V1.0 June 2006 First edition
V1.1 July 2006 First release, Soft 3
V1.2 October 2006 Revision
Note: Distribution and maintenance information
SCHILLER has an international network of customer service, sales and consulting
agencies. For details about your local representative, please contact the SCHILLER
subsidiary near you. You will find a complete list of all the representatives and
subsidiaries on our website at
Sales information is also available from sales@schiller.ch
Manufacturer
SCHILLER Medical S.A.S. Tel: +33 (0) 3 88 63 36 00
4, rue Louis Pasteur Fax: +33 (0) 3 88 94 12 82
F- 67162 Wissembourg Email:
Website: www.schiller.fr
This manual shall be deemed to be an integral part of the
described unit.
Compliance with its content is a prerequisite for proper device
performance and for patients and operators safety .
The manufacturer disclaims all responsibility for the safety,
reliability and performance of the device if:
- assembly, extensions, adjustments, modifications and
repairs are not performed by the manufacturer or by
manufacturer authorised persons.
- the electrical installation of the premises does not comply
with locally applicable requirements.
- the device is not used in compliance with this instructions
for use.
This manual describes the device at the printing time.
Upon request, the supplier will provide circuit diagrams, lists of
components, descriptions, calibration instructions or any other
information required by the user’s qualified technical personnel
to repair those parts of the device that have been stated as
“repairable” by the device manufacturer. The supply of such
information shall not in any event constitute permission or
approval to modify or repair a device.
All rights reserved for the devices, circuits, processes and
names appearing in this manual.
The device is not designed for any use that is not specifically
provided in this manual, which may be hazardous.
Foreword
Part number: 0-48-0080 Page 3
Page 4
MAGLIFE light
This MAGLIFE light manual provides the information required for proper device performance.
Knowledge of monitoring and the understanding of the characteristics and functions of MAGLIFE light are
required for proper use of the device.
Do not use the monitor before you read these instructions.
Device maintenance information is provided in the service manual of the MAGLIFE light. For more
information, please contact your nearest SCHILLER representative.
MAGLIFE light bears the CE-0459 mark in accordance with directive 93/42/EEC relating to medical devices
and fulfils the essential requirements of annex I of that directive.
This product complies with the electromagnetic immunity requirements of standard EN 60601-1-2
"Electromagnetic Compatibility-Medical Electrical Equipment » .
The radio interference emitted by this device is within the limits specified in the standard EN 55011, class B.
When the device is connected to a printer, it is within the limits specified in the standard EN 55011 – class A.
Part number: 0-48-0080 Page 4
Page 5
MAGLIFE light
WARNINGS, PRECAUTIONS AND NOTES
Please read and adhere to the following list of warnings, precautions and notes. Some of them have been
repeated at appropriate areas throughout this manual.
Follow the warnings and precautions stated on the labels affixed on the monitor. Throughout this manual we
have added specific notifications that offer additional information. These notations are designated as:
A Note is provided when extra general information is applicable.
A Caution is provided when special care is to be exercised by the user and/or patient, to avoid injury to the
patient, damage to the device or damage to other property.
A Warning is provided when actions may result in a serious outcome (i.e. injury, serious adverse effect,
death) to the patient or user.
Caution: MAGLIFE light is a monitor designed solely for use close to Magnetic
Resonance Imagers of 0.2 to 3 T for monitoring of patients undergoing an MRI
examination.
Caution: The continuous presence of a qualified person is imperative throughout the
examination.
Warning:Before using the MAGLIFE light monitor, follow the safety instructions below:
MAGLIFE light may only be used by trained healthcare workers who are
familiar with its instructions for use.
Make sure that the voltage and frequency of the electricity system match
those stated on the identification plate.
Warning: Because the device is a class I device, it may only be used in premises with
an electrical system comprising an earth connection.
Warning: Connecting other devices or sensors to the patient could lead to leakage
currents that may be harmful to the patient. Consult SCHILLER before
interconnecting with other equipment.
Warning: Take account of the physiological effect of the other devices connected to the
patient.
Warning: This device may not be installed or operated in explosive environment.
Warning: This device is not designed for use with inflammable anaesthetic agents.
Caution: Use only sensors supplied by SCHILLER. They have been designed specially
for use in MRI environments.
Warning:MAGLIFE light is designed for use with MRI systems with a magnetic field
between 0.2 and 3 Tesla. Contact SCHILLER for use beyond that magnetic
field range.
Warning: It is imperative the MAGLIFE light be installed in the area around the magnet
where the magnetic leakage field is less than or equal to 40 mT (400 Gauss)
(see typical installation drawings and labels on the device).
Part number: 0-48-0080 Page 5
Page 6
MAGLIFE light
Note: The ambient temperature must be located within the following limits:
15°C ≤ Amb. T ≤ 35°C
60°F ≤ Amb. T ≤ 96°F Note: The instructions specific to the installation and use of the MAGLIFE light
monitor have been provided in section 4 of this manual, which must be read
carefully before initial use of the monitor.
Caution:The safety instructions relating to the MRI environment must be followed at all
times (during installation, use, repairs etc.). Any work on MAGLIFE light (e.g.
repairs) must be done outside the area where the risk of magnetic attraction is
non-existent.
Caution: No equipment that hampers blood flow (e.g. use of a cuff type
sphygmomanometer) may be used on the limb of a patient undergoing oxygen
saturation monitoring (SpO2), as it could disrupt the correct determination of
the data.
Caution:Sensors and probes may not come in contact with conductive parts, including
the earth.
Caution:Do not defibrillate on the accessories. Place the defibrillation electrodes as far
away from the other accessories as possible.
Caution: The monitor is adapted for the MRI configuration upon power up. Any
subsequent change in the MRI system can alter the performance of the
MAGLIFE light monitor. Inform SCHILLER of any such change.
Caution: The device is designed to operate with electrosurgical devices. Place the
accessories as far away from the electrodes of the electrosurgical device.
That device must be installed and used in accordance with the manufacturer’s
instructions.
Caution:The device is designed to operate without any equipotential connection.
1. Description of MAGLIFE light ............................................................................................ 8
2. Location and description of displays and controls.........................................................9
2.1 Front of MAGLIFE light..................................................................................................................... 9
2.2 Rear of MAGLIFE light...................................................................................................................... 9
2.3 Mains power unit................................................................................................................................. 9
3. Device symbol identification............................................................................................10
4.5.1 Operating during a mains failure................................................................................................... 17
4.6 Use of the menu ................................................................................................................................ 18
4.7 Description of menus ........................................................................................................................ 18
4.7.1 Main menu .................................................................................................................................... 18
4.8 Structure of menus............................................................................................................................ 39
4.8.1 User menus.................................................................................................................................... 39
7.1 General errors.................................................................................................................................... 47
7.2 Errors of modules.............................................................................................................................. 48
MAGLIFE light may be configured with all combinations of the above parameters
MAGLIFE light is fitted with batteries as standard and can move with the patient (e.g. to and
from the examination room and the adjacent preparation room).
MAGLIFE light is designed to be placed on a non-magnetic mobile stand.
.
Part number: 0-48-0080 Page 8
Page 9
MAGLIFE light
2. Location and description of displays and controls
2.1 Front of MAGLIFE light
(see photo in section 11 )
1. Main device on/off key.
2. Device “ON” indicator.
3. “MAINS” connection indicator.
4. “BATTERY” Indicator, showing that the battery is being charged (flashing), or is fully charged
(steady).
5. Main “MENU” key. Access to the main menu and for exiting a menu from any other place.
6. Navigation button for menu selection.
7. Display for curves, parameters, menus and messages.
8. “NIBP” key, for starting up or stopping a measurement or a series of measurements of the
blood pressure via a cuff.
9. “Alarm silence” key, enables or disables the two-minute or permanent audio alarm function
(physiological and technical alarms).
10. Device “STANDBY” key.
11. SPO
12. NIBP cuff connector.
optical oximeter sensor connector.
2
2.2 Rear of MAGLIFE light
(see photo in section 12 )
13. Connector for the low voltage cord for connecting to the mains power unit.
14. Loudspeaker
15. USB connector
16. RS 232 connector
17. Auxiliary connector compartment cover.
18. Identification label.
2.3 Mains power unit
(see photo in section 12 )
19. Mains plug.
20. Low voltage plug for connecting the MAGLIFE light.
Part number: 0-48-0080 Page 9
Page 10
MAGLIFE light
3. Device symbol identification
3.1 MAGLIFE light symbols
Main device on/off key
Caution! Read the instructions for use of the device.
Electrical and electronic device identification symbol. The device
Notified body for CE certification (G-MED)
Mains power on
Battery charge
Entry into main menu or exit from any menu
Audio alarm enable/disable
Non-invasive blood pressure start/stop
CF type device, is protected against defibrillation shocks (device designed
for direct heart applications).
components must be disposed of separately and the relevant parts must be
sent to available recycle centres. Inappropriate disposal may be harmful to
the environment and to public health as a result of the presence of
hazardous materials in electrical and electronic devices.
3.2 Power symbols
Power input
Part number: 0-48-0080 Page 10
Power connected
Page 11
MAGLIFE light
4. Use of MAGLIFE light
4.1 Installation
MAGLIFE lighthas been designed to operate at the patient’s bedside. It is installed in the
Faraday cage, i.e. in the room with the MRI system.
Minimum distances must be kept in relation to the measuring tunnel entrances. These
depend on the magnetic field of the magnet and the type of magnet.
As a result, it is imperative the MAGLIFE light be kept outside the area around the magnet demarcated by the 40 mT (400 G) line. A magnetic field
detector sets off an alarm if that value is reached. Beyond the 40 mT limit, the
monitor is exposed to a force of attraction (due to the magnetic field) that
rises rapidly as the distance from the magnet decreases. This area must
be indicated by markings on the floor.
Caution: The monitor may be installed on a non-magnetic trolley supplied as
an option by SCHILLER. The trolley has wheels for turning and
moving the monitor. If the minimum distance is not kept, the
performance of the device may be disrupted. The trolley bearing
the monitor may not be placed in or moved to an area where the
field is greater than 40 mT (400G), regardless of whether the
monitor is on or off. The force of attraction of the magnet could pull
the monitor against the walls of the tunnel.
The wheels of the trolley can be locked when the trolley is installed
in the permitted area.
Important note: The magnetic field is always present, even when the imager is not
being used for an examination.
Part number: 0-48-0080 Page 11
Page 12
MAGLIFE light
Typical plan of an MRI system with a MAGLIFE light monitor
1) Faraday cage
2) MAGLIFE light monitor
3) Mains connector
4) Marking on the floor (40 mT line)
40 mT(400G) line
3
Figure 1
4
2
Forbidden area
1
Authorised area
Part number: 0-48-0080 Page 12
Page 13
MAGLIFE light
f
N
a
ea
t
N
4.2 Display screens
4.2.1 SPO2 option only
Time
Date
Quality o
SPO2 signal
SPO2 value
Alarm Thresholds
Patient type
Sound alarm
status
Plethysmogram
Pulse rate value
Pulse area
4.2.2 NIBP option only
IBP are
Uni
Mode
Previous measurements
IBP values
Pulse ar
Part number: 0-48-0080 Page 13
Page 14
MAGLIFE light
a
N
a
4.2.3 SPO2 and NIBP options
SPO2 are
IBP are
4.3 Placing of probes and sensors
Warning: The recommendations relate to the type of sensors and probes to be
used, the positioning of the sensors and probes on the patient and the
placing of the cables that connect the sensors and probes on the
patient and the device.
The following rules must be followed precisely to avoid the problems described below:
Status area
4.3.1 Absolute rule
Use only the cables, sensors and cuffs supplied by SCHILLER.
4.3.2 Probes and sensors to be used and positioning instructions
4.3.2.1 SpO2 sensor
1. Make sure that the Oximeter function is on and that the oximeter
parameters are set correctly.
Caution:Use only the sensors from the list of accessories supplied by
SCHILLER.
2. Connect the oximeter patient cable to the oximeter connector and
give the connector a quarter turn to the right-hand side to lock it in
place (to disconnect, turn the connector to the left and pull it out).
3. Place the sensor over the nail of the index finger or equivalent site;
place the cable on the top of the hand and fix it using Velcro™ strap
or secure to the patient’s wrist without tightening it.
Warning:Be cautious in placing the cable correctly so as to prevent damage
or injury (tangling and/or strangulation).
Part number: 0-48-0080 Page 14
Page 15
MAGLIFE light
Warning:
Warning:
Caution:
Caution:
Warning:
Warning:
Warning:
Warning:
Warning:
Caution:
Do not place the sensor on an extremity with an invasive probe or
Do not place the cuff on the same limb that the oxygen saturation
Do not use the oximeter alone while monitoring vital parameters.
Remove nail polish or false nails before placing the sensor on the
patient’s finger, as they could lead to inaccurate oximeter
measurements. Cut long nails, as they could hinder the placing of
the sensor.
If the sensor is secured with tape, do not tighten the tape
excessively. If the tape is too tight, it may affect the measurement
accuracy of the device and blisters may form on the patient’s skin
(due to the lack of circulation in the skin and not because of a
source of heat).
a blood pressure cuff.
is being measured, as it could disrupt the correct determination of
the data (resistance that hinders the blood flow).
The oximeter measurement may be modified by the presence of
strong ambient light. Cover the sensor (e.g. with a surgical sheet) if
required.
The incorrect application or improper use of the sensor can lead to
measurement inaccuracies, as can the presence of significant
levels of dysfunctional haemoglobins (e.g. carboxyhaemoglobin or
methaemoglobin) or intravascular dyes such as indocyanine green
or methylene blue, the exposure to excessive light such as with
surgical lamps (especially lamps with xenon sources), bilirubin
lights, fluorescent lamps, infrared heating lamps or direct sunlight,
excessive movement by the patient, vein pulsations, the installation
of a sensor on a limb with a blood pressure cuff, an arterial probe
or an intravascular line.
In some cases, when the perfusion and signal are weak, e.g. with
patients with a thick or dark skin, the device may produce
abnormally low oximeter readings. Oxygenation should be verified
before starting therapies and interventions, especially in preterm
babies and patients with chronic lung diseases.
It often happens that patients suffer from low peripheral perfusion
due to hypothermia, hypovolaemia, serious vasoconstriction or
reduced cardiac output etc. These symptoms may lead to a loss in
the oximeter readings.
The temperature of the patient and the room must not be too low
1. Make sure that the NIBP function is on and that the NIBP
parameters are set correctly.
2. Connect the NIBP tube connector to the NIBP measurement
connector (12). Use only the cuffs listed in section 8.
Warning:
Warning:
Caution:
Caution:
Warning:
Caution:
Warning:
Caution:
If the cuff is placed too tightly on the limb, the results will be
excessively high readings. Among other considerations, the use of
a cuff suited to the patient has a direct bearing on the accuracy of
the NIBP measurements obtained. Select the cuff according to the
circumference of the patient’s limb.
Cuffs can soften over time and get permanent creases that may
mark the limb temporarily. Replace any cuffs which have that
characteristic.
Make sure that the pressure connectors are not flattened or
blocked.
Position the cuff slightly above the elbow, the Velcro™ turned up.
Since the cuff does not contain a microphone, specific positioning
is not necessary.
The cuff must be tightened around the arm, but must not apply
pressure on the blood vessels before the measurement. Roll the
cuff around the arm and close the Velcro™.
Check to ensure the hose is not tangled (check any incorrect
placement of the arms, legs or body of the patient, any clamps
etc.), as device operation will be disrupted by a flattened or
restricted hose.
Check that the blood is flowing correctly in the relevant limb (arm,
leg) (except during the measurement).
A cuff type sphygmomanometer must not be used on the patient’s
limb undergoing oxygen saturation measurement as it could disrupt
the correct determination of the data.
Several cuff models are available, including:
- Adult cuff
- Child cuff
- Neonate cuff
Extension hoses are also available.
Part number: 0-48-0080 Page 16
Page 17
MAGLIFE light
4.4 Start up
Connect the mains cord to the mains connector (19) and the low-voltage connecting
cord to connectors (20) and (13). Indicator (3) will light up to show that the device is
connected to the mains.
Caution:
Press key (1); the associated indicator (2) will illuminate.
After a few seconds, the screen will activate and the system initialised. The initialisation
sequence will last about 10 seconds. After that time, the parameters will be displayed.
For battery operation (with the mains cord disconnected), just press key (1) (the battery
is integrated in the device). If the battery is charged, associated indicator (2) illuminate
and the device will start up. With a fully charged battery, the device can operate for at
least two hours.
Note:
Indicator (4) will illuminate and the battery will be charge automatically
The operating of key (1) is disabled during the initialisation phase.
while the device is connected to the mains, even if it is not in operation.
4.5 Battery charge
MAGLIFE lightmay be powered by the internal battery to accompany the patient as the
patient is moved, or if the mains power is absent. The battery charges automatically
whenever the device is connected to the mains, whether or not it is operating. While the
battery is charging, indicator (4) flashes. Once the battery is fully charged, it stays on
and steady.
MAGLIFE light can operate on the battery (new battery, fully charged) for two hours
(one hour if intensive use is made of the NIBP function).
An alarm message is displayed about 5 to 10 minutes before the battery is completely
discharged and the automatic device shutdown occurs,
4.5.1 Operating during a mains failure
MAGLIFE light manages its power supply automatically. If the mains supply fails or is
of poor quality, the battery will automatically start supplying power to the device
regardless of the duration of the disruption.
Indicators (3) and (4) will go off and the mains present indicator (see section
the battery charge indicator will reflect the new situation in the Status zone after three
seconds.
4.2 ) and
Part number: 0-48-0080 Page 17
Page 18
MAGLIFE light
4.6 Use of the menu
All the functions other than those accessible via the controls on the unit are selected
from the menus displayed on the screen.
You can enter the main menu in two ways:
− by pressing rotary button (6) or
− by pressing key (5).
Caution:
The rotary button is used to:
− confirm a choice in the menu (when the line is highlighted in blue)
− make a choice
− switch to the digital entry mode by making the cursor flash
− move from one selected field to another without making any change
− select a status in a toggle function
Double clicking (two clicks in rapid succession) is used to exit any menu. It has the
same function as key (5).
Button rotation is used to:
− move in a menu
− increase or decrease the value of an entry field
Caution:
The displayed menus depend on the parameters defined as active in the
Menus disappear if no action is taken for more than 10 seconds.
configuration (see paragraph 4.7.2 ).
4.7 Description of menus
4.7.1 Main menu
The main menu is as follows:
It is used to select submenus.
Note:
Only the labels corresponding to the installed parameters are displayed.
The main menu contains the following submenus:
Part number: 0-48-0080 Page 18
Page 19
MAGLIFE light
4.7.1.1 NIBP (Non Invasive Blood Pressure) menu
Windows displayed
− with the NIBP parameter disabled (select NIBP On/Off to enable)
− with the NIBP parameter enabled (select NIBP On/Off to disable)
• Cycle selection
to select a cycle value from 1 minute to 90 minutes, activate symbol ,
then select the appropriate value by rotating button (6) and confirm by
pressing the button.
Part number: 0-48-0080 Page 19
Page 20
MAGLIFE light
• Patient type selection
Note:
For subsequent measurements, the inflating pressure is automatically
• Measuring unit selection
Depending on the type of patient, the initial inflating pressure is
automatically selected (180mmHg for Adult/Child and 120mmHg for
Neonate.
selected (about 50mmHg above the measured systolic pressure for
Adult/Child and 30mmHg for Neonate.
The Unit button is used to switch between the two units - kPa and mmHg.
Part number: 0-48-0080 Page 20
Page 21
MAGLIFE light
• Use the Alarm limits button to manage thresholds. For more details, refer
to section
• NIBP alarm limit setting
Note:
At the switch on time, the alarm levels corresponds to the standard values
4.7.1.3 Alarm limits and 4.9 Alarms
Select a limit by rotating the navigation knob (6), press to confirm
Select the limit value by rotating the navigation knob (6) again, press to
confirm the selected value.
of the selected patient. The operator selection is done in general alarm limit
management window refer to section
4.7.1.3 Alarm limits
Part number: 0-48-0080 Page 21
Page 22
MAGLIFE light
• The Calibration submenu is used to check if the NIBP pressure module is
measuring the pressure correctly. Connect the external verification system
with an expansion vessel of at least 75 cl and a pressure measuring device.
If required, add an inflating bulb with a deflating valve. Activate Calibration;
the display on the screen will show the pressure value measured by the
NIBP module. Make several comparisons between the measurements by
MAGLIFE light and those by the external measuring device over the entire
measurement range.
Note:
Calibration is regulated in some countries. Comply with
applicable laws and regulations.
Note:
The Calibration submenu is locked by a password.
The password is communicated to the people who followed
the technical training.
Part number: 0-48-0080 Page 22
Page 23
MAGLIFE light
4.7.1.2 Oximeter (SpO2) submenu
Windows displayed
− with the SpO2 parameter disabled (select SpO2 On/Off to enable)
− with the SpO2 parameter enabled (select SpO2 On/Off to disable)
• Select the averaging time to be taken into account while calculating
the SpO2 value and the pulse rate. The choice between the two values
- 8 s and 16 s - is made by activating the Average tab.
• Patient type selection. The patient type selection only affects the
thresholds.
Part number: 0-48-0080 Page 23
Page 24
MAGLIFE light
• Use the Alarm limits tab to manage thresholds. For more details, refer
• SPO2 alarm limits setting
Note:
At the power up time, the alarm levels corresponds to the standard values
to section 4.7.1.3
Select a limit by rotating the navigation knob (6), press to confirm
Select the limit value by rotating the navigation knob (6) again, press to
confirm the selected value.
of the selected patient. The operator selection is done in general alarm limit
management window refer to section
Alarm limits and 4.9 Alarms
4.7.1.3 Alarm limits
Part number: 0-48-0080 Page 24
Page 25
MAGLIFE light
4.7.1.3 Alarm limits menu
The Alarm limits menu is used to set the thresholds for triggering alarms
relating to the different parameters.
Caution:
Caution:
The following submenus are available:
When the device is powered up and when it comes out of the standby
mode, the threshold values are the default values for the type of patient
selected (Adult, Child or Neonate).
. See section 4.7.1.3.1
Three different sets of default values are available for each type of patient.
These values can be set from the Device configuration menu (see
section 4.7.2 )
The Thresholds menu only displays the thresholds for the
parameters present in the configuration.
If one or more alarms have been disabled, an icon
with a crossed-out bell and the word OFF is
displayed in the upper right-hand corner of the
screen.
Power up and standby mode
Part number: 0-48-0080 Page 25
Page 26
MAGLIFE light
The Operator 1 choice gives access to the following submenu, which is
used to set the thresholds from the values selected by the operator when
the function was last accessed.
• The Quick selection is used to set all the values in relation to the
measured physiological values of the patient. The upper values are
set to 20% above and the lower values are set to 20% below those
physiological values.
• The Default selection is used to go back to the default values for
the selected patient type.
• provides access to the following submenu:
Is used to go back to the previous submenu
• OFF is used to disable all the audio and visual physiological
alarms
Part number: 0-48-0080 Page 26
Page 27
MAGLIFE light
Caution:
4.7.1.3.1 Power up and standby mode
When the device is powered up, the threshold values are the default values
for the type of patient selected, and therefore not necessarily the values
used when the device was last used for monitoring. (The default thresholds
are factory set and can be adapted from the configuration menu). Patient
type selection is offered for 12 seconds after device power up and also
when the device comes out of standby mode. If the operator does not
select the patient type, the default values of the neonatal type are selected
automatically. The possibility to select the values used last is available with
the Previous button.
Note:
Failure to monitor the condition of the patient may lead to
death.
• Selection Operator 2 provides access to the second set of values
defined by the operator when the function was last accessed. The
values can then be edited.
• Exit is used to close this submenu.
Selecting the Previous option makes it possible to obtain
trend continuity, as all the other selections result in the
insertion of a separating mark between patients in the trends.
Part number: 0-48-0080 Page 27
Page 28
MAGLIFE light
4.7.1.4 Magnetic field display
The B0 display button is used to display the magnetic field measurement
on the screen. Values Bx, By and Bz show the value of the magnetic field
in mT along the three orthogonal directions in space. The B value shows
the corresponding vector modulus.
4.7.1.5 Trends menu
The Trends menu contains two main submenus - Graphs and Tables - and
two options. It enables the user to erase the trends and select the period
for trend tables.
Part number: 0-48-0080 Page 28
Page 29
MAGLIFE light
4.7.1.5.1 Trend graph submenu
The Trend graph menu is used to select the trend curves to be displayed
and contains the following tabs:
Note:
Only two waveforms can be shown at the same time.
Navigation buttons
Back one page
Forward one page
Reduce time line from 24 hours to 1 hour
Increase time line from 1 hour to 24 hours
Adapt graph scale
Exit the trend graphs display
Part number: 0-48-0080 Page 29
Page 30
MAGLIFE light
4.7.1.5.2 Trend table submenu
Trend table show the numerical values of the physiological parameters
according to the frequency that has been defined earlier from the trend
type selection menu.
The display varies according to the selected parameters
Note:
If a non-invasive pressure measurement does not strictly
Navigation buttons
Forward to the end of the table (most recent value)
Forward to next line
Back to previous line
Back to the start of the table (oldest value)
Exit the trend tables display
follow the table frequency rules, it is inserted at the
appropriate location.
Part number: 0-48-0080 Page 30
Page 31
MAGLIFE light
m
m
Patient change
Technical alar
Physiological alar
Special displays
4.7.1.6 Edit menu
(To be written )
Part number: 0-48-0080 Page 31
Page 32
MAGLIFE light
4.7.1.7 Settings menu
The Settings menu is used to set the general parameters of MAGLIFE
light and to access the demonstration mode.
4.7.1.7.1 Sound submenu
The sound submenu is used to set the level of the sound for alarm signals,
pulse bip or keyboard clics
Sounds may be High, Medium, Low or Off, with the exception of alarms,
which can never be Off.
Part number: 0-48-0080 Page 32
Page 33
MAGLIFE light
4.7.1.7.2 Clock submenu
The various buttons are used to set the date and time. The daylight saving
option can be disabled. The clock change date is pre-set.
Example of date setting. Only plausible values can be selected.
Note:
A time or date setting will erase all trends.
4.7.1.7.3 Screen brightness setting
You can set the screen brightness from 50 to 100%.
Part number: 0-48-0080 Page 33
Page 34
MAGLIFE light
4.7.2 Device configuration menu
The configuration menus are hidden during normal operation. Access is
initiated by pressing the navigation button (6) when the device is switched
on and keeping it pressed until the configuration menu is displayed.
Note:
To exit the Device configuration mode, the device must be
switched off.
4.7.2.1 Alarm limits submenu
The default thresholds are set for the three type of patients - Adult, Child
and Neonate. Select one type of patient, press button (6) and make the
setting in the following tab
Select the parameter by navigation button (6) counter clockwise and
confirm by pressing button (6). The first value will flash. Turn as required to
modify the value. Save the value by pressing button (6). The next value is
automatically selected. When the last parameter value has been entered,
the selection cursor goes back to the parameter selection column.
Part number: 0-48-0080 Page 34
Page 35
MAGLIFE light
4.7.2.2 Monitor submenu
Note: The Language, Colour, Alarms, NIBP and Patients
selection list buttons are used for settings, whilst the Serial
number and Hardware number buttons are provided for
information only and cannot be modified by the user.
• A certain number of languages are available. The
language selection does not affect the operating of
MAGLIFE light.
• The screen colour combination can be selected with
button .You cannot define a customised colour
combination.
• The allowed patient selection button is used to
restrict or extend the patient selection list.
Part number: 0-48-0080 Page 35
Page 36
MAGLIFE light
Button: is used for special alarm configuration. It opens the
following submenu:
Button:
Caution:
Caution:
Enabling or disabling the alarm function if the SPO
more than 4%.
A physiological alarm may be latched, i.e. confirmation by
pressing key (9) is required to stop it, even when the value of
the parameter has exceeded the threshold and is back in the
non-alarm range. If the alarm is not latched, it stops as soon
as the value is back within the non-alarm range.
This option is used to disable audio alarms for two minutes
after the device is switched on, allowing for time required to
apply the sensors for the measured values to stabilise.
This button is used to permit or prevent the ability to mute
audible alarms permanently with key (9). If the alarms are
permanently muted, a reminding audio signal (double beep) is
emitted after every two minutes.
is used to define the parameters of the NIBP
measurement in the continuous mode.
Is used to select the duration of the
continuous mode.
To ensure the correct vascularization of the member on whom
the blood pressure is measured, the repetition of the
continuous mode is not advised.
Is used to select the frequency of
automatic measurements after the end
of the continuous mode.
These configuration settings must be in accordance with the
laws applicable in the country of use.
drops
2
Part number: 0-48-0080 Page 36
Page 37
MAGLIFE light
4.7.2.3 Options submenu
The Options menu is used to enable or disable one or both parameters of
MAGLIFE light
4.7.2.4 Software update submenu
The person performing the update must have the requisite skills to make
the relevant functional and safety checks and is entirely responsible for such checking.
The software is updated by means of a USB key, standard 1.1 or above.
Make sure that the software on the USB key in the root directory (only that
software) is compatible with the device you want to update.
Insert the USB key in the connector (15) provided, select the update button
and loading will start automatically. Follow the instructions. At the end of
the loading process, the device will go off automatically.
Note:
The procedure has been described in detail in the Service
Manual of the device. For more information, please contact
the technical service network of Schiller Medical.
Part number: 0-48-0080 Page 37
Page 38
MAGLIFE light
4.7.2.5 Version submenu
The Version tab will show detailed info about used software versions and
releases in MAGLIFE light
4.7.2.6 Reserved SCHILLER submenu
This submenu is only for technical purpose and is described in a separate
manual.
Part number: 0-48-0080 Page 38
Page 39
MAGLIFE light
4.8 Structure of menus
4.8.1 User menus
Part number: 0-48-0080 Page 39
Page 40
MAGLIFE light
Part number: 0-48-0080 Page 40
Page 41
MAGLIFE light
4.8.2 Configuration Menus
Same for Child
Same for Adult
Part number: 0-48-0080 Page 41
Page 42
MAGLIFE light
4.9 Alarms
4.9.1 Screen alarm symbols
S1
S2 Alarm sound disabled!
S3 Alarm sound disabled for 2 minutes.
S4 Alarm sound disabled permanently.
Alarm muting function
Key (9) is used to mute the alarms.
• When this key is pressed briefly, alarms stay disabled for two
minutes and the symbol (S3) shows the time remaining in minutes up
the end of the disabling of the alarm.
• If you press key (9) for three seconds or more, alarm sounds remain
disabled permanently or until the key (9) is pressed again. Symbol
(S4) is displayed, the symbol “∞” flashes and a beep is emitted once
in every two minutes to remind the user.
Caution:
Audible alarm enabling/disabling
The permanent disabling of audible alarms is not permitted in some
medical facilities. That is why that function can be configured. (See
Permanent muting section 4.7.2.2 )
Audible alarms enabled
Displayed when at least one monitoring threshold is disabled.
Technical alarms will go off all the same.
Displayed when you press key (9) (for less than three
seconds). The remaining time is displayed under the symbol.
Displayed when the alarm off key (9) is held down for more
than three seconds.
4.9.2 Physiological alarms
If the measured value of a parameter exceeds a threshold for over three
seconds, an alarm goes off and:
• the display of the measured value flashes in red
• an intermittent audible alarm sounds (sequence of four digital sounds that
is repeated)
• the window of the parameter over the threshold flashes on a red
background and the display colour are inversed
• Depending on the choice made while configuring (see section 4.7.2.2) the
alarm stops:
- As soon as its cause stops (non latched)
- After the causes disappears and key (9) is pressed (latched)
Part number: 0-48-0080 Page 42
Page 43
MAGLIFE light
4.9.3 Technical alarms
When a technical alarm is triggered:
• an error message is displayed in the display field of the parameter in
question
• an intermittent audible alarm goes off (sequence of two audible
signals that is repeated); it depends on the audible alarm
configuration criteria
• a question mark (-?-) is displayed instead of the measured value.
Note: going beyond of the measurement limits of the device.
In this case no error message is displayed:
• technical alarm sounds;
• 3 fixed indents (- - -) are displayed in the place of the measured value.
These alarms go off automatically when their causes disappear.
5. Technical specifications
5.1 System specifications
Manufacturer SCHILLER Medical S.A.S.
Device name MAGLIFE light
Dimensions
main unit 270 x 216 x 116 mm; 10.6” x 8.5” x 4.6”
power supply 180 x 84 x 68 mm; 7” x 3.3” x 2.7”
Weight
main unit 6 Kg
power supply 1.3 Kg
Protection class IP 21
of the enclosure
Electricity supply 100, 115, 230 VAC, 50/60Hz
Factory defined voltage
Power consumption 25 VA
Fuses 2 x 100 mA (T) - 230 VAC; 2 x 200 mA (T) - 100 - 115
VAC
Battery12 V, 2 Ah Lead
Autonomy 2 hours
Environmental conditions
Operating
Temperature 15°C - 35°C; 60°F - 96°F
Relative humidity 30 – 95% non condensing
Pressure 500 - 1060 hPa
Magnetic field ≤ 40mT
Storage
Temperature -10°C - 50°C; 13°F - 124°F
Relative humidity 30 – 95% non condensing
Pressure 500 - 1060 hPa
Magnetic field ≤ 40mT
Part number: 0-48-0080 Page 43
Page 44
MAGLIFE light
Display Colour TFT screen: 6.8”; 98 x 132mm; 480 x 640 dots
Connections SPO
Interfaces RS232, USB 1.1 for connecting the equipment
specified by SCHILLER
Safety standards IEC 60601-1
EMC IEC 60601-1-2
Static discharge up to 8 kV
Radio frequency 10 V/m (80 – 2500 MHz, 5 Hz
modulating)
CE marking According to directive 93/42/EEC class IIb
Protection class Class I according to IEC 60601-1
and NIBP
2
CISPER 11 Class B;
with printer connected, class A
The device may be subject to the following
interference without being affected:
5.2 Technical specifications of modules
5.2.1 Pulse oximeter
OEM module BCI
Connection Fibre optic
Class CF
SPO2 accuracy ± 2 % from 70 to 99 %
± 3 % from 50 to 69 %
display range 0 – 99%
SPO
2
Pulse accuracy 5 b/min
Pulse display range30 – 250 b/min
Blocking by defibrillator shock 10 seconds maximum
HF protection protection from electro surgery devices
5.2.2 NIBP Non-Invasive Blood Pressure
OEM module CAS
Connection Fast snap
Class CF
Measuring principle Oscillometric
Mode Manual, Automatic, Continuous
Part number: 0-48-0080 Page 44
Page 45
MAGLIFE light
Types of patient Neonate, Child, Adult
Sensor accuracy ± 3mmHg or ± 2 %
Pulse accuracy 5 b/min
RF protection protection from electro surgery devices
5.3 Threshold ranges
The upper and lower alarm thresholds are adjustable according to the values in the
table below.
The lower limit setting may never be greater than the upper limit setting and vice versa.
Parameter Lower limit Upper limit
SPO2
NIBP
SPO2
NIBP
SPO2
NIBP
Neonate patient
Pulse 30-245 35-250
Saturation 50-98 51-99
SYS 30-130 35-135
DIA 15-105 20-110
MAP 20-120 25-125
Child patient
Pulse 30-245 35-250
Saturation 50-98 51-99
SYS 30-250 35-255
DIA 15-115 20-220
MAP 20-230 25-235
Adult patient
Pulse 30-245 35-250
Saturation 50-98 51-99
SYS 30-250 35-255
DIA 15-215 20-220
MAP 20-230 25-235
Part number: 0-48-0080 Page 45
Page 46
MAGLIFE light
6. Cleaning
MAGLIFE light can be cleaned with common cleaning and disinfecting agents (BURATON,
INCIDIN. GG. KORSOLIN or LYSO FORMIN 2000).
Follow the manufacturer’s instructions for use.
Switch the device off and disconnect from the mains before cleaning.
Do not remove any covers.
If any liquid does penetrate into the device, the device must be cleaned completely and
inspected.
Do not expose the device to temperatures above 50 °C.
The use of an autoclave is not permitted.
The device and its accessories need to be decontaminated (risk of pathological contamination)
before disposal.
The batteries of MAGLIFE light must be disposed of using a special procedure and not merely
scrapped.
Part number: 0-48-0080 Page 46
Page 47
MAGLIFE light
7. Troubleshooting
7.1 General errors
Error Cause Remedy
The screen does not activate
when the device is switched on
The screen activates but the
device initialise sequence is not
carried out
The device shuts off
automatically
No printing
• Mains not connected
• Battery discharged
• Magnetic field too great
• Device faulty
• Device faulty • Replace the device
• Battery discharged
• Magnetic field too great
• Non specific printer
• No power to printer
• Connecting cable loose
• Check and correct the
connections - LED (3) must
be on
• Connect the mains - LED
(4) flashes
• Place the device outside the
field (<40mT)
• Replace the device
• Connect the mains
• Place the device outside the
field
• Replace the printer with a
printer specified by
SCHILLER
• Connect the printer to the
power source and switch on
• Check and remedy the
connections
Part number: 0-48-0080 Page 47
Page 48
MAGLIFE light
7.2 Errors of modules
Error Cause Remedy
No SPO2 measurement
SPO2 signal disturbed
The cuff will not inflate
Doubtful NIBP measurements
• Optical probe faulty
• Ambient light artefacts
• No perfusion
• Module faulty.
(no light in the SPO2
connector)
• Incorrect sensor installation
• Compression of the
measurement location
• Poor perfusion
• Pump motor magnetised,
because the device has
been subject to an
excessive magnetic field
• Faulty cuff or tube
• Incorrect cuff installation
• Inappropriate cuff
• System not airtight
• Not locatable
• Replace the probe
• Cover the sensor holder on
the finger
• Change the signal taking
location
• Replace the device
• Move the sensor
• Remove the compression
• Move the sensor
• Replace the device
• Replace and check
tightness
• Check and adjust the cuff
• Replace with an appropriate
cuff
• Replace the leaking part
• Replace the device
8. Maintenance
• Before each use:
- check all the device functions
- simulate an alarm
- check if all the connectors and cable insulators are in good condition
• Once a year:
- check the leakage current according to paragraph 19 of standard IEC 601-1.
8.1 Alarm simulation
To simulate an alarm for verification, proceed as follows:
• connect a device that simulates a physiological signal to the device (e.g. SPO
wait for the value to stabilise and take the reading
• select the corresponding parameter with button (6) and select the Thresholds
button
• modify one of the threshold values so that the value of the physiological parameter
read earlier exceeds it
Amagnetic trolley with storage basket for MAGLIFE light
Adult finger SpO
Child finger SpO2 sensor, 5.2 m - fibre optic
Universal Y SpO2 sensor, 5.2 m - fibre optic
Adult finger SpO2 sensor, 4.2 m - fibre optic
Child finger SpO2 sensor, 4.2 m - fibre optic
Universal Y SpO2 sensor, 4.2 m - fibre optic
Adult finger SpO2 sensor, 3.5 m - fibre optic
Child finger SpO2 sensor, 3.5 m - fibre optic
Universal Y SpO2 sensor, 3.5 m - fibre optic
5 m cuff tube
3.5m m cuff tube
2.5 cm neonate cuff
4 cm neonate cuff
6 cm child cuff
7 cm child cuff
9 cm child cuff
12 cm adult cuff
14 cm adult cuff
16 cm adult cuff
Shielded MAGLIFE light power cable
sensor, 5.2 m - fibre optic
2
10. Disposal
10.1 Battery disposal
Important
Caution:
Caution: Acid burn hazard! Never open or overheat the battery!
10.2 Device disposal
In normal use, the battery requires no maintenance.
After five years, the battery must be replaced, whether or not the device has
been used.
Explosion hazard! The battery may not be incinerated or disposed of
with household waste.
In accordance with national law, the battery may only be disposed of in an approved
disposal facility or sent back to SCHILLER.
The device components must be disposed of separately and the relevant parts must be
sent to appropriate recovery and disposal centres.
If you do not know of such a recovery and disposal system, you may return the device
to the distributor or manufacturer, who will take charge of disposing of the device in
accordance with applicable regulations. In that way, you will contribute to the recycling
and recovery of old electrical and electronic devices and their reuse in other forms.
Inappropriate disposal may be harmful to the environment and to public health as a
result of the presence of hazardous materials in electrical and electronic devices.
Part number: 0-48-0080 Page 49
Page 50
11. Front
8
10
11
7
6
9
4
3
1
2
12
Part number: 0-48-0080 Page 50
Page 51
12. Rear and power supply
13
19
18
14
15
20
Part number: 0-48-0080 Page 51
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.