Siemens Mammomat 3000 Planning Guide

Page 1
MAMMOMAT 300/1000/3000
Planning Guide
SP
© Siemens AG 1999
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: SPB7-120.021.01.01.02 Doc. Gen. Date: 03.99
Replaces: RXB7-120.021.01.05.02
Page 2
0 - 2 Revision
Chapter Page Revision
all all 01 all all 01 all all 01 all all 01 all all 01
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 4 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 3
Contents 0 - 3
Page
1 _______General Notes _________________________________________________1 - 1
General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
2 _______Room Planning ________________________________________________2 - 1
Dimensions for the MAMMOMAT 300/ 1000/ 3000 . . . . . . . . . . . . . . . . . . .2 - 1
Attaching the remote control panel to the wall or on a table . . . . . . . . . . . . . . 2 - 3
Measurements of the column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Measurements for the generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Generator with integrated radiat ion protection shield and operating console . . . . 2 - 5
Measurements for the generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Generator for external operating console . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Measurements for the radiation protection shield (optional) . . . . . . . . . . . . . . 2 - 7
Range of the swivel arm system . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 8
Stereotactic biopsy attachment (option for MAMMOMAT 3000) . . . . . . . . . . . . 2 - 9
3 _______Preparation for Installation_______________________________________3 - 1
Recommendation for On-site Electro Installation . . . . . . . . . . . . . . . . . . . . 3 - 1
Recommendation for On-site Electro Installation . . . . . . . . . . . . . . . . . . . . 3 - 2
4 _______Technical Data _________________________________________________4 - 1
Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Environmental conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Weight and heat dissipati o n. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Packing and transport routes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Surface colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Additional data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 2
5 _______Project Management ____________________________________________5 - 1
Responsibility of the project manager towards the service contractor . . . . . . . . . 5 - 1
Preparation for installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Siemens AG SPB7-120.021.01 Page 3 of 4 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 4
0 - 4 Contents
This page intentionally left blank.
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 4 of 4 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 5

General Notes 1

1 - 1

General notes 1

- With distributi on of this revision level, all precedi ng planning guides, Speed - Infos (PG’s ) and drafts lose their validity.
- All layouts i ssued by the Planning Departments must bear a note ref erring to the installati on and delivery conditions of Siemens Medical Engineering Group. The installation and del ivery conditions must be submitted with the layouts.
- Unless otherwise spe cified, all dimensions are indi cated in "mm".
- The symbol indicates a change (see revision status).
- Orientation poi nts Points specific to system components to whi ch reference is made when positioning syst em components to each other or in the room. The isocenter of a radiographic syste m is always illustrated as the orien tation point.
- Fixpoints Clearly marked points on syst em components, installation ceil ing, walls or floor on which cable outlets are located. Illustration in the dra wings: circle with letter /number-combination. The cable lengths establish the maximum fixpoi nt distances and thus the maximum distances between the individual system components.
- Room height The room height is the distance measured f rom the top surface of the floor to the bot tom surface of the ceiling structural element s (Unistrut rails) (bottom surfac e of drop ceiling).

Safety 1

- The provisions of t he relevant fire protection reg ulations must be observed for the premi ses.
- The system has been developed acc ording to EN 60601 - 1.
- Minimum dimensions (e. g. room heights, safety distances) i ndicated in the planning guides are marked "min."
- Basic resistanc e to electromagnetic sources of inter ference. Result of lightning discharges. The protection targets of the diff erent lightning protection ar eas up to the unit connection are also specified in the IEC 1024, DIN 48810, VDE 0675 and in th e DEMVT recommendations.
Hotline Tel. (+49) (0) 9191/18 - 8080
Siemens AG SPB7-120.021.01 Page 1 of 4 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 6
1 - 2 General Notes
This Planning Guide is valid for:
MAMMOMAT 300 beginning with serial number 04 001 ...
MAMMOMAT 3000 beginning with serial number 03001 .. .
- Stereotactic biopsy attachment beginning with serial number 03001 .. .
MAMMOMAT 3000 beginning with serial number 05001 .. .
The MAMMOMAT 3000 beginning with serial number 05001 is available in 8 versions as the customer can make 3 choises when ordering the MAMMOMAT.
Common parts
Stand base module
Generator base module
X-ray tube P40 Mo/W-100G
1 A
2 A
3 A
W and Mo focus
Collimator with 3 filters
(30 µm Mo, 25 µm Rh and
50 µm Rh)
"Flying wing"
2 support plates with AEC
Generator mounted control
panel and radiation
protection
Selection 1
Selection 2
Selection 3
Mo focus, W focus inhibited
Collimator with 2 filters
(30 µm Mo and 25 µm Rh)
One table
1 support plate with AEC
Remote control panel
1 B
2 B
3 B
Options Free-standing radiation shield with lead equivalence of 0,3 mm Pb/110 kV, Free-standing radiation shield with lead equivalence of 0,1 mm Pb/110 kV,
Bus Installation Kit, Printer Interface, Stereo-tactic biobsy attachment, Opdima
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 4 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 7
General Notes 1 - 3
This Planning Guide is valid for:
MAMMOMAT 3000 beginning with serial number 06501 ...
In addition to the 8 versions of the MAMMOMAT 3000 beginning with Serial number 05001, there are 4 versions beginning with Serial number 06501 of the MAMMOMAT 1000 available. The customer can make 2 more choises when ordering the MAMMOMAT.
Common parts
Stand base module
Generator base module
X-ray tube P40 Mo/W-100G
Mo focus,
4 A 4 BSelection 4
5 A 5 BSelection 5
W focus inhibited
Collimator with 1 filter
(30 µm Mo)
One table
1 support plate with AEC
Generator mounted control
panel and radiation
protection
Mo focus, W focus inhibited
Collimator with 2 filters
(30 µm Mo and 25 µm Rh)
Remote control panel
Options Free-standing radiation shield with lead equivalence of 0,3 mm Pb/110 kV, Free-standing radiation shield with lead equivalence of 0,1 mm Pb/110 kV,
Bus Installation Kit, Printer Interface, Opdima spot mode
Siemens AG SPB7-120.021.01 Page 3 of 4 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 8
1 - 4 General Notes

System configuration 1

(schematic)
figure shows MAMMOMAT 3000
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 4 of 4 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 9
NOTICE

Room Planning 2

2 - 1

Dimensions for the MAMMOMAT 300/ 1000/ 3000 2

MAMMOMAT 300 beginning with serial number 04001...
MAMMOMAT 3000 beginning with serial number 03001...
MAMMOMAT 1000/ 3000 beginning with serial number 05001... in the configurati on:
Generator with integrated operat ing control and radiation protect ion shield.
Radation
shield Generator with control panel
Focus mark
*1 The cable conduit and generator can be installed eithe r to the left or right of the stand. The maxi-
mum cable length between the stand and the generator is 3.5 m. Two cable conduits, each 2 m long, are included in the shipment. The cables can also be laid below the floor (i.e. on the ceiling of the room below). To install or remove the rear stand cover panel, a minimum space of 180 mm between the wall and stand is required.
*2 Center the radiation protection shield at a ri ght angle to the swivel arm system and the focus
marking as shown in the Lead equivalent value of the radiation protection shield: 0.3 mm Pb/110 kV.
*3 Upon customer request the recommended minimum distance between the base plate of the gen-
erator and the wall can be reduced if in compliance with local requirements.
*4 Cable of the evaluation unit is routed along the wall.
Remember to provide sufficient storage space for the stereotactic biopsy attachment when planning MAMMOMAT 3000 space requirements. Refer to page 2-10.
figure
above.
Siemens AG SPB7-120.021.01 Page 1 of 10 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 10
2 - 2 Room Planning
MAMMOMAT 1000/ 3000 beginning with seri al number 05001 in the configuration:
External operating console attached t o the optional radiation protect ion shield.
Generator
Radation shield
Focus mark
with control panel
*1 The cable conduit and generator can be installed either to the left or right of the stand. The maxi-
mum cable length between the stand and the generator is 3.5 m. Two cable conduits, each 2 m long, are included in the shipment. The cables can also be laid below the floor (i.e. on the ceiling of the room below). To install or remove the rear stand cover panel, a minimum space of 180 mm between the wall and stand is required.
*2 Center the radiation protection shield with the attached operating console at a right angle to the
swivel arm system and the focus marking as shown in the
figure
above.
Lead equivalent value of the radiation protection shield: 0.3 mm Pb/110 kV or 0.1 mm Pb/110 kV.
NOTICE
Order additional cable ducts, item number 64 38 795 X041E, if required between the generator and the operating console. Each cable duct is 2 m long.
*3 Upon customer request the recommended minimum distance between the base plate of the gen-
erator and the wall can be reduced if in compliance with local requirements.
*4 Cable of the evaluation unit is routed along the wall.
NOTICE
Remember to provide sufficient storage space for the stereotactic biopsy attachment when planning MAMMOMAT 3000 space requirements. Refer to page 2-10.
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 10 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 11
Room Planning 2 - 3
NOTICE
NOTICE
Attaching the remote control panel to the wall or on a table 2
If the system is not equipped with the optional radiation protection shield, attach the remote control panel to the wall or on a table located behind an on-site radiation protection shield. Max. cable length to the generator: 10 m No additional cable ducts are delivered for this configuration. Refer to "Notice" on page 2-2.
Dimensions of the external control console:
L 520 mm x W 200 mm x H 80 mm
Attach the remote control panel at the wall or on the table, as shown in the remote control panel either vertically or horizontally to the wall are included in the delivery volume.
We do not supply anchors or screws. Please obtain these locally and ensure that they are suitable for the material used in the wall.
Siemens AG SPB7-120.021.01 Page 3 of 10 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
figures.
The parts for attaching
Page 12
2 - 4 Room Planning

Measurements of the column 2

Film plane
* The measurements shown in brackets apply to the MAMMOMAT 300 only.
NOTICE
Mounting plate of the stand
The figure shows the maximum configuration of the MAMMOMAT 3000 with "flying wing ".
Opening to optionally lay the cable below the floor
Front
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 4 of 10 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 13
Room Planning 2 - 5

Measurements for the generator 2

Generator with integrated radiation protection shield and operating console 2
Mounting plate of the generator (with integrat ed radiation protection shield and operating console)
Opening to optionally lay the cable below the floor (either to the left or right)
Operator’s side
Siemens AG SPB7-120.021.01 Page 5 of 10 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 14
2 - 6 Room Planning

Measurements for the generator 2

Generator for external operating console 2

NOTICE
Mounting plate of the generator (for external operating console)
The base plate for a generator without integrated radiation or operating con­sole is shorter to allow for closer installation to the wall. For this reason the generator has to wall support are included in the delivery volume. Anchors or screws are not supplied, since their selection depends on the material of the wall or the floor, please obtain them locally. We recommend that you use Liebig anchors M6 S10/35 for bolting the generator to the floor.
Opening to optionally lay the cable below the floor (either to the left or right)
be bolted to the wall or the floor. The mounting brackets for
Operator’s side
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 6 of 10 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 15
Room Planning 2 - 7
NOTICE
NOTICE
Measurements for the radiation protection shield (optional) 2
The free-standing radiation shield has to be bolted to the floor. Anchors or screws are not supplied, since their selection depends on the material of the floor, please obtain them locally. We recommend that you use Liebig anchors M6 S10/35.
Mounting plate of the radiation protection shi eld
Opening to optionally lay the cable below the floor
The radiation protection wall can be installe d eit her to the right or to the left of the column.
Siemens AG SPB7-120.021.01 Page 7 of 10 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 16
2 - 8 Room Planning

Range of the swivel arm system 2

MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 8 of 10 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 17
Room Planning 2 - 9
Stereotactic biopsy attachment (option for MAMMOMAT 3000) 2
Biopsy unit
Evaluation unit
Dimensions of the biopsy unit (W x D x H): 405 mm x 370 mm x max. 410 mm Weight of biopsy unit 12 kg
Dimensions of the evalution unit (W x D x H): 390 mm x 355 mm x 145 mm Weight of the evaluation unit 12 kg
Siemens AG SPB7-120.021.01 Page 9 of 10 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 18
2 - 10 Room Planning
Biopsy unit and evaluation unit on a shelf mounted to the wall. The storage space on the shelf should be 500 mm x 900 mm. The maximum height of the biopsy unit is 410 mm.
Line voltage 230 V, 50/60 Hz
NOTICE
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 10 of 10 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Max. cable length between the column and the evaluation unit is 5 m. Depend­ing on the installation, the maximum useful cable length may be limited to 4.5 m.
Page 19

Preparation for Installation 3

3 - 1

Recommendation for On-site Electro Installation 3

Proposal for building power distribution to DIN VDE 0107 or applicable country - specific regulations
1-phase
Power line for X - ray
SIEMENS
To extraneous conductive parts FI - switch 63 A/I
nating and pulsing direct current faults (recommenda­tioning: Siemens FI 5SZ3 466 0KG00 All-cu rrent sensitive or ABB No. F 804 - 63 /0,03. Order from ABB Stotz - contact Heidelberg Tel. 06221 701-00). For other power line voltages, an approriate Fi - switch must be obtained locally.
System breaker
NOTICE
Reduce the output (refer to the installations instructions), if the internal line
30 mA, UN = 400/415 V for alter-
N
must be provided by the customer
with U=11 0 V; 50 A slow-blowing fuse Emergency shutdown button with lock SIEMENS On/off button with pilot lamp
Figures are free lengths in m
impedance Ri exceeds the values given (refer to page 4-1).
This is a dedicated power line. Therefore do not connect the following:
Electrical systems
Heating/air conditioning units
Elevators
General electrical devi ces
Siemens AG SPB7-120.021.01 Page 1 of 2 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 20
3 - 2 Preparation for Installation

Recommendation for On-site Electro Installation 3

Proposal for building power distribution to DIN VDE 0107 or applicable country - specific regulations
2-phase
Power line for X - ray
SIEMENS
To extraneous conductive parts FI - switch 63 A/I
nating and pulsing direct current faults (recommenda­tioning: Siemens FI 5SZ3 466 0KG00 All-current sensitive or ABB No. F 804 - 63 /0,03. Order from ABB Stotz - contact Heidelberg Tel. 06221 701-00). For other power line voltages, an approriate Fi - switch must be obtained locally.
System breaker
NOTICE
Reduce the output (refer to the installations instructions), if the internal line
30 mA, UN = 400/415 V for alter-
N
must be provided by the customer
Emergency shutdown button with lock SIEMENS On/off button with pilot lamp
Figures are free lengths in m
impedance Ri exceeds the values given (refer to page 4-1).
This is a dedicated power line. Therefore do not connect the following:
Electrical systems
Heating/air conditioning units
Elevators
General electrical devices
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 2 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 21

Technical Data 4

4 - 1

Electrical data 4

1~110 V ±10 % 50/60 Hz
Line voltage connection
Power line Recommended cross section: 6 mm
1~/2 ~208, 230, 240, 277 V ±10% 50/60 Hz 2~400 V ±10 % 50/60 Hz
2
Internal line impdance Ri
Fuses (internal) 20 A at 208, 230, 240, 277 and 400 V/ 35 A at 110 V I max. 40 A at 230 V, 80 A at 110 V, 35 A at 400 V (2-phase)
Power consumption Rating (cos ϕ) 0.6 at 230 V
Connection value 4.6 kVA at 230 V
Short term consumption approx. 10 kVA
UN[V]
110 208 230 240 277 400
[Ω]
Ri
max
0.25
0.45
0.50
0.60
0.65
0.85
Continous load approx. 0.8 kVA

Environmental conditions 4

Operation Transport/ Storage
accept. ambient temperature + 10 accept. relative humidity 15 % ... 75 % 10 % ... 100 %
° ... + 40° C- 10° ... + 70° C
Siemens AG SPB7-120.021.01 Page 1 of 2 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 22
4 - 2 Technical Data

Weight and heat dissipation 4

Weight [kg] Heat dissipation [W]
Stand with X-ray tube assembly including base plate
Generator including base plate and radiation protection shield
approx. 285 approx. 200
approx. 139 approx. 680

Packing and transport routes 4

largest crate L 2140 x W 800 x H 1375 mm Weight(including stand and accessories) approx. 470 kg

Surface colors 4

Basic color
Combination color
White pebbled, Med surface no. 3111 similar to RAL gray - white 9002
neutral grey pebbled, Med surface no. 4426 similar to RAL gray 7036 (staubgrau)
medical blue, Med surface no. 564E

Additional data 4

Bus installation RXB7-120.021.02. OPDIMA installation SPB7-230.021.01.
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 2 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering
Page 23

Project Management 5

5 - 1
Responsibility of the project manager towards the service contractor 5
The scope of the project manager’s responsibilities requi res that he
is at the installation site when the system arrives
supports the service contractor in solving problems
clarifies the final location of the individual components
checks to ensure that the installation is proceeding as specifi ed
clarifies problems together with the service contract or prior to the delivery of t he system, e. g.
- establishes the transport route of the truck
- establishes the transport route within the bui lding

Preparation for installation 5

Activities that have to be completed before the system’s arrival or installation
None of the construction workers may be still working in the room
Walls have been finished and pai nted
The floor (possibly pre-installation plate) must be prese nt and finished in the system room
Ceilings are in place
Room lighting has been insta lled and is ready for use
All electrical installation has been completed
The rooms have been swept clean with a broom
A room that can be locked is available as an intermediate storage area for t he components
Siemens AG SPB7-120.021.01 Page 1 of 2 MAMMOMAT 300/1000/3000 Medical Engineering Rev. 01 03.99 TD SD 21
Page 24
5 - 2 Project Management

Protocols 5

This protocol has to be completed and signed by the installation team (servi ce contractor, SIEMENS
service engineer or project manager).
NOTICE
The supervising SIEMENS project manager project management. Furthermore, he is responsible for perfect and proper installation of the system. Perform the further work according to the technical documentation (customer service instructions, installation instructions, etc.).
is responsible for the entire
MAMMOMAT 300/1000/3000 SPB7-120.021.01 Page 2 of 2 Siemens AG Rev. 01 03.99 TD SD 21 Medical En gineering