Siemens MAG 1100, MAG 3100, MAG 5100 W, MAG 5000, MAG 6000 Handbook

Page 1
Handbook
Order no.: FDK:521H0723
DKFD.PS.027.W3.02 - A5E00253199
SITRANS F M MAGFLO
Electromagnetic flowmeters Sensor types MAG 1100, MAG 3100, MAG 5100 W Signal converter types MAG 5000, MAG 6000
[]
s
Page 2
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
Siemens Flow Instruments range of electromagnetic flowmeters
2
MAG 5000 MAG 6000 MAG 3000
Ex-d
Outputs 1 current output 1 current output 1 current output
1digital output 1digital output 1 frequency/pulse output 1 relay output 1 relay output 1 relay output
Flow direction Uni/bidirectional Uni/bidirectional Uni/bidirectional
Communication Optional Add-on modules
HART
HART
Display 3 lines 3 lines 2 lines
20 characters 20 characters 16 characters
(optional without display) (optional without display)
Meter uncertainty ±0,5% o.r. ±0,25% o.r. ±0,25% o.r.
Enclosure IP 67, IP 20 IP 67, IP 20 IP 65, IP 67
Custody transfer PTB PTB approval (cold water) OIML R75
OIML R117
Ex-version [EEx ia] IIC [EEx ia/ib] IIB EEx de [ia/ib] IIB T6 Safety barrier [EEx ia] IIC
Power supply 12-24 V a.c./d.c. 12-24 V a.c./d.c. 24 V a.c./d.c.
115-230 V a.c. 115-230 V a.c.
Batch No Yes No
MAG 1100 MAG 1100 MAG 3100 MAG 3100 W MAG 5100 W
FOOD
Size [mm] DN 6-100 DN 10-100 DN 15-2000 DN 25-1200 DN 25-1200
Connection Flangeless Weld-in adapter, Flange Flange Flange
(Sandwich design) clamp adapter,
thread adapter
Pressure [bar] Max. 40 Max. 40 Max. 100 Max. 40 Max. 40
Temperature [
°°
°°
°C] 20 to 200 30 to 150 40 to180 10 to 95 5 to 90
Liner Ceramic (Al
2 O3
) Ceramic (Al2 O3) Neoprene, EPDM, Neoprene and DN 25-40 &
PFA PFA Teflon (PTFE), EPDM D N 350-1200
Polyurethane, hard elastomer
Ebonite, DN 50-300 Linatex
composite elastomer
Electrodes Platinum Platinum AISI 316 Ti, AISI 316 Ti, AISI 316 Ti,
Hastelloy C276 Hastelloy Hastelloy C, PE electrodes PE electrodes
Platinum/Iridium,
Titanium, Tantalum,
PE electrodes
Enclosure IP 67 IP 67 IP 67/IP 68 IP 67/IP 68 IP 67/IP 68
Ex-version EEx ia/ib IIB T4-T6 EEx ia/ib IIB T4-T6
EEx ia/e IIC T3-T6
Page 3
3
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
1.1 Product introduction ................................................................................................................................................. 4
1.2 Mode of operation .................................................................................................................................................... 5
1.3 Pressure Equipment Directive 97/23ECs ................................................................................................................. 6
1.3.1 Exclusions ............................................................................................................................................................... 7
1.3.2 Product marking ...................................................................................................................................................... 7
2.1 Sensor MAG 1100 and MAG 1100 Ex ..................................................................................................................... 8
2.2 Sensor MAG 1100 FOOD ........................................................................................................................................ 9
2.3 Sensor MAG 3100, MAG 3100 Ex and MAG 3100 W ............................................................................................ 10
2.4 Sensor MAG 5100 W ............................................................................................................................................. 12
2.5.1 Signal converter MAG 5000 (DN 6 to DN 1200) .................................................................................................... 13
2.5.2 Signal converter MAG 6000 ................................................................................................................................... 14
2.5.3 Safety barrier (ia/ib) DN 300 ............................................................................................................................... 15
2.5.4 Safety barrier (ia) DN 350 ................................................................................................................................... 15
2.5.5 Cleaning unit .......................................................................................................................................................... 15
2.6 Meter uncertainty ................................................................................................................................................... 16
2.7 Output characteristics MAG 5000 and MAG 6000 ................................................................................................. 17
2.8.1 Sensor cables and conductivity of medium ............................................................................................................ 18
2.8.2 Minimum accept data for cable .............................................................................................................................. 18
2.9 HART
communication add-on module ................................................................................................................. 18
2.10 Cable data (Supplied by Siemens Flow Instruments) ........................................................................................... 18
3.1 Sizing table (DN 6 to DN 2000) .............................................................................................................................. 19
3.2.1 Minimum conductivity ............................................................................................................................................ 20
3.2.2 Liner selection guide .............................................................................................................................................. 20
3.2.3 Electrode selection guide ....................................................................................................................................... 20
3.3 Installation conditions ............................................................................................................................................ 20
3.4 Cleaning unit .......................................................................................................................................................... 24
3.5 Custody transfer approval ...................................................................................................................................... 25
3.6 Signal converter MAG 5000 CT, MAG 6000 CT Sealing ........................................................................................ 25
3.7 Ex installations ....................................................................................................................................................... 26
4.1 Sensor MAG 1100 ................................................................................................................................................. 27
4.2 Sensor MAG 1100 FOOD ...................................................................................................................................... 28
4.3 Sensor MAG 5100 W ............................................................................................................................................. 31
4.4 Sensor MAG 3100 and MAG 3100 W .................................................................................................................... 33
4.5 Signal converter ..................................................................................................................................................... 34
5.1 Potential equalization ............................................................................................................................................. 37
5.2 Inlet protection MAG 3100 ..................................................................................................................................... 38
5.3 Cathodic protected piping ...................................................................................................................................... 38
6.1 Compact installation MAG 5000 and MAG 6000.................................................................................................... 39
6.2.1 Add-on modules MAG 6000 only ........................................................................................................................... 41
6.2.2 Remote installation. At the sensor ......................................................................................................................... 41
6.2.3 Remote installation. Wall mounting ........................................................................................................................ 42
6.2.4 Remote installation. Signal conver ter in 19" insert ................................................................................................. 44
6.2.5 Add-on modules MAG 6000 only ........................................................................................................................... 45
6.2.6 Installation in IP 65 wall mounting enclosure ......................................................................................................... 46
6.2.7 Installation in IP 65 panel mounting enclosure (front of panel) ............................................................................... 47
6.2.8 Installation into the back of a panel ........................................................................................................................ 48
6.3 Signal converter Safety barrier .............................................................................................................................. 49
6.4 Signal converter Cleaning unit ............................................................................................................................... 50
7.1 Signal converter MAG 5000 and MAG 6000 connection diagram .......................................................................... 51
7.2 Wiring diagram for signal converter and sensor ..................................................................................................... 52
8.1 Keypad and display layout ..................................................................................................................................... 56
8.2 Menu build-up ........................................................................................................................................................ 57
8.2.1 Password ............................................................................................................................................................... 57
8.3.1 MAG 5000 and MAG 6000 ..................................................................................................................................... 58
8.3.2 MAG 6000 CT ........................................................................................................................................................ 59
8.4.1 Basic settings ........................................................................................................................................................ 60
8.4.2 Outputs .................................................................................................................................................................. 61
8.4.3 Digital and relay outputs ........................................................................................................................................ 61
8.4.4 Relay outputs ......................................................................................................................................................... 62
8.4.5 Exter nal input ......................................................................................................................................................... 62
8.4.6 Sensor characteristics ........................................................................................................................................... 63
8.4.7 Reset mode ........................................................................................................................................................... 63
8.4.8 Service mode ........................................................................................................................................................ 64
8.4.9 Operator menu setup ............................................................................................................................................. 65
8.4.10 Product identity ...................................................................................................................................................... 66
8.4.11 Change password .................................................................................................................................................. 66
8.4.12 Language mode ..................................................................................................................................................... 67
8.4.13 HART
communication MAG 5000 HART or as add-on module ........................................................................... 67
8.5.1 Flow rate ................................................................................................................................................................ 68
8.5.2 Totalizer ................................................................................................................................................................. 68
8.5.3 Batch ..................................................................................................................................................................... 68
8.6.1 Settings available ................................................................................................................................................... 69
8.6.2 Dimension dependent factory settings MAG 5000 and MAG 6000 ........................................................................ 70
8.6.3 Dimension dependent batch and pulse output settings .......................................................................................... 70
8.6.4 MAG 5000 CT and MAG 6000 CT settings ............................................................................................................ 71
8.7.1 Error handling ........................................................................................................................................................ 72
8.7.2 List of error numbers .............................................................................................................................................. 73
9.1 Converter check list ............................................................................................................................................... 74
9.2 Trouble shooting MAG converter ............................................................................................................................ 75
9.3 Check list MAG sensor .......................................................................................................................................... 76
9.4 Coil resistance table .............................................................................................................................................. 77
10.1 Sensor MAG 1100 ................................................................................................................................................. 78
10.2 Adapter, MAG 1100 FOOD (contains 2 adapters, 2 clamp rings and 2 gaskets) ................................................... 80
10.3 Sensor MAG 3100 and MAG 3100 Ex ................................................................................................................... 82
10.4 Sensor MAG 3100 W ............................................................................................................................................. 84
10.5 Sensor MAG 5100 W ............................................................................................................................................. 85
10.6 Signal converter ..................................................................................................................................................... 86
10.7 Signal converter 19" incl. back plate ...................................................................................................................... 87
10.8 Accessories ........................................................................................................................................................... 88
10.9 Calibration ............................................................................................................................................................. 89
2. Technical data
1. Product introduction
3. Project guidance
4. Dimensions and weight
6. Installation of signal converter
5. Installation of sensor
7. Electrical connection
8. Commissioning
10. Ordering
9. Service
Technical data
Sensor
CommissioningOrdering D & WInstallation of signal conv.
E. c.
Project guidance
Page 4
4
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
All SITRAN S F M MAGFLO electromagnetic flowmeters feature a unique SENSORPROM
®
memory unit which stores sensor calibration data and signal converter settings for the lifetime of the product. At commissioning the flowmeter commences measurement without any initial programming.
1.1 Product introduction
1. Product introduction
SITRANS F M MAGFLO
electromagnetic flowmeters offer reliable, precise and inexpensive flow measurement on all electrically conductive liquids. Typical applications are found in all industries. E.g.:
Water sector: Potable water, treatment of chemicals, waste water and sludge.
Food sector: Dairy products, beer, wine, soft-drinks and fruit juices.
Chemical sector: Detergents, pharmaceuticals, acids and alkalies.
Other sectors: District heating, paper pulp and mining slurries.
SITRANS F M MAGFLO
electromagnetic flowmeters are characterised by simplicity:
Simple to installSimple to commissionSimple to operateSimple to maintain
SITRANS F M MAGFLO
electromagnetic flowmeters are manufactured by Siemens Flow
Instruments A/S - one of the worlds leading makers of flowmeters.
The factory settings matching the sensor are stored in the SENSORPROM
®
unit. Also customer specified
settings are downloaded to the SENSORPROM
®
unit. Should the signal converter be replaced, the new converter will upload all previous settings and resume measurement without any need for re­programming. Furthermore, the "fingerprint" used in connection with the Siemens Flow Instruments Verificator is stored during the sensor calibration.
USM II "Plug & Play" add-on communication modules.
USM II - the Universal Signal Module with "Plug & Play" simplicity makes it easy to access and integrate the flow measurement with almost any system. It ensures the flowmeter will be easy to upgrade to new communication platforms in the future, too.
Page 5
5
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
1. Product introduction
U
i
= When an electrical conductor of length L is moved at velocity v, perpendicular to the lines of flux
through a magnetic field of strength B, the voltage U
i
is induced at the ends of the conductor
Ui = L x B x v
Ui= Induced voltage L = Conductor length = Inner pipe diameter = k
1
B = Magnetic field strength = k
2
v = Velocity of conductor (media) k=k1 x k
2
Ui = k x v, the electrode signal is directly proportional to the fluid velocity
SENSOR
The sensor converts the flow into an electrical voltage (Ui) proportional to the velocity of the flow. The sensor is built up of a stainless steel pipe, 2 coils, electrodes, an isolating liner, housing and where applicable, connecting flanges.
SIGNAL CONVERTER
The signal converter consists of a number of function blocks which convert the sensor voltage into flow readings.
Power supply
2 different types of power supply are available. A 12 - 24 V a.c./d.c. and a 115 - 230 V a.c. switch mode type.
Coil current module generates a pulsating magnetizing current that drives the coils in the sensor. The current is permanently monitored and corrected. Errors or cable faults are registered by the self­monitoring circuit.
Input circuit amplifies the flow proportional signal from the electrodes. The input impedance is extremely high: >10
14
which allows flow measurements on fluids with conductivities as low as
1 mS/cm. Measuring errors due to cable capacitance are eliminated due to active cable screening.
Digital signal processor converts the analog flow signal to a digital signal and suppresses electrode noise through a digital filter. Inaccuracies in the signal converter as a result of long-term drift and temperature drift are monitored and continuously compensated for via the self-monitoring circuit. The analog to digital conversion takes place in an ultra low noise ASIC with 23 bit signal resolution. This has eliminated the need for range switching. The dynamic range of the signal converter is therefore unsurpassed with a turn down ratio of minimum 3000:1.
CAN communication
The signal converter operates internal via an internal CAN communication bus. Signals are transferred to/from a signal conditioner to the display module, internal/external option modules and the dialog module.
Dialog module
The display unit consists of a 3-line display and a 6-key keypad. The display shows a flow rate or a totalizer value as a primary reading.
Output module converts flow data to an analog, a digital and a relay output. The outputs are galvanically isolated and can be individually set to suit a particular application.
1.2 Mode of operation
The flow measuring principle is based on Faraday’s law of electromagnetic induction. The flowmeter consists of a sensor type MAG 1100, MAG 3100 or MAG 5100 W and a signal converter type MAG 5000 or 6000.
Page 6
6
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
12345678901
1
234567890
1
1
234567890
1
1
234567890
1
1
234567890
1
1
234567890
1
12345678901
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
2345678901234567890
1
1. Product introduction
1.3 Pressure Equipment Directive 97/23ECs
Flange PN 6 PN 10 PN 16 PN 25 PN 40 PN 64 PN 100 150 lb 300 lb AWWA
mm
25 N/A N/A N/A N/A SEP N/A SEP SEP SEP N/A 40 N/A N/A N/A N/A SEP N/A PED SEP SEP N/A 50 N/A N/A N/A N/A SEP PED PED SEP PED* N/A 65 SEP N/A SEP N/A PED PED PED SEP PED* N/A
80 SEP N/A SEP N/A PED PED PED SEP PED* N/A 100 SEP N/A SEP N/A PED PED PED SEP PED* N/A 125 SEP N/A SEP N/A PED PED PED PED* PED* N/A 150 SEP N/A PED N/A PED PED PED PED* PED* N/A 200 SEP SEP PED PED PED PED PED PED* PED* N/A 250 SEP LVD PED PED PED PED PED PED* PED* N/A 300 SEP LVD PED PED PED PED PED PED* PED* N/A 350 LVD LVD PED PED PED PED PED PED* PED* N/A 400 LVD LVD PED PED PED PED N/A PED* PED* N/A 450 LVD LVD PED PED PED N/A N/A PED* PED* N/A 500 LVD LVD PED PED PED N/A N/A PED* PED* N/A 600 LVD LVD PED PED PED N/A N/A PED* PED* N/A 700 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 750 N/A N/A N/A N/A N/A N/A N/A N/A N/A PED* 800 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 900 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED*
1000 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1050 N/A N/A N/A N/A N/A N/A N/A N/A N/A PED* 1100 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1200 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1400 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1500 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1600 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 1800 LVD LVD PED* N/A N/A N/A N/A N/A N/A PED* 2000 LVD LVD PED* N/A N/A N/A N/A N/A PED*
MAG 3100 & MAG 3100 W
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
234567890123456789
0
From May 30th 2002 the "Pressure Equipment Directive" is mandatory for all pressure equipment sold within the EU and EFTA.
The approach that Siemens Flow Instruments has taken is outlined in the tables below.
MAG 5100 W
Flange
PN 10 PN 16 PN 40 150 lb 300 lb
mm
25 N/A N/A SEP SEP N/A 40 N/A N/A SEP SEP N/A 50 N/A SEP N/A SEP N/A 65 N/A SEP N/A SEP N/A
80 N/A SEP N/A SEP N/A 100 N/A SEP N/A SEP N/A 125 N/A SEP N/A PED* N/A 150 N/A PED N/A PED* N/A 200 SEP PED N/A PED* N/A 250 LVD PED N/A PED* N/A 300 LVD PED N/A PED* N/A 350 LVD PED N/A PED* N/A 400 LVD PED N/A PED* N/A 450 LVD PED N/A PED* N/A 500 LVD PED N/A PED* N/A 600 LVD PED N/A PED* N/A 700 LVD PED* N/A N/A PED* 750 N/A N/A N/A N/A PED* 800 LVD PED* N/A N/A PED* 900 LVD PED* N/A N/A PED*
1000 LVD PED* N/A N/A PED* 1050 N/A N/A N/A N/A PED* 1100 N/A N/A N/A N/A PED* 1200 LVD PED* N/A N/A PED*
Page 7
7
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
234567890
1
MAG 3100 high temperature PTFE
Flange PN 6 PN 10 PN 16 PN 25 PN 40 150 lb 300 lb
mm
15 N/A N/A N/A N/A SEP SEP SEP 25 N/A N/A N/A N/A SEP SEP SEP 40 N/A N/A N/A N/A PED LVD PED* 50 N/A N/A N/A N/A PED PED* PED* 65 LVD N/A PED N/A PED PED* PED*
80 LVD N/A PED N/A PED PED* PED* 100 LVD N/A PED N/A PED PED* PED* 125 PED N/A PED N/A PED PED* PED* 150 PED N/A PED N/A PED PED* PED* 200 PED PED PED PED PED PED* PED* 250 PED PED PED PED PED PED* PED* 300 PED PED PED PED PED PED* PED* 350 PED PED PED PED PED PED* PED* 400 PED PED PED PED PED PED* N/A 450 PED PED PED PED PED PED* N/A 500 PED PED PED PED PED PED* N/A 600 PED PED PED PED PED*
1. Product introduction
12345678901234567890123456789012123456
1
234567890123456789012345678901212345
6
1
234567890123456789012345678901212345
6
1
234567890123456789012345678901212345
6
12345678901234567890123456789012123456
1234567890
1
23456789
0
1
23456789
0
1
23456789
0
1234567890
MAG 1100
Flange Ceramic Ceramic Ceramic Ceramic Ceramic PFA PFA PFA
mm 150° 200° Ex Ex-d FOOD Ex FOOD
6 SEP N/A SEP SEP N/A N/A N/A N/A 10 SEP N/A SEP SEP SEP SEP SEP SEP 15 SEP SEP SEP SEP SEP SEP SEP SEP 25 SEP SEP SEP SEP SEP SEP SEP SEP 40 PED PED PED PED PED LVD LVD LVD 50 PED PED PED PED PED PED PED PED 65 PED N/A PED PED PED PED PED PED 80 PED PED PED PED PED PED PED PED
100 PED PED PED PED PED PED PED PED
The key to the above tables is as follows.
PED Product covered by PED and only available as fully PED conforming
PED* Product covered by PED but available as either conforming or non conforming to PED
SEP Excluded from PED under Sound Engineering Practice
LVD Excluded from PED under the Low Voltage Directive
1.3.1 Exclusions
All products sold outside of EU and EFTA are excluded from the directive, also products sold into certain market sectors are also excluded. These include
1) Meters used in networks for the supply, distribution and discharge of water.
2) Meters used in pipelines for the conveyance of any fluid from offshore to onshore.
3) Meters used in the extraction of petroleum or gas, including christmas tree and manifold equipment.
4) Any meter mounted on a ship or mobile offshore platform.
1.3.2 Product marking
All meters will now carry either a CE mark or a CE mark followed by 0086
CE
0086: This indicates that the product conforms to PED 97/23/EC, LVD 73/23/EEC + Amendment
93/68/EEC & EMC 89/336 EEC
CE: This indicates that the product conforms to LVD 73/23/EEC + Amendment 93/68/EEC &
EMC 89/336 EEC
Page 8
8 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
Type Flangeless sensor (Sandwich design)
Nominal size
mm
DN 6, 10, 15, 25, 40, 50, 65, 80, 100 DN 10, 15, 25, 40, 50, 65, 80, 100 DN 6, 10, 15, 25, 40, 50, 65, 80, 100
Operating pressure DN 6-65: 40 bar, DN 80: 37.5 bar, DN 6-65: 40 bar, DN 80: 37.5 bar,
DN 100: 30 bar 20 bar DN 100: 30 bar
Vacuum: 1 × 10-6 bar Vacuum: 0.02 bar Vacuum: 1 × 10-6 bar
Temperature of medium
PFA
30°C to +130°C
Ceramic
20°C to +150°C 20°C to +120°C
High temp.
20°C to +200°C Suitable for steam sterilization at 150°C
Temperature shock (Duration > 1 min.): Max. ±100°C momentarily (Duration > 1 min.):
(Ceramic liner) DN 6, 10, 15, 25: DN 6, 10, 15, 25:
Max. T 15°C/min. Max. T 15°C/min.
DN 40, 50, 65: Max. ∆T ≤ 10°C/min. DN 40, 50, 65: Max. ∆T ≤ 10°C/min.
DN 80, 100: Max. ∆T ≤ 5°C/min. DN 80, 100: Max. ∆T ≤ 5°C/min.
(Duration 1 min., (Duration 1 min., followed by 10 min. rest): followed by 10 min. rest):
DN 6, 10, 15, 25: Max. ∆T ≤ 80°C DN 6, 10, 15, 25: Max. ∆T ≤ 80°C
DN 40, 50, 65: Max. ∆T ≤ 70°C DN 40, 50, 65: Max. ∆T ≤ 70°C
DN 80, 100: Max. ∆T ≤ 60°C DN 80, 100: Max. ∆T ≤ 60°C
Ambient temperature Remote signal converter: −40°C to +100°C
Compact signal converter: −20°C to +50°C
Liner Aluminium oxide Al2O3 (ceramics) Reinforced PFA (Teflon) Aluminium oxide Al2O3 (ceramics)
Electrodes Platinum with gold/titanium brazing Hastelloy C-276 Platinum with gold/titanium brazing
alloy alloy
Enclosure Stainless steel AISI 316 L (1.4404) Stainless steel AISI 316 (1.4404) Stainless steel AISI 316 L (1.4404)
Terminal box
Standard
Fibre glass-reinforced polyamide Fibre glass-reinforced polyamide Stainless steel AISI 316 (1.4436)
(not compact) High temp.
Stainless steel AISI 316 (1.4436) Stainless steel AISI 316 L (1.4436)
Fixing studs Stainless steel AISI 304 (1.4301) Stainless steel AISI 304 (1.4301)
Number and size to EN 1092-1:2001 Number and size to EN 1092-1:2001
Mating flanges EN 1092-1:2001, ANSI B16.5
class 150 and 300 or equivalent
EN 1092-1:2001, ANSI B16.5 class 150 and 300 or equivalent
Option
DN 6/10: ½" pipe connection adapters thread: ½" tappered ISO 7-1
Gaskets
Standard
EPDM (max. 150°C, PN 40) EPDM (max. 150°C, PN 40)
Option
Graphite (max. 200°C, PN 40) Graphite (max. 200°C, PN 40)
Option
PTFE (max. 130°C, PN 25) PTFE (max. 130°C, PN 25)
Cable entries 4 Pg 13.5
Enclosure rating
Standard
IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.)
Option
IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.)
Mechanical load (vibration) 18-1000 Hz random, 3.17 G rms in all directions to EN 60068-2-36 18-1000 Hz random in all directions
to EN 60068-2-36
Sensor: 3.17 G/Compact Ex-d: 1.14 G
Test pressure 80 bar (2 × PN) 40 bar (2 × PN) 80 bar (2 × PN)
Approvals 3A EEx ia/ib IIB T4-T6/
DEMKO, No. 97D.121909X EEx de [ia/ib] IIB T4-T6/ DEMKO 94C.115327X
Excitation frequency DN 6-65: 12.5 Hz DN 10-65: 12.5 Hz DN 6-65: 6.25 Hz
DN 80-100: 6.25 Hz DN 80-100: 6.25 Hz DN 80-100: 3.125 Hz
Conforms to PED, LVT, EMC PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX
2. Technical data
2.1 Sensor MAG 1100 and MAG 1100 Ex
MAG 1100
MAG 1100 PFA
MAG 1100 Ex & Ex-d
MAG 1100 and MAG 1100 Ex
Page 9
9
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2. Technical data
Technical data
Adapter Stainless steel AISI 316 Pressure Pipe connection/ Adapter for direct welding into dairy pipe:
Operating
Tri-Clover
ISO 2037, DIN 11850, SMS 3008, BS 4825-1
pressure DN 10, 15, 25, 40, 50, 65, 80 PN 40
DN 100 PN 25
Clamp adapter:
Tri-Clamp
ISO 2852, DIN 32676, SMS 3016, BS 4825-3
DN 10, 15, 25, 40, 50 PN 16 DN 65, 80, 100 PN 10
Thread adapter:
DIN 11851: DN 10, 15, 25, 40 PN 40
DN 50, 65, 80, 100 PN 25
ISO 2853, SS 3351, BS 4825-4: DN 10, 15, 25, 40, 50, 65, 80 PN 16 SMS 1145: DN 25, 40, 50, 65, 80 PN 6
Gasket
Standard
EPDM (−20 °C to 150 °C)
Option
NBR (−20 °C to 100 °C)
Clamp Stainless steel AISI 304, ISO 2852
Type Hygienic sensor Nominal size
mm
DN 10, 15, 25, 40, 50, 65, 80, 100
Process connection Hygienic adapters available for:
Direct welding into dairy pipe Clamp fitting Threaded fitting
Operating pressure DN 10-65: 40 bar, DN 80: 37.5 bar, DN 100: 30 bar 20 bar
Vacuum
1 × 10-6 bar 0.02 bar
Temperature of medium −20°C to +150°C 30°C to +130°C
Suitable for steam sterilization Suitable for steam sterilization at 150°C
Temperature shock (Duration > 1 min.): Max. ±100°C momentarily
DN 10, 15, 25 Max. ∆T ≤ 15°C/min. DN 40, 50, 65 Max. ∆T ≤ 10°C/min. DN 80, 100 Max. ∆T ≤ 5°C/min. (Duration 1 min., followed by 10 min. rest):
DN 10, 15, 25 Max. ∆T ≤ 80°C DN 40, 50, 65 Max. ∆T ≤ 70°C DN 80, 100 Max. ∆T ≤ 60°C
Ambient temperature Remote signal converter: −40°C to +100°C Remote signal converter: −40°C to +100°C
Compact signal converter: −20°C to +50°C Compact signal converter: −20°C to +50°C
Liner Aluminium oxide Al2O3 (ceramic) Reinforced PFA (Teflon) Electrodes Platinum with gold/titanium brazing alloy Hastelloy C-276 Enclosure Stainless steel AISI 316 L (1.4404) Stainless steel AISI 316 L (1.4404) Terminal box
Standard
Fibre glass-reinforced polyamide Fibre glass-reinforced polyamide
(not compact) Option
Stainless steel AISI 316 (1.4436) Stainless steel AISI 316 (1.4436)
Cable entries 4 Pg 13.5 4 Pg 13.5 Enclosure rating
Standard
IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.) IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.)
Option
IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.) IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.)
Mechanical load (vibration) 18-1000 Hz random, 3.17 G rms in all directions 18-1000 Hz random, 3.17 G rms in all directions
to EN 60068-2-36 to EN 60068-2-36
Test pressure 80 bar (2 × PN) 40 bar (2 × PN) Approvals 3A, EHEDG 3A Excitation frequency DN 10-65: 12.5 Hz DN 10-65: 12.5 Hz
DN 80-100: 6.25 Hz DN 80-100: 6.25 Hz
Conforms to PED, LVT, EMC PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX
2.2 Sensor MAG 1100 FOOD
MAG 1100 FOOD MAG 1100 FOOD PFA
Accessories MAG 1100 FOOD
Note
When combined sensor and adapter, the working pressure is the lower rated of the pair.
MAG 1100 FOOD
Page 10
10 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
Type Sensor with flanges Sensor with flanges Sensor with flanges Nominal size
mm
DN 15-2000 DN 15-2000 / 15-300 DN 25-1200
Temperature of medium Temperature classification
Liner:
T3 + T4 T5 T6 Neoprene (standard) 0 to 70°C 0 to 70°C 0 to 70°C 0 to 70°C 0 to 70°C EPDM1) 10 to 95°C 10 to 95°C 10 to 90°C 10-75°C 10 to 95°C Linatex rubber 40 to 70°C2) 20 to 70°C 20 to 70°C 20 to 70°C Ebonite1) 0 to 95°C 0 to 95°C 0 to 90°C 0 to 75°C PTFE 20 to 100°C −20 to 100°C −20 to 90°C −20 to 75°C PTFE high temperature 20 to 180°C
Ambient temperature
Remote signal converter −40°C to 100°C −20°C to 50°C −40°C to 100°C Compact signal converter −20°C to 50°C −20°C to 50°C −20°C to 50°C
Operating pressure3)
[[
[[
[abs.bar
]]
]]
]
Liner:
Neoprene 0.01 to 100 bar 0.01 to 100 bar 0.01 to 40 bar EPDM 0.01 to 40 bar 0.01 to 40 bar 0.01 to 40 bar Linatex
0.01 to 40 bar 0.01 to 40 bar
Ebonite 0.01 to 100 bar 0.01 to 100 bar
PTFE teflon:
DN 15 to 600 Max. 100°C: 0.3 to 50 bar 0.3 to 40 bar DN 15 to 300 Max. 180°C: 0.6 to 50 bar
Excitation frequency DN 15 - 65: 12.5 Hz DN 15 - 65: 6.25 Hz All sizes 3.125 Hz
DN 80 - 150: 6.25 Hz DN 80/100: 3.125 Hz DN 200 - 1200: 3.125 Hz DN 125 - 300: 1.5625 Hz DN 1400 - 2000: 1.5625 Hz DN 350 - 1200: 3.125 Hz
Enclosure rating
Standard
IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.)
Option
IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.)
Cable entries 4 Pg 13.5 Mechanical load 18-1000 Hz random, 3.17 G rms in all directions to EN 60068-2-36 Test pressure 1.5 × PN Conforms to PED, LVT, EMC PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX
2.3 Sensor MAG 3100, MAG 3100 Ex and MAG 3100 W
MAG 3100 MAG 3100 Ex / Ex-d MAG 3100 W
1
) With WRC (Water Research Council, UK) approval
2
) For temperature below −20°C AISI 304 or 316 flanges must be used
3
) Maximum operating pressure decreases with increasing operating temperature and with stainless steel flanges
MAG 3100, MAG 3100 Ex and MAG 3100 W
Page 11
11
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2. Technical data
Technical data
Flanges
Standard
DN 15-50: PN 40 DN 25-50: PN 40 EN 1092-1:20011) DN 65-150: PN 16 DN 65-150: PN 16 Rased face DN 200-1000: PN 10 DN 200-1200: PN 10
DN 1100 -2000: PN 6
Option
DN 65-1000: PN 6 DN 200-600: PN 16
DN 1200-2000: PN 10
DN 200-2000: PN 16
DN 200-600: PN 25
DN 65-600: PN 40
DN 50-400 PN 63
DN 25-350 PN 100 ANSI B 16.5 3/4"-24": Class 150 (20 bar) 3/4"-24": Class 150 (20 bar) (BS 1560) 3/4"-24": Class 300 (50 bar) AS 2129 3/4"-48": Table D/E AS 4087 Class 14 (DN 50 - 1200, 14 bar)
Class 21 (DN 50 - 600, 21 bar)
Class 35 (DN 50 - 600, 35 bar) AWWA C-207 28"-78": Class D (10 bar) 28"-48": Class D (10 bar)
Electrodes
Standard
AISI 316 Ti (1.4571) AISI 316 Ti (1.4571)
Option
Hastelloy C-276, Platinum / Iridium, Titanium,
AISI 316 Ti Ceramic Coated, Tantalum
PE - electrodes
Standard
As measuring electrodes (except PTFE) AISI 316 Ti (1.4571)
Measuring pipe
Standard
AISI 304 (1.4301) AISI 304 (1.4301)
Option
AISI 316 L (1.4404)
Flange and
Standard
Carbon steel Carbon steel
housing material Corrosion-resistant two-component coating (min. 150 µm) Corrosion-resistant two-compo-
nent coating (min. 150 µm)
Option
AISI 304 (1.4301) flanges and carbon steel housing. Coating as above
Option
AISI 316 L (1.4404) flanges and housing
Ex-approval
Remote
DN 15-25 EEx ia/ib IIB T4-T6
DN 40-300 EEx ia/ib IIB T4-T6
DN 350-2000 EEx e ia IIC T3-T6
Compact
DN 15-300 EEx de [ia/ib] IIB T4-T6
MAG 3000 Ex-d
Conforms to PED, LVT, EMC PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX
2.3 Sensor MAG 3100, MAG 3100 Ex and MAG 3100 W
(continued)
MAG 3100 MAG 3100 Ex / Ex-d MAG 3100 W
MAG 3100, MAG 3100 Ex and MAG 3100 W
1
) EN 1092-1, DIN 2501 & BS 4504 have the same mating dimensions
Page 12
12 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
Type Sensor with flanges Design Straight Coned 1 DN reduction Straight Nominal size
mm
25-40 50-300 350-1200
Liner Hard elastomer Composite elastomer Hard elastomer
(hard rubber) (hard & soft rubber) (hard rubber)
Liner approvals WRc WRc WRc Medium temperature 5 to 90°C Ambient temperature
Remote signal converter 40 to 90°C Compact signal converter 20 to 50°C
Operating pressure 0.01 to 40 bar 0.03 to 20 bar 0.01 to 16 bar Excitation frequency 12.5 Hz 50-65 mm: 12.5 Hz 3.125 Hz
80-150 mm: 6.25 Hz
200-300 mm: 3.125 Hz
Enclosure rating
Standard
IP 67 to EN 60529 1 m w.g. for 30 minutes
Option
IP 68 to EN 60529 10 m w.g. continuously
Cable entries 4 Pg 13.5 Mechanical load 18-1000 Hz random, 3.17 G rms in all directions to EN 60068-2-36 Test pressure 1.5 × nominal pressure Flanges
EN 1092-1
Standard
PN 40 50-150 mm: PN 16 PN 10
200-300 mm: PN 10
Option
200-300 mm: PN 16 PN 16
ANSI B16.5
Standard
Class 150 lb Class 150 lb 14"-24": Class 150 lb
AWWA C-207 Standard 28"-48": Class D
Pressure drop at 3 m/sec. As straight pipe Max. 25 mbar As straight pipe Electrodes AISI 316 Ti (1.4571) PE/grounding electrodes
Standard
AISI 316 Ti (1.4571)
Measuring pipe/meter body AISI 304 (1.4301) Composite elastomer AISI 304 (1.4301) Flanges Carbon steel Housing Carbon steel Surface finish Two component epoxy Polyester powder coat Two component epoxy
min. 150 microns min. 100 microns min. 150 microns
Colour RAL 7035 pale grey Approvals
Conforms to
PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX1)
2.4 Sensor MAG 5100 W
1
) For sizes greater than 600 mm PED conformity is available as a cost added option, the basic unit will only carry the LVD (Low Voltage Directive)
and EMC approval.
MAG 5100 W
Page 13
13
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2. Technical data
Technical data
Current output
Current 0-20 mA, 4-20 mA or 4-20 mA + alarm Load < 800 ohm Time constant 0.1-30 s adjustable
Digital output
Frequency 0-10 kHz, 50% duty cycle Time constant 0.1-30 s adjustable
Active 24 V d.c., 30 mA, 1 K ≤ R
load
10 K, short-circuit-protected
Passive 3-30 V d.c., max. 110 mA, 200 Ω ≤ R
load
10 K
Relay Time constant Changeover relay, time constant same as current time constant
Load 42 V a.c./2 A, 24 V d.c./1A
Digital input 11-30 V d.c., Ri = 4.4 K
Activation time 50 ms Current I
11 V d.c.
= 2.5 mA, I
30 V d.c.
= 7 mA
Functions Flow rate, 2 totalizers, low flow cut-off, empty pipe cut-off, flow direction, error system, operating time,
uni/bidirectional flow, limit switches, pulse output, control for cleaning unit
Galvanic isolation All inputs and outputs are galvanically isolated Cut-off Low flow 0-9.9% of maximum flow
Empty pipe Detection of empty pipe1)
Totalizer Two eight-digit counters for forward, net or reverse flow Display Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults
Reverse flow indicated by negative sign
Time constant Time constant as current output time constant
Zero point adjustment Automatic Electrode input impedance > 1 x 1014 Excitation frequency Sensor size depending pulsating d.c. current (125 mA) Ambient temperature Display version during operation:20 to +50°C
Blind version during operation:20 to +60°C
During storage:40 to +70°C (RH max. 95%)
Custody transfer approval PTB MAG 5000 CT (cold water)
Communication
Standard
Without serial communication
Optional
HART
Compact
Enclosure material Fibre glass-reinforced polyamide Enclosure rating IP 67 to EN 60529 and DIN 40050 (1 m w.g. for 30 minutes) Mechanicalload 18-1000 Hz random, 3.17 G rms in all directions to EN 60068-2-36
19" insert
Enclosure material Standard 19" insert of aluminium/steel (DIN 41494)
Width: 21 TE
Height: 3 HE
Enclosure rating IP 20 to EN 60529 and DIN 40050 Mechanical load Version: 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance Emission: EN 50081-1 (Light industry)
Immunity: EN 50082-2 (Industry)
Supply voltage 115-230 V a.c. +10% to 15%, 50-60 Hz
11-30 V d.c. or 11-24 V a.c.
Power consumption 230 V a.c.: 9 VA
24 V d.c.: 9 W, IN = 380 mA, IST = 8 A (30 ms)
12 V d.c.: 11 W, IN = 920 mA, IST = 4 A (250 ms)
2.5.1 Signal converter MAG 5000 (DN 6 to DN 1200)
Accuracy 0.5%
MAG 5000 and MAG 5000 CT
1
) Special cable required in separate mounted installation
6.221
99.19
Page 14
14 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
Current output
Current 0-20 mA, 4-20 mA or 4-20 mA + alarm Load < 800 ohm Time constant 0.1-30 s adjustable
Digital output
Frequency 0-10 kHz, 50% duty cycle Time constant 0.1-30 s adjustable
Active 24 V d.c., 30 mA, 1 K ≤ R
load
10 K, short-circuit-protected
Passive 3-30 V d.c., max. 110 mA, 200 Ω ≤ R
load
10 K
Relay Time constant Changeover relay, time constant same as current time constant
Load 42 V a.c./2 A, 24 V d.c./1A
Digital input 11-30 V d.c., Ri = 4.4 K
Activation time 50 ms Current I
11 V d.c.
= 2.5 mA, I
30 V d.c.
= 7 mA
Functions Flow rate, 2 totalizers, low flow cut-off, empty pipe cut-off, flow direction, error system, operating time,
uni/bidirectional flow, limit switches, pulse output, control for cleaning unit and batch
Galvanic isolation All inputs and outputs are galvanically isolated Cut-off Low flow 0-9.9% of maximum flow
Empty pipe Detection of empty pipe1)
Totalizer Two eight-digit counters for forward, net or reverse flow Display Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults Reverse flow indicated by negative sign
Time constant Time constant as current output time constant
Zero point adjustment Automatic Electrode input impedance > 1 x 1014 Excitation frequency Sensor size depending pulsating d.c. current (125 mA) Ambient temperature Display version during operation:20 to +50°C
Blind version during operation:20 to +60°C
During storage:40 to +70°C (RH max. 95%) Custody transfer approval PTB DANAK OIML R75 DANAK OIML R117 MAG 6000 CT (cold water) (hot water) (cold water/milk, beer etc.)
Communication
Standard
Prepared for client mounted add-on modules
Optional
HART, Profibus PA, Profibus DP, CANopen, DeviceNet as add-on module
Compact
Enclosure material Fibre glass-reinforced polyamide Enclosure rating IP 67 to EN 60529 and DIN 40050 (1 m w.g. for 30 minutes) Mechanicalload 18-1000 Hz random, 3.17 G rms in all directions to EN 60068-2-36
19" insert
Enclosure material Standard 19" insert of aluminium/steel (DIN 41494)
Width: 21 TE
Height: 3 HE
Enclosure rating IP 20 to EN 60529 and DIN 40050 Mechanical load Version: 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance Emission: EN 50081-1 (Light industry)
Immunity: EN 50082-2 (Industry) Supply voltage 115-230 V a.c. +10% to 15%, 50-60 Hz
11-30 V d.c. or 11-24 V a.c. Power consumption 230 V a.c.: 9 VA
24 V d.c.: 9 W, IN = 380 mA, IST = 8A (30 ms)
12 V d.c.: 11 W, IN = 920 mA, IST = 4A (250 ms)
2.5.2 Signal converter MAG 6000
Accuracy 0.25%
MAG 6000 and MAG 6000 CT
6.221
99.19
1
) Special cable required in separate mounted installation
Page 15
15
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2. Technical data
Technical data
Safety barrier & cleaning unit
Application As combined unit with MAG 6000 only and MAG 1100 Ex/3100 Ex in the size
range DN 6-300
Ex approval [EEx ia/ib] IIB Cable parameter Group Capacity in µF Inductance in mH
Electrode cable IIB ≤ 31 ≤ 80 Coil cable IIB 0.5 8
Ambient temperature During operation: 20 to +50°C
During storage: 20 to +70°C
19" insert
Enclosure material Standard 19" insert in aluminium/steel (DIN 41494)
Width: 21 TE
Height: 3 HE Enclosure rating IP 20 to EN 60529 and DIN 40050 Mechanical load 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance
Emission EN 50081-1 (Light industry)
Immunity EN 50082-2 (Industry)
2.5.3 Safety barrier (ia/ib) DN
≤≤
≤≤
300
Application For use with MAG 5000/6000 19 insert and MAG 3100 Ex in the size
range DN 350-2000
Ex approval [EEx ia] IIC Cable parameter Group Capacity in µF Inductance in mH
IIC 4.1 1.5
IIB 45 87
IIA 45 87
Ambient temperature During operation: 20 to +50°C
During storage: 20 to +70°C
19" insert
Enclosure material Standard 19" insert in aluminium/steel (DIN 41494)
Width: 21 TE
Height: 3 HE Enclosure rating IP 20 to EN 60529 and DIN 40050 Mechanical load 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance
Emission EN 50081-1 (Light industry)
Immunity EN 50082-2 (Industry)
2.5.4 Safety barrier (ia) DN
≥≥
≥≥
350
Application For use together with MAG 5000 and 6000 19" insert to clean the electrodes
on MAG 1100, MAG 3100 or MAG 5100 W.
NB Must not be used with intrinsically safe systems
Cleaning voltage (unloaded)
a.c. cleaning 60 V a.c. d.c. cleaning 30 V d.c.
Cleaning period 60 sec. + 60 sec. pause period Relay Switch relay activated when cleaning is in progress
Load 42 V/2 A
Operation
Automatic Yes Manual No
Indicator lamps LEDs: "ON" and "CLEANING" Supply voltage and 115-230 V a.c. +10% to 15%, 50-60 Hz, 7 VA cleaning, 5 VA stand by power consumption Ambient temperature During operation: 20 to +50°C
During storage: 20 to +70°C
19" insert
Enclosure material Standard 19" insert in aluminium/steel (DIN 41494)
Width: 21 TE
Height: 3 HE Enclosure rating IP 20 to EN 60529 and DIN 40050 Mechanical load 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
2.5.5 Cleaning unit
Page 16
16 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
MAG 5000 or MAG 6000 used with MAG 3100 W or MAG 1100 PFA2.6
Meter uncertainty
V: Actual flow velocity [m/s] E: Meter uncertainty as a percentage of actual flow
MAG 6000 used with MAG 3100, MAG 1100 Ceramic or MAG 5100 W
V: Actual flow velocity [m/s] E: Meter uncertainty as a percentage of actual flow
Reference conditions (ISO 9104 and DIN/EN 29104)
Temperature of medium 20°C ±5 K Ambient temperature 20°C ±5 K Supply voltage Un ±1% Warming-up time 30 minutes Incorporation in pipe section Inlet section 10×DN (DN 1200), 5×DN (DN > 1200)
Outlet section 5×DN (DN 1200), 3×DN (DN > 1200)
Flow conditions Fully developed flow profile
Additions in the event of deviations from reference conditions
Current output As pulse output ±(0.1% of actual flow +0.05% FSO) Effect of ambient temperature Display/frequency/pulse output: < ±0.003% / K act.
Current output: < ±0.005% / K act. Effect of supply voltage < 0.005% of measuring value on 1% change Repeatability ±0.1% of actual flow for V 0.5 m/s
Page 17
17
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
2. Technical data
Technical data
2.7 Output characteristics MAG 5000 and MAG 6000
Output characteristics Bidirectional mode Unidirectional mode 0-20 mA
4-20 mA
Frequency
Pulse output
Relay
Power down Active
Error relay No error Error
Limit switch or 1 set point 2 set points direction switch
Low flow Intermediate flow (Reverse flow) High flow High flow/ (Forward flow) Low flow
Batch on digital output
Batch on relay Hold Batch
Page 18
18 DKFD.PS.027.W3.02
2. Technical data
Technical data
SITRANS F M MAGFLO
Coil cable Electrode cable
Basic data No. of conductors 2 3
Min. sqr. area 0.5 mm
2
0.2 mm
2
Screen Yes Yes Max. capacitance N.A. 350 pF/m
Max. cable loop Media temperature: < 100°C 40 N.A. resistance < 200°C6 N.A.
2.8.1 Sensor cables and conductivity of medium
Conductivity of Compact installation: medium Liquids with an electrical conductivity 5 µS/cm.
For a conductivity between 5 and 10 µS/cm, the repeatability may degrade to ±0.5% of actual flow.
Remote installation:
Standard cable
Note
For detection of empty sensor the min. conductivity must always be
20 µS/cm and the max. length
of electrode cable when remote mounted is 50 metres. Special cable must be used. For remote mounting in Ex applications special cable cannot be used, empty sensor cannot be detected and the electrically conductivity must be
30 µS/cm.
For remote mounted CT installations the max. cable length is 200 metres.
2.8.2 Minimum accept data for cable
2.9 HART


communication
add-on module
Application MAG 6000, MAG 6000 CT
Optional available as factory mounted in MAG 5000
Communication standard Bell 202 frequency shift keying (f.s.k.) standard Communication modes Single loop mode
Multi-drop mode, 15 slave devices
Communicator Rosemount Hand-held communicator type 275
Cable specification
Communication mode /
Single loop
Q [mm2] CU 0.2 mm2/AWG 24 Screen Yes (Overall screen) Loop resistance
Min.
230
Max.
800
Cable capacity 400 pF/m Cable length 1500 m Twisted pair Yes
HART is a registered trademark of the HART Communication Foundation.
2.10 Cable data (Supplied by Siemens Flow Instruments)
Standard cable Special cable (electrode/coil) (electrode)
Basic data No. of conductors 3 3
Sqr. area 1.5 mm
2
0.25 mm
2
Screen Yes Double Color code Brown, blue, black Brown, blue, black Outside color Grey Grey Ext. diameter 7.8 mm 8.1 mm Conductor Flexible CU Flexible CU Isolation material PVC PVC
Amb. temperature Flexible installation 5 to 70°C 5 to 70°C
Non flexible installation 30 to 70°C 30 to 70°C
Cable parameter Capacity 161.50 pF/m N.A.
Inductance 0.583 µH/m N.A. L/R 43.83 µH/ N.A.
Special cable
Page 19
19
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
3. Project guidance
3.1 Sizing table (DN 6 to DN 2000)
The table shows the relationship between flow velocity V, flow quantity Q and sensor dimension DN.
Guidelines for selection of sensor
Min. measuring range: 0-0.25 m/s Max. measuring range: 0-10 m/s Normally the sensor is selected so that V lies within the measuring range 1-2 m/s.
Flow velocity calculation formula:
V =
1273.24 x Q [l/s] [m/s]
DN
2
[mm]
or V =
[m/s]
DN2 [mm]
353.68 x Q [m
3
/h]
Page 20
20
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
3.2.1 Minimum conductivity
Applications Min. conductivity Compact mounted 5 µS/cm Remote mounted 5 µS/cm With empty pipe detection 20 µS/cm Ex-installations
(Remote mounted only)
30 µS/cm
District heating systems
(Without DC cleaning unit)
250 µS/cm
3.2.2 Liner selection guide
Liner Applications Ceramics Al2O
3
General purpose, agressive chemicals
PFA General purpose, dairy, food and beverage Neoprene General purpose, sewage EPDM Drinking water, sea water PTFE Agressive chemicals, paper and pulp, high temperature applica-
tions
Linatex


Abrasive media and mining slurries
Ebonite Drinking water
3.2.3 Electrode selection guide
Electrodes Applications AISI 316 Ti General purpose, water, sewage and district heating AISI 316 Ti Ceramic coated High content of fibres, paper pulp Hastelloy C-276 Good chemical proporties, sea water Titanium Chlorine, chlorite, nitric and chromic acids
Textile bleaching industry
Tantalum Almost any acid solution Platinum and platinum/irridium The ultimate electrode material. Unaffected by most liquids
3.3 Installation conditions
Reading and operating the flowmeter is possi­ble under almost any installation conditions because the display can be oriented in relation to the sensor. To ensure optimum flow mea­surement, attention should be paid to the fol­lowing:
Page 21
21
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
The sensor must be mounted as shown in the upper figure. Do not mount the sensor as shown in the lower figure. This will position the electrodes at the top where there is possibility for air bubbles and at the bottom where there is possibility for mud, sludge, sand etc. If using empty pipe detection, the sensor can be tilted 45°, as shown in the upper figure.
Recommended installation is in a vertical/in­clined pipe to minimize the wear and deposits in the sensor.
Installation in horizontal pipes
Measuring abrasive liquids and liquids containing particles
The sensor must always be completely full with liquid.
Therefore avoid:
Installation at the highest point in the pipe system
Installation in vertical pipes with free outlet
For partially filled pipes or pipes with downward flow and free outlet the flowmeter should be located in a U-tube.
Recommended flow direction: upwards. This minimizes the effect on the measurement of any gas/air bubbles in the liquid.
Installation in vertical pipes
3.3 Installation conditions
(continued)
Page 22
22
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
To achieve accurate flow measurement it is es­sential to have straight lengths of inlet and outlet pipes and a certain distance between pumps and valves. It is also important to centre the flowmeter in relation to pipe flanges and gaskets.
The electrical potential of the liquid must always be equal to the electrical potential of the sensor. This can be achieved in different ways depending on the application: A. Wire jumper between sensor and adjacent
flanges. (MAG 1100 and MAG 3100).
B. Direct metallic contact between sensor and
fittings. (MAG 1100 FOOD).
C. Built-in earthing electrodes. (MAG 3100 and
MAG 3100 W).
D. Optional earthing/protection flanges/rings.
(MAG 1100 and MAG 3100).
E. Optional graphite gaskets on MAG 1100.
(Standard for MAG 1100 High temperature).
Inlet and outlet condi­tions
Potential equalization
Avoid a vacuum in the measuring pipe, since this can damage certain liners. See "Technical data", chapter 2.
The flowmeter can be installed between two reducers (e.g. DIN 28545). Assuming that at 8° the following pressure drop curve applies. The curves are applicable to water.
Example:
A flow velocity of 3 m/s (V) in a sensor with a diameter reduction from DN 100 to DN 80 (d
1/d2
= 0.8) gives a pressure drop of 2.9 mbar.
Vacuum
Installation in large pipes
Page 23
23
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
The sensor and signal converter can be installed either compact or remote.
With compact installation the temperature of medium must be according to the graph.
With remote installation, the cable length and type described under "Technical data", chapter 2 must be used.
Compact/remote installa­tion
IP 68 applications only
If the sensor is buried or permanently sub­merged, the terminal box must be encapsu­lated with silicon dielectric gel.
Mix the two components well and pour the contents into the terminal box. The material is a non-toxic, transparent, self­healing gel which cures in approx. 24 hours. The gel can be penetrated with test instru­ments or be removed in case of cable replace­ment.
Horizontal installation
Vertical installation
Suggestions for the direct burial of MAG 3100 & MAG 5100 W sensors
If MAG 3100 or MAG 5100 W sensors are buried directly into the ground we would suggest the following precautions are observed.
The SENSORPROM
unit should be removed from the terminal box on the sensor and relocated in the signal converter remote mounting prior to burying the sensor. (See product manual LS.27.V2.02).
All the sensor data plate information and serial number should be recorded for each sensor prior to burying. This will ensure correct matching with the SENSORPROM unit.
The sensor should be potted to IP 68 and suitable coil and electrode cables used prior to burying.
The use of pea gravel at least 300 mm all around the sensor. This provides some drainage and also avoids caking the sensor with earth. It also helps located the sensor should excavation take place.
Before covering the pea gravel with earth we suggest the use of electrical cable identification tape laid above the gravel.
The sensor should not be subject to heavy vehicles applying excessive weight above the sensor or pipeline.
Page 24
24
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
The Siemens Flow Instruments cleaning unit can be used with MAG 5000 or 6000 in 19" insert version. The cleaning unit can be used in applications where the liner material and subsequently the electrodes may be coated with deposits. If the coating is electrically insulating, the electrode signal will be reduced. If the coating is electrically conductive, the electrode signal will be partly short­circuited and in both cases the accuracy of the meter will decrease (dependent on the type and thickness of the coating).
Note
The cleaning unit cannot be used for inflammable or explosive media! Empty pipe detection and cleaning facility cannot be used at the same time.
The cleaning unit cleans the electrodes electro-chemically by applying a voltage to the electrodes for approx. 60 sec. While cleaning, the signal converter stores and holds the latest measured flow reading on the display and also the signal outputs. After an additional pausing period of 60 sec. the flowmeter resumes normal measurement and the cleaning is now completed.
The relay in the signal converter activates the cleaning cycle. In the relay output menu (under cleaning) the cleaning interval can be set between 1 hour and 24 hours.
Cleaning should only take place with liquid in the pipe. This can be detected via the empty pipe function. It is therefore recommended to select “empty pipe detection” ON when using the cleaning unit.
The cleaning sequence can also be controlled manually through the electrical input of the signal converter. Before this is done, ensure that the measuring pipe is full.
Mode of operation
3.4 Cleaning unit
AC-cleaning
DC-cleaning
AC-cleaning is used to remove fatty deposits on the electrodes. These fatty deposits are seen in waste water applications, in abbatoirs and water applications with oil residuals. During the cleaning process, the surface of the electrodes get warmer, which tends to soften grease particles and the gas bubbles generated mechanically lift deposits away from the surface of the electrodes.
DC-cleaning is used to eliminate electrically conductive deposits in the measuring pipe influencing the measuring accuracy. Particularly in district heating applications an electrically conductive deposit (magnetite) may occur and short-circuit the electrode signal. In this case the accuracy of the meter decreases and the signal/ noise conditions of the meter become inferior. The problem only arises if the conductivity of the water is less than approx. 250 µS/cm.
During DC-cleaning, electrolysis takes place where the flow of electrons removes the particle deposits from the electrode area.
Note Do not use DC-cleaning on sensors with tantalum electrodes.
Page 25
25
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
A signal converter can be supplied in a version tested and approved for custody transfer (CT ). The internal counter can accordingly be used for charging. This requires verification, sealing and setting of the signal converter together with the sensor for a specific flow range. After sealing the data on the signal converter must not be changed.
The sealing of the signal converter is done by placing sealing marks on the signal converter and on the connection plate in the terminal box.
3.5 Custody transfer ap­proval
3.6 Signal converter MAG 5000 CT, MAG 6000 CT
Sealing
The final sealing should be carried out as shown:
MAG 6000 CT is installed like a standard MAG 6000 except for the final sealing. Calibration sealing has been carried out at calibration.
Page 26
26
SITRANS F M MAGFLO
3. Project guidance
DKFD.PS.027.W3.02
Project guidance
Signal converters
The signal converter can be one of the following types:
MAG 5000/6000 19” with safety barrier (ia/ib) for separate mounting in safe area
Approval [EEx ia/ib] llB. The safety barrier is to be used with sensors MAG 1100 Ex and MAG 3100 Ex, DN 6 - 300. When this safety barrier is used, the coil circuit is intrinsic safety “ib” and the electrode circuit is intrinsic safety “ia”.
MAG 5000/6000 19” with safety barrier (ia) for separate mounting in safe area
Approval [EEx ia] llC. The safety barrier is to be used with sensors MAG 3100 Ex, DN 350 - 2000. When the safety barrier is used, the coil circuit is increased safety “e” and the electrode circuit is intrinsic safety “ia”.
Sensors
The sensors can be one of the following type.
MAG 1100 Ex for mounting in Ex areas
Approval EEx ia/ib llB T4...T6. DEMKO no. 97D.121909X. DN 6 - 100.
MAG 3100 Ex for mounting in Ex areas
The sensor carries the approval: DN 15 - 25 EEx ia/ib llB T4...T6, DEMKO no. 98E.123914X DN 40 - 300 EEx ia/ib llB T4...T6, DEMKO no. 98E.123915X DN 350 - 2000 EEx e ia llC T3...T6, SIRA no. Ex 92C3107X.
The electrode circuit in the sensors is manufactured to an intrinsically safe category “ia” and the coil circuit to an intrinsically safe category “ib”, achieved by an integrated and patented protection circuit. For DN 350 - 2000 the coil circuit is increased safety “e”.
Marking
The marking has the following meaning according to European Norm EN 50014.
E: Certified to CENELEC standard.
Ex: Designates explosion proof material and indicates that the apparatus has been approved
in accordance with a certificate issued.
i: “Intrinsic safety” is a protection ensuring that the energy in the electric circuit is too small to ignite
the explosive atmosphere. There are two categories of intrinsic safety: “ia” and “ib”.
ia: In intrinsic safety category “ia”, the circuit must remain safe, even in the event of two
simultaneous errors occurring that are independent of one another.
ib: In intrinsic safety category “ib” the circuit must remain safe if an error occurs.
d: The enclosure of the of the signal converter is so strong that it can resist an explosion in-
side the enclosure. The enclosure is dimensioned in a way so that an explosion will not effect the surroundings.
e: “Increased safety” is a constructional safeguard which ensures the apparatus does not
contain normally arcing or sparking devices, or hot surfaces that will cause ignition.
ll: Designates that the apparatus may be used in all areas (Except mining).
B: Indicates the gas group in which the unit may be used.
T4...T6 The temperature class describes the maximum temperature which any exposed surface of the equipment may reach. The sensor can have temperature class T3, T4, T5 or T6 depending on the temperature of the media. Please see technical data for the sensor.
T3: Max. surface temperature 200 °C. => (Max. media temperature 180 °C). T4: Max. surface temperature 135 °C. => (Max. media temperature 120 °C). T5: Max. surface temperature 100 °C. => (Max. media temperature 90 °C). T6: Max. surface temperature 85 °C. => (Max. media temperature 75 °C).
3.7 Ex installations
Page 27
27
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
MAG 1100, compact/separate4. Dimensions and
weight
4.1 Sensor MAG 1100
DN A1)B1)A1B
1
DDiD
i
D
p
DGWeight2)
(Al2O3) (PFA)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [kg]
6 156 181 309 334 48.3 6 17.3 34 2.2 10 156 181 309 334 48.3 10 10 13.6 34 2.2 15 156 181 309 334 48.3 15 16 17.3 40 2.2 25 164 196 317 349 63.4 25 26 28.5 56 2.7 40 176 218 329 371 84.0 40 38 43.4 75 3.4 50 184 235 337 388 101.6 50 50 54.5 90 4.2 65 194 254 347 407 120.0 65 66 68.0 112 5.5 80 200 266 353 419 133.0 80 81 82.5 124 7.0
100 213 292 366 445 159.0 100 100 107.1 150 10.0
1
) 13 mm shorter when the AISI terminal box is used. (Ex and high temperature 200°C).
2
) With signal converter MAG 5000 or MAG 6000 installed, weight is increased by approx. 0.8 kg.
The total built-in length "L" [mm] before assembling depends on the gasket selected.
DN EPDM Graphite PTFE(Teflon) Without gasket Earthing ring
6 64 66 70 64 77 10 64 66 70 64 77 15 65 66 70 64 77 25 80 81 85 79 92 40 95 96 100 94 107 50 105 106 110 104 117 65 130 131 135 129 142 80 155 156 160 154 167
100 185 186 190 184 197
The MAG 1100 DN 6 and DN 10 are prepared for assembly with the 1/2" pipe connection.
The length "L" varies dependent on the gasket choice:
Without gasket EPDM Graphite Teflon
L [mm] 150 150 152 156
Page 28
28
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
4.2 Sensor MAG 1100 FOOD
MAG 1100 FOOD, compact and separate
DN L A A
1
BB1DDiD
i
Weight 1)
(Al2O3) (PFA)
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[mm
]]
]]
]
[[
[[
[kg
]]
]]
]
10 64 156 309 188.0 341.0 64.0 10 10 2.2 15 64 156 309 188.0 341.0 64.0 15 16 2.2 25 79 164 317 202.8 355.8 77.5 25 26 2.7 40 94 176 329 221.5 374.5 91.0 40 38 3.4 50 104 184 337 243.5 396.5 119.0 50 50 4.2 65 131 194 347 259.0 412.0 130.0 65 66 5.5 80 156 200 353 277.5 430.5 155.0 80 81 7.0
100 186 213 366 304.5 457.5 183.0 100 100 10.0
1
) With signal converter MAG 5000 or MAG 6000 installed, weight is increased by approx. 0.8 kg.
Built-in length
DN A L 1)
[mm] [mm]
10 99 146 15 99 146 25 113 161 40 126 176 50 154 186 65 165 223 80 200 258
100 225 288
1
) The total built-in length "L" is independent of the
adapter type selected.
Page 29
29
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
Accessories MAG 1100 FOOD
Weld-in type
DIN 11850 DS/ISO 2037 SMS 3008 BS4825-1 Tri-Clover
®
Di Do Di Do Di Do Di Do Di Do
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
10 10 40 10.0 13.0 10.0 13.0 10.0 13.0 10.0 13.0 9.4 12.7 15 15 40 16.0 19.0 16.0 19.0 16.0 19.0 16.0 19.0 - -
15.9 15 40 --------15.75 19.05 20 15 40 20.0 23.0 20.0 23.0 20.0 23.0 20.0 23.0 - -
25 25 40 - - 22.6 25.6 22.6 25.6 22.6 25.6 22.1 25.4 25 25 40 26.0 29.0 - - - - - - - -
28 25 40 - - 25.6 28.6 - - - - - ­322540----29.6 32.0 - - - ­32 25 40 32.0 35.0 - - - - - - - -
33.7 25 40 - - 31.3 34.3 31.3 34.3 - - - ­38 40 40 - - 35.6 38.6 35.6 38.6 35.6 38.6 34.8 38.1 40 40 40 - - 37.6 40.6 - - - - - -
40 40 40 38.0 40.0 - - - - - - - ­50 50 40 - - 48.6 51.6 48.6 51.6 48.6 51.6 47.5 50.8 50 50 40 50.0 53.0 - - - - - - - -
63.5 65 45 - - 60.3 64.1 60.3 64.1 60.3 64.1 60.2 63.5 65 65 45 66.0 70.0 - - - - - - - ­70 65 45 - - 66.8 70.6 - - - - - -
766545----72.0 76.0 - - - -
76.1 80 50 - - 72.9 76.7 72.9 76.7 72.9 76.7 72.9 76.2 80 80 50 81.0 85.0 - - - - - - - -
88.9 80 50 - - 84.9 89.8 84.9 89.8 - - - ­100 100 50 100 104 - - - - - - - -
101.6 100 50 - - 97.6 102.5 97.6 102.5 97.6 102.6 97.38 101.6
114.3 100 50 - - 110.3 115.6 110.3 115.6 110.3 115.6 - -
Adapter
DN
Sensor
DN
Clamp type
DIN 32676 ISO 2852 SMS 3016 BS4825-3 Tri-Clamp
®
Di Do Di Do Di Do Di Do Di Do
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
10 10 40 10.0 34.0 10.0 34.0 10.0 34.0 - - 9.4 25.0 15 15 40 16.0 34.0 16.0 34.0 16.0 34.0 - - 15.75 25.0 20 15 40 20.0 34.0 20.0 34.0 - - - - - -
25 25 40 - - 22.6 50.5 22.6 50.5 22.6 50.5 22.1 50.5 25 25 40 26.0 50.5 26.0 50.5 - - - - - -
33.7 25 40 31.3 50.5 31.3 50.5 31.3 50.5 - - - ­38 40 40 - - 35.6 50.5 35.6 50.5 35.6 50.5 34.8 50.5
40 40 40 38.0 50.5 38.0 50.5 - - - - - ­50 50 40 50.0 64.0 - - - - - - - -
51 50 40 48.6 64.0 48.6 64.0 48.6 64.0 47.5 64.0
63.5 65 45 - - 60.3 77.5 60.3 77.5 60.3 77.5 60.2 77.5 65 65 45 66.0 91.0 - - - - - - - -
76.1 80 50 - - 72.9 91.0 72.9 91.0 72.9 91.0 72.9 91.0 80 80 50 81.0 106.0 - - - - - - - -
100 100 50 100 119.9 - - - - - - - -
101.6 100 50 - - 97.6 119.0 97.6 119.0 97.6 119.0 97.38 119.0
Adapter
DN
Sensor
DN
L
L
Tri-Clover and Tri-Clamp are registered trademarks for Ladish Co.
Page 30
30
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
Accessories MAG 1100 FOOD
(continued)
Threaded type
DIN 11851
Di Do
[mm] [mm] [mm] [mm] [mm]
10 10 40 10.0 28.0 15 15 40 16.0 34.0 20 15 40 20.0 44.0
25 25 40 26.0 52.0 32 25 40 32.0 58.0 40 40 40 38.0 65.0 50 50 40 50.0 78.0 65 65 45 66.0 95.0 80 80 50 81.0 110.0
100 100 50 100.0 130.0
Adapter
DN
Sensor
DN
Threaded type
ISO 2853 SS 3351 BS 4825-4 (IDF)
Di Do Di Do Di Do
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
25 25 40 22.6 37.0 22.6 37.0 22.6 37.0 38 40 40 35.6 51.0 35.6 51.0 35.6 51.0 51 50 40 48.6 64.0 48.6 64.0 48.6 64.0
63.5 65 45 60.3 78.0 60.3 78.0 60.3 78.0
76.1 80 50 72.9 91.0 72.9 91.0 72.9 91.0
101.6 100 50 - - - - 97.6 126.0
101.6 100 50 97.6 118.0 97.6 118.0 - -
Adapter
DN
Threaded type
SMS 1145
Di Do
[mm] [mm] [mm] [mm] [mm]
25 25 40 22.6 40.0 32 25 40 29.6 48.0 38 40 40 35.6 60.0
51 50 40 48.6 70.0
63.5 65 45 60.3 85.0 76 65 45 72.0 98.0
Adapter
DN
Sensor
DN
L
L
Sensor
DN
L
Page 31
31
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
mm inch mm inch mm inch mm inch mm inch mm inch mm inch 25 1" 187 7.4 N/A N/A N/A N/A 200 7.9 200 7.9 N/A N/A 40 1½” 197 7.8 N/A N/A N/A N/A 200 7.9 200 7.9 N/A N/A 50 2" 188 7.4 N/A N/A 200 7.9 N/A N/A 200 7.9 N/A N/A 65 2½” 194 7.6 N/A N/A 200 7.9 N/A N/A 200 7.9 N/A N/A 80 3" 200 7.9 N/A N/A 200 7.9 N/A N/A 200 7.9 N/A N/A 100 4" 207 8.1 N/A N/A 250 9.8 N/A N/A 250 9.8 N/A N/A 125 5" 217 8.5 N/A N/A 250 9.8 N/A N/A 250 9.8 N/A N/A 150 6" 232 9.1 N/A N/A 300 11.8 N/A N/A 300 11.8 N/A N/A 200 8" 257 10.1 350 13.8 350 13.8 N/A N/A 350 13.8 N/A N/A 250 10" 284 11.2 450 17.7 450 17.7 N/A N/A 450 17.7 N/A N/A 300 12" 310 12.2 500 19.7 500 19.7 N/A N/A 500 19.7 N/A N/A 350 14" 362 14.3 550 21.7 550 21.7 N/A N/A 550 21.7 N/A N/A 400 16" 387 15.2 600 23.6 600 23.6 N/A N/A 600 23.6 N/A N/A 450 18" 418 16.5 600 23.6 600 23.6 N/A N/A 600 23.6 N/A N/A 500 20" 443 17.4 625 24.6 625 24.6 N/A N/A 680 26.8 N/A N/A 600 24" 494 19.4 750 29.5 750 29.5 N/A N/A 820 32.3 N/A N/A 700 28" 544 21.4 875 34.4 875 34.4 N/A N/A N/A N/A 875 34.4 750 30" 571 22.5 N/A N/A N/A N/A N/A N/A N/A N/A 937 36.9 800 32" 606 23.9 1000 39.4 1000 39.4 N/A N/A N/A N/A 1000 39.4 900 36" 653 25.7 1125 44.3 1125 44.3 N/A N/A N/A N/A 1125 44.3 1000 40" 704 27.7 1250 49.2 1250 49.2 N/A N/A N/A N/A 1250 49.2 42" 704 27.7 N/A N/A N/A N/A N/A N/A N/A N/A 1250 49.2 1100 44" 755 29.7 N/A N/A N/A N/A N/A N/A N/A N/A 1375 54.1 1200 48" 810 31.9 1500 59.1 1500 59.1 N/A N/A N/A N/A 1500 59.1
4.3 Sensor MAG 5100 W
Dimensions
AWWA
Nominal
size
A
L
PN 10 PN 16 PN 40 Class 150
Page 32
32
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
mm inch kgs lbs kgs lbs kgs lbs kgs lbs kgs lbs 25 1" N/A N/A N/A N/A 4 9 4 9 N/A N/A 40 1½” N/A N/A N/A N/A 7 15 6 13 N/A N/A 50 2" N/A N/A 9 20 N/A N/A 8 20 N/A N/A 65 2½” N/A N/A 10.7 24 N/A N/A 11 24 N/A N/A 80 3" N/A N/A 11.6 26 N/A N/A 13 28 N/A N/A 100 4" N/A N/A 15.2 33 N/A N/A 19 41 N/A N/A 125 5" N/A N/A 20.4 45 N/A N/A 24 52 N/A N/A 150 6" N/A N/A 26 57 N/A N/A 29 64 N/A N/A 200 8" 48 106 48 106 N/A N/A 56 124 N/A N/A 250 10" 64 141 69 152 N/A N/A 79 174 N/A N/A 300 12" 76 167 86 189 N/A N/A 110 243 N/A N/A 350 14" 100 220 116 255 N/A N/A 131 289 N/A N/A 400 16" 127 280 144 317 N/A N/A 165 364 N/A N/A 450 18" 152 335 178 393 N/A N/A 176 388 N/A N/A 500 20" 184 405 232 512 N/A N/A 235 518 N/A N/A 600 24" 258 568 343 736 N/A N/A 345 761 N/A N/A 700 28" 315 693 350 772 N/A N/A N/A N/A 309 681 750 30" N/A N/A N/A N/A N/A N/A N/A N/A 480 1058 800 32" 410 904 442 975 N/A N/A N/A N/A 421 928 900 36" 512 1129 550 1213 N/A N/A N/A N/A 539 1188 1000 40" 650 1433 732 1614 N/A N/A N/A N/A 670 1477 42" N/A N/A N/A N/A N/A N/A N/A N/A 700 1544 1100 44" N/A N/A N/A N/A N/A N/A N/A N/A 1100 2426 1200 48" 990 2183 1106 2439 N/A N/A N/A N/A 1030 2271
AWWA
Nominal size
PN 10 PN 16 PN 40 Class 150
MAG 5100 W weight
The effect of temperature on working pressure MAG 5100 W
Metric (Pressures in bar) Sizes 25 mm, 40 mm & > 600 mm Flange spec. Flange Temperature
°°
°°
°C
rating −510 5090 EN 1092-1 PN 10 10.0 10.0 9.7 9.4 PN 16 16.0 16.0 15.5 15.1 PN 40 40.0 40.0 38.7 37.7 ANSI B16.45 150 lb 19.7 19.7 19.3 18.0 AWWA C-207 Class D 10.3 10.3 10.3 10.3
Sizes 50 mm to 600 mm
EN 1092-1 PN 10 10.0 10.0 10.0 8.2 PN 16 10.0 16.0 16.0 13.2 PN 40 10.0 40.0 40.0 32.9 ANSI B16.45 150 lb 10.0 19.7 19.7 16.2
Imperial (Pressures in Psi) Sizes 1", 1½”, & > 24" Flange spec. Flange Temperature
°°
°°
°F
rating 23 50 120 200 EN 1092-1 PN 10 145 145 141 136 PN 16 232 232 225 219 PN 40 580 580 561 547 ANSI B16.45 150 lb 286 286 280 261 AWWA C-207 Class D 150 150 150 150
Sizes 2" to 24"
EN 1092-1 PN 10 145 145 145 119 PN 16 145 232 232 191 PN 40 145 580 580 477 ANSI B16.45 150 lb 145 286 286 235
Page 33
33
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
DN A
1)
A
1
BD
1
L
2)
AS T
C
3)
T
E
3)
2129 E, AWWA AS 4087 C -207
PN PN PN PN PN Class Class Class Class
6, 25 40 64 100 150 300 14- D
10, 21-
16 35
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [kg]
15 187 338 59 104 200 200 200 - - 200 200 200 - 6 4 25 187 338 59 104 200 200 200 - 260 200 200 200 1.2 6 5 40 197 348 82 124 200 200 200 - 280 200 200 200 1.2 6 8 50 205 356 72 139 200 200 200 276 300 200 200 200 1.2 6 9 65 212 363 72 154 200 200 200 320 350 200 272 200 1.2 6 11
80 222 373 72 174 200 272 272 323 340 272 272 200 1.2 6 12 100 242 393 85 214 250 250 250 380 400 250 310 250 1.2 6 16 125 255 406 85 239 250 250 250 420 450 250 335 250 1.2 6 19 150 276 427 85 282 300 300 300 415 450 300 300 300 1.2 6 27 200 304 455 137 338 350 350 350 480 530 350 350 350 1.2 8 40 250 332 483 137 393 450 450 450 550 620 450 450 450 1.2 8 60 300 357 508 137 444 500 500 500 600 680 500 500 500 1.6 8 80 350 362 513 270 462 550 550 550 700 800 550 550 550 - 1.6 8 110 400 387 538 270 512 600 600 600 750 - 600 600 600 - 1.6 10 125 450 418 569 310 563 600 600 600 - - 600 640 600 - 1.6 10 175 500 443 594 350 614 625 625 680 - - 680 730 625 - 1.6 10 200 600 494 645 430 715 750 750 750 - - 820 860 750 - 1.6 10 300 700 544 695 500 816 875 - - - - - - 875 875 2.0 - 350 750 571 722 556 869 - - - - - - - 937 937 2.0 - 380 800 606 757 560 939 1000 - - - - - - 1000 1000 2.0 - 475 900 653 804 630 1042 1125 - - - - - - 1125 1125 2.0 - 560
1000 704 906 670 1146 1250 - - - - - - 1250 1250 2.0 - 700 1100 755 906 770 1248 1375 - - - - - - - - 2.0 - 1200 1200 810 961 792 1348 1500 - - - - - - 1500 1500 2.0 - 1250 1400 925 1076 1000 1675 1750 - - - - - - - - 3.0 - 1753 1500 972 1123 1020 1672 - - - - - - - 1875 1875 3.0 - 2600 1600 1025 1176 1130 1915 2000 - - - - - - - - 3.0 - 2341 1800 1123 1274 1250 1974 2250 - - - - - - - - 3.0 - 3253 2000 1223 1374 1375 2174 2500 - - - - - - - - 3.0 - 4060
4.4 Sensor MAG 3100 and MAG 3100 W
MAG 3100 & MAG 3100 W, compact/separate
1
) 13 mm shorter with AISI terminal box (Ex and high temperature)
2
) When earthing flanges are used, the thickness of the earthing flange must be added to the built-in length
3
)TC = Type C grounding ring, TE = Type E grounding ring
4
) Weights are approx. and for PN 16 without signal converter
D = Outside diameter of flange, see flange tables
Weight
4)
EN 1092-1-2001
BS 1560/ ANSI 16.5
Page 34
34
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
4.5 Signal converter
Signal converter installed in compact mode
Weight: MAG 6000 and MAG 5000: 0.75 kg
Weight: Wall bracket: 0.9 kg
Signal converter installed remote
19"insert, standard unit
Weight incl. back print:
MAG 5000: 0.8 kg MAG 6000: 0.8 kg Safety barrier (ia/ib): 1.0 kg Safety barrier (ia): 0.8 kg Cleaning unit: 0.9 kg
Compact polyamide
Earthing/protection flange
DN
t
1
t
2
Weight
[mm] [mm] [kg]
25-250 1.2 15 0.03-0.4 300-600 1.6 20 0.6-2.6 700-1200 2.0 25 3-5 1400-2000 3.0 40 9-16
DN
t
1
Weight
[mm] [kg]
15 6 0.07
25-150 6 0.3-1.4 200-350 8 1.7-4.1 400-600 10 6.5-13.0
Type C flanges for liners of neoprene, EPDM, Linatex and ebonite. Type E flanges for liners of PTFE.
Note
MAG 3100 high temperature (PTFE) is always equipped with 2 pcs. type E earthing flanges.
Page 35
35
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
Wall mounting box 21 TE
Weight excl. signal converter: 2.3 kg
Wall mounting box 42 TE
Weight excl. signal converter: 2.9 kg
Panel front unit 21 TE
Weight excl. signal converter: 1.2 kg
Page 36
36
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
4. Dimensions and weight
D & W
Back of panel unit 21 TE
Back of panel unit 42 TE
Weight: 0.9 kg
Weight: 0.7 kg
Panel front unit 42 TE
Weight excl. signal converter: 1.6 kg
Page 37
37
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
5. Installation of sensor
Sensor
To obtain optimum results from the measuring system, the chassis body of the sensor must have the same electrical potential as the liquid being measured.
5. Installation of sensor
5.1 Potential equalization
MAG 1100
Graphite gaskets EPDM or PTFE gaskets
Electrically conductive piping
A: Potential equalization with electri-
cally conductive graphite gaskets
B: Potential equalization using earth
strap supplied.
C: Potential equalization with electri-
cally conductive graphite gaskets
D: Potential equalization using sepa-
rate potential equalization ring
Electrically non-conductive piping
MAG 1100 FOOD
The sensor must be installed between two adap­ters. Potential equalization with the liquid oc­curs automatically via these adapters and through the adjacent pipe.
MAG 3100 W / MAG 3100 (except PTFE liner)
Potential equalization is carried out with the built-in earthing electrodes. No further action need to be taken.
Page 38
38 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
5. Installation of sensor
Sensor
MAG 3100 (PTFE liner)
Electrically conductive piping
Use an earth straps on one side.
Potential equalization, electrically conductive pipe
Non-conductive piping
Use an earthing flange. Place the flange between flowmeter and the adjacent pipe flange. Selection of earthing flange depends on the medium, liner material and application, see figure.
Special attention must be given to systems with cathodic protection.
Compact installation:
The signal converter must be supplied through an isolation transformer. The terminal "PE" must never be connected.
Remote installation:
The screen must only be connected at the sensor end via a 1.5 µF condensator. The screen must never be connected at both ends.
Isolated sensor:
If above mentioned connections are unaccept­able, the sensor must be isolated from the pipe work.
Liner Suitable
material earthing flange
PTFE Type E
Type of earthing flange depending on liner material
With abrasive liquids, flowmeter inlet protection may be necessary. Here type C and E earthing flanges are used. Type C (for all liners except PTFE) is clamped between the flanges. Type E (for PTFE liner only) is fitted to the flange. When using an earthing flange, gaskets must always be used between the adjacent pipe flange and the earthing flange.
5.2 Inlet protection MAG 3100
5.3 Cathodic protected piping
Page 39
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
39
6. Installation of signal converter
6.1
Compact installation MAG 5000 and MAG 6000
Compact polyamide
Step 1
Remove and discard the terminal box lid of the sensor.
Fit the PG 13.5 cable glands for the supply and output cables.
Step 2
Remove the two black plug assemblies for coil and electrode cables in the terminal box and connect them to their corresponding terminal numbers on the connection board.
Step 3
Connect an earth wire between PE on connec­tion board and bottom of connection box. Connect the 2 pin connector and 3 pin connector as shown.
Note
In earlier version the 3 pin connector was a 5 pin connector.
Step 4
Mount the connection plate in the terminal box. The SENSORPROM
®
unit connections will be established automatically when the connection plate is mounted in the terminal box.
Note
Check that your connection board lines up with the SENSORPROM
®
unit, if not, move the SENSORPROM
®
unit to the other side of the
terminal box.
Step 5
Fit the supply and output cables respectively and tighten the cable glands to obtain optimum sealing. Please refer to the wiring diagram in chapter 7 for the electrical connections.
Mount the signal converter on the terminal box.
Note
System will not register flow if black plugs are not connected to connection board
Page 40
40
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
1. Use a screw driver to remove the outer frame.
2. Loosen the 4 screws retaining the control pad.
3. Withdraw the control pad and turn it to the required orientation.
4. Tighten the 4 screws until a mechanical stop is felt in order to obtain IP 67 enclo­sure rating.
5. Snap-lock the outer frame onto the control pad (click).
The signal converter can be mounted in either direction as the arrow indicates without turning the terminal box.
The terminal box can be rotated ±90° in order to optimize the viewing angle of the signal con­verter display/keypad: Unscrew the four screws in the bottom of the terminal box. Turn the terminal box to the re­quired position and retighten the screws firmly.
Turning the control pad
Turning the signal converter
1
2
4
3
5
Page 41
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
41
Locate the add-on module in the bottom of the MAG 6000 signal converter.
6.2.1 Add-on modules MAG 6000 only
Press the add-on module forwards as far as possible.
The add-on module has now been installed and the signal converter is ready to be installed on the terminal box. Communication to the operator menu and electrically inputs and outputs is automatically established by power on.
6.2.2 Remote installation -
At the sensor
Fit and connect the electrode and coil cables as shown in chapter 7 “Electrical connections”. The unscreened cable ends must be kept as short as possible. The electrode cable and the coil cable must be kept separate to prevent interference. Tighten the cable glands well to obtain opti­mum sealing.
Remove the SENSORPROM® unit from the sensor and mount it on the connection plate in the signal converter.
Page 42
42
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
6.2.3 Remote installation -
Wall mounting
MAG 6000
Mount wall bracket on a wall or into the back of a panel.
Mount wall bracket on a vertical or horizontal pipe using an ordinary hose clip or a duct strap.
Vertical pipe mounting
Horizontal pipe mounting
Take the SENSORPROM® memory unit from the sensor. Mount the SENSORPROM
®
unit in the wall mounting unit as shown. The text on the SENSORPROM
®
unit must face towards the
wall bracket.
Mount an earth wire between PE on connection board and bottom of connection box.
Page 43
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
43
6.2.3 Remote installation -
Wall mounting
(continued)
Mount the connection plate in the terminal box. Fix the connection plate with the two diagonal opposite screws.
Fit the coil, electrode, supply and output cables respectively and tighten the cable glands to obtain optimum sealing. Please see the wiring diagram in chapter 7 for the electrical connec­tions.
Mount the signal converter on the terminal box.
Page 44
44
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
6.2.4 Remote installation -
Signal converter in 19" insert
1. Fit the SENSORPROM® memory unit on the connection board supplied with the signal converter. The SENSORPROM® unit is supplied with the sensor in the terminal box.
2. Mount the guide rails into the rack system as shown. Distance between guide rails is 20 TE. Guide rails are supplied with the rack system and not with the signal converter.
3. Mount the connection board as shown.
4. Connect the cables as shown under "Electrical connection", chapter 7.
5. Insert the signal converter into the rack system.
Page 45
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
45
Locate the add-on module in the bottom of the MAG 6000 signal converter.
6.2.5 Add-on modules MAG 6000 only
Press the add-on module forwards as far as possible.
The add-on module has now been installed and the signal converter is ready to be installed on the terminal box. Communication to the operator menu and electrically inputs and outputs is automati-cally established by power on.
Page 46
46
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
6.2.6 Installation in IP 65 wall mounting enclosure
1. Mount the IP 65 enclosure on the wall with four screws.
2. Mount the SENSORPROM® memory unit on the connection board as shown.
The SENSORPROM® unit is supplied with the sensor in the terminal box. The connection board for IP 65 wall mounting boxes must be used.
3. Connect the cables to the terminals, see "Electrical connection", chapter 7.
4. Insert the signal converter and close the cover.
Page 47
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
47
6.2.7 Installation in IP 65 panel mounting enclosure (front of panel)
1. Mount the SENSORPROM
®
memory unit on the connection board as shown.
The SENSORPROM
®
unit is supplied with the sensor in the terminal box.
2. Fit the enclosure in a cut out at the front of a panel. Fasten the four screws accessible at the front.
3. Connect the cables as shown under "Electrical connection", chapter 7.
4. Insert the signal converter and close the cover.
Page 48
48
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
6.2.8 Installation into the back of a panel
1. Mount the SENSORPROM
®
memory unit on the connection board as shown.
The SENSORPROM
®
unit is supplied with the sensor in the terminal box.
2. Mount the connection board into the back of the enclosure.
3. Connect the cables as shown under "Electrical connection", chapter 7.
4. Mount the enclosure into the back of a panel with four screws.
5. Insert the signal converter.
Page 49
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
49
6.3 Signal converter
Safety barrier
1. Fit the SENSORPROM® memory unit on the connection board supplied with the safety barrier. The SENSORPROM
®
unit is delivered mounted in the terminal box of the sensor. The
connection board supplied with the signal converter is not used.
2. Mount the guide rails into the rack system as shown. Distance between guide rails is 20 TE. Guide rails are supplied with the rack system and not with the signal converter.
3. Mount the connection board as shown. The mounting screw must be installed just in line with the guide rails.
4. Connect the cables as shown under "Electrical connection", chapter 7.
5. Insert the signal converter and the safety barrier into the rack system.
Page 50
50
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
6. Installation of signal converter
Installation of signal conv.
6.4 Signal converter
Cleaning unit
1. Fit the SENSORPROM
®
memory unit on the connection board supplied with the cleaning unit.
The SENSORPROM
®
unit is delivered mounted in the terminal box of the sensor. The
connection board supplied with the signal converter is not used.
2. Mount the guide rails into the rack system as shown. Distance between guide rails is 20 TE. Guide rails are supplied with the rack system and not with the signal converter.
3. Mount the connection board as shown. The mounting screw must be installed just in line with the guide rails.
4. Connect the cables as shown under "Electrical connection", chapter 7.
5. Select AC-cleaning or DC-cleaning mode at the switch located on the base of the cleaning unit.
6. Insert the cleaning unit and the signal converter into the rack system.
Page 51
51
SITRANS F M MAGFLO
7. Electrical connection
DKFD.PS.027.W3.02
E. c.
7. Electrical connection
7.1 Signal converter MAG 5000 and MAG 6000 connection diagram
Grounding
PE must be connected due to safety class 1 power supply.
Mechanical counters
When mounting a mechanical counter to terminals 57 and 58 (active output), a 1000 µF capacitor must be connected to the terminals 56 and 58. Capacitor + is connected to terminal 56 and capacitor − to terminal 58.
Output cables
If long cables in noise environment, we recommend to use screened cable.
Electrodes cables
Dotted connections only to be when using special electrode cable.
Note
Special cable with individual wire shields (shown as dotted lines) are only requried when using empty pipe function or long cables. (See “Technical data” chapter 2 for further details.)
Page 52
52
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
7. Electrical connection
E. c.
7.2 Wiring diagram for signal converter and sensor
Compact installation
Compact
Remote IP 67 wall mounting
Remote installation
Cathodic protected piping
Compact installation:
The signal converter must be supplied through an isolation transformer. The terminal "PE" must not be connected.
Remote installation:
The screen must only be connected at the sensor end via a 1.5 µF capacitor. The screen must never be connected at both ends.
Sensor cables
Unscreened cable ends must be as short as possible and the two cables must be kept separate. Cables must be in one length and must not be taken to a distribution box or similar terminal arrangement.
Terminals 81 and 84 are only connected when special electrode cable with double screening is used.
Coil cable screen must be connected in both ends. Electrode cable screen must be connected at sensor side only.
Note
See 5.3 when using cathodic protection.
Note
Mount the grounding wire from connection box to PE to ensure sufficient grounding.
Page 53
53
SITRANS F M MAGFLO
7. Electrical connection
DKFD.PS.027.W3.02
E. c.
19” IP 20 version
19” IP 66 version
19” IP 20 version EEx (ia/ib) DN
≤ ≤
≤ ≤
300
Page 54
54
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
7. Electrical connection
E. c.
19” IP 20 version EEx e (ib) DN
≥≥
≥≥
350
19” IP 66 version EEx (ia/ib) DN
≤ ≤
≤ ≤
300
19” IP 66 version EEx e (ib) DN
≥≥
≥≥
350
Page 55
55
SITRANS F M MAGFLO
7. Electrical connection
DKFD.PS.027.W3.02
E. c.
19” IP 20 version with cleaning
19” IP 66 version with cleaning
Page 56
56 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Communication mode Basic settings Operator active
Service mode Output Operator inactive
Operator menu External input
Product identity Sensor characteristics
Language mode Reset mode
8. Commissioning
8.1 Keypad and display layout
Keypad
The keypad is used to set the flowmeter. The function of the keys is as follows:
TOP UP KEY This key (hold 2 sec.) is used to switch between operator menu
and setup menu. In the converter setup menu, a short press will cause a return to the previous menu.
FORWARD KEY This key is used to step forward through the menus. It is the only
key normally used by the operator.
BACKWARD KEY This key is used to step backward through the menus.
CHANGE KEY This key changes the settings or numerical values.
SELECT KEY This key selects the figures to be changed.
LOCK/UNLOCK KEY This key allows the operator to change settings and gives access
to submenus.
Display
The display is alphanumerical and indicates flow values, flowmeter settings and error messages.
The upper line is for primary flow readings and will always show either flow rate, totalizer 1 or totalizer 2. The line is divided into 3 fields.
S: Sign field P: Primary field for numerical value U: Unit field
The centre line is the title line (T) with individual information according to the selected operator or setup menu.
The lowest line is the subtitle line (ST) which either will add information to the title line or keep individual information independent of the title line.
F: The alarm field. Two flashing triangles will appear by a fault condition.
M: The mode field. The symbols indicate the following.
Ready for change Access to submenu
Value locked RESET MODE: Zero setting of
totalizers and initialization of setting
L: The lock field. Indicates the function of the lock key.
Page 57
57
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.2 Menu build-up
The menu structure of a specific signal converter type is shown in a menu overview map. Details of how a specific parameter is set is shown in a menu detail map for the specific parameter. A detail map is valid for each type of signal converter if not indicated otherwise. The menu structure is valid for the title and subtitle line only. The upper line is for primary readings only and will always be active with either flowrate, totalizer 1 or totalizer 2.
The menu is built up in two parts. An operator menu and a setup menu.
Operator menu
The operator menu is for daily operation. The operator menu is customised in the operator menu setup. The signal converter always starts in operator menu No. 1. The page forward and page backward keys are used to step through the operator menus.
Setup menu
The setup menu is for commissioning and service only.
Access to the setup menu is gained by pressing the top up key for 2 seconds. The setup menu operates in two modes:
View mode
Setup mode
View mode
is a read only mode. The pre-selected settings can only be scanned.
Setup mode
is a read and write mode. The pre-selected settings can be scanned and changed.
Access to the setup mode is password protected. The factory set password is 1000.
Access to a submenu in the set up menu is gained by the lock key. A short press on a top up key will bring you back to the previous menu. A long press (2 sec.) on the top up key will exit the setup menu and bring you back to operator menu No. 1.
8.2.1 Password
The SETUP MENU can be operated in two different modes:
VIEW MODE (Read only) CHANGE MODE (Read and write mode)
Access to view mode is always gained by pressing the forward key when in the password menu.
Access to change mode is password protected. The password is factory set to 1000, but can be changed to any value between 1 and 9999 in the change password menu.
The factory setting of 1000 can be re-established as follows:
Switch off power suppply
Press the TOP UP key and switch on the power supply
Release the key after ROM and RAM tests are completed
The user code is now reset to 1000.
Page 58
58 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.3.1 MAG 5000 and MAG 6000
Menu overview
Page 59
59
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.3.2 MAG 6000 CT
Menu overview
Page 60
60 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.4.1 Basic settings
Menu detail
Comma for flow rate, totalizer 1 and totalizer 2 can be individually positioned.
open the respective window.
ensure that the cursor is positioned below the comma. Use the SELECT KEY .
move the comma to the requested position. Use the CHANGE KEY .
Units are changed by means of the CHANGE KEY with the cursor placed below the unit selected. Select units (cursor moved) by means of the SELECT KEY .
Totalizer 2 is not visible when batch is selected as digital output.
Q
max.
2 - is only visible when it has been choosen as external input.
Main frequency
To select the main power supply frequency corresponding to the country in which the flowmeter is installed. (US = 60 Hz)
Flow direction
Select the correct flow direction in the pipe
Q
max.
Sets the measuring range, the analog outputs and the frequency output. Value, decimal point, unit and time can be set individually (setting is dimension dependent).
Q
max.
2
Sets the measuring range, the analog outputs and the frequency output. Value, decimal point, unit and time can be set individually (setting is dimension dependent). Only visible when it has been choosen as external digital input.
Totalizers
To set unit and decimal point.
Low flow cutt off
To set a % of selected Q
max.
. To filter noise in the
installation. Influences display and all outputs.
Error level
To select which error level, the flowmeter will detects an error.
Page 61
61
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.4.3 Digital and relay outputs
Error level
Error number
8.4.2 Outputs
Digital output Frequency Proportional to flowrate (Terminal 56, 57, 58)
The current output must be set off when not used.
Current output Proportional to flowrate (Terminal 31 and 32)
4 - 20 mA + alarm: Current output gives the following mA, depending on what is selected as error level in basic settings. Fatal: 1 mA, permanent: 2 mA, warning: 3 mA
Digital output Pulse/volume (Terminal 56, 57, 58)
Menu detail
Page 62
62 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Menu detail
Limit/direction
Limit switches are available for both digital as well as relay output.
Direction mode: 1 set point at 0% flow; hysteresis 5%. If 2 set points must activate 2 separate outputs, a single set point has to be selected individually for digital as well as relay outputs.
Batch
(MAG 6000 only)
(Possible through relay and digital output)
8.4.4 Relay outputs
Cleaning
8.4.5 External input
Batch control is available on MAG 6000 only.
The relay output must always be used to operate the cleaning unit when a cleaning unit has been installed together with the signal converter. The relay output cannot be used for other purposes.
Page 63
63
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Menu detail
8.4.6 Sensor characteristics
8.4.7 Reset mode
If “SENSORPROM not installed” is shown, refer to chapter 6 (depending on type of mounting configura­tion).
Page 64
64 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.4.8 Service mode
All previous settings are reinitialised when service mode is exited using the top up key.
The error system
The error system is divided into an error pending list and a status log list. Time is gained as days, minutes and hours since the error has occurred.
The first 9 standing errors are stored in error pending. When an error is removed it is removed from error pending.
The latest 9 errors are stored in the status log. When an error is removed it is still kept in status log. Errors in status log is stored for 180 days.
Error pending and status log are accessible when enabled in the operator menu.
Menu detail
1) Standard
2) If digital output is set to frequency
Page 65
65
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Menu setup
8.4.9 Operator menu setup
The upper line is always active and can never be deselected.
The two lower lines are for individual operator information. Information which the operator can scroll through with the forward key.
A closed lock key in the operator menu setup, means that the menu is enabled when viewing the operator menu.
An open lock key symbol, means that the menu is not available in the operator menu.
The middle line can either be used as a heading “Text line” for the lower line, or as a flow reading. A flow reading can be individually selected for each menu.
The lower line may be used for an additional flow reading to the reading already available in the upper line.
Text means that the text for the chosen measured value is shown. For example if text is chosen in line 2 and flow rate is chosen in line 3, the text “flow rate” is shown in line 2 and the measured flow rate value is show in line 3.
Page 66
66 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.4.10 Product identity
Software version of add-on module is only available if the add-on module has been installed.
Menu detail
8.4.11 Change password
Page 67
67
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.4.12 Language mode
Menu detail
8.4.13 HART


communication MAG 5000 HART or as add-on module
Page 68
68 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.5.3 Batch
(Only available on MAG 6000)
A batch can be started, paused or stopped from the operator menu, in addition to the externally operated batch control. The batch is controlled using the lock and the top up keys.
The lock key:
Starts the batch
Holds the batch (pause) when pressed during batching
Restarts the batch to continue when pressed during a pause.
The top up key resets a batch completely during a pause.
The accumulated number of performed batches can be viewed when enabled in the operator menu setup.
The batch cycle counter is reset by pressing the lock key in the “batch cycle cnt reset” menu.
Batch cycle counter
Batch cycle counter reset
8.5.1 Flow rate
Operator menu
The 1
st
display line is always active and shows the value enabled in the operator menu setup.
Flow rate
Totalizer 1
Totalizer 2
The 2
nd
and 3rd display lines are individually set in the operator menu. The page forward key steps
through the enabled settings.
Flow rate
Totalizer
Totalizer reset
Batch control
Batch cycle counter
Batch cycle counter reset
Pipe size
Sensor type
Pending errors
Status log
Tag No.
8.5.2 Totalizer
A totalizer is reset by pressing the lock key when the corresponding totalizer reset window is open.
Page 69
69
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
8.6.1 Settings available
Parameter Factory settings Settings available Password
Default value 1000 Password 1000 1000 - 9999
Basic settings
Flow direction Positive Positive, negative Q
max.
Dim. dependent Dim. dependent
- Volume units Dim. dependent m3, ml, l, kl, hl, Ml, ft3, in3, USG, USkG, USMG, UKG, UKMG
- Time units Dim. dependent Sec., min., hour, day
Totalizer 1 Forward Forward, reverse, net
- Totalizer 1 units Dim. dependent m3, ml, l, kl, hl, Ml, ft3, in3, USG, USkG, USMG, UKG, UKMG
Totalizer 2 Reverse Forward, reverse, net
- Totalizer 2 units Dim. dependent m3, ml, l, kl, hl, Ml, ft3, in3, USG, USkG, USMG, UKG, UKMG
Low flow cut-off 1.5 % 0 - 9.9 % Empty pipe Off Off, on Error level Warning Fatal, permanent, warning
Output
Current output Off On/off, uni-/bidirectional, 0/4 - 20 mA
- Time constant 5 s 0.1 - 30 s
Digital output Pulse Error, direction/limit, batch1), frequency, pulse, error no.,
off
Relay output Error Error, direction/limit, cleaning, error No., off Direction/limit switch Off 1 set point/2 set points, 100 - 100%
- Hysteresis 5% 0.0 - 100%
Batch
1)
Off
- Batch quantity 0 Dim. dependent
- Batch compensation 0 100 - 100 m
3
- Batch counter Down Up/down
- Time constant 0.1 s 0.1 - 30 s
Frequency Off 500 Hz, 1 kHz, 5 kHz, 10 kHz
- Time constant 5 s 0.1 - 30 s
Pulse On
- Pulse polarity Positive Positive/negative
- Pulse width 66 ms 64 µs, 130 µs, 260 µs, 510 µs, 1.0 ms, 2.0 ms,
4.1 ms, 8.2 ms, 16 ms, 33 ms, 66 ms, 130 ms, 260 ms, 520 ms, 1.0 s, 2.1 s, 4.2 s.
- Volume/pulse Dim. dependent Dim. dependent
- Time constant 0.1 s 0.1 - 30 s
Electrode cleaning Off Off/cleaning
- Cleaning cycle time 24 h 1 - 240 h
External input
External input Off Batch, reset totalizer, freeze output, forced output, off
- Batch Start, hold/continue, stop, Q
max.
2
Sensor characteristics
Correction factor 1 0.85 - 2.00 Language English English, German, French, Danish, Swedish, Finnish,
Spanish, Russian, Italian, Portuguese
Operator menu
Primary field Flow rate Flow rate, Totalizer 1, Totalizer 2 Title/subtitle line Flow rate Flow rate, Flow rate %, Q
max.
, Totalizer 1, Totalizer 2, Totalizer 1 reset, Totalizer 2 reset, Batch start/ paused/stop, Batch cycle counter, Batch cycle counter reset, Sensor size, Sensor type, Error pending, Status log, Tag No.
1)
Batch is available on MAG 6000 only
The signal converter is delivered with factory settings ready to measure the actual flow.
Page 70
70 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
DN Q
max.
MAG 5100 W MAG 1100,
3100, 3100 W Volume/ Pulse Totalizer
mm [inches] fac.set. min. max. min. max. unit pulse unit unit
6
5
/
16
300 - - 25.5 1017 l/h 1 l l
10
3
/
8
900 - - 70.7 2827 l/h 1 l l
15
1
/
2
2000 - - 159.1 6361 l/h 1 l l 25 1 5000 442.0 17671 442.0 17671 l/h 10 l l 40 11/
2
12 1.2 45 1.2 45 m3/h 10 l l 50 2 20 1.6 63 1.8 70 m3/h 10 l l 65 21/
2
30 2.5 100 3.0 119 m3/h 100 l l 80 3 50 4.0 160 4.6 180 m3/h 100 l l
100 4 120 6.3 250 7.1 282 m3/h 100 l l 125 5 180 10.0 400 11.1 441 m3/h 100 l m
3
150 6 250 15.7 629 16.0 636 m3/h 100 l m
3
200 8 400 24.9 997 28.3 1130 m3/h 1 m
3
m
3
250 10 700 40.0 1600 44.2 1767 m3/h 1 m
3
m
3
300 12 1000 62.5 2500 63.7 2544 m3/h 1 m
3
m
3
350 14 1200 86.6 3463 86.6 3463 m3/h 1 m
3
m
3
400 16 1800 113.1 4523 113.1 4523 m3/h 1 m
3
m
3
450 18 2000 143.2 5725 143.2 5725 m3/h 1 m
3
m
3
500 20 3000 176.8 7068 176.8 7068 m3/h 1 m
3
m
3
600 24 4000 254.5 10178 254.5 10178 m3/h 10 m
3
m
3
700 28 5000 346.4 13854 346.4 13854 m3/h 10 m
3
m
3
750 30 6000 397.7 15904 397.7 15904 m3/h 10 m
3
m
3
800 32 7000 452.4 18095 452.4 18095 m3/h 10 m
3
m
3
900 36 9000 573.0 22902 573.0 22902 m3/h 10 m
3
m
3
1000 40 12000 707.0 28274 707.0 28274 m3/h 10 m
3
m
3
1100 44 14000 855.3 34211 855.3 34211 m3/h 10 m
3
m
3
1200 48 15000 1018.0 40715 1018.0 40715 m3/h 10 m
3
m
3
1400 54 25000 - - 1385.5 55417 m3/h 10 m
3
m
3
1500 60 30000 - - 1590.5 63617 m3/h 10 m
3
m
3
1600 66 35000 - - 1809.6 72382 m3/h 10 m
3
m
3
1800 72 40000 - - 2290.3 91608 m3/h 10 m
3
m
3
2000 78 45000 - - 2827.5 113097 m3/h 10 m
3
m
3
8.6.2 Dimension dependent factory settings MAG 5000 and MAG 6000
Volume/pulse or batch quantity
min. max.
DN 15 147 µl 3.86 m
3
DN 25 409 µl 10.7 m
3
DN 40 1.05 ml 27.5 m
3
DN 50 1.64 ml 42.9 m
3
DN 65 2.77 ml 72.5 m
3
DN 80 4.19 ml 110 m
3
DN 100 6.54 ml 172 m
3
DN 125 10.2 ml 268 m
3
DN 150 14.7 ml 386 m
3
DN 200 26.2 ml 686 m
3
DN 250 40.9 ml 1072 m
3
DN 300 58.9 ml 1544 m
3
DN 350 80.2 ml 2102 m
3
DN 400 105 ml 2745 m
3
DN 450 133 ml 3474 m
3
DN 500 164 ml 4289 m
3
DN 600 236 ml 6177 m
3
DN 700 321 ml 8407 m
3
DN 800 419 ml 10981 m
3
DN 900 530 ml 13897 m
3
DN 1000 654 ml 17157 m
3
DN 1200 942 ml 24706 m
3
DN 2000 2.62 l 68629 m
3
8.6.3 Dimension dependent batch and pulse output settings
Page 71
71
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Internal totalizers
Depending on type of approval it is possible to reset the internal totalizers. The type of approval is selected in the reset menu, with the hardware key mounted. It is possible to select between:
Hot/cold water
Other liquids
Resetting of totalizers by electrical input is not possible.
Hot/cold water
Totalizer 1 is allocated to forward flow (cannot be reset).
Totalizer 2 is allocated to reverse flow (cannot be reset).
Other liquids
Both totalizer 1 and totalizer 2 are allocated to measure the net flow, i.e. any reverse flow will make the totalizers count backwards.
Totalizer 1 cannot be reset.
Totalizer 2 can be reset if the flow velocity in the meter pipe is <0.25 m/s. When the totalizer is
reset, the pulse output register will also be reset.
Setting primary operating parameters such as Q
max.
, low flow cut-off, units, approvals, etc. is blocked
during normal operation. See menu setup.
These settings are made in connection with commissioning or calibration by mounting a hardware key on the connection plate of the signal converter. When the key is mounted, there is access to all menu items. When the key is removed, the primary settings are blocked in accordance with the requirements in the authorisation.
8.6.4 MAG 5000 CT and MAG 6000 CT settings
Output
When selecting hot water, the output settings are not allowed and menu will not be shown on display.
When selecting cold water or other liquids, all output settings can be changed.
Page 72
72 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Error system
The converter system is equipped with an error and status log system with 4 groups of information.
Information without a functional error involved
Warnings which may cause malfunction in the application. The cause of the error may disappear
on its own
Permanent errors which may cause malfunction in the application. The error requires an operator
Fatal error which is essential for the operation of the flowmeter
2 menus are available in service and operator menus for registration of information and errors
Error pending
Status log
Error pending
The first 9 standing errors are stored in “error pending”. When an error is removed it is removed from “error pending”. The acceptance level for “error pending” can be individually configured to a particular application. The acceptance level is set in the “basic settings” in the converter setup menu.
Acceptance levels
Fatal error: Fatal errors are registered as errors
Permanent errors: Permanent and fatal errors are registered as errors
Warning (Default value): Warnings, permanent and fatal errors are registered as errors
The error information is displayed in the title and subtitle line. The title line will show the time since occurrence of error. The subtitle line will flash between an error text and a remedy text. The error text will indicate type of error (I, W, P or F), error No. and error text. The remedy text will inform the operator of the action to take to remove the error.
Status Log
Like “error pending” except that information, warnings, permanent and fatal errors is always stored in the “status log”. The “status log” stores the latest 9 messages received/registered during the last 180 days.
Alarm field
The alarm field on the display will always flash with an error pending.
Error output
The digital and relay output can individually be activated error by error (error level). The relay output is default selected to error level. An output can also be selected to activate on a single error number. The alarm field, error output and error pending always operate together. The analog output turns to a 1 mA level when in the 4-20 mA mode.
Operator menu
Error pending and status log are as default enabled in the operator menu.
8.7.1 Error handling
Page 73
73
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
8. Commissioning
Commissioning
Error Error text #Comment Outputs Input
No. Remedy text status status
1 I1 -
Power on
OK Power on has happened Active Active
2 I2 -
Add-on module
Applied A new module has been applied to the system Active Active
3 I3 -
Add-on module
Install An add-on module is defect or has been removed.
This can be an internal add-on module Active Active
4 I4 -
Param. corrected
OK A less vital parameter in the converter has been re-
placed by its default value Active Active
20 W20 -
Totalizer 1
Reset manually During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the saved totalizer value anymore. The totalizer value must be reset manually in order to rely on future readings Active Active
20 W20 -
Totalizer 2
Reset manually During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the saved totalizer value anymore. The totalizer value must be reset manually in order to rely on future readings Active Active
21 W21 -
Pulse overflow
Adj. pulse settings Actual flow is too big compared with pulse width and Reduced
volume/pulse pulse width Active
22 W22 -
Batch timeout
Check installation Duration of batching has exceeded a predefined Batch out-
max. time put on zero Active
23 W23 -
Batch overrun
Check installation Batch volume has exceeded a predefined maximum Batch out-
overrun volume put on zero Active
24
W24 - Batch neg. flow
Check flow direction Negative flow direction during batch Active Active
30 W30 -
Overflow
Adj. Q
max.
Flow is above Q
max.
settings Max. 120 % Active
31 W31 -
Empty pipe
Pipe is empty Zero Active
40 P40 -
SENSORPROM
Insert/change SENSORPROM unit not installed Active Active
41 P41 -
Parameter range
Switch off and on A parameter is out of range. The parameter could not
be replaced by its default value. The error will dis­appear at the next power-on Active Active
42 P42 -
Current output
Check cables Current loop is disconnected or the loop resistance
is too big Active Active
43 P43 -
Internal error
Switch off and on Too many errors occured at the same time
Some errors are not detected correctly Active Active
44 P44 -
CT SENSORPROM
SENSORPROM unit has been used as CT version Active Active
60 F60 -
CAN comm. error
Converter/AOM CAN bus communication error. An add-on module, the
display module or the converter is defect Zero Inactive
61 F61 -
SENSORPROM error
Replace It is not possible to rely on the data in SENSOR-
PROM unit anymore Active Active
62 F62 -
SENSORPROM ID
Replace The SENSORPROM unit ID does not comply with the
product ID. The SENSORPROM
unit is from another type of
product MASSFLO, SONOFLO etc. Zero Inactive
63 F63 -
SENSORPROM
Replace It is not possible to read from the SENSORPROM
unit anymore. Active Active
70 F70 -
Coil current
Check cables Coil excitation has failed Active Active
71 F71 -
Internal error
Replace converter Internal convertion error in ASIC Active Active
8.7.2 List of error numbers
Page 74
74 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
9. Service
Commissioning
9. Service
Power on converter, display light on
YES
NO
NO
Check cables/connections Check connection board Check pins in converter multiplug - OK
Correct fault
YES
NO
Output readings OK
Converter defective
YES
Display defect Change display
Output and display readings OK ?
NO
NO
Check cables/connections Check connection board Check pins in converter multiplug - OK
Correct fault
Error triangles flashing
NO
YES
Check error table
Converter OK ­Check settings/application Check installation/sensor/ earthing connection etc.
YES
YES
Often problems with unstable/wrong measurements occur due to insufficient/wrong earthing or potential equalization. Please check this connection. If OK, the SITRANS F M MAGFLO con­verter can be checked as described under 9.1 and sensor under 9.3.
9.1 Converter check list
When checking SITRANS F M MAGFLO installations for malfunction the easiest method to check the signal converter is to replace it with another MAG 5000/6000 converter with a similar power supply. A replacement can easily be done as all settings are stored in and downloaded from the SENSORPROM
unit - no extra settings need to be made.
If no spare converter is available - then check converter according to check table.
Page 75
75
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
9. Service
Commissioning
9.2 Trouble shooting MAG converter
Symptom Output Error Cause Remedy
signals code
Empty display Minimum 1. No power supply Power supply
Check MAG 5000/6000 for bended pins on the connector
2. MAG 5000/6000 defective Replace MAG 5000/6000
No flow signal Minimum 1. Current output disabled Turn on current output
2. Digital output disabled Turn on digital output
3. Reverse flow direction Change direction
F70 Incorrect or no coil current Check cables/connections
W31 Measuring pipe empty Ensure that the measuring
pipe is full
F60 Internal error Replace MAG 5000/6000
Undefined P42 1. No load on current output Check cables/connections
2. MAG 5000/6000 defective Replace MAG 5000/6000
P41 Initializing error Switch off MAG 5000/6000,
wait 5 s and switch on again
Indicates flow Undefined Measuring pipe empty Select empty pipe cut-off with no flow
Empty pipe cut-off is OFF Ensure that the measuring
in pipe
pipe is full Electrode connection missing/ Ensure that electrode cable electrode cable is insufficiently is connected and sufficiently screened screened
Unstable Unstable 1. Pulsating flow Increase time constant flow signal
2. Conductivity of medium Use special electrode cable
too low
3. Electrical noise potential Ensure sufficient potential
between medium and equalization
sensor
4. Air bubbles in medium Ensure medium does not
contain air bubbles
5. High concentration of par- Increase time constant
ticles or fibres
Measuring error Undefined Incorrect installation Check installation
P40 No SENSORPROM unit Install SENSORPROM unit P44 CT SENSORPROM unit Replace SENSORPROM unit
or reset SENSORPROM unit
with MAG CT converter
F61 Deficient SENSORPROM unit Replace SENSORPROM unit F62 Wrong type of SENSORPROMReplace SENSORPROM unit
unit
F63 Deficient SENSORPROM unit Replace SENSORPROM unit F71 Loss of internal data Replace MAG 5000/6000
Maximum W30 Flow exceeds 100% of Q
max.
Check Q
max.
(Basic Settings)
W21 Pulse overflow
Volume/pulse too small Change volume/pulse
Pulse width too large Change pulse width
Measuring Missing one electrode Check cables approx. 50%
connection
Loss of totalizer OK W20 Initializing error Reset totalizer manually data ##### OK Totalizer roll over Reset totalizer or increase
Signs in display
totalizer unit
Page 76
76 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
9. Service
Commissioning
ATTENTION!
If there is leakage from MAG 1100/3100/3100 W or MAG 5100 W and the unit has been used to measure inflammable/explosive liquids, there might be a risk of explosion when checking with a megger.
9.3 Check list MAG sensor
Disconnect all leads to MAG 1100/3100/3100 W or MAG 5100 W
MAG 1100/3100/3100 W or MAG 5100 W installed and filled with the medium:
Check coil resistance with a multimeter R 85-86 Nominal value: See table
Check coil insulation with a megger (500 V) R 85 - Gnd Nominal value: Infinite resistance
Check electrode circuit with a moving-coil instru­ment without amplifier (measurement unstable) R 82 - Gnd Nominal value: 5-500 kohm R 83 - Gnd Nominal value: 5-500 kohm
Sensor apparently OK. Remove sensor, dry it and check
Sensor may be defective. Contact Siemens Flow Instruments
Dirty electrodes or conductivity too low
NO
NO
> 500 kohm
< 500 kohm
YES
YES
MAG 1100/3100/3100 W or MAG 5100 W removed from system – empty and dry:
Check coil resistance with a multimeter R 85-86 Nominal value: See table
Check coil insulation with a megger (500 V) R 85 - Gnd Nominal value: Infinite resistance
Check electrode insulation with megger (500 V) R 82 - Gnd Nominal value: Infinite resistance R 83 - Gnd Nominal value: Infinite resistance
NO
NO
YES
YES
MAG 1100/3100/3100 W or MAG 5100 W OK
YES
Check continuity between electrode and electrode terminal R 82 - Electrode Nominal value: 0 kohm R 83 - Electrode Nominal value: 0 kohm
YES
Sensor may be defective. Contact Siemens Flow Instruments
NO
NO
Page 77
77
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
9. Service
Commissioning
9.4 Coil resistance
Coil resistance for MAG 1100, MAG 1100 PFA = 98 ohms +/ 4
Note
On MAG 1100 DN 15 produced as from May 1999 the coil resistance must be 86 ohm, +8/4 ohm.
Coil resistance
MAG 3100 MAG 3100 W MAG 5100 W
DN Resistance Size mm Ohms Tolerance Ohms Tolerance
15 104 15 104 +/ 2 104 +/ 2 25 104 25 104 +/ 2 104 +/ 2 40 92 40 92 +/ 292+/− 2 50 92 50 92 +/ 2 124 +/ 4 65 100 65 100 +/ 2 127 +/ 4
80 94 80 94 +/ 2 126 +/ 4 100 92 100 92 +/ 2 125 +/ 4 125 92 125 92 +/ 2 126 +/ 4 150 94 150 94 +/ 2 116 +/ 4 200 90 200 90 +/ 2 109 +/ 4 250 92 250 92 +/ 2 104 +/ 4 300 100 300 100 +/ 2 108 +/ 4 350 112 350 112 +/ 2 112 +/ 2 400 100 400 100 +/ 4 100 +/ 4 450 108 450 108 +/ 4 108 +/ 4 500 122 500 122 +/ 4 122 +/ 4 600 115 600 114 +/ 4 114 +/ 4 700 128 700 112 +/ 4 112 +/ 4 750 133 800 128 800 127 +/ 4 127 +/ 4 900 131 900 93 +/ 493+/− 4
1000 131 1000 103 +/ 4 103 +/ 4 1100 126 1200 130 1200 124 +/ 4 124 +/ 4 1400 130 1500 124 1600 133 1800 133 2000 147
All resistance values are at 20 °C. The resistance changes proportionally 0.4% / °C.
Page 78
78
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Size Code No. Symbol
MAG 1100 DN 6 FDK:083G4044 Ceramic Al2O
3
DN 10 FDK:083G4046 Temperature of medium max. 150°C DN 15 FDK:083G4047 Included: DN 25 FDK:083G4049 2 EPDM gaskets, studs and nuts DN 40 FDK:083G4051
DN 50 FDK:083G4052
DN 65 FDK:083G4053
DN 80 FDK:083G4054
DN 100 1) FDK:083G4055
MAG 1100 DN 10 FDK:083G5046 PFA-liner DN 15 FDK:083G5047 Temperature of medium max. 150°C DN 25 FDK:083G5049 Included: DN 40 FDK:083G5051 2 EPDM gaskets, studs and nuts DN 50 FDK:083G5052
DN 65 FDK:083G5053
DN 80 FDK:083G5054
DN 100 1) FDK:083G5055
MAG 1100 (High temperature) DN 15 FDK:083G4057 Ceramic Al2O
3
DN 25 FDK:083G4059 Temperature of medium max. 200°C DN 40 FDK:083G4061 Included: DN 50 FDK:083G4062 2 graphite gaskets, studs and nuts DN 80 FDK:083G4064
DN 100 1) FDK:083G4065
MAG 1100 Ex DN 6 FDK:083G4024 Ceramic Al2O
3
DN 10 FDK:083G4026 Temperature of medium max. 120°C DN 15 FDK:083G4027 Included: DN 25 FDK:083G4029 2 EPDM gaskets, studs and nuts DN 40 FDK:083G4031
DN 50 FDK:083G4032
DN 65 FDK:083G4033
DN 80 FDK:083G4034
DN 100 1) FDK:083G4035
MAG 1100/3000 Ex-d DN 6 FDK:083G3054 Ceramic Al2O
3
DN 10 FDK:083G3056 Temperature of medium max. 120°C DN 15 FDK:083G3057 Included: DN 25 FDK:083G3059 2 EPDM gaskets, studs and nuts DN 40 FDK:083G3061
DN 50 FDK:083G3062
DN 65 FDK:083G3063
DN 80 FDK:083G3064
DN 100 1) FDK:083G3065
MAG 1100 FOOD DN 10 FDK:083G2016 Ceramic Al2O
3
DN 15 FDK:083G2017 Temperature of medium max. 150°C DN 25 FDK:083G2019 Enclosure IP 67 DN 40 FDK:083G2021
DN 50 FDK:083G2022
DN 65 FDK:083G2023
DN 80 FDK:083G2024
DN 100 FDK:083G2025
MAG 1100 FOOD DN 10 FDK:083G5066 PFA DN 15 FDK:083G5067 Temperature of medium max. 130°C DN 25 FDK:083G5069 Enclosure IP 67 DN 40 FDK:083G5071
DN 50 FDK:083G5072
DN 65 FDK:083G5073
DN 80 FDK:083G5074
DN 100 FDK:083G5075
10.1 Sensor MAG 1100
1
) The studs supplied for DN 100 are for PN 10/16 only. For use with PN 25/40, see "Accessories".
10. Ordering
Page 79
79
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Material DN Code No. Symbol Pipe connection 1/2” external thread AISI 316 6, 10 FDK:083G0080
2 pipe connections (1.4436) 2 gaskets EPDM 12 M4 screws (12 mm)
Grounding ring AISI 316 6, 10 FDK:083G0686 1 potential equalizing ring (1.4436) 15 FDK:083G0687 3 teflon gaskets 25 FDK:083G0689 1 earth strap 40 FDK:083G0691 1 M6 screw 50 FDK:083G0692
65 FDK:083G0693 80 FDK:083G0694
100 FDK:083G0695
Grounding ring Hastelloy C22 6, 10 FDK:083G3256 1 potential equalizing ring 15 FDK:083G3257 3 teflon gaskets 25 FDK:083G3259 1 earth strap 40 FDK:083G3261 1 M6 screw 50 FDK:083G3262
65 FDK:083G3263 80 FDK:083G3264
100 FDK:083G3265
Studs and nuts for DN 100 PN 25/40 AISI 304 100 FDK:083G0226 8 M20 studs (1.4305) 16 M20 nuts
Accessories
Description Material DN Code No. Symbol EPDM gaskets EPDM 10 FDK:083G2206
2 gaskets 15 FDK:083G2207
25 FDK:083G2209 40 FDK:083G2211 50 FDK:083G2212 65 FDK:083G2213 80 FDK:083G2214
100 FDK:083G2215
NBR gaskets NBR 10 FDK:083G2216 2 gaskets 15 FDK:083G2217
25 FDK:083G2219 40 FDK:083G2221 50 FDK:083G2222 65 FDK:083G2223 80 FDK:083G2224
100 FDK:083G2225
Description Material DN Code No. Symbol EPDM gaskets EPDM 6, 10 FDK:083G3116
2 gaskets 15 FDK:083G3117 2 earth straps 25 FDK:083G3119 3 M6 screws 40 FDK:083G3121
50 FDK:083G3122 65 FDK:083G3123 80 FDK:083G3124
100 FDK:083G3125
PTFE gaskets PTFE 6, 10 FDK:083G0156 2 PTFE gaskets 15 FDK:083G0157 2 earth straps 25 FDK:083G0159 3 M6 screws 40 FDK:083G0161
50 FDK:083G0162 65 FDK:083G0163 80 FDK:083G0164
100 FDK:083G0165
Graphite gaskets Graphite 6, 10 FDK:083G0116 2 gaskets 15 FDK:083G0117
25 FDK:083G0119 40 FDK:083G0121 50 FDK:083G0122 65 FDK:083G0123 80 FDK:083G0124
100 FDK:083G0125
Gaskets for MAG 1100 FOOD
Gaskets for MAG 1100
Page 80
80
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Adapter DoDiSensor Clamp type: Matching standard
DN DN DIN 32676 DS/ISO 2852 SMS 3016 BS 4825-3 Tri-Clamp
[mm] [mm] [mm] [mm] Code No. Code No. Code No. Code No. Code No.
10 34.0 10.0 10 FDK:083G2186 FDK:083G2186 FDK:083G2186 - FDK:083G2286 15 34.0 16.0 15 FDK:083G2187 FDK:083G2187 FDK:083G2187 - FDK:083G2287 20 34.0 20.0 15 FDK:083G2188 FDK:083G2188 ---
25 50.5 22.6 25 - FDK:083G2189 FDK:083G2189 FDK:083G2189 FDK:083G2289 25 50.5 26.0 25 FDK:083G2179 FDK:083G2179 ---
33.7 50.5 31.3 25 FDK:083G2190 FDK:083G2190 FDK:083G2190 --
38 50.5 35.6 40 - FDK:083G2191 FDK:083G2191 FDK:083G2191 FDK:083G2291 40 50.5 38.0 40 FDK:083G2181 FDK:083G2181 --­50 64.0 50.0 50 FDK:083G2182 ----
51 64.0 48.6 50 FDK:083G2192 FDK:083G2192 FDK:083G2192 FDK:083G2292
63.5 77.5 60.3 65 - FDK:083G2193 FDK:083G2193 FDK:083G2193 FDK:083G2293 65 91.0 66.0 65 FDK:083G2183 ----
76.1 91.0 72.9 80 - FDK:083G2194 FDK:083G2194 FDK:083G2194 FDK:083G2294 80 106.0 81.0 80 FDK:083G2184 ----
100 119.0 100.0 100 FDK:083G2185 ---FDK:083G2295
101.6 119.0 97.6 100 - FDK:083G2195 ---
Adapter DoDiSensor Weld-in type: Matching standard
DN DN DIN 11850 DS/ISO 2037 SMS 3008 BS 4825-1 Tri-Clover
[mm] [mm] [mm] [mm] Code No. Code No. Code No. Code No. Code No.
10 13.0 10.0 10 FDK:083G2116 FDK:083G2116 FDK:083G2116 FDK:083G2116 FDK:083G2276 15 19.0 16.0 15 FDK:083G2117 FDK:083G2117 FDK:083G2117 -
15.9 15.9 13.5 15 - - - - FDK:083G2277
20 23.0 20.0 15 FDK:083G2118 FDK:083G2118 FDK:083G2118 ­25 25.6 22.6 25 - FDK:083G2109 FDK:083G2109 FDK:083G2109 FDK:083G2279 25 29.0 26.0 25 FDK:083G2119 ----
28 28.6 25.6 25 - FDK:083G2100 --­32 32.0 29.6 25 - - FDK:083G2102 -­32 35.0 32.0 25 FDK:083G2120 ----
33.7 34.3 31.3 25 - FDK:083G2110 FDK:083G2110 -­38 38.6 35.6 40 - FDK:083G2111 FDK:083G2111 FDK:083G2111 FDK:083G2281 40 40.6 37.6 40 - FDK:083G2101 ---
40 41.0 38.0 40 FDK:083G2121 ---­50 53.0 50.0 50 FDK:083G2122 ---­51 51.6 48.6 50 - FDK:083G2112 FDK:083G2112 FDK:083G2112 FDK:083G2282
63.5 64.1 60.3 65 - FDK:083G2113 FDK:083G2113 FDK:083G2113 FDK:083G2283 65 70.0 66.0 65 FDK:083G2123 ---­70 70.6 66.8 65 - FDK:083G2103 ---
76 76.0 72.0 65 - - FDK:083G2106 --
76.1 76.7 72.9 80 - FDK:083G2114 FDK:083G2114 FDK:083G2114 FDK:083G2284 80 85.0 81.0 80 FDK:083G2124 ----
88.9 88.9 84.9 80 - FDK:083G2104 FDK:083G2104 --
100 104.0 100.0 100 FDK:083G2125 ---FDK:083G2285
101.6 102.5 97.6 100 - FDK:083G2115 ---
114.3 115.6 110.3 100 - FDK:083G2105 ---
10.2 Adapter, MAG 1100 FOOD (contains 2 adapters, 2 clamp rings and 2 gaskets)
Symbol
Symbol
Page 81
81
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Adapter D
o
DiSensor Threaded type: Matching standard
DN DN ISO 2853 SS 3351 BS 4825-4 (IDF)
[mm] [mm] [mm] [mm] Code No. Code No. Code No.
25 37.0 22.6 25 FDK:083G2149 FDK:083G2149 FDK:083G2149 38 51.0 35.6 40 FDK:083G2151 FDK:083G2151 FDK:083G2151 51 64.0 48.6 50 FDK:083G2152 FDK:083G2152 FDK:083G2152
63.5 78.0 60.3 65 FDK:083G2153 FDK:083G2153 FDK:083G2153
76.1 91.0 72.9 80 FDK:083G2154 FDK:083G2154 FDK:083G2154
101.6 118.0 97.6 100 FDK:083G2155 FDK:083G2155 -
101.6 126.0 97.6 100 - - FDK:083G2145
Threaded type: Matching standard
Adapter D
o
D
i
Sensor DIN 11851
DN DN
[mm] [mm] [mm] [mm] Code No.
10 28.0 10.0 10 FDK:083G2156 15 34.0 16.0 15 FDK:083G2157 20 44.0 20.0 15 FDK:083G2158
25 52.0 26.0 25 FDK:083G2159 32 58.0 32.0 25 FDK:083G2160 40 65.0 38.0 40 FDK:083G2161
50 78.0 50.0 50 FDK:083G2162 65 95.0 66.0 65 FDK:083G2163 80 110.0 81.0 80 FDK:083G2164
100 130.0 100.0 100 FDK:083G2165
Symbol
10.2 Adapter, MAG 1100 FOOD
(continued)
Symbol
Symbol
Threaded type: Matching standard
Adapter D
o
D
i
Sensor SMS 1145
DN DN
[mm] [mm] [mm] [mm] Code No.
25 40.0 22.6 25 FDK:083G2139 32 48.0 29.6 25 FDK:083G2140 38 60.0 35.6 40 FDK:083G2141
51 70.0 48.6 50 FDK:083G2142
63.5 85.0 60.3 65 FDK:083G2143 76 98.0 72.0 65 FDK:083G2144
Tri-Clover and Tri-Clamp are registered trademarks for Ladish Co.
Page 82
82
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Type No.: MAG 3100
−−
−−
1. Nominal size DN
15 ......................................................................... 04
25 ......................................................................... 06
40 ......................................................................... 08
50 ......................................................................... 09
65 ......................................................................... 10
80 ......................................................................... 11
100 ......................................................................... 12
125 ......................................................................... 13
150 ......................................................................... 14
200 ......................................................................... 15
250 ......................................................................... 16
300 ......................................................................... 17
350 ......................................................................... 18
400 ......................................................................... 19
450 ......................................................................... 30
500 ......................................................................... 20
600 ......................................................................... 21
700 ......................................................................... 22
750 ......................................................................... 34
800 ......................................................................... 23
900 ......................................................................... 24
1000 ......................................................................... 25
1050 ......................................................................... 35
1100 ......................................................................... 26
1200 ......................................................................... 27
1400 ......................................................................... 28
1500 ......................................................................... 29
1600 ......................................................................... 31
1800 ......................................................................... 32
2000 ......................................................................... 33
2. Liner material
Neoprene ............................................................................. 1
EPDM ................................................................................... 2
PTFE (DN 300, PN ≤ 50 bar), ........................................... 3
PTFE (DN ≥ 350 ≤ 600, PN ≤ 40 bar)
Ebonite ................................................................................ 6
Linatex
(PN 40 bar) ....................................................... 7
3. Flanges
EN 1092-1, PN 6 (DN 65 - 2000) ................................................... A
EN 1092-1, PN 10 (DN 200 - 2000) ............................................... B
EN 1092-1, PN 16 (DN 65 - 2000) conform to PED 97/23/EC ...... C
EN 1092-1, PN 16 (DN 700 - 2000) not conform to PED 97/23/EC W
EN 1092-1, PN 25 (DN 200 - 600) ................................................. D
EN 1092-1, PN 40 (DN 15 - 600) .................................................... E
EN 1092-1, PN 63 (DN 50 - 400) not with PTFE ........................... Q
EN 1092-1, PN 100 (DN 25 - 350) not with PTFE ......................... R
ANSI class 150 (DN 15 - 600, 20 bar) ........................................... F
ANSI class 300 (DN 15 - 600, 50 bar) ........................................... G
AS 2129, Table E (DN 15 - 1200, 14 bar) ...................................... N
AWWA C-207, class D (DN 700 - 2000, 10 bar) ........................... O
AS 4087 Class 14 (DN 50 - 1200, 14 bar) ..................................... P
AS 4087 Class 21 (DN 50 - DN 600, 21 bar) ................................. T
AS 4087 Class 35 (DN 50 - DN 600, 35 bar) ................................. V
4. Electrode material/PE electrodes (not PTFE)
AISI 316 Ti, ceramic coated (PE electrodes not coated) .......................... 1
AISI 316 Ti ................................................................................................... 2
Hastelloy C-276 ........................................................................................... 4
Platinum/iridium (80/20) .............................................................................. 5
Titanium ....................................................................................................... 6
Tantalum ..................................................................................................... 8
5. Flange and housing material
Carbon steel (grey coating) .................................................................................... 1
AISI 304 flanges/carbon steel (grey coating), ....................................................... 2
AISI 316 L flanges/housing (polished) .................................................................... 3
6. Temperature/Ex-version
Standard ............................................................................................................................................. 1000
High temperature (PTFE only, DN 300 mm, PN 50 bar) .............................................................. 2100
EEx e ia IIC T4-T6 (DN 350 - 2000) .................................................................................................... 4000
EEx ia ib IIB T4-T6 (DN 15 - 300) ....................................................................................................... 4100
Compact EEx de (DN 15 - 300) ......................................................................................................... 5000
10.3 Sensor MAG 3100 and MAG 3100 Ex
Page 83
83
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Earthing/protection flange type E (AISI 316) for PTFE liner only
Accessories
Description Code No. Symbol
Submersible kit (IP 68, 10 m w.g.) For use with standard MAG 3100 when sensor is buried FDK:085U0220 or permanently submerged.
Earthing/protection flange type C (AISI 304) for all liners except PTFE for MAG 3100 and 3100 W only
Flange
PN 6 PN 10 PN 16 PN 25 PN 40 150 lb 300 lb
FDK: FDK: FDK: FDK: FDK: FDK: FDK:
DN 083N... 083N... 083N... 083N... 083N... 083N... 083N...
25 8361 8361 8361 40 8362 8362 8362
50 8344 8344 8344 65 8345 8345 8345 8345 8345 80 8347 8347 8347 8347 8347 100 8070 8025 8025 8025 8025 125 8071 8071 8071 8071 8071 150 8072 8008 8008 8008 8073 200 8074 8011 8011 8011 8075 8011 8076 250 8078 8013 8013 8013 8079 8013 8079 300 8080 8012 8012 8081 8082 8012 8082 350 8083 8039 8039 8084 8085 8039 8085 400 8099 8100 8100 8101 8102 8100 8102 450 8103 8103 8104 8104 8105 8104 8106 500 8107 8107 8108 8108 8109 8107 8110 600 8111 8111 8112 8112 8113 8114 700 8300 8294 8294 800 8303 8304 8304 900 8306 8307 8307 1000 8309 8310 8310 1200 8312 8313 8313
1400 8349 8353 8357 1600 8350 8354 8358 1800 8351 8355 8359 2000 8352 8356 8360
ANSI B 16.5EN 1092-1
Pressure
stage
Flange
PN 6 PN 10 PN 16 PN 25 PN 40 150 lb 300 lb
stage FDK: FDK: FDK: FDK: FDK: FDK: FDK:
DN 083N... 083N... 083N... 083N... 083N... 083N... 083N...
15 8365 8365 8365 25 8271 8272 8272 40 8278 8275 8275 50 8282 8283 8283 65 8284 8285 8286 8287 8287 80 8288 8289 8290 8291 8292 100 8116 8117 8118 8118 8119 125 8120 8121 8122 8122 8123 150 8124 8125 8126 8126 8127 200 8129 8130 8130 8131 8132 8370 8133 250 8135 8136 8137 8138 8139 8140 8141 300 8144 8144 8145 8146 8147 8148 8149 350 8152 8153 8154 8155 8156 8157 8158 400 8160 8161 8162 8163 8164 8165 8166 450 8168 8169 8170 8171 8172 8173 8174 500 8177 8178 8179 8180 8181 8182 8183 600 8186 8187 8188 8189 8190 8191
ANSI B 16.5EN 1092-1
Pressure
Page 84
84
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
DN PN Code No.
EN 1092-1 flanges Neoprene EPDM
liner liner
Liner:
Neoprene or EPDM 25 40 FDK:083Z8000 FDK:083Z8100
Flanges:
Carbon steel, EN 1092-1 40 40 FDK:083Z8001 FDK:083Z8101
Electrodes:
AISI 316 Ti 50 40 FDK:083Z8002 FDK:083Z8102
Grounding electrodes:
AISI 316 Ti 65 16 FDK:083Z8003 FDK:083Z8103
Enclosure:
IP 67 80 16 FDK:083Z8004 FDK:083Z8104
100 16 FDK:083Z8005 FDK:083Z8105 125 16 FDK:083Z8053 FDK:083Z8055 150 16 FDK:083Z8007 FDK:083Z8107 200 10 FDK:083Z8008 FDK:083Z8108 200 16 FDK:083Z8208 FDK:083Z8308 250 10 FDK:083Z8009 FDK:083Z8109 250 16 FDK:083Z8209 FDK:083Z8309 300 10 FDK:083Z8010 FDK:083Z8110 300 16 FDK:083Z8210 FDK:083Z8310 350 10 FDK:083Z8058 FDK:083Z8060 350 16 FDK:083Z8059 FDK:083Z8061 400 10 FDK:083Z8012 FDK:083Z8065 400 16 FDK:083Z8063 FDK:083Z8066 450 10 FDK:083Z8068 FDK:083Z8070 450 16 FDK:083Z8213 FDK:083Z8071 500 10 FDK:083Z8014 FDK:083Z8114 500 16 FDK:083Z8214 FDK:083Z8314 600 10 FDK:083Z8015 FDK:083Z8115 600 16 FDK:083Z8215 FDK:083Z8315 700 10 FDK:083Z8016 FDK:083Z8116 800 10 FDK:083Z8017 FDK:083Z8117
900 10 FDK:083Z8018 FDK:083Z8118 1000 10 FDK:083Z8019 FDK:083Z8119 1200 10 FDK:083Z8021 FDK:083Z8121
10.4 Sensor MAG 3100 W
Nominal Nominal Flange type PN Code No. ANSI/AWWA size size Neoprene EPDM flanges DN inch bar liner liner
25 1" ANSI 150 20 FDK:083Z8600 FDK:083Z8650 40 1½" ANSI 150 20 FDK:083Z8601 FDK:083Z8651 50 2" ANSI 150 20 FDK:083Z8602 FDK:083Z8652 65 2½" ANSI 150 20 FDK:083Z8603 FDK:083Z8653
80 3" ANSI 150 20 FDK:083Z8604 FDK:083Z8654 100 4" ANSI 150 20 FDK:083Z8051 FDK:083Z8052 125 5" ANSI 150 20 FDK:083Z8054 FDK:083Z8656 150 6" ANSI 150 20 FDK:083Z8056 FDK:083Z8057 200 8" ANSI 150 20 FDK:083Z8608 FDK:083Z8658 250 10" ANSI 150 20 FDK:083Z8609 FDK:083Z8659 300 12" ANSI 150 20 FDK:083Z8610 FDK:083Z8660 350 14" ANSI 150 20 FDK:083Z8611 FDK:083Z8661 400 16" ANSI 150 20 FDK:083Z8064 FDK:083Z8067 450 18" ANSI 150 20 FDK:083Z8613 FDK:083Z8663 500 20" ANSI 150 20 FDK:083Z8614 FDK:083Z8664 600 24" ANSI 150 20 FDK:083Z8615 FDK:083Z8665 700 28" AWWA 10 FDK:083Z8616 FDK:083Z8666 750 30" AWWA 10 FDK:083Z8622 FDK:083Z8672 800 32" AWWA 10 FDK:083Z8617 FDK:083Z8667 900 36" AWWA 10 FDK:083Z8618 FDK:083Z8668
1000 40" AWWA 10 FDK:083Z8619 FDK:083Z8669 1050 42" AWWA 10 FDK:083Z8673 1200 48" AWWA 10 FDK:083Z8621 FDK:083Z8671
Liner:
Neoprene/
EPDM
Flange:
Carbon steel ANSI Class 150 or AWWA C-207, Class D
Electrodes:
AISI
316 Ti
Grounding electrode:
AISI 316 Ti
Enclosure:
IP 67
Page 85
85
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
10.5 Sensor MAG 5100 W
EN 1092-1 flanges DN PN PED Code No.
Liner:
Composite elastomer or 25 40 Yes FDK:082Z8500
hard elastomer 40 40 Yes FDK:082Z8502
Flanges:
Carbon steel, EN 1092-1 50 40*) Yes FDK:082Z8504
Electrodes:
AISI 316 Ti 65 16 Yes FDK:082Z8506
Grounding electrodes:
AISI 316 Ti 80 16 Yes FDK:082Z8508
Enclosure:
IP 67 100 16 Yes FDK:082Z8510
125 16 Yes FDK:082Z8512 150 16 Yes FDK:082Z8514 200 10 Yes FDK:082Z8516
16 Yes FDK:082Z8517
250 10 Yes FDK:082Z8519
16 Yes FDK:082Z8520
300 10 Yes FDK:082Z8522
16 Yes FDK:082Z8523
350 10 Yes FDK:082Z8525
16 Yes FDK:082Z8526
400 10 Yes FDK:082Z8528
16 Yes FDK:082Z8529
450 10 Yes FDK:082Z8531
16 Yes FDK:082Z8532
500 10 Yes FDK:082Z8534
16 Yes FDK:082Z8535
600 10 Yes FDK:082Z8537
16 Yes FDK:082Z8538
700 10 Yes FDK:082Z8540
16 No FDK:082Z8541 16 Yes FDK:082Z8558
800 10 Yes FDK:082Z8544
16 No FDK:082Z8545 16 Yes FDK:082Z8559
900 10 Yes FDK:082Z8547
16 No FDK:082Z8548 16 Yes FDK:082Z8560
1000 10 Yes FDK:082Z8550
16 No FDK:082Z8551 16 Yes FDK:082Z8561
1200 10 Yes FDK:082Z8555
16 No FDK:082Z8556 16 Yes FDK:082Z8562
*) max. 20 bar
Page 86
86
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Version Enclosure Code No. Symbol
Signal converter MAG 5000 Blind 11-30 V d.c./ IP 67, fibre- for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5006
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5005
polyamide
Signal converter MAG 5000 11-30 V d.c./ IP 67, fibre- for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5002
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5001
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5011
HART
polyamide
10.6 Signal converter
Compact polyamide
Description Version Enclosure Code No. Symbol
Signal converter MAG 6000 Blind 11-30 V d.c./ IP 67, fibre-
for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5008
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5007
polyamide
Signal converter MAG 6000 11-30 V d.c./ IP 67, fibre- for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5004
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5003
polyamide
Description Version Enclosure Code No. Symbol
Signal converter MAG 6000 CT 11-30 V d.c./ IP 67, fibre- for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5010
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5009
polyamide
Accessories MAG 5000 and MAG 6000
Description Code No. Symbol
Wall mounting unit for IP 67 version
Wall bracket, 4 Pg 13.5 cable glands FDK:085U1001
Spare parts
Description Version Code No. Symbol
Connection plate 12-24 V FDK:083F4149
115-230 V FDK:083F4148
Add-on module MAG 6000 only
Description Code No. Symbol
HART


FDK:085U0226
Profibus PA FDK:085U0227 CANopen FDK:085U0228 DeviceNet FDK:085U0229 Profibus DP FDK:085U0230
Description Version Enclosure Code No. Symbol
Signal converter MAG 5000 CT 11-30 V d.c./ IP 67, fibre-
for compact and wall mounting 11-24 V a.c. glass reinforced FDK:083F5046
polyamide
115/230 V a.c. IP 67, fibre-
50/60 Hz glass reinforced FDK:083F5044
polyamide
Page 87
87
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Version Code No. Symbol
Signal converter MAG 5000 for rack 11-30 V d.c./ FDK:083F5021 and panel mounting 11-24 V a.c.
115-230 V a.c. FDK:083F5020
50/60 Hz
Signal converter MAG 6000 for rack 11-30 V d.c./ FDK:083F5023 and panel mounting 11-24 V a.c.
115-230 V a.c. FDK:083F5022
50/60 Hz
Signal converter MAG 6000
(DN ≤ 300)
11-30 V d.c./
insert with safety barrier [EEx ia/ib] IIB 11-24 V a.c. FDK:083F5041
Signal converter MAG 6000
(DN ≤ 300)
115-230 V a.c.
insert with safety barrier [EEx ia/ib] IIB 50/60 Hz FDK:083F5040
Wall boxes (Without back plates)
Description Code No. Symbol
IP 66 wall mounting enclosure for IP 20 19" inserts (21 TE) FDK:083F5037
IP 66 wall mounting enclosure for IP 20 19" inserts
(42 TE) FDK:083F5038
Description Version Code No. Symbol Cleaning unit for electrode cleaning 115-230 V a.c.
19" insert (21 TE) incl. back plate 50/60 Hz FDK:083F5036 Cleaning unit for electrode cleaning 11-30 V d.c.
19" insert (21 TE) incl. back plate 11-24 V a.c. FDK:083F5039
Safety barrier [EEx ia] IIC for MAG 3100 Ex DN 350-2000 19" insert (21 TE) incl. back plate FDK:083F5034
Panel mounting kit for 19" insert (21 TE) IP 65 enclosure in ABS plastic for panel-front mounting FDK:083F5030
Panel mounting kit for 19" insert (42 TE) IP 65 enclosure in ABS plastic for panel-front mounting FDK:083F5031
Back of panel mounting kit for 19" insert (21 TE) IP 20 enclosure in aluminium FDK:083F5032
Back of panel mounting kit for 19" insert (42 TE) IP 20 enclosure in aluminium FDK:083F5033
Front cover (7 TE) FDK:083F4525
Accessories
Description Version Code No. Symbol
MAG 6000 with IP 66 enclosure 115/230 V a.c.
50/60 Hz FDK:083F5026
MAG 6000 19" insert and cleaning unit 115/230 V a.c.
complete mounted with 50/60 Hz FDK:083F5029 IP 66 wall mounting enclosure
MAG 6000 19" insert and cleaning unit 11-30 V d.c. complete mounted with 11-24 V a.c. FDK:083F5047 IP 66 wall mounting enclosure
MAG 6000 19" insert and safety barrier
(DN ≤ 300)
115/230 V a.c. complete mounted with IP 66 50/60 Hz FDK:083F5028 wall mounting enclosure, [EEx ia/ib] IIB
MAG 6000 19" insert and safety barrier
(DN ≥ 350)
115/230 V a.c. complete mounted with IP 66 50/60 Hz FDK:083F5027 wall mounting enclosure, [EEx ia/e] IIC
Wall mounting units complete
10.7 Signal converter 19" incl. back plate
Page 88
88
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Code No. Symbol
Length
Cable 10 m FDK:083F0121 for standard electrode or coil, 20 m FDK:083F0210 3 x 1.5 mm2 with shield 40 m FDK:083F0211 PVC 60 m FDK:083F0212
100 m FDK:083F0213 150 m FDK:083F3052 200 m FDK:083F3053 500 m FDK:083F3054
Cable 20 m FDK:083F3095 required for empty pipe or low conductivity, 40 m FDK:083F3094 double shielded, 3 x 1.5 mm
2
60 m FDK:083F3093
PVC 100 m FDK:083F3092
150 m FDK:083F3056 200 m FDK:083F3057 500 m FDK:083F3058
Standard Pg 13.5 screwed cable entries for above cables (nickel-plated brass) 2 pcs. FDK:083G3140
Standard Pg 13.5 screwed cable entries for above cables in black polyamide (100°C) 2 pcs. FDK:083G0228
Sealing screws for sensor/signal converter 2 pcs. FDK:085U0221
Stainless steel (AISI 316) terminal box with lid FDK:085U1000
Polyamide terminal box FDK:085U1002
Complete incl. terminals incl. lid
Polyamide lid for terminal box FDK:085U1003
2 kB SENSORPROM


unit FDK:085U1005
(Sensor serial No. must be specified when ordering)
Adapter for ½ NPT brass FDK:083N4394
Adapter for ½ NPT stainless steel FDK:083N4395
Adapter for ½ NPT polyamide FDK:083N4396
10.8 Accessories
Back plates
Description Enclosure Version Code No. Symbol
Signal converter 19" 12-24 V FDK:083F4117
115-230 V
Signal converter & ia, safety barrier 19" 12-24 V FDK:083F4118
115-230 V
Signal converter & ia/ib, safety barrier 19" 12-24 V FDK:083F4119
115-230 V
Signal converter & cleaning unit 19" 12-24 V FDK:083F4123
115-230 V
Signal converter Wall unit 12-24 V FDK:083F4121
115-230 V
Signal converter & ia, safety barrier Wall unit 12-24 V FDK:083F4122
115-230 V
Signal converter & ia/ib, safety barrier Wall unit 12-24 V FDK:083F4120
115-230 V
Signal converter & cleaning unit Wall unit 12-24 V FDK:083F4124
115-230 V
Page 89
89
SITRANS F M MAGFLO
DKFD.PS.027.W3.02
10. Ordering
Ordering
Description Code No. Code No. Code No. Code No.
DN 3-65 DN 80-150 DN 200-500 DN 600-1200
Standard calibration
Each sensor calibrated twice at two calibration points Included in sensor scope of delivery
Matched pair
Sensor and signal converter calibrated together with standard calibration
Customer specified matched pair
Sensor with signal converter calibrated in max. 10 customer specified points
Accredited Siemens Flow Instruments matched pairs Calibrations acc. to EN 45001
Sealing and labeling instruction must follow the order
- PTB cold water
- PTB other media than water
- OIML R75 hot water
- OIML R117 other media than water
- OIML R117 cold water
Accredited Delft matched pair Calibration acc. to EN 45001
Sealing and labeling instruction as above
Wittness inspection
Any of above mentioned calibrations
10.9 Calibration
Initial calibration
Re-calibration
Description Code No. Code No. Code No. Code No.
DN 3-65 DN 80-150 DN 200-500 DN 600-1200
Matched pair
Sensor and signal converter calibrated FDK:085F7302 FDK:085F7303 FDK:085F7304 FDK:085F7305 together with standard calibration
Customer specified matched pair
Sensor with signal converter calibrated FDK:085F7377 FDK:085F7378 FDK:085F7379 FDK:085F7380 in max. 10 customer specified points
Accredited Siemens Flow Instruments matched pairs Calibrations acc. to EN 45001
Sealing and labeling instruction must follow the order
- PTB cold water FDK:085F7387 FDK:085F7388 FDK:085F7389 FDK:085F7390
- PTB other media than water
- OIML R75 hot water
- OIML R117 other media than water
- OIML R117 cold water
Accredited Delft matched pair Calibration acc. to EN 45001 N/A FDK:085F7393 FDK:085F7394 FDK:085F7395
Sealing and labeling instruction as above
Add-on
Code No.: On application form to be filled in and sent to SFI-GB
Page 90
90 DKFD.PS.027.W3.02
SITRANS F M MAGFLO
Page 91
DKFD.PS.027.W3.02
SITRANS F M MAGFLO
91
Page 92
We have checked the contents of this manual for agreement with the hardware and software described. Since deviations cannot be precluded entirely, we cannot guarantee full agreement. However, the data in this manual are reviewed regularly and any necessary corrections included in subsequent editions. Suggestions for improvement are always welcomed.
Technical data subject to change without prior notice.
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
Copyright © Siemens AG 2002 All Rights Reserved
SIEMENS Flow Instruments A/S Nordborgvej 81 DK-6430 Nordborg