Siemens Mammomat 3000 Repair Instructions

Page 1
MAMMOMAT 3000
Service
Repair Instructions
SP
Power Upgrade Energy Level 12Power Upgrade Energy Level 12
Power Upgrade Energy Level 12
Reinforcement for "flying wing" locking lever
Repair kit; 64 32 939 X041E
Register 5 English Print No.: SPB7-230.091.06.02.02 Doc. Gen. Date: 08.04
Replaces: SPB7-230.091.06.01.02
© Siemens AG 2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
(64 59 619)
Page 2
2

Document revision level 1

The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer 1

The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro cedures.
1
-
MAMMOMAT 3000 Register 5 SPB7-230.091.06 Page 2 of 8 Siemens AG Service Rev. 02 08.04 CS PS 24 Medical Solutions
Page 3
3

General 2

Purpose 2

This is an instruction on how to reinforce the attachment of the locking lever for the cassette format changer "pivoting Bucky" a.k.a. "flying wing" system if its fastening gets loose.
Explanation
The locking attachment, holding the locking lever, is fixed to the flange by means of two Allen screws M 6 x 25 mm. On Figure 1 one of these two screws is clearly visible under the spanner used to push away the cables obstructing the view. The other screw is in a distance of just about 15 mm, but on this Figure it is covered by cables. During manufacturing the thread-locker LOCTITE 270 is applied and these two bolts are tightened with a torque of 10 Nm. The correct tightening and fixation of these two Allen screws is essential. If they get loose despite of the above mentioned measures, their re-tightening in the field is rather difficult. The approach with a suitable tool is not easy, attempts can be made to use ball-headed Allen key. It is impossi ble to engage a regular Allen key in axial direction. Realising the difficulties the factory offers an additional reinforcement by the repair kit Material number: 64 32 939.
-
NOTE
Even if the kit 64 32 939 is going to be installed, first the described two screws have to be removed (always keep one screw in, to keep the lever in place). LOCTITE 270 has to be applied and the best possible tightening has to be attempted. (See the Work step no: 10.)
Fig. 1
Siemens AG Register 5 SPB7-230.091.06 Page 3 of 8 MAMMOMAT 3000 Medical Solutions Rev. 02 08.04 CS PS 24 Service
Page 4
4

Documents required 2

Installation and Start-Up Instructions for MAMMOMAT 3000; RX B7-230.033.01.....

Tools required 2

Standard service tools.
Set of punches.
Locking ring pliers.
Set of ball-headed Allen keys.
Thread-locker: LOCTITE 270.

Time required 2

2 hours/ 1 person

Contents of update kit 2

The repair kit Material Number 064 32 939 X041E contains: (the repair kit does not contain the spare part Mat No: 64 21 528 - Lever Assembly).
Amount Name
1 Reinforcement for locking lever 64 32 285
2 Screw M6 × 25
1 Locking pin
2 Locking ring
1 Repair Instructions (this document)
NOTE
As of August 2004 the design of the reinforcement bracket MatNo: 64 32 285 has changed. The diameter of the pre-drilled openings has decreased from former 8 mm to actual 7 mm. At the same time the holes aren't strictly circular anymore. They are made in a slotted (elongated) shape so they equally fit if the distance of the threaded holes is 55mm (as it was in Mammomat with serial num ber lower 5000), or if their distance is 60 mm (as can be found in Mammomat with serial numbers higher 5000).
If observing Figure 2 one can see the two Allen screws holding the bracket for fastening the plastic covers. If performing repair according to this instruction those two screws will have to be removed (see work step 4). Follow the procedure of attaching the reinforcement bracket as described further, however, always start with attempting to fasten the two Allen screws mentioned in the Para Explanations above.
-
MAMMOMAT 3000 Register 5 SPB7-230.091.06 Page 4 of 8 Siemens AG Service Rev. 02 08.04 CS PS 24 Medical Solutions
Page 5
Fig. 2

Training of field service engineers 2

5
Due to the technology, used in this equipment, the modification is only allowed to be performed by a field service engineer trained on MAMMOMAT 3000.

Safety notes 2

When performing work and tests the following must be observed;
- the safety notes RA0-000-012
- the general safety notes contained in ARTD Part 2 (CD ROM)
Siemens AG Register 5 SPB7-230.091.06 Page 5 of 8 MAMMOMAT 3000 Medical Solutions Rev. 02 08.04 CS PS 24 Service
Page 6
6

Procedure 3

Dismounting 3

1 Turn the whole swivel arm system -180° for easier service handling;
switch
turn the arm to -180°,
switch
2 Remove the covers around the swivel arm axis, one on each side of the lever.
3 Remove the locking rings and the locking pin.
4 Remove the two screws of the cover mounting angle.
ON
the MAMMOMAT,
OFF
the MAMMOMAT.
Locking pin
Locking rings
Allen screws
Kit mounting
New longer Allen screws
and
Cover mounting angle
3
The picture shows the when the swivel arm system is turned -180°
Lever attachment
4
Flange
The picture shows the parts when the
6
8 Locking ring
swivel arm system is turned -180°.
Reinforcement 5
Edge with corner cut
5 Place the new reinforcement into the cover mounting angle.
cut corner as shown in the figure above.
6 Attach, but not tighten, the two new longer Allen screws.
7 Insert the new locking pin into the holder through the lever package.
8 Mount the locking rings.
9 Tighten the two Allen screws.
10 From opposite side of flange; Tighten the two screws, securing the lever attachment to
the flange.
MAMMOMAT 3000 Register 5 SPB7-230.091.06 Page 6 of 8 Siemens AG Service Rev. 02 08.04 CS PS 24 Medical Solutions
Lever attachment
7 Locking pin
8 Locking ring
Position the edge with the
Page 7
Final procedures
11 Fit the covers.
12 Turn the swivel arm back to 0°.
13 Check the operation of the lever for both positions of the flying wing.
7
Siemens AG Register 5 SPB7-230.091.06 Page 7 of 8 MAMMOMAT 3000 Medical Solutions Rev. 02 08.04 CS PS 24 Service
Page 8
8
This page intentionally left blank.
MAMMOMAT 3000 Register 5 SPB7-230.091.06 Page 8 of 8 Siemens AG Service Rev. 02 08.04 CS PS 24 Medical Solutions