voltage and S700 is ON, then the mains
input converter is probably defective and must be
replaced. Remember to switch off the input voltage
before disconnecting this unit.
If only one or some of the LEDs are off, switch off
S700 and disconnect the cables from X700, X701,
X705 and X706. If the LEDs still are off when you
switch on S700, then the mains input converter is
probably defective and must be replaced. Remember
to switch off the input voltage before disconnecting
this unit.
If all LEDs are on when you switch on S 700, then the
problem cable has to be identified. If the cable connected to X 700 is causing the failure, then the problem is either the HV inverter or the tube power
supply. If the failure remains after the cable to connector X720 on the tube power supply has been disconnected, the tube power supply is probably
defective and must be replaced. If the failure disappears after the cable to connector X720 has been
disconnected, then the HV inverter is probably defective and must be replaced. If the cable connected to
X701 is causing the failure, then the master board is
probably defective and must be replaced. If the cable
connected to X706 is causing the failure, then the
problem is either the AC inverter or the tube power
supply. If the cable to connector X723 on the tube
power supply is disconnected and the failure
remains, then the AC inverter is probably defective
and must be replaced.
If the failure disappears after the cable to connector
X723 has been disconnected, then the tube power
supply is probably defective and must be replaced.
Attention: connectors X706 and X723 and related
cables are galvanic and connected to mains voltage.
Control panel
not working
at switch-on
LED
“DC_BUS_OK”
on mains input
converter
If neither LED “DC_BUS_OK” or LED
“DC_BUS_FAILURE” is lit, there is probably no signal
from the master board to the mains input converter or
from the control panel to master board.
If LED “DC_BUS_FAILURE” is lit, then the mains
input converter is probably defective and must be
replaced. Remember to switch off the input voltage
before disconnecting this unit.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 58 of 100
Medical Solutions
Page 59
Generator checks59
FailureCheckMeasures to be taken if not OK
Control panel
not working
at switch-on
Control panel
not working
at switch-on
LED “400VDC
live H3”
LED “AC ON” on
AC inverter
If LED H4 is lit but not “H3”, fuse F2 in the mains input
converter is probably defective. Switch off the mains
to the input, wait until LED “H4” is off, then remove
the mains input converter and replace fuse F2. Put
back the unit and connect the cables but before
switching on the the mains; remove the cable from
connector X712 on the AC inverter. Switch on the
mains and the control panel and check if LED “H3” is
lit.
If it is lit, switch off the mains to the input, wait until
LED “H4” is off, then reattach the cable to X712 on
the AC inverter. Switch on the mains and the control
panel. If “H3” is still off, the AC inverter and fuse F2
have to be replaced. But remember to wait until “H4”
is off before disconnecting cables.
If “H3” is off in the first place, the tube power supply is
probably defective and and must be replaced along
with fuse F2. But remember to wait until “H4” is off
before disconnecting cables.
The AC inverter is probably defective and must be
replaced.
If “AC ON” is OK, the failure must be located in the
stand electrical devices.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 59 of 100
Page 60
60Generator checks
Signals on D7500
The actual values of the generator can be measured on board D750 or checked with the
aid of the Mammomat Novation, Software System, SPB7-250.816.03 service program.
Refer to software system Main menu-Normal mode->Actual values.
Measurement points on board D750 are shown in following picture:
Fig. 20: D750
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 60 of 100
Medical Solutions
Page 61
Generator checks61
Test pointDescriptionStatus
5VD_1Logic supply voltage: 5 volt digital
0VD_1,2,3,4,5Logic supply voltage: 0 volt digital
P15VALogic supply voltage: Positive 15 volt analog
0VA_1,2,3Logic supply voltage: 0 volt analog
N15VA Logic supply voltage: Negative 15 volt analog
5V_RS232Logic supply voltage: 5 volt; for communication
with PXCM unit (Brick unit)
0V_RS232Logic supply voltage: 0 volt; for communication
with PXCM unit (Brick unit)
5VI_RS232Logic supply voltage: 5 volt; for communication
with service PC, Opdima WS and printer
0VI_RS232Logic supply voltage: 0 volt; for communication
with service PC, Opdima WS and printer
M_WD_RESET_Signal that resets the master CPU and causes
the entire Mammomat system to restart.
GEN_POWER_ONSignal to the generator that turns on the genera-
tor and the Mammomat system.
DC_BUS_OKSignal from the generator that indicates the
DC-bus voltage is within its allowed voltage window.
HV_OFF_INT_Indicates that the signal M_PLD_HV_OFF_ from
the PLD circuit has detected an event that will
terminate or prevent high voltage generation. It
causes the signal HV_ON from the master board
to go inactive, which immediately prohibits high
voltage generation.
REQ_HV_ONSignal from PLD and CPU that generates high
voltage and thus starts the exposure. Together
with the signal GRID_O_HV_ON, it sets the signal HV_ON.
HV_ON Signal set by the signals REQ_HV_ON and
GRID_O_HV_ON that are output to the generator. One of two necessary signals, together with
the signal EN_HV_ON, that generates high voltage and thus starts the exposure.
EN_HV_ONSignal from the exposure release buttons at the
control panel to the generator. One of two necessary signals, together with the signal HV_ON, to
generate high voltage and thus start the exposure.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 61 of 100
Page 62
62Generator checks
Test pointDescriptionStatus
GEN_POWERED_ONSignal to the PXCM unit (Brick unit) that the
power to the Mammomat system has been
switched on by the operator.
PREP_DR_DET_Signal to the PXCM unit (Brick unit). The positive
edge indicates to the Brick that the generator
preparation phase to perform an exposure has
started and requests that the DR detector prepare for image acquisition.
The negative edge of the signal comes 5 ms
after the negative edge of the signal
GEN_RDY_FOR_DR_EXP at normal exposure
termination; if the hardware detects an error, the
negative edge of the signals will coincide. If a
tube arc occurs, the signal will not go inactive.
GEN_RDY_FOR_
DR_EXP
Signal to the PXCM unit (Brick unit). The generator preparation phase has finished and the generator is ready to generate high voltage to the
tube (i.e., X-ray). The inactivation of the signal
will coincide with the negative edge of the signal
X_RAY_ON at normal termination of the exposure. If the exposure terminates abnormally, the
inactivation of the signal coincides with the inactivation of signal PREP_DR_DET. If a tube arc
occurs, the signal will not go inactive.
0 V <=> OK
4.3 V <=>
Standby
0 V <=>
Acquisition
4.3 V <=>
Standby
0 V Acquisition ON
X_RAY_ONSignal to the PXCM unit (Brick unit). The expo-
sure is in progress.
Note: The signal stays active - at the grid turning
points when X-ray is actually off.
- at the first two tube arcs when X-ray is actually
off.
DR_DET_RDY_F
The DR detector is acquiring image data.5 V <=>
OR_EXP
DR_SYST_RDYThe DROC of the DR system indicates that it is
ready for an exposure.
DR_DET_ERR_The DR system indicates to the generator that it
has detected an error.
4.3 V <=>
Standby
0 V <=> X-ray
ON
Acquisition
ON
5 V <=> OK
0 V <=>Not
ready
0 V <=> OK
5 V <=>Error
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 62 of 100
Medical Solutions
Page 63
Generator checks63
Tube high voltage, current and mAs values0
This section describes how to check the X-Ray tube in regard to the:
High voltage, tube current, mAs value with and without AEC and with a small or large
focus.
These values can be checked by measuring HV_ACT/MA_ACT on board D750 and/or by
using the service PC -> SW -> Normal -> Actual values. Only one of them is necessary,
but both methods are described in the following.
Preparation
•Switch the system OFF (e.g., S700 on main input)
•Connect the oscilloscope as follows:
Channel 1 to measurement point HV_ACT (1V=5 kV) on D750
Channel 2 to measurement point MA_ACT (1V= 40 mA) on D750.
•Connect the service PC to the stand.
•Switch the system ON.
•Start the Service program and check in the anode menu whether the tungsten anode is
enabled.
•In the Service program, select Normal->Main menu ->Normal mode->Actual values
Measurement method
The mAs value must be calculated as the product of the tube current and exposure time.
The tube current and the exposure time can be obtained from the oscillograms. The diagram may have glitches, in which case you have to measure the exposure time t1 and t2.
Example:
Fig. 21:
In the figure, the tube current is 150 mA and the exposure time is t1+t2 in seconds.
The mAs is 150mA*(t1+t2)
NOTE
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
Be aware that with the AEC mode on, a pre-exposure of 50 ms is
done.
Select the correct trigger on the scope so you get the pre-exposure
and all the exposure times.
Page 63 of 100
10.05CS SD 24
Page 64
64Generator checks
The following oscillograms and values should be of assistance to you in troubleshooting
the high-voltage circuit or radiation release.
Checks with a large focus and without AEC
Prerequisites
•Deselect D and H on the control console; the Mammomat is in mAs mode without AEC.
•Make sure the magnification table is removed so that a large focus is selected.
•The check can be performed for each measurement of the table.
MeasurementkVmAsAnode/Filter
13020Mo/Mo
23020W/Rh
330100Mo/Mo
430100W/Rh
•Select the Examination tab card and load a patient for examination
•Set the exposure values on the control console according to the table above.
•Start an exposure.
•Check and compare the values with the following diagrams.
•The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
30 kV 20mAs Mo/Mo large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
40 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 170 ms
Fig. 22: 30 kV 20 mAs MoMo large
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 64 of 100
Medical Solutions
Page 65
Generator checks65
30 kV 20mAS W/Rh large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T (1T<=>40 mA)
40 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 140 ms
Fig. 23: 30 kV 20 mAs W large
30 kV 100mAs Mo/Mo large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 950 ms
Fig. 24: 30 kV 100 mAs MoMo large
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 65 of 100
Page 66
66Generator checks
30 kV 100mAs W/Rh large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 790 ms
Fig. 25: 30 kV 100 mAs W large
Checks with a large focus and with AEC
Prerequisites
•Mount the collimator-mounted plexi phantom
•Register a patient
•Make sure the magnification table is removed so that a large focus is selected.
•Select H on the control console; the Mammomat is now in AEC mode.
•The check can be performed for each measurement of the table.
MeasurementkVmAsAnode/Filter
130--Mo/Mo
230--W/Rh
•Adjust the exposure values on the control console according to the table above.
•Start an exposure.
•Check and compare the values with the following diagrams.
•The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 66 of 100
Medical Solutions
Page 67
Generator checks67
30 kV AEC Mo/Mo large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
120 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 118 mA
Act. exp. time 360 ms
Fig. 26: 30 kV AEC MoMo large
30 kV AEC W/Rh large focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
150 mA 1V/T(1T<=> 40 mA)
200 ms/T
Service->Normal->Act. values
Act. tube current 148 mA
Act. exp. time 330 ms
Fig. 27: 30 kV AEC W large
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 67 of 100
Page 68
68Generator checks
Checks with a small focus
NOTE
Prerequisites
These checks are possible only when a magnification table is
available. Select the small focus. The magnification table is needed. This table is a option.
•Make sure the magnification table is mounted so that a small focus is selected.
•Set the exposure values on the control console according to the table below.
•Start an exposure.
•Check and compare the values with the following diagrams.
•The accuracy of the kV is +/- 5%, for tube current +/- 10% and for mAs product +/- 10%.
MeasurementkVmAsAnode/Filter
13020Mo/Mo
23020W/Rh
330AECMo/Mo
430AECW/Rh
30 kV 20mAs Mo/Mo small focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
24 mA 0.2V/T(1T<=> 8 mA)
825 ms 100 ms/T
Service->Normal->Act. values
Act. tube current 24 mA
Act. exp. time 830 ms
Fig. 28: 30 kV 20 mAs MoMo small
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 68 of 100
Medical Solutions
Page 69
Generator checks69
30 kV 20mAs W/Rh small focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
30 mA 0.2V/T(1T<=> 8 mA)
680 ms 100 ms/T
Service->Normal->Act. values
Act. tube current 30 mA
Act. exp. time 690 ms
Fig. 29: 30 kV 20 mAs W small
30 kV AEC Mo/Mo small focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
24 mA 0.2V/T(1T<=> 8 mA)
400 ms/T
Service->Normal->Act. values
Act. tube current 24 mA
Act. exp. time 1250 ms
Fig. 30: 30 kV AEC MoMo small
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 69 of 100
Page 70
70Generator checks
30 kV AEC W/Rh small focusValues
D750 HV_ACT /mA_ACT
30 kV 1V/T (1T<=> 5 kV)
30 mA 0.2V/T(1T<=> 8 mA)
400 ms/T
Service->Normal->Act. values
Act. tube current 30 mA
Act. exp. time 1090 ms
Fig. 31: 30 kV AEC W small
If there is an error in the high-voltage circuit, one of the following components could be
defective:
•X-ray tube
•HV tank
•HV inverter module
•AC inverter module
Inverter current
This is an analog signal proportional to the output inverter current fed to the HV tank.
I-OUT 30 kV 100mAs W/Rh large focusValues
Module: HV inverter
Measurement point I_OUT
I-OUT
0.5 V/T
100 ms/T
Fig. 32: I-out, 30 kV 100 mAs W large
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 70 of 100
Medical Solutions
Page 71
Generator checks71
If no signal can measured at I_OUT due to an error, then one of the following components
could be defective:
•X-ray tube
•HV tank
•HV inverter
•Mains input converter
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 71 of 100
Page 72
72Generator checks
Filament circuit0
The Service SW, software system, is used for setting the filament current values.
Work steps
•Connect the Service PC.
•Start the SW.
•Turn the system ON.
•Select Main menu->Configuration->Filament.
Fig. 33: Filament
Indicates the adapted filament current of available anode materials for large and small
focus. Adaptation exposures for the filament current are also made in this menu. Pressing
F3 sets the exposure parameters on the control panel, and subsequently pressing Alt-F6
starts the exposure. Repeat this procedure until the tube current deviation is within ±5%. If
the tube has been replaced, program the filament current value to be 200 mA less than
the value stated in the test certificate as a pre-adjustment. (e.g., if tube value is 6900, program 6700.)
If a new filament current has been entered, you must save the new value (press F2)
before learn filament is performed.
NOTE
The Dynamic Learn Filament may change this adjustment. But the
DLF will learn only when the exposure time is longer than 60 ms.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 72 of 100
Medical Solutions
Page 73
Generator checks73
Fig. 34: Miscellaneous, DLF switch
Activation and deactivation of the filament current adaptation. When activated, the adaptation takes place after a total of ten exposures have been made with the actual focus, if
the deviation of the average value from the nominal value is not within -15% to + 5%.
NOTE
Test points on D750
The filament current values can also be measured on board D750.
FIL_CURR_SET Analog output signal to the generator for setting the magnitude of the filament current. The relation between the filament current and the analog control signal
is1A/V.
FIL_CURR_ACT Analog input signal FIL_CURR_ACT, from the generator, that is proportional to the true filament current. The relation between the filament current and the analog signal is 1A/V.
Independent of the selected focus, the current value is 3.5 A (At test point D750
FIL_CURR_ACT = 3.5 V).
During preparation or exposure, the value is increased to approx. 6.6 A to 6.9 A according
to selected focus.
If no filament current is present, there could be an error on one of the following components:
Filament adaptations are possible only when the detector wing is
selected. "dr" is displayed on the control panel.
•X-ray tube
•Tube power module
•AC inverter module
•Cable between X-ray tube and tube power module
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 73 of 100
Page 74
74Image quality
5Image quality
6-
General0
The detector system is an important component of the Mammomat Novation and is key to
image quality. There are 2 detector calibration procedures that should be used if image
artifacts (pixel failures, bad lines, etc.) occur.
•Gain calibration
•Bad pixel mapping
In case of image quality problems, defined test images and files are needed for analysis:
•MPXXXX.map, MPXXXX.cfg, quality.log, MPXXXX.smj
•Test image: 4 cm Plexi, 28 kV, AEC, QC Raw
•Calibration image
These images and files have to be supplied to USC/HSC for support.
Documents required
•Startup Instructions, SPB7-250.815....
•Rep. of parts, SPB7-250.841.01...
•Rep. of parts, SPB7-250.841.02..
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 74 of 100
Medical Solutions
Page 75
Image quality75
Gain calibration0
A gain calibration should be performed first. If the image artifacts persist, then perform
bad pixel mapping.
Gain calibration
•Creates the offset for each pixel value.
•Filename:<Serial_Number>.map
•If there is no gain calibration file that matches the current array serial number, only cali-
bration scans and service scans are allowed.
Furthermore, the gain calibration procedure must be performed weekly. This procedure
can be performed by the customer or the service engineer.
For SW version VA10x, see the user manual.
Fig. 35: _select_calibration
Eight exposures (accepted) have to be taken.
Work steps for SW VA11x
•Select Examination->Calibration
•Select 28 kV and 250 mAs.
•Select the Anode/Filter combination the customer uses most frequently.
•Mount the collimator-mounted plexi phantom.
•Be sure that no additional objects are in the radiation field.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 75 of 100
Page 76
76Image quality
During the calibration procedure, an image can be seen on the monitor, shown in the following example:
Fig. 36: Normal Calibration
This image is normal during the calibration and must not be interpreted as an error.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 76 of 100
Medical Solutions
Page 77
Image quality77
Bad pixel mapping0
A new pixel map should be created if excessive pixel dropout occurs repeatedly on
images displayed in the Examination tab card(Fig. 35 / p. 75). The procedure has to be
performed by service and only when a defective line or pixel occurs.
Work steps
•Select Options -> Service -> Local Service
•Enter the service key and press OK.
•Select Acquisition System.
•Select Examination -> Calibration in the menu bar.
•Select Bad Pixel Mapping.
•Select 28 kV and 250 mAs.
•Select the Anode/Filter combination the customer uses most frequently.
•Insert the collimator-mounted plexi phantom.
•One exposure has to be performed.
•Be sure that no additional objects are in the radiation field.
•Once accepted, the defective line(s) or pixel(s) is/are not released again.
If the line/pixel is still visible after bad pixel mapping calibration, try to lower the mAs value
by 50 mAs, e.g., 200 mAs.
NOTE
NOTE
NOTE
NOTE
Bad pixel mapping cannot be repeated without restoring the old
mapping files. Bad pixel mapping is a continuous procedure. Bad
results remain in the mapping file. There is no undo procedure.
Contact the Uptime Support Center.
The bad pixel mapping procedure cannot be performed until the
detector has been powered on for at least one hour to allow it to
warm up.
To go back, the original mapping files from the supplied detector-specific CD must be loaded.
Bad pixels after bad pixel mapping can result from an incorrect adjustment of the collimator setting or wrong X-ray tube settings. In
case of bad pixel(s), also check the collimator adjustment and
X-ray tube settings.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 77 of 100
Page 78
78Image quality
Qualify log0
Description
An important file of the detector system for image quality is the qualify.log. This provides
information about possible pixel failures in the detector. After every bad pixel mapping,
this file is overwritten with the current data of the detector. This includes possible pixel failures, bad areas or defective lines.
The qualify.log can be found in folder C:\AWS\drul\data or using the Service SW,
Home->Acquisition System->Bad Pixel Mapping.
Images defects in the raw, uncorrected image inside the imageable area must meet the
following performance criteria:
•Not to exceed a total of 2936 defective pixels exclusively or dead lines which corre-
spond to 0.02% of the total amount of pixels.
•Have zero 3x3 pixel areas containing more than 8 defective pixels
•Have zero 5x5 pixel areas containing more than 15 defective pixels.
•Not to exceed a total number of 80 pixel clusters (2x2) where all 4 pixels are defective.
•Not to exceed a total number 20 defective lines.
•Have zero adjacent defective lines (e.g., two defective lines must be separated by at
the least one good line)
The following examples show qualify logs for good and defective detectors:
Fig. 37: Qualify.log
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 78 of 100
Medical Solutions
Page 79
Image quality79
Fig. 38: Bad qualify.log
If the check is not successful (fails) and most dropouts are at the image edges, check the
collimator adjustment. Refer to Collimator Adjustment, System Startup,
SPB7-250.815.05...
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 79 of 100
Page 80
80Image quality
Image examples0
The following image examples show different effects that can occur on the Mammomat
Novation. If one of these effects or artifacts can be seen, perform a gain calibration and, if
required, bad pixel mapping. For grid stripes, check the mechanical grid settings. In some
image examples, the details cannot be recognized quite accurately (pixel failures, grid
stripes). The online version (HTML version) of this document shows higher perceptibility
compared to the printed version. Should the recommended measures be unsuccessful,
then contact USC/HSC. The detector may need to be replaced.
Collimator problemsDescriptionActions
Fig. 39: Bad collimator
Shadow on the
chest wall side
1. Bad line
Perform collimator adjustment according to Startup
Instructions,
SPB7-250.815... Collimator Adjustment chapter.
1. Perform gain calibration
procedure
Perform bad pixel mapping
correction.
2. Perform collimator
2. Shadow on
chest wall side
Fig. 40: Bad line 1 cm
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 80 of 100
adjustment according to
Startup Instructions,
SPB7-250.815... Collimator Adjustment chapter.
Medical Solutions
Page 81
Image quality81
Line problemsDescriptionActions
1 cm line (1)Perform gain calibration
procedure.
If not successful, perform
bad pixel mapping correction.
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
Fig. 41: Bad line 1 cm
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 81 of 100
Page 82
82Image quality
Line problemsDescriptionActions
1 mm line (1)Perform gain calibration
procedure.
If not successful, perform
bad pixel mapping correction.
If not successful, repeat
bad pixel mapping with
lower mAs values (e.g.:
200mAs).
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
Fig. 42: Bad line 1 mm
Bad line (1)Perform an exposure with-
out the phantom and
check whether the line(s)
or bad pixels are located at
the same position.
Fig. 43: Bad line into RMI156
If yes, perform the gain
calibration procedure.
If not successful, perform
bad pixel mapping correction.
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 82 of 100
Medical Solutions
Page 83
Image quality83
Quadrant failureDescriptionActions
Fig. 44: Quadrant failure 1
Quarter panel
defect (lower right
side)
Quarter panel
defect (upper left
side)
Restart the whole system
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
Restart the whole system.
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
Fig. 45: Quadrant failure 4
Fig. 46: Quadrant failure
Quarter panel
defect (lower right
side)
Restart the whole system.
If not successful, the
detector may need to be
replaced. Contact
USC/HSC.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 83 of 100
Page 84
84Image quality
Problems during calibrationDescriptionActions
Fig. 47: Bad lines during calibration
Bad lines
appear during
calibration
Bad lines
appear during
calibration
After gain calibration, turn
the whole system OFF.
Turn the system ON and
repeat the gain calibration.
After gain calibration, turn
the whole system OFF.
Turn the system ON and
repeat the gain calibration.
Fig. 48: Bad lines during calibration
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 84 of 100
Medical Solutions
Page 85
Image quality85
Pixel dropoutsDescrip-
tion
Defective
pixels or
lines on
image
Actions
Perform an exposure
without the phantom
and check whether the
line(s) or bad pixels are
located at the same
position.
Perform gain calibration procedure.
If not successful, perform bad pixel mapping
correction.
Check qualify.log file
whether map file is
within specification
(pass).
If not, the detector may
need to be replaced.
Contact USC/HSC.
Fig. 49: Bad pixel 1
Fig. 50: Pixel problem
Defective
pixels (1) on
image
Perform an exposure
without the phantom
and check whether the
line(s) or bad pixels are
located at the same
position.
Perform gain calibration procedure.
If not successful, perform bad pixel mapping
correction.
Check qualify.log file
whether map file is
within specification
(pass).
If not, the detector may
need to be replaced.
Contact USC/HSC.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 85 of 100
Page 86
86Image quality
Grid linesDescriptionActions
Fig. 51: Raster6
Grid lines
are visible
on image.
Check exposure parameters: If
exposure time is very short (e.g.:
pre-shot), grid lines can be visible
on image. In this case, select the
appropriate exposure parameters.
To eliminate grid lines, check the
grid adjustment according to the
Startup Instructions,
SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the “wheel”
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replacement of Parts,
SPB7-250.841.02.01.02.
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 86 of 100
Medical Solutions
Page 87
Image quality87
Grid linesDescriptionActions
Fig. 52: Raster8
Grid lines
are visible
on image.
Grid lines
are visible
on image.
Check exposure parameters: If
exposure time is very short (e.g.:
pre-shot), grid lines can be visible
on image. In this case, select the
appropriate exposure parameters.
To eliminate grid lines, check the
grid adjustment according to the
Startup Instructions,
SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the “wheel”
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replacement of Parts,
SPB7-250.841.02.01.02.
Check exposure parameters: If
exposure time is very short (e.g.:
pre-shot), grid lines can be visible
on image. In this case, select the
appropriate exposure parameters.
Fig. 53: Raster5
To eliminate grid lines, check the
grid adjustment according to the
Startup Instructions,
SPB7-250.815..,
Chapter 9.:
Adjusting the grid switch
Adjusting the grid location
Adjusting the grid speed
If not successful, refer to Replacement of Parts,
SPB7-250.841.02.01.02.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 87 of 100
Page 88
88Image quality
Bad paddleDescriptionActions
Fig. 54: Bad paddle
Edge of compression paddle visible on
image
Replace compression paddle
Adjust the whole compression
unit according to Replacement
of Parts SPB7-250.841.01...
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 88 of 100
Medical Solutions
Page 89
Image quality89
Heel effectDescriptionActions
Decrease of
beam intensity
to the stand
side.
Problems during “Bad Pixel
Mapping”.
Perform an
exposure with
the other compression paddle.
If not successful, check collimator
adjustment.
If not successful, the X-ray
tube may need
to be replaced.
Contact
USC/HSC.
Fig. 55: Heel effect
Fig. 56:
Decrease of
beam intensity
to the stand
side.
Perform an
exposure with
the other compression paddle
If not successful, check collimator
adjustment.
If not successful, the X-ray
tube may need
to be replaced.
Contact
USC/HSC.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 89 of 100
Page 90
90Image quality
MicrophonyDescriptionActions
Fig. 57: Raster7
Lines and artifacts visible on
image. Detector motions during acquisition
of image data
Lines and artifacts visible on
image. Detector motions during acquisition
of image data
Repeat image acquisition
and be sure there is no
motion during image data
transfer.
If successful, inform the
customer about handling
during data acquisition.
Repeat image acquisition
and be sure there is no
motion during image data
transfer.
If successful, inform the
customer about handling
during data acquisition.
Fig. 58: Microphony
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 90 of 100
Medical Solutions
Page 91
Image quality91
Starry skyDescriptionActions
Fig. 59: Starry sky 1
Many spots are
visible on image
Many spots are
visible on image
Reboot the whole system.
If not successful, perform an
exposure with the collimator-mounted plexi-phantom
and QC-Raw.
If not successful, perform
gain calibration procedure.
If not successful, perform
bad pixel mapping correction.
If not successful, the detector may need to be
replaced. Contact
USC/HSC.
Reboot the whole system.
If not successful, perform an
exposure with the collimator-mounted plexi-phantom
and QC-Raw.
If not successful, perform
gain calibration procedure.
Fig. 60: Starry sky 2
If not successful, perform
bad pixel mapping correction.
If not successful, the detector may need to be
replaced. Contact
USC/HSC.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 91 of 100
Page 92
92Image quality
“Scrambled” ImageDescriptionActions
Fig. 61:
Image information
“scrambled” on
image.
Reboot the whole system
(including Brick)
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 92 of 100
Medical Solutions
Page 93
Image quality93
AEC/Dose problemsDescriptionActions
Fig. 62: Bad exposure 2
Dose for exposure too
low, or acquisition
aborted by user, or
acquisition was terminated before exposure
was completed.
Dose for exposure too
low, or acquisition
aborted by user, or
acquisition was terminated before exposure
was completed.
Check exposure parameters. Exposure time probably too short
Check generator values
according to the Troubleshooting Guide, “Generator checks” chapter,
(these instructions.)
Check exposure parameters. Exposure time probably too short
Check generator values
according to the Troubleshooting Guide, “Generator checks” chapter,
(these instructions.)
Fig. 63: Bad exposure dose1
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 93 of 100
Page 94
94Compression Unit
6Compression Unit
7-
General0
These instructions describe the work steps for troubleshooting the compression unit:
•Check the adjustments (see Replacement of Parts, SPB7-250.841.01..)
•Check the potentiometer R863 (Position transducer of the compression unit)
•Check the potentiometer R861 (compression force)
•Check the movements(up/down)
•Check the detection of the compression paddles
Documents required0
•Replacement of Parts, SPB7-250.841.01..
•Wiring Diagrams, SPB7-250.844.01...
•Service SW Stand, SPB7-250.816.03..
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 94 of 100
Medical Solutions
Page 95
Compression Unit95
Checks0
Check the adjustments0
For the compression unit to function properly, it is very important that it is adjusted
according to the instructions described in Replacement of Parts, SPB7-250.841.01...
Check the potentiometer R863 (position transducer)0
•Attach a compression paddle to the compression unit.
•Move the compression paddle up/down.
The display for thickness changes the values during movement.
•Place a defined object thickness onto the detector, e.g.: 40 mm Plexi-phantom.
•Move down the compression paddle by means of the foot switch until the paddle reach-
es the object on the detector.
The value on the thickness display (D803) has to be the same as the object thickness.
If not, adjust the compression unit according to Replacement of Parts, SPB7-250.841.01..
If the potentiometer still does not work properly after the adjustment, the potentiometer
might be defective and need to be replaced.
Check the potentiometer R861 (compression force)0
•Turn the potentiometer clockwise to a defined position, e.g.: 8 kp (see display on the
lower position of stand).
•Place a scale with a digital display onto the detector and move down the compression
paddle until the movement stops automatically.
•Read the force on the display of the scale. If a deviation >0.3 kp between the values oc-
curs, perform a force adjustment according to Replacement of Parts,
SPB7-250.841.01..
If adjustment is not possible, potentiometer R861 might be defective and need to be
replaced or the compression unit is defective.
Check the movements (up/down)0
•Press the compression pedals (up/down) for electrical movement of the compression
paddle. The paddle moves up/down.
If electrical movement is not possible, check the power supply:
•Check 14 volt supply on board D802
If this voltage is not present, fuse F1 (5AT) on D805 or F803 might be defective or D805 is
defective or motor of compression unit is defective.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 95 of 100
Page 96
96Compression Unit
Check the detection of compression paddle.0
One condition for an exposure release is the detection of a compression paddle. This will
occur by attaching the compression paddle in the guide shaft of the compression unit. The
detection of compression paddles can be checked using the Service SW Stand,
SPB7-250.816.03.. Chapter Service=>Stand test.
Work steps
•Connect service PC and start Service SW of stand, see Software System Stand,
SPB7-250.250.816.03..
Fig. 64: Stand test paddleID
Select Main menu=>Service=>Stand test PaddleID
•Attach a compression paddle to the compression unit and start the test with the service
SW.
A paddle ID will be shown on PC display.
•Repeat this procedure with each compression paddle.
A list with the corresponding PaddleID is shown below:
Compression plate Art.noPlate IDObject Table
Comp. plate simulator1004852700Detector
Comp. plate 20x22H1004852001Detector
Comp. plate 20x22H (Flex)1004852102Detector
Comp. plate 18x24 (High Edge)1004851903Detector
Comp. plate 18x24 (Low Edge)1004851604Detector
Comp. plate 18x24 (Flex)05Detector
Comp. plate 24x30 (High Edge)1004851506Detector
Comp. plate 24x30 (Low Edge)1004851807Detector
Comp. plate 24x30 (Flex)1004851708Detector
Comp. plate Mag.1.51004852209Mag. Table 1.5
Comp. plate Mag. 1.81004852310Mag. Table 1.8
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 96 of 100
Medical Solutions
Page 97
Compression Unit97
Compression plate Art.noPlate IDObject Table
Comp. plate Spot 1.01004851311Detector
Comp. plate Spot 1.51004852412Mag. Table 1.5
Comp. plate Spot 1.81004852513Mag. Table 1.8
Shadow cross14Detector
Comp. plate Stereo1004851415Stereo
Biopsy Hole Plate1004667604Detector
No plate attached99Detector
If one of the paddles cannot be detected (ID 99), the paddle might be defective and need
to be replaced.
If none of the paddles can be detected, a problem occurred on the compression unit or
D805 or the ribbon cable between both:
Fig. 65: D805
•Check the ribbon cable between guide shaft and D805 X831(1/Fig. 65 / p. 97). The
connector on the ribbon cable (2/Fig. 65 / p. 97) has to be plugged in so that the "nose"
on the connector (1/Fig. 66 / p. 98), and the pins are visible from above; they are on the
upper side.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 97 of 100
Page 98
98Compression Unit
Fig. 66: X831
Measurement on D801
If none of the paddles can be detected, a pulse sequence can be checked at TP
SMART_RX on D801:
•When no paddle is attached on the compression unit, a short pulse sequence can be
measured (Fig. 67 / p. 98).
D801 TP SMART_RX
1 V/T
10 ms/T
Fig. 67: Pulse sequence without compression
paddle
When a paddle is attached on the compression unit, a longer pulse sequence can be
measured than without a paddle (Fig. 68 / p. 99)
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 98 of 100
Medical Solutions
Page 99
Compression Unit99
D801 TP SMART_RX
1V/T
10 ms/T
Fig. 68: Pulse sequence with compression plate
simulator
When no pulse sequence occurs, board D805 may be defective or there is a bad
connection between the guide shaft (compression unit) and D805 (X831).
•Err 831 appears when connection D801 <=> is defective or J10 (D805) is defective.
Siemens AGSPB7-250.840.01.01.02MAMMOMAT Novation
Medical Solutions
10.05CS SD 24
Page 99 of 100
Page 100
100Changes to the previous version
7Changes to the previous version
8-
ChapterPageChange
n.a.n.a.Initial version
MAMMOMAT NovationSPB7-250.840.01.01.02Siemens AG
10.05CS SD 24
Page 100 of 100
Medical Solutions
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.