This document contains proprietary information that is protected by copyright. All Rights Reserved.
Reproduction, adaptation, or translation without prior written permission is prohibited, except as
allowed under the copyright laws.
Warranty
The information contained in this document is subject to change without notice.
Schiller makes no warranty of any kind with regard to this material, including, but not limited to, the
implied warranties or merchantability and fitness for a particular purpose.
Schiller shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
Revision History
The documentation part number and revision number indicate its current edition. The revision
number changes when a new edition is printed in accordance with the revision history of the
documentation. Minor corrections and updates which are incorporated at reprint do not cause the
revision number to change. The document part number changes when extensive technical changes are
incorporated.
Related Documents.......................................................................................................................................... 3
Description of the ARGUS VCM monitor .................................................................................................... 3
Service Mode ..................................................................................................................................................25
How to Download .........................................................................................................................................33
How to Use This Section ...............................................................................................................................35
Who Should Perform Repairs.......................................................................................................................35
General ............................................................................................................................................................43
Prior to Disassembly...................................................................................................................................... 45
Parts List.......................................................................................................................................................... 58
ii ARGUS VCM Service Manual
ARGUS VCM
Contents
Packing For Shipment 61
General Instructions ......................................................................................................................................61
Returning the ARGUS VCM......................................................................................................................... 61
Repacking In Original Carton ...................................................................................................................... 61
Repacking In a Different Carton.................................................................................................................. 62
System Processing Description 63
System Overview ...........................................................................................................................................63
Temperature Processing ................................................................................................................................ 70
ARGUS VCM Service Manual iii
Contents
ARGUS VCM
Figures
Figure 1. ARGUS VCM Front Panel............................................................................................................................................... 4
Figure 3. ARGUS VCM Left Side Panel.......................................................................................................................................... 6
Figure 4. ARGUS VCM Right Side Panel....................................................................................................................................... 6
Figure 5. System Version............................................................................................................................................................... 25
Figure 6. Battery Voltage Level..................................................................................................................................................... 26
Figure 18. Front Case Disassembly.............................................................................................................................................. 47
Figure 19. Main BD, SpO2, FND modules Disassembly .............................................................................................................. 48
Figure 20. Front Case Disassembly – Overlay, NIBP cuff hose connector................................................................................... 49
Figure 21. Rear Case Disassembly- Battery, SMPS...................................................................................................................... 50
Figure 22. Rear Case Disassembly- NIBP module ....................................................................................................................... 51
Figure 23. Rear Case Disassembly- Printer ................................................................................................................................. 52
Figure 24. Rear Case Disassembly – Speaker, Handle, etc........................................................................................................... 53
Figure 25. Temperature Case Disassembly – Temperature Module.............................................................................................. 54
Figure 26. Temperature Case Disassembly ................................................................................................................................... 55
Table 4. Patient Leakage Current Values ...................................................................................................................................... 22
Table 5. Patient Leakage Current Values—Mains Voltage on Applied Part.................................................................................. 23
Table 6. Test Lead Combinations .................................................................................................................................................. 24
Table 11. Part Descriptions – Monitor Assembly.......................................................................................................................... 46
Table 12. Part Descriptions – Front Case Assembly..................................................................................................................... 47
Table 13. Part Descriptions – Main BD, SpO2, FND Modules Assembly..................................................................................... 48
Table 15. Part Descriptions – Battery, SMPS Assembly ............................................................................................................... 50
Table 16. Part Descriptions – NIBP Module Assembly................................................................................................................. 51
Table 17. Part Descriptions – Printer Assembly........................................................................................................................... 52
Table 18. Part Descriptions – Speaker, Handle, etc Assembly...................................................................................................... 53
Table 19. Part Descriptions – Temperature Module Assembly...................................................................................................... 54
Table 20. Part Descriptions – Temperature Case Assembly.......................................................................................................... 55
Table 21. ARGUS VCM Parts List ................................................................................................................................................ 59
iv ARGUS VCM Service Manual
Introduction
Manual Overview
Related Documents
Description of the ARGUS VCM monitor
Warnings
Warnings are identified by the WARNING symbol shown above.
Warnings alert the user to potential serious outcomes (death, injury, or adverse
events) to the patient or user.
WARNING: Explosion hazard. Do not use ARGUS VCM in the presence of
flammable anesthetics or gases.
WARNING: Do not spray, pour, or spill any liquid on ARGUS VCM, its accessories,
connectors, switches, or openings in the chassis.
WARNING: Do not immerse ARGUS VCM or its accessories in liquid or clean with
caustic or abrasive cleaners.
WARNING: Ensure that conductive portions of the cables do not come into contact
with any other conductive parts.
WARNING: Electrical shock hazard. Disconnect the power cord from ARGUS VCM
before attempting to open or disassemble ARGUS VCM.
WARNING: The use of accessories, transducers and cables other than those specified
may result in increased emission and/or decreased immunity of ARGUS VCM
monitor.
WARNING: Do not silence ARGUS VCM audible alarm or decrease its volume if
patient safety could be compromised.
WARNING: During safety tests, AC mains voltage will be present on the applied
part terminals. Exercise caution to avoid electrical shock hazard.
WARNING: Do not place ARGUS VCM into operation after repair or maintenance
until Performance, Safety Tests and NIBP Calibration listed in this service manual
have been performed. Failure of these tests could result in erroneous readings.
ARGUS VCM Service Manual 1
Introduction
Cautions
ARGUS VCM
Cautions are identified by the Caution symbol shown above.
Cautions alert the user to exercise care necessary for the safe and effective use of
ARGUS VCM monitor.
CAUTION: Observe ESD (electrostatic discharge) precautions when working within
the unit and/or when disassembling and reassembling ARGUS VCM monitor and
when handling any of the components of ARGUS VCM monitor.
CAUTION: When reassembling ARGUS VCM, over-tightening could strip out the
screw holes in the cases, rendering it unusable.
CAUTION: If any problem with ARGUS VCM built in an optional printer, check a
printer’s door is closed well. Operating error may be caused if the cover is not closed
correctly.
CAUTION: If internal battery cable has been disconnected, pay particular attention to
polarity of the cable before reattaching. If battery cable polarity is reversed, it is likely
that circuit damage will occur.
CAUTION: Ferrite Cores are used for electromagnetic compatibility. Please do not
remove Ferrite Cores while disassembling or reassembling, otherwise the monitor can
be affected by electromagnetic interference and measure inaccurate data to be
displayed or stored.
2 ARGUS VCM Service Manual
ARGUS VCM
Manual Overview
This manual contains information for servicing
The monitor subsequently referred to as
qualified service personnel should service this product. Before servicing
read the operator’s manual carefully for a thorough understanding of safe operation.
Read and understand all safety warnings and service notes printed in this service
manual and the operator’s manual.
This manual contains information about ARGUS VCM monitor. ARGUS VCM monitor
includes the following configuration:
Config. Features Config. Features
N Standard (NIBP + Pulse Rate) NP Standard + Printer
NS Standard + SpO2NSPStandard + SpO2 + Printer
NT Standard + Temperature NTPStandard + Temperature+ Printer
NSTStandard + SpO2 + Temperature NSTPStandard + SpO2 + Temperature + Printer
All information in this manual, including the illustrations, is based on the monitor
configured with Temperature, SpO
of these options, some information in this manual does not apply.
2 and Printer options. If your monitor lacks any
Related Documents
To perform test and troubleshooting procedures and to understand the principles of
operation and circuit analysis sections of this manual, you must know how to operate
the monitor. Refer to
To understand the various blood pressure cuffs, SpO
that work with the monitor, refer to the individual directions for use that accompany
these accessories.
ARGUS VCM operator’s manual.
Description of the ARGUS VCM monitor
The purpose and function of Schiller
blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial
oxygen saturation, pulse rate, and temperature for adult, pediatric and neonate patients
in all hospital areas and hospital-type facilities. It may be used during hospital
transport and in mobile, land-based environments, such as ambulances, within the
specification of the environmental characteristics. Monitor users should be skilled at
the level of qualified health care professionals, such as a technician, doctor, nurse or
medical specialist.
Note: Hospital use typically covers such areas as general care floors, operating rooms,
special procedure areas, intensive and critical care areas, within the hospital plus
hospital-type facilities. Hospital-type facilities include physician office based facilities,
sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
Note: Intra-hospital transport includes transport of a patient within the hospital or
hospital-type facility.
ARGUS VCM monitor is to monitor noninvasive
Introduction
ARGUS VCM monitor.
ARGUS VCM throughout this manual. Only
ARGUS VCM,
2 sensors and temperature probes
ARGUS VCM Service Manual 3
Introduction
ARGUS VCM
The physical and operational characteristics of the monitor are described in the
operator’s manual and in the Specification section of this manual.
Figure 1 and 2 identify the displays, controls, indicators and symbols of the front and
rear panels.
Note: Systolic and diastolic blood pressure measurements determined with this device are equivalent to
those obtained by a trained observer using the cuff/stethoscope auscultation method, within the limits
prescribed by the American National Standard, Electronic or automated sphygmomanometers.
WARNING: Do not spray or pour any liquid on the monitor or its accessories. Do
not immerse ARGUS VCM or its accessories in liquid or clean with caustic or
abrasive cleaners.
Cleaning
To clean ARGUS VCM, dampen a cloth with a commercial, nonabrasive cleaner and wipe
the exterior surfaces lightly.
CAUTION: Do not allow any liquids to come in contact with the power connector or
switches or to penetrate connectors or openings in the instrument.
Note: For cuffs, sensors and probes, follow the cleaning instructions in the directions
for use that accompany these accessories.
Note: If liquid is spilled on the monitor, clean and dry thoroughly before reuse.
Note: If in doubt about the monitor safety, refer the unit to qualified service personnel.
For surface-cleaning, follow your institution’s procedures or:
ARGUS VCM may be surface-cleaned by using a soft cloth dampened with
either a commercial or nonabrasive cleaner, and lightly wiping the top,
bottom, and front surfaces of the monitor lightly.
Periodic Safety and Functional Checks
ARGUS VCM requires no routine service or calibration other than cleaning and battery
maintenance. The following performance verification tests may be used following repair
or during routine maintenance (if required by your local institution).
1. Inspect the exterior of
2. Inspect labels for legibility. If the labels are not legible, contact Schiller Technical
Services Department.
3. Verify that the unit performs properly as described in Performance Verification
section.
4. Perform the electrical safety tests detailed in Performance Verification section. If the
unit fails these electrical safety tests, do not attempt to repair. Contact Schiller
Technical Services Department.
ARGUS VCM for damage.
ARGUS VCM Service Manual 11
Routine Maintenance
Functional Checks
The following checks should be performed at least every 2 years by a qualified service
technician.
1. If the monitor has been visibly damaged or subjected to mechanical shock (for
example, if dropped), perform the performance tests as described in Performance Verification section.
2. Perform the electrical safety tests as described in Safety Tests section. If the unit fails
these electrical safety tests, refer to Troubleshooting section.
3. Inspect the fuses for proper value and rating
- qty 2, 2.0 A, 250 volts for AC mains (L/N: F1, F2)
- qty 1, 6.3 A, 250 volts for internal battery (L/N: F3)
Batteries
If ARGUS VCM has not been used for a long period of time, more than 2 months, the
battery will need charging. To charge the battery, connect
as described in Performance Verification section of this service manual or the Battery Operation section of the operator’s manual.
Schiller recommends replacing the instrument’s battery every 2 years. When
VCM
is going to be stored for 2 months or more, it is recommended to remove the
battery prior to storage. To replace or remove the battery, refer to Disassembly Guide.
Note: Storing
the battery capacity. The battery may require a full charge/discharge cycle to
restore normal capacity. A full charge of a dead battery takes approximately 12
hours while the monitor is turned off. Schiller recommends that
sealed, lead acid batteries be replaced at 2-year intervals. Refer to Disassembly
Guide Section.
CAUTION: If ARGUS VCM is to be stored for a period of 2 months or longer, it is
recommended to notify service personnel to remove the battery from the monitor prior
to storage. Recharging the battery is strongly recommended when the battery has not
been recharged for 2 or more months.
CAUTION: If the battery shows any signs of damage, leakage, or cracking, it must be
replaced immediately.
CAUTION: Discarded battery may explode during incineration. Recycle used batteries
properly. Do not dispose of batteries in refuse containers.
ARGUS VCM for a long period without charging the battery may degrade
Environmental Protection
Follow local governing ordinances and recycling plans regarding disposal or recycling
batteries and other device components.
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing
All tests except battery charge and battery performance tests must be performed as the
last operation before the monitor is returned to the user.
If
ARGUS VCM fails to perform as specified in any test, repairs must be made to correct
the problem before the monitor is returned to the user.
Equipment Needed
Table 1 lists the equipment required for performance verification.
Table 1. Required Test Equipments
ARGUS VCM covers.
EquipmentDescription
Digital Multi Meter (DMM) Fluke Model 87 or Equivalent
NIBP Cuff Durable, Adult 11cm
NIBP Cuff Durable, Neonatal 5cm
NIBP Hose Adult, 8 feet
NIBP Hose Neonatal, 6 feet
NIBP Rigid PVC Vessel 9cm diameter
NIBP Rigid PVC Vessel 5cm diameter
SpO2 Sensor Extension Cable DOC-10
Temperature Probe Oral Probe
NIBP Simulator Bio-Tek BP Pump 2 or Equivalent
SpO2 Simulator NELLCOR SRC-MAX
Temperature Simulator WelchAllyn 9600 Calibration Tester
Safety Analyzer METRON QA-90 or Equivalent
Stopwatch Manual or electronic
ARGUS VCM Service Manual 13
Performance Verification
Performance Tests
The battery charge and battery performance test should be performed before the
monitor repairs whenever the battery is suspected as being a source of the problems.
All other tests may be used following repairs or during routine maintenance (if
required by your local institution). Before performing the battery performance test,
ensure that the battery is fully charged. This section is written using factory defaults set
as power-up. If your institution has preconfigured custom defaults, those values will be
displayed.
Power
1. Connect the monitor to AC power source using proper power cord.
2. Verify Charging/AC in indicator is lit (or flashing).
3. Press Power button over 1 second, and then verify that the monitor is turned on.
4. After the monitor operates in normal mode, disconnect the power cord.
5. Verify Battery indicator is lit instead of Charging/AC in indicator.
6. Press Power button over 1 second, and then verify that the monitor is turned off.
Battery Charge
1. Connect the monitor to AC power source using proper power cord. (the monitor is
turned off)
2. Verify Charging/AC in indicator is lit (or flashing).
ARGUS VCM
Note: If Charging/AC in indicator is flashing, the battery needs fully charging.
3. Charge the battery fully until Charging/AC in indicator is no more flashing. It takes
for at least 12 hours.
4. To check for a full charge, disconnect the power cord. Press Power button and the
NIBP start/stop button simultaneously over 1 second to enter the Service mode. Then
select battery voltage level displayed.
5. Verify that current battery voltage display in Service mode indicates ‘6.0V’ or greater.
Note: The battery may require a complete charge/discharge cycle to restore its normal
capacity, depending on its previous usage.
Battery Discharge
1. Disconnect the power cord from the monitor with fully charged battery.
2. Turn on the monitor by pressing Power button over 1 second.
3. Connect NIBP simulator to the monitor. Set NIBP simulator:
SYS of 120mmHg, DIA of 80mmHg, PR of 80 bpm.
4. Set NIBP Auto interval of the monitor to 15 minutes.
5. After more than 2 hours of the monitor operating, verify an audible alarm is sounding
and the battery level of the monitor is under 5.4V.
6. Allow the monitor to operate until it automatically powers down due to low battery
condition. Verify that high priority alarm occurs 5 seconds before the monitor
automatically shuts down.
7. If the monitor passes this test, immediately recharge battery.
14 ARGUS VCM Service Manual
ARGUS VCM
Performance Verification
Power-On Self-Test
1. Connect the monitor to AC power source and verify Charging/AC in indicator is lit.
2. Observe the monitor front panel. With the monitor off, press Power button. The
Note: Power-on self-test takes approximately 7 seconds to complete.
Note: If an error code is displayed, please refer to Troubleshooting section.
monitor must perform the following sequence.
a. All FNDs and indicators are illuminated.
b. Power-on beeps sounds consecutively.
c. Upon successful completion of power-on self-test, the display will enter the
normal mode.
General Operation Tests
Alarms and Alarm Silence
1. Connect the monitor to an AC power source.
2. Press the monitor Power button to turn the monitor on.
3. Connect NIBP simulator to the monitor.
4. Set NIBP simulator as follows:
Systolic of 120mmHg, Diastolic of 80mmHg and pulse rate of 150 bpm.
5. Press NIBP start/stop button.
6. Verify following the monitor reaction after measurement completes:
a. The monitor displays NIBP and pulse rate as specified by the simulator.
b. Audible alarm sounds and Pulse rate display flashes indicating the parameter
has violated default alarm limits.
7. Press Alarm silence button on the front panel of the monitor.
8. Verify the following:
a. An audible alarm is temporarily silenced.
b. Pulse rate display continues flashing.
c. Alarm silence indicator is lit.
d. Audible alarm returns in approximately 90 seconds.
9. Set NIBP simulator as follows:
Systolic of 120mmHg, Diastolic of 80mmHg and pulse rate of 80 bpm.
10. Press NIBP start/stop button.
11. Verify following the monitor reaction after measurement completes:
a. The monitor displays NIBP and pulse rate as specified by the simulator.
b. The alarm is terminated.
12. Disconnect NIBP simulator from the monitor.
13. Kink NIBP cuff hose connector of monitor and then press NIBP start/stop button.
14. Verify that the monitor displays an error code, E23 and sounds audible alarm.
15. Press Alarm silence button.
16. Verify that the alarm is terminated.
ARGUS VCM Service Manual 15
Performance Verification
Pulse Tone Volume Control
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Connect SpO
monitor.
4. Set SpO
5. Verify SpO2/ pulse rate values and pulse amplitude indicator correctly display, and
pulse tone sounds.
6. Press the mode button to enter setting mode.
7. Press Up (+) selection button to adjust pulse tone volume from 1 to 8. Return to the
monitoring screen.
8. Verify beeping pulse tone increases.
9. Press the mode button to enter setting mode.
10. Press Down (-) selection button to adjust pulse tone volume from 8 to 1. Return to
the monitoring screen.
11. Verify beeping pulse tone decreases.
12. Set pulse tone volume to 0, and return to the monitoring screen. Verify pulse tone is
no longer audible.
2 simulator as follows: SpO2 of 90% and pulse rate of 60 bpm.
ARGUS VCM
2 simulator to the pulse oximetry cable and connect the cable to the
13. Return pulse tone volume to a comfortable level.
Alarm Volume Control
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Kink NIBP cuff hose connector of the monitor and then press NIBP start/stop button.
4. Verify that the monitor ‘E23’ displays and sounds audible alarm.
5. Press the mode button to enter setting mode.
6. Press Up (+) selection button to adjust alarm volume from 1 to 8. Return to the
monitoring screen.
7. Verify alarm tone increases.
8. Press the mode button to enter setting mode.
9. Press Down (-) selection button to adjust alarm volume from 8 to 1. Return to the
monitoring screen.
10. Verify alarm tone decreases.
11. Return alarm volume to a comfortable level.
Real-Time Clock
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Verify that Time display updates every second.
16 ARGUS VCM Service Manual
ARGUS VCM
Printer Test
Flash Memory
Performance Verification
If Printer option installed, the following test procedures will verify printer performance.
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Connect all necessary simulators to the monitor.
4. Press Printer button when the monitor displays the measurements.
5. Verify the measurement values are printed out correctly.
6. Press the mode button over 3 seconds (configuration mode) to select Stream print.
7. Kink NIBP cuff hose connector of the monitor and then press NIBP start/stop button.
8. Verify that the monitor ‘E23’ displays and also prints out.
9. Open the printer door, and then press Print button.
10. Verify that the monitor sounds ‘invalid button tone’.
Perform the following procedure to test the flash memory.
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Change alarm limits and other settings.
Temperature mode C M
Alarm Limit, SYS HI 160
Alarm Limit, DIA LO 15
Alarm Limit, PR HI 180
Alarm Limit, SpO
4. Kink NIBP cuff hose connector of the monitor to generate an error ‘E23’.
5. Turn the monitor off.
6. Press Power button to turn the monitor on again.
7. Verify that previously changed alarm limits and settings are saved/displayed.
8. Press Review button. Verify that ‘E23’ event is saved/displayed.
9. Turn the monitor off.
10. Press Power button & NIBP start/stop button simultaneously to enter Service mode.
11. Set ‘DEFAULT RESET’ to ‘YES’. The monitor will return to the factory defaults.
12. Turn the monitor off.
13. Press Power button to turn the monitor on again.
14. Verify that alarm limits and settings of the monitor are set to the factory defaults.
Refer to Table 8 Factory defaults.
15. Press Review button. Verify invalid button tone sounds and all events of trend data
has been removed.
2 LO 85
ARGUS VCM Service Manual 17
Performance Verification
Measurement Parameter Operation Tests
Pneumatic System Operation
These tests verify the functionality of
1. Place the neonatal cuff with a rigid PVC vessel (5cm diameter). Connect the cuff to
NIBP cuff hose connector via the neonatal hose.
2. Connect the monitor to an AC power source, and then press Power button to turn the
monitor on.
3. Press NIBP start/stop button.
4. Verify that Systolic blood pressure display indicates an error code E23 and an audible
alarm sounds.
5. Press the mode button (setting mode) to change Patient type to Neonatal. Return to
the monitoring screen.
6. Press NIBP start/stop button. Verify that the monitor inflates.
7. Press NIBP start/stop button. Verify that the monitor deflates.
8. Place the adult cuff with a rigid PVC vessel (9cm diameter). Connect the cuff to NIBP
cuff hose connector via the adult hose.
9. Press NIBP start/stop button.
10. Verify that Systolic blood pressure display indicates an error code E22 (or E24) and
an audible alarm sounds.
11. Press the mode button (setting mode) to change Patient type to Adult. Return to the
monitoring screen.
12. Press NIBP start/stop button. Verify that the monitor inflates.
13. Press NIBP start/stop button. Verify that the monitor deflates.
NIBP air leakage test
1. Connect the monitor to an AC power source.
2. Press Power button and NIBP start/stop button to enter Service mode.
3. Select NIBP Pressure Test by pressing Mode button (see figure 12).
4. Set a desired target pressure between 50and 300mmHg by pressing Up/Down (+/-)
selection buttons.
5. Place the adult cuff with a rigid PVC vessel (9cm diameter). Connect the cuff and the
adult hose to NIBP simulator. Then connect the adult hose to the monitor.
6. Set NIBP simulator to Pressure test or Leakage test mode.
7. Select the set point of NIBP simulator as the same point of target pressure of the
monitor.
8. Press NIBP start/stop button of the monitor as simultaneously pressing start button of
NIBP simulator. Current pressure appears on the diastolic pressure display.
9. After current inflating pressure comes up to the desired target pressure, the monitor
starts deflating. Ensure that air leakage rate displays.
10. Verify that air leakage rate is within 6mmHg/min.
ARGUS VCM
ARGUS VCM pneumatic system.
18 ARGUS VCM Service Manual
ARGUS VCM
Performance Verification
NIBP overpressure test
1. Connect the monitor to an AC power source, and then press Power button to turn the
monitor on.
2. Press Mode button. Set to Adult patient type and NIBP target pressure, 270mmHg.
3. Place the adult cuff with a rigid PVC vessel (9cm diameter). Connect the cuff and the
adult hose to NIBP simulator. Then connect the adult hose to the monitor.
4. Set NIBP simulator to Pressure relief or Overpressure test mode.
5. Press NIBP start/stop button of the monitor as simultaneously pressing start button of
NIBP simulator. NIBP simulator pressurizes the system until the monitor’s overpressure relief system activates.
6. Verify that peak point displayed on NIBP simulator (point of protection pressure) is
within 300mmHg.
Also this point of protection pressure may be verified at the moment of the monitor’s
NIBP valve of relieved.
NIBP transducer accuracy test
1. Follow the procedure of NIBP air leakage test specified above.
2. During the deflation of the cuff pressure, verify that the difference is within ±3mmHg
between monitor’s current pressure and simulator’s current pressure.
Pulse Oximetry Operation
2 option installed, the following test procedures will verify SpO2 performance.
If SpO
1. Connect the monitor to an AC power source.
2. Turn on the monitor by pressing Power button.
3. Connect SpO
monitor.
4. Set SpO
5. The monitor will:
- sound an audible alarm
- display an SpO
- display a pulse rate of 60 ±3 bpm
- display Pulse amplitude indicator
6. Disconnect the pulse oximetry cable from the monitor.
7. Verify that the monitor displays an error code, E42 and sounds audible alarm.
2 simulator as follows: SpO2 of 75% and pulse rate of 60 bpm.
2 simulator to the pulse oximetry cable and connect the cable to the
2 of 75 ±2 digits (flashing)
ARGUS VCM Service Manual 19
Loading...
+ 54 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.