Microvit MT-101 Holter and
MT-200 Evaluation Software
ECG CHANNEL 1
SPEED x1 CHAN2
t
w
f
o
a
S
R
E
L
L
I
H
C
S
r
o
f
s
t
h
g
i
r
y
t
r
e
p
o
r
p
l
a
i
c
r
e
m
m
o
c
d
n
a
l
a
i
r
t
s
u
d
n
i
l
l
A
.
s
e
m
a
N
e
SDS-200 2.01
d
a
r
T
SDS-104 2.01
d
n
SEMA-200 1.81
a
sk
SEMA-COMM 1.80
ra
M
MT-190/200 1.80
ed
MS-3 2.03
a
r
T
BR-102 2.40
d
e
r
Demo Sema-200
e
t
s
i
Demo MT-200
g
e
r
System Software
e
r
Release Notes
Ea
P
Acrobat Reader 4.0
O
C
S
I
N
I
M
d
n
a
A
For further information please visit our homepage
M
E
S
,
www.schiller.ch or send an e-mail to sales@schiller.ch
S
U
G
R
A
,
T
I
V
O
N
O
S
,
T
I
V
O
R
C
I
M
,
T
I
V
O
R
I
P
SCHILLER
SWITZERLAND
S
,
T
I
V
O
I
D
R
A
C
,
R
E
L
L
I
H
C
S
r
e
o
n
t
h
i
s
C
D
b
e
l
o
n
g
t
o
S
C
H
I
L
LE
R
A
G
,
S
w
i
t
z
e
r
l
a
n
d
.
A
l
l
r
i
g
h
t
s
r
e
s
e
r
v
e
d
.
Ó
S
C
H
I
L
L
E
R
A
G
,
C
H
-
6
3
4
Part No. 2.100256
Version x.xx
e
n
a
m
e
s
b
e
l
o
n
g
t
o
t
h
e
i
r
r
e
s
p
e
c
t
i
v
e
h
o
l
d
e
r
s
.
1
B
a
a
r
,
S
w
i
t
z
e
r
l
a
n
d
.
A
l
l
o
t
h
e
r
s
t
a
t
e
m
e
n
t
s
a
n
d
t
r
a
d
Art. no.: 2.510492 rev.: b *2.510492*
User Guide
Page 2
Sales and Service Information
The SCHILLER sales and service centre network is world-wide. For the address
of your local distributor, contact your nearest SCHILLER subsidiary. A complete
list of all distributors and subsidiaries is provided on our Internet site: http://
V This Holter Recorder and PC program is provided for the exclusive use of
qualified physicians or trained personnel under their direct supervision.
V The numerical and graphical results as well as any interpretation suggested by
the device must be examined with respect to the patient's overall clinical condition
and the quality of the recorded data.
V The responsibilities of the personnel for the operation and maintenance of the
device must be specified.
V Make sure that the personnel have read and understood the user guide, and
especially these safety notes.
V Damaged or missing parts must be replaced immediately.
V It is the owner's responsibility that the valid regulations for safety and prevention
of accidents are observed.
1.2Intended Use
V The MT-101/MT-200 Holter and evaluation software is designed to record long-
term electrocardiograms for the diagnosis of symptomatic and asymptomatic
arrhythmias, i.e. bradycardia or tachycardia, and for patients after resuscitation or
suffering from diseases such as cardiomyopathy, high blood pressure or long QT
syndrome.
V There is no danger when using the device for a patient with a pacemaker fitted.
V Always observe the indicated technical data when operating the device.
V The device is not designed for sterile use.
V Do not use the device in areas where there is any danger of explosion or in the
presence of flammable gases such as anaesthetic agents.
VThe device is CF classified. It is defibrillation protected when the original
SCHILLER patient cable is used. However, as a safety precaution when possible,
remove the electrodes before defibrillation.
V The device is not designed for direct cardiac application.
1.3Organisational Measures
V Before using the device, ensure that an introduction regarding its functions and
the safety precautions has been provided by a product representative.
Art. no.: 2.510492 rev.: b
V Always store the user guide near the device. Make sure that the user guide is
always complete and readable.
V Observe the safety notes for devices connected to the MT-101/MT-200.
V In addition to this user guide, also legal and other binding regulations for the
prevention of accidents and for environment protection must be observed.
Page 5
Page 8
1General and Safety Notes
1.4Operational Precautions
1.4Operational Precautions
1.5Safety Equipment
MT-101/MT-200
V This user guide, and especially these safety notes, must be read and observed.
V Do not touch the unit casing during defibrillation.
V It must be ensured that neither the patient nor the electrodes come into contact
with other persons or conducting objects (even if these are earthed).
V Changes, including operators behaviour, affecting safety must be immediately
reported to the responsible person.
V Operating this device without safety equipment or with damaged cables can
endanger the health or life of the patient or the person operating the device! For
this reason:
– Damaged cables and connections must immediately be replaced.
1.6Precautions for Operation with other
Devices
V Use only accessories and other parts recommended or supplied by SCHILLER
AG. The use of other than recommended or supplied parts may result in injury,
inaccurate information and/or damage to the device.
V Accessory equipment connected to the analogue and digital interfaces must be
certified according to the respective IEC standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment).
Furthermore, all configurations shall comply with the valid version of the system
standard IEC/EN 60601-1-1. Everyone who connects additional equipment to the
signal input part or signal output part configures a medical system, and is
therefore responsible that the system complies with the requirements of the valid
version of the system standard IEC/EN 60601-1-1. If in doubt, consult the
technical service department or your local representative.
V Special care must be exercised when the unit is used with high frequency
equipment. To prevent the display of incorrect ECG signals, only use special
SCHILLER ECG cables protected against high frequency radiation.
V There is no danger when using this device simultaneously with electrical
stimulation equipment. However, the stimulation units should only be used at a
sufficient distance from the electrodes. If in doubt, disconnect the patient from the
recorder.
Page 6
1.7Maintenance
V Do not use high temperature sterilisation processes (such as autoclaving). Do not
use e-beam or gamma radiation sterilisation.
V Do not use aggressive or abrasive cleaners.
V Do not, under any circumstances, immerse the device or cable assemblies in
liquid.
Page 9
MT-101/MT-200
General and Safety Notes1
User GuideSafety Symbols and Pictograms1.8
1.8Safety Symbols and Pictograms
The safety level is classified according ANSI Z535.4. The following overview shows
the safety symbols and pictograms used in this handbook.
For a direct danger which could lead to severe personal injury or to death.
For a possibly dangerous situation, which could lead to bodily injury or to death.
For a possibly dangerous situation which could lead to personal injury. This symbol is
also used to indicate possible damage to property.
For general safety notes as listed in this chapter. When this symbol is displayed on the
unit, it means that the user should refer to the user guide.
Note for possible dangerous situations which could lead to damage to property or system failure. Important or helpful user information.
Reference to other guidelines.
Potential equalization.
CF symbol. This unit is classified safe for direct cardiac application. Only defibrillation
protected when used with the original SCHILLER patient cable.
The unit/component can be recycled.
Art. no.: 2.510492 rev.: b
Notified body of the CE certification (TÜV P.S.).
Is intended for infants weighing less then 10 kg.
Page 7
Page 10
2Introduction
2.1MT-101/200 Range of Application
2Introduction
2.1MT-101/200 Range of Application
The MICROVIT MT-101 Holter is designed to record long-term electrocardiograms for
the diagnosis of symptomatic and asymptomatic arrhythmias, i.e. bradycardia and
tachycardia, and for patients after resuscitation or suffering from diseases such as
cardiomyopathy, high blood pressure or long QT syndrome.
The recording can also be used to help examine palpitations or syncopes and
dizziness, to verify medical therapies, and to carry out subsequent treatments after a
bypass operation or a PTCA. The ST segment analysis of an ECG recording allows
the detection of a symptomatic or asymptomatic ischemia.
MT-101/MT-200
The SCHILLER Holter system comprises two main parts. The MT-101 Holter recorder
and the MT-200 program. Recordings made by the MT-101 unit are downloaded to
the MT-200 for display, storage and analysis.
o
S
R
E
L
L
I
H
C
S
r
o
f
s
t
h
g
i
r
y
t
r
e
p
o
r
p
l
a
i
c
r
e
m
m
o
c
d
n
a
l
a
i
r
t
s
u
d
n
i
l
l
A
.
s
e
m
a
N
e
SDS-200 2.01
d
a
r
T
SDS-104 2.01
d
n
SEMA-200 1.81
a
s
k
SEMA-COMM 1.80
r
a
M
MT-190/200 1.80
e
d
MS-3 2.03
a
r
T
BR-102 2.40
d
e
r
Demo Sema-200
e
t
s
i
Demo MT-200
g
e
r
System Software
e
r
Release Notes
Ea
P
Acrobat Reader 4.0
O
C
S
I
N
I
M
d
n
a
A
For further information please visit our homepage
M
E
S
,
www.schiller.chor send an e-mail to sales@schiller.ch
S
U
G
R
A
,
T
I
V
O
N
O
S
,
T
I
V
O
R
C
I
M
,
T
I
V
O
R
I
P
S
,
T
I
V
O
I
SCHILLER
SWITZERLAND
D
R
A
C
,
R
E
L
L
I
H
C
S
t
f
w
a
r
e
s
.
o
n
t
h
i
s
C
e
h
o
l
d
e
r
D
b
e
l
o
n
g
t
o
S
C
H
I
L
Part No. 2.100256
g
t
o
t
h
e
i
r
r
e
s
p
e
c
t
i
v
LE
R
A
G
,
S
w
i
t
z
e
r
l
a
n
Version x.xx
d
e
n
a
m
e
s
b
e
l
o
n
Good signal quality is vital for the success of a recording. The built-in Holter display,
enables the ECG signal quality to be checked before starting, and the recording
commenced directly from the device. This gives a high degree of reliability.
At the end of a recording, the data is transferred from the Holter recorder to a PC. The
transfer of a recording typically only requires a few minutes.
d
.
A
l
l
r
i
g
h
t
s
r
e
s
e
r
v
e
d
.
Ó
S
C
H
I
L
L
E
R
A
G
,
C
H
6
3
4
1
B
a
a
r
,
S
w
i
t
z
e
r
l
a
n
d
.
A
l
l
o
t
h
e
r
s
t
a
t
e
m
e
n
t
s
a
n
d
t
r
a
The MT-200 is a PC based ECG evaluation program. An ECG is recorded using the
SCHILLER MICROVIT MT-101 Holter. Two or 3-channel ECG recordings can be
recorded over a period up to 72 hours. After the transfer of the recording data into the
MT-200 program, the data can be displayed, saved, analysed and printed. The MT200 program enables quick access to the recording data and displays the ECG and
analysis data in a logical and understandable way for diagnosis.
The MT-101 Holter can be programmed simply using the two keys following the menu
guidance on the LCD display.
MT-101/MT-200
Data transmission
The ECG data transmission to the PC can be realised in two ways:
– directly via standard USB connection (3)
– by removing the memory card (9) from the battery housing and transmitting the
data to the PC by means of a memory card reader. The advantage resulting from
this procedure is that the Holter can be equipped with a new memory card and
is immediately available for the next patient.
Page 10
Page 13
MT-101/MT-200
Introduction2
User GuideOperating and Display Elements2.3
2.3Operating and Display Elements
The Microvit MT-101 is operated with the two buttons and the menu guidance on the
LCD. The green key with the switching-on symbol is used additionally to switch the
device on and off.
Functions
green button:
On/off
NEXT
EVENT
SPEED 1x - 3x
NO
2.3.1Switching on
Press the green button. The display shows the name and version number of the device before the main menu is displayed.
2.3.2Switching off
Keep the green button pressed for five seconds. When the button is released, the device is switched off. If an ECG recording is running, first stop it following the same procedure.
If no recording is running, the device will be switched off automatically after five minutes.
Status
NEXT OK
15:45
Functions of the
blue button:
OK
CHANGE
EVENT
YES
CHANN2/3
2.3.3Battery display
The battery symbol indicates the battery's load status. If the battery is full, the symbol
is solid - also see para. 2.5.2 Inserting/changing the battery, page 13.
2.3.4Status display
The operating status is displayed in the left upper corner of the LCD. REC for record-
Art. no.: 2.510492 rev.: b
ing, USB if the MT-101 is connected to a PC, USB* for data transmission to a PC.
Page 11
Page 14
2Introduction
2.4MT-101 Menu Structure
Main MenuSub-Menu 1Value/InfoSub-Menu 2Value/Info Sub-Menu 3Value/Info
RECORD STARTECG SignalChan1speedx1, x2 or x3
ECG SignalChan2speedx1, x2 or x3
ECG SignalChan3speedx1, x2 or x3
Start record?Yes/No
ECG-recording start-
ed!
RECORD SETUP
LAST RECORDPatient ID*
SYSTEM INFOSerNo.
SYSTEM SETUPContrast1...8
Patient ID > Voice-record
PM Det.*On/off
Duration24, 48, 72
Sampling 125 Hz
Events* >Events
LanguageENG ...
Bat
Date/time > Year2000....2099
2.4MT-101 Menu Structure
The menus are selected with the green button (NEXT). A selected menu is opened
with the blue button. Depending on the called menu, the button functions may
change.)
MAIN 15:45
RECORD START
RECORD SETUP
LAST RECORD
SYSTEM INFO
SYSTEM SETUP
NEXT OK
2.4.1Menu Overview
Ye s > Event buttonEvent saved!
> Record info
> Stop recording?
> Play ID
Version
Bat Type
SD-Card
Alkaline
NiMH2100
> Month01....12
> Day01....31
> Hour00.23
> Minute00....59
ECG recording
stopped
>
StopStop Recording
>
MT-101/MT-200
Main
StartStart Recording
StopStop Playing
Page 12
* To record pacemaker pulses it is important that ‘PM Det‘ is set to on - see para. 5.1.1
Pacemaker, page 21.
Page 15
MT-101/MT-200
Introduction2
User GuideInitial Operation2.5
2.5Initial Operation
2.5.1Unpacking
Check that all ordered items are present and free of shipping damage. Immediately
report any damage to SCHILLER AG.
2.5.2Inserting/changing the battery
Open the battery compartment and insert the supplied battery or accumulator. Observe the polarity!
Note
The delivered battery is of alkaline type AA/LR6. If you use an NiMH 2100 mAh accumulator, make sure that BAT NiHM is selected in the SYSTEM SETUP menu. If the
wrong type is selected, the battery capacity will not be displayed correctly.
On closing the battery cover, pay attention that the two lugs (A) are inserted correctly.
The cover is closed in the direction indicated by the arrow (B). In order to engage the
cover, press it down (at position C) until it clicks in place.
V Attention - danger of explosion Do not dispose of batteries by fire or incinera-
tor.
V Attention - danger of acid burn Do not open the battery casing.
Only dispose of batteries in official recycling centres or municipally approved areas.
Switch on the device and check the battery charge capacity. The battery symbol must
Full
Half full
Empty
Art. no.: 2.510492 rev.: b
be fully black. This corresponds to a maximum recording time of 24 hours.
An audible and visual indication is given during recording when battery capacity is
limited. The time will vary according to the type of battery installed (alkaline or
NiMH2100) - but is normally between 1 and 2 hours. When the alarm is given and
recording is to be continued, we recommend that the battery is replaced at the first
opportunity - see para. 2.5.2 Inserting/changing the battery, page 13
When recording is stopped because of low battery capacity, and the battery is
replaced within 5 hours of the unit switching off, the recording will continue - see para.
3.2.1 During the Recording and Patient Information, page 17.
Page 13
Page 16
3Preparing a Holter Recording
1
2
3
4
5
K2 +
K2 -
3.1Position of the Electrodes
3Preparing a Holter
3.1Position of the Electrodes
MT-101/MT-200
Recording
Typical electrode position for a 4-lead cable (2-channel recording)
The recommended electrode placement for a 2-channel recording is shown below.
Channel 1 positive (K1+) = green
Channel 1 negative (K1-) = red
Channel 2 positive (K2+) = white
Channel 2 negative (K2-) = yellow
K1-
K2 +
K3 -
Typical electrode placement for a 6-lead cable (3-channel recording)
The recommended electrode placement for a 3-channel recording is shown below.
Channel 1 positive (K1+) = green
1
2
3
4
5
K2 -
K3 +
K1 +
Channel 1 negative (K1-) = red
Channel 2 positive (K2+) = white
Channel 2 negative (K2-) = yellow
Channel 3 positive (K3+) = orange (positioned on the patient’s back)
Channel 3 negative (K3-) = blue
Page 14
Page 17
MT-101/MT-200
Preparing a Holter Recording3
User GuidePosition of the Electrodes3.1
Electrode Placement
Form a stress loop in every cable and secure them with adhesive strips to relieve the
electrodes (strain relief). In order to ensure good data evaluation, the ECG amplitudes
should be examined in the sitting, lying and standing position of the patient.
Holter ECGs use a bipolar lead system (one positive and one negative lead) for each
channel. Channel 1 approximates to modified lead V
modified lead V
and channel 3 approximates to modified lead V3.
2
, channel 2 approximates to
5
Channel 1 Place the RED negative electrode under the clavicle on the right Sternal margin.
Place the GREEN positive electrode in the fifth left intercostal space on the anterior
axillary line (position approximately equates to V
).
5
Channel 2Place the YELLOW negative electrode under the clavicle on the left sternal margin.
Place the WHITE positive electrode in the fourth left intercostal space on the anterior
axillary line (position approximately equates to V
).
2
Channel 3Place the BLUE negative electrode in the fourth left intercostal space near the ster-
num.
Place the ORANGE positive electrode on the back in the fifth left intercostal space,
between the spine and the scapula (position approximately equates to V
).
3
• The above electrode placement is suggested; other electrode configurations are
possible.
• Ensure that the QRS complex is bigger than the T wave.
• Ensure that the trace is larger than 1mV. See 1mV reference (4) on following page.
Art. no.: 2.510492 rev.: b
• To avoid artifacts in women patients, the red and white electrodes can be placed
lower if necessary.
Page 15
Page 18
3Preparing a Holter Recording
3.2Commencing a Holter Recording
3.2Commencing a Holter Recording
3
ECG
4
SPEED x1
5
1
2
MT-101/MT-200
The ECG recording can be started without the MT-200 PC software. The most important data can be entered in the MT-101 directly, and the ECG signal examined directly
on the LCD.
A recording can also be started from the MT-200 program where all channels can be
viewed simultaneously before commencing - see para. 5.1 Starting a Recording from
the MT-200 Program, page 19.
Preparing the patient
1. Attach electrodes to patient.
Setting up the Holter MT-101
2. Press button (1) to switch on MT-101. Check battery charge capacity. If the symbol is only half filled out, change battery.
3. Choose NEXT (1) to select RECORD SETUP menu to make recording settings:
– Patient ID - record patient ID using the microphone and playback facility
– Select Pacemaker detection on or off - see para. 5.1.1 Pacemaker, page 21
– Define period of recording - 24, 48 or 72 hours
Checking the ECG signal
4. Connect patient cable to MT-101 (3).
5. Confirm RECORD START menu with “OK” (2) and check ECG signal Channel 1.
Press CHAN2/3 to select and check channel 2/3. Press “OK” (2) to access
START RECORDING panel.
The signal's max. amplitude corresponds to the height of the moving line (5). The 1mV
amplitude reference is the vertical line on the left (4). Ensure the signal amplitude is
greater than 1mV.
Starting an ECG recording
6. Confirm the start of the recording with YES (2).
Stopping the ECG recording
7. Press and hold button (1) for 5 seconds. You will be prompted if you wish to stop
the recording. Confirm with YES (2).
NOTE: If no confirmation is received to cease recording (button (2) pressed), within
15 seconds, the unit returns to recording mode.
Page 16
Switching off the MT-101
8. If the unit is recording, first stop the recording step (7).
9. Ensure the main menu is displayed and that the cursor is at the RECORD START
position.
10. Press and hold button (1) for 5 seconds to switch the device off.
Page 19
MT-101/MT-200
Preparing a Holter Recording3
User Guide Taking an Extended Recording (Longer than 24 hours)3.3
3.2.1During the Recording and Patient Information
Inform the patient about the use of the MT-101.
Event record
• Every event should be entered in the diary, together with the time, the activities at
the time of occurrence and the symptoms.
• Instruct the patient to press the EVENT button at any time during the recording to
register an event as follows:
1. Press button (1 or 2).
2. Record event in the patient diary.
Note:
The template for the patient diary is stored on the software CD as Word or pdf file. An
example is given at the end of the book - see para. 13 Patient Diary, page 89.
No ECG signal or lead-off
1. Check cable connection on device.
2. Check cable connection on electrodes.
3. Re-attach electrodes to body.
General information
The device is not waterproof. The patient should be advised not to take a bath or
shower during the recording.
Battery replacement during the recording.
Change battery when an audible indication is given and the message ‘BATTERY LOW
- change battery’ is displayed the MT-101. - this will occur approximately 1-2 hours
before switch off (dependent on battery type). Proceed as follows:
RE C 8 :56
ECG recording
(00:02)
1 2
1. Press EVENT button (1 or 2) and make an entry in your diary.
– DO NOT SWITCH THE DEVICE OFF
2. Open battery compartment and replace battery with a new one of the same type.
Observe correct polarity, and replace battery cover - see para. 2.5.2 Inserting/
changing the battery, page 13.
3. Switch the device on by pressing button (1). After a few seconds the message
‘ECG recording restart’ is displayed while the unit re-initialises. This is followed by
the message ‘ECG recording’ and ECG recording automatically resumes.
When a recording is stopped (because of low battery capacity or because of
battery removal), the battery must be replaced within 5 hours of the unit
switching off for the recording to continue.
3.3Taking an Extended Recording (Longer than
24 hours)
The MT-101 can record up to 72 hours of Holter data if required. To make a recording
Art. no.: 2.510492 rev.: b
longer than 24 hours, the battery in the MT-101 must be changed as detailed above.
An audible alarm and visual indication will be given when the battery must be
changed.
Page 17
Page 20
4Transferring a Recording to the PC
4.1Data Transmission to PC from MT-101
4Transferring a Recording
to the PC
4.1Data Transmission to PC from MT-101
1. Connect the USB cable between the MT-101 and PC (see picture below - the rubber cover on the MT-101 must be removed - the message "MT-101 connected to
PC" appears.
2. Start MT-200 software on PC.
3. Click on the data transfer icon and select "Request Holter Data". The dialogue box
shows the transferred data in per cent.
The data is stored automatically. If no patient data is entered, the file will be saved with
the date and time.
MT-101/MT-200
4.2Data transmission to PC with Memory Card
Reader
1. Connect the card reader to the USB. The memory card reader appears as a physical drive on your desktop.
2. The path name for this drive must be entered in the menu Option/System/Path/
SD-Card path - see para. 8.5.3 Directories, page 76.
3. Insert the memory card in the card reader or PCMCIA adapter.
4. Execute the function “Read SD-Card”. The data is read into the indicated path (for
path location see point 2)
Data can also be imported from an SD card - see para. 7.2.3 Importing recordings,
page 65
Page 18
Page 21
MT-101/MT-200
Displaying an ECG Signal5
User GuideStarting a Recording from the MT-200 Program5.1
5Displaying an ECG
Signal
5.1Starting a Recording from the MT-200
Program
1. Start the MT-200 program on your PC /CS-200. The welcome page is displayed.
1
2. Click on the “New recording” icon (1). The patient data screen is displayed:
Art. no.: 2.510492 rev.: b
3. Enter the patient data and confirm with the "OK" button. To enter the data, click
with the mouse cursor into the fields or jump from entry field to entry field using
the tab key.
4. If pacemaker detection is required check the pacemaker box (1) - see para. 5.1.1
Pacemaker, page 21 for notes on pacemaker detection
1
Page 19
Page 22
5Displaying an ECG Signal
5.1Starting a Recording from the MT-200 Program
5. When the patient data has been entered and Ok is clicked, enter the duration of
the recording
Note: The data displayed or entered on the other pages of patient information selected by clicking on the tabs at the top of this window (Recording data, Assignment
etc.) - are only available after a recording has been made.
6. Place the electrodes as indicated in the dialogue box.
MT-101/MT-200
If a 4-lead patient cable was used for the Holter recording, only two channels will be
displayed in the MT-200.
Page 20
7. Check the signal and re-apply the electrodes if necessary.
Page 23
MT-101/MT-200
Displaying an ECG Signal5
User GuideStarting a Recording from the MT-200 Program5.1
8. Click on START RECORDING to commence the recording. The dialogue box re-
minds you to check the battery load capacity.
The recording has now commenced. The LCD of the MT-101 indicates that ECG recording has been started.
Remove the USB from the MT-101 Holter recorder and close the connector again with
the protective cover.
5.1.1Pacemaker
The MT-200 cannot determine pacemaker spikes itself and it is not possible to detect
pacemaker signals by later analysing the recording in the program if the pacemaker
detection function was not enabled during the original recording. So that the MT-200
can detect pacemaker spikes, pacemaker detection must be activated before the start
of the recording.
Activating Pacemaker Detection
Pacemaker detection can be activated in either the MT-200 program or the MT-101:
« If the recording is started from the MT-200 program, then pacemaker activation
can be carried out in the MT-200 by ticking the pacemaker box in the patient data
window - see para. 5 Displaying an ECG Signal, page 19.
« If the recording is started from the MT-101 itself, then pacemaker activation must
be carried out in the MT-101 menu > RECORD SETUP > PM Detection > ON, see para. 2.4.1 Menu Overview, page 12.
«
Detection and Recording of Pacemaker Spikes
The MT-101 only detects pacemaker signals in long-term ECG recordings when the
slopes and amplitudes of the signal exceed the preset limits and when pacemaker
detection is enabled. As the MT-101 uses a sampling frequency of 500 Hz for a
recording (i.e. a digitalisation interval of 2 ms), complete digital processing is
impossible due to the short duration of some pacemaker signals (less than 1 ms).
Therefore, analogue processing of the ECG signals is applied by the MT-101 for
pacemaker detection.
Analogue Pacemaker Detection in Channel 1
Analogue pacemaker detection is confined to the first channel of the MT-101. It is
therefore optimal when the amplitude of the pacemaker signal for the first ECG channel is greater than that for the second channel. In some instances, this is not the case.
In this case, it is recommended that the real-time ECG traces are viewed on the
screen before starting the long-term recording. Pacemaker detection is automatically
enabled on the real-time display. The pacemaker signals, however, are not always
detected. If this is the case and the amplitude of the pulse is greater in the second
channel, simply exchange the electrodes of channels 1 and 2. After the real-time display, pacemaker detection must be enabled for long-term recording in the MT-200!
Evaluation and Display of Pacemaker Spikes in the MT-200
Art. no.: 2.510492 rev.: b
Program
Pacemaker signals are marked in the MT-200 program by vertical lines in the ECG
after evaluation of the recording. Note that these lines are correctly positioned in relation to time but are not proportional in either amplitude (voltage) or duration of the
pacemaker pulse, nor do they indicate the polarity. The pacemaker representation is
always positive but the actual pacemaker spike may be positive or negative.
Page 21
Page 24
5Displaying an ECG Signal
5.2Transmission Problems
5.2Transmission Problems
If an error message appears either before starting a recording or when attempting to
transfer a recording from the MT-101 to the PC, possible causes are as follows:
ErrorCauseRemedy
MT-101/MT-200
Communication Error (error
message displayed)
Display of DEMO VERSION
No error message. The program
cannot find the Holter.
• The USB cable assembly is not inserted
correctly in either the MT-101 or the PC.
• The SCHILLER USB driver has not been
installed
• The device is not connected.
• The SD memory card is not inserted.
• Unit not switched on
• Battery exhausted.
• Another device is connected to the USB.« Disconnect the device and connect the
• The hard-lock key is not present or incorrectly inserted.
• There is no license.
• With an USB installation, the program can
occasionally "hang" and not recognise
the USB connection.
« Check that both the cable connectors
are securely placed.
« Install USB driver (on SCHILLER soft-
ware CD.
« Connect the device.
« Check the memory card.
« Switch unit on
« Change batteries
MT-101.
« Check the hard-lock key on the PC.
« Contact SCHILLER for network license
1. Close the MT-200 program.
2. Disconnect the USB connector to your
PC. Wait circa three seconds and replace the connector.
3. Open the MT-200 program again.
Page 22
Page 25
MT-101/MT-200
Displaying an ECG Signal5
User GuideTransmission Problems5.2
5.2.1Checking the connection
A test function is available to check the integrity of the connection between the MT101 and the PC. To carry out the test function, proceed as follows:
1. In the "Options" menu, select "System".
2. Click the "Holter" tab.
3. Check the correct box for your installation (USB or AT card), and click the "Test
Connection" icon. Follow the instructions on the screen.
4. When "OK" is clicked, the software sends a test message to/from the MT-101.
5. A message box indicates the success of the transmission.
If the problem could not be solved, check all connections, ensure the MT-101 is
switched on, close the MT-200 application and restart the software.
Art. no.: 2.510492 rev.: b
Page 23
Page 26
6Viewing and Editing a Recording
6.1Icons
6Viewing and Editing a
6.1Icons
MT-101/MT-200
Recording
The MT-200 program gives different views for the presentation of a Holter recording.
Every view offers various data and analytical information. Besides display icons, the
toolbar contains additional function icons enabling the quick and easy activation of the
most frequently used functions. All icons can be activated any time and in any view.
In the ECG and zoom views, additional function icons are given to the left of the
screen to change the size of the traces and/or the time segment of the recording.
Function and View icons
Page 24
Function and View icons
in the ECG overview
All icons are selected by mouse-click (position the cursor on the icon and click with
the left mouse button).
When an icon is dimmed, it means that this function is not available for the currently
displayed screen and cannot be selected. For example, the zoom function is not
available in the ST view so the "Zoom" icon is dimmed. The patient name is always
displayed at the top of the page when an ECG recording is displayed.
Page 27
MT-101/MT-200
Viewing and Editing a Recording6
User GuideIcons6.1
6.1.1View icons
The function and view icons are only active (selectable) when a patient recording is
displayed. When an icon function cannot be selected, the icon is dimmed.
Event View*This gives an overview of all events in the 24 hour
period. Use this view to quickly identify and select
a specific time segment for display.
Analysis Summary ViewThis provides a tabular overview of all important
measurements for the entire 24 hour period.
Event Samples ViewThis allows three event samples from every event
category to be displayed, i.e. selected or replaced.
The user event samples can also be selected for
printing in this view.
ECG View*This zooms in on a specific time segment of ECG
for closer analysis. The next page can also be
displayed using the automatic scrolling function.
Analysis SummaryThis provides a Tabular overview of the recording.
ST Trend View(Option) This provides a graphical overview of the
ST trend with tabular measurements for ST
episodes.
Template View(Option) This provides a graphical overview of the
different types of averaged QRS waveforms with
classification, detected over the entire recording.
Pacemaker View(Option) This provides a graphical overview of
QRS templates measured in relation to the
pacemaker pulse.
Heart Rate Variability(Option) This provides a graphical overview of the
heart rate variability.
Heart Rate TrendThis provides a graphical overview of the heart
rate trend. The maximum/minimum HR and NN
interval can be manually defined
Zoom ViewThis gives a zoom view of a selected ECG
segment. Specific QRS complexes can also be
reclassified in this view. See "Reclassifying/Editing
a QRS complex" later in this section. The two leads
displayed in the zoom view are the two leads that
have been analysed. The display will always
contain two channels, even if only one channel has
been analysed.
HR TrendThis displays the heart rate trend over the entire
recording.
Art. no.: 2.510492 rev.: b
If only one lead has been analysed, channel 1 is the second lead in the display. If
channel 1 has been analysed, channel 2 is the second lead displayed.
Options
All options are enabled with a programmed hard-lock key. This hard-lock key can be
obtained from SCHILLER AG.
*In event and ECG view modes, the lower part of the display is divided to give either
ECG zoom or heart rate trend.
Page 25
Page 28
6Viewing and Editing a Recording
6.1Icons
6.1.2Function icons
MT-101/MT-200
The following function icons are available when a patient recording is viewed:
Open RecordingOpen a recording. Click the icon to list all available
files.Click the arrow to the right of the icon, to list
the last four opened recordings.
New RecordingEnter the data of a new patient and start a new
recording.
Save RecordingSave the current recording. Click the arrow to the
right of the icon, to display further options to save
as a pdf file. If saved as PDF file, it is possible to
delete the original if desired.
E-MailSend the currently displayed recording by e-mail.
Click the arrow to the right of the icon, to give
further options to send as a pdf file. If “Send EMail
as PDF File” is selected, a pdf is generated and
automatically attached to the e-mail.
Patient and Recording
Data
Print PreviewSelect and display the pages to be printed (before
View / edit patient and recording data. View
analysis settings and diagnosis. General recording
settings and options.
printing)
PrintPrint (user defined) recording data. Click the arrow
to the right of the icon to select specific data for
print.
Request Holter Data Load data from the connected MT-101 Holter
recorder or via a memory card reader by selecting
"Read SD Card". In Win 95/98, a third function is
available to load data from a tape.
AnalyseClick the Analyse icon to analyse the currently
displayed recording to the defined analysis
parameters. Click the arrow to the right of the icon
to analyse multiple recordings.
Scroll BackScroll backwards (in time) of the zoom ECG
currently displayed.
Scroll ForwardScroll forwards (in time) of the zoom ECG currently
displayed.
Scroll Event Back Go to previous event.
Scroll Event Forward Go to next event.
Time Scale for ViewDisplay 3, 6, 12, 24 hours of analysed data or, for
recordings longer than 24 hours, all the data.
When 3, 6, 12, or 24 hour is selected, the arrow
icons at the side of the box, enable the user to
jump to the next time segment.
Page 26
Page 29
MT-101/MT-200
Viewing and Editing a Recording6
User GuideIcons6.1
6.1.3Tool icons in rhythm and zoom views
The tool icons are displayed in the rhythm and zoom views on the left hand side. Use
these icons to:
• decrease or increase the amplitude and speed
• move up or down a line and page
• immediately print a selected half hour segment of the recording
• change the channel (ECG view)
• select and analyse specific channels (ECG view)
• select a zoom section of the recording for printing
The page up/down and line up/down icons are not applicable and not displayed in the
zoom view
CentreIn the ECG view, centres the selected (highlighted)
ECG section in the middle of the screen.
In the zoom view, positions the selected QRS
complex (cursor above and below QRS complex)
slightly to the left of centre in the zoom screen.
PrintIn the ECG view , immediately prints a 30 min.
segment of the recording (1/4 hour before and 1/4
after the selected section).
In the zoom view, marks the displayed zoom
section as "selected".
Page UpMoves to the previous page. Each page displays
between approximately 1 and 24 minutes of
Page DownMoves to the next page
Line UpShifts the display up one line)
Line DownShifts the display down one line
Decrease SpeedDecreases the ECG scale (curves closer).
Increase SpeedIncreases the ECG scale (curves wider).
Increase AmplitudeCurves bigger
Decrease Amplitude Curves smaller
Art. no.: 2.510492 rev.: b
Select ChannelSelect any combination of one, two or three
recording dependent on the speed selected.
channels (for display)
Select Channel for
Analysis
Auto Scroll DownAutomatically scrolls down through the recording.
Select and analyse one or two channel
Subsequent time segments are displayed
automatically
Page 27
Page 30
6Viewing and Editing a Recording
6.2Accessing and Opening Files
6.2Accessing and Opening Files
MT-101/MT-200
Use the open file icon to:
• Open a file
• Review stored files
• Show all exported/not exported or validated/not validated recordings
To display (and select) the last four viewed recordings, click on the arrow by the side
of the icon.
This lists all files (and subdirectories) in the last folder selected (usually ‘ data’. Click
on the desired file to open it.
Page 28
The "V" and the "E" columns indicate if the recording has been validated or exported.
If you wish to display only all validated/non-validated or all exported/not exported
recordings, click on the arrow at the side of the "Files of type" field to define.
Page 31
MT-101/MT-200
Viewing and Editing a Recording6
User GuideEvent & Zoom Views6.3
6.3Event & Zoom Views
« Click on the "Event view" icon.
1
2
3
4
5
Event View
The time duration of the event view is selected (1) between 3, 6 12, 24 (hours), or all.
When a section is displayed and earlier/later data is available, the left/right arrows are
active to go to the previous/next time segment or first/last time segment of the
recording.
The time bar (2) shows the position in the recording. The first figure gives the hour
and the second figure the day.
NOTE: When the next/previous time section is selected (1), the displayed section (2)
will overlap the one before by an hour.
Art. no.: 2.510492 rev.: b
Click anywhere in the event view. A vertical cursor line is displayed and a zoom view
(or HR trend) for the selected recording section is displayed in the bottom part of the
screen. To view another section, move the cursor line by clicking on another point in
the event table. The vertical line is repositioned and the zoom ECG view (or HR trend)
in the bottom part of the screen displays the ECG at the selected time.
The heart rate trend can be displayed instead of the ECG zoom display in the bottom
window (see “Heart Rate View”).
The numbers at the right of the event table give the total number of specific events
detected in the entire recording
Page 29
Page 32
6Viewing and Editing a Recording
6.3Event & Zoom Views
MT-101/MT-200
Zoom View
Two blue measurement lines (4) are displayed in the zoom view when the
measurement box, in the bottom left (3) of the zoom screen, is checked. The
amplitudes of the two channels and the time can be measured by moving the
measurement lines. A small window gives the differences between the values of the
two lines. Reposition the lines by moving with the cursor. The measurement values
will change instantly.
Both measurement lines (4) can be moved together to maintain the same time
interval. To do this, press the control key while moving the cursor.
The “ms“ and/or “min“ boxes in the bottom left of the zoom screen display the
measured time and/or heart rate when checked and indicate:
• the time in milliseconds between respective heart beats
• the heart rate calculated beat to beat.
ECG editing and reclassification is carried out in the zoom view - see para. 6.14
Reclassifying/Editing a QRS Complex, page 52.
When a file is opened, the event view is given. This screen displays all the events in
the entire recording. The time scale of the recording is displayed below the events.
The black up and down pointing arrows positioned in the bradycardia and tachycardia
events indicate the lowest and highest heart rates (with reference to time) measured
in the complete recording.
To display the zoom view of the respective ECG sections, select the corresponding
option in the "View" menu:
2/3-Channel Analysis Indications
The number of channels recorded, and the channels that have been analysed, are
indicated in every view in the bottom right of the display (5). The indication states the
number of channels recorded, followed by the channels that have been analysed. The
following combinations are possible:
Selection of channels for analysis and display are detailed later in this book, see para.
6.6 ECG View, page 35. For analysis options see para. 6.15 Analysing/Re-analysing
the Recording, page 54.
Page 33
MT-101/MT-200
3
Viewing and Editing a Recording6
User GuideAnalysis Summary6.4
6.4Analysis Summary
« Click on the "Analysis summary" icon.
1
2
Art. no.: 2.510492 rev.: b
Page 31
Page 34
6Viewing and Editing a Recording
6.4Analysis Summary
MT-101/MT-200
In the ‘Heartrate / VES‘ view and the ‘VES/signal quality‘ view, the analysis summary
can be displayed for day 1, 2, 3 or all (3).
An analysis summary of all important data over the entire Holter recording is given
over three pages.
« To display one page, click on the respective "Holter Report", "Heartrate/SVES" or
"VES/Signal Quality" tab at the lower edge of the table.
« In the table, click on any of the categories (1) highlighted blue (e.g. HR [/min]).
The zoom view of the desired ECG section is immediately displayed.
« Click on the arrows on the scroll bar (2) to the right to display the upper or lower
part of the table.
« Click on the arrows on the scroll bar at the bottom to move the table further to the
right or left.
ECG editing and reclassification is carried out in the zoom view - see para. 6.14
Reclassifying/Editing a QRS Complex, page 52.
Page 32
Page 35
MT-101/MT-200
Event 1
Viewing and Editing a Recording6
User GuideEvent Samples View6.5
6.5Event Samples View
« Click on the Event samples icon.
123
Event 2
Event 3
4
Art. no.: 2.510492 rev.: b
Page 33
Page 36
6Viewing and Editing a Recording
6.5Event Samples View
MT-101/MT-200
Up to three event samples can be selected for every event category. After the analysis
of a recording, the three event samples are automatically displayed (the first event
within an eight-hour cycle, respectively):
Event 1: first detected event in the recording
Event 2: first detected event after eight hours
Event 3: first detected event after 16 hours
If three or less events in a category are detected, then all of the events detected are
displayed, regardless of where in the recording they occurred. The event samples
given here are also the samples given on the printout when the event samples box is
checked in the print menu. The event categories are selected by clicking on the arrow
(1) next to the type indication in the top right of the screen.
All events in the selected category are indicated by a vertical line. Click on any event
(2) on this line to display a zoom view of the selected segment.
To display the next or previous event in the zoom window, click on the icons (3) in the
toolbar.
Click on an event sample (4) to display the sequence in the zoom window.
ECG editing and reclassification is carried out in the zoom view - see para. 6.14
Reclassifying/Editing a QRS Complex, page 52.
Page 34
Page 37
MT-101/MT-200
1
2
Viewing and Editing a Recording6
User GuideECG View6.6
6.6ECG View
« Click on the "ECG View" icon.
3
The time of the ECG recording is displayed on the left of each ECG line. In the upper
window, you can see a frame showing an ECG segment. This framed segment is
zoomed in the lower window. When the cursor is positioned in the ECG view (in the
Art. no.: 2.510492 rev.: b
upper screen), it changes to a zoom cursor.
To zoom in on a particular segment of the displayed ECG, move the zoom cursor to
the required ECG area in the upper screen and click with the left mouse button. The
area highlighted on the upper screen is then shown in zoom mode in the bottom
section.
The maximum time span for a page in ECG view is approximately 60 minutes of ECG
recording. If you wish to jump hours, it is easier to go into the event view, select the
time segment and then return to the ECG view.
Page 35
Page 38
6Viewing and Editing a Recording
6.6ECG View
6.6.1Selecting Channels for Display
MT-101/MT-200
« Click on the "Channel number" icon (1) to display the required channel(s). In the
case of:
• a 2-channel recording, the following window is displayed:
• a 3-channel recording, the following window is displayed:
6.6.2Selecting Channels for Analysis
« Click on the "Analyse channels" icon (2) to define the channels to be analysed.
The channel(s) for analysis are the same the options above
6.6.3Auto Scrolling
« Click on the “Auto page down” icon to automatically page scroll through the Holter
recording. When Selected, a speed setting (3) is displayed to define the scrolling
speed.
Zoom
« To zoom in on a selected section during the scrolling, click with the zoom cursor.
This displays the zoom view of the selected section.
Pause
« If you click on "Pause" during the automatic scrolling, manual page up and page
down is possible with the "Page up" and "Page down" icons.
Page 36
Page 39
MT-101/MT-200
Viewing and Editing a Recording6
User GuideEvent Chart6.7
6.7Event Chart
« Click on the "Event-chart view" icon.
1
This display mode provides an overview of all events in the 24 hour recording in the
form of bar charts. The X axis gives the period from the beginning to the end of the
recording. The height of the bars (along the Y axis) corresponds to the frequency of
events at a certain point in time. The Y axis is automatically adapted to the number of
events.
V Ventricular arrhythmia - Displaying graphs for VES, couplet, triplet, Vtach.,
bigeminy, trigeminy
SSupraventricular arrhythmia - Displaying graphs for SVES, couplet, triplet,
SVtach., bigeminy, trigeminy
MMiscellaneous - Displaying graphs for marker (event button pressed during
« Click on the "Heartrate trend" icon.
The heart rate view gives the heart rate trend over the entire recording.
Note that the heart rate view can only be displayed from the ECG or event view
1
« Click on the cursor line (1) with the left mouse button to move the line.
The heart rate for the entire recording time is displayed in the lower half of the screen.
In the bottom line, you can find the time axis.
The heart rate is averaged over 1, 2, 5 or 10 minutes (for settings, refer to section see
para. 8.2 Heart Rate Trend, page 72). The red line gives the maximum heart rate, the
black line the average and the blue line the lowest heart rate.
When the heart rate trend is displayed in the ECG view, the time segment is indicated
in the upper window which corresponds to the current position of the vertical cursor
line in the HR trend view. When the vertical cursor line is moved in the HR trend view,
the corresponding ECG segment is changed in the upper window. To move the cursor
line, click on any point in the HR trend view with the left mouse button.
« Return to the zoom view by clicking on the zoom icon.
Page 38
Page 41
MT-101/MT-200
Viewing and Editing a Recording6
User GuideST Trend View6.9
6.9ST Trend View
« Click on the "ST trend" icon.
1
2
The measurements table (2) indicates the measured values belonging to the current
position of the vertical line (1) in the upper window. However, these values are ONLY
displayed when the vertical line is positioned on an ST event (as in the ST-Episode 1
line in the above example). Click on "Begin", "Max ST" or "End" in the measurements
table to display the corresponding section in the zoom view with the respective
measured values.
Art. no.: 2.510492 rev.: b
Page 39
Page 42
6Viewing and Editing a Recording
6.9ST Trend View
ST EpisodesThe top section (1) gives the incidents of artifacts and ST episodes for each of the two
channels (any combination of the three channels - see page 36, “Selecting channels
for analysis”)
ST TrendThe middle section gives a graphical representation of the ST trends for both
channels. The ST level is measured in mm and the slope is measured in mV/s (taken
at the point (J-point + 10 to 100 ms) defined in the menu option “Analysing Options” -
see para. 6.16 Analysing Options, page 56).
The scale of the level (mm) and slope (mV/s) are given on the left hand side of the
trend graph.
MT-101/MT-200
The thicker line (in the example: +
episode detection in the analysis options.
The ST values are coloured green in channel 1 and pink in channel 2. The ST level is
the thicker line; the ST slope is the thinner line for both channels. Any of these four
trends can be displayed or removed by checking the boxes to the right of the ST trend
graph (level 1, slope 1 and level 2, slope 2).
Heart Rate Trend See previous page.
ZoomThis zooms in the time segment corresponding to the current position of the vertical
pointer in the trend graph view (ST and HR). Move within the zoom view with the left
and right keyboard arrow keys, or the forward/backward scroll icons in the toolbar.
ST Measurement Table (2)The "ST-Episode-Properties" table (2) is displayed in the bottom right corner. When
the vertical pointer (line) is positioned at a point where an ST episode occurred, the
table gives the following measured values:
• the start and end times when the ST values exceeded the user defined value
• the ST value in mm at the start and the end of the ST episode sequence and the
highest ST level
• the ST slope in mV/s at the start and the end of the ST episode sequence
• the heart rate at the start and the end of the ST episode sequence and the highest
heart rate
• the duration of the ST episode sequence
1, +2 or +3 mm) indicates the limits defined for ST
Page 40
Moving the vertical pointer:
Position the cursor on any spot in the HR graph, the ST trend graph or on an ST event
in the upper part of the screen and click with the left mouse button. The vertical pointer
will move to the new cursor position.
The limits for ST event detection and the measurement point where the ST values are
measured is defined in the “Analysing Options” menu option - see para. 6.16
Analysing Options, page 56.
Page 43
MT-101/MT-200
Viewing and Editing a Recording6
User GuideTemplate Matching6.10
6.10Template Matching
« Click on the "template" icon.
It is also possible to select a classified QRS complex ("N", "V", "S" etc.) in the ECG
zoom view and to open the template view via right mouse click. The template class of
the selected complex is highlighted.
1
2
3
When an ECG is measured over several hours, differences will naturally occur in the
QRS complexes. QRS complexes with similar shapes can be grouped as one class
(or template). There can be several classes in which the QRS complex is considered
normal, several classes with an abnormal QRS complex (ventricular extrasystoles:
VES classes) and also classes with disturbed QRS complexes. This gives the user a
direct overview of the different kinds of electrical activity in the heart. It is also possible
to immediately reclassify a complete group of QRS complexes that are considered to
be incorrectly classified.
The SCHILLER software can differentiate between twelve normal templates, a
maximum of 300 abnormal VES templates, one artifact template and up to twelve
Art. no.: 2.510492 rev.: b
supraventricular templates. The templates are shown in frequency (number of
occurrences) order. The number of occurrences is displayed in the shaded box in the
bottom right of the template. The first templates displayed are the twelve normal
templates, followed by the VES templates. If less than twelve normal QRS forms fit
the templates, then only those measured are displayed .
Page 41
Page 44
6Viewing and Editing a Recording
6.10Template Matching
Disturbed QRS ComplexesDisturbed complexes that are still considered to be normal and complexes that do not
fit into any other templates are collected under "Normal Various". When the user
manually reclassifies a QRS complex as normal, it is shown in the template as
"Normal Manual".
Template MatchingThe SCHILLER template matching function subdivides the QRS complexes
recognised as normal or SVES into different subgroups. This very exact analysis
differentiates the normal QRS complexes when relatively small shape differences
appear. This enables the reclassification of abnormal complexes that appear very
similar to the normal QRS complexes and may therefore be incorrectly labelled as
normal by the algorithm. They can be reclassified as VES in one group, using a single
command.
Highly Disturbed QRS ComplexesQRS complexes which have such extreme disturbances that no clear allocation as
normal or VES is possible are collected in the artifact template.
Reclassifying TemplatesUsing the right mouse button, each template class can be reclassified as normal,
supraventricular (SVES), VES or artifact. A template class can also be deleted.
Merging TemplatesWhen two templates are assessed to be equal, one template can be "dragged and
dropped" into the other. Hold the left mouse button on the template to be merged. Still
holding the mouse button, move the mouse to position the QRS in the template area
of the similar template. Release the mouse button to merge the two templates. The
following special icons are associated with template matching on the left of the
template view:
MT-101/MT-200
Change the screens with the Page icons on the left of the screen. Including (from the
top:
« previous / next page of templates (last or first page of templates) (1)
« Increase / decrease speed
« Increase / decrease amplitude
« Increase/ decrease distance between the two leads shown on each template
« display 16 / 24 templates per page (2)
To limit the number of VES templates after evaluation, click on the `limit VES`
template icon (3). The templates which are no longer required are collected under the
templates `VES Diverse`. Note that the default number of VES templates is set in the
analysis menu - see para. 6.16 Analysing Options, page 56.
Page 42
Page 45
MT-101/MT-200
Viewing and Editing a Recording6
User GuideTemplate Matching6.10
6.10.1Detailed Overview of the Template Classes
To obtain a detailed view (2) of the template classes click on the number of
occurrences box (1) at the bottom of the templates.
1
2
3
4
Art. no.: 2.510492 rev.: b
A reference QRS complex of the template group is displayed in the bottom left (3). In
this way, the zoomed QRS complex can be visually compared with the reference
complex of the template. The reference complex number is displayed in the bottom
right corner of each template. To select a previous or subsequent complex, click on
the right or left arrow (4). When the template view is exited, the same reference
complex is set on re-entering.
5
Page 43
Page 46
6Viewing and Editing a Recording
6.10Template Matching
MT-101/MT-200
Overview of QRS Complexes in the
Upper Screen
Zoom View of QRS Complexes in the
Lower Screen
Reclassifying a QRS Complex (4)The selected QRS complex in either the top section of the screen or the zoom view
"Measure" Field (5)A measurement box is displayed in the zoom view when the “Measure” box (5) in the
The upper part of the screen displays all complexes of the selected template. A
maximum of 100 QRS complexes can be shown on one page. Select the next/
previous page(s) with the icons on the left (see previous page).
A zoom view of any QRS complex is displayed in the bottom section. To select
another zoom view, position the cursor on any complex in the top section of the
screen and click. The selected section is zoomed in the bottom section.
can be reclassified as normal, SVES, VES or artifact or deleted by clicking with the
right mouse button. It can also be reclassified/deleted directly using the keys “N“, “V“,
“S“, "A", or “D“ (delete).
To reclassify the whole template, position the mouse on the reference QRS
complex (or template class in the template overview), and click with the right
mouse button to display the reclassify window.
All events (couplet, triplet, bigeminy etc.) that have been reclassified manually are
automatically rearranged when the template view is exited and re-entered.
bottom left of the zoom screen is checked. This gives the difference between the two
horizontal blue lines. These values indicate the differences (amplitude and time)
between the two blue lines. The two blue lines can be positioned anywhere on the
zoom view by moving the cursor. The measurement values change instantly (see
section see para. 6.3 Event & Zoom Views, page 29).
The two settings (5) below the "Measure" field - in the bottom left of the zoom screen
- display the measured time and heart rate when checked. They indicate, depending
on the respective selection:
• the time in milliseconds between the respective heart beats.
• the heart rate calculated beat to beat.
Classifying Multiple or Contiguous Complexes
To select multiple complexes press the Control key on your computer (CTRL), and
select the complexes with the mouse.
To select contiguous complexes press the Shift key on your computer (), and select
the complexes with the mouse.
When the complexes have been selected, they can be classified as described. Note
however that cannot be deleted.
Page 44
Page 47
MT-101/MT-200
Viewing and Editing a Recording6
User GuidePacemaker Templates6.11
6.11Pacemaker Templates
« Click on the “Pacemaker View” icon. The pacemaker templates are displayed:
«
I
individual PM templates or the complete class of PM template can be moved into
another PM template using the right mouse key
Art. no.: 2.510492 rev.: b
User defined names can be defined for all pacemaker templates (see para. 8.4
Pacemaker Templates, page 74.)
Page 45
Page 48
6Viewing and Editing a Recording
6.11Pacemaker Templates
MT-101/MT-200
6.11.1Template classes
The pacemaker template option provides an evaluation of the pacemaker impulses
with presentation in several so-called pacemaker templates. Eight different
pacemaker templates are defined:
Template 1:Two pacemaker pulses, P1 and P2, are detected. The distance between the two PM
actions is greater than 100 ms and less than 220 ms. The QRS trigger T is after the
second PM pulse P2.
Template 1a
P1
Trigger
P2
Template 1b
P1
P2
Trigger
Template 2:One pacemaker pulse P1 is detected, and the QRS trigger T is after P1. The distance
between PM pulse and QRS trigger is greater than 150 ms and less than 350 ms.
Template 2c
P1
T
Template 2a
P1
T
Template 2b
P1
T
Template 3:One pacemaker pulse P1 is detected, and the QRS trigger T is after P1. The distance
between PM pulse and QRS trigger is less than 150 ms.
Template 3a
T
P1
Template 3b
P1
T
Page 46
Page 49
Viewing and Editing a Recording6
MT-101/MT-200
Template 4:Two pacemaker pulses, P1 and P2, are detected. The first PM pulse P1 is before the
User GuidePacemaker Templates6.11
QRS trigger T (their distance is greater than 150 ms and less than 350 ms), and the
second PM pulse P2 is after the QRS trigger (at a distance of max. 0.35 times the
average heart rate).
.
Template 4b
P1
T
P2
Template 4a
P1
T
P2
Template 4c
P1
T
P2
Template 5:A pacemaker pulse P1 is detected after the QRS trigger T at a distance of up to 0.35
times the average heart rate (in the ST segment).
Template 5a
T
P1
Template 5b
T
P1
Art. no.: 2.510492 rev.: b
Page 47
Page 50
6Viewing and Editing a Recording
6.11Pacemaker Templates
Template 6:Pacemaker pulse(s) detected between two QRS triggers: the distance of the PM
pulse to the previous QRS trigger is greater than 350 ms and the distance to the next
RS trigger is greater than 0.35 times the average heart rate.
MT-101/MT-200
Template 6
T1
P1
T2
Template PM RestThis template contains all the pacemaker pulses that could not yet be assigned to any
of the other PM templates.
Template ArtifactThe user can assign here all triggers in which a pacemaker action was incorrectly
recognised by the system due to artifacts.
Pacemaker settings and options are described in the settings section - see para. 8
System Settings and Options, page 68.
Page 48
Page 51
MT-101/MT-200
2
1
4
Viewing and Editing a Recording6
User GuideHeart Rate Variability6.12
6.12Heart Rate Variability
« Click on the "HRV" icon.
Select between three different graphical representations of the HRV. Choose one of
the three top icons in the left toolbar (1) as shown below
3
1
2
3
4
5
Art. no.: 2.510492 rev.: b
6
7
8
9
10
1. NN/dNN Tachogram Distribution over the entire recording of absolute and relative intervals
2. SDANN/SDNNidx Graph Distribution over the entire recording
3. NN Distribution Bar graph showing the percentage of NN beats against the
percentage deviation from the mean
4. Update of the graphical representation after, for example, changing a setting
5. Scaling of the X-axis
6. Scaling of the Y-axis
7. Move to the point in the recording of the minimum (shortest) NN interval
8. Move to the point in the program of the next longer NN interval
9. Move to the point in the recording of the maximum (longest) NN interval
10. Move to the point in the program of the next shorter NN interval
Page 49
Page 52
6Viewing and Editing a Recording
6.12Heart Rate Variability
MT-101/MT-200
Heart Rate Variability Analysis Notes
The following parameters are analysed in tabular form for the entire recording (day 1,
2, 3 or all (4), and separately for day and night time (2). The night time is also shaded
on the distribution charts (3). The night time can be individually defined for every
patient - see para. 6.17 Editing Patient Data/Recording, page 58.
• The RR intervals before and after a VES are not considered for the calculation of
the heart rate variability. Only time intervals between normal QRS complexes, i.e.
so-called NN intervals, are used.
• The recording is subdivided into five-minute intervals for the HRV analysis. When
a compliance ratio is less than 80% of NN intervals within a five-minute interval,
this time interval is not used for the calculation.
• When an absolute arrhythmia is detected, the QRS complexes during this absolute
arrhythmia are not accepted for HRV analysis.
• For a correct evaluation, it is important that no normal QRS complexes are overlooked (falsely recognised as abnormal beats). Also, disturbances or VES should
not be incorrectly designated as normal QRS complexes (falsely recognised as
normal beats).
• To enable the user to check this, the following functions are provided in the screen
display: maximum NN interval, subsequent maximum NN interval, minimum NN interval, subsequent minimum NN interval.
Valid NN The percentage of valid NN used in the analysis.
Mean NN The average value of all intervals between consecutive normal QRS complexes within
the entire ECG recording (also specifically for day and night).
SDNN (standard deviation of mean NN)The standard deviation of all intervals between consecutive normal QRS complexes
within the entire ECG recording (also specifically for day and night).
SDANN (standard average deviation of
NN)
SDNNidx (standard deviation of NN intervals)
rMSSD (root mean square of successive differences)
NN50 The sum of all congruent NN with variations greater than 50 ms. The intervals
The standard deviation of calculated average values. After splitting up the entire
recording into 5-minute segments, the average values of the intervals between
consecutive normal QRS complexes within these 5-minute segments are calculated.
After splitting up the entire recording into 5-minute segments, the standard deviations
of the intervals between consecutive normal QRS complexes within these 5-minute
segments are calculated. The average value of these standard deviations is referred
to as SDNNidx.
All differences between consecutive NN intervals are raised to the power of 2 and
summed up, and the sum is normalised to the number of actual differences. The
square root taken from this value gives the rMSSD.
*Parameters analysed according to: Standards of Measurement, Physiological Interpretation, and Clinical Use of Heart Rate Variability; Task Force of The European Society of Cardiology and The North American Society of Pacing and Electrophysiology,
Confidential Draft, July 1994.
between consecutive normal QRS complexes are determined from the entire ECG
recording (NN interval). The differences between successive NN intervals are
calculated and all differences greater than 50 ms summed to give the NN50 value.
pNN50 (proportion of adjacent NN-intervals with more than 50 ms difference)
NN100/NN200Same as NN50 but with a time difference greater than 100 ms and 200 ms
Page 50
Proportion (percentage) of adjacent NN intervals with variations greater than 50 ms
calculated against the total number of NN intervals.
respectively.
Page 53
MT-101/MT-200
Viewing and Editing a Recording6
User GuideHeart Rate Trend6.13
6.13Heart Rate Trend
« Click on the "HRX" icon.
The following screen is displayed:
2
3
The screen shows the HR trend over the selected time period. When the mouse key
is clicked (in the trend view), the instantaneous HR values are displayed (1), and the
HR segment is displayed in the zoom view.
1
6.13.1Jumping to the max/min Heart Rate or max/min NN Interval
To jump to the maximum or minimum HR or NN interval, click on the appropriate icon
(2).
To go the ‘next’ highest or next lowest HR or NN interval, click the relevant icons (Abs
>> HR, Abs << HR, Abs >> NN, Abs << NN).
Art. no.: 2.510492 rev.: b
6.13.2Redefining the Max/Min Heart Rate and NN Interval
The maximum and minimum heart rate, and the maximum and minimum NN interval,
can be manually set. To do this, position the cursor (1) and click the relevant icon (set
max HR, set min HR, set max NN, set min NN). When the Maximum and minimum
HR is manually set, the Max and Min HR indication (3) are reset.
The edited maximum / minimum values are taken over in all other screens e.g templates, event view etc. If the recording is re-analysed, the edited values are lost and
replaced by the analysis program values.
Page 51
Page 54
6Viewing and Editing a Recording
6.14Reclassifying/Editing a QRS Complex
6.14Reclassifying/Editing a QRS Complex
In the zoom view (which can be selected when in rhythm view and event view and is
always displayed in ST trend and event samples view), it is possible to reclassify a
QRS complex. A complex can be selected by means of two horizontal thick blue lines
that are positioned in the zoom view above and below the two channels. To reclassify
an individual complex, proceed as follows:
MT-101/MT-200
1
1. In the zoom view, click on the QRS complex you want to reclassify. The position
is marked by blue horizontal lines (1) above and below.
2. Choose from one of the following :
– double click with the left mouse button to display this screen
–or
– display the "Classify" menu by clicking the right mouse button or opening the
"Classify" menu in the main menu bar.
3. Select the classification applying to the selected complex.
4. The reclassification is carried out immediately, i.e. the colour will be changed and
the complex will be assigned to the new class.
Page 52
Page 55
MT-101/MT-200
Viewing and Editing a Recording6
User GuideReclassifying/Editing a QRS Complex6.14
Reclassifying a Group of QRS Complexes
A group of QRS complexes can be classified as follows:
1. Enter the ECG view and in the zoom view click the right mouse button at the be-
ginning of the event. Select "Begin Event" in the displayed menu.
2. Go to the position in the recording where the event ended.
3. In the zoom view click the right mouse button and select "End Event". Select
"Normal", "VES", "SVES", "Artifact" etc.
The "VES > Artifact, SVES > Normal" option will reclassify all VES to artifacts and all
SVES to normal QRS complexes in the selected time period.
Direct Keys
1. Position the two blue horizontal lines above and below the complex (as detailed
above).
2. Press one of the following direct keys to reclassify the complex.
– KEY `N`Classify the selected QRS complex as NORMAL
– KEY `V`Classify the selected QRS complex as VES
– KEY `S`Classify the selected QRS complex as SVES
– KEY `A`Classify the selected QRS complex as artifact
– KEY `D`Delete the selected QRS complex
– KEY `B`Mark the selected QRS as the beginning of an event
– KEY `E`Mark the selected QRS as the end of an event
– KEY ‘DEL’ delete the selected QRS complex
– KEY ‘+’ jump to next event
– KEY ‘-’ jump to previous event
Art. no.: 2.510492 rev.: b
Page 53
Page 56
6Viewing and Editing a Recording
6.15Analysing/Re-analysing the Recording
MT-101/MT-200
6.15Analysing/Re-analysing the Recording
Analysis is initiated with the "Analyse" icon.
Further analysing options are given in the "Analyse" menu
.
Analysis is only possible on one or two channels - see para. 6.6 ECG View, page 35.
Analyse ECGSelect this option to analyse a new recording or to re-analyse the current file after
changing any settings. The analysis of a complete recording may take several
minutes. During analysis a progress bar is displayed
Analyse ECG-EventsSelect this option to recalculate all event information based on the defined analysis
parameters. Selecting this analysing option is quicker than the analysis of the
complete ECG. QRS complexes which have been manually reclassified remain
unchanged.
Analyse ST-Trends/EpisodesSelect this option to recalculate all ST-trend data (levels and slopes) and episodes.
Selecting this analysing option is quicker than the analysis of the complete ECG
V When ‘Analyse ECG’ is selected manually edited complexes are lost. When
‘analyse ECG events or ‘analyse ST-trends/episode’ are selected, manually
edited complexes remain.
Analyse Multiple RecordingsThis option gives you the possibility of analysing selected recordings using the same
analysis settings. When this menu option is selected, all patients in the MT-200
directory are displayed. Check all recordings that you wish to analyse. At the end of
the analysis, a report is displayed with information about the number of successful
recording analysed.
Page 54
Page 57
MT-101/MT-200
Viewing and Editing a Recording6
User GuideAnalysing/Re-analysing the Recording6.15
Analysis of a single recording can take many minutes. The multiple analyse option
can be used therefore, to analyse many recordings at quiet times e.g overnight
Art. no.: 2.510492 rev.: b
Page 55
Page 58
6Viewing and Editing a Recording
6.16Analysing Options
6.16Analysing Options
6.16.1Arrhythmias
MT-101/MT-200
When ‘Analyse Options‘ is selected from the Analyse menu (previous page), the
options are displayed which define the calculation analysis parameters used for all
new recordings, and when re-analysing a previous recording. Settings are given for
arrhythmia analysis options, for ST episodes and for the number of VES templates
that will be generated.
The following arrhythmia analysis options can be set:
Heart rate
Define the settings for the detection of bradycardias, tachycardias and pauses. The
program detects a bradycardia when the heart rate falls below a set value. It detects
a tachycardia when the heart rate lies above a set value. The program detects a
pause when the set time interval is exceeded.
Abs. Arrhythmia
V Absolute Arrhythmia can be an indication of atrial flutter or atrial fibrillation.
V When absolute arrhythmia detection and supraventricular detection is required,
the percentage setting must be the same for both.
Absolute arrhythmia - Enable absolute arrhythmia detection by checking the
"Detection" box. The percentage difference of the RR intervals can be set in the range
from 15 % to 50%.
Minimal Duration - The time duration above which absolute arrhythmia will be
classified. The minimal duration can be set from 10s to 10 minutes.
Important
When absolute arrhythmia is detected by the evaluation software, this can be a sign
of atrial fibrillation or flutter.
Supraventricular Arrhythmias
Supraventricular arrhythmia detection is enabled when the “Detection” box is
checked. The percentage difference at which a supraventricular arrhythmia is defined
is set in the range of 15% to 50%.
6.16.2Manually Defining Arrhythmias
Arrhythmias can be defined as atrial flutter or atrial fibrillation as follows. Events can
also be deleted at any time
Page 56
1. Enter the ECG view and in the zoom view click the right mouse button at the beginning of the arrhythmia. In the displayed menu, select "Begin Event".
2. Go to the position in the recording where the arrhythmia ends.
3. In the zoom view click the right mouse button. In the displayed menu select "End
Event". A number of classification options are available. Select "Atrial Flutter" or
"Atrial Fibrillation" as required.
4. Select "Delete Event" to delete the event.
Page 59
MT-101/MT-200
User GuideAnalysing Options6.16
6.16.3ST-episodes
The following analysis options are available for ST episodes:
•ST-Point
• ST-Episodes Ch.1
• ST-Episodes Ch.2
• ST-Episodes Ch.3
The ST point and ST-level limit can only be defined for the selected channels. As
channels 2 and 3 have been analysed in the figure to the right, the option is not active
for channel 1.
Note: In the MT-200 analysis software, the J-point is located 56 ms after the trigger
point (the point in the QRS complex at which the slope is greatest). The ST-point is
located x ms after the J-point, where x is a value in the range of 10 ms to 100 ms (see
above). Thus, when the heart rate changes significantly, correct values of ST-segment
shifts can only be obtained with an adaptation of the ST measurement point (the
higher the heart rate, the smaller the value x ms after the J-point).
6.16.4Templates
The number of templates to be defined are set here.
Viewing and Editing a Recording6
6.16.5Mode
The SAM program is currently under test and this option is only displayed when
the SAM dll file is installed.
This defines the analyse program used for analysing a recording. The Monec program
is the older program and the SAM the newer program currently under development.
Using either program will make no difference to the analysed data. However at the
current time the Monec software is the only authorised analysis software and must be
selected.
Art. no.: 2.510492 rev.: b
Page 57
Page 60
6Viewing and Editing a Recording
6.17Editing Patient Data/Recording
6.17Editing Patient Data/Recording
MT-101/MT-200
Enter the "Patient/Recording Data" window to carry out the following:
• Enter or edit patient data - name, weight, height, DOB etc.
• Enter the date/time of the recording
• View the settings that were used for the analysis of the current recording
• Enter medication, diagnosis, comments etc.
To view the above data, select in the "Patient" menu the option "Patient/Recording
Data".
Six entry options for editing patient data are available in the window, as shown on the
following pages.
Patient datIn this screen the patient details at the time of the recording can be viewed, entered
or edited
1
– Last name, first name
– Pat. number
– Date of birth*
– Gender
– Height, weight
– Pacemaker - fitted or not]
– Station
– Address and telephone number
The play button (1) is active when the patient ID has been entered before the
recording was commenced. Click the button to hear the recorded ID.
If pacemaker pulses are to be displayed in the recording, the “Pacemaker” box must
be checked before the recording is made - see para. 5.1.1 Pacemaker, page 21.
Page 58
• * If the year is entered using two digits only, the program calculates the date of birth
automatically based on the current PC date settings. The patient's age is in this
case always between 1 and 99 years.
Page 61
Viewing and Editing a Recording6
MT-101/MT-200
Recording dataIn this screen, you can enter/edit the date and time of the recording. On the right hand
AssignmentIn this screen, you can enter/edit referral details and medication.
User GuideEditing Patient Data/Recording6.17
• If the year is entered using four digits, patients can be 100 years or older.
In both cases, the program stores the date of birth as a four digit value.
side, you can also see the analysis settings for the recording analysis.
DiagnosisThis gives a text entry field for entry of diagnostic analysis and/or comments. When a
printout of the analysis summary is obtained, the diagnosis and comments are
included on the same page.
HRV Day/NightThis defines the start of the day and night periods for HRV analysis.
Art. no.: 2.510492 rev.: b
Page 59
Page 62
6Viewing and Editing a Recording
6.17Editing Patient Data/Recording
Cover LettersThis defines the recipients and addresses of the cover letter and the information
categories to be included in the letter. Note that the office address printed with the
cover letter, is entered in system setting - see para. 8.5 System Settings, page 75
MT-101/MT-200
Page 60
Page 63
MT-101/MT-200
Viewing and Editing a Recording6
User GuidePrinting6.18
6.18Printing
6.18.1Print preview / Printing a specific page
This function enables the display of the entire printout page by page before printing it.
« Click on the "Print Preview " icon or
« select in the "Patient" menu the option "Print Preview (User-defined)"
« Select the required data and click on "OK" The selected data can then be viewed
page by page in printout format.
Art. no.: 2.510492 rev.: b
« For other printout options, see para. 8.1 Print Formats, page 68
« For a description of the printout data see para. 8.1.4 User-defined print formats,
page 70.
Page 61
Page 64
6Viewing and Editing a Recording
6.18Printing
6.18.2Obtaining a printout:
MT-101/MT-200
To obtain a printout of the displayed recording,
« Click on the "Print" icon or
« select in the "Patient" menu the option "Print"
The printout will contain all data defined as standard. The standard printout settings
are defined in the menu “Options” > “Print Formats” see para. 8.1 Print Formats, page
68). There is also a preview option for all print layouts for individual selection before
printing. This function is called “Show print-formats before printing” and when
checked enables individual selection of data before printing.
Several print settings are available depending on the installed printer drivers. The MT200 program adopts the driver print settings. The various options are mostly self
explanatory and are fully explained in the handbook for your computer or operating
system.
6.18.3Printing a selected half hour ECG segment
1. In the ECG view, use the mouse to highlight an ECG segment (1).
2. First click on the Centre icon (2), and then on the Print direct (1/2 h) icon (3).
A half hour segment of the ECG recording is printed with the segment selected in the
middle of the printed page; i.e. approximately a 1/4 hour on either side of the selected
segment is printed
2
3
1
Page 62
Page 65
MT-101/MT-200
Miscellaneous Functions7
User GuideE-Mail and PDF Functions7.1
7Miscellaneous Functions
7.1E-Mail and PDF Functions
7.1.1PDF files with Acrobat Reader
The MT-200 pdf function will only work with Acrobat Reader version 4.0 or later. This
is available on the SCHILLER software CD and must be installed before pdf files can
be generated or viewed. If an earlier version of Acrobat Reader is installed on your
computer, it must first be uninstalled before installing V4.0 Proceed as follows:
1. Access the Windows uninstall program under Desktop/My computer/control panel/ > add/remove programs.
2. In the list, select Acrobat and click the "add/remove" icon. You will be prompted
to confirm deletion.
3. When the old version has been uninstalled, the new version can be installed from
the SCHILLER CD-ROM in the required language.
7.1.2Editing PDF files
PDF files generated with the MT-200 program can be edited with Adobe Acrobat
(version 4.0). To do this, the pdf file must be opened in Adobe Acrobat. Close the file
again in Adobe Acrobat before opening the MT-200 program. If Adobe Acrobat is
opened when using the MT-200 software, it is not possible to open the pdf files in the
MT-200 software.
Art. no.: 2.510492 rev.: b
Page 63
Page 66
7Miscellaneous Functions
7.2Saving a Recording
7.2.1Saving a recording in MT-200 or PDF format
MT-101/MT-200
7.2Saving a Recording
Recordings can be saved in original MT-200 or in pdf format. The advantage of pdf
format is that it is compressed and therefore the stored recordings take significantly
less space on the harddisk than normally stored recordings.
Three methods are available to store a recording as follows:
1. With the mouse click on the "Save recording" icon.
Clicking on the icon itself saves the recording in original MT-200 format. Clicking on
the arrow to the right gives you the options of "Save" (in original MT-200 format) and
"Save as a PDF File".
2. Open the "Patient" menu with the mouse and click on "Save" or "Save as PDF
File".
3. Or hold the Control key pressed and press the "S" key on your keyboard. This
saves the files in original MT-200 format.
When a recording is saved as a pdf file, you are prompted to select specific data for
saving.
The Adobe Acrobat program is opened automatically and the recording is saved as a
pdf file. You are then prompted to discard the original file if required.
Page 64
Page 67
MT-101/MT-200
Miscellaneous Functions7
User GuideSaving a Recording7.2
7.2.2Sending a recording by e-mail
Recordings (original MT-200 or pdf format) can be sent over the Internet by e-mail.
The compressed pdf files are much smaller and can therefore be sent in less time.
Only the currently displayed recording can be sent by e-mail. Click on ‘Send EMail‘
(original MT-200 file) or "Send EMail as PDF File" in the Patient menu. You are then
prompted to enter the recipient's address. Click on "OK" to send.
7.2.3Importing recordings
Recordings can be imported from floppy disk, CD-ROM, network drives etc. Select
‘Import‘ (see above), to import the data from the source to the target directory. The
target directory is defined under "Options/System/Directories".
Select "Import/MT-101 RAW - Data Files" to import the files from the SD card.
7.2.4Exporting recordings
Recordings can be exported to SD card, CD-ROM, network drives, or any other
storage medium. Select the "Export" option in the "Patient" menu. There are three
different export options: "Current recording", "Not yet exported recordings" and
"Select recordings". The data is exported to the drive and directory defined under
"Options > System > Directories > Export Directory.
When the option "Export/RR distances" is selected, a text file containing the
numerical values of the RR distance is exported.
When the option "Export/Enabled cover letters" is selected, the cover letter can be
exported as an rtf file and processed in a word processing system.
Art. no.: 2.510492 rev.: b
Page 65
Page 68
7Miscellaneous Functions
7.3Deleting a Recording
MT-101/MT-200
7.3Deleting a Recording
Recordings can be automatically deleted when between 5 and 100 have been stored.
The auto deletion function is detailed in the settings section - see para. 8.5.4 Data
storage mode (auto delete), page 76.
A complete 3-channel 24 hour recording requires approximately 16 Mbytes of storage
space on the harddisk. This means that your harddisk can quickly fill if recordings are
not deleted when no longer required.
For every recording, there are three files (*.cmp, *.qrs and *.bak). All files have to be
deleted. Proceed as follows:
1. Open Windows Explorer.
2. Go to “Folders“ and search for the MT-200 folder. Open the "Data" sub-folder
containing the patient data. (If you have specified another location for the storage
or recordings, open that folder).
3. Select in turn the *.cmp, *.qrs and *.bak files of the recording you wish to delete.
4. Press the right mouse button to display the Explorer menu.
5. Select the "Delete" option and click with the right mouse button.
6. If you are sure you no longer need the recording, click on "Yes". Empty the
Recycle Bin.
Page 66
Page 69
Miscellaneous Functions7
MT-101/MT-200
User GuideAccelerator Keys7.4
7.4Accelerator Keys
A number of keys and key sequences exist to enable you to quickly carry out
commonly used program functions. An overview of these keys is given below:
Key Function
Ctrl + NNew patient - entry of patient data
Ctrl + OOpening the list of saved recordings
Ctrl + SSave of the currently displayed recording
Ctrl + GMove to the beginning of the recording
F8Move to previous event.
F9Move to next event.
ArrowScrolling back/forward in the recording (right/left keys)
F5Display of event view
F6ECG (24 h) view
F7Display of event chart (graph) view
Ctrl + F11
Ctrl + F12
Alt + F12
Alt + F11
Shift + F11Move to the segment in which the minimum HR has been detected
Shift + F12Move to the segment in which the maximum HR has been detected
F12
F11
NClassification of the selected QRS complex as normal
VClassification of the selected QRS complex as VES
SClassification of the selected QRS complex as SVES
AClassification of the selected QRS complex as artifact
BIdentification of the highlighted complex as the beginning of an event
EIdentification of the highlighted complex as the end of an event
+Jump to next event
Art. no.: 2.510492 rev.: b
-Jump to previous event
Move to the segment in which the minimum sinus HR has been detected
Move to the segment in which the maximum sinus HR has been detected
Move to the segment in which the minimum NN interval has been detected
Move to the segment in which the maximum NN interval has been detected
Move to the segment in which the minimum RR interval has been detected
Move to the segment in which the maximum RR interval has been detected
Page 67
Page 70
8System Settings and Options
8.1Print Formats
8System Settings and
8.1Print Formats
MT-101/MT-200
Options
In the "Options/Print Formats" menu, you can define the default print data, i.e. the
information defined here is remembered when the program is next opened
If you wish to display the data options for individual definition before printout, this must
be set in the system menu: Options > System > Print Setup, check the "Show printformats before printing" box.
Four tabs are given at the top of the "Print Formats" window as follows
« Templates
« Event Samples
« Full Disclosure (1 channel, 2 channels or 3 channels)
« User defined
Click on the tab required to define the desired settings.
8.1.1Templates
When the "Templates" tab is clicked, the following options are given.
Select the required template classes to be printed.
Page 68
Page 71
MT-101/MT-200
System Settings and Options8
User GuidePrint Formats8.1
8.1.2Event samples
In this window, event categories are defined for sample event printing.
A maximum of three events per category can be printed. If more than three events
have been recorded for an event category, the first recorded event of every eight-hour
cycle will be printed. Event-samples selected by the user (in the event-samples view)
are printed as defined.
Check the relevant boxes. The amplitude and speed refer to the print size. If “Auto” is
selected for the amplitude, the program automatically selects the optimal setting.
8.1.3Full disclosure (1, 2 or 3 channel)
Here you can select to print the complete 24 hour recording or selected segments.
Click on "Whole Recording" or on "Selected periods of time". With the latter selection,
the half hour segments to be printed can be selected.
Note
The recorded half hour segments apply to the currently displayed ECG. When no
Holter recording is displayed, no time segments can be selected.
In the "Channels" box at the bottom right, you can select channel 1, channel 2 or both
channels for the ECG full disclosure.
Art. no.: 2.510492 rev.: b
Page 69
Page 72
8System Settings and Options
8.1Print Formats
8.1.4User-defined print formats
MT-101/MT-200
In this window, the data sections to be printed are defined.
Note
A header is given on every printed page, giving the following information:
• User identification - entered in the "Options" menu"
• Date and time of the recording
• Patient name and number
The following data can be selected for print:
Cover LetterThe cover letter see para. 3.2.1 During the Recording and Patient Information, page
17.
Holter reportThis gives a statistical overview of the recording and includes:
• Name, sex, date of birth, age, weight and height of the patient
• Start/stop time and duration of the recording
• Summary of the heart rate analysis: Total number of beats; max., average and min.
heart rate with time of occurrence; total number, duration and time of the respective
bradycardia, tachycardia and pause; number of min. and max. RR intervals with
time of occurrence
• Sinus rhythm analysis: time of occurrence of the max. and min. heart rate
• SVES analysis (SVES, SV tachycardia): total number and max. number in one
hour and other statistical data; occurrence and duration of the longest and highest
SV tachycardia
• VES analysis (couplet, triplet, V tachycardia, bigeminy, trigeminy): total number
and max. number in one hour and other statistical data; occurrence and duration
of the longest and highest V tachycardia, bigeminy, trigeminy.
• Manually entered diagnosis and assignment information (see section see para.
6.17 Editing Patient Data/Recording, page 58.
• HR Trend/Events
This page corresponds to the event view and gives:
• An overview of all recorded events of the 24 hour recording in tabular form
• The heart rate trend over the entire recording giving the max., min. and average
heart rates
Analysis SummaryStatistical overview for every hour of the recording for:
• Heart rate - total number of beats, min./max. rate, number of tachycardias, brady-
cardias and pauses
• Supraventricular arrhythmias - number of couplets, triplets, SV tachycardias,
bigem. and trigem.
• Ventricular arrhythmias - number of couplets, triplets, SV tachycardias, bigem. and
trigem., R on T
Page 70
Page 73
System Settings and Options8
MT-101/MT-200
Event Chart ReportGraphical view for all event categories giving the number of events in every hour of
Min/Max HR/RRZoom views of the recording segments in which the max. and min. heart rates as well
Event SamplesFor each event category, up to three events are printed. If more than three events
User SelectedZoom print of a section of the recording that has been selected by the user.
User GuidePrint Formats8.1
• Manual event recording/lead-off/artifacts - number
the recording
as the max. and min. RR intervals were detected
have been recorded for an event category, the first recorded event of every eight-hour
cycle will be printed. Event-samples selected by the user (in the event samples view)
are printed as defined.
As many or as few event categories can be printed as required according to the
definitions for event samples (see "Event Samples" on previous pages).
The user selection is made:
• in the "Edit" menu - "Select User Sample" or
• in the zoom view, by positioning the cursor on the desired complex and clicking on
the "Select" (print) icon.
To undo the selection, click on the menu "Edit/Delete all user samples".
Full DisclosurePrintout of the complete recording or of selected 30 minute sections (one page per 30
minutes).
You can select channel 1, 2 or 3 or any combination.
ST Analysis Overview and TrendThe following information will be printed:
• Measurement table of all ST episodes
• Trend of the ST amplitude and ST slope for channels 1 and 2
• Heart rate trend (max., min. and average HR)
ST Episode SamplesThe four ST episodes with the highest recorded ST levels are given (with the highest
recorded ST level printed first). Each sample gives the maximum level reached during
the episode, and the slope, the heart rate and the time of occurrence.
MarkerPrints all manual markers (set during the recording when the patient presses the
Art. no.: 2.510492 rev.: b
event button on the MT-101).
Page 71
Page 74
8System Settings and Options
8.2Heart Rate Trend
8.2Heart Rate Trend
MT-101/MT-200
The heart rate is calculated over 4, 8 or 16 beats and averaged over 1, 2, 5 or 10
minutes. To define the corresponding settings, select the "Heart Rate/Trend" option in
the "Options" menu.
1. Select the relevant settings.
2. Click "OK" to confirm, or "Cancel" to keep the current settings.
The definitions for the calculation are applied to the entire recording.
The heart rate is calculated as follows:
• The heart rate is calculated (as beats/minute) over either 4, 8 or 16 contiguous continuously averaged heart beats as shown in the example below.
• The calculated maximum and minimum of these averaged heart rates are given as
the highest and lowest heart rates measured over 1, 2, 5 or 10 minutes in the HR
trend diagram.
• The average heart rate is calculated from the individual heart rates in the trend
graph.
Example
The heart rate setting is 8; the average setting is 2 minutes:
Page 72
Page 75
MT-101/MT-200
Event samples
System Settings and Options8
User GuideAmplitude/Speed8.3
8.3Amplitude/Speed
At the top of the window are four icons which select sub screens for data entry:
This refers to the event samples view. Three different speeds and three different
amplitudes are available.
ECG window
Zoom window
Live data
Art. no.: 2.510492 rev.: b
This refers to the ECG view.
Four different speeds and three different amplitudes are available. Note that the larger
two amplitudes (2.5 & 5 mm/mV) can only be set when one of the two higher speeds
is set (i.e. 6.25 or 12.5 mm/s). Similarly, the smaller two amplitudes (1.25 and 2.5 mm/
mV) can only be set when a speed of 3 mm/s or 4 mm/s is selected.
This refers to the Zoom view
Three different speeds and three amplitudes are available.
This refers to the monitor screen displayed when the Holter unit is connected to the
PC before commencing a 24 hour recording. Only two amplitudes can be set.
Page 73
Page 76
8System Settings and Options
8.4Pacemaker Templates
8.4Pacemaker Templates
MT-101/MT-200
Here the user can define more meaningful individual names to the pacemaker
templates in place of the defaults, such as "DDD pacemaker template" instead of "PM
Template 1 A". Any title up to 20 characters can be defined for each template.
Page 74
Page 77
MT-101/MT-200
System Settings and Options8
User GuideSystem Settings8.5
8.5System Settings
Access the system options by clicking on "System" in the "Options" menu.
Seven tabs are given at the top of the System window as follows
« Print Setup
« Language/ Units
« Directories
« Data Storage Mode
« Holter
« GDT
« Office Address
Click on the tab required to defined the desired settings.
8.5.1Print setup
Also see see para. 8.5.1 Print setup, page 75.
"Print Setup" gives the following options:
MarginsHere, the free space between the top of the page and the top of
the printing and between the left margin of the page and the left
margin of the printing can be defined.
Font SizeSet to your preference. The actual print size depends on your
printer. Some experiment may be necessary to determine your
preference. The font size defined here does not affect the size
of the data printout.
Show print formats Check the ‘Show print-formats before printing‘ box to display the
print format dialogue before printing. When the box is checked,
the user can select the data groups to be printed before every
printout.
8.5.2Units and language
Use this screen to define the units that you wish to use in the MT-200 program for
patient data. The language setting refers to the printout and program.
Art. no.: 2.510492 rev.: b
UnitsSelect between cm/kg or ins/lbs.
LanguageSelect the desired language
Cable setSelect the correct patient cable
Page 75
Page 78
8System Settings and Options
8.5System Settings
8.5.3Directories
8.5.4Data storage mode (auto delete)
MT-101/MT-200
Use the directories screen to define the directory/folder where you wish to store
recordings and where you want to export recording data. You can of course select SD
card or any another storage medium if required. Enter the new directory paths as
required.
A complete 3-channel 24 hour recording requires approximately 16 Mbytes of storage
space which is stored automatically after downloading. This means that your harddisk
can quickly fill if recordings are not deleted when they are no longer required. The auto
delete function allows the number of saved recordings to be automatically limited (in
the range of 5 to 100). When the set limit (of e.g. 20) has been reached and a new
recording is made, the oldest saved recording is deleted automatically.
« Check the "Auto-Delete" box and select the number of recordings that you wish
to have stored on your system before overwriting.
8.5.5USB / AT-card Connection and Test transmission Mode
The type of installation (USB converter or AT card) is defined on installation see para.
10 Installation, page 81. The address of the AT card is also defined on installation. Use
this screen to check the setting and redefine if necessary.
Checking the Connection between the Holter and PC
Ensure that the correct box is checked for the type of installation (USB Converter or
AT card), and click the `Test Connection` tab to check the integrity of the opto cable
transmission. Follow the instruction on the screen.
Page 76
Page 79
MT-101/MT-200
System Settings and Options8
User GuideSystem Settings8.5
8.5.6GDT
Use the GDT screen to define the default directory where you wish to store recordings
and export recordings when in a GDT network. Check the “GDT Directory” box, and
enter the directory and abbreviations.
8.5.7Office Address
Define the address that will be printed on the cover letter
Art. no.: 2.510492 rev.: b
Page 77
Page 80
8System Settings and Options
8.6User Identification
8.6User Identification
MT-101/MT-200
The user identification is printed at the top of every page. This can be the name of the
physician, the user, the department etc.
1. Select "User identification" in the "Options" menu.
2. Enter the desired user identification.
3. Click "OK" to confirm, or "Cancel" to keep the current settings.
Page 78
Page 81
Maintenance9
MT-101/MT-200
User GuideVisual Inspection9.1
9Maintenance
All maintenance work must be carried out by a qualified technician authorised by
SCHILLER AG. Only maintenance procedures given in this book, for example battery replacement, visual inspection, etc., may be carried out by the user.
The following table indicates the maintenance intervals, the maintenance requirement, and the person authorised to carry out the procedure.
IntervalMaintenanceResponsible
Every 6 months
Every 12 months
Every 24 months
• Visual inspection of the unit and cables (see below)
• All maintenance work performed at the six month interval.
• Function inspections according to the test instructions P300-01.
• Safety test according to EN 60601-1 (1990), clause 20 according to the test
instructions P300-02.
• All maintenance work performed at the twelve month interval.
• All measurement inspections and calibration according to the test instructions.
« User
« Service staff author-
ised by SCHILLER AG
« Service staff author-
ised by SCHILLER AG
9.1Visual Inspection
Visually inspect the unit and cable assemblies for the following:
« Device casing not broken or cracked
« LCD screen not broken or cracked
« Electrode cable sheathing and connectors undamaged. No kinks in the cable.
« USB cable sheathing and connectors undamaged. No kinks in the cable.
« Input/output connector undamaged.
In addition, at the same time as the visual inspection, the MT-101 should be switched
on, the menu scrolled through, and some sample functions tested - see para. 2.4.1
Menu Overview, page 12. This will:
« provide a basic software integrity check
« check the LCD display
« ensure correct operation of the two control keys
« Defective units or damaged cables must be replaced immediately.
Art. no.: 2.510492 rev.: b
Page 79
Page 82
9Maintenance
9.2Cleaning the device and cable assemblies
9.2Cleaning the device and cable assemblies
V Do not immerse the unit or the cable and sensors in liquid.
V Do not use aggressive cleaners.
9.2.1Cleaning the device, electrode cable, and USB cable
« Wipe with a dampened cloth and a mild cleaning solution. The manufacturer
recommends a 70% alcohol solution.
MT-101/MT-200
Page 80
Page 83
MT-101/MT-200
User GuideSystem Requirements10.1
10Installation
This section is only applicable for initial installation. If the MT-200 program is already
installed on the CS-200 diagnosis system or on your computer, skip this section.
10.1System Requirements
To install the MT-200 program, the system must fulfil the following minimum
requirements:
• BM-PC or compatible system (800 MHz or more) and at least 256 MB RAM.
• At least 1 GB of free disk space.
• XGA 1024 x 786 resolution or higher.
• USB port version 1.1.
• Windows
The MT-200 program package comprises the following:
• MT-200 CD-ROM
• Hard-lock key (special code connector that must be connected to the parallel port
for the program to work)
®
95 operating system or newer version.
Installation10
10.2Installation of MT-200 General Network
License
• The general network license enables full use of the program. A demo version of
the software is installed when the MT-200 program is installed without the license
number or hardlock key.
• The installation of MT-200 licenses is possible under Windows 98, Windows 2000
and Windows XP for MT-200 software version 1.52 (or above). A correct installation requires an installed Internet Explorer version 4.0 (or above).
• First install the MT-200 software (on each PC). After this, install the MT-200 gen-
eral license on each PC of the network, from the special general license disk.
• The general license number is given in the menu "Help" > "About MT-200". Addi-
tionally, it is indicated on the bottom of each page of print previews and printouts.
10.3Network Licence Option
• The software for the second user works like the complete MT-200 Holter version.
However, the software network license cannot read data from the MT-101 Holter
recorder.
Art. no.: 2.510492 rev.: b
10.4Unpacking
Refer to the software license agreement. By opening the package, you accept the
content of the agreement.
Page 81
Page 84
10Installation
10.4Unpacking
MT-101/MT-200
To prevent any damage of the hard-lock key caused by electrostatic charge,
neutralise the static of your body before removing the card from its cover by:
• touching the computer's metal casing or
• wearing around your wrist a special band connected with the computer casing for
isolation against the earth.
1. Remove the hard-lock key from the box.
2. Remove the antistatic protection cover.
3. Put the package material back into the box and keep it at a safe place.
If you have to return the CD-ROM or the hard-lock key to SCHILLER, the original
package must be used as otherwise, the warranty will be void.
4. Examine the CD-ROM and the hard-lock key for visible failings.
Report any damage immediately to your SCHILLER representative.
Page 82
Page 85
MT-101/MT-200
Installation10
User GuideInstalling the Hard-Lock Key10.5
10.5Installing the Hard-Lock Key
The hard-lock key is a plug which contains a code to enable the program to be used.
It is connected to the parallel port of your computer. The hard-lock key can be inserted
at any time but we recommend that it is inserted before installing the program.
« Before switching your computer on, insert the hard-lock key in the parallel port of
your computer (on the back panel).
If your printer is connected to the parallel port, the hard-lock key is inserted between
the computer's printer connector and the printer cable.
The picture given is an example only. For the exact location of the parallel port
connector on your computer, refer to the computer's handbook.
Art. no.: 2.510492 rev.: b
Page 83
Page 86
10Installation
10.6Installing the MT-200 Program from the CD
10.6Installing the MT-200 Program from the CD
1. Insert the CD into the CD-ROM drive.
2. On the CD open the MT-200 program folder and ‘double click‘ elect the start
program MT-200_x_xx.exe. (where x is the software release)
3. Follow the instructions on the screen to install the program
« If you work with an MT-101 Holter, check "USB". MT1-4 AT Card is only selected
if you install this software as an update of a PC with a previous version.
« When the computer reopens, the MT-200 icon appears on the desktop.
« Double click on the icon. The system only requires a few seconds to open the
program.
« To enable the connection of the MT-101 to the USB, the required driver must be
installed from the CD. See next section.
10.7Installing a USB Driver
When the MT-101 is first connected to a computer, the operating system will
automatically detect the connection and open the installation program.
MT-101/MT-200
1. Turn on the computer.
2. Switch on the MT-101 Holter and connect the USB cable to the USB connector.
The dialogue assistant for new hardware is displayed.
3. Check the "Search for a suitable driver..." box and confirm with "Next".
4. Check the "Specify a location" box and select the file "UBcMt100" in the folder
"MT-200/982k" on the CD and confirm. The driver will be copied to your PC.
Page 84
Page 87
Technical Data11
MT-101/MT-200
User GuideMicrovit MT-10111.1
11Technical Data
11.1Microvit MT-101
ManufacturerSCHILLER AG
Device nameMicrovit MT-101 Holter
Dimensions94 x 61 x 20 mm
Weight110 grams (with battery)
Battery type1 x AA/LR6 alkaline 1.5 V or 1 x accumulator NiMH 1.2 V, > 2100 mAh
Power consumption95 mW (recording)
Data acquisitionSimultaneous sampling of two or three channels, differential
Data transmissionBuilt-in USB1.1 interface
Memory mediumSD memory card 64 MB
Environmental conditionsTemperature, operating:10 °C to 40 °C
Temperature, storage:10 °C to 50°
Relative humidity:25 % to 95% (non condensing)
Atmospheric pressure:700 hPa to 1060 hPa
Digital resolution:2.5 µV 12 bit
Dynamic range:±5.12 mV AC
Max. electrode potential: ±300 mV DC
Frequency response:0.05 to 150 Hz (-3 dB)
Input impedance:>10 MOhms
Pacemaker detection± 2...200 mV/0.1...2 ms
Common mode rejection:>80 dB
DisplayLCD 98 x 64 dots
Display speed12.5/20/35 mm/s
Max. displayed amplitude5 mVpp
Dialogue languageGerman, English, French, Italian, Spanish, Portuguese, Russian, Swedish
Patient input protectionFully floating and isolated, defibrillation protection 5 KV
ConformityCE according the directive 93/42/EEC, class IIa
Page 85
Page 88
11Technical Data
11.2MT-200 Software
11.2MT-200 Software
Requirements on the PC
Operating systemWINDOWS 95, 98 SE, 2000 or XP
ProcessorPentium 800 MHz or higher
Harddisk capacityMin. 4 GB (10 GB recommended)
Memory space for software10 MB free memory
Memory space for recordingsOne record needs 64 MB of free memory
RAM128 MB (256 MB recommended)
Monitor resolutionXGA 1024 x 768
MT-101/MT-200
Page 86
Page 89
Options, Accessories and Disposables12
MT-101/MT-200
User GuideComplete Systems12.1
12Options, Accessories
and Disposables
12.1Complete Systems
0.300000PC Connection Kit MT-200 including:MT-101 Holter USB connection cable for PC,
hard-lock key, SCHILLER MT-200 software CD, user guide
1.300000Holter Kit MT-101 2-channel: Solid state Holter recorder MT-101, 4-lead patient
cable for 2-channel recordings, reusable Holter carrying pouch, short user guide and
starter kit.
1.300010Holter Kit MT-101 3-channel: Solid state Holter recorder MT-101, 6-lead patient
cable for 2-channel recordings, reusable Holter carrying pouch, short user guide and
starter kit.
12.1.1Software and Hardware Options
5.270001ST-measurement software for MT-200 Holter software
5.270002Template matching software for MT-200 Holter software
5.270003Network licence software for second user for MT-200 Holter software 5.270000, incl.
hard-lock key.
Note: The network license software cannot read data from the MT-101 Holter
recorder. Otherwise, it has the same functionality as the MT-200 Holter software full
version.
5.270000Network license software for each further user for MT-200 Holter software 5.270000
incl. hard-lock key (please indicate the total number of additional users).
5.270008General network license software (incl. 10 operation manuals) fully functioning from
SW version 1.52.
Note: Name and address incl. phone number must be specified for each individual
order of a general network license software MT-200. Please ask for the special order
form.
Every time an event is registered, stressful activities for example, going
up stairs, carrying groceries, bicycling etc., all meals and medications,
bowel movement and urination. Also describe your dominant sleep
position, e.g. back, right side etc.
Changing the battery during a 48/72 hour recording
•
Change battery when an audible beep is heard and the message
‘BATTERY LOW - change battery’ is displayed.
•
Press button (1) or (2), and make an entry into your diary. DO NOT
SWITCH THE DEVICE OFF.
•
Open battery compartment.
•
Replace battery with a new one of the same type. Observe correct
polarity! Insert the lugs (A), close cover (B) and firmly press down (Pos. C)
until cover clicks in place.
•
ECG recording is automatically resumed by pressing button (1). After a
few seconds the message ‘ECG recording restart’ is displayed while the
unit re-initialises. This is followed by the message ECG recording’.
•
The battery must be replaced within 5 hours for the recording to continue.
Page 90
Replacing the electrodes during a 48/72 hour recording
•
The electrode body pads should not need
replacing during the recording, however, the
cable electrodes can be removed temporarily
during a 48 or 72 hour recording to enable you
to wash.
•
Do not get the electrodes or device wet.
•
Remove cable electrodes but not the electrode
pads.
•
Re-attach the electrode cables to the body
pads and secure the leads in position with
tape.
•
Recording resumes when the electrodes are replaced