Siemens Mammomat 3000 Replacement of parts

MAMMOMAT 3000
Service
Replacement of parts
SP
Exchange and calibration of the Compression unit
© Siemens AG 2005
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 5 English Print No.: SPB7-230.841.08.01.02 Doc. Gen. Date: 06.05
Replaces: n.a.
0 - 2 Revision
Chapter Page Revision
all all 01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi­sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
MAMMOMAT 3000 Register 5 SPB7-230.841.08 Page 2 of 4 Siemens AG Service Rev. 01 06.05 CS PS 24 Medical Solutions
Contents 0 - 3
Page
1 _______MAMMOMAT compression unit ___________________________________ 1 - 1
Calibration after an exchange (replacement) . . . . . . . . . . . . . . . . . . . . . . 1 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Compression Unit (CU) - Mounting . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Calibration of the compression unit . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Calibration of the compression unit . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Calibration of the "Preset force 200 N" (corresponds to 20 kp) . . . . . . . . . . . 1 - 5
Calibration of the "Thickness bottom" . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Calibration of the "Thickness bottom mm" . . . . . . . . . . . . . . . . . . . . . 1 - 5
Calibration of the "Thickness top" . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Calibration of the "Thickness top mm" . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Calibration of the "Force = 0 N" . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Calibration of the "Force = 200 N" (corresponds to 20 kilopond) . . . . . . . . . . 1 - 6
Optimized Compression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Check of the compression unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Maximal compression - Lowest position of the paddle . . . . . . . . . . . . . . . 1 - 7
Lower Endpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Upper Endpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Concluding steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
2 _______Changes to previous revision ____________________________________ 2 - 1
Siemens AG Register 5 SPB7-230.841.08 Page 3 of 4 MAMMOMAT 3000 Medical Solutions Rev. 01 06.05 CS PS 24 Service
0 - 4 Contents
Page
MAMMOMAT 3000 Register 5 SPB7-230.841.08 Page 4 of 4 Siemens AG Service Rev. 01 06.05 CS PS 24 Medical Solutions

MAMMOMAT compression unit 1

Calibration after an exchange (replacement) 1

General Remarks 1

The mechanical work not described here includes the removal of the swivel arm covers, the disconnection of the connectors X 872, X 873 and X874, and the removal of the bolts holding the compression unit in its place (there are three such bolts in MAMMOMAT 3000/ 3000 Nova, but in MAMMOMAT 1000 there are four of those bolts holding it). This work can be done basically without an additional written instruction.
It is important, however, that the newly installed compression unit be fixed within the pos­sibilities given by the tolerances of the holes in such a manner that the front edge of the compression plates is parallel with the breast-wall edge (thoracic edge) of the object table.
At the same time one should comply with the adjustment procedure as is described in the corresponding Chapter 3 of the currently valid instructions "Replacement of parts - Compression plates", Print no: SPB7-230.841.01. The adhesive gauge block mentioned there has the material number 6483775.
1 - 1
NOTE
When positioning the exchange (replacement) compression unit to its place, especially at the MAMMOMAT 3000/3000Nova, make sure that the two little isolating shrunk-on hoses fixed on the backside of the electromotor do not get entangled in the upper part of the arc (or dome, also called "church"). The electromotor has to move freely under that arc. Several checks moving it by hand are necessary! See attached pictures.
Fig. 1 Cable laying around the CU
Siemens AG Register 5 SPB7-230.841.08 Page 1 of 8 MAMMOMAT 3000 Medical Solutions Rev. 01 06.05 CS PS 24 Service
Loading...
+ 9 hidden pages