MAMMOMAT 2 / 3 / -Stereo
Replacement Instruction
Replacement of the lifting carrige - rope
SP
© Siemens AG 1999
The reproduction, transm ission or
use of this document or itscontents
is not permitted witho ut express
written author ity. O ffender s will be
liable for damages. All rights,
including rights create d by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: RXB7-211.841.01.01.02 Doc. Gen. Date: 12.99
Replaces: n.a.
2 Replacement of the lifting carrige - rope1
Safety information and protective measures 1
Prior to performing any work inside the unit, the system must be disconnected from the
main power by switching off the main power switch. When the sysem is switched off via
the operating panel, voltage is still present at the generator of the Mammomat.
When performing repair work and tests, adhere to:
- the product-specificsafetyinformation in thentechnicaldocumentation(system
binder)
- as well as the general safety information in the ARTD Part 2
Tools required 1
Standard assemby tool
Ratchet with 24 mm hex. insert
Parts required 1
Rope part-no. 11 85 680 G2116
MAMMOMAT 2 / 3 / -Stereo RXB7-211.841.01 Page 2 of 8 Siemens AG
Rev. 01 12.99 TD PS 24 Medical Engineering
Replacement of the lifting carrige - rope 3
Fig. 1
Fig. 2 Fig. 3
Removing the housing 1
• Re move thecovers (1/Fig.1) frombothsides of the front of thecolumn.
• Remove the bellows (2/Fig.1).
• Only Mammomat2: remove the front panel (3/Fig.1) from the generatorcabinet.
Attaching the safety bracket 1
• Take the red safetybracket from its park position (1/Fig.2) and secure it (2/Fig.2) so it
blocks the upward movementof the carrige.
• Move the carrige up (see arrow)to the point where it contacts the block(Fig.3).
• S witch Mammomat and Installation contactor off.
Siemens AG RXB7-211.841.01 Page 3 of 8 MAMMOMAT2 / 3 / -Stereo
Medical Engineering Rev. 01 12.99 TD PS 24