The SCHILLER sales and service centre network is worldwide. For the address of your local distributer,
contact your nearest Schiller subsidiary . In case of difficulty a complete list of all distributers and subsidiaries
is provided on our internet site or can be supplied from our head office.
S CHILLER AG, A ltgasse 68, CH-6341 Baar, Switzerland
Tel.: + 4141 766 4242Fax : + 4141 761 0880
Home Page : www.schiller.ch
Art.No. 2.540022
3
Warranty
Terms of Warranty
The SCHILLER AT-1 / AT-1 smartprint is warranted against defects in material and
manufacture for the duration of one year (as from date of purchase). Excluded from
this guarantee is damage caused by an accident or as a result of improper handling.
The warranty entitles free replacement of the defective part. Any liability for subsequent
damage is excluded. The warranty is void if unauthorized or unqualified persons attempt
to make repairs.
In case of a defect, contact your dealer or the manufacturer.
The manufacturer can only be held responsible for the safety , reliability , and performance
of the apparatus if:
* assembly operations, extensions, readjustments, modifications, or repairs are carried
out by persons authorized by him, and
* the ECG unit and approved attached equipment are used in accordance with the
manufacturers instructions.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES
HEREINABOVE SET FORTH. SCHILLER MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS
FOR A PARTIC ULAR PURPOSE WITH RESPECT TO THE PRODUCT OR P ARTS THE REOF.
Physicians responsibility
THIS UNIT IS PROVIDED FOR THE EXCLUSIVE USE OF QUALIFIED PHYSICIANS OR PERSONNEL
UNDER THEIR DIRECT SUPERVISION. THE NUMERICAL AND GRAPHICAL RESULTS AND ANY
INTERPRETATION DERIVED FROM A RECORDING MUST BE EXAMINED WITH RESPECT TO THE
PATIENTS OVERALL CLINICAL CONDITION. THE RECORDING PREP ARATION QUALITY AND THE
GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DA TA ACCURACY,
MUST ALSO BE TAKEN INTO ACCOUNT.
IT IS THE PHYSICIANS RESPONSIBILITY TO MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT
OPINION ON THE RESULTS, AND TO INSTITUTE CORRECT TREATMENT IF INDICATED.
4
Safety Notices
DO NOT USE SOLVENT CLEANERS.
THIS UNIT COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS
PROTECTION AGAINST EMISSIONS AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE
MUST BE EXERCISED WHEN THE UNIT IS USED WITH HIGH FREQUENCY EQUIPMENT.
BEFORE USING THE UNIT, ENSURE THAT AN INTRODUCTION REGARDING THE UNIT FUNCTIONS
AND THE SAFETY PRECAUTIONS HAS BEEN PROVIDED BY A PRODUCT REPRESENTATIVE.
SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS.
THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION
IS SUSPECT OR IF THE MAINS LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
NOTES
CAUTIONS
USE ONLY A CCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY SCHILLER AG.
USE OF OTHER THAN RECOMMENDED OR SUPPLIED PARTS MA Y RESULT IN INJURY INACCURATE
INFORMATION AND / OR DAMAGE TO THE UNIT.
IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE
NEUTRAL ELECTRODE) COME INTO CONTACT WITH OTHER PERSONS OR CONDUCTING
OBJECTS (EVEN IF THESE ARE EARTHED).
THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH
SIMULTANEOUS USE OF OTHER ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE
STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT DISTANCE FROM THE
ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE
RECORDER.
IF SEVERAL UNITS ARE COUPLED THERE IS A DANGER OF SUMMATION OF LEAKA GE CURRENTS.
WHEN OPERATING SEVERAL DEVICES FOR MEDICAL AND NON-MEDICAL APPLICATION DO
NOT USE ANY EXTENSION CABLES OR DISTRIBUTION BOXES FOR THE CONNECTION.
EARTH MUST BE CONNECTED WHEN THE EXTERNAL PRINTER IS USED.
Art.No. 2.540022
5
WARNINGS
SERVICING IS TO BE CARRIED OUT BY QUALIFIED PERSONNEL ONLY .
DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE
PRESENCE OF FLAMMABLE GASES SUCH AS ANAESTHETIC AGENTS.
THIS PRODUCT IS NOT DESIGNED FOR STERILE USE.
THIS PRODUCT IS NOT DESIGNED FOR OUTDOOR USE.
DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID.
DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO
NOT USE E-BEAM OR GAMMA RADIATION STERILISATION.
THIS UNIT IS CF
PATIENT CONNECTION IS FULLY ISOLATED AND DEFIBRILLATION PROTECTED. SCHILLER CAN
ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION VOLTAGE HOWEVER, WHEN THE
ORIGINAL SCHILLER PATIENT CABLE IS USED.
DO NOT TOUCH THE CASING DURING DEFIBRILLATION.
IF THE PATIENT CABLE SHOULD BECOME DEFECTIVE AFTER DEFIBRILLATION, LEAD OFF WILL
BE DISPLAYED AND AN ACOUSTIC ALARM GIVEN.
CLASSIFIED ACCORDING TO IEC EN 60601-1. THIS MEANS THAT THE
6
What is in this book
THE SERVICE PHILOSOPHY FOR ALL SCHILLER UNITS IS F AULT FINDING TO MODULE LEVEL ONL Y.
THE PURPOSE OF THIS BOOK IS TO PROVIDE ALL THE INFORMATION NECESSARY TO ENABLE
THE SERVICE ENGINEER TO EFFICIENTL Y LOCA TE AND REPLACE A FAUL TY MODULE. THIS BOOK
ASSUMES NO DETAILED KNOWLEDGE OF THE ECG UNIT BUT DOES REQUIRE THA T THE SERVICE
ENGINEER IS F AMILIAR WITH STANDARD WORKSHOP PRACTICES.
The book is divided into the following chapters:
Chapter 1 - Operating Elements
The purpose of this chapter is to provide an easy reference for all main operator functions
and to give a basic introduction to the AT-1 / AT -1smartprint. This chapter gives details of
the operator controls with the operation and function of each key briefly explained.
The information in this chapter provides a background to the operating functions only .
Complete operating information is provided in the SCHILLER AT-1 / AT-1 smartprint User
Guide.
Chapter 2 - Functional Overview
This chapter provides a functional overview of the AT-1 / AT-1 smartprint. The description
is supported by functional block diagrams.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Chapter 3 - Fault Diagnosis
This chapter provides a guide to locate a fault to module level. The diagnostics are
presented in a logical sequence of fault finding algorithms and procedures. Illustrations
are provided to support the text where needed.
Chapter 4 - Module Removal and Replacement
This chapter gives an overview of the physical construction of the AT-1 / AT-1 smartprint
with the main physical attributes of the unit briefly described. The physical description
is supported by illustrations showing the internal location of all modules. Removal and
replacement instructions for all removable modules are also provided in this chapter.
Each procedure is autonomous with details of tools, jumper settings, adjustments and
settings or special requirements that are required before and after replacement.
Functional checks that must be carried out after replacing a module are also provided.
Chapter 5 - Adjustments
This chapter provides all adjustments and settings. Also detailed in this chapter are
basic functional test procedures that can be performed to check the functioning of
the unit.
Chapter 6 - Spare Parts
This chapter provides the part numbers and reordering information for all replaceable
modules. Also included in this chapter are details of any special test equipment or
special tools required for adjustment or fault finding procedures.
Art.No. 2.540022
7
What's in this book
Chapter 7 - Technical Data
The full technical specification of the AT-1 / AT-1 smartprint is given in this chapter .
Index
Circuit Diagrams & Board Layouts
The circuit diagrams and component layouts are provided for all boards. These details
are provided for information only .
General Symbols
8
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Contents
Associated Documents2
Where to Obtain Service3
Warranty4
Safety Notices5
What is in this book7
CHAPTER 1OPERATING ELEMENTS11
Introduction12
Location & Power13
The Keyboard14
AT-1/AT-1smartprint Short Form Operating Instructions15
with external printer connected and switched on:1 5
with external printer NOT connected or switched off:15
Modes of Operation16
Automatic Mode1 6
Manual Mode1 6
Automatic Mode17
a. without external printer connected or printer switched off17
b. with external printer18
Settings20
Default Settings22
Language23
Filters24
Baseline Filter24
Mains Filter24
Myogram Filter2 5
Defining Lead Sequence & Printout2 6
Acoustic QRS Indication2 7
Time / Date27
Automatic Mode (ECG) Settings, Internal Printer28
Average Cycles2 8
Measurements and Markings (M and C versions only)2 9
Interpretation (C version only)29
Automatic Mode (ECG) Settings, External Printer30
Average Cycles3 0
Measurements and Markings (M and C versions only)3 1
Interpretation (C version only)31
Art.No. 2.540022
9
Automatic Mode (ECG) Settings32
Interpretation Settings (C version only)32
Selecting Rhythm Leads33
Service Printout34
Installing Software Options35
Care & Maintenance36
Replacing the Recording Paper in the Internal Printer37
Thermal Paper Handling38
CHAPTER 2FUNCTIONAL OVERVIEW39
Main Board MK 11-1042
CHAPTER 3FAULT DIAGNOSIS45
Fault Diagnosis Chart46
General Functional Check Procedures48
Internal Printer Check49
CHAPTER 4MODULE REMOVAL AND REPLACEMENT51
Warnings and Cautions54
Physical Overview55
Opening the Case56
Top Housing Replacement57
Physical Overview58
ECG Interface Board MK 11-259
Printer Tray Assembly and Thermal Printer60
Thermal Printer Removal61
Control and Power Supply Board MK 11-1062
Battery Pack64
Keyboard66
10
CHAPTER 5ADJUSTMENTS67
Introduction68
Warnings, Cautions and Conditions69
Test Equipment70
Battery Charge Voltage71
ECG Amplifier Reference Voltage72
CHAPTER 6SPARE PARTS75
Ordering Information76
Spare Parts77
CHAPTER 7TECHNICAL DATA79
Technical Data80
Index83
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Chapter1
Operating Elements
Art.No. 2.540022
11
Introduction
The AT-1 / AT-1 smartprint is a 3-channel ECG recorder with all (12) ECG signals
simultaneously processed to provide instant ECG recordings. Three automatic recording
modes - two for the internal printer and one for the external printer - can be individually
preset to enable one button ECG recording of preferred print formats.
Individual lamps are provided to give power , paper error , filter , lead group and lead off
indications. In addition, any detected disturbance (i.e. loose electrode or end of paper),
gives an audible alarm and the corresponding indicator lamp flashes.
The AT-1 / AT-1 smartprint includes the following features:
• Low weight and compact dimensions
• Printout from integrated quality thermal printer - OR - from external A4 inkjet
printer
• Built-in rechargeable battery for mains-independent use - 2 hrs normal use on one
battery charge
• Simple one key operation for main functions
• Automatic or manual recording modes
• Selectable printing formats - external and internal printer
• Interpretation program option (including measurements) for children and adults
• Choice of 10 languages (AT-1) and 9 languages (AT -1 smartprint) for printing.
12
Location & Power
Location
Do not keep or operate the device in a wet, moist, or dusty environment. Also avoid
exposure to direct sunlight or heat from other sources. Do not allow the unit to come
into contact with acidic vapours or liquids, as such contact may cause irreparable
damage. The unit should not be placed near X-ray or diathermy units, large transformers
or motors. The unit must be placed on a flat surface and must not be operated in areas
where there is any danger of explosion.
Power Supply
The mains connection is on the rear of the unit. The mains indicator lamp on the keyboard
is always lit when the unit is connected to the mains supply. The unit can either be
operated from the mains supply or from the built-in rechargeable battery. The power
source is indicated by the respective indicator lamp. When battery capacity is limited,
the battery symbol flashes on and off.
To recharge the battery, connect the device to the mains supply by means of the
supplied power cable. A totally discharged battery needs less than 15 hours to be fully
recharged (60% in less than 3 hours, 90% in less than 7 hours). A fully charged battery
gives approximately 2 hours of normal use. The unit can remain connected to the mains
supply without any danger of damage to either the battery or the unit.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Switching On and Off
The AT-1 / AT-1 smartprint is switched on with the green ON key
and off by means of the red OFF key
The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is
limited) if no key is pressed and the patient cable is not connected.
Potential Equalisation
If the A T-1 / AT-1 smartprint is used in conjunction with other patient connected equipment,
we recommend that the potential equalisation stud on the rear of the unit is connected
to the building common ground with the yellow/green ground cable (Part-no. 2.310005).
When working from an emergency vehicle, the vehicle common ground can be used.
WARNING !
THE UNIT MUST BE GROUNDED WHEN THE EXTERNAL PRINTER IS CONNECTED.
Art.No. 2.540022
13
The Keyboard
14
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
AT-1/AT-1smartprint Short Form Operating Instructions
Automatic ECG Recording
with external printer connected and switched on:
Prepare skin, hook up patient.
Switch unit on, press ON
Press AUTO to record and print externally
with external printer NOT connected or switched
off:
To record and print on internal printer:
For AUTO Format 1 press
Press
electrode dc offsets.
Best results are obtained when the electrode voltage
readings (right column) are between +50mV and - 50mV .
Filter On/Off
Press
System Configuration
Press
settings.
to switch the (Myogram) filter On / Off.
for print-out of
to print system
15
Modes of Operation
Automatic Mode
a. without external printer connected or printer switched off
Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads
with a choice of 2 different formats (only for S version).
The following can be programmed freely for each of the 2 formats before recording:
• Lead Format
• Chart Speed
• With the optional interpretation program installed it is also possible to select the
measurement table, average cycles with optional markings and interpretation
statements for the printout.
For further information see paragraph `Automatic Mode (ECG) Settings` on page 28.
b. with external printer connected and switched on
Only one Automatic Mode format possible. Other details as above.
For further information see paragraph `Automatic Mode (ECG) Settings` on page 30.
Manual Mode
Manual Mode provides a real time print-out of 3 selected leads and is only possible on
the internal thermal printer.
The following can be freely selected before or during recording:
• Lead Group
• Chart Speed
• Sensitivity
• Myogram Filter
For further information see paragraph `Manual Mode` on page 19.
16
Automatic Mode
a. without external printer connected or printer switched off
In automa tic m ode, a full 12-lead ECG is printed in one of two predefined formats
with selectable sensitivity. These formats (see Settings) are selected by the user to suit
his specific needs and requirements.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
When the AUTO GAIN key
unit detects very large waveform amplitudes and sets the sensitivity for the extremity
and/or precordial leads accordingly to reduce the overlapping of traces.
To start the automatic ECG recording in Format 1, press the AUTO START
key:
To start the automatic recording in Format 2, press the ALT key followed by
the AUTO START key:
Depending upon which option is included, the printout gives the following:
• ECG recording of all leads in either Standard or Cabrera format according to
selection
• Sensitivity
• Heart Rate
• Speed
• Filter Settings
• Interpretation statements (with option C)
• Average Cycles (with options M, C)
• Intervals (with options M, C)
• Axes (with options M, C)
• Detailed Measurement Table (with options M, C)
To obtain an extra printout of the ECG recording in Format 1, simply press the COPY key:
is pressed before recording in automatic mode, the
+
To obtain an extra printout of the second format, press the ALT key followed by the
COPY key:
Note:The Auto mode settings for the two formats are detailed in the paragraph
Art.No. 2.540022
-
entitled ' Automatic Mode (ECG) Settings` later in this chapter .
17
Automatic Mode
b. with external printer
With the external printer connected and switched on, only one Auto Mode format
(selectable) is available.
To start the automatic ECG recording in Automatic Mode, press the AUTO
START key:
For further information see paragraph `Automatic Mode (ECG) Settings` on page 30.
18
Manual Mode
Manual mode provides a real-time ECG printout of 3 selected leads with full control
of parameter selection. This is only possible on the internal thermal printer.
To start the manual recording of a real-time ECG, press the MAN START
Printout key
To stop the manual recording (printout) press the STOP key
The printout provides you with the following:
• Three (selected) leads with lead identification.
• On the lower edge, the chart speed and filter settings (if on).
• At the top, the heart rate as current average of 4 beats, trace sensitivity.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
The following can be freely chosen during or before the recording:
Lead Groupby means of the LEAD FORWARD
and LEAD BACKWARD key
The following lead groups are selectable:
•I, II, IIIaVR, aVL, aVF
(Cabrera: aVL, I, -aVR / II, aVF, III)
•V1, V2, V3/V4, V5, V6
Chart SpeedSelect speed 5, 25 or 50 mm/s by means of the SPEED keys:
SensitivitySelect 5, 10 or 20 mm/mV by means of the SENSITIVITY keys:
Myogram FilterSwitch the filter ON or OFF with the FILTER key:
25 Hz or 35 Hz is displayed on the bottom line of the printout
when the filter is switched on.
RecenteringT o re-centr e the ECG traces, press the 1mV key
Art.No. 2.540022
19
Settings
Each parameter is set by means of a code. This code comprises a combination of keys
starting with the ALT key followed by two or three numbers. The setting is confirmed with
the ST OP key. As soon as the ALT key is pressed, the keyboard is dedicated to the
programming function.
Note:The Alternative (ALT) function is only active for 4 seconds. If a programming
The setting is remembered and the keyboard released for other functions when the
STOP key is pressed. Once a setting has been confirmed, it is stored in the memory
even when the unit is switched off.
Example
If you want to set the language on your ECG unit to English, the key sequence given in
the table (see page 23), is
key is not pressed within 4 seconds, the unit reverts to standard mode. The
ALT key must again be pressed to activate the programming mode
ALT - 0 - 2 - 2 - STOP.
20
On the following pages the programmable parameters and the programming sequences
are described in detail.
Settings (cont.)
The defined formats and settings for your unit can be checked as follows:
A printout of the defined settings will be produced and gives the following information,
depending on the installed software.
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
ALT - 0 - 1 - any number
Unit designationSoftware version, Software option installed (C =
Interpretation) and interpretation version
Serial numberSerial number of the unit
LeadsStandard (S) or Cabrera (C)
ECG FormatLong (ooo), Short (o) or Suppressed (-)
M ECGAverage cycles as defined in auto ECG recording setup
(e.g. 4 * 3 (25 mm/s) + 2)
MeasurementsEnabled (+) or Suppressed (-)
MarksE n a b l e d (+ ) o r Su ppressed (-)
InterpretationEnabled (+) or Suppressed (-)
Selected Rhythm leadsLeads selected for R1 and R2
Automatic CenteringEnabled (+) or Suppressed (-)
Paper typeZ-folded
Pr intout of signal sSequential or Simultaneous
Baseline Filter0. 0 5 , 0 . 1 5 o r 0 .3 0 H z
Mains Filter50, 6 0 Hz o r OFF ( -)
Myogram Filter25 or 35 H z, O N (+ ) or OFF (-)
Interpretation settings:N/A: + /-‘no rmal/a bno rma l’ i s wr itten (+) or
suppressed (-)
U:+/-‘unconfirmed report’ is written (+) or
suppressed (-)
A30:+/-patient age is assumed to be < 30 (-)
or >30 (+)
S: +/-low (-) or high (+) sensitivity
Art.No. 2.540022
21
Settings (cont.)
Default Settings
To reset the unit to the basic default settings, proceed as follows:
ALT - 0 - 6 - 6
SETTINGSS = STANDARDM = MEASUREMENTS
LANGUAGEAS SETAS SETAS SET
LEADSSTANDARD (S)STANDARD (S)STANDARD (S)
ECG: 25MM/S, SHORT (O)
AUTO FORMAT 1
PATIENT DATA FORM
ENABLED +
ECG: 25MM/S, LONG (OOO)
AUTO FORMAT 2
PATIENT DATA FORM
ENABLED +
RHYTHM LEADSV1, IIV1, II
AUTOM.
CENTERING
PAPERZ - FOLDZ - FOLDZ - FOLD
PRINTOUT OF
SIGNALS
BASELINE FILTER
SETTING
MAINS FILTER
SETTINGS
MYOGRAM
FILTER SETTING
INTERPRETATION
SETTINGS
EXTERNAL
PRINTER
ENABLED (+)ENABLED (+)ENABLED (+)
SEQUENTIALSEQUENTIALSEQUENTIAL
0.05HZ0.05HZ0.05HZ
50HZ (60HZ)50HZ (60HZ)50HZ (60HZ)
35HZ, OFF (-)35HZ, OFF (-)35HZ, OFF (-)
ECG: 25MM/S, SHORT (O)
PATIENT DATA FORM
ENABLED +
ECG : 25MM/S, SHORT (O)ECG : 25MM/S, SHORT (O)
M ECG :6*2 (50MM/S + 1)M ECG: 6*2 (50MM/S + 1)
MEASUREMENTS:
SUPRESSED (-)
MARKS: ENABLED (+)MARKS: ENABLED (+)
ECG : 25MM/S, LONG
(OOO)
M ECG: - -M ECG: NONE
MEASUREMENTS:
SUPRESSED (-)
MARKS: ENABLED (+)MARKS: ENABLED (+)
ECG : 25MM/S, SHORT (O)ECG : 25MM/S, SHORT (O)
M ECG: 6*2 (50MM/S) + 2M ECG: 6*2 (50MM/S) + 2
MEASUREMENTS:
SUPRESSED (-)
MARKS: ENABLED (+)MARKS: ENABLED (+)
C = WITH
INTERPRETATION
MEASUREMENTS:
SUPRESSED (-)
INTERPRETATION:
ENABLED (+)
ECG : 25MM/S, LONG
(OOO)
MEASUREMENTS:
SUPRESSED (-)
INTERPRETATION:
DISABLED (-)
N/A: SUPRESSED (-)
U: ENABLED (+)
A30: UNDER THIRTY (- )
S: LOW (-)
MEASUREMENTS:
SUPRESSED (-)
INTERPRETATION:
ENABLED (+)
22
Settings (cont.)
Language
The language is selected as follows:
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Language Selection AT-1
Entry Key SequenceLanguageConfirm
1German
2English
3French
4Swedish
ALT02
5American
6Italian
7Spanish
8Portugese
9Dutch
0Russian
STOP key
Press
Language S election AT-1 smartprint
Entry Key SequenceLanguageConfirm
1German
2English
3French
4Swedish
ALT02
5American
6Italian
7Spanish
8Portugese
9Dutch
Confirm the selection by pressing
Notes: The difference between English and American is the mains filter setting -
English = 50Hz; American = 60Hz and the physical units.
Press
STOP key
Art.No. 2.540022
23
Settings (cont.)
Filters
There are three different filters which can be set individually as follows:
• Baseline filter
• Mains filter
• Myogram filter
Baseline Filter
The digital Baselin e fi lte r suppresses excessive baseline drifts. The setting options are
as follows:
Baseline Filter
Entry Key SequenceFilter SettingConfirm
ALT5
00.05 H z (default)
10.15 Hz
30.30 Hz
Press STOP
key
Confirm the selection by pressing
Note:The set value is the lower limit of the frequency range and is normally set to
0.05 Hz. The settings 0.15 and 0.30 Hz should only be used when absolutely
necessary , as the possibility exists that they could affect the original ECG
signal, especially the ST segments.
Mains Filter
The Mains filter is an adaptive digital interference filter designed to suppress AC
interference without attenuating or distorting the ECG.
Set the mains filter in accordance with the frequency of your local mains supply as
follows:
Mains Filter
Entry Key
Sequence
ALT8
Filter SettingConfirm
5Mains Filter 50 Hz
6Mains Filter 60 Hz
9Mains Filter Off
Press STOP
key
24
Settings (cont.)
Myogram Filter
The Myogram filter suppresses disturbances caused by strong muscle tremor. The set
value will be the new upper limit of the frequency range as soon as the FILT ER key is
pressed on or programmed as default when the unit is switched on. When the Myogram
filter is on, the value, i.e. 35 Hz is displayed on the bottom line of the printout.
Entry Key SequenceSettingConfirm
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Myogram Filter
2Myogram Filter 25 Hz
3Myogram Filter 35 Hz
ALT8
Myogram Filter active when the unit is first
1
switched on (marked on printout with +)
Myogram Filter off when the unit is first
8
switched on (marked on printout with - )
Press STOP
key
Confirm the selection by pressing the STO P key
The myogram filter is switched on and off manually with the FILTE R KEY
Note:An ECG recorded in auto mode is stored unfiltered. It is therefore possible to
print the stored ECG either with or without passing the myogram filter. Filter
ON is indicated in the bottom information line of the printout. When theFILTER
key is pressed again, the filter is switched off and the `35 Hz` indication on
the bottom information line of the printout is removed. The cutoff frequency
of the myogram filter is set to either 25 or 35 Hz.
Art.No. 2.540022
25
Settings (cont.)
Defining Lead Sequence & Printout
The required settings can be selected as follows:
Sequences, Print & Auto-centering
Entry Key SequenceDefinitionConfirm
1Standard Lead Sequence
2Cabrera Lead Sequence
3Simultaneous Print
ALT7
Confirm the selection by pressing the STOP key
4Se quential Print
5Auto-centering ON
6Auto-centering OFF
7Z-fold paper
Press STOP
key
The selectable printout forms are:
SimultaneousAll ECG le ads are printed in the same time segment ( in
automatic mode only).
SequentialEach group is a contiguous time segment of approximately
2.5 or 5 seconds (in automatic mode only).
Auto-Centering ONAll ECG traces are centred dynamically for optimal use of
paper width.
Auto-Centering OFFECG traces are set to a fixed baseline position and may
possibly overlap.
The Standard and Cabrera lead groups available for the AT-1 / AT-1smartprint are:
Lea d Groups
StandardCabrera
IaVRV1V4aVLIIV1V4
IIaVLV2V5IaVFV2V5
IIIaVFV3V6-aVRIIIV3V6
26
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.