Siemens A60 User Manual 2

SOMATOM Sensation Cardiac
Application Guide
Software Version A60
2
The information presented in this application guide is for illustration only and is not intended to be relied upon by the reader for instruction as to the practice of medicine. Any health care practitioner reading this information is reminded that they must use their own learning, training and expertise in dealing with their individual patients.
This material does not substitute for that duty and is not intended by Siemens Medical Solutions Inc., to be used for any purpose in that regard. The drugs and doses mentioned herein were specified to the best of our knowledge. We assume no responsibility what is over for the correctness of this information. Variations may prove necessary for individual patients.
The treating physician bears the sole responsibility for all of the parameters selected. The pertaining oper­ating instructions must always be strictly followed when operating the SOMATOM Sensation Cardiac. The statutory source for the technical data are the corresponding data sheets. To improve future versions of this application guide, we would highly appreciate your questions, suggestions and comments.
Please contact us: CT Application Hotline: Tel. no. +49-9191-18 80 88 (outside Germany)
0130-18 53 91 (in Germany)
Fax no. +49-9191-18 99 98
email: somatom.examinationprotocol@siemens.com
We express our sincere gratitude to the many customers who contributed valuable input.
In addition I would like to stress at this point that the SOMATOM Sensation Cardiac Application Guide is based very much on the other SOMATOM Application Guides which have been written by Dr. Xiao-Yan Chen, Loke-Gie Haw and Bettina Klingemann.
Special thanks to Dr. med. Martin Heuschmid, Universitätsklinikum Tübingen.
Editor: Ute Feuerlein
3
3
Overview
General 8
HeartView CT 22
Bolus Tracking 54
WholeBodyAngio 62
LungCARE 64
CT Colonography 66
Children 68
Head 88
Neck 98
Thorax 102
Abdomen 112
Extremities 124
4
Content
General 8
· Concept 8
· Scan Set Up 8
· Scan and Reconstruction 9 – Slice Collimation and Slice Width 9 – Pitch 10 – Recon Job 11 – Kernels 11 – Image Filter 11
· Effective mAs 12
· Dose Information 13
· CARE Dose 16
· How does it work 16
· Workflow 18
· Contrast Medium IV Injection 19
· How to Create your own Scan Protocols 20
5
Content
HeartView CT 22
· The Basics 22 – Important Anatomical Structures
of the Heart 22 – Cardiac Cycle and ECG 25 – Temporal Resolution 25 – Technical Principles 26 – Preview Series Reconstruction 27 – ECG Trace Editor 28 – ECG Pulsing 28 – CardioCARE 29 – CardioSharp 29
· How to do it 30 – Calcium Scoring 30 – Hints in General 30 – Placement of ECG Electrodes 31
· CaScoreSpiStd 32
· CaScoreSpiStd0.5s 33
· CaScoreSeqStd 34 – Coronary CTA 35 – General Hints 36
· CoronaryStd 37
· CoronaryStd0.5s 38
· CorStd_LowHeartRate 39
· CoronaryCARE 40
· CoronaryCARE0.5s 41
· CoronarySharp 42
· CoronarySharp0.5s 44
· ECGTrigCTA 46 – Aortic and Pulmonary Studies 48 – General Hints 48
· PulmonaryECG 49
· Additional Important Information 50
6
Content
Bolus Tracking 54
· The Basics 54
· How to do it 56
· CARE Bolus 56 – General Hints 56
· Test Bolus 58 – Application Procedures 59
· Additional Important Information 60
WholeBodyAngio 62
LungCARE 64
CT Colonography 66
Children 68
· Hints in General 69
· HeadSpi05s 72
· HeadSeq0.5s 74
· HeadAngio 76
· CarotidCTA 78
· NeonateBody 80
· ChestRoutine 82
· AbdPelRoutine 84
· AbdCTA 86
Head 88
· Overview 88
· Hints in General 89
· Angio Head 90
· AngioThinSlice 92
· HeadSeq0.5s 94
· HeadSpi0.5s 96
Neck 98
· Overview 98
· Hints in General 99
· AngioCarotid 100
7
Content
Thorax 102
· Overview 102
· Hints in General 103
· LungLowDose 104
· PulmonaryEmboli 106
· ThoraxCombi 108
· ThoraxRoutine 110
Abdomen 112
· Overview 112
· Hints in General 113
· AbdRoutine 116
· AngioFast 118
· AngioRoutine 120
· AbdomenSeq 122
Extremities 124
· Overview 124
· Hints in General 125
· AngioRunOff 126
8
General
Concept
The scan protocols for adult are defined according to body regions – Head, Neck,Thorax, Abdomen,
Extremities.
The pediatric scan protocols are defined under the folder “Children”.
The protocols for special applications are defined under “Special”.
The general concept is as follows: “Fast”: uses 1.5 mm slice collimation and a higher pitch for fast acquisition for CT Angios or trauma cases. “Routine”: uses either 0.75 mm or 1.5 mm slice collimation depending on the region of interest for routine studies. “ThinSlice”: uses 0.75 mm slice collimation for thin slice studies. “Combi”: uses 0.75 mm as slice collimation and reconstruct images as both thinner slice width for HiRes or CTA and thicker slice width for soft tissue studies. “Seq”: stands for Sequence. E. g. “HeadSeq” means the sequence mode for the head; “Spi” stands for “Spiral”. E. g. “HeadSpi” means the spiral mode for the head.
Scan Set Up
Scans can be simply set up by selecting a predefined examination protocol. To repeat any mode, just click the chronicle with the right mouse button for “repeat”. To delete it, select “cut”. Each chronicle is rewriteable before “load”.
Multiple ranges can be run either automatically with “auto range”, or separately with a “pause” in between.
9
General
Scan and Reconstruction
Slice Collimation and Slice Width
Slice collimation is the slice thickness collimated by the tube collimator, which determines the Z-coverage per rotation. In Multislice CT, this is divided by the number of active detector channels (e.g. 16 x 0.75 mm).
Slice width is the true thickness of the reconstructed image.
With the SOMATOM Sensation Cardiac, you select the slice collimation together with the slice width desired, the slice width is independent of pitch and algorithm, i. e. what you select is always what you get. Actually, you do not need to care about the algorithm any more; the software does it for you.
On the SOMATOM Sensation Cardiac some slice widths are marked as “fast” (blue background). These images will be reconstructed with highest performance (up to 6-10 images per second). All others will be recon­structed with up to 3 images per second.
The reconstruction time depends on slice collimation and the reconstructed slice width. To get the fast performance, slice width has to be at least 3 times the slice collimation.
During scanning the user normally will get “real time” reconstructed images in full image quality, if the “fast” slice has been selected.
In some cases – this depends also on Scan range, Feed/ Rotation and Reconstruction increment – the Recon icon on the chronicle will be labeled with “RT”. This indi­cates the real time display of images during scanning. The real time displayed image series has to be recon­structed afterwards.
10
General
The following tables show you the possibilities of image reconstruction in spiral and sequential scanning.
Slice Collimation and Slice Width for Spiral Mode
0.75 mm 0.75, 1, 1.5, 2, 3, 4, 5, 6, 7, 8, 10 mm
1.5 mm 2, 3, 4, 5, 6, 7, 8, 10 mm
Cardio Spiral Modes
0.75 mm 0.75, 1.0, 1.5, 2,3 mm
1.5 mm 2, 3, 4, 5 mm
Slice Collimation and Slice Width for Sequence Mode
0.75 mm 0.75, 1.5, 3, 4.5, 9 mm
1.5 mm 1.5, 3, 4.5, 6, 9 mm 5 mm 5, 10 mm
ECG triggered Modes
0.75 mm 0.75, 1.5, 3 mm
1.5 mm 1.5, 3, 6 mm
Pitch
In single slice CT: Pitch = table movement per rotation/slice collimation E. g.: slice collimation = 5 mm, table moves 5 mm per rotation, then pitch = 1.
With the SOMATOM Sensation Cardiac, in Siemens Multislice CT, we differentiate between:
Feed/Rotation, the table movement per rotation Volume Pitch, the table movement per rotation/single
slice collimation E. g.: single slice collimation = 1.5 mm, table moves 24 mm per rotation, then the Volume Pitch = 16 Pitch Factor, the table movement per rotation/ collimation E. g.: slice collimation = 16 x 1.5 mm, table moves 24 mm per rotation, then the Pitch Factor = 1
With the SOMATOM Sensation Cardiac, you do not need to select pitch. Once the scan range, scan time, slice collimation, and rotation time is defined, the software will adapt the table feed per rotation accor­dingly. The Pitch Factor can be freely adapted from
0.5 to 1.5.
11
General
Recon Job
In the Recon card, you can define up to 3 reconstruc­tion jobs with different parameters either before or after you acquire the data. When you click on “Recon”, they will all be done automatically. In case you want to add another recon job, simply click the little icon on the chronicle with the right mouse button and select “delete recon job” to delete the one which has been completed, and then one more recon job will be avail­able in the Recon card (Note: what you delete is just the job from the display, not the images that have been reconstructed).
You can also reconstruct images for all scan modes completed at once – do not select any chronicle before you click “Recon”.
Kernels
There are 5 different types of kernels: “H” stands for Head, “B” stands for Body and “C” stands for ChildHead. The image sharpness are defined by the numbers – the higher the number, the sharper the image; the lower the number, the smoother the image. Note: Do not use different kernels for other body parts than designed for.
Image Filter
There are 4 different filters available: PFO: To reduce beam-hardening artifacts in head images, particularly in the base of the skull, use the Posterior Fossa Optimization (PFO) filter.
ASA: The Advanced Smoothing Algorithm (ASA) filter reduces noise in soft tissue while edges with high contrast are preserved.
LCE: The Low-contrast enhancement (LCE) filter enhances low-contrast detectability. It reduces the image noise.
HCE: The High-contrast enhancement (HCE) filter enhances high-contrast detectability. It increases the image sharpness.
12
General
Effective mAs
In sequential scanning, the dose (D
seq
) applied to the patient is the product of the tube current-time (mAs) and the CTDI
w
per mAs:
D
seq
= D
CTDI
w
x mAs
In spiral scanning, however, the applied dose (D
spiral
) is influenced by the mAs (mA x Rot Time) and in addi­tion by the Pitch Factor. For example, if a Multislice CT scanner is used, the actual dose applied to the patient in spiral scanning will be decreased when the Pitch Factor is larger than 1, and increased when the Pitch Factor is smaller than 1. Therefore, the dose in spiral scanning has to be corrected by the Pitch Factor:
D
spiral
= (D
CTDI
w
x mA x Rot Time)/Pitch Factor
To make it easier for the users, the concept of the effective mAs was introduced with the SOMATOM Multislice scanners.
The effective mAs takes into account the influence of pitch on both the image quality and dose:
Effective mAs = mAs/Pitch Factor To calculate the dose on the SOMATOM Sensation
Cardiac, you simply have to multiply the CTDI
w
per mAs
with the effective mAs of the scan: D
spiral
= D
CTDI
w
x effective mAs
For spiral scan protocols, the indicated mAs is the effective mAs per image. The correlation between tube current mA and effective mAs of spiral scans on a Multi­slice CT scanner is given by the following formula:
Effective mAs = mA x RotTime/Pitch Factor
Pitch Factor =
Feed/Rot
nrow x Slice collimation
mA =
effective mAs
x Pitch Factor
RotTime
where collimated Slice refers to the collimation of one detector row, and nrow is the number of used detector rows.
13
General
Dose Information
The dose as described by CTDIwis displayed on the user interface for the selected scan parameters. The CTDI
w
is measured in the dedicated plastic phantoms – 16 cm diameter for head and 32 cm diameter for body (as defined in IEC 60601-2-44). This dose number gives a good estimate for the average dose applied in the scanned volume as long as the patient size is similar to the size of the respective dose phantoms.
Since the body size can be smaller or larger than 32 cm, the CTDI value displayed can deviate from the dose in the scanned volume.
The CTDI
w
value does not provide the entire information of the radiation risk associated with CT examination. For the purpose, the concept of the “Effective Dose“ was introduced by ICRP (International Commission on Radiation Protection). The effective dose is expressed as a weighted sum of the dose applied not only to the organs in the scanned range, but also to the rest of the body. It could be measured in whole body phantoms (Alderson phantom) or simulated with Monte Carlo techniques.
The calculation of the effective dose is rather com­plicated and has to be done by sophisticated programs. These have to take into account the scan parameters, the system design of individual scanner, such as x-ray filtration and gantry geometry, the scan range, the organs involved in the scanned range and the organs affected by scattered radiation. For each organ, the respective dose delivered during the CT scanning has to be calculated and then multiplied by its radiation risk factor. Finally the weighted organ dose numbers are added up to get the effective dose.
14
General
Fig. 1: User interface of the PC program WinDose. All parameters necessary for the effective dose calcu­lation have to be specified.
** The Calculation of Dose from External Photon Expo-
sures Using Reference Human Phantoms and Monte Carlo Methods. M. Zankl et al. GSF report 30/91
** Radiation Exposure in Computed Topography, edited
by Hans Dieter Nagel, published by COCIR c/o ZVEI, Stresemannallee 19, D-60596, Frankfurt, Germany.
The concept of effective dose would allow the com­parison of radiation risk associated with different CT or x-ray exams, i. e. different exams associated with the same effective dose would have the same radiation risk for the patient. It also allows comparing the applied x-ray exposure to the natural background radiation, e. g. 2 – 3 mSv per year in Germany.
For most of our scan protocols, we calculated the effective dose numbers for standard male* and female* and listed the result in the description of each scan protocol.
The calculation was done by the commercially avail­able program “WinDose“ (Wellhoefer Dosimetry) – as shown in figure 1-3. For pediatric protocols, we used the WinDose calculation and the correction factors published in “Radiation Exposure in Computed Topo­graphy“**, in which there only the conversion factors for the age of 8 weeks and 7 years old are available.
15
General
Fig. 2: A graphic interface of WinDose allows to specify the anatomical scan range.
Fig.3: Results as output of WinDose with the organ dose readings and the effective dose according to ICRP26 (previous version) and ICRP60 (currently valid).
Please note: Effective dose values were calculated with the WinDose™ Software Version 2.1, which was optimized for SOMATOM Volume Zoom scanners.
SOMATOM Sensation Cardiac has modified shaped filters which might cause slightly different effective dose numbers ( difference not expected to be more than 5%). An update of WinDose™ for the SOMATOM Sensation Cardiac is currently developed.
16
General
Fig. 1: Example of scanning in the region of shoulder.
CARE Dose
CARE Dose is a clinical application package that pro­vides real-time tube current modulation for Spiral and Sequential Scanning.
CARE Dose reduces patient dose significantly, espe­cially in the regions of shoulder and pelvis. It decreases tube load, which extends the capacity for volume scanning with thinner slices, larger volumes or Multi­phase studies.
It can also improve image quality by increasing mA and thus reducing image noise on the lateral views.
How does it work
It reduces the mA for low attenuation views up to 90% and keeps the nominal higher mA for high attenuation views, e.g. in the lateral projection (Fig. 1). This is done “on-the-fly”, i.e. the scanner adapts the mA in real-time, according to the patient’s attenuation profile (Fig. 2).
High attenuation, high mA
Low attenuation, low mA
17
General
Fig. 2: Principle of CARE Dose tube current adaptation.
• CARE Dose is pre-selected by default for all standard
protocols, except head and cardiac protocols. It can be switched on/off in the scan card.
• The application of CARE Dose does not require any
changes in the scan parameters. The mean value of the mAs applied will be lower than what you have selected.
• The mean value of the effective mAs applied is shown
in the image text.
lateral
1.00
0.75
0.50
0.25
rel. units
a.p.
time
Object attenuation Modulated tube current
18
General
Workflow
Auto Load in 3D and Postprocessing Presets
You can activate the “Auto load in 3D” function on the Examination Card/Auto Tasking and link it to a recon job. For example, the 2
nd
recon job with thinner slice width in some of the examination protocols. If the post­processing type is chosen from the pull down menu, the reconstructed images will be loaded automatically into the 3D Card on the Navigator with the corres­ponding postprocessing type.
On the 3D Card you have the possibility to create for MPR and MIPthin Range Parallel and Radial protocols which can be linked to a special series.
For example, if you always do some sagittal Multipla­nar Reconstructions for a Spine examination, you load once a Spine examination into the 3D Card. Select the image type (MPR, MIPthin), select the orientation and open the Range Parallel function. Adapt the range settings (Image thickness, Distance between the images etc.) and hit the link button. From now on, you have a predefined postprocessing protocol, linked to the series description of a Spine examination.
Exactly the same can be done for VRT presets. In the main menu, under Type/VRT Definition, you can link VRT presets with a series description.
19
General
Some of the Scan protocols, mainly for Angio exami­nations, are already delivered with Auto load in 3D. If you do not like, please deselect the Auto load in 3D and save your scan protocol.
Some of the Scan protocols are delivered with links to a postprocessing protocol. If you do not like our suggestions, please delete the Range Parallel preset or overwrite them with your own settings.
Contrast Medium IV Injection*
The administration of a contrast medium depends on the indication and on the delay times to be used during the examination. The patients weight and circulatory situation also play a role. In general, no more than 3 ml per kg of body weight for adults and 2 ml per kg of body weight for children should be applied.
As a rule of thumb, the contrast medium injection should be stopped when the scan (or acquisition) is finished. Keep this in mind, you may save contrast medium in your routine study since the Multislice spiral scan can be up to 32 times faster than a 1 second, single slice spiral scan.
For CTA study (arterial phase), the principle is to keep contrast injection for the whole scan. Thus, the total amount of contrast medium needed should be calcu­lated with the following formula:
CM = (start delay time + scan time) x flow rate. CARE Bolus or Test Bolus may be used for optimal
contrast bolus timing. Please refer to the Application Guide for special protocols.
* For more information regarding the general use
of drugs and doses mentioned in this guide, please refer to page 2.
20
General
How to Create your own Scan Protocols
User-specific scan protocols can be saved with the following basic procedure:
– Register a test patient, patient position must be
head-supine. – Select a scan protocol. – Set the table position to 0 (either at the gantry panel,
or on the topogram routine card). – Modify the scan protocol, change parameters,
add new ranges etc. – Do not load the scan protocol. – Select Edit/Save Scan Protocol in the main menu. – Select the organ region and the scan protocol name
in the pop-up dialog. You can either use the same
name to modify the existing scan protocol, or enter
a new name.
21
General
Tips:
– It is recommended that you save your own scan
protocol with a new name in order to avoid over­writing the default scan protocol.
– You may use preceding numbers (e.g. 1_Abdomen)
for user specific scan protocols to make them appear on top of the list and to distinguish them from the Siemens defaults.
– Do not use special characters like “/“, “.“ or “\“ within
scan protocol names.
– Don’t rename scan protocol files on Windows NT
level – this will lead to inconsistencies.
– Do not mix head and body scan protocols: e. g. do
not save a head mode in the abdomen directory.
– System/Run offers the tool “Restore Default Scan
Protocols“ which allows one to remove user specific scan protocols and to restore the Siemens default
settings. – System/Edit offers save/delete Scan Protocols. – System/Run also offers the tool “List Scan Protocols”
which generates an HTML table of all available scan
protocols. This list can be printed or saved on Floppy
(“File/Save As…”).
22
HeartView CT
HeartView CT
HeartView CT is a clinical application package specifically tailored to cardiovascular CT studies.
The Basics
Important Anatomical Structures of the Heart
Four chambers:
• Right atrium – receives the deoxygenated blood back from the body circulation through the superior and inferior vena cava, and pumps it into the right ventricle
• Right ventricle – receives the deoxygenated blood from the right atrium, and pumps it into the pulmo­nary circulation through the pulmonary arteries
• Left atrium – receives the oxygenated blood back from the pulmonary circulation through the pulmo­nary veins, and pumps it into the left ventricle
• Left ventricle – receives the oxygenated blood from the left atrium, and pumps it into the body circulation through the aorta.
23
HeartView CT
Fig. 1: Blood fills both atria
Fig. 2: Atria contract, blood enters ventricles
A: Aorta P: Pulmonary Artery RV: Right Ventricle LV: Left Ventricle RA: Right Atrium LA: Left Atrium
Fig. 3: Ventricles contract, blood enters into aorta and pulmonary arteries
RA
LA
A
P
RV
LV
24
HeartView CT
Coronary arteries:
• Right coronary artery (RCA) Right coronary artery supplies blood to the right atrium, right ventricle, a small part of the ventricular septum.
• Left coronary artery (LCA) Left coronary artery supplies blood to the left atrium, left ventricle and a large part of the ventricular septum.
SVC:Superior Vena Cava IVC: Inferior Vena Cava RA: Right Atrium RV: Right Ventricle A: Aorta PA: Pulmonary Artery
Fig. 4: Front view Fig. 5: Conventional
Angiography
LM: Left Main Artery LAD: Left Anterior Descending Artery Cx: Circumflex Artery
Fig. 6: Front view Fig. 7: Conventional
Angiography
RV
PA
RA
A
IVC
SVC
LM
Cx
LAD
25
HeartView CT
Cardiac Cycle and ECG
The heart contracts when pumping blood and rests when receiving blood. This activity and lack of activity form a cardiac cycle, which can be illustrated by an Electrocardiograph (ECG) (Fig. 8).
To minimize motion artifacts in cardiac images, the following two requirements are mandatory for a CT system:
• Fast gantry rotation time in order to achieve fast image acquisition time
• Prospective synchronization of image acquisition or retrospective reconstruction based on the ECG recording in order to produce the image during the diastolic phase when the least motion happens.
Temporal Resolution
Temporal resolution, also called time resolution, represents the time window of the data that is used for image reconstruction. It is essential for cardiac CT imaging – the higher the temporal resolution, the fewer the motion artifacts. With the SOMATOM Sensation Cardiac, temporal resolution for cardiac imaging can be achieved at down to 105 ms.
Fig. 8
R
T
U
S
Q
P
Ventricular contraction
Systolic phase
Atrial contraction
Diastolic phase
Relaxation
Diastolic phase
26
HeartView CT
Technical Principles
Basically, there are two different technical approaches for cardiac CT acquisition:
• Prospectively ECG triggered sequential scanning.
• Retrospectively ECG gated spiral scanning. In both cases, an ECG is recorded and used to either
initiate prospective image acquisition (ECG triggering), or to perform retrospective image reconstruction (ECG gating). A given temporal relation relative to the R-waves is predefined and can be applied with the following possibilities:
Relative – delay: a given percentage of R-R interval (_RR) relative to the onset of the previous or the next R-wave (Fig. 9, 10).
Time
Time
ECG (t)
ECG (t)
50 % of R-R
-50 % R-R
Scan/
Recon
Scan/
Recon
Fig. 9
Fig. 10
27
HeartView CT
Absolute – delay: a fixed time delay after the onset of the R-wave (Fig. 11).
Absolute – reverse: a fixed time delay prior to the onset of the next R-wave (Fig. 12).
Preview Series Reconstruction
Preview series can be used to define the optimal time window before the full series is reconstructed.
Click on the preview series button in the Trigger card. The slice position of the preview series is based on the currently displayed image in the tomogram segment, which has to be chosen by the user.
Fig. 11
Time
Estimated R-Peak
ECG (t)
-400 msec
Scan/
Recon
Fig. 12
Time
ECG (t)
400 msec
Scan/
Recon
28
HeartView CT
ECG Trace Editor
The ECG trace editor is used for adaptation of image reconstruction to irregular heart rates. This editing tool can be used after the scan is acquired. By using the right mouse menu on the Trigger card you can use several modification tools for the ECG Sync, such as Delete, Disable, Insert. To reset the ECG curve select the check box Original ECG.
ECG Pulsing
ECG Pulsing is a dedicated technique used for online dose modulation for Cardiac imaging. The tube current is ECG-controlled and reduced during systolic phases of the cardiac cycle while maintained normal during diastolic phases when best image quality is required. As shown in figure 13, essential dose reduction up to 50% can be achieved. It can be switched on/off by the user (Fig. 14).
Fig. 13: Dose modulation with ECG pulsing.
29
HeartView CT
Fig. 14
CardioCARE
This is a dedicated cardiac filter which can reduce image noise thus provides the possibility of dose reduc­tion. It is applied in a pre-defined scan protocol called “CoronaryCARE.”
CardioSharp
This is a dedicated reconstruction kernel used for better edge definition in coronary artery imaging. It is applied in a pre-defined cardiac scan protocol called “CoronarySharp.”
Image examples are shown in figure 15.
Fig. 15: Image reconstruction with (15b) and without (15a) Cardio Sharp kernel.
15a
15b
30
HeartView CT
How to do it
Calcium Scoring
This application is used for identification and quanti­fication of calcified lesions in the coronary arteries. It can be performed with both ECG triggering (sequen­tial scanning) and gating (spiral scanning) techniques. The following scan protocols are predefined:
• CaScoreSpiStd
– Standard spiral scanning protocol with ECG gating
and a 0.42 s Rotation Time
• CaScoreSpiStd0.5s
– Standard spiral scanning protocol with ECG gating
and a 0.5 s Rotation Time
• CaScoreSeqStd
– Sequential scanning protocol with ECG triggering
Hints in General:
• Kernel B35f is dedicated to calcium scoring studies. To ensure the best image quality and correlation to known reference data, other kernels are not recom­mended.
• Use the ECG triggered protocol generally except for patients with arrhythmia. Use the ECG gated protocol when accuracy and/or reproducibility are essential, e.g. follow-up studies of calcium scoring or compari­son studies with conventional angiography.
• The protocol with 0.42 s Rotation Time can be applied to all examinations for HeartView CT. Temporal resolution for cardiac imaging can be achieved at down to 105 ms. The temporal resolution for the protocol with 0.5 s Rotation Time can be achieved at down to 125 ms.
• We recommend a tube voltage of 120 kV. If the tube voltage is lowered to 80 kV please use at least effec­tive mAs 250. The use of 80 kV is not advised for large patients.
Loading...
+ 98 hidden pages