The SCHILLER sales and service centre network is world-wide. For the address
of your local distributor, contact your nearest SCHILLER subsidiary. In case of difficulty a complete list of all distributors and subsidiaries is provided on our internet
site: http://www.schiller.ch.
Operating InstructionResponsibility of the user1.1
1Safety notes
1.1Responsibility of the user
V This device must only be used by qualified doctors or trained medical personnel
under their direct supervision.
V The numerical and graphical results and any interpretation given must be exam-
ined with respect to the overall clinical condition of the patient and the general recorded data quality.
V The indications given by this equipment are not a substitute for regular checking
of vital functions.
V Specify the competencies of the personnel for operation and repair.
V Ensure that the personnel have read and understood these operating instructions
and in particular this chapter “safety notes".
V Have damaged or missing components replaced immediately.
V The operator is responsible for compliance with all applicable accident prevention
and safety regulations.
1.2Intended use
V The ARGUS LCM/PLUS is a patient monitoring device used for the measuring of
, CO
the parameters of a patient, including ECG, SpO
sure, temperature and respiration.
V The system is suitable for internal hospital transport.
V There is no danger for patients with pacemaker.
V The device is only intended for single patient use.
V Only operate the device in accordance with the specified technical data.
V The system is not designed for sterile use nor is it designed for outdoor use.
V Do not use this unit in areas where there is any danger of explosion or in the pres-
ence of flammable gases such as anaesthetic agents.
VThis unit is CF classified. It is defibrillation protected only when the
SCHILLER original patient cable is used. However, as a safety precaution when
possible, remove electrodes before defibrillation.
V This product is not designed for direct cardiac application.
non invasive blood pres-
2
2,
1.3Organisational measures
V Before using the unit, ensure that an introduction regarding the unit functions and
the safety precautions has been provided by a medical product representative.
V Keep these operating instructions in an accessible place for reference when re-
quired. Make sure that they are always complete and legible.
V Observe the operating instructions and maintenance instructions.
Art. no.: 2.510474 rev.: d
V These operating instructions do not override any statutory or local regulations, or
procedures for the prevention of accidents and environmental protection.
Page 5
1Safety notes
1.4Safety-conscious operation
1.4Safety-conscious operation
1.5Safety facilities
ARGUS LCM/PLUS
V Make sure that the staff has read and understood the operating instructions - par-
ticularly the "Safety Notes" chapter.
V Position the device so that there is no possibility of it falling on the patient or floor.
V Do not touch the unit casing during defibrillation.
V To grant the patient's safety, it must be ensured that neither the electrodes, in-
cluding the neutral electrode, nor the patient, or persons touching the patient,
come into contact with conducting objects (e.g. RS-232 interface - see Fig. 3.1 on
page 15), even if these are earthed.
V Immediately report any changes that impair safety (including operating behav-
iour) to the person responsible.
V Do not place any liquids on the unit. If liquid should be spilled over the device,
immediately disconnect the device from the mains and wipe it. The device must
be serviced before reusing.
V Operating the device without the correctly rated fuse, or with defective cables,
constitutes a danger to life. Therefore:
– Do not operate the unit if the earth connection is suspect or if the mains lead is
damaged or suspected of being damaged.
– Damaged cable connections and connectors must be replaced immediately.
– The electrical safety devices, such as fuses, must not be modified.
– Ruptured fuses must only be replaced with the same type and rating as the orig-
inal.
1.6Operation with other devices
V Use only accessories and other parts recommended or supplied by SCHILLER
AG. Use of other than recommended or supplied parts may result in injury, inaccurate information and/or damage to the unit.
V Ancillary equipment connected to the analogue and/or digital interfaces must be
certified according to the respective IEC standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard IEC/
EN 60601-1-1. Everybody who connects additional equipment to the signal input
part or signal output part configures a medical system, and is therefore responsible that the system complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult the technical service department or your local representative.
V Any other equipment used with the patient must use the same common earth as
the ARGUS LCM.
V Precautions must be observed when using high frequency devices. Use the fre-
quency SCHILLER patient cable to avoid possible signal interference during ECG
acquisition.
V There is no danger when using the ECG unit simultaneously with electrical stim-
ulation equipment. However, the stimulation units should only be used at a sufficient distance from the electrodes. If in doubt, the patient should be disconnected
from the monitor.
V If the patient cable should become defective after defibrillation, a lead-off indica-
tion is displayed in the upper right part of the screen and an audible alarm is issued.
Art. no.: 2.510474 rev.: d
Page 6
ARGUS LCM/PLUS
Safety notes1
Operating InstructionMaintenance1.7
1.7Maintenance
V Danger of electric shock! Do not open the device. No serviceable parts inside.
Servicing may only be carried out by a qualified technician authorised by Schiller
AG.
V Before cleaning and to isolate the mains power supply, switch the unit off and dis-
connect it from the mains by removing the plug.
V Do not use high temperature sterilisation processes
(such as autoclaving). Do not use E-beam or gamma radiation sterilisation.
V Do not use solvent or abrasive cleaners on either the unit or cable assemblies.
V Do not, under any circumstances, immerse the unit or cable assemblies in liquid.
Art. no.: 2.510474 rev.: d
Page 7
1Safety notes
1.8Safety symbols and pictograms
1.8Safety symbols and pictograms
1.8.1Symbols used in this user guide
ARGUS LCM/PLUS
The safety level is classified according ANSI Z535.4. The following overview shows
the safety symbols and pictograms used in this manual
For a direct danger which could lead to severe personal injury or death.
For a possibly dangerous situation, which could lead to heavy bodily injury or death.
For a possibly dangerous situation which could lead to personal injury. This symbol
is also used to indicate possible damage to property.
For general safety notes as listed in this chapter.
Used for electrical dangers, warnings and other notes in regarding operation with
electricity.
NOTE For possibly dangerous situations, which could lead to damages to property
or system failure or IMPORTANT for helpful user information.
Reference to other guidelines
Page 8
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Safety notes1
Operating InstructionAdditional terms1.9
1.8.2Symbols used on the device
Potential equalisation
CF symbol. This unit is classified safe for internal and external use. However, It is
only defibrillation protected when used with the original SCHILLER patient cable!
The unit/component can be recycled.
Notified body of the CE certification (TÜV P.S.)
Note: Follow the instructions in the documentation.
1.9Additional terms
1.9.1Implied authorisation
Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would alone, or in combination with this device, fall within the scope of one or more patents relating to this device.
1.9.2Terms of Warranty
Your SCHILLER ARGUS LCM/PLUS is warranted against defects in material and
manufacture for the duration of one year (as from date of purchase). Excluded from
this guarantee is damage caused by an accident or as a result of improper handling.
The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorised or unqualified persons attempt to make repairs.
In case of a defect, send the apparatus to your dealer or directly to the manufacturer.
The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if:
• assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorised by him, and
• the ARGUS LCM/PLUS and approved attached equipment is used in accordance
with the manufacturer's instructions.
There are no express or implied warranties which extend beyond the warranties
hereinabove set forth. SCHILLER makes no warranty of merchantability or fitness for
a particular purpose with respect to the product or parts thereof.
Art. no.: 2.510474 rev.: d
Page 9
2Introduction
2.1Version overview
ARGUS LCM/PLUS
2Introduction
The ARGUS LCM/PLUS (Low Weight and Compact Monitor) is a flexible patient
monitoring device for the comprehensive monitoring of vital parameters in adults,
children and neonates.
Mains power supply (115/230 VAC) is used for stationary use. The LCM also provides full vital data monitoring during transport with the built in battery (one hour),
or using a DC voltage supply (11 - 30 VDC).
2.1Version overview
2.1.1ARGUS LCM (basic) from serial number 781.000-781.999
The following table shows the equipment of the ARGUS LCM. This basic device cannot be upgraded.
ARGUS LCMECGSpO
Basic deviceOnly with 3 leads/display one lead (II)xx
Options: printer, additional battery (can be updated in the field)
2
NIBP
2.1.2ARGUS LCM (basic) from serial number 781.1000 and higher
The ARGUS LCM BASIC with the microprocessor MK19-11 has a new ECG amplifier
(3p). It is now possible to to display 6 leads with a 3p-lead cable. The 3p-lead cable
is marked with a black connector housing instead of green.
This works only with software 1.24 and higher.
ARGUS LCM version
Axxxx1x
Bxxxx1xx
Dxxxx1xx
Exxxxxx
a
Monitoring amplifier only with 3p- and 5 lead cable (6 or 7 lead display)
Options for all versions: printer, additional battery and nurse call or vehicle power supply
Version A can be updated with etCO
Version B/C/D/E can be updated with printer and additional battery
a
ECG
SpO2 NIBPRespirationTemperatureetCO
module, printer and additional battery.
2
IBP
2
Page 10
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Introduction2
Operating InstructionVersion overview2.1
2.1.3ARGUS LCM PLUS (from 780.001and higher)
ARGUS LCM PLUS version
a
ECG
SpO2 NIBPRespirationTemperatureetCO2IBP
AxxxX1x
Bxxxx1xx
Dxxxx1xx
Exxxxxx
a
Diagnostic amplifier only with 3- or 5- or 10 lead cable (1- or 7- or 12 lead display)
Options for all versions: printer, additional battery and nurse call or vehicle power supply
Version A can be upgrade with etCO
module, printer and additional battery.
2
Version B/C/D/E can be upgrade with printer and additional battery
Art. no.: 2.510474 rev.: d
Page 11
2Introduction
2.2Functional overview
ARGUS LCM/PLUS
2.2Functional overview
2.2.1Buttons of the Argus LCM PLUS
i
Audible Alarm
suppress/acknowledge
ECG
Alarm OFF
This is the LCM plus
version E.
1
2
Patient connections
CO
IBP
SPO
NIBP
2
Access to the menu IBP and NIBP alarm
on/off and IBP calibration.
Direct buttons for :
1. NIBP s
tart/stop
2. Trend display
3
4
3. Print
4. Setting alarm limits
Menu access button
LED indicating battery
operation
LCM ON/OFF button
LED indicating main supply operation
°C/°F
2/
Navigation buttons down/up,/leftright
Program and enter button
IBP connection instead of Temperature
This is the LCM plus
version D.
Page 12
Fig. 2.1ARGUS LCM front view
IBP connection instead of CO
2
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Introduction2
Operating InstructionFunctional overview2.2
2.2.2Description of buttons
ON/OFF
The LED indicates if the LCM is running from mains power or battery.
Button for suppression/acknowledgement of audible alarms
The alarm can be suppressed or acknowledged in two ways. This is defined in set-
tings About+/Alarm setup menu.
See detailed description in paragraph 3.7.3
Direct buttons for:
(1) NIBP start/stop or, if pressed for 2 s, switch between auto and manual
(2) Trend and alarm displays
(3) Printout of the current display
(4) Alarm limit settings
Art. no.: 2.510474 rev.: d
Enter
Navigation
Alarm off
These buttons disable alarms of the individual parameters. Acknowledgement is
displayed on the top left e.g. ECG Alarm OFF. The LED above the corresponding
button is lit (for details refer to section 3.7.4).
Menu access
Opens the main menu. Navigate through the menus with the right/left buttons.
Press the menu button to again leave the menu.
The navigation and enter buttons have different functions dependent on operating
mode:
Normal single lead display:
Pressing the enter button shows the first 3 leads instead of the single lead display.
By pressing the up/down buttons, the next lead, or lead group is displayed.
Press the left button to adjust the loudness of the audible alarm and the QRS signal.
Programming mode:
1. Activating the programming mode with the menu button.
2. Selecting the menu with the left/right buttons.
3. Selecting a value with the up/down buttons.
4. Press the enter button.
5. Change a value using the up/down buttons.
6. Retrieve a value by pressing enter.
Trend display mode:
Activate the trend display mode with the trend display button.
Use the
left/right buttons to move forward/backward in the trend display.
Use the trend button to toggle between the trend display, alarm display and normal
monitor display.
Page 13
2Introduction
2.2Functional overview
ARGUS LCM/PLUS
2.2.3Description of display
Alarms
(1) Status field for physiological alarms
(2) Status field for technical alarms
Curve field
(3) 1 or 3 channel leads display. With enter display 1 or 3 leads, with
up/down select the following/previous lead/lead group.
System status field
(4) System status field
(5) Symbol for battery operation.
(6) Loudness (3-step) for alarm- and QRS-sound. Adjust with left button.
(7) Symbol of the selected patient type.
Blood pressure
(8) < 0:12 A = remaining time (h/min) to the next automatic blood pressure meas-
urement.
> 0:12 = Time since the last manual blood pressure measurement
Remark: During blood pressure measuring the actual system pressure in mmHg
will be displayed.
Measurement field
(9) Various displayed values e.g. HR, Sp02, IBP, CO2, RRi, temperature.
I
aVR, III,...
3
I,II, …
4
1
Alarm HR LowECG lead off
I
II
I
II
III
III
P
P
R
11 :0428-10-04
NIBP
mmHg
<0:12A
>0:12
127/77
SYS/DIA
2
(100
(MAP)
PM detection
HR
/ min
PM ON
SpO
%
PP
/ min
RRi
/ min
T
1
)
°C
RRi Alarm
switch off
60
2
98
60
10
36.8
Art. no.: 2.510474 rev.: d
Page 14
56
78
9
ARGUS LCM/PLUS
Operation3
Operating InstructionStart-up3.1
3Operation
3.1Start-up
V Danger of electrical shock. Do not operate the unit if the earth connection is sus-
pect or if the mains lead is damaged or suspected of being damaged.
3.1.1Connecting and power on
(1) RS-232
(2) Mains connection (100 -115 or 220 -240 VAC)
(3) Potential equalisation
(4) Nurse call or DC in 11 - 30 VDC (option)
Fig. 3.1ARGUS LCM back panel
1. Voltage setting (2) 115 or 230 V. Refer to chapter 5.3 for the mains voltage. Connect the power cable at the rear of the unit (2).
2. Connect the potential equalisation cable (3) to the central potential equalisation
socket.
3. Press the on/off button.
4. Check that all LEDs flash shortly and there is a
beep on start-up.
5. Check the settings according to sections 3.4 and 7.1.
3.1.2Battery operation
Important
The Battery operation is indicated by the LED below the battery symbol.
When the battery charge is low, the alarm message Battery low appears or
Art. no.: 2.510474 rev.: d
1
– the LED (1) blinks
– the Battery symbol in the bottom left display field blinks
for Battery recharging refer to chapter 5.2
Page 15
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.