Siemens Mammomat 3000 Installation recommendations

Page 1
MAMMOMAT 300/3000
Installation
Installation recommendations
External exposure lamp
SP
© Siemens Elema AB 1997
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 3 English Print No.: RX B7-120.038.04.03.02 Release Date:11.97
Replaces: RX B7-120.038 .04.02.02 64 60 112
Page 2
Chapter 0 Revisions
Page 2 of 15
This page intentionally left blank
Letzte Seite des Kapitel 0
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN 08.12.1997 Installation
Page 3
Contents Chapter 0
Page 3 of 15
Page
1__General__________________________________4
Scope of this document. . . . . . . . . . . . . . . . . . . . . . 4
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Time required . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2__Installation information ____________________5 3__Installation - M300_________________________6
Cables and components - M300 . . . . . . . . . . . . . . . . . 6
Preparation of cable for M300 . . . . . . . . . . . . . . . . . . 7
Removal of covers - M300 . . . . . . . . . . . . . . . . . . . . 8
Connecting and clamping - M300 . . . . . . . . . . . . . . . . 8
Cable placement - M300.. . . . . . . . . . . . . . . . . . . . . 9
Strain relieving - M300 . . . . . . . . . . . . . . . . . . . . . . 10
Connecting the relay - M300 . . . . . . . . . . . . . . . . . . . 10
Final procedure - M300 . . . . . . . . . . . . . . . . . . . . . . 10
4__Installation - M3000_______________________11
Cables and components - M3000 . . . . . . . . . . . . . . . . 11
Preparation of cable for M3000. . . . . . . . . . . . . . . . . . 12
Removal of covers - M3000 . . . . . . . . . . . . . . . . . . . 13
Connection and clamping - M3000. . . . . . . . . . . . . . . . 13
Cable laying - M3000. . . . . . . . . . . . . . . . . . . . . . . 14
Strain relieving - M3000 . . . . . . . . . . . . . . . . . . . . . 15
Connecting the lamp - M3000 . . . . . . . . . . . . . . . . . . 15
Final procedure - M3000 . . . . . . . . . . . . . . . . . . . . . 15
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN 08.12.1997 Installation
Page 4

Chapter 1 General

Page 4 of 15
General 1

Scope of this document 1

The scope of this document is to describe the pro cedure of instal ling an external exposure lamp to indicate preparation and exposure on the MAMMOMAT 300/3000.
All components and electrical data are recommendations and Siemens Elema takes no responsibility whatsoever for installation other than described in this document.

Function 1

MAMMOMAT 300/3000 enables an external connection of a lamp to indicate risk of radiation. The warning lamp lig h ts up under preparation and goes out when the exposure is through.

Limitations 1

Some MAMMOMAT 300 are not prepared for external exposure lamp (serial number 1006 to 1021).
Verify that the MAMMOMAT in question is equipped with connectors X20, X21 and X22 on the p.c. board D700.
All MAMMOMAT 3000 are prepared for external exposure lamp.
If not, installation will not be possi ble.

Tools 1

Standard tools for installation.

Time requi red 1

1h
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
Page 5

Installation information Chapter 2

Page 5 of 15
Installation information 2
1.5m min.
Figure 1
. At preparation and exposure the lamp lights.
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
Page 6

Chapter 3 Installation - M300

Page 6 of 15
Installation - M300 3

Cables and components - M300 3

Relay:
Cable between generator and relay:
24VDC rated 5A, UL-CE-CSA, e.g. Omron MK-I/-S series.
2
4 x 0.5mm CE-CSA, e.g.Kromberg & Schubert, Schaltflex 321464.
shielded, 6A, 300V, max 150nF/km,7.0 mm, UL-
Cable between relay and power supply:
UL-CE-CSA power supply cable, FSchHARSVT, 3 x 18 A W G.
Diodes:
Connectors:
Fuse:
Fuseholder:
3 ×1N4004
3 × FASTON 6.3 x 0.8 with insulation.
3.15 AT
Fulfilling national standards.
Lamp:
AC voltage Lamp
7.3 mm, e.g. FELLER
Lampholder:
Fulfilling national standards.
Cable tie:
Strain relieve bushing:
Labels:
Clamp:
Ferrite core:
100VAC 40W
110-120VAC 60W
220 - 280VAC 100W
1
×
× 1
× 6 1 × 6 mm
× 1
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
Page 7
Installation - M300 Chapter 3
Page 7 of 15

Preparation of cable for M300 3

Cable according to Figure 2.
Note: Min. dimensions.
Figure 2.
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
Page 8
Chapter 3 Installation - M300
Page 8 of 15

Removal of covers - M300 3

Switch off mains.
Remove the generator front cover (1/Fi gure 3).
Remove the cable outlet cover (2/ Figure 3).
Remove the cable duct covers.
Cable protection
1
2
Figure 3.

Connecting and clamping - M300 3

Connect the cable conductors marked X20, X21 and X22 t o corresponding generator
connectors X20, X21 and X22 on the p.c. board D700 according to Fi gure 4.
F1
F2
F3 F4 F5
x22
x21
x20
x22
x21
x20
F10 F20
D700
9504 0018
Figure 4.
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
Page 9
Installation - M300 Chapter 3
Page 9 of 15
Clamp the cable to the generator floor or below the D700 p.c. board, see Figure 5.
To X20, X21 and X22 on D700
Figure 5.
Attach a ferrite core to the cabl e.

Cable placement - M300. 3

Lay the cable between generator and stand , in
one of the outer cable duct compartments, see Figure 6.
Figure 6.
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
Page 10
Chapter 3 Installation - M300
Page 10 of 15

Strain relieving - M300 3

The cable for external exposure lamp shall be strain relieved at the stand, see Figures 7 and 8.
Figure 7.
Figure 8.
CE-version
Place the strain relief bushing and c able tie in position.
Tighten the cable tie, cut of the remai ns and push the bushing over the cable tie to
achieve proper relieving.
Place the bushing in the notch and tighten t he bushing.

Connecting the relay - M300 3

The cable shall be connected to a relay according to Figure 9.
Note!
PCB. D700
Do not use power from the generator
X20
+24V
1N4004
X21
DS_
X22
SS_
1N4004
1N4004
.
40 - 100W
N
RL 1
3.15AT
100 - 280V
L
Figure 9.
Connect the lamp holder with fuse to the relay.

Final procedure - M300 3

Close the generator box, the cable outlet cover and the cable duct covers.
Switch on mains.
Connect power to the lamp and perform a test expos ure and check that the lamp lights
under preparation as well as under expos ure.
Note!
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
The lamp blinks briefly between preparation and exposure.
Page 11

Installation - M3000 Chapter 4

Page 11 of 15
Installation - M3000 4

Cables and components - M3000 4

Cable between generator and power supply:
UL-CE-CSA power supply cable, FSchHARSVT, 3 x 18 AWG.
Connector:
Fuse:
Fuseholder:
Transformer:
Lamp:
Lampholder:
Cable tie:
Strain relieve bushing:
Fulfilling national standards.
AMP Metrimate part No. 207360-1
3.15 AT Primary fuse per transformer specification
Fulfilling national standards.
24 VAC, 25 VA, The transformer shall provide safety isolation from the mains and fulfill nationall standards.
24 V, 25 W max
1
×
1
×
7.3 mm, e.g. FELLER
Labels:
Clamp:
Ferrite core:
1
×
1 × 6 mm
1
×
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
Page 12
Chapter 4 Installation - M3000
Page 12 of 15

Preparation of cable for M3000 4

Cable according to Figure 10.
Note: Min. dimensions.
Figure 10.
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
Page 13
Installation - M3000 Chapter 4
Page 13 of 15

Removal of covers - M3000 4

Switch off mains.
Remove the generator front cover ( 1/Figure 11).
Remove the cable outlet cover (2/ Figure 11).
Remove the cable duct covers .
Cable protection
1
2
Figure 11.

Connection and clamping - M3000. 4

Connect the connector to X41 on the
D706 p.c. board, see Figure 12.
x41
9504 0018
Figure 12.
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
Page 14
Chapter 4 Installation - M3000
Page 14 of 15
Clamp the cable to the generator floor or below the D700 p.c. board, see Figure 13.
To X41
Figure 13.
Attach a ferrite core to the cable.

Cable laying - M3000. 4

Lay the cable between generator and stand, in
one of the outer cable duct compartments, see Figure 14.
Figure14.
MAMMOMAT 300/3000 Register 3 RX B7-120.038.04 SPS-UD Siemens Elema AB Installation Page Rev. 01/ 06.96 Solna, SWEDEN
Page 15
Installation - M3000 Chapter 4
Page 15 of 15

Strain relieving - M3000 4

The cable for external exposure lamp shall be strain relieved at the stand, see Figures15 and 16.
Figure 15.
Figure 16.
CE-version
Place the strain relief bushing and c able tie in position.
Tighten the cable tie, cut of the remai ns and push the bushing over the cable tie to
achieve proper relieving.
Place the bushing in the notch and tighten t he bushing.

Connecting the lamp - M3000 4

Connect cable to the lamp holder in series wi th the power supply according to Figu re 17.
Note!
Do not use power from the generator
PCB. D706
1
3
X41
.
25W
24 V
3.15AT
per transformer specification
Mains
Figure17.

Final procedure - M3000 4

Close the generator box, the cable outlet cover and the cable duct covers.
Switch on mains.
Perform a test exposure and check that th e lamp lights under preparation as well as
under exposure.
Note!
Siemens Elema AB SPS-UD RX B7-120.038.04 Register 3 MAMMOMAT 300/3000 Solna, SWEDEN Page Rev. 01/ 06.96 Installation
The lamp blinks briefly between preparation and exposure. The lamp blinks at system turn off.