Schiller Argus LCM User Manual

ARGUS LCM
Patient monitor
ARGUS LCM and ARGUS LCM PLUS
Art. no.: 2.510474 rev.: d
Operating Instruction
Sales and Service Information
The SCHILLER sales and service centre network is world-wide. For the address of your local distributor, contact your nearest SCHILLER subsidiary. In case of dif­ficulty a complete list of all distributors and subsidiaries is provided on our internet site: http://www.schiller.ch.
Sales information can also be obtained from:
sales@schiller.ch
Address Headquarters
SCHILLER AG Phone: +41 (0) 41 766 42 42 Altgasse 68 Fax: +41 (0) 41 761 08 80, CH-6341 Baar, Switzerland E-mail: sales@schiller.ch Web: www.schiller.ch
Article no.: 2.510474 rev.: d Issue date: 10.04.06
ARGUS LCM/PLUS
Operating Instruction
Contents
1 Safety notes ..............................................5
1.1 Responsibility of the user ................................................... 5
1.2 Intended use ........................................................................ 5
1.3 Organisational measures..................................................... 5
1.4 Safety-conscious operation ................................................ 6
1.5 Safety facilities ..................................................................... 6
1.6 Operation with other devices .............................................. 6
1.7 Maintenance.......................................................................... 7
1.8 Safety symbols and pictograms ......................................... 8
1.8.1 Symbols used in this user guide ........................................................ 8
1.8.2 Symbols used on the device .............................................................. 9
1.9 Additional terms ................................................................... 9
1.9.1 Implied authorisation.......................................................................... 9
1.9.2 Terms of Warranty ............................................................................. 9
2 Introduction ............................................10
2.1 Version overview ................................................................ 10
2.1.1 ARGUS LCM (basic) from serial number 781.000-781.999............. 10
2.1.2 ARGUS LCM (basic) from serial number 781.1000 and higher ....... 10
2.1.3 ARGUS LCM PLUS (from 780.001and higher)................................ 11
2.2 Functional overview ........................................................... 12
2.2.1 Buttons of the Argus LCM PLUS...................................................... 12
2.2.2 Description of buttons ..................................................................... 13
2.2.3 Description of display ....................................................................... 14
3 Operation ................................................ 15
3.1 Start-up................................................................................ 15
3.1.1 Connecting and power on ................................................................ 15
3.1.2 Battery operation.............................................................................. 15
3.1.3 Operation with external dc voltage................................................... 16
3.2 Switching off and disconnecting from mains.................. 16
3.2.1 Interruption of the mains supply ....................................................... 16
3.3 Inserting printing paper .................................................... 17
3.4 Initial settings ..................................................................... 18
3.4.1 Selecting the language..................................................................... 18
3.4.2 Saving and restoring default values ................................................ 18
3.4.3 Load factory defaults........................................................................ 18
3.4.4 Alarm limit setting............................................................................. 19
3.4.5 Setting loudness of audible alarm and QRS sound ......................... 20
3.5 Trend or alarm display ....................................................... 21
3.6 Display of additional leads ................................................ 21
Art. no.: 2.510474 rev.: d
3.7 Procedure in case of an alarm .......................................... 22
3.7.1 Display of alarms.............................................................................. 22
3.7.2 Switching off an alarm...................................................................... 22
3.7.3 Suppressing/acknowledging an audible alarm................................. 23
3.7.4 Preventive alarm suppression.......................................................... 24
3.7.5 Overview of physiological alarms..................................................... 24
Page 1
ARGUS LCM/PLUS
4 Start monitoring .....................................25
4.1 Connecting the cables for monitoring.............................. 25
4.2 ECG monitoring .................................................................. 25
4.2.1 RRi monitoring with HF-ECG cable ................................................. 25
4.2.2 Pacemaker monitoring..................................................................... 26
4.2.3 Pacemaker monitoring with HF-ECG cable ..................................... 26
4.3 NIBP monitoring ................................................................. 27
4.4 SPO
monitoring................................................................. 28
2
4.5 IBP monitoring .................................................................... 29
4.5.1 Preparing IBP measurement............................................................ 29
4.5.2 IBP calibration.................................................................................. 30
4.6 CO2 monitoring ................................................................... 30
4.7 Temperature monitoring .................................................... 30
5 Maintenance ............................................31
5.1 Maintenance interval .......................................................... 31
5.1.1 Visual unit check.............................................................................. 31
5.1.2 Button test........................................................................................ 31
5.2 Maintenance interval for the battery ................................. 32
5.2.1 Charging the battery ........................................................................ 32
5.2.2 Battery disposal ............................................................................... 32
5.3 Changing the fuse and mains voltage .............................. 33
5.4 Cleaning the device, cable and sensors........................... 34
5.4.1 Cleaning the device, cables and sensors ........................................ 34
5.5 Zero and two point CO2 calibration .................................. 35
5.5.1 Starting the zero point calibration .................................................... 35
5.5.2 Starting the two point calibration...................................................... 35
5.6 Trouble shooting ................................................................ 36
5.6.1 Accessories and disposables........................................................... 37
6 Technical Data ........................................38
6.1 System data......................................................................... 38
6.2 Technical data - measured values .................................... 40
6.2.1 ECG ................................................................................................. 40
6.2.2 Respiration....................................................................................... 40
6.2.3 Temperature .................................................................................... 41
6.2.4 NIBP - non-invasive blood pressure ................................................ 41
6.2.5 IBP - invasive blood pressure .......................................................... 41
6.2.6 SpO
6.2.7 etCO
- pulsoximetry ........................................................................ 42
2
- capnography....................................................................... 42
2
7 Appendix .................................................43
7.1 Settings menu ..................................................................... 43
7.1.1 Main menu ....................................................................................... 43
7.1.2 Special menu “About+” .................................................................... 44
7.2 Factory settings .................................................................. 45
7.3 Connecting the ECG patient cable ................................... 46
7.3.1 3-leads cable for children and neonates.......................................... 46
7.3.2 3-lead cable for respiration .............................................................. 46
7.3.3 3-lead cable for adults...................................................................... 46
7.3.4 3p-lead cable ................................................................................... 48
Art. no.: 2.510474 rev.: d
Page 2
ARGUS LCM/PLUS
Operating Instruction
7.3.5 Three- and five-lead cables for adults and children ......................... 48
7.3.6 10-lead ECG patient cable............................................................... 49
7.3.7 Electrodes identification and colour code IEC/AHA ......................... 50
7.4 CO2 settings........................................................................ 51
7.4.1 Description of CO2 parameters........................................................ 51
7.4.2 Compensation settings..................................................................... 51
7.4.3 Combination of N
7.4.4 Environmental pressure compensation............................................ 52
O and O2 compensation .................................... 52
2
8 Index ........................................................ 53
Art. no.: 2.510474 rev.: d
Page 3
ARGUS LCM/PLUS
Page 4
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Safety notes 1
Operating Instruction Responsibility of the user 1.1

1 Safety notes

1.1 Responsibility of the user

V This device must only be used by qualified doctors or trained medical personnel
under their direct supervision.
V The numerical and graphical results and any interpretation given must be exam-
ined with respect to the overall clinical condition of the patient and the general re­corded data quality.
V The indications given by this equipment are not a substitute for regular checking
of vital functions.
V Specify the competencies of the personnel for operation and repair. V Ensure that the personnel have read and understood these operating instructions
and in particular this chapter “safety notes".
V Have damaged or missing components replaced immediately. V The operator is responsible for compliance with all applicable accident prevention
and safety regulations.

1.2 Intended use

V The ARGUS LCM/PLUS is a patient monitoring device used for the measuring of
, CO
the parameters of a patient, including ECG, SpO sure, temperature and respiration.
V The system is suitable for internal hospital transport. V There is no danger for patients with pacemaker. V The device is only intended for single patient use. V Only operate the device in accordance with the specified technical data. V The system is not designed for sterile use nor is it designed for outdoor use. V Do not use this unit in areas where there is any danger of explosion or in the pres-
ence of flammable gases such as anaesthetic agents.
V This unit is CF classified. It is defibrillation protected only when the
SCHILLER original patient cable is used. However, as a safety precaution when possible, remove electrodes before defibrillation.
V This product is not designed for direct cardiac application.
non invasive blood pres-
2
2,

1.3 Organisational measures

V Before using the unit, ensure that an introduction regarding the unit functions and
the safety precautions has been provided by a medical product representative.
V Keep these operating instructions in an accessible place for reference when re-
quired. Make sure that they are always complete and legible.
V Observe the operating instructions and maintenance instructions.
Art. no.: 2.510474 rev.: d
V These operating instructions do not override any statutory or local regulations, or
procedures for the prevention of accidents and environmental protection.
Page 5
1 Safety notes

1.4 Safety-conscious operation

1.4 Safety-conscious operation

1.5 Safety facilities

ARGUS LCM/PLUS
V Make sure that the staff has read and understood the operating instructions - par-
ticularly the "Safety Notes" chapter.
V Position the device so that there is no possibility of it falling on the patient or floor. V Do not touch the unit casing during defibrillation. V To grant the patient's safety, it must be ensured that neither the electrodes, in-
cluding the neutral electrode, nor the patient, or persons touching the patient, come into contact with conducting objects (e.g. RS-232 interface - see Fig. 3.1 on page 15), even if these are earthed.
V Immediately report any changes that impair safety (including operating behav-
iour) to the person responsible.
V Do not place any liquids on the unit. If liquid should be spilled over the device,
immediately disconnect the device from the mains and wipe it. The device must be serviced before reusing.
V Operating the device without the correctly rated fuse, or with defective cables,
constitutes a danger to life. Therefore:
– Do not operate the unit if the earth connection is suspect or if the mains lead is
damaged or suspected of being damaged. – Damaged cable connections and connectors must be replaced immediately. – The electrical safety devices, such as fuses, must not be modified. – Ruptured fuses must only be replaced with the same type and rating as the orig-
inal.

1.6 Operation with other devices

V Use only accessories and other parts recommended or supplied by SCHILLER
AG. Use of other than recommended or supplied parts may result in injury, inac­curate information and/or damage to the unit.
V Ancillary equipment connected to the analogue and/or digital interfaces must be
certified according to the respective IEC standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall comply with the valid version of the system standard IEC/ EN 60601-1-1. Everybody who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore responsi­ble that the system complies with the requirements of the valid version of the sys­tem standard IEC/EN 60601-1-1. If in doubt, consult the technical service depart­ment or your local representative.
V Any other equipment used with the patient must use the same common earth as
the ARGUS LCM.
V Precautions must be observed when using high frequency devices. Use the fre-
quency SCHILLER patient cable to avoid possible signal interference during ECG acquisition.
V There is no danger when using the ECG unit simultaneously with electrical stim-
ulation equipment. However, the stimulation units should only be used at a suffi­cient distance from the electrodes. If in doubt, the patient should be disconnected from the monitor.
V If the patient cable should become defective after defibrillation, a lead-off indica-
tion is displayed in the upper right part of the screen and an audible alarm is is­sued.
Art. no.: 2.510474 rev.: d
Page 6
ARGUS LCM/PLUS
Safety notes 1
Operating Instruction Maintenance 1.7

1.7 Maintenance

V Danger of electric shock! Do not open the device. No serviceable parts inside.
Servicing may only be carried out by a qualified technician authorised by Schiller AG.
V Before cleaning and to isolate the mains power supply, switch the unit off and dis-
connect it from the mains by removing the plug.
V Do not use high temperature sterilisation processes
(such as autoclaving). Do not use E-beam or gamma radiation sterilisation.
V Do not use solvent or abrasive cleaners on either the unit or cable assemblies. V Do not, under any circumstances, immerse the unit or cable assemblies in liquid.
Art. no.: 2.510474 rev.: d
Page 7
1 Safety notes

1.8 Safety symbols and pictograms

1.8 Safety symbols and pictograms

1.8.1 Symbols used in this user guide

ARGUS LCM/PLUS
The safety level is classified according ANSI Z535.4. The following overview shows the safety symbols and pictograms used in this manual
For a direct danger which could lead to severe personal injury or death.
For a possibly dangerous situation, which could lead to heavy bodily injury or death.
For a possibly dangerous situation which could lead to personal injury. This symbol is also used to indicate possible damage to property.
For general safety notes as listed in this chapter.
Used for electrical dangers, warnings and other notes in regarding operation with electricity.
NOTE For possibly dangerous situations, which could lead to damages to property or system failure or IMPORTANT for helpful user information.
Reference to other guidelines
Page 8
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Safety notes 1
Operating Instruction Additional terms 1.9

1.8.2 Symbols used on the device

Potential equalisation
CF symbol. This unit is classified safe for internal and external use. However, It is only defibrillation protected when used with the original SCHILLER patient cable!
The unit/component can be recycled.
Notified body of the CE certification (TÜV P.S.)
Note: Follow the instructions in the documentation.

1.9 Additional terms

1.9.1 Implied authorisation

Possession or purchase of this device does not convey any express or implied li­cense to use the device with replacement parts which would alone, or in combina­tion with this device, fall within the scope of one or more patents relating to this de­vice.

1.9.2 Terms of Warranty

Your SCHILLER ARGUS LCM/PLUS is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subse­quent damage is excluded. The warranty is void if unauthorised or unqualified per­sons attempt to make repairs.
In case of a defect, send the apparatus to your dealer or directly to the manufacturer. The manufacturer can only be held responsible for the safety, reliability, and perform­ance of the apparatus if:
• assembly operations, extensions, readjustments, modifications, or repairs are car­ried out by persons authorised by him, and
• the ARGUS LCM/PLUS and approved attached equipment is used in accordance with the manufacturer's instructions.
There are no express or implied warranties which extend beyond the warranties hereinabove set forth. SCHILLER makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof.
Art. no.: 2.510474 rev.: d
Page 9

2 Introduction

2.1 Version overview

ARGUS LCM/PLUS
2 Introduction
The ARGUS LCM/PLUS (Low Weight and Compact Monitor) is a flexible patient monitoring device for the comprehensive monitoring of vital parameters in adults, children and neonates.
Mains power supply (115/230 VAC) is used for stationary use. The LCM also pro­vides full vital data monitoring during transport with the built in battery (one hour), or using a DC voltage supply (11 - 30 VDC).
2.1 Version overview

2.1.1 ARGUS LCM (basic) from serial number 781.000-781.999

The following table shows the equipment of the ARGUS LCM. This basic device can­not be upgraded.
ARGUS LCM ECG SpO
Basic device Only with 3 leads/display one lead (II) x x Options: printer, additional battery (can be updated in the field)
2
NIBP

2.1.2 ARGUS LCM (basic) from serial number 781.1000 and higher

The ARGUS LCM BASIC with the microprocessor MK19-11 has a new ECG amplifier (3p). It is now possible to to display 6 leads with a 3p-lead cable. The 3p-lead cable is marked with a black connector housing instead of green.
This works only with software 1.24 and higher.
ARGUS LCM version
Axxxx1x B x x x x 1x x Dxxxx1xx E x x x x x x
a
Monitoring amplifier only with 3p- and 5 lead cable (6 or 7 lead display) Options for all versions: printer, additional battery and nurse call or vehicle power supply Version A can be updated with etCO
Version B/C/D/E can be updated with printer and additional battery
a
ECG
SpO2 NIBP Respiration Temperature etCO
module, printer and additional battery.
2
IBP
2
Page 10
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Introduction 2
Operating Instruction Version overview 2.1

2.1.3 ARGUS LCM PLUS (from 780.001and higher)

ARGUS LCM PLUS version
a
ECG
SpO2 NIBP Respiration Temperature etCO2IBP
AxxxX1x B x x x x 1x x Dxxxx1xx E x x x x x x
a
Diagnostic amplifier only with 3- or 5- or 10 lead cable (1- or 7- or 12 lead display) Options for all versions: printer, additional battery and nurse call or vehicle power supply Version A can be upgrade with etCO
module, printer and additional battery.
2
Version B/C/D/E can be upgrade with printer and additional battery
Art. no.: 2.510474 rev.: d
Page 11
2 Introduction

2.2 Functional overview

ARGUS LCM/PLUS
2.2 Functional overview

2.2.1 Buttons of the Argus LCM PLUS

i
Audible Alarm suppress/acknowledge
ECG
Alarm OFF
This is the LCM plus version E.
1
2
Patient connections
CO IBP
SPO
NIBP
2
Access to the menu IBP and NIBP alarm on/off and IBP calibration.
Direct buttons for :
1. NIBP s
tart/stop
2. Trend display
3
4
3. Print
4. Setting alarm limits
Menu access button
LED indicating battery operation
LCM ON/OFF button
LED indicating main supply operation
°C/°F
2/
Navigation buttons down/up, /leftright
Program and enter button
IBP connection instead of Temperature
This is the LCM plus version D.
Page 12
Fig. 2.1 ARGUS LCM front view
IBP connection instead of CO
2
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
Introduction 2
Operating Instruction Functional overview 2.2

2.2.2 Description of buttons

ON/OFF
The LED indicates if the LCM is running from mains power or battery.
Button for suppression/acknowledgement of audible alarms The alarm can be suppressed or acknowledged in two ways. This is defined in set-
tings About+/Alarm setup menu.
See detailed description in paragraph 3.7.3
Direct buttons for:
(1) NIBP start/stop or, if pressed for 2 s, switch between auto and manual
(2) Trend and alarm displays
(3) Printout of the current display
(4) Alarm limit settings
Art. no.: 2.510474 rev.: d
Enter
Navigation
Alarm off
These buttons disable alarms of the individual parameters. Acknowledgement is displayed on the top left e.g. ECG Alarm OFF. The LED above the corresponding button is lit (for details refer to section 3.7.4).
Menu access
Opens the main menu. Navigate through the menus with the right/left buttons. Press the menu button to again leave the menu.
The navigation and enter buttons have different functions dependent on operating mode:
Normal single lead display:
Pressing the enter button shows the first 3 leads instead of the single lead display.
By pressing the up/down buttons, the next lead, or lead group is displayed.
Press the left button to adjust the loudness of the audible alarm and the QRS sig­nal.
Programming mode:
1. Activating the programming mode with the menu button.
2. Selecting the menu with the left/right buttons.
3. Selecting a value with the up/down buttons.
4. Press the enter button.
5. Change a value using the up/down buttons.
6. Retrieve a value by pressing enter.
Trend display mode:
Activate the trend display mode with the trend display button. Use the left/right buttons to move forward/backward in the trend display. Use the trend button to toggle between the trend display, alarm display and normal monitor display.
Page 13
2 Introduction
2.2 Functional overview
ARGUS LCM/PLUS

2.2.3 Description of display

Alarms
(1) Status field for physiological alarms (2) Status field for technical alarms
Curve field
(3) 1 or 3 channel leads display. With enter display 1 or 3 leads, with
up/down select the following/previous lead/lead group.
System status field
(4) System status field (5) Symbol for battery operation. (6) Loudness (3-step) for alarm- and QRS-sound. Adjust with left button. (7) Symbol of the selected patient type.
Blood pressure
(8) < 0:12 A = remaining time (h/min) to the next automatic blood pressure meas-
urement. > 0:12 = Time since the last manual blood pressure measurement Remark: During blood pressure measuring the actual system pressure in mmHg will be displayed.
Measurement field
(9) Various displayed values e.g. HR, Sp02, IBP, CO2, RRi, temperature.
I
aVR, III,...
3
I,II, …
4
1
Alarm HR Low ECG lead off
I
II
I II III
III
P P
R
11 :04 28-10-04
NIBP
mmHg <0:12A
>0:12
127/77
SYS/DIA
2
(100
(MAP)
PM detection
HR
/ min
PM ON
SpO
%
PP
/ min
RRi
/ min
T
1
)
°C
RRi Alarm switch off
60
2
98 60
10
36.8
Art. no.: 2.510474 rev.: d
Page 14
56
78
9
ARGUS LCM/PLUS
Operation 3
Operating Instruction Start-up 3.1

3Operation

3.1 Start-up

V Danger of electrical shock. Do not operate the unit if the earth connection is sus-
pect or if the mains lead is damaged or suspected of being damaged.

3.1.1 Connecting and power on

(1) RS-232 (2) Mains connection (100 -115 or 220 -240 VAC) (3) Potential equalisation (4) Nurse call or DC in 11 - 30 VDC (option)
Fig. 3.1 ARGUS LCM back panel
1. Voltage setting (2) 115 or 230 V. Refer to chapter 5.3 for the mains voltage. Con­nect the power cable at the rear of the unit (2).
2. Connect the potential equalisation cable (3) to the central potential equalisation socket.
3. Press the on/off button.
4. Check that all LEDs flash shortly and there is a beep on start-up.
5. Check the settings according to sections 3.4 and 7.1.

3.1.2 Battery operation

Important
The Battery operation is indicated by the LED below the battery symbol.
When the battery charge is low, the alarm message Battery low appears or
Art. no.: 2.510474 rev.: d
1
– the LED (1) blinks – the Battery symbol in the bottom left display field blinks
for Battery recharging refer to chapter 5.2
Page 15
Loading...
+ 39 hidden pages