Siemens Mammomat 3000 Collimator Service manual [gb]

Page 1
SIEMENS
MAMMOMAT 3000
SP
Service Instructions
Collimator
0 Siemens AG 1997
The reproduction, transmission or
use of this document or its contents is not permitted without express
written authority. Offenders will be
liable for damages. All rights, including rights created by patent
grant or registration of a utility
model or design, are reserved.
Register 5 Print No.: SPB7-230.061.04.01.02
Replaces: n.a.
English Dot. Gen. Date: 05.97
Page 2
o-2

Revision

Chapter
0
1 2 3
Page
l-4 l-2
l-10
l-6
Rev.
01 01 01 01
MAMMOMAT 3000
Register
5 SPB7-230.061.04
Rev. 01 05.97
Page 2 of 4
TDSP2
Medical Engineering
Siemens AG
Page 3

Contents

o-3
Page
1 Preliminary steps
Remove the swivel arm covers Remove the collimator cover
2
Replacing components and assemblies
Replacing the light localizer lamp or the lamp holder
Replacingthelamp ................................
Replacing the lamp holder (if necessary) Closing the collimator
Replacing the mirror (including the drive mechanics)
General ......................................
Requiredparts. ..................................
Required documents.
Replacing the rotating magnet M7 (for the light localizer mirror)
General ......................................
Requiredparts. ..................................
Required documents. Replacing the spring (for the mirror). Replacing the filter disk
General ......................................
Requiredparts. ..................................
Required documents. Replacing the motor M4 (Format selection) andthelightbarriersS865andS866.
Replacing the motor.
Replacing light barriers S865 and S866.
Functiontest. ...................................
Replacing the motor M5 (Filter selection) andthelightbarriersS867
Replacing the motor.
Replacing light barrier S867.
Functiontest. ...................................
Replacing the motor M6 (Focus correction).
Removing the focus correction device.
Replacing the motor.
Perform a function test. Replacing light barrier S868 (Focus correction).
Preliminary steps .................................
Replacing the light barrier
Perform a function test. Replacing the microswitch.
...........................
............................
...............................
...............................
...............................
.........................
................................
...............................
........................
...............................
..............................
...............................
...........................
...............................
..............................
.............................
..............................
..............................
..................
.....................
..................
.....................
.....................
......................
...................
............
l-1
1 - 1 1 - 1
2-1
2 - 1
.2-l .2 - 1 .2 - 1
2 - 2
.2-2 .2-2 .2 - 2
2 - 2
.2-2 .2-2 .2 - 2
.2 - 3 .2 - 3 .2-3
.2-3 .2 - 3
.2-4 .2 - 4
.2 - 5 .2-5
.2-6 .2 - 6
.2 - 6 .2-7 .2 - 8
.2 - 8 .2 - 8 .2 - 8 .2 - 9
.2 - 9 .2 - 9 .2 - 9
2 - 10
Siemens AG Medical Engineering
Register 5
SPB7-230.061.04 Rev. 01 05.97
Page 3 of 4
TDSP2
MAMMOMAT 3000
Page 4
o-4
Contents
Page
3
Radiation field, enlargement and position
Retrofit: Larger radiation field.
General. ......................................
Required parts
Required documents Reference to legal regulations Checking the light field / radiation field limitation.
General. ......................................
Measurement procedure
Evaluation.
Coincidence of light and radiation fields
Radiation field limitation
Correction.
Finalsteps Update for adjusting the position of the radiation field
General. ......................................
Required parts
Required documents
...................................
.....................................
.....................................
.....................................
...................................
.............................
................................
.............................
...................
..............................
......................
..............................
................................
.................
3-1
3 - 1 3 - 1
3 - 1 3 - 1 3 - 1
3 - 1 3 - 1
3 - 2 3 - 3 3 - 3 3 - 3 3 - 4 3-4
3 - 5 3 - 5
3 - 5 3 - 5
MAMMOMAT
3000
Register
5
SPB7-230.061.04 Rev. 01 05.97
Page 4 of 4
TDSP2
Siemens AG
Medical Engineering
Page 5

Preliminary steps

Fig.
1
l-l
Remove the swivel arm covers
l
Switch off the system.
l
Remove the screws from the X-ray tube covers at the front, as well as the left and right sides (1, 2 and 3/Fig. 1).
Fig. 2
Remove the collimator cover
l
Remove the four screws 5/Fig. 2).
piiq
l
Allow the cover to hang down (4/Fig.2) by the collimator cable and, if necessary, secure it to the swivel head with adhesive tape or cable ties.
Siemens AG Medical Engineering
Remove the four screws shown in Fig. 2 only!
Register
5
SPB7-230.061.04 Rev. 01 05.97
Page 1 of
TDSP2
2
MAMMOMAT
3000
Page 6
1-2
Preliminary steps
This page intentionally left blank.
MAMMOMAT 3000
Register 5
SPB7-230.061.04 Rev. 01 05.97
Page 2 of 2
TDSP2 Medical Engineering
Siemens AG
Page 7

Replacing components and assemblies

Fig. 1 Fig. 2
Replacing the light localizer lamp or the lamp holder
l
Remove the swivel arm and collimator covers.
2-1
Replacing the lamp
l
Switch the system on.
l
Select the MO focus. The carriage (l/Fig.l)will move to the bottom.
l
Switch the system off.
l
Remove the two screws shown in (3/Fig.l).
l
Remove the lamp holder as shown in (2/Fig.l, l/Fig.2).
Be careful not to damage the filter disk!
l
Replace the lamp (4/Fig.2).
Do not touch the lamp with your fingers!
Wear gloves or use a soft cloth when replacing the lamp.
Replacing the lamp holder (if necessary)
l
Remove the springs (5/Fig.2).
l
Unsolder the connection cable from the lamp holder.
l
Remove the lamp from the old lamp holder and install it in the new one.
Closing the collimator
l
Reinstall the lamp holder.
l
Check the light field / radiation field as described in chapter 3.
l
Reinstall all covers.
Siemens AG Medical Engineering
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page1 of10
TDSP2
MAMMOMAT 3000
Page 8
2-2
Replacing components and assemblies
Replacing the mirror (including the drive mechanics)
General
In some systems, it may happen that the mirror sporadically does not snap back into
place after the exposure. If this is the case, the following procedure must be performed. When a defective mirror (or the associated drive mechanics) must be replaced, proceed
according to these instructions.
Required parts
Upgrade kit, complete 63 96 605 X041 E Mirror only 62 45 042 X 041 E
Required documents
Update instructions “Replacing the mirror (including the drive mechanics)” RXB7-230.092.05.02
Replacing the rotating magnet M7 (for the light localizer mirror)
General
The preceding section describes the procedure for replacing the mirror including the drive mechanics.
This procedure should also be followed when replacing a defective rotating magnet (or the associated drive mechanics).
Required parts
Rotating magnet Item 62 46 842 X041 E No.
Required documents
Update instructions “Replacing the mirror (including the drive mechanics)” RXB7-
230.092.05.02...
MAMMOMAT 3000
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page2of 10
TDSP2
Medical Engineering
Siemens AG
Page 9
Replacing components and assemblies
Fig. 3 Fig. 4
Replacing the spring (for the mirror)
l
Remove the swivel arm and collimator covers.
2-3
l
Remove the 4 screws shown in (l/Fig.3).
l
Carefully lower the collimator assembly (2/Fig.3).
l
Replace the spring (5/Fig.4).
l
Reattach the collimator assembly with the screws.
l
Switch the system on.
l
Switch on object illumination by actuating the footswitch (for compression)
0 the mirror must correspondingly snap in or out when switching object illumination
on or off.
l
Check the coincidence of the light field/radiation field as described in chapter 3.
l
Reattach all covers.
Replacing the filter disk
General
The following instructions describe how to replace the filter disk.
Required parts
Filter disk 64 20 694 X041 E
Required documents
Repair instructions “Replacing the filter disk” RXB7-230.091.02.01...
Siemens AG Medical Engineering
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page3of 10
TDSP2
MAMMOMAT 3000
Page 10
2-4
Fig. 5 Fig. 6
Replacing components and assemblies
Replacing the motor M4 (Format selection)
and the light barriers S865 and S866
l
Remove the covers from the swivel arm and the collimator.
l
Remove the 4 screws (l/Fig. 5) from the collimator.
l
Carefully lower the collimator unit.
Replacing the motor
l
Remove connector X854 on D805.
l
Remove the nut (3/Fig.6).
l
Remove the motor (5/Fig.6) from the drive (4/Fig.6).
l
Remove the 2 screws (6/Fig.6) and take out the old motor.
l
Install the new motor.
MAMMOMAT 3000
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page4of 10
TDSP2
Medical Engineering
Siemens AG
Page 11
Replacing components and assemblies
Fig. 7
Replacing light barriers S865 and S866
l
Replace light barrier S865 (7/Fig.7).
2-5
l
Replace light barrier S866 (8/Fig.7).
l
Close the collimator.
Function test
l
Alternately move each table 18x24 and 24x30 into the exposure position. The beam limiting device must move freely and must shut off in the correct end positions.
l
Check the position of the radiation field and the coincidence of light and radiation fields according to the procedure described in chapter 3. If necessary, adjust them.
l
Reinstall all covers.
Siemens AG Medical Engineering
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page5of 10
TDSP2
MAMMOMAT 3000
Page 12
2-6
Replacing components and assemblies
Fig. 8
Fig. 9
Replacing the motor M5 (Filter selection)
and the light barriers S867
l
Remove the swivel arm and collimator covers.
Replacing the motor
l
Remove connector X867 on D805.
l
Remove the 4 screws shown in (l/Fig. 8) (if necessary, remove the screws from the X-
ray tube and lift it up).
l
Take the old motor out (2/Fig.8).
l
Install the new motor using the reverse procedure and connect it.
l
Check that the filter rotates easily. If necessary, adjust the position of the motor.
l
Secure the motor by tightening the screws.
Replacing light barrier S867
l
Remove the 4 screws (3/Fig.9) on the collimator.
l
Carefully lower the collimator (4/Fig.9) (this improves access to the light barrier).
MAMMOMAT 3000
Register 5 SPB7-230.061.04
Rev. 01
05.97
Page6of 10
TDSP2
Medical Engineering
Siemens AG
Page 13
Replacing components and assemblies
Fig. 10
l
Remove the 2 screws as shown in (5/Fig.l0).
l
Replace light barrier S867 (G/Fig.1 0).
2-7
l
Reattach the collimator.
Function test
l
Select each filter consecutively on the control panel. The filter disk must stop in the exact positions, respectively.
l
Reattach all covers.
Siemens AG Medical Engineering
Register 5 SPB7-230.061.04
Rev. 01
05.97
Page7of 10
TDSP2
MAMMOMAT 3000
Page 14
2-8
Fig. 11 Fig. 12
Replacing components and assemblies
Replacing the motor M6 (Focus correction)
l
Remove the swivel arm and collimator covers.
Removing the focus correction device
l
Secure the focus correction device according to Fig.1 1 (e.g. with cable ties)
l
Loosen the 4 screws of the focus correction device, remove it and allow it to hang down
by the cable ties.
Replacing the motor
l
Remove connector X879 on D805 and secure the carriage (l/Fig.1 2).
l
Remove the nut (2/Fig.l2) and the 2 screws (3/Fig.l2).
l
Remove the old motor (4/Fig.l2). Install the new motor using the reverse procedure and
connect it.
l
Reattach the focus correction device.
Perform a function test
l
Switch the system on and select all four foci consecutively.
(To switch to the small focus, insert the magnification table.)
Check that the focus correction device travels smoothly.
l
Check the size and position of the radiation field as described in chapter 3.
l
Correction procedure: - connect the service PC.
- Using the module “Configuration Beam limiting device” slightly change the size of the radiation field.
piiq
l
MAMMOMAT 3000
This will influence the position of the radiation field.
Switch the system off and reattach all covers.
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page8of 10
TDSP2
Medical Engineering
Siemens AG
Page 15
Replacing components and assemblies
Detail
Fig. 13
Replacing light barrier S868 (Focus correction)
Preliminary steps
2-9
l
Remove the swivel arm and collimator covers.
l
Switch the system on.
l
Select the MO focus.
l
Select the small focus (insert the magnification table).
l
Switch the system off.
Replacing the light barrier
l
Replace light barrier S868 (1 /Fig. 13).
Perform a function test
l
Switch the system on.
l
Select all four foci consecutively (to switch to the small focus, insert the magnification table)
l
Check the size and position of the radiation field as described in chapter 3.
l
Correction suggestion: - Connect the service PC.
- Using the module “Configuration/Beam limiting device”, slightly change the size of the radiation field.
l
Switch the system off.
l
Reattach all covers.
Siemens AG Medical Engineering
This will influence the position of the radiation field.
Register 5
SPB7-230.061.04 Rev. 01
05.97
Page9of 10
TDSP2
MAMMOMAT 3000
Page 16
2-10 Replacing components and assemblies
Fig. 14
Fig. 15
Replacing the microswitch
l
Remove the swivel arm and collimator covers. Remove the four screws (l/Fig.1 4) at the inside of the collimator cover.
l
l
Remove the holder.
l
Replace the switch (2/Fig.l5).
l
Reassemble the collimator cover.
l
Reattach all covers.
MAMMOMAT 3000 Register 5
SPB7-230.061.04 Rev. 01
05.97
Page 10of 10
TDSP2
Medical Engineering
Siemens AG
Page 17

Radiation field, enlargement and position

Retrofit: Larger radiation field
General
Many customers would like the entire film to be exposed. There is an upgrade kit available
for the corresponding change required to the collimator.
Required parts
3-1
Retrofit kit
63 96 985 X041 E
Required documents
Installation instructions “Larger radiation field ” RXB7-230.031.03.01....
Reference to legal regulations
Depending on local regulations, a partial test for the position of the radiation field, as well
as the coincidence of light and radiation fields may be required. The results must be documented in the corresponding protocols.
Checking the light field / radiation field limitation
Adjustment and radiation field limitation
General
- The nominal radiation field width is 248 mm for the 18 cm x 24 cm format and 308 mm for the 24 cm x 30 cm format.
- The radiation field may not exceed the front edge of the film marking device (G/Fig.l).
- On the side close to the chest wall, the radiation field may not exceed the film edge by more than 9.5 mm.
- The light and radiation fields must be within the following tolerances:
Siemens AG Medical Engineering
left edge: right edge: back edge: front edge:
+/- 6.5 mm +/- 6.5 mm +/- 6.5 mm +/- 6.5 mm
Register 5 SPB7-230.061.04 Page 1 of 6 MAMMOMAT 3000
Rev. 01 05.97 TDSP2
Page 18
3-2
Radiation field, enlargement and position
Measurement procedure
I \\ \-
Fig. 1
-5
l
Insert a film packet, centering it in the cassette (l/Fig.l).
l
Insert the cassette (l/Fig.1 8) in the object table (2/Fig.l).
l
Place the film bag on the object table (3/Fig.l). It must extend a minimum of 10 mm beyond the edge of the chest wall.
If there is no film bag available, two film cassettes may be placed next to each other on
the object table.
l
Place the centering cross (4/Fig.l) on the film bag or on the cassettes.
l
Switch on the light localizer and mark the edges of the light field with four washers (e.g. coins) (5/Fig.l). Record the distance of the centering cross (scale values) to the side edges of the object table and the front edge of the film marking device (G/Fig.l).
l
Release radiation (MO/MO) and develop the film.
MAMMOMAT 3000 Register 5 SPB7-230.061.04 Page 2 of 6 Siemens AG
Rev. 01 05.97 TDSP2 Medical Engineering
Page 19
Radiation field, enlargement and position
Evaluation
3-3
Fig. 2
Centering cross
4.
left _
Chest-wall side Cassette film
\
--
1 back* ,
--) -- right
*\
front
\
[----.!--
3 *
b-
Light field
Allowed overradiation
1.5 - 9.5 mm
Radiation field
*
left: +/- 6.5 mm
*
right: +/- 6.5 mm
.front: +I- 6.5
*
back: +I- 6.5
mm
mm
Coincidence of light and radiation fields
Evaluation of the film in the film bag:
l
Draw the position of the light field by connecting the outer edges of the coin markings as
shown in Fig. 2.
l
Compare the outer edges of the light field with those of the radiation field.
l
The maximum deviation between the radiation field and the light field may not exceed
6.5 mm on any one side.
Radiation field limitation
Compare the film from the cassette with the one from the film bag using the scale on the centering cross:
l
The cassette film must be completely exposed on the side opposite the chest wall.
However, the radiation field must not extend beyond the front edge of the film marking
device.
l
The overexposure on the chest wall side must fall between a minimum of 1.5 mm and a maximum of 9.5 mm.
l
The radiation field must not extend over the side edges of the object table.
l
The distance to the edges must be at least 7 mm. To check this, compare the scale values of the film from the film bag with the values recorded on page 19.
Refer to the next page for correction procedures.
l
Repeat the test with all object tables and both format selectors (basic tables).
Siemens AG Medical Engineering
Register 5 SPB7-230.061.04
Rev. 01 05.97
Page 3 of 6 MAMMOMAT 3000
TDSP2
Page 20
3-4
I
Fig. 3
Radiation field, enlargement and position
Correction
The light field can be adjusted in the horizontal direction (X-direction) with the adjustment screw shown in (l/Fig.3). The light field can be adjusted in the Y-direction with the screw shown in (2/Fig.3)
Adjust the radiation field as follows:
l
Loosen the screws (2,3/Fig. 4) on both sides of the collimator.
l
Adjust the radiation field in the chest wall /stand wall direction (Y-direction) by shifting the collimator.
If a horizontal adjustment is necessary, refer to Speed Info RX 768-95 (Update kit 63 82
886).
I
“0 8 8 8
Fig. 4
MAMMOMAT
Final steps
l
Reattach the collimator and X-ray tube covers.
l
Complete the test protocol (if required), e. g. radiation field limitation check.
3000
Register 5
SPB7-230.061.04 Rev. 01 05.97
Page 4 of
TDSP2
6
Medical Engineering
Siemens AG
Page 21
Radiation field, enlargement and position
Update for adjusting the position of the radiation field
General
- The following Update Instructions describe how to adjust the position of the radiation field only.
- The size of the field cannot be adjusted.
- If the customer wishes to have a larger radiation field, the complete format selection system must be replaced, as previously described.
On some systems, collimation is not symmetrical even with the CC (cranio/caudal) beam.
- An upgrade kit is available which allows this to be readjusted in the field (Item No. 63 82 886 X041 E). This kit contains instructions in English and auxiliary measurement devices for centering the radiation field.
- The main modification is the enlargement of the holes for the collimator holder.
The parts can be used for all affected units in your region.
Required parts
3-5
Update kit Item No. 63 82 886 X041 E
Required documents
Update Instructions “Collimator holder” RXB7-230.092.01 .Ol . . . . .
TD SP 2 / Guggenmoss TD SP 1 / Schlee SMS lselin / O’Donnell
Siemens AG Medical Engineering
Register 5 SPB7-230.061.04 Page 5 of 6 MAMMOMAT 3000
Rev. 01 05.97 TDSP2
Page 22
3-6
Radiation field, enlargement and position
This page intentionally left blank.
MAMMOMAT 3000
Register 5 SPB7-230.061.04
Rev. 01 05.97
Page 6 of 6 Siemens AG
TDSP2 Medical Engineering